LEICA DM IRE2 MICROSCOPE MANUAL Neuroscience Imaging Core Rightmire Hall Ohio State University Director: Tony Brown Rightmire 060 292-1205 brown.2302@osu.edu Facility Manager: Paula Monsma Rightmire 062 292-3025 monsma.1@osu.edu This manual prepared by Tony Brown. 1 CMN Imaging Core January 16, 2013 TABLE OF CONTENTS LEICA DM IRE2 MICROSCOPE MANUAL ................................ 1 Getting to know the microscope .............................................. 3 Microscope control panel ......................................................... 7 LCD display panel ...................................................................... 8 LCD display panel buttons ....................................................... 9 Changing filters ....................................................................... 10 Changing objectives................................................................ 11 To switch between dry and immersion objectives ............... 12 Adjust halogen lamp brightness ............................................ 13 To focus using the focusing knobs ....................................... 14 Using the focusing buttons .................................................... 15 Changing the coarseness of the focusing ............................ 16 Eyepieces ................................................................................. 17 Placing a slide on the Z-galvo stage...................................... 17 Transmitted light detector selection knob ............................ 18 Bright field observation .......................................................... 18 Koehler illumination ................................................................ 19 Phase contrast ......................................................................... 20 Differential interference contrast (DIC) ................................. 21 Switching stage holders ......................................................... 22 Using immersion objectives ................................................... 23 Cleaning objectives ................................................................. 24 Applying oil to objective without removing slide ................. 25 Specimen preparation ............................................................. 26 DOs and DON’Ts ...................................................................... 27 2 CMN Imaging Core January 16, 2013 Getting to know the microscope VIEW FROM LEFT Transmitted light illumination column (tilts back for better access to stage) Transmitted light detector Eyepieces Condenser lens Z- galvo stage (detachable) Filter cube turret Laser scan head with 3 fluorescence detectors Side port Focus knob Halogen lamp intensity adjustment wheel 3 Microscope control panel CMN Imaging Core January 16, 2013 VIEW FROM RIGHT Halogen lamp housing (transmitted light) Filter holders (empty) Transmitted light detector selection knob Condenser turret Mercury lamp housing (epi-fluorescence) Objective turret Tube lens module with Bertrand lens Focus knob X-Y stage movement 4 CMN Imaging Core January 16, 2013 OBJECTIVE TURRET (DETAILED VIEW) objective prism turret objective turret Epifluorescence Illumination diaphragm (controls brightness of illumination) Epi-fluorescence field diaphragm (controls area of illumination) filter cube turret cuvesfou 5 DIC analyzer CMN Imaging Core January 16, 2013 OBJECTIVE PRISM TURRET POSITIONS Objective turret position Description Matching objectives BF Empty (for bright field or phase contrast) C DIC objective prism D DIC objective prism E DIC objective prism 20x multi-immersion 63x water immersion 100x oil immersion 40x oil immersion 63x oil immersion CONDENSER TURRET POSITIONS Condenser turret position Description Matching objectives BF Empty (for bright field) PH1 phase ring 10x dry PH2 phase ring 40x dry 20 40/63 63/100 DIC condenser prism DIC condenser prism DIC condenser prism 6 20x mulit-immersion 40x oil immersion 63x water immersion 63x oil immersion 100x oil immersion CMN Imaging Core January 16, 2013 Microscope control panel LCD display LCD display panel buttons Red light shows current filter selection switch filter cube GFP: green fluorochromes RED: red fluorochromes CFP/: far red fluorochromes SCAN=empty slot select port VIS=light to eyepieces SIDE=light to scanner BOTTOM: disabled Mercury lamp shutter (red light is on when shutter is closed) 7 switch tube lens (disabled) CMN Imaging Core January 16, 2013 LCD display panel arrow indicates lower limit has been set Indicates