Below you will find brief information for Revolution ACTS. This document provides technical specifications for the Revolution ACTS CT scanner. Revolution ACTS supports various network services, including CT Image Storage, MR Image Storage, Secondary Capture Image Storage, Grayscale Softcopy Presentation State Storage, Enhanced SR, X-Ray Radiation Dose SR, RT Structure Set Storage, Positron Emission Tomography Image Storage as well as a number of others. It also supports various media storage application profiles which can be used to read and write files on a variety of media formats, including CD-R, CD-RW, DVD-R, DVD-RW, and USB.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Revolution ACTS DICOM Conformance Statement | Manualzz

GE Healthcare

Direction DOC1906666 Rev 1

Technical

Publications

Direction DOC1906666

Revision 1

REVOLUTION ACTs DICOM CONFORMANCE STATEMENT

Revolution ACTs

Dicom Conformance Statement

DOC1906666Rev 1

© 2016, General Electric Company, All Rights Reserved

GE Healthcare

Direction DOC1906666 Rev 1 REVOLUTION ACTs DICOM CONFORMANCE STATEMENT

Page Left Intentionally Blank

DOC1906666Rev 1

© 2016, General Electric Company, All Rights Reserved

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

LEGAL NOTES

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

TRADEMARKS

All products and their name brands are trademarks of their respective holders.

OMISSIONS & ERRORS

Customers, please contact your GE Sales or Service representatives.

GE personnel, please use the GE Healthcare PQR Process to report all omissions, errors, and defects in this publication.

Copyrights

All Material Copyright (c) 2016 by the General Electric Company, All rights reserved.

3 of 181

GE H

EALTHCARE

DOC1906666, R

EV

1 R

E

VOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

REVISON HISTORY

REV/V

ER

1/1

DATE

18-October-

2016

REASON FOR CHANGE

Initial Revision

4 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Page Left Intentionally Blank

5 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

CONFORMANCE STATEMENT OVERVIEW

This DICOM Conformance Statement captures the DICOM capabilities of the GEHC CT scanners identified below. This document applies to the following software releases for each product:

Revolution ACTs 16IW40.7 (ayush.7) or later

Table 0.1 – NETWORK SERVICES

Table 0.1 provides an overview of the network services supported by GEHC CT products identified above

.

SOP Classes User of Service

(SCU)

Provider of

Service (SCP)

CT Image Storage

MR Image Storage

Secondary Capture Image Storage

Grayscale Softcopy Presentation State Storage

Enhanced SR

X-Ray Radiation Dose SR

Transfer

RT Structure Set Storage

Positron Emission Tomography Image Storage

Query/Retrieve

Study Root Query/Retrieve Information Model – FIND

Study Root Query/Retrieve Information Model – MOVE

Print Management

Basic Grayscale Print Management Meta SOP Class

Basic Color Print Management Meta SOP Class

Print Job SOP Class

Printer SOP Class

Workflow Management

Storage Commitment Push Model SOP Class

Modality Performed Procedure Step SOP Class

Basic Modality Worklist Information Model – FIND SOP

Class

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

No

Yes

No

Yes

Yes

Yes

Yes

No

No

No

No

No

No

No

6 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Table 0.2 provides an overview of the Media Storage Application Profiles supported by the CT products identified above.

Table 0.2 - MEDIA SERVICES

Media Storage Application Profile Write Files

(FSC or FSU)

Read Files

(FSR)

Compact Disk - Recordable

General Purpose CD-R Yes (FSC) Yes

DVD

General Purpose JPEG DVD

Yes (FSC) No

USB

General Purpose JPEG USB

Yes (FSC) Yes

7 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

8 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

TABLE OF CONTENTS

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

1 INTRODUCTION .............................................................................................................................. 15

1.1

O

VERVIEW ................................................................................................................................... 15

1.2

O

VERALL

C

ONFORMANCE

S

TATEMENT

D

OCUMENTATION

S

TRUCTURE

.......................................................... 17

1.3

I

NTENDED

A

UDIENCE

...................................................................................................................... 18

1.4

S

COPE AND

F

IELD OF

A

PPLICATION .................................................................................................... 18

1.5

I

MPORTANT

R

EMARKS

..................................................................................................................... 19

1.6

R

EFERENCES ................................................................................................................................ 20

1.7

D

EFINITIONS

................................................................................................................................ 20

1.8

S

YMBOLS AND

A

BBREVIATIONS ......................................................................................................... 21

2 NETWORK CONFORMANCE STATEMENT ......................................................................................... 23

2.1

I

NTRODUCTION

............................................................................................................................. 23

2.2

I

MPLEMENTATION

M

ODEL ................................................................................................................ 23

2.2.1

2.2.2

2.2.3

Application Data Flow Diagram ..................................................................................... 24

Functional Definition of AE’s ........................................................................................... 26

Sequencing of Real-World Activities ............................................................................... 28

2.3

AE S

PECIFICATIONS

....................................................................................................................... 28

2.3.1

DICOM Server AE Specification ....................................................................................... 28

2.3.1.1

Association Establishment Policy ................................................................................................ 29

2.3.1.1.1

2.3.1.1.2

2.3.1.1.3

2.3.1.1.4

General ........................................................................................................................... 29

Number of Associations ................................................................................................... 29

Asynchronous Nature ...................................................................................................... 29

Implementation Identifying Information ............................................................................ 30

2.3.1.2

Association Initiation by Real-World Activity ................................................................................ 30

2.3.1.2.1

Real World Activity: Push Images ...................................................................................... 30

2.3.1.2.1.1

Associated Real-World Activity ...................................................................................... 30

2.3.1.2.1.2

Proposed Presentation Contexts .................................................................................... 30

Application Data Flow Diagram ........................................................................................................ 33

2.3.1.2.1.2.2.1.1 GSPS Acquisition System with MWL data ............................................................................. 33

2.3.1.2.1.2.2.2.3 GSPS Acquisition System with MWL data ............................................................................. 34

2.3.1.2.2

Real World Activity: Manual Query .................................................................................... 34

2.3.1.2.2.1

Associated Real-World Activity ...................................................................................... 34

2.3.1.2.2.2

Proposed Presentation Contexts .................................................................................... 35

2.3.1.2.3

Real World Activity: Manual Retrieve ................................................................................. 39

2.3.1.2.3.1

Associated Real-World Activity ...................................................................................... 39

2.3.1.2.3.2

Proposed Presentation Contexts .................................................................................... 39

2.3.1.2.4

Real-World Activity: Verify (DICOM Ping)............................................................................. 41

2.3.1.2.4.1

Associated Real-World Activity ...................................................................................... 41

2.3.1.2.4.2

Proposed Presentation Context Table ............................................................................ 41

2.3.1.2.5

Real-World Activity: Remote Archive ................................................................................. 42

2.3.1.2.5.1

Associated Real World Activity:

“Choose Archive Save Option” ................................. 42

2.3.1.2.5.2

Associated Real World Activity: “Auto Archive Exam/Series” ..................................... 42

2.3.1.2.5.3.2 Storage Commitment Push Model SOP Class Request Processing ........................................... 44

2.3.1.2.5.3.3 Storage Commitment Push Model SOP Class Response Processing ........................................ 45

2.3.1.2.5.3.3.1 Commit response with SUCCESS status .......................................................................... 45

2.3.1.2.5.3.3.2 Commit response with FAILURE status ........................................................................... 45

2.3.1.2.6

Real-World Activity: Modality Worklist Query ..................................................................... 46

2.3.1.2.6.1 Associated Real-World Activity ............................................................................................ 46

2.3.1.2.6.2

Proposed Presentation Context Table ............................................................................ 47

2.3.1.2.7

Real-World Activity: Modality Performed Procedure Step Notification ................................. 50

2.3.1.2.7.1

Sequencing of Real-World Activities ........................................................................... 50

2.3.1.2.7.2 Proposed Presentation Context Table .................................................................................. 52

2.3.1.2.8

Real-World Activity: Print .................................................................................................. 54

2.3.1.2.8.1

Associated Real-World Activity ...................................................................................... 54

2.3.1.2.8.2

Proposed Presentation Context ..................................................................................... 56

9 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

2.3.1.3

Association Acceptance Policy ................................................................................................... 67

2.3.1.3.1

Real-World Activity: Receive Images.................................................................................. 68

2.3.1.3.1.1

Associated Real-World Activity ...................................................................................... 68

2.3.1.3.1.2

Presentation Context Table ........................................................................................... 68

2.3.1.3.2

Real-World Activity: Verification Request from Remote AE .................................................. 70

2.3.1.3.2.1

Associated Real-World Activity This AE is indefinitely listening for associations. No operator action is required to respond to a verification message. The Real-World Activity associated with the verification request is to send a C-ECHO response message with a status of “success” to the requesting AE. ........................................................................................... 70

2.3.1.3.2.2

Presentation Context Table ........................................................................................... 70

2.3.1.3.3

Real-World Activity: Search Local Database (Query Request from Remote AE) ..................... 71

2.3.1.3.3.1

Associated Real-World Activity ...................................................................................... 71

2.3.1.3.3.2

Presentation Context Table ........................................................................................... 71

2.3.1.3.4

Real-World Activity: Move Images (Retrieve Request from Remote AE) ................................ 74

2.3.1.3.4.1

Associated Real-World Activity ...................................................................................... 74

2.3.1.3.4.2

Presentation Context Table ........................................................................................... 75

2.3.1.3.4.3

Presentation Context Acceptance Criteria ...................................................................... 76

2.3.1.3.4.4

Transfer Syntax Selection Policy .................................................................................... 77

2.3.1.3.5

Real-World Activity: Listen to remote Storage Commitment SCP ......................................... 77

2.3.1.3.5.1

Associated Real-World Activity ...................................................................................... 77

2.3.1.3.5.2

Presentation Context Table ........................................................................................ 77

2.3.1.3.5.3

Presentation Context Acceptance Criterion .................................................................... 78

2.3.1.3.5.4

Transfer Syntax Selection Policies .............................................................................. 79

2.4

C

OMMUNICATION

P

ROFILES

............................................................................................................. 79

2.4.1

2.4.2

Supported Communication Stacks (parts 8) ..................................................................... 79

TCP/IP Stack.................................................................................................................. 79

2.4.2.1

Physical Media Support .............................................................................................................. 79

2.4.3

Others .......................................................................................................................... 79

2.5

E

XTENSIONS

/ S

PECIALIZATIONS

/ P

RIVATIZATIONS

................................................................................. 79

2.5.1

2.5.2

Standard Extended Elements ......................................................................................... 79

Private Data Elements ................................................................................................... 79

2.6

C

ONFIGURATION

........................................................................................................................... 79

2.6.1

2.6.2

AE Title/Presentation Address Mapping .......................................................................... 79

Configurable Parameters ............................................................................................... 80

2.6.2.1

Print Configuration .................................................................................................................... 80

2.6.2.2

PPS Configuration ...................................................................................................................... 81

2.6.2.3

Modality Worklist Query Configuration ........................................................................................ 81

2.6.2.4

GSPS Configuration ................................................................................................................... 82

2.7

S

UPPORT OF

E

XTENDED

C

HARACTER

S

ETS

............................................................................................ 82

2.8

C

ODES AND

C

ONTROLLED

T

ERMINOLOGY ............................................................................................. 83

2.8.1

Mapped Coded Terminology .......................................................................................... 83

2.9

S

ECURITY

P

ROFILES

....................................................................................................................... 83

3 MEDIA STORAGE CONFORMANCE STATEMENT ............................................................................... 84

3.1

I

NTRODUCTION

............................................................................................................................. 84

3.2

I

MPLEMENTATION

M

ODEL ................................................................................................................ 84

3.2.1

3.2.2

3.2.3

3.2.4

Application Data Flow Diagram ..................................................................................... 84

Functional Definitions of AE's ......................................................................................... 85

Sequencing of Real World Activities ................................................................................ 85

File Meta Information for Implementation Class and Version............................................ 86

3.3

AE S

PECIFICATIONS ....................................................................................................................... 86

3.3.1

DICOM CD-R/CD-RW/DVD-R/DVD-RW/USB Media Interchange AE Specification ............... 86

3.3.1.1

File Meta Information for the CD-R/CD-RW/DVD-R/DVD-RW/USB DICOM Media Interchange Application Entity

87

3.3.1.2

Real World Activities for the CD-R/CD-RW/DVD-R/DVD-RW/USB DICOM Media Interchange Application Entity

3.3.1.2.1

Real World Activity: Record Media - CREATE (CD/DVD/USB) .............................................. 87

3.3.1.2.1.1

Application Profiles for the RWA: Record Media - CREATE (CD/DVD/USB) .......................... 88

87

3.3.1.2.2

Real World Activity: Display Directory – QUERY (CD/USB) .................................................... 89

3.3.1.2.2.1

Application Profiles for the RWA: Display Directory ......................................................... 89

3.3.1.2.2.2

Media Storage Application Profile for the RWA: Display Directory:.................................... 89

3.3.1.2.3

Real World Activity (RWA): Local Storage – RESTORE (CD/USB) .......................................... 89

3.3.1.2.3.1

Application Profiles for the RWA: Local Storage – RESTORE (CD/USB) ............................... 90

10 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

3.4

A

UGMENTED AND

P

RIVATE

A

PPLICATION

P

ROFILES ................................................................................. 90

3.4.1

3.4.2

Augmented Application Profiles ..................................................................................... 90

Private Application Profiles ............................................................................................ 90

3.5

E

XTENSIONS

, S

PECIALIZATIONS AND

P

RIVATIZATIONS OF

SOP C

LASSES AND

T

RANSFER

S

YNTAX .......................... 91

3.5.1

3.5.1.1

SOP Specific Conformance Statement for Basic Directory SOP Class ............................................ 91

3.5.2

Extensions, Specialization’s and Privatizations of SOP Classes ......................................... 91

Private Transfer Syntax Specification .............................................................................. 91

3.6

C

ONFIGURATION

........................................................................................................................... 91

3.7

S

UPPORT OF

E

XTENDED

C

HARACTER

S

ETS ............................................................................................ 91

4 PRINT MANAGEMENT IMPLEMENTATION ........................................................................................ 93

4.1

I

NTRODUCTION ............................................................................................................................. 93

T

HIS SECTION OF CONFORMANCE

S

TATEMENT SPECIFIES THE IMPLEMENTATION FOR THE SPECIFIC

SOP C

LASSES SUPPORTED IN THE

B

ASIC

G

RAYSCALE AND

C

OLOR

P

RINT

M

ANAGEMENT

M

ETA

SOP C

LASSES

,

THE ATTRIBUTES SUPPORTED FOR BOTH

IOD

S AND SERVICES

,

AND THE VALID RANGE OF VALUES FOR

MANDATORY AND OPTIONAL ATTRIBUTES

. ........................................................................................................ 93

4.2

B

ASIC

F

ILM

S

ESSION

SOP C

LASS

....................................................................................................... 93

4.2.1

Basic Film Session N-Create Attributes ........................................................................... 93

4.3

B

ASIC

F

ILM

B

OX

SOP C

LASS

............................................................................................................ 94

4.3.1

4.3.2

4.3.3

Basic Film Box N-Create Attributes ................................................................................. 94

Basic Film Box N-Action Attributes ................................................................................. 95

Basic Film Box N-Delete ................................................................................................. 96

4.4

B

ASIC

G

RAYSCALE

I

MAGE

B

OX

SOP C

LASS ........................................................................................... 96

4.5

B

ASIC

C

OLOR

I

MAGE

B

OX

SOP C

LASS

................................................................................................ 97

4.6

P

RINTER

SOP C

LASS ...................................................................................................................... 98

5 STORAGE COMMITMENT PUSH MODEL IMPLEMENTATION ............................................................ 100

5.1

S

TORAGE

C

OMMITMENT

P

USH

M

ODEL

I

NFORMATION

O

BJECT

D

EFINITION .................................................. 100

5.1.1

5.1.2

Storage Commitment Module for N-ACTION ................................................................. 100

Storage Commitment Module for N-EVENT-REPORT ...................................................... 100

5.1.2.1

Processing of Failure Reason when received in a N-Event-Report .............................................. 101

6 MODALITY WORKLIST INFORMATION MODEL DEFINITION ............................................................ 102

6.1

M

ODALITY

W

ORKLIST

I

NFORMATION

M

ODEL

D

EFINITION

........................................................................ 102

6.1.1

6.1.2

6.1.3

Introduction ................................................................................................................ 102

Modality Worklist Information Model Description .......................................................... 102

Modality Worklist Information Model Entity-Relationship Model ..................................... 102

6.1.3.1

Entity Descriptions ................................................................................................................... 103

6.1.3.1.1

6.1.3.1.2

6.1.3.1.3

6.1.3.1.4

Scheduled Procedure Step .............................................................................................. 103

Requested Procedure Entity Description .......................................................................... 104

Imaging Service Request Entity Description ..................................................................... 104

Visit Entity Description .................................................................................................... 104

6.1.4

6.1.3.1.5

Patient Entity Description ............................................................................................... 104

ModalityWorklist Mapping of DICOM Entities ................................................................. 104

6.1.5

6.1.6

Modality Worklist Information Model Module ................................................................ 104

Information Model Keys ............................................................................................... 105

6.1.6.1

Supported Matching ................................................................................................................ 106

6.1.6.2

Scheduled Procedure Step Entity .............................................................................................. 106

6.1.6.2.1

SOP Common Module .................................................................................................... 106

6.1.6.2.2

Scheduled Procedure Step Module .............................................................................. 106

6.1.6.3

Requested Procedure Entity ..................................................................................................... 108

6.1.6.3.1

Requested Procedure Module ......................................................................................... 108

6.1.6.4

Imaging Service Request Entity ................................................................................................ 109

6.1.6.4.1

Imaging Service Request Module .................................................................................... 109

6.1.6.5

Visit Entity ............................................................................................................................... 110

6.1.6.5.1

6.1.6.5.2

6.1.6.5.3

6.1.6.5.4

Visit Identification .......................................................................................................... 110

Visit Status ..................................................................................................................... 110

Visit Relationship ............................................................................................................ 111

Visit Admission ............................................................................................................... 111

11 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

6.1.6.6

Patient Entity ........................................................................................................................... 111

6.1.6.6.1

6.1.6.6.2

Patient Relationship ....................................................................................................... 111

Patient Identification ...................................................................................................... 111

6.1.6.6.3

Patient Demographic .................................................................................................... 112

6.1.6.6.4

Patient Medical ............................................................................................................. 112

6.2

P

RIVATE

D

ATA

D

ICTIONARY

............................................................................................................ 112

6.3

C-FIND R

EQUEST

M

ESSAGE .......................................................................................................... 113

6.4

U

SE OF

S

PECIFIC

DICOM D

ATA

...................................................................................................... 115

6.5

S

ETTING

U

SER

P

REFERENCES

.......................................................................................................... 117

6.5.1

6.5.2

Setting “Use MWL Study UID “ Option ........................................................................... 117

Setting Custom Query Option ....................................................................................... 117

7 PERFORMED PROCEDURE STEP CONFORMANCE STATEMENT ....................................................... 118

7.1

I

NTRODUCTION

........................................................................................................................... 118

7.2

N-CREATE & N-SET R

EQUEST

M

ESSAGE ......................................................................................... 118

7.3

MODALITY PERFORMED PROCEDURE STEP MODULE DEFINITIONS ............................................. 120

7.4

U

SE OF

S

PECIFIC

DICOM D

ATA

...................................................................................................... 123

7.4.1

7.4.2

7.4.3

Patient Level ............................................................................................................. 123

Study Level ............................................................................................................... 123

Series Level .............................................................................................................. 123

8 GRAYSCALE SOFTCOPY PRESENTATION STATE .............................................................................. 125

8.1

I

NTRODUCTION

........................................................................................................................... 125

8.2

CT M

APPING OF

DICOM E

NTITIES ................................................................................................... 125

8.3

IOD M

ODULE

T

ABLE

.................................................................................................................... 125

8.4

I

NFORMATION

M

ODULE

D

EFINITIONS ................................................................................................ 126

8.4.1

Patient Entity Module ................................................................................................... 126

8.4.1.1

Patient Module ........................................................................................................................ 126

8.4.2

Study Entity Module ..................................................................................................... 126

8.4.2.1

General Study Module .............................................................................................................. 126

8.4.2.2

Patient Study Module ............................................................................................................... 127

8.4.3

Series Entity Module .................................................................................................... 127

8.4.3.1

General Series Module ............................................................................................................. 127

8.4.3.2

Presentation Series Module ...................................................................................................... 129

8.4.4

Equipment Module ...................................................................................................... 129

8.4.4.1

General Equipment .................................................................................................................. 129

8.4.5

Presentation State Entity Module .................................................................................. 129

8.4.5.1

Presentation State Identification Module .................................................................................. 129

8.4.5.2

Presentation State Relationship ................................................................................................ 130

8.4.5.3

Presentation State Shutter ....................................................................................................... 130

8.4.5.4

Displayed Area ........................................................................................................................ 130

8.4.5.5

Modality LUT Module ............................................................................................................... 131

8.4.5.6

Softcopy VOI LUT Module ......................................................................................................... 131

8.4.5.7

Softcopy Presentation LUT Module ........................................................................................... 131

8.4.5.8

SOP Common Module .............................................................................................................. 131

8.5

I

MAGE

H

EADER

C

HANGES

S

UPPORTING

GSPS .................................................................................... 131

8.5.1

Request Attributes Sequence ....................................................................................... 131

9 STRUCTURED REPORT OBJECT IMPLEMENTATION ........................................................................ 133

9.1

IOD M

ODULE

T

ABLE .................................................................................................................... 133

9.2

SR D

OCUMENT

S

ERIES

M

ODULE

...................................................................................................... 133

9.3

SR D

OCUMENT

G

ENERAL

M

ODULE .................................................................................................. 133

9.4

SR D

OCUMENT

C

ONTENT

M

ODULE .................................................................................................. 135

9.4.1

9.4.2

9.4.3

9.4.4

TID 10011 – CT Radiation Dose .................................................................................... 135

TID 10012 – CT Accumulated Dose Data ....................................................................... 136

TID 10013 – CT Irradiation Event Data .......................................................................... 137

TID 10014 - Scanning Length ....................................................................................... 140

12 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

9.4.5

9.4.6

9.4.7

9.4.8

TID 10015 – CT Dose Check Details............................................................................... 140

TID 1002 – Observer Context ........................................................................................ 142

TID 1004 – Device Observer Identifying Attributes ......................................................... 142

TID 1020 – Person Participant ...................................................................................... 142

9.5

C

ONFIGURATION

......................................................................................................................... 143

10 SECURITY CONFORMANCE STATEMENT ........................................................................................ 145

10.1

D

E

-

IDENTIFICATION

..................................................................................................................... 145

APPENDIX A: CT IMAGE AND SECONDARY CAPTURE MODULES/ATTRIBUTES .................................. 148

A.1

CT I

MAGE

IOD ........................................................................................................................... 148

A.1.1

CT Image IOD Modules ................................................................................................ 148

A.1.1.1

Implementation Specific details ................................................................................................ 149

A.2

SC I

MAGE

IOD ........................................................................................................................... 149

A.2.1

SC Image IOD Modules ................................................................................................ 149

CT Dose Report SC Image Details ................................................................................. 150 A.2.2

A.2.2.1

Implementation Specific details ................................................................................................ 151

A.3

C

OMMON

M

ODULES

.................................................................................................................... 152

A.3.1

A.3.2

A.3.3

A.3.4

A.3.5

A.3.6

A.3.7

A.3.8

Patient Module ............................................................................................................ 152

General Study Module ................................................................................................. 152

Patient Study Module ................................................................................................... 153

General Series Module ................................................................................................. 153

General Equipment Module .......................................................................................... 155

General Image Module ................................................................................................ 155

VOI LUT Module ........................................................................................................... 156

SOP Common Module .................................................................................................. 156

A.4

CT I

MAGE

M

ODULES

.................................................................................................................... 157

A.4.1

A.4.2

A.4.3

A.4.4

A.4.5

A.4.6

A.5

SC I

MAGE

M

ODULES

.................................................................................................................... 162

A.5.1

A.5.2

A.5.3

Image Plane Module .................................................................................................... 157

Image Pixel Module ..................................................................................................... 157

Contrast Bolus Module................................................................................................. 158

CT Image Module ........................................................................................................ 158

Frame of Reference Module ......................................................................................... 161

Other Attributes ........................................................................................................... 161

SC Equipment Module.................................................................................................. 162

Image Pixel Module ..................................................................................................... 162

Modality LUT Module ................................................................................................... 162

APPENDIX B: PRIVATE DATA ELEMENTS .......................................................................................... 163

B.1

CT I

MAGE

IOD P

RIVATE

D

ATA

E

LEMENTS

D

EFINITION

............................................................................ 163

B.1.1

B.1.2

B.1.3

B.1.4

B.1.5

B.1.6

B.1.7

B.1.8

B.1.9

B.1.10

B.1.11

Private Creator Identification (GEMS_IDEN_01) .............................................................. 163

Private Creator Identification (GEMS_ACQU_01) ............................................................ 163

Private Creator Identification (GEMS_RELA_01) ............................................................. 164

Private Creator Identification (GEMS_STDY_01) ............................................................. 164

Private Creator Identification (GEMS_IMAG_01) ............................................................. 164

Private Creator Identification (GEMS_0039) ................................................................... 164

Private Creator Identification (GEMS_CT_FLRO_1) ......................................................... 164

Private Creator Identification (GEMS_PARM_01) ............................................................ 165

Private Creator Identification (GEMS_HELIOS_01) .......................................................... 165

Private Creator Identification (GEMS_CT_CARDIAC_001) ................................................ 166

Private Creator Identification (GEHC_CT_ADVAPP_001) ................................................. 167

APPENDIX C: PRIVATE DATA ELEMENTS DICOMDIR DIRECTORY INFORMATION .............................. 169

C.1

B

ASIC

D

IRECTORY

IOD D

EFINITION .................................................................................................. 169

C.2

F

ILE

S

ET

I

DENTIFICATION

M

ODULE ................................................................................................... 169

C.3

D

IRECTORY

I

NFORMATION

M

ODULE

.................................................................................................. 169

13 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

C.4

D

IRECTORY

R

ECORD

S

ELECTION

K

EYS ............................................................................................... 171

C.4.1

C.4.1.1

C.4.2

C.4.2.1

C.4.2.2

C.4.3

C.4.3.1

C.4.4

C.4.4.1

C.4.5

C.4.6

PATIENT KEYS ............................................................................................................. 171

Methods to set Patient ID ............................................................................................. 171

STUDY KEYS ................................................................................................................ 172

Method to set Study Date and Time .............................................................................. 173

Method to set Study ID................................................................................................. 173

SERIES KEYS ................................................................................................................ 174

Method to set Series Number ....................................................................................... 175

IMAGE KEYS ................................................................................................................ 175

Method to set Image Number....................................................................................... 177

PRESENTATION KEYS ................................................................................................... 177

SR DOCUMENT KEYS .................................................................................................... 178

APPENDIX D: IMPLEMENTATION UID FOR PRODUCT VERSIONS ...................................................... 180

14 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

1 INTRODUCTION

Note: Throughout this entire document the term “GEHC CT” refers to the following products:

Revolution ACTs

This document applies to the following software releases for each product:

Revolution ACTs 16IW40.7 (ayush.7) or later

Note: The GEHC CT product lines present an Image Works desktop, which contains advanced applications offered on the Advantage Workstation. The DICOM Conformance Statements related to these applications can be found at the following website: http://www3.gehealthcare.com/en/Products/Interoperability/DICOM/Workstations_DICOM_Conformance_Statements

Advanced Application DICOM Conformance Statement Reference on website

CardIQ Xpress 2.0 Reveal

CT Perfusion 4D Multi-organ / Neuro

Reformat

Volume Viewer

Volume Viewer and its Applications (Release 10.x) 5404107-

100 Rev. 2

Volume Viewer and its Applications (Release 10.x) 5404107-

100 Rev. 2

Volume Viewer and its Applications (Release 10.x) 5404107-

100 Rev. 2

Volume Viewer and its Applications (Release 10.x) 5404107-

100 Rev. 2

Advantage CTC Pro3D EC

AVA Xpress

Volume Viewer and its Applications (Release 10.x) 5404107-

100 Rev. 2

Volume Viewer and its Applications (Release 10.x) 5404107-

100 Rev. 2

1.1 Overview

Section 1, Introduction, which describes the overall structure, intent, and references for this

Conformance Statement.

Section 2, Network Conformance Statement, which specifies GEHC CT compliance to the DICOM requirements for the implementation of networking features.

Section 3, Media Storage Conformance Statement, which specifies GEHC CT compliance to the

DICOM requirements for the implementation of Media Storage features.

Section 4, Print Management Implementation, which specifies GEHC CT compliance to DICOM requirements for the implementation of the Print management service.

15 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Section 5, Storage Commitment Push Model Implementation, which specifies the GEHC CT compliance to DICOM requirements for the implementation of the Storage Commitment service, to store the images using remote DICOM entity, which is Storage Commitment SCP

Section 6, Modality Worklist Query Implementation - which specifies GEHC CT compliance to DICOM requirements for the implementation of the Modality Worklist service. ModalityWorklist is providing the DICOM C-FIND service as a service class user (SCU).

Section 7, Modality Performed Procedure Step Implementation - which specifies GEHC CT compliance to DICOM requirements for the implementation of the Modality Performed Procedure Step service.

The PPS option allows a Modality Performed Procedure Step to be communicated to the

Hospital/Radiology information system. The PPS feature is providing the DICOM Modality Performed

Procedure Step service as a service class user (SCU).

Section 8, Gray Scale Softcopy Presentation State Implementation - which specifies GEHC CT compliance to DICOM requirements for the implementation of the Grayscale Softcopy Presentation

State IOD.

Section 9, Structured Report Information Object Implementation - CT X-ray Radiation Dose / Enhanced

Structured Report Object Implementation.

Section 10 specifies the de-identification for CT images provided by GEHC CT system.

Appendix A specifies the CT IOD information object.

Appendix B specifies the private data element definition for CT IOD.

Appendix C specifies the DICOMDIR directory information.

Appendix D specifies Implementation UIDs for different product versions.

16 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

1.2 Overall Conformance Statement Documentation Structure

The Documentation Structure of the GEHC DICOM Conformance Statements is shown in the Illustration below.

17 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

This document specifies the DICOM implementation. It is entitled:

Revolution ACTs DICOM Conformance Statement

Conformance Statements

Direction DOC1906666

This DICOM Conformance Statement documents the DICOM Conformance Statement and Technical

Specification required interoperating with the GEHC network interface.

The GEHC Conformance Statement, contained in this document, also specifies the Lower Layer communications, which it supports (e.g., TCP/IP). However, the Technical Specifications are defined in the DICOM Part 8 standard.

For more information regarding DICOM, copies of the Standard may be obtained on the Internet at http://medical.nema.org

. Comments on the Standard may be addressed to:

DICOM Secretariat

NEMA

1300 N. 17 th

Street, Suite 1752

Rosslyn, VA 22209

USA

Phone : +1.703.841.3200

1.3 Intended Audience

The reader of this document is concerned with software design and/or system integration issues. It is assumed that the reader of this document is familiar with the DICOM Standard and with the terminology and concepts, which are used in that Standard.

1.4 Scope and Field of Application

It is the intent of this document to provide an unambiguous specification for GEHC implementations.

This specification, called a Conformance Statement, includes a DICOM Conformance Statement and is necessary to ensure proper processing and interpretation of GEHC medical data exchanged using

DICOM. The GEHC Conformance Statements are available to the public.

The reader of this DICOM Conformance Statement should be aware that different GEHC devices are capable of using different Information Object Definitions. For example, a GEHC CT Scanner may send images using the CT Information Object, MR Information Object, Secondary Capture Object, etc.

Included in this DICOM Conformance Statement are the Module Definitions, which define all data elements, used by this GEHC implementation. If the user encounters unspecified private data elements while parsing a GEHC Data Set, the user is well advised to ignore those data elements (per the DICOM standard). Unspecified private data element information is subject to change without notice. If, however, the device is acting as a "full fidelity storage device", it should retain and retransmit all of the private data elements which are sent by GEHC devices.

18 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

1.5 Important Remarks

The use of these Conformance Statements, in conjunction with the DICOM Standards, is intended to facilitate communication with GE imaging equipment. However, by itself, it is not sufficient to

insure that inter-operation will be successful. The user (or user’s agent) needs to proceed with caution and address at least four issues:

Integration

The integration of any device into an overall system of interconnected devices goes beyond the scope of standards (DICOM), and of this introduction and associated Conformance

Statements when interoperability with non-GE equipment is desired. The responsibility to analyze the applications requirements and to design a solution that integrates GE imaging equipment with non-GE systems is the user’s responsibility and should not be underestimated. The user is strongly advised to ensure that such integration analysis is correctly performed.

Validation

Testing the complete range of possible interactions between any GE device and non-GE devices, before the connection is declared operational, should not be overlooked. Therefore, the user should ensure that any non-GE provider accepts full responsibility for all validation required for their connection with GE devices. This includes the accuracy of the image data once it has crossed the interface between the GE imaging equipment and the non-GE device and the stability of the image data for the intended applications.

Such a validation is required before any clinical use (diagnosis and/or treatment) is performed. It applies when images acquired on GE imaging equipment are processed/displayed on a non-GE device, as well as when images acquired on non-GE equipment is processed/displayed on a GE console or workstation.

Future Evolution

GE understands that the DICOM Standard will evolve to meet the user’s growing requirements. GE is actively involved in the development of the DICOM Standard. DICOM will incorporate new features and technologies and GE may follow the evolution of the Standard.

Evolution of the Standard may require changes to devices that have implemented DICOM. In addition, GE reserves the right to discontinue or make changes to the support of communications features (on its products) reflected on by these DICOM Conformance

Statements. The user should ensure that any non-GE provider, which connects with GE devices, also plans future evolution of the DICOM standard. Failures to do so will likely result in the loss of function and/or connectivity as the DICOM Standard changes and GE products are enhanced to support these changes.

Interaction

It is the sole responsibility of the non-GE provider to ensure that communications with the interfaced equipment does not cause degradation of GE imaging equipment performance and/or function.

19 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

1.6 References

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

NEMA PS3 Digital Imaging and Communications in Medicine (DICOM) Standard, available free at http://medical.nema.org/

1.7 Definitions

Informal definitions are provided for the following terms used in this Conformance Statement. The

DICOM Standard is the authoritative source for formal definitions of these terms.

Abstract Syntax – the information agreed to be exchanged between applications, generally equivalent to a Service/Object Pair (SOP) Class. Examples: Verification SOP Class, Modality Worklist

Information Model Find SOP Class, Computed Radiography Image Storage SOP Class.

Application Entity (AE) – an end point of a DICOM information exchange, including the DICOM network or media interface software; i.e., the software that sends or receives DICOM information objects or messages. A single device may have multiple Application Entities.

Application Entity Title – the externally known name of an Application Entity, used to identify a

DICOM application to other DICOM applications on the network.

Application Context – the specification of the type of communication used between Application

Entities. Example: DICOM network protocol.

Association – a network communication channel set up between Application Entities.

Attribute – – a unit of information in an object definition; a data element identified by a tag. The information may be a complex data structure (Sequence), itself composed of lower level data elements. Examples: Patient ID (0010,0020), Accession Number (0008,0050), Photometric

Interpretation (0028,0004), Procedure Code Sequence (0008,1032).

Information Object Definition (IOD) – the specified set of Attributes that comprise a type of data object; does not represent a specific instance of the data object, but rather a class of similar data objects that have the same properties. The Attributes may be specified as Mandatory (Type 1),

Required but possibly unknown (Type 2), or Optional (Type 3), and there may be conditions associated with the use of an Attribute (Types 1C and 2C). Examples: MR Image IOD, CT Image IOD, Print Job IOD.

Joint Photographic Experts Group (JPEG) – a set of standardized image compression techniques, available for use by DICOM applications.

Media Application Profile – the specification of DICOM information objects and encoding exchanged on removable media (e.g., CDs)

Module – a set of Attributes within an Information Object Definition that are logically related to each other. Example: Patient Module includes Patient Name, Patient ID, Patient Birth Date, and Patient Sex.

Negotiation – first phase of Association establishment that allows Application Entities to agree on the types of data to be exchanged and how that data will be encoded.

20 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Presentation Context – the set of DICOM network services used over an Association, as negotiated between Application Entities; includes Abstract Syntaxes and Transfer Syntaxes.

Protocol Data Unit (PDU) – a packet (piece) of a DICOM message sent across the network. Devices must specify the maximum size packet they can receive for DICOM messages.

Security Profile – a set of mechanisms, such as encryption, user authentication, or digital signatures, used by an Application Entity to ensure confidentiality, integrity, and/or availability of exchanged

DICOM data

Service Class Provider (SCP) – role of an Application Entity that provides a DICOM network service; typically, a server that performs operations requested by another Application Entity (Service Class

User). Examples: Picture Archiving and Communication System (image storage SCP, and image query/retrieve SCP), Radiology Information System (modality worklist SCP).

Service Class User (SCU) – role of an Application Entity that uses a DICOM network service; typically, a client. Examples: imaging modality (image storage SCU, and modality worklist SCU), imaging workstation (image query/retrieve SCU)

Service/Object Pair (SOP) Class – the specification of the network or media transfer (service) of a particular type of data (object); the fundamental unit of DICOM interoperability specification.

