Nivus Pipe Mounting System RMS is designed for temporary sensor installation in pipes, primarily for portable flow measurements. It's compatible with various NIVUS wedge and air-ultrasonic sensors (CS2, CSM, KDA, POA, DSM, OCL) across a range of pipe diameters (DN200 - DN2000), utilizing a robust stainless steel (1.4404) construction. Multiple system configurations (RMS 2, 3, 4, 5) cater to different pipe sizes and sensor combinations.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

NIVUS Pipe Mounting System RMS Installation Manual | Manualzz
Installation Instruction for
Pipe Mounting Systems RMS
Rev. 00 / 21.04.2017
(Original Manual: German)
Branches
NIVUS AG
NIVUS Sp. z o.o.
Burgstraße 28
CH - 8750 Glarus
Tel.: +41 (0)55 6452066
Fax: +41 (0)55 6452014
E-mail: [email protected]
Internet: www.nivus.de
ul. Hutnicza 3 / B-18
PL - 81-212 Gdynia
Tel.: +48 (0) 58 7602015
Fax: +48 (0) 58 7602014
E-mail: [email protected]
Internet: www.nivus.pl
NIVUS Austria
NIVUS Middle East (FZE)
Mühlbergstraße 33B
A-3382 Loosdorf
Tel.: +43 (2754) 567 63 21
Fax: +43 (2754) 567 63 20
E-mail: [email protected]
Internet: www.nivus.de
Building Q 1-1 ap. 055
P.O. Box: 9217
Sharjah Airport International
Free Zone
Tel.: +971 6 55 78 224
Fax: +971 6 55 78 225
E-mail: [email protected]
Internet: www.nivus.com
NIVUS France
14, rue de la Paix
F - 67770 Sessenheim
Tel.: +33 (0)3 88071696
Fax: +33 (0)3 88071697
E-mail: [email protected]
Internet: www.nivus.fr
NIVUS U.K. Ltd
Head office UK:
Wedgewood Rugby Road
Weston under Wetherley
Royal Leamington Spa
CV33 9BW, Warwickshire
Tel.: +44 (0)8445 3328 83
E-mail: [email protected]
Internet: www.nivus.com
page 2
NIVUS Korea Co. Ltd.
#2502, M Dong, Technopark IT Center,
32 Song-do-gwa-hak-ro,
Yeon-su-gu,
INCHEON, Korea 406-840
Tel. +82 32 209 8588
Fax. +82 32 209 8590
E-Mail: [email protected]
Internet: www.nivus.com
NIVUS Vietnam
21 Pho Duc Chinh, Ba Dinh,
Hanoi, Vietnam
Mobile (VN) 012 0446 7724
[email protected]
Pipe mounting systems / Rev. 00 /21.04.2017
Installation instruction
Pipe Mounting Systems
Copyrights and property rights
This document and its contents are proprietary to NIVUS GmbH and are not to be reproduced or copied without the express written permission of NIVUS GmbH.
Violations oblige to compensation.
Important Note
This manual may exclusively - even in parts - be copied or translated in any other way
with the express written consent of NIVUS GmbH.
Translation
If the device is sold to a country in the European Economic Area (EEA) this manual must
be translated into the language of the country in which the device is to be used.
Should the translated text be unclear, the original manual (German) must be consulted or
the manufacturer contacted for clarification.
Copyright
No part of this publication may be reproduced, transmitted, sold or disclosed without prior
permission. Damages will be claimed for violations.
All rights reserved.
Names
The use of general descriptive names, trade names, trade-marks and the like in this manual does not entitle the reader to assume they may be used freely by everyone. They are often protected registered trademarks even if not marked as such.
Pipe mounting systems / Rev. 00 / 21.04.2017
page 3
Table of contents
Table of contents
Copyrights and property rights
3
Table of contents
4
General
5
1
Applicable documentation ....................................................................... 5
2
Required Tools/Utilities ........................................................................... 5
3
Signs and definitions used ...................................................................... 6
3.1
Abbreviations used ..................................................................... 6
Safety Instructions
4
Used symbols and signal words ............................................................. 7
4.1
Protective conductor................................................................... 8
5
Use in accordance with the requirements ............................................... 8
5.1
User’s Responsibilities ............................................................... 9
5.1.1
5.1.2
Manual Keeping ................................................................................... 9
Manual Handover................................................................................. 9
5.2
Liability Disclaimer...................................................................... 9
Product specification
6
7
10
Overview Pipe Mounting Systems ........................................................10
6.1
Overview Pipe Mounting System RMS 2 .................................11
6.2
Overview Pipe Mounting System RMS 3 .................................12
6.3
Overview Pipe Mounting System RMS 4 .................................13
6.4
Overview Pipe Mounting System RMS 5 .................................14
Assembly
16
Assembly of the Pipe Mounting Systems..............................................16
7.1
Principles of Installation ...........................................................16
7.2
Principles of Sensor Fastening ................................................17
7.3
Assembly of pipe mounting system RMS 2 .............................18
7.4
Assembly of pipe mounting system RMS 3 .............................20
7.5
Assembly of pipe mounting system RMS 4 .............................23
7.6
Assembly of pipe mounting system RMS 5 .............................24
7.6.1
7.6.2
7.6.3
Base plate RMS 5 .............................................................................. 24
Extension sheet RMS 5 ..................................................................... 26
Assembling the clamping device ........................................................ 27
7.7
Air-Ultrasonic sensors in the RMS 5 pipe mounting system ....28
Maintenance
page 4
7
31
8
Maintenance ..........................................................................................31
8.1
Spare parts and accessories ....................................................31
9
Dismantling/Disposal.............................................................................31
Pipe mounting systems / Rev. 00 /21.04.2017
Installation instruction
Pipe Mounting Systems
General
Important Note
READ CAREFULLY BEFORE USE!
KEEP IN A SAFE PLACE FOR LATER REFERENCE.
This manual is an original instruction for the pipe mounting systems RMS and is for the intended use of the device. This manual is oriented exclusively to qualified expert personnel.
Read this manual carefully and completely prior to installation and connection since it contains relevant information on this product. Observe the notes and particularly follow the
warning notes and safety instructions.
If you should have problems to understand information contained within this manual either
contact the manufacturer or one of the distributors for further support. The manufacturer
cannot be held responsible for damage to persons or material due to incorrectly understood information in this instruction.
1 Applicable documentation
For the installation and operation of the complete system extra instruction manuals or
technical descriptions may be required apart from this manual.

