Solu Comp® II Dual Input Analyzer

Solu Comp® II Dual Input Analyzer
Product Data Sheet
PDS 71-1055/rev.O
February 2006
Model 1055
Solu Comp® II Dual Input Analyzer
• BROAD SELECTION OF MEASUREMENT CHOICES includes
pH/ORP, Resistivity/Conductivity, % Concentration, Total and Free
Chlorine, Dissolved Oxygen, Dissolved Ozone, Flow, and
Temperature.
• SINGLE OR DUAL MEASUREMENT with dual 4-20 mA outputs.
• FULL COMPLEMENT OF MEASUREMENT COMBINATIONS CAN
BE COMMISSIONED IN THE FIELD.
• THREE FULLY PROGRAMMABLE ALARMS.
• CLEAR, EASY-TO-READ, two-line, back-lit display easily customized to read in English,
French, German, Italian, Spanish, or Portuguese.
• CHOICE OF ENCLOSURES for pipe, surface, and panel mounting meet NEMA 4X/CSA 4
(IP 65) requirements.
FEATURES AND APPLICATIONS
The Solu Comp II analyzers offer the choice of single
or dual sensor input with measurement choices of
pH/ORP, resistivity/conductivity/TDS, % concentration, ratio conductivity, total and free chlorine, dissolved oxygen, dissolved ozone, flow and temperature. Dual measurement analyzers offer a wide
choice of measurement combinations thus reducing
the cost per loop and needed panel space.
FIELD COMMISSION OPTION: The Solu Comp II
can be ordered with the ability to commission measurements in the field. This added flexibility can greatly
reduce the number of spare instruments required for
field servicing.
QUICK START PROGRAMMING: Exclusive Quick
Start screens appear the first time the Solu Comp II is
powered up. Screen prompts direct the user to register the number of sensors, the measurement unit(s)
and the language to display. Some measurement
specific prompts are also displayed. The measurement loop is ready for use in a matter of minutes.
MENUS: Menu screens for calibrating and registering
choices are simple and intuitive. Plain language
prompts guide the user through the procedures.
There are no service codes to enter before gaining
access to menus.
DUAL SENSOR INPUT AND OUTPUT: The Solu
Comp II accepts single or dual sensor input. The two
4-20 mA outputs can be independently programmed
to correspond to any selected measurement or temperature. Output damping and linear or log output
may also be field selected.
ALARMS: The Solu Comp II has three fully programmable alarm relays that can be assigned to any
selected measurement or temperature. Alarms can be
configured as high, low, or USP1. The third relay has
the additional choice of fault alarm operation. When
selected, a fault alarm will activate the relay when a
sensor or analyzer fault occurs.
ENCLOSURE: The panel mount version fits standard
½ DIN panel cutouts, and its shallow depth is ideally
suited for easy mounting in Hoffman-type enclosures.
A panel mount gasket is included to maintain the
weather rating of the panel. Surface/pipe mount
enclosure includes self-tapping screws for surface
mounting. A pipe mounting accessory kit is available
for mounting to a 2-inch pipe.
1USP alarm applies to conductivity/resistivity only.
Continued on page 2
DISPLAY: The two-line, 16-character, back-lit display
can be customized to meet user requirements. All
operations and descriptive messages can be field
selected for English, French, German, Italian,
Spanish, or Portuguese. Informative screens, which
permit data not shown in the regular display, may be
seen at the push of a button.
TEMPERATURE: Most measurements (except ORP
and flow) require temperature compensation. The
Solu Comp II will automatically recognize either a
Pt100 or Pt1000 RTD, normally built into the sensor.
When this RTD is present, the Solu Comp II can be
set up to display the temperature in °C or °F as well
as set any one or more of the alarms and/or outputs
to respond to this sensor input. If two measurements
with temperature are present either can be chosen for
each alarm and output selected.
SPECIFICATIONS - General
Code -68 only:
Case: ABS (panel mount), polycarbonate (surface/pipe
mount). All versions are NEMA 4X/CSA 4 (IP65).
Dimensions
Panel (code -10): 6.10 x 6.10 x 3.72 in. (155 x
155 x 94.5 mm)
Surface/Pipe (code -11): 6.23 x 6.23 x 3.23 in.
