Data Sheet
K Series
421.D.101.01
Switchable
Analog Meters with
MovingāĆāIron Movement
EQ 72 SWT
EQ 96 SWT
with SlideāĆāInāĆāDial
Application
Measuring Ranges
The movingāĆāiron panel meters EQ 72/96 SWT (K series) housed in
moulded thermoplastic cases are used for the measurement of AC voltĆ
ages in the usual 3āĆāphase system.
AC voltage
500 V
operating voltage
600 V
for use on VT
N/100 V *)
operating voltage
150 V
Please state transformer ratio when ordering.
*) full scale value = 1.2 times rated value (overload scaling)
Moving-iron meters indicate rms-values practically independent of
wave form even of high harmonics. Error of indication may occur for
extreme wave forms (e.g. phase gating controls) and / or frequencies
above 100 Hz.
With the builtāĆāin 6 position switch, all voltages in a 4āĆāwire 3āĆāphase syĆ
stem can be measured.
Scaling
Movements
Moving-iron movement with pivot suspension, spring loaded shock
absorbing jewel bearings and silicon oil damping.
Mechanical Data
case details
material of case
material of window
colour of bezel
position of use
panel fixing
mounting
moulded square case suitable to be mounted
in control / switchgear panels, machine tool
consoles or mosaic panels, stackable
polycarbonate thermoplastics,
flame retardant with UL rating of 94 V - 0
glass '
black (similar to RAL 9005) '
vertical ±5° '
screw clamps
or spring clamps
stackable next to each other
panel thickness
40 mm
terminals
hexagon studs, M4 screws and
wire clamps E3
dimensionsĂ(in mm)
EQ 72 SWT
EQ 96 SWT
bezel
V 72
V 96
case
V 66
V 90
depth
53
53
panel cutout
V 68+0.7
V 92+0.8
weight approx.
0.19 kg
0.23 kg
Electrical Data
measuring unit
AC voltage / current
frequency range
162/3 ... 100 Hz
power consumption
<4.5 VA
overload capacity (acc. to DIN EN 60 051)
continuously
1.2 times rated voltage
5 s max.
2 times rated voltage, 1000 V max.
measurement category
operating voltage
pollution level
enclosure code
switch positions in a 4āĆāwire 3āĆāphase system
6 switch positions
L1L3; L2L3; L1L2; L1N; L2N; L3N
CAT III
refer to Measuring Ranges
2
IP 52 case
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact '
' also refer to "Options"
pointer
pointer deflection
scale characteristics
bar / knife-edge pointer
0 ... 90°
practically linear
above 10% of rated full-scale value
scale division
coarse-fine
scale length
EQ 72 SWT
EQ 96 SWT
61 mm
97 mm
overload scaling
1.2 times rated voltage
(voltmeters for use on voltage transformers)
Accuracy at Reference Conditions
accuracy class
1.5 according to DIN EN 60 051
reference conditions
ambient temperature
position of use
input
wave form
frequency
others
23°C±1K
nominal position ±1° '
rated measuring value
sinusoidal, distortion factor
45 ... 65 Hz
DIN EN 60 051
influences
ambient temperature
position of use
frequency
stray magnetic field
-10°C ... +23°C ... +55°C
nominal position ±5°
15 ... 100 Hz
0.5 mT
5%
Environmental
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
shock resistance
vibration resistance
climatic class 3 acc. to VDE/VDI 3540 sheet 2
-10 ... +55°C
-25 ... +65°C
75% annual average, non-condensing
15 g, 11 ms
2.5 g, 5 ... 55 Hz
Data Sheet
K Series
421.D.101.01
Switchable
Analog Meters with
MovingāĆāIron Movement
Rules and Standards
Connections
DIN 43 700
EQ 72 SWT
DIN 43 701
DIN 43 718
DIN 43 802
measuring and control instruments
for panel mounting;
nominal case and cutout dimensions
electrical switchboard instruments
bezels and front panels
scales and pointers for electrical measuring
instruments
nominal position of use and
position symbols
applicable for measuring instruments
DIN 40 050
enclosure codes;
protection of electrical equipment against
ingress of solid foreign bodies and of water
DIN EN 60 051
direct acting indicating electrical measuring
instruments and their accessories
DIN EN 61 010
safety requirements for electrically operated
measuring, control and laboratory
equipment
VDE/VDI 3540 sheet 2 reliability of measuring and control
equipment (classification of climates)
EQ 96 SWT
DIN 16 257
Options
L1
L1
L2
L2
L3
L3
N
N
non-glaring glass
colour of bezel
gray (similar to RAL 7037)
position of use
on request 15°...165°
terminal safety protection
fullāĆāsized rear cover or protective sleeves to go on hexagon studs and
M4 screws with wire clamps E3
protection against accidental contact (hand and fingers)
acc. to VBG 4 / DIN 57 106, sec. 100
dial
non-calibrated
with dial symbols
blank dial
pencil-marked on initial and end values
scale division
and figuring
0 ... 100%
linear scale division
non-standard captions on request
additional lettering
on request e.g. "generator"
additional figuring
on request
coloured marks
red, green or blue for important scale values
coloured sector
red, green or blue within scale division
logo on the dial
none or on request
b
case
window
a
Dimensions
c
dimensions (in mm)
a
b
c
d
EQ 72 SWT
72
66
53
13
d
EQ 96 SWT
96
90
53
13
Ordering Information
type
EQ
front dimensions
72 SWT
96 SWT
measuring ranges
window
colour of bezel
position of use
panel fixing
terminal safety
protection
dial
logo
*)
**)
movingāĆāiron panel meter, switchable
72 mm x 72 mm
96 mm x 96 mm
refer to preceding table
glass *)
non-glaring glass
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
vertical *)
on request 15 ... 165° **)
screw clamps *)
spring clamps
none *)
full-sized rear cover
protective sleeves
according to measuring range resp.
standard series for use on transformers *)
non-calibrated, with dial symbols
blank dial
scale division and figuring
0 ... 100%
according to standard series
additional lettering on request **)
additional figuring on request **)
coloured marks red, green or blue **)
coloured sector red, green or blue **)
WEIGEL *)
none
OEM logo **)
standard
Please clearly add the desired specifications.
ordering example
EQ 72 SWT, measuring range 0 ... 500 V, window non-glaring glass,
WEIGEL logo
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
- specifications subject to change without notice; date of issue 03/07 -
Data Sheet
K Series
420.D.101.07
Analog Meters with
MovingāĆāIron Movement
90°ā-ĂDial
EQ 48
EQ 72
EQ 96
EQ 144
K
K
K
K
with SlideāĆāInāĆāDial
Application
Electrical Data
The moving-iron panel meters EQ 48/72/96/144 K (K series) housed
in moulded thermoplastic cases are mainly used for the measurement
of AC currents and voltages in the usual technical frequency range of
162/3 ... 100 Hz.
measuring unit
AC voltage / current
frequency range
162/3 ... 100 Hz
power consumption
voltmeters
<4.5 VA
ammeters 15 A
<0.5 VA
ammeters >15 A
<0.8 VA
overload capacity (acc. to DIN EN 60 051)
continuously
1.2 times rated voltage / current
voltmeters
5 s max.
2 times, 1000 V max.
ammeters
EQ 48 K
EQ 72/96/144 K
5 s max.
10 times, 200 A max. 10 times
1 s max.
40 times, up to 250 A
measurement category CAT III
operating voltage
refer to Measuring Ranges
pollution level
2
enclosure code
IP 52 case
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact '
Moving-iron meters indicate rms-values practically independent of
wave form even of high harmonics. Error of indication may occur for
extreme wave forms (e.g. phase gating controls) and / or frequencies
above 100 Hz.
These meters are not suitable for use with shunts or tachogenerators
due to their high power consumption.
The K-series meters offer application-oriented advantages in switchĆ
board and generating set production, also suitable for mosaic panel
mounting. The bezel, the glass window and the dial can be easily exĆ
changed on site.
Movements
Moving-iron movement with pivot suspension, spring loaded shock
absorbing jewel bearings and silicon oil damping.
Mechanical Data
case details
material of case
material of window
colour of bezel
position of use
panel fixing
mounting
moulded square case suitable to be mounted
in control / switchgear panels, machine tool
consoles or mosaic panels, stackable
polycarbonate thermoplastics,
flame retardant with UL rating of 94 V - 0
glass '
black (similar to RAL 9005) '
vertical ±5° '
screw clamps
or spring clamps (except EQ 144 K)
stackable next to each other
panel thickness
40 mm
terminals
voltmeters and
hexagon studs, M4 screws and
ammeters up to 30 A wire clamps E3
40 A
threaded studs M6 with nuts
ammeters >60 A
threaded studs M8 with nuts
ammeters
dimensionsĂ(in mm)
EQ 48 K
EQ 72 K
EQ 96 K
EQ 144 K
bezel
V 48
V 72
V 96
V 144
case
V 42.5
V 66
V 90
V 136
depth
53
53
53
53
panel cutout
V 45+0.6
V 68+0.7
V 92+0.8
V 138+1
weight approx.
0.1 kg
0.15 kg
0.2 kg
0.25 kg
Measuring Ranges
measuring ranges
operating voltage
AC current
EQ 48 K EQ 72 K EQ 96 K EQ 144 K
1; 1,5; 2,5; 4; 5; 6; 10;
1
15; 25 A )
300 V
300 V
300 V
600 V
40; 60; 100 A 1) *)
300 V
300 V
600 V
AC voltage
EQ 48 K EQ 72 K EQ 96 K EQ 144 K
60 V, 100 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
250 V
300 V
600 V
600 V
600 V
400 V, 500 V
300 V
600 V
600 V
600 V
*
600 V )
600 V
600 V
600 V
for use on VT/CT
EQ 48 K EQ 72 K EQ 96 K EQ 144 K
N/1 A, N/5 A 1)
150 V
150 V
150 V
150 V
N/100 V, N/110 V 2)
150 V
150 V
150 V
150 V
Please state transformer ratio when ordering.
*) not for EQ 48 K
1) full scale value = 2 times rated value (overload scaling) '
2) full scale value = 1.2 times rated value (
-" - )
Scaling
pointer
pointer deflection
scale characteristics
scale division
scale length
bar / knife-edge pointer
0 ... 90°
practically linear
above 10% of rated full-scale value
coarse-fine
EQ 48 K EQ 72 K EQ 96 K EQ 144 K
41 mm
61 mm
97 mm
146 mm
overload scaling
ammeters
2 times rated current '
voltmeters for use on 1.2 times rated voltage
voltage transformers
' also refer to "Options"
Data Sheet
K Series
420.D.101.07
Analog Meters with
MovingāĆāIron Movement
90°ā-ĂDial
Accuracy at Reference Conditions
Options
accuracy class
1.5 according to DIN EN 60 051
reference conditions
ambient temperature
position of use
input
wave form
frequency
others
23°C±1K
nominal position ±1° '
rated measuring value
sinusoidal, distortion factor
45 ... 65 Hz
DIN EN 60 051
case
window
colour of bezel
index marking pointer
position of use
marine application
influences
ambient temperature
position of use
frequency
stray magnetic field
5%
-10°C ... +23°C ... +55°C
nominal position ±5°
15 ... 100 Hz (voltage)
15 ... 400 Hz (current)
0.5 mT
Environmental
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
shock resistance
vibration resistance
climatic class 3 acc. to VDE/VDI 3540 sheet 2
-10 ... +55°C
-25 ... +65°C
75% annual average, non-condensing
15 g, 11 ms
2.5 g, 5 ... 55 Hz
Rules and Standards
DIN 43 700
measuring and control instruments
for panel mounting;
nominal case and cutout dimensions
DIN 43 701
electrical switchboard instruments
DIN 43 718
bezels and front panels
DIN 43 802
scales and pointers for electrical measuring
instruments
DIN 16 257
nominal position of use and
position symbols
applicable for measuring instruments
DIN 40 050
enclosure codes;
protection of electrical equipment against
ingress of solid foreign bodies and of water
DIN EN 60 051
direct acting indicating electrical measuring
instruments and their accessories
DIN EN 61 010
safety requirements for electrically operated
measuring, control and laboratory
equipment
VDE/VDI 3540 sheet 2 reliability of measuring and control
equipment (classification of climates)
non-glaring glass
gray (similar to RAL 7037)
red, front adjustable
on request 15°...165°
non-certified or
with approbation by "Germanischer Lloyd"
(except EQ 48 K)
terminal safety protection
full-sized rear cover (except direct-connected ammeters > 25 A),
protective sleeves to go on hexagon studs and M4 screws with wire
clamps E3
protection against accidental contact (hand and fingers)
acc. to VBG 4 / DIN 57 106, sec. 100
dial
non-calibrated
blank dial
scale division
and figuring
linear scale division
additional lettering
additional figuring
coloured marks
coloured sector
logo on the dial
overload scaling
(ammeters)
with dial symbols
pencil-marked on initial and end values
0 ... 100%
non-standard captions on request
on request e.g. "generator"
on request
red, green or blue for important scale values
red, green or blue within scale division
none or on request
no overload range or
overload range 5 times rated current
Connections
Ordering Information
type
EQ
front dimensions
48 K
72 K
96 K
144 K
measuring ranges
window
AC voltage (direct-connected)
17
18
48 mm x 48 mm
72 mm x 72 mm
96 mm x 96 mm
144 mm x 144 mm
refer to preceding table
glass *)
non-glaring glass
colour of bezel
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
index marking pointer none *)
red, front adjustable
position of use
vertical *)
on request 15 ... 165° **)
marine application
none *)
non-certified
with approbation by
"Germanischer Lloyd" (except EQ 48 K)
panel fixing
screw clamps *)
spring clamps (except EQ 144 K)
terminal safety
none *)
protection
full-sized rear cover
protective sleeves
dial
according to measuring range resp.
standard series for use on transformers *)
non-calibrated, with dial symbols
blank dial
scale division and figuring
0 ... 100%
according to standard series
additional lettering on request **)
additional figuring on request **)
coloured marks red, green or blue **)
coloured sector red, green or blue **)
overload scaling
no overload range
(ammeters)
2 times rated current *)
5 times rated current
logo
WEIGEL *)
none
OEM logo **)
L1
N(L2)
AC current (direct-connected)
17
18
L1
N(L2)
AC current (for use on current transformer)
17
L1
18
k
l
K
L
N(L2)
b
a
Dimensions
c
dimensions (in mm)
a
b
c
EQ 48 K
48
42.5
53
EQ 72 K
72
66
53
moving-iron panel meter
EQ 96 K EQ 144 K
96
144
90
136
53
53
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
*)
**)
standard
Please clearly add the desired specifications.
ordering example
EQ 72 K, measuring range 0 ... 150 V, window non-glaring glass,
dial 0 ... 100%, no logo
- specifications subject to change without notice; date of issue 06/06 -
Data Sheet
K Series
410.D.101.10
Analog Meters with
MovingāĆāCoil Movement
90°ā-ĂDial
PQ 48
PQ 72
PQ 96
PQ 144
K
K
K
K
with SlideāĆāInāĆāDial
Application
Measuring Ranges
The movingāĆācoil panel meters PQ 48/72/96/144 K (K series) housed in
moulded thermoplastic cases are suitable for the measurement of DC
currents and voltages.
The instruments are suitable to be mounted in switchboards, control
panels, machine tool consoles and mosaic panels. The bezel, the glass
window and the dial can be easily exchanged onāĆāsite.
For mains use
Movements
SelfāĆāshielding movingāĆācoil movements with coreāĆātype magnet and
pivot suspension. Spring loaded jewel bearings for vibration and shock
resistance.
Mechanical Data
case details
material of case
material of window
colour of bezel
position of use
panel fixing
mounting
panel thickness
terminals
voltmeters and
ammeters 4 A
ammeters 60 A
ammeters 100 A
dimensionsĂ(in mm)
bezel
case
depth
panel cutout
weight approx.
moulded square case suitable to be mounted
in control / switchgear panels, machine tool
consoles or mosaic panels, stackable
polycarbonate thermoplastics,
flame retardant with UL rating of 94 V - 0
glass '
black (similar to RAL 9005) '
vertical ±5° '
screw clamps
or spring clamps (except PQ 144 K)
stackable next to each other
40 mm
hexagon studs, M4 screws and
wire clamps E3
threaded studs M6 with nuts
threaded studs M8 with nuts
PQ 48 K
V 48
V 42.5
53
V 45+0.6
0.11 kg
PQ 72 K
V 72
V 66
53
V 68+0.7
0.15 kg
PQ 96 K
V 96
V 90
53
V 92+0.8
0.2 kg
voltage drop approx.
PQā48āK PQā72/96/144āK
100 mA
1 mA
1.5 mA
2.5 mA
4 mA
5 mA
6 mA
10 mA
15 mA
20 mA
25 mA
40 mA
60 mA
1A
1.5 A
2.5 A
4A
6A
10 A
15 A
25 A
40 A 2)
60 A 2)
100 A 2)
270 mV
30 mV
90 mV
90 mV
90 mV
100 mV
100 mV
100 mV
15 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
-
400 mV
40 mV
200 mV
200 mV
200 mV
200 mV
200 mV
200 mV
15 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
DC voltage >5V
sensitivity1)'
6V
10 V
15 V
25 V
40 V
60 V
100 V
150 V
250 V
400 V 2)3)
500 V 2)3)
600 V 2)3)
1 kW/V
1 kW/V
1 kW/V
1 kW/V
1 kW/V
1 kW/V
1 kW/V
1 kW/V
1 kW/V
1 kW/V
1 kW/V
1 kW/V
for use with external shunt
60 mV current consumption 15 mA approximately,
150 mV a total lead resistance of 0.035 W is considered in the
calibration of the indicator for interconnecting leads 1 m, 2x 1 mm2
Not for mains use
sensitivity1)'
DC voltage ≤5V
PQ 144 K
V 144
V 136
53
V 138+1
0.25 kg
Electrical Data
measuring unit
DC voltage or current
overload capacity acc. to DIN EN 60 051āĆā1
continuously
1.2 times rated voltage / current
5 s max.
2 times rated voltage,
10 times rated current
overvoltage category CAT III
operating voltage
refer to Measuring Ranges
pollution level
2
enclosure code
IP 52 case (front side)
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact '
' also refer to "Options"
DC current
60 mV; 100 mV; 150 mV; 250 mV; 400 mV; 600 mV
1 V; 1.5 V; 2.5 V; 4 V; 5 V
1 kW/V
1 kW/V
for use on transducer
4 ... 20 mA mechanically suppressed zero, without zero adjustment,
voltage drop approx. 60 mV
Operating Voltages
measuring ranges
operating voltage '
DC current
PQ 48 K PQ 72 K PQ 96 K PQ 144 K
100 mA
1; 1.5; 2.5; 4; 5; 6; 10;
15; 20; 25; 40; 60 mA 150 V
150 V
150 V
150 V
1; 1.5; 2.5; 4; 6; 10;
15; 25 A
150 V
150 V
150 V
150 V
40; 60; 100 A 2)
150 V
150 V
150 V
DC voltage
PQ 48 K PQ 72 K PQ 96 K PQ 144 K
60; 100; 150; 250;
400; 600 mV
150 V
150 V
150 V
150 V
1; 1.5; 2.5; 4; 6; 10;
15; 25; 40; 60; 100 V 150 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
250 V
300 V
300 V
300 V
600 V
400; 500; 600 V 2)3)
600 V
600 V
1) the resistance values are limited to a tolerance of ±20%
2) except PQ 48 K
3) except PQ 72 K
Data Sheet
K Series
410.D.101.10
Analog Meters with
MovingāĆāCoil Movement
90°ā-ĂDial
Scaling
pointer
pointer deflection
scale characteristics
scale division
scale length
DIN EN 61 010āĆā1
bar / knife-edge pointer
0 ... 90°
linear
coarse-fine
PQ 48 K PQ 72 K PQ 96 K
41 mm
61 mm
97 mm
PQ 144 K
146 mm
Accuracy at Reference Conditions
accuracy class
1.5 according to DIN EN 60 051āĆā1
reference conditions
ambient temperature
position of use
input
others
23°C
nominal position ±1° '
rated measuring value
DIN EN 60 051āĆā1
influences
ambient temperature
position of use
stray magnetic field
23°C±2K
nominal position ±5°
0.5 mT
Environmental
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
shock resistance
vibration resistance
climatic class 3 acc. to VDE/VDI 3540 sheet 2
-10 ... +55°C
-25 ... +65°C
75% annual average, non-condensing
15 g, 11 ms
2.5 g, 5 ... 55 Hz
Rules and Standards
DIN 43 718
DIN 43 802
DIN 16 257
DIN EN 60 051
-1
-2
-9
DIN EN 60 529
Measurement and control; frontāĆāframes and
frontpanels of measurement and control
equipment; principal dimensions
Line scales and pointers for indicating
electrical measuring instruments;
general requirements
Nominal positions and position symbols used
for measuring instruments
Direct acting indicating analogue electrical
measuring instruments and their accessories
Part 1: Definitions and general requirements
common to all parts
Part 2: special requirements for ammeters
and voltmeters
Part 9: recommended test methods
Enclosure codes by housings (IPāĆācode)
Safety requirements for electrical measuring,
control and laboratory equipment
Part 1: General requirements
DIN EN 61 326āĆā1
Electrical equipment for measurement, conāĆ
trol and laboratory use - EMC requirements
Part 1: General requirements
DIN IEC 61 554
Panel mounted equipment Electrical measuring instruments Dimensions for panel mounting
VDE/VDI 3540 sheet 2 reliability of measuring and control
equipment (classification of climates)
Options
case
window
colour of bezel
index marking pointer
position of use
marine application
dial
nonāĆācalibrated
blank dial
scale division
and figuring
linear scale division
additional lettering
additional figuring
coloured marks
coloured sector
logo on the dial
dial illumination
others
zero position
increased sensitivity
adjustment of
resistance
operating voltage
nonāĆāglaring glass
gray (similar to RAL 7037)
red, front adjustable
on request 15°...165°
nonāĆācertified or
with approbation by "Germanischer Lloyd"
(except EQ 48 K)
with dial symbols
pencilāĆāmarked on initial and end values
0 ... 100%
nonāĆāstandard captions on request
on request e.g. "generator"
on request
red, green or blue for important scale values
red, green or blue within scale division
none or on request
on request
centre zero or offāĆāset zero
4 kW/V for voltmeters 1 ... 600 V
10 kW/V for voltmeters 1.5 ... 150 V
to ±1% at 23°C
higher operating voltage on request
terminal protection against accidental contact
fullāĆāsized rear cover (not for directly connected ammeters >5 A),
protective sleeves (for meters with hexagon studs and M4 screws with
wire clamps)
Connections
Ordering Information
DC voltage
17
type
PQ
front dimensions
48 K
72 K
96 K
144 K
measuring ranges
window
18
L+
L-
DC current
17
18
L+
L-
b
a
Dimensions
c
dimensions (in mm)
a
b
c
PQ 48 K
48
42.5
53
PQ 72 K
72
66
53
PQ 96 K PQ 144 K
96
144
90
136
53
53
moving-coil panel meter
48 mm x 48 mm
72 mm x 72 mm
96 mm x 96 mm
144 mm x 144 mm
refer to preceding table
glass *)
nonāĆāglaring glass
colour of bezel
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
position of use
vertical *)
on request 15 ... 165° **)
panel fixing
screw clamps *)
spring clamps (except PQ 144 K)
marine application
none *)
nonāĆācertified
with approbation by
"Germanischer Lloyd" (except PQ 48 K)
terminal protection
none *)
fullāĆāsized rear cover
protective sleeves
index marking pointer none *)
red, front adjustable
zero position
left hand zero *)
centre or offāĆāset zero **)
increased sensitivity 1 kW/V *)
4 kW/V for voltmeters 1 ... 600 V
10 kW/V for voltmeters 1.5 ... 150 V
adjustment of
±20% *)
resistance
to ±1% at 23°C
dial
scale division & measuring range alike *)
no dial
nonāĆācalibrated, with dial symbols
blank dial
scale division and figuring 0 ... 100%
linear scale division **)
additional lettering on request **)
additional figuring on request **)
coloured mark red, green or blue **)
coloured sector red, green or blue **)
logo
WEIGEL *)
none
OEM logo **)
*)
**)
standard
Please clearly add the desired specifications.
ordering example
PQ 72 K, measuring range 0 ... 20 mA, window nonāĆāglaring glass,
dial with linear scale division 0 ... 100°C, red mark at 37°C,
no logo
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
- specifications subject to change without notice; date of issue 04/10 -
Data Sheet
K Series
411.D.101.07
Analog Meters with
MovingāĆāCoil Movement
240°ā-ĂDial
LSP 72
LSP 96
LSG 72
LSG 96
K
K
K
K
with SlideāĆāInāĆāDial
Application
Measuring Ranges
The 240° moving-coil panel meters LSP 72/96 K (K series) are used
for the measurement of DC currents or DC voltages; the moving-coil
rectifier meters LSG 72/96 K measure sinusoidal AC currents or
AC voltages.
The instruments cases are injection moulded in flame retardant thermoĆ
plastic. The meters are suitable to be mounted in switchboards, control
panels, machinery and/or mosaic grid panels.
The bezel, the glass window and the dial can easily be exchanged onsite.
current
Movements
Moving-coil movement with a swivel coil system; dual spring loaded
jewel bearings for vibration and shock resistance.
The models LSG 72/96 K are AC rectified.
Mechanical Data
case details
material of case
material of window
colour of bezel
position of use
panel fixing
mounting
panel thickness
terminals
square case suitable to be mounted
in control / switchgear panels, machine tool
consoles or mosaic grid panels, stackable
polycarbonate thermoplastics,
self-extinguishing, non-dripping
with UL rating of 94 V - 0
glass '
black (similar to RAL 9005) '
vertical ±5° '
screw clamps
stackable next to each other
40 mm
hexagon studs with
M3 (LSP/G 72 K) or M4 (LSP/G 96 K) screws
and wire clamps E3 '
dimensions
LSP/G 72 K
LSP/G 96 K
bezel
V 72 mm
V 96 mm
case
V 66 mm
V 90 mm
depth
53 mm
53 mm / 104 mm *)
+0.7
mm
V 92+0.8 mm
panel cutout
V 68
weight approx.
0.25 kg
0.3 kg
*) LSG 96 K including builtāĆāin miniature current transformer
Electrical Data
measuring unit
LSP DC current or DC voltage
LSG AC current or AC voltage
frequency range LSG voltage 40 Hz ... 50 Hz ... 10 kHz
current 50 Hz (others on request)
overload capacity (acc. to DIN EN 60 051)
continuously
1.2 times rated voltage / current
5 s max.
voltmeters
2 times rated voltage
ammeters
10 times rated current
measurement category CAT III
operating voltage
refer to Measuring Ranges
pollution level
2
enclosure code
IP 52 case
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact '
' also refer to "Options"
100 mA
150 mA
250 mA
400 mA
600 mA
1 mA
1.5 mA
2.5 mA
4 mA
5 mA
6 mA
10 mA
15 mA
20 mA
25 mA
40 mA
60 mA
100 mA
150 mA
250 mA
400 mA
600 mA
1 A **)
1.5 A **)
2.5 A **)
4 A **)
'
internal resistance1) / voltage
voltage dropapprox.
LSP
LSG
6500 W
4900 W
2500 W
2500 W
1700 W
270 W
225 W
135 W
85 W
12 W
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
60 mV
1.5 V
1.5 V
1.5 V
1.5 V
1.5 V
1.5 V
1.5 V
1.5 V
1.5 V
1.5 V
1.5 V
1.5 V
1.5 V
1.5 V
1.5 V
1.5 V
1.5 V
1.5 V
1.5 V
1.5 V
1.5 V
1.5 V
0.2 V
0.2 V
0.2 V
0.3 V
LSG for use on transformer
0.2 V
N/1 A **)
0.3 V
N/5 A **)
(scaling without overload range) '
sensitivity1)
LSP
LSG
60 mV
200 W/V
100 mV
200 W/V
150 mV
200 W/V
250 mV
200 W/V
400 mV
1 kW/V
600 mV
1 kW/V
1V
1 kW/V
1.5 V
1 kW/V
2.5 V
1 kW/V
4V
1 kW/V
6V
1 kW/V
10 V
1 kW/V
15 V
1 kW/V
25 V
1 kW/V
40 V
1 kW/V
60 V
1 kW/V
100 V
1 kW/V
150 V
1 kW/V
250 V
1 kW/V
*
1 kW/V
400 V )
1 kW/V
500 V *)
600 V *)***) 1 kW/V
'
'
900 W/V
900 W/V
900 W/V
900 W/V
900 W/V
900 W/V
900 W/V
900 W/V
900 W/V
900 W/V
900 W/V
900 W/V
900 W/V
900 W/V
900 W/V
sec. 100 V
sec. 110 V
900 W/V
900 W/V
-
LSP for use on transducer ("live zero")
0/4 ... 20 mA
electrically suppressed zero with zero adjustment,
voltage drop approx. 900 mV
LSP for use with external shunt
60 mV
150 mV
sensitivity 200 W/V 1)
sensitivity 200 W/V 1)
a total lead resistance of 0.050 W is considered in the calibration
of the indicator for interconnecting leads 1 m, 2x 0.75 mm2 '
measuring ranges
operating voltage
current
150 V (LSP/LSG 72/96 K)
voltage
60; 100; 150; 250;
400; 600 mV; 1 V
1.5; 2.5; 4; 6; 10;
15; 25; 40; 60; 100 V
150 V
250 V
400; 500 V *)
600 V *)***)
LSP 72 K LSP 96 K LSG 72 K
1)
*)
**)
LSG 96 K
150 V
150 V
-
-
150 V
150 V
300 V
-
150 V
150 V
600 V
600 V
600 V
150 V
150 V
300 V
300 V
-
150 V
150 V
600 V
600 V
600 V
the resistance values are limited to a tolerance of ±20% '
not for LSP 72 K
separate (LSG 72 K) resp. builtāĆāon (LSG 96 K)
miniature current transformer 50 Hz, 10 mA sec. included
***) not for LSG 72 K
Data Sheet
K Series
411.D.101.07
Analog Meters with
MovingāĆāCoil Movement
240°ā-ĂDial
Scaling
scale
pointer
colour of pointer
pointer deflection
colour of dial
scale characteristics
scale division
scale length
Options
flat dial
bar / knife-edge pointer
black '
0 ... 240°
white '
linear
coarse-fine
LSP/G 72 K
LSP/G 96 K
106 mm
142 mm
Accuracy at Reference Conditions
accuracy class
1.5 according to DIN EN 60 051 '
reference conditions
ambient temperature
position of use
input
frequency
LSG
wave form
LSG
others
23°C±1K
nominal position ±1° '
rated measuring value
50±2 Hz
sinusoidal, distortion factor <5%
DIN EN 60 051
influences
ambient temperature
position of use
frequency
LSG
stray magnetic field
-25°C ... +23°C ... +40°C
nominal position ±5°
40 Hz ... 50 Hz ... 10 kHz
0.5 mT
Environmental
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
shock resistance
vibration resistance
climatic class 2 '
according to VDE/VDI 3540 sheet 2 '
-25 ... +40°C '
-25 ... +65°C
75% annual average, non-condensing
15 g, 11 ms
2.5 g, 5 ... 55 Hz
measuring range
special measuring
range
measuring range
adjustment
accuracy class
adjustment
lead resistance
increased sensitivity
case
window
colour of bezel
position of use
terminals
performance
climatic suitability
with operating
temperature range
marine application
enclosure code
measuring and control instruments
for panel mounting;
nominal case and cutout dimensions
DIN 43 701
electrical switchboard instruments
DIN 43 718
bezels and front panels
DIN 43 802
scales and pointers for electrical measuring
instruments
DIN 16 257
nominal position of use and
position symbols
applicable for measuring instruments
DIN 40 050
enclosure codes;
protection of electrical equipment against
ingress of solid foreign bodies and of water
DIN EN 60 051
direct acting indicating electrical measuring
instruments and their accessories
DIN EN 61 010
safety requirements for electrically operated
measuring, control and laboratory
equipment
VDE/VDI 3540 sheet 2 reliability of measuring and control
equipment (classification of climates)
adjustment potentiometer installed in
voltmeters LSP/LSG 96 K, adjustment range
approx. ±10% or ±20 ... 50%;
ammeters and LSP/G 72 K on request
1.0 with fine scale division
(as far as practicable)
of internal resistance to ±1% at 23°C
calibration to >0.05W
to 2 kW/V, 5 kW/V or 10 kW/V for voltmeters
1 ... 600 V (as far as practicable)
non-glaring glass
gray (similar to RAL 7037)
horizontal or to be specified 15°...165°
connector blades 6.3 x 0.8
limited use in the tropics
climatic class 3 according to VDE/VDI 3540
-25 ... +55°C
non-certified
at least IP 53
(front of case with meter mounted)
terminal protection
full-sized rear cover or protective sleeves
safe against accidental backhanded and fingertip contact
complying with VBG 4 / DIN 57 106, section 100
dial
blank dial
scale division
and figuring
Rules and Standards
DIN 43 700
deviating from standard
2nd scale division
coloured dial
additional lettering
additional figuring
coloured marks
coloured sector
logo on the dial
zero position
overload scaling
LSG ammeters
for use on CT
expanded scale
dial illumination
pencil marked initial and end values
0 ... 100%,
linear, full-scale values acc. to DIN series
(1āĆā1.2āĆā1.5āĆā2āĆā2.5āĆā3āĆā4āĆā5āĆā6āĆā7.5 and
any decimal multiple of these numbers
e.g. 150 m3/h) or deviating from DIN series;
special calibration from customer's
non-linear graph or chart;
scaling of voltmeters in ohms;
captions optional
including figuring
dial plate black; pointer, DIN scale division
and figuring yellow or white
non-glaring glass window included
to be specified e.g. "generator"
to be specified
red, green or blue for important scale values
red, green or blue within scale division
none or to be specified
centre zero or off-set zero
2 to 5 times rated current
expanded initial scale division (electronically)
up to approx. 5% of
full-scale value in centre scale
by two lamps 6 V, 12 V or 24 V, dial translucent
(LSP/LSG 96 K only;
LSP/LSG 72 K on request)
Connections
accuracy class
DC current
adjustments
DC voltage
17
18
17
L+
L+
L-
L-
18
sensitivity
(voltmeters)
window
AC current
17
18
17
L1
L1
N (L2)
N (L2)
colour of bezel
18
k
l
K
L
18
climatic suitability
marine application
AC voltage
17
position of use
17
18
enclosure code
terminal safety
protection
L1
L1
N (L2)
N (L2)
terminals
dial
b
a
Dimensions
c
dimensions (in mm) LSP/G 72 K
LSP/G 96 K
a
72
96
b
66
90
c
53
53 / 104 *)
*) LSG 96 K including builtāĆāin miniature current transformer
Ordering Information
type
LSP
LSG
front dimensions
72 K
96 K
measuring ranges
special measuring
range
measuring range
adjustment
(voltmeters)
240° moving-coil panel meter
measuring DC current or DC voltage
measuring AC current or AC voltage
72 mm x 72 mm
96 mm x 96 mm
refer to preceding table
to be specified **)
none *)
adjustment range ±10% approx. ***)
adjustment range ±20 ... 50% approx. ***)
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
webmaster@weigel-messgeraete.de
logo on dial
zero position
expanded initial
scale
overload scaling
LSG ammeters
for CT use
dial illumination
*) standard
**) Please clearly add
***) LSP/G 96 K only
1.5 *)
1.0 with fine scale division
internal resistance to ±20% *)
internal resistance to ±1% at 23°C
lead resistance >0.05 W
1 kW/V *)
increased to 2 kW/V
increased to 5 kW/V
increased to 10 kW/V
glass *)
non-glaring glass
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
vertical *)
to special order 15 ... 165°
class 2, -25 ... +40°C *)
class 3, -25 ... +55°C
none *)
non-certified
IP 52 *)
IP 53 at least
none *)
full-sized rear cover ***)
protective sleeves
screws and wire clamps *)
connector blades 6.3 x 0.8
scale division and measuring range alike
resp. full-scale values to DIN series
blank dial
scale division and figuring
0 ... 100%
deviating from DIN series **)
calibration fr. non-linear graph or chart **)
scaling in ohms for voltmeters **)
2nd scale division **)
black dial, yellow scale division
black dial, white scale division
additional lettering to be specified **)
additional figuring to be specified **)
coloured marks red, green or blue **)
coloured sector red, green or blue **)
WEIGEL *)
none
OEM logo **)
left-hand zero *)
centre zero or off-set zero **)
none *)
up to approx. 5%, electronically
without overload range *)
2 to 5 times rated current
none *)
by 2 lamps 6 V, 12 V or 24 V ***)
the desired specifications.
ordering example
LSP 72 K, measuring range 0 ... 20 mA, scale 0 ... 100 A,
window non-glaring glass, WEIGEL logo
- specifications subject to change without notice; date of issue 02/07 -
Data Sheet
K Series
415.D.101.08
Analog Meters with
MovingāĆāCoil Movement
and Rectifier
90°ā-ĂDial
VQ 48
VQ 72
VQ 96
VQ 144
K
K
K
K
with SlideāĆāInāĆāDial
Application
Measuring Ranges
The moving-coil rectifier meters VQ 48/72/96/144 K (K series) housed
in moulded thermoplastic cases are used for the measurement of sinuĆ
soidal AC currents and voltages.
Moving-coil rectifier instruments measure average values and are
scaled to indicate r.m.s., assuming a sinusoidal wave form.
AC current
voltage drop approx. AC voltage
10 mA
15 mA
25 mA
40 mA
60 mA
100 mA
The instruments are suitable to be mounted in switchboards, control
panels, machine tool consoles and mosaic panels. The bezel, the glass
window and the dial can be easily exchanged on-site.
1.7 V
1.7 V
1.7 V
1.9 V
1.9 V
2.0 V
6V
10 V
15 V
25 V
40 V
60 V
100 V
150 V
250 V
400 V
500 V
600 V *)
Movements
Self-shielding moving-coil movements with core-type magnet and
pivot suspension. Spring loaded jewel bearings for vibration and shock
resistance.
Mechanical Data
case details
moulded square case suitable to be mounted
in control / switchgear panels, machine tool
consoles or mosaic panels, stackable
material of case
polycarbonate thermoplastics,
flame retardant with UL rating of 94 V - 0
material of window
colour of bezel
glass '
black (similar to RAL 9005) '
position of use
panel fixing
vertical ±5° '
screw clamps
or spring clamps (except VQ 144 K)
mounting
panel thickness
stackable next to each other
40 mm
terminals
hexagon studs, M4 screws and
wire clamps E3
dimensionsĂ(in mm)
VQ 48 K
VQ 72 K
VQ 96 K
VQ 144 K
bezel
V 48
V 72
V 96
V 144
case
V 42.5
V 66
V 90
V 136
depth
53
53
53
53
panel cutout
V 45+0.6
V 68+0.7
V 92+0.8
V 138+1
weight approx.
0.11 kg
0.15 kg
0.2 kg
0.25 kg
Electrical Data
measuring unit
frequency range
AC voltage or current
voltage 40 Hz ... 50 Hz ... 10 kHz
current 50 Hz (others on request)
overload capacity (acc. to DIN EN 60 051)
continuously
1.2 times rated voltage / current
5 s max.