current step size setting for focusing Indicates position of objective relative to upper limit Indicates current objective Indicates desired DIC objective prism arrow indicates upper limit has been set Indicates current filter cube D= dry mode I= immersion mode Indicates actual DIC objective prism in light path 8 CMN Imaging Core January 16, 2013 LCD display panel buttons Learn mode (DO NOT USE!) Toggle between voltage readout and objective readout Set lower limit of travel for objective (DO NOT USE!) Set upper limit of travel for objective (DO NOT USE!) Adjust step size for coarseness of focusing Do not press the LEARN button (this will enter you into learn mode, which you must not play with). If you accidentally press this button, you will see the word “EXIT “ flashing. Press the LEARN button again to exit the learn mode. 9 CMN Imaging Core January 16, 2013 Changing filters Use the motorized fluorescent filter cube changer on the microscope control panel: Press left button to rotate filter cube turret left Press right button to rotate filter cube turret right Red light shows current filter cube selection The filter cube turret contains three filters plus an empty slot GFP : cube for green fluorochromes RED : cube for red fluorochromes CFP/ : cube for far red fluorochromes or cube for CFP (see Paula for details) SCAN : empty slot Notes: The cube in the CFP/ position is normally the far red cube, suitable for fluorochromes such as TOTO-3 and Cy5 The SCAN position is used for confocal imaging and for transmitted light observation through the eyepieces 10 CMN Imaging Core January 16, 2013 Changing objectives Use the objective turret control buttons on the left side of the microscope Press upper key to increase magnification Press lower key to decrease magnification Current objective indicated on LCD display panel 11 CMN Imaging Core January 16, 2013 To switch between dry and immersion objectives The standard objectives on our microscope are grouped into three blocks or modes: 100x oil immersion 63x oil immersion oil immersion mode 40x oil immersion 20x oil/water/glycerol immersion 40x dry 10x dry multi immersion mode dry mode To reduce the chance of immersion medium getting on to a dry objective or the chance of mixing of different immersion media, the microscope will not allow you to move freely between these modes using the objective turret control buttons. To switch from one mode to another: simultaneously press the “upper limit” and “lower limit” buttons on the microscope control panel the words “CHANGE OBJECTIVE” flash on the LCD display panel now you can change the objective using the objective turret control buttons Press these two buttons simultaneously 12 CMN Imaging Core January 16, 2013 Adjust halogen lamp brightness Use the dial on the front left side of the microscope stand near the base Halogen lamp brightness control The lamp voltage will display automatically on the microscope control panel display when the intensity dial is adjusted To switch off transmitted light illumination adjust lamp intensity to 2.5V then continue rotating dial beyond this point 0V on control panel display indicates illumination is off To switch on transmitted light illumination, rotate briefly in the opposite direction 13 CMN Imaging Core January 16, 2013 To focus using the focusing knobs One way to focus is using the focusing knobs located on the left and right side of the microscope. Turning the knob so that your thumb moves away from you focuses down Turning the knob so that your thumb moves toward from you focuses up Focusing knobs (turn in direction of arrows to focus objective down) 14 CMN Imaging Core January 16, 2013 Using the focusing buttons Another way to focus is to use the focusing buttons located on the right side of the microscope Press upper focus key to focus up Press lower focus key to focus down Note about upper limits: If an upper limit is set for the objective (see LCD display panel) then the objective will not move above that limit if you are focusing using the focusing buttons. The only way to focus above the upper limit is to use the focusing knobs This is a safety feature to prevent accidental damage of the objective when focusing using the focusing buttons. 