Examples: Ultrasound Image Storage Service, Basic Grayscale Print Management.

Service/Object Pair (SOP) Instance – information object; a specific occurrence of information exchanged in a SOP Class. Examples: a specific x-ray image.

Tag – a 32-bit identifier for a data element, represented as a pair of four digit hexadecimal numbers, the “group” and the “element”. If the “group” number is odd, the tag is for a private (manufacturerspecific) data element. Examples: (0010,0020) [Patient ID], (07FE,0010) [Pixel Data], (0019,0210)

[private data element]

Transfer Syntax – the encoding used for exchange of DICOM information objects and messages.

Examples: JPEG compressed (images), little endian explicit value representation.

Unique Identifier (UID) – a globally unique “dotted decimal” string that identifies a specific object or a class of objects; an ISO-8824 Object Identifier. Examples: Study Instance UID, SOP Class UID, SOP

Instance UID.

Value Representation (VR) – the format type of an individual DICOM data element, such as text, an integer, a person’s name, or a code. DICOM information objects can be transmitted with either explicit identification of the type of each data element (Explicit VR), or without explicit identification

(Implicit VR); with Implicit VR, the receiving application must use a DICOM data dictionary to look up the format of each data

1.8 Symbols and Abbreviations

AE Application Entity

CD-R Compact Disk Recordable

21 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

CT Computed Tomography

DICOM Digital Imaging and Communications in Medicine

FSC

FSU

File-Set Creator

File-Set Updater

FSR File-Set Reader

GSPS Grayscale Softcopy Presentation State

HIS

IHE

Hospital Information System

Integrating the Healthcare Enterprise

IOD Information Object Definition

JPEG Joint Photographic Experts Group

MPPS Modality Performed Procedure Step

MSPS Modality Scheduled Procedure Step

MTU Maximum Transmission Unit (IP)

MWL Modality Worklist

NTP

O

Network Time Protocol

Optional (Key Attribute)

PACS Picture Archiving and Communication System

PET

R

RIS

SC

SCP

Positron Emission Tomography

Required (Key Attribute)

Radiology Information System

Secondary Capture

Service Class Provider

Service Class User SCU

SOP

SR

Service-Object Pair

Structured Reporting

TCP/IP Transmission Control Protocol/Internet Protocol

U

VR

Unique (Key Attribute)

Value Representation

22 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

2 NETWORK CONFORMANCE STATEMENT

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

2.1 Introduction

This Conformance Statement(CS) specifies the GEHC CT compliance to DICOM. It details the DICOM

Service Classes and roles that are supported by this product.

The GEHC CT product uses DICOM services to support the following functionalities:

• Import images for possible further analysis and/or processing

• Export images to other DICOM-compliant machines

• Confirm that a DICOM image has been permanently stored (archived) by a device

• Query for and display DICOM modality worklist information from a remote hospital or radiology department information systems

• Communicate Modality Performed Procedure Step to the Hospital/Radiology information systems

• Print images on DICOM Compliant Printers

2.2 Implementation Model

All DICOM functionality on the GEHC CT product is handled by the DICOM Server Application Entity

(AE). The DICOM Server AE is commanded to perform DICOM services through the buttons and menu selections on the main user interface panel. The DICOM Server AE is also listening to a pre-defined port for incoming connections.

23 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

2.2.1 Application Data Flow Diagram

ILLUSTRATION 2-0

IMPLEMENTATION MODEL DATA FLOW DIAGRAM

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

24 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

There are several Real-World Activities that will cause the DICOM Server Application Entity (DICOM

Server AE) to initiate a DICOM association to a remote DICOM Application Entity. Illustration 2.0 above shows basic Real-World Activities. In addition, Sections 3, 4, 5, 6, and 7 cover Media Storage, Print,

Modality Worklist, Performed Procedure Step, and Storage Commit in more detail.

The following paragraphs describe the Real-World Activities of Illustration 2.0:

1. Push Images

This Real-World Activity consists of an operator selecting one or more studies, series, or images in the local database browser. The operator then clicks on the destination button in the network panel at the bottom of the local database browser.

2. Manual Query

The operator queries one or a set of remote DICOM databases to obtain a list of data at

Study/Series/Image level by clicking on the icon that represents the wanted remote DICOM AE.

3. Manual Retrieve

This Real-World activity will be available once the Manual Query activity is performed. The operator can now select one or more studies (series or images) and ask the DICOM Server AE to retrieve the selected image(s) from the Remote DICOM AE by clicking on the “Local DB” button at the bottom of the local database browser.

4. Verify Connectivity (DICOM Ping)

This Real-World activity consists of an operator selecting a Remote DICOM AE from the

“Network Configuration” window and clicking on “Ping” on the right side-bar. This is to check the status of the selected remote DICOM AE.

5. Remote Archive

This Real-World activity consists of an operator choosing a remote DICOM AE that supports

Storage Commitment as provider from the archive panel at the bottom of the local database browser. The operator chooses the exam or series in the local database browser and clicks on the archival destination from the archive list at the bottom of the local database browser.

The Commitment request for the transferred image instances is sent after the complete image transfer. The Commitment response can come on same association or on a different association. Refer to Section 5 for details on the Storage Commitment implementation.

6. Modality Worklist Query

The operator or the system initiates a modality worklist query to the modality worklist SCP with a given set of query parameters. The modality worklist SCP returns responses matching the query parameters. Worklist items from the returned worklist query responses are presented to the user. The user then chooses the desired worklist item and begins the image acquisition process.

7. Modality Performed Procedure Step Notification

When the user begins the image acquisition process and generates the first image, the DICOM

25 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

SERVER AE sends N-CREATE message to the configured MPPS SCP to indicate that the image acquisition process has been started for the requested procedure.

The operator can close the acquisition session either by completing the acquisition process or discontinuing the ongoing scan. On closing the acquisition session, the DICOM SERVER AE sends N-SET message to the configured MPPS SCP to indicate the acquisition state of the requested procedure, with appropriate MPPS status (COMPLETED/DISCONTINUED).

8. Print

The Film Composer allows the operator to select printers and it also allows the user to drag and drop the images (from viewer application) into the film. When user presses the "Print" Button, the DICOM SERVER AE tries to establish the association with requested printer and sends the images for printing.

9. Receive Images

When remote DICOM hosts send DICOM images to DICOM SERVER AE, images are installed in the local database. The browser displays the content of the local database.

10. Image Installation

The DICOM Server AE will perform this Real-World activity after the remote AE sends an image to the GEHC CT product.

11. Receive Verification request

No operator action is required to respond to a verification message. The Real-World Activity associated with the verification request is to send a C-ECHO response message with a status of

“success” to the requesting AE.

12. Search Local Database

For this operation, a remote DICOM AE asks to obtain the list of data at Study/Series/Image level. Once a Query request is received, the DICOM Server AE will search the local database for all entries that match the keys requested by the Remote DICOM AE and send back the list of matches.

13. Move Images

For this operation, a remote DICOM AE asks to send data at Study/Series/Image level from the local AE to another DICOM Remote AE. The Remote DICOM AE shall be declared locally on the system. The declaration of remote DICOM AE is done through a configuration tool.

14. Listen to remote Storage Commitment SCP

The DICOM SERVER AE is indefinitely listening for association requests. No operator action is required to receive a Storage Commitment notofication (N-EVENT-REPORT)

2.2.2 Functional Definition of AE’s

DICOM Server Application Entity initiates the following operations:

Initiate an association to a Remote AE for the purpose of sending images to the Remote AE.

Remote AE accepts the presentation context applicable to the image(s) being sent, the DICOM

26 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Server AE will send the image(s) to the receiving Remote AE by invoking the C-STORE-RQ operation for each image on the same association.

Initiate an association to a Remote AE for the purpose of committing images previously sent successfully to the Remote AE for the purpose of the remote AE to commit to the storage of those images. If the Remote AE accepts the presentation context, a storage commitment will be established with the Remote AE with the DICOM Server AE sending the N-ACTION Request. The

Remote AE completes the commitment by sending the N-EVENT REPORT. The DICOM Server AE updates the archive flag in the image browser for successful instances.

Initiate an association with a Remote AE to query for images on the remote host. A Study-Root

Study-Level C-FIND-RQ request will be sent to the Remote AE once an association has been established. Once all the responses have been received, the operator needs to select an exam in the local database browser, on selection of the exam the DICOM Server AE will issue a Series-

Level C-FIND-RQ request to get the series for a study in the list. Similarly the Image-Level C-FIND-

RQ will be issued for the series selected from the series list.

Send a C-MOVE-RQ request to a Remote AE for retrieve of images after successful establishment.

The DICOM Server AE’s Storage SCP will receive the images over a separate association.

Initiate an association with a Remote AE to verify its status with a C-ECHO-RQ. The Remote AE will report its status in a C-ECHO-RSP

Initiate an association with a Remote AE to query for and display DICOM modality worklist information. Once the remote AE accepts the presentation context, CT system will issue a modality worklist query request using the C-FIND service

Initiate an association with a Remote AE to create and update DICOM Modality Performed

Procedure Step SOP instance in the remote AE. Once the remote AE accepts the presentation context, SOP instance is created using N-CREATE service and updated using N-SET service.

Initiates as association with a remote printer for printing the images. If the Remote AE accepts the presentation context applicable to the print job, CT system will send the print job to the receiving remote AE.

The DICOM Server AE waits for association requests from Remote AEs that wish to perform the following operations:

Verification: If a C-ECHO-RQ message is received, the DICOM Server AE will send back a C-ECHO-

RSP message with a status of “success”.

Image Storage: If a C-STORE-RQ message is received, the DICOM Server AE will receive the image and try to update the local database. If the image is stored successfully on storage media

 and the database updated a status of “success” will be returned in a C-STORE-RSP message.

Query: If a C-FIND-RQ message is received the DICOM Server AE will search the database for the requested attributes and send back a C-FIND-RSP message containing a match and a status of

27 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

“pending”. After all matching records have been sent, a status of “success” will be returned in a

C-FIND-RSP message. The Remote AE can terminate the query by sending a C-CANCEL-FIND-RQ message.

Retrieve: If a C-MOVE-RQ message is received the DICOM Server AE will look up its list of configured Remote AEs for the Destination AE. If the Destination AE is configured, the DICOM

Server AE will open a new association to the Destination AE and use C-STORE-RQ to send the image(s). The DICOM Server AE will send a C-MOVE-RSP message with a status of “pending” after every five images are sent. When all images are sent or if DICOM Server AE receives a C-CANCEL-

MOVE-RQ a final C-STORE-RSP will be sent back with an appropriate status.

Storage Commitment Response: If a N-EVENT-REPORT request is received from the Remote

Storage Commitment SCP the DICOM Server AE will update the Archive flag information for successful instances.

2.2.3 Sequencing of Real-World Activities

Real-World Activity Manual Query must be performed before Manual Retrieve can be performed.

2.3 AE Specifications

2.3.1 DICOM Server AE Specification

This Application Entity provides Standard Conformance to the following DICOM SOP Classes as an

SCU:

SOP Class Name (SCU)

Verification (Echo)

CT Image Information Storage

MR Image Information Storage

PET Image Information Storage

RT Structure Set Storage

Secondary Capture Image Storage

SOP Class UID

1.2.840.10008.1.1

1.2.840.10008.5.1.4.1.1.2

1.2.840.10008.5.1.4.1.1.4

1.2.840.10008.5.1.4.1.1.128

1.2.840.10008.5.1.4.1.1.481.3

1.2.840.10008.5.1.4.1.1.7

Grayscale Softcopy Presentation State Storage 1.2.840.10008.5.1.4.1.1.11.1

X-Ray Radiation Dose SR Image Storage 1.2.840.10008.5.1.4.1.1.88.67

Enhanced SR Image Storage 1.2.840.10008.5.1.4.1.1.88.22

Study Root Query/Retrieve – FIND

Study Root Query/Retrieve – MOVE

Storage Commitment Push Model

1.2.840.10008.5.1.4.1.2.2.1

1.2.840.10008.5.1.4.1.2.2.2

1.2.840.10008.1.20.1

Note: As to GSPS as an SCU, refer to section 8 for further details.

This Application Entity provides Standard Conformance to the following DICOM SOP classes as an

SCP:

SOP Class Name (SCP)

Verification (Echo)

CT Image Information Storage

MR Image Information Storage

28 of 181

SOP Class UID

1.2.840.10008.1.1

1.2.840.10008.5.1.4.1.1.2

1.2.840.10008.5.1.4.1.1.4

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

PET Image Information Storage

RT Structure Set Storage

Secondary Capture Image Storage

Enhanced SR

Study Root Query/Retrieve – FIND

Study Root Query/Retrieve – MOVE

2.3.1.1 Association Establishment Policy

2.3.1.1.1 General

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

1.2.840.10008.5.1.4.1.1.128

1.2.840.10008.5.1.4.1.1.481.3

1.2.840.10008.5.1.4.1.1.7

1.2.840.10008.5.1.4.1.1.88.22

1.2.840.10008.5.1.4.1.2.2.1

1.2.840.10008.5.1.4.1.2.2.2

The DICOM Application Context Name (ACN), which is always proposed, is:

Application Context Name 1.2.840.10008.3.1.1.1

The Maximum Length PDU negotiation is included in all association establishment requests. The maximum length PDU for association initiated by the DICOM Server AE is:

Maximum Length PDU 64 Kbytes

SOP class Extended Negotiation is not supported.

The maximum number of Presentation Context Items that is supported is 60. Note that the same

Abstract Syntax may be offered multiple times with different Transfer Syntax.

The user information items sent by this product are:

Maximum PDU Length

Implementation UID

The table in Appendix D identifies the Implementation UID for this product version.

2.3.1.1.2 Number of Associations

The DICOM Server AE (SCU) will initiate only one DICOM association at a time for each of the following services:

-Push image to a remote host

- Query/Retrieve

- Storage Commitment

- Modality Worklist

- PPS

- Print service

The DICOM Server AE (SCP) can have a maximum of four DICOM associations open simultaneously to

Receive and Store image store, N-EVENT-REPORT of Storage Commitment or Respond to an Echo or

Resond to a Query/Retrieve.

2.3.1.1.3 Asynchronous Nature

Asynchronous mode is not supported. All operations will be performed synchronously.

29 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

2.3.1.1.4 Implementation Identifying Information

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

The Implementation UID allows unique identification of a set of products that share the same implementation.

The table in Appendix D identifies the Implementation UID for this product version.

2.3.1.2 Association Initiation by Real-World Activity

2.3.1.2.1 Real World Activity: Push Images

2.3.1.2.1.1 Associated Real-World Activity

The operator must first select the exam/series/image on the local database browser and click on the

Remote DICOM AE in the network panel at the bottom of the local database browser to which the operator desires to send the exam/series/image to.

Note: If multiple study, series, or images are chosen to be pushed, one association will be established for each of the studies, series, or images.

2.3.1.2.1.2 Proposed Presentation Contexts

The following table shows the proposed presentation contexts for the DICOM Server AE after Real-

World Activity “Push Images” Operation has been performed.

Name

CT Image

Storage

MR Image

Storage

PET Image

Storage

Abstract Syntax

RT Structure Set

Storage

Presentation Context Table – Proposal

Transfer Syntax

UID

1.2.840.10008.5.1.4.1.1.

2

1.2.840.10008.5.1.4.1.1.

4

1.2.840.10008.5.1.4.1.1

.128

1.2.840.10008.5.1.4.1.1.

48

1.3

Name List

Implicit VR Little

Endian

Explicit VR Little

Endian

Explicit VR Big Endian

Implicit VR Little

Endian

Explicit VR Little

Endian

Explicit VR Big Endian

Implicit VR Little

Endian

Explicit VR Little

Endian

Explicit VR Big Endian

Implicit VR Little

Endian

Explicit VR Little

Endian

Explicit VR Big Endian

UID List

1.2.840.10008.1.2

1.2.840.10008.1.2.

1

1.2.840.10008.1.2.

2

1.2.840.10008.1.2

1.2.840.10008.1.2.

1

1.2.840.10008.1.2.

2

1.2.840.10008.1.2

1.2.840.10008.1.2.

1

1.2.840.10008.1.2.

2

1.2.840.10008.1.2

1.2.840.10008.1.2.

1

1.2.840.10008.1.2.

2

Role

SCU

SCU

SCU

SCU

Extended

Negotiation

None

None

None

None

30 of 181

GE

D

H

EALTHCARE

IRECTION

DOC1906666,

Name

Secondary

Capture Image

Storage

Grayscale

Softcopy

Presentation

State Storage

Enhanced SR

Abstract Syntax

X-Ray Radiation

Dose SR

R

EV

1

7

1.2.840.10008.5.1.4.1.1.

88.67

UID

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Presentation Context Table – Proposal

Transfer Syntax Role

1.2.840.10008.5.1.4.1.1.

1.2.840.10008.5.1.4.1.1.

11

.1

1.2.840.10008.5.1.4.1.1.

88.22

Name List

Implicit VR Little

Endian

Explicit VR Little

Endian

Explicit VR Big Endian

Implicit VR Little

Endian

Explicit VR Little

Endian

Explicit VR Big Endian

Implicit VR Little

Endian

Explicit VR Little

Endian

Explicit VR Big Endian

Implicit VR Little

Endian

Explicit VR Little

Endian

Explicit VR Big Endian

UID List

1.2.840.10008.1.2

1.2.840.10008.1.2.

1

1.2.840.10008.1.2.

2

1.2.840.10008.1.2

1.2.840.10008.1.2.

1

1.2.840.10008.1.2.

2

1.2.840.10008.1.2

1.2.840.10008.1.2.

1

1.2.840.10008.1.2.

2

1.2.840.10008.1.2

1.2.840.10008.1.2.

1

1.2.840.10008.1.2.

2

SCU

SCU

SCU

SCU

Extended

Negotiation

None

None

None

None

2.3.1.2.1.2.1 SOP Specific Conformance Statement for all storage SOP Classes

This implementation can perform multiple C-STORE operations over a single association.

Upon receiving a C-STORE confirmation containing a successful status, this implementation will perform the next C-STORE operation. The association will be maintained if possible.

Upon receiving a C-STORE confirmation containing a Refused status, this implementation will terminate the association.

Upon receiving a C-STORE confirmation containing any status that is not Success or Warning, this implementation will consider the current request to be a failure and will terminate the association except if the C-STORE is invoked from a C-MOVE SCP. In this case it will continue to attempt to send the remaining images in the request on the same association.

Following are the status codes that are more specifically processed when receiving messages from

Storage SCP equipment:

Service

Status

Status Code Further Meaning Application Behavior When Receiving Status

Code

Failure A700-A7FF

A900-A9FF

Refused: Out of resources Error displayed in the Network status and browser job manager queue.

Error: Data Set does not match

SOP Class

Error displayed in the Network status and browser job manager queue.

31 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

C000-CFFF

0122

Warning B000

B006

B007

Success 0000

Error: Cannot Understand

SOP Class Not Supported

Coercion of Data Elements

Elements Discarded

Data Set does not match SOP

Class

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Error displayed in the Network status and browser job manager queue.

Error displayed in the Network status and browser job manager queue.

Log-files updated

Log-files updated

Log-files updated

Any other status code

Success status displayed in the Network status and browser job manager queue.

Treated as Failure * *

Each C-STORE operation supports an “Association Timer”. This timer starts when the association request is sent or received and stops when the association is established. The time-out is configurable in network-cfg.xml file as StoreAssociationTimeout. Default is 60 seconds.

Each C-STORE operation also supports an “Operation Inactivity Timer”. This time-out starts once the first C-STORE request has been issued (on association) or received and is reset each time a C-STORE response has been received or when subsequent C-STORES are sent. This time-out is configurable in networkcfg.xml file as StoreResponseTimeout. Default is 5 minutes.

If any of the two timers mentioned above expires, the connection is closed and the operation in progress is considered failed.

Note: The time-outs are configurable.

When DICOM Server AE initiates an association to issue a C-STORE, the image will be transmitted by the DICOM Server AE with the same elements as was originally received or created locally.

2.3.1.2.1.2.2 SOP Specific Conformance Statement for Grayscale Softcopy Presentation State Storage SOP Class

To create Grayscale Softcopy Presentation State, there are 2 sequences - Exam Split and Save State

2.3.1.2.1.2.2.1 Exam Split

Exam Split option have the two options i.e. Virtual Exam Split and Hard Exam Split.

Virtual Exam Split will create Gray Scale Presentation State and Performed Presentation State.

In Hard Exam Split, application will send the user selected images from grouped Images that acquired during acquisition. PPS will not be sent along with these images.

The DICOM Modality ‘Gray Scale Presentation State’ service is provided by the Exam Split. The DICOM

AE is commanded to perform transmission of Gray Scale Presentation State object through the user

32 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

interface. The DICOM AE is commanded to perform transmission of Hard Split Images through the user interface.

Application Data Flow Diagram

The basic Application models for the feature are shown in the following illustration:

Scanner receives MWL Requested

Procedures from HIS/RIS.

Operator associates images, display settings (Window Level,

Window Width, Pan, Zoom and

Image Orientation) with a requested procedure

Operator selects multiple requested procedures and scans images as a single exam

GSPS object

PPS object

After post processing Exam Split application presents user selected images and Requested procedures for split to create PPS and GSPS objects

GSPS Remote

Viewer

PPS

Manager

2.3.1.2.1.2.2.1.1 GSPS Acquisition System with MWL data

The system has a Modality Work-list Server AE installed. Work-List information is obtained from HIS/RIS system through the use of Basic Work-list Management Service. Use of the information retrieved in the creation of Image SOP instance is described in the Modality Work-list Conformance statement. Use of the information retrieved in MPPS SOP instances is described later in this document.

After Post processing Exam Split application presents associated requested Procedure(s) along with selected acquired Images.

Exam Split application includes the necessary information related to Requested Procedure, scheduled Procedure Steps and the Performed Procedure Step of the images acquired during acquisition.

After User modifications on Image(s), user has to click on Send button to create Gray Scale

Presentation State and Performed Procedure Step objects to transmit to remote host. If there is any Image Orientation, user will be notified about the orientation before sending to Remote

Host.

Exam Split will provide Host Selection user interface to select remote host to transmit of Gray

Scale Presentation State object. Performed Procedure Step object will be sent to default

Performed Procedure Step host.

After successful transmission of Gray Scale Presentation State and Performed Procedure Step objects to remote host, user will be notified with success.

2.3.1.2.1.2.2.2 Save State

2.3.1.2.1.2.2.2.1 Implementation Model

33 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Save State will create Grayscale Presentation State (GSPS) including user annotation.

2.3.1.2.1.2.2.2.2 Application Data Flow Diagram

The basic Application models for the feature are shown in the following illustration:

Scan a series.

Select the scanned series from image browser

.

Associates images, display settings

(Window Level, Window Width,

Pan, Zoom and Image Orientation) with a requested.

Press Save State.

GSPS series is created with new series

2.3.1.2.1.2.2.2.3 GSPS Acquisition System with MWL data

Operator takes some scans. After the scan finished, images shall be selected from image browser and select Viewer.

Operator adds annotation or ROI on images, change WW or WL, zoom or roam images.

After changing settings, operator shall select “Save State” on Viewer menu. Then, GSPS object shall be created with a new series number.

2.3.1.2.2 Real World Activity: Manual Query

2.3.1.2.2.1 Associated Real-World Activity

The operator must select the Remote DICOM AE from the “Source” pull-down on the local database browser and select the hostname of the source from where the images are to be retrieved.

Provided that the operator had set "Custom search" to "Off" when setting the Remote host parameters, the “Query” operation will cause the DICOM Server AE to initiate an association (with zero length Patient name, Patient id, Study date, Accession number, and Study id) to the selected

Remote AE when the “Hostname” entry is selected from the “Source” pull-down menu.

Otherwise, if the operator had set "Custom search" to "On" when setting the Remote host parameters, the “Query” operation will cause a Customize search parameters menu to appear. The operator can

34 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

enter values for Patient name, Patient id, Study date, Accession number, and Study id. Not entering a value means match on any value for that field. Patient name will match on any patient name that contains what the operator entered. Patient id, Study id, and Accession number will match on what the operator enters.

For Study date, the operator selects a range type from the "Exam Date" pull down menu, where the choices are “Equals“, “Before“, “Between“, or “After“. Once a range type is selected, the correct number of fields appears and the operator enters dates into those fields. Once the desired parameters are entered the operator chooses "OK" and that will cause the DICOM Server AE to initiate an association to the selected Remote AE.

Once a list of Study/Series/Image is retrieved, the operator can invoke the “Pull” operation by clicking on the “Local DB” button at the bottom of the local database browser.

2.3.1.2.2.2 Proposed Presentation Contexts

When the Real-World activity “Manual Query” is initiated all presentation contexts shown in the following table are proposed during association establishment, but only the Query/Retrieve-FIND related contexts are applicable to this activity.

Abstract Syntax

Presentation Context Table – Proposal

Transfer Syntax Role

Name

Study Root

Query/Retrieve

FIND

Study Root

Query/Retrieve

MOVE

UID

1.2.840.10008.5.1.4.1.2.2.1

1.2.840.10008.5.1.4.1.2.2.2

Name List

Implicit VR Little Endian

Implicit VR Little Endian

UID List

1.2.840.10008.1.2

1.2.840.10008.1.2

SCU

SCU

Extended

Negotiation

None

None

2.3.1.2.2.2.1 SOP Specific Conformance Statement for C-FIND SCU

After the Query operation is initiated, the DICOM Server AE will perform a study-root C-FIND-RQ request at each of the three levels (Study, Series, and Image) in succession.

The Initial Study-Level request will ask for all studies in the Remote database that match the user specified Patient name, Patient Id, Study ID, Accession number, and Study Date. Zero length data in any of those fields means match on any value. Zero length data is the default value if the user does not specify a value.

The user specified Patient name will come surrounded by single asterisks (asterisk is a wild card that matches any number of characters).

Caret (“^”) character cannot be used as a separator of Last name and First name.

The Study date is a range of dates where a date is exactly 8 characters long and the format of the

35 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

range is:

"" is any date,

"yyyymmdd-" is any date on or after that year/month/day,

"-yyyymmdd" is any date on or before that year/month/day,

"yyyymmdd-yyyymmdd" is any date on or between those dates,

"yyyymmdd" is only on that date.

The C-FIND SCU will not perform any extended negotiation and so will only perform hierarchical query. Relational Queries are not supported.

Each C-FIND SCU operation supports an “Association Timer” and “Operation Inactivity Timer” using

QueryAssociationTimeout and QueryResponseTimeout, in network-cfg.xml configuration file.

Default values are 30 seconds, 5 minutes respectively.

The user can cancel the current C-FIND request by changing the source listing to Local DB or any other remote host. If a query operation is in progress, the DICOM SERVER AE will issue a C-FIND-

CANCEL message to the remote AE to discontinue the ongoing C-FIND operation.

If a “Cancel” or “Refused” status is returned from the Remote AE the association is closed and the operation terminated.

The user can cancel the C-FIND request by moving to another network node or clicking on the local database.

The DICOM Server AE will parse each matching C-FIND-RSP reply and ignore the entries that do not contain a valid DICOM data stream. No VR validation is performed which means that syntax errors will be ignored. Tables 2.3.1.2.2.2.1-1 - 2.3.1.2.2.2.1-3 shows the various fields that are requested at the Study, Series, and Image levels of the C-FIND request.

Query results are filtered based on the Modality field. Only CT/MR Screen Save images are supported.

Table 2.3.1.2.2.2.1-1: Requested Study Level Keys

Description

Study date

Study time

Accession number

Type

R

R

R

Tag

0008,0020

0008,0030

0008,0050

Value

Zero length for any Study date, or "yyyymmddyyyymmdd" matches inclusive range of 'from-to'

Study dates. 'From' or 'To' date can be zero length meaning that side of the range is any date.

Zero length

Zero length for any

Accession number, or

Accession number matches this value

36 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

Description

Patient’s name

Patient ID

Study id

Study Instance UID

Type

R

R

R

U

Study description

Private Creator Identification

Suite Id

O

P

P

Table 2.3.1.2.2.2.1-2: Requested Series Level Keys

Description

Modality

Series number

Series Instance UID

Type

R

R

U

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Tag

0010,0010

0010,0020

0020,0010

0020,000D

0008,1030

0009,00xx

0009,xx02

Value

Zero length for any Patient name, or Patient's name matches this value

Zero length for any Patient id, or Patient id matches this value

Zero length for any Study id, or Study id matches this value

Zero length for Study level query. Study Instance UID of study for which matches are requested at a lower level

Zero length

GEMS_IDEN_01

Zero Length

Tag

0008,0060

0020,0011

0020,000E

0008,103E

0008,0070

0020,1002

Value

Zero length

Zero length

Zero length for Series level query. Series instance UID of series for which matches are requested at a lower level

Zero length

Zero length

Zero length

Series description

Manufacturer

Images in series

O

O

O

Table 2.3.1.2.2.2.1-3: Requested Image Level Keys

Description

Type

Image number R

Image Instance UID

Image type

Rows

Columns

Image position

Image orientation

Slice thickness

Slice spacing

Gantry tilt

Convolution kernel

Reconstruction diameter

O

O

O

O

U

O

O

O

O

O

O

37 of 181

Tag

0020,0013

0008,0018

0008,0008

0028,0010

0028,0011

0020,0032

0020,0037

0018,0050

0018,0088

0018,1120

0018,1210

0018,1100

Value

Zero length

Zero length

Zero length

Zero length

Zero length

Zero length

Zero length

Zero length

Zero length

Zero length

Zero length

Zero length

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

Description

Data collection diameter

Flip angle

Echo number

Echo time

Inversion time

Repetition time

Private Creator Identification

Dfov Rect

Midscan Time

Azimuth

Number of Echo

Private Creator Identification

Scout Anref

Private Creator Identification

Location RAS

Location

Center R Coordinate

Center A Coordinate

Table Start Location

Table End Location

RAS Letter for Side of Image

RAS Letter for Anterior/Posterior

RAS Letter for Scout Start Location

RAS Letter for Scout End Location

Image Dimension X

Image Dimension Y

P

P

P

P

P

P

P

P

P

P

P

P

P

P

O

O

O

P

P

P

P

P

P

Type

O

O

O

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Tag

0018,0090

0018,1314

0018,0086

0018,0081

0018,0082

0018,0080

0019,00xx

0019,xx1E

0019,xx24

0019,xx26

0019,xx7E

0021,00xx

0021,xx4A

0027,00xx

0027,xx40

0027,xx41

0027,xx42

0027,xx43

0027,xx50

0027,xx51

0027,xx52

0027,xx53

0027,xx54

0027,xx55

0027,xx60

0027,xx61

Value

Zero length

Zero length

Zero length

Zero length

Zero length

Zero length

GEMS_ACQU_01

Zero Length

Zero Length

Zero Length

Zero Length

GEMS_RELA_01

Zero Length

GEMS_IMAG_01

Zero Length

Zero Length

Zero Length

Zero Length

Zero Length

Zero Length

Zero Length

Zero Length

Zero Length

Zero Length

Zero Length

Zero Length

Note1: Refer to section 2.3.1.2.2.1 for Custom Search/Query option.

Note2: Type P refers to a private DICOM element.

During the C-FIND, the following status values are supported:

 0xFF00: Study/Series/Image items contained in identifier is collected for later display or further processing and wait for the next response from the remote host.

 0xFF01: Study/Series/Image items contained in identifier is collected for later display or further processing and wait for the next response from the remote host.

Following are the status codes that are more specifically processed when receiving messages from

Query SCP equipment:

38 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

Service

Status

Status Code

Failure A700

A900

Cancel

Success

Pending

C000-CFFF

FE00

0000

FF00

Further Meaning

REVOLUTION ACT

S

DICOM C

ONFORMANCE

Code

S

TATEMENT

Application Behavior When Receiving Status

Refused: Out of resources

Error: Identifier does not match SOP Class

Error: Unable to process

Browser displays appropriate error to user

Browser displays appropriate error to user

Matching terminated due to cancel

Matching is complete - No final identifier is supplied

Browser displays appropriate error to user

Processed and gracefully exits C-FIND request processing

Processed and gracefully exits C-FIND request processing

Matches are continuing -

Current Match is supplied and any Optional Keys were supported in the same manner as Required Keys.

Processed and the data is displayed in the

Browser

*

FF01

*

Matches are continuing -

Warning that one or more

Optional Keys were not supported for existence for this Identifier

Any other status code.

Processed and the data is displayed in the

Browser

Treated as Failure

2.3.1.2.3 Real World Activity: Manual Retrieve

2.3.1.2.3.1 Associated Real-World Activity

The operator must select the Remote DICOM AE from the “Source” pull-down on the local database browser and select the hostname of the source from where the images are to be retrieved. Once a list of Study/Series/Image is retrieved, the operator can invoke the “Retrieve” operation by clicking on the “Local DB” button at the bottom of the local database browser.

2.3.1.2.3.2 Proposed Presentation Contexts

When the Real-World activity “Manual Retrieve” is initiated all presentation contexts shown in the following table are proposed during association establishment, but only the Query/Retrieve-MOVE related contexts are applicable to this activity.

Abstract Syntax

Presentation Context Table – Proposal

Transfer Syntax Role

Name

Study Root

Query/Retrieve

MOVE

UID

1.2.840.10008.5.1.4.1.2.2.2

Name List

Implicit VR Little Endian

UID List

1.2.840.10008.1.2 SCU

Extended

Negotiation

None

39 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

Name

Study Root

Query/Retrieve

FIND

Abstract Syntax

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Presentation Context Table – Proposal

Transfer Syntax Role

UID

1.2.840.10008.5.1.4.1.2.2.1

Name List

Implicit VR Little Endian

UID List

1.2.840.10008.1.2 SCU

Extended

Negotiation

None

2.3.1.2.3.2.1 SOP Specific Conformance Statement for C-MOVE SCU

When the operator starts a Move operation at any level (Study, Series, Image) the DICOM Server AE will initiate a C-MOVE-RQ request with UIDs at the appropriate level (Study, Series, Image) to the

Remote AE with the DICOM Server AE as the Destination AE. The Storage SCP (“Receive Images”) will handle the incoming images as described in section 2.3.1.3.1.

C-MOVE supports list of Studies (through list of Study UIDs), list of Series (through list of Series UIDs) and list of Images (through list of Instances UIDs).

Each C-MOVE SCU operation supports an “Association Timer” and “Operation Inactivity Timer” with time out values of 30 seconds and 15 seconds respectively.

If the C-MOVE SCU receives a status different from success (0x0000) or pending (0xFF00) during the association, the DICOM SERVER AE will release the association. This information will be logged in the system log files and the network queue will be updated accordingly.

During Manual Retrieve, the DICOM SERVER AE is able to generate a C-MOVE-CANCEL if the operator

Pauses or Deletes the retrieve job.

When receiving a Cancel request response (0xFE00), the DICOM SERVER AE will release the association. This information will be logged in the system log files and the network queue will be updated accordingly.

Following are the status codes that are more specifically processed when receiving messages from

Retrieve SCP equipment:

Service

Status

Status Code Further Meaning Application Behavior When Receiving Status

Code

Failure A701

A702

A801

A900

Refused: Out of resources -

Unable to calculate number of matches

System will display the message in job manager user interface and retry the operation after configured time is elapsed.

Refused: Out of resources -

Unable to perform suboperations

System will log the message and retry the operation after configured time is elapsed.

Refused: Move Destination

Unknown

Error: Identifier does not match SOP Class

System will display the message in job manager user interface.

System will display the message in job manager user interface.

40 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

C000-CFFF

Cancel FE00

Warning B000

Success 0000

Pending FF00

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Error: Unable to process

Sub-operations terminated due to a Cancel indication

System will display the message in job manager user interface.

Operation will be terminated and the system will display the message in job manager user interface.

System will log the information.

Sub-operations Complete -

One or more Failures.

Sub-operations Complete - No

Failure.

Processed and gracefully exits C-MOVE request processing

Sub-operations are continuing

-

System processes the information and displays the progress in the job manager UI

Any other status code. Treated as Failure * *

The DICOM Server AE will send a C-CANCEL-MOVE-RQ to the Remote AE if the operator “Pauses” or

“Clears” the job from the local database manager Network queue.

2.3.1.2.4 Real-World Activity: Verify (DICOM Ping)

2.3.1.2.4.1 Associated Real-World Activity

The operator shall select a Remote DICOM AE from the “Network Configuration” window and click on

“Ping” on the right side-bar. The DICOM server will initiate an association with the remote DICOM AE in order to verify communication at the application level. The success or failure of the verification process is displayed to the user.

If the C-ECHO response is received with a success, the DICOM Server will post a pop-up to the operator indicating that the remote device is alive.

2.3.1.2.4.2 Proposed Presentation Context Table

Refer to the following table for the Proposed Presentation Contexts for DICOM Server AE and Real-

World activity Verification

Abstract Syntax

Presentation Context Table – Proposal

Transfer Syntax Role

Name

Verification

UID

1.2.840.10008.1.1

Name List

Implicit VR Little

Endian

UID List

1.2.840.10008.1.2 SCU

Extended

Negotiation

None

41 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

2.3.1.2.4.2.1 SOP Specific DICOM Conformance C_ECHO SCU

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

The GEHC CT DICOM Server AE provides standard conformance to the DICOM Verification Service

Class.

Each ECHO operation supports “Association Timer” with EchoAssociationTimeout, in networkcfg.xml configuration file. Default values are 20 seconds.