Installation Instruction for Correlation and Doppler Sensors

Technical Instructions for Correlation Sensors and external Electronic Box

Technical Instruction for Doppler Sensors
These manuals are provided with the sensors.
2 Required Tools/Utilities
The following tools/utilities are required to install the pipe mounting systems

hammer (300 g)

cable ties

ring or open-end spanner, spanner gap 13 (for RMS 2)

Phillips screw driver Pozidriv size 2

wire cutter
RMS 5 only
The following tools are required to modify the air-ultrasonic sensors:

ring or open-end spanner, spanner gap 22

Phillips screw driver Phillips size 1
The tools are not part of the standard delivery.
Pipe mounting systems / Rev. 00 / 21.04.2017
page 5
General
3 Signs and definitions used
Image
Meaning
Remark

(Action) Step
Action to be performed by you
Note the numbering of action steps. Observe
the order of the working steps!
3.1

Cross-reference
Reference to further or detailed information
>Text<
Parameter or Menu
Indicates selected or described parameters or
menus
Reference to document
Refers to an accompanying documentation
Abbreviations used
Item designation
page 6

CS2

CSM Cross correlation sensors of the Mini sensor family

DSM Air-Ultrasonic sensor of the Mini sensor family

KDA
Doppler sensor

OCL
Air-Ultrasonic sensor

POA
Cross correlation sensors for use in part filled and full pipes and open
channels