15 CMN Imaging Core January 16, 2013 Changing the coarseness of the focusing The focusing is electronic and has five settings: Setting Step size S0 S1 Fine Medium fine S2 Medium coarse S3 Coarse You can use any step size with any objective, but when you first select an objective the default step size will be as follows: Objective Setting 63,100x 40x S0 S1 20x S2 10x S3 Press STEP button to switch between S0, S1, S2 and S3 STEP button 16 CMN Imaging Core January 16, 2013 Eyepieces Adjusting interpupillary distance: Adjust eyepieces to match interpupillary distance by moving the eyepieces closer together or further apart Adjusting parfocality Focus on specimen using electronic focusing controls Close right eye and adjust the left eyepiece so that the image appears in focus to your left eye Close the left eye and adjust the right eyepiece so that the image appears in focus to your right eye The eyepieces are now parfocal Placing a slide on the Z-galvo stage To avoid touching or damaging the objective, lower the objective turret all the way using the focusing buttons 2 2 1 1 1 Position the slide holder clips as shown above Insert slide into the holder in a front-to-back motion (1) Slide clips inward onto slide to secure the slide (2) 1 17 CMN Imaging Core January 16, 2013 Transmitted light detector selection knob For any transmitted light observation (bright field, DIC, phase contrast), this knob should be in the vis position. Bright field observation Bright field observation means observation with transmitted light using no contrast enhancement method (e.g. no phase contrast or DIC) For bright field observation: rotate condenser turret to BF (empty) position rotate objective prism turret to BF (empty) position switch filter cube turret to SCAN (empty) position Condenser turret in bright field (BF) position 18 CMN Imaging Core January 16, 2013 Koehler illumination Koehler illumination is essential to obtain good transmitted light images Select objective Open the condenser aperture diaphragm (move lever to the right) Focus on specimen Close the field aperture diaphragm (move lever to left) Focus condenser using condenser focusing knob until the image of the aperture is sharp If necessary center the field diaphragm in the field of view using the two centering screws located on the front of the condenser Open field aperture diaphragm until it just disappears from the field of view Close down the condenser aperture diaphragm until the desired contrast is achieved Field aperture diaphragm centering screws Condenser focusing knob DIC polarizer Condenser aperture diaphragm Condenser turret Phase ring centering keys insert here 19 CMN Imaging Core January 16, 2013 Phase contrast For phase contrast, you need: a phase contrast objective a matching phase ring in the condenser turret To set up phase contrast: Select a phase contrast objective (i.e. either of the dry objectives) rotate objective prism turret to BF (empty) position Rotate condenser turret to select phase ring PH1 for 10x PH2 for 40x Focus on specimen Insert the Bertrand lens into the optical path using by rotating the tube lens module from “SCAN” to “B” Focus on the phase ring using the Bertrand lens focus slider on the tube lens module If condenser phase ring (dark ring) is not centered on the objective phase ring (light ring), center it using the insertable phase ring centering keys (see Paula for instructions) Set Koehler illumination Tube lens module in B position Bertrand lens focus slider 20 CMN Imaging Core January 16, 2013 Differential interference contrast (DIC) For DIC, you need: A DIC objective (i.e. any of the immersion objectives) A polarizer above the condenser (no need to insert this – it is kept in the light path at all times) A condenser prism in the condenser turret An analyzer beneath the objective prism turret An objective prism in the objective prism turret To set up DIC: Select a DIC objective (any of the immersion objectives) The objective DIC prism required for that objective is displayed on the microscope control panel Manually rotate condenser turret to select the appropriate objective DIC prism Condenser prism 63/100 63/100 Objective prism D E 63x water 40/63 D 40x 20x multi-immersion 40/63 20 E C Objective 100x oil 63x oil Manually rotate objective prism turret to select objective prism Insert analyzer into light path 21 CMN Imaging Core January 16, 2013 Switching stage holders The microscope has two stage holders: Z- galvo stage (permits rapid and precise movement in Z axis for acquisition of Z stacks) Universal stage holder (holds a wider range of dishes) To remove Z-galvo stage: Unscrew the two thumbail screws (circled in red): Place stage (still attached to the cable) on top of the laser scan head as shown: Take care not to touch condenser lens surface when removing and replacing the stages! 