Upon receiving a C-ECHO-RSP, containing a successful status, a message will be posted to the operator indicating success.

2.3.1.2.5 Real-World Activity: Remote Archive

2.3.1.2.5.1 Associated Real World Activity:

“Choose Archive Save Option”

The operator must first set remote hosts from Network Configuration from “Tools“ pull-down menu. If the remote node is already present, select one remote host, click “Edit“ on the right side-bar and make a check on “Archive Node” check box. If the remote host is not present, add the remote host with check on “Archive Node” check box.

Note: The remote node should be a Storage Commitment SCP. STC AE Title can be different from

STORE AE title. The user selects the exam/series to be committed and click on the archival destination from the archive list at the bottom of the local database browser.

All the images currently in the selected exam/series will be sent to the selected remote archive node

(which is also the Storage commitment SCP) using DICOM C-STORE operations. Once all the images are transferred the commitment request will be sent on a separate association.

2.3.1.2.5.2 Associated Real World

Activity: “Auto Archive Exam/Series”

The Auto archive API’s are used to archive the exams/series onto local archive media or the remote archive node (which shall be a Storage Commitment SCP) without manual interface. If the default device selected for Auto Archive is a remote Storage Commitment SCP then all the images currently in the specified exam/series will be sent to the selected Storage commitment SCP using C-STORE operations. On successful transfer of all the images the Storage Commitment request will be sent.

2.3.1.2.5.3 Proposed Presentation Context

The Proposed Presentation Context table for the DICOM Storage Commitment SCU is as shown in following Table.

Presentation Context Table

Abstract Syntax

Name UID

Storage

Commitment

Push Model SOP

Class

1.2.840.10008.1.

20.1

Transfer Syntax

Name UID

DICOM Implicit VR Little

Endian Transfer Syntax

1.2.840.10008.1.

2

Role Extended

Negotiatio n

SCU None

42 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

2.3.1.2.5.3.1 SOP Specific DICOM Conformance Statement for the Storage Commitment Push Model SOP Class SCU (N-

ACTION)

The Storage Commitment SCU can send the commitment request for following

DICOM SOP classes.

NAME

CT Image Storage SOP Class

UID

1.2.840.10008.5.1.4.1.1.2

PET Image Storage SOP Class

MR Image storage SOP Class

Secondary Capture Storage SOP Class

X-Ray Radiation Dose SR

Enhanced SR

RT Structure Set Storage

1.2.840.10008.5.1.4.1.1.128

1.2.840.10008.5.1.4.1.1.4

1.2.840.10008.5.1.4.1.1.7

1.2.840.10008.5.1.4.1.1.88.67

1.2.840.10008.5.1.4.1.1.88.22

1.2.840.10008.5.1.4.1.1.481.3

Grayscale Softcopy Presentation State Storage SOP Class 1.2.840.10008.5.1.4.1.1.11.1

The images in the selected exam/series are sent to the remote commitment provider entity using

DICOM C-STORE operations. If there are any failures in the image transfers the Storage

Commitment request will not be sent. The corresponding job will be marked as failed and user will be notified of the status of job.

If all the images are successfully transferred then the commitment request will be sent on a different association with the list of SOP instances.

If the N-Action request fails, the job will be marked as failed otherwise following sequence of actions will take place.

The SCU waits for N_ACTION response from the provider on the same association for a configurable amount of time. Default time-out is 600 seconds. If it does not receive N_ACTION response during this time it closes the association and marks the job as failed.

If the received N-ACTION Response from the Storage Commitment Provider has a success status, the DICOM SERVER AE waits 10 seconds for an N-EVENT-REPORT on the same association. This time-out is not configurable. If the N-EVENT-REPORT request is not received in this time, it closes the association and changes the Job state to ”Waiting” indicating the job is waiting for the response from commitment provider.

The DICOM SERVER AE can receive N-EVENT-REPORT from the Storage Commitment Provider at any time (See section 2.3.1.3.5 - Real-World Activity: Listen to remote Storage Commitment SCP”).

A New transaction UID will be created for each retry by user. The old transaction uid is not applicable for these requests.

Following are the status codes that are more specifically processed when receiving N-Action responses from Storage Commitment SCP equipment:

43 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Status Codes Received by DICOM SERVER AE for Activity Remote Archive

Service

Status

Status

Code

Further Meaning

Failure 0119 Class-instance conflict

Application Behavior When Receiving

Status Code

System displays the appropriate error message in job manager user interface

0115 Invalid argument value

0117

0212

0114

0118

0110

Invalid SOP Instance

Mistyped argument

No such argument

No such SOP Class

Processing failure

System displays the appropriate error message in job manager user interface

System displays the appropriate error message in job manager user interface

System displays the appropriate error message in job manager user interface

System displays the appropriate error message in job manager user interface

System displays the appropriate error message in job manager user interface

System displays the appropriate error message in job manager user interface

Success

*

0213 Resource limitation

0211

0000

*

Unrecognized operation

Any other status code.

System logs the appropriate error message and retries the operation after the configured time is elapsed.

System displays the appropriate error message in job manager user interface

The request for storage comment is considered successfully sent. System displays “waiting” message in job manager user interface.

System displays the appropriate error message in job manager user interface.

2.3.1.2.5.3.2 Storage Commitment Push Model SOP Class Request Processing

The Following DIMSE service Elements are supported for the Storage Commitment request and processing.

N-ACTION – Requests the remote Storage Commitment SCP to commit to storing the image instances.

The following attributes are sent as part of the DATA Set for the N-ACTION request.

Attribute

Transaction UID

Referenced SOP Sequence

SOP Class UID

SOP Instance UID

Tag

(0008,1195)

(0008,1199)

(0008,1150)

(0008,1155)

Value

Transaction UID

Referenced Study Component sequence attribute is not sent.

Storage Media File-Set ID and Storage Media File-Set UID attributes are not supported.

44 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

2.3.1.2.5.3.3 Storage Commitment Push Model SOP Class Response Processing

The following DIMSE Service Elements are supported for the Storage Commitment response processing

N-EVENT-REPORT – The Response sent by the remote Storage Commitment SCP.

Once the N-EVENT REPORT is received, the following actions will be taken depending on the status of the response.

2.3.1.2.5.3.3.1 Commit response with SUCCESS status

The Archive flag information in the browser for all the successful instances will be updated. The status will be changed to “Y”.

The job queue entry will be removed

The following attributes are expected as part of DATA Set for N-Event-REPORT from SCP

Attribute

Transaction UID

Referenced SOP Sequence

SOP Class UID

SOP Instance UID

Tag

(0008,1195)

(0008,1199)

(0008,1150)

(0008,1155)

Value

Value received from SCP

Value received from SCP

Value received from SCP

Value received from SCP

N-EVENT-REPORT-RSP will be sent on the same association itself. No DATA Set will be sent along with the response.

2.3.1.2.5.3.3.2 Commit response with FAILURE status

The following attributes are expected as part of DATA Set for N-EVENT-REPORT from SCP:

Attribute

Transaction UID

Failed SOP Sequence

> SOP Class UID

> SOP Instance UID

>Failure Reason

Tag

(0008,1195)

(0008,1198)

(0008,1150)

(0008,1155)

(0008,1197)

Value

Value received from SCP

Value received from SCP

Value received from SCP

Value received from SCP

Value received from SCP

In case of complete/partial failure the user will be notified about the status and the job entry will be paused. There is no attempt made to retry automatically the failed sop instances. However the user can manually retry the failed jobs. Such requests will be treated as new requests. This will go through the whole sequence of operations once again.

45 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

The failure reason is ignored.

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Failed SOP instances will have their archive flag information unaltered.

Note: The archive status flag in the browser is a shared flag with local archive. When the status is

“Y”, it means that the images are archived but doesn’t specify whether on local archive device or remote archive device. It is left to the user’s discretion whether the local sop instances (with their archive flag set to “Y”) are to be deleted.

N-EVENT-REPORT-RSP will be sent on the same association itself. No DATA Set will be sent along with the response.

Note: (0008, 1199) Reference SOP Sequence is not handled.

Refer Section - 5 Storage Commitment Push Model Implementation.

2.3.1.2.6 Real-World Activity: Modality Worklist Query

2.3.1.2.6.1 Associated Real-World Activity

The operator of the system initiates a query for a modality worklist by either opening the Schedule screen or by opening the Schedule screen and pressing the Update button. The choice of which of these two behaviors occurs is user configurable. The Worklist Server will then initiate an association with the remote AE in order to query for the worklist

A user can configure a number of parameters that directly control the worklist query request. The user can request worklist items that are intended for the scanner the user is working at, all items that apply to the modality of the scanner the user is working at or all worklist items available. These selections and their effects on worklist query parameters are given below:

This Scanner:

(CT)

All Systems

Modality, (0008,0060) – set to CT

Scheduled Station AE Title, (0040,0001) – set to local AE title

Modality, (0008,0060) - set to CT

Scheduled Station AE Title, (0040,0001) - zero-length (universal matching)

Modality, (0008,0060) - zero-length (universal matching)

Scheduled Station AE Title, (0040,0001) - zero-length (universal matching)

The scheduled dates of procedures of interest can be specified for query by selecting a specific date range. The date ranges available are Today, Days Before Today, Days After Today and All Days. These selections and their effects on worklist query parameters are given below:

Today: Scheduled Procedure Step Start Date (0040,0002) - set to

YYYYMMDD, where this date is the current date.

Days Before Today and

Days After Today:

Scheduled Procedure Step Start Date (0040,0002) - set to

YYYYMMDD-YYYYMMDD, where this date range represents the specified number of days before today and/or after today. Note

46 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

that number of days both before and after can be specified in the same query and that each always includes today.

All Days: Scheduled Procedure Step Start Date (0040,0002) - zero-length

(universal matching)

The system also supports a Custom Query option where the user can filter based on Patient Name,

Patient ID, Accession Number or Requested Procedure Step ID using a User Interface. Refer sec

6.5.2 Setting Custom Query Option

2.3.1.2.6.2 Proposed Presentation Context Table

The following table shows the proposed presentation contexts for the Worklist Server AE after realworld activity “Worklist Query” has been initiated:

Presentation Context Table – Proposed by DICOM SERVER AE for Activity

Modality Worklist Query

Abstract Syntax Transfer Syntax

Name

Modality Worklist

Information

Model – FIND

UID

1.2.840.10008.5.1.4.

31

Name List

Implicit VR Little

Endian

Explicit VR Little

Endian

Explicit VR Big Endian

UID List

1.2.840.10008.1.

2

1.2.840.10008.1.2.

1

1.2.840.10008.1.2.

2

Role

Extended

Negotiatio n

SCU None

2.3.1.2.6.2.1 SOP Specific DICOM Conformance Statement for the Modality Worklist Information Model FIND SOP Class

If the remote AE does not support the proposed Presentation Context, an appropriate error is logged and the operator is notified.

This implementation can receive multiple C-FIND results over a single association. Only one association is opened at a time.

Each C-FIND response received from the remote AE is parsed to verify the length/type of the items in the response. Upon detecting any error in the response data, the response is discarded and the next response (if any) is considered.

Each C-FIND SCU operation supports an “Association Timer” and “Operation Inactivity Timer” using

QueryAssociationTimeout and QueryResponseTimeout configuration parameters. Default values are

30 seconds, 5 minutes respectively.

All errors and failures detected by the DICOM SERVER AE are logged to the service log files.

In case of ERROR, persisted worklist item(s) got from previous request to the configured SCP would be

47 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

shown.

User can cancel the ongoing worklist query by clicking on the Cancel button in the UI. Upon cancel, the DICOM SERVER AE will send C-FIND-CANCEL request to the Modality Worklist SCP.

Following are the status codes that are more specifically processed when receiving messages from a

Modality Worklist SCP equipment :

Service Status Status Code

Failure A700

Further Meaning Application Behavior

When Receiving

Status Code

Refused: Out of resources Logs the error and displays all the persisted worklists

Cancel

A900

C001

FE00

Error: Identifier does not match SOP Class

Error: Unable to process

Logs the error and displays all the persisted worklists

Logs the error and displays all the persisted worklists

Closes the association

Success

Pending

*

0000

FF00

FF01

*

Matching terminated due to cancel

Matching is complete -

No final identifier is supplied

Matches are continuing -

Current Match is supplied and any Optional Keys were supported in the same manner as

Required Keys.

Matches are continuing -

Warning that one or more Optional Keys were not supported for existence for this

Identifier

Any other status code.

Displays all the worklists items obtained from the SCP

Continue to wait for worklists items from

SCP

Continue to wait for worklists items from

SCP

Worklist query operation is considered as failure and the error message is displayed to the user. Persisted worklist item(s) got from previous request to the SCP will be shown

2.3.1.2.6.2.2 Record Acceptance Policy

The GEHC CT implementation adheres to strict value checking of incoming query responses from the remote AE. Each response received is examined to verify that all Type 1 attributes are present with non-zero length, that all Type 2 attributes are present (possibly with zero length) and that the data for all attributes is consistent with respect to the attributes’ value representation (VR).

48 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Any inconsistencies in the response data, with respect to the categories described above, are considered errors. Upon detecting any such errors in the response data, the Worklist Server AE will issue a C-FIND-CANCEL and, upon receipt of a C-FIND-RSP (or if an applicable timer expires), will abort the association. All previously received worklist items are retained. Note that the absence of requested Type 3 attributes is not considered an error.

Fields considered Type 1 by the Worklist Server include:

 (0010,0010), Patient Name

 (0010,0020), Patient ID

 (0020,000D), Study Instance UID

 (0040,0001), Scheduled Station AE Title

 (0040,0002), Scheduled Procedure Step Start Date

1

 (0040,0003), Scheduled Procedure Step Start Time

1

 (0040,0009), Scheduled Procedure Step ID

 (0040,1001), Requested Procedure ID

Start Date must be of the form YYYYMMDD, exactly eight numeric characters, and Start Time must be of the form HHMMSS, exactly six numeric characters.

Fields considered Type 2 by Worklist Server include:

 (0008,0050), Accession Number

 (0008,0060), Modality

 (0008,0090), Referring Physician Name

 (0010,0030), Patient Date of Birth

 (0010,0040), Patient Sex

 (0010,1030), Patient Weight in kg

 (0010,2000), Medical Alerts

 (0010,2110), Contrast Allergies

 (0010,21C0), Pregnancy Status

 (0032,1032), Requesting Physician

 (0032,1070), Requested Contrast Agent

 (0038,0010), Admission ID

 (0038,0050), Special Needs

 (0038,0300), Current Patient Location

 (0038,0500), Patient State

 (0040,0006), Performing Physician

 (0040,0010), Scheduled Station Name

 (0040,0011), Scheduled Procedure Step Location

 (0040,0012), Pre-order Medication

 (0040,1003), Requested Procedure Priority

 (0040,1004), Patient Transport Arrangements

 (0040,3001), Confidentiality Constraint

Refer Section.6 Modality Worklist Query Implementation

49 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

2.3.1.2.7 Real-World Activity: Modality Performed Procedure Step Notification

The PPS Server AE is implemented as an application process on the scanner host computer. It runs as a daemon serving requests from other applications to send the PPS information to the remote AE and return the results to the requesting application.

The PPS Server AE initiates the following functions.

Start PPS: Initiates a DICOM association in order to create a DICOM Modality Performed Procedure

Step SOP instance in the remote AE. If the remote AE accepts a presentation context applicable to

Modality performed Procedure Step, the PPS Server AE will issue a request to create the SOP instance in the remote AE via the N-CREATE service.

Complete PPS: Initiates a DICOM association in order to update a DICOM Modality Performed Step instance that is already created with the remote AE. If the remote AE accepts a presentation context applicable to Modality performed Procedure Step, the PPS Server AE will issue a request to update the SOP instance in the remote AE via the N-SET service. The PPS Status is set to

‘COMPLETED’.

Discontinue PPS: Initiates a DICOM association in order to update a DICOM Modality Performed

Step instance that is already created with the remote AE. If the remote AE accepts a presentation context applicable to Modality performed Procedure Step, the PPS Server AE will issue a request to update the SOP instance in the remote AE via the N-SET service. The PPS Status is set to

‘DISCONTINUED’.

2.3.1.2.7.1 Sequencing of Real-World Activities

2.3.1.2.7.1.1 PPS from Acquisition System with MWL data

The system has a Modality Work-list Server AE installed. Work-List information is obtained from HIS/RIS system through the use of Basic Work-list Management Service. Use of the information retrieved in the creation of Image SOP instance is described in the Modality Work-list Conformance statement. Use of the information retrieved in MPPS SOP instances is described later in this document.

 The system initiates a ‘Start PPS’ before starting a scan, i.e. when the image acquisition is started.

The system retrieves necessary information related to the Scheduled Procedure Step from Modality

Work-list Server. PPS Server AE initiates a MPPS (Modality Performed Procedure Step) N-CREATE request to the remote AE (MPPS SCP), in-order to create a MPPS SOP instance at the remote AE.

 The MPPS SCP returns response indicating the success/failure of the request execution. The PPS state information is updated in the system based on the response data, and is presented to the user. The DICOM association is closed.

 System includes the necessary information related to Scheduled Procedure Steps and the

Performed procedure Step in the image instances created.

 At the end of image acquisition, system initiates a ‘Complete PPS’ or ‘Discontinue PPS’ based on the choice selected by the user using the user interface provided. The user is also given a choice

‘Defer PPS’ which is described below. PPS Server AE initiates a MPPS N-SET request to the remote

50 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

AE, in-order to update the MPPS SOP instance that is already created. The N-SET is sent over a

new DICOM association

 The remote AE returns response indicating the success/failure of the request execution. The PPS state information is updated in the system based on the response data, and is presented to the user.

 At the end of image acquisition, if the user has chosen ‘Defer PPS’, the user is provided with an interface to ‘Complete PPS’ or ‘Discontinue PPS’ at any later time. The user might wish to alter the image set generated through acquisition, before invoking these operations. Note that the user explicitly uses the user interface provided to invoke this operation, as in the case of PPS generated for post-processing, which is described in the following section. PPS messages N-CREATE ( if

applicable) and N-SET will be sent over the same DICOM association

 The remote AE returns response indicating the success/failure of the request execution. The PPS state information is updated in the system based on the response data, and is presented to the user.

2.3.1.2.7.1.1.2 PPS from acquisition system without MWL data

The system either does not have a Modality Work-list Server AE installed or a Modality Work-list Server

AE installed but no Work-List information is obtained from HIS/RIS system for the current procedure that is being performed. The information required for performing the procedure is supplied through the user interface of the system. The use of this information in MPPS SOP instances is described later in this document.

 The system initiates a ‘Start PPS’ before starting a scan, i.e. when the image acquisition is started.

PPS Server AE initiates a MPPS (Modality Performed Procedure Step) N-CREATE request to the remote AE (MPPS SCP), in-order to create a MPPS SOP instance at the remote AE.

 The MPPS SCP returns response indicating the success/failure of the request execution. The PPS state information is updated in the system based on the response data, and is presented to the user.

 System includes the necessary information related to Scheduled Procedure Steps and the

Performed procedure Step in the image instances created.

 At the end of image acquisition, system initiates a ‘Complete PPS’ or ‘Discontinue PPS’ based on the choice selected by the user using the user interface provided. The user is also given a choice

‘Defer PPS’ which is described below. PPS Server AE initiates a MPPS N-SET request to the remote

AE, in-order to update the MPPS SOP instance that is already created.

 The remote AE returns response indicating the success/failure of the request execution. The PPS state information is updated in the system based on the response data, and is presented to the user.

 At the end of image acquisition, if the user has chosen ‘Defer PPS’, the user is provided with an interface to ‘Complete PPS’ or ‘Discontinue PPS’ at any later time. The user might wish to alter the image set generated through acquisition, before invoking these operations. Note that the user

51 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

explicitly uses the user interface provided to invoke this operation, as in the case of PPS generated for post-processing, which is described in the following section.

 The remote AE returns response indicating the success/failure of the request execution. The PPS state information is updated in the system based on the response data, and is presented to the user.

2.3.1.2.7.1.1.3 PPS from post-processing system

 The user initiates post-processing on the images generated through acquisition.

 The system creates a Modality performed Procedure Step instance locally in the system. If the source image instance has the Scheduled Procedure Step information, it is copied into the image instances created. Also the system includes the necessary information related to the Modality

Performed Procedure Step into the image instance.

 At the end of (one or more) post-processing, the user initiates ‘Complete PPS’ or ‘Discontinue PPS’ through the user interface provided. PPS Server AE initiates a MPPS (Modality Performed Procedure

Step) N-CREATE request to the remote AE (MPPS SCP), in-order to create a MPPS SOP instance at the remote AE (which is actually a replica of the locally created MPPS SOP instance).

 The remote AE returns response indicating the success/failure of the request execution. If the response indicates success, PPS Server AE initiates a MPPS N-SET request to the remote AE, inorder to update the MPPS SOP instance that is already created, with the additional information.

 The remote AE returns response indicating the success/failure of the request execution. The PPS state information is updated in the system based on the response data, and is presented to the user.

2.3.1.2.7.2 Proposed Presentation Context Table

The following table shows the proposed presentation contexts for the PPS Server AE after any of the real-world activity listed in section 2.3.1.6.1 Sequencing of Real-World Activities, is initiated.

Presentation Context Table – Proposed

Abstract Syntax

Name

Modality

Performed

Procedure Step

UID

Transfer Syntax

Name List

1.2.840.10008.3.1.2.3.3 Implicit VR

Little Endian

Explicit VR

Little Endian

Explicit VR

Big Endian

UID List

1.2.840.10008.1.2

1.2.840.10008.1.2.1

1.2.840.10008.1.2.2

Role Extended

Negotiation

SCU None

2.3.1.2.7.2.1 SOP Specific DICOM Conformance Statement for MPPS SOP class

Only one association is opened at a time. The association is open until the response message is returned from the SCP or a network error occurs. No other configurable timers are available.

52 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

If the remote AE does not support the proposed Presentation context, an appropriate error message is logged.

Following are the status codes that are more specifically processed when receiving an N-CREATE and N-SET response from an MPPS SCP equipment:

Service Status Status Code Further Meaning Application Behavior

When Receiving

Status Code

Success

Failure

0000

0105

0106

0110

0112

0115

0117

0118

0119

0120

Success

No such attribute

Invalid attribute value

Processing failure

No such SOP Instance

Invalid argument value

Invalid SOP Instance

No such SOP Class

Class-instance conflict

Missing attribute

System updates the state and indicates the same on the user interface

Error message is logged into system log-file and indicates the status on the user interface

Error message is logged into system log-file and indicates the status on the user interface

Error message is logged into system log-file and indicates the status on the user interface. Retries the operation after the configured time is elapsed

Error message is logged into system log-file and indicates the status on the user interface

Error message is logged into system log-file and indicates the status on the user interface

Error message is logged into system log-file and indicates the status on the user interface

Error message is logged into system log-file and indicates the status on the user interface

Error message is logged into system log-file and indicates the status on the user interface

Error message is logged into system log-file and indicates the status on the user interface

53 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

*

0144

0210

0211

0212

0213

C002

*

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

No such argument

Duplicate invocation

Unrecognized operation

Mistyped argument

Resource limitation

Operation failed

Any other status Code

Error message is logged into system log-file and indicates the status on the user interface

Error message is logged into system log-file and indicates the status on the user interface

Error message is logged into system log-file and indicates the status on the user interface

Error message is logged into system log-file and indicates the status on the user interface

Error message is logged into system log-file and indicates the status on the user interface. Retries the operation after the configured time is elapsed

Error message is logged into system log-file and indicates the status on the user interface. Retries the operation after the configured time is elapsed

The operation is deemed

‘Failed’. Detailed message is logged into system logfile and indicates the status on the user interface. Retries the operation after the configured time is elapsed

Refer Section.7 Modality Performed Procedure Step Implementation.

2.3.1.2.8 Real-World Activity: Print

2.3.1.2.8.1 Associated Real-World Activity

The Film Composer allows the user to select printers and it also allows the user to drag and drop the images (from the viewer application) into the film. It also allows the user to manipulate some print parameters like film format and number of copies to print. When the user presses the "Print" Button, the Film composer communicates this request to Print SCU, which then tries to establish the association with the requested printer and transmits the images for printing.

54 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

2.3.1.2.8.1.1 Sequencing of Real-World Activities

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

The DICOM printer is installed through a camera installation process. The DICOM printer is selected from the Film Composer Interface for Manual Filming. Autofilming is enabled using the camera setup via the camera installation process.

Images may be acquired for printing in the following manner:

– Images to be printed may be manually dropped from Viewing applications into the manual film composer slots.

Images to be printed may be automatically dropped from Viewing applications into the manual film composer slots.

Images to be printed may be manually dropped from Viewing applications into the auto film composer slots.

Image series to be printed may be automatically dropped from the

Print Series application into the manual film composer slots.

Images to be printed may be automatically dropped from the

Scanning Application into the auto film composer slots.

Based upon the filming mode (and the attributes) used to acquire images into the film composer, films will be automatically printed when the film session is full, or the user must press the Print Button to print the images.

The Print SCU will start the Print Session. The Print Session involves establishing association with the printer followed by the next sequence of activities:

-

The Print SCU gets the Print SCP status using N-GET service. If the Printer returns FAILURE status the print session will be terminated and the requester will be notified of the printer status.

-

The film session is created using N-CREATE service. In case of error, the print session will be terminated. The attribute values for the Film session will be specified with the film session request.

-

The film box is created using N-CREATE service. The print session will be terminated if the printer fails to create the film box. The film box attribute values will be sent in the film box create request

-

The image attributes for the images to be printed in this session will be set using the N-SET service. If the printer fails to accommodate the images in the image set, the print session will be terminated. The film will be printed using the N-ACTION service. Only film box printing is supported. In case of error, the print session will be terminated.

-

The Film Box instance will be deleted using the N-DELETE service.

-

The SCU does not wait for N-EVENT-REPORT from the Printer after deleting the film box instance. The N-EVENT-REPORT received when the association was still active is handled, but the data received will be ignored.

55 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

-

Finally the association will be terminated and if all the above operations are successful the requester will be notified of the successful print session. This status just indicates that the images to be printed have been successfully sent to the printer.

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

2.3.1.2.8.2 Proposed Presentation Context

The Proposed Presentation Context Table for the Print SCU is as shown in following Table.

Table 2.3.1.2.8.2-1 Presentation Context– Proposed by DICOM SERVER AE for Activity Print

Abstract Syntax

Presentation Context Table – Proposed

Transfer Syntax

Name

Basic Grayscale

Print

Management

Meta SOP Class

SOP CLASS UID

1.2.840.10008.5.1.1.

9

Basic Color Print

Management

Meta SOP Class

1.2.840.10008.5.1.1.

18

Printer SOP Class 1.2.840.10008.5.1.1.

16

Name

Implicit VR

Little Endian

Explicit VR

Little Endian

Explicit VR Big

Endian

Implicit VR

Little Endian

Explicit VR

Little Endian

Explicit VR Big

Endian

Implicit VR

Little Endian

Explicit VR

Little Endian

Explicit VR Big

Endian

UID

1.2.840.10008.1.

2

1.2.840.10008.1.

2.1

1.2.840.10008.1.

2.2

1.2.840.10008.1.

2

1.2.840.10008.1.

2.1

1.2.840.10008.1.

2.2

1.2.840.10008.1.

2

1.2.840.10008.1.

2.1

1.2.840.10008.1.

2.2

Role Extended

Negotiatio n

SCU None

SCU

SCU

None

None

Note:

Support of the Basic Grayscale Print Management Meta SOP Class as an SCU mandates support for the Basic Film Session, Basic Film Box, Basic Grayscale Image Box, and Printer SOP Classes as an

SCU.

Support of the Basic Color Print Management Meta SOP Class as an SCU mandates support for the

Basic Film Session, Basic Film Box, Basic Color Image Box, and Printer SOP Classes as an SCU.

56 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

2.3.1.2.8.2.1

SOP Classes

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

SOP Specific DICOM Conformance Statement for Basic Grayscale and Color Print Management

The DICOM SERVER AE uses the following DIMSE services of the supported SOP Classes for Activity

Print:

Table 2.3.1.2.8.2.1-1 DIMSE Services used by DICOM SERVER AE for Activity Print

SOP Class SOP Class UID DIMSE Service

Element

SCU

Usage

Basic Film Session 1.2.840.10008.5.1.1.1 N-CREATE

Basic Film Box

Basic Grayscale

Image Box

1.2.840.10008.5.1.1.2

1.2.840.10008.5.1.1.4

N-SET

N-DELETE

N-ACTION

N-CREATE

N-ACTION

N-DELETE

N-SET

N-SET

Used

(Mandatory)

Not Used

Not Used

Not Used

Used

(Mandatory)

Used

(Mandatory)

Used

Not Used

Used

(Mandatory)

N-SET Basic Color Image

Box

Printer

1.2.840.10008.5.1.1.4.1

1.2.840.10008.5.1.1.16 N-EVENT-

REPORT

N-GET

Used

(Mandatory)

Used

(Mandatory)

Used

57 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

2.3.1.2.8.2.1.1 Basic Film Session SOP Class

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

The DICOM SERVER AE supports the following DIMSE Service Elements for the Basic Film Session SOP

Class.

- N-CREATE - Requests the Print SCP to create an instance of Basic Film Session.

Only one association is opened at a time. The association is open until the response message is returned from the SCP or a network error occurs. No other configurable timers are available.

Following are the status codes that are more specifically processed when receiving messages from a

Print SCP equipment for the Basic Film Session SOP Class N-CREATE:

Table 2.3.1.2.8.2.1.1-1 Status Codes Received by DICOM SERVER AE for Basic Film Session SOP Class N-

CREATE

Service

Status

Status

Codes

Further Meaning

Failure 0119 Class-instance conflict

0210 Duplicate invocation

0111 Duplicate SOP Instance

0106 Invalid attribute value

0117 Invalid SOP instance

0120 Missing attribute

0121 Missing attribute value

0212 Mistyped argument

0105 No such attributes

0118 No such SOP Class

Application Behavior When receiving

Status Codes

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

58 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

0112

0110

0213

0211

No such SOP Instance

Processing failure

Resource limitation

Unrecognized operation

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI.

Resource limitation error details is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Warning B600 Memory allocation not supported Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Success 0000 Film session successfully created Logs film session instance UID and proceeds for creating film box.

* * Any other status code. Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

If a Failure/Warning status is returned by the Print SCP during the N-CREATE operation of the Film

Session, the association will be terminated.

2.3.1.2.8.2.1.2 Basic Film Box SOP Class

The DICOM SERVER AE supports the following DIMSE Service Elements for the Basic Film Box SOP

Class:

- N-CREATE - Requests the Print SCP to create an instance of Film Box.

- N-ACTION - Requests the Print SCP to print the Film Box onto Printer.

- N-DELETE - Requests the Print SCP to delete the Film Box Instance.

Only one association is opened at a time. The association is open until the response message is returned from the SCP or a network error occurs. No other configurable timers are available.

Following are the status codes that are more specifically processed when receiving messages from a

Print SCP equipment for the Basic Film Box SOP Class N-CREATE:

Table 2.3.1.2.8.2.1.2-1 Status Codes Received by DICOM SERVER AE for Basic Film Box SOP Class N-CREATE

Service

Status

Status

Codes

Further Meaning

Failure C616 There is an existing Film Box that has not been printed and N-

ACTION at the Film Session level is not supported. A new Film Box will not be created when a previous

Film Box has not been printed.

Application Behavior When receiving

Status Codes

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

59 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

0119 Class-instance conflict

0210

0111

0106

0117

0120

0121

0212

0105

0118

0112

0110

0213

0211

Duplicate invocation

Duplicate SOP Instance

Invalid attribute value

Invalid SOP instance

Missing attribute

Missing attribute value

Mistyped argument

No such attributes

No such SOP Class

No such SOP Instance

Processing failure

Resource limitation

Unrecognized operation

Warning B605 Requested Min Density or Max

Density outside of printer’s operating range. The printer will use its respective minimum or maximum density value instead.

Success 0000 Film box successfully created

* * Any other status code.

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI.

Resource limitation error details is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on film composer UI. Error details along with the status code is logged.

Proceed for setting image box.

Pauses print job and displays print failure message on the film composer UI. Error

60 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

details along with the status code is logged.

If a Failure or Warning status is returned during the N-CREATE operation of the Film Box, the association will be terminated.

Following are the status codes that are more specifically processed when receiving messages from a

Print SCP equipment for the Basic Film Box SOP Class N-ACTION:

Table 2.3.1.2.8.2.1.2-2 Status Codes Received by DICOM SERVER AE for Basic Film Box SOP Class N-ACTION

Service Status Further Meaning Application Behavior When receiving

Status

Failure

Codes

C602 Unable to create Print Job SOP

Instance; print queue is full

Status Codes

Logs “unable to create print job” error.

Pauses print job and displays print failure message on the film composer UI.

C603 Image size is larger than image box size (by using the specified magnification value)

C604 Image position collision : multiple images assigned to single image position

C613 Combined Print Image size is larger than the Image Box size

0119 Class-instance conflict

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

0210 Duplicate invocation

0115 Invalid argument value

0117 Invalid SOP Instance

0212 Mistyped argument

0123 No such action

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

0114 No such argument

0118 No such SOP Class

0112 No such SOP Instance

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

61 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

0110 Processing failure

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

0213 Resource limitation

0211 Unrecognized operation

Warning B603 Film Box SOP Instance hierarchy does not contain Image Box SOP

Instances (empty page)

B604 Image size is larger than image box size, the image has been demagnified.

B609 Image size is larger than the Image

Success

*

Box size. The Image has been cropped to fit.

B60A Image size or Combined Print Image size is larger than the Image Box size. Image or Combined Print

Image has been decimated to fit.

0000 Film accepted for printing; if supported, the Print Job SOP

Instance is created

* Any other status code.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Proceed in parsing N-Action response.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

If a Failure or Warning status is returned during the N-ACTION operation of the Film Box, the association will be terminated.

Following are the status codes that are more specifically processed when receiving messages from a

Print SCP equipment for the Basic Film Box SOP Class N-DELETE:

Table 2.3.1.2.8.2.1.2-3 Status Codes Received by DICOM SERVER AE for Basic Film Box SOP Class N-DELETE

Service

Status

Status

Codes

Further Meaning

Failure 0119 Class-instance conflict

0210 Duplicate invocation

0117 Invalid SOP instance

Application Behavior When receiving

Status Codes

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error

62 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

0212 Mistyped argument

0118 No such SOP Class

0112 No such SOP Instance

0110 Processing failure

0213 Resource limitation

0211 Unrecognized operation

Success 0000 Film box successfully deleted

* * Any other status code.

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Logs the success message.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

If a Failure or Warning status is returned during the N-DELETE operation of the Film Box, the association will be terminated.

2.3.1.2.8.2.1.3 Basic Grayscale and Color Image Box SOP Classes

The DICOM SERVER AE supports the following DIMSE Service Elements for Basic Grayscale and Color

Image Box SOP Classes :

- N-SET - Requests the Printer to set the image box attributes.

Only one association is opened at a time. The association is open until the response message is returned from the SCP or a network error occurs. No other configurable timers are available.

Following are the status codes that are more specifically processed when receiving messages from a

Print SCP equipment for the Basic Grayscale and Color Image Box SOP Classes N-SET:

63 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Table 2.3.1.2.8.2.1.3-1 Status Codes Received by DICOM SERVER AE for Basic Grayscale and Color Image Box

Service

Status

Status

Codes

SOP Classes N-SET

Further Meaning Application Behavior When receiving

Status Codes

Failure C603 Image size is larger than image box size

C605

C613

Insufficient memory in printer to store the image

Combined Print Image size is larger than the Image Box size

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Log printer memory full error.

Pauses print job and displays print failure message on the film composer UI.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

0119 Class-instance conflict

0210

0106

0212

Duplicate invocation

Invalid attribute value

Mistyped argument

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

0117 Invalid SOP instance

0121

0105

0118

0112

Missing attribute value

No such attributes

No such SOP Class

No such SOP Instance

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

64 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

0110 Processing failure

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

0213

0211

Warning B604

* *

Resource limitation

Unrecognized operation

Image size larger than image box size, the image has been demagnified.

Any other status code.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

B605 Requested Min Density or Max

Density outside of printer’s operating range. The printer will use its respective minimum or maximum density value instead.

B609 Image size is larger than the Image

Box size. The Image has been cropped to fit.

B60A Image size or Combined Print

Image size is larger than the Image

Box size. The Image or Combined

Print Image has been decimated to fit.

Success 0000 Image successfully stored in Image

Box

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Continues setting image box if more images are there, else sends N-Action request.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

If a Failure/Warning status is returned during the N-SET operation of the Image Box, the association will be terminated.

2.3.1.2.8.2.1.4 Printer SOP Class

The N-GET DIMSE service is supported for the Printer SOP Class. If an N-EVENT-REPORT DIMSE service is received when the association is active, the DICOM SERVER AE handles the relevant states, but the data received is ignored.

65 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Following are the status codes the Application may send back in the N-Event-Report response command to the Printer SOP Class SCP Equipment that sent the N-Event-Report request:

Table 2.3.1.2.8.2.1.4-1 Status Codes Returned by DICOM SERVER AE for Printer SOP Class N-Event-Report

Service

Status

Status

Code

Further Meaning Status Code Explanation Related Fields

Sent Back to the

SCU

Success 0000 N-EVENT-REPORT received successfully

None

The DICOM SERVER AE issues the N-GET service to retrieve the printer status. The statuses are processed as follows:

 If Printer status (2110,0010) is NORMAL, the film is printed.

