Surveyor Plus Surveyor LC Pump Upgrade Guide

Surveyor Plus Surveyor LC Pump Upgrade Guide
Surveyor Plus
Surveyor LC Pump
Upgrade Guide
60053-97119 Revision B
January 2009
© 2009 Thermo Fisher Scientific Inc. All rights reserved.
Xcalibur is a registered trademark and ChromQuest and Surveyor are trademarks of Thermo Fisher Scientific
Inc. in the United States.
The following are registered trademarks in the United States and other countries: Windows is a registered
trademark of Microsoft Corporation.
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the
product operation. This document is copyright protected and any reproduction of the whole or any part of this
document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document supersede
all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or errorfree and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might
result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This
document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of
Sale shall govern all conflicting information between the two documents.
Release history: Revision A released in March 2006. Revision B released in January 2009
Software version: ChromQuest 4.0 SP2, ChromQuest 4.1 SP1, or Xcalibur 1.4 SP1
For Research Use Only. Not regulated for medical or veterinary diagnostic use by U.S. Federal Drug
Administration or other competent authorities.
WEEE Compliance
This product is required to comply with the European Union’s Waste Electrical & Electronic
Equipment (WEEE) Directive 2002/96/EC. It is marked with the following symbol:
Thermo Fisher Scientific has contracted with one or more recycling or disposal companies in each
European Union (EU) Member State, and these companies should dispose of or recycle this product.
See www.thermo.com/WEEERoHS for further information on Thermo Fisher Scientific’s compliance
with these Directives and the recyclers in your country.
WEEE Konformität
Dieses Produkt muss die EU Waste Electrical & Electronic Equipment (WEEE) Richtlinie 2002/96/EC
erfüllen. Das Produkt ist durch folgendes Symbol gekennzeichnet:
Thermo Fisher Scientific hat Vereinbarungen mit Verwertungs-/Entsorgungsfirmen in allen EUMitgliedsstaaten getroffen, damit dieses Produkt durch diese Firmen wiederverwertet oder entsorgt
werden kann. Mehr Information über die Einhaltung dieser Anweisungen durch Thermo Fisher
Scientific, über die Verwerter, und weitere Hinweise, die nützlich sind, um die Produkte zu
identifizieren, die unter diese RoHS Anweisung fallen, finden sie unter www.thermo.com/
WEEERoHS.
Conformité DEEE
Ce produit doit être conforme à la directive européenne (2002/96/EC) des Déchets d'Equipements
Electriques et Electroniques (DEEE). Il est marqué par le symbole suivant:
Thermo Fisher Scientific s'est associé avec une ou plusieurs compagnies de recyclage dans chaque état
membre de l’union européenne et ce produit devrait être collecté ou recyclé par celles-ci. Davantage
d'informations sur la conformité de Thermo Fisher Scientific à ces directives, les recycleurs dans votre
pays et les informations sur les produits Thermo Fisher Scientific qui peuvent aider la détection des
substances sujettes à la directive RoHS sont disponibles sur www.thermo.com/WEEERoHS.
CAUTION Symbol
CAUTION
VORSICHT
ATTENTION
PRECAUCION
AVVERTENZA
Electric Shock: This instrument uses
high voltages that can cause personal
injury. Before servicing, shut down the
instrument and disconnect the instrument
from line power. Keep the top cover on
while operating the instrument. Do not
remove protective covers from PCBs.
Elektroschock: In diesem Gerät werden
Hochspannungen verwendet, die
Verletzungen verursachen können. Vor
Wartungsarbeiten muß das Gerät
abgeschaltet und vom Netz getrennt
werden. Betreiben Sie Wartungsarbeiten
nicht mit abgenommenem Deckel. Nehmen
Sie die Schutzabdeckung von Leiterplatten
nicht ab.
Choc électrique: L’instrument utilise des
tensions capables d’infliger des blessures
corprelles. L’instrument doit être arrêté et
débranché de la source de courant avant
tout intervention. Ne pas utiliser
l’instrument sans son couvercle. Ne pas
elensver les étuis protecteurs des cartes de
circuits imprimés.
Descarga eléctrica: Este instrumento
utiliza altas tensiones, capaces de
producir lesiones personales. Antes de
dar servicio de mantenimiento al
instrumento, éste debera apagarse y
desconectarse de la línea de alimentacion
eléctrica. No opere el instrumento sin sus
cubiertas exteriores quitadas. No remueva
las cubiertas protectoras de las tarjetas
de circuito impreso.
Shock da folgorazione. L’apparecchio è
alimentato da corrente ad alta tensione
che puo provocare lesioni fisiche. Prima di
effettuare qualsiasi intervento di
manutenzione occorre spegnere ed isolare
l’apparecchio dalla linea elettrica. Non
attivare lo strumento senza lo schermo
superiore. Non togliere i coperchi a
protezione dalle schede di circuito
stampato (PCB).
Chemical: This instrument might contain
hazardous chemicals. Wear gloves when
handling toxic, carcinogenic, mutagenic,
or corrosive or irritant chemicals. Use
approved containers and proper
procedures to dispose waste oil.
Chemikalien: Dieses Gerät kann
gefährliche Chemikalien enthalten. Tragen
Sie Schutzhandschuhe beim Umgang mit
toxischen, karzinogenen, mutagenen oder
ätzenden/reizenden Chemikalien.
Entsorgen Sie verbrauchtes Öl
entsprechend den Vorschriften in den
vorgeschriebenen Behältern.
Chimique: Des produits chemiques
dangereux peuven se trouver dans
l’instrument. Proted dos gants pour
manipuler tous produits chemiques
toxiques, cancérigènes, mutagènes, ou
corrosifs/irritants. Utiliser des récipients
et des procédures homologuées pour se
débarrasser des déchets d’huile.