RMS Pipe mounting system
Cross correlation sensors for use in part filled and full pipes and open
channels
Pipe mounting systems / Rev. 00 /21.04.2017
Installation instruction
Pipe Mounting Systems
Safety Instructions
4 Used symbols and signal words
The general warning symbol indicates the risk of personal injuries or death. In the
text section the general warning symbol is used in conjunction with the signal words
described below.
WARNING
Warnings at medium level risk and risk of personal injury
Indicates a possible danger with medium risk which may result in a life-threatening situation or (severe) personal injury if not avoided.
CAUTION
Warnings at low level risk or property damages
Indicates a possible danger with moderate risk which may result in minor or moderate
personal injury or material damage if not avoided.
Important Note
Contains information that should be highlighted.
Indicates a potentially damaging situation which can result in a damage of the product or
an object in its environment if not avoided.
Note
Contains information and facts.
Pipe mounting systems / Rev. 00 / 21.04.2017
page 7
Safety Instructions
4.1
Protective conductor
WARNING
Germ contamination
Please note that due to the operation in the waste water field the measurement system
and cables may be loaded with dangerous disease germs. Respective precautionary
measures must be taken to avoid damage to one’s health.
Wear protective clothing.
WARNING
Observe occupational safety regulations
Before starting installation work, check whether the occupational safety regulations are
observed.
Disregarding may lead in personal injury.
CAUTION
Risk of injury
The pipe mounting system sheets have very sharp edges. There is a risk of injury from
cuts.
Wear protective gloves.
5 Use in accordance with the requirements
Important note
The pipe mounting systems and all corresponding components are intended solely for the
purpose described below.
The pipe mounting systems are designed for the temporary installation of sensors in pipes.
Modifying or using the instruments for any other purposes without the manufacturer’s written consent will not be considered as use in accordance with the requirements.
The manufacturer cannot be held responsible for any damage resulting from improper use.
The user alone bears any risk.
page 8
Pipe mounting systems / Rev. 00 /21.04.2017
Installation instruction
Pipe Mounting Systems
5.1
User’s Responsibilities
Important Note
In the EEA (European Economic Area) national implementation of the frame-work directive 89/391/EEC and corresponding individual directives, in particular the directive
2009/104/EC concerning the minimum safety and health requirements for the use of work
equipment by workers at work, as amended, are to be observed and adhered to.
In Germany the Industrial Safety Ordinance must be observed.
5.1.1
Manual Keeping
Keep this manual in a safe place and make sure it is available for the user of this product
at any time.
5.1.2
Manual Handover
In case of selling the instruments this manual shall be provided to the purchaser since it is
a part of the standard delivery.
5.2
Liability Disclaimer
The manufacturer reserves the right to change the contents of this document including this
liability disclaimer without prior notice and cannot be held responsible in any way for possible consequences resulting from such changes.
For installation of the pipe mounting systems the following information and higher legal
regulations of the respective country (e. g. VDE regulations in Germany) such as applicable safety requirements and regulations in order to avoid accidents shall be observed.
All operations on the instruments which go beyond installation shall be carried out by
NIVUS staff or personnel authorised by NIVUS due to reasons of safety and guarantee.
The manufacturer is not responsible for failures resulting from improper use.
Pipe mounting systems / Rev. 00 / 21.04.2017
page 9
Product specification
Product specification
This installation manual exclusively refers to the installation of pipe mounting systems in
pipelines. The pipe mounting systems are mainly used for portable flow measurements.
All pipe mounting systems are made of stainless steel 1.4404.
The pipe mounting systems are installation aids for the following wedge sensors:

CS2

CSM

KDA

POA
Air-ultrasonic sensors below can be installed too:

DSM

OCL
The technical data of the sensors are listed in the corresponding manuals.
6 Overview Pipe Mounting Systems
Four different sets are available:

RMS 2 – Pipe mounting set for pipes DN200 to DN800

RMS 3 – Pipe mounting set for pipes DN150 to DN300

RMS 4 – Combination of RMS 2 and RMS 3.
Pipe mounting set for pipes DN150 to DN800

RMS 5 – Pipe mounting set for pipes DN700 to DN2000
Pipe mounting sets consist of the following elements:

Clamping device

Base plate

Clamps

Extension sheets

Completion plate (optional)
The individual parts of the RMS 2, RMS 3 and RMS 4 pipe mounting systems are compatible with each other.
The RMS 5 system uses thicker materials and hence cannot be combined with components of other sets.
Details on the scope of delivery can be found in the according overview of the individual
mounting sets.
page 10
Pipe mounting systems / Rev. 00 /21.04.2017
Installation instruction
Pipe Mounting Systems
6.1
Overview Pipe Mounting System RMS 2
Scope of Delivery set RMS 2:

1x Carrying bag, blue

1x Base plate

1x Clamping device with rotary toggle

2x Extension sheet V5

2x Extension sheet V10

2x Extension sheet V15

8x Clamp
1
2
3
4
5
6
Clamp
Extension sheet V5
Extension sheet V10
Base plate
Extension sheet V15
Clamping device with rotary toggle
Fig. 6-1
Parts of RMS 2
Fig. 6-2
Completion plates (optional) of RMS 2 and RMS 3
Pipe mounting systems / Rev. 00 / 21.04.2017
page 11
Product specification
Insideø [mm]
BST
Base plate
SPV
Clamping
device
200
1
Hole outside
1
250
1
Hole inside
1
2
300
1
Hole outside
1
2
350
1
Hole inside
1
2
400
1
Hole outside
1
2
450
1
Hole inside
1
2
2
500
1
Hole outside
1
2
2
600
1
Hole outside
1
2
700
1
Hole outside
1
800
1
Hole outside
1
Fig. 6-3
6.2
V5
Extension
sheet
V10
Extension
sheet
2
V15
Extension
sheet
2
2
2
2
2
List of required plates and sheets RMS 2
Overview Pipe Mounting System RMS 3
Scope of Delivery set RMS 3:

1x Carrying bag, blue

1x Base plate

1x Clamping device

1x Extension sheet V5

1x Extension sheet V10

4x Clamp
1
2
3
4
5
Clamp
Base plate BST
Extension sheet V5
Extension sheet V10
Clamping device
Fig. 6-4
page 12
Parts of RMS 3
Pipe mounting systems / Rev. 00 /21.04.2017
Installation instruction
Pipe Mounting Systems
Inside-ø [mm]
BST
Base plate
SPV
Clamping
device
V5
Extension sheet
150
1
Hole inside
1
1
200
1
Hole inside
1
250
1
Hole inside
1
1
1
300
1
Hole outside
1
1
1
Fig. 6-5
6.3
V10
Extension sheet
1
List of required plates RMS 3
Overview Pipe Mounting System RMS 4
The pipe mounting system RMS 4 is combined out of RMS 2 and RMS 3.

See tables in Fig. 6-3 and Fig. 6-5.
Scope of Delivery set RMS 4:

1x Carrying bag, blue

1x Base plate RMS 2

1x Base plate RMS 3

1x Clamping device RMS 2

1x Clamping device RMS 3

2x Extension sheet V5

2x Extension sheet V10

2x Extension sheet V15
Pipe mounting systems / Rev. 00 / 21.04.2017
page 13
Product specification
6.4
Overview Pipe Mounting System RMS 5
Scope of Delivery set RMS 5:

1x Carrying bag, blue

1x Base plate

1x Clamping device

4x Extension sheet VS

4x Extension sheet VM

4x Extension sheet VL

10x Clamp

1x Distance plate for DSM-L0 (two parts)

