1.13 Software and Firmware Upgrades
1.13.1 Viewing Software Revision
The software revisions running on the Druck DPI 620 Genii
can be viewed by selecting:
DASHBOARD >>
STATUS >> SOFTWARE BUILD
Note: If the software revision number is highlighted red then
an upgrade is available.
1.13.2 Upgrading the Software
Follow the website instructions to download the files onto
a USB flash memory device.
www.ge-mcs.com
1. Select:
DASHBOARD >>
ADVANCED
2. Enter the calibration PIN: 5487
3. Select the
button.
4. Select:
UPGRADE
5. Continue with one of these operations:
28 April 2014
K0541 issue 2
Page 26 of 200
User Manual
Druck DPI620 Genii
• Upgrade the Application software and SDC625
Application.
1. Copy the ‘AMC’ application folder into the root of
a USB flash memory device.
2. Put the USB flash memory device in the USB
type A connector.
3. Select:
APPLICATION
4. Follow the on-screen instructions.
Note: The SDC625 HART® Application can only be
upgraded as part of an application upgrade.
• Upgrade the Operating System and Bootloader
software.
1. Create a folder named ‘OS’ in the root of a USB
flash memory device.
2. Copy the files ‘DK418.nb0’ and ‘DK419.nb0’ into
the ‘OS’ folder.
3. Put the USB flash memory device in the USB
type A connector.
4. Select:
OPERATING SYSTEM
5. Follow the on-screen instructions.
Note: The bootloader can only be upgraded as part of
an operating system upgrade.
28 April 2014
K0541 issue 2
Page 27 of 200
User Manual
Druck DPI620 Genii
• Upgrade the HART processor Application and Boot
Loader
1. Create a folder named ‘HART’ in the root of a
USB flash memory device.
2. Copy the files ‘DK416.s19’ and ‘DK417.s19’ into
the ‘HART’ folder.
3. Put the USB flash memory device in the USB
type A connector.
4. Select:
HART APPLICATION
5. Follow the on-screen instructions.
Note: The HART bootloader can only be upgraded as
part of an HART application upgrade.
• Upgrade the CH1 FPGA
1. Create a folder named ‘FPGA’ in the root of a
USB flash memory device.
2. Copy the files ‘DK413.bin’’ into the ‘FPGA’ folder.
3. Put the USB flash memory device in the USB
type A connector.
4. Select:
CH1 FPGA
5. Follow the on-screen instructions.
Note: The CH2 cannot be remotely upgraded .
28 April 2014
K0541 issue 2
Page 28 of 200
User Manual
Druck DPI620 Genii
• Upgrade the HART Device Library
By default the HART device library is stored on the
micro SD Card.
1. Set the DPI620 Genii USB client port to Storage
Device mode by selecting:
DASHBOARD >>
DEVICES >> USB CLIENT PORT
2. Locate the self-extracting file
‘DPI620_DD_library_20**_*.exe’
3. Connect DPI620 Genii Client USB port to PC
USB port. Device will connect to PC as a
Removable Disk.
4. Run ‘DPI620_DD_library_20**_*.exe’ and
extract files to the Removable Disk (this will
take several minutes due to the large file size).
**_* indicates the DD release version from the
HART foundation.
The required directory structure on the micro SD
card is shown in Figure 1-7 Hart DD Directory
Structure.
28 April 2014
K0541 issue 2
Page 29 of 200
User Manual
Druck DPI620 Genii
Figure 1-7 Hart DD Directory Structure
Note:
• If a mistake is made during upgrade and there are
no files to upload, follow the on-screen instructions
and complete the procedure.
• When an upgrade completes normally, the initial
operation of the touch screen may be slower (a
period of approximately 30 seconds).
• To make sure the upgrade completed correctly,
use the Status menu.
28 April 2014
K0541 issue 2
Page 30 of 200