2 times rated voltage,
10 times rated current
measurement category CAT III
operating voltage
2
enclosure code
IP 52 case
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact '
900 W/V
900 W/V
900 W/V
900 W/V
900 W/V
900 W/V
900 W/V
900 W/V
900 W/V
900 W/V
900 W/V
900 W/V
measuring ranges
operating voltage
AC current
10; 15; 20; 25; 40; 60;
100 mA
AC voltage
6; 10; 15; 25; 40; 60;
100; 150 V
250 V
400; 500 V
600 V *)
VQ 48 K
VQ 72 K
VQ 96 K
VQ 144 K
150 V
VQ 48 K
150 V
VQ 72 K
150 V
VQ 96 K
150 V
VQ 144 K
150 V
300 V
300 V
-
150 V
300 V
300 V
600 V
150 V
600 V
600 V
600 V
*)
1)
150 V
300 V
300 V
-
not for VQ 48/72 K
the resistance values are limited to a tolerance of ±20%
Scaling
pointer
pointer deflection
scale characteristics
scale division
scale length
bar / knife-edge pointer
0 ... 90°
linear
coarse-fine
VQ 48 K VQ 72 K VQ 96 K
41 mm
61 mm
97 mm
VQ 144 K
146 mm
Accuracy at Reference Conditions
accuracy class
1.5 according to DIN EN 60 051 '
reference conditions
ambient temperature
position of use
input
frequency
wave form
others
23°C±1K
nominal position ±1° '
rated measuring value
45 ... 50 ... 65 Hz
sinusoidal, distortion factor <5%
DIN EN 60 051
influences
ambient temperature
position of use
frequency
stray magnetic field
-10°C ... +23°C ... +55°C
nominal position ±5°
40 ... 50 ... 100 Hz
0.5 mT
refer to Measuring Ranges
pollution level
sensitivity1)
' also refer to "Options"
Data Sheet
K Series
415.D.101.08
Analog Meters with
MovingāĆāCoil Movement
and Rectifier
90°ā-ĂDial
Environmental
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
shock resistance
vibration resistance
climatic class 3 acc. to VDE/VDI 3540 sheet 2
-10 ... +55°C
-25 ... +65°C
75% annual average, non-condensing
15 g, 11 ms
2.5 g, 5 ... 55 Hz
terminal protection against accidental contact
full-sized rear cover (except direct-connected ammeters > 25 A),
protective sleeves to go on hexagon studs and M4 screws with wire
clamps E3
protection against accidental contact (hand and fingers)
acc. to VBG 4 / DIN 57 106, sec. 100
Connections
AC voltage
Rules and Standards
18
17
18
DIN 43 700
Options
dial
non-calibrated
blank dial
scale division
and figuring
linear scale division
additional lettering
additional figuring
coloured marks
coloured sector
logo on the dial
others
increased sensitivity
adjustment of
resistance
N(L2)
AC current
L1
N(L2)
Dimensions
non-glaring glass
gray (similar to RAL 7037)
red, front adjustable
on request 15°...165°
non-certified or
with approbation by "Germanischer Lloyd"
(except VQ 48 K)
with dial symbols
pencil-marked on initial and end values
0 ... 100%
non-standard captions on request
on request e.g. "generator"
on request
red, green or blue for important scale values
red, green or blue within scale division
none or on request
4 kW/V for voltmeters 6 ... 600 V
10 kW/V for voltmeters 6 ... 150 V
to ±1% at 23°C
a
case
window
colour of bezel
index marking pointer
position of use
marine application
L1
b
measuring and control instruments
for panel mounting;
nominal case and cutout dimensions
DIN 43 701
electrical switchboard instruments
DIN 43 718
bezels and front panels
DIN 43 802
scales and pointers for electrical measuring
instruments
DIN 16 257
nominal position of use and
position symbols
applicable for measuring instruments
DIN 40 050
enclosure codes;
protection of electrical equipment against
ingress of solid foreign bodies and of water
DIN EN 60 051
direct acting indicating electrical measuring
instruments and their accessories
DIN EN 61 010
safety requirements for electrically operated
measuring, control and laboratory
equipment
VDE/VDI 3540 sheet 2 reliability of measuring and control
equipment (classification of climates)
17
c
dimensions (in mm)
a
b
c
VQ 48 K
48
42.5
53
VQ 72 K
72
66
53
VQ 96 K VQ 144 K
96
144
90
136
53
53
Ordering Information
type
VQ
front dimensions
48 K
72 K
96 K
144 K
measuring ranges
window
moving-coil rectifier panel meter
48 mm x 48 mm
72 mm x 72 mm
96 mm x 96 mm
144 mm x 144 mm
refer to preceding table
glass *)
non-glaring glass
colour of bezel
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
position of use
vertical *)
on request 15 ... 165° **)
panel fixing
screw clamps *)
spring clamps (except VQ 144 K)
marine application
none *)
non-certified
with approbation by
"Germanischer Lloyd" (except VQ 48 K)
terminal protection
none *)
full-sized rear cover
protective sleeves
index marking pointer none *)
red, front adjustable
increased sensitivity 4 kW/V for voltmeters 6 ... 600 V
10 kW/V for voltmeters 6 ... 150 V
adjustment of
resistance
dial
logo
*)
**)
none *)
to ±1% at 23°C
scale division & measuring range alike *)
no dial
non-calibrated, with dial symbols
blank dial
scale division and figuring 0 ... 100%
linear scale division **)
additional lettering on request **)
additional figuring on request **)
coloured marks red, green or blue **)
coloured sector red, green or blue **)
WEIGEL *)
none
OEM logo **)
standard
Please clearly add the desired specifications.
ordering example
VQ 72 K, measuring range 0 ... 100 mA, window non-glaring glass,
dial 0 ... 100%, red mark at 90%, no logo
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
- specifications subject to change without notice; date of issue 06/06 -
Data Sheet
K Series
432.D.101.07
Analog PointerāĆāType
Frequency Meters
90° or 240° Dial
ZQ 72
ZQ 96
ZQ 144
LSZ 96
K
K
K
K
with SlideāĆāInāĆāDial
Application
Measuring Ranges
The pointer-type frequency meters ZQ 72/96/144 K with 90° dial and
LSZ 96 K with 240° dial (K series) housed in moulded thermoplastic
cases are used for measurement of frequencies in the range of 50, 60,
200 or 400 Hz.
frequency range
For maximum accuracy, the meteers show only a small range around
the selected nominal frequency. The nonāĆāinteresting measuring range
is suppressed electronically.
The bezel, the front window and the dial can be exchanged easily.
Functional Principle
SelfāĆāshielding movingāĆācoil movements with coreāĆātype magnet and
pivot suspension (ZQ) or with swivel coil (LSZ). Spring loaded jewel
bearings for vibration and shock resistance.
Electronic measuring unit based on condensor reloading.
Mechanical Data
case details
material of case
material of window
colour of bezel
position of use
panel fixing
mounting
panel thickness
terminals
moulded square case suitable to be mounted
in control / switchgear panels, machine tool
consoles or mosaic panels, stackable
polycarbonate thermoplastics,
flame retardant with UL rating of 94 V - 0
glass '
black (similar to RAL 9005) '
vertical ±5° '
screw clamps
stackable next to each other
40 mm
hexagon studs, M4 screws and
wire clamps E3
45 ... 50 ...
48 ... 50 ...
55 ... 60 ...
58 ... 60 ...
180 ... 200 ...
360 ... 400 ...
380 ... 400 ...
rated voltage
55 Hz
52 Hz
65 Hz
62 Hz
220 Hz
440 Hz
420 Hz
57.7 V
63.5 V
100 V
110 V
115 V
120 V
127 V
208 V
230 V
289 V
400 V
415 V
440 V
500 V
operating voltage
ZQ 72 K
ZQ 96 K
ZQ 144 K
LSZ 96 K
150 V
150 V
150 V
150 V
150 V
150 V
150 V
300 V
300 V
300 V
300 V
300 V
300 V
300 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
600 V
600 V
600 V
600 V
600 V
600 V
600 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
600 V
600 V
600 V
600 V
600 V
600 V
600 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
300 V
300 V
300 V
300 V
300 V
300 V
300 V
Scaling
dial
pointer
pointer deflection
scale characteristics
scale division
scale length
flat dial
bar / knife-edge pointer
0 ... 90° (ZQ)
0 ... 240° (LSZ)
linear
coarse-fine
ZQ 72 K
ZQ 96 K
ZQ 144 K
61 mm
97 mm
146 mm
dimensions
ZQ 72 K
ZQ/LSZ 96 K
ZQ 144 K
bezel
V 72 mm
V 96 mm
V 144 mm
case
V 66 mm
V 90 mm
V 136 mm
depth
53 mm
53 mm
53 mm
accuracy class
0.5 according to DIN EN 60 051
panel cutout
V 68+0.7 mm V 92+0.8 mm
V 138+1 mm
weight approx.
0,3 kg
0,3 kg
reference conditions
ambient temperature
position of use
input
voltage
others
23°C±1K
nominal position ±1° '
rated measuring value
nominal voltage
DIN EN 60 051
influences
ambient temperature
position of use
stray magnetic field
-10°C ... +23°C ... +55°C
nominal position ±5°
0.5 mT
0,3 kg
Electrical Data
measuring unit
frequency
admissible voltage
variation
-15% ... +10% of rated voltage
3 VA
power consumption
measurement category CAT III
operating voltage
refer to Measuring Ranges
pollution level
2
enclosure code
IP 52 case
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact '
' also refer to "Options"
LSZ 96 K
142 mm
Accuracy at Reference Conditions
Data Sheet
K Series
432.D.101.07
Analog PointerāĆāType
Frequency Meters
90° or 240° Dial
Environmental
climatic class 3 acc. to VDE/VDI 3540 sheet 2
-10 ... +55°C
measuring and control instruments
for panel mounting;
nominal case and cutout dimensions
DIN 43 701
electrical switchboard instruments
DIN 43 718
bezels and front panels
DIN 43 802
scales and pointers for electrical measuring
instruments
DIN 16 257
nominal position of use and
position symbols
applicable for measuring instruments
DIN 40 050
enclosure codes;
protection of electrical equipment against
ingress of solid foreign bodies and of water
DIN EN 60 051
direct acting indicating electrical measuring
instruments and their accessories
DIN EN 61 010
safety requirements for electrically operated
measuring, control and laboratory
equipment
VDE/VDI 3540 sheet 2 reliability of measuring and control
equipment (classification of climates)
Options
non-glaring glass
gray (similar to RAL 7037)
red, front adjustable (ZQ 72/96/144 K only)
on request 15°...165°
non-certified or
with approbation by "Germanischer Lloyd"
(ZQ 96/144 K only)
with dial symbols
pencil-marked on initial and end values
0 ... 100%
non-standard captions on request
on request e.g. "generator"
on request
red, green or blue for important scale values
red, green or blue within scale division
none or on request
L1
N(L2)
Dimensions
b
75% annual average, non-condensing
15 g, 11 ms
2.5 g, 5 ... 55 Hz
DIN 43 700
dial
non-calibrated
blank dial
scale division
and figuring
linear scale division
additional lettering
additional figuring
coloured marks
coloured sector
logo on the dial
18
-25 ... +65°C
Rules and Standards
case
window
colour of bezel
index marking pointer
position of use
marine application
17
a
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
shock resistance
vibration resistance
Connections
c
dimensions (in mm)
a
b
c
ZQ 72 K
72
66
53
ZQ/LSZ 96 K
96
90
53
ZQ 144 K
144
136
53
Ordering Information
type
ZQ
LSZ (96 K only)
front dimensions
72 K
96 K
144 K
measuring ranges
rated voltages
window
pointer-type frequency meter, 90° dial
pointer-type frequency meter, 240° dial
72 mm x 72 mm
96 mm x 96 mm
144 mm x 144 mm
refer to preceding table
refer to preceding table
glass *)
non-glaring glass
colour of bezel
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
index marking pointer none *)
red, front adjustable ****)
position of use
vertical *)
on request 15 ... 165° **)
marine application
none *)
non-certified
with approbation by
"Germanischer Lloyd" ***)
terminal protection
none *)
full-sized rear cover
protective sleeves
dial
scale division & measuring range alike *)
no dial
non-calibrated, with dial symbols
blank dial
scale division and figuring 0 ... 100%
linear scale division **)
additional lettering on request **)
additional figuring on request **)
coloured marks red, green or blue **)
coloured sector red, green or blue **)
logo
WEIGEL *)
none
OEM logo **)
*)
standard
**) Please clearly add the
***) ZQ 96/144 K only
****) ZQ 72/96/144 K only
desired specifications.
ordering example
ZQ 96 K, measuring range 45 ... 50 ... 55 Hz, rated voltage AC 230 V,
window non-glaring glass, WEIGEL logo
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
- specifications subject to change without notice; date of issue 06/06 -
Data Sheet
Analog Meters
K Series
450.D.101.08
Maximum Demand Indicators
with Bimetallic Movement,
Combined M.D.I. and
Moving-Iron Ammeter
BIQ
BIEQ
BIQ
BIEQ
BIQ
48 K
72 K
72 K
96 K
96 K
with Slideā-āInā-āDial
Application
Electrical Data
Bimetallic maximum demand indicators monitor the most economic
use of transformer stations and L.T. distribution feeders by indicating
the thermal/time characteristics of the load.
The bimetallic movements are thermally inert. They indicate the mean
rms-value over 15 or 8 minute periods enabling to evaluate continuous
loads rather than short-time current peaks.
The high torque of the thermal movement offers the possibility to drive a
red slave pointer linked to the instrument pointer. Thereby, the highest
current reached in the circuit can be read off at any time. The slave
pointer is reset to the position of the indicator pointer by means of a
sealable reset knob.
Where the instantaneous and maximum demand currents are required,
the BIEQ 72/96 K instruments combine a thermal bimetallic and a
moving-iron movement installed diametrically in one case.
The maximum demand indicators are suitable to be installed in switchĆ
boards, mosaic grid panels (except model BIQ 48 K) or machine tool
consoles. The bezel, the glass window and the dial can easily be
exchanged on-site.
measuring unit
frequency range
power consumption
VA ratings
at 1 A rated current
at 5 A rated current
pollution level
enclosure code
case details
material of case
material of window
colour of bezel
position of use
panel fixing
mounting
terminals
dimensionsĂ(inĂmm)
bezel
case
mounting depth
panel cutout
panel thickness
weight approx.
dimensionsĂ(inĂmm)
bezel
case
mounting depth
panel cutout
panel thickness
weight approx.
square case suitable to be mounted in
control / switchgear panels or mosaic grid
panels (except model BIQ 48 K), stackable
polycarbonate thermoplastics,
self-extinguishing with UL rating of 94 V - 0
glass '
black (similar to RAL 9005) '
vertical ±5° '
swivel screw clamps
or plate springs on top and bottom (except
BIEQ 72 K ')
stackable next to each other
hexagon studs, M4 screws and
wire clamps E3
terminal safety protection '
BIQ 48 K
V 48 mm
V 45
48
V 45.2+0.3
1 ... 15
0.1 kg
BIQ 72 K
V 72
V 66
53
V 68+0.7
40
0.2 kg
BIEQ 72 K
V 72
V 66
53
V 68+0.7
40
0.2 kg
BIQ 96 K
V 96
V 90
60
V 92+0.8
40
0.26 kg
BIEQ 96 K
V 96
V 90
60
V 92+0.8
40
0.3 kg
BIEQ
72 K
<1.6
<2.7
96 K
<1.6
<3.4
2
IP 52 case
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact '
Measuring Ranges
measuring ranges AC current
bimetallic
moving iron
Mechanical Data
96 K
<1
<2.5
overload capacity (acc. to DIN EN 60 051)
continuously
1.2 times rated current
1 s max.
10 times rated current
Saturating current transformers shall be used to protect the movements
against overloads exceeding specified overload ratings.
measurement category CAT III
operating voltage
BIQ
BIEQ
48 K
72 K
96 K
72 K
96 K
600 V 600 V 150 V
150 V 150 V
Functional Principle
Bimetallic movement with resetable red slave pointer and a thermally
delayed indication enabling to measure the mean rms-value within a
time lag of 8 min or 15 min.
Moving iron movement with shell-type system, silicon oil damping,
pivot and spring-loaded jewel bearings (response time approx. 1 s).
AC current
50 ... 100 Hz
BIQ
48 K
72 K
<0.5 <1
<2.2 <2.5
0 ... 1 / 1.2 A
0 ... 1 / 2 A
or
0 ... 5 / 6 A
0 ... 5 / 10 A
for use on current transformer (scaling to DIN series)
0 ... N/5 / 6 A
bimetallic
0 ... N/1 / 1.2 A or
moving iron
0 ... N/1 / 2 A
0 ... N/5 / 10 A
(with overload scaling)
movements available BIQ
48 K
72 K
96 K
bimetallic 1 A
moving iron 1 A
-
-
-
bimetallic 5 A
moving iron 5 A
-
-
-
BIEQ
72 K
96 K
Scaling
pointer
pointer deflection
scale characteristics
overload scaling
scale division
scale length
bimetallic
moving-iron
thermal time delay '
bimetallic movem.
response time
moving iron movem.
bar / knife-edge pointer
0 ... 90°
bimetallic
moving-iron
quadratic;
practically linear;
scales are calibrated down to
1/ th rated current.
5
bimetallic '
moving-iron
1.2 times
2 times
rated current
rated current
coarse-fine
BIQ
BIEQ
48 K
72 K
96 K
72 K
96 K
44Ămm 62Ămm 98Ămm
44Ămm 71Ămm
62Ămm 98Ămm
BIQ
48 K
72 K
96 K
15Ămin 15Ămin 15Ămin
' also refer to "Options"
-
-
-
BIEQ
72 K
96 K
15Ămin 15Ămin
approx. 1 s
Data Sheet
K Series
450.D.101.08
Analog Meters
Maximum Demand Indicators
with Bimetallic Movement,
Combined M.D.I. and
Moving-Iron Ammeter
Accuracy at Reference Conditions
Options
accuracy class
3 (bimetallic movement
acc. to DIN EN 60 051
referred to slave pointer)
1.5 (moving-iron movement)
case
window
colour of bezel
position of use
marine application
panel fixing
reference conditions
ambient temperature
position of use
input
others
23°C±1K
nominal position ±1°
rated measuring value
DIN EN 60 051
influences
ambient temperature
position of use
stray magnetic field
-10°C ... +23°C ... +55°C
nominal position ±5°
0.5 mT
dial
blank dial
scale division
and figuring
additional lettering
additional figuring
coloured marks
coloured sector
overload scaling
bimetallic
logo on the dial
to be specified e.g. "generator"
to be specified
red, green or blue for important scale values
red, green or blue within scale division
no overload range
or overload range 1.5 times rated current
none or to be specified
-25 ... +65°C
others
calibration
thermal time delay
for a definite frequency 100 ... 1000 Hz
8 min or 15 min
75% annual average, non-condensing
15 g, 11 ms
2.5 g, 5 ... 55 Hz
terminal protection against accidental contact
full-sized rear cover (except model BIQ 48 K) or protective sleeves
safe against back-handed and fingertip contact acc. to VBGĂ4Ă/
DIN 57 106, sec. 100
Environmental
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
shock resistance
vibration resistance
climatic class 3 acc. to VDE/VDI 3540 sheet 2
-10 ... +55°C
Rules and Standards
DIN 43 700
DIN 43 701
DIN 43 718
DIN 43 802
DIN 16 257
non-glaring glass
gray (similar to RAL 7037)
to be specified 15°...165°
non-certified
plate springs for BIEQ 72 K on request
measuring and control instruments
for panel mounting;
nominal case and cutout dimensions
electrical switchboard instruments
bezels and front panels
scales and pointers for electrical measuring
instruments
nominal position of use and
position symbols
applicable for measuring instruments
DIN 40 050
enclosure codes;
protection of electrical equipment against
ingress of solid foreign bodies and of water
DIN EN 60 051
direct acting indicating electrical measuring
instruments and their accessories
DIN EN 61 010
safety requirements for electrically operated
measuring, control and laboratory
equipment
VDE/VDI 3540 sheet 2 reliability of measuring and control
equipment (classification of climates)
pencil-marked initial and end values
0 ... 100%
saturating current transformer
saturating transformer accuracy class 3, 50 Hz to protect the moveĆ
ments against overloads up to 100 times rated current (1 s max).
with base fixing attachment for panel projection mounting
ESW 1/5 A, 4.25 VA
ESW 5/5 A, 4.25 VA
Connections
Ordering Information
type
BIQ
direct-connected
17
18
BIEQ
front dimensions
48 K
72 K
96 K
measuring ranges
window
L1
N(L2)
for use on current transformer
17
L1
colour of bezel
18
position of use
k
l
K
L
marine application
dial
N(L2)
Dimensions
b
a
overload scaling
bimetallic
calibration
thermal time delay
c
dimensionsā(ināmm) BIQ 48 K
a
48
b
45
c
48
d
11
d
BIQ/BIEQĂ72ĂK BIQ/BIEQĂ96ĂK
72
96
66
90
53
60
11 (BIQ 72 K)
20
20 (BIEQ 72 K)
logo
terminal safety
protection
saturating current
transformer
*)
**)
1)
2)
maximum demand indicator with
bimetallic movement
combined M.D.I. & moving-ironāammeter
48 mm x 48 mm
72 mm x 72 mm
96 mm x 96 mm
refer to table inside
glass *)
non-glaring glass
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
vertical *)
to be specified 15 ... 165° **)
none *)
non-certified
scale division & measuring range alike
resp. acc. to DIN series if used on C.T. *)
no dial
blank dial
scale division and figuring 0 ... 100%
additional lettering to be specified **)
additional figuring to be specified **)
coloured marks red, green or blue **)
coloured sector red, green or blue **)
no overload range
1.2 times rated current *)
1.5 times rated current
50 Hz *)
for a definite frequency 100 ... 1000 Hz
8 min 1)
15 min 2)
WEIGEL *)
none
to be specified **)
none *)
full-sized rear cover
protective sleeves
none *)
ESW 1/5 A, 4.25 VA
ESW 5/5 A, 4.25 VA
standard
Please clearly add the desired specifications.
standard for models BIQ 48 K, BIEQ 72 K
standard for models BIQ 72/96 K, BIEQ 96 K
ordering example
BIEQ 96 K for use on current transformer 300/5 A,
thermal time delay 15 min, WEIGEL logo
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
webmaster@weigel-messgeraete.de
- specifications subject to change without notice; date of issue 10/06 -
Data Sheet
019.D.001.03
Analog Meters with
MovingāĆāCoil Movement
for use with
Thermocouples
PQ 72 RS
PQ 96 RS
PQ 144 RS
P 72 PrS
P 96 PrS
P 144 PrS
Application
terminals
The moving-coil panel meters PQ 72/96/144 RS (M series) and the
profile models P 72/96/144 PrS are used with thermocouples to meaĆ
sure and indicate temperature.
The moving-coil movements, manufactured to advanced engineering
standards, are characterized by a low power consumption, high preciĆ
sion and excellent damping.
The square and profile meter-models are housed in pressed steel
cases (except P 144 PrS with a thermoplastic case). They are suitable to
be mounted in switchboards, control panels, machine tool consoles or
mosaic grid panels.
hexagon studs, M3 screws and wire
clamps (PQ 72/96/144 RS, P 72/96 PrS), '
connector blades 6.3 x 0.8 (P 144 PrS)
dimensions
bezel
case
depth
panel cutout
panel thickness
weight approx.
PQā72āRS
V 72 mm
V 66 mm
60 mm
V68.3+0.4Ămm
1 ... 15 mm
0.5 kg
PQā96āRS
V 96 mm
V 90 mm
62 mm
V92+0.8Ămm
1 ... 15 mm
0.6 kg
PQā144āRS
V 144 mm
V 137 mm
60 mm
V138+1Ămm
1 ... 15 mm
0.9 kg
dimensions (in mm)
bezel
case
depth
panel cutout
P 72 PrS
72 x 36
67.5 x 32
94 mm
68+0.7Ămm
x 33+0.6Ămm
1 ... 25 mm
0.5 kg
P 96 PrS
96 x 48
90.5 x 42.5
107 mm
92+0.8Ămm
x 45+0.6Ămm
1 ... 12 mm
0.7 kg
P 144 PrS
144 x 72
137 x 67
192 mm
138+1.0Ămm
x 68+0.7Ămm
40 mm
1.3 kg
Functional Principle
Moving-coil movement comprising a core-type magnetic system,
pivot suspension. Dual spring loaded, shock absorbing jewel bearĆ
ings.
Functional Circuit Diagram
Thermocouple Compensation lead
Measurement
junction
Copper lead
Electrical Data
+
Connection point
Reference juncĆ
tion
Adjustment resistor
The resistance of the connecting lead may substantially influence the
temperature indication.
Consequently, the lead resistance Ra (=connecting lead resisĆ
tance+compensation lead resistance+thermocouple resistance) will
have to be considered in the calibration of the meter.
It shall be stated when ordering.
The internal resistance Ri of the meter plus the calibrated lead resisĆ
tance Ra is printed on the dial. The actual lead resistance will have to be
adjusted to the calibrated Ra-value.
Thermoelectric voltages up to 30 mV
External (PQ 72 RS, P 72/96 PrS) resp.
built-in (PQ 96/144 RS, P 144 PrS) amplifier.
The amplifier output current is not linearized, it rather follows the therĆ
moelectric voltage. Input and output are electrically not insulated.
input impedance
approx. 1 kW/mV
temperature influence <0,1%/°C
maximum lead resistance Ra=50 W requires no calibration,
over 50 W up to max. 200 W requires calibration of the meter.
Thermoelectric voltages higher than 30 mV
direct-connected, requires no amplifier.
A total lead resistance value Ra=2 W for the thermocouple, copper lead
and compensation lead is considered in the calibration of the indicator.
If possible, non-standard resistance values may be considered (up to
10 W maximum).
Mechanical Data
case details
material of case
material of window
colour of bezel
position of use
panel fixing
mounting
panel thickness
weight approx.
square (PQ 72/96/144 RS) resp.
edgewise (P 72/96/144 PrS) case
suitable to be mounted in switchboards
or mosaic grid panels
pressed steel (PQ 72/96/144 RS, P 72/96 PrS)
self-extinguishingthermoplastics(P144PrS)
glass '
black (similar to RAL 9005) '
vertical ±5° '
screw clamps
stackable next to each other
(except P 144 PrS)
measuring unit
thermoelectric voltage (DC voltage)
measurement category CAT III
operating voltage
PQā72/96/144āRS Pā72/96āPrS Pā144āPrS
150 V
50 V
150 V
pollution level
2
enclosure code
IP 52 case '
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact
Measuring Ranges
Thermoelectric voltages
from 11.5 mV upwards
for thermocouples acc. to IEC 584 resp. DIN 43 710
measuring range sensor
type
thermoelectric voltage
0 ... 250°C Fe - CuNi
L
13.75 mV
0 ... 400°C Fe - CuNi
L
22.16 mV
0 ... 600°C Fe - CuNi
L
33.67 mV
0 ... 800°C Fe - CuNi
L
46.22 mV
0 ... 900°C Fe - CuNi
L
53.14 mV
0 ... 250°C Fe - CuNi
J
13.55 mV
0 ... 400°C Fe - CuNi
J
21.85 mV
0 ... 600°C Fe - CuNi
J
33.10 mV
0 ... 800°C Fe - CuNi
J
45.50 mV
0 ... 900°C Fe - CuNi
J
51.88 mV
0 ... 600°C NiCr - Ni
K
24.90 mV
0 ... 900°C NiCr - Ni
K
37.33 mV
0 ... 1000°C NiCr - Ni
K
41.27 mV
0 ... 1200°C NiCr - Ni
K
48.83 mV
0 ... 1300°C NiCr - Ni
K
52.40 mV
0 ... 1200°C Pt10Rh - Pt
S
11.95 mV
0 ... 1400°C Pt10Rh - Pt
S
14.37 mV
0 ... 1600°C Pt10Rh - Pt
S
16.77 mV
reference temperature 0°C
In case of external reference junction,
state reference temperature 0°C, 20°C or 50°C.
' also refer to "Options"
Data Sheet
019.D.001.03
Analog Meters with
MovingāĆāCoil Movement
for use with
Thermocouples
Scaling
pointer
scale arrangement
scale characteristics
scale division
scale length
DIN 40 050
bar / knife-edge pointer
horizontal, left-hand zero (P 72/96/144 PrS) '
practically linear
coarse-fine
PQā72āRS
PQā96āRS
PQā144āRS
69 mm
94 mm
146 mm
P 72 PrS
P 96 PrS
P 144 PrS
46 mm
67 mm
92 mm
Auxiliary Supply
for thermoelectric voltages up to 30 mV
auxiliary voltage
AC 230 V -20 ... +5%, 48 ... 62 Hz
power consumption
approx. 3 VA
Electrical insulation between measuring circuit and
auxiliary supply circuit.
Accuracy at Reference Conditions
accuracy class
1.5 according to DIN EN 60 051
reference conditions
ambient temperature
position of use
input
auxiliary voltage
others
23°C±1K
nominal position ±1° '
rated measuring value
within specified limits
DIN EN 60 051
influences
ambient temperature
position of use
stray magnetic field
-25°C ... +23°C ... +40°C
nominal position ±5°
0.5 mT
Environmental
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
shock resistance
vibration resistance
climatic class 2 '
according to VDE/VDI 3540, sheet 2
-25 ... +40°C '
-25 ... +65°C
75% annual average, non-condensing
15 g, 11 ms '
2.5 g, 5 ... 55 Hz '
Rules and Standards
DIN 43 700
DIN 43 701
DIN 43 718
DIN 16 257
DIN EN 60 051
DIN EN 61 010
measuring and control instruments
for panel mounting;
nominal case and cutout dimensions
electrical switchboard instruments
bezels and front panels
nominal position of use and
position symbols
applicable for measuring instruments
direct acting indicating electrical measuring
instruments and their accessories
safety requirements for electrically operated
measuring, control and laboratory
equipment
enclosure codes;
protection of electrical equipment against
ingress of solid foreign bodies and of water
DIN 43 710
Electrical Temperature Sensors
DIN IEC 584-1
Thermocouples; deviation limits of
thermoelectric voltages
VDE/VDI 3540 sheet 2 reliability of measuring and control
equipment (classification of climates)
Options
case
window
colour of bezel
position of use
non-glaring glass
gray (similar to RAL 7037)
horizontal or to be specified 15°...165°
performance
increased mechanical shock 30 g, 11 ms
loads
vibration 5 g, 5 ... 55 Hz
climatic suitability
limited use in the tropics, climatic class 3
according to VDE/VDI 3540, sheet 2
with operating
-10 ... +55°C
temperature range
marine application
non-certified
enclosure code
IP 54 splash-water protected front
accessories
terminal protection against accidental contact
full-sized rear cover (PQ 72/96/144 RS only) or protective sleeves SW6
terminals
connector blades 6.3 x 0.8
dial
scale arrangement
vertical, bottom zero (P 72/96/144 PrS)
blank dial
pencil marked initial and end values
scale division and figuring 0 ... 100%
additional lettering
to be specified e.g. "generator"
additional figuring
to be specified
coloured marks
red, green or blue for important scale values
coloured sector
red, green or blue within scale division
logo on the dial
none or to be specified
dial illumination by one (PQ 72/96 RS) resp. two (PQ 144 RS) lamps
6 V, 12 V or 24 V to be installed from the rear, dial translucent
Connections
up to 30 mV, external amplifier (PQ 72 RS, P 72/96 PrS)
L1
N
auxiliary voltage
2
5
external amplifier
10 11
+
17
-
+
18
-
thermocouple
up to 30 mV, built-in amplifier (PQ 96/144 RS, P 144 PrS)
thermocouple
+
17
-
18
L1
N
auxiliary
voltage
2
5
higher than 30 mV, direct-connected, no amplifier required
+ 17
- 18
Dimensions
Ordering Information
a
b
PQā72/96/144āRS
c
dimensions (in mm)
a
b
c
PQā72āRS
72
66
60
PQā96āRS
96
90
62
PQā144āRS
144
137
60
P 72/96 PrS
colour of bezel
d
b
type
PQ
front dimensions
72 RS
96 RS
144 RS
type
P
front dimensions
72 PrS
96 PrS
144 PrS
measuring ranges
window
c
a
position of use
P 144 PrS
d
b
performance loads
climatic suitability
c
a
dimensions (in mm)
a
b
c
d
P 72 PrS
72
36
94
32
external amplifier
P 144 PrS
144
72
192
67
1
10
terminal safety
protection
scale arrangement
(P 72/96/144 PrS)
dial
70
25
enclosure code
terminals
4.8
30
case for
clamping
to DIN rail
P 96 PrS
96
48
107
43
marine application
18
10
60
110
scales and position of use (P 72/96/144 PrS)
II
I
a
a
III
IV
VI
V
vertical mounting
I
horizontal scale (standard)
II vertical scale
inclined mounting
III horizontal scale
IV vertical scale
horizontal mounting
V horizontal scale
VI vertical scale
ordering example
P 72 PrS, measuring range 0 ... 600°C for use with thermocouple
Fe-CuNi, type L, 33.67 mV, reference temperature 0°C, lead resisĆ
tance Ra=2 W, horizontal scale 0 ... 600°C, vertical position of use
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
logo
dial illumination
moving-coil panel meter
to measure thermoelectric voltages
72 mm x 72 mm
96 mm x 96 mm
144 mm x 144 mm
profile moving-coil panel meter
to measure thermoelectric voltages
72 mm x 36 mm
96 mm x 48 mm
144 mm x 72 mm
refer to preceding table
glass *)
non-glaring glass
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
vertical *)
horizontal
to be specified 15 ... 165° **)
shock 15 g, vibration 2.5 g *)
shock 30 g, vibration 5 g
class 2, -25 ... +40°C *)
class 3, -10 ... +55°C
none *)
non-certified
IP 52 *)
IP 54 splash-water protected front
none *)
full-sized rear cover ***)
protective sleeves
screws and wire clamps *)
connector blades 6.3 x 0.8 (P 144 PrS *) )
horizontal, left-hand zero *)
vertical, bottom zero
°C (DIN range) *)
blank dial
scale division and figuring 0 ... 100%
additional lettering to be specified **)
additional figuring to be specified **)
coloured marks red, green or blue **)
coloured sector red, green or blue **)
WEIGEL *)
none
OEM logo **)
none
with 1 lamp 6, 12 or 24 V
(models PQ 72/96 RS only)
with 2 lamps 6, 12 or 24 V
(model PQ 144 RS only)
*) standard
**) Please clearly add the
***) PQ 72/96/144 only
desired specifications.
- specifications subject to change without notice; date of issue 04/07 -
Data Sheet
018.D.001.03
Analog Meters with
MovingāĆāCoil Movement
arranged in a
Bridge Circuit
PBQ
72
PBQ
96
PBQ 144
PB 72 PrS
PB 96 PrS
PB 144 PrS
Application
pollution level
enclosure code
The moving-coil panel meters PBQ 72/96/144 (M series) in pressed
steel cases as well as PB 72/96/144 PrS (edgewise series) have two
main fields of application. They are used with RTD (resistance therĆ
mometers) Pt or Ni according to DIN 43 760 to measure and indicate
temperature.
If used with resistance sensors they indicate position, e.g. transformer
tap position, hoist or valve position, transformer winding temperature
and any similar function where the position can be related to the moveĆ
ment of a potentiometer.
The indicators are suitable to be mounted in switchboards, control panĆ
els or mosaic grid panels.
Functional Principle
Self-shielding moving-coil movement with a core-type magnet,
pivot suspended. Spring loaded jewel bearings for vibration and shock
resistance.
The measuring system comprises a stabilized power source, a movĆ
ing-coil indicator arranged in a bridge circuit and a remotely mounted
potentiometer resp. a RTD (resistance thermometer) both to be supĆ
plied by customer. (Separate bridge circuit for PB 72 PrS model.)
Mechanical Data
case details
material of case
material of window
colour of bezel
position of use
panel fixing
mounting
terminals
square (PBQ 72/96/144) resp.
edgewise (PB 72/96/144 PrS) case
suitable to be mounted in switchboards
or mosaic grid panels
pressed steel (PBQ 72/96/144, PB 72/96 PrS)
self-extinguishing thermoplastics
(PB 144 PrS)
glass '
black (similar to RAL 9005) '
vertical ±5° '
screw clamps
stackable next to each other
(except PB 144 PrS)
hexagon studs, M3 screws and wire
clamps C6 (PBQ 72/96/144, PB 72/96 PrS), '
connector blades 6.3 x 0.8 (PB 144 PrS)
dimensions
bezel
case
depth
panel cutout
panel thickness
weight approx.
PBQā72
V 72 mm
V 66 mm
60 mm
V68.3+0.4Ămm
1 ... 15 mm
0.3 kg
PBQā96
V 96 mm
V 90 mm
62 mm
V92+0.8Ămm
1 ... 15 mm
0.4 kg
PBQā144
V 144 mm
V 137 mm
60 mm
V138+1Ămm
1 ... 15 mm
0.7 kg
dimensions (in mm)
bezel
case
depth
panel cutout
PB 72 PrS
72 x 36
67.5 x 32
91 mm
68+0.7Ămm
x 33+0.6Ămm
1 ... 25 mm
0.2 kg
PB 96 PrS
96 x 48
90.5 x 42.5
100 mm
92+0.8Ămm
x 45+0.6Ămm
1 ... 12 mm
0.45 kg
PB 144 PrS
144 x 72
137 x 67
180 mm
138+1.0Ămm
x 68+0.7Ămm
40 mm
0.7 kg
panel thickness
weight approx.
Electrical Data
measuring unit
resistance (DC)
measurement category CAT III
operating voltage
PBQ 72/96/144
150 V
PB72/96 PrS
50 V
PB 144 PrS
100 V
2
IP 52 case '
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact
Measuring Ranges
for RTD (resistance thermometer) acc. to DIN 43 760
measuring range
sensor
-220 ... +50°C
-100 ... +50°C
-20 ... +20°C
0 ... +40°C
-30 ... +60°C
0 ... +60°C
0 ... +100°C
0 ... +150°C
+50 ... +150°C
0 ... +200°C
0 ... +300°C
0 ... +400°C
0 ... +550°C
+200 ... +400°C
+300 ... +550°C
Pt 100
Pt 100
Pt 100,
Pt 100,
Pt 100,
Pt 100,
Pt 100,
Pt 100,
Pt 100,
Pt 100,
Pt 100
Pt 100
Pt 100
Pt 100
Pt 100
Ni 100
Ni 100
Ni 100
Ni 100
Ni 100
Ni 100
Ni 100
Ni 100
-
for resistance sensors
please state
when ordering:
- measuring range and scaling
- auxiliary voltage
- total resistance of sensor
- variation range of sensor
- maximum lead resistance
(standard 2x 10 W)
Note
Indication of the meter is influenced by the lead resistance. ConseĆ
quently, the lead resistance will have to be considered in the calibration
of the meter. It should be stated when ordering.
The lead resistance will be calibrated to 10 W for RTD (resistance therĆ
mometer) in a 2 wire system, to 2x 10 W for RTD in a 3 wire system and
to 2x 10 W for resistance sensors. If possible, varying resistance values
may be considered.
The lead resistance calibrated is printed on the dial. The actual resisĆ
tance will have to be adjusted to this value.
Scaling
pointer
dial position
scale characteristics
scale division
scale length
bar / knife-edge pointer
horizontal dial (PB 72/96/144 PrS) '
linear
coarse-fine
PBQā72
PBQā96
PBQā144
69 mm
94 mm
146 mm
PB 72 PrS
PB 96 PrS
PB 144 PrS
46 mm
67 mm
92 mm
Auxiliary Supply
auxiliary voltage
DC 24 V ±10% '
current consumption approx. 40 mA
residual ripple
<3%
Measuring input and auxiliary supply are not electrically insulated.
' also refer to "Options"
Data Sheet
018.D.001.03
Analog Meters with
MovingāĆāCoil Movement
arranged in a
Bridge Circuit
Accuracy at Reference Conditions
marine application
enclosure code
accuracy class
1.5 according to DIN EN 60 051
reference conditions
ambient temperature
position of use
input
auxiliary voltage
others
23°C±1K
nominal position ±1° '
rated measuring value
within the limits specified
DIN EN 60 051
accessories
terminal protection against accidental contact
full-sized rear cover (PBQ 72/96/144 only) or protective sleeves
terminals
connector blades 6.3 x 0.8
influences
ambient temperature
position of use
stray magnetic field
-25°C ... +23°C ... +40°C
nominal position ±5°
0.5 mT
Environmental
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
shock resistance
vibration resistance
climatic class 2 '
according to VDE/VDI 3540, sheet 2
-25 ... +40°C '
-25 ... +65°C
75% annual average, non-condensing
15 g, 11 ms '
2.5 g, 5 ... 55 Hz '
Rules and Standards
DIN 43 700
measuring and control instruments
for panel mounting;
nominal case and cutout dimensions
DIN 43 701
electrical switchboard instruments
DIN 43 718
bezels and front panels
DIN 16 257
nominal position of use and position symbols
applicable for measuring instruments
DIN 40 050
enclosure codes;
protection of electrical equipment against
ingress of solid foreign bodies and of water
DIN EN 60 051
direct acting indicating electrical measuring
instruments and their accessories
DIN EN 61 010
safety requirements for electrically operated
measuring, control and laboratory
equipment
VDE/VDI 3540, sheet 2 reliability of measuring and control
equipment (classification of climates)
Options
auxiliary voltage
case
window
colour of bezel
position of use
AC 230 V -20% ... +5%, 48 ... 62 Hz
electrically insulated
applying to PBQ 96/144, PB 144 PrS only
non-glaring glass
gray (similar to RAL 7037)
horizontal or to be specified 15°...165°
performance
increased mechanical shock 30 g, 11 ms
loads
vibration 5 g, 5 ... 55 Hz
climatic suitability
limited use in the tropics, climatic class 3
according to VDE/VDI 3540, sheet 2
with operating
-10 ... +55°C
temperature range
' also refer to "Options"
non-certified
IP 54 splash-water protected front
dial
dial position
blank dial
scale division
and figuring
vertical dial (PB 72/96/144 PrS)
pencil marked initial and end values
0 ... 100%,
linear, full-scale values acc. to standardized
series (1 Ć 1.2 Ć 1.5 Ć 2 Ć 2.5 Ć 3 Ć 4 Ć 5 Ć 6 Ć 7.5
and their decimal multiples e.g. 150 m3/h) or
deviating from standard;
captions optional
to be specified e.g. "generator"
to be specified
red, green or blue for important scale values
red, green or blue within scale division
none or to be specified
additional lettering
additional figuring
coloured marks
coloured sector
logo on the dial
Attachment
Power Supply
please refer to accessories data sheets
Lead Adjustment Resistor 10 W coil-type with soldering tags
Test Resistor for RTD (resistance thermometer) to adjust the
measuring circuit.