1x Distance plate for OCL-L1
1
2
3
4
5
6
7
8
Clamping device
Distance plate for OCL-L1
Distance plate for DSM-L0 (two parts)
Base plate BST
Extension sheet VS
Extension sheet VM
Extension sheet VL
Clamp
Fig. 6-6
page 14
Parts of RMS 5
Pipe mounting systems / Rev. 00 /21.04.2017
Installation instruction
Pipe Mounting Systems
Number of plates
Inside-ø
[mm]
Base plate
700
1
Without Air-Ultrasonic sensor
short
middle
long
(VS)
(VM)
(VL)
4
800
1
2
3
1
900
1
2
2
2
2
1000
1
4
2
1
3
1100
1
4
3
3
1200
1
2
4
2
4
1300
1
2
2
2
1400
1
4
2
1
1
2
1500
1
2
2
3
1
2
1600
1
2
2
2
2
2
2
1700
1
4
2
2
3
3
2
1800
1
4
2
3
3
2
1900
1
2
4
2
2
4
2
2000
1
2
4
2
Fig. 6-7
With Air-Ultrasonic sensor
short
middle
long
(VS)
(VM)
(VL)
1
1
2
4
List of required plates RMS 5
Pipe mounting systems / Rev. 00 / 21.04.2017
page 15
Assembly
Assembly
7 Assembly of the Pipe Mounting Systems
WARNING
Check hazards caused by explosive gases!
Prior to executing mounting works in sewer systems always check compliance with all
regulations on safety at work.
Check possible hazards caused by explosive gases by using a gas detector.
Disregarding may lead to personal injury.
CAUTION
Risk of injury
The pipe mounting system sheets have very sharp edges. There is a risk of injury from
cuts.
Wear protective gloves.
Note
In larger pipe diameters it may be necessary to additionally secure the pipe mounting
system from being washed away due to high flow velocities.
Use stainless steel screws for this purpose.
7.1
Principles of Installation
The number of sheets depends on the existing inner pipe diameter.
The mounting steps for the individual pipe mounting systems are described in chapter
>Mounting<.
The following applies for the mounting of pipe mounting systems:
page 16

Observe the selection list of mounting sheets.

Place the clamping device always in the pipe crown (except when using an airultrasonic sensor).

Place the base plate on the pipe bottom.

Install the same number of required extension plates left and right between clamping
device and base plate (except when using an air-ultrasonic sensor).

Install plates parallel to the pipe walls.

Insert clamps towards the flow direction flush with the mounting plates.

Observe appropriate contact pressure on the pipe walls to prevent the mounting system from getting loose.

In larger channel diameters it may be necessary to additionally fix the pipe mounting
system on the channel walls by using stainless steel screws.

No gap shall remain between mounting plate and sensor as well as between base
plate and channel bottom (risk of build-up).
Pipe mounting systems / Rev. 00 /21.04.2017
Installation instruction
Pipe Mounting Systems
7.2
Principles of Sensor Fastening
The following applies for using air-ultrasonic sensor and wedge sensor at the same
time:

Place air-ultrasonic sensor on pipe crown.

Align air-ultrasonic sensor plane parallel to water surface.

Install clamping device immediately adjacent to air-ultrasonic sensor.

Use completion plate (optional for RMS 2-4) for wedge sensor (Fig. 6-2).
Fig. 7-1
Basic sensor fastening
With heavy dirt load or sedimentation:

Place wedge sensor off centre (towards step irons).

Lay the sensor cable to the same side as the sensor. Do not lay cables across
channel bottom.

Fix sensor cables to RMS using cable ties and lay upwards.
1
2
3
4
5
6
Pipe wall
Pipe mounting system
Sedimentation
Sensor
Cable
Cable tie
Fig. 7-2
Install sensor off centre
Pipe mounting systems / Rev. 00 / 21.04.2017
page 17
Assembly
7.3
Assembly of pipe mounting system RMS 2
Measure the pipe diameter before you begin installation.
The table in Fig. 6-3 shows which sheets to use for which pipe diameter.