Química: El instrumento puede contener
productos quimicos peligrosos. Utilice
guantes al manejar productos quimicos
tóxicos, carcinogenos, mutagenos o
corrosivos/irritantes. Utilice recipientes y
procedimientos aprobados para
deshacerse del aceite usado.
Prodotti chimici. Possibile presenza di
sostanze chimiche pericolose
nell’apparecchio. Indossare dei guanti per
maneggiare prodotti chimici tossici,
cancerogeni, mutageni, o
corrosivi/irritanti. Utilizzare contenitori
aprovo e seguire la procedura indicata per
lo smaltimento dei residui di olio.
Heat: Before servicing the instrument,
allow any heated components to cool.
Hitze: Warten Sie erhitzte Komponenten
erst nachdem diese sich abgekühlt haben.
Haute Temperature: Permettre aux
composants chauffés de refroidir avant
tout intervention.
Altas temperaturas: Permita que lop
componentes se enfríen, ante de efectuar
servicio de mantenimiento.
Calore. Attendere che i componenti
riscaldati si raffreddino prima di
effetturare l’intervento di manutenzione.
Fire: Use care when operating the system
in the presence of flammable gases.
Feuer: Beachten Sie die einschlägigen
VorsichtsmaBnahmen, wenn Sie das
System in Gegenwart von entzündbaren
Gasen betreiben.
Incendie: Agir avec précaution lors de
l’utilisation du système en présence de
gaz inflammables.
Fuego: Tenga cuidado al operar el
sistema en presencia de gases
inflamables.
Incendio. Adottare le dovute precauzioni
quando si usa il sistema in presenza di gas
infiammabili.
Eye Hazard: Eye damage could occur
from splattered chemicals or flying
particles. Wear safety glasses when
handling chemicals or servicing the
instrument.
Verletzungsgefahr der Augen:
Verspritzte Chemikalien oder kleine
Partikel können Augenverletzungen
verursachen. Tragen Sie beim Umgang mit
Chemikalien oder bei der Wartung des
Gerätes eine Schutzbrille.
Danger pour les yeux: Dex projections
chimiques, liquides, ou solides peuvent
être dangereuses pour les yeux. Porter des
lunettes de protection lors de toute
manipulationde produit chimique ou pour
toute intervention sur l’instrument.
Peligro par los ojos: Las salicaduras de
productos químicos o particulas que
salten bruscamente pueden causar
lesiones en los ojos. Utilice anteojos
protectores al mnipular productos
químicos o al darle servicio de
mantenimiento al instrumento.
Pericolo per la vista. Gli schizzi di
prodotti chimici o delle particelle presenti
nell’aria potrebbero causare danni alla
vista. Indossare occhiali protettivi quando
si maneggiano prodotti chimici o si
effettuano interventi di manutenzione
sull’apparecchio.
General Hazard: A hazard is present that
is not included in the above categories.
Also, this symbol appears on the
instrument to refer the user to instructions
in this manual.
Allgemeine Gefahr: Es besteht eine
weitere Gefahr, die nicht in den
vorstehenden Kategorien beschrieben ist.
Dieses Symbol wird im Handbuch
auBerdem dazu verwendet, um den
Benutzer auf Anweisungen hinzuweisen.
Danger général: Indique la présence
d;un risque n’appartenant pas aux
catégories citées plus haut. Ce symbole
figure également sur l’instrument pour
renvoyer l’utilisateur aux instructions du
présent manuel.
Peligro general: Significa que existe un
peligro no incluido en las categorias
anteriores. Este simbolo también se utiliza
en el instrumento par referir al usuario a
las instrucciones contenidas en este
manual.
Pericolo generico. Pericolo non
compreso tra le precedenti categorie.
Questo simbolo è utilizzato inoltre
sull’apparecchio per segnalare all’utente
di consultare le istruzioni descritte nel
presente manuale.
When the safety of a procedure is
questionable, contact your local Technical
Support organization for Thermo Fisher
Scientific San Jose Products.
Wenn Sie sich über die Sicherheit eines
Verfahrens im unklaren sind, setzen Sie
sich, bevor Sie fortfahren, mit Ihrer
lokalen technischen
Unterstützungsorganisation für Thermo
Fisher Scientific San Jose Produkte in
Verbindung.
Si la sûreté d’un procédure est incertaine,
avant de continuer, contacter le plus
proche Service Clientèle pour les produits
de Thermo Fisher Scientific San Jose.
Cuando la certidumbre acerca de un
procedimiento sea dudosa, antes de
proseguir, pongase en contacto con la
Oficina de Asistencia Tecnica local para
los productos de Thermo Fisher Scientific
San Jose.
Quando e in dubbio la misura di sicurezza
per una procedura, prima di continuare, si
prega di mettersi in contatto con il
Servizio di Assistenza Tecnica locale per i
prodotti di Thermo Fisher Scientific San
Jose.
CAUTION Symbol
CAUTION
Electric Shock: This instrument uses
high voltages that can cause personal
injury. Before servicing, shut down the
instrument and disconnect the instrument
from line power. Keep the top cover on
while operating the instrument. Do not
remove protective covers from PCBs.
Chemical: This instrument might contain
hazardous chemicals. Wear gloves when
handling toxic, carcinogenic, mutagenic,
or corrosive or irritant chemicals. Use
approved containers and proper
procedures to dispose waste oil.
Heat: Before servicing the instrument,
allow any heated components to cool.
Fire: Use care when operating the system
in the presence of flammable gases.
Eye Hazard: Eye damage could occur
from splattered chemicals or flying
particles. Wear safety glasses when
handling chemicals or servicing the
instrument.
General Hazard: A hazard is present that
is not included in the above categories.