Connections
RTD (resistance thermometer)
in a 2 wire system
in a 3 wire system
Resistance Sensor
A
E
S
AW1
AW1
AW2
AW2
3
1
4
5
4
5
+
~
~
+
~
~
1
3
1
4
5
+
~
~
auxiliary voltage
AW1
2
auxiliary voltage
2
3
auxiliary voltage
Dimensions
type
PB
front dimensions
72 PrS
96 PrS
144 PrS
measuring ranges
wiring
a
b
PBQā72/96/144āRS
auxiliary voltage
c
dimensions (in mm)
a
b
c
PBQā72
72
66
60
PBQā96
96
90
62
PBQā144
144
137
60
PB 72/96 PrS
window
colour of bezel
b
d
position of use
performance loads
c
a
climatic suitability
PB 144 PrS
b
d
marine application
enclosure code
c
a
e
dimensions
(in mm) PB 72 PrS PB 96 PrS PB 144 PrS
a
72
96
144
b
36
48
72
c
91
100
180
d
32
43
67
e
67.5
90.5
137
terminal safety
protection
terminals
dial position
(PB 72/96/144 PrS)
dial
scales and position of use (PB 72/96/144 PrS)
II
I
a
a
III
IV
VI
V
vertical mounting
I
horizontal scale (standard)
II vertical scale
inclined mounting
III horizontal scale
IV vertical scale
horizontal mounting
V horizontal scale
VI vertical scale
Ordering Information
type
PBQ
front dimensions
72
96
144
square moving-coil panel meter
arranged in a bridge circuit
72 mm x 72 mm
96 mm x 96 mm
144 mm x 144 mm
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
logo
profile moving-coil panel meter
arranged in a bridge circuit
72 mm x 36 mm
96 mm x 48 mm
144 mm x 72 mm
refer to preceding table
RTD in 2 wire system
RTD in 3 wire system
resistance sensor
DC 24 V *)
AC 230 V
glass *)
non-glaring glass
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
vertical *)
horizontal
to special order 15 ... 165° **)
shock 15 g, vibration 2.5 g *)
shock 30 g, vibration 5 g
class 2, -25 ... +40°C *)
class 3, -10 ... +55°C
none *)
non-certified
IP 52 *)
IP 54 splash-water protected front
none *)
full-sized rear cover ***)
protective sleeves
screws and wire clamps *)
connector blades 6.3 x 0.8 (PB 144 PrS *) )
horizontal dial *)
vertical dial
scale division and measuring range alike *)
blank dial
scale division and figuring
0 ... 100%
linear to standardized series **)
linear deviating from standard **)
additional lettering to be specified **)
additional figuring to be specified **)
coloured marks red, green or blue **)
coloured sector red, green or blue **)
WEIGEL *)
none
OEM logo **)
*) standard
**) Please clearly add the
***) PBQ 72/96/144 only
desired specifications.
ordering example
PB 72 PrS, measuring range -30 ... 60°C on Pt 100, 3 wire system,
auxiliary voltage DC 24 V, horizontal scale -30 ... 60°C,
vertical mounting, window non-glaring glass, WEIGEL logo
- specifications subject to change without notice; date of issue 04/07 -
Data Sheet
M Series
017.D.101.03
Analog Meters with
Dual MovingāĆāCoil
Movement
PQ 48 /2
Application
Measuring Ranges
The square moving-coil instruments PQ 48 /2 (M series) having two
meter movements are used for process control measurement of DC
currents and/or DC voltages. They are scaled to determine control deviĆ
ations and regulating link positions in control systems.
Movement I (outer)
control deviation internal resistance*) pointer deflection
The instruments are suitable to be mounted in switchboards, control
panels, machine tool consoles or mosaic grid panels.
Movements
Self-shieldingmoving-coilmovementswithcore-typemagneticsysĆ
tems, pivot suspended. Twin spring loaded jewel bearings for vibration
and shock resistance.
Mechanical Data
case details
square case suitable to be mounted
in control / switchgear panels or
mosaic grid panels, stackable
material of case
thermoplastics, flame retardant
material of window
glass '
colour of bezel
black (similar to RAL 9005) '
position of use
any position permissible
panel fixing
screw clamps '
panel thickness
1 ... 40 mm
mounting
stackable next to each other
terminals
connector blades 6.3 x 0.8
dimensions
bezel
case
depth
panel cutout
weight approx.
V 48 mm
V 41 mm
72 mm
V 45.0+0.6 mm
0.09 kg
Electrical Data
measuring unit
measurement category
operating voltage
pollution level
enclosure code
-20 ... 0 ... +20 mA
6 kW
±22.5° -10ā...ā0ā...ā+10
-20 ... 0 ... +20 mA
6 kW
±22.5° -20ā...ā0ā...ā+20
-20 ... 0 ... +20 mA
13 kW
±22.5° -10ā...ā0ā...ā+10
-20 ... 0 ... +20 mA
13 kW
±22.5° -20ā...ā0ā...ā+20
-20 ... 0 ... +20 mA
50 kW
±22.5° -10ā...ā0ā...ā+10
-20 ... 0 ... +20 mA
50 kW
±22.5° -20ā...ā0ā...ā+20
-50 ... 0 ... +50 mA
2 kW
±22.5° -10ā...ā0ā...ā+10
-50 ... 0 ... +50 mA
2 kW
±22.5° -20ā...ā0ā...ā+20
-300ā...ā0ā...ā+300ĂmA
325 W
±22.5° -10ā...ā0ā...ā+10
-300ā...ā0ā...ā+300ĂmA
325 W
±22.5° -20ā...ā0ā...ā+20
regulating link
internal resistance*) pointer deflection scale '
position
0 ... 600 mA
0 ... 20 mA '
0 ...
3V
0 ...
10 V
CAT III
100 V
2
IP 52 case '
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact
' also refer to "Options"
325 W
3W
30 kW
100 kW
90°
90°
90°
90°
0 ... 100%
0 ... 100%
0 ... 100%
0 ... 100%
Movement II (inner)
regulating link
internal resistance*) pointer deflection
position
scale '
0 ... 600 mA
325 W
90°
0 ... 100%
0 ... 20 mA '
3W
90°
0 ... 100%
0 ...
3V
30 kW
90°
0 ... 100%
0 ...
10 V
100 kW
90°
0 ... 100%
*) the resistance values are limited to a tolerance of ±20%
Scaling
pointers
pointer deflection
scale characteristics
scale division
scale length
DC voltages or DC currents
overload capacity (acc. to DIN EN 60 051)
continuously
1.2 times rated voltage / current
5 s max.
voltmeters
2 times rated voltage
ammeters
10 times rated current
scale '
bar / knife-edge pointers
control deviation regulating link position
±22.5°
0 ... 90°
linear
coarse-fine
control deviation regulating link position
movement I
movement I movement II
15.5 mm
31 mm
28 mm
Accuracy at Reference Conditions
accuracy class
1.5 according to DIN EN 60 051
reference conditions
ambient temperature 23°C±1K
input
rated measuring value
others
DIN EN 60 051
influences
ambient temperature
-25°C ... +23°C ... +40°C
Data Sheet
M Series
017.D.101.03
Analog Meters with
Dual MovingāĆāCoil
Movement
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
shock resistance
vibration resistance
climatic class 2 '
according to VDE/VDI 3540 sheet 2
-25 ... +40°C '
-25 ... +65°C
75% annual average, non-condensing
15 g, 11 ms '
2.5 g, 5 ... 55 Hz '
Rules and Standards
DIN 43 700
measuring and control instruments
for panel mounting;
nominal case and cutout dimensions
DIN 43 701
electrical switchboard instruments
DIN 43 718
bezels and front panels
DIN 16 257
nominal position of use and
position symbols
applicable for measuring instruments
DIN 40 050
enclosure codes;
protection of electrical equipment against
ingress of solid foreign bodies and of water
DIN EN 60 051
direct acting indicating electrical measuring
instruments and their accessories
DIN EN 61 010
safety requirements for electrically operated
measuring, control and laboratory
equipment
VDE/VDI 3540 sheet 2 reliability of measuring and control
equipment (classification of climates)
dial
blank dial
scale division
and figuring
additional lettering
additional figuring
coloured marks
coloured sector
logo on the dial
Connections
+
movement I
movement II
+
Dimensions
case
window
colour of bezel
panel fixing
48
Options
measuring range
special measuring
range
zero position
pencil marked initial and end values
0 ... 100%,
linear, full-scale values acc. to standardized
series (1 - 1.5 - 2.5 - 4 - 6 and any decimal
multiple of these numbers e.g. 150 m3/h) or
deviating from standard series,
captions optional
to be specified e.g. "generator"
to be specified
red, green or blue for important scale values
red, green or blue within scale division
none or to be specified
41
Environmental
deviating from ranges as specified in table
3.5
suppressed for range 4 ... 20 mA
mechanically suppressed zero
non-glaring glass
gray (similar to RAL 7037)
plate springs
performance
increased mechanical shock 30 g, 11 ms
loads
vibration 5 g, 5 ... 55 Hz
climatic suitability
limited use in the tropics climatic class 3
according to VDE/VDI 3540 sheet 2
with operating
-10 ... +55°C
temperature range
marine application
non-certified
enclosure code
IP 54 splash-water protected front
accessories
terminal protection against accidental contact
protective sleeves
63
72
(scaled in mm)
5.5
Ordering Information
type
PQ
front dimension
48
type identification
/2
measuring ranges
special
measuring range
zero position
window
colour of bezel
panel fixing
mechanical loads
climatic suitability
marine application
enclosure code
terminal safety
protection
dial
logo
*)
**)
moving-coil instrument
for DC voltage and/or DC current
48 mm x 48 mm
2 movements
refer to preceding table
on special order **)
electrical zero = mechanical zero *)
mechanically suppressed zero
for range 4 ... 20 mA
glass *)
non-glaring glass
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
screw clamps *)
plate springs
shock 15 g, vibration 2.5 g *)
shock 30 g, vibration 5 g
class 2, -25 ... +40°C *)
class 3, -10 ... +55°C
none *)
non-certified
IP 52 *)
IP 54 splash-water protected front
none *)
protective sleeves
refer to table inside *)
blank dial
scale division and figuring
0 ... 100%
linear acc. to standardized series
linear deviating from standard **)
additional lettering to be specified **)
additional figuring to be specified **)
coloured marks red, green or blue **)
coloured sector red, green or blue **)
WEIGEL *)
none
OEM logo **)
standard
Please clearly add the desired specifications.
ordering example
PQ 48 /2, measuring range movement I: -20 ... 0 ...+20 mA,
internal resistance 13 kW, scale -20 ... 0 ...+20,
movement II: 4 ... 20 mA, scale 0 ... 100 %,
panel fixing by plate springs, no logo on the dial
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
- specifications subject to change without notice; date of issue 12/06 -
Data Sheet
Edgewise Series
015.D.301.06
Analog Meters
Slim Edgewise with
MovingĆCoil Movement
and Rectifier
MG 48x24
MG 72x24
MG 96x24 K
G 144x36
MG 96x24 K
with SlideāĆāInāĆāDial
Application
Measuring Ranges
The edgewise movingāĆācoil panel meters MG 48x24, MG 72x24,
MG 96x24 K and G 144x36 with a slimāĆāline dial are used for measureĆ
ment of sinusoidal AC currents or voltages.
MovingāĆācoil rectifier instruments measure average values and are
scaled to indicate r.m.s., assuming a sinusoidal wave form.
AC current
The movingāĆācoil movement is manufactured to newest findings and
distinguishes in a small power consumption, a high accuracy and a
very good damping.
The instruments are suitable to be mounted in switchboards, control
panels, machine tool consoles and mosaic panels (except G 144x36).
With MG 96x24 K the front panel, the front window and the dial as well
as the dial illumination (optional) can be exchanged easily. On the rear
side a cable grip for the connecting leads is integrated.
100 mA, 150 mA, 250 mA, 400 mA, 600 mA, 1 mA, 1.5 mA, 2.5 mA, 4 mA,
6 mA, 10 mA, 15 mA, 25 mA, 40 mA, 60 mA, 100 mA, 150 mA,
250 mA, 400 mA, 600 mA (voltage drop approx. 1.5 V)
with separate miniature current transformer 50 Hz, 10 mA sec.
1 A, 1.5 A, 2.5 A (voltage drop approx. 0.2 V)
4 A, 5 A, 6 A, 10 A, 15 A, 25 A (voltage drop approx. 0.3 V)
for use on current transformer (scale without overload range)
separate miniature current transformer 50 Hz, 10 mA sec. inclusive
N/1 A (voltage drop approx. 0.2 V)
N/5 A (voltage drop approx. 0.3 V)
operating
MG 48x24 MG 72x24 MG 96x24 K G 144x36
voltage
300 V
600 V
300 V
300 V
Movements
AC voltage
SelfāĆāshielding movingāĆācoil movements with swivel coil, pivot susĆ
pended. SeriesāĆāconnected rectifier incorporated. Spring loaded jewel
bearings for vibration and shock resistance.
Mechanical Data
case details
dimensions (in mm)
bezel
edgewise case suitable to be mounted
in control panels, machine tool consoles
or mosaicpanels(exceptG144x36),stackable
thermoplastics
glass '
black (similar to RAL 9005) '
vertical ±5° '
screw clamps '
stackableānextātoāeachāother
1 ... 25 mm
connector blades 6.3 x 0.8
or connector blades 2.8 x 0.8 (MG 96x24 K)
MGĂ48x24 MGĂ72x24 MGĂ96x24K GĂ144x36
48 x 24
72 x 24
96 x 24
144 x 36
case
43 x 17
66 x 17
92 x 18
137 x 32
depth
75
98
108
173
panel cutout
68+0.7
92+0.8
45+0.6
x 22.2+0.3 x 22.2+0.3 x 22.2+0.3
weight approx.
0.08 kg
material of case
material of window
colour of bezel
position of use
panel fixing
mounting
panel thickness
terminals
0.1 kg
0.12 kg
138+1.0
x 33+0.6
0.5 kg
Electrical Data
measuring unit
frequency range
AC voltage or AC current
40 Hz ... 50 Hz ... 10 kHz
overload capacity (acc. to DIN EN 60 051)
continuously
1.2 times rated voltage / current
5 s max. voltmeters 2 times rated voltage
ammeters 10 times rated current
measurement category CAT III
operating voltage
refer to Measuring Ranges
pollution level
2
enclosure code
IP 52 case '
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact
' for other ratings refer to "Options"
operating voltage
MG 48x24 MG 72x24 MG 96x24 K G 144x36
2.5 V
50 V
100 V
300 V
100 V
4V
50 V
100 V
300 V
100 V
6V
50 V
100 V
300 V
100 V
10 V
50 V
100 V
300 V
100 V
15 V
50 V
100 V
300 V
100 V
25 V
50 V
100 V
300 V
100 V
40 V
50 V
100 V
300 V
100 V
60 V
300 V
100 V
300 V
100 V
100 V
300 V
100 V
300 V
100 V
150 V
300 V
600 V
300 V
300 V
250 V
300 V
600 V
300 V
300 V
400 V
300 V
600 V
300 V
300 V
500 V
300 V
600 V
300 V
300 V
600 V
600 V
600 V
600 V
(600 V in MG 48x24 on request)
sensitivity 900 W/V '
the resistance values are limited to a tolerance of ±20% '
Scaling
pointer
zero adjustment
response time
scale arrangement
scale characteristics
scale division
scale length
bar pointer
front accessible '
1 s for full-scale deflection
horizontal (left-hand zero) '
linear
coarse-fine
MG 48x24 MG 72x24 MG 96x24 K G 144x36
30 mm
52 mm
60 mm
95 mm
Data Sheet
Edgewise Series
015.D.301.06
Analog Meters
Slim Edgewise with
MovingĆCoil Movement
and Rectifier
Accuracy at Reference Conditions
accuracy class
1.5 according to DIN EN 60 051
reference conditions
ambient temperature
position of use
input
frequency
wave form
others
23°C±1K
nominal position ±1°
rated measuring value
50±2 Hz
sinusoidal, distortion factor <5%
DIN EN 60 051
influences
ambient temperature
position of use
frequency
stray magnetic field
-25°C ... +23°C ... +40°C
nominal position ±5°
40 Hz ... 45 ... 60 Hz ... 10 kHz
0.5 mT
Environmental
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
shock resistance
vibration resistance
climatic class 2 '
according to VDE/VDI 3540 sheet 2
-25 ... +40°C '
position of use
mounting
(MG 96 x24 K)
performance
climatic suitability
with operating
temperature range
increased mechanical
loads
marine application
enclosure code
75% annual average, non-condensing
15 g, 11 ms '
2.5 g, 5 ... 55 Hz '
DIN 43 700
measuring and control instruments
for panel mounting;
nominal case and cutout dimensions
DIN 43 701
electrical switchboard instruments
DIN 43 718
bezels and front panels
DIN 16 257
nominal position of use and
position symbols
applicable for measuring instruments
DIN 40 050
enclosure codes;
protection of electrical equipment against
ingress of solid foreign bodies and of water
DIN EN 60 051
direct acting indicating electrical measuring
instruments and their accessories
DIN EN 61 010
safety requirements for electrically operated
measuring, control and laboratory
equipment
VDE/VDI 3540 sheet 2 reliability of measuring and control
equipment (classification of climates)
dial
scale arrangement
blank dial
scale division
and figuring
case
window
colour of bezel
adjustment to ±1% at 23°C
for voltmeters
to 2 kW/V, 5 kW/V, 10 kW/V or 20 kW/V
(as far as possible)
non-glaring glass
gray (similar to RAL 7037)
shock 30 g, 11 ms
vibration 5 g, 5 ... 55 Hz
non-certified
IP 54 splash-water protected front
(without zero adjustment or with zero
adjustment rear accessible for MG 96x24 K)
vertical (bottom zero)
pencil-marked on initial and end values
0 ... 100%,
linear, full-scale values acc. to standardized
series (1 Ć 1.2 Ć 1.5 Ć 2 Ć 2.5 Ć 3 Ć 4 Ć 5 Ć 6 Ć 7.5
and their decimal multiples e.g. 150 m3/h) or
deviating from standard;
captions on request
including figuring (on request)
scale in black; pointer, scale division and
lettering yellow or white
on request e.g. "generator"
on request
red, green or blue for important scale values
red, green or blue within scale division
none or on request
mechanically suppressed zero, no zero
adjustment, max. 40% of full-scale value
for ammeters 100 mA, voltmeters 1,5 V
rear accessible
2nd scale division
coloured scale
(MG 96x24 K only)
additional lettering
additional figuring
coloured marks
coloured segment
logo on the dial
zero position
zero adjustment
(MG 96x24 K only)
expanded scale
expanded initial scale division
by means of electronic circuits up to approx.
5% of full-scale value in centre of scale
for MG 72x24 / 96x24 K, P 144x36
by 2 lamps 6 V, 12 V or 24 V
dial translucent
dial illumination
(MG 96x24 K only)
Connections
AC voltage
Options
deviating from standard
limited use in the tropics climatic class 3
according to VDE/VDI 3540 sheet 2
-10 ... +55°C
terminal protection against accidental contact
protective sleeves B6 for connector blades 6.3 x 0.8
protective sleeves 110 for connector blades 2.8 x 0.8
-25 ... +65°C
Rules and Standards
measuring range
special measuring
range
internal resistance
increased sensitivity
horizontal or on request 15°...165°
blade springs on the narrow sides
17
18
17
18
L1
N (L2)
AC current
17
L1
L1
N (L2)
N (L2)
18
k
l
K
L
Dimensions
window
colour of bezel
dial carrier
dial
a
b
zero adjustment front
position of use
connector
blades
6.3/2.8āxā0.8
cable grip
d
mounting
climatic suitability
e
mechanical loads
marine application
enclosure code
6
terminal protection
lamp for dial illuĆ
mination (optional)
c
screw clamp blade
spring
(optional)
scale arrangement
zero adjustment rear (optional)
dimensions (in mm) MG 48x24 MG 72x24 MG 96x24 K G 144x36
a
48
72
96
144
b
24
24
24
36
c
43
66
92
137
d
17
17
18
32
e
75
98
108
173
scales and position
of use
vertical mounting
I
horizontal scale (standard)
II
I
II vertical scale
inclined mounting
III horizontal scale
IV
IV vertical scale
III
VI
horizontal mounting
V
V horizontal scale
dial
a
a
VI vertical scale
Ordering Information
type
(M)G
front dimensions
48x24
72x24
96x24 K
144x36
measuring ranges
sp. measuring range
adjustment
sensitivity
voltmeters
slim edgewise moving-coil panel meter
for AC voltage or current
48 mm x 24 mm
72 mm x 24 mm
96 mm x 24 mm
144 mm x 36 mm
refer to preceding table
on request **)
internal resistance ±20% *)
internal resistance ±1% at 23°C
900 W/V *)
2 kW/V
5 kW/V
10 kW/V
20 kW/V as far as possible
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
logo
zero position
zero adjustment
expanded scale
(not for MG 48x24)
dial illumination
*)
**)
***)
glass *)
non-glaring glass
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
vertical *)
horizontal
on request 15 ... 165° **)
screw clamps *)
spring blades
class 2, -25 ... +40°C *)
class 3, -10 ... +55°C
shock 15 g, vibration 2.5 g *)
shock 30 g, vibration 5 g
none *)
non-certified
IP 52 *)
IP 54 splash-water protected front
none *)
protective sleeves B6 (for 6.3 x 0.8)
protective sleeves 110 (for 2.8 x 0.8)
horizontal *)
vertical
scale division & measuring range alike *)
blank dial
scale division and figuring
0 ... 100%
deviating from standard **)
2 scale divisions on request **)
yellow on black dial **)
white on black dial **)
additional lettering on request **)
additional figuring on request **)
coloured marks red, green or blue **)
coloured sector red, green or blue **)
WEIGEL *)
none
OEM logo **)
leftāĆāhand or bottom zero *)
mechanically suppressed zero **)
front accessible *)
rear accessible ***)
none *)
up to approx. 5% of fullāĆāscale value
in center scale electronically **)
none *)
with 1 lamp 6, 12 or 24 V ***)
standard
Please clearly add the desired specifications.
MG 96x24 K only
ordering example
MG 72x24, measuring range 0 ... 250 V, horizontal scale,
vertical mounting, window non-glaring glass, WEIGEL logo
- specifications subject to change without notice; date of issue 09/06 -
Data Sheet
Edgewise Series
015.D.201.05
Analog Meters
Edgewise with
MovingĆCoil Movement
and Rectifier
G 48 PrS
G 72 PrS
G 96 PrS
G144 PrS
Application
enclosure code
The edgewise moving-coil rectifier instruments G 48/72/96/144 PrS
with a curved dial are used for the measurement of sinusoidal AC curĆ
rents and voltages.
Moving-coil rectifier instruments measure average values and are
scaled to indicate r.m.s., assuming a sinusoidal wave form.
Measuring Ranges
The movingāĆācoil movement is manufactured to newest findings and
distinguishes in a small power consumption, a high accuracy and a
very good damping.
These instruments are suitable to be mounted in switchboards, control
panels, machine tool consoles and mosaic panels.
Movements
Self-shielding moving-coil movements with core-type magnet
(P 72/96/144 PrS) resp. swivel coil (P 48 PrS), pivot suspended.
Series-connectedrectifierincorporated.Springloadedjewelbearings
for vibration and shock resistance.
Mechanical Data
case details
material of case
material of window
colour of bezel
position of use
panel fixing
mounting
terminals
voltmeters and
ammeters 3 A
edgewise case suitable to be mounted
in control / switchgear panels, machine tool
consoles or mosaic panels, stackable
pressed steel (G 72/96 PrS)
thermoplastics (G 48/144 PrS)
glass '
black (similar to RAL 9005) '
vertical ±5° '
screw clamps
stackableānextātoāeachāotherā(exceptāG144āPrS)
hexagon studs, M3 screws and
wire clamps C6 (G 72/96 PrS),
connector blades 6.3 x 0.8 (G 48/144 PrS)
ammeters >3 A
hexagon studs, M5 screws and
wire clamps C10
voltmeters 60 ... 150, 600 V (G 72/96 PrS)
connector blades 6.3 x 0.8 for protective wire
dimensions (in mm) G 48 PrS G 72 PrS G 96 PrS G 144 PrS
bezel
48 x 24
72 x 36
96 x 48
144 x 72
case
43 x 17
66 x 32
91 x 43
137 x 67
depth
75
94
107
192
panel cutout
45+0.6
68+0.7
92+0.8
138+1.0
x 22.2+0.3 x 33+0.6 x 45+0.6 x 68+0.7
panel thickness
1 ... 25
1 ... 25
1 ... 12
40
weight approx.
0.08 kg
0.2 kg
0.45 kg
0.6 kg
Electrical Data
measuring unit
frequency range
AC voltage or AC current
40 Hz ... 50 Hz ... 10 kHz
overload capacity (acc. to DIN EN 60 051)
continuously
1.2 times rated voltage / current
5 s max.
2 times rated voltage,
10 times rated current
measurement category CAT III
operating voltage
refer to Measuring Ranges
pollution level
2
IP 52 case '
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact
AC current
100 mA, 150 mA, 250 mA, 400 mA, 600 mA, 1 mA, 1.5 mA, 2.5 mA, 4 mA,
6 mA, 10 mA, 15 mA, 25 mA, 40 mA, 60 mA, 100 mA, 150 mA,
250 mA, 400 mA, 600 mA
(voltage drop approx. 1.5 V)
(voltage drop approx. 0.2 V)
1 A, 1.5 A, 2.5 A *)
(voltage drop approx. 0.3 V)
4 A, 5 A, 6 A, 10 A, 15 A, 25 A *)
'
for use on current transformer (scale without overload range)
N/1 A *)
(voltage drop approx. 0.2 V)
(voltage drop approx. 0.3 V)
N/5 A *)
*) separate (G 48/72/96 PrS) resp. buitāĆāin (G 144 PrS)
miniature current transformer 50 Hz, 10 mA sec. inclusive
operating
G 48 PrS G 72 PrS G 96 PrS
G 144 PrS
voltage
300 V
300 V
300 V
600 V
AC voltage
operating voltage
G 48 PrS G 72 PrS
G 96 PrS
G 144 PrS
1.5 V
50 V
50 V
50 V
100 V
2.5 V
50 V
50 V
50 V
100 V
4V
50 V
50 V
50 V
100 V
6V
50 V
50 V
50 V
100 V
10 V
50 V
50 V
50 V
100 V
15 V
50 V
50 V
50 V
100 V
25 V
50 V
50 V
50 V
100 V
40 V
50 V
50 V
50 V
100 V
60 V
300 V
100 V
100 V
100 V
100 V
300 V
100 V
100 V
100 V
150 V
300 V
100 V
100 V
100 V
250 V
300 V
300 V
300 V
600 V
400 V
300 V
300 V
300 V
600 V
500 V
300 V
300 V
300 V
600 V
600 V
600 V
600 V
600 V
600 V
'
for use on voltage transformer (scale without overload range)
N/100 V
300 V
100 V
100 V
100 V
N/110 V
300 V
100 V
100 V
100 V
sensitivity 900 W/V '
the resistance values are limited to a tolerance of ±20% '
Scaling
pointer
response time
scale arrangement
scale characteristics
scale division
scale length
bar / knife-edge pointer
1 s for full-scale deflection
horizontal (left-hand zero) '
practically linear for voltages >20 V
initial scale compressed for voltages 20 V
coarse-fine
G 48 PrS G 72 PrS G 96 PrS G 144 PrS
30 mm
45 mm
67 mm
92 mm
' also refer to "Options"
Data Sheet
Edgewise Series
015.D.201.05
Analog Meters
Edgewise with
MovingĆCoil Movement
and Rectifier
Accuracy at Reference Conditions
accuracy class
adjustment of
resistance
increased sensitivity
accuracy class
1.5 according to DIN EN 60 051 '
reference conditions
ambient temperature
position of use
input
frequency
wave form
others
23°C±1K
nominal position ±1°
rated measuring value
50±2 Hz
sinusoidal, distortion factor <5%
DIN EN 60 051
case
window
colour of bezel
position of use
influences
ambient temperature
position of use
frequency
stray magnetic field
-25°C ... +23°C ... +40°C
nominal position ±5°
40 Hz ... 50 Hz ... 10 kHz
0.5 mT
with operating
temperature range
marine application
enclosure code
Environmental
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
shock resistance
vibration resistance
climatic class 2 '
according to VDE/VDI 3540 sheet 2
-25 ... +40°C '
-25 ... +65°C
75% annual average, non-condensing
15 g, 11 ms
2.5 g, 5 ... 55 Hz
performance
climatic suitability
measuring range
special measuring
range
measuring range
adjustment
2nd
measuring range
additional measuring
ranges
limited use in the tropics climatic class 3
according to VDE/VDI 3540 sheet 2
-10 ... +55°C
non-certified
IP 54 splash-water protected front
(without zero adjustment)
accessories
terminal protection against accidental contact
protective sleeves B6 for G 48 PrS
SW6, SW10 (ammeters >3A) for G 72/96 PrS
dial
scale arrangement
blank dial
scale division
and figuring
DIN 43 700
Options
to 2 kW/V, 5 kW/V, 10 kW/V or 20 kW/V
for voltmeters 1 V (as far as possible)
non-glaring glass
gray (similar to RAL 7037)
horizontal or on request 15°...165°
Rules and Standards
measuring and control instruments
for panel mounting;
nominal case and cutout dimensions
DIN 43 701
electrical switchboard instruments
DIN 43 718
bezels and front panels
DIN 16 257
nominal position of use and position symbols
applicable for measuring instruments
DIN 40 050
enclosure codes;
protection of electrical equipment against
ingress of solid foreign bodies and of water
DIN EN 60 051
direct acting indicating electrical measuring
instruments and their accessories
DIN EN 61 010
safety requirements for electrically operated
measuring, control and laboratory
equipment
VDE/VDI 3540 sheet 2 reliability of measuring and control
equipment (classification of climates)
1.0 with fine scale division (as far as possible)
to ±1% at 23°C
2nd scale division
additional lettering
additional figuring
coloured marks
coloured segment
logo on the dial
zero position
expanded scale
for G 72/96/144 PrS
vertical (bottom zero)
pencil-marked on initial and end values
0 ... 100%,
linear, full-scale values acc. to standardized
series (1 Ć 1.2 Ć 1.5 Ć 2 Ć 2.5 Ć 3 Ć 4 Ć 5 Ć 6 Ć 7.5
and their decimal multiples e.g. 150 m3/h) or
deviating from standard;
special calibration from customer's
non-linear graph or chart;
scaling of voltmeters in ohms;
captions on request
linear including figuring,
non-linear including figuring
on request e.g. "generator"
on request
red, green or blue for important scale values
red, green or blue within scale division
none or on request
mechanically suppressed zero, no zero
adjustment, max. 40% of full-scale value
for ammeters 100 mA, voltmeters 1.5 V
electrically suppressed zero
for voltmeters 6 V
expanded initial scale division by means of
electronic circuits up to approx. 5% of
full-scale value in centre of scale
Connections
deviating from standard
adjustment potentiometer installed in
voltmeters, adjustment range approx.
±10% or ±20 ... 50% (except G 48/72 PrS)
with 3rd terminal for voltmeters,
2nd figuring and 1 or 2 scale divisions
(except G 48/72 PrS)
on request
DC voltage
*
L1
N (L2)
DC current
17
18
17
for use on transformer
17
18
L1
L1
N (L2)
N (L2)
* G 72/96 PrS voltmeters 60 ... 150, 600 V
18
k
l
K
L
Dimensions
accuracy class
adjustments
G 48 PrS
b
d
sensitivity,
voltmeters
c
a
G 72/96 PrS
window
b
d
colour of bezel
c
a
G 144 PrS
position of use
climatic suitability
b
d
marine application
enclosure code
c
a
dimensions (in mm) G 48 PrS G 72 PrS G 96 PrS G 144 PrS
a
48
72
96
144
b
24
36
48
72
c
75
94
107
192
d
17
32
43
67
scales and position of use
II
I
a
a
III
IV
VI
V
vertical mounting
I
horizontal scale (standard)
II vertical scale
inclined mounting
III horizontal scale
IV vertical scale
horizontal mounting
V horizontal scale
VI vertical scale
Ordering Information
type
G
front dimensions
48 PrS
72 PrS
96 PrS
144 PrS
measuring ranges
special measuring
range
measuring range
adjustment
edgewise-type moving-coil
rectifier instrument
48 mm x 24 mm
72 mm x 36 mm
96 mm x 48 mm
144 mm x 72 mm
refer to preceding table
on request **)
none *)
voltage ±10%
voltage ±20 ... 50%
2nd measuring range none *)
1 scale division, 2nd figuring
2 scale divisions, 2 figurings
WEIGEL - MESSGERÄTE GmbH
PostfachĂ720ā154 S DāĆā90241 Nürnberg S Telefon: 0911ā/ā4ā23ā47āĆā0
Erlenstraße 14 S DāĆā90441 Nürnberg S Telefax: 0911ā/ā4ā23ā47āĆā39
Vertrieb:
Telefon: 0911ā/ā4ā23ā47āĆā94
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
terminal protection
scale arrangement
dial
logo
zero position
expanded scale
1.5 *)
1.0 with fine scale division ***)
none *)
internal resistance ±1% at 23°C
900 W/V *)
increased to approx. 2 kW/V
increased to approx. 5 kW/V
increased to approx.10 kW/V
increased to approx.20 kW/V ***)
glass *)
non-glaring glass
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
vertical *)
horizontal
on request 15 ... 165° **)
class 2, -25 ... +40°C *)
class 3, -10 ... +55°C
none *)
non-certified
IP 52 *)
IP 54 splash-water protected front
none *)
protective sleeves B6, SW6 resp. SW10
horizontal *)
vertical
scale division & measuring range alike *)
blank dial
scale division and figuring
0 ... 100%
acc. to standardized series **)
deviating from standard **)
calibration fr. non-linear graph or chart **)
scaling in ohms for voltmeters **)
2 scale divisions **)
additional lettering on request **)
additional figuring on request **)
coloured marks red, green or blue **)
coloured sector red, green or blue **)
WEIGEL *)
none
OEM logo **)
electrically suppressed zero **)
mechanically suppressed zero **)
none *)
electrically up to approx.
5% full-scale value*****)
*)
standard
**) Please clearly add the
***) as far as possible
****) G 72/96/144 PrS only
desired specifications.
ordering example
G 72 PrS, measuring range 0 ... 25 mA, horizontal scale 0 ... 100%,
vertical mounting, window non-glaring glass, WEIGEL logo
- specifications subject to change without notice; date of issue 02/07 -
Data Sheet
Edgewise Series
010.D.301.06
Analog Meters
Slim Edgewise with
MovingĆCoil Movement
MP 48x24
MP 72x24
MP 96x24 K
P 144x36
MP 96x24 K
with SlideāĆāInāĆāDial
Application
Measuring Ranges
The edgewise movingāĆācoil panel meters MP 48x24, MP 72x24,
MP 96x24 K and P 144x36 with a slimāĆāline dial are used for measureĆ
ment of DC currents and voltages.
The movingāĆācoil movement is manufactured to newest findings and
distinguishes in a small power consumption, a high accuracy and a
very good damping.
The instruments are suitable to be mounted in switchboards, control
panels, machine tool consoles and mosaic panels (except P 144x36).
With MP 96x24 K the front panel, the front window and the dial as well
as the dial illumination (optional) can be exchanged easily. On the rear
side a cable grip for the connecting leads is integrated.
DC current
Movements
SelfāĆāshielding movingāĆācoil movements with swivel coil, pivot susĆ
pended. Spring loaded jewel bearings for vibration and shock resisĆ
tance.
Mechanical Data
case details
dimensions (in mm)
edgewise case suitable to be mounted
in control panels, machine tool consoles
or mosaic panels (except P144x36), stackable
thermoplastics
glass '
black (similar to RAL 9005) '
vertical ±5° '
screw clamps '
stackableānextātoāeachāother
1 ... 25 mm
connector blades 6.3 x 0.8
or connector blades 2.8 x 0.8 (MP 96x24 K)
MPĂ48x24 MPĂ72x24 MPĂ96x24ĂK PĂ144x36
bezel
48 x 24
72 x 24
96 x 24
144 x 36
case
43 x 17
66 x 17
92 x 18
137 x 32
depth
75
98
108
173
panel cutout
68+0.7
92+0.8
45+0.6
x 22.2+0.3 x 22.2+0.3 x 22.2+0.3
weight approx.
0.08 kg
material of case
material of window
colour of bezel
position of use
panel fixing
mounting
panel thickness
terminals
0.1 kg
0.12 kg
138+1.0
x 33+0.6
0.5 kg
Electrical Data
measuring unit
DC voltage or DC current
overload capacity (acc. to DIN EN 60 051)
continuously
1.2 times rated voltage / current
5 s max. voltmeters 2 times rated voltage
ammeters 10 times rated current
measurement category CAT III
operating voltage
refer to Measuring Ranges
pollution level
2
enclosure code
IP 52 case '
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact
' for other ratings refer to "Options"
internal resistance*) / voltage drop approx.
MP 48x24 MP 72x24 MP 96x24 K P 144x36
100 mA
1000 W
680 W
2000 W
550 W
150 mA
835 W
480 W
1500 W
420 W
250 mA
500 W
300 W
900 W
250 W
400 mA
310 W
205 W
560 W
155 W
600 mA
210 W
110 W
390 W
105 W
1 mA
32 mV
31 W
47 W
52 W
1.5 mA
46 mV
24 W
40 W
45 W
2.5 mA
46 mV
20 W
25 W
27 W
4 mA
46 mV
17 W
15 W
17 W
5 mA
46 mV
16 W
11 W
14 W
6 mA
46 mV
60 mV
60 mV
60 mV
10 mA
46 mV
60 mV
60 mV
60 mV
15 mA
46 mV
60 mV
60 mV
60 mV
20 mA
46 mV
60 mV
60 mV
60 mV
25 mA
46 mV
60 mV
60 mV
60 mV
40 mA
46 mV
60 mV
60 mV
60 mV
60 mA
46 mV
60 mV
60 mV
60 mV
100 mA
46 mV
60 mV
60 mV
60 mV
150 mA
46 mV
60 mV
60 mV
60 mV
250 mA
46 mV
60 mV
60 mV
60 mV
400 mA
46 mV
60 mV
60 mV
60 mV
600 mA
46 mV
60 mV
60 mV
60 mV
1A
46 mV
60 mV
60 mV
60 mV
'
DC current with separate amplifier 10 mA, 15 mA, 25 mA, 50 mA, 60 mA
(power supply AC 230 V ā48 ... 62 Hz)
for use with external shunt 60 mV, 150 mV **)
a total lead resistance of 0.05 W is considered in the calibration of
the indicator for interconnecting leads 1 m, 2 x 0.75 mm2 '
for use on transducer ("live zero") 4 ... 20 mA
mechanically suppressed zero, no zero adjustment,
voltage drop approx. 46 mV '
operating
MP 48x24 MP 72x24 MP 96x24 K P 144x36
voltage
300 V
600 V
300 V
300 V
DC voltage **)
60 mV, 100 mV, 150 mV, 250 mV, 400 mV, 600 mV,
1 V, 1.5 V, 2.5 V, 4 V, 6 V, 10 V, 15 V, 25 V 40 V
operating
MP 48x24 MP 72x24 MP 96x24 K P 144x36
voltage
50 V
100 V
300 V
100 V
60 V, 100 V
operating
MP 48x24 MP 72x24 MP 96x24 K P 144x36
voltage
300 V
100 V
300 V
100 V
150 V, 250 V
operating
MP 48x24 MP 72x24 MP 96x24 K P 144x36
voltage
300 V
600 V
300 V
300 V
400 V, 500 V, 600 V (MP 48x24 on request)
operating voltage 600 V
*) the resistance values are limited to a tolerance of ±20%
**) sensitivity*) 1000 W/V '
Scaling
pointer
zero adjustment
response time
scale arrangement
scale characteristics
scale division
scale length
bar pointer
front accessible '
1 s for full-scale deflection
horizontal (left-hand zero) '
linear
coarse-fine
MP 48x24 MP 72x24 MP 96x24 K P 144x36
30 mm
52 mm
60 mm
95 mm
Data Sheet
Edgewise Series
010.D.301.06
Analog Meters
Slim Edgewise with
MovingĆCoil Movement
Accuracy at Reference Conditions
accuracy class
1.5 according to DIN EN 60 051
reference conditions
ambient temperature
position of use
input
others
23°C±1K
nominal position ±1°
rated measuring value
DIN EN 60 051
influences
ambient temperature
position of use
stray magnetic field
-25°C ... +23°C ... +40°C
nominal position ±5°
0.5 mT
Environmental
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
shock resistance
vibration resistance
climatic class 2 '
according to VDE/VDI 3540 sheet 2
-25 ... +40°C '
-25 ... +65°C
75% annual average, non-condensing
15 g, 11 ms '
2.5 g, 5 ... 55 Hz '
case
window
colour of bezel
position of use
mounting
(MP 96 x24 K)
non-glaring glass
gray (similar to RAL 7037)
horizontal or on request 15°...165°
blade springs on the narrow sides
performance
climatic suitability
with operating
temperature range
increased mechanical
loads
marine application
enclosure code
measuring and control instruments
for panel mounting;
nominal case and cutout dimensions
DIN 43 701
DIN 43 718
electrical switchboard instruments
bezels and front panels
DIN 16 257
nominal position of use and
position symbols
applicable for measuring instruments
DIN 40 050
enclosure codes;
protection of electrical equipment against
ingress of solid foreign bodies and of water
DIN EN 60 051
direct acting indicating electrical measuring
instruments and their accessories
safety requirements for electrically operated
measuring, control and laboratory
equipment
DIN EN 61 010
VDE/VDI 3540 sheet 2 reliability of measuring and control
equipment (classification of climates)
Options
measuring range
live zero"
(MP 96x24 K)
dial
scale arrangement
blank dial
scale division
and figuring
special measuring
range
internal resistance
increased sensitivity
lead resistance
2nd scale division
coloured scale
(MP 96x24 K only)
additional lettering
additional figuring
coloured marks
coloured segment
logo on the dial
zero position
zero adjustment
(MP 96x24 K only)
expanded scale
dial illumination
(MP 96x24 K only)
measuring range 0/4 ... 20 mA
electrically suppressed zero with zero
adjustment, voltage drop 900 mV approx.
deviating from standard
adjustment to ±1% at 23°C
for voltmeters 1 V and higher
to 2 kW/V, 5 kW/V, 10 kW/V or 20 kW/V
(as far as possible)
calibration of total value >0.05W
shock 30 g, 11 ms
vibration 5 g, 5 ... 55 Hz
non-certified
IP 54 splash-water protected front
(without zero adjustment or with zero
adjustment rear accessible for MP 96x24 K)
terminal protection against accidental contact
protective sleeves B6 for connector blades 6.3 x 0.8
protective sleeves 110 for connector blades 2.8 x 0.8
Rules and Standards
DIN 43 700
limited use in the tropics climatic class 3
according to VDE/VDI 3540 sheet 2
-10 ... +55°C
vertical (bottom zero)
pencil-marked on initial and end values
0 ... 100%,
linear, full-scale values acc. to standardized
series (1 Ć 1.2 Ć 1.5 Ć 2 Ć 2.5 Ć 3 Ć 4 Ć 5 Ć 6 Ć 7.5
and their decimal multiples e.g. 150 m3/h) or
deviating from standard;
special calibration from customer's
non-linear graph or chart;
scaling of voltmeters in ohms;
captions on request
including figuring (on request)
scale in black; pointer, scale division and
lettering yellow or white
on request e.g. "generator"
on request
red, green or blue for important scale values
red, green or blue within scale division
none or on request
centre zero or off-set zero,
mechanically suppressed zero, no zero
adjustment, max. 40% of full-scale value
for ammeters 100 mA, voltmeters 60 mV
rear accessible
expanded initial scale division
by means of electronic circuits up to approx.