22 CMN Imaging Core January 16, 2013 Using immersion objectives OIL immersion: Clean coverslips before using immersion objectives Apply oil to the objective or to the coverslip before placing your slide on the stage Use only LEICA IMMERSION OIL Use the minimum amount of oil necessary If you use too much oil it may run down the side of the objective and damage the optics! To apply oil to an objective: dip oil applicator in immersion oil bottle allow excess to drain off apply oil to objective by touching to the metal next to the lens DO NOT TOUCH THE LENS DIRECTLY! WATER immersion: The 63x water immersion objective (not normally installed on the microscope) uses water rather than oil as the immersion medium. Contact Paula for instructions on using this objective. 23 CMN Imaging Core January 16, 2013 Cleaning objectives This microscope is equipped with the highest quality objectives, with a total value exceeding $30,000! You must exercise great care to preserve these objectives To clean oil off immersion objectives: Remove specimen from stage Blot (not wipe!) off excess oil using lens tissue Use a fresh area of the lens tissue for each blot Repeat blotting until no more oil comes off onto the lens tissue Wipe the metal housing around the lens using lens tissue It typically takes 2 or 3 sheets of lens tissue to properly clean a single objective What to do if you accidentally get oil on a dry objective: Blot off excess with lens tissue Then contact Paula, who will clean the lens with special cleaning fluid (do not attempt this yourself!) Lens tissue alone will not get all the oil off the lens and this will interfere with the optics 24 CMN Imaging Core January 16, 2013 Applying oil to objective without removing slide FOR ADVANCED USERS ONLY! You can use the following procedure if you are observing your slide and you wish to switch to an oil immersion objective without removing the slide from the stage holder. Focus objective turret all the way down using the focus down button In the LCS software, click on obj button and select the objective diametrically opposite in the objective turret from the objective you want to apply oil to (i.e. 3 positions away from the current objective) Objective you want to use 100x oil 63x oil Objective in opposite position 20x multi-immersion 40x dry 40x oil 10x dry 20x multi-immersion 40x dry 10x dry 100x oil 63x oil 40x oil This will rotate the objective that you want to apply oil to so that it is accessible from the right of the microscope Apply oil to the objective In the LCS software, click on obj button and select the objective that you applied oil to This will rotate the objective back into the light path Focus up on your specimen 25 CMN Imaging Core January 16, 2013 Specimen preparation Some general tips: Select fluorochromes that are optimally excited by the confocal laser lines For multiple labeling, the less overlap between the excitation and emission spectra the better Always use #1.5 coverslips Mount coverslip to slide securely and seal with nail enamel or mounting medium that solidifies Do not observe on the microscope until the nail enamel or mounting medium has hardened! Secure the slide with the clips when acquiring Z series 26 CMN Imaging Core January 16, 2013 DOs and DON’Ts DOs Use only lens tissue to clean objectives Use Kimwipes to clean coverslips and slides DONTs Never touch any optical surface with anything other than lens tissue Never clean the objectives with any kind of solvent Never use Kimwipes to clean objectives or any other optics Never use any liquid cleaners or solvents to clean the microscope optics or any part of the microscope! When to ask for Paula’s assistance Changing mercury or halogen bulbs Aligning mercury and halogen lamps Adding or removing objectives Adding or removing filter cubes Using the 63x water immersion objective Cleaning oil off 10x and 40x dry objectives Using the Leica objective warmer Using the Bioptechs heated chamber Any other time you are unsure what you are doing! 27 CMN Imaging Core January 16, 2013 28 CMN Imaging Core January 16, 2013 Extras Setting Step size Objective S0 S1 Finest 63,100x 40x S2 20x S3 SC 10x Coarsest Note about the coarsest focusing setting (SC): Press both upper and lower focus keys simultaneously to switch to the coarsest focusing setting (SC) Press both focus keys simultaneously to switch back to fine focusing setting (S0-S3) 29 CMN Imaging Core January 16, 2013
* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project
advertising