Also, this symbol appears on the
instrument to refer the user to instructions
in this manual.
When the safety of a procedure is
questionable, contact your local Technical
Support organization for Thermo Fisher
Scientific San Jose Products.
C
Contents
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ix
Related Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ix
Safety and Special Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ix
Contacting Us . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . x
Chapter 1
Upgrade Kit Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Chapter 2
Installing the Interconnect Printed Circuit Board. . . . . . . . . . . . . . . . . . . . . . . . . . .3
Chapter 3
Downloading Firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Step 1: Preparing the LC Pump for a Firmware Download . . . . . . . . . . . . . . . . . 8
Step 2: Downloading the Instrument Control Firmware . . . . . . . . . . . . . . . . . . . 9
Step 3: Downloading the Communication Firmware . . . . . . . . . . . . . . . . . . . . 14
Step 4: Returning the Pump to Normal Operation . . . . . . . . . . . . . . . . . . . . . . 19
Chapter 4
Upgrading the Chromatography Data System . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Chapter 5
Adding the Surveyor LC Pump Plus Labels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Chapter 6
Configuring the Surveyor Plus System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Configuring the Instrument in ChromQuest. . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Configuring the Surveyor LC Pump Plus . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Configuring the Signal Polarities for the Surveyor AS . . . . . . . . . . . . . . . . . . 30
Configuring the Instrument in Xcalibur . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Configuring the LC Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Configuring the Signal Polarities for the Surveyor AS . . . . . . . . . . . . . . . . . . 32
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
Thermo Scientific
Surveyor LC Pump Upgrade Guide
vii
P
Preface
This installation guide describes how to upgrade the Surveyor™ LC Pump to a Surveyor LC
Pump Plus.
To increase the retention time reproducibility of the LC system, the upgraded LC pump issues
two output signals to the autosampler at the start of a run. The LC pump issues the first
output signal when it senses a stable backpressure. It issues the second output signal when its
cams return to the start position. The autosampler waits for these two signals before it
performs an injection.
Related Documentation
In addition to this guide, Thermo Fisher Scientific provides the following documents for the
Surveyor LC Pump Plus:
• Surveyor LC Pump Plus Hardware Manual
• Surveyor Plus Getting Started with ChromQuest Guide
• Surveyor Plus Getting Started with Xcalibur Guide
Safety and Special Notices
Make sure you follow the precautionary statements presented in this guide. The safety and
other special notices appear in boxes.
Safety and special notices include the following:
CAUTION Highlights hazards to humans, property, or the environment. Each CAUTION
notice is accompanied by an appropriate CAUTION symbol.
IMPORTANT Highlights information necessary to prevent damage to software, loss of
data, or invalid test results; or might contain information that is critical for optimal
performance of the system.
Thermo Scientific
Surveyor LC Pump Upgrade Guide
ix
Preface
Note Highlights information of general interest.
Tip Highlights helpful information that can make a task easier.
Contacting Us
There are several ways to contact Thermo Fisher Scientific for the information you need.
Y To contact Technical Support
Phone
800-532-4752
Fax
561-688-8736
E-mail
us.techsupport.analyze@thermofisher.com
Knowledge base
www.thermokb.com
Find software updates and utilities to download at mssupport.thermo.com.
Y To contact Customer Service for ordering information
Phone
800-532-4752
Fax
561-688-8731
E-mail
us.customer-support.analyze@thermofisher.com
Web site
www.thermo.com/ms
Y To copy manuals from the Internet
Go to mssupport.thermo.com and click Customer Manuals in the left margin of the
window.
Y To suggest changes to documentation or to Help
• Complete a brief survey about this document by clicking the link below. Thank you
in advance for your help.
• Send an e-mail message to the Technical Publications Editor at
techpubs-lcms@thermofisher.com.
x
Surveyor LC Pump Upgrade Guide
Thermo Scientific
1
Upgrade Kit Contents
Table 1 lists the items supplied in the Surveyor LC Pump upgrade kit (P/N 60053-62045).
In addition to these items, you must also have the items listed in Table 2 on page 7 to upgrade
the communication firmware. The items listed in Table 2 are included in the Surveyor LC
Pump accessory kit.
To upgrade the Surveyor LC Pump to a Surveyor LC Pump Plus, do the following:
1. Install the new interconnect PCB by following the instructions in Chapter 2, “Installing
the Interconnect Printed Circuit Board.”
2. Download the new CPU and communication firmware by following the instructions in
Chapter 3, “Downloading Firmware.”
3. Install the service pack for your chromatography data system by following the
instructions in Chapter 4, “Upgrading the Chromatography Data System.”
4. Attach the back panel labels for the Surveyor LC Pump Plus by following the instructions
in Chapter 5, “Adding the Surveyor LC Pump Plus Labels.”
5. Configure the autosampler to recognize the output signals from the upgraded LC pump
by following the instructions in Chapter 6, “Configuring the Surveyor Plus System.”
Table 1. Parts list for the Surveyor LC Pump Upgrade kit
Thermo Scientific
Description
Part Number
LC Pump Plus interconnect PCB assembly
00950-01-00022
ChromQuest 4.0 SP1 software CD-ROM
CHROM-98002
ChromQuest 4.1 SP1 software CD-ROM
CHROM-98003
Label set, Surveyor LC Pump Plus
60053-50105
Label, Surveyor LC Pump Plus upgrade
60053-50110
Xcalibur Surveyor LC Pump 1.4 SP1 software
XCALI-64069
Manual CD-ROM
60053-64000
Surveyor LC Pump Upgrade Guide
1
2
Installing the Interconnect Printed Circuit Board
The new interconnect PCB contains the circuitry that enables the Pump RDY output signal
and the INJ HOLD output signal, generated from pins 5 and 6, respectively, on the back
panel of the LC pump. See Figure 24 on page 23.