5% of full-scale value in centre of scale
for MP 72x24 / 96x24 K, P 144x36
by 2 lamps 6 V, 12 V or 24 V
dial translucent
Connections
DC voltage
17
DC current
18
17
L+
L+
L-
L-
18
Dimensions
window
colour of bezel
dial carrier
dial
a
b
zero adjustment front
position of use
connector
blades
6.3/2.8āxā0.8
cable grip
d
mounting
climatic suitability
e
mechanical loads
marine application
enclosure code
6
terminal protection
lamp for dial illuĆ
mination (optional)
c
screw clamp blade
spring
(optional)
scale arrangement
zero adjustment rear (optional)
dimensions (in mm) MP 48x24
a
48
b
24
c
43
d
17
e
75
scales and position
of use
III
a
IV
VI
V
vertical mounting
I
horizontal scale (standard)
II vertical scale
inclined mounting
III horizontal scale
IV vertical scale
horizontal mounting
V horizontal scale
VI vertical scale
Ordering Information
type
(M)P
front dimensions
48x24
72x24
96x24 K
144x36
measuring ranges
sp. measuring range
live zero"
MP 96x24 K
adjustment
sensitivity
voltmeters
dial
MP 96x24 K P 144x36
96
144
24
36
92
137
18
32
108
173
II
I
a
MP 72x24
72
24
66
17
98
slim edgewise moving-coil panel meter
for DC voltage or current
48 mm x 24 mm
72 mm x 24 mm
96 mm x 24 mm
144 mm x 36 mm
refer to preceding table
on request **)
mechanically suppressed zero *)
electrically suppressed zero
internal resistance ±20% *)
internal resistance ±1% at 23°C
lead resistance >0,05 W
1 kW/V *)
2 kW/V
5 kW/V
10 kW/V
20 kW/V as far as possible
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
logo
zero position
zero adjustment
expanded scale
(not for MP 48x24)
dial illumination
*)
**)
***)
glass *)
non-glaring glass
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
vertical *)
horizontal
on request 15 ... 165° **)
screw clamps *)
spring blades
class 2, -25 ... +40°C *)
class 3, -10 ... +55°C
shock 15 g, vibration 2.5 g *)
shock 30 g, vibration 5 g
none *)
non-certified
IP 52 *)
IP 54 splash-water protected front
none *)
protective sleeves B6 (for 6.3 x 0.8)
protective sleeves 110 (for 2.8 x 0.8)
horizontal *)
vertical
scale division & measuring range alike *)
blank dial
scale division and figuring
0 ... 100%
linear acc. to standardized series **)
linear deviating from standard **)
calibration f. nonāĆālinear graph or chart**)
scaling in ohms for voltmeters **)
2 scale divisions on request **)
yellow on black dial **)
white on black dial **)
additional lettering on request **)
additional figuring on request **)
coloured marks red, green or blue **)
coloured sector red, green or blue **)
WEIGEL *)
none
OEM logo **)
leftāĆāhand or bottom zero *)
centre zero or offāĆāset zero **)
mechanically suppressed zero **)
front accessible *)
rear accessible ***)
none *)
up to approx. 5% of fullāĆāscale value
in center scale electronically **)
none *)
with 1 lamp 6, 12 or 24 V ***)
standard
Please clearly add the desired specifications.
MP 96x24 K only
ordering example
MP 72x24, measuring range 0 ... 20 mA, horizontal scale 0 ... 100%,
vertical mounting, window non-glaring glass, WEIGEL logo
- specifications subject to change without notice; date of issue 09/06 -
Data Sheet
Edgewise Series
010.D.201.05
Analog Meters
Edgewise with
MovingĆCoil Movement
P 48 PrS
P 72 PrS
P 96 PrS
P 144 PrS
Application
Measuring Ranges
The edgewise moving-coil panel meters P 48/72/96/144 PrS with a
curved dial are used for measurement of DC currents and voltages. The
edgewise case styles provide a high ratio of scale length to panel area.
These instruments are suitable to be mounted in switchboards, control
panels, machine tool consoles and mosaic panels.
DC current
Movements
Self-shielding moving-coil movements with core-type magnet
(P 72/96/144 PrS) resp. swivel coil (P 48 PrS), pivot suspended. Spring
loaded jewel bearings for vibration and shock resistance.
Mechanical Data
case details
material of case
material of window
colour of bezel
position of use
panel fixing
mounting
terminals
voltmeters and
ammeters 3 A
ammeters >3 A
voltmeters ≥300 V
dimensions (in mm)
bezel
case
depth
panel cutout
panel thickness
weight approx.
edgewise case suitable to be mounted
in control / switchgear panels, machine tool
consoles or mosaic panels, stackable
pressed steel (P 72/96 PrS)
thermoplastics (P 48/144 PrS)
glass '
black (similar to RAL 9005) '
vertical ±5° '
screw clamps
stackableānextātoāeachāotherā(exceptāPā144āPrS)
hexagon studs, M3 screws and
wire clamps C6 (P 72/96 PrS),
connector blades 6.3 x 0.8 (P 48/144 PrS)
hexagon studs, M5 screws and
wire clamps C10
(P 72/96 PrS)
connector blades 6.3 x 0.8 for protective wire
P 48 PrS P 72 PrS P 96 PrS P 144 PrS
48 x 24
72 x 36
96 x 48
144 x 72
43 x 17
66 x 32
91 x 43
137 x 67
75
94
107
192
68+0.7
92+0.8
138+1.0
45+0.6
x 22.2+0.3 x 33+0.6 x 45+0.6 x 68+0.7
1 ... 25
1 ... 25
1 ... 12
40
0.08 kg
0.2 kg
0.45 kg
0.6 kg
Electrical Data
measuring unit
DC voltages or DC currents
overload capacity (acc. to DIN EN 60 051)
continuously
1.2 times rated voltage / current
5 s max.
voltmeters
2 times rated voltage
ammeters
10 times rated current
measurement category CAT III
operating voltage
refer to Measuring Ranges
pollution level
2
enclosure code
IP 52 case '
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact
' also refer to "Options"
internal resistance *) / voltage drop approx.
P 48 PrS P 72 PrS P 96 PrS P 144 PrS
50 mA
1000 W
6500 W
6500 W
4600 W
60 mA
1040 W
5500 W
5500 W
4400 W
100 mA
1000 W
4900 W
4900 W
2800 W
150 mA
835 W
3600 W
3600 W
2200 W
250 mA
500 W
2200 W
2200 W
740 W
400 mA
310 W
1300 W
1300 W
630 W
600 mA
210 W
250 W
250 W
260 W
1 mA
32 mV
48 W
48 W
48 W
1.5 mA
46 mV
60 mV
60 mV
60 mV
2.5 mA
46 mV
60 mV
60 mV
60 mV
4 mA
46 mV
60 mV
60 mV
60 mV
5 mA
46 mV
60 mV
60 mV
60 mV
6 mA
46 mV
60 mV
60 mV
60 mV
10 mA
46 mV
60 mV
60 mV
60 mV
15 mA
46 mV
60 mV
60 mV
60 mV
20 mA
46 mV
60 mV
60 mV
60 mV
25 mA
46 mV
60 mV
60 mV
60 mV
40 mA
46 mV
60 mV
60 mV
60 mV
60 mA
46 mV
60 mV
60 mV
60 mV
100 mA
46 mV
60 mV
60 mV
60 mV
150 mA
46 mV
60 mV
60 mV
60 mV
250 mA
46 mV
60 mV
60 mV
60 mV
400 mA
46 mV
60 mV
60 mV
60 mV
600 mA
46 mV
60 mV
60 mV
60 mV
1A
46 mV
60 mV
60 mV
60 mV
1.5 A
60 mV
60 mV
60 mV
2.5 A
60 mV
60 mV
60 mV
4A
60 mV
60 mV
60 mV
6A
60 mV
60 mV
60 mV
10 A
60 mV
60 mV
60 mV
15 A
60 mV
60 mV
60 mV
25 A
60 mV
60 mV
60 mV
40 A
60 mV
60 mV
60 A
60 mV
'
DC current with separate amplifier 10 mA, 15 mA, 25 mA
(power supply AC 230 V ā48 ... 62 Hz)
for use with external shunt 60 mV, 150 mV **)
a total lead resistance of 0.05 W is considered in the calibration of
the indicator for interconnecting leads 1 m, 2 x 0.75 mm2 '
for use on transducer ("live zero")
0/4 ... 20 mA electrically suppressed zero, with zero adjustment
(P 72/96/144 PrS), voltage drop approx. 900 mV
4 ... 20 mA mechanically suppressed zero, no zero adjustment
(P 48 PrS), voltage drop approx. 46 mV
operating
P 48 PrS P 72 PrS P 96 PrS
P 144 PrS
voltage
300 V
300 V
300 V
600 V
DC voltage
P 48 PrS
P 72/96 PrS
P 144 PrS
40 mV
3300 W/V
2000 W/V
sensitivity*)
operating voltage
300 V
600 V
60 mV, 100 mV, 150 mV, 250 mV, 400 mV, 600 mV, **)
1 V, 1,5 V, 2,5 V, 4 V, 6 V, 10 V, 15 V, 25 V, 40 V, **)
operating voltage
50 V
50 V
100 V
60 V, 100 V, **)
operating voltage
300 V
100 V
100 V
150 V, 250 V, **)
operating voltage
300 V
300 V
600 V
400 V, 500 V, 600 Vā **)
operating voltage
600 V
600 V
600 V
'
*) the resistance values are limited to a tolerance of ±20%
**) sensitivity*) 1000 W/V '
Data Sheet
Edgewise Series
010.D.201.05
Analog Meters
Edgewise with
MovingĆCoil Movement
Scaling
pointer
response time
scale arrangement
scale characteristics
scale division
scale length
Options
bar / knife-edge pointer
1 s for full-scale deflection
horizontal (left-hand zero)
linear
coarse-fine
P 48 PrS P 72 PrS P 96 PrS
30 mm
45 mm
67 mm
measuring range
special measuring
range
measuring range
adjustment
P 144 PrS
92 mm
Accuracy at Reference Conditions
accuracy class
1.5 according to DIN EN 60 051 '
reference conditions
ambient temperature
position of use
input
others
23°C±1K
nominal position ±1°
rated measuring value
DIN EN 60 051
influences
ambient temperature
position of use
stray magnetic field
-25°C ... +23°C ... +40°C
nominal position ±5°
0.5 mT
2nd measuring range
additional measuring
ranges
accuracy class
adjustment of
internal resistance
increased sensitivity
case
window
colour of bezel
position of use
non-glaring glass
gray (similar to RAL 7037)
horizontal or on request 15°...165°
climatic suitability
with operating
temperature range
marine application
enclosure code
-25 ... +65°C
75% annual average, non-condensing
15 g, 11 ms
2.5 g, 5 ... 55 Hz
Rules and Standards
DIN 43 700
measuring and control instruments
for panel mounting;
nominal case and cutout dimensions
DIN 43 701
electrical switchboard instruments
DIN 43 718
bezels and front panels
DIN 16 257
nominal position of use and position symbols
applicable for measuring instruments
DIN 40 050
enclosure codes;
protection of electrical equipment against
ingress of solid foreign bodies and of water
DIN EN 60 051
direct acting indicating electrical measuring
instruments and their accessories
DIN EN 61 010
safety requirements for electrically operated
measuring, control and laboratory
equipment
VDE/VDI 3540 sheet 2 reliability of measuring and control
equipment (classification of climates)
1.0 with fine scale division (as far as possible)
to ±1% at 23°C
lead resistance
performance
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
shock resistance
vibration resistance
adjustment potentiometer installed in
voltmeters, adjustment range approx.
±10% or ±20 ... 50% (except P 48/72 PrS),
ammeters on request
with 3rd terminal for voltmeters,
2nd figuring and 1 or 2 scale divisions
(except P 48/72 PrS)
on request
to 2 kW/V, 5 kW/V, 10 kW/V or 20 kW/V
for voltmeters 1 V (as far as possible)
calibration of a total value >0.05W
Environmental
climatic class 2 '
according to VDE/VDI 3540 sheet 2
-25 ... +40°C '
deviating from standard
limited use in the tropics, climatic class 3
according to VDE/VDI 3540 sheet 2
-10 ... +55°C
non-certified
IP 54 splash-water protected front
(without zero adjustment)
terminal protection against accidental contact
protective sleeves B6 for P 48 PrS
SW6, SW10 (ammeters >3A) for P 72/96 PrS
dial
scale arrangement
blank dial
scale division
and figuring
2nd scale division
additional lettering
additional figuring
coloured marks
coloured sector
logo on the dial
zero position
expanded scale
for P 72/96/144 PrS
vertical (bottom zero)
pencil-marked on initial and end values
0 ... 100%,
linear, full-scale values acc. to standardized
series (1 Ć 1.2 Ć 1.5 Ć 2 Ć 2.5 Ć 3 Ć 4 Ć 5 Ć 6 Ć 7.5
and their decimal multiples e.g. 150 m3/h) or
deviating from standard;
special calibration from customer's
non-linear graph or chart;
scaling of voltmeters in ohms;
captions on request
linear including figuring,
non-linear including figuring
on request e.g. "generator"
on request
red, green or blue for important scale values
red, green or blue within scale division
none or on request
centre zero or off-set zero,
mechanically suppressed zero, no zero
adjustment, max. 40% of full-scale value
for ammeters 100 mA, voltmeters 60 mV
electrically suppressed zero
for voltmeters 6 V
expanded initial scale division by means of
electronic circuits up to approx. 5% of
full-scale value in centre of scale
Connections
DC voltage
*
measuring ranges
sp. measuring range
measuring range
adjustment
DC current
17
18
17
L+
L+
L-
L-
18
* P 72/96 PrS voltmeters ≥300 V
Dimensions
b
d
P 48 PrS
c
a
b
d
P 72/96 PrS
c
a
b
d
P 144 PrS
c
a
dimensions (in mm) P 48 PrS P 72 PrS P 96 PrS P 144 PrS
a
48
72
96
144
b
24
36
48
72
c
75
94
107
192
d
17
32
43
67
scales and position of use
II
I
a
a
III
IV
VI
V
vertical mounting
I
horizontal scale (standard)
II vertical scale
inclined mounting
III horizontal scale
IV vertical scale
horizontal mounting
V horizontal scale
VI vertical scale
Ordering Information
type
P
front dimensions
48 PrS
72 PrS
96 PrS
144 PrS
refer to preceding table
on request **)
none *)
voltage ±10%
voltage ±20 ... 50%
2nd measuring range none *)
1 scale division, 2nd figuring
2 scale divisions, 2 figurings
accuracy class
1.5 *)
1.0 with fine scale division as far as possible
adjustment
internal resistance to ±20% *)
internal resistance to ±1% at 23°C
lead resistance >0.05 W
sensitivity,
1 kW/V *)
voltmeters
to approx. 2 kW/V
to approx. 5 kW/V
to approx. 10 kW/V
to approx. 20 kW/V as far as possible
window
glass *)
non-glaring glass
colour of bezel
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
position of use
vertical *)
horizontal
on request 15 ... 165° **)
climatic suitability
class 2, -25 ... +40°C *)
class 3, -10 ... +55°C
marine application
none *)
non-certified
enclosure code
IP 52 *)
IP 54 splash-water protected front
terminal protection
none *)
protective sleeves B6, SW6 resp. SW10
scale arrangement
horizontal *)
vertical
dial
scale division & measuring range alike *)
blank dial
scale division and figuring
0 ... 100%
linear acc. to standardized series **)
linear deviating from standard **)
calibration fr. non-linear graph or chart **)
scaling in ohms for voltmeters **)
2 scale divisions **)
additional lettering on request **)
additional figuring on request **)
coloured marks red, green or blue **)
coloured sector red, green or blue **)
logo
WEIGEL *)
none
OEM logo **)
zero position
centre zero or off-set zero **)
electrically suppressed zero **)
mechanically suppressed zero **)
*) standard
**) Please clearly add the desired specifications.
edgewise-type moving-coil panel meter
48 mm x 24 mm
72 mm x 36 mm
96 mm x 48 mm
144 mm x 72 mm
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
ordering example
P 72 PrS, measuring range 0 ... 20 mA, horizontal scale 0 ... 100%,
vertical mounting, window non-glaring glass, WEIGEL logo
- specifications subject to change without notice; date of issue 02/07 -
Application
Connections
The wall brackets WA 96/144 are being supplied either empty or
equipped as synchronizing wall bracket.
Synchronizing Wall Bracket including Synchronoscope
The wall brackets can hold three each panel meters of the standard
square form 96 mm x 96 mm resp. 144 mm x 144 mm with a mounting
depth up to 150 mm resp. 190 mm.
L1
L2
supply system
L3
Synchronizing Wall Brackets
Voltage, frequency and phase balance have to coincide when connectĆ
ing an alternator to the supply system or to a second alternator, when
they are to be operated in parallel.
Four types of indicating meters are available
for synchronizing application:
(see data sheet No.)
-
(080.D.101.##) or
(022.D.101.##)
(023.D.101.##) and
(031.D.101.##)
Synchronoscope
Null Voltmeter
Dual Voltmeter
Dual Frequency Meter
SY
WQ
WQ
FQ
96/144 S
96/144 /0S
96/144 /2S
96/144 /2
or
form B
wall bracket suspended, support on top
form C
wall bracket upright,
support on top and bottom (WA 144 only)
form D
bench-type bracket, support on bottom
All bracket forms have a swivel range of 180°.
material
case
support
pressed steel
AL-die casting
colour
gray (similar to RAL 7037)
fixing
by 4 each screws ø10 mm
dimensions
see dimensional drawings
weight approx.
(wall bracket empty)
WA 96-A/B/D WA 144-A/B/D WA 144-C
4.5 kg
6.5 kg
7.5 kg
IP 52
additional data
refer to data sheets
relating to sole instruments
L2
II
L1
L1 L2
L1
L2
supply system
L3
17
L2
18
I
II
L1
L2
L2
I
G
null voltmeter
dual frequency meter
L1
L2
measuring and control instruments
for panel mounting;
nominal case and cutout dimensions
enclosure codes;
protection of electrical equipment against
ingress of solid foreign bodies and of water
connection leads
approx. 1 m free rubber-insulated
cable 5 x 1 mm2
alternator to be connected
II
L1
L1
L2 L1
L1 L2
DIN 57 410Ă/ĂVDEĂ0410 safety requirements for electrically operated
measuring, control and laboratory
equipment
DIN 40 050
dual voltmeter
Synchronizing Wall Bracket including Null Voltmeter
Rules and Standards
DIN 43 700
dual frequency meter
L1
L2 L1
G
synchronoscope
L2
II
L1
rubber-insulated cable approx. 1 m,
5 x 1 mm2
enclosure code
I
I
Synchronizing Wall Brackets
weight approx.
WA 96-A/B/D WA 144-A/B/D WA 144-C
(wall bracket equipped)7 kg
10 kg
11 kg
connection
3
L2
Mechanical Data
wall bracket upright, bottom support
1
L1
Null Voltmeter
Dual Voltmeter
Dual Frequency Meter
form A
4
L2
Three each indicating meters can be combined to a synchronizing
group in a wall bracket WA 96/144:
Synchronoscope
Dual Voltmeter
Dual Frequency Meter
2
alternator to be connected
dual voltmeter
connection leads
approx. 1 m free rubber-insulated
cable 5 x 1 mm2
Data Sheet
M - Series
100.D.101.03
Wall Brackets, empty
Synchronizing
Wall Brackets
Dimensions
form A, B
(form B: support will be mounted on upper side of wall bracket)
b
c
form D
c
11
100
140
d
b
a
a
30
100
11
dimensions (in mm)
a
b
c
30
170
140
form C (WA 144 only)
b
d
a
c
100
11
170
140
dimensions (in mm)
a
b
c
d
30
WA 96-A/ B
345
150
160
414
WA 144-A/ B/ C
487
200
200
556
WA 96-D
345
150
160
WA 144-D
487
200
200
Ordering Information
type
WA
dimensions
96
144
form
A
B
C
D
Wall Bracket empty
accommodating instruments
96 mm x 96 mm
accommodating instruments
144 mm x 144 mm
bracket upright, bottom support *)
bracket suspended, support on top
bracket upright,
support on top and bottom **)
bench-type bracket, bottom support
*)
**)
standard
WA 144 only
ordering example 1
Wall Bracket WA 144-B, empty
type
WA
dimensions
96
144
form
A
B
C
D
combination
SY 96/144 S
WQ 96/144 /2S
FQ 96/144 /2
or
WQ 96/144 /0S
WQ 96/144 /2S
FQ 96/144 /2
assembly
Synchronizing Wall Bracket
accommodating instruments
96 mm x 96 mm
accommodating instruments
144 mm x 144 mm
bracket upright, bottom support *)
bracket suspended, support on top
bracket upright,
support on top and bottom **)
bench-type bracket, bottom support
meter
refer to data sheet No.
synchronoscope
080.D.101.##
dual voltmeter
023.D.101.##
dual frequency meter
031.D.101.##
null voltmeter
022.D.101.##
dual voltmeter
023.D.101.##
dual frequency meter
031.D.101.##
mount and connect instruments,
install connection leads
*)
**)
standard
WA 144 only
ordering example 2
Synchronizing Wall Bracket WA 96-A
comprising instrument combination:
Synchronoscope SY 96 S,
Dual Voltmeter WQ 96 /2S and
Dual Frequency Meter FQ 96 /2, 2x 21 reeds
rated voltage 400 V, rated frequency 50 Hz,
installed and wired
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 DĆ90241 Nürnberg Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 DĆ90441 Nürnberg Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
- specifications subject to change without notice; date of issue 04/03 -
Data Sheet
Phase Sequence
Indicator
DFQ 96
M Series
090.D.101.04
Application
Indication
The phase sequence indicator DFQ 96 (M series) is used to determine
the phase sequence in three-phase systems up to 500 V by a direct
connection.
A disc marked with two arrows rotates clockwise (in arrow direction)
when pressing a button accessible on indicator front, provided the
three phases are logically connected in accordance with the indicator
terminal markings, otherwise the disc will rotate anticlockwise.
In case of an incorrect phase sequence, the correct direction of rotation
is obtained by interchanging any of two phases.
Phase sequence indicators are housed in pressed steel cases suitable
to be mounted in switchboards, control panels, machinery and/or moĆ
saic grid panels.
Disc marked with two arrows rotates behind dial plate with an arrow
indicating the correct phase sequence.
disc diameter
36 mm
Functional Principle
Induction-movement with a freely rotating disc.
Mechanical Data
case details
material of case
material of window
colour of bezel
position of use
panel fixing
panel thickness
mounting
terminals
square case suitable to be mounted
in switchboards or mosaic grid panels,
stackable '
pressed steel
glass '
black (similar to RAL 9005) '
vertical ±5° '
screw clamps
1 ... 15 mm
stackable next to each other
hexagon studs, M3 screws and
wire clamps C6
dimensions
DFQ 96
bezel
case
depth
panel cutout
weight approx.
V 96 mm
V 90.5 mm
62 mm
V 92+0.8 mm
0.4 kg
Electrical Data
measuring unit
frequency range
voltage range
power consumption
safe operational
period
measurement category
operating voltage
pollution level
enclosure code
phase sequence in three-phase systems
40 ... 100 Hz '
100 ... 500 V
at 100 V approx. 0.5 VA per phase
at 500 V approx. 2 VA per phase
5 min max.
CAT III
300 V
2
IP 52 case '
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact
' also refer to "Options"
Environmental
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
shock resistance
vibration resistance
climatic class 2 according to VDE/VDI 3540,
sheet 2 '
-25 ... +40°C '
-25 ... +65°C
75% annual average, non-condensing
15 g, 11 ms '
2.5 g, 5 ... 55 Hz '
Rules and Standards
DIN 43 700
measuring and control instruments
for panel mounting;
nominal case and cutout dimensions
DIN 43 701
electrical switchboard instruments
DIN 43 718
bezels and front panels
DIN 16 257
nominal position of use and
position symbols
applicable for measuring instruments
DIN 40 050
enclosure codes;
protection of electrical equipment against
ingress of solid foreign bodies and of water
DIN EN 60 051
direct acting indicating electrical measuring
instruments and their accessories
DIN EN 61 010
safety requirements for electrically operated
measuring, control and laboratory
equipment
VDE/VDI 3540 sheet 2 reliability of measuring and control
equipment (classification of climates)
Data Sheet
M Series
090.D.101.04
Phase Sequence
Indicator
Options
case
portable type
window
colour of bezel
position of use
400 Hz on request
L1 (R)
mains
L2 (S)
on request
non-glaring glass
gray (similar to RAL 7037)
horizontal or to be specified 15°...165°
performance
increased mechanical shock 30 g, 11 ms
loads
vibration 5 g, 5 ... 55 Hz
climatic suitability
limited use in the tropics
climatic class 3 according to VDE/VDI 3540,
sheet 2
with operating
-10 ... +55°C
temperature range
marine application
non-certified
enclosure code
IP 54 splash-water protected front
L3 (T)
S
T
L1
L2
L3
Dimensions
a
accessories
terminal protection against accidental contact
full-sized rear cover or protective sleeves
terminals
connector blades 6.3 x 0.8
dial
custom logo
R
b
frequency
Connections
push button
none or as specified
c
dimensions (in mm)
a
b
c
DFQ 96
96
90
62
15
Ordering Information
type
DFQ
front dimensions
96
frequency
Phase Sequence Indicator
version
window
colour of bezel
position of use
mechanical loads
climatic suitability
marine application
enclosure code
terminal protection
terminals
logo
*) standard
**) Please clearly
***) on request
96 mm x 96 mm
40 ... 100 Hz *)
400 Hz ***)
panel type *)
portable type ***)
glass *)
non-glaring glass
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
vertical *)
to be specified 15 ... 165° **)
shock 15 g, vibration 2.5 g *)
shock 30 g, vibration 5 g
class 2, -25 ... +40°C *)
class 3, -10 ... +55°C
none *)
non-certified
IP 52 *)
IP 54 splash-water protected front
none *)
full-sized rear cover ***)
protective sleeves
screws M3 x 6 *)
connector blades 6.3 x 0.8
WEIGEL *)
none
OEM logo **)
add the desired specifications.
ordering example
DFQ 96 50 Hz, panel type indicator,
window non-glaring glass, WEIGEL logo
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
- specifications subject to change without notice; date of issue 02/07 -
Data Sheet
M Series
080.D.101.05
Analog Meters for
Phase Comparison
Synchronoscopes
SY 96 S
SY 144 S
Application
Mechanical Data
The synchronoscopes SY 96/144 S (M series) are used to measure
phase difference of two AC voltages, e.g. those of a generator and the
mains.
The dial carries a zero marker, a red arrow marked with +" and a black
arrow marked with -"
When the frequencies of the two voltages differ less than approx. 2 Hz,
they can be compared approximately:
The pointer rotates according to frequency ratio (and to meter connecĆ
tion) clockwise or anticlockwise. The pointer stands still when the freĆ
quencies get equal. The pointer rests at the zero marker when the
phases coincide also; the AC voltages can be connected to each other.
The connection of the meter can be chosen so that the pointer rotates
clockwise or counterclockwise when the frequency f1 (e.g. generator)
is higher.
When the meter is not in work, the pointer cannot be seen.
The instruments are suitable to be mounted in generating sets, power
supply control panels, switchboards or mosaic panels.
case details
Movements
The synchronoscopes SY 96 S and SY 144 S are supplied with an ironĆ
less electrodynamical quotient movement.
The pointer is able to rotate over 360° in both directions.
material of case
material of window
colour of bezel
terminals
terminal protection
against accidental
contact
position of use
dimensions
bezel
case
depth
panel cutout
panel thickness
panel fixing
SY 96 S
V 96 mm
V 90 mm
119 mm
V 92+0.8 mm
1...15 mm
4 pieces WEIGEL
screw clamps
weight approx.
1.0 kg
measuring unit
for voltages 100/110 V
generator coil
mains
phase angle of two sinusoidal AC voltages
Measuring Ranges
frequency
mains coils
generator
SY 144 S
V 144 mm
V 137 mm
117 mm
V 138+1 mm
1...40 mm
2 pieces type B
screw clamps
acc. to DIN 43 835
1.1 kg
Electrical Data
Principle Circuit Diagrams
U
V
W
square case suitable to be mounted
in switchboards or mosaic grid panels,
stackable '
pressed steel
glass '
black (similar to RAL 9005) '
hexagon studs, M3 screws and
wire clamps C6 '
connector blades 6.3 x 0.8 for protective wire
full-sized terminal cover (SY 96 S) /
protective sleeves SW6 (SY 144 S)
included
vertical ±5° '
L1
L2
L3
for voltages 230/400/500 V
mains coils
generator coil
rated frequency
frequency range
50 Hz '
48.5 ... 51.5 Hz
voltage
rated voltage UN operating voltage
60; 100; 110 V
300 V
230; 400; 415 V
300 V
400 V
300 V
440; 500 V '
300 V
power consumption approx.
at rated voltage
generator side
mains side
110 V
0.7 VA
4.0 VA
230 V
1.5 VA
5.3 VA
400 V
3.2 VA
4.8 VA
500 V
3.5 VA
6.7 VA
operating range
UN ±10%
overload capacity
UN +20%
measurement category CAT III
operating voltage
refer to Measuring Ranges
pollution level
2
enclosure code
IP 52 case '
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact
Scaling
U
V
W
generator
mains
L1
L2
L3
pointer
pointer deflection
zero
accuracy class
bar pointer
360°
triangle on dial top
1 according to DIN EN 60 051
' also refer to "Options"
Data Sheet
M Series
080.D.101.05
Analog Meters for
Phase Comparison
Synchronoscopes
Environmental
dial
logo on the dial
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
shock resistance
vibration resistance
stray magnetic field
climatic class 2 according to VDE/VDI 3540 '
-10 ... +40°C
Connections
-25 ... +65°C
Note: The rotation direction depends on connection.
none or on request
75% annual average, non-condensing
15 g, 11 ms
2.5 g, 5 ... 55 Hz
<0.5 mT according to DIN EN 60 051
Rules and Standards
DIN 43 700
voltage
voltage range
Dimensions
g
f
rated frequency
frequency range
16 1/3 ... 17 Hz
16 2/3 Hz
60 Hz
58.5 ... 61.5 Hz
rated voltage UN : 60 V, 415 V, 440 V
others (>57.8 V ... <500 V) on request
100 ... 120 V : 3
100 ... 120 V
208 ... 230 V
380 ... 400 V
case
window
non-glaring glass
colour of bezel
gray (similar to RAL 7037)
position of use
on request 15°...165°
increased mechanical shock 30 g, 11 ms
loads
vibration 5 g, 5 ... 55 Hz
climatic suitability
climatic class 3 according to VDE/VDI 3540
with operating
-10 ... +55°C
temperature range
75% annual average, non-condensing)
(relative humidity
climatic suitability
limited use in the tropics"
with operating
-25 ... +55°C
temperature range
75% annual average, non-condensing)
(relative humidity
marine application
non-certified
enclosure code
IP 54 splash-water protected front
terminals
connector blades 6.3 x 0.8
terminal protection
rubber nozzles
against accidental contact
d
a
Options
electrical data
frequency
See also meter label
a
b
measuring and control instruments
for panel mounting;
nominal case and cutout dimensions
DIN 43 701
electrical switchboard instruments
DIN 43 718
bezels and front panels
DIN 16 257
nominal position of use and position symbols
applicable for measuring instruments
DIN 40 050
enclosure codes;
protection of electrical equipment against
ingress of solid foreign bodies and of water
DIN EN 60 051
direct acting indicating electrical measuring
instruments and their accessories
DIN EN 61 010
safety requirements for electrically operated
measuring, control and laboratory
equipment
VDE/VDI 3540 sheet 2 reliability of measuring and control
equipment (classification of climates)
c
dimensions (in mm)
e
SY 96 S
a
96
b
92
c
100
d
5
e
6.5
f
19 *)
g
*) including cover for external series resistor
SY 144 S
144
138
103
8
3
14
Ordering Information
type
SY
front dimensions
96 S
144 S
rated voltage
rated frequency
window
colour of bezel
position of use
mechanical loads
climatic suitability
marine application
enclosure code
terminals
logo
*)
**)
synchronoscope
96 mm x 96 mm
144 mm x 144 mm
60 V
100 V
110 V
230 V *)
400 V
415 V
440 V
500 V
others **) (>57.8 V ... <500 V)
16 2/3 Hz
50 Hz *)
60 Hz
glass *)
non-glaring glass
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
vertical *)
on request 15 ... 165° **)
shock 15 g, vibration 2.5 g *)
shock 30 g, vibration 5 g
class 2, -10 ... +40°C *)
class 3, -10 ... +55°C
limited use in the tropics", -25 ... +55°C
none *)
non-certified
IP 52 *)
IP 54 splash-water protected front
screws M3 x 6 and wire clamps *)
connector blades 6.3 x 0.8
additional lettering on request **)
WEIGEL *)
none
OEM logo **)
standard
Please clearly add the desired specifications.
ordering example
SY 96 S, rated voltage AC 230 V, rated frequency 50 Hz,
window non-glaring glass, no logo
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstrasse 14 S DĆ90441 Nürnberg S Fax: 0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
- specifications subject to change without notice; date of issue 12/06 -
Data Sheet
M Series
031.D.101.05
Double Frequency
Meters with Vibrating
Reed Movements
FQ 96 /2
FQ 144 /2
Application
Measuring Ranges
The double frequency indicators FQ 96 /2 and FQ 144 /2 (M series)
comprise two electrically independant vibrating reed movements for
simultaneously measuring the frequencies of two AC voltages, that of a
generator and of a power supply system for instance.
two identical
measuring ranges
Double frequency indicators are also available with consecutive freĆ
quency ranges to one voltage.
The instruments are suitable to be mounted in generating sets, power
supply control panels, switchboards or mosaic panels.
Functional Principle
Vibrating reed-type movements with two lines of steel reeds. Each
reed is tuned to a different value in the frequency span.
Mechanical Data
case details
square case suitable to be mounted
in control / switchgear panels, machine tool
consoles or mosaic panels, stackable
material of case
pressed steel
material of window
glass
colour of bezel
black (similar to RAL 9005)
position of use
vertical ±5°
panel fixing
WEIGEL screw clamps
panel thickness
1 ... 15 mm
mounting
stackable next to each other
terminals
hexagon studs, M3 x 6 screws and
wire clamps
dimensions
FQ 96 /2
FQ 144 /2
bezel
V 96 mm
V 144 mm
case
V 90 mm
V 137 mm
depth
66 mm
58 mm
panel cutout
V 92+0.8 mm
V 138+1 mm
weight approx.
0.65 kg
1.0 kg
Electrical Data
measuring units
two frequencies
overload capacity (acc. to DIN EN 60 051)
continuously
1.2 times rated voltages
5 s max.
2 times rated voltages
measurement category
operating voltage
pollution level
enclosure code
CAT III
refer to Measuring Ranges
2
IP 52 case '
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact
' for other ratings refer to "Options"
2x 47 ... 50 ... 53 Hz
2x 45 ... 50 ... 55 Hz
2x 57 ... 60 ... 63 Hz
2x 55 ... 60 ... 65 Hz
two consecutive
measuring ranges
left
right
left
right
number of steel reeds
2x 13
2x 21
2x 13
2x 21
number of steel reeds
47 ... 50 ... 53 Hz and
57 ... 60 ... 63 Hz
45 ... 50 ... 55 Hz and
55 ... 60 ... 65 Hz
rated voltage
100 V
110 V
230 V
400 V
500 V
600 V
2x 13
2x 21
power consumption
each movement
FQ 96 /2
FQ 144 /2
FQ 96 /2
FQ 144 /2
< 1.1 VA
< 1.1 VA
< 1.2 VA
< 2.0 VA
< 2.5 VA
< 2.9 VA
300 V
300 V
300 V
300 V
300 V
600 V
300 V
300 V
300 V
300 V
300 V
600 V
< 1.1 VA
< 1.1 VA
< 1.2 VA
< 1.8 VA
< 2.4 VA
< 2.9 VA
operating voltage
Scaling
scale division
reed arrangement
coarse-fine
vertical
Accuracy at Reference Conditions
accuracy class
0.5 acc. to DIN EN 60 051
reference conditions
ambient temperature
position of use
input
others
23°C±1K
nominal position ±1°
rated measuring value
DIN EN 60 051
influence quantities
ambient temperature
position of use
stray magnetic field
-25°C ... +23°C ... +40°C
nominal position ±5°
0.5 mT
Environmental
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
shock resistance
vibration resistance
climatic class 2
according to VDE/VDI 3540 sheet 2
-25 ... +40°C
-25 ... +65°C
75% annual average, non-condensing
15 g, 11 ms
2.5 g, 5 ... 55 Hz
Data Sheet
M Series
031.D.101.05
Double Frequency
Meters with Vibrating
Reed Movements
Rules and Standards
Connections
DIN 43 700
two AC voltages
measuring and control instruments
for panel mounting;
nominal case and cutout dimensions
DIN 43 701
electrical switchboard instruments
DIN 43 718
bezels and front panels
DIN 16 257
nominal position of use and
position symbols
applicable for measuring instruments
DIN 40 050
enclosure codes;
protection of electrical equipment against
ingress of solid foreign bodies and of water
DIN EN 60 051
direct acting indicating electrical measuring
instruments and their accessories
DIN EN 61 010
safety requirements for electrically operated
measuring, control and laboratory
equipment
VDE/VDI 3540 sheet 2 reliability of measuring and control
equipment (classification of climates)
L1
L2
mains supply
L3
L1
L2
I
L1
II
L2
L1
generator
L2
L3
one AC voltage
17
18
Options
L1
case
window
non-glaring glass
colour of bezel
gray (similar to RAL 7037)
position of use
on request 15°...165°
N(L2)
Dimensions
performance
limited use in the tropics
climatic class 3 according to VDE/VDI 3540
with operating
temperature range
-10 ... +55°C
marine application
non-certified
enclosure code
IP 54 splash-water protected front
b
climatic suitability
a
increased mechanical shock 30 g, 11 ms
loads
vibration 5 g, 5 ... 55 Hz
c
dial
reed arrangement
horizontal
additional lettering
on request e.g. "generator"
additional figuring
on request
coloured marks
red, green or blue for important scale values
coloured sector
red, green or blue within scale division
logo on the dial
none or on request
accessories
terminal protection against accidental contact
full-sized rear cover or protective sleeves SW6
terminals
connector blades 6.3 x 0.8
dimensions (in mm)
a
b
c
FQ 96 /2
96
90
66
FQ 144 /2
144
137
58
Ordering Information
type
FQ
front dimensions
96
144
version
/2
measuring ranges
rated voltage
window
reed-type frequency meter
96 mm x 96 mm
144 mm x 144 mm
two movements
refer to preceding table
refer to preceding table
glass *)
non-glaring glass
colour of bezel
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
position of use
vertical *)
on request 15 ... 165° **)
increased mechanical shock 30 g, 11 ms
loads
vibration 5 g, 5 ... 55 Hz
climatic suitability
class 2, -25 ... +40°C *)
class 3, -10 ... +55°C
marine application
none *)
non-certified
enclosure code
IP 52 *)
IP 54 splash-water protected front
terminal protection
none *)
full-sized rear cover ***)
protective sleeves
terminals
screws M3 x 6 and wire clamps *)
connector blades 6.3 x 0.8
reed arrangement
vertical *)
horizontal
dial
scale division & measuring range alike *)
additional lettering on request **)
additional figuring on request **)
coloured marks red, green or blue **)
coloured sector red, green or blue **)
logo
WEIGEL *)
none
OEM logo **)
*) standard
**) Please clearly
***) FQ 96 /2 only
add the desired specifications.
ordering example
FQ 96 /2. measuring range 2x 45 ... 50 ... 55 Hz,
rated voltage AC 230 V, window non-glaring glass, no logo
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
- specifications subject to change without notice; date of issue 06/06 -
Data Sheet
M Series
023.D.101.05
Analog Meters with
Two Moving-Iron
Movements
WQ 96 /2S
WQ 144 /2S
Application
Measuring Ranges
The dual voltmeters WQ 96 /2S and WQ 144 /2S (M series) comprise
two electrically independent moving-iron movements for simultaĆ
neously measuring two AC voltages, one to indicate the generator voltĆ
age, the other to indicate the power supply voltage.
AC voltages
The dual voltmeters are housed in pressed steel cases. They are suitĆ
able to be mounted in generating sets, power supply control panels,
switchboards or mosaic panels.
Functional Principle
Two electrically independent moving-iron movements, pivot susĆ
pended. Spring loaded jewel bearings and silicon oil damping for
vibration and shock resistance.
Mechanical Data
operating voltage
power consumption
approximately
WQ 96 2/S WQ 144 2/S
2 x 0 ... 100 V *)
300 V
1.8 VA
2 x 0 ... 110 V *)
300 V
2.0 VA
2 x 0 ... 150 V
300 V
2.2 VA
2 x 0 ... 250 V
300 V
2.3 VA
2 x 0 ... 500 V
300 V
2.1 VA
2 x 0 ... 600 V
600 V
2.5 VA
*) also for use on voltage transformer,
please state primary rating of voltage transformer.
