Single Phase or Three Phase Comparator
Primary Standard with Accuracy Class 0.01
COM1003/COM3003
The System
Concept
Comparator COM1003/COM3003 is part of the
ZERA high precision measuring instrument
series. To meet the requirements a single phase
COM1003 or a three phase COM3003 is
available.
As AC/DC transfer standard comparator serves
for testing current and voltage meters as well
as single phase or poly phase* power and
energy meter testing systems in metrological
institutes, official testing places and other
testing fields of energy suppliers or
manufacturers of electricity meters.
Via soft keys, 6.4 colour display and
integrated user software MT3701 inputs,
measurements and visualisations will be
performed during the operation. An additional
external control including evaluation and report
generation is achieved via windows based user
software SSM3000 and external Windows PC.
3
1
2
Front view COM3003
Three-phase Comparator
1 USB connection
2 Softkeys
3 Alphanumeric keypad, Enter and cursor keys
* only COM3003
Features
-
High accuracy (class 0.01), independent of
measuring mode
Excellent long-term stability by using of DCcapable current transformers
Recalibration period by PTB can possibly be
set for > 2 years
RS232 and IEEE 488 interface
SCPI compatible IEEE488 interface commands
Automatic measuring range selection
Only one current input for the complete
measuring range
Direct traceability of measuring accuracy by
connection of DC- and frequency standard
devices
Functions
The following functions are available via
softkeys:
- Actual values
- Vectorial diagram
- Curve diagram
- Harmonic measurement
- Error measurement
- Reference measurement
User Software
Data management
Individual functions of the integrated user
software are described in the following.
Measurement data can be stored on an
especially configured USB stick.
For evaluation and reporting of the measurement results data management software MTVis
can be used on external PC.
COM3003 with suitcase
1
2 3
6
7
4
5
Rear side
COM3003
1 Interface: RS232, Remote control, IEEE, VGA and
external keypad
2 Scanning head input f. comparative P measurement
3 Quartz output for integration time
4 Power proportional pulse outputs (2x)
Pulse inputs for comparative P measurement (1x)
5 1 V and 10 V DC reference input
6 Mains supply 115/230 V
7 Current and voltage inputs
Display with softkeys
COM1003/COM3003
Actual Values
Measurement
The following system parameters are displayed
as averages values over an adjustment
integration time:
- RMS values of phase voltages and currents
and their DC component
- All angles between currents and voltages
calculated from the fundamental components
- Active, reactive and apparent power, per
phase or total
- Frequency and direction of rotating field
Vector Display
Actual values can be displayed in table form or
as vectorial graphic.
Waveform Display
The waveforms of voltage and currents can be
measured and displayed. The user can choose
between display as curves with indication of
individual values and harmonic displaying with
individual distortion values.
Error Measurement
Static and electromechanical power meters as
well as all kind of measuring instruments with
power proportional frequency output can be
tested in the menu “Error measurement”.
The user can select between scanning head
input or frequency input.
Harmonic
Measurement
Harmonic spectrum measurement in voltage
and current up to the 40th THD conforming to
EN50160. The harmonic spectrum can be
displayed in a chart or in a diagram. All
measured harmonic values can be stored
according to the customer information data on
USB stick. The system has also the ability to
scan the measured harmonic by using a cursor
and to display the scanned values referring to
the cursor position on the screen.