Put the RMS components on a plane surface.
Assemble base plate and extension plate.
Insert the spigots of the base plate into the
holes of the extension sheet.
Insert the clamp towards the flow direction
flush with the mounting plate.
If the clamp cannot be fastened easily use a
hammer to carefully hammer the clamp into
position.
page 18
Pipe mounting systems / Rev. 00 /21.04.2017
Installation instruction
Pipe Mounting Systems
Place flow velocity sensor
Put the flow velocity sensor onto the base
plate with the slotted holes down.
Pull sensor backwards on the base plate.
Sensor shall be arrested flush with the base
plate.
Assemble RMS completely
When the inner pipe diameter is reached:
Assemble clamping device and extension
sheet.
Insert the spigots of the clamping device into
the holes of the extension sheet.
Fasten clamping device to extension sheets
on both sides using clamps.
Prior to installation in the channel fully close
the clamping device by rotating clockwise.
Pipe mounting systems / Rev. 00 / 21.04.2017
page 19
Assembly
Insert the complete pipe mounting system
into the pipe and align accordingly.
Open the clamping device by rotating
counterclockwise until the pipe mounting
system is clamped firmly within the pipe.
Note
Always fix the cable to the RMS (small holes) by using cable ties. Do not lay the sensor
cable across the invert centre.
Fig. 7-3
7.4
RMS 2 with mounted POA and air-ultrasonic sensor
Assembly of pipe mounting system RMS 3

Proceed as described under RMS 2.
1. Measure the pipe diameter.
2. Select the required plates from the set (see table in Fig. 6-5).
3. Put the components on a plane surface.
4. Assemble base plate and extension sheets as seen with RMS 2.
page 20
Pipe mounting systems / Rev. 00 /21.04.2017
Installation instruction
Pipe Mounting Systems
Insert Mini Sensor family flow velocity sensor
CSM:
Put the CSM flow velocity sensor with the
slotted holes down onto the base plate.
Pull sensor backwards on the base plate until
arrested.
Important:
The cable side of the sensor is not flush with
the base plate.
CSM-D:
Put the CSM-D flow velocity sensor with the
slotted holes down onto the base plate.
Pull sensor backwards on the base plate
until arrested.
Pipe mounting systems / Rev. 00 / 21.04.2017
page 21
Assembly
Flow velocity sensor and air-ultrasonic sensor
CSM-D on completion plate:
Put the flow velocity sensor with the slotted
holes down onto the completion plate.
Pull sensor backwards on the completion
plate until arrested.
Put the DSM air-ultrasonic sensor alongside
the extension sheet.
Push extension sheet through the slot
between sensor mounting plate and cover
plate.
Slot see Fig. 7-10.
page 22
Pipe mounting systems / Rev. 00 /21.04.2017
Installation instruction
Pipe Mounting Systems
RMS 3 pipe mounting system with
completion plate for common installation of
wedge sensor and air-ultrasonic sensor.
The clamping device is placed laterally to
the air-ultrasonic sensor.
7.5
Assembly of pipe mounting system RMS 4
The RMS 4 is a combination of RMS 2 and RMS 3.
The selection of base plate and clamping device depends on the sensor used.
Extension sheets are determined based on the pipe diameter.

For installation proceed as described under RMS 2 and RMS 3.
Pipe mounting systems / Rev. 00 / 21.04.2017
page 23
Assembly
7.6
Assembly of pipe mounting system RMS 5
Note
Since the dimensions of the RMS 5 complicate the handling of the individual components
NIVUS recommends 2 persons to assemble and to install the RMS 5.
The RMS 5 pipe mounting system is designed for large inner pipe diameters. This is why
the sheets are much larger than those of the other pipe mounting systems.
7.6.1

Firstly equip the base plate with sensors and wedge supports if required.

