Configuring the 20xx Inbuilt Ethernet Device.

Configuring the 20xx Inbuilt Ethernet Device.
Grant Instruments
Inbuilt Ethernet Configuration
Configuring the 20xx Inbuilt Ethernet Device.
V1.1
-1-
Grant Instruments
Inbuilt Ethernet Configuration
Contents
Configuring the 20xx Inbuilt Ethernet Device. ........................................................................... 1
Contents .................................................................................................................................... 2
Overview.................................................................................................................................... 3
Prerequisites .......................................................................................................................... 3
Setting the 20xx Loggers IP Address. ....................................................................................... 4
20xx Discovery....................................................................................................................... 4
Email Notification Setup ............................................................................................................ 6
The SQ20xx Applet. .................................................................................................................. 9
Prerequisites .......................................................................................................................... 9
Checking The Browser JVM Version ..................................................................................... 9
Updating the Java JVM.......................................................................................................... 9
Execution Over an Intranet .................................................................................................... 9
Execution Over the Web ........................................................................................................ 9
V1.1
-2-
Grant Instruments
Inbuilt Ethernet Configuration
Overview
This document describes the process for configuring the 20xx range of loggers inbuilt Ethernet adaptor.
Prerequisites
Before proceeding with the installation you will require all of the following:
- A 20xx logger with inbuilt Ethernet
- A networked PC that is able to access the network on which you intend to connect the 20xx
logger
- An Ethernet connection on the network you wish the SQ20xx logger to reside.
- Before proceeding you must speak to your network administrator to get the following information regarding the network you will be connecting the SQ20xx logger to:
○ The network gateway
○ The network subnet mask
○ An available IP address.
• This should ideally be a static IP address.
• If a static address cannot be found then the next best would be an assigned address.
• The worst case would be a leased address with a significant timeout.
NOTE: One of the above must be found. Failure to do so could result in problems communicating
with the logger in the event of a reset caused by a power failure.
-
When you have the details from your network administrator as detailed above note them
down here as they will be required when setting up communications from within SquirrelView
and the logger.
○
○
○
V1.1
IP address.
_____._____.______._____
Subnet mask
_____._____.______._____
Network gateway _____._____.______._____
-3-
Grant Instruments
Inbuilt Ethernet Configuration
Setting the 20xx Loggers IP Address.
20xx Discovery
The 20xx Discovery program will allow you to find and select the 20xx logger(s) to be setup.
To run:
1. Go to the following directory:
a. \Program Files\SquirrelView\Tools\SQ20xx_Ethernet
2. Double click on the file Finder.exe. The 20xx Discovery program should appear as below after
completing a search for connected 20xx loggers.
Next click on the logger you wish to configure in the list of devices:
V1.1
-4-
Grant Instruments
Inbuilt Ethernet Configuration
Then click on the ‘Configure IP Settings’ button to display the configuration dialog
From this dialog you can type in the settings provided by your network administrator by selecting the
‘Manually configure network settings’ radio button.
After filling in the three fields type dbps into the password field.
Finally click ‘Apply’ to send your settings to the logger.
-
V1.1
Configuration of the IP settings is now complete.
-5-
Grant Instruments
Inbuilt Ethernet Configuration
Email Notification Setup
The logger may be setup to send an email when a new alarm is raised. Before starting to setup the
alarm notification you need to ensure you have the following:
- Access to a web browser which can be connected to the 20xx device
- Mail Server Address
- Logger IP address
- Email Address for recipient
To setup email notification perform the following steps:
- Open web browser
- Type in the address of the 20xx logger followed by /home.htm
○ e.g. http://192.168.1.237/home.htm
- Enter username and password
○ Username – ‘root’ (default)
○ Password – ‘dbps’ (default)
- Select alarms
-
V1.1
Tick the ‘Enable alarm notifications’ tick box
Type in the IP address of your companies mail server
Type in the email address of the logger
Tick ‘Enable’ tick box of the alarm you wish to use
Click on the ‘Apply Button’ at the bottom of the window
-6-
Grant Instruments
-
V1.1
Inbuilt Ethernet Configuration
Click on the alarm you wish to configure, this should be the same alarm that you enabled in
the previous step.
-7-
Grant Instruments
-
V1.1
Inbuilt Ethernet Configuration
In the ‘Pin 2’ pull down box select the ‘Low’ option
○ Ensure that all other pins are set to ‘Ignore’
Tick the ‘Send E-mail to the following recipients when the alarm occurs’ tick box
Type in the email address you wish the notification to be sent to
Type in the text you wish to appear in the subject field of the email.
Finally click on the ‘Apply Button’ then logout of the 20xx device setup page.
-8-
Grant Instruments
Inbuilt Ethernet Configuration
The SQ20xx Applet.
Prerequisites
Before proceeding with the installation you will require all of the following:
-
You must complete the previous section(s) in this document i.e. the device must be configured and communicating with SquirrelView.
-
The users browser must have the latest Java Virtual Machine from Sun Microsystems installed. (See following section)
Checking The Browser JVM Version
Before running the SQ20xx Applet the Java Virtual Machine (JVM) of the browser should be upgraded to the latest version from Sun Microsystems. To check the JVM on a machine running Microsoft Internet Explorer perform the following steps:
1.
2.
3.
4.
Start Microsoft Internet Explorer browser.
Select the ‘Tools’ menus and then the ‘Internet Options..’ menu item.
In the next dialog click on the ‘Advanced’ tab.
Scroll through the list of items and check for an entry with the title ‘Java (Sun)’. If this does not
exist or the version number is less than ‘Java 2 v1.4.2_04’ then you should carry out the steps
in the ‘Updating the Java JVM’ section.
Updating the Java JVM
The latest JVM can be downloaded from the web site www.java.com. From this site select the
‘Manual Download’ link and then download the appropriate file for your operating system. If you are
using Microsoft Windows use the ‘Windows (Offline Installation)’ link and double click on the
downloaded file to install the JVM.
Execution Over an Intranet
You may now run the SQ20xx applet by entering the IP address of the SQ20xx device in a browser.
This will bring up the SQ20xx Ethernet Configuration and Management window.
- Enter username and password
○ Username – ‘root’ (default)
○ Password – ‘dbps’ (default)
Click on the ‘Set as Default’
This will enable a connection straight to the SQ20xx applet in future when the IP address is
entered into the browser. The Username and Password will need to be enter each time.
To enter the SQ20xx Ethernet Configuration and Management window after this you will need to type
in the IP address of the 20xx logger followed by /home.htm to the browser.
○ e.g. http://192.168.1.237/home.htm
Execution Over the Web
Once the SQ20xx Applet is working over the intranet the user must get access to this from the internet to view the applet from the web. To do this the users I.T. department should allow access to the
devices IP address from an IP address or URL external to the companies.
V1.1
-9-