(158 x 158 x 82 mm); see page 5 for dimensions
of pipe mounting bracket.
Conduit Openings: Accepts PG13.5 or 1/2 in. conduit fittings
Display: Two line, 16-character, back-lit display.
Character height: 4.8 mm. Display can be customized to meet individual requirements.
Depending on number of sensors, as many as 14
display screens are available.
Ambient Temperature and Humidity: 0 to 50°C, (32
to 122°F) RH 5 to 95% (non-condensing)
Note: The analyzer is operable from -20 to 60°C
(-4 to 140°F) with some degradation in display
performance.
Ambient Temperature Effect: ±0.05% of reading/ °C
Power:
Code -01: 115/230 Vac ±15%, 50/60 Hz ±6%, 8.0 W
Code -02*: 24 Vdc ±15%, 6.0 W
Installation Category II
* For +24Vdc Power Supply use only devices meeting
NEC Class II or UL recognized (UL 1950).
Equipment protected throughout by double insulation.
Hazardous Location:
Class I, Division 2,
Groups A, B, C, & D
12RN
POLLUTION DEGREE 2: Normally only non-conductive pollution occurs. Occasionally, however, a
temporary conductivity caused by condensation
must be expected.
Input: Choice of single or dual sensor input with
measurement choices of pH/ORP, conductivity/
resistivity, toroidal conductivity, flow, chlorine, dissolved oxygen, and dissolved ozone. Field-commissioned units allow user to change measurements on either or both inputs. See combination
guide for valid combinations. For contacting conductivity measurements, temperature element
must be a Pt 1000 RTD. For other measurements, use either a Pt100 RTD, Pt1000 RTD, or
22k NTC (D.O. only).
Outputs: Two 4-20 mA or 0-20 mA isolated outputs.
Continuously adjustable. Linear or logarithmic.
Maximum load 500 ohms. Output dampening with
time constant of 5 sec is user-selectable.
Analog Output Accuracy: ±0.05 mA
Alarms: Three alarm relays for process measurement(s) or temperature. Alarm 3 can be
configured as a fault alarm, instead of a
process alarm. Each relay can be configured
independently. Alarm logic (high or low activation
or USP*) and deadband are user-programmable.
The USP* alarm can be programmed to activate
when the conductivity is within a user-selectable
percentage of the limit.
*conductivity/resistivity measurement only
Relays: Form C, single pole double throw, epoxy sealed
POLLUTION DEGREE 4: Extended Environment
Outdoor use where conductive contamination
such as rain, snow, or dust may be present.
(Hazardous Location only)
RFI/EMI:
LVD:
EN-61326
EN-61010-1
28 Vdc
115 Vac
230 Vac
Resistive
5.0 A
5.0 A
5.0 A
Inductive
3.0 A
3.0 A
1.5 A
Terminal Connections Rating: 26-14 AWG wire size
Weight/Shipping weight (rounded up to nearest lb or
nearest 0.5 kg): 3 lb (1.5 kg)/4 lb (2.0 kg)
2 Rosemount Analytical
MILLIMETER
INCH
PANEL MOUNT INSTALLATION
Access to the wiring terminals is through the rear cover. Four screws hold the cover in place.
Rosemount Analytical 3
MILLIMETER
INCH
SURFACE MOUNT INSTALLATION
The front panel is hinged at the bottom. The panel swings down for access to the wiring terminals.
4 Rosemount Analytical
MILLIMETER
INCH
PIPE MOUNT INSTALLATION
The front panel is hinged at the bottom. The panel swings down for access to the wiring terminals.
Rosemount Analytical 5
CONTACTING CONDUCTIVITY (Codes -20 and/or -30)
Measures conductivity in the range 0 to 20,000 µS/cm.
Display choices are conductivity, resistivity, and TDS
(total dissolved solids). Three temperature corrections
are available: high purity water (dilute sodium chloride), cation conductivity (dilute hydrochloric acid),
and adjustable linear temperature coefficient (0 to
5.00%/°C). Temperature correction can be disabled,
allowing the analyzer to display raw conductivity.
RECOMMENDED SENSORS FOR
CONDUCTIVITY:
The Solu Comp II is intended for use with the
ENDURANCE Model 400 series conductivity sensor
(Pt 1000 RTD).