Y To install the new PCB
1. Turn off the power to the Surveyor LC Pump by pressing the power button below the left
door of the pump, and then unplug the pump from line power.
CAUTION Because the Surveyor LC Pump contains a high voltage supply, you must
turn off the power to the pump and unplug it from line power before you pull the
chassis out of the housing.
2. Gain access to the interconnect PCB as follows:
a. Detach the LED cable from the inner right-door of the pump.
Figure 1.
Detaching the LED cable
LED cable
b. Pull the drainage funnel forward off the housing of the pump. See Figure 2.
c. Use a #2 Phillips screwdriver to remove the four screws that secure the chassis of the
pump to its housing. See Figure 2.
Thermo Scientific
Surveyor LC Pump Upgrade Guide
3
2
Installing the Interconnect Printed Circuit Board
Figure 2.
View of the LC pump, showing the chassis pulled out of the housing
Screws
Drainage
funnel
Screws
d. Pull the chassis forward out of the housing. See Figure 3.
Figure 3.
View of the LC pump without its housing
Interconnect
PCB
3. Remove the interconnect PCB as follows:
a. Detach the eight cables connected to the interconnect PCB. See Figure 4.
CAUTION Because it is easy to drop the small screws that secure the interconnect
PCB to the chassis, take care as you remove them.
4
Surveyor LC Pump Upgrade Guide
Thermo Scientific
2
Installing the Interconnect Printed Circuit Board
b. Using a #1 Phillips screwdriver, carefully remove the four screws that secure the
interconnect PCB to the chassis. See Figure 4.
Figure 4.
View of the interconnect PCB, showing the cable connections
Screws (4x)
4. Install the new interconnect PCB as follows:
a. Use the screws you removed in step 3b to secure the new interconnect PCB to the
chassis of the pump.
b. Reattach the eight cables shown in Figure 4.
5. Push the chassis back into the housing. Then secure it to the housing with the four screws
you removed in step c on page 3.
6. Reconnect the LED cable to the inner right-door of the pump. See Figure 1 on page 3.
7. Reinstall the drainage funnel. See Figure 2 on page 4.
8. Reconnect the LC pump to line power, but do not turn the pump on.
Thermo Scientific
Surveyor LC Pump Upgrade Guide
5
3
Downloading Firmware
Table 2 lists the hardware that you must have to download firmware to the CPU board.
Table 2. Hardware required to download firmware to the CPU board
Figure
Description
Part Number
8-pin, minicombicon connector 00004-02511
RJ45 to DB9 cable adapter
A3538-010
Serial communication cable
60053-63003
To download the instrument control firmware and the communication firmware, perform
these steps:
• Step 1: Preparing the LC Pump for a Firmware Download
• Step 2: Downloading the Instrument Control Firmware
• Step 3: Downloading the Communication Firmware
• Step 4: Returning the Pump to Normal Operation
Thermo Scientific
Surveyor LC Pump Upgrade Guide
7
3
Downloading Firmware
Step 1: Preparing the LC Pump for a Firmware Download
Step 1: Preparing the LC Pump for a Firmware Download
Y To prepare the LC pump for a firmware download
1. Ensure that the LC Pump is turned off.
2. Use a flathead screwdriver to set the rotary switches located on the back panel of the LC
pump to 00. See Figure 5.
Figure 5.
View of rotary switches, showing the Unit ID set to 00
9
5 6
7 8
9
4
5 6
0 1
2 3
4
0 1
2 3
7 8
3. Plug an 8-pin, 3.81 mm pitch, minicombicon connector (P/N 00004-02511) into
positions 9 through 16 on the back panel of the LC pump. Then create a jumper across
pins 9 and 15 with a piece of wire, such as a paperclip. See Figure 6.
Figure 6.
View of the LC pump back panel, showing a jumper between pins 9 and 15
Shorting wire
[connecting pin 9 (GND) to pin 15 (FW Download)]
Minicombicon plug
RS232
UNIT ID ENET
NC
FW DOWNLOAD
PRESSURE PRESSURE+
TF1
TF2
+5V @ 15OmA
GND
PUMP ON
GND
NC
NC
PROG. START
NC
+5V @ 150mA
GND
4. Plug an RJ-45 to DB9 adapter (P/N A3538-010) into a COM port on your computer
and record the port number.
5. Using a serial communication cable (P/N 60053-63003), connect the COM port on the
back panel of the computer to the RS232 connection located on the back panel of the
LC pump. See Figure 7.
8
Surveyor LC Pump Upgrade Guide
Thermo Scientific
3 Downloading Firmware
Step 2: Downloading the Instrument Control Firmware
Figure 7.
Back panel connections between the computer and the Surveyor LC Pump
Ethernet switch
Serial port with
DB9 to RJ45 adapter
Serial communication cable
Ethernet
port
Ethernet port
NC
FW DOWNLOAD
PRESSURE PRESSURE+
NC
NC
+5V @ 15OmA
GND
PUMP ON
GND
Computer
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
NC
NC
+5V @ 150mA
GND
RS232
UNIT ID
ENET
Back panel of Surveyor LC Pump
6. Turn on the power to the LC pump.
7. Verify that the Power LED is green, the Comm LED is flashing amber, and the other two
LEDs are not illuminated. See Figure 8.
Figure 8.
Surveyor LC Pump LEDs
Power
Comm
Run
Degas
Step 2: Downloading the Instrument Control Firmware
The ChromQuest software CD-ROM contains the Update.exe utility that allows you to
download the instrument control firmware.