 If Printer status (2110,0010) is FAILURE the print job is terminated. The Printer Status Info

(2110,0020) attribute is not processed.

 If Printer status (2110,0010) is WARNING one of two things will happen: o

If the Printer Status Info (2110,0020) is “RECEIVER FULL” or “SUPPLY EMPTY”, or “FILM

JAM” the status is displayed to the user and the print job is aborted. o

For all other Printer Status Info (2110,0020) values, the status is ignored and the print job continues.

Only one association is opened at a time. The association is open until the response message is returned from the SCP or a network error occurs. No other configurable timers are available.

For the attribute list requested by this product and for the behavior of this product on each returned

value, refer to Section 4.6 Printer SOP Class.

Following are the status codes that are more specifically processed when receiving messages from a

Print SCP equipment for the Printer SOP Class N-GET:

2.3.1.2.8.2.1.4-2 Status Codes Received by DICOM SERVER AE for Printer SOP Class N-GET

Service

Status

Status

Codes

Further Meaning

Failure 0107 Attribute list error

0119 Class-instance conflict

0210 Duplicate invocation

Application Behavior When receiving

Status Codes

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error

66 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

0117 Invalid SOP instance

0212 Mistyped argument

0118 No such SOP Class

0112 No such SOP Instance

0110 Processing failure

0213 Resource limitation

0211 Unrecognized operation

Success 0000 Success

* * Any other status code.

Refer Section.4 Print Management Implementation.

2.3.1.3 Association Acceptance Policy

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

Proceed with parsing dataset returned from N-Get response and perform appropriate operations. Refer section 4.2.5

Pauses print job and displays print failure message on the film composer UI. Error details along with the status code is logged.

The DICOM Server AE places limitations on who may connect to it.

If the Remote AE needs to “Push Images” or “Query/Retrieve Images”, to the local system then it has to be configured in the Local system to do the same.

67 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

When the DICOM Server AE accepts an association for image storage, it will receive any images transmitted on that association and store the images on disk.

It will also respond to queries from Remote AEs by sending matching entries. Any Remote AE can request and receive a list of images on the local database. The Remote AE must be configured in the local database manager’s list of Remote AE for it to be able to retrieve images from DICOM Server AE.

Any remote AE can open an association to the DICOM Server AE for the purpose of verification.

2.3.1.3.1 Real-World Activity: Receive Images

This AE is indefinitely listening for associations. No operator action is required to receive an image.

2.3.1.3.1.1 Associated Real-World Activity

The Real-World Activity associated with the Receive Images operation is the storage of the image on the disk drive of the GEHC CT.

2.3.1.3.1.2 Presentation Context Table

Table 2.3.1.3.1.2-1: Accepted Presentation Contexts for DICOM Server AE and Real-World Activity

Receive Images

Name

Verification

Abstract Syntax

UID

1.2.840.10008.1.1

MR Image Storage 1.2.840.10008.5.1.4.1.1.

PET Image Storage

RT Structure Set

Storage

2

4

481

.3

128

Presentation Context Table – Accepted

Transfer Syntax

CT Image Storage 1.2.840.10008.5.1.4.1.1.

1.2.840.10008.5.1.4.1.1.

1.2.840.10008.5.1.4.1.1.

Name List

Implicit VR Little

Endian

Implicit VR Little

Endian

Explicit VR Little

Endian

Explicit VR Big

Endian

Implicit VR Little

Endian

Explicit VR Little

Endian

Explicit VR Big

Endian

Implicit VR Little

Endian

Explicit VR Little

Endian

Explicit VR Big

Endian

Implicit VR Little

Endian

Explicit VR Little

Endian

Explicit VR Big

Endian

1.2.840.10008.1.2

1.2.840.10008.1.2.

1

1.2.840.10008.1.2.

2

1.2.840.10008.1.2

1.2.840.10008.1.2.

1

1.2.840.10008.1.2.

2

1.2.840.10008.1.2

1.2.840.10008.1.2.

1

1.2.840.10008.1.2.

2

1.2.840.10008.1.2

1.2.840.10008.1.2.

1

1.2.840.10008.1.2.

2

UID List

1.2.840.10008.1.2

Role

SCP

SCP

SCP

SCP

SCP

Extended

Negotiation

None

None

None

None

None

68 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

Secondary

Capture Image

Storage

1.2.840.10008.5.1.4.1.1.

7

Enhanced SR 1.2.840.10008.5.1.4.1.1.

88

.22

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

SCP None Implicit VR Little

Endian

Explicit VR Little

Endian

Explicit VR Big

Endian

Implicit VR Little

Endian

Explicit VR Little

Endian

Explicit VR Big

Endian

1.2.840.10008.1.21

.2.840.10008.1.2.1

1.2.840.10008.1.2.

2

1.2.840.10008.1.2

1.2.840.10008.1.2.

1

1.2.840.10008.1.2.

2

SCP None

Note: The SCP does not have a default acceptance policy if more than one acceptable transfer syntaxes are proposed by the SCU. It is the responsibility of the SCU to make a selection from more than one transfer syntaxes accepted. If multiple transfer syntaxes are proposed in the same presentation context, the SCP will accept the first one.

2.3.1.3.1.2.1

SOP Specific Conformance to C_STORE SCP

The DICOM Server AE conforms to the SOP’s of the Storage Service Class at level 1 (base).

Each C-STORE SCP operation supports an “Association Timer” and “Operation Inactivity Timer” with time out values of 30 seconds, 15 seconds respectively.

Association Timer – duration for SCP to respond to an association request.

Inactivity Timer – duration between two commands after the association.

Image Reception

If the DICOM Server AE returns one of the following status codes, then the C-STORE operation was unsuccessful and no image will be installed:

Service

Status

Failure

Status

Code

A700

Further Meaning Status Code Explanation

Refused: Out of resources Not enough disk space to store DICOM image.

Related

Fields Sent

Back to the

SCU

(0000,0902)

A710

A711

A900

Refused: Out of Resources Remote AE is not given permission to store on this AE

(0000,0902)

Refused: Out of resources Unable to connect to local database for storage.

(0000,0902)

Error: Data Set does not match SOP

Class

Storage of the DICOM object in localdb failed due to invalid data set.

(0000,0901)

(0000,0902)

69 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

C000

REVOLUTION ACT

S

DICOM C

ONFORMANCE

Error: Cannot understand Storage of the DICOM object in localdb failed.

S

TATEMENT

(0000,0901)

(0000,0902)

In the event of a successful C-STORE operation, the image has successfully been written to disk. The image will then be accessed in the same manner as any other image by the applications on the GEHC CT system.

Images may be deleted when instructed to do so by the user. Thus the duration of the storage of the image is determined by the users of the GEHC CT system.

Image Installation

If the image installation is unsuccessful, a message will appear in the Message Log informing the user of the failure and the image will be removed.

If the image installation process finds that an element is not encoded according to the DICOM standard, it will fail to install the image and the file will be removed.

Image Installation of non-GE Created MR or CT Images

Images received from non-GE products are installed as the appropriate image object. Their private data elements will be maintained if the negotiated transfer syntax is Explicit Little or

Big Endian. Also if any critical fields (mandatory) are missing, then the image will not be installed.

This AE is indefinitely listening for associations. No operator action is required to respond to a

verification message.

2.3.1.3.2 Real-World Activity: Verification Request from Remote AE

2.3.1.3.2.1

Associated Real-World Activity

This AE is indefinitely listening for associations. No operator action is required to respond to a verification message. The Real-World Activity associated with the verification request is to send a C-ECHO response message with a status of “success” to the requesting AE.

2.3.1.3.2.2 Presentation Context Table

Table 2.3.1.3.2.2-1: Acceptable Presentation Contexts for DICOM Server AE and

Real-World Activity Verification Request

Abstract Syntax

Presentation Context Table

Transfer Syntax

Name UID

Verification 1.2.840.10008.1.

1

Name List

Implicit VR Little

Endian

Explicit VR Little Endian

Explicit VR Big Endian

UID List

1.2.840.10008.1.2

1.2.840.10008.1.2.

1

1.2.840.10008.1.2.

2

Role

SCP

Extended

Negotiation

None

70 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

2.3.1.3.2.2.1 SOP Specific Conformance to C_ECHO SCP

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

The DICOM Server AE provides standard conformance to the DICOM Verification Service Class.

Each ECHO operation supports an “Association Timer” and “Operation Inactivity Timer” with time out values of 30 seconds and 15 seconds respectively.

2.3.1.3.3 Real-World Activity: Search Local Database (Query Request from Remote AE)

This AE is indefinitely listening for associations. No operator action is required to respond to a query request.

2.3.1.3.3.1 Associated Real-World Activity

The Real-World Activity associated with “Search Local Database” is to search the local database for entries that match the request and send a C-FIND response message with a status of “pending” for each matching entry and send a C-FIND response message with a status of "success" after the last

"pending" response.

If the C-FIND SCP receives a C-FIND-CANCEL request, it sends a C-FIND response message with a status of "cancel (FE00H)".

2.3.1.3.3.2 Presentation Context Table

Table 2.3.1.3.3.2-1: Acceptable Presentation Contexts for DICOM Server AE and Real-World Activity

Query Request

Presentation Context Table

Name

Study Root

Query/Retrieve

FIND

Abstract Syntax

UID

1.2.840.10008.5.1.4.1.2.2.1

Transfer Syntax

Name List

Implicit VR Little Endian

Explicit VR Little Endian

Explicit VR Big Endian

UID List

1.2.840.10008.1.2

1.2.840.10008.1.2.1

1.2.840.10008.1.2.2

Role

SCP

Extended

Negotiation

None

2.3.1.3.3.2.1 SOP Specific Conformance to C-FIND SCP

Each C-FIND SCP operation supports an “Association Timer” and “Operation Inactivity Timer” with time out values of 30 seconds and 15 seconds respectively.

All Required (R) and Unique (U) study, series, and image level keys for the Study-Root Query/Retrieve information model are supported. Some optional (O) keys are also supported as described in the following tables.

Table 2.3.1.3.3.2.1-1 Supported study level keys

Description

Study date

Study time

Accession number

Patient’s name

Type

R

R

R

R

Tag

0008,0020

0008,0030

0008,0050

0010,0010

Usage

Matched

Matched

Matched

Matched

71 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

Description

Patient id

Study id

Study Instance UID

Study description

Suite Id

Type

R

R

U

O

P

Table 2.3.1.3.3.2.1-2 Supported series level keys

Description

Modality

Series number

Series Instance UID

Series description

Manufacturer

Images in series

Type

R

R

U

O

O

O

Table 2.3.1.3.3.2.1-3 Supported image level keys

Description

Image number

Image Instance UID

Image type

Rows

Columns

Image position

Image orientation

Slice thickness

Slice spacing

Gantry tilt

Convolution kernel

Reconstruction diameter

Data collection diameter

Flip angle

Echo number

Echo time

Inversion time

Repetition time

Trigger time

Dfov Rect

Midscan Time

Azimuth

Number of Echo

Scout Anref

Type

R

O

O

O

O

O

O

O

O

O

P

P

P

P

P

O

O

O

O

O

U

O

O

O

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Tag

0010,0020

0020,0010

0020,000D

0008,1030

0009,0002

Usage

Matched

Matched

Matched

Returned

Returned

Tag

0008,0060

0020,0011

0020,000E

0008,103E

0008,0070

0020,1002

Tag

Usage

Matched

Matched

Matched

Returned

Returned

Returned

Usage

0020,0013 Matched

0008,0018 Matched

0008,0008 Returned

0028,0010 Returned

0028,0011 Returned

0020,0032 Returned

0020,0037 Returned

0018,0050 Returned

0018,0088 Returned

0018,1120 Returned

0018,1210 Returned

0018,1100 Returned

0018,0090 Returned

0018,1314 Returned

0018,0086 Returned

0018,0081 Returned

0018,0082 Returned

0018,0080 Returned

0018,1060 Returned

0019,001E Returned

0019,0024 Returned

0019,0026 Returned

0019,007E Returned

0021,004A Returned

72 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

Description

Location RAS

Location

Center R Coordinate

Center A Coordinate

Table Start Location

Table End Location

RAS Letter for Side of Image

RAS Letter for Anterior/Posterior

RAS Letter for Scout Start Location

RAS Letter for Scout End Location

Image Dimension X

Image Dimension Y

Type

P

P

P

P

P

P

P

P

P

P

P

P

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Tag Usage

0027,0040 Returned

0027,0041 Returned

0027,0042 Returned

0027,0043 Returned

0027,0050 Returned

0027,0051 Returned

0027,0052 Returned

0027,0053 Returned

0027,0054 Returned

0027,0055 Returned

0027,0060 Returned

0027,0061 Returned

Note: In the above tables the type field has the following meaning:

R = Required U = Unique O = Optional P = Private

Only keys with Usage type Matched will be matched against values in the database.

Values in keys of type Returned will be ignored and will be filled in with data from the database.

If an optional key is requested that does not appear in any of the tables above, that key will be ignored and no corresponding element will be returned.

If the database does not have a value corresponding to any requested optional key a zero-length element will be returned.

Sequence matching is not supported.

Range matching is supported for attributes of type date and time.

Only hierarchical query is supported.

Special character “?” can be used to match any single character and special character * can be used to match any character or set of characters for (0008, 0050) Accession Number, (0010, 0010)

Patient’s Name, (0010, 0020) Patient ID and (0020, 0010) Study ID.

Except sequence matching all other matching are supported. (i.e. wildcard (“*”, “?”) and range (“-“) matching is supported as defined in DICOM PS3.4 Section C.2 Query/Retrieve Information Model

Definition.)

Following are the status codes the Application may send back to the SCU Equipment while performing the requested Query:

73 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

Service

Status

Status

Code

Further Meaning

Failure A900 Error: Invalid dataset

C001 Error: Unable to process

Cancel FE00 Matching terminated due to cancel

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Status Code Explanation

This status code is sent for the following reasons

Related

Fields Sent

Back to the

SCU

(0000,0902)

1.When does not have required information, like query retrieve level

This status code is sent for the following reasons

(0000,0902)

1.When remote AE is not given permission to store on this AE.

2. Not able to connect to local database, because of max connection limit reached.

3. Database query fails.

This status code will be returned when

DICOM SERVER AE receives C-FIND-CANCEL from remote AE

None

Success 0000 Matching is complete - No final identifier is supplied

Matching complete None

Pending FF00 Matches are continuing -

Current Match is supplied and any Optional Keys were supported in the same manner as Required

Keys.

For pending messages Identifier

2.3.1.3.4 Real-World Activity: Move Images (Retrieve Request from Remote AE)

This AE is indefinitely listening for associations. No operator action is required to respond to a retrieve request.

2.3.1.3.4.1 Associated Real-World Activity

The Real-World Activity associated with “Move Images” is to send all images corresponding to the C-

MOVE request to the destination AE through a separate association.

If the C-MOVE SCP receives a C-MOVE-CANCEL request, it discontinues the move operation and closes the separate association.

74 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

2.3.1.3.4.2 Presentation Context Table

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Table 2.3.1.3.4.2-1: Acceptable Presentation Contexts for DICOM Server AE and

Real-World Activity “Move Images

Name

Study Root

Query/Retrieve

MOVE

Abstract Syntax

UID

1.2.840.10008.5.1.4.1.2.2.2

Presentation Context Table

Transfer Syntax

Name List

Implicit VR Little Endian

Explicit VR Little Endian

Explicit VR Big Endian

UID List

1.2.840.10008.1.2

1.2.840.10008.1.2.1

1.2.840.10008.1.2.2

Role

SCP

Extended

Negotiation

None

2.3.1.3.4.2.1 SOP Specific Conformance to C-MOVE SCP

The DICOM Server AE provides standard conformance to the baseline Study-root C-MOVE Service

Class SCP.

Each C-MOVE SCP operation supports an “Association Timer” and “Operation Inactivity Timer” with time out values of 30 seconds and 15 seconds and 60 minutes respectively. All images requested in a C-MOVE-RQ will be sent over a single association. A C-MOVE-RSP with a “pending” status will be returned to the requester every five images.

C-MOVE supports list of Studies (through list of Study UIDs), list of Series (through list of Series UIDs) and list of Images (through list of Instances UIDs).

To send C-MOVE to a 3rd party system is supported if 3rd party destination AE details are configured in the system.

The C-MOVE SCP will invoke C-STORE requests for the following SOP classes:

SOP Class Name

CT Image Information Storage

SOP Class UID

1.2.840.10008.5.1.4.1.1.2

MR Image Information Storage

PET Image Information Storage

RT Structure Set Storage

Secondary Capture image storage

1.2.840.10008.5.1.4.1.1.4

1.2.840.10008.5.1.4.1.1.128

1.2.840.10008.5.1.4.1.1.481.3

1.2.840.10008.5.1.4.1.1.7

Grayscale Softcopy Presentation State Storage 1.2.840.10008.5.1.4.1.1.11.1

Enhanced SR 1.2.840.10008.5.1.4.1.1.88.22

X-Ray Radiation Dose SR 1.2.840.10008.5.1.4.1.1.88.67

Following are the status codes the Application may send back to the SCU Equipment while performing the requested Retrieve:

75 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

Service

Status

Status

Code

Failure A701

A702

A801

Further Meaning

Refused: Out of resources -

Unable to calculate number of matches

Refused: Out of resources -

Unable to perform suboperations

Error: Move Destination

Unknown

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Status Code Explanation

This status code is sent when local database query fails

Related

Fields Sent

Back to the

SCU

(0000,0902)

This status code is sent to Retrieve SCU when the association is rejected by the C-STORE SCP

(0000,0902)

This status code is sent when the destination AE named in the C-MOVE request is unknown to

DICOM SERVER AE.

(0000,0902)

A900 Error: Invalid dataset

C001 Error: Unable to process

This status code is sent for the following reasons (0000,0902)

1.When the retrieve request does not contain required information, like query retrieve level

This status code is sent for the following reasons (0000,0902)

1.When Remote AE does not have permissions to use DICOM SERVER AEs retrieve service

2. Not able to connect to local database, because of max connection limit reached.

3. Database query fails.

Cancel

Success 0000

Pending

FE00

FF00

Sub-operations terminated due to a Cancel indication

Sub-operations Complete -

No Failure.

Sub-operations are continuing -

This status code is sent when DICOM SERVER AE receives C-MOVE-CANCEL request from the

Remote AE.

(0000,1021)

(0000,1022)

(0000,1023)

(0000,0902))

All the composite SOP instances have been successfully sent to the C-MOVE destination

(0000,1021)

(0000,1022)

(0000,1023)

For pending sub operation to remote AE (0000,1020)

(0000,1021)

(0000,1022)

(0000,1023)

2.3.1.3.4.3 Presentation Context Acceptance Criteria

No criterion.

76 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

2.3.1.3.4.4 Transfer Syntax Selection Policy

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

All protocols detailed above in the tables are supported; no priority selection is given.

2.3.1.3.5 Real-World Activity: Listen to remote Storage Commitment SCP

The DICOM SERVER AE is indefinitely listening for associations. No operator action is required to receive a Storage Commitment notification (N-EVENT-REPORT).

2.3.1.3.5.1 Associated Real-World Activity

The Real-World Activity associated consists into:

• Flag the exams/series that have been committed (transfer of ownership) in the database

• Display the error when some images of a patient existing in the database have not been committed

2.3.1.3.5.2 Presentation Context Table

Presentation Context Table – Accepted by DICOM SERVER AE for Activity Storage commit notification

from Remote AE

Abstract Syntax Transfer Syntax

Storage

Commitment Push

Model

Name UID Name List UID List

Role Extended

Negotiatio n

SCU None

Explicit VR Little

Endian

Explicit VR Big Endian

1.2.840.10008.1.2

1.2.840.10008.1.2.

1

1.2.840.10008.1.2.

2

2.3.1.3.5.2.1 SOP Specific DICOM Conformance Statement for the Storage Commitment Push Model SOP Class

(N-EVENT-REPORT)

The DICOM SERVER AE monitors an “Operation Inactivity” timer. The connection with the SCU will be terminated if it is inactive for the configured time-out interval. Default time-out is 15 seconds and is configurable.

Once the N_EVENT_REPORT response is received from the SCP, the following actions will be taken depending on the status of response.

2.3.1.3.5.2.1.1 Commitment response with SUCCESS status

The “Archived” flag information in the browser for all the successfully archived exam/series will be updated. The archive status column in the browser will be changed to display “Archived” icon to indicate that the exam/series has been archived successfully. The job queue entry will be removed.

N_EVENT_REPORT response will be sent on the same association as N_EVENT_REPORT request. No data set will be sent along with the response.

Following are the status codes the Application may send back to the SCP Equipment after receiving the N-EVENT-REPORT:

77 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

Servic e

Status

Error

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Status

Codes

Further Meaning Status Code sending explanation Related Fields sent back to the SCU

0110 Processing

Failure

Indicates that an internal error occured while processing.

None

Succe ss

0000 Success The storage commitment result received successfully.

None

2.3.1.3.5.2.1.2 Commitment response with FAILURE status

In the event of complete/partial failure the user will be notified about the status and the job entry will be paused. There is no attempt made to automatically retry the failed SOP instances. However the user can manually retry the failed jobs. Such requests will be treated as new requests. This will go through the complete sequence of operations once again.

The failure reason is ignored. Failed SOP instances will have their “Archived” flag information unaltered. Failed SOP instances are logged.

Note: The archived status flag in the browser is a shared flag with local archive. When the status is

“Archived”, it means that the images are archived but doesn’t specify whether on local archive device or remote archive device. It is left to the user’s discretion whether the local SOP instances are to be deleted.

N_EVENT_REPORT response will be sent on the same association as N_EVENT_REPORT request. No data set will be sent along with the response.

Please see section 5.1.2.1

for the complete list of N_EVENT_REPORT failure statuses processed by the

system.

Following are the status codes the Application may send back to the SCP Equipment after receiving the N-EVENT-REPORT:

Service Status Further Meaning Status Code sending explanation Related

Status Codes Fields sent back to the

SCU

Error 0110 Processing Failure Indicates that an internal error occurred while processing.

None

Success 0000 Success The storage commitment result received successfully.

None

2.3.1.3.5.3 Presentation Context Acceptance Criterion

The DICOM SERVER AE evaluates each Presentation Context independently, and accepts any

Presentation Context that matches an Abstract Syntax for any Real-World Activity.

78 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

2.3.1.3.5.4 Transfer Syntax Selection Policies

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Within each Presentation Context, the DICOM SERVER AE will select Transfer Syntaxes according to the following priority (highest priority first):

1. Explicit VR Little Endian

2. Implicit VR Little Endian

3. Explicit VR Big Endian

2.4 Communication Profiles

2.4.1 Supported Communication Stacks (parts 8)

DICOM Upper Layer (Part 8) is supported using TCP/IP.

2.4.2 TCP/IP Stack

The TCP/IP stack is inherited from a UNIX Operating System.

2.4.2.1 Physical Media Support

Ethernet v2.0, IEEE 802.3

2.4.3 Others

DHCP is not supported.

IPv6 is not supported.

2.5 Extensions / Specializations / Privatizations

2.5.1 Standard Extended Elements

Refer to Appendix A for supported Standard Extended Elements.

2.5.2 Private Data Elements

Refer to Appendix B for a complete listing of private data elements used with this implementation.

2.6 Configuration

2.6.1 AE Title/Presentation Address Mapping

The GEHC CT system allows the user to “add”, “Remove”, or “Update the mapping of remote AE Titles to IP Addresses and Ports. These options can be selected from “Network Configuration” dialog box displayed by choosing “Network Configuration” from the “Tools” pull-down menu from the local database browser.

79 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

2.6.2 Configurable Parameters

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

The following fields are configurable for this AE (local):

Local AE Title

Local IP Address

Local Listening Port Number

Local IP Netmask

Note: The local port on which the GEHC CT system receives DICOM incoming TCP connections is configured by default to port 4006.

The following fields are configurable for every remote DICOM AE:

Remote AE Title

Remote IP Address

Listening TCP/IP Port Number

The following fields are configurable:

Association Establishment Timer time-out

Store, Find, Move Timers time-out

Inactivity Timers time-out

Maximum PDU length

Note: Timeouts are set in network-cfg.xml and mergecom.pro files

Note: All configurations should be performed by a GE Field Service Engineer.

2.6.2.1 Print Configuration

The following fields are configurable for DICOM Print. The valid ranges are as described in Section 6.2

Attribute Name DICOM Tag

Medium Type

(2000, 0030)

Film Destination (2000, 0040)

Magnification Type

Min Density

Max Density

Empty Image Density

Border Density

(2010, 0060)

(2010, 0120)

(2010, 0130)

(2010, 0110)

(2010, 0100)

Configuration Information (2010, 0150)

Smoothing Type (2010, 0080)

The following network timers are supported and are configurable. When these timers expire the association is terminated. The default values are defined in dprint.cfg as below

80 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

Timeout Description

Association Time out

Session Time out

Echo Time out

N-SET Time out

N-ACTION Time out

N-CREATE Time out

N-DELETE Time out

N-GET Time out

2.6.2.2 PPS Configuration

Default Value

120 seconds

1200 seconds

200 seconds

300 seconds

300 seconds

300 seconds

300 seconds

100 seconds

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Configuration Parameter Name

bi_assoc_tio bi_session_tio bi_echo_tio bi_nset_tio bi_naction_tio bi_ncreate_tio bi_ndelete_tio bi_nget_tio

The parameters Remote MPPS AE IP Address, Remote AE(HIS/RIS) IP Port and Remote MPPS AE Title define where the MPPS requests will be directed.

The default value of Local AE Title is host name of the scanner appended with string “_PPS”. If the length exceeds 16 characters the AE Title is truncated to a length of 16.)

Configurable DICOM Data:

The following DICOM sequences are currently defaulted to be automatically sent in the PPS N-

CREATE and PPS N-SET however PPS can be configured to not send them:

(0008, 1032) Procedure Code Sequence – sent in the PPS N-CREATE. If not sent, the PPS N-

CREATE will contain an empty Procedure Code Sequence.

(0040, 0260) Performed Protocol Code Sequence – sent in the PPS N_SET. If not sent, the PPS N-

SET will not contain a Performed Protocol Code Sequence.

The configuration is performed by GEHC Field Service Engineers changing parameters in configuration file Dicom.cfg

2.6.2.3 Modality Worklist Query Configuration

HIS/RIS is configured as a remote AE. These parameters define where worklist queries will be directed. Configuration of these parameters is performed by GEHC Field Service engineers using the

ModalityWorklist installation facilities.

The following parameters are configurable for the DICOM Worklist Server AE:

Local (Worklist Server) AE Title (automatically set to host name of scanner)

Local IP address

Local IP netmask

Local IP gateway

The following parameters are configurable by changing their values in the configuration file

/usr/g/config/WLdcm.cfg. Note that these parameters typically need not be changed. Furthermore,

81 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

no support is provided for retaining changed settings: the values will require changing again after a system software upgrade.

Implementation UID

PDU size

Association time-out period

Session time-out period

C-FIND time-out period

2.6.2.4 GSPS Configuration

Exam Split allows for the configuration of the following parameters that pertain to the remote AE.

Remote GSPS AE IP Address IP Address used to contact the remote GSPS AE

Remote GSPS IP Port:

Remote GSPS AE Title

IP Port used to contact the remote GSPS AE

Application Entity Title used to contact the remote

GSPS AE

Remote HES AE IP Address

Remote HES IP Port:

IP Address used to contact the remote HES AE

IP Port used to contact the remote HES AE

Remote HES AE Title

The following are configurable for the GSPS.

Remote Host IP address

Remote Host port

Remote AE Title

Changesplit_mode <HES/VES>

Application Entity Title used to contact the remote

HES AE

2.7 Support of Extended Character Sets

In addition to the DICOM default character set, GEHC CT supports the ISO IR 100 Latin alphabet #1 supplementary set for the purpose of interchange .

As a Storage SCP, the product will accept SOP Instances with any value of Specific Character Set

(0008,0005). As a Query SCU, it will similarly accept response items with any value of Specific

Character Set. However, it will display in the user interface only characters specified as within ISO_IR

6 (ASCII) or the configured extended character set.

The product user interface will allow the user to enter characters from the console keyboard that is within ASCII or the configured extended character set. If any such extended characters are included

82 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

in SOP Instances or in query identifier matching fields, the product will appropriately specify the extended character set in Specific Character Set (0008,0005).

The generated MPPS messages will also specify the character set in (0008,0005) if the referenced images contain extended characters.

As a Modality Worklist SCU, it only uses ISO_IR 100 Specific Character Set in the worklist query requests.

2.8 Codes and Controlled Terminology

The product uses coded terminology as defined below .

2.8.1 Mapped Coded Terminology

The product maps, without change, coded terminology values supplied in Modality Worklist

Scheduled Procedure Steps into Image SOP Instance and Modality Performed Procedure Step attributes, as described in Sections 6 and 7.

2.9 Security Profiles

The product does not conform to any defined DICOM Security Profiles.

It is assumed that the product is used within a secured environment. It is assumed that a secured environment includes at a minimum:

1. Firewall or router protections to ensure that only approved external hosts have network access to the product.

2. Firewall or router protections to ensure that the product only has network access to approved external hosts and services.

3. Any communications with external hosts and services outside the locally secured environment use appropriate secure network channels (such as a Virtual Private

Network (VPN))

83 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

3 MEDIA STORAGE CONFORMANCE STATEMENT

3.1 Introduction

This conformance statement specifies the GEHC CT conformance to the DICOM Media Interchange. It details the DICOM Media Storage Application Profiles and roles that are supported by this product.

GEHC CT system provides capabilities to DICOM interchange on CD-R (Compact Disc - Recordable),

CD-RW (Compact Disc - ReWritable), DVD-R (Digital Video Disc - Recordable), DVD-RW (Digital Video

Disc - ReWritable) and USB (Universal Serial Bus) memory.

GEHC CT system works with most of the IOD’s like Computed Tomography (CT), Magnetic Resonance

(MR), Computed Radiography (CR), Positron Emission Tomography (PET).

3.2 Implementation Model

The DICOM Interchange Archive Server Application Entity (AE) handles all DICOM media storage functionality on the GEHC CT system.

The DICOM Interchange Archive Server AE is commanded to perform DICOM services through the buttons and menu selections on the user interface. It also uses the appropriate recorders to provide the service Interchange Media profiles.

3.2.1 Application Data Flow Diagram

The Basic and Specific Application models for the CD-R/CD-RW/DVD-R/DVD-RW/USB devices are shown in the following Illustrations.

The DICOM Interchange Archive Server Application Entity (AE) handles the DICOM CREATE CD

(/DVD/USB) / RESTORE CD (USB) / QUERY CD (USB) functionality for the CD/DVD/USB media. The

DICOM Interchange Archive Server Application Entity (AE) is commanded by the user to perform

DICOM services operating on the DICOM media through the use of buttons and menu selections on the graphical user interface of the platform.

The user requests the creation of a DICOM File Set and the writing of this DICOM File Set on blank

Interchange Media by selecting images in the local Browser and selecting the Interchange Media as being the selected device. Then, the iso9660 image of the CD/DVD/USB to write will be generated.

Once the generation has been done, it writes the complete set of data on the selected Interchange

Media.

The Media Interchange Application Model for the GEHC CT system is shown in Illustration 3-1:

84 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

Illustration 3-1: Specific AE Application Model

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Display

Directory

QUERY

Local

Storage

RESTORE

DICOM

AE

Archive Server

FSC

Media

Storage

FSR

Record Media

CREATE

The DICOM Interchange Archive Server AE has a local storage that may contain various SOP instances. These may have been obtained by original creation, network (DICOM or proprietary) or by removable media using other application entities. These instances are external to this conformance claim and the origin of SOP instances is outside the scope of this claim.

The DICOM Interchange Archive Server AE can initialize Media by acting as an FSC to create a new

DICOM File-set on a blank interchange media. It initializes the DICOM File-set and writes the specified

SOP instances onto the Interchange Media at once. The SOP instances written will be limited to instances that match the criteria of one of the Application Profiles that is supported.

Updating the media is not supported.

3.2.2 Functional Definitions of AE's

The CD-R/ CD-RW/DVD-R/DVD-RW/USB DICOM Interchange Archive Server Application Entity support the following functions:

· Generate and write a DICOM File Set (FSC) in a one shot activity. (CREATE).

· Read a DICOM File Set (FSR) on an Interchange Media (QUERY).

· It can copy SOP instances from the media onto local storage. (RESTORE).

3.2.3 Sequencing of Real World Activities

The display function (QUERY) can only be performed on a piece of media that already had a DICOM

File-set created. With no SOP instances having been added, the directory will be displayed empty.

85 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

The write function (CREATE) can only be performed on a blank (unused) Interchange Media. Updates to an already recorded Interchange Medium are not supported.

There are no other sequencing requirements.

3.2.4 File Meta Information for Implementation Class and Version

The File Meta-Information for this implementation is:

File Meta-Information Version .................................1

Implementation Version Name ...............................Software Revision (See Appendix D)

The table in Appendix D identifies the Implementation UID for this product version.

3.3 AE Specifications

3.3.1 DICOM CD-R/CD-RW/DVD-R/DVD-RW/USB Media Interchange AE Specification

The DICOM CD-R/CD-RW/DVD-R/DVD-RW/USB Media Interchange AE provides standard conformance to DICOM Media Storage Service Class. The Application Profiles and roles are listed in

Table 3.3.1-1.

Table 3.3.1-1: DICOM CD-R/CD-RW/DVD-R/DVD-RW/USB Supported Application Profiles

Application Profiles

Supported

STD-GEN-CD

STD-GEN-CD

STD-GEN-CD

STD-GEN-DVD-JPEG

STD-GEN-USB-JPEG

STD-GEN-USB-JPEG

STD-GEN-USB-JPEG

Real World Activity

CREATE CD

QUERY CD

RESTORE CD

CREATE DVD

CREATE USB

QUERY USB

RESTORE USB

Role

FSC

FSR

FSR

FSC

FSC

FSR

FSR

SC Option

Interchange

Interchange

Interchange

Interchange

Interchange

Interchange

Interchange

Note: JPEG compressed images are supported for USB restore only.

.

86 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

3.3.1.1 File Meta Information for the CD-R/CD-RW/DVD-R/DVD-RW/USB DICOM Media Interchange

Application Entity

Following are the values set in the File Meta Information for this AE. Please refer to Appendix D for details on specific GEHC CT system information.

Implementation UID : See Appendix D

Implementation version name : See Appendix D

3.3.1.2 Real World Activities for the CD-R/CD-RW/DVD-R/DVD-RW/USB DICOM Media Interchange

Application Entity

3.3.1.2.1 Real World Activity: Record Media - CREATE (CD/DVD/USB)

The DICOM Media Interchange Application acts as an FSC using the interchange option when requested to copy SOP Instances from the local database to a CD-R / CD-RW / DVD-R / DVD-RW /

USB.

The user selects the entries in the local database that he/she wants the DICOM Media Interchange

Application to copy onto Interchange Media.

The graphic interface allows the user to select the entries (studies, series or images) in the local database to be copied onto to the selected Interchange Media.

The DICOM Media Interchange Application creates one File Set per generated Interchange Media.

The graphic interface allows the user to select the entries (studies, series or images) in the local database to be copied onto to the selected Interchange Media.

A user cannot create CD/DVD/USB while restore CD/USB is in process.

A DICOM Media Viewer is provided along with the selected object instances on the interchange media. This viewer can be loaded on a standard PC running Windows XP, Windows Vista or

Windows 7.

Before writing on the Interchange Media, the DICOM Media Interchange Application checks for the following condition:

Note:

The inserted media is blank and write-able. If the condition is not met, an error is displayed and the user needs to replace it with a blank media.

Since JPEG compression is not enabled, the transfer syntax for SOP Instances will always be Explicit

VR Little Endian (ELE) Uncompressed Transfer Syntax, UID 1.2.840.10008.1.2.1. For any other IODs with Transfer syntax other than ELE, the corresponding SOP instance data are converted to ELE and then copied to CD/DVD/USB. During the conversion to ELE, the private tags are stored with VR as UN if the syntax is implicit to facilitate fidelity.

87 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

The Object SOP instances have to be encoded with the ISO_IR 100 Specific Character set or DICOM

Default Character Set.

3.3.1.2.1.1

Application Profiles for the RWA:

R ecord Media - CREATE (CD/DVD/USB)

Refer to Table 3.3.1-1 for the list of Application Profiles that invoke this AE for the Create CD/DVD/USB

RWA.

3.3.1.2.1.1.1 Options for STD-GEN-CD, STD-GEN-DVD-JPEG and STD-GEN-USB-JPEG Application Profile:

Following are the SOP Classes supported by the RWA “Create CD/DVD/USB”. All SOP Instances use the Explicit VR Little Endian Uncompressed Transfer Syntax, UID 1.2.840.10008.1.2.1 for creating

CD/DVD/USB.