Do not assemble the RMS 5 completely before you are on site.
Base plate RMS 5
Several mounting aids can be found on the base plate.
1
2
3
4
5
6
Fastening holes for extension sheets
Fastening holes for outer sensors or wedge supports
Specification of nominal diameter according to Gauss. Positioning aid for
three flow velocity sensors
Fastening for centre flow velocity sensor sensor
Indication of flow direction
Optional / additional fastening with cable ties
Fig. 7-4
page 24
Mounting aids on the RMS 5 base plate
Pipe mounting systems / Rev. 00 /21.04.2017
Installation instruction
Pipe Mounting Systems
Place sensors on the base plate
Up to 3 flow velocity sensors can be fixed on the base plate.
The centre flow velocity sensor shall be installed so that the slotted holes lock into the
spigots on the base plate. Here the sensor tip must face the arrow on the base plate (Fig.
7-4, point 5).
When using the flow velocity sensors below observe to use both rear slotted holes for locking:

CS2

KDA

POA
The position of both outer sensors depends on the inner pipe diameter and is based on the
Gauss specifications. The according sensor positions are marked on the base plate specifying a DN figure.
Mount outer sensors horizontally
To mount both outer flow velocity sensors horizontally two wedge supports (30°) must be
installed first.
The wedge supports are available from NIVUS.

See accessories on page 31.
The wedge supports shall be fastened at the according positions (DN markings) on the
base plate by using the accompanying countersunk head screws.
Then fix the sensor on the wedge support by using the accompanying countersunk head
screws.
RMS 5 with three sensors.
For horizontal mounting:
Mount both outer sensors on wedge
supports (30°).
Pipe mounting systems / Rev. 00 / 21.04.2017
page 25
Assembly
Fig. 7-5
7.6.2
RMS 5 mounted horizontally (before installation)
Extension sheet RMS 5
Insert the spigots of the RMS 5 extension
sheet into the holes of the base plate and fix
the sheet with a clamp.
Use the front pairs of holes to assemble extension sheets and base plate if an airultrasonic sensor is to be installed at the pipe crown.
1
2
Fastening of extension sheets using an air-ultrasonic sensor
Fastening of extension sheets using flow velocity sensors only
Fig. 7-6
page 26
Fastening the extension sheets
Pipe mounting systems / Rev. 00 /21.04.2017
Installation instruction
Pipe Mounting Systems
7.6.3
Assembling the clamping device

Proceed as described in the RMS 2 mounting instructions.
1. Assemble clamping device and extension sheet.
2. Insert the spigots of the clamping device into the holes of the extension sheet.
3. Fix the clamping device at the extension sheets on both sides by using clamps.
Fig. 7-7
Clamping device fully assembled
Pipe mounting systems / Rev. 00 / 21.04.2017
page 27
Assembly
7.7
Air-Ultrasonic sensors in the RMS 5 pipe mounting system
The type OCL and DSM air-ultrasonic sensors are conceived for temporary clamped fastening in a pipe mounting system.
As per default both air-ultrasonic sensors feature slots to slide the RMS sheets through.
The installation of air-ultrasonic sensors is described on page 23.
Fig. 7-8
Pre-installed air-ultrasonic sensor on RMS sheet
Exception:
Air-ultrasonic sensors in RMS 5. Here the slot needs to be enlarged.
Air-Ultrasonic sensor in RMS 5
Note
Type OCL or DSM air-ultrasonic sensors must be modified to be used with a RMS 5 pipe
mounting system.
The RMS 5 pipe mounting system is conceived for larger pipe diameters. This is why the
individual sheets and plates are designed with extra material thickness.
Due to this the slots of the air-ultrasonic sensor types OCL and DSM need to be larger.
The distance plates for the according air-ultrasonic sensor are part of the standard delivery
of the RMS 5.