Y To download the control firmware
1. Use Windows™ Explorer to locate the Update.exe utility on the ChromQuest software
CD-ROM. As Figure 9 shows, the Update.exe utility and the firmware file (*lcb) are
located in the following folder:
CD-ROM drive:\ChromQuest\Firmware\60053-98020 LPump v2.00 \LPump Control
Firmware
Thermo Scientific
Surveyor LC Pump Upgrade Guide
9
3
Downloading Firmware
Step 2: Downloading the Instrument Control Firmware
Figure 9.
View of the contents of the LPump Control Firmware folder
2. Double-click the Update.exe icon to start the utility.
The Update Windows Application – Update dialog box appears. See Figure 10.
Figure 10. Update Windows Application – Update dialog box
3. Set up the appropriate communication settings as follows:
a. Choose Communication > Settings to open the COM Port Setting dialog box. See
Figure 11.
10
Surveyor LC Pump Upgrade Guide
Thermo Scientific
3 Downloading Firmware
Step 2: Downloading the Instrument Control Firmware
Figure 11. COM Port Setting dialog box
b. In the COM Port Setting dialog box, do the following:
• In the Port list, select the COM port that the pump is connected to.
• In the Baud Rate list, select 19200 bps.
• In the Time out (ms) box, type 2000.
• In the Retry count box, type 3.
c. Click OK to close the COM Port Settings dialog box.
4. In the Update Windows Application – Update dialog box, click Connect.
The ROM Program Update Utility dialog box appears. See Figure 12.
Figure 12. ROM Program Update Utility dialog box
5. In the ROM Program Update Utility dialog box, click OK to return to the Update
Windows Application – Update dialog box.
The Current box displays the current firmware version. See Figure 13.
Thermo Scientific
Surveyor LC Pump Upgrade Guide
11
3
Downloading Firmware
Step 2: Downloading the Instrument Control Firmware
Figure 13. Update Windows Application – Update dialog box, showing the current firmware
version
6. In the Update Windows Application – Update dialog box, choose File > Change
Directory to open the Change Directory dialog box.
7. In the Change Directory dialog box, change the directory to the LPump Control
Firmware folder as follows (see Figure 14):
a. Double-click [chromq’’’c].
b. Double-click [firmware].
c. Double-click [60053-’’’7.00].
d. Double-click [lcpump’’’2].
e. Click OK.
Figure 14. Change Directory dialog box
The Update Windows Application – Update dialog box appears with the name of the
firmware file in the Files box. See Figure 15.
12
Surveyor LC Pump Upgrade Guide
Thermo Scientific
3 Downloading Firmware
Step 2: Downloading the Instrument Control Firmware
Figure 15. Update Windows Application – Update dialog box, showing the new file name in
the Files box
8. In the Update Windows Application – Update dialog box, click the file name to highlight
it as Figure 16 shows. Then click Update.
Figure 16. Update Windows Application – Update dialog box, showing the new file
name highlighted
The utility displays the status of the download activity at the bottom of the dialog box.
When the download is complete, the message shown in Figure 17 appears.
Figure 17. ROM Program Update Utility
Thermo Scientific
Surveyor LC Pump Upgrade Guide
13
3
Downloading Firmware
Step 3: Downloading the Communication Firmware
9. Click OK to close the ROM Program Update Utility dialog box and return to the Update
Windows Application dialog box.
The Current read-only box displays the current firmware version. See Figure 18.
Figure 18. Update Windows Application – Update dialog box, showing the updated firmware
listed in the Current box
10. After verifying a successful firmware download, click Exit to close the program.
Step 3: Downloading the Communication Firmware
The Surveyor Firmware Upgrade Utility allows you to download firmware files to the ethernet
converter board of the Surveyor instrument modules. Communication firmware upgrade files
have a .bin extension.
The installer for the Xcalibur data system automatically installs the firmware files and the
Surveyor Firmware Upgrade utility to your computer during the installation of the data
system. These files can be found in the following folder: [drive]:\Xcalibur\system\Surveyor
Firmware. See Figure 19.
14
Surveyor LC Pump Upgrade Guide
Thermo Scientific
3 Downloading Firmware
Step 3: Downloading the Communication Firmware
Figure 19. Surveyor Firmware folder installed with Xcalibur
The installer for the ChromQuest data system does not install the firmware files and the
Surveyor Firmware Upgrade Utility to your computer during the installation of the data
system. These files reside on the ChromQuest CD, and you must copy them to an
appropriate folder on your computer.
Y To download the communication firmware
1. Do one of the following:
• For the Xcalibur data system, go to step 3.
• For the ChromQuest data system, go to step 2.
2. Install the firmware files onto your computer as follows:
a. Insert the ChromQuest CD into your CD-ROM drive.
b. From the Windows desktop, double-click My Computer.
c. Right-click the CQ icon to open the shortcut menu.
d. Choose Explore from the shortcut menu to access the contents of the CD.
e. Copy the contents of the Firmware folder to your ChromQuest directory.
The Firmware folder contains folders for all the Surveyor instruments and the Surveyor
Firmware Upgrade Utility. See Figure 20.
Thermo Scientific
Surveyor LC Pump Upgrade Guide
15
3
Downloading Firmware
Step 3: Downloading the Communication Firmware
Figure 20. CD Drive:\ChromQuest\Firmware folder
3. Before you start the Surveyor Firmware Upgrade Utility application, ensure that the LC
pump is connected to the data system computer with the standard Ethernet connection.
See the Ethernet connections in Figure 7 on page 9.
4. Start the Surveyor Firmware Upgrade Utility by double-clicking the Surveyor Firmware
Upgrade Utility application.
The Surveyor Firmware Upgrade Utility window appears. See Figure 21.