TABLE 3.3.1.2.1.1.1-1

SOP Class SOP Class UID

CT Image Storage

MR Image Storage

1.2.840.10008.5.1.4.1.1.2

1.2.840.10008.5.1.4.1.1.4

PET Image Information Storage

RT Structure Set Storage

Secondary Capture Image Storage

Grayscale Softcopy Presentation State Storage

X-Ray Radiation Dose SR Image Storage

Enhanced SR Image Storage

1.2.840.10008.5.1.4.1.1.128

1.2.840.10008.5.1.4.1.1.481.3

1.2.840.10008.5.1.4.1.1.7

1.2.840.10008.5.1.4.1.1.11.1

1.2.840.10008.5.1.4.1.1.88.67

1.2.840.10008.5.1.4.1.1.88.22

3.3.1.2.1.1.2 FSC Directory Options

Refer to Appendix C for a complete listing of all modules and attributes used in the DICOMDIR definition.

NOTE:

Modality attribute (0008,0060) in the DICOMDIR Series records should be “CT” for Secondary

Capture SOP Class Images. Otherwise the GEHC CT system’s DICOM media display browser will not list-up the series for contents rendering the retrieval to local storage impossible.

For USB, only one DICOM file set shall be stored in the first partition of a partitioned device. If the device is not partitioned, only one DICOM file set shall be stored on the device.

88 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

3.3.1.2.2 Real World Activity: Display Directory – QUERY (CD/USB)

The DICOM Media Interchange AE acts as an FSR using the interchange option when requested to browse the Interchange Media.

When the DICOM Media Interchange AE is requested to provide a directory listing, it reads the Fileset and displays the DICOMDIR directory entries, according to the STUDY Root paradigm.

If the media is not blank, then Interchange Media gets mounted. To remove the media, the user has to select the eject button on user interface for the appropriate drive.

A user cannot display the directory when create CD/DVD/USB or restore CD/USB are in progress.

3.3.1.2.2.1 Application Profiles for the RWA: Display Directory

For the list of Application Profiles that invoke this AE for the Display Directory CD/USB RWA,, see Table

3.3.1-1.

There are no extensions or specializations.

3.3.1.2.2.2 Media Storage Application Profile for the RWA: Display Directory:

Following are the SOP Classes supported by the RWA “Display Directory of CD/USB”:

Information

Object

Definition

Media

Storage

Directory

Storage

SOP Class UID

1.2.840.10008.1.3.10

Transfer Syntax

Explicit VR Little Endian

Transfer Syntax UID

1.2.840.10008.1.2.1

3.3.1.2.3 Real World Activity (RWA): Local Storage – RESTORE (CD/USB)

The CD/USB DICOM Media Interchange AE acts as an FSR using the interchange option when requested to copy SOP instances from the CD/USB to the local database.

The user selects the SOP instances that he wants the DICOM Media Interchange AE to copy on the local database by selecting appropriate Study/Series/Image instances and clicking on the suitable restore buttons. Once selected, the SOP instances are copied from the media to the local database.

If the media is not blank, then the Interchange Media gets mounted. To remove the media, the user has to select the eject button on the User Interface for the appropriate drive.

A user cannot restore CD/USB while create CD/DVD/USB is in process.

A user can only restore selected composite objects at a time from a media; any other attempt of selections to restore CD/USB on media in same drive will wait until the first one is completed.

89 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

3.3.1.2.3.1 Application Profiles for the RWA: Local Storage – RESTORE (CD/USB)

For the list of Application Profiles that invoke this AE for the Restore RWA, see Table 3.3.1-1. For extensions and specialization’s see section 3.5.

3.3.1.2.3.1.1 Options for STD-GEN-CD and STD-GEN-USB-JPEG Application Profile:

Following are the SOP Classes supported by the RWA “

Local Storage – RESTORE (CD/USB)

”.

Information

Object

SOP Class UID Transfer Syntax

Transfer Syntax UID

Definition

Media

Storage

Directory

Storage

See Table

3.3.1.2.1.1.1-1

See Table

3.3.1.2.1.1.1-1

1.2.840.10008.1.3.10

See Table

3.3.1.2.1.1.1-1

See Table

3.3.1.2.1.1.1-1

Explicit VR Little Endian

Explicit VR Little Endian

1.2.840.10008.1.2.1

1.2.840.10008.1.2.1

1.2.840.10008.1.2.4.70

See Table

3.3.1.2.1.1.1-1

See Table

3.3.1.2.1.1.1-1

See Table

3.3.1.2.1.1.1-1

See Table

3.3.1.2.1.1.1-1

JPEG Lossless Process 14

(selection value 1)

(STD-GEN-USB-JPEG only)

JPEG Lossy, Baseline

Sequential with Huffman

Coding

(Process 1)

(STD-GEN-USB-JPEG only)

JPEG Extended (Process 2 & 4):

Default Transfer Syntax for

Lossy JPEG 12 Bit Image

Compression (Process 4 only)

(STD-GEN-USB-JPEG only)

1.2.840.10008.1.2.4.50

1.2.840.10008.1.2.4.51

3.4 Augmented and Private Application Profiles

3.4.1 Augmented Application Profiles

The CD/DVD/USB Media Archive Interchange AE does not support any augmented Application

Profiles.

3.4.2 Private Application Profiles

The CD/DVD/USB Media Interchange AE does not support any private Application Profiles.

90 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

3.5 Extensions, Specializations and Privatizations of SOP Classes and Transfer Syntax

3.5.1 Extensions, Specialization’s and Privatizations of SOP Classes

3.5.1.1 SOP Specific Conformance Statement for Basic Directory SOP Class

Following attributes are Standard extensions for each Directory Record type.

The CT/MR SOP Class Images have definitions extended for Defined Terms and include GE specific

Private Data elements. The following sections describe the details for these SOP classes.

Image Type

Key Attribute

Manufacturer’s ID

Series Description

Manufacturer’s Model Name

SOP Instance UID

Acquisition Time

Slice Thickness

Spacing Between Slices

Data Collection Diameter

Reconstruction Diameter

Gantry/Detector Tilt

Convolution Kernel

Slice Location

Tag

(0008,0008)

(0008,0070)

(0008,103E)

(0008,1090)

(0008,0018)

(0008,0032)

(0018,0050)

(0018,0088)

(0018,0090)

(0018,1100)

(0018,1120)

(0018,1210)

(0020,1041)

Directory Record Type

SERIES

SERIES

SERIES

SERIES

IMAGE

PRESENTATION

SR DOCUMENT

IMAGE

IMAGE

IMAGE

IMAGE

IMAGE

IMAGE

IMAGE

IMAGE

Refer to Appendix B for private data elements.

3.5.2 Private Transfer Syntax Specification

None specified.

3.6 Configuration

For the CD/DVD/USB Interchange Media Application, the source AE Title encoded in the File Meta-

Information cannot be modified.

3.7 Support of Extended Character Sets

The CD/DVD/USB Interchange Media Application will support copy of SOP instances containing the

ISO IR 100 (Latin alphabet No. 1, supplementary set) and DICOM default character sets as defined in

PS3.5.

91 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Any incoming SOP instances encoded using another extended character set will not be installed in the GEHC CT system database.

92 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

4 Print Management Implementation

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

4.1 Introduction

This section of conformance Statement specifies the implementation for the specific SOP Classes supported in the Basic Grayscale and Color Print Management Meta SOP Classes, the attributes supported for both IODs and services, and the valid range of values for mandatory and optional attributes.

4.2 Basic Film Session SOP Class

The Print SCU supports the following DIMSE Service Elements for the Basic Film Session SOP Class.

N-CREATE – Requests the Print SCP to create an instance of Basic Film Session.

4.2.1 Basic Film Session N-Create Attributes

The following Attribute values are supported by the N-CREATE:

Attribute name

Specific Character Set

*Number of Copies

*Print Priority

*Medium Type

*Film Destination

Film Session Label

Memory Allocation

Owner ID

Tag

(0008,0005)

(2000,0010)

(2000,0020)

(2000,0030)

(2000,0040)

(2000,0050)

(2000,0060)

(2100,0160)

Use

Not used

Set by user.

Valid range is : 1-99

Set when a printer device is added using the

Film Composer Tool. Default value is HIGH. Set in Configuration file

Valid Range is :

HIGH or MED or LOW

Set when a printer device is added using the

Film Composer Tool. Set in Configuration File

Valid values are :

PAPER or CLEAR FILM or BLUE FILM

Set when a printer device is added using the

Film Composer Tool. Set in Configuration File

Valid values are :

MAGAZINE or PROCESSOR

Not used.

Not used.

Not used.

Note: * denotes that the attribute is optional for the SCU. However, we do provide values for all of these optional attributes and if the SCP does not support the requested value it may choose to either return a failure status or ignore the value provided and use its default value.

93 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

4.3 Basic Film Box SOP Class

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

The Print SCU supports the following DIMSE Service Elements for the Basic Film Box SOP Class.

N-CREATE – Requests the Print SCP to create an instance of Film Box.

N-ACTION – Requests the Print SCP to print the Film Box onto Printer.

N-DELETE – Requests the Print SCP to delete the Film Box Instance.

4.3.1 Basic Film Box N-Create Attributes

This table lists the attributes that are sent to the SCP in the Basic Film Box N-Create Request, and that are received in the Basic Film Box N-Create Response from the SCP

Attribute Name Tag Use

Image Display Format (2010,0010) Set in User Interface.

Valid Range: STANDARD\C,R

(2010,0500) Sent in the request Referenced Film Session

Sequence

>Referenced SOP Class

UID

>Referenced SOP

Instance UID

Referenced Image Box

Sequence

>Referenced SOP Class

UID

>Referenced SOP

Instance UID

*Film Orientation

(0008,1150) Film session SOP Class UID (1.2.840.10008.5.1.1.1)

(0008,1155)

(2010,0510)

(0008,1150)

(0008,1155)

Film session SOP instance UID

Used in Image Box N-SET request

Film Size ID

(2010,0040) Set when a printer device is added using the Film

Composer Tool. Set in Configuration File.

Valid value is : PORTRAIT.

(2010,0050) Set when a printer device is added using the Film

Composer Tool.

Valid values are :

14INX17IN,

A3,

A4,

8INX10IN,

8.5INX11IN,

10INX12IN,

10INX14IN,

11INX14IN,

11INX17IN,

14INX14IN,

24CMX24CM,

94 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

*Magnification Type

*Max Density

Illumination

Reflected Ambient Light

Requested Resolution ID

ICC Profile

(2010,015E)

(2010,0160)

(2020,0050)

(0028,2000)

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

24CMX30CM.

(2010,0060) Set when a printer device is added using the Film

Composer Tool. Set in Configuration File.

Valid values are:

REPLICATE

BILINEAR

CUBIC

NONE

(2010,0130) Set when a printer device is added using the Film

Composer Tool. Set in Configuration File.

Valid range is : 0-4095.

Configuration Information (2010,0150) Set when a printer device is added using the Film

Composer Tool

Annotation Display Format

ID

(2010,0030) Not Supported.

*Smoothing Type (2010,0080) Set when a printer device is added using the Film

Composer Tool. Set in Configuration File.

*Border Density

*Empty Image Density

*Min Density

(2010,0100) Set when a printer device is added using the Film

Composer Tool. Set in Configuration File.

Valid values are:

BLACK or WHITE or Not sent

(2010,0110) Set when a printer device is added using the Film

Composer Tool. Set in Configuration File.

Valid values are:

BLACK or WHITE or Not sent

(2010,0120) Set when a printer device is added using the Film

Composer Tool. Set in Configuration File.

Valid values are:

0-4095 or Not sent

*Trim (2010,0140) Valid values are:

YES or NO or Not Sent

Default value : NO

Not Supported.

Not Supported.

Not Supported.

Not Supported.

4.3.2 Basic Film Box N-Action Attributes

Following are the Action reply arguments that are supported if present in the N-Action response of the

Basic Film Box SOP Class

95 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

Action Type Name

Print

Action Type ID

1

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Attribute

Referenced Print

Job Sequence

>Referenced SOP

Class UID

>Referenced SOP

Instance UID

>Print Job ID

Tag

(2100,0500)

(0008,1150)

(0008,1155)

(2100,0010)

4.3.3 Basic Film Box N-Delete

The DICOM SERVER AE requests the Print SCP to delete the Film Box Instance.

Usage SCU

Ignored

Ignored

Ignored

Not supported

Note: * denotes that the attribute is optional for the SCU. However, we do provide values for all of these optional attributes and if the SCP does not support the requested value it may choose to either return a failure status or ignore the value provided and use its default value.

4.4 Basic Grayscale Image Box SOP Class

The Print SCU supports the following DIMSE Service Elements for Grayscale Image Box SOP Class.

N-SET – Requests the Printer to set the image box attributes.

This table lists the attributes that are sent in the Basic Grayscale Image Box N-Set Request:

Attribute Name

Image Position

Basic Grayscale Image

Sequence

>Samples Per Pixel

>Photometric

Interpretation

>Rows

>Columns

>Pixel Aspect Ratio

>Bits Allocated

>Bits Stored

>High Bit

>Pixel Representation

>Pixel Data

*Polarity

Tag

(2020,0010)

(2020,0110)

(0028,0002)

(0028,0004)

(0028,0010)

(0028,0011)

(0028,0034)

(0028,0100)

(0028,0101)

(0028,0102)

(0028,0103)

(7FE0,0010)

(2020,0020)

Use

Based on Image Display Format

Sent in the request

1

MONOCHROME2

Image Dependent

Image Dependent

1\1

8 (if Bits Stored=8) or

16 (if Bits Stored=12)

8 or 12. Default value: 8

7 (if Bits Stored=8) or

11 (if Bits Stored=12)

0 (unsigned integer)

Pixel data

NORMAL

96 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

Magnification Type (2010,0060)

Smoothing Type

Min Density

(2010,0080)

(2010,0120)

Max Density (2010,0130)

Configuration Information (2010,0150)

Requested Image Size (2020,0030)

Requested Decimate/Crop

Behavior

(2020,0040)

Referenced Presentation

LUT Sequence

> Referenced SOP Class

UID

(2050,0500)

(0008,1150)

> Referenced SOP Instance

UID

(0008,1155)

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

BILINEAR

CUBIC

REPLICATE

NONE

Printer Dependent

0-4095, Not Sent

0-4095

Printer Dependent

Not used

Not used

Not supported.

Not supported.

Not supported.

Note:

* denotes that the attribute is optional for the SCU. However, we do provide values for all of these optional attributes and if the SCP does not support the requested value it may choose to either return a failure status or ignore the value provided and use its default value.

4.5 Basic Color Image Box SOP Class

The Print SCU supports the following DIMSE Service Elements for Basic Color Image Box SOP Class.

N-SET - Requests the Printer to set the image box attributes.

This table lists the attributes that are sent in the Basic Color Image Box N-Set Request:

Attribute Name

Image Position

Basic Color Image

Sequence

>Samples Per Pixel

>Photometric

Interpretation

>Planar Configuration

>Rows

>Columns

>Pixel Aspect Ratio

>Bits Allocated

>Bits Stored

Tag

(2020,0010)

(2020,0111)

(0028,0002)

(0028,0004)

(0028,0006)

(0028,0010)

(0028,0011)

(0028,0034)

(0028,0100)

(0028,0101)

Use

Based on Image Display Format

3

RGB

8

8

1

Image Dependent

Image Dependent

1/1

97 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

>High Bit

>Pixel Representation

>Pixel Data

*Polarity

Magnification Type

Smoothing Type

Requested Image Size

Requested Decimate/Crop

Behavior

(0028,0102)

(0028,0103)

(7FE0,0010)

(2020,0020)

(2010,0060)

(2010,0080)

(2020,0030)

(2020,0040)

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

7

0 (unsigned integer)

Pixel data

NORMAL

BILINEAR

CUBIC

REPLICATE

NONE

Printer Dependent

Not used

Not used

4.6 Printer SOP Class

N-GET DIMSE service is supported for the Printer SOP Class. If an N-EVENT-REPORT DIMSE service is received when the association is active, Print SCU handles the relevant states but the data received is ignored.

This table defines the set of attributes that this product may request using the Printer N-Get service.

It also describes the product behavior when receiving the N-Get response from the Printer SCP:

Attribute Name

Printer Status

Printer Status Info

Tag

(2110,0010)

(2110,0020)

Use

NORMAL: Status is logged and sends

Film Session N-CREATE request.

WARNING: Status is logged. Pauses the print job and displays status code on the film composer UI.

FAILURE: Status is logged. Pauses the print job and displays print failure message on the film composer UI.

SUPPLY LOW: Logs the message and continues printing.

SUPPLY EMPTY: print job is paused and the message is logged.

RECEIVER FULL: print job is paused and the message is logged.

FILM JAM: print job is paused and the message is logged.

98 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

Printer Name

Manufacturer

(2110,0030)

(0008,0070)

Manufacturer Model Name (0008,1090)

Device Serial Number

Software Versions

(0018,1000)

(0018,1020)

Date Of Last Calibration

Time Of Last Calibration

(0018,1200)

(0018,1201)

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Logged if sent by SCP

Logged if sent by SCP

Logged if sent by SCP

Logged if sent by SCP

Logged if sent by SCP

Ignored

Ignored

99 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

5 Storage Commitment Push Model Implementation

5.1 Storage Commitment Push Model Information Object Definition

Please refer to DICOM Part 3 (Information Object Definitions) for a description of each of the attributes contained within the Storage Commitment Information Object.

The Storage Commitment Information Object is used both for N-ACTION Storage Commitment

Requests and N-EVENT-REPORT Storage Commitment Notifications by the SCU.

5.1.1 Storage Commitment Module for N-ACTION

Table – 5.1.1-1 Storage Commitment Module for N-ACTION

Attribute Name Tag

Transaction UID

Storage Media File-Set ID

SCU Use

(0008,1195) Generated for each retry

(0088,0130) Not supported

Storage Media File-Set UID

Referenced SOP Sequence

>Referenced SOP Class UID

>Referenced SOP Instance UID

>Storage Media File-Set ID

>Storage Media File-Set UID

(0088,0140) Not supported

(0008,1199)

(0008,1150)

(0008,1155)

(0088,0130) Not supported

(0088,0140) Not supported

5.1.2 Storage Commitment Module for N-EVENT-REPORT

Table – 5.1.2-1 Storage Commitment Module for N-EVENT-REPORT

Attribute Name Tag SCU Use

Transaction UID

Retrieve AE Title

Storage Media File-Set ID

Storage Media File-Set UID

Referenced SOP Sequence

>Referenced SOP Class UID

(0008,1195) Value received from SCP

(0008,0054) Not used

(0088,0130) Not used

(0088,0140) Not used

(0008,1199) When status is SUCCESS, the

“Archived” flag value for the referenced SOP instances is changed to “Yes” in the browser.

In case of partial failure, the archival status of a series is updated only if all the images of the series are archived successfully. Archive status of series is not updated if archiving of one or more images of the series failed. Image level archive status updation is not supported.

(0008,1150)

100 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

>Referenced SOP Instance UID

>Retrieve AE Title

>Storage Media File-Set ID

>Storage Media File-Set UID

Failed SOP Sequence

>Referenced SOP Class UID

>Referenced SOP Instance UID

>Failure Reason

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

(0008,1155)

(0008,0054) Not used.

(0088,0130) Not used

(0088,0140) Not used

(0008,1198) “Archived” flag value for the failed

SOP instance is unaltered. Failed

SOP instances are logged.

(0008,1150)

(0008,1155)

(0008,1197) Not used.

5.1.2.1 Processing of Failure Reason when received in a N-Event-Report

When receiving a N-Event-Report request with Event Type ID equal to 2, meaning that Storage

Commitment is complete, but failure exists, following is the set of value that this Storage

Commitment SCU AE is able to process.

Meaning Failure

Reason

Application Behavior When Receiving Reason

Code

0110H Processing Failure Log file updated: Processing Failure. Job failed.

0112H No such object instance

0213H Resource limitation

0122H Referenced SOP Class not supported

0119H Class / Instance conflict

0131H Duplicate transaction UID

* Other Failure Reason code values

Log file updated: No such object instance. Job failed.

Log file updated: resource limitation. Job failed.

Log file updated: reference SOP class not supported. Job failed.

Log file updated: class/instance conflict. Job failed.

Log file updated: duplicate transaction UID. Job failed.

Log file updated: unknown failure. Job failed.

101 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

6 Modality Worklist Information Model Definition

6.1 Modality Worklist Information Model Definition

6.1.1 Introduction

This section specifies the use of the DICOM Modality Worklist Information Model used to organize data and against which a Modality Worklist Query will be performed.

6.1.2 Modality Worklist Information Model Description

In order to serve as a Service Class Provider (SCP) of the Modality Worklist Service Class, a DICOM

Application Entity (AE) possesses information about the attributes of a number of managed worklist items. These items are organized into Modality Worklist Information Modules. In this Service Class, the Information Model plays a role similar to an Information Object Definition of most other DICOM

Service Classes.

6.1.3 Modality Worklist Information Model Entity-Relationship Model

The Entity-Relationship diagram for the Modality Worklist Information Model schema is shown in

Illustration 6.1.3-1. It represents the information that composes a Worklist Item. In this figure, the following diagrammatic convention is established to represent the information organization:

 Each entity is represented by a rectangular box

 Each relationship is represented by a diamond shaped box.

 The fact that a relationship exists between two entities is depicted by lines connecting the corresponding entity boxes to the relationship boxes.

102 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Illustration 6.1.3-1 – Modality Worklist Information Model E/R DIAGRAM

Scheduled

Procedure Step

Worklist

Item contained in

1

Requested

Procedure requested for

1

Imaging Service

Request done for

Patient

1 is included

0,1

Visit

6.1.3.1 Entity Descriptions

Please refer to DICOM Standard PS 3.3. (Information Object Definitions) and PS 3.4 (Service Class

Specifications) for a description of each of the Entities contained within the Modality Worklist

Information Model.

6.1.3.1.1 Scheduled Procedure Step

A Scheduled Procedure Step is an arbitrarily defined scheduled unit of service that is specified by the

Procedure Plan for a Requested Procedure. It specifies one or more Action Items (events) involving equipment (i.e. imaging modality equipment), human resources, location and time (i.e. start time, stop time, duration).

103 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

6.1.3.1.2 Requested Procedure Entity Description

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

A Requested Procedure is an instance of a Procedure of a given Procedure Type. An instance of a

Requested Procedure includes all of the items of information that are specified by an instance of a

Procedure Plan that is selected for the Requested Procedure by the imaging service provider.

6.1.3.1.3 Imaging Service Request Entity Description

An Imaging Service Request is a set of one or more Requested Procedures selected from a list of

Procedure Types. An Imaging Service Request is submitted by one authorized imaging service requester to one authorized imaging service provider in the context of one Service Episode.

6.1.3.1.4 Visit Entity Description

A Visit is the context in which the treatment or management of an arbitrary subset of a Patient’s medical conditions occurs. A Visit is limited to the description of a Patient’s activities at a single facility.

6.1.3.1.5 Patient Entity Description

A Patient is a person receiving, or registered to receive healthcare services.

6.1.4 ModalityWorklist Mapping of DICOM Entities

Table 6.1.4-1 —Mapping of DICOM Entities to ModalityWorklist Entities

DICOM

Scheduled Procedure Step

Requested Procedure

Imaging Service Request

Visit

Patient

ModalityWorklist Entity

Exam

Exam

Exam

Exam

Patient

6.1.5 Modality Worklist Information Model Module

Within an entity of the DICOM Modality Worklist Information Model, attributes are grouped together into related set of attributes called modules. A module facilitates the understanding of the semantics concerning the attributes and how the attributes relate to one another. A module grouping does not infer any encoding of information into datasets.

Table 6.1.5-1 identifies the defined modules within the entities that comprise the DICOM Modality

Worklist Information Model. Modules are identified by Module Name.

See DICOM PS 3.3 and PS 3.4 for a complete definition of the entities, modules, and attributes.

104 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Table 6.1.5-1 – Modality Worklist Information Model Modules

Entity Name Module Name

Scheduled Procedure Step SOP Common

Requested Procedure

Scheduled Procedure Step

Requested Procedure

Imaging Service Request Imaging Service Request

Visit Visit Identification

Visit Status

Patient

Visit Relationship

Visit Admission

Patient Relationship

Patient Identification

Patient Demographic

Patient Medical

Reference

6.1.6.2.1

6.1.6.2.2

6.1.6.3.1

6.1.6.4.1

6.1.6.5.1

6.1.6.5.2

6.1.6.5.3

6.1.6.5.4

6.1.6.6.1

6.1.6.6.2

6.1.6.6.3

6.1.6.6.4

6.1.6 Information Model Keys

Please refer to DICOM Standard PS 3.3. (Information Object Definitions) and PS 3.4 (Service Class

Specifications) for a description of each of the Entities contained within the Modality Worklist

Information Model.

The following Module descriptions contain the attributes that are present in a C-FIND request message sent by the Worklist Server AE to a remote AE. It should be noted that they are the same as those defined in the DICOM Standard, PS 3.4 (Service Class Specifications) and include:

Attribute Name

Tag: group and element numbers

Expected Matching Key Type: R-required, O-optional

Expected Return Key Type:

 1 - non-zero value required

 1C - conditionally of type 1

 2 - required to be present, possibly with zero-length value

 3 - optional

Mapped into The Image: whether this data is mapped into subsequently acquired images

Notes: clarification of this implementation’s use/treatment of this attribute

All data elements in the following Module descriptions are requested by the Worklist Server AE.

Values of data elements that are not mapped into images, and are not otherwise dealt with

(displayed on the user interface, etc.), are not used and are, thus, discarded upon receipt. See Table

B-1 for further information.

105 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Data elements for which values can be sent for matching purposes are described as such. Data elements for which values are not sent are sent with zero length and universal matching will apply.

This is the default case if no other description to the contrary is provided.

6.1.6.1 Supported Matching

The following are the types of matching that can be request by the implementation:

Single Value matching

Universal Matching

Range of date/time

6.1.6.2 Scheduled Procedure Step Entity

6.1.6.2.1 SOP Common Module

Table 6.1.6.2.1-1 – SOP Common Module Attributes

Attribute Name

Specific Character

Set

Tag

(0008,0005)

Expected

Matching

Key Type

O

Expected

Returned

Key Type

1C

Mapped into the

Image/

MPPS

Note

No/No This value is always sent.

ISO_IR_100 is the default value sent during querying if not set explicitly.

Product uses the value returned if it is set in the response.

6.1.6.2.2 Scheduled Procedure Step Module

Table 6.1.6.2.2-1 – Scheduled Procedure Step Module Attributes

Attribute Name Tag

(0040,0100)

Expected

Matchin g Key

Type

R

Expected

Returned

Key Type

1

Mapped into the

Image/

MPPS

Note

No/No Requested in the MWL queries. Scheduled Procedure

Step Sequence

Scheduled Station AE

Title

(0040,0001) R 1 No/No Matching is supported as follows: either no AE title is supplied

(universal matching), or the scanner’s Worklist Server AE title is supplied for matching; this is user selectable.

106 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

Scheduled Procedure

Step Start Date

Scheduled Procedure

Step Start Time

Modality

Scheduled

Performing

Physician's Name

Scheduled Procedure

Step Description

Scheduled Station

Name

Scheduled Procedure

Step Location

Scheduled Protocol

Code Sequence

Code Value

Coding Scheme

Designator

Code Meaning

(0040,0002)

(0040,0003)

(0008,0060)

(0040,0006)

(0040,0007)

(0040,0010)

(0040,0011)

(0040,0008)

(0008,0100)

(0008,0102)

(0008,0104)

R

R

R

R

O

O

O

O

O

O

O

Pre-Medication

Scheduled

Procedure Step ID

(0040,0012)

(0040,0009)

O

O

107 of 181

1C

1C

3

1

1

1

2

1C

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

No/No Matching is supported as one of the following; this is user selectable:

 all days, today only, today and a number of days before today, today and a number of days after today, today and a number of days before today and a number of days after today.

Number of days before/after is specified by the user.

Returned values must be exactly 8 numeric characters in YYYYMMDD format.

No/No This attribute is sent with zerolength. Returned values must be exactly 6 numeric characters in

HHMMSS format.

Yes/Yes Matching is supported as follows: either no Modality is supplied

(universal matching), or the scanner’s Modality is supplied for matching; this is user selectable.

Yes/Yes This attribute is sent with zerolength. This value is mapped into

(0008, 1050) in the image header.

Yes/Yes

2 No/No

2

1C

2C

1

No/No

Yes/Yes Up to 5 Scheduled Protocol Code

Sequence Items are mapped into the image if they system is configured to support GSPS.

Yes/Yes

Yes/Yes

Yes/Yes NOTE: though type 3, the scanner expects Code Meaning to be sent if

(0040,0008) is sent.

No/No

Yes/Yes

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

Requested

Contrast Agent

(0032,1070) O 2C

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

No/No Displayed on “More Info...” screen.

6.1.6.3 Requested Procedure Entity

6.1.6.3.1 Requested Procedure Module

Table 6.2.6.3.1-1 Requested Procedure Module Attributes

Attribute Name

Requested Procedure ID

Requested Procedure

Description

Tag

(0040,1001)

(0032,1060)

(0032,1064)

Expected

Matching

Key Type

O

O

O

Expected

Returned

Key Type

1

1C

1C

Mapped into the

Image/

MPPS

Note

Yes/Yes User can enter the value for

Requested Procedure Id prior to query. If user has entered the value then that value will be sent as part of the query.

The value returned in the response shall be mapped to the image.

Yes/Yes Truncated to 22 characters by default however this is configurable to allow 64 characters.

Yes/No Requested Procedure Code

Sequence

Code Value

Coding Scheme

Designator

Code Meaning

(0008,0100)

(0008,0102)

(0008,0104)

O

O

O

1C

1C

3

Yes/No

Yes/No

Yes/No NOTE: thought type 3, the scanner expects Code

Meaning to be sent if

(0032,1064) is sent.

108 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

Attribute Name

Study Instance UID

Referenced Study Sequence

Tag

(0020,000D)

(0008,1110)

Expected

Matching

Key Type

O

O

O

O

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Expected

Returned

Key Type

1

2

1C

1C

Mapped into the

Image/

MPPS

Note

Yes

(Based on User option)/Y es

User ( Modality ) will be able to configure the following. If the user has set the option to

“Use MWL Study UID”, then study Instance UID will be copied into the final DICOM image header. If this option is not set then a new Study instance UID is generated locally on the scanner.

The default setting is to generate a new study instance uid if a worklist entry is re-used however it can be configured to reuse the same study instance uid.

Yes/Yes Only 1 Referenced Study

Sequence is mapped into the image.

Yes/Yes

Yes/Yes

Referenced SOP Class UID (0008,1150)

Referenced SOP Instance

UID

(0008,1155)

Requested Procedure Priority (0040,1003)

(0040,1004) Patient Transport

Arrangements

Requested Procedure

Location

Confidentiality Code

(0040,1005)

(0040,1008)

O

O

O

O

2

2

3

3

No/No

No/No

No/No

No/No

6.1.6.4 Imaging Service Request Entity

6.1.6.4.1 Imaging Service Request Module

Table 6.1.6.4.1-1 – Imaging Service Request Module Attributes

Attribute Name Tag Expected

Matching

Key Type

Expected

Returned

Key Type

Mapped into the

Image/

MPPS

Note

109 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

Accession Number (0008,0050)

Requesting Physician (0032,1032)

Referring Physician's Name (0008,0090)

Requesting Service (0032,1033)

O

O

O

O

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

2

2

2

3

Yes/Yes User will be able to enter value for Accession Number prior to query to be sent as part of C-FIND request.

Supports maximum of 16 characters.

No/No

Yes/Yes Truncated to 32 characters by default however this is configurable to allow 64 characters.

No/No

6.1.6.5 Visit Entity

6.1.6.5.1 Visit Identification

Table 6.1.6.5.1-1 – Visit Identification Module Attributes

Attribute Name

Admission ID

Institution Name

Tag

(0038,0010)

(0008.0080)

Expected

Matching

Key Type

O

O

Expected

Returned

Key Type

2

3

Mapped into the

Image/

MPPS

No/No

No/No

Note

6.1.6.5.2 Visit Status

Table 6.1.6.5.2-1 – Visit Status Module Attributes

Attribute Name

Current Patient Location

Tag

(0038,0300)

Expected

Matching

Key Type

O

Expected

Returned

Key Type

2

Mapped into the

Image/

MPPS

Note

No/No Displayed on “More Info...” screen.

110 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

6.1.6.5.3 Visit Relationship

Table 6.1.6.5.3-1 – Relationship Module Attributes

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Attribute Name Tag

Referenced Patient Sequence

Referenced SOP Class UID

(0008,1120)

(0008,1150)

Referenced SOP Instance UID (0008,1155)

Expected

Matching

Key Type

O

O

O

Expected

Returned

Key Type

2

2

2

Mapped into the

Image/

MPPS

No/Yes

No/Yes

No/Yes

Note

6.1.6.5.4 Visit Admission

No data elements are requested from the Visit Admission Module.

6.1.6.6 Patient Entity

6.1.6.6.1 Patient Relationship

No data elements are requested from the Patient Relationship Module.

6.1.6.6.2 Patient Identification

Table 6.1.6.6.2-1 – Patient Identification Module Attributes

Attribute

Name

Patient's

Name

Patient ID

Tag

(0010,0010)

(0010,0020)

(0010, 1000)

Expected

Matching

Key Type

Expected

Returned

Key Type

R

R

O

1

1

3

Mapped into the

Image/

MPPS

Note

Yes/Yes Supports up to 32 characters by default however this is configurable to allow 64 characters. See also Note1.

Yes/Yes Supports up to 16 characters by default however this is configurable to 64 characters.

Yes/No Supports 64 characters. Other

Patient ID

Note 1 :

Modality Worklist server supports 32 characters for patient Name

Supports DICOM format for patient Name ( with “^” as delimiters )

If patient name in worklist has more than 32 characters then

Worklist will be accepted by the server

Worklist Browser will display only the first 32 characters

“More Info” screen will display the full patient name

Only the FIRST 32 characters is copied into the final DICOM image header by default

however this is configurable to allow 64 characters.

111 of 181

GE H

EALTHCARE

D

IRECTION

DOC1906666, R

EV

1

6.1.6.6.3 Patient Demographic

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Table 6.1.6.6.3-1 – Patient Demographic Module Attributes

Attribute Name

Patient’s Birth Date

Patient's Sex

Patient's Weight

Tag

(0010,0030)

(0010,0040)

(0010,1030)

(0040,3001)

Expected

Matching

Key Type

O

O

O

O

Expected

Returned

Key Type

2

2

2

2

Mapped into the

Image/

MPPS

Note

Yes/Yes This value is also used to calculate the Patient’s Age.

Yes/Yes

Yes/No Limited to maximum value of

999 kg.

No/No Confidentiality constraint on patient data

Patient’s Size

Patient’s Address

Patient’s Telephone

Numbers

(0010,1020)

(0010,1040)

(0010,2154)

O

O

O

3

3

3

No/No

No/No

No/No

6.1.6.6.4 Patient Medical

Table 6.1.6.6.4-1 – Patient Medical Module Attributes

Attribute Name

Patient State

Pregnancy Status

Medical Alerts

Contrast Allergies

Special Needs

Tag

(0038,0500)

(0010,21C0)

(0010,2000)

(0010,2110)

(0038,0050)

Additional Patient History (0010,21B0)

Expected

Matching

Key Type

O

O

O

O

O

O

Expected

Returned

Key Type

2

2

2

2

2

3

Mapped into the

Image/

MPPS

Note

No/No

No/No Displayed on “More Info...” screen.

No/No Displayed on “More Info...” screen.

No/No Displayed on “More Info...” screen.

No/No Displayed on “More Info...” screen.

Yes/No Displayed on “More Info...” screen.

6.2 Private Data Dictionary

The ModalityWorklist implementation does not define any Private Attributes within the Modality

Worklist Information Model.

112 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

6.3 C-FIND Request Message

This section provides a detailed description of the C-FIND request message data that is provided to the remote AE during a worklist query operation. The dump in Table 6.4-1 below lists, in exact message order, the fields transferred as part of the C-FIND request message for a typical query.

In this particular dump, no values are specified for the Scheduled Procedure Step Start and

End Dates (the attributes are sent with zero length). In DICOM this is interpreted as meaning all dates (i.e. universal matching). The Modality is also not specified in this particular dump, meaning all modalities. Note that the user, through the use of the GEHC CT user interface, can submit a worklist query that will cause non-zero values to be sent for these attributes.