Scaling
pointer
pointer deflection
scale characteristics
bar / knife-edge pointer
0 ... 60°
practically linear
down to 1/5 th of full-scale rating
coarse-fine
WQ 96 /2S
WQ 144 /2S
60 mm
97 mm
case details
square case suitable to be mounted
in control / switchgear panels
or mosaic panels, stackable
material of case
pressed steel
material of window
glass '
colour of bezel
black (similar to RAL 9005) '
position of use
vertical ±5° '
mounting
stackable next to each other
accuracy class
1.5 acc. to DIN EN 60 051
terminals
hexagon studs, M3 x 6 screws and
wire clamps
connector blades 6.3 x 0.8 for protective wire
(WQ 96/2 only)
reference conditions
ambient temperature
position of use
input
others
23°C±1K
nominal position ±1°
rated measuring value
DIN EN 60 051
influence quantities
ambient temperature
position of use
stray magnetic field
-25°C ... +23°C ... +40°C
nominal position ±5°
0.5 mT
dimensions
WQ 96 /2S
WQ 144 /2S
bezel
V 96 mm
V 144 mm
case
V 90 mm
V 137 mm
depth
115 mm
121 mm
panel cutout
V 92+0.8 mm
V 138+1 mm
panel thickness
1...15 mm
1...40 mm
panel fixing
4 pieces WEIGEL
screw clamps
2 pieces type B
screw clamps
acc. to DIN 43 835
weight approx.
1.2 kg
1.5 kg
Electrical Data
measuring units
two AC voltages
frequency range
50 ... 100 Hz
overload capacity (acc. to DIN EN 60 051)
continuously
1.2 times rated voltages
5 s max.
2 times rated voltages
enclosure code
IP 52 case
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact
measurement category CAT III
operating voltage
refer to Measuring Ranges
pollution level
2
' also refer to "Options"
2.5 VA
2.3 VA
2.6 VA
3.9 VA
3.8 VA
4.5 VA
scale division
scale length
Accuracy at Reference Conditions
Environmental
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
shock resistance
vibration resistance
climatic class 2 '
according to VDE/VDI 3540 , sheet 2
-25 ... +40°C
-25 ... +65°C
75% annual average, non-condensing
15 g, 11 ms
2.5 g, 5...55 Hz
Data Sheet
M Series
023.D.101.05
Analog Meters with
Two Moving-Iron
Movements
Rules and Standards
DIN 43 700
measuring and control instruments
for panel mounting;
nominal case and cutout dimensions
DIN 43 701
electrical switchboard instruments
DIN 43 718
bezels and front panels
DIN 16 257
nominal position of use and
position symbols
applicable for measuring instruments
DIN 40 050
enclosure codes;
protection of electrical equipment against
ingress of solid foreign bodies and of water
DIN EN 60 051
direct acting indicating electrical measuring
instruments and their accessories
DIN EN 61 010
safety requirements for electrically operated
measuring, control and laboratory
equipment
VDE/VDI 3540 sheet 2 reliability of measuring and control
equipment (classification of climates)
Connections
L1
L2
L3
mains
supply
generator
L1
L2
I
L1
II
L2
L1
(WQ 96/2 only)
L2
L3
Dimensions
performance
climatic suitability
with operating
temperature range
marine application
enclosure code
dielectric test
dial
blank dial
scale division
and figuring
additional lettering
additional figuring
coloured marks
coloured sector
logo on the dial
non-glaring glass
gray (similar to RAL 7037)
on request 15°...165°
b
case
window
colour of bezel
position of use
a
Options
c
limited use in the tropics
climatic class 3 according to VDE/VDI 3540
-10 ... +55°C
non-certified
IP 54 splash-water protected front
3 kV based on 50 Hz, 1 min acc. to DIN 57 410
pencil-marked on initial and end values
0 ... 100% or
deviating from standard,
captions on request
on request e.g. "generator"
on request
red, green or blue for important scale values
red, green or blue within scale division
none or on request
accessories
terminal protection against accidental contact
full-sized rear cover (WQ 96 /2S only) or protective sleeves SW6
terminals
connector blades 6.3 x 0.8
dimensions (in mm)
a
b
c
WQ 96 /2S
96
90
115
WQ 144 /2S
144
137
121
Ordering Information
type
WQ
front dimensions
96
144
version
/2S
measuring ranges
window
colour of bezel
position of use
climatic suitability
marine application
enclosure code
terminal protection
terminals
dial
logo
*) standard
**) Please clearly add
***) WQ 96 /2S only
dual moving-iron voltmeter
96 mm x 96 mm
144 mm x 144 mm
two movements
refer to preceding table
glass *)
non-glaring glass
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
vertical *)
on request 15 ... 165° **)
class 2, -25 ... +40°C *)
class 3, -10 ... +55°C
none *)
non-certified
IP 52 *)
IP 54 splash-water protected front
none *)
full-sized rear cover ***)
protective sleeves
screws M3 x 6 and wire clamps *)
connector blades 6.3 x 0.8
scale division & measuring range alike *)
blank dial
scale division and figuring 0 ... 100%
deviating from standard **)
additional lettering on request **)
additional figuring on request **)
coloured marks red, green or blue **)
coloured sector red, green or blue **)
WEIGEL *)
none
OEM logo **)
the desired specifications.
ordering example
WQ 96 /2S, measuring range 2x 0 ... 150 V,
window non-glaring glass, dial 0 ... 100%, no logo
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
- specifications subject to change without notice; date of issue 09/06 -
Data Sheet
M Series
022.D.101.04
Analog Meters
Null Voltmeters with
MovingāĆāIron Movement
90°āĆāDial
WQ 96 /0S
WQ 144 /0S
Application
Measuring Ranges
The null voltmeters WQ 96/144 /0S (M series), housed in pressed steel
cases, contain a moving-iron movement which scale is extremely exĆ
panded in its initial value range. Even small voltage differences of two
AC voltages can be detected, e.g. those of a generator and of the mains
supply.
The meters are suitable to be mounted in switchboards, synchronizing
wall brackets or mosaic grid panels.
rated voltage
power consumption
WQ 96 /0S
power consumption
WQ 144 /0S
0 ... 115/230 V
0 ... 230/460 V
0 ... 400/800 V
0 ... 500/1000 V
approx. 4.2 VA
approx. 4.5 VA
approx. 5.4 VA
approx. 5.9 VA
approx. 4.9 VA
approx. 5.4 VA
approx. 5.8 VA
approx. 7.2 VA
for use on voltage transformer
0 ... 100/200 V
approx. 4.2 VA
0 ... 110/220 V
approx. 4.2 VA
approx. 4.9 VA
approx. 4.9 VA
Functional Principle
Moving-iron movement; spring loaded jewel bearings and silicon oil
damping for vibration and shock resistance.
Mechanical Data
case details
panel thickness
panel fixing
square case suitable to be mounted
in control / switchgear panels or
mosaic grid panels, stackable
pressed steel
glass '
black (similar to RAL 9005) '
vertical ±5° '
stackable next to each other
hexagon studs, M3x6 screws and
wire clamps C6
connector blades 6.3 x 0.8 for protective wire
(WQ 96/0S only)
1 ... 15 mm
4 pieces WEIGEL screw clamps
dimensions (in mm)
WQ 96 /0S
WQ 144 /0S
bezel
V 96 mm
V 144 mm
case
V 90 mm
V 137 mm
depth
60 mm
60 mm
panel cutout
V 92+0.8 mm
V 138+1 mm
weight approx.
0.6 kg
0.7 kg
material of case
material of window
colour of bezel
position of use
mounting
terminals
Electrical Data
measuring unit
difference in voltages of two AC supplies
frequency range
50 ... 100 Hz
overload capacity (acc. to DIN EN 60 051)
continuously
1.2 times
5 s max.
2 times
measurement category CAT III
operating voltage
300 V
pollution level
2
enclosure code
IP 52 case '
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact
' for other ratings refer to "Options"
Please state primary rating of voltage transformer.
Scaling
pointer
pointer deflection
scale characteristics
scale division
scale length
bar / knife-edge pointer
0 ... 90°
initial scale range greatly expanded
coarse-fine
WQ 96 /0S
WQ 144 /0S
94 mm
146 mm
Accuracy at Reference Conditions
accuracy class
1.5 according to DIN EN 60 051
reference conditions
ambient temperature
position of use
input
frequency
wave form
others
23°C±1K
nominal position ±1°
rated measuring value
50 Hz
sinusoidal, distortion factor <5%
DIN EN 60 051
influences
ambient temperature
position of use
frequency
stray magnetic field
-25°C ... +23°C ... +40°C
nominal position ±5°
50 ... 100 Hz
0.5 mT
Environmental
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
shock resistance
vibration resistance
climatic class 2 '
according to VDE/VDI 3540 sheet 2
-25 ... +40°C '
-25 ... +65°C
75% annual average, non-condensing
15 g, 11 ms '
2.5 g, 5 ... 55 Hz '
Data Sheet
M Series
022.D.101.04
Analog Meters
Null Voltmeters with
MovingāĆāIron Movement
90°āĆāDial
Rules and Standards
DIN 43 700
measuring and control instruments
for panel mounting;
nominal case and cutout dimensions
DIN 43 701
electrical switchboard instruments
DIN 43 718
bezels and front panels
DIN 43 802
scales and pointers for electrical measuring
instruments
DIN 16 257
nominal position of use and
position symbols
applicable for measuring instruments
DIN 40 050
enclosure codes;
protection of electrical equipment against
ingress of solid foreign bodies and of water
DIN EN 60 051
direct acting indicating electrical measuring
instruments and their accessories
DIN EN 61 010
safety requirements for electrically operated
measuring, control and laboratory
equipment
VDE/VDI 3540 sheet 2 reliability of measuring and control
equipment (classification of climates)
Connections
L1
L2
mains supply
L3
17
18
*
L1
generator
L2
L3
* WQ 96 /0S
Dimensions
case
window
colour of bezel
position of use
non-glaring glass
gray (similar to RAL 7037)
horizontal or to be specified 15°...165°
performance
increased mechanical shock 30 g, 11 ms
loads
vibration 5 g, 5 ... 55 Hz
climatic suitability
limited use in the tropics climatic class 3
according to VDE/VDI 3540 sheet 2
with operating
-10 ... +55°C
temperature range
marine application
non-certified
enclosure code
IP 54 splash-water protected front
accessories
terminal protection against accidental contact
full-sized rear cover (WQ 96 /0S only) or protective sleeves SW6
terminals
connector blades 6.3 x 0.8
dial
blank dial
scale division
and figuring
additional lettering
additional figuring
coloured marks
coloured sector
logo on the dial
pencil-marked initial and end values
0 ... 100/200%
to be specified e.g. "generator"
to be specified
red, green or blue for important scale values
red, green or blue within scale division
none or as specified
b
a
Options
c
dimensions (in mm)
a
b
c
WQ 96 /0S
96
90
60
WQ 144 /0S
144
137
60
Ordering Information
type
WQ
front dimensions
96
144
version
/0S
measuring ranges
window
colour of bezel
position of use
mechanical loads
climatic suitability
marine application
enclosure code
terminal protection
terminals
dial
logo
*) standard
**) Please clearly add
***) WQ 96 /0S only
moving-iron panel meter
96 mm x 96 mm
144 mm x 144 mm
null voltmeter
refer to preceding table
glass *)
non-glaring glass
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
vertical *)
to special order 15 ... 165° **)
shock 15 g, vibration 2.5 g *)
shock 30 g, vibration 5 g
class 2, -25 ... +40°C *)
class 3, -10 ... +55°C
none *)
non-certified
IP 52 *)
IP 54 splash-water protected front
none *)
full-sized rear cover ***)
protective sleeves
screws and wire clamps *)
connector blades 6.3 x 0.8
scale division & measuring range alike *)
blank dial
scale division and figuring 0 ... 100/200%
additional lettering to be specified **)
additional figuring to be specified **)
coloured marks red, green or blue **)
coloured sector red, green or blue **)
WEIGEL *)
none
OEM logo **)
the desired specifications.
ordering example
WQ 96 /0S, measuring range and dial 0 ... 230/460 V,
window non-glaring glass, no logo
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
- specifications subject to change without notice; date of issue 12/06 -
Data Sheet
Edgewise Series
010.D.231.01
Analog Edgewise Meters
with MovingĆCoil
Movement for
Railway Applications
P 96 PrS Bahn
Application
enclosure code
The edgewise movingāĆācoil panel meters P 96 PrS Bahn with a curved
dial are used for measurement of DC currents or DC voltages.
The meters for railway application are specially used in rail vehicles
where robustness and reliability are mandatory.
Typical applications are display of rotational speed, temperature,
pressure, velocity or other physical quantities.
The movingāĆācoil movements are characterized by a low power
consumption, high precision and excellent damping, also in extreme
enviromental conditions.
Movements
MovingāĆācoil movement with coreāĆātype magnet and bilateral pivot
suspensions.
Mechanical Data
desiggn
case
front window
colour of bezel
position of use
panel fixing
mounting
terminals
voltmeters and
ammeters 3 A
ammeters >3 A
up to 25 A
ammeters >25 A
protection ground
edgewise case suitable to be mounted
in control panels, machine tool consoles
or mosaic panels, stackable
material pressed steel
surface thickāĆāfilm passivated
nonāĆglaring glass '
black (similar to RAL 9005) '
any
screw clamps DINāĆB
stackable next to each other
hexagon studs, M3 screws and wire clamps
hexagon studs, M5 screws and wire clamps
hexagon studs, M6 screws and wire clamps
connector blades 6.3 x 0.8
dimensions
bezel
96 mm x 48 mm
case
91 mm x 43 mm
case depth
99 mm with M3, 102 mm with M5, M6
including hexagon studs
panel depth
107 ... 118 mm
including clamps
(depends on panel thickness)
panel cutout
92+0.8 mm x 45+0.6 mm
panel thickness
1 ... 12 mm
weight approx.
0.45 kg
Electrical Data
measuring unit
DC voltage or DC current
overload capacity (according to DIN EN 60 051)
continuously
1.2 times rated voltage / current
5 s max.
voltmeters
2 times rated voltage
ammeters
10 times rated current
measurement category CAT III
operating voltage
refer to Measuring Ranges
pollution level
2
dial illumination
IP 52 case front '
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact
LED 24 V DC dimmable, approx. 1.2 VA
colour white '
Measuring Ranges
DC current '
voltage drop approx. operating voltage
1 mA
48 mV
1.5 mA
60 mV
2.5 mA
60 mV
4 mA
60 mV
5 mA
60 mV
6 mA
60 mV
10 mA
60 mV
15 mA
60 mV
20 mA
60 mV
25 mA
60 mV
40 mA
60 mV
60 mA
60 mV
100 mA
60 mV
150 mA
60 mV
250 mA
60 mV
400 mA
60 mV
600 mA
60 mV
1A
60 mV
1.5 A
60 mV
2.5 A
60 mV
4A
60 mV
6A
60 mV
10 A
60 mV
15 A
60 mV
25 A
60 mV
40 A
60 mV
for use on transducer
4 ... 20 mA
60 mV
mechanically suppressed zero
150 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
DC voltage '
operating voltage
sensitivity '
60 mV
1000 W/V ±20%
100 mV
1000 W/V ±20%
150 mV
1000 W/V ±20%
250 mV
1000 W/V ±20%
400 mV
1000 W/V ±20%
600 mV
1000 W/V ±20%
1V
1000 W/V ±20%
1.5 V
1000 W/V ±20%
2.5 V
1000 W/V ±20%
4V
1000 W/V ±20%
6V
1000 W/V ±20%
10 V
1000 W/V ±20%
15 V
1000 W/V ±20%
25 V
1000 W/V ±20%
40 V
1000 W/V ±20%
60 V
1000 W/V ±20%
72 V
1000 W/V ±20%
for use with external shunt
60 mV
1000 W/V ±20%
150 mV
1000 W/V ±20%
150 V
50 V
50 V
50 V
50 V
50 V
50 V
50 V
50 V
50 V
50 V
50 V
50 V
50 V
50 V
50 V
100 V
100 V
150 V
150 V
A total lead resistance of 0.05 W for interconnecting leads 1 m,
2 x 0.75 mm2 is considered in the calibration. '
' also refer to "Options"
Data Sheet
Edgewise Series
010.D.231.01
Analog Edgewise Meters
with MovingĆCoil
Movement for
Railway Applications
Scaling
DIN EN 61ā010āĆā1
pointer
pointer colour
response time
scale arrangement
scale characteristics
scale division
scale length
dial colour
scale figuring
bar pointer
orange to RAL 2007 '
1 s for fullāĆāscale deflection
vertical (bottom zero) '
linear
coarseāĆāfine
67 mm
black '
white '
DIN EN 61ā373
DIN IEC 61ā554
DIN 16 257
DIN 43 718
Accuracy at Reference Conditions
DIN 60ā529
accuracy class
Options
1.5 according to DIN EN 60 051āĆā1
reference conditions
ambient temperature 23°C±1K
input
rated measuring value
others
DIN EN 60 051āĆā1
influences
ambient temperature
stray magnetic field
-25°C ... +23°C ... +40°C
0.5 mT
Environmental
climatic suitability
operating
temperature range
storage
temperature range
category 1, class B
according to DIN EN 61 373
-25 ... +55°C
-25 ... +65°C
Rules and Standards
DIN EN 50ā121 ...
āĆā1
āĆā3āĆā2
āĆā4
DIN EN 50155
DIN EN 60ā051 ...
āĆā1
āĆā2
DIN EN 60ā068āĆā1
DIN EN 60068 ...
āĆā2āĆā1
āĆā2āĆā2
āĆā2āĆā27
āĆā2āĆā30
āĆā2āĆā47
āĆā2āĆā64
Railway applications Electromagnetic compatibility
Part 1:
General
Part 3āĆā2: Rolling stock - Apparatus
Part 4:
Emission and immunity
of the signalling and
telecommunications apparatus
Railway applications Electronic equipment used on rolling stock
Direct acting indicating analogue electrical
instruments and their accessories
Part 1:
Definitions and general
requirements common to all parts
Part 2:
Special requirements for ammeters
and voltmeters
Environmental testing Part 1: General and guidance
Environmental testing - Tests Part 2āĆā1: Test A: Cold
Part 2āĆā2: Test B: Dry heat
Part 2āĆā27:Test Ea and guidance: Shock
Part 2āĆā30:Test Db: Damp heat,
cyclic (12 h + 12 h cycle)
Part 2āĆā47 Mounting of specimens for
vibration, impact and
similar dynamic tests
Part 2āĆā64:Test Fh: Vibration,
broadband random and guidance
Safety requirements for electrical equipment
for measurement, control, and laboratory use
Part 1: General requirements
Railway applications Rolling stock equipment Shock and vibration tests
category 1 class B
Devices for mounting in control stands
Electrical measuringg devices
Dimensions for mounting in control stands
Nominal position of use and position symbols
applicable for measuring instruments
Measuring, control, regulate; bezels and
front panels for MSR devices;
main dimensions
Enclosure codes by housings (IPāĆācode)
measuring range
special measuring
range
sensitivity adjustment
lead resistance
to ±1% at 23°C
calibration of a total value >0.05W
case
front window
colour of bezel
glass
gray (similar to RAL 7037)
performance
enclosure code
on request
IP 55 splashāĆāwater protected front
(with rear zero adjustment)
accessories
terminal protection
against accidental
contact
protective sleeves for hexagon studs
SW6 (for M3 screws)
SW10 (for M5/M6 screws)
pointer
other colour on request
dial
scale arrangement
dial background
scale figuring
dial illumination
scale division
and figuring
additional lettering
additional figuring
coloured marks
coloured sector
logo on the dial
zero position
horizontal (leftāĆāhand zero)
other colour on request
other colour on request
other colour on request
0 ... 100%,
linear, fullāĆāscale values acc. to standardized
series (1 Ć 1.2 Ć 1.5 Ć 2 Ć 2.5 Ć 3 Ć 4 Ć 5 Ć 6 Ć 7.5
and their decimal multiples e.g. 150 m3/h) or
deviating from standard;
special calibration from
nonāĆālinear graph or chart;
scaling of voltmeters in ohms;
captions on request
on request e.g. "generator"
on request
red, green or blue for important scale values
red, green or blue within scale division
none or on request
centre zero or offāĆāset zero,
mechanically suppressed zero, no zero
adjustment, max. 40% of fullāĆāscale value
' for other ratings refer to "Options"
Connections
Ordering Information
DC voltage
Type
P 96 PrS Bahn
DC current
illumination
measuring input
ilumination
measuring input
-
-
-
-
+
+
+
+
14
15
+
17
-
+
18
14
-
+
15
17
-
+
18
-
b
d
Dimensions
c
a
dimensions P 96 PrS Bahn
DINāĆāB panel fixing
a
96 mm
b
48 mm
c
99 mm (M3)
102 mm (M5, M6)
d
43 mm
DINāĆāB panel fixing mounting details
7 ... 18 mm
1 ... 12 mm
107 ... 118 mm
scales and position of use
II
I
a
a
III
IV
VI
V
vertical mounting
I
horizontal scale
II vertical scale
inclined mounting
III horizontal scale
IV vertical scale
horizontal mounting
V horizontal scale
VI vertical scale
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
edgewise movingāĆācoil panel meter
for railway applications
for DC voltage or DC current
96 mm x 48 mm
measuring ranges
refer to preceding table
sp. measuring range on request **)
adjustment
to internal resistance to ±20% *)
to internal resistance to ±1% at 23°C
for lead resistance >0.05 W
front window
nonāĆāglaring glass *)
glass
colour of bezel
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
enclosure code
IP 52 *)
(case front)
IP 55 splashāĆāwater protected front
terminal protection
none *)
protective sleeves SW6 or SW10
scale arrangement
vertical *)
horizontal
dial
scale division & measuring range alike *)
scale division and figuring
according to standardized series **)
0 ... 100%
linear deviating from standard **)
calibration fr. nonāĆālinear graph or chart **)
scaling in ohms for voltmeters **)
additional lettering on request **)
additional figuring on request **)
coloured marks red, green or blue **)
coloured sector red, green or blue **)
ponter colour
orange *)
other colour on request **)
dial colour
black *)
other colour on request **)
figuring colour
white *)
other colour on request **)
dial illumination
white *)
other colour on request **)
logo
WEIGEL *)
none
OEM logo **)
zero position
bottom or leftāĆāhand zero *)
centre zero or offāĆāset zero **)
mechanically suppressed zero **)
*)
**)
standard
Please clearly add the desired specifications.
ordering example
P 96 PrS Bahn, measuring range 0 ... 20 mA, horizontal scale 0 ... 100%,
front window nonāĆāglaring glass, OEM logo (template included)
- specifications subject to change without notice; date of issue 02/10 -
Data Sheet
M Series
030.D.101.03
Frequency Meters
with Vibrating Reed
Movement
FQ 48
FQ 72
FQ 96
FQ 144
Application
Measuring Ranges
The frequency indicators FQ 48/72/96/144 (M series) are used to meaĆ
sure frequencies of power supplies and/or generator sets in spans of
rated frequencies 50 Hz or 60 Hz.
frequency ranges
The indicators are suitable to be mounted in switchboards, generator
sets, machine tool consoles or mosaic grid panels.
Movement
Vibrating reed movement.
Each reed is tuned to a different value in the frequency span.
Mechanical Data
case details
material of case
material of window
colour of bezel
position of use
panel fixing
panel thickness
mounting
terminals
square case suitable to be mounted
in control / switchgear panels, machine tool
consoles or mosaic grid panels, stackable
thermoplastics, flame retardant (FQ 48)
pressed steel (FQ 72/96/144)
glass '
black (similar to RAL 9005) '
vertical ±5° '
plate springs (FQ 48)
WEIGEL screw clamps (FQ 72/96/144)
1 ... 15 mm
stackable next to each other
hexagon studs, M3 x 6 screws and
wire clamps C6 '
dimensions (in mm)
FQ 48
FQ 72
FQ 96
FQ 144
bezel
V 48
V 72
V 96
V 144
case
V 45
V 66.5
V 90.5
V 137
depth
61
52
58
58
panel cutout
V45.2+0.3 V68.3+0.4 V92+0.8
V138+1
weight approx.
0.15 kg
0.8 kg
0.3 kg
0.4 kg
Electrical Data
measuring unit
frequency
variation permissible rated voltage ±10%
overload capacity (acc. to DIN EN 60 051)
continuously
1.2 times rated voltage
measurement category CAT III
operating voltage
refer to Measuring Ranges
pollution level
2
enclosure code
IP 52 case '
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact
' for other ratings refer to "Options"
47 ... 50 ... 53 Hz
47 ... 50 ... 53 Hz
44 ... 50 ... 56 Hz
45 ... 50 ... 55 Hz
57 ... 60 ... 63 Hz
57 ... 60 ... 63 Hz
54 ... 60 ... 66 Hz
55 ... 60 ... 65 Hz
resolution
number of reeds
1 Hz
½ Hz
1 Hz
½ Hz
1 Hz
½ Hz
1 Hz
½ Hz
7
13
13
21
7
13
13
21
rated voltage
operating voltage
approx. power FQ 48
FQ 72
consumption
100 V
110 V
230 V
400 V
500 V
600 V
0.4 VA
0.5 VA
1.0 VA
1.5 VA
<3 VA
<3 VA
600 V
600 V
600 V
600 V
600 V
600 V
100 V
600 V
600 V
600 V
600 V
600 V
(FQ 48 only)
(except FQ 48)
(except FQ 48)
(except FQ 48/72)
(FQ 48 only)
(except FQ 48)
(except FQ 48)
(except FQ 48/72)
FQ 96
FQ 144
300 V
300 V
300 V
300 V
300 V
600 V
300 V
300 V
300 V
300 V
300 V
600 V
Scaling
scale division
reed arrangement
coarse-fine
horizontal
Accuracy at Reference Conditions
accuracy class
0.5 acc. to DIN EN 60 051
reference conditions
ambient temperature
position of use
input
others
23°C±1K
nominal position ±1°
rated measuring value
DIN EN 60 051
influences
ambient temperature
position of use
stray magnetic field
-25°C ... +23°C ... +40°C
nominal position ±5°
0.5 mT
Environmental
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
shock resistance
vibration resistance
climatic class 2 acc. to VDE/VDI 3540 sheet 2 '
-25 ... +40°C '
-25 ... +65°C
75% annual average, non-condensing
15 g, 11 ms '
2.5 g, 5 ... 55 Hz '
Data Sheet
M Series
030.D.101.03
Frequency Meters
with Vibrating Reed
Movement
DIN 43 700
17
18
L1
N(L2)
Dimensions
a
measuring and control instruments
for panel mounting;
nominal case and cutout dimensions
DIN 43 701
electrical switchboard instruments
DIN 43 718
bezels and front panels
DIN 16 257
nominal position of use and
position symbols
applicable for measuring instruments
DIN 40 050
enclosure codes;
protection of electrical equipment against
ingress of solid foreign bodies and of water
DIN EN 60 051
direct acting indicating electrical measuring
instruments and their accessories
DIN EN 61 010
safety requirements for electrically operated
measuring, control and laboratory
equipment
VDE/VDI 3540 sheet 2 reliability of measuring and control
equipment (classification of climates)
Connections
b
Rules and Standards
Options
case
window
colour of bezel
position of use
non-glaring glass
gray (similar to RAL 7037)
to be specified 15°...165°
performance
increased mechanical shock 30 g, 11 ms
loads
vibration 5 g, 5 ... 55 Hz
climatic suitability
limited use in the tropics climatic class 3
according to VDE/VDI 3540 sheet 2
with operating
-10 ... +55°C
temperature range
marine application
non-certified
enclosure code
IP 54 splash-water protected front
accessories
projection mounting outfit
terminal protection against accidental contact
full-sized rear cover or protective sleeves SW6
terminals
connector blades 6.3 x 0.8
dial
additional lettering
additional figuring
coloured marks
coloured sector
logo on the dial
to be specified e.g. "generator"
to be specified
red, green or blue for important scale values
red, green or blue within scale division
none or to be specified
c
dimensions (in mm)
a
b
c
FQ 48
48
45
61
FQ 72
72
66.5
52
FQ 96
96
90.5
58
FQ 144
144
137
58
Ordering Information
type
FQ
front dimensions
48
72
96
144
measuring ranges
rated voltage
window
colour of bezel
position of use
performance loads
climatic suitability
marine application
enclosure code
terminal safety
protection
terminals
dial
logo
*)
**)
reed type frequency indicator
48 mm x 48 mm
72 mm x 72 mm
96 mm x 96 mm
144 mm x 144 mm
refer to preceding table
refer to preceding table
glass *)
non-glaring glass
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
vertical *)
to special order 15 ... 165° **)
shock 15 g, vibration 2.5 g *)
shock 30 g, vibration 5 g
class 2, -25 ... +40°C *)
class 3, -10 ... +55°C
none *)
non-certified
IP 52 *)
IP 54 splash-water protected front
none *)
full-sized rear cover ***)
protective sleeves SW6
screws M3 x 6 and wire clamps *)
connector blades 6.3 x 0.8
scale division & measuring range alike *)
additional lettering to be specified **)
additional figuring to be specified **)
coloured marks red, green or blue **)
coloured sector red, green or blue **)
WEIGEL *)
none
OEM logo **)
standard
Please clearly add the desired specifications.
ordering example
FQ 144, measuring range 47 ... 50 ... 53 Hz, 13 reeds,
rated voltage 230 V~ , window non-glaring glass,
WEIGEL logo
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
- specifications subject to change without notice; date of issue 06/06 -
Data Sheet
K Series
470.D.101.07
Analog
Watt and VAr Meters,
Electronic,
90° or 240°ĂDial
LQ 96 K
LQ 144 K
LSL 96 K
with SlideāĆāInāĆāDial
Application
Electrical Data
The electronic Watt and VAr movingāĆācoil meter models LQ 96/144 K
with 90° dial or LSL 96 K with 240° dial of the K series are offered for the
following AC systems:
- single phase,
- 3 phase balanced load, 3 or 4 wire,
- 3 phase unbalanced load,3 or 4 wire.
These wattmeters are suitable to indicate forward (export) and reverse
(import) power flow as well as inductive and capacitive reactive power.
They can be used both on sinusoidal and non-sinusoidal current.
The instruments are suitable to be mounted in switchboards, control
panels, machine tool consoles and mosaic panels. The bezel, the glass
window and the dial can be easily exchanged on-site.
measuring unit
active or reactive power
response time
4s
overload capacity (acc. to DIN EN 60 051)
continuously
1.2 times rated voltage / current
5 s max.
2 times rated voltage,
10 times rated current
power consumption
0.2 VA / each
current path
voltage path types
EW 1, DW 1, DB 1,
3.0 VA / each
VW 1, VB 1
3.5 VA / each
EB 1
3.4 VA / each
DW 2, DB 2
3.9 VA / each
VW 3
4.3 VA / each
VB 3
pollution level
2
operating voltage
refer to Measuring Ranges
measurement category CAT III
enclosure code
IP 52 case
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact '
Functional Principle
The meters consist of a movingāĆācoil movement with coreāĆāmagnet (LQ)
or pivot suspended spring loaded jewel bearings (LSL) system and a
power converter. Both devices are included in a common plastic case.
4
block circuit diagram
=
2
U
J
~
1
U. I = P
1
T
T
3
P dt
0
The power converter uses one, two or three multiplier systems 2 deĆ
pending on the measurement of balanced or unbalanced load AC sysĆ
tems. Current transformers 1 adapt the input current to the multiplier
electronics.
The multipliers form the product of the instantaneous values of current
and voltage (TDM principle). Subsequently, the product resultant is inĆ
tegrated, thereby suppressing the AC ripple. A DC voltage output sigĆ
nal is fed to the movingāĆācoil movement 3.
Power supply is obtained from voltage input in block 4.
Mechanical Data
case details
material of case
material of window
colour of bezel
position of use
panel fixing
mounting
panel thickness
terminals
dimensions
bezel
case
depth
depth
panel cutout
weight approx.
moulded square case suitable to be mounted
in control / switchgear panels, machine tool
consoles or mosaic panels, stackable
polycarbonate thermoplastics,
flame retardant with UL rating of 94 V - 0
glass '
black (similar to RAL 9005) '
vertical ±5° '
screw clamps
stackable next to each other
40 mm
hexagon studs with M4 screws
LQ 96 K
LQ 144 K
LSL 96 K
V 96 mm
V 144 mm
V 96 mm
V 90 mm
V 136 mm
V 90 mm
129 mm
129 mm
129 mm
VW/B 3 versions
all versions
104 mm
104 mm
EW/B1, DW/B 1, VW/B 1, DW/B 2 versions
V92+0.8 mm
V92+0.8 mm V138+1 mm
1.1 kg
1.1 kg
1.1 kg
Measuring Ranges
type
active
power
reactive
power
single phase system
3 phase 3 wire system balanced load
3 phase 4 wire system balanced load
3 phase 3 wire system unbalanced load
3 phase 4 wire system unbalanced load
EW 1
DW 1
VW 1
DW 2
VW 3
EB 1
DB 1
VB 1
DB 2
VB 3
selection of measuring range
The apparent power PS is calculated from the primary ratings of current
transformers and voltage transformers:
single phase
PS = U . I
3 phase
PS = 3 . U . I
Select full-scale values between 0.5 and 1.2 times the calculated apĆ
parent power preferably from DIN series (acc. to DIN 43 701): 1 - 1.2 1.5 - 2 - 2.5 - 3 - 4 - 5 - 6 -7.5 - 8 and their decimal multiples.
single phase system
3 phase 3 wire system
rated
operating voltage
voltage
LQ 96 K 144 K
LSL
96 K
57.7 V (100 V :
63.5 V (110 V :
100 V
115 V
120 V
127 V (220 V :
230 V (400 V :
289 V (500 V :
400 V
440 V
500 V
rated current
3 phase 4 wire system
rated
operating voltage
voltage
LQ 96 K 144 K
LSL
96 K
3 ) 150 V 150 V
57.7/100 V
150 V 150 V
3 ) 150 V 150 V
63.5/110 V
150 V 150 V
150 V 150 V
150 V 150 V
150 V 150 V
3 ) 150 V 150 V
127/220 V
150 V 150 V
3 ) 300 V 600 V
3 ) 300 V 600 V
600 V 600 V
230/400 V
600 V 600 V
600 V 600 V
254/440 V
600 V 600 V
600 V 600 V
289/500 V
600 V 600 V
1 A or 5 A
If used on current transformer, please state
transformer ratio on the order.
' also refer to "Options"
Data Sheet
K Series
470.D.101.07
Analog
Watt and VAr Meters,
Electronic,
90° or 240°ĂDial
Scaling
Rules and Standards
DIN 43 700
bar / knife-edge pointer
0 ... 90° (LQ)
0 ... 240° (LSL)
linear
coarse-fine
LQ 96 K
LQ 144 K
97 mm
146 mm
scale characteristics
scale division
scale length
LSL 96 K
142 mm
Accuracy at Reference Conditions
accuracy class
1.5 according to DIN EN 60 051
reference conditions
ambient temperature
position of use
input
23°C±2K
nominal position ±1° '
full-scale power value PN
calibration factor
l = PN / PS
cos
=l / 0.6 resp. sin
for 0.3 l < 0.6
power factor
cos
=1 resp. sin
for 0.6 l 1.5
voltage
frequency
warm-up
others
=l / 0.6
=1
rated voltage
50 Hz ±2%
15 min
DIN EN 60 051
influences
ambient temperature
position of use
stray magnetic field
power factor
-10°C ... +23°C ... +55°C
nominal position ±5°
0.5 mT
1 ind ... 0 ... 1 cap
Environmental
climatic suitability
climatic class 3 acc. to VDE/VDI 3540 sheet 2
operating
temperature range
storage
temperature range
relative humidity
shock resistance
vibration resistance
-10 ... +55°C
-25 ... +65°C
75% annual average, non-condensing
15 g, 11 ms
2.5 g, 5 ... 55 Hz
Options
case
window
colour of bezel
index marking pointer
position of use
marine application
non-glaring glass
gray (similar to RAL 7037)
red, front adjustable
on request 15°...165°
non-certified or with approbation
by "Germanischer Lloyd" (LQ 96/144 K only)
terminal protection against accidental contact
protective sleeves
protection against accidental contact (hand and fingers)
acc. to VBG 4 / DIN 57 106, sec. 100
dial
additional lettering
additional figuring
coloured marks
coloured sector
logo on the dial
on request e.g. "generator"
on request
red, green or blue for important scale values
red, green or blue within scale division
none or on request
Dimensions
b
pointer
pointer deflection
measuring and control instruments
for panel mounting;
nominal case and cutout dimensions
DIN 43 701
electrical switchboard instruments
DIN 43 718
bezels and front panels
DIN 43 802
scales and pointers for electrical measuring
instruments
DIN 16 257
nominal position of use and
position symbols
applicable for measuring instruments
DIN 40 050
enclosure codes;
protection of electrical equipment against
ingress of solid foreign bodies and of water
DIN EN 60 051
direct acting indicating electrical measuring
instruments and their accessories
DIN EN 61 010
safety requirements for electrically operated
measuring, control and laboratory
equipment
VDE/VDI 3540 sheet 2 reliability of measuring and control
equipment (classification of climates)
b
flat dial
a
dial
c
LQ/LSL 96 K
c
LQ 144 K
dimensions (in mm) LQ 96 K
LQ 144 K
a
96
144
b
90
136
c
104
104
(EW/B1, DW/B 1, VW/B 1, DW/B 2 versions)
c
129
129
(VW/B 3 versions)
LSL 96 K
96
90
129
129
Connections
Ordering Information
LQ/LSL 96/144 K EW1/EB1
LQ/LSL 96/144 K DW1
LQ/LSL 96/144 K VW1
LQ/LSL 96/144 K DB1
LQ/LSL 96/144 K VB1
LQ/LSL 96/144 K DW2/DB2
LQ/LSL 96/144 K VW3
LQ/LSL 96/144 K VB3
type
LQ
LSL (96 K only)
front dimensions
96 K
144 K
type
EW1, EB1
DW1, DB1
VW1, VB1
DW2, DB2
VW3, VB3
Watt and VAr meters, electronical
with movingāĆācoil movement. 90° dial
with movingāĆācoil movement. 240° dial
96 mm x 96 mm
144 mm x 144 mm
single phase system
3 phase 3 wire system balanced load
3 phase 4 wire system balanced load
3 phase 3 wire system unbalanced load
3 phase 4 wire system unbalanced load
measuring ranges
rated voltages
rated currents
refer to preceding table
refer to preceding table
1A
5A
window
glass *)
non-glaring glass
colour of bezel
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
index marking pointer none *)
red, front adjustable
position of use
vertical *)
on request 15 ... 165° **)
marine application
none *)
non-certified
with approbation by
"Germanischer Lloyd" ***)
terminal protection
none *)
protective sleeves
dial
scale division & measuring range alike *)
additional lettering on request **)
additional figuring on request **)
coloured marks red, green or blue **)
coloured sector red, green or blue **)
logo
WEIGEL *)
none
OEM logo **)
*) standard
**) Please clearly add
***) LQ 96/144 K only
the desired specifications.
ordering example
LQ 96 K VW3 for active power, 3 phase 4 wire system, unbalanced load,
measuring range 0 ... 400 kW, rated voltage AC 230/400 V,
for use on current transformer 600/5 A,
window non-glaring glass, no logo
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
- specifications subject to change without notice; date of issue 12/05 -
Data Sheet
K Series
460.D.101.06
Analog
Power Factor Meters,
Electronic,
90° or 240°ĂDial
CQ 96 K
CQ144 K
LSC 96 K
with SlideāĆāInāĆāDial
Application
Electrical Data
The movingāĆācoil meter models CQ 96/144 K with 90° dial or LSC 96 K
with 240° dial of the K series are suitable to measure the power factor as
a ratio of active and reactive power in single phase AC or in balanced 3
phase systems:
The instruments are suitable to be mounted in switchboards, control
panels, machine tool consoles and mosaic panels. The bezel, the glass
window and the dial can be easily exchanged on-site.
Functional Principle
The meters consist of a movingāĆācoil movement with coreāĆāmagnet (CQ)
or pivot suspended spring loaded jewel bearings (LSC) system and a
measuring converter. Both devices are included in a common plastic
case.
block circuit diagram
4
)
overload capacity (acc. to DIN EN 60 051)
continuously
1.2 times rated voltage / current
5 s max.
2 times rated voltage,
10 times rated current
power consumption
current path
0.1 VA
voltage path
3.0 VA
pollution level
operating voltage
measurement category
enclosure code
2
refer to Measuring Ranges
CAT III
IP 52 case
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact
Measuring Ranges
1
type
Mechanical Data
material of window
colour of bezel
position of use
panel fixing
mounting
panel thickness
terminals
moulded square case suitable to be mounted
in control / switchgear panels, machine tool
consoles or mosaic panels, stackable
polycarbonate thermoplastics,
flame retardant with UL rating of 94 V - 0
glass '
black (similar to RAL 9005) '
vertical ±5° '
screw clamps
stackable next to each other
40 mm
hexagon studs with M4 screws
dimensions
bezel
case
depth
panel cutout
weight approx.
CQ/LSC 96 K
V 96 mm
V 90 mm
104 mm
V92+0.8 mm
0,55 kg
material of case
49 ... 50 ... 51 Hz (single phase system)
45 ... 50 ... 65 Hz (3 phase system)
3
A current transformer 1 of the phase angle converter adapts the input
current to the electronic circuit. Both the input voltage and the current
are passed to a bistable flip-flop stage 2.
The pulse duty cycle of the flip-flop is proportional to the phase
angle . A low-pass filter forms the mean value which is fed to the
moving-coil movement 3. The standard dial is scaled with the cosinus
of the phase angle .