Model 400 Screw-in/Insertion
Model 400VP Screw-in/Insertion with 6.0 VP connector
Model 401 Screw-in/Insertion (except 401-15)
PERFORMANCE SPECIFICATIONS Range
Cell constant (/cm)
Accuracy1,2
0.055 - 9.99 µS/cm
0.01
0.9% of reading
or ±0.002 µS/cm
10 - 50 µS/cm
0.01
±2% of reading
0.055 - 500 µS/cm
0.1
±2% of reading
or ±0.1 µS/cm
0.055 - 5000 µS/cm
1.0
±2% of reading
or ±1 µS/cm
0 - 5 mS/cm
1.0
±2% of reading
or ±0.001 mS/cm
0 - 20 mS/cm
10
±2% of reading
or ±0.01 mS/cm
ANALYZER (CONDUCTIVITY)
Accuracy (Resistivity)2: 0.9% of reading
Accuracy (Temperature)2: ±0.1°C between 5°C and
100°C; ±1°C between 101°C and 200°C
Stability: 0.5% of reading/month
Temperature Correction: High purity water (dilute
sodium chloride), cation conductivity (dilute
hydrochloric acid), linear temperature coefficient
(0.0 to 5.00%/°C), or none. High purity water and
cation conductivity temperature correction apply
between 0 and 100°C. Linear temperature coefficient can be applied between -5 and 200°C.
Measurement Range: 0.0 to 20,000 µS/cm, 0.05 to
20 MΩ-cm, or 0 to 10,000 ppm TDS
Model 402 Retractable
Model 402VP Retractable with 6.0 VP connector
Model 403 Sanitary Flanged
Model 403VP Sanitary Flanged with 6.0 VP connector
Model 404 Flow-Through
The analyzer can also be used with Rosemount
Analytical conductivity sensor Models 140, 141, 142, and
150 having a Pt 100 RTD.
Ratio Conductivity (Codes -20-30):
The dual conductivity Solu Comp II can function as a
ratio analyzer or recovery device (% passage or %
rejection). Product sensor 2’s conductivity reading is
always displayed.
Ratio
Ratio
S2
.3325
4.621 µS/cm
%Pass
%Passage
S2
12.1
4.621 µS/cm
Temperature Range: -5°C to 200°C (23°F to 392°F)
1
2
12.34 µS/cm
40.3 C
7.34pH
25.3 C
whichever is greater
Accuracy values pertain to Endurance Model 400
Series conductivity sensors only
6 Rosemount Analytical
%Reject
%Reject
S2
87.9
4.621 µS/cm
TOROIDAL CONDUCTIVITY (Codes -21 and/or -31)
When used with Model Series 200 Toroidal
Conductivity Sensors, display choices are conductivity, resistivity, and percent concentration. The percent concentration selection includes the choice of
four common solutions (0-12% NaOH, 0-15% HCl,
and 0-25% or 96-99.7% H2SO4). The conductivityconcentration algorithms for these solutions are fully
temperature compensated. For other solutions, a
simple-to-use menu allows the customer to enter his
own data. The analyzer accepts as many as five (5)
data points and fits either a linear (two [2] points) or
a quadratic function (three [3] or more points) to the
data. Reference temperature and linear temperature
slope may also be adjusted for optimum results.
Model 222 Flow-through conductivity sensor
Model 225 Clean-in-place conductivity sensor
Model 226 Large bore conductivity sensor
Model 228 Toroidal conductivity sensor
Model 242 Flow-through conductivity sensor
Model 247 Economy conductivity sensor
PERFORMANCE SPECIFICATIONS Measurement Range: see table below
Accuracy: ± 1% of reading and ± 0.01 mS/cm
Repeatability: ± 0.5% of reading and ± 0.005 mS/cm
S1 1027mS/cm
100 C
S2 847.1µS/cm
100 C
Stability: ± 0.25% of reading and ± 0.005 mS/cm/month,
noncumulative
Temperature Compensation: -15 to 200°C (5 to
392°F) automatic or manual. Automatic requires
a Pt100/1000 RTD
Temperature Correction: Linear temperature
coefficient (0.0 to 5.00%/°C) neutral salt (dilute
sodium chloride) or none
RECOMMENDED SENSORS:
RECOMMENDED RANGES FOR TOROIDAL SENSORS
Conductivity Sensor Model
Nominal Cell Constant
µS/cm)
Minimum Conductivity (µ
µS/cm)
Maximum Conductivity (µ
226
1.0
50
1,000,000
228
3.0
200
2,000,000
225
3.0
200
2,000,000
222 (1in.)