Table 6.3-1 – C-FIND Request Message Dump

(0008,0000) UL 108 # 4, 1 IdentifyingGroupLength

(0008,0005) CS [ISO_IR 100] # 12, 1 SpecificCharacterSet

(0008,0050) SH (no value available) # 0, 0 AccessionNumber

(0008,0080) LO (no value available) # 0, 0 InstitutionName

(0008,0090) PN (no value available) # 0, 0 ReferringPhysicianName

(0008,1110) SQ (Sequence with explicit Length #=1) # 24, 1 ReferencedStudySequence

(fffe,e000) na (Item with explicit Length #=2) # 16, 1 Item

(0008,1150) UI (no value available) # 0, 0 ReferencedSOPClassUID

(0008,1155) UI (no value available) # 0, 0 ReferencedSOPInstanceUID

(fffe,e00d) na (ItemDelimitationItem for re-encoding) # 0, 1 ItemDelimitationItem

(fffe,e0dd) na (SequenceDelimitationItem for re-enc.) # 0, 1 SequenceDelimitationItem

(0008,1120) SQ (Sequence with explicit Length #=1) # 24, 1 ReferencedPatientSequence

(fffe,e000) na (Item with explicit Length #=2) # 16, 1 Item

(0008,1150) UI (no value available) # 0, 0 ReferencedSOPClassUID

(0008,1155) UI (no value available) # 0, 0 ReferencedSOPInstanceUID

(fffe,e00d) na (ItemDelimitationItem for re-encoding) # 0, 1 ItemDelimitationItem

(fffe,e0dd) na (SequenceDelimitationItem for re-enc.) # 0, 1 SequenceDelimitationItem

(0010,0000) UL 96 # 4, 1 PatientGroupLength

(0010,0010) PN (no value available) # 0, 0 PatientName

(0010,0020) LO (no value available) # 0, 0 PatientID

(0010,0030) DA (no value available) # 0, 0 PatientBirthDate

(0010,0040) CS (no value available) # 0, 0 PatientSex

(0010,1000) LO (no value available) # 0, 0 Other Patient IDs

(0010,1020) DS (no value available) # 0, 0 PatientSize

(0010,1030) DS (no value available) # 0, 0 PatientWeight

(0010,1040) LO (no value available) # 0, 0 PatientAddress

(0010,2000) LO (no value available) # 0, 0 MedicalAlerts

(0010,2110) LO (no value available) # 0, 0 ContrastAllergies

(0010,2154) SH (no value available) # 0, 0 PatientTelephoneNumber

(0010,2160) SH (no value available) # 0, 0 Ethinc Group

(0010,21b0) LT (no value available) # 0, 1 AdditionalPatientHistory

(0010,21c0) US (no value available) # 0, 0 PregnancyStatus

(0010,4000) LT (no value available) # 0, 0 Patient Comments

(0020,0000) UL 8 # 4, 1 ImageGroupLength

(0020,000d) UI (no value available) # 0, 0 StudyInstanceUID

(0032,0000) UL 64 # 4, 1 StudyGroupLength

(0032,1032) PN (no value available) # 0, 0 RequestingPhysician

(0032,1033) LO (no value available) # 0, 0 RequestingService

(0032,1060) LO (no value available) # 0, 0 RequestedProcedureDescription

(0032,1064) SQ (Sequence with explicit Length #=1) # 32, 1 RequestedProcedureCodeSequence

(fffe,e000) na (Item with explicit Length #=3) # 24, 1 Item

(0008,0100) SH (no value available) # 0, 0 CodeValue

113 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

(0008,0102) SH (no value available) # 0, 0 CodingSchemeDesignator

(0008,0104) LO (no value available) # 0, 0 CodeMeaning

(fffe,e00d) na (ItemDelimitationItem for re-encoding) # 0, 1 ItemDelimitationItem

(fffe,e0dd) na (SequenceDelimitationItem for re-enc.) # 0, 1 SequenceDelimitationItem

(0038,0000) UL 32 # 4, 1 VisitGroupLength

(0038,0010) LO (no value available) # 0, 0 AdmissionID

(0038,0050) LO (no value available) # 0, 0 SpecialNeeds

(0038,0300) LO (no value available) # 0, 0 CurrentPatientLocation

(0038,0500) LO (no value available) # 0, 0 PatientState

(0040,0000) UL 192 # 4, 1 ModalityWorklistGroupLength

(0040,0100) SQ (Sequence with explicit Length #=1) # 136, 1 ScheduledProcedureStepSequence

(fffe,e000) na (Item with explicit Length #=12) # 128, 1 Item

(0008,0060) CS (no value available) # 0, 0 Modality

(0032,1070) LO (no value available) # 0, 0 RequestedContrastAgent

(0040,0001) AE (no value available) # 0, 0 ScheduledStationAETitle

(0040,0002) DA (no value available) # 0, 0 ScheduledProcedureStepStartDate

(0040,0003) TM (no value available) # 0, 0 ScheduledProcedureStepStartTime

(0040,0006) PN (no value available) # 0, 0 ScheduledPerformingPhysiciansName

(0040,0007) LO (no value available) # 0, 0 ScheduledProcedureStepDescription

(0040,0008) SQ (Seq with explicit Length #=1) # 32, 1 ScheduledActionItemCodeSequence

(fffe,e000) na (Item with explicit Length #=3) # 24, 1 Item

(0008,0100) SH (no value available) # 0, 0 CodeValue

(0008,0102) SH (no value available) # 0, 0 CodingSchemeDesignator

(0008,0104) LO (no value available) # 0, 0 CodeMeaning

(fffe,e00d) na (ItemDelimitationItem for re-encoding) # 0, 1 ItemDelimitationItem

(fffe,e0dd) na (SequenceDelimitationItem for re-enc.) # 0, 1 SequenceDelimitationItem

(0040,0009) SH (no value available) # 0, 0 ScheduledProcedureStepID

(0040,0010) SH (no value available) # 0, 0 ScheduledStationName

(0040,0011) SH (no value available) # 0, 0 ScheduledProcedureStepLocation

(0040,0012) LO (no value available) # 0, 0 PreMedication

(fffe,e00d) na (ItemDelimitationItem for re-encoding) # 0, 1 ItemDelimitationItem

(fffe,e0dd) na (SequenceDelimitationItem for re-enc.) # 0, 1 SequenceDelimitationItem

(0040,1001) SH (no value available) # 0, 0 RequestedProcedureID

(0040,1003) SH (no value available) # 0, 0 RequestedProcedurePriority

(0040,1004) LO (no value available) # 0, 0 PatientTransportArrangements

(0040,1005) LO (no value available) # 0, 0 RequestedProcedureLocation

(0040,1008) LO (no value available) # 0, 0 ConfidentialityCode

(0040,1010) PN (no value available) # 0, 0 Names of Intended Recipients

(0040,1400) LT (no value available) # 0, 0 Requested Procedure Comments

(0040,2400) LT (no value available) # 0, 0 Imaging Service Request Comm.

(0040,3001) LO (no value available) # 0, 0 ConfidentialityConstraintOnPatientData

S

TATEMENT

If the query is for a particular date range, the ScheduledProcedureStepStartDate will be filled with a valid date range. If either the start or end date are left blank by the user, they will simply be blank in the query.

Below is an example of a date range for August 30, 1997 through October 12, 1997.

(0040,0002) DA [19970830-19971012] # 18, 1 ScheduledProcedureStepStartDate

Below is an example of a date range for August 30, 1997 through the end of time.

(0040,0002) DA [19970830-] # 18, 1 ScheduledProcedureStepStartDate

Below is an example of a date range from the beginning of time through August 30, 1997.

(0040,0002) DA [-19970830] # 18, 1 ScheduledProcedureStepStartDate

If the query is for records for this modality, the Modality will be filled in as follows:

(

0008,0060) CS [CT] # 2, 1 Modality

114 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

If the query is for records for this Scanner, the Modality will be filled in with CT as above and the

Scheduled Station AE Title will be filled in with the value configured for this system. For example, this station was configured as CTRoom1.

(0040,0001) AE [CTRoom1] # 8, 1 ScheduledStationAETitle

User will be able to enter the values for “Accession Number” prior to the query. If value is entered then that value will be sent as part of the query. For example, if “1234” is entered then

(0008,0050) SH [1234] # 4, 1 AccessionNumber

User will be able to enter the values for “Requested Procedure Id” prior to the query. If value is entered then that value will be sent as part of the query. For example, if “3456” is entered then

(0040,1001) SH [3456] # 4, 1 RequestedProcedureID

User will be able to enter the values for “Patient Id” prior to the query. If value is entered then that value will be sent as part of the query. For example, if “6789” is entered then

(0010,0020) LO [6789] # 4, 1 PatientID

User will be able to enter the values for “Patient Name” prior to the query. If value is entered then that value will be sent as part of the query. For example, if “Lastname^Firstname” is entered then

(0010,0010) PN [Lastname^Firstname] # 18, 1 PatientName

6.4 Use of Specific DICOM Data

This section details the use of the DICOM data returned by remote AEs during worklist queries.

The GEHC CT user interface fields that display the data, along with the data’s mapping into resulting acquired and transferred DICOM images, are presented in following table 6.5-1.

Table 6.4-1 – Specific Data Usage

DICOM Worklist Data

Element

Accession Number

(0008,0050)

Patient Schedule Screen Field GEHC CT DICOM Image Data Element

Req Number

Patient ID (0010,0020) Patient ID

Accession Number

Supports maximum of 16 characters.

Can be configured so the top level accession number tag in the image header will be empty if worklists are grouped and have different accession numbers. Default behavior is to fill in the top level accession number in the grouped case.

Patient ID

Supports maximum of 16 characters.

Can be configured to allow 64 characters.

115 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Other Patient ID (0010,1000) Other Patient Ids (only displayed on the More Info screen)

Patient Name (0010,0010) Patient Name

Patient ID

Supports maximum of 64 characters.

Patient’s Birth Date

(0010,0030)

Patient's Sex (0010,0040)

Patient Age (Patient Birth Date user to calculate age)

Sex

Patient's Weight (0010,1030) Weight in Kg

Referring Physician’s Name

(0008,0090)

Referring Physician

Exam Description Requested Procedure

Description (0032,1060)

Scheduled Procedure Step

Start Date

(0040,0002)

Scheduled Procedure Step

Start Time

(0040,0003)

Scheduled Performing

Physician’s Name (0040,0006)

Study Instance UID

(0020,000d)

Date

Time

Performing Physician (only displayed on the “More Info..” screen)

Study instance UID (only displayed on the “More Info...” screen)

Patient Name

Supports maximum of 32 characters.

Can be configured to allow 64 characters.

Patient Birth Date

Patient’s Sex

Patient's Weight

Referring Physician’s Name

Study Description.

Not available.

Not available.

This value is mapped into (0008, 1050) in the image header.

Requested Procedure Id

(0040,1001)

Requested Proc ID

Pregnancy Status (0010,21C0) Pregnancy Status (only displayed on the “More Info...” screen)

Medical Alerts (0010,2000)

Contrast Allergies (0010,2110) Contrast Allergies (only displayed on the “More Info...” screen)

Special Needs (0038,0050)

Medical Alerts (only displayed on the “More Info...” screen)

Special Needs (only displayed on the “More Info...” screen)

Requested Contrast Agent

(0032,1070)

Requested Contrast Agent (only displayed on the “More Info...” screen)

User ( Modality ) will be able to configure the following. If the user has set the option to “Use MWL Study UID”, then study Instance UID will be copied into the final DICOM image header. If this option is not set then a new Study instance UID is generated locally on the scanner.

Requested Procedure Id.

Not available.

Not available.

Not available.

Not available.

Not available.

116 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

Current Patient Location

(0038,0300)

Additional Patient History

(0010,21B0)

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Current Patient Location (only displayed on the “More Info...” screen)

Additional Patient History (only displayed on the “More Info...” screen)

Not available.

Patient History.

Note that the display of a specific data item on the “More Info...” screen is contingent on the item being enabled for display. Depending on the preferences of each specific site, data can either be displayed or not. A GE field service engineer can assist in setting these site preferences.

6.5 Setting User Preferences

6.5.1 Setting “Use MWL Study UID “ Option

Setting this option to “Yes” copies the Study Instance UID from Worklist into the final DICOM

Image header. If this option is set to “No” then a new study instance uid will be generated locally.

1. Click on “Patient Schedule”

2. Click on “Preferences” button

3. Set the option “Use Study UID?”, to either “Yes” or “No”

6.5.2 Setting Custom Query Option

This option allows the user to enter values for “Accession Number” and / or “ Requested

Procedure Id” , which are used for Custom Query.

1. Click on “Patient Schedule”

2. Click on “Preferences” button

3. Set the option “Show Update Parameters? ”, to “Yes”

4. To do a query click on “Update” button.

A User Interface appears, with provision to enter values for: a) Accession Number b) Requested Proc. ID

117 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

7 PERFORMED PROCEDURE STEP CONFORMANCE STATEMENT

7.1 Introduction

This section of the DICOM Conformance Statement specifies the compliance to DICOM conformance requirements for the Modality Performed Procedure Step feature on this GEHC product. Note that the format of this section strictly follows the format defined in DICOM

Standard PS 3.2 (Conformance). Please refer to that part of the standard while reading this section. The PPS option for GEHC CT allows a Modality Performed Procedure Step to be communicated to the Hospital/Radiology information system. The PPS feature is providing the

DICOM Modality Performed Procedure Step service as a service class user (SCU).

This feature works in conjunction with DICOM Modality Worklist feature, if installed. However the conformance of this feature is independent of Modality Worklist feature. For information on conformance of Modality Worklist feature to DICOM standard please refer to the appropriate section or document.

7.2 N-CREATE & N-SET Request Message

PPS Feature for GEHC CT supports all named attributes listed in Table F.7.2.1 in PS3.4 of DICOM standard. That is, attributes that are not explicitly referenced by name in the table are not supported. (Example is last row in the table reads “All other attributes from Radiation Dose

Module and Billing and Material Code Module”. The attributes referenced here are not supported).

For the MPPS associated with an acquisition, the following attributes are copied from the

Modality Worklist SCU into the MPPS request Message, if procedure performed corresponds to the SPS information retrieved through the Modality Worklist.

GEHC CT supports the selection of single or multiple SPS for a scan. The following are applicable.

Single SPS selection results in single PPS message

Multiple SPS selection results in single PPS message

Multiple SPS selection is allowed only if they all correspond to same patient id

A maximum of 15 SPS’s can be selected

Referenced Study Sequence - a maximum of ten Item’s is supported. This attribute will be

present only if SPS information is available from Modality Worklist SCU.

At the end of acquisition the user might choose to ‘Defer PPS’ and later choose to

‘Complete PPS’ or ‘Discontinue PPS’ from the user interface provided in the system. In this case, the date and time when user chooses to ‘Complete PPS’ or ‘Discontinue PPS’ is taken as the Performed Procedure Step End Date and Performed Procedure Step End Time respectively (Not the actual end date and end time of acquisition)

Mapping of SPS data to MPPS SOP instance is explained in section 7.3

118 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Mapping of specific SPS data to CT DICOM IMAGE HEADER , for PPS is explained in section

7.4

For the MPPS associated with a post-processing the following restrictions apply on the attributes listed below.

Referenced Study Sequence – The sequence is not sent in the MPPS message

Scheduled Step Attribute Sequence – a maximum of ONE item is supported. The attribute will be send only if SPS information is available in the image instance.

Referenced Patient Sequence – This sequence is not added

Scheduled Protocol Code Sequence – The sequence is not send in the MPPS message

Performed Procedure Step Start date & Performed Procedure Step start time – The exam date and exam time that is the Start date and Start Time of the associated Study

Component (Exam) is used, not the actual time when post-processing started.

Performed Procedure Step end date & Performed Procedure Step end time - The date and time when user chooses to ‘Complete PPS’ or ‘Discontinue PPS’ is taken as the Performed

Procedure Step End Date and Performed Procedure Step End Time respectively (Not the actual end date and end time of post-processing).

Procedure Code Sequence – This sequence is sent with ZERO items in the MPPS message

Performed Protocol Code Sequence – This sequence is sent with ZERO items in the MPPS message.

Referenced Standalone SOP Instance Sequence – The sequence is sent with ZERO items in the MPPS message.

119 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

7.3 MODALITY PERFORMED PROCEDURE STEP MODULE DEFINITIONS

The following table gives specific usage of some of attributes in the MPPS SOP instance created, with reference to each of the real-world scenarios mentioned in 2.3.1.2.7.1.1 of this document.

Attribute Name Tag

Specific Character Set

Scheduled Step Attribute

Sequence

>Study Instance UID

(0008,0005)

(0040,0270)

(0020,000D)

Acquisition with

MWL data

Not used, but copied into image header

Up to 15 items

Usage in MPPS Instance

Acquisition without

MWL data

Post-Processing

Not used Not used, but copied into image header

Up to 15 items

Copied from source image

>Referenced Study

Sequence

>Accession Number

(0008,1110)

(0008,0050)

Copied from SPS, if option to copy is selected or else created at the scanner

Copied from SPS, if selected

Copied from SPS, if selected

Not sent as part of image header

Created at the scanner

Not sent as part of image header

User input on the scanner

Not sent as part of image header

Not sent as part of image header

Copied from source image

Not sent as part of image header

>Placer Order

Number/Imaging Service

Request

(0040,2016)

>Filler Order

Number/Imaging Service

Request

(0040,2017)

>Requested Procedure ID (0040,1001)

>Requested Procedure

Description

(0032,1060)

>Placer Order

Number/Procedure

>Filler Order

Number/Procedure

(0040,1006)

(0040,1007)

>Scheduled Procedure ID (0040,0009)

> Scheduled Performing

Physician’s Name

>Scheduled Procedure

Step Description

>Scheduled Protocol Code

Sequence

Patient's Name

(0040,0006)

(0040,0007)

(0040,0008)

(0010,0010)

Patient ID (0010,0020)

Not sent as part of image header

Not sent as part of image header

Copied from SPS, if selected

Copied from SPS, if selected

Not sent as part of image header

Not sent as part of image header

Copied from SPS, if selected

Copied from SPS, if selected

Copied from SPS, if selected

Copied from SPS, if selected

Copied from SPS, if selected

Copied from SPS, if selected

User input on the scanner

Not used

Not used

Not used

Not used

Not used

Not used

Not used

User input on the scanner

User input on the scanner

Not sent as part of image header

Copied from source image

Not sent as part of image header

Not sent as part of image header

Not sent as part of image header

Copied from source image

Copied from source image

Copied from source image

Not sent as part of image header

Copied from source image

Copied from source image

120 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

Attribute Name

Patient's Birth Date

Patient's Sex

Tag

(0010,0030)

(0010,0040)

Referenced Patient

Sequence

Performed Procedure Step

ID

(0008,1120)

(0040,0253)

Performed Station AE Title (0040,0241)

Performed Station Name (0040,0242)

Performed Location (0040,0243)

Performed Procedure Step

Start Date

Performed Procedure Step

Start Time

Performed Procedure Step

Description

Performed Procedure Step status

(0040,0244)

(0040,0245)

(0040,0254)

(0040,0252)

Performed Procedure Type

Description

(0040,0255)

Procedure Code Sequence (0008,1032)

Performed Procedure Step

End Date

Performed Procedure Step

End Time

Modality

(0040,0250)

(0040,0251)

(0008,0060)

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Acquisition with

MWL data

Copied from SPS, if selected

Copied from SPS, if selected

Copied from SPS, if selected

Created at the scanner. Will have the following

“PPS_ID_<exam

number”

Local system host-name

Local system suite id

Local system suite id

Same as exam start date

Same as exam start time

Copied from SPS, if selected

See Note 1.

Usage in MPPS Instance

Acquisition without

MWL data

Post-Processing

User input on the scanner

User input on the scanner

Not used

Copied from source image

Copied from source image

No item

Created at the scanner. Will have the following

“PPS_ID_<exam

number”

Local system hostname

Local system hostname

Local system suite id Local system hostname

Local system suite id Not used

Same as exam date

Same as exam time

Not sent as part of image header

See Note 1.

New generated, may not be unique

Same as exam date

Same as exam time

Not sent as part of image header

See Note 1.

Not sent as part of image header

Not sent as part of image header.

Not sent in MPPS

N-Create.

Date when all images got installed

Time when all images got installed

Value “CT” is stored in image header

Not sent as part of image header

Not sent as part of image header. Not sent in MPPS N-

Create.

Date when all images got installed

Time when all images got installed

Value “CT” is stored in image header

Not sent as part of image header

Not sent as part of image header. Not sent in MPPS N-

Create.

The date “Complete

PPS” or

“Discontinue PPS” is invoked, not the actual end of postprocessing

The time “Complete

PPS” or

“Discontinue PPS” is invoked, not the actual end of postprocessing

Value “CT” is stored in image header

121 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

Attribute Name

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Study ID

Performed Protocol Code

Sequence

Performed Series

Sequence

>Performing Physician's

Name

>Protocol Name

>Operator's Name

>Retrieve AE Title

>Referenced Image

Sequence

Tag

(0020,0010)

(0040,0260)

(0040,0340)

(0008,1050)

(0018,1030)

(0008,1070)

(0008,0054)

(0008,1140)

>Referenced Standalone

SOP Instance Sequence

>All other attributes from

Performed Series

Sequence (which Table

F.7.2.1 of DICOM standard

PS3.4 does not explicitly list)

All other attributes from

Radiation Dose Module and Billing and Material

Code Module (which Table

F.7.2.1 of DICOM standard

PS3.4 does not explicitly list)

(0040,0220)

Acquisition with

MWL data

Same as exam number

Derived from

(0040, 0008)

One item for each series created

Copied from SPS, if selected

The name of the protocol selected on the scanner

Copied from SPS if selected and present otherwise from user input on the scanner

Local system host-name

One item for each image created within the series

Not sent as part of image header

Not sent as part of image header

Usage in MPPS Instance

Acquisition without

MWL data

Post-Processing

Same as exam number

Not used

Copied from source image

Not used

One item for each series created

User input on the scanner

The name of the protocol selected on the scanner

User input on the scanner

Local system hostname

One item for each image created within the series

Not sent as part of image header

Not sent as part of image header

One item for each series created with post-processing

Not sent as part of image header

Copied from source image

Not sent as part of image header host-name of the system

One item for each image generated by post-processing

Not sent as part of image header

Not sent as part of image header

Not sent as part of image header

Not sent as part of image header

Not sent as part of image header

Note 1:

 When PPS start ( N-CREATE) message is sent, this element will have the value

“IN PROGRESS”

 When PPS end ( N-SET) message is sent, this element will have either

“COMPLETED” or “DISCONTINUED” based on user selection

122 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

7.4 Use of Specific DICOM Data

7.4.1 Patient Level

Attribute Name

Patient Name

Patient ID

Patient Birthdate

Patient Sex

Referenced Patient Sequence

Ref. SOP class uid

Ref. SOP Instance uid

Tag

(0010,0010)

(0010,0020)

(0010,0030)

(0010,0040)

(0008,1120)

(0008,1150)

(0008,1155)

7.4.2 Study Level

Attribute Name

Study Instance UID

Study ID

Referring Physicians name

Accession Number

Tag

(0020,000D)

(0020,0010)

(0008,0090)

(0008,0050)

(0008,1110)

(0008,1150)

(0008,1155)

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Usage in CT DICOM Image Header

Copied from SPS, if selected

Copied from SPS, if selected

Copied from SPS, if selected

Copied from SPS, if selected

Copied from SPS, if selected

Usage in CT DICOM Image Header

Copied from SPS, if selected

Scanner generated study ID

Copied from SPS, if selected

If multiple SPS’s are selected, then accession number from the first selection (determined by the user) is used, however, it is configurable to be set to zero length. Default is to use the accession number from the first selection.

Copied from SPS, if selected Referenced Study Sequence

Ref. SOP class uid

Ref. SOP Instance uid

7.4.3 Series Level

Attribute Name

Modality

Protocol Name

Operator Name

Referenced Performed

Procedure Step Sequence

Ref. SOP class uid

Ref. SOP Instance uid

Requested Attribute Sequence

Tag

(0008,0060)

(0018,1030)

(0008,1070)

(0008,1111)

(0008,1150)

(0008,1155)

(0040,0275)

Usage in CT DICOM Image Header

Value “CT” is copied

Name of the selected protocol is copied

Copied from SPS if selected and present otherwise generated from user input on scanner

Will be present only if SPS , obtained from HIS/RIS is selected for scanning

Value of MPPS SOP class UID

Scanner generated unique UID

Will be present only if SPS obtained from HIS/RIS is selected for scanning. If multiple SPS’s are selected then this will contain multiple items one for each

123 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

Attribute Name

Requested Procedure Id

Scheduled Procedure Step

Id

Scheduled Procedure Step description

Scheduled Protocol Code

Sequence

Performed Procedure Step Id

Performed Procedure Step start date

Performed Procedure Step start time

Performed Procedure Step description

Tag

(0040,1001)

(0040,0009)

(0040,0007)

(0040,0008)

(0040,0253)

(0040,0244)

(0040,0245)

(0040,0254)

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Usage in CT DICOM Image Header

SPS. For all other cases this element will not be present

Copied from SPS, if selected

Copied from SPS, if selected

Copied from SPS, if selected

Copied from SPS, if selected

Created at the scanner. The string “PPS_ID_<exam

number” is used.

Same as exam start date

Same as exam start time

Exam description is used

124 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

8 Grayscale Softcopy Presentation State

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

8.1 Introduction

This section specifies the use of the DICOM Grayscale Softcopy Presentation State (GSPS) IOD to represent the information included in GSPSs produced by this product.

8.2 CT Mapping of DICOM Entities

The CT system maps DICOM Information Entities to local Information Entities in the product’s database and user interface.

DICOM IE CT Entity

Patient Patient

Study

Series

Presentation State

Exam

Series

Presentation State

8.3 IOD Module Table

The Grayscale Softcopy Presentation State Information Object Definition comprises the modules of the following table.

Entity Name Module Name Reference Usage

Patient

Study

Series

Equipment

Presentation State

Patient

General Study

Patient Study

General Series

Presentation Series

General Equipment

8.4.1.1

8.4.2.1

8.4.2.2

8.4.3.1

8.4.3.2

8.4.4.1

Presentation State

Identification

Presentation State

Relationship

Presentation State

Shutter

8.4.5.1

8.4.5.2

Exam Split : Not sent (attribute condition not met)

Save State : 8.4.5.3

Presentation State Mask Not sent (attribute condition not met)

Mask Not sent (no multi-frame present)

Display Shutter Not sent (no display shutter)

Bitmap Display Shutter Not sent (no bitmap data present)

Overlay Plane

Overlay Activation

Not sent (no overlay data present)

Not sent (no overlay data present)

M

C

C

C

C

C

M

M

U

M

M

M

M

M

M

125 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Displayed Area 8.4.5.4

Graphic Annotation Not sent (no annotation data present)

Spatial Transformation Not sent (no flip/rotate being applied)

Graphic Layer

Modality LUT

Softcopy VOI LUT

Softcopy Presentation

LUT

SOP Common

Not sent (no graphics layers are added)

8.4.5.5

8.4.5.6

8.4.5.7

8.4.5.8 M

C

C

C

M

M

C

C

8.4 Information Module Definitions

Please refer to DICOM Part 3 (Information Object Definitions) for a description of each of the entities, modules, and attributes contained within the GSPS Information Object.

The following modules are included to convey Enumerated Values, Defined Terms, and Optional

Attributes supported. Type 1 & Type 2 Attributes are also included for completeness and to define what values they may take and where these values are obtained from when generating the instance. It should be noted that they are the same ones as defined in the DICOM Standard Part

3 (Information Object Definitions). Also note that Attributes not present in tables are not supported

8.4.1 Patient Entity Module

8.4.1.1 Patient Module

Attribute Name

Patient’s Name

Patient ID

Patient’s Birth Date

Patient’s Sex

Other Patient ID

Tag

(0010,0010)

(0010,0020)

(0010,0030)

(0010,0040)

(0010,1000)

Type Notes

2

2

2

2

3

Copied from original images

Copied from original images

Copied from original images

Copied from original images

Copied from original images

8.4.2 Study Entity Module

8.4.2.1 General Study Module

Attribute Name

Study Instance UID

Study Date

Study Time

Accession Number

Referring Physician’s

Name

Tag

(0020,000D)

(0008,0020)

(0008,0030)

(0008,0050)

(0008,0090)

Type Notes

1

2

Copied from original images

Copied from original images

2

2

2

Copied from original images

Copied from original images

Copied from original images

126 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Attribute Name

Referenced Study

Sequence

> Referenced SOP

Class UID

> Referrenced SOP

Instance UID

Study ID

Study Description

Tag

(0008,1110)

(0008,1150)

(0008,1155)

(0020,0010)

(0008,1030)

Type Notes

3 Exam Split : Copied from original images

Save State : Not sent

1

1

2

3

Value copied from original Images corresponding to the Requested Procedure this GSPS is responding to.

Value copied from original Images corresponding to the Requested Procedure this GSPS is responding to

Copied from original images

Exam Split : Copied from original images

Save State : Not Sent

Note: If two or more of the accession numbers in the grouped case are different, the top level Accession Number

(0008, 0050) shall contain 1 of the accession numbers if the relevant CT Image Objects contain a value in the top level Accession Number (0008, 0050) otherwise the top level Accession Number shall be zero length.

8.4.2.2 Patient Study Module

Attribute Name

Patient’s Age

Patient’s Size

Patient’s Weight

Additional Patient’s

History

Tag

(0010,1010)

(0010,1020)

(0010,1030)

(0010,21b0)

Type Notes

3

3

3

3

Copied from original images

Copied from original images

Copied from original images

Not Sent.

8.4.3 Series Entity Module

8.4.3.1 General Series Module

Attribute Name

Series Number

Laterality

Series Date

Series Time

Modality

Performing Physician’s

Name

Protocol Name

Series Description

Tag

(0020,0011)

(0020,0060)

(0008,0021)

(0008,0031)

(0008,0060)

(0008,1050)

(0018,1030)

(0008,103E)

2C

3

3

1

3

Type Notes

2

Set to(original Series Number + 100000)

Always sent zero-length.

Not Sent.

Not Sent.

“PR”

Not Sent.

3

3

Not Sent.

Exam Split : This will be set as Requested

Procedure Code Meaning. If Code Meaning absent, it will be Requested Procedure

Description

127 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

Operators Name

Series Instance UID

Performed Procedure

Step Start Date

Performed Procedure

Step Start Time

Performed Procedure

Step ID

Performed Procedure

Step Description

Requested Attributes

Sequence

(0008,1070)

(0020,000E)

(0040,0244)

(0040,0245)

(0040,0253)

(0040,0254)

(0040,0275)

3

1

3

3

3

3

3

> Requested

Procedure ID

> Accession Number

> Referenced Study

Sequence

> Study Instance UID

(0040,1001)

(0008,0050)

(0008,1110)

(0020,000D)

> Requested

Procedure Desc

> Requested

Procedure Code

Sequence

> Scheduled Procedure

Step Desc

> Scheduled Protocol

Code Sequence

> Scheduled Procedure

Step ID

> Reason for

Requested Procedure

> Reason for

Requested Procedure

Code sequence

(0032,1060)

(0032,1064)

(0040,0007)

(0040,0008)

(0040,0009)

(0040,1002)

(0040,1004)

3

3

3

3

3

3

3

128 of 181

3

3

1C

3

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Save State : “Presentation Series”

Not sent

Generated new number for each series and always sent.

Exam Split : Calculated and sent.

Save State ; Not sent

Exam Split : Calculated and sent.

Save State : Not sent.

Exam Split : Will be set as Requested

Procedure ID (0040,1001) from (0040,0275)

Save State : Not sent.

Exam Split : This will be set as Requested

Procedure Code Meaning with a “PGP-“ prefix. If Code Meaning absent, it will be

Requested Procedure Description. The length will be truncated to 64 if necessary.

Save State : Not sent.

Exam Split : Sent when MPPS option is enabled and filled in based on what RIS provides to scanner. Number of items relates to number of items selected from

Patient Schedule.

Save State : Not sent.

Exam Split : Always sent, copied from RIS

Save State : Not sent.

Exam Split : Value from worklist sent

Save State : Not sent.

Exam Split : Value from worklist sent

Save State : Not sent.

Exam Split : Value from worklist sent

Save State : Not sent.

Exam Split : Value from worklist sent

Save State : Not sent.

Exam Split : Value from worklist sent.

Save State : Not sent.

Exam Split : Value from worklist sent

Save State : Not sent.

Exam Split : Value from worklist sent

Save State : Not sent.

Exam Split : Value from worklist sent

Save State : Not sent.

Exam Split : Value from worklist sent

Save State : Not sent.

Exam Split : Value from worklist sent

Save State : Not sent.

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Referenced Performed

Procedure

Step Sequence

>Referenced SOP Class

UID

>Referenced SOP

Instance UID

(0008,1111)

(0008,1150)

(0008,1155)

3

1C

1C

Exam Split : Value used for performed procedure step

Save State : Not Sent.

Copied from original images

Generated for each series and always sent.

Note: If scheduled procedures are grouped, the accession number for each procedure is stored in the Request

Attributes Sequence (0040, 0275) in the image header.

8.4.3.2 Presentation Series Module

Attribute Name Tag

Modality (0008,0060)

Type Notes

1

“PR“

8.4.4 Equipment Module

8.4.4.1 General Equipment

Attribute Name

Manufacturer

Institution Name

Station Name

Manufacturers Model

Name

Software Versions

Tag

(0008,0070)

(0008,0080)

(0008,1010)

(0008,1090)

(0018,1020)

Type Notes

2

3

Always sent as “GE MEDICAL SYSTEMS”

Copied from original images

3

3

Copied from original images

Copied from original images

3 Copied from original images

8.4.5 Presentation State Entity Module

8.4.5.1 Presentation State Identification Module

Attribute Name Tag Type Notes

(0070,0082) 1

Calculated and sent.

Presentation Creation

Date

Presentation Creation

Time

(0070,0083) 1

Calculated and sent.

Instance Number

Content Label

(0020,0013)

(0070,0080)

1

1

Set to 23 and sent.

Exam Split: This will be set as “PGP” plus

Requested Procedure Code Meaning or

Requested Procedure Description converted to all capital letters and truncated to 16 characters.

Save State : “GEGSPS_0”

129 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

Content Description (0070,0081) 2

2

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Exam Split : String input on Comments box in

Exam Split screen

Save State : “SE (Series Number) IM (First

Image Number) TO IM (Last Image Number)”

Same as Station Name(0008,1010) Content Creator’s

Name

(0070,0084)

8.4.5.2 Presentation State Relationship

Attribute Name Tag

(0008,1115) Referenced Series

Sequence

> Referenced Image

Sequence

>> Reference SOP

Class UID

>> Reference SOP

Instance UID

> Series Instance UID

(0008,1140)

(0008,1150)

(0008,1155)

(0020,000E)

8.4.5.3 Presentation State Shutter

Attribute Name

Shutter Presentation

Value

Tag

(

0018,1622

)

8.4.5.4 Displayed Area

Attribute Name

Displayed Area

Selection Sequence

Tag

(0070,005A)

>Referenced Image

Sequence

>Displayed Area Top

Left Hand Corner

>Displayed Area

Bottom Right Hand

Corner

>Presentation Size

Mode

>Presentation Pixel

Spacing

(0008,1140)

(0070,0052)

(0070,0053)

(0070,0100)

(0070,0101)

1

1

1

Type Notes

1 Calculated and sent.

1 Always sent, references images selected.

Always sent

Always sent

Always sent, references series selected

Type Notes

1 Exam Split : Not sent.

Save State : Always sent.

Type Notes

1

1C

Exam Split : calculated values for pan and zoom being displayed.

Save State : the number of images in the original series

Always sent.

1

1

Always sent.

Always sent.

1

1C

Always sent.

Always sent.

130 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

>Presentation Pixel

Magnification Ratio

8.4.5.8 SOP Common Module

Attribute Name

SOP Class UID

SOP Instance UID

Specific Character Set

Instance Creation Date

Instance Creation Time

Instance Number

(0070,0103) 1C

8.4.5.5 Modality LUT Module

Attribute Name

Rescale slope intercept

Rescale slope

Rescale type

8.4.5.6 Softcopy VOI LUT Module

Attribute Name

Softcopy VOI LUT Sequence

> Reference Image Sequence

> Window Center

> Window Width

> Window Center & Width

Explanation

Tag

(0028,1052)

(0028,1053)

(0028,1054)

Tag

(0028,3110)

(0008,1140)

(0028,1050)

(0028,1051)

(0028,1055)

8.4.5.7 Softcopy Presentation LUT Module

Attribute Name

Presentation LUT sequence

Tag

(2050,0010)

Presentation LUT shape (2050,0020)

Tag

(0008,0016)

(0008,0018)

(0008,0005)

(0008,0012)

(0008,0013)

(0020,0013)

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Always sent.

Type

1C

1C

1C

Type

1C

1C

Type

1

1

1C

3

3

1

Type

1

1C

1C

1C

3

Notes

Always sent

Always sent

Always sent as “HU”

Notes

Always sent

Always sent

Always sent

Always sent

Always sent

Notes

Not sent, condition not met.

Always sent as “IDENTITY”

Notes

Always sent.

Always sent.

“ISO_IR 100“

Always sent.

Always sent.

Set to 23 and sent.

8.5 Image Header Changes Supporting GSPS

8.5.1 Request Attributes Sequence

131 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

The (0040, 0275) Request Attributes Sequence has been expanded to include DICOM tags that the Virtual Exam Split (VES) application requires to complete the VES GSPS and PPS.

The Requested Attributes Sequence may contain a maximum of 15 sequence items.

This sequence will only appear in the image header if the VES/HES option is installed or PPS is enabled..

132 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

9 Structured Report Object Implementation

GEHC CT system supports 2 types of SRs, Enhanced SR and X-ray Radiation Dose SR. Both of them support following attributes.

9.1 IOD Module Table

Entity Name

Patient

Study

Series

Equipment

Document

9.2 SR Document Series Module

Attribute Name

Modality

Series Instance UID

Series Number

Referenced Performed Procedure

Step Sequence

Module Name

Patient

General Study

Patient Study

SR Document Series

General Equipment

SR Document General

SR Document Content

SOP Common

Tag

(0008,0060)

(0020,000E)

(0020,0011)

(0008,1111)

Reference

A.3.1

A.3.2

A.3.3

9.2

A.3.5

9.3

9.4

A.3.8

Type

1

1

1

2

Notes

Sent as “SR“

Sent.

Sent as “997”

Sent.