Power supply is obtained from voltage input in block 4.
case details
power factor (phase angle
=
2
U
I
~
measuring unit
frequency range
' also refer to "Options"
CQ 144 K
V 144 mm
V 136 mm
104 mm
V138+1 mm
0,75 kg
E
D
single phase system
3 phase 3 wire system balanced load
measuring ranges
cos
cos
cos
cap 0.5 ... 1 ... 0.5 ind
cap 0.8 ... 1 ... 0.3 ind
cap 0.8 ... 1 ... 0.8 ind
single phase system
rated
operating voltage
voltages
CQ 96 K 144 K
LSC
96 K
57.7 V (100 V :
63.5 V (110 V :
100 V *)
110 V *)
115 V
120 V
127 V (220 V :
208 V
230 V
289 V (500 V :
400 V
*)
3 ) 150 V
3 ) 150 V
150 V
150 V
150 V
150 V
3 ) 150 V
300 V
300 V
3 ) 600 V
600 V
150 V
150 V
150 V
150 V
150 V
150 V
150 V
600 V
600 V
600 V
600 V
also for use on voltage transformer
rated currents
1A
5A
3 phase system
rated
operating voltage
voltages
CQ 96 K 144 K
LSC
96 K
100 V *)
110 V *)
115 V
120 V
127 V
208 V
230 V
289 V
400 V
415 V
440 V
500 V
150 V
150 V
150 V
150 V
150 V
300 V
300 V
600 V
600 V
600 V
600 V
600 V
150 V
150 V
150 V
150 V
150 V
600 V
600 V
600 V
600 V
600 V
600 V
600 V
Data Sheet
K Series
460.D.101.06
Analog
Power Factor Meters,
Electronic,
90° or 240°ĂDial
Scaling
dial
pointer
pointer deflection
scale characteristics
scale division
scale length
Rules and Standards
flat dial
bar / knife-edge pointer
0 ... 90° (CQ)
0 ... 240° (LSC)
nonāĆālinear
coarse - fine
CQ 96 K
CQ 144 K
97 mm
146 mm
DIN 43 700
LSC 96 K
142 mm
Accuracy at Reference Conditions
accuracy class
1.5 according to DIN EN 60 051
reference conditions
ambient temperature
position of use
voltage
frequency
wave form
distortion factor
current
warmāĆāup
others
23°C±2K
nominal position ±1° '
rated voltage
50 Hz ±0.1%
sine wave
0.1%
95 ... 100 % rated current
5 min
DIN EN 60 051
influences
ambient temperature
position of use
stray magnetic field
-10°C ... +23°C ... +55°C
nominal position ±5°
0.5 mT
Environmental
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
shock resistance
vibration resistance
climatic class 3 acc. to VDE/VDI 3540 sheet 2
-10 ... +55°C
-25 ... +65°C
75% annual average, non-condensing
15 g, 11 ms
2.5 g, 5 ... 55 Hz
measuring and control instruments
for panel mounting;
nominal case and cutout dimensions
DIN 43 701
electrical switchboard instruments
DIN 43 718
bezels and front panels
DIN 43 802
scales and pointers for electrical measuring
instruments
DIN 16 257
nominal position of use and position symbols
applicable for measuring instruments
DIN 40 050
enclosure codes;
protection of electrical equipment against
ingress of solid foreign bodies and of water
DIN EN 60 051
direct acting indicating electrical measuring
instruments and their accessories
DIN EN 61 010
safety requirements for electrically operated
measuring, control and laboratory
equipment
VDE/VDI 3540 sheet 2 reliability of measuring and control
equipment (classification of climates)
Options
case
window
colour of bezel
index marking pointer
position of use
marine application
non-glaring glass
gray (similar to RAL 7037)
red, front adjustable
on request 15°...165°
non-certified or with approbation
by "Germanischer Lloyd" (CQ 96/144 K only)
terminal protection against accidental contact
protective sleeves
protection against accidental contact (hand and fingers)
acc. to VBG 4 / DIN 57 106, sec. 100
dial
additional lettering
additional figuring
coloured marks
coloured sector
logo on the dial
on request e.g. "generator"
on request
red, green or blue for important scale values
red, green or blue within scale division
none or on request
Connections
Ordering Information
type
CQ
LSC (96 K only)
front dimensions
96 K
144 K
type
E
D
CQ 96/144 K E, LSC 96 K E
power factor meter, electronical
with movingāĆācoil movement. 90° dial
with movingāĆācoil movement. 240° dial
96 mm x 96 mm
144 mm x 144 mm
single phase system
3 phase system balanced load
measuring ranges
cap 0.5 ... 1 ... 0.5 ind
cap 0.8 ... 1 ... 0.3 ind
cap 0.8 ... 1 ... 0.8 ind
rated voltages
refer to preceding table
rated currents
1A
5A
window
glass *)
non-glaring glass
colour of bezel
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
index marking pointer none *)
red, front adjustable ***)
position of use
vertical *)
on request 15 ... 165° **)
marine application
none *)
non-certified
with approbation by
"Germanischer Lloyd" ***)
safety terminal
none *)
touch protection
protective sleeves
dial
scale division & measuring range alike *)
additional lettering on request **)
additional figuring on request **)
coloured marks red, green or blue **)
coloured sector red, green or blue **)
logo
WEIGEL *)
none
OEM logo **)
CQ 96/144 K D, LSC 96 K D
b
a
b
Dimensions
c
CQ/LSC 96 K
dimensions (in mm)
a
b
c
CQ/LSC 96 K
96
90
104
c
CQ 144 K
CQ 144 K
144
136
104
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
*) standard
**) Please clearly add
***) CQ 96/144 K only
the desired specifications.
ordering example
CQ 96 K D for 3 phase system balanced load,
measuring range (cos ) cap 0.5 ... 1 ... 0.5 ind, rated voltage AC 230 V,
rated current 1 A, window non-glaring glass, no logo
- specifications subject to change without notice; date of issue 12/05 -
Data Sheet
Edgewise Series
020.D.201.03
Analog Meters
Edgewise with
Moving-Iron Movement
W 72 PrS
W 96 PrS
W144 PrS
Application
Electrical Data
The edgewise movingāĆāiron panel meters W 72/96/144 PrS with a
curved dial are mainly used for the measurement of AC currents and
voltages in the usual technical frequency range of 162/3 ... 100 Hz. SpeĆ
cial calibration for a definite frequency up to 1000 Hz on request.
MovingāĆiron meters indicate rms values practically independent of
wave form even of high harmonics. Error of indication may occur for
extreme wave forms (e.g. phase gating controls) and / or frequencies
above 100 Hz.
measuring unit
AC voltages or AC currents
frequency range
162/3 ... 100 Hz
power consumption
W 72/96 PrS
W 144 PrS
voltmeters
<4 VA
<4 VA
ammeters 15 A
<0.5 VA
<1,7 VA
ammeters >15 A
<0.8 VA
overload capacity (acc. to DIN EN 60 051)
continuously
1.2 times rated voltage / current
5 s max.
2 times rated voltage,
10 times rated current
measurement category CAT III
operating voltage
refer to Measuring Ranges
pollution level
2
enclosure code
IP 52 case '
IP 00 for terminals without protection against
accidental contact
IP 20 for terminals protected against
accidental contact
These meters are not suitable for use with shunts or tachogenerators
due to their high power consumption.
They are suitable to be mounted in switchboards, control panels, maĆ
chine tool consoles and mosaic panels.
Movements
MovingāĆiron movement with pivot suspension. Spring loaded jewel
bearings and silicon oil damping for vibration and shock resistance.
Mechanical Data
case details
material of case
material of window
colour of bezel
position of use
panel fixing
mounting
terminals
voltmeters and
ammeters 3 A
ammeters >3 A
voltmeters 600 V
edgewise case suitable to be mounted
in control / switchgear panels, machine tool
consoles or mosaic panels, stackable
pressed steel (W 72/96 PrS)
thermoplastics (W 144 PrS)
glass '
black (similar to RAL 9005) '
vertical ±5°
screw clamps
stackable next to each other
(except W 144 PrS)
hexagon studs, M3 screws and
wire clamps C6 (W 72/96 PrS),
connector blades 6.3 x 0.8 (W 144 PrS)
hexagon studs, M5 screws and
wire clamps C10
connector blades 6.3 x 0.8 for protective wire
(W 72/96 PrS)
dimensions (in mm)
W 72 PrS
W 96 PrS
W 144 PrS
bezel
72 x 36
96 x 48
144 x 72
case
66 x 32
91 x 43
137 x 67
depth
94
107
192
panel cutout
68+0.7āxā33+0.6 92+0.8āxā45+0.6 138+1.0
x 68+0.7
panel thickness
weight approx.
1 ... 25
0.28 kg
1 ... 12
0.45 kg
40
1.0 kg
Measuring Ranges
measuring ranges
operating voltage
AC current 1)
W 72 PrS W 96 PrS W 144 PrS
100; 150; 250; 400; 600 mA
1; 1.5; 2.5; 4; 5; 6; 10; 15; 25*) A 300 V
300 V
600 V
AC voltage
W 72 PrS W 96 PrS W 144 PrS
6; 10; 15; 25; 40; 60; 100; 150 V 100 V
100 V
100 V
250 V, 400 V, 500 V
300 V
300 V
600 V
600 V
600 V
600 V
600 V
for use on VT/CT
W 72 PrS W 96 PrS W 144 PrS
N/1 A, N/5 A 1)
100 V
100 V
600 V
N/100 V, N/110 V 2)
100 V
100 V
100 V
Please state transformer ratio when ordering.
*) W 72/96 PrS only
1) full scale value = 2 times rated value (overload scaling) '
2) full scale value = 1.2 times rated value (
-" - )
Scaling
pointer
response time
scale arrangement
scale characteristics
scale division
scale length
bar / knife-edge pointer
1 s for full-scale deflection
horizontal (left-hand zero)
practically linear
down to 1/5 th of rated full-scale value.
The initial scale is compressed.
coarse-fine
W 72 PrS W 96 PrS W 144 PrS
45 mm
67 mm
92 mm
overload scaling
ammeters
2 times rated current
voltmeters for use on 1.2 times rated voltage
voltage transformers
' for other ratings refer to "Options"
Data Sheet
Edgewise Series
020.D.201.03
Analog Meters
Edgewise with
Moving-Iron Movement
Accuracy at Reference Conditions
Options
accuracy class
measuring range
special measuring
range
accuracy class
calibration
1.5 according to DIN EN 60 051
reference conditions
ambient temperature
position of use
input
frequency
wave form
23°C±1K
nominal position ±1°
rated measuring value
162/3 ... 100 Hz
sinusoidal, distortion factor <5%
others
DIN EN 60 051
influences
ambient temperature
position of use
frequency
-25°C ... +23°C ... +40°C
nominal position ±5°
100 Hz
stray magnetic field
0.5 mT
Environmental
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
shock resistance
vibration resistance
climatic class 2 '
according to VDE/VDI 3540 sheet 2
-25 ... +40°C '
-25 ... +65°C
75% annual average, non-condensing
15 g, 11 ms '
2.5 g, 5 ... 55 Hz '
Rules and Standards
DIN 43 700
measuring and control instruments
for panel mounting;
nominal case and cutout dimensions
DIN 43 701
electrical switchboard instruments
DIN 43 718
bezels and front panels
DIN 16 257
nominal position of use and position symbols
applicable for measuring instruments
DIN 40 050
enclosure codes;
protection of electrical equipment against
ingress of solid foreign bodies and of water
DIN EN 60 051
direct acting indicating electrical measuring
instruments and their accessories
DIN EN 61 010
safety requirements for electrically operated
measuring, control and laboratory
equipment
VDE/VDI 3540 sheet 2 reliability of measuring and control
equipment (classification of climates)
case
window
colour of bezel
position of use
performance
climatic suitability
with operating
temperature range
marine application
enclosure code
deviating from standard
1.0 with fine scale division (as far as possible)
for a definite frequency 100 ... 1000 Hz
non-glaring glass
gray (similar to RAL 7037)
horizontal or on request 15 ... 165°
limited use in the tropics climatic class 3
according to VDE/VDI 3540 sheet 2
-10 ... +55°C
non-certified
IP 54 splash-water protected front
(without zero adjustment)
accessories
terminal protection against accidental contact
protective sleeves SW6, SW10 (ammeters >3A) for W 72/96 PrS
dial
scale arrangement
blank dial
scale division
and figuring
additional lettering
additional figuring
coloured marks
coloured segment
logo on the dial
zero position
overload scaling
(ammeters)
expanded scale
vertical (bottom zero)
pencil-marked on initial and end values
0 ... 100%,
deviating from standard;
special calibration from customer's
non-linear graph or chart;
captions on request
on request e.g. "generator"
on request
red, green or blue for important scale values
red, green or blue within scale division
none or on request
mechanically suppressed zero, no zero
adjustment, max. 40% of full-scale value
no overload scale
on request
Connections
Ordering Information
type
W
front dimensions
72 PrS
96 PrS
144 PrS
measuring ranges
sp. measuring range
accuracy class
AC voltage
*
17
*
18
L1
L1
N (L2)
N (L2)
17
18
* W 72/96 PrS voltmeters 600 V
calibration
AC current
17
18
17
L1
L1
N (L2)
N (L2)
18
k
l
K
L
colour of bezel
position of use
climatic suitability
Dimensions
marine application
W 72/96 PrS
enclosure code
b
d
terminal protection
c
a
scale arrangement
dial
b
d
W 144 PrS
c
a
dimensions (in mm)
a
b
c
d
W 72 PrS
72
36
94
32
W 96 PrS
96
48
107
43
W 144 PrS
144
72
192
67
scales and position of use
II
I
a
window
a
III
IV
VI
V
vertical mounting
I
horizontal scale (standard)
II vertical scale
inclined mounting
III horizontal scale
IV vertical scale
horizontal mounting
V horizontal scale
VI vertical scale
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
logo
zero position
overload scaling
(ammeters)
expanded scale
*)
**)
***)
edgewise-type moving-iron panel meter
72 mm x 36 mm
96 mm x 48 mm
144 mm x 72 mm
refer to preceding table
on request **)
1.5 *)
1.0 with fine scale division ***)
none *)
for a definite frequency 100 ... 1000 Hz **)
glass *)
non-glaring glass
black (similar to RAL 9005) *)
gray (similar to RAL 7037)
vertical *)
horizontal
on request 15 ... 165° **)
class 2, -25 ... +40°C *)
class 3, -10 ... +55°C
none *)
non-certified
IP 52 *)
IP 54 splash-water protected front
none *)
protective sleeves SW6 / SW10
horizontal *)
vertical
scale division & measuring range alike *)
blank dial
scale division and figuring
0 ... 100%
deviating from standard **)
calibration fr. non-linear graph or chart **)
additional lettering on request **)
additional figuring on request **)
coloured marks red, green or blue **)
coloured sector red, green or blue **)
WEIGEL *)
none
OEM logo **)
mechanically suppressed zero **)
2 times rated current *)
no overload scale
on request **)
standard
Please clearly add the desired specifications.
as far as possible
ordering example
W 72 PrS, measuring range 0 ... 250 mA, horizontal scale,
vertical mounting, window non-glaring glass, WEIGEL logo
- specifications subject to change without notice; date of issue 02/07 -
Data Sheet
Shunts
Class 0.5
60 mV
100 mV
150 mV
300 mV
800.D.101.05
Application
rated current '
Shunts provide an accurate DC millivolt signal to drive ammeter indicaĆ
tors, overload protection and control devices, especially for higher amĆ
perage. They supply a voltage drop proportional to the DC current
which is measured and indicated by a moving-coil meter with the dial
calibrated in amps.
In accordance with DIN standard 43 703 shunts are available from 1 A
up to 15,000 A with an accuracy of 0.5%. Standard voltage drop is
60 mV or 150 mV. Intermediate current ratings, other voltage outputs,
better accuracy and purpose-built shunts can be supplied.
Shunts are manufactured in three different format versions depending
on current ratings.
60 A
0.12 kg
0.14 kg
0.17 kg
80 A 1) 2)
0.12 kg
0.15 kg
0.18 kg
100 A
0.12 kg
0.17 kg
0.20 kg
150 A
0.13 kg
0.20 kg
0.23 kg
1)
2
200 A
)
0.61 kg
0.65 kg
0.68 kg
250 A
0.61 kg
0.65 kg
0.68 kg
300 A 1) 2)
0.61 kg
0.68 kg
0.72 kg
400 A
0.83 kg
1.00 kg
1.05 kg
500 A 1) 2)
0.83 kg
1.10 kg
1.15 kg
600 A
0.85 kg
1.11 kg
1.16 kg
800 A 1) 3)
0.90 kg
1.12 kg
1.30 kg
1,000 A
1.45 kg
2.00 kg
2.15 kg
1,200 A 1) 2)
1.45 kg
2.10 kg
2.25 kg
1,500 A
1.96 kg
2.50 kg
3.10 kg
2,000 A 1)
2.30 kg 3) 2.80 kg 3) 5.10 kg 2)
2,500 A
2.90 kg
3.20 kg
5.20 kg
3,000 A 1)
3.00 kg 3) 3.50 kg 3) 9.80 kg 2)
4,000 A
4.25 kg
5.80 kg 10.5 kg
5,000 A 1)
4.30 kg 3) 7.30 kg 3) 13.4 kg 2)
6,000 A
10.5 kg
12.0 kg
15.0 kg
8,000 A 1)
12.0 kg 3) 25.4 kg 2)
10,000 A
21.0 kg
28.0 kg
15,000 A
32.0 kg
20,000 A 1)
44.0 kg
1) ratings deviating from DIN standard
2) dimensions equal to next higher current rating
3) dimensions equal to next lower current rating
overload range (according to DIN EN 60 051)
continuously
1.2 times rated current
2,000 A 5 times rated current
5 s max.
>2,000 ... 10,000 A 2 times rated current
Operating Principle
The current passing through the shunt produces a proportional voltage
drop. A moving coil instrument connected to the shunt measures the
voltage drop across the shunt terminals.
Shunts are calibrated in such a way that they produce an accurately
defined voltage drop (60 mV, 150 mV or other).
General Data
format version A
format version B
format version C
material
resistance bars
end blocks
format version A
format version B
format version C
base material
format version A
connections
current
voltage
mounting
format version A
enclosure code
dimensions
weight
rated current '
1A
1.5 A
2A
2.5 A
3A
4A
5A
6A
8A
10 A
12 A
15 A
20 A
25 A
30 A
40 A
50 A
1) 2)
1) 2)
1) 2)
1) 2)
1) 2)
1) 2)
1) 2)
1) 2)
insulating base mounted shunts clamping
to DINā mounting rail or wall mounting
(up to 25 A / 60, 100, 150 or 300 mV);
without insulating base (30 ... 150 A) '
L-profile end blocks
T-profile end blocks
manganin
high conductivity brass
high conductivity brass/solid copper
solid copper
Lexan, black
selfāĆāextinguishing to UL rating 94 V-0
thread screws
please refer to "Dimensions"
M5x8
screw mounting (M8 max.) or
clamping to DINā mounting rail
(to DIN EN 50 022Ă-Ă35)
IP 00
please refer to "Dimensions"
please refer to table below
weight approx. for rated voltage drop '
60 mV
100 mV1) 150 mV 300 mV1)
0.12 kg
0.13 kg
0.13 kg
0.12 kg
0.12 kg
0.13 kg
0.12 kg
0.12 kg
0.13 kg
0.13 kg
0.13 kg
0.13 kg
0.13 kg
0.13 kg
0.12 kg
0.12 kg
0.12 kg
0.12 kg
0.12 kg
0.12 kg
0.12 kg
0.12 kg
0.12 kg
0.12 kg
0.12 kg
0.13 kg
0.13 kg
0.13 kg
0.13 kg
0.14 kg
0.14 kg
0.13 kg
0.14 kg
0.14 kg
0.12 kg
0.12 kg
0.12 kg
0.12 kg
0.12 kg
0.12 kg
0.12 kg
0.13 kg
0.13 kg
0.13 kg
0.13 kg
0.13 kg
0.14 kg
0.14 kg
0.15 kg
0.16 kg
0.16 kg
0.14 kg
0.14 kg
0.14 kg
0.14 kg
0.14 kg
0.14 kg
0.14 kg
0.14 kg
0.15 kg
0.15 kg
0.15 kg
0.15 kg
0.16 kg
0.16 kg
0.20 kg
0.20 kg
0.20 kg
weight approx. for rated voltage drop '
60 mV
100 mV1) 150 mV 300 mV1)
0.20 kg
0.20 kg
0.25 kg
0.30 kg
0.80 kg
0.80 kg
0.90 kg
1.30 kg
1.40 kg
1.60 kg
1.80 kg
2.80 kg
3.10 kg
3.70 kg
6.40 kg 2)
6.00 kg
11.7 kg 2)
13.1 kg
16.8 kg 2)
17.7 kg
-
Accuracy at Reference Conditions
accuracy
class 0.5 '
reference conditions
ambient temperature 23°C±1K
Environmental
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
climatic class 3 acc. to VDE/VDI 3540
-10 ... +55°C
-25 ... +65°C
75% annual average, non-condensing
Rules and Standards
DIN 43 703
DIN EN 60 051
DIN EN 50 022āĆā35
Shunts
direct acting indicating electrical measuring
instruments and their accessories
mounting rails
Options
rated voltage drop
rated current
accuracy
insulating base
purpose built shunts
on request
on request
class 0.2
suitable for shunts 30 ... 150 A / 60 mV
others on request
on request
' for other ratings refer to "Options"
Data Sheet
800.D.101.05
Shunts
Class 0.5
Accessory
cover
Dimensions
for shunts with insulating base
1 ... 25 A / 60 mV - 100 mV - 150 mV
30 ... 150 A / 60 mV
10
d
Dimensions
format version B
format version A with insulating base
8
d
5
8
15
g
f
c
9
30
e
c
a
b
b
115
a
122
rated voltage drop 60 mV
135
dimensions
200 A
400 A
(in mm)
250 A
600 A
a
145
145
b
105
105
c
30
40
d
30
30
e
15
20
f
g
M12x40 M16x45
33,5
number of
2x 1
current connections
28
30
38,5
format version A with insulating base and cover
a
b
c
d
e
f
g
123
135
rated voltage drop
60 ... 300 mV
dimensions (in mm)
a
b
c
d
1 ... 25 A
90
78
20
M5x12
60 mV
30 .... 150 A
100
80
20
M8x16
Disattaching
2
format version A without insulating base
100 mV
dimensions (in mm) 30 .... 150 A
a
145
b
125
c
25
d
M8x16
165
115
90
30
21
48
M16x45
165
115
120
30
30
60
M20x50
2x 1
2x 2
2x 2
150 mV
30 .... 150 A
225
205
25
M8x16
190
150
40
30
20
M16x45
210
160
60
30
30
M20x50
210
160
120
30
30
60
M20x50
number of
2x 1
current connections
2x 1
2x 1
2x 2
rated voltage drop 150 mV
300 mV
30 .... 150 A
384
364
25
M8x16
270
230
30
50
15
M12x40
270
230
40
50
20
M16x45
290
240
70
60
35
M20x50
number of
2x 1
current connections
2x 1
2x 1
rated voltage drop 300 mV
a
b
c
d
e
g
2.500 A
165
115
60
30
30
M20x50
190
150
30
30
15
M12x40
a
b
c
d
e
g
1
3
rated voltage drop
1.500 A
rated voltage drop 100 mV
115
Attaching
2x 1
1.000 A
429
389
30
50
15
M12x40
429
389
40
50
20
M16x45
449
399
70
60
35
M20x50
number of
2x 1
current connections
2x 1
2x 1
Dimensions
rated voltage drop 150 mV
dimensions
1,500 A 2,500 A 4,000 A 6,000 A 10,000A
(in mm)
a
290
290
300
300
310
b
240
240
250
250
260
c
90
120
120
154
206
d
60
60
130
130
170
e
21
30
30
25
25
f
48
60
60
52
52
g
M16x60 M20x60
M20x75 M20x75 M20x80
h
15
15
25
25
30
number of
2x 2
2x 2
2x 2
2x 3
2x 4
current connections
h
d
format version C
f
e
c
f
g
b
a
rated voltage drop 60 mV
dimensions
4.000 A 6.000 A 10.000 A 15.000A 20.000A
(in mm)
a
165
175
185
185
185
b
115
125
135
135
135
c
120
154
206
310
362
d
60
130
170
170
170
e
30
25
25
25
25
f
60
52
52
52
52
g
M20x50 M20x75
M20x80 M20x80 M20x80
h
15
25
30
30
30
number of
2x 2
2x 3
2x 4
2x 6
2x 7
current connections
rated voltage drop 100 mV
dimensions (in mm)
a
b
c
d
e
f
g
h
number of
current connections
2,500 A 4,000 A 6,000 A
210
220
220
160
170
170
120
120
154
60
130
130
30
30
25
60
60
52
M20x50
M20x50 M20x75
15
25
25
2x 2
2x 2
2x 3
rated voltage drop 300 mV
dimensions
1,500 A 2,500 A 4,000 A 6,000 A
(in mm)
a
449
449
459
459
b
399
399
409
409
c
90
120
120
154
d
60
60
130
130
e
21
30
30
25
f
48
60
60
52
g
M16x60 M20x60
M20x75 M20x75
h
15
15
25
25
number of
2x 2
2x 2
2x 2
2x 3
current connections
Ordering Information
type
rated voltage
drop
shunt
60 mV
100 mV
150 mV
300 mV
purpose built on request **)
rated current
please refer to table inside
purpose built on request **)
class 0.5 *)
class 0.2
included (up to 25 A) *)
not included (more than 25 A) *)
included (more than 25 A)
none *)
for shunts with insulating base
on request **)
accuracy
insulating base
cover
purpose built
*)
**)
standard
Please clearly add the desired specifications.
ordering example
shunt, rated voltage drop 60 mV, rated current 1,000 A,
accuracy class 0.5
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
- specifications subject to change without notice; date of issue 06/07 -
=
Data Sheet
MultiāĆāFunctional
Transducer for
Currents, Voltages,
Power
MMU 3.0
055.8e
Application
Operating Principle
The multiāĆāfunctional transducer MMU 3.0 accepts any measurable
quantity in existing singleĆāphase or threeĆāphase power supply sysĆ
tems, converts these input signals into a load independent DC current
and/or impressed DC voltage (current and voltage are synchronous on
analog output 1) and issues the measured values parametrically to an
interface RS 232 and RS 485. A digital output is also available in the
basic version. Transducers with additional analog outputs (voltage or
current programmable) and/or four resp. eight additional digital outĆ
puts are optionally available.
Inputs (exept 10 V measuring input) are galvanically isolated from outĆ
puts and the auxiliary voltage input. The outputs are shortāĆācircuit proof
and safe against idling.
The transducers comply with the safety requirements and are tested for
interference immunity.
The transducers are designed to be mounted in machines/systems.
Regulations for installation of electrical systems and equipment have to
be observed.
Transformers in the current and voltage circuits galvanically isolate the
power inputs from the electroniccircuitry.HoldĆā/sequenceĆācircuitsproĆ
cess the input signals and transfer them via a multiplexer and a ADĆāconĆ
verter to the microprocessor which processes the signals and comĆ
putes all important measuring quantities.
The transducer is connected to the PC via a commercially available
RS232 cable (9āĆācontact 1:1 connection, socketāĆāplug. An optional conĆ
nection will be a 3āĆācontact cable provided the signals DTR and DSR as
well as RTS and CTS will be twoāĆāway bridged.)
Analog Outputs
Any of the measurable quantities (current, voltage, activeĆ, reactiveāĆ
power, frequency etc.) can be allocated to each of the analog outputs.
The analog output available in the basic version synchronously
provides voltage and current (4 terminals). The output signal of each of
the optional analog outputs can be parametrized freely (0/4 ... 20 mA,
0/2 ... 10 V, -10 ... 10 mA; linear or buckled characteristic curve).
Selection between current or voltage output is effected by software.
It is possible to connect more than one indicator, recorder or controller
to the output circuit provided the total impedance does not exceed the
rating.
RS 232/485
The transducers are equipped with both a RS 232 and a RS 485 interĆ
face enabling to request measured values and to perform adjustments.
When using the RS 485 interface, up to 32 devices can be networked
and read out via a 2Ćāwire line (1000 m maximum length).
Digital Outputs
The digital outputs can be used as switching contacts for setpoint conĆ
trolling.
Auxiliary Supply
Power supply is effected by a separate auxiliary voltage input.
Software
The software WSoft ready for execution on Windows® 95/98/2000/XP
is available for control of functions and for readāĆāout of measured valĆ
ues. The control is effected by the widespread machine language
SCPI.
Sequence -/
Hold Circuit
INP 10 V Measuring Input
Analog Output 1
D
UL1
IA
A
RA
UL2
UA
Multiplexer
Measurement
The multiāĆāfunctional transducer processes input currents up to 5 A and
input voltages up to 519 V at rated frequencies of 50 Hz and 60 Hz.
Depending on the measurement task, input terminals not required reĆ
main idle.
Measurement is effected in true RMSāĆāvalues including wave forms up
to the 50th harmonics.
Block Circuit Diagram
UL3
IL1
RA
A
D
mP
Digital Output 1
(galvanically
isolated)
IL2
additional galvanically
isolated voltage-,
current - resp.
digital outputs
(optional)
IL3
UH
Interfaces
RS 232
RS 485
~
=
General Technical Data
case details
material of case
terminals
wire crossāĆāsection
enclosure code
dielectric test
2.2 kV
3.7 kV, 50 Hz
sinusoidal
projecting case clamping to TH 35 DIN
rail according to DIN EN 60 715
Cycoloy C2950 black
selfāĆāextinguishing to UL rating 94 V-0
screwāĆāterminals, maximum torque 0.8 Nm
4 mm2 max.
IP 40 case
IP 20 terminals
all circuits to case
currents to each other and to voltages;
inputs (exept 10 V measuring input) to
outputs, auxiliary voltage and interfaces;
auxiliary voltage to outputs and interfaces;
1000 V=
outputs to each other (the analog output 1 is
galvanically connected to 10 V measuring
input and to interfaces.)
operating voltage
300 V (rated voltage phase to zero)
class of protection
II
measurement category CAT III
pollution level
2
dimensions
basic version: 3 modules in singleāĆāphase
system resp. 4 modules in threeāĆāphase
systems,
optional outputs: additional 1 to 3 modules
each module WxHxL 22.5 mm x 80 mm x 115 mm
weight
approx. 0.6 kg (basic version)
=
Data Sheet
055.8e
MultiāĆāFunctional
Transducer for
Currents, Voltages,
Power
Inputs
input quantities
voltages
currents
auxiliary supply
10 V measurement
input
rated input current
IEN
rated input voltage
UEN
operating voltage
modulation range
overload limits
frequency range
power consumption
AC current and AC voltage
L1, L2, L3 (3 terminals), N (1 terminal)
I1, I2, I3 (6 terminals)
UH (2 terminals)
e.g. connecting an analog converter
N/5 A '
519 V (interĆāconnected) '
519 V max.
1.2 UEN and 1.2 IEN
1.2 UEN, 1.6 IEN continuously
2 UEN, 10 IEN max. 1 s
50 ... 60 Hz
2 mA ±10% each voltage circuit
≤ 0.1 VA each current circuit for IEN = 1 A
≤ 1.6 VA each current circuit for IEN = 5 A
Measuring Quantities
Measuring Quantity
Total
L1
L2
L3
voltage (U)
U
U1
U2
U3
current (I)
I
1)
I1
I2
1)
I3
active power (P)
P
P1
P2 1)
P3
reactive power (Q)
Q
Q1
Q2 1)
Q3
apparent power (S)
S
S1
S2 1)
S3
active factor (PF)
PF
PF1
PF2 1)
PF3
reactive factor (QF)
QF
QF1
QF2 1)
QF3
phase angle (PH)
PH
PH1
PH2 1)
PH3
frequency (f)
F
Depending on power system, it will not be possible to measure all these
values.
10 V measuring input
INP
current output
output current
rated current
IA
IAN
load range
RA
load error
residual ripple
idling voltage
current limitation
voltage output
output voltage UA
load independent DC voltage
rated voltage
UAN 0 (2) ... 10 V (parameterizable)
load
RA
≥ 4 kW (based on UAN)
load error
≤ 0.1% based on 50% load change
residual ripple
≤ 1%rms of UAN with load RA = UAN / 2 mA
idling voltage
≤ 16 V
voltage limitation
up to 12 V
Inputs (exept 10 V measuring input) and outputs are galvanically isoĆ
lated.
Conversion Characteristics
examples
standard
live zero"
IA, UA
IA, UA
IA, UA
measuring
quantity
IA, UA
measuring
quantity
IA, UA
IA, UA
(±10 V)
voltage & current synchronous
(2 terminals each)
interfaces
RS 232 (SUB-D jack)
RS 485 (2 terminals)
(All outputs listed above and the analog input have one and the same
potential.)
digital output
contactāĆāfree via opto coupler,
max. 230 V / 100 mA,
internal resistance 25 ... 35 W,
insulation voltage 2.3 kV,
switching frequency admissible ≤2 Hz
1, 2, or 3 additional analog outputs (galvanically isolated) and
up to 8 additional digital outputs (galvanically isolated) are optional '
response time
≤ 500 ms,
based on 50 Hz
exception for 3āĆāphase 3āĆāwire unbalanced
load system for quantities marked with 1)
(see table Measuring Quantities): ≤ 750 ms
additional response
20 ms for each value
time for serial output (RS 232/485, 19,200 baud)
' refer also to Extras
incoming and outgoing
measuring
quantity
measuring
quantity
Outputs
Outputs '
analog output 1
load independent DC current
0 (4) ... 20 mA or
0
... 10 mA (parameterizable)
0 ... 500 W (based on 20 mA)
0 ... 1000 W (based on 10 mA)
≤ 0.1% based on 50% load change
≤ 1%rms of IAN with load RA
≤ 16 V
up to 24 mA
measuring
quantity
measuring
quantity
buckled characteristic curve
IA, UA
IA, UA
IA, UA
measuring
quantity
measuring
quantity
measuring
quantity
Interfaces
type
Baudā rate
data bit
parity
stop bit
RSā232 (V.24) and
RSā485 (SCPI commands)
19200 Baud
8
none
2
Auxiliary Supply
Environmental
auxiliary voltage UHN wideā Ćā range supply
20 ... 90 V DC resp. 15 ... 65 V AC,
90 ... 357 V DC resp. 65 ... 253 V AC
power consumption
< 10 VA
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
Accuracy at Reference Conditions
accuracy
better than class 0.5 (±0.5% of end value)
exception for 3āĆāphase 3āĆāwire
unbalanced load system for quantities
marked with 1)
(see table Measuring Quantities)
These ratings are calculated values
(Aron circuit):
class 1.5 (±1.5% of end value)
temperature coefficient ≤ 0.06%/K
valid for standard products and a lifeāĆāperiod of 1 year maximum
reference conditions
input current
input voltage
power factor
frequency
wave form
auxiliary voltage
load
ambient temperature
warmāĆāup
IEN ±0.5%
UEN ±0.5%
cos ϕ =1
50 Hz
sine wave, distortion factor ≤ 1%
UHN ±1%, 48 ... 62 Hz
0.5 RA max ±1% based on current
10 kW ±1% based on voltage
23°C ±1K
≥5 min
climatic class 3 to VDE/VDI 3540 sheet 2
-10 ... +55°C
-25 ... +65°C
≤ 75% annual average, nonāĆācondensing
Rules and Standards
DIN EN 60 529
DIN EN 60 688
Enclosure codes by housings (IPāĆācode)
Electrical measuring transducers
converting AC quantities into analog or
digital signals
DIN EN 60 715
Dimensions of low voltage switching devices:
standardized DIN rails for mechanical fixation
of electrical devices in switchgears
DIN EN 61 010āĆā1
Safety requirements for electrical measuring,
control and laboratory equipment
Part 1: General requirements
DIN EN 61 326āĆā2āĆā1
Electrical equipment for measurement, conāĆ
trol and laboratory use - EMC requirements
Part 2āĆā1: Particular requirements Test configurations, operational conditions
and performance criteria for sensitive test and
measurement equipment for EMC
unprotected applications
(IEC 61 000āĆā4āĆā3 evaluation criterion B)
VDE/VDI 3540 sheet 2 Reliability of measuring and control
equipment (classification of climates for
equipment and accessories)
Data Sheet
=
055.8e
MultiāĆāFunctional
Transducer for
Currents, Voltages,
Power
Connections
Terminals
input
D61 D62
D21 D22
– A2 +
D71 D72
D31 D32
– A3 +
D81 D82
D41 D42
– A4 +
D91 D92
D51 D52
1
3
D11 D12
4
6
14 13
7
9
20 19
active and reactive power, singleāĆāphase
front view
active and reactive power, 3āĆāphase, 3āĆāwire, balanced load
digital
outputs
6 to 9
terminal
digital
outputs
2 to 5
analogā
outputs
2 to 4
2
11
A
B
5 8
17 16
E– E+
basic version for 3āĆāphase network (one
module inapplicable for singleāĆāphase
network)
MMU 3.0
1 IE L1
2 U E L1
3 IE L1
4 IE L2
5 U E L2
6 IE L2
active and reactive power, 3āĆāphase, 4āĆāwire, balanced load
7 IE L3
8 U E L3
9 IE L3
11 UE N
13 UA1 (+)
14 UA1 (–)
16 UH L1 (+)
17 UH N (–)
active and reactive power, 3āĆāphase, 3āĆāwire, unbalanced load
19 IA1 (+)
20 IA1 (–)
E+ UE (+)
E– UE (–)
A RS 485
B RS 485
SUBāĆāD RS 232
Dn1 digital output n, contact 1, (n = 1 ... 9)
active and reactive power, 3āĆāphase, 4āĆāwire, unbalanced load
Dn2 digital output n, contact 2, (n = 1 ... 9)
Amā– analog output m, negative pole, (m = 2 ... 4)
Am+ analog output m, positive pole, (m = 2 ... 4)
Depending on the measurement task, inputāĆ resp. outputāĆāterminals
remain idle.
current input
IE
voltage input
UE
The numbers on the terminals correspond to details
in connection diagrams (refer to DIN 43 807).
