Rack, for 50 ml tubes Rack, for 250 ml bottles

Rack, for 50 ml tubes Rack, for 250 ml bottles
GE Healthcare
Life Sciences
Instructions 28-9904-45 AB
ÄKTA™ avant
Rack, for 50 ml tubes
Rack, for 250 ml bottles
Scope
This document contains information about Rack for 50 ml tubes and Rack for 250 ml
bottles, for use with ÄKTA avant 25 and 150:
•
product description
•
tube information for Rack for 50 ml tubes
•
bottle information for Rack for 250 ml bottles
•
required software updates
•
preparing the Fraction collector
•
cleaning
Software updates
Before a new rack is used the first time, ensure that the latest software is installed:
•
UNICORN™ 6.1 or later. Contact your local GE Healthcare representantive for software
update.
•
ÄKTA avant Instrument Configurations:
- version 2.1, or later for ÄKTA avant 25
- version 1.1, or later for ÄKTA avant 150
These are mandatory when using Rack for 250 ml bottles, and recommended when
using Rack for 50 ml tubes. See: www.gelifesciences.com/AKTAavant-software
Description
Rack for 50 ml tubes:
Rack for 250 ml bottles:
Code number: 28-9803-19
Code number: 28-9818-73
3
1
4
4
3
1
2
Part
2
2
Name
Description
Rack for 50 ml
tubes:
Rack for 250 ml
bottles:
1
Compartments for
tubes/bottles.
Holds 55 tubes.
Holds 18 bottles.
2
Numbers for identifying
tube/bottle position.
Numbers 1 to 11
written on the top.
Numbers 1 to 6
written on the side.
3
Letters for identifying
tube/bottle position.
Letters A to E
written on the top.
Letters A to C
written on the side.
4
Type code.
Used by the code reader in the Fraction
collector to determine the type of
cassette/rack.
Note: The type code is fragile and must be
handled with care.
ÄKTA avant Instructions 28-9904-45 AB
Tubes and bottles
Max
volume
Size [mm]
Height [mm]
Examples of
manufacturers
Min
Max
Min
Max
50 ml
Ø: 28
Ø: 30
110
116
BD Biosciences,
Falcon™ Conical Tubes 50 ml
250 ml
L: 55
L: 64.5
121
Nalgene, 2110-0008
W: 55
W: 64.0
See
note
below
Kautex, 303-770531
Note: If using other tubes or bottles, ensure that they fulfil the requirements above.
Also make sure they fit properly into the rack, can not tip, stand steadily on the
bottom plate, and have a wide opening.
Fraction collector
The maximum flow rate is 150 ml/min for the Fraction collector when using 50 ml tubes
or 250 ml bottles.
Note: The Cassette tray shall not be used in the Frac drawer when using Rack for 50 ml
tubes or Rack for 250 ml bottles. A cassette tray is integrated in their design.
For more information about the Fraction collector, see ÄKTA avant and UNICORN 6.1
User Manual:
•
section Prepare the Fraction collector, regarding how to place the racks.
•
section Maintenance/Maintenance when required/Clean the Fraction collector,
regarding cleaning.
Cleaning the rack
When required, clean the rack with water (max 60°C) and a mild cleaning agent.
Note: The type code must be handled with care when the rack is cleaned.
Note: It is not recommended to disassemble the tray for the cleaning.
ÄKTA avant Instructions 28-9904-45 AB
3
For local office contact information,
visit
www.gelifesciences.com/contact
GE Healthcare Bio-Sciences AB
Björkgatan 30
751 84 Uppsala
Sweden
www.gelifesciences.com/AKTA
GE, imagination at work and GE monogram are trademarks of General
Electric Company.
ÄKTA and UNICORN are a trademarks of GE Healthcare companies.
© 2010-2011 General Electric Company – All rights reserved.
First published Feb. 2010
All goods and services are sold subject to the terms and conditions of
sale of the company within GE Healthcare which supplies them. A copy
of these terms and conditions is available on request. Contact your
local GE Healthcare representative for the most current information.
UNICORN: Any use of this software is subject to GE Healthcare Standard
Software End-User License Agreement for Life Sciences Software
Products.
GE Healthcare Europe GmbH
Munzinger Strasse 5, D-79111 Freiburg, Germany
GE Healthcare UK Limited
Amersham Place, Little Chalfont, Buckinghamshire, HP7 9NA, UK
GE Healthcare Bio-Sciences Corp.
800 Centennial Avenue, P.O. Box 1327, Piscataway, NJ 08855-1327,
USA
GE Healthcare Japan Corporation
Sanken Bldg.3-25-1, Hyakunincho Shinjuku-ku, Tokyo 169-0073, Japan
imagination at work
28-9904-45 AB 04/2011