6.0
500
2,000,000
222 (2 in.)
4.0
500
2,000,000
242
*
100*
2,000,000*
247
3.5
500
2,000,000
* Model 242 values depend on sensor configuration and wiring.
Rosemount Analytical 7
pH/ORP (Codes -22 and/or -32)
For use with any standard pH or ORP sensor and all
Uniloc sensors and junction boxes with built-in diagnostic style preamplifiers, display choices are pH,
ORP or Redox. The automatic buffer recognition feature uses stored buffer values and their temperature
curves for the most common buffer standards available worldwide. The analyzer will recognize the value
of the buffer being measured and perform a self stabilization check on the sensor before completing the
calibration. Manual or automatic temperature compensation is keypad selectable. Change in pH due to
process temperature can be compensated using a
programmable temperature coefficient or isopotential
point. Measurement and display of pH glass and reference impedance helps alert the user to sensor
maintenance needs.
*reference impedance is suppressed with amperometric/pH
combinations (-24, -25, -26)
PERFORMANCE SPECIFICATIONS ANALYZER (pH INPUT)
Measurement Range [pH]: 0 to 14 pH
RECOMMENDED SENSORS FOR pH:
Model 320B Flow Through pH
Model 320HP High Purity pH
Model 328A Steam Sterilizable pH
Model 370 and 371 EuroSenz pH
Model 381+ Insertion/Submersion/Flow Through pH
Model 385+ Insertion/Submersion/Retractable pH
Model 389 Insertion/Submersion pH
Model 396 Insertion/Submersion pH
Model 396VP Insertion/Submersion pH with VP 6.0 connector
Model 396P Insertion/Submersion pH
Model 396PVP Insertion/Submersion pH with VP 6.0 connector
Model 396R Retractable pH
Model 396RVP Retractable pH with VP 6.0 connector
Model 397 Quik Disconnect pH
Model 398 Insertion/Submersion pH
Model 398VP Insertion/Submersion with VP 6.0 connector
Model 398R Retractable pH
Model 398RVP Retractable pH with VP 6.0 connector
Model 399 Insertion/Submersion pH
Model Hx338 Steam Sterilizable pH
Model Hx348 Steam Sterilizable pH
Accuracy: ±0.01 pH
Repeatability: ±0.01 pH
RECOMMENDED SENSORS FOR ORP:
Stability: ±0.01 pH/month, non-cumulative
Model 330 Flow Through ORP
Model 371 EuroSenz ORP
Model 381+ Insertion/Submersion/Flow Through ORP
Model 385+ Insertion/Submersion/Retractable pH
Model 389 Insertion/Submersion ORP
Model 396P Insertion/Submersion ORP
Model 396PVP Insertion/Submersion ORP with VP 6.0 connector
Model 396R Retractable ORP
Model 398 Insertion/Submersion ORP
Model 398VP Insertion/Submersion with VP 6.0 connector
Model 398R Retractable ORP
Model 398RVP Retractable ORP with VP 6.0 connector
Temperature Coefficient: ±0.003 pH/°C
Temperature Compensation: Pt100/Pt1000 RTD,
Automatic or Manual -15 to 100°C (5 to 212°F)
Temperature Correction: Choose from standard
measurement compensation, solution temperature
correction for high purity or dilute base solutions,
and custom temperature correction.