9.3 SR Document General Module

Attribute Name

Instance Number

Completion Flag

Tag

(0020,0013)

(0040,A491)

Completion Flag Description

Verification Flag

(0040,A492)

(0040,A493)

Content Date

Content Time

Verifying Observer Sequence

>Verifying Observer Name

>Verifying Observer

Identification Code Sequence

(0008,0023)

(0008,0033)

(0040,A073)

(0040,A075)

(0040,A088)

133 of 181

Type Notes

1 Sent

1 COMPLETE sent

3 Sent empty

1

UNVERIFIED sent

1 Sent

1 Sent

1C

Not sent per conditional

1 Not sent

2 Not sent

Usage

M

M

U

M

M

M

M

M

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

>>Include 'Code Sequence Macro' Table 8.8-1

>Verifying Organization

>Verification DateTime

(0040,A027)

(0040,A030)

Not sent

1 Not sent

1 Not sent

Author Observer Sequence (0040,A078) 3 Not sent

>Include 'Identified Person or Device Macro' Table C.17-3b

Participant Sequence

>Participation Type

(0040,A07A)

(0040,A080)

3 Not sent

1

Not sent

>Participation DateTime (0040,A082) 2 Not sent

>Include 'Identified Person or Device Macro' Table C.17-3b

Custodial Organization

Sequence

(0040,A07C) 3 Not sent

>Institution Name

>Institution Code Sequence

>>Include 'Code Sequence Macro' Table 8.8-1

Predecessor Documents

Sequence

(0008,0080)

(0008,0082)

(0040,A360)

2 Not sent

2 Not sent

1C

Not sent per condition

>Include ' Hierarchical SOP Instance Reference Macro' Table C.17-3

Identical Documents

Sequence

(0040,A525) 1C Not sent per condition

>Include ' Hierarchical SOP Instance Reference Macro' Table C.17-3

Referenced Request

Sequence

>Study Instance UID

(0040,A370)

(0020,000D)

1C

1

Not sent per condition

Not Sent

>Referenced Study Sequence (0008,1110) 2 Not Sent

>> Include ‘SOP Instance Reference Macro' Table 10-11

>Accession Number (0008,0050)

>Placer Order

Number/Imaging Service

Request

>Filler Order Number/Imaging

Service Request

>Requested Procedure ID

(0040,2016)

(0040,2017)

(0040,1001)

>Requested Procedure

Description

>Requested Procedure Code

Sequence

(0032,1060)

(0032,1064)

>>Include 'Code Sequence Macro' Table 8.8-1

2 Not Sent

2

2

2

2

2

Not Sent

Not Sent

Not Sent

Not Sent

Not Sent

No Baseline Context ID Number is specified.

(0040,1002) 3 Not Sent >Reason for the Requested

Procedure

>Reason for Requested

Procedure Code Sequence

(0040,100A) 3 Not Sent

134 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

>>Include 'Code Sequence Macro' Table 8.8-1

Performed Procedure Code

Sequence

(0040,A372)

>Include 'Code Sequence Macro' Table 8.8-1

Current Requested Procedure

Evidence Sequence

(0040,A375)

No Baseline Context ID Number is specified.

2

Sent empty

No Baseline Context ID Number is specified.

1C Not Sent

>Include ' Hierarchical SOP Instance Reference Macro' Table C.17-3

Pertinent Other Evidence

Sequence

(0040,A385) 1C Not sent

>Include ' Hierarchical SOP Instance Reference Macro' Table C.17-3

Referenced Instance

Sequence

(0008,114A) 1C Not sent

>Include ‘SOP Instance Reference Macro' Table 10-11

>Purpose of Reference Code

Sequence

(0040,A170) 1 Not sent

>>Include 'Code Sequence Macro' Table 8.8-1

9.4 SR Document Content Module

The tables below capture notes concerning the values stored.

9.4.1 TID 10011 – CT Radiation Dose

NL Rel with

Parent

1

2 > HAS

CONCEPT

MOD

VT

CONTA

INER

Concept Name VM Req

Type

EV (113701,

DCM, “X-Ray

Radiation Dose

Report”)

CODE EV (121058,

DCM, ”Procedure reported”)

1

1

M

M

Notes

Sent

Code sequence (P5-08000, SRT,

Computed Tomography X-Ray) sent

3 >> HAS

CONCEPT

MOD

4 >

CODE EV (G-C0E8, SRT,

“Has Intent”)

INCLU

DE

DTID (1002)

Observer

Context

1 M Code sequence (R-408C3, SRT,

Diagnostic Intent) sent

1-n M See TID 1002 Observer context

(sent with Device context values equal to Implementation UID.

Device Role in Procedure code sequences (113859, DCM,

Irradiating Device) and (121097,

DCM, Recording) sent. Single item sent.)

135 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

5 >

6 >

7 >

9 >

HAS OBS

CONTEXT

8 >> HAS

PROPERTIES

CODE EV (113705,

DCM, “Scope of

Accumulation”)

UIDRE

F

CONTAINS INCLU

DE

DCID (10001) UID

Types

DTID (10012) CT

Accumulated

Dose Data

10 >

11 >

CONTAINS INCLU

DE

DTID (10013) CT

Irradiation Event

Data

CONTAINS TEXT EV (121106,

DCM,

“Comment”)

12 >

13 >

HAS OBS

CONTEXT

HAS OBS

CONTEXT

DATETI

ME

EV (113809,

DCM, “Start of X-

Ray Irradiation”)

DATETI

ME

EV (113810,

DCM, “End of X-

Ray Irradiation”)

CONTAINS CODE EV (113854,

DCM,

“Source of Dose

Information”)

CONTAINS INCLU

DE

DTID (1020)

Person

Participant

1

1

1

1

M

M

M

Sent

Sent

Study Instance UID sent

1 M See TID 10012 CT Accumulated

Dose Data table below for details

1-n M See TID 10013 CT Irradiation

Event Data table below for details

1 U Not sent

1-n

1

M

M

U

Code sequence (113014, DCM,

Study) sent

Code sequence (113856, DCM,

Automated Data Collection) sent

Not sent.

9.4.2 TID 10012 – CT Accumulated Dose Data

NL Rel with

Parent

VT Concept Name VM Req

Type

1

Notes

M Sent 1 CONTA

INER

EV (113811,

DCM, ”CT

Accumulated

Dose Data”)

2 > CONTAINS NUM EV (113812,

DCM, “Total

Number of

Irradiation

Events”)

3 > CONTAINS NUM EV (113813,

DCM, “CT Dose

Length Product

Total”)

1

1

M Sent

M Sent

136 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

4 > CONTAINS NUM EV (113814,

DCM, “CT

Effective Dose

Total”)

5 >> HAS

PROPERTIES

TEXT EV (121406,DCM,

“Reference

Authority”)

6 >> HAS

PROPERTIES

CODE EV (121406,DCM,

“Reference

Authority”)

7 >> HAS

CONCEPT

MOD

CODE EV (G-C036,SRT,

“Measurement

Method”)

8 >> HAS

PROPERTIES

TEXT EV (113815,DCM,

“Patient Model”)

9 >> HAS

PROPERTIES

CONTA

INER

EV (113816,

DCM, “Condition

Effective Dose measured”)

10 >>> CONTAINS TEXT EV (113817,DCM,

“Effective Dose

Phantom Type”)

11 >>> CONTAINS TEXT EV (113818,

DCM, “Dosimeter

Type”)

12 > CONTAINS TEXT EV (121106,

DCM,

“Comment”)

13 > CONTAINS INCLU

DE

DTID(1021)

Device

Participant

1

1

1

1

1

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

U Not sent

1 MC Not sent

1 MC Not sent

M

1 MC Not sent

1 MC Not sent

M

M

U

Not sent

Not sent

Not sent

Not sent

1 MC Not sent

9.4.3 TID 10013 – CT Irradiation Event Data

NL Rel with

Parent

VT Concept Name VM Req

Type

Notes

1 M Sent 1 CONTA

INER

EV (113819, DCM,

”CT Acquisition”)

2 > CONTAINS TEXT EV (125203, DCM,

“Acquisition

Protocol”)

3 > CONTAINS CODE EV (123014 ,

DCM, ”Target

Region”)

1 U Not sent

1 M Sent. See note below for specifics.

137 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

4 > CONTAINS CODE EV (113820, DCM,

“CT Acquisition

Type”)

5 > CONTAINS CODE (G-C32C, SRT,

“Procedure

Context”)

6 > CONTAINS UIDRE

F

EV (113769, DCM,

“Irradiation Event

UID”)

1

1

M

U

Sent

Not sent

1 M Sent; all images from this irradiation event contain same

Irradiation Event UID (0x0008,

0x3010)

1 M Sent 7 > CONTAINS CONTA

INER

EV (113822, DCM,

“CT Acquisition

Parameters”)

8 >> CONTAINS NUM EV (113824, DCM,

“Exposure Time”)

9 >> CONTAINS INCLU

DE

DTID(10014)

Scanning Length

10 >> CONTAINS NUM EV (113826, DCM,

“Nominal Single

Collimation

Width”)

11 >> CONTAINS NUM EV (113827, DCM,

“Nominal Total

Collimation

Width”)

12 >> CONTAINS NUM EV (113828, DCM,

“Pitch Factor”)

13 >> CONTAINS NUM EV (113823, DCM,

“Number of X-

Ray Sources”)

1

1

1

1

1

1

M

M

M

M

See TID 10014 Scanning Length

table below for details

MC Sent per conditional

M

Sent

Sent

Sent

Sent

14 >> CONTAINS CONTA

INER

EV (113831, DCM,

“CT X-Ray Source

Parameters”)

15 >>> CONTAINS TEXT EV (113832, DCM,

“Identification of the X-Ray

Source”)

1-n M Single item sent.

1 M Sent

1 M Sent 16 >>> CONTAINS NUM EV (113733, DCM,

“KVP”)

17 >>> CONTAINS NUM EV (113833, DCM,

“Maximum X-Ray

Tube Current”)

18 >>> CONTAINS NUM EV (113734, DCM,

“X-Ray Tube

Current”

1

1

M

M

Sent

Sent

138 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

19 >>> CONTAINS NUM EV (113834, DCM,

“Exposure Time per Rotation”)

20 >>> CONTAINS NUM EV (113821, DCM,

“X-Ray Filter

Aluminum

Equivalent”)

21 > CONTAINS CONTA

INER

EV (113829, DCM,

“CT Dose”)

22 >> CONTAINS NUM EV (113830, DCM,

“Mean CTDIvol ”)

23 >> CONTAINS CODE EV (113835, DCM,

“CTDIw Phantom

Type”)

1

1

1

1

MC Sent per conditional

U

MC Sent per conditional

M

Not sent

Sent

1 M Sent (IEC Head Dosimetry

Phantom or IEC Body

Dosimetry Phantom)

1 U Not sent 24 >> CONTAINS NUM EV (113836, DCM,

“CTDIfreeair

Calculation

Factor”)

25 >> CONTAINS NUM EV (113837, DCM,

“Mean

CTDIfreeair”)

26 >> CONTAINS NUM EV (113838, DCM,

“DLP”)

27 >> CONTAINS NUM EV (113839, DCM,

“Effective Dose”)

28 >>> HAS

CONCEPT

MOD

CODE EV (G-C036, SRT,

“Measurement

Method“)

29 >>>> HAS

PROPERTIES

NUM EV (113840, DCM,

“Effective Dose

Conversion

Factor”)

1

1

1

1

1

U

M

U

Not sent

Sent

Not sent

MC Not sent

MC Not sent

30 >> CONTAINS INCLU

DE

DTID(10015) CT

Dose Check

Details

31 > CONTAINS TEXT EV (113842, DCM,

“X-Ray

Modulation

Type”)

1

1

M

U

See TID 10015 CT Dose Check

Details table below for details

Not sent

32 > CONTAINS TEXT EV (121106, DCM,

“Comment”)

33 >

34 >

CONTAINS INCLU

DE

CONTAINS INCLU

DE

DTID (1020)

Person

Participant

DTID (1021)

Device

1

1-n U Not sent

1

U Not sent

MC Not sent.

139 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Participant

9.4.4 TID 10014 - Scanning Length

NL Rel with

Parent

1

2

3

4

5

6

7

8

VT

NUM

NUM

Concept Name VM Req

Type

EV(113825, DCM,

“Scanning

Length”)

1 M

EV(113893, DCM,

“Length of

Reconstructable

Volume”)

1 U

Notes

Sent

Sent, except not sent for Scout

1 UC Sent per conditional NUM

NUM

NUM

NUM

NUM

EV(113899, DCM,

“Exposed

Range”)

EV(113895, DCM,

“Top Z Location of

Reconstructable

Volume”)

EV(113896, DCM,

“Bottom Z

Location of

Reconstructable

Volume”)

EV(113897, DCM,

“Top Z Location of Scanning

Length”)

EV(113898, DCM.

“Bottom Z

Location of

Scanning

Length”)

UIDREF EV(112227, DCM,

“Frame of

Reference UID”)

1

1

1

1

1

U

U

U

U

MC

Sent, except not sent for Scout

Sent, except not sent for Scout

Sent

Sent

Sent

9.4.5 TID 10015 – CT Dose Check Details

1

NL Rel with

Parent

VT

CONTAINER

Concept Name

2 > CONTAINS CODE

EV(113900, DCM,

“Dose Check Alert

Details”)

EV(113901, DCM,

“DLP Alert Value

Configured”)

VM Req

Type

Notes

1 MC Sent

1 M Sent

140 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

3 > CONTAINS CODE

4

5

6

7

8

9

10

>

>

>

>

>

>

11 >

12 >

13 >

14 >

15 >

16 >

17 >

CONTAINS

CONTAINS

CONTAINS

CONTAINS

CONTAINS

CONTAINS

CONTAINS

CONTAINS

CONTAINS

CONTAINS

CONTAINS

CONTAINS

CONTAINS

NUM

NUM

NUM

NUM

TEXT

INCLUDE

CONTAINER

CODE

CODE

NUM

NUM

NUM

NUM

TEXT

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

EV(113908, DCM,

“Dose Check

Notification

Details”)

EV(113909, DCM,

“DLP Notification

Value Configured”)

EV(113910, DCM,

“CTDIvol

Notification Value

Configured”)

EV(113911, DCM,

“DLP Notification

Value”)

EV(113912, DCM,

“CTDIvol

Notification

Value”)

EV(113913, DCM,

“DLP Forward

Estimate”)

EV(113914, DCM,

“CTDIvol Forward

Estimate”)

EV(113907, DCM,

“Reason for

Proceeding”)

EV(113902, DCM,

“CTDIvol Alert

Value Configured”)

EV(113903, DCM,

“DLP Alert Value”)

EV(113904, DCM,

“CTDIvol Alert

Value”)

EV(113905, DCM,

“Accumulated DLP

Forward

Estimate”)

EV(113906, DCM,

“Accumulated

CTDIvol Forward

Estimate”)

EV(113907, DCM,

“Reason for

Proceeding”)

DTID(1020) Person

Participant

1 M

1 UC

Sent

1 MC Sent per conditional

1 MC Sent per conditional

1 MC Sent per conditional

1 MC Sent per conditional

Sent per conditional

1 MC Sent per conditional.

See TID 1020 Person

Participant table below for details

1 MC Sent

1

1

1

1

1

1

1

M

M

MC

MC

MC

MC

UC

Sent

Sent

Sent per conditional

Sent per conditional

Sent per conditional

Sent per conditional

Sent per conditional

141 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

18 > CONTAINS INCLUDE DTID(1020) Person

Participant

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

1 UC Not sent

9.4.6 TID 1002 – Observer Context

NL Rel with

Parent

VT

1

2

3

HAS OBS

CONTEXT

HAS OBS

CONTEXT

HAS OBS

CONTEXT

Concept Name

CODE EV (121005,DCM,

“Observer Type”)

INCLUDE DTID (1003) Person observer identifying attributes

INCLUDE DTID (1004) Device observer identifying attributes

VM Req

Type

Note

1 MC Sent as “Device“.

1 MC Not sent.

1 MC See TID 1004 Device

Observer Identifying

Attributes table below for details.

9.4.7 TID 1004 – Device Observer Identifying Attributes

NL Rel with

Parent

VT Concept Name VM Req

Type

1

2

3

4

5

6

7

Note

UIDREF

TEXT

TEXT

TEXT

TEXT

TEXT

CODE

EV (121012,DCM, “Device

Observer UID”)

EV (121013,DCM, “Device

Observer Name”)

EV (121014,DCM, “Device

Observer Manufacturer”)

EV (121015,DCM, “Device

Observer Model Name”)

1 M Sent. Same as

Implementation UID

(0002,0012)

See Appendix.D.

1

1

U Sent. Same as Station

Name (0008,1010)

U Sent as “GE Medical

Systems”.

EV (121016,DCM, “Device

Observer Serial Number”)

EV (121017,DCM, “Device

Observer Physical

Location during observation”)

EV (113876, DCM,

“Device Role in

Procedure”)

1 U Sent. Same as

Manufacturer’s Model

Name (0008,1090)

1

1

U Sent. Same as Device

Serial Number (0018,1000)

U Not Sent.

1-n U Sent as “Irradiating

Device” and “Recording”.

9.4.8 TID 1020 – Person Participant

142 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

NL Rel with

Parent

VT Concept Name VM Req

Type

Note

1

2 > HAS

PROPERTIES

3 > HAS

PROPERTIES

4 > HAS

PROPERTIES

5 > HAS

PROPERTIES

PNAME

CODE

TEXT

TEXT

TEXT

EV (113870,DCM,

“Person Name”)

EV (113875,DCM,

“Person Role in

Procedure”)

EV (113871,DCM,

”Person ID”)

EV (113872,DCM,

”Person ID Issuer”)

EV (113873,DCM,

“Organization Name”)

1 M Sent

Person’s name is free text, not PNAME format.

1 M Sent as “Irradiation

Authorizing”.

1 U Sent.

1 U Not Sent.

1 U Not Sent.

6 > HAS

PROPERTIES

CODE EV (113874,DCM,

“Person Role in

Organization”)

1 U Not Sent.

Note: Target region is filled in per the following table:

Protocol Category Value stored in Target Region

Head

Orbit

Neck

Upper Extremity

Chest

Abdomen

Spine

Pelvis

SRT T-D1100 Head

SRT T-D0801 Eye region

SRT T-D1600 Neck

SRT T-02220 Shoulder

SRT T-D3000 Chest

SRT T-D4000 Abdomen

SRT T-11503 Lumbar spine

SRT T-D6000 Pelvis

Lower Extremity SRT T-D0300 Extremity

Note: CT X-Ray Radiation Dose SR Record adheres to the DICOM 2011 edition.

9.5 Configuration

The product is configured by default to not generate the CT X-Ray Radiation Dose SR Record.

Creation of the CT Dose Record can be enabled via reconfig.

The product implements the CT Radiation Dose Report template (TID 10011) per DICOM Supplement

127.

The DICOM standard says that the SOP class shall be the “X-Ray Radiation Dose SR” SOP Class

(1.2.840.10008.5.1.4.1.1.88.67).

However, since many products may not support this SOP class, the product supports the option via reconfig to create the CT Radiation Dose Report as an “Enhanced SR” SOP Class

(1.2.840.10008.5.1.4.1.1.88.22).

143 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

SR setting can be configured in “Dose Report” in reconfig menu as below.

“Full” will produce the X-Ray Radiation Dose SR SOP class

“On” will produce the Enhanced SR SOP class.

“Off” will not produce any SOP class.

144 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

10 Security Conformance Statement

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

10.1 De-identification

The GEHC CT system satisfies the following requirements for de-identification

De-identification occurs, for example, when an anonymous patient operation is performed. This deidentification only covers DICOM attributes that contain protected patient information. The encrypted data capability is not supported, so once an image has been de-identified no facility is provided to recover the lost information.

The de-identification specified in this section only applies to GE generated images. All third party images have only the standard DICOM elements de-identified, all private tags (non-GE private tags) within these images are dropped.

De-identification supports the following object types:

· CT Image Objects

De-identification does not support the following object types:

· Secondary capture images

· Structured reports

· Grayscale presentation state objects

This de-identification specification applies when the GEHC CT system is configured for full deidentification. If the system is configured for partial de-identification, some attributes will remain unchanged.

Table 10.1-1 provides the list of attributes and the expected action when de-identifying images.

Table 10.1-1

DICOM Attributes De-Identified

DICOM Tag

0008,0014

0008,0018

0008,0050

0008,0080

0008,0081

0008,0090

0008,0092

0008,0094

0008,1010

0008,1030

0008,103E

0008,1040

0008,1048

0008,1050

Attribute Name

Instance Creator UID

Action

element dropped

SOP Instance UID

Accession Number

Institution Name

Institution Address assign new UID made zero length made zero length element dropped

Referring Physician’s Name made zero length

Referring Physician’s Address element dropped

Referring Physician’s Telephone

Numbers element dropped

Station Name

Study Description

“ANONYMIZED”

“ANONYMIZED”

Series Description “ANONYMIZED”

Institutional Department Name element dropped

Physicians of Record element dropped

Performing Physicians’ Name element dropped

Comments

Refer to Note 1

Refer to Note 1

Refer to Note 1

Refer to Note 1

Refer to Note 1

Refer to Note 1

Refer to Note 1

145 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

0010,0010

0010,0020

0010,0030

0010,0032

0010,0040

0010,1000

0010,1001

0010,1010

0010,1020

0010,1030

Other Patient Names

Patient’s Age

Patient’s Size

Patient’s Weight

0010,1090

0010,2160

Medical Record Locator

Ethnic Group

0010,2180 Occupation

0010,21B0 Additional Patient’s History

0010,4000

0018,1030

Patient Comments

Protocol Name

0020,000D Study Instance UID

0020,000E

0020,0010

0020,0052

Patient’s Name

Patient ID

Patient’s Birth Date

Patient’s Birth Time

Patient’s Sex

Other Patient Ids

Series Instance UID

Study ID

Frame of Reference UID

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

DICOM Tag

0008,1060

Attribute Name

Name of Physicians Reading

Study

Action

made zero length

0008,1070

0008,1080

Operators’ Name

Admitting Diagnoses

Description made zero length element dropped

0008,1110 Referenced Study Sequence

>0008,1150 Referenced SOP Class UID drop sequence element dropped

>0008,1155 Referenced SOP Instance UID element dropped

0008,1140 Referenced Image Sequence drop sequence

>0008,1150 Referenced SOP Class UID element dropped

>0008,1155 Referenced SOP Instance UID element dropped

“ANONxxxx”

“ANONxxxx” made zero length element dropped made zero length made zero length element dropped made zero length made zero length made zero length element dropped element dropped element dropped made zero length element dropped made zero length assign new UID assign new UID retain Study ID assign new UID element dropped

Comments

Refer to Note 1

Special handling is performed when the entire exam in deidentified in one operation. See Table

10.1-2 for details. xxxx = study id xxxx = study id

Refer to Note 1

Refer to Note 1

Refer to Note 1

Refer to Note 1

Refer to Note 1

Refer to Note 1

This value is used for all images in exam

This value is used for all images in series

This value is used in patient ID

This value is used for all images sharing this UID

Refer to Note 1 0020,0200 Synchronization Frame of

Reference UID

0040,A124 UID

0040,A730 Content Sequence

0088,0140

3006,0024

Storage Media File-set UID

Referenced Frame of Reference

UID

3006,00C2 Related Frame of Reference

UID element dropped drop sequence element dropped element dropped element dropped

Refer to Note 1

Refer to Note 1

Refer to Note 1

Refer to Note 1

Refer to Note 1

146 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Note 1: these elements are not currently present in GE images, however, the described action will occur on 3 rd

party images.

The following special processing is performed based on image type:

Table 10.1-2

De-identification Processing by Image Type

Image Type

CT Image

SC Image

GSPS

SR

Processing

de-identified based on the tables above image discarded object discarded object discarded

10.2 ASSOCIATION LEVEL SECURITY

The GEHC CT system provides association level security for Query Retrieve operations.

The system will only allow remote DICOM servers to pull images if the remote DICOM server is defined on the GEHC CT system. The remote DICOM server’s IP address, AE title, and Port number must be configured on the GEHC CT system for the networking pull to begin.

Note that this association level security only applies to the retrieve request; query operations will be accepted from any connected DICOM client.

147 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

APPENDIX A: CT Image and Secondary Capture Modules/Attributes

The tables below specify the attributes of the CT Image and Secondary Capture Image transmitted by the Storage SCU AE.

A.1 CT Image IOD

A.1.1 CT Image IOD Modules

Entity Name

Patient

Study

Series

Frame of Reference

Equipment

Image

Module Name

Patient

General Study

Patient Study

General Series

Frame of Reference

General Equipment

General Image

Image Plane

Image Pixel

Contrast/Bolus

CT Image

VOI LUT

SOP Common

A.4.5

A.3.5

A.3.6

A.4.1

A.4.2

A.4.3

A.4.4

A.3.7

A.3.8

Reference

A.3.1

A.3.2

A.3.3

A.3.4

M

C

M

U

M

M

M

M

M

Usage

M

M

U

M

148 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

A.1.1.1 Implementation Specific details

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

For all CT Images created by GEHC CT patient birth date element (0010,0030) is sent, if the operator has entered the details. If the operator has not entered the details it will be sent as a

“Zero length element”

The length of Accession number element (0008,0050) is a maximum of 16 characters

Window Center (0028,1050) and Window Width (0028,1051) are sent for all CT Images created by GEHC CT

Based on whether contrast was used or not the following applies for CT Images created by

GEHC CT

Mode

No Contrast

Oral Contrast

IV Contrast

Oral and IV contrast

(0018,0010)

Not sent

Oral Contrast Agent name

IV Contrast agent name

Oral contrast agent “ & “

IV Contrast agent

(0018,1040)

Not sent

The string “Oral”

The String “IV”

The string “Oral & IV”

Supported field length for DICOM fields

DICOM Tag Field Name

( 0008, 0050 ) Accession Number

( 0010, 0010 ) Patient Name

( 0010, 0020 ) Patient Id

Field Length

16 characters

32 characters

(configurable to 64)

16 characters

(configurable to 64)

A.2 SC Image IOD

A.2.1 SC Image IOD Modules

Entity Name

Patient

Study

Series

Equipment

Image

Module Name

Patient

General Study

Patient Study

General Series

General Equipment

SC Equipment

General Image

149 of 181

Reference

A.3.1

A.3.2

A.3.3

A.3.4

A.3.5

A.5.1

A.3.6

Usage

M

M

U

M

U

M

M

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Image Pixel

SC Image

Modality LUT

VOI LUT

SOP Common

A.5.2

Not sent (consists entirely of type

3 element).

A.5.3

A.3.7

A.3.8

M

M

U

U

M

A.2.2 CT Dose Report SC Image Details

The product is configured by default to generate a CT X-Ray Radiation Dose Report SC Image summarizing the study dose. The pixel data contains a textual, viewable report of the dose information. The next section identifies the additional standard DICOM attributes added to capture the same information as reported in the CT X-Ray Radiation Dose Record, described in Section 9.

150 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

A.2.2.1 Implementation Specific details

Attribute Name

Total Number of

Exposures

Exposure dose sequence

>kVp

>X-Ray Tube Current

>Exposure Time

>Acquisition Type

>CTDIvol

>CTDI Phantom Type

Code Seq

>Single Collimation

Width

>Total Collimation

>Spiral Pitch Factor

>Body Part Examined

Comments on

Radiation Dose

Tag

(0040,0301)

(0040,030E)

(0018,0060)

(0018,8151)

(0018,1150)

(0018,9302)

(0018,9345)

(0018,9346)

(0018,9306)

(0018,9307)

(0018,9311)

(0018,0015)

(0040,0310)

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Type Notes

3 Total number of exposures made during this

Performed Procedure Step.

3 Exposure Dose Sequence will contain Total

Number of Exposures (0040,0301) items.

3

3

3

3

3

Peak kilo voltage output

X-Ray tube current in microA. NOTE: the X-

Ray Radiation Dose SR records this in mA.

Time of x-ray exposure in msec.

3

3

Computed Tomography Dose Index in mGy according to IEC 60601-2-44.

The type of phantom used for CTDI measurement according to IEC 60601-2-44.

The width of a single row of acquired data in mm.

3

3

3

3

The width of the total collimation in mm over the area of active x-ray detection.

Ratio of the Table Feed per Rotation

(0092,9310) to the Total Collimation Width

(0018,9307)

Text description of the part of the body examined. See the values for target region at the end of section 9.4.3 for the values stored.

Used to record total and per exposure DLP.

The format is:

TotalDLP=xxx

DLP=xxx

DLP=xxx

151 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

A.3 Common Modules

A.3.1 Patient Module

Attribute Name

Patient’s Name

Tag

(0010,0010)

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Patient ID

Patient’s Birth Date

Patient’s Sex

Referenced Patient

Sequence

>Referenced SOP

Class UID

>Referenced SOP

Instance UID

Other Patient ID

(0010,0030)

(0010,0040)

(0008,1120)

(0008,1150)

(0008,1155)

(0010,1000)

A.3.2 General Study Module

Attribute Name

Study Instance UID

Tag

(0020,000D)

Study Date

Study Time

Accession Number

Referring Physician’s

Name

(0010,0020)

(0008,0020)

(0008,0030)

(0008,0050)

(0008,0090)

Type Notes

2 As entered at user interface or from worklist.

Supports 5 different components delimited by

“^”. Supports a maximum length of 32

characters including the delimiter. Can be configured to allow 64 characters.

2

2

2

3

As entered at user interface or from worklist.

Supports maximum of 16 characters. Can be configured to allow 64 characters.

As entered at user interface or from worklist.

As entered at user interface or from worklist.

Value from the worklist

1C Value from the worklist

1C

3

Value from the worklist

Value from the worklist

Type Notes

1 Value from worklist, if present, is used.

Otherwise, the scanner creates a unique value for each exam.

If worklist is reused, system can be configured to reuse the study instance uid.

Default behavior is to generate a new study instance uid if the worklist is reused.

2

2

2

2

Generated for each exam and always sent.

Generated for each exam and always sent.

Value from user interface or worklist sent.

If worklists with different accession numbers are grouped, the default behavior is to fill in the top level accession number in the image header. The accession number can be configured to be empty in the grouped case.

Value from user interface sent. Accepts 32 characters by default but is configurable to accept 64 characters.

152 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

Attribute Name

Study ID

Study Description

Tag

(0020,0010)

(0008,1030)

Name of Physician(s)

Reading Study

Referenced Study

Sequence

>Referenced SOP Class

UID

>Referenced SOP

Instance UID

(0008,1060)

(0008,1110)

(0008,1150)

(0008,1155)

A.3.3 Patient Study Module

Attribute Name

Patient’s Age

Patient’s Size

Patient’s Weight

Additional Patient’s

History

Tag

(0010,1010)

(0010,1020)

(0010,1030)

(0010,21b0)

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Type Notes

2 Generated for each exam on the scanner and always sent..

3

3

Value from user interface or worklist sent.

Accepts 22 characters by default but is configurable to accept 64 characters.

Sent if entered at the user interface.

3 Value from the worklist

1C

1C

Value from the worklist

Value from the worklist

Type Notes

3 Calculated from Date of Birth entered at user interface.

3

3

3

Value from user interface or worklist sent

Value from user interface or worklist sent.

Value from user interface or worklist sent

A.3.4 General Series Module

Attribute Name

Series Number

Laterality

Series Date

Series Time

Modality

Performing Physician’s

Name

Protocol Name

Series Description

Operator’s Name

Patient Position

Tag

(0020,0011)

(0020,0060)

(0008,0021)

(0008,0031)

(0008,0060)

(0008,1050)

(0018,1030)

(0008,103E)

(0008,1070)

(0018,5100)

Type Notes

2 Generated sequentially, always sent.

2C

3

Always sent zero-length.

Generated for each series and always sent.

3

1

3

Generated for each series and always sent.

Always sent as “CT”

Mapped from (0040, 0006) in the worklist.

3

3

3

2C

Sent if entered at user interface.

Value from user interface is sent.

Value from user interface is sent.

Sent. As selected by operator when patient is positioned. Defined terms are:

HFP = Head-First Prone

HFS = Head-First Supine

HFDR = Head-First Decubitus Right

HFDL = Head-First Decubitus Left

153 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

Series Instance UID

Performed Procedure

Step Start Date

Performed Procedure

Step Start Time

Performed Procedure

Step ID

Performed Procedure

Step Description

Requested Attributes

Sequence

(0020,000E)

(0040,0244)

(0040,0245)

(0040,0253)

(0040,0254)

(0040,0275)

1

3

3

3

3

3

> Requested

Procedure ID

> Accession Number

> Referenced Study

Sequence

> Study Instance UID

> Requested

Procedure Desc

> Requested

Procedure Code

Sequence

> Scheduled Procedure

Step Desc

> Scheduled Protocol

Code Sequence

> Scheduled Procedure

Step ID

> Reason for

Requested Procedure

> Reason for

Requested Procedure

Code sequence

Body Part Examined

Referenced Performed

Procedure

Step Sequence

>Referenced SOP Class

UID

>Referenced SOP

Instance UID

(0040,1001)

(0008,0050)

(0008,1110)

(0020,000D)

(0032,1060)

(0032,1064)

(0040,0007)

(0040,0008)

(0040,0009)

(0040,1002)

(0040,1004)

(0018,0015)

(0008,1111)

(0008,1150)

(0008,1155)

3

3

1C

1C

154 of 181

1C

3

3

3

3

3

3

3

3

3

3

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

FFDR = Feet-First Decubitus Right

FFDL = Feet-First Decubitus Left

FFP = Feet-First Prone

FFS = Feet-First Supine

Always sent

Sent when MPPS option is enabled

Sent when MPPS option is enabled

Sent when MPPS option is enabled

Sent when MPPS option is enabled

Sent when MPPS option is enabled and filled in based on what RIS provides to scanner.

Number of items relates to number of items selected from Patient Schedule.

Always sent, copied from RIS

Value from worklist sent

Value from worklist sent

Value from worklist sent

Value from worklist sent

Value from worklist sent

Value from worklist sent

Value from worklist sent

Value from worklist sent

Value from worklist sent

Value from worklist sent

Only sent for GSI CT images.

Value used for performed procedure step

Value used for performed procedure step

Value used for performed procedure step

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

A.3.5 General Equipment Module

Attribute Name

Manufacturer

Institution Name

Station Name

Manufacturers Model

Name

Device Serial Number

Software Versions

Pixel Padding Value

Tag

(0008,0070)

(0008,0080)

(0008,1010)

(0008,1090)

(0018,1000)

(0018,1020)

(0028,0120)

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Type Notes

2

3

Always sent as “GE MEDICAL SYSTEMS”

Sent. Value is configurable.

3

3

Sent. Value is configurable.

Sent.

3

3

Sent, Using Gantry Serial number from

Product Locator tool.

Sent.

3 Sent.

A.3.6 General Image Module

Attribute Name

Image Number

Patient Orientation

Image Date

Image Time

Image Type

Tag

(0020,0013)

(0020,0020)

(0008,0023)

(0008,0033)

(0008,0008)

Type Notes

2 Generated sequentially, always sent.

2C

2C

Always sent zero length for SC images.

Generated for each image, always sent.

2C

3

Generated for each image, always sent.

Always sent.

Value 3: CT Image IOD specific specializations

AXIAL

LOCALIZER

SEGMENTED

REFORMATTED

PROCESSED

COMBINED

CTINTERVENTION

Value 4: GE CT Image implementation specific

MIN IP

MIP

AVERAGE

VOLREN

INTEGRAL

HD MIP

RAYSUM

SURFACE

MINMAX

155 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

Acquisition Number

Acquisition Date

Acquisition Time

(0020,0012)

(0008,0022)

(0008,0032)

Referenced Image

Sequence

>Reference SOP Class

UID

>Reference SOP

Instance UID

Irradiation Event UID

A.3.7 VOI LUT Module

Attribute Name

Window Center

Window Width

(0008,1140)

(0008,1150)

(0008,1155)

(0008,3010)

Tag

(0028,1050)

(0028,1051)

1

1

3

3

3

3

3

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

FLUORO

GSI MONO

GSI HIGH KV

GSI LOW KV

GSI MD

GSI QC

GSI EFF Z

GSI CLR OVRLY

DIGITALTILT

Generated for each acquisition, always sent.

Generated for each acquisition, always sent.

Generated for each acquisition, always sent.

Format is in fractional seconds as small as 1 millionth of a second.

Always sent.

Type

1C

1C

Notes

Window Center for display. Always sent.

Window Width for display. Always sent.

A.3.8 SOP Common Module

Attribute Name

SOP Class UID

SOP Instance UID

Specific Character Set

Instance Creation Date

Instance Creation Time

Instance Number

Tag

(0008,0016)

(0008,0018)

(0008,0005)

(0008,0012)

(0008,0013)

(0020,0013)

Type

1

1

1C

3

3

3

Notes

Always sent.

Always sent.

ISO_IR 100

Always sent.

Always sent.

Always sent.

156 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

A.4 CT Image Modules

A.4.1 Image Plane Module

Attribute Name

Slice Thickness

Image Slice Location

Image Position

Image Orientation

Tag

(0018,0050)

(0020,1041)

(0020,0032)

(0020,0037)

(0028,0030)

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Type Notes

2

3

1

1

1

Value always sent.