COM1003/COM3003
Technical Data
Comparator
COM1003
COM3003
General
Power supply
Power consumption
Temperature range
Relative humidity (not condensing)
Dimensions (HxWxD)
Weight
Max. height above sea level
Reference Meter
Voltage measurement
Voltage ranges
Current measurement
Current ranges
Reference voltage ranges
Fundamental frequency
Bandwidth
Measuring modes
Accuracy class rating according to PTB for
measuring power and energy 1
115/230 V + 10 % -15 %, 50…60 Hz
approx. 80 VA
15° … 40° C
max. 95 %
172 x 465 x 460 mm
16 kg
2000 m
115/230 V + 10 % -15 %, 50…60 Hz
approx. 120 VA
15° … 40° C
max. 95 %
172 x 465 x 460 mm
25 kg
2000 m
30 V … 500 V
60 - 120 - 240 - 480 V
1 mA … 160 A
5-10-20-50-100-200-500 mA
1-2-5-10-20-50-100-200 A
1 V and 10 V DC
15 … 70 Hz
DC … 3500 Hz
2 wire active and reactive
2 wire DC 3
mixed signals (AC+DC) in all
measuring modes
< 100 x 10-6
30 V … 500 V
60 - 120 - 240 - 480 V
1 mA … 160 A
5-10-20-50-100-200-500 mA
1-2-5-10-20-50-100-200 A
1 V and 10 V DC
15 … 70 Hz
DC … 3500 Hz
4 wire active and apparent, reactive
true, cross and Q60
4 wire DC 3
3 wire active and reactive true, cross
connected A and B
2 wire active and reactive
2 wire DC 3
mixed signals (AC+DC) in all
measuring modes
< 100 x 10-6
> 2 years
< 30 x 10-6
< 50 x 10-6 (30 V … 500 V)
< 15 x 10-6 / year
< 50 x 10-6 (50 mA … 160 A)
< 70 x 10-6 (10 mA … < 50 mA)
< 150 x 10-6 (1 mA … <10 mA)
< 350 x 10-6 (50 mA … 160 A)
< 25 x 10-6 / year
< 80 x 10-6 (50 mA … 160 A)
< 100 x 10-6 (10 mA … < 50 mA)
> 2 years
< 30 x 10-6
< 50 x 10-6 (30 V … 500 V)
< 15 x 10-6 / year
< 50 x 10-6 (50 mA … 160 A)
< 70 x 10-6 (10 mA … < 50 mA)
< 150 x 10-6 (1 mA … < 10 mA)
< 350 x 10-6 (50 mA … 160 A)
< 25 x 10-6 / year
< 80 x 10-6 (50 mA … 160 A)
< 100 x 10-6 (10 mA … < 50 mA)
< 180 x 10-6 (1 mA … < 10 mA) 7
< 400 x 10-6 (50 mA … 160 A)
< 30 x 10-6 / year
< 180 x 10-6 (1 mA … < 10 mA) 7
< 400 x 10-6 (50 mA … 160 A)
< 30 x 10-6 / year
< 0.005°
< 20 x 10-6
< 0.005°
< 20 x 10-6
< 5 x 10-6 / year
< 5 x 10-6 / year
U <
0.5 x 10-6 / K
I <
0.5 x 10-6 / K
P <
1 x 10-6 / K
DC reference input < 1 x 10-6 / K
Quartz time base < 0.1 x 10-6 / K
U<
0.5 x 10-6 / K
I <
0.5 x 10-6 / K
P <
1 x 10-6 / K
DC reference input < 1 x 10-6 / K
Quartz time base < 0.1 x 10-6 / K
Independent of measuring mode
Recalibration period according to PTB 4
Voltage measurement accuracy
Voltage measurement accuracy DC 3
Voltage measurement long term stability
Current measurement accuracy
Current measurement accuracy DC 3
Current measurement long term stability
Power/energy measurement accuracy 2
Independent of measuring mode
Power/energy measurement accuracy DC
Power/energy measurement long term
stability
Phase angle measurement error
DC reference voltage measurement
accuracy
DC reference voltage measurement long
term stability
Temperature drift
3
Status:4th February 2016
ZERA GmbH
Hauptstraße 392
53639 Königswinter
Germany
Phone: +49 (0) 2223 704-0
Fax: +49 (0) 22 23 704-70
Email: zera@zera.de
Web: www.zera.de
1 Within the range of 30 V … 500 V and 50 mA … 160 A
2 Related to the active power with power factors from 1 to 0.5 i and from 1 to 0.8 c in the range of 30 V … 500 V
3 Only with option DC measurement
4 Period defined by PTB (national metrological authority), corresponding operation time and stability of the device provided.
5 Up to the 40th harmonic for current and voltage
7 Related to 10 mA
Subjects to alteration.