IA
current output
voltage output
UA
UH
auxiliary voltage input
RSā232āĆāInterconnection
Ordering Guide
type
DSR 6
RTS 7
CTS 8
RI 9
1 DCD
2 RxD
3 TxD
4 DTR
5 GND
multiāĆāfunctional transducer
for currents, voltages, power
MMU 3.0
9āĆācontact
interconnection
E
basic version for singleāĆāphase AC
D
basic version for 3āĆāphase network
outputs *)
analog output 1 (voltage & current synchronous)
digital output 1
3āĆācontact
interconnection
analog outputs
Extras
A1
1 additional analog output with basic version
A2
2 additional analog outputs with basic version
outputs
1, 2 or 3 additional
analog outputs
A3
3 additional analog outputs with basic version
Ax
additional analog outputs with basic version**)
4 or 8 additional
digital outputs
rated input current
IEN
rated input voltage
UEN
can be parametrized via software between
20 mA (load<500 W) and 10 V (load >4 kW);
galvanically isolated,
power supply unit integrated
(width: 1 module)
230 V, galvanically isolated
(width: 1 resp. 2 modules)
N/1.2 A (also programmable for N/1 A,
with same accuracy)
N/120 V (interĆāconnected)
(also programmable for N/100 V or N/110 V,
with same accuracy)
Dimensions
digital outputs
D4
4 additional digital outputs with basic version
D8
8 additional digital outputs with basic version
auxiliary supply
H4
DC 20 ... 90 V / AC 15 ... 65 V
H5
DC 90 ... 357 V / AC 65 ... 253 V
programming
P0
by user *)
P1
by factory
accessory
example: basic version with 3 modules, width: each module 22.5 mm
front view
67.5
side view
WSoft
software on CD for configuration and
readāĆāout of measured values
RS 232 cable serial connection cable
APĆRS232/485 RSā232āĆā485 converter
*)
**)
standard
on request
Note: Data relating to input, measuring range and to the output assignĆ
ment are not required, as the transducers are suitable to be configured
with a PC or laptop.
ordering example
115
80
MMU 3.0
D D4 H5 P0 WSoft
multiāĆāfunctional transducer for use on 3Ćāphase network
(1 analog output and 1 digital output included),
not any additional analog outputs,
4 additional digital outputs,
auxiliary voltage DCā90ā...ā357āVā/āACā65ā...ā253āV,
userĆāprogramming; software WSoft
(dimensions in mm)
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
- specifications subject to change without notice; date of issue 03/09 -
Product Guide
x
800.U.001.04
y
Transducers
J/U
=
P/Q
=
P+Q
=
f
=
=
=
RS232
RS485
=
=
=
=
RS232
RS485
RS232
RS485
✓ for AC Current or AC Voltage:
A1U / V1U 2.2 – also self-powered: AU / VU 2.0
A1U / V1U 2.3
✓ for AC Current (Co–generation):
A1U 2.2 E/D
✓ for Active or Reactive Power
EW/B / DGW/B / DUW/B / VGW/B / VUW/B 2.2
new: EW/B / DGW/B / DUW/B / VGW/B / VUW/B 2.3
✓ for Active and Reactive Power
EW+B / DGW+B / DUW+B / VGW+B / VUW+B 2.2
✓ Programmable Multi-Functional Transducer: MMU 3.0
✓ for Frequency: FU 2.2
for Phase Angle (cos ϕ): CU 2.2
for DC Current or DC Voltage: AUD / VUD 2.2
for RMS Current or RMS Voltage: AUE / VUE 2.2
for Temperature: PTU 2.0
Isolation Transducers for Standard Signals: TUA 2.2
also self-powered: TUP 2.0
✓ Standard Signal Interface Converter: MU-RS232/485
✓ RS232-RS485 Converter: AP-RS232/485
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
SMD and microprocessor controlled technology
Accuracy: class 0.5
Mounting on DIN - rail 35 mm
Case width: 22.5 mm each module
Auxiliary supply AC 115/230 V (50/60 Hz), DC 24 V
or wide range supply
General Data
x
Short Form Data
J/U
=
=
Transducers
Application
Transducers convert power current and voltage quantities or proccess
control inputs into proportional load independent DC current or voltage
outputs.
for measuring
AC current or AC voltage,
active or reactive power,
frequency,
phase angle (ϕ ) / power factor (cos ϕ ),
DC current or DC voltage,
standard signals,
temperature
Transducers for
AC Current or
AC Voltage,
Self-Powered
AU 2.0
VU 2.0
Technical Data
case details
projecting case clamping to TH35 mounting
rail according to DIN EN 60 715
material of case
ABS/PC black
self - extinguishing to UL rating 94 V–0
terminals
screw - terminals
wire cross - section
4 mm2 max.
enclosure code
IP 40 case
IP 20 terminals
class of protection
II
measuring category
CAT III
pollution level
2
operating voltage
300 V (rated mains voltage phase to zero)
climatic suitability
climatic class 3 acc. to VDE/VDI 3540 sheet 2
operating temperature –10 ... +55oC
range
storage temperature
–25 ... +65oC
range
relative humidity
≤ 75% annual average, non - condensing
Input Ratings
input quantity
type
rated input
Output Ratings
current output
output current
IA
load range
RA
current limitation
load independent DC current 0 ... 20 mA
or optionally 0 ... 10 mA, 0 ... 5 mA,
4 ... 20 mA (”live zero”), –20 ... 0 ... 20 mA
(bipolar output only with wide range supply)
0 ... 10 V / IAN
to approx.120% of full–scale value,
other ratings on request
voltage output
output voltage
UA
measuring range
modulation range
overload limit
frequency range
power consumption
sinusoidal AC current (AU 2.0)
sinusoidal AC voltage (VU 2.0)
AU 2.0
VU 2.0
voltage UEN
current IEN
1 A *), 1.2 A,
57.7 V, 63.5 V,
5 A *), 6 A
100 V *), 110 V *),
150 V, 250 V, 400 V,
other rating on request 500 V
*) also for use on transducer
0 ... IEN
0 ... UEN
1.2 IEN
1.2 UEN
1.5 IEN continuously 1.2 UEN continuously
2 UEN 1 s max.
10 IEN 1 s max.
48 ... 62 Hz
voltage transformer
< 3 VA
current transformer 5A
< 4 VA
current transformer 1A
< 2 VA
Output Ratings
load independent DC voltage 0 ... 10 V or
optionally 2 ... 10 V (”live zero”), –10 ... 0 ... 10 V
(bipolar output only with wide range supply)
≥ 4 kW
load
RA
load error
≤ 0.1% based on 50% load change
residual ripple
≤ 1%rms
response time
approx. 500 ms
idling voltage
≤ 15 V
(Ratings do not apply completely to self - powered transducers.)
current output
output current
IA
rated current
IAN
load range
RA
load error
idling voltage
accuracy
Auxiliary Supply
Other Ratings
auxiliary voltage UHN 230 V AC (195 ... 253 V), 48 ... 62 Hz
optional
115 V AC (98 ... 126 V), 48 .. 62 Hz
24 V DC (20 ... 72 V)
wide range supply
20 ... 100 V DC resp. 20 ... 70 V AC
90 ... 357 V DC resp. 65 ... 253 V AC
galvanic isolation between input, output and auxiliary supply circuits
auxiliary voltage
dimensions WxHxL
weight
load independent DC current
0 ... 20 mA
0 ... 500 W
≤ 0.4% based on 50% load change
≤ 20 V
class 0.5 (±0.5% of end value)
none required
22.5 mm x 80 mm x 115 mm
approx. 0.35 kg
for detailed information refer to Data Sheet No. 045.##
Short Form Data
J/U
Short Form Data
J/U
=
=
Transducers for
AC Current or
AC Voltage
A1U 2.2
V1U 2.2
Transducers for
AC Current or
AC Voltage
A1U 2.3
V1U 2.3
Input Ratings
Input Ratings
input quantity
input quantity
type
meeasuring unit
rated input
in the range
measuring range
modulation range
overload limit
frequency range
power consumption
sinusoidal AC current or
sinusoidal AC voltage
A1U 2.2
V1U 2.2
AC current
AC voltage
voltage UEN
current IEN
0 ... 200 mA up to 5 A 0 ... 60 mV up to 519 V
(also for use on transformer)
0 ... IEN
0 ... UEN
current input
voltage input
1.2 UEN
1.2 IEN
1.2 IEN continuously 1.2 UEN continuously
2 UEN 1 s max.
10 IEN 1 s max.
48 ... 62 Hz or
162/3 Hz, 100 Hz, other ratings on request
≤ 0.4 VA
Output Ratings
output
current or voltage output
refer to General Data
Other Ratings
accuracy
auxiliary voltage
dimensions WxHxL
weight
class 0.5 (±0.5% of end value)
refer to General Data
22.5 mm x 80 mm x 115 mm
approx. 0.12 kg
type
meeasuring unit
rated input
measuring range
modulation range
overload limit
frequency range
power consumption
sinusoidal AC current or
sinusoidal AC voltage
A1U 2.3
V1U 2.3
AC current
AC voltage
voltage UEN
current IEN
1 A*) / 5 A*)
100 V*) / 250 V / 500 V
*) also for use on transformer
0 ... IEN
0 ... UEN
current input
1.2 IEN
1.2 IEN continuously
10 IEN 1 s max.
48 ... 62 Hz
≤ 0.4 VA
voltage input
1.2 UEN
1.2 UEN continuously
2 UEN 1 s max.
Output Ratings
current output
output current
IA
rated current
IAN
load range
RA
current limitation
load independent DC current
0 ... 20 mA or 4 ... 20 mA
0 ... 600 W
to 120 ... 140% of end value
or voltage output
impressed DC voltage
output voltage
UA
rated voltage
UAN 0 ... 10 V or 2 ... 10 V
load
RA
≥ 4 kW
load error, residual ripple, response time refer to General Data
idling voltage
≤ 20 V
Other Ratings
accuracy
class 0.5 (±0.5% of end value)
auxiliary voltage UHN 230 V AC (195 ... 253 V), 48 ... 62 Hz
optional
115 V AC (98 ... 126 V), 48 .. 62 Hz
galvanic isolation between input, output and auxiliary supply circuits
dimensions WxHxL
22.5 mm x 80 mm x 115 mm
weight
approx. 0.16 kg
for detailed information refer to Data Sheet No. 061.##
for detailed information refer to Data Sheet No. 068.##
Short Form Data
J/U
Short Form Data
P/Q
=
=
Transducers for
AC Current
(Co-generation)
Transducers for
Active or
Reactive Power,
Fixed Calibration
EW
EB
DGW
VGW
DUW
VUW
DGB
VGB
DUB
VUB
A1U 2.2 E
A1U 2.2 D
2.3
2.3
2.3
2.3
2.3
2.3
2.3
2.3
2.3
2.3
with mP
Input Ratings
Input Ratings
input quantity
input quantity
sinusoidal AC current and voltage
measuring unit PE or QE
active / reactive power
single - phase AC system
EW 2.3
EB 2.3
3 - phase 3 - wire balanced load system
DGW 2.3 DGB 2.3
3 - phase 4 - wire balanced load system
VGW 2.3 VGB 2.3
3 - phase 3 - wire unbalanced load system DUW 2.3 DUB 2.3
3 - phase 4 - wire unbalanced load system VUW 2.3 VUB 2.3
measuring range
0 ... PN
PN=calibration factor . PS
single phase AC
PS = U . I
(calibration factor=0.87)
3 - phase system
PS = 3 . U . I (calibration factor=0.72)
rated input voltage
UEN 0 ... 230 V / 0 ... 400 V
rated input current
IEN 0 ... 1 A / 0 ... 5 A (also for use with CT)
modulation range
1.2 UEN and 1.2 IEN
overload limits
1.2 UEN, 1.2 IEN continuously
2 UEN, 10 IEN 1 s max.
frequency range
48 ... 62 Hz
power consumption
<1 mA each voltage circuit
≤ 0.1 VA each current circuit for IEN = 1 A
≤ 1.6 VA each current circuit for IEN = 5 A
sinusoidal AC current
type / meeasuring unit
A1U 2.2 E
A1U 2.2 D
rated input current
IEN
1A
5A
AC current with co - generation indication
single - phase AC system
3 - phase 3 - or 4 - wire balanced load system
measuring range
–IEN ... 0 ... +IEN
–1 A (input) ... 0 ... +1 A (output)
–5 A (input) ... 0 ... +5 A (output)
(also for use on transformer)
rated input voltage UEN ranging from 57.7 V to 500 V
modulation range
1.2 IEN
overload limit
1.2 IEN continuously
10 IEN 1 s max.
frequency range
50 ... 60 Hz
power consumption
<1 mA each voltage circuit
≤ 0.1 VA each current circuit for IEN = 1 A
≤ 1.6 VA each current circuit for IEN = 5 A
Output Ratings
outputs
current and voltage outputs
refer to General Data
Other Ratings
outputs
accuracy
auxiliary voltage
dimensions WxHxL
weight
refer to General Data
class 0.5 (±0.5% of end value)
refer to General Data
45 mm x 80 mm x 115 mm
approx. 0.27 kg
for detailed information refer to Data Sheet No. 065.##
Output Ratings
outputs
current and voltage output
refer to A1U/V1U 2.3
Other Ratings
accuracy
class 0.5 (±0.5% of end value)
auxiliary voltage UHN 230 V AC (195 ... 253 V), 48 ... 62 Hz
optional
115 V AC (98 ... 126 V), 48 .. 62 Hz
24 V DC (20 ... 72 V)
dimensions WxHxL
45 mm x 80 mm x 115 mm
weight approx.
EW/B 2.3
DGW/B 2.3
DUW/B 2.3
VGW/B 2.3
VUW/B 2.3
0.18 kg
0.20 kg
0.22 kg
for detailed information refer to Data Sheet No. 069.##
Short Form Data
P/Q
Short Form Data
P+Q
=
=
Transducers for
Active or
Reactive Power
EW
EB
DGW
VGW
DUW
VUW
DGB
VGB
DUB
VUB
2.2
2.2
2.2
2.2
2.2
2.2
2.2
2.2
2.2
2.2
Transducers for
Active and
Reactive Power
EW+B 2.2
DGW+B 2.2
VGW+B 2.2
DUW+B 2.2
VUW+B 2.2
with mP
with mP
Input Ratings
Input Ratings
input quantity
sinusoidal AC current and voltage
measuring unit PE / type
active / reactive power
single - phase AC system
EW 2.2
EB 2.2
3 - phase 3 - wire balanced load system
DGW 2.2 DGB 2.2
3 - phase 4 - wire balanced load system
VGW 2.2 VGB 2.2
3 - phase 3 - wire unbalanced load system DUW 2.2 DUB 2.2
3 - phase 4 - wire unbalanced load system VUW 2.2 VUB 2.2
measuring range
0 ... PN or –PN ... 0 ... PN
PN= (0.3 ... 1.5) . PS
PS = U . I (single - phase AC system)
PS = 3 . U . I (3 - phase systems)
rated input voltage UEN ranging from 50 V to 519 V
rated input current IEN 1 A or 5 A (also for use on transformer)
or ranging from 0.5 to 5 A
modulation range
1.2 UEN and 1.2 IEN
overload limits
1.2 UEN, 1.2 IEN continuously
2 UEN, 10 IEN 1 s max.
frequency range
48 ... 62 Hz
power consumption
<1 mA each voltage circuit
≤ 0.1 VA each current circuit for IEN = 1 A
≤ 1.6 VA each current circuit for IEN = 5 A
input quantity
sinusoidal AC current and voltage
measuring unit PE / type
active and reactive power
single - phase AC system
EW +B 2.2
3 - phase 3 - wire balanced load system
DGW+B 2.2
3 - phase 4 - wire balanced load system
VGW+B 2.2
3 - phase 3 - wire unbalanced load system
DUW+B 2.2
3 - phase 4 - wire unbalanced load system
VUW+B 2.2
equal active and reactive power ranges, optonally also not equal
measuring range
0 ... PN or –PN ... 0 ... PN
PN= (0.3 ... 1.5) . PS
PS = U . I (single - phase AC system)
PS = 3 . U . I (3 - phase systems)
rated input voltage UEN ranging from 50 V to 519 V
rated input current IEN 1 A or 5 A (also for use on transformer)
modulation range
1.2 UEN and 1.2 IEN
overload limits
1.2 UEN, 1.2 IEN continuously
2 UEN, 10 IEN 1 s max.
frequency range
48 ... 62 Hz
power consumption
<1 mA each voltage circuit
≤ 0.1 VA each current circuit for IEN = 1 A
≤ 1.6 VA each current circuit for IEN = 5 A
Output Ratings
Output Ratings
outputs
outputs
current and voltage output
refer to General Data
current or voltage outputs
refer to General Data
Other Ratings
Other Ratings
accuracy
auxiliary voltage
dimensions WxHxL
weight approx.
accuracy
auxiliary voltage
dimensions WxHxL
weight approx.
class 0.5 (±0.5% of end value)
refer to General Data
45 mm x 80 mm x 115 mm
EW/B 2.2
DGW/B 2.2
DUW/B 2.2
VGW/B 2.2
VUW/B 2.2
0.27 kg
0.29 kg
0.31 kg
for detailed information refer to Data Sheet No. 062.##
class 0.5 (±0.5% of end value)
refer to General Data
45 mm x 80 mm x 115 mm
EW+B 2.2
DGW+B 2.2 DUW+B 2.2
VGW+B 2.2
VUW+B 2.2
0.27 kg
0.29 kg
0.31 kg
for detailed information refer to Data Sheet No. 066.##
Short Form Data
f
Short Form Data
=
=
Transducer
for Frequency
FU 2.2
Transducer for
Phase Angle (cos )
CU 2.2 E
CU 2.2 D
with mP
with mP
Input Ratings
Input Ratings
input quantity
measuring unit
input quantity
type / measuring unit
CU 2.2 E
CU 2.2 D
measuring ranges
–37° ... 0 ... 37°
–60° ... 0 ... 60°
optional
–180° ... 0 ... 180°
measuring ranges
rated input voltage
overload limit
current consumption
AC voltage
frequency fE
fEmin ≥ 14 Hz
fEmax ≤ 500 Hz
Df
class
fEmin ... fN ... fEmax
45 ... 50 ... 55 Hz
10 Hz
0,2
48 ... 50 ... 52 Hz
4 Hz
0,3
55 ... 60 ... 65 Hz
10 Hz
0,2
58 ... 60 ... 62 Hz
4 Hz
0,5
360 ...400 ... 440 Hz
80 Hz
0,2
380 ...400 ... 420 Hz
40 Hz
0,2
other ratings on request (Df = fEmax – fEmin)
UEN
100 V, 110 V, 115 V, 120 V, 230 V,
240 V, 380 V, 400 V, 415 V, 440 V
1.2 UEN continuously
2 UEN 1 s max.
≤ 1 mA
sinusoidal AC current / voltage
phase angle ϕ (power factor) in
single - phase AC system or
3 - phase 3 - wire balanced load system
corresponds to cos ϕ : cap 0.8 ... 1 ... 0.8 ind
corresponds to cos ϕ : cap 0.5 ... 1 ... 0.5 ind
to be specified in the range
corresponds to cos ϕ :
ind. (output) –1 ... 1 ... –1 cap. (output)
(unique measuring range –175° to + 175°)
rated input voltage UEN ranging from 50 V to 519 V
rated input current IEN 1 A or 5 A (also for use on transformer)
or ranging from 0.5 to 5 A
modulation range
1.2 UEN and 1.5 IEN
overload limits
1.2 UEN, 1.6 IEN continuously
2 UEN, 10 IEN 1 s max.
frequency range
48 ... 62 Hz
power consumption
<1 mA each voltage circuit
≤ 0.1 VA each current circuit for IEN = 1 A
≤ 1.6 VA each current circuit for IEN = 5 A
Output Ratings
Output Ratings
outputs
outputs
current and voltage output
refer to General Data
current and voltage output
refer to General Data
Other Ratings
Other Ratings
accuracy
auxiliary voltage
dimensions WxHxL
weight
accuracy
auxiliary voltage
dimensions WxHxL
weight
refer to measuring ranges
refer to General Data
45 mm x 80 mm x 115 mm
approx. 0.22 kg
for detailed information refer to Data Sheet No. 064.##
class 0.5 (±0.5% of end value)
refer to General Data
45 mm x 80 mm x 115 mm
approx. 0.27 kg
for detailed information refer to Data Sheet No. 063.##
=
RS232
RS485
Short Form Data
Short Form Data
P/Q
=
Programmable
Multi-Functional Transducer for AC Currents,
AC Voltages and Powers
MMU 3.0
Transducers for
Active Power or
Reactive Power
DUW 2.1
DUB 2.1
VUW 2.1
VUB 2.1
with mP
Input Ratings
Input Ratings
input quantities
input rating
voltages
currents
AC current and AC voltage
in single phase or 3 phase system
L1, L2, L3 (3 terminals), N (1 terminal)
519 V (inter - connected) or optionally
N/120V (also for N/100V or N/110V)
I1, I2, I3 (6 terminals)
N/5 A or optional N/1.2 A (also for N/1 A)
Measuring Units
Total
L1
L2
L3
voltage (U)
U
U1
U2
U3
current (I)
I
I1
I2
I3
active power (P)
P
P1
P2
P3
reactive power (Q)
Q
Q1
Q2
Q3
apparent power (S)
S
S1
S2
S3
active factor (PF)
PF
PF1
PF2
PF3
reactive factor (QF)
QF
QF1
QF2
QF3
phase angle (PH)
PH
PH1
PH2
PH3
frequency (f)
F
Depending on power system, not all these values van be measured.
10 V measuring input
INP
(±10 V)
Other Ratings
voltage & current synchronous
(2 terminals each) refer to General Data
interfaces
RS 232 (SUB–D jack), RS 485 (2 terminals)
digital output
contact - free via opto coupler
1, 2, or 3 additional analog outputs (galvanically isolated) and
up to 8 additional digital outputs (galvanically isolated) are optional
accuracy
class 0.5 (±0.5% of end value)
auxiliary voltage
wide range supply refer to General Data
dimensions
basic version: 3 modules in single - phase
resp. 4 modules in three - phase systems,
optional outputs: additional 1 to 3 modules
each module WxHxL 22.5 mm x 80 mm x 115 mm
weight
approx. 0.6 kg (basic version)
sinusoidal AC current and
sinusoidal AC voltage
measuring unit PE / type
active / reactive power
3 - phase 3 - wire unbalanced load system
DUW 2.1 DUB 2.1
3 - phase 4 - wire unbalanced load system
VUW 2.1 VUB 2.1
measuring range
0 ... PN or –PN ... 0 ... PN
PN= (0.3 ... 1.5) . PS
PS = 3 . U . I
rated input voltage UEN 65 V, 100 V, 110 V, 240 V, 400 V, 415 V, 440 V,
500 V or deviating from standard inputs
ranging from 0 ... (60 V ... UEN ... 519 V)
rated input current IEN N/1 A, N/5 A or deviating from standard inputs
ranging from 0 ... (0,5 A ... IEN ... 5 A)
modulation range
1.2 UEN and 1.2 IEN
overload limits
1.2 UEN, 1.2 IEN continuously
2 UEN, 10 IEN 1 s max.
frequency range
50 Hz (48 ... 52 Hz) or
162/3 Hz, 60 Hz, 100 Hz, others on request
power consumption
<1 mA each voltage circuit
≤ 0.1 VA each current circuit for IEN = 1 A
≤ 1.6 VA each current circuit for IEN = 5 A
Output Ratings
outputs
analog output 1
for detailed information refer to Data Sheet No. 055.##
current and voltage output
refer to General Data
Other Ratings
accuracy
auxiliary voltage
dimensions WxHxL
weight
class 0.5 (±0.5% of end value)
refer to General Data
45 mm x 80 mm x 115 mm
DUW/DUB 2.1
VUW/VUB 2.1
approx. 0.29 kg
approx. 0.31 kg
for detailed information refer to Data Sheet No. 051.##
Short Form Data
J/U
Short Form Data
=
=
=
Transducers for
DC Current/Voltage,
RMS Current/Voltage,
Isolating Transducers
Isolating Transducer
for Standard Signals,
Self-Powered
TUP 2.0
AUD2.2
VUD 2.2
AUE 2.2
VUE 2.2
TUA 2.2
Single or
Dual Channel
Input Ratings
measuring range
modulation range
overload limit
power consumption
current input
0 ... IEN
1.2 IEN
1.2 IEN continuously
10 IEN max. 1 s
≤ 0.4 VA
Input Ratings
voltage input
0 ... UEN
1.2 UEN
1.2 UEN continuously
2 UEN max. 1 s
model
input quantities
rated input value
AUD 2.2
DC current
IEN = 200 mA – 5 A
VUD 2.2
DC voltage
UEN = 60 mV – 300 V
AUE 2.2
nonāĆāsinusoidal AC current
(true RMS value) *)
IEN = 200 mA – 5 A
VUE 2.2
nonāĆāsinusoidal AC voltage
(true RMS value) *)
UEN = 60 mV – 519 V
TUA 2.2
DC standard signals
IEN = 20 mA
UEN = 60 mV, 10 V
*) also for use on transformer
AUE/VUE 2.2:
frequency range
crest factor
48 ... 62 Hz or
162/3 Hz, 100 Hz, other ratings on request
≤ 4 (peak value / rms value)
Output Ratings
output
current or voltage output
refer to General Data
input quantity
IE
rated input
IEN
current
measuring range
modulation range
overload limit
continuously
max. input voltage
permissible
power consumption
DC current
20 mA
0 ... IEN
1.2 IEN
2 IEN max.
16 V
2.4 V based on 20 mA
Output Ratings
current output
load independent DC current
output current
IA
rated current
IAN 0 ... 20 mA
0 ... 500 W (rated load 250 W)
load range
RA
load error
≤ 0.1% based on 50% load change
residual ripple
≤ 30 mVss
idling voltage
≤ 25 V
response time
≤ 0.05 s based on RA max
Input and output are galvanically isolated.
Other Ratings
as an option
accuracy
auxiliary voltage
dimensions WxHxL
weight
dual channnel unit
for detailed information refer to Data Sheet No. 061.##
for detailed information refer to Data Sheet No. 049.##
Other Ratings
selectable from standard output ratings
via jumpers located behind front panel
accuracy
class 0.5 (±0.5% of end value)
class 0.2 (±0,2% of end value) only for DC models on request
auxiliary voltage
refer to General Data
dimensions WxHxL
22.5 mm x 80 mm x 115 mm
weight
approx. 0.12 kg
class 0.2 (±0.2% of end value)
none required
22.5 mm x 80 mm x 115 mm
approx. 0.12 kg
on request
Short Form Data
=
RS232
RS485
RS232
RS485
Standard Signal
Interface Converter
MURS232/485
Short Form Data
RS232-RS485
Converter
APRS232/485
with mP
Input Ratings
Function
input quantity
The converter is directly plugged to a 9 - contact serial interface of a PC.
It converts the RS232 signals of the PC into standardized RS485
signals. This conversion enables to simultaneously connect several
devices by a 2 - wire junction to one interface and to inquire one after
another.
The device is protocol - transparent, i.e. the RS485 signals are translated 1:1 into the signals of the RS232 interface.
Depending on the 2 - wire junction of the RS485, a half- duplex operation only will be possible. This means, only one device may send in the
network at a time. By that a simultaneous operation of sending and
receiving will not be possible.
IE
UE
DC current or
DC voltage
rated input
current
IEN 20 mA
voltage
UEN 10 V / 1 V
measuring range
current input
0 ... IEN
“live zero” option
modulation range,
1.2 IEN
admissible
overload limit
continuously
1.2 IEN
1 s max.
2 IEN
input resistance
50 W
1 MW / 100 kW
voltage input
0 ... UEN
1.2 UEN
1.2 UEN
2 UEN
Interfaces
type
Baud - rate
as an option:
switching output
open collector
switching output
MOS FET
Ratings
case details
terminals
RS 232 (V.24) and
RS 485 (SCPI commands)
19200 Baud
8 ... 40 V DC / 10 ... 30 mA
insulation voltage 1 kV
for voltages up to 230 V AC/DC
and currents up to 100 mA
insulation voltage 3 kV
auxiliary supply
dimensions
weight
metalized thermoplastic case
SUB - D 9 contact (RS232 interface on the PC)
screw - terminals 2 contacts,
wire - cross section 4 mm2 max.
(RS485 interface)
power adapter prim. AC 230 V, ±10%, 50 Hz,
sec. DC 9 V / 250 mA
included with supply
73 mm x 34 mm x 12 mm
approx. 50 g
Other Ratings
accuracy
auxiliary voltage
dimensions WxHxL
weight
±0.1% and ±1 count
(for 0 ... IEN resp. 0 ... UEN )
refer to General Data
22.5 mm x 80 mm x 115 mm
approx. 0.12 kg
120 W
RS 232
RS 485
MU - RS232/485
UH
for detailed information refer to Data Sheet No. 052.##
~
=
120 W
for detailed information refer to Data Sheet No. 054.##
Short Form Data
=
Transducers for
Temperature (Pt 100)
PTU 2.0 L
Input Ratings
Other Ratings
input quantity
temperature (for RTD Pt 100)
Spans DT
Initial Temperature TE1
accuracy
auxiliary voltage
dimensions WxHxL
weight
– 200 °C
– 150 °C
– 100 °C
– 50 °C
0 °C
+ 50 °C
+100 °C
+150 °C
+200 °C
100 K
150 K
200 K
300 K
400 K
500 K
600 K
700 K (for TE1
100°C only)
800 K (for TE1
0°C only)
900 K (for TE1 –100°C only)
1000 K (for TE1 = –200°C only)
or deviating from standard values in the range of 100 ... 1000 K
measuring range
TE1 ... TE2 = TE1 + DT
input
potential–free differential input
connection
2–, 3– or 4–wire system
±0.5% referred to the span DT
refer to General Data
22.5 mm x 80 mm x 115 mm
approx. 0.12 kg
Output Ratings
current output
output current
IA
rated current
IAN
load range
RA
load error
residual ripple
idling voltage
response time
load independent DC current
4 ... 20 mA
0 ... 500 W (based on 20 mA)
≤ 0.1% based on 50% load change
≤ 1%rms of IAN with load RAN
≤ 16 V
≤ 1 s based on RA max
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
for detailed information refer to Data Sheet No. 050.##
– specifications subject to change without notice; date of issue 02/10 –
Operating Manual
679.B.421.01
Digital Multifunctional
Power Meter
with LCD Display:
Software
WPM 735 E
WPM 735 P
Contents
1.
General Information ..................................................................................................... 2
1.1. System requirements ................................................................................................... 2
1.2. Connection................................................................................................................... 2
1.3. Starting the software .................................................................................................... 3
2.
Displaying Measured Values ....................................................................................... 4
2.1. Realtime Data .............................................................................................................. 4
2.2. Common Data.............................................................................................................. 5
2.3. Harmonics.................................................................................................................... 6
2.4. Demand ....................................................................................................................... 7
2.4.1. Reading demand data ............................................................................................... 7
2.4.2. Clearing demand values............................................................................................ 7
2.5. Maximum/minimum values .......................................................................................... 8
2.5.1. Reading maximum/minimum values.......................................................................... 8
2.5.2. Clearing maximum/minimum values.......................................................................... 8
2.6. Energy counters........................................................................................................... 9
2.6.1. Reading energy counters .......................................................................................... 9
2.6.2. Resetting energy counters......................................................................................... 9
2.7. SOE events................................................................................................................ 10
2.7.1. Updating the event list ............................................................................................. 10
-1-
2.7.2. Clearing the event list.............................................................................................. 10
2.7.3. Displaying date and time of the meter..................................................................... 10
2.7.4. Setting date and time of the meter .......................................................................... 11
3.
Programming ............................................................................................................. 11
3.1. Displaying meter settings .......................................................................................... 11
3.2. Changing meter settings............................................................................................ 12
4.
Status Inputs.............................................................................................................. 12
1. General Information
1.1. System requirements
The WPM 735 software is suitable to run in Windows 98/2000/XP. The software can be
copied into any directory, and then, it can by started by double-clicking without installation.
Minimum hardware requirements:
CPU: Pentium 50 Mhz
RAM: 256MB
Screen resolution: min. 1024x768, recommended 1280x1024 or higher
Hard disk: min. 1MB free disk space
Mouse
The PC should have at least one serial port.
RS485-RS232 converter with cables
1.2. Connection
♦ Connect the RS485 bus according to the following figure.
Converter
RS232
735
RS 485
Terminal
15 RS485A+
16 RS485B–
17 SHLD
735
735
♦ Connect the RS232 cable to the PC and the converter.
-2-
WEIGEL-MESSGERÄTE GmbH
P.O.B. 720 154 D–90241 Nürnberg Phone: 0911 / 4 23 47–0
Erlenstraße 14 D–90441 Nürnberg
Fax: 0911 / 4 23 47–39
Management:
Phone: 0911 / 4 23 47–94
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
Description
RS485+
RS485 –
Shield
Operating Manual
679.B.421.01
Digital Multifunctional
Power Meter
with LCD Display:
Software
1.3. Starting the software
♦ Double-click WPM735_en.exe
Note: With a screen resolution of 1024x768 the card stack on top will not be shown.
♦ Enter the meter address inte the ID field top left.
♦ Select the serial port to which the RS485 converter is connected from the Port selection
list.
♦ Select the boud rate set on the meter from the Baud selection list.
♦ Select none from the Check selection list.
-3-
2. Displaying Measured Values
2.1. Realtime Data
♦ Click the
symbol to the left of the window and then click Realtime Data.
In the table on the right part of the window, the real-time measured values Ua, Ub, Uc, etc.
are displayed and updated continuously. (a = Phase L1, b = Phase L2, c = Phase L3)
♦ To update data manually, click Load.
-4-
WEIGEL-MESSGERÄTE GmbH
P.O.B. 720 154 D–90241 Nürnberg Phone: 0911 / 4 23 47–0
Erlenstraße 14 D–90441 Nürnberg
Fax: 0911 / 4 23 47–39
Management:
Phone: 0911 / 4 23 47–94
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
Operating Manual
679.B.421.01
Digital Multifunctional
Power Meter
with LCD Display:
Software
2.2. Common Data
The “Common Data” are the most important measured values, which are summarized on
one page.
♦ Click Common Data to the left of the window or click the Common Data tab.
In the table on the right part of the window, the most important measured values are
displayed and updated continuously.
-5-
2.3. Harmonics
♦ Click Harmonic to the left of the window or click the Harmonic tab.
nd
st
In the table on the right part of the window, the harmonics (2 to 31 harmonic), the total
harmonic distortion THD and the K factor of L1 phase current are displayed.
♦ Click the I L2 Harmonic or I L3 Harmonic button to display the L2 or L3 phase current
harmonics.
♦ Click the U L2 Harmonic, U L2 Harmonic, or U L3 Harmonic button to display the L1,
L2, or L3 phase voltage harmonics.
-6-
WEIGEL-MESSGERÄTE GmbH
P.O.B. 720 154 D–90241 Nürnberg Phone: 0911 / 4 23 47–0
Erlenstraße 14 D–90441 Nürnberg
Fax: 0911 / 4 23 47–39
Management:
Phone: 0911 / 4 23 47–94
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
Operating Manual
679.B.421.01
Digital Multifunctional
Power Meter
with LCD Display:
Software
2.4. Demand
♦ Click Demand to the left of the window or click the Demand tab.
2.4.1. Reading demand data
In the table on the right part of the window, demand data are displayed.
• Underneath Realtime Data, the actual demand values (15 min values) for current, active
power, reactive power, and apparent power are displayed.
• Underneath Max record, the maximum demand values for current, active power, reactive
power, and apparent power are displayed with time stamp (date and time).
2.4.2. Clearing demand values
♦ Click Clear Demand.
All maximum demand values are reset to 0.
-7-
2.5. Maximum/minimum values
♦ Click Max/Min Data to the left of the window or click the Max/Min Data tab.
2.5.1. Reading maximum/minimum values
In the table on the right part of the window, maximum and minimum values of the currents,
active power, reactive power, and apparent power, as well as the total harmonic distortions
THD of currents and voltages are displayed.
2.5.2. Clearing maximum/minimum values
♦ Click Clear max./min. real time data.
All limit values are deleted.
-8-
WEIGEL-MESSGERÄTE GmbH
P.O.B. 720 154 D–90241 Nürnberg Phone: 0911 / 4 23 47–0
Erlenstraße 14 D–90441 Nürnberg
Fax: 0911 / 4 23 47–39
Management:
Phone: 0911 / 4 23 47–94
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
Operating Manual
679.B.421.01
Digital Multifunctional
Power Meter
with LCD Display:
Software
2.6. Energy counters
♦ Click Energy to the left of the window or click the Energy tab.
2.6.1. Reading energy counters
In the table on the right part of the window, the imported and exported active and reactive
energies of the four quadrants are displayed.
♦ To update data, click Load.
2.6.2. Resetting energy counters
♦ Click Clear.
All the energy counters are reset.
-9-
2.7. SOE events
♦ Click SOE to the left of the window or click the SOE tab.
In the table on the right part of the window, the list of events is displayed.
2.7.1. Updating the event list
♦ Click SOE.
The event list is loaded again from the meter.
2.7.2. Clearing the event list
♦ Click Clear SOE.
All events are deleted in the meter and from the list.
2.7.3. Displaying date and time of the meter
♦ Click Meter Time.
Date and time of the meter are shown bottom left.
- 10 -
WEIGEL-MESSGERÄTE GmbH
P.O.B. 720 154 D–90241 Nürnberg Phone: 0911 / 4 23 47–0
Erlenstraße 14 D–90441 Nürnberg
Fax: 0911 / 4 23 47–39
Management:
Phone: 0911 / 4 23 47–94
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
Operating Manual
679.B.421.01
Digital Multifunctional
Power Meter
with LCD Display:
Software
2.7.4. Setting date and time of the meter
♦ Click Set Time.
Date and time of the connected PC are taken over to the meter.
3. Programming
3.1. Displaying meter settings
♦ Click Meter Set to the left of the window or click the Meter Set tab.
The communication parameters, the transformer ratings, the system mode, and the settings
for the analog output and pulse outputs are displayed.
♦ To update display, click Load.
- 11 -
3.2. Changing meter settings
♦ Change the settings as required.
(for details see Operating Manual No. 679.B.121.##)
♦ To transfer the settings to the meter, click Set.
Note: If communication parameters have been changed, you must set the same parameters
in the window top left.
♦ Check the setting by clicking Load.
4. Status Inputs
♦ Click I/O to the left of the window or click the I/O tab.
does not apply
Status inputs
In the right part of the window, the the status of switch contact inputs are indicated on
bottom.
- Specifications subject to change without notice; date of issue 07/10 - 12 -
WEIGEL-MESSGERÄTE GmbH
P.O.B. 720 154 D–90241 Nürnberg Phone: 0911 / 4 23 47–0
Erlenstraße 14 D–90441 Nürnberg
Fax: 0911 / 4 23 47–39
Management:
Phone: 0911 / 4 23 47–94
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
Data Sheet
679.D.121.03
Digital Multifunctional
Power Meter
with LCD Display
WPM 735 E
WPM 735 P
panel thickness
weight
Application
The digital Weigel Power Meters have been designed to display the
electrical parameters in low voltage systems. The meters are suitable for
use in 3 - phase 3 - wire or 3 - phase 4 - wire systems 230/400 V and for
use with current transformers N/5 A.
D WPM 735 E - H - T - AO - V3 is equipped with harmonics analysis, SOE
(event monitoring) as well as 8 status inputs, 2 pulse outputs and
1 analog output
D WPM 735 P - H - V3 is equipped with harmonics analysis and Profibus
The panel meters display the electrical parameters in the network:
voltages phase to neutral and phase to phase, currents of the 3 phases,
neutral current; unbalance rates for currents and voltages; active power,
reactive power, apparent power and power factor of the 3 phases and in
total; frequency; active energy and reactive energy in 4 quadrants (imported/exported, inductive/capacitive); demand and maximum demand
of the currents and powers; harmonics of current and voltage up to the
31st harmonics as well as total harmonic distortion (THD). Various ratings are displayed at the same time and can be selected by pressing
buttons.
Special events SOE (Status of Events) are recorded with date and time
stamp.
Using the integrated RS485 or Profibus interface, all measured ratings
and the events can be read out and settings can be made.
Software for read out and configuration is available via download from
www.weigel-messgeraete.de
Using the front - side buttons, the voltage and current transformer ratios,
the system type and communication parameters can be set and the energy counters and demand values can be reset.
Functional Principle
WPM 735 is a microprocessor-controlled digital measuring device for
measuring, calculating, and indicating electrical ratings.
The meter has 3 voltage inputs and 3 current inputs to enable simultaneous evaluation of voltage, current and power for all three phases.
The meter must be connected via current transformers.
General Technical Data
Case
case details
material of case
panel fixing
enclosure code
insulation voltage
suitable to be mounted in switch gear panels
plastics, black
plastic clips
IP 50 front of case
IP 20 rear of case
2 kV AC
inputs – outputs – auxiliary supply
5 kV (peak), 1.2/50 mS
pulse withstand
voltage
insulation resistance
MTBF
operating elements
≥50 MW
≥50000 h
6 buttons
Terminals
wire-cross section
3 plugable screw-terminal barrier strips
2.5 mm2 max.
Dimensions
bezel
bezel height
panel cutout
mounting depth
V 96 mm
18 mm
V 90+0.5 mm
80 mm minus panel thickness
5 mm
approx. 0,7 kg
Electrical Data
system type
rated voltage
PT ratio
rated current
current connection
CT ratio
overload capacity
current
voltage
frequency range
3 - phase 3 - wire or 3 - phase 4 - wire system,
unbalanced loads
230/400 V
adjustable 1 to 2500
5A
via CTs N/5 A
adjustable 1 to 10000
1.2 times continously
10 times for 1 sec.
1.2 times continously
2 times for 1 sec.
35 ... 65 Hz
Auxiliary Supply
auxiliary voltage
power consumption
85 ... 265 V AC or 80 ... 300 V DC
<3 VA
Display/Measuring/Accuracy
display
LCD display, green, 65 mm x 65 mm
with backlight (when pressing a button)
Measured quantities
voltages
U
L-L and L-N
currents
I
L1, L2, L3 and N
active power
P
L1, L2, L3 and SP
reactive power Q
L1, L2, L3 and SQ
apparent power S
L1, L2, L3 and SS
power factor
PF
L1, L2, L3 and in total
frequency
F
F
active energy
imported and exported
reactive energy
induktive and capacitive
demand and maximum demand of currents and powers
demad time
15 min (fixed)
unbalance rates of voltages und currents
recording of maximum and minimum ratings of I, P, Q, S
individual harmonics of current and voltage up to the 31st harmonics
THD
total harmonic distortion
Measuring unit
Measuring range Resolution Accuracy
voltages (L-L, L-N)
0 ... 600 V
0.01 V
0.2%
currents (L1, L2, L3, N)
0 ... 50 kA
0.001 A
0.2%
active power
0 ... ±100 MW
0.1 W
0.2%
reactive power
0 ... ±100 MVar
0.1 Var
0.2%
apparent power
0 ... ±100 MVA
0.1 VA
0.2%
power factor
–1 ... +1 ... –1
0.001
0.2%
frequency
35 ... 65 Hz
0.01 Hz
0.1%
active energy*) 0 ... 99,999,999.9 kWh
0.1 kWh
1.0%
reactive energy*) 0 ... 99,999,999.9 kVarh 0.1 kVarh
1.0%
unbalance rate
0 ... 100%
0.001
1.0%
individual harmonics
0 ... 100%
0.001
1.0%
THD
0 ... 100%
0.001
1.0%
*) The accuracy for energy measurement is class 1
according to IEC 62053–21.
Data Sheet
679.D.121.03
Digital Multifunctional
Power Meter
with LCD Display
4 Quadrants Measurement
Reactive energy
Q2
capacitive
generator/
export
-1
Q3
0
+90°
±180°
inductive
generator/
export
inductive
motor/
import
ϕ
type
protocol
baud rate
address
RS485
MODBUS–RTU
2400, 4800, 9600, 19200
1 ... 247
+1
0°
-90°
Q1
RS485 Interface
Active energy
cos ϕ
Q4
capacitive
motor/
import
0
WPM 735
Environmental
WPM 735
operating temperature –25 ... +70°C
range
storage temperature
–40 ... +85°C
range
relative humidity
5% ... 95% non-condensing
WPM 735
example
Status Inputs
WPM 735 E is equipped with 8 status inputs which can be used, e.g. for
recording switch positions.
Standards
EN 55022
EN 55024
EN 60950–1
EN 61000–3
–2
internal supply
30 V
–3
Pulse Outputs
WPM 735 E is equipped with two pulse outputs:
D Output P1 provides pulses for total active energy.
D Output P2 provides pulses for total reactive energy.
supply+
to receptor
supply –
WPM 735 E
type
pulse weight
pulse width
IEC 61000–4
–2
–3
–4
external devices
optocoupler open–collector, max. 50 V, 50 mA
adjustable 1000 ... 5000,
depending on CT ratio,
i.e. with 1 kWh/kvarh X CT
the set pulse count will be output
max. 200 ms
Analog Output
WPM 735 E is equipped with one analog output, to which a measured
rating can be assigned.
type
4 ... 20 mA
burdan range
0 ... 500 W
scale factor
1 ... 10 adjustable
–5
–6
–8
–11
Information technology equipment –
Radio disturbance characteristics –
Limits and methods of measurement
Information technology equipment –
Immunity characteristics –
Limits and methods of measurement
Information technology equipment – Safety –
Part 1: General requirements
Electromagnetic compatibility (EMC) – Limits –
Part 3–2: Limits for harmonic current emissions
(equipment input current <= 16 A per phase)
Part 3–3: Limitation of voltage changes,
voltage fluctuations and flicker in public
low–voltage supply systems, for equipment
with rated current <= 16 A per phase and
not subject to conditional connection
Electromagnetic compatibility (EMC) –
Testing and measurement techniques –
Part 4–2: Electrostatic discharge
immunity test.