PERFORMANCE SPECIFICATIONS ANALYZER (ORP INPUT)
Measurement Range [ORP]: -1400 to +1400 mV
Accuracy: ± 2.0 mV
Repeatability: ± 1.0 mV
Stability: ± 1.0 mV/month, non-cumulative
Temperature Coefficient: ± 0.2 mV/°C
Temperature Measurement: -15 to 100°C
(5 to 212°F)
Temperature Correction: none required
S1 4.34pH
25 C
S2 12.34pH
27 C
8 Rosemount Analytical
When used with conductivity (-20-32 or -22-30):
Model 320HP High Purity ORP
Model 381+ Insertion/Submersion/Flow Through ORP
Model 385+ Insertion/Submersion/Retractable ORP
Model 396P Insertion/Submersion ORP
Model 396PVP Insertion/Submersion ORP with VP 6.0 connector
Model 396R Retractable ORP
Model 396RVP Retractable ORP with VP 6.0 connector
FLOW (Code -23 and/or -33 or Suite 2)
For use with most pulse signal flow sensors, the Solu
Comp II's user selectable units of measure include
flow rates in GPM (Gallon per minute), LPM (liters per
minute), or m3/hr (cubic meters per hour), and velocity in ft/sec or m/sec. When configured to measure
flow, the unit also acts as a totalizer in the chosen unit
(gallons, liters, or cubic meters).
Dual flow instruments can be configured as a %
recovery device or a flow difference device.
RECOMMENDED SENSORS*
+GF+ Signet 515 Rotor-X Flow sensor Model
515/8510-XX (PN P51530-PO)
Fluidyne Flow Sensor Model 2300A
(PN Hydro-Flow-2300-A-10-5R-3-1-1)
* Input voltage not to exceed ±36V
Consult factory for other pulse type sensor compatibility.
PERFORMANCE SPECIFICATIONS
Frequency Range: 0.5 - 4000 Hz
Flow Rate: 0 - 9999 GPM, LPM, m3/hr
Totalized Flow: 0 - 9,999,999 Gallons;
37,850,000 Liters; 37,850 m3
Accuracy: ±1% (±1.5% from 3000 to 4000 Hz)
Repeatability: ±1%
S2 12.34 GPM
S2 47.25K Gal
FREE AND TOTAL CHLORINE (Code -24)
The Solu Comp II is compatible with the Model 499ACL-01
free chlorine sensor and the Model 499ACL-02 total
chlorine sensor. The Model 499ACL-02 sensor must be
used with the Model SCS921 sample conditioning system. Both sensors are membrane covered amperometric
sensors. For more information concerning the use and
operation of chlorine amperometric sensors, refer to the
sensor product data sheets.
The Solu Comp II fully compensates free and total chlorine readings for changes in membrane permeability
caused by temperature changes. Temperature is measured by a Pt 100 RTD in the sensor.
For free chlorine measurements, both automatic and
manual pH correction are available. pH correction is necessary because amperometric chlorine sensors measure
only hypochlorous acid. To measure free chlorine
(hypochlorous acid plus hypochlorite ion) most competing analyzers add acid to the sample. Acid lowers the pH
and converts hypochlorite to hypochlorous acid. The
Solu Comp II eliminates messy and expensive chemicals
by using the measured pH to correct the chlorine sensor
signal. If the pH is relatively constant, a fixed pH correction can be used. If the pH is greater than 7 and fluctuates more than about 0.2 units, continuous measurement
of pH and automatic pH correction is necessary.
Corrections are valid to pH 9.5. For automatic pH correction select code -32 and an appropriate pH sensor.
An input filter allows the user to configure the analyzer
for rapid response or low noise. The low noise option is
recommended for samples containing less than 0.1 ppm
chlorine.
Chlorine solutions are unstable, so chlorine standards
are not generally available. Instead, chlorine sensors
must be calibrated against the results of a chemical test
run on a grab sample.
PERFORMANCE SPECIFICATIONS
Measurement Range: 0-20 ppm (mg/L) chlorine
(as Cl2)
Resolution: 0.001 ppm
Automatic pH Correction (requires Code -32): 5.0
to 9.5 pH
Temperature Correction: Automatic (with Pt100 RTD
in sensor) or manual 0-50°C. Can be disabled if
desired.
Input filter: time constant 1 - 999 sec
RECOMMENDED SENSORS
Chlorine: Model 499A CL-01 Free Chlorine or Model
499A CL-02 Total Residual Chlorine (requires
sample conditioning)
pH: The following pH sensors are recommended for
automatic pH correction of chlorine readings:
Models: 399-09-62, 399-14, and 399VP-09
12.34 ppm
26.3 C
8.34pH
Rosemount Analytical 9
DISSOLVED OXYGEN (Code -25)
The Solu Comp II is compatible with the Model
499ADO, 499ATrDO, Hx438, and Gx438 dissolved
oxygen sensors. The sensors are membrane-covered
amperometric sensors. For more information concerning the use and operation of the amperometric
oxygen sensors, refer to the product data sheets.