RMS 5 standard delivery see Fig. 6-6.
Modification of Air-Ultrasonic sensor

To modify a type OCL air-ultrasonic sensor proceed as follows:
1. Loosen screws of the cover plate (2) with Phillips screwdriver Phillips size 1.
2. Remove cover plate.
3. Screw on the OCL-L1 distance plate.
page 28
Pipe mounting systems / Rev. 00 /21.04.2017
Installation instruction
Pipe Mounting Systems
1
2
3
Mounting plate
Cover plate
Slot for pipe mounting system
Fig. 7-9

Mounting plates of type OCL air-ultrasonic sensor
To modify a type DSM air-ultrasonic sensor proceed as follows:
1. Loosen screws of the cover plate (5) with Phillips screwdriver Phillips size 1.
2. Remove cover plate and both distance plates (3).
3. Replace distance plates with DSM distance plates from RMS 5.
4. Screw on distance plates together with cover plate.
1
2
3
4
5
Mounting plate
Extra plate
Distance plate (two parts) short and long
Slot for pipe mounting system
Cover plate
Fig. 7-10
Mounting plates of type DSM air-ultrasonic sensor
Pipe mounting systems / Rev. 00 / 21.04.2017
page 29
Assembly
Clamping sheets installed with CS2
sensors
Clamping device installed with airultrasonic sensor
Fig. 7-11
page 30
Mounting plates and sheets fully installed with sensors
Pipe mounting systems / Rev. 00 /21.04.2017
Installation instruction
Pipe Mounting Systems
Maintenance
8 Maintenance
WARNING
Germ contamination
Please note that due to the operation in the waste water field the measurement system
and cables may be loaded with dangerous disease germs. Respective precautionary
measures must be taken to avoid damage to one’s health.
Wear protective clothing.
The pipe mounting systems are made of stainless steel and thus virtually maintenancefree.
8.1
Spare parts and accessories
Wedge supports 30°
ZUB0 KS00 L30V4A
Wedge supports 30°
Wedge support left 30° for POA and CS2 sensor; for horizontal sensor installation in pipes.
Material: stainless steel 1.457
ZUB0 KS00 R30V4A
Wedge support right 30° for POA and CS2 sensor; for horizontal sensor installation in pipes.
Material: stainless steel 1.457
Distance plate DSM
Extra distance plate to enlarge the RMS 5 slot
Distance plate OCL
Extra distance plate to enlarge the RMS 5 slot
Base plate RMS 2
Spare part for pipe mounting system RMS 2
ZUB0 RMS2 BSB
Base plate RMS 3
Spare part for pipe mounting system RMS 3
ZUB0 RMS3 BSB
Base plate RMS 5
Spare part for pipe mounting system RMS 5
ZUB0 RMS5 BSB
Completion plate
ZUB0 RMS2 EB
RMS 2 completion plate for parallel installation of POA,
CS2 and KDA wedge sensors and air-ultrasonic sensors,
suitable for pipelines starting at DN250
The accessories are optionally available.
More accessories for our pipe mounting systems can be found in our current price list.
9 Dismantling/Disposal

Remove sensors from base plate.

Remove wedge supports from the RMS 5 base plate.

The stainless steel sheets and plates of the pipe mounting systems can be recycled
as regular scrap metal.
Pipe mounting systems / Rev. 00 / 21.04.2017
page 31

advertisement

Key Features

  • Versatile pipe diameter compatibility
  • Stainless steel construction
  • Modular design for various sensors
  • Portable flow measurement solutions

Frequently Answers and Questions

What tools are needed to install the RMS 2?
A 300g hammer, cable ties, a 13mm ring or open-end spanner, a Pozidriv size 2 Phillips screwdriver, and wire cutters are required.
What sensors are compatible with the RMS systems?
The systems are compatible with CS2, CSM, KDA, POA, DSM, and OCL sensors.
Can RMS components be interchanged between different RMS sets?
RMS 2, 3, and 4 components are compatible; however, RMS 5 uses thicker materials and is not compatible with other sets.
What is the material of the pipe mounting systems?
All pipe mounting systems are made of stainless steel 1.4404.
How should I handle the RMS 5 assembly?
Due to the size of RMS 5 components, NIVUS recommends two people for assembly and installation.