16
Surveyor LC Pump Upgrade Guide
Thermo Scientific
3 Downloading Firmware
Step 3: Downloading the Communication Firmware
Figure 21. Surveyor Firmware Upgrade Utility
5. In the Surveyor Firmware Upgrade Utility window, initiate communication between the
utility and the Surveyor LC Pump Plus.
a. Under Device Type, select the Surveyor LC Pump option (see Figure 21).
CAUTION Do not interrupt the firmware download process. Do not turn off the
power to the LC pump or close the Surveyor Firmware Upgrade Utility after you
click Connect and establish communication. If you turn off the power to the
LC pump or close the utility at this point, you will have to return the pump to
the factory for reprogramming.
b. Click Connect.
The status display at the bottom of the Surveyor Firmware Upgrade Utility window
indicates that a connection has been made.
6. In the Surveyor Firmware Upgrade Utility window, select the firmware files (see
Figure 22) for the Surveyor LC Pump Plus as follows:
a. Under File Names, click Browse to the right of the ROM box to select the ROM file
(“ROM*.bin”).
The filename appears in the ROM box.
b. Under File Names, click Browse to the right of the App1 box to select the APP1 file
(“LC Pump*.bin”).
Thermo Scientific
Surveyor LC Pump Upgrade Guide
17
3
Downloading Firmware
Step 3: Downloading the Communication Firmware
The filename appears in the App1 box.
IMPORTANT If you do not select both a ROM file and an APP1 file, the
Surveyor LC Pump will not work properly.
Figure 22. Surveyor Firmware Upgrade Utility, showing selection of Surveyor LC Pump
Plus firmware files
7. After you select the appropriate firmware files, click Download. Then wait for the
download to finish.
The utility notifies you when the download is complete.
8. Click Disconnect to terminate the connection with the Surveyor LC Pump Plus, and
then close the Surveyor Firmware Upgrade Utility.
Note Now that you have terminated the connection, it is safe to turn off the power to the
LC pump.
18
Surveyor LC Pump Upgrade Guide
Thermo Scientific
3 Downloading Firmware
Step 4: Returning the Pump to Normal Operation
Step 4: Returning the Pump to Normal Operation
Y To return the LC pump to its operational configuration
1. Turn off the power to the LC pump.
2. Disconnect the serial cable from the back of the LC pump and the computer.
3. Disconnect the adapter from the computer.
4. Remove the minicombicon connector from the back panel of the LC pump.
IMPORTANT It is important to turn the Surveyor LC Pump Plus power off before
adjusting the rotary switches.
5. Use a small flathead screwdriver to reset the right-rotary switch to 1. See Figure 23.
Figure 23. View of Unit ID set to 01
9
5 6
7 8
9
4
5 6
0 1
2 3
4
0 1
2 3
7 8
6. Turn on the power to the LC pump, and then verify the following status of the LEDs:
• Power LED illuminates green
• Comm LED illuminates amber
• Run LED illuminates amber
• Degas LED illuminates amber. The Degas LED turns green within one minute after
the power is turned on if the degasser is functioning properly.
Thermo Scientific
Surveyor LC Pump Upgrade Guide
19
4
Upgrading the Chromatography Data System
If you are operating the Surveyor LC Plus system from the ChromQuest™ 4.0,
ChromQuest 4.1, or Xcalibur™ 1.4 data systems, you must install one of the following service
packs included in Surveyor LC Pump upgrade kit:
• ChromQuest 4.0 SP1
• ChromQuest 4.1 SP2
• Xcalibur 1.4 SP1
Insert the appropriate CD-ROM into your computer, and follow the installation wizard
instructions.
Thermo Scientific
Surveyor LC Pump Upgrade Guide
21
5
Adding the Surveyor LC Pump Plus Labels
After you complete the hardware and software installation, affix the self-adhesive signal label
below the signal pins on the back panel of the LC pump. Affix the self-adhesive upgrade label
above the line rating sticker. See Figure 24.
Figure 24. Back panel of upgraded Surveyor LC Pump, showing the signal label and the upgrade label
New signal label
1 2 3 4 5 6 7 8 9 10 111213141516
NC
FW DOWNLOAD
PRESSURE PRESSURE+
NC
NC
+5V @ 15OmA
GND
PUMP ON
GND
INJ HOLD
PUMP RDY
PROG START
NC
+5V @ 150mA
GND
RS232
UNIT ID
ENET
This device complies with FCC Rules, part 15.
Operation is subject to the following two conditions:
(1) This device may not cause harmful interference and
(2) This device must accept any interference received,
including interference that may cause undesired operation.
This product is covered by one or more of the following patents:
Upgraded to
Lpump Plus specifications
Upgrade label
Thermo Scientific
Surveyor LC Pump Upgrade Guide
23
6
Configuring the Surveyor Plus System
This chapter describes how to add the Surveyor LC Pump Plus to your instrument
configuration and how to configure the Surveyor Autosampler Plus to recognize the active
high input signals issued from the instrument-name.
IMPORTANT If the signal polarities for the Surveyor Autosampler Plus are not configured
correctly, the system status will remain indefinitely at “Waiting for Trigger.”
The instrument-name can be controlled from either the Xcalibur data system or the
ChromQuest data system. To configure your LC system appropriately for the
instrument-name, follow the instructions for your data system:
• Configuring the Instrument in ChromQuest
• Configuring the Instrument in Xcalibur
Configuring the Instrument in ChromQuest
To control the instrument-name from the ChromQuest data system, you must add the pump
to the instrument configuration. You must also configure the Surveyor Autosampler Plus
signal polarities to recognize the output signals from the instrument-name.