The Solu Comp II displays dissolved oxygen in ppm,
ppb, or percent saturation.
The Solu Comp II fully compensates oxygen readings
for changes in membrane permeability caused by
temperature changes. In the Model 499ADO and
499ATrDO sensors, temperature is measured by a
Pt 100 RTD. The Hx438 and Gx438 sensors use a
22kNTC.
Calibration is easy. Simply expose the sensor to
water saturated air. Wait until readings are stable and
press a few keys. The analyzer measures the temperature and barometric pressure and automatically
completes the calibration. If removing the sensor
from the process liquid is impractical, the analyzer
can be calibrated against a standard instrument.
Calibration can be corrected for process salinity.
10.34 ppm
29.3 C
PERFORMANCE SPECIFICATIONS
Measurement Range: 0-20 ppm (mg/L) dissolved
oxygen; 0- 250% saturation
Resolution: 0.01 ppm; 0.1 ppb for 499A TrDO sensor
(when O2 <1.00 ppm); 0.1%
Temperature Correction for Membrane Permeability:
Automatic (with Pt100 RTD in sensor) or manual
0-50°C. Can be disabled if desired.
Input filter: time constant 1 - 999 sec
RECOMMENDED SENSORS
Model 499A DO Dissolved Oxygen Sensor
Model Hx438 Steam Sterilizable Dissolved Oxygen
Sensor
Model Gx438 Steam Sterilizable Dissolved Oxygen
Sensor
Model 499A TrDO Trace Dissolved Oxygen Sensor
NOTE: Suite S2A is not compatible with steam-sterilizable Dissolved Oxygen sensors (Models Hx438,
Gx448, Bx438). Order Suite S1A for installations
with steam sterilizable sensors.
12.34mA
DISSOLVED OZONE (Code -26)
The Solu Comp II is intended for use with the Model
499AOZ sensor. The Model 499AOZ sensor is a
membrane-covered amperometric sensor. For more
information concerning the use and operation of the
sensor, refer to the sensor product data sheet.
The Solu Comp II fully compensates ozone readings
for changes in membrane permeability caused by
temperature changes. Temperature is measure by a
Pt 100 RTD in the sensor.
An input filter allows the user to configure the analyzer for rapid response or low noise. The low noise
option is recommended for samples containing less
than 0.1 ppm ozone.
Ozone solutions are unstable; commercial standards
are not available. Instead, ozone sensors must be calibrated against the results of a chemical test run on a
grab sample.
10 Rosemount Analytical
PERFORMANCE SPECIFICATIONS
Measurement Range: 0-10 ppm (mg/L)
Resolution: 0.001 ppm dissolved ozone
Temperature Correction for Membrane Permeability:
Automatic (with Pt100 RTD in sensor) or
manual 0-35°C. Can be disabled if desired.
Input filter: time constant 1 - 999 sec
RECOMMENDED SENSOR
Model 499A OZ Dissolved Ozone Sensor
10.34 ppm
29.3 C
12.34mA
ORDERING INFORMATION
COMBINATION GUIDE
The Solu Comp II analyzers offer the choice of single or dual sensor input with measurement choices of pH/ORP, conductivity/resistivity, toroidal conductivity, flow, chlorine, dissolved oxygen, and dissolved ozone. See combination guide for valid combinations.
Standard features include two isolated outputs, three alarm relays, customizable two-line
display, and temperature correction.
MODEL 1055
SOLU COMP II ANALYZER
CODE
01
02
POWER
115/230 Vac, 50/60 Hz
24 Vdc
CODE
10
11
MOUNTING
Panel mounting enclosure
Pipe/Surface mounting enclosure (Pipe mounting requires accessory kit PN 23820-00)
CODE
20
21
22
23
24
25
26
MEASUREMENT 1 (Required Selection)
Contacting Conductivity
Toroidal Conductivity
pH/ORP
Flow
Chlorine
Dissolved Oxygen
Ozone
CODE
30
31
32
33
MEASUREMENT 2 (Optional)
Contacting Conductivity
Toroidal Conductivity
pH/ORP
Flow
CODE
68
OPTIONAL
UL Approval
Field Commisssioned Suites option offers the user the ability to commission the Solu
Comp II to any valid measurement combination. This feature provides the benefit of a
reduction in the number of spare instruments required to meet emergency inventory needs.