Value always sent. Note: Slice Location reflects the gantry ISO center table location value for the image in mm. If the gantry is tilted, this value will differ from the image position (0020, 0032) Z value which reflects the upper left-hand voxel (center of the first voxel transmitted) of the grid, in mm in the registered Frame of Reference. The system can be configured to save the upper lefthand corner Z coordinate value into (0020,

1041) Slice Location instead of the gantry ISO center table location.

Always sent. System configuration allows images to be flipped and rotated. Default behavior is to not allow flip and rotate. Note that the Image Position reflects the x, y and z coordinate of the upper left corner of the first voxel transmitted instead of the center of the first voxel transmitted.

Always sent. System configuration allows images to be flipped and rotated. Default behavior is to not allow flip and rotate.

Always sent Pixel Spacing

A.4.2 Image Pixel Module

Attribute Name

Samples per Pixel

Photometric

Interpretation

Rows

Columns

Bits Allocated

Bits Stored

High Bit

Pixel Representation

Pixel Data

Tag

(0028,0002)

(0028,0004)

(0028,0010)

(0028,0011)

(0028,0100)

(0028,0101)

(0028,0102)

(0028,0103)

(7FE0,0010)

Type Notes

1 Always sent with value = 1

1 Always sent.

1

1

1

1

1

1

1

Always sent

Always sent

Always sent with value = 16

Always sent with value = 16

Always sent with value = 15

Always sent with value = 1

Always sent

157 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

A.4.3 Contrast Bolus Module

Attribute Name

Contrast/Bolus Agent

Contrast/Bolus Route

A.4.4 CT Image Module

Attribute Name

Image Type

Samples per Pixel

Photometric Interpretation

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Tag

(0018,0010)

(0018,1040)

Type

2

3

Notes

Sent if contrast exam, as entered in user interface.

Sent if contrast exam, as entered in user interface.

Tag

(0008,0008)

(0028,0002)

(0028,0004)

158 of 181

Type

1

1

1

Notes

Always sent.

Defined terms:

Value 3:

AXIAL

LOCALIZER

SEGMENTED

REFORMATTED

PROCESSED

COMBINED

CTINTERVENTION

Value 4:

MIN IP

MIP

AVERAGE

VOLREN

INTEGRAL

HD MIP

RAYSUM

SURFACE

MINMAX

FLUORO

GSI MONO

GSI HIGH KV

GSI LOW KV

GSI MD

GSI QC

GSI EFF Z

GSI CLR OVRLY

DIGITALTILT

Always sent with value = 1

Always sent

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

Attribute Name

Bits Allocated

Bits Stored

High Bit

Rescale Intercept

Rescale Slope

Rescale Type

KV

Acquisition Number

Scan Options

Data Collection Diameter

Reconstruction Diameter

Distance Source to Detector

Distance Source to Patient

Gantry / Detector Tilt

Table Height

Rotation Direction

Exposure Time

X-Ray Tube Current

Exposure

Filter Type

Tag

(0028,0100)

(0028,0101)

(0028,0102)

(0028,1052)

(0028,1053)

(0028,1054)

(0018,0060)

(0020,0012)

(0018,0022)

(0018,0090)

(0018,1100)

(0018,1110)

(0018,1111)

(0018,1120)

(0018,1130)

(0018,1140)

(0018,1150)

(0018,1151)

(0018,1152)

(0018,1160)

3

3

3

3

3

3

3

3

3

3

3

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Type

1

1

1

1

1

1C

2

2

3

Notes

Always sent with value = 16

Always sent with value = 16

Always sent with value = 15

Always sent

Always sent

Always sent with value = HU

Value always sent.

Value always sent.

Value always sent.

Defined Terms :

AXIAL MODE

SCOUT MODE

AXIAL XRON MODE

AXIAL XROFF MODE

STATIC XRON MODE

STATIC XROFF MOD

TUBE HEAT MODE

DAS MODE

TUBE CAL MODE

BIOPSY MODE

CINE

HELICAL

ROTGENCAL MODE

FLUORO MODE

Value always sent.

Value sent for all images except scouts.

Value always sent.

Value always sent.

Value always sent.

Value always sent.

Not sent for scout or axial.

Value always sent.

Value always sent.

Value always sent. (See Note below)

Sent.

Defined terms:

BODY FILTER

MEDIUM FILTER

159 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

Attribute Name

Generator Power

Focal Spot

Convolution Kernel

Revolution Time

Single Collimation Width

Total Collimation Width

Table Speed

Table Feed per Rotation

CT Pitch Factor

CAUTION

Tag

(0018,1170)

(0018,1190)

(0018,1210)

(0018, 9305)

(0018, 9306)

(0018, 9307)

(0018, 9309)

(0018, 9310)

(0018, 9311)

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Type

3

3

3

3

3

3

3

3

3

Notes

HEAD FILTER

Always sent

Sent. Fixed value of 0.7 or

1.2.

Not sent for scouts.

Defined terms:

SMOOTH

SOFT

STANDARD

STD+

DETAIL

BONE

BONEPLUS

CHST

EDGE

SHARP

LUNG

ULTRA

HD LUNG

STANDARD2

DETAIL2

BONE2

BONEPLUS2

EDGE2

STANDARDPLUS2

DETAILPLUS2

HD ULTA

HD SOFT

Sent for spiral scan only

Sent for spiral scan only

Sent for spiral scan only

Sent for spiral scan only

Sent for spiral scan only

Sent for spiral scan only

It is possible for the operator of GEHC CT system to change the table height while scanning a series of images. Therefore, implementations must use the Frame of Reference UID

(0020,0052) in conjunction with the Table Height (0018,1130) to determine if two images are spatially related.

160 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM

Decsciption of how to calculate the Dicom Exposure field (0018,1152)

prescribed image coverage

(avg mA for image)

C

ONFORMANCE

S

TATEMENT

mA image thickness views to make image

Total coverage exposure = (exposure time) * (image avg mA)* (slice thickness)/(total coverage)

Exposue.ppt

T. Toth 03-Aug-04 exposure time = total x-ray on time helical total coverage= (exposure time * table velocity) axial or cine total coverage = macro row thickness * no of active rows

Slice thickness = nominal prospective reconstructed slice selection

Note:

A.4.5 Frame of Reference Module

Attribute Name

Frame of Reference UID

Position Reference Indicator

Tag

(0020,0052)

(0020,1040)

Type

1

2

Notes

See Caution below.

Value as entered at the user interface.

CAUTION

It is possible for the operator of GEHC CT system to change the table height while scanning a series of images. Therefore, implementations must use the Frame of Reference UID

(0020,0052) in conjunction with the Table Height (0018,1130) to determine if two images are spatially related.

A.4.6 Other Attributes

Attribute Name

Temporal Position Index

Tag

(0020,9128)

Type

3

Nominal Percentage of Cardiac

Phase

Creator Version UID

(0020,9241)

(0008,9123)

3

3

Notes

Filled into CT image for some applications. Pass number for shuttle images.

Filled into CT image for some applications. Prescribed percentage of cardiac phase.

Unique identifier filled into CT GSI images to capture the version of raw data.

161 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

A.5 SC Image Modules

A.5.1 SC Equipment Module

Attribute Name

Conversion Type

Modality

Tag

(0008,0064)

(0008,0060)

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Type

1

3

Notes

Always sent with value WSD

Modality of original image.

This will not be sent, if its value is

NULL.

A.5.2 Image Pixel Module

Attribute Name

Samples per Pixel

Photometric

Interpretation

Rows

Columns

Bits Allocated

Bits Stored

High Bit

Pixel Representation

Pixel Data

A.5.3 Modality LUT Module

Attribute Name

Rescale intercept

Rescale Slope

Rescale Type

Tag

(0028,0002)

(0028,0004)

(0028,0010)

(0028,0011)

(0028,0100)

(0028,0101)

(0028,0102)

(0028,0103)

(7FE0,0010)

Tag

(0028,1052)

(0028,1053)

(0028,1054

Type Notes

1

1

Always sent with value = 1

Always sent.

1

1

1

1

1

1

1

Always sent

Always sent

Always sent with value = 16

Always sent with value = 16

Always sent with value = 15

Always sent with value = 1

Always sent

Type

1C

1C

1C

Notes

Always sent

Always sent

Always sent with value = HU

162 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

APPENDIX B: Private Data Elements

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Enclosed is a listing of private data elements used in this implementation for CT Image IOD definition.

B.1 CT Image IOD Private Data Elements Definition

B.1.1 Private Creator Identification (GEMS_IDEN_01)

Attribute Name

Full fidelity

Suite id

Product id

Image actual date

Equipment UID

Tag

(0009,1001)

(0009,1002)

(0009,1004)

(0009,1027)

(0009,10E3)

VR

LO

SH

SH

SL

UI

VM

1

1

1

1

1

Note: For all images created by GEHC CT system (0009,1001) element will have the value

“CT_LIGHTSPEED”.

B.1.2 Private Creator Identification (GEMS_ACQU_01)

Attribute Name

Number of cells I in Detector

Cell number at Theta

Cell spacing

Horiz. Frame of ref.

Series contrast

First scan ras

Last scan ras

Table speed

Mid scan time

Mid scan flag

Degrees of azimuth

Gantry period

Number of triggers

Angle of first view

Trigger frequency

Scan FOV type

Segment number

Total segments requested

View compression factor

Recon post proc. Flag

Dependent on #views processed

Tag

(0019,1002)

(0019,1003)

(0019,1004)

(0019,100F)

(0019,1011)

(0019,1018)

(0019,101A)

(0019,1023)

(0019,1024)

(0019,1025)

(0019,1026)

(0019,1027)

(0019,102C)

(0019,102E)

(0019,102F)

(0019,1039)

(0019,1042)

(0019,1043)

(0019,1047)

(0019,1052)

(0019,106A)

VR

LO

DS

DS

SS

SL

DS

SL

SL

DS

DS

DS

SS

LO

DS

DS

SS

SS

SS

SS

SS

SS

VM

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

163 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

B.1.3 Private Creator Identification (GEMS_RELA_01)

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Attribute Name

Series from which Prescribed

Series Prescribed From

Image Prescribed From

Biopsy position

Biopsy T location

Biopsy ref location

Tag

(0021,1003)

(0021,1035)

(0021,1036)

(0021,1091)

(0021,1092)

(0021,1093)

VR VM

SS

SS

SS

SS

FL

FL

1

1

1

1

1

1

B.1.4 Private Creator Identification (GEMS_STDY_01)

Attribute Name

Start time(secs) in first axial

Tag

(0023,1070)

B.1.5 Private Creator Identification (GEMS_IMAG_01)

Attribute Name

Scout Type

Vma mamp

Vma mod

GE Noise Index * 10

Smart scan ON/OFF flag

Plane Type

Center R coord of plane image

Center A coord of plane image

Center S coord of plane image

Normal R coord

Normal A coord

Normal S coord

Table start location

Table end location

Tag

(0027,1010)

(0027,101C)

(0027,101E)

(0027,101F)

(0027,1020)

(0027,1035)

(0027,1042)

(0027,1043)

(0027,1044)

(0027,1045)

(0027,1046)

(0027,1047)

(0027,1050)

(0027,1051)

VR

FD

VR

SS

SS

FL

FL

FL

FL

SL

SL

SL

SS

FL

FL

FL

FL

VM

1

VM

1

1

1

1

1

1

1

1

1

1

1

1

1

1

B.1.6 Private Creator Identification (GEMS_0039)

Attribute Name

SR Application Name

Tag

(0039,1095)

B.1.7 Private Creator Identification (GEMS_CT_FLRO_1)

Attribute Name

CT Int Fluoro

Image Precise Location

Tag

(0029,1001)

(0029,1002)

VR

LO

VR

SS

DS

VM

1

VM

1

1

164 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

B.1.8 Private Creator Identification (GEMS_PARM_01)

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Attribute Name

Window value

X-ray chain

Number of overranges

Delta start time

Max overranges in a view

Corrected after glow terms

Reference channels

No views ref chans blocked

Scan pitch ratio

Unique image iden

Private Scan Options

RA cord of target recon center

Trigger on position

Degree of rotation

DAS trigger source

DAS fpa gain

DAS output source

DAS ad input

DAS cal mode

Start scan to X-ray on delay

Duration of X-ray on

Prospective image filter

Motion correction

Tag

(0043,1010)

(0043,1012)

(0043,1016)

(0043,101E)

(0043,101F)

(0043,1021)

(0043,1025)

(0043,1026)

(0043,1027)

(0043,1028)

(0043,102B)

(0043,1031)

(0043,1040)

(0043,1041)

(0043,1042)

(0043,1043)

(0043,1044)

(0043,1045)

(0043,1046)

(0043,104D)

(0043,104E)

(0043,1064)

(0043,1065)

VR

OB

SS

DS

FL

FL

SL

SL

SL

SL

SL

FL

FL

CS

US

SL

SS

SS

US

SH

US

SS

SS

DS

VM

1

1

1

1

1

4

2

1

1

1

1

1

8

2

1

1

6

4

1

1

3

1

1

B.1.9 Private Creator Identification (GEMS_HELIOS_01)

Note: Dicom elements (0045, 1030-1034) and (0045, 1036-1039) and (0045, 103B) are present only if the appropriate cardiac option is installed on the scanner.

Attribute Name Tag VR VM

Number of Macro Rows in Detector

Macro width at ISO Center

DAS type

DAS gain

DAS Temprature

Table Direction

Z smoothing Factor

View Weighting Mode

Sigma Row number

Minimum DAS value

Maximum Offset Value

Number of Views shifted

Z tracking Flag

Mean Z error

Z tracking Error

(0045, 1001)

(0045, 1002)

(0045, 1003)

(0045, 1004)

(0045, 1005)

(0045, 1006)

(0045, 1007)

(0045, 1008)

(0045, 1009)

(0045, 100A)

(0045, 100B)

(0045, 100C)

(0045, 100D)

(0045, 100E)

(0045, 100F)

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

SS

CS

FL

SS

SS

SS

FL

SS

SS

FL

FL

SS

SS

FL

FL

165 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

Attribute Name

Start View 2A

Number of Views 2A

Start View 1A

Sigma Mode

Number of Views 1A

Start View 2B

Number Views 2B

Start View 1B

Number of Views 1B

Iterbone Flag

Perisstaltic Flag

CardiacReconAlgorithm

AvgHeartRateForImage

TemporalResolution

PctRpeakDelay

ActualPctRpeakDelay

EkgFullMaStartPhase

EkgFullMaEndPhase

EkgModulationMaxMa

EkgModulationMinMa

NoiseReductionImageFilterDesc

RPeakTimeDelay

ActualRPeakTimeDelay

CardiacScanOptions

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Tag

(0045, 1010)

(0045, 1011)

(0045, 1012)

(0045, 1013)

(0045, 1014)

(0045, 1015)

(0045, 1016)

(0045, 1017)

(0045, 1018)

(0045, 1021)

(0045, 1022)

(0045, 1030)

(0045, 1031)

(0045, 1032)

(0045, 1033)

(0045, 1034)

(0045, 1036)

(0045, 1037)

(0045, 1038)

(0045, 1039)

(0045, 103B)

(0045, 103F)

(0045, 1044)

(0045, 1045)

VR VM

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

CS

CS

CS

CS

CS

CS

LO

IS

IS

ST

SS

SS

CS

CS

FL

SS

SS

SS

SS

SS

SS

SS

SS

SS

B.1.10 Private Creator Identification (GEMS_CT_CARDIAC_001)

Note: Private Group 49 is present only if the appropriate cardiac option is installed on the scanner.

Attribute Name

CT Cardiac Sequence

Tag

(0049, 1001)

VR

SQ

VM

1

HeartRateAtConfirm

AvgHeartRatePriorToConfirm

MinHeartRatePriorToConfirm

MaxHeartRatePriorToConfirm

StdDevHeartRatePriorToConfirm

NumHeartRateSamplesPriorToConfirm

AutoHeartRateDetectPredict

SystemOptimizedHeartRate

EkgMonitorType

NumReconSectors

RpeakTimeStamps

EkgGatingType

EkgWaveTimeOffFirstDataPoint

TemporalAlg

PhaseLocation

PreBlendedCycle1

(0049, 1002)

(0049, 1003)

(0049, 1004)

(0049, 1005)

(0049, 1006)

(0049, 1007)

(0049, 1008)

(0049, 1009)

(0049, 100A)

(0049, 100B)

(0049, 100C)

(0049, 1016)

(0049, 101B)

(0049, 1022)

(0049, 1023)

(0049, 1024)

US

CS

CS

ST

CS

FL

SH

CS

FL

CS

CS

FL

FL

CS

CS

OW

1

1

1

1

1

1-256

1

1

1

1

1

1

1

1

1

1

166 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

PreBlendedCycle2

CompressionAlg

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

(0049, 1025)

(0049, 1026)

OW

CS

1

1

B.1.11 Private Creator Identification (GEHC_CT_ADVAPP_001)

Note: Private tags (0053, 1001) through (0053,1003) and (0053,1070) through (0053,109F) are only present for multi-energy CT images.

Attribute Name Tag VR VM

MultiEnergyNoiseRedBlendingFactor

MultiEnergyNoiseRedScaleFactor

MultiEnergyMDTransformEnergies

ShuttleFlag

TableSpeedNotReachesTargetFlag

IterativeReconAnnotation

IterativeReconMode

IterativeReconConfiguration

IterativeReconLevel reconFlipRotateAnno

HiResMode

RespiratoryFlag

Image Position Patient Setting

Shutter Mode

Shutter Mode Percent

Image Browser Annotation

Overlapped Recon Flag

Row Number Anotation Flag

ImageCheckAnnotation

ODMFlag

ODMReductionPercent multiEnergySourceCount multiEnergyScanType multiEnergyReconType multiEnergyImageType multiEnergyMaterialType monchromaticEnergy

MultiEnergyWeightedSubtractionWeight1

MultiEnergyWeightedSubtractionWeight2

MultiEnergyWeightedSubtractionType

MultiEnergyAcqMethod

MultiEnergyFeatAnnotName

MultiEnergyNoiseReduced

MultiEnergyNoiseReducdeMethod

SubOptimalIQString

MultiEnergyHighLowRatio

MultiEnergyDutyCycle

MeasuredEffectiveHighkVp

MeasuredEffectiveLowkVp

(0053, 1070)

(0053, 1071)

(0053, 1072)

(0053, 1073)

(0053, 1074)

(0053, 1075)

(0053, 1076)

(0053, 1077)

(0053, 1078)

(0053, 1079)

(0053, 107A)

(0053, 107B)

(0053, 107C)

(0053, 107D)

(0053, 107E)

(0053, 107F)

(0053, 1080)

(0053, 1081)

(0053, 1001)

(0053, 1002)

(0053, 1003)

(0053, 1020)

(0053, 1021)

(0053, 1040)

(0053, 1041)

(0053, 1042)

(0053, 1043)

(0053, 1060)

(0053, 1061)

(0053, 1062)

(0053, 1063)

(0053, 1064)

(0053, 1065)

(0053, 1066)

(0053, 1067)

(0053, 1068)

(0053, 1069)

(0053, 106A)

(0053, 106B)

IS

LO

LO

LO

LO

DS

DS

DS

LO

LO

SH

SH

LO

LO

DS

DS

DS

DS

SH

SH

LO

SH

SH

SH

SH

FL

FL

IS

IS

IS

IS

IS

LO

IS

IS

CS

IS

IS

LO

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1-n

1-n

1-n

1

1

1

1

1

1

1

1

1

1

1

167 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

MeasuredEffectiveMeankVp

MeasuredEffectiveMeanmA

CommandedFirstkVp

CommandedFirstmA

CommandedSecondkVp

CommandedSecondmA

MultiEnergyKVAnnotName

MultiEnergyKVUnitLabel

MaterialType#1

MaterialType#2

GSIScanModePreset

MonoWindowLow

MonoWindoHigh

MD1AttenuationCurve

MD1intercept

MD1slope

MD1supportData

MD2supportData

NM1data

MD2intercept

MD2slope

NM2data

MD2attenuationCurve

GSIdataVersion

MARsAnnotation

MultiEnergyNumNoiseRedPairs

MultiEnergyNoiseRedString

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

(0053, 1082)

(0053, 1083)

(0053, 1084)

(0053, 1085)

(0053, 1086)

(0053, 1087)

(0053, 1088)

(0053, 1089)

(0053, 108A)

(0053, 108B)

(0053, 108C)

(0053, 108D)

(0053, 108E)

(0053, 108F)

(0053, 1092)

(0053, 1093)

(0053, 1095)

(0053, 1096)

(0053, 1097)

(0053, 1098)

(0053, 1099)

(0053, 109A)

(0053, 109B)

(0053, 109C)

(0053, 109D)

(0053, 109E)

(0053, 109F)

DS

DS

DS

DS

DS

DS

SH

SH

LO

LO

LO

IS

IS

FL

DS

DS

OW

OW

OW

DS

DS

OW

FL

SH

LO

IS

LO

1

1

1

110

1

1

1

1-n

2

110

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

2

168 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

APPENDIX C: Private Data Elements DICOMDIR Directory Information

Enclosed here is a listing of only the optional (conditional) modules and optional attributes used by this implementation in the DICOMDIR definition. All standard attributes as defined in Part 3 Addendum

(Basic Directory Information Object) are supported by this implementation but not listed here.

C.1 Basic Directory IOD Definition

Module

Directory Information

File Set Identification

Usage Notes

U

M

Sent

Sent

C.2 File Set Identification Module

Attribute Name

File Set ID

C.3 Directory Information Module

Attribute Name

Offset of the First Directory Record of the Root Directory Entity

Offset of the Last Directory Record of the Root Directory Entity

File-set Consistency Flag

Directory Record Sequence

>Offset of the Next Directory Record

>Record In-use Flag

>Offset of Referenced Lower-Level

Directory Entity

>Directory Record Type

>Referenced File ID

>Referenced SOP Class UID in file

>Referenced SOP Instance UID in File

Tag

(0004,1130)

Tag

(0004,1200)

(0004,1202)

(0004,1212)

(0004,1220)

(0004,1400)

(0004,1410)

(0004,1420)

(0004,1430)

(0004,1500)

(0004,1510)

(0004,1511)

Type Notes

2 Set by application

Type Notes

1 Set by application

1

1

2

1

1

1

1

1C

1C

1C

Set by application

0000H: no known inconsistencies.

Supported.

Set by application

FFFFH: record is in use

Set by application

“PATIENT”, “STUDY”, “SERIES”,

“IMAGE”, “PRESENTATION” and

“SR DOCUMENT”

Generated only for Image,

Presentation and SR Document

Directory Records

, starting with A/A/A/A/Z01.

Generated only for Image,

Presentation and SR Document

Directory Records

Generated only for Image,

Presentation and SR Document

169 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

>Referenced Transfer Syntax UID in File

>Record Selection Keys

(0004,1512)

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

1C

Directory Records. Set to SOP

Instance UID (0008,0018) during save to media.

Generated only for Image,

Presentation and SR Document

Directory Records

See C.4

170 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

C.4 Directory Record Selection Keys

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

STD-GEN-CD, STD-GEN-DVD-JPEG and STD-GEN-USB-JPEG Application Profiles will have PATIENT,

STUDY, SERIES, IMAGE, PRESENTATION and SR DOCUMENT directory record types. Given below are the attributes supported under each of these directories.

Additional Keys column in each list specifies whether each key is an additional key for STD-GEN-CD,

STD-GEN-DVD-JPEG and STD-GEN-USB-JPEG or not.

C.4.1 PATIENT KEYS

Attribute Name Tag Type Notes

Specific Character Set

Patient’s Name

(0008,0005)

(0010,0010)

Additional

Keys

No

No

Patient ID

Patient Birth Date

Patient Sex

(0010,0020)

(0010,0030)

(0010,0040)

No

Yes

(DVD/USB)

Yes

(DVD/USB)

1C ISO_IR 100

2 If present in composite object instances it will be set to same value, otherwise set to NULL

1 If present in composite object instances it will be set to same value, otherwise method as stated in section C.4.1.1 is followed to set the value.

1C If present in composite object instances it will be set to same value, otherwise not present

1C If present in composite object instances it will be set to same value, otherwise not present

C.4.1.1 Methods to set Patient ID

dummyPatID = 0;

} if (patientID is present in the image)

{ put patientID value in DICOMDIR

} else

{ if (current image is of same study)

{

{ put "no_patID" + dummyPatID

} else if (current image is not of same study)

} put "no_patID" + (++dummyPatID)

171 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

Later point of time, if we get the patient ID then the above dummy patient ID should be updated with the actual patient ID.

Scenario:

=========

For example if we are adding 8 images, and 4 images are of Series S1 and 4 images are of series

S2 belonging to the same Study as S1. Series S1 images are not having any value for patient ID but series S2 images are having the value of patient ID. Now while DICOMDIR creation, initially we would put dummy patient ID value when iterating through series S1 but when we reach to the second series image then previously filled dummy patient ID should be filled with the actual patient

ID found in series S2 images.

Another example: if several series belong to a same Study without Patient ID in any of the images, a single Patient ID (i.e. a single Directory Record) is assigned to this Study, and all Series are attached to the Study they belong to. Do not duplicate Study UIDs (i.e. do not generate two

Directory Records with the same Study UID).

C.4.2 STUDY KEYS

Attribute Name Tag Type Notes

Specific Character Set

Study Date

(0008,0005)

(0008,0020)

Additional

Keys

No

No

Study Time

Accession Number

Study Description

Study Instance UID

Study ID

(0008,0030)

(0008,0050)

(0008,1030)

(0020,000D)

(0020,0010)

No

No

No

No

No

1C ISO_IR 100

1

1

2

2

If present in composite object instances it will be set to same value, otherwise method as stated in sec C.4.2.1 is followed to set the value.

If present in composite object instances it will be set to same value, otherwise method as stated in sec C.4.2.1 is followed to set the value.

If present in composite object instances it will be set to same value, otherwise set to NULL

If present in composite object instances it will be set to same value, otherwise set to NULL

1C This is set to value that is present in the composite object instances, otherwise composite object instances are not

Archived.

1 If present in composite object instances it will be set to same value, otherwise method as stated in section

C.4.2.2 is followed to set the value

172 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

C.4.2.1 Method to set Study Date and Time

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

}

} if (studyDateTime is present)

{ put its value in DICOMDIR else

{ look for seriesDateTime

} if (seriesDateTime is present)

{ put seriesDateTime value in DICOMDIR else

{

} look for acquisitionDateTime (either ((0008,0022),(0008,0032)) or (0008,002A))

} if (acquisitionDateTime is present)

{ put acquisitionDateTime value in DICOMDIR else

{ look for contentDateTime

} if(contentDateTime is present)

{ put contentDateTime value in DICOMDIR

} else

{ look for instanceCreationDateTime if(instanceCreationDateTime is present)

{

} put instanceCreationDateTime value in DICOMDIR else

{

} put a DummyDateTime

}

Ideally, the oldest Date-Time found among all images belonging to the Study should be stored in

DICOMDIR.

C.4.2.2 Method to set Study ID

if (studyID is present in the image)

{ put studyID value in DICOMDIR

173 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

} else

{

}

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

put a dummy studyID value in DICOMDIR. please take care of the length (< 16)

Later point of time, if correct Study ID is found in any of the image present in the Study then dummy Study ID has to be replaced with correct one.

C.4.3 SERIES KEYS

Attribute Name Tag Type Notes

Specific Character Set

Image Type

(0008,0005)

(0008,0008)

Additional

Keys

No

No

Modality

Manufacturer’s Id

Institution Name

Institution Address

Series Description

Performing Physician’s

Name

Manufacturer’s Model

Name

Series Instance UID

(0008,0060)

(0008,0070)

(0008,0080)

(0008,0081)

(0008,103E)

(0008,1050)

(0008,1090)

(0020,000E)

No

No

Yes

(DVD / USB)

Yes

(DVD / USB)

No

Yes

(DVD / USB)

No

No

1C ISO_IR 100

3

1

If present in composite object instances it will be set to same value, otherwise not sent

If present in composite object instance it will be set to same value, otherwise an error is returned and the object not put on media

3 If present in composite object instances it will be set to same value, otherwise not sent

1C If present in composite object instances it will be set to same value, otherwise not sent

1C If present in composite object instances it will be set to same value, otherwise not sent

3 If present in composite object instances it will be set to same value, otherwise not sent

1C If present in composite object instances it will be set to same value, otherwise not sent

3 If present in composite object instances it will be set to same value, otherwise not sent

1 This is set to value that is present in the composite object instance, otherwise check for presence of

Referenced SOP Instance UID in

File (0004,1511), if both not present, instance is not put on media.

174 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

Series Number (0020,0011)

No

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

1 If present in composite object instance, it will be set to same value, otherwise method as stated in section C.4.3.1 is followed to set the value.

C.4.3.1 Method to set Series Number

}

{

} dummySeriesNum = 0; if (seriesNum is present in image) put seriesNum value in DICOMDIR else

{ if (same series)

{ put dummySeriesNum value in DICOMDIR

} else (different series)

{

} put (++dummySeriesNum) value in DICOMDIR

Later point of time, if correct series Num is found in any of the image present in the series then dummy series number has to be replaced with correct one.

C.4.4 IMAGE KEYS

Attribute Name Tag Type Notes

Specific Character Set

Image Type

(0008,0005)

(0008,0008)

Additional

Keys

No

Yes

(CD/DVD/USB)

1C

1C

SOP Instance UID

Acquisition Datetime

Acquisition Time

(0008,0018)

(0008,002A)

(0008,0032)

No

No

No

3

1C

3

ISO_IR 100

If present in composite object instances it will be set to same value, otherwise not sent

Same as (0004,1511)

If present in composite object instances it will be set to same value, otherwise not sent.

If present in composite object instances it will be set to same value, otherwise not sent

175 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

Referenced Image

Sequence

(0008,1140)

>Reference SOP Class

UID

>Reference SOP

Instance UID

Slice Thickness (0018,0050)

Spacing Between Slices (0018,0088)

(0018,0090) Data Collection

Diameter

Reconstruction

Diameter

Gantry/Detector Tilt

(0018,1100)

Convolution Kernel

Acquisition Time

Synchronized

(0018,1120)

(0018,1210)

(0018,1800)

Image Number

Image Position (Patient) (0020,0032)

Image Orientation

(Patient)

Frame of Reference UID (0020,0052)

Synchronization Frame of Reference UID

(0008,1150)

(0008,1155)

(0020,0013)

(0020,0037)

(0020,0200)

No

No

No

No

No

No

No

Yes

(CD/DVD/USB)

Yes

(CD/DVD/USB)

Yes

(CD/DVD/USB)

No

Yes

(CD/DVD/USB)

Yes

(CD/DVD/USB)

Yes

(CD/DVD/USB)

Yes

(CD/DVD/USB)

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

1C

1C

1C

3

3

3

3

3

3

1C

1

1C

1C

1C

1C

If present in composite object instances it will be set to same value, otherwise not sent

Required if Referenced

Image

Sequence (0008,1140) is present

Required if Referenced

Image

Sequence (0008,1140) is present

Included

Included

Included

Included

Included

Included

If present in composite object instances it will be set to same value, otherwise not sent.

If present in composite object instances it will be set to same value. otherwise method as stated in section

C.4.4.1 is followed to set the value.

If present in composite object instances it will be set to same value, otherwise not sent

If present in composite object instances it will be set to same value, otherwise not sent

If present in composite object instances it will be set to same value, otherwise not sent

If present in composite object instances it will be set to same value, otherwise not sent

176 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

Slice Location

Number of Frames

(0020,1041)

(0028,0008)

Rows (0028,0010)

Columns

Pixel Spacing

Bits Allocated

Bits Stored

High Bit

Lossy Image

Compression Ratio

Calibration Image

(0028,0011)

(0028,0030)

(0028,0100)

(0028,0101)

(0028,0102)

(0028,2112)

(0050,0004)

No

Yes

(DVD/USB)

Yes

(CD/DVD/USB)

Yes

(CD/DVD/USB)

Yes

(CD/DVD/USB)

No

No

No

Yes

(DVD/USB)

Yes

(DVD/USB)

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

3

1C

1

1

1C

1

1

1

1C

1C

Included

Not included

If present in composite object instances it will be set to same value, otherwise an error is returned

If present in composite object instances it will be set to same value, otherwise an error is returned

If present in composite object instances it will be set to same value, otherwise not sent

Not included

Not included

Not included

If present in composite object instances it will be set to same value, otherwise not sent

If present in composite object instances it will be set to same value, otherwise not sent

C.4.4.1 Method to set Image Number

{ dummyImageNum = 0; if (imageNum is present in image)

} put imageNum value in DICOMDIR

} else

{ put ++ dummyImageNum in DICOMDIR

C.4.5 PRESENTATION KEYS

Attribute Name Tag Additional

Keys

Type Notes

177 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

Specific Character Set

SOP Instance UID

Referenced Series

Sequence

> Referenced Image

Sequence

>> Reference SOP Class

UID

>> Reference SOP Instance

UID

> Series Instance UID

(0008,0005)

(0008,0018)

(0008,1115)

(0008,1140)

(0008,1150)

(0008,1155)

(0020,000E)

Image Number (0020,0013)

Content Label

Content Description

Presentation Creation

Date

Presentation Creation

Time

Content Creator’s Name

(0070,0080)

(0070,0081)

(0070,0082)

(0070,0083)

(0070,0084)

No

No

No

No

No

No

No

No

No

No

No

No

No

1

1

2

1

1

1

1

1C

3

1C

1

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

1

2

ISO_IR 100

Same as (0004,1511)

Calculated and sent.

Always sent, references images selected.

Always sent

Always sent

Always sent, references series selected

If present in composite object instances it will be set to same value. otherwise method as stated in section

C.4.4.1 is followed to set the value.

Set to 23 and sent.

Exam Split : This will be set as “PGP” plus Requested Procedure Code

Meaning or Requested Procedure

Description converted to all capital letters and truncated to 16 characters.

Save State : “GEGSPS_0”

Exam Split : String input on Comments box in Exam Split screen

Save State : “SE (Series Number) IM

(First Image Number) TO IM (Last

Image Number)”

Calculated and sent.

Calculated and sent.

Same as Station Name(0008,1010)

C.4.6 SR DOCUMENT KEYS

Attribute Name

Specific Character Set

SOP Instance UID

Tag

(0008,0005)

(0008,0018)

Additional

Keys

No

No

Type Notes

1C

3

ISO_IR 100

Same as (0004,1511)

178 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1

Image Date

Image Time

Image Number

Concept Name Code

Sequence

>Code Value

>Coding Scheme

Designator

>Code Meaning

Completion Flag

Verification Flag

Content Sequence

>Relationship Type

(0008,0023)

(0008,0033)

(0020,0013)

(0040,A043)

(0008,0100)

(0008,0102)

(0008,0104)

(0040,A491)

(0040,A493)

(0040,A730)

No

No

No

No

No

No

No

No

No

No

(0040,A010) No

1

1

1

1

1

1

1C

1

1

1

REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

1

Generated for each image, always sent.

Generated for each image, always sent.

If present in composite object instances it will be set to same value. otherwise method as stated in section

C.4.4.1 is followed to set the value.

A coded representation of the document title.

Always sent

Always sent

Always sent

COMPLETE sent

UNVERIFIED sent

Contains the Target Content Items that modify the Concept Name Code

Sequence of the root Content Item

(Document Title).

One or more Items shall be included in this Sequence.

All, and only, Content Items with the

HAS CONCEPT MOD relationship from the root Content Item shall be included in this Sequence.

Required if the root Content Item is the

Source Content Item of HAS CONCEPT

MOD relationships.

Enumerated Value:

HAS CONCEPT MOD

179 of 181

GE H

EALTHCARE

D

IRECTION

DOC1673565, R

EV

1 REVOLUTION ACT

S

DICOM C

ONFORMANCE

S

TATEMENT

APPENDIX D: Implementation UID for Product Versions

Product Model

Revolution ACTs

Software

Revision

16IW38.5

Implementation UID

1.2.840.113619.2.415

180 of 181

GE Healthcare

3000 N. Grandview Blvd., Waukesha, WI 53188 U.S.A. www.gehealthcare.com

Imagination at work

advertisement

Key Features

  • Network services support
  • Media storage application profiles
  • Structured report object implementation
  • Private data elements
  • Modality worklist information model
  • Print management
  • Security profiles
  • DICOM conformance

Frequently Answers and Questions

What network services does Revolution ACTS support?
Revolution ACTS supports a wide range of network services, including CT image storage, MR image storage, secondary capture image storage, grayscale softcopy presentation state storage, enhanced SR, X-ray Radiation Dose SR, RT Structure Set Storage, Positron Emission Tomography Image Storage, Study Root Query/Retrieve Information Model - FIND, Study Root Query/Retrieve Information Model - MOVE, Basic Grayscale Print Management Meta SOP Class, Basic Color Print Management Meta SOP Class, Print Job SOP Class, Printer SOP Class, Storage Commitment Push Model SOP Class, and Modality Performed Procedure Step SOP Class.
What media storage application profiles does Revolution ACTS support?
Revolution ACTS support various media storage application profiles, including Compact Disk - Recordable (CD-R), DVD, and USB. These profiles can be used to read and write files on a variety of media formats.
What are the key security features of Revolution ACTS?
Revolution ACTS includes security features like de-identification protocols to protect patient data.
What is the purpose of the DICOM conformance statement?
The DICOM Conformance Statement defines the DICOM capabilities of the Revolution ACTS CT scanner. It details the specific SOP classes, information object definitions, and private data elements supported by the device.

Related manuals

Download PDF

advertisement

Table of contents