This section contains the following topics:
• Configuring the Surveyor LC Pump Plus
• Configuring the Signal Polarities for the Surveyor AS
Configuring the Surveyor LC Pump Plus
The following procedure assumes that you are adding the instrument-name to a Surveyor
instrument that already exists in the ChromQuest Enterprise. If you have not yet added an
instrument to your Enterprise, refer to your Surveyor Plus Getting Started with ChromQuest
Guide.
Thermo Scientific
Surveyor LC Pump Upgrade Guide
25
6 Configuring the Surveyor Plus System
Configuring the Instrument in ChromQuest
Y To add the instrument-name to your ChromQuest configuration
1. Double-click the ChromQuest icon on the Windows desktop to open the Main Menu
window. See Figure 25. Alternatively, from the Windows taskbar, choose Start > All
Programs > Chromatography > ChromQuest to open ChromQuest.
Figure 25. Main Menu window (for CQ 4.2)
Enterprise Login button for CQ 4.2
(For CQ 4.0 and CQ.4.1,
the login button looks like a key)
System Administration Wizard
button
2. Look at the toolbar in the Main Menu window.
• If the System Administration Wizard button is unlocked, go to step 4.
• If the System Administration Wizard button is locked, the security feature is enabled
and you are required to login. Go to step 3.
3. To log in to the ChromQuest data system, do the following:
a. Click the Enterprise Login button to open the Login dialog box shown in Figure 26.
b. Type your user name in the User Name box.
c. Type your password in the Password box.
d. If available, select the appropriate domain in the Domain list.
Note If your workstation is not networked to a domain controller, the Enable
Administration login box does not contain the Domain list box.
e. Click OK.
26
Surveyor LC Pump Upgrade Guide
Thermo Scientific
6
Configuring the Surveyor Plus System
Configuring the Instrument in ChromQuest
Figure 26. Enable Administration dialog box
4. Open the Surveyor Modules dialog box as follows:
a. In the Main Menu window, right-click the icon of the instrument that you want to
configure to display a shortcut menu.
b. Choose Configure from the shortcut menu to open the Instrument Configuration
dialog box. See Figure 27.
Figure 27. Instrument Configuration dialog box
5. To add the Surveyor LC Pump to your Surveyor instrument, click Configure to open the
Surveyor dialog box shown in Figure 28.
Thermo Scientific
Surveyor LC Pump Upgrade Guide
27
6 Configuring the Surveyor Plus System
Configuring the Instrument in ChromQuest
Figure 28. Surveyor dialog box, showing an instrument containing a PDA Plus and a
Surveyor AS
6. Double-click the Surveyor LC Pump icon in the Available Modules pane to add the
pump to your configuration.
The Surveyor LC Pump moves to the Configured Modules pane as shown in Figure 29.
Figure 29. Surveyor dialog box, showing added modules
28
Surveyor LC Pump Upgrade Guide
Thermo Scientific
6
Configuring the Surveyor Plus System
Configuring the Instrument in ChromQuest
7. Double-click the Surveyor LC Pump icon in the Configured Modules pane to open the
Surveyor LC Pump Configuration dialog box. See Figure 30.
Figure 30. Surveyor LC Pump Configuration dialog box
8. In the Surveyor LC Pump Configuration dialog box, make the following selections and
entries:
a. In the Pressure list, select the Pressure units (MPa, Bar, or psi) that you prefer to use
to display the backpressure of your system.
(1 MPa = 10 Bar = 145 psi.)
Note The data system displays the system backpressure on the Surveyor LC
Pump Instrument Status page and on the Surveyor LC Pump Instrument Setup
page.
b. (Optional) In the ID Number box, type an identification number for your pump.
c. In the Stack ID spin box, type or select the Unit ID value for your pump.
The Unit ID on the back panel of the pump consists of two rotary switches. Each
switch has ten positions. The arrow on the left switch points to the “tens” digit of the
Unit ID. The arrow on the right switch points to the “ones” digit of the Unit ID.
d. Select the Pause Sequence Following Degasser Error check box if you want the
sequence to pause after the data system detects a degasser error.
e. Click OK to exit the dialog box.
9. Do not exit the Surveyor dialog box. Go to the next procedure to modify the signal
polarities in the Surveyor Autosampler configuration.
Thermo Scientific
Surveyor LC Pump Upgrade Guide
29
6 Configuring the Surveyor Plus System
Configuring the Instrument in ChromQuest
Configuring the Signal Polarities for the Surveyor AS
Y To configure the signal polarities for the Surveyor Autosampler
1. Double-click the Surveyor AS icon (Figure 29 on page 28) in the Configured Modules
pane of the Surveyor dialog box.
The Surveyor AS Configuration dialog box appears. See Figure 31.
2. In the Surveyor AS Configuration dialog box, select the Run Starts when Pump Ready
check box and the Inject when Inject Hold signal is “Off ” check box if they are not
already selected.
3. Click OK to close the Surveyor AS Configuration dialog box.
4. Click OK to close the Surveyor dialog box.
5. Click OK to close the Instrument Configuration dialog box and return to the Main
Menu window in ChromQuest.
Note Refer to the Surveyor Plus Getting Started with ChromQuest Guide for detailed
instructions on configuring your Surveyor Autosampler Plus.
Figure 31. Surveyor AS Configuration dialog box
30
Surveyor LC Pump Upgrade Guide
Thermo Scientific
6
Configuring the Surveyor Plus System
Configuring the Instrument in Xcalibur
Configuring the Instrument in Xcalibur
To control the instrument-name from the Xcalibur data system, you must add it to the
instrument configuration. You must also configure the Surveyor Autosampler Plus signal
polarities to recognize the output signals from the instrument-name.
This section contains the following topics:
• Configuring the LC Pump
• Configuring the Signal Polarities for the Surveyor AS
Configuring the LC Pump
Y To add the instrument-name to your Xcalibur configuration
1. Choose Start > All Programs > Xcalibur > Instrument Configuration to open the
Instrument Configuration window.