Please refer to the Suites tables (right) for valid measurement combinations. Suites include
a complete set of instrument wiring labels.
MODEL 1055
SOLU COMP II ANALYZER
CODE
01
02
POWER
115/230 Vac, 50/60 Hz
24 Vdc
CODE
10
11
MOUNTING
Panel mounting enclosure
Pipe/Surface mounting enclosure (Pipe mounting requires accessory kit PN 23820-00)
CODE
S1
S1A
S2
S2A
FIELD-COMMISSIONED SUITES (Optional) see tables below
Suite 1 - Field Commissioned Measurement (basic)
Suite 1 - Field Commissioned Measurement (includes amperometric)
Suite 2 - Field Commissioned Measurement (basic)
Suite 2 - Field Commissioned Measurement (includes amperometric)
CODE
DM
OPTIONAL
Dual Measurement
CODE
68
OPTIONAL
UL Approval
NOTE: Do not order Model 1055 Suite S2A for use with steam-sterilizable Dissolved
Oxygen sensors (Models Hx438, Gx448, Bx438). Order Suite S1A for installations
with steam sterilizable sensors.
* For D.O. sensors with 22k thermistor, use Suite 1 only.
Rosemount Analytical 11
ACCESSORIES (Weights are rounded up to nearest whole lb or 0.5 kg)
PART NUMBER
DESCRIPTION
WEIGHT
SHIPPING WT.
23820-00
Pipe mounting kit, includes U-bolts, mounting bracket, nuts,
washers, and screws (complete)
2 lb (1.0 kg)
4 lb (2.0 kg)
23554-00
Gland fittings, PG 13.5, 5 per package
1 lb (0.5 kg)
2 lb (1.0 kg)
9240048-00
Tag, stainless steel (specify marking)
1 lb (0.5 kg)
1 lb (0.5 kg)
pH INPUT
905-3506
Buffer Solution, 4.01 pH, 1 pt.
1 lb (0.5 kg)
2 lb (1.0 kg)
905-3501
Buffer Solution, 7.0 pH, 1 pt.
1 lb (0.5 kg)
2 lb (1.0 kg)
905-3505
Buffer Solution, 10.0 pH, 1 pt.
1 lb (0.5 kg)
2 lb (1.0 kg)
CONDUCTIVITY INPUT
SS-1
Conductivity Standard, 1409 µS/cm at 25°C, 1 quart (945 mL)
2 lb (1.0 kg)
4 lb (2.0 kg)
SS-1A
Conductivity Standard, 1409 µS/cm at 25°C, 1 gallon (3785 mL)
9 lb (4.0 kg)
11 lb (5.0 kg)
SS-5
Conductivity Standard, 1000 µS/cm at 25°C, 1 quart (945 mL)
2 lb (1.0 kg)
4 lb (2.0 kg)
SS-5A
Conductivity Standard, 1000 µS/cm at 25°C, 1 gallon (3785 mL)
9 lb (4.0 kg)
11 lb (5.0 kg)
SS-6
Conductivity Standard, 200 µS/cm at 25°C, 1 quart (945 mL)
2 lb (1.0 kg)
4 lb (2.0 kg)
SS-6A
Conductivity Standard, 200 µS/cm at 25°C, 1 gallon (3785 mL)
9 lb (4.0 kg)
11 lb (5.0 kg)
SS-7
Conductivity Standard, 5000 µS/cm at 25°C, 1 quart (945 mL)
2 lb (1.0 kg)
4 lb (2.0 kg)
SS-7A
Conductivity Standard, 5000 µS/cm at 25°C, 1 gallon (3785 mL)
9 lb (4.0 kg)
11 lb (5.0 kg)
The right people,
the right answers,
right now.
ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
http://www.raihome.com
Specifications subject to change without notice.
Credit Cards for U.S. Purchases Only.
Emerson Process Management
Liquid Division
2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 757-8500
Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2005