2. In the Device Types list, select All to show all the instruments. The installed device
drivers are displayed as buttons in the Available Devices pane. See Figure 32.
3. Double-click the Surveyor LC Pump button in the Available Devices pane to copy the
Surveyor LC Pump to the Configured Devices pane.
Figure 32. Xcalibur Instrument Configuration dialog box
Thermo Scientific
Surveyor LC Pump Upgrade Guide
31
6 Configuring the Surveyor Plus System
Configuring the Instrument in Xcalibur
4. Double-click the Surveyor LC Pump button in the Configured Devices pane.
The Surveyor LC Pump Configuration dialog box appears. See Figure 33.
Figure 33. Surveyor LC Pump Configuration dialog box
5. In the Surveyor LC Pump Configuration dialog box, do the following:
a. In the Pressure Units list, select the units for the backpressure display.
b. In the Stack Number box, type the value for the Unit ID on the back panel of the
instrument-name.
The Unit ID on the back panel of the pump consists of two rotary switches. Each
switch has ten positions. The arrow on the left switch points to the “tens” digit of the
Unit ID. The arrow on the right switch points to the “ones” digit of the Unit ID.
6. Click OK to close the Surveyor LC Pump dialog box.
7. Leave the Instrument Configuration view open. Go to the next procedure.
Configuring the Signal Polarities for the Surveyor AS
During an injection sequence, the instrument-name issues a pump ready signal and an
injection hold release signal to the autosampler.
After you submit a run, the Surveyor AS waits for a pump ready signal from the LC pump.
The instrument-name issues this signal after its pressure transducer monitors a stable
backpressure. When the autosampler receives this signal, it begins an injection sequence.
After the autosampler begins the injection sequence, it waits for an injection hold release
signal from the pump. The instrument-name issues this release signal when its piston cam
returns to the start position. When the autosampler receives the injection hold release signal,
it switches the injection valve to the inject position, allowing the mobile phase stream to
sweep the contents of the sample loop onto the LC column.
For the instrument-name, select the Pump ready active high check box and the Injection hold
release active high check box. See Figure 34. If you do not select these two check boxes, the
autosampler does not recognize the input signals from the instrument-name and does not
begin an injection sequence.
32
Surveyor LC Pump Upgrade Guide
Thermo Scientific
6
Configuring the Surveyor Plus System
Configuring the Instrument in Xcalibur
Y To check the input signal polarities for the Surveyor Autosampler Plus
1. From the Windows XP taskbar, choose Start > All Programs > Xcalibur > Instrument
Configuration to open the Instrument Configuration window if it is not already open.
2. Double-click the Surveyor Autosampler button in the Configured Devices pane to open
the Surveyor Autosampler Configuration dialog box.
3. Click the Signal polarity tab to open the Signal polarity page. See Figure 34.
4. Make sure that the Pump ready active high and the Injection hold release active high
check boxes are selected.
5. Click OK to close the Surveyor Autosampler Configuration dialog box.
6. Click Done to close the Instrument Configuration view.
Figure 34. Configuring the input signal polarities for the Surveyor Autosampler
Thermo Scientific
Surveyor LC Pump Upgrade Guide
33
I
Index
A
E
autosampler
configuring the signal polarities in ChromQuest 30
configuring the signal polarity settings in Xcalibur 33
Ethernet cables (figure) 9
B
back panel connections for firmware downloads 9
backpressure 29
baud rate 10
bin files 14
F
Firmware
cable connections for a firmware download 16
file types 14
files for the pump ethernet converter board 18
upgrading with Surveyor Upgrade Utility 14
I
C
cables
connected to the Interconnect PCB 4
LED, connecting 5
LED, detaching (figure) 3
LED, disconnecting 3
serial communication 8
Caution about interrupting the firmware download process
Instrument Configuration dialog box (figure) 27
interconnect PCB (figure) 5
internal components of pump (figure) 4
J
jumper used for shorting pins 9 and 15 on the back panel of
the pump (figure) 8
17
Change Directory dialog box 12
chassis, removing screws 4
ChromQuest
CD-ROM 9
configuring instruments 27
configuring the signal polarities for the autosampler 30
location of firmware files 15
monitoring the system backpressure 29
opening 26
upgrade utilities 15
Comm LED, flashing amber 9
communication firmware files 14
communication settings 10
configuration in Xcalibur, pump 31
K
D
parts list for the Surveyor LC Pump Upgrade kit 1
Port setting 10
pressure units 29
drainage cone, removing 3
Thermo Scientific
kit contents 1
L
labels 23
LED cable, detaching 3
LPump Control Firmware folder contents 10
M
Main Menu window in ChromQuest, accessing 26
minicombicon connector 8
P
Surveyor LC Pump Upgrade Guide
35
Index: R
pressure, monitoring in ChromQuest 29
pump, configuring in Xcalibur 31
R
retention time reproducibility ix
retry count 10
S
screws
securing chassis to pump housing (figure) 4
securing the interconnect PCB 5
Signal label, placement 23
software CD-ROMs, part numbers 1
Stack ID 29
status of LEDs
after the pump is turned on for normal operation 19
during a firmware download 9
Surveyor dialog box (figure) 28
Surveyor Firmware folder, location for Xcalibur 15
Surveyor Firmware Upgrade Utility (figure) 17
Surveyor LC Pump Configuration dialog box (figure) 29
T
Time out setting 10
U
Unit ID
set to 00 (figure) 8
set to 01 (figure) 19
units, configuring the pressure display in ChromQuest 29
Upgrade label, placement 23
X
Xcalibur
configuring the pump 31
configuring the signal polarities for the autosampler 33
location of firmware folder 15
36