Part 4–3: Radiated, radio–frequency,
electromagnetic field immunity test
Part 4–4: Electrical fast transient/burst
immunity test
Part 4–5: Surge immunity test
Part 4–6: Immunity to conducted disturbances,
induced by radio–frequency fields
Part 4–8: Power frequency magnetic field
immunity test
Part 4–11: Voltage dips, short interruptions
and voltage variations immunity tests
3–phase 3–wire system, 3 CTs
Terminals
addon module
auxiliary
supply
base unit
currents
voltages
No.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
Term.
L/+
NC
N/–
NC
FG
NC
NC
NC
I32
I31
I22
I21
I12
I11
Description
auxiliary supply
not connected
auxiliary supply
not connected
protective ground
not connected
not connected
not connected
current L3
current L3
current L2
current L2
current L1
current L1
RS485A+ RS485+
RS485B– RS485 –
SHLD shield
NC
not connected
NC
not connected
A1
analog output+
AG
analog output –
VN
neutral
NC
not connected
V3
voltage L3
NC
not connected
V2
voltage L2
NC
not connected
V1
voltage L1
No.
29
30
31
32
33
Term.
P2–
P2+
P1–
P1+
Scom
WPM 735 E only
pulse output
reactive energy
pulse output
active energy
supply +30V
status inputs common
34 S8
status input 8
35 S7
status input 7
36 S6
status input 6
37 S5
status input 5
38 S8
status input 4
39 S7
status input 3
40 S6
status input 2
41 S5
status input 1
42 NC
not connected
No. Term. WPM 735 P only
29 RB
termination resistor B
30 P–
Profibus –
31 P+
Profibus +
32 RA
termination resistor A
33 SHLD shield
34 NC
not connected
35 NC
not connected
36 NC
not connected
37 NC
not connected
38 NC
not connected
39 NC
not connected
40 NC
not connected
41 NC
not connected
42 NC
not connected
RS485
Ground
3–phase 3–wire system, 2 CTs
auxiliary
supply
currents
voltages
RS485
Ground
Dimensions
addon
module
Connections
3–phase 3–wire system
Ordering Information
Multifunctional Power Meter with LCD Display, 96 mm x 96 mm
WPM 735 E - H - T - AO - V3 with harmonics analysis, SOE, 8 status
inputs, 2 pulse outputs and 1 analog output
WPM 735 P - H - V3 with harmonics analysis and Profibus
auxiliary
supply
currents
voltages
RS485
Ground
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
Ordering Example
WPM 735 P - H - V3
Multifunctional Power Meter with LCD Display, harmonics analysis, and
Profibus
- specifications subject to change without notice; date of issue 08/10 -
Data Sheet
679.D.101.04
Digital Multifunctional
Power Meter
with LED Display
WPM 600B-Z
Application
Display
The digital Weigel Power Meters have been designed to display electrical parameters in low voltage systems.
WPM 600 B - Z is suitable for use in 3 - phase 4 - wire systems 230/400 V.
This panel meter displays the following electrical parameters: voltage,
current, active power, reactive power, power factor, frequency, active
energy and reactive energy. The ratings are displayed one after the
other or can be selected by pressing a button.
Using the front - side buttons, the current transformer ratio can be set
and the energy values can be reset.
display
Functional Principle
Microprocessor-controlled digital measuring device that measures,
calculates and indicates electrical ratings.
WPM 600 B - Z has 3 voltage inputs and 3 current inputs to enable simultaneous evaluation of voltage, current and power for all three phases.
The meter must be connected via current transformers.
General Technical Data
Case
case details
material of case
panel fixing
enclosure code
insulation voltage
insulation resistance
MTBF
operating elements
Terminals
wire-cross section
2 screw-terminal barrier strips
2.5 mm2 max.
Dimensions
bezel
bezel height
panel cutout
mounting depth
panel thickness
weight
WPM 600 B - Z
V 96 mm
12 mm
V 90+0.5 mm
58 mm minus panel thickness
4 mm
approx. 0,5 kg
Electrical Data
voltage
frequency range
Measured Quantity
voltage
current
active power
reactive power
power factor (cos ϕ)
frequency
active energy
reactive energy
LED
U
I
P
Q
PF
F
EP
EQ
WPM 600 B - Z
Unit
4 digits
V
4 digits
A
4 digits
kW
4 digits
kvar
3 digits
–
4 digits
Hz
9 digits *)
kWh
9 digits*)
kvarh
*) spread over 3 displays
Auxiliary Supply
auxiliary voltage
power consumption
85 ... 265 V AC or 80 ... 300 V DC
<2 VA
Accuracy at Reference Conditions
suitable to be mounted in switch gear panels
plastics, black
plastic clips
IP 50 front of case
IP 20 rear of case
2 kV
50 MW
50000 h
4 membrane keys
system type
rated voltage
rated current
current connection
CT ratio
overload capacity
current
digit height
3x 4 - digit 7 - segment LED display, red,
and 8 red LEDs for measuring unit
15 mm
3 - phase 4 - wire system, unbalanced loads
230/400 V
5A
via CTs N/5 A
adjustable 1 to 1999
1.2 times continously
10 times for 1 sec.
1.2 times continously
2 times for 1 sec.
35 ... 65 Hz
voltage
current
power
active energy
0.5%
0.5%
1.0%
class 1
Environmental
operating temperature –25 ... +70°C
range
storage temperature
–40 ... +85°C
range
relative humidity
5% ... 95% non-condensing
Standards
DIN EN 61010–1
DIN EN 61326–2–1
Safety requirements for electrical equipment
for measurement, control, and laboratory use –
Part 1: General requirements
Electrical equipment for measurement, control
and laboratory use – EMC requirements –
Part 2–1: Particular requirements –
Test configurations, operational conditions
and performance criteria for sensitive test
and measurement equipment for
EMC unprotected applications
Data Sheet
679.D.101.04
Digital Multifunctional
Power Meter
with LED Display
Terminals/Connection Diagram
1 L/+ auxiliary supply
3 N/– auxiliary supply
6–14
reserved
0.5 A
Dimensions
15
16
17
18
19
20
I32
I31
I22
I21
I12
I11
current L3
current L3
current L2
current L2
current L1
current L1
22
24
26
28
N
L3
L2
L1
neutral
phase L3
phase L2
phase L1
Ordering Information
Type
Multifunctional Power Meter with LED Display
WPM 600 B - Z for use in 3 - phase 4 - wire systems, for voltage,
current, active power, reactive power, power factor,
frequency, active energy, reactive energy,
96 mm x 96 mm
Ordering Example
WPM 600B - Z
Multifunctional Power Meter with LED Display
for use in 3 - phase systems
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
- specifications subject to change without notice; date of issue 06/10 -
Product Guide
DG – Series
730.U.001.02
Digital Large-Sized
Displays and Meters,
programmable
192x 96
240x 96
288x 96
336x 96
384x 96
336x144
432x144
528x144
624x144
720x144
✓ for DC Current and Voltage, Standard Signals
✓ for Parallel BCD-Input
✓ for Serial Input
RS 232 C / V.24, 20 mA / TTY, RS 422 and RS 485
✓ Digit Height 57 or 100 mm
✓ LED red or green
✓ 3, 4, 5 Digits
✓ Choice of Caption, e.g. C, km/h
✓ Auxiliary Supply 100 ... 240 V AC/DC or 18 ... 35 V DC
✓ Wall-Mounting Case (optional)
General Data
Digital Large-Sized
Displays and Meters,
programmable
dimensions (in mm)
Application
DGA/DGP/DGS
used to display information to be viewed a great distance off, e.g. to
indicate operating conditions, malfunctions, alarm signals, nominal/
actual values, order - / production line numbers, stock numbers, quantities, weights, rate, speed, power output and other units to be measured
in industrial use.
Additional fields of application: automation of production, display of
date, time of day, display of local weather conditions / environmental
influence.
Digital large - sized indicators using LED - technology are utmost robust. They are suitable under nearly all environmental conditions.
Moreover, they are a low - priced solution in the line of large - sized indicators.
Custom - made designs (size of display, digit heights, number of digits
and arrangement, LED - colour, inputs, additional lettering, case configuration (e.g. wall - mounting case) are available on request.
for mounting in
switchboards,
machine tools,
mosaic grid panels (optional)
wall -, ceiling - or outdoor mounting
for measuring
DC current or DC voltage,
standard signals
display of
BCD or serial data,
display of date / time of day (optional)
data format
analog inputs
direct or for use on transducer, shunt,
transformer
digital inputs
parallel or multiplex,
SPS 24 V active high
DGU
front bezel
panel cutout
mounting depth
bezel height
DGA/DGP/DGS
DGU
front bezel
panel cutout
mounting depth
bezel height
DGA/DGP/DGS
DGU
front bezel
panel cutout
mounting depth
bezel height
DGA/DGP/DGS
DGU
front bezel
panel cutout
mounting depth
bezel height
digit height / number of digits
–/D : without/with caption (e.g. C)
57–30
57–40
57–50
57–30/D
57–40/D
57–30/D1
57–40
192 x 96
240 x 96
288 x 96
186 x 90
234 x 90
282 x 90
94
94
94
4
4
4
57–60
57–50/D
57–60/D
57–40/D1
57–50/D1
57–60/D1
57–60
336 x 96
384 x 96
432 x 96
330 x 90
378 x 90
426 x 90
94
94
94
4
4
4
100–30
100–40
100–50
100–30/D
100–40/D
100–30/D1
100–40
336 x 144
432 x 144
528 x 144
330 x 138
426 x 138
522 x 138
94
94
94
4
4
4
100–50/D
100–40/D1
100–50/D1
100–60
624 x 144
720 x 144
618 x 138
714 x 138
94
94
4
4
Mechanical Data
Display
case details
material of case
front bezel
terminals
enclosure code
display
7–segment LED
DGA/DGP/DGS/DGU
digit height
viewing distance max.
display colour
57
100
57 mm
100 mm
approx. 20 – 25 m approx. 40 – 45 m
red or green
panel fixing
rectangular format
aluminium, black, anodized
aluminium, black, anodized, mat
screw terminal barrier strip
IP 65 front of case
IP 20 rear of case
screw clamps
mask lettering
caption, custom logo or extra
lettering to special order
operating elements
on rear side of case
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
0oC ... +23oC ... +50oC
–20oC ... +23oC ... +70oC
75% annual average, non - condensing
Extras
lettering
custom logo
special measuring
range
factory calibration
non - standard auxiliary
voltage
display
please specify
please specify
on request
please specify
on request
tendency symbols or special arrangement
of digits on request
on request
non - standard set
of symbols
code - table
enclosure code IP 65
for overall product
housed in sheet metal case for wall -, ceiling or outdoor mounting on request
additional options
additional data
prices
on request
on request
refer to Price Sheet No. 730.P.000.##
Short Form Data
Short Form Data
BCD
mA, V
Large-Sized Indicators
for DC Current and
DC Voltage,
Standard Signals
Digits
30
3 digits
40
4 digits
50
5 digits
DGA 57
192 x
240 x
288 x
336 x
384 x
96 mm
96 mm
96 mm
96 mm
96 mm
Display Span
–199 ... 999
–1999 ... 9999
–19999 ... 30000
Input
A
DC Current 20 / 200 mA and
DC Voltage 2 / 10 / 40 / 200 V
Auxiliary Voltage
1
18 ... 35 V DC, electr. insulated
2
100 ... 240 V AC/DC, electr. insulated
Options
0
none
x
on request
DGA 100
336 x 144 mm
432 x 144 mm
528 x 144 mm
624 x 144 mm
720 x 144 mm
Digital Large-Sized
Displays
with BCD-Input
Digits
30
3 digits
40
4 digits
50
5 digits
DGP 57
192 x
240 x
288 x
336 x
384 x
96 mm
96 mm
96 mm
96 mm
96 mm
Input
A
DGP 100
336 x 144 mm
432 x 144 mm
528 x 144 mm
624 x 144 mm
720 x 144 mm
Display Colour
R
red
G
green
Caption
–
none
D
with 2 digit measuring unit
D1
with 4 digit measuring unit
Electrical Data
Electrical Data
measuring ranges
DC current
±20 mA
±200 mA
DC voltage
±2 V
±10 V
±40 V
±200 V
measuring principle
conversion rate
error, max.
auxiliary voltage
inputs
power consumption
input resistance
100 10 100 k
560 k
2.2 M
12 M
dual - slope, integrating
approx. 1 read per second
±0.01% of reading, ±1 count
DC 24 V (18 ... 35 V), electrically insulated or
AC/DC 100 ... 240 V, electrically insulated
12 VA max. (DGA 57), 18 VA max. (DGA 100)
BCD Input, 24 V active high
Auxiliary Voltage
1
18 ... 35 V DC, electr. insulated
2
100 ... 240 V AC/DC, electr. insulated
Display Colour
R
red
G
green
Caption
–
none
D
with 2 digit measuring unit
D1
with 4 digit measuring unit
24 V (35 V max.) active high,
parallel or multiplex
low: < 7 V, high: >10 V
15 k
2 ms min.
ST display test
DT display blanking
DP decimal point
LE latch enable
DC 24 V (18 ... 35 V), electrically insulated or
AC/DC 100 ... 240 V, electrically insulated
12 VA max. (DGP 57), 18 VA max. (DGP 100)
input resistance
pulse width
functional inputs
auxiliary voltage
power consumption
Connection
100...240 V
AC/DC
0V
overrange indication
center segments will flash
mean value set - up
adjustable from 1 up to 99 readings
round - off last digit
adjustable: 2nd, 5th, 10th steps, none
zero point and end value programmable
18...35 V
DC
0V
plug - in terminal barrier strip, double -row
Display
Connection
data - and functional inputs
input assignment
plug - in terminal barrier strip
Input
200 mA
20 mA
ordering example
DGA
Gnd
2V
10 V
40 V
200 V
200 mA
DC
0V
20 mA
100...240 V
AC/DC
0V
18...35 V
DC 0 V
switches
jumper when
measuring
current
100 – 30 A 1 0 R –
ordering example
DGP
100 – 40 A 1 G D
Short Form Data
Short Form Data
DCF 77
V.24
Digital Large-Sized
Indicators
for Date/Time
Quartz Clocks / DCF 77
Digital Large-Sized
Displays
with Serial Input
192 x
240 x
288 x
336 x
384 x
432 x
96 mm
96 mm
96 mm
96 mm
96 mm
96 mm
Input
A
DGS 100
3 digits
4 digits
5 digits
6 digits (DGS 57 only)
RS 232 C / V.24 and 20 mA / TTY
and RS 422 resp. RS 485
Caption
–
none
D
with 2 digit measuring unit
D1
with 4 digit measuring unit
Electrical Data
inputs
RS 232 C / V.24, 20 mA / TTY and
RS 422 resp. RS 485
150 / 300 / 600 / 1200 /
2400 / 4800 / 9600 / 19200 baud
with/without parity bit,
even / odd parity,
7 or 8 data bits
with/without STX/ETX, CR, echo
adjustable by DIP - switch
100 ms min. between two telegrams
display test, fore - running zeros, fading - out,
fading - out of up to 98 fore - running symbols,
flashing sign, display annunciator
direction left→right / right→left
01 ... 99
DC 24 V (18 ... 35 V), electrically insulated or
AC/DC 100 ... 240 V, electrically insulated
12 VA max. (DGS 57), 18 VA max. (DGS 100)
on request
data transfer
interval
software functions
device addresses
auxiliary voltage
power consumption
set of symbols
DGU 100
528 x 144 mm
720 x 144 mm
0V
plug - in terminal barrier strip
passive
Input
D
for DCF 77 reception module
Auxiliary Voltage
1
18 ... 35 V DC, electr. insulated
2
100 ... 240 V AC/DC, electr. insulated
Options
0
none
x
on request
Electrical Data
time - of - day clock
time - base accuracy
battery backup
input
auxiliary voltage
power consumption
quartz clock, battery -buffered
±5 ppm max. / year
lithium torch battery 3 V
for DCF 77 reception module DCF – EBG
with remote supply
DC 24 V (18 ... 35 V), electrically insulated or
AC/DC 100 ... 240 V, electrically insulated
12 VA max. (DGU 57), 18 VA max. (DGU 100)
Display
display modes
intensity
24 - hour clock operation only (separated by
colons) or date only (separated by decimal
points) or change time of day / date
each 1 ... 99 s programmable
adjustable
Connection
plug - in terminal barrier strip
ordering example
line
termination
Digits
40
4 digits
60
6 digits
Display Colour
R
red
G
green
18...35 V
DC 0 V
Connection
100...240 V
AC/DC
288 x 96 mm
384 x 96 mm
Auxiliary Voltage
1
18 ... 35 V DC, electr. insulated
2
100 ... 240 V AC/DC, electr. insulated
Display Colour
R
red
G
green
336 x 144 mm
432 x 144 mm
528 x 144 mm
624 x 144 mm
720 x 144 mm
18...35 V
DC 0 V
DGU 57
100...240 V
AC/DC
0V
Digits
30
40
50
60
DGS 57
DGU
minute
cycle
output
DCF 77
reception
module
57 – 60 D 2 0 G
DCF 77 Reception Module
jumper
active
ordering example
DGS
57 – 60 A 2 G –
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 DĆ90241 Nürnberg Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 DĆ90441 Nürnberg Telefax: 0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
reception module for DCF 77 - transmitter in Braunschweig
case
thermoplastic case for wall - mounting
dimensions (lxwxd)
120 mm x 80 mm x 60 mm
enclosure code
IP 65
ordering example
DCF
– EBG
– specifications subject to change without notice; date of issue 08/03 –
Product Guide
L Series
720.U.001.04
Digital Bargraph
Indicators,
programmable
72x24
96x24
144x48
✓
✓
✓
✓
✓
✓
✓
✓
for DC Current and Voltage, Standard Signals
72x24, 96x24 or 144x48 mm Bezel
Vertical or Horizontal Display
20, 30, 40 or 50 Segments
Red or Green LEDs
One or two Bargraphs
With or without additional 7-Segment Display
Auxiliary Supply 115/230 V AC, 50 Hz or 24 V DC
General Data
Short Form Data
mA, V
Digital Bargraph
Indicators,
programmable
Application
Bargraph Indicator
for DC Current and
DC Voltage,
programmable
LEA 20
for standard applications in switchboard -, control equipment -,
process control -, and machine tool industry.
for mounting in
switchboards, machine tools or
mosaic grid panels (optional)
for measuring
DC current or DC voltage, standard signals
connection
direct or for use on transducer, shunt
72 x 24 mm
Bargraph
20
20 Segments
Input
A
DC Current / 200 mA and
DC Voltage 2 / 10 / 40 / 200 V
Auxiliary Voltage
1
24 V DC electrically insulated
Mechanical Data
Options none
Display Colour
R
red
G
green
case details
material of case
bezel
Appearance
H
Vertical Display
Q
Horizontal Display
terminals
enclosure code
panel fixing
mask lettering
rectangular format
glass - fibre reinforced thermoplastics, black
black, mat or optionally
gray varnished (similar to RAL 7037) or
siliceous gray varnished (similar to RAL 7032)
screw terminal barrier strip
IP 50 front of case
IP 20 rear of case
screw clamps
0 ... 100% standard, extra
caption, custom logo or
special lettering to special order
behind removable display lens
operating elements
climatic suitability
operating
0oC ... +23oC ... +50oC
temperature range
storage
–20oC ... +23oC ... +70oC
temperature range
relative humidity
75% annual average, non - condensing
dimensions (in mm) LEA 20
LEA 23/30
bezel
72 x 24
96 x 24
+0.7
+0.3
x 22.2
92+0.8 x 22.2+0.6
panel cutout
68
panel thickness
5
5
mounting depth
120
120
bezel height
5.2
7.5
weight approx.
0.15 kg
0.25 kg
dimensions (in mm) LEA/LZA 43/50
bezel
144 x 48
panel cutout
138+1.0 x 45+0.6
panel thickness
45
mounting depth
120
bezel height
8.5
weight approx.
0.6 kg
mounting in mosaic grid panels
please specify type / manufacturer of mosaic grid when ordering
Electrical Data
measuring ranges
DC current
±20 mA
±200 mA
DC voltage
±2 V
±10 V
±40 V
±200 V
measuring principle
conversion rate
error, max.
auxiliary voltage
power consumption
input resistance
100 W
10 W
100 kW
560 kW
2.2 MW
10 MW
dual - slope, integrating
approx. 1 read per second
±0.01% of reading ±1 count/segment
24 V DC (20 ... 30 V), electrically insulated
2.5 VA max.
Display
scale length
display colour
appearance
overrange indication
zero point/end value
scaling
50 mm, 20 segments
red or green
vertical or horizontal display
each 2nd scale segment will flash
programmable
0 ... 100 %
Connection
24 V
0V
plug - in terminal barrier strip
DC
jumper when
measuring current
options
refer to last page
ordering example
LEA
20
A 1 0 R H
96 x 24 mm
Short Form Data
Bargraph Indicators
for DC Current and
DC Voltage,
programmable
Bargraph Indicator
for DC Current and
DC Voltage,
programmable
Bargraph
23
20 Segments &
3 - digit Display
30
30 Segments
Input
A
LEA 43
LEA 50
144 x 48 mm
DC Current 20 / 200 mA and
DC Voltage 2 / 10 / 40 / 200 V
Bargraph
43
40 Segments &
3 - digit Display
50
50 Segments
Input
A
Auxiliary Voltage
1
24 V DC electrically insulated
2
115/230 V AC, 50 Hz electr.
insulated
Auxiliary Voltage
1
24 V DC electrically insulated
2
115/230 V AC, 50 Hz electr.
insulated
Options refer to last page
Display Colour
R
red
G
green
Options refer to last page
Display Colour
R
red
G
green
Appearance
H
Vertical Display
Q
Horizontal Display
Appearance
H
Vertical Display
Q
Horizontal Display
Electrical Data
Electrical Data
measuring ranges
DC current
±20 mA
±200 mA
DC voltage
±2 V
±10 V
±40 V
±200 V
measuring principle
conversion rate
error, max.
auxiliary voltage
measuring ranges
DC current
±20 mA
±200 mA
DC voltage
±2 V
±10 V
±40 V
±200 V
measuring principle
conversion rate
error, max.
auxiliary voltage
power consumption
input resistance
100 W
10 W
100 kW
560 kW
2.2 MW
10 MW
dual - slope, integrating
approx. 1 read per second
±0.01% of reading, ±1 count/segment
24 V DC (20 ... 30 V), electrically insulated or
115/230 V AC, 50 Hz, electrically insulated
2.5 VA max.
Display
scale length
Display
display colour
appearance
overrange indication
mean value set -up
round - off last digit
zero point/end value
scaling
power consumption
input resistance
100 W
10 W
100 kW
560 kW
2.2 MW
10 MW
dual - slope, integrating
approx. 1 read per second
±0.01% of reading, ±1 count/segment
24 V DC (20 ... 30 V), electrically insulated or
115/230 V AC, 50 Hz, electrically insulated
5 VA max.
Display
LEA 30
75 mm, 30 segments
LEA 23
50 mm, 20 segments
LEA 23
7- segment LED,
digit height
7 mm
display span
–99 ... 999
red or green
vertical or horizontal display
each 2nd scale segment and
center segments will flash
adjustable from 1 to 500 reads
adjustable: 2nd, 5th, 10th step, none
programmable
0 ... 100 %, caption on request
scale length
Display
display colour
appearance
overrange indication
mean value set -up
zero point/end value
scaling
LEA 50
125 mm, 50 segments
LEA 43
100 mm, 40 segments
LEA 43
7- segment LED,
digit height
7 mm
display span
–99 ... 999
red or green
vertical or horizontal display
each 2nd scale segment and
center segments will flash
adjustable from 1 to 500 reads
programmable
0 ... 100 %, caption on request
DC
0V
24 V
230 V
0V
plug - in terminal barrier strip
115 V
plug - in terminal barrier strip
24 V
Connection
0V
Connection
DC
AC
jumper when
measuring current
jumper when
measuring current
ordering example
DC Current 20 / 200 mA and
DC Voltage 2 / 10 / 40 / 200 V
LEA
30
A 2 0 G H –M
ordering example
LEA
43
A 2 0 G Q
0V
115 V
230 V
LEA 23
LEA 30
Short Form Data
AC
LZA 43
LZA 50
144 x 48 mm
Short Form Data
Options
Bargraph Indicator
for DC Current, DC
Voltage with 2 Inputs,
programmable
Digital Bargraph
Indicators,
programmable
Bargraphs
43
2x 40 Segments &
2x 3 - digit Display
50
2x 50 Segments
LEA 23
LEA 30
LEA 43
LEA 50
LZA 43
LZA 50
Inputs
A
DC Current 20 / 200 mA and
DC Voltage 2 / 10 / 40 / 200 V
Auxiliary Voltage
1
24 V DC, electrically insulated
2
115/230 V AC, 50 Hz electr.
insulated
Options refer to right - side column
Display Colour
R
red
G
green
Appearance
H
Vertical Display
Q
Horizontal Display
Options
0
none
1*
analog output 0 ... 10 V
(load ≥1000 W)
2*
analog output 0/4 ... 20 mA
(load ≤500 W)
7*
1 control output open collector
30 V max., 0.5 A max., 12 W max.
8
2 control outputs open collector
30 V max. , 0.5 A max., 12 W max.
D*
display blanking input
active high, 24 V
H*
display hold input
active high, 24 V
R
2 setpoints
relay N.O. / N.C.
S*
serial output RS232
*LEA 23 / LEA 30 only
Additional Option
–M
min./max. value storage (LEA 23/43,
LZA 43 with reference measurement)
Electrical Data
Options
measuring ranges
DC current
±20 mA
±200 mA
DC voltage
±2 V
±10 V
±40 V
±200 V
measuring principle
conversion rate
error, max.
auxiliary voltage
R option
setpoints
LEA 23/30
LEA/LZA 43/50
response time
hysteresis
status indication
power consumption
input resistance
100 W
10 W
(input 1 & 2)
(input 1 only)
100 kW
(input 1 & 2)
560 kW
(input 1 & 2)
2.2 MW
(input 1 only)
10 MW
(input 1 only)
dual - slope, integrating
approx. 1 read per second
±0.01% of reading, ±1 count/segment
24 V DC (20 ... 30 V), electrically insulated or
115/230 V AC, 50 Hz, electrically insulated
5 VA max.
Display
same as LEA 43, LEA 50 (refer to preceding page)
Connection
plug - in terminal barrier strip
input 1
input 2
jumper when
measuring current
S option
RS232 serial output
Further options
scaling
extra lettering
custom logo
special measuring
range
factory calibration
non-standard
auxiliary voltage
mosaic grid mounting
0V
115 V
AC
230 V
0V
24 V
DC
jumper when
measuring current
LZA
baud rate (150 ... 19200 baud), parity,
data bits, device address, write direction
and protocol adjustable
non-standard (0 ... 100%)
please specify
please specify
on request
please specify
on request
on request
Connection
analog output
0 ... 10 V /
0/4 ... 20 mA
10 V /
20 mA
serial output
RS232
ordering example
2 relays N.O. / N.C.
max. 200V/0.05A, max. 20V/0.5A, max. 10 W
max. 250V/0.8A, max. 50V/4A, max. 200 W
approx. 1 ms
adjustable, internal adjustment
small, single - digit 7- segment LED
50
1 control
output o.c.
2 control outputs
LEA 23/30
2 control outputs
LEA/LZA 43/50
input blanked
control/ display
hold
setpoints
LEA 23/30
setpoints
LEA/LZA 43/50
A 2 0 R Q
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Telefax: 0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
– specifications subject to change without notice; date of issue 05/07 –
Product Guide
DA Series
713.U.001.04
Digital Remote
Displays,
parallel, serial
48x24
48x48
72x24
72x48
96x24
96x48
144x48
144x72
✓ for Parallel BCD Input
✓ for Serial Input
RS 232 C / V.24, 20 mA / TTY, RS 422 and RS 485
✓ 48x24, 48x48, 72x24, 72x48, 96x24, 96x48, 144x48
or 144x72 mm Bezel
✓ 7.62 / 10 / 14 / 20 or 25 mm Digit Height
✓ Red or Green LEDs
✓ 2, 3, 3½, 4, 4½, 5, 6 Digits
✓ Auxiliary Supply 5/15/24 V DC or 115/230 V AC, 50 Hz
General Data
Short Form Data
BCD
Digital Remote
Displays,
parallel, serial
Digital Remote
Displays
with BCD Input
Application
Digits
30
31
40
41
50
DAP 08
DAP 10
for standard applications in switchboard -, control equipment -,
process control -, and machine tool industry.
for mounting in
switchboards, machine tools or
mosaic grid panels (optional)
display of
BCD or serial data
data format
digital BCD inputs
parallel or multiplex,
TTL 5 V or SPS 24 V,
active high or active low
digital inputs (serial)
RS 232 C / V.24, 20 mA /TTY, RS 422, RS 485
–
–4
–7
48 x 24 mm
48 x 48 mm
72 x 24 mm
Inputs
A
multiplex active high 24 V
B
multiplex active low 24 V
C
parallel active high 24 V
D
parallel active low 24 V
E
multiplex active high 5 V
F
multiplex active low 5 V
G
parallel active high 5 V
H
parallel active low 5 V
Auxiliary Voltage
1
24 V DC
3
15 V DC
4
5 V DC
Mechanical Data
case details
material of case
bezel
terminals
enclosure code
panel fixing
mask lettering
operating elements
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
dimensions (in mm)
bezel
panel cutout
panel thickness
mounting depth
bezel height
dimensions (in mm)
bezel
panel cutout
panel thickness
mounting depth
bezel height
dimensions (in mm)
bezel
panel cutout
panel thickness
mounting depth
bezel height
dimensions (in mm)
bezel
panel cutout
panel thickness
mounting depth
bezel height
rectangular format
glass - fibre reinforced thermoplastics, black
black, mat or optionally
gray varnished (similar to RAL 7037) or
siliceous gray varnished (similar to RAL 7032)
screw terminal barrier strip
IP 50 front of case
IP 20 rear of case
screw clamps
caption, custom logo or extra
lettering to special order
on rear side of case
Display Colour
R
red
G
green
Electrical Data
inputs
TTL 5 V or
SPS 24 V, (input resistance 10 kW)
2 ms min.
ST display test
DT display blanking
DP decimal point
LE latch enabling
24 DC (20 ... 30 V) or
15 DC (12.5 ... 17.5 V) or
5 DC (4.75 ... 5.25 V)
electrically not insulated
2.5 VA max.
pulse width
functional inputs
0oC ... +23oC ... +50oC
–20oC ... +23oC ... +70oC
75% annual average, non - condensing
DAP 08/10
...– 4
...–7
48 x 24
48 x 48
72 x 24
45+0.6 x 22.2+0.3 45+0.6 x 45+0.6 68+0.7 x 22.2+0.3
3 digits
3½ digits
4 digits
4½ digits *
5 digits *
* DAP 08 only
auxiliary voltage
power consumption
5
120 incl. terminal barrier strip
5.25
DAP/DAS 14 ...– 4
(2/3/4 digits)
72 x 24
72 x 48
Display
5
120 incl. terminal barrier strip
5.25
DAP/DAS 14 ...– 4
(5/6 digits)
96 x 24
96 x 48
92+0.8 x 22.2+0.6 92+0.8 x 45+0.6
5
120 incl. terminal barrier strip
7.5
DAS 20/25
...– 4
144 x 48
144 x 72
display colour
display
digit height
7 - segment LED
DAP 08
DAP 10
red or green
68+0.7 x 22.2+0.6 68+0.7 x 45+0.6
7.62 mm
10 mm
Connection
refer to last page
ordering example
1
2
DAP
08
3
4 5 6 7
– 30 D 1 G –
138+1.0 x 45+0.6 138+1.0 x 68+0.7
45
120 incl. terminal barrier strip
8.5
Options
lettering, custom logo please specify
mosaic grid panel mounting, extra housing, non - standard auxiliary
voltage, non - standard inputs, software, connection cable with socket
connector on request, enclosure code IP 54 up to IP 65 front of case
using rubber cover attachment 96x48.
additional options
additional data
on request
on request
mounting in mosaic grid panel (all models)
please specify type / manufacturer of mosaic grid when ordering
ordering information
1
2
type
digit
height
3
#
– digits
4
5
6
7
input
power
supply
colour
case
Short Form Data
Short Form Data
BCD
V.24
Digital Remote
Displays with
BCD Input
DAP 14
–
–4
–
–4
72 x 24 mm
72 x 48 mm
96 x 24 mm
96 x 48 mm
Digital Remote
Displays with
Serial Input
Digits
20
30
40
50
60
2 digits (72x24/48)
3 digits
-”4 digits
-”5 digits (96x24/48)
6 digits
-”-
Input
A
multiplex, 24 V active high
DAS 14
–
–4
–
–4
72 x 24 mm
72 x 48 mm
96 x 24 mm
96 x 48 mm
Auxiliary Voltage
1
24 V DC
Input
20
30
40
50
60
2 digits (72x24/48)
3 digits
-”4 digits
-”5 digits (96x24/48)
6 digits
-”-
Input
A
B
D
RS 232 C / V.24 and 20 mA / TTY
RS 422
RS 485
Auxiliary Voltage
1
24 V DC
Display Colour
R
red
G
green
Display Colour
R
red
G
green
Electrical Data
Electrical Data
inputs
inputs
24 V (35 V max.) active high, multiplex
low: < 7 V, high: >10 V
15 kW
2 ms min.
ST display test
DT display blanking
DP decimal point
LE latch enable
24 V DC (20 ... 30 V) electrically not insulated
2.5 VA max.
input resistance
pulse width
functional inputs
auxiliary voltage
power consumption
devices addresses
data transfer
software functions
Display
display
digit height
display colour
7- segment LED
14 mm
red or green
auxiliary voltage
power consumption
Display
Connection
display
digit height
display colour
plug - in terminal barrier strip, double- row
72x24
RS 232 C / V.24 and 20 mA / TTY or
RS 422 or
RS 485
0 ... 9 or 00 ... 99
1200 / 2400 / 4800 / 9600 / 19200 baud
with / without parity bit,
even / odd parity,
7 or 8 data bits
with / without STX/ETX
programmable
display test, show/hide pre - zeros,
flashing characters/display,
write direction left→right / right→left
suppression of up to 99 characters,
timout function
24 V DC (20 ... 30 V) electrically not insulated
2.5 VA max.
1st Digit
BCD Input
7 - segment LED
14 mm
red or green
Connection
plug - in terminal barrier strip
DAS 14 A Input
2nd Digit
96x24
4th Digit
3rd Digit
DC
1st Digit
BCD Input
DAS 14 B Input
2nd Digit
3rd Digit
DAS 14 D Input
4th Digit
5th Digit
6th Digit
DC
NOTE: Close mounting on top of each other will generally not be
possible due to both terminal barrier strips and cables.
ordering example
1
2
DAP
14
3
4 5 6 7
– 50 A 1 R – 4
ordering example
1
2
DAS
14
3
4 5 6 7
– 40 B 1 R -
Short Form Data
V.24
Short Form Data
BCD
Digital Remote
Displays with
Serial Input
DAS 20
DAS 25
–
–4
144 x 48 mm
144 x 72 mm
Digital Remote
Displays with
BCD-Input
(continued)
Digits
30
3 digits
40
4 digits
50
5 digits *
* DAS 20 only
Input
A
DAP 08
DAP 10
RS 232 C / V.24 and 20 mA / TTY
and RS 485
Auxiliary Voltage
1
24 V DC electrically insulated
2
115/230 V AC, 50 Hz electr.
insulated
Display Colour
R
red
G
green
Electrical Data
Connection
inputs
device addresses
data transfer
Sub - D Connector Pin Assignment, 26 poles
Input multiplex
Input parallel
software functions
auxiliary voltage
power consumption
RS 232 C / V.24 and 20 mA / TTY and RS 485
01 ... 98
1200 / 2400 / 4800 / 9600 baud
with / without parity bit,
even / odd parity,
7 or 8 data bits
with / without STX/ETX
programmable
display test, show/hide pre - zeros,
flashing characters/display,
write direction left→right / right→left
suppression of up to 99 characters
24 V DC (20 ... 30 V), electrically insulated or
115/230 V AC, 50 Hz, electrically insulated
5 VA max.
Display
display
digit height
display colour
BCD Input
1st Digit
1st Digit
2nd Digit
2nd Digit
3rd Digit
4th Digit
3rd Digit
5th Digit*
7 - segment LED
DAS 20
20 mm
DAS 25
25 mm
red or green
4th Digit
latch
Connection
plug - in terminal barrier strip
2
DAS
20
3
DC
230 V
0V
0V
ordering example
1
115 V
AC
24 V
5th Digit*
*
4 5 6 7
– 40 A 1 R –
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
*DAP 08 only
Accessories
connector plug with cover
cable 1 m with plug
cable 2 m with plug
cable 3 m with plug
– specifications subject to change without notice; date of issue 04/07 –
Product Guide
DP Series
712.U.001.04
Digital Panel Meters,
programmable
48x24
48x48
72x24
96x24
96x48
96x96
144x48
144x72
✓
✓
✓
✓
✓
✓
✓
✓
✓
for DC Current and Voltage
for AC Current and Voltage
for Standard Signals
for Temperature (RTD Pt 100, Pt 1000)
48x24, 48x48, 72x24, 96x24, 96x48, 96x96, 144x72 or
144x48 mm Bezel
7.62 / 10 / 14 / 20 or 25 mm Digit Height
Red or Green LEDs
3, 3½, 4, 4½ or 5 Digits
Auxiliary Supply AC 115/230 V, 50 Hz or DC 24 V
General Data
mA, V
mV
Digital Panel Meters,
programmable
Application
for standard applications in switchboard -, control equipment -,
process control -, and machine tool industry.
for mounting in
switchboards,
machine tools or mosaic grid panels (optional)
for measuring
DC current or DC voltage,
AC current or AC voltage TRMS,
standard signals,
temperature
connection
analog inputs
direct or for use on transducer, shunt,
transformer
DPA 08
DPA 10
Input E
Input D
–
48 x 24 mm
–4 48 x 48 mm
–7 72 x 24 mm
Mechanical Data
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
dimensions (in mm)
bezel
panel cutout
panel thickness
mounting depth
bezel height
dimensions (in mm)
bezel
panel cutout
panel thickness
mounting depth
bezel height
dimensions (in mm)
bezel
panel cutout
panel thickness
mounting depth
bezel height
0oC ... +23oC ... +50oC
3
digits
3½ digits
4
digits
4½ digits *
5
digits *
*DPA 08 only
Display Span
–199 ...
999
–1999 ... +1999
–1999 ... 9999
–19999 ... +19999
–19999 ... 30000
Input
E
DC Current –19 ... 20 mA and
DC Voltage –10 ... 10 V
D
DC Voltage 0 ... 60 – 150 mV
Auxiliary Voltage
1
24 V DC electrically insulated
Options
0
none
F
Action Input: Display Dark/Hold/
ON/OFF/HELP, Segment Test
Display Colour
R
red
G
green
Electrical Data
measuring range
DC current
0/4 ... 20 mA
DC voltage
0 ... ±10 V
0 ... ±60 – 150 mV
measuring principle
conversion rate
error, max.
overrange indication
auxiliary voltage
power consumption
input resistance
100 W
input resistance
220 kW
470 kW
dual - slope integration
approx. 1 read per second
±0.01% of reading, ±1 count
center segments will flash
DC 24 V (20 ... 30 V), electrically insulated
2.5 VA max.
Display
–20oC ... +23oC ... +70oC
75% annual average, non - condensing
DPA 08/10
...– 4
...–7
48 x 24
48 x 48
72 x 24
45+0.6 x 22.2+0.3 45+0.6 x 45+0.6 68+0.7 x 22.2+0.3
5
120 incl. terminal barrier strip
5.25
DPA 14
...– 4
...–9
96 x 24
96 x 48
96 x 96
+0.8
+0.6
+0.8
+0.6
92
x 22.2
92
x 45
92+0.8 x 92+0.8
5
120 incl. terminal barrier strip
7.5
DPA 20/25
...– 4
144 x 48
144 x 72
display
digit height
7 - segment LED
DPA 08
7.62 mm
DPA 10
10 mm
display span
see above
display colour
red or green
overrange indication
center segments will flash
mean value set - up
adjustable from 1 up to 99 readings
zero point, end value, polarity sign and decimal point programmable
Connection
plug - in terminal barrier strip
E input
D input
138+1.0 x 45+0.6 138+1.0 x 68+0.7
45
120 incl. terminal barrier strip
8.5
mounting in mosaic
grid panels
please specify type / manufacturer
of mosaic grid when ordering
options
additional data
refer to last page
on request
20mA: -19 ... +20 mA (Ri = 50 W)
10V: -10 ... +10 V (Ri = 150 kW)
ordering information
1
2
type
digit
height – digits
3
Ri =
47 kW
Action Input (F option)
operating elements
Digits
30
31
40
41
50
isolated
power supply
panel fixing
mask lettering
Digital Panel Meters,
for Standard Signals
Action Input (F option)
terminals
enclosure code
rectangular format
glass - fibre reinforced thermoplastics, black
black, mat or optionally
gray varnished (similar to RAL 7037)
or siliceous gray (similar to RAL 7032)
screw terminal barrier strip
IP 50 front
operating elements are accessible behind
removable display lens.
IP 20 front
operating elements are passed through front
mask
IP 20 rear of case
screw clamps
caption, custom logo or extra
lettering to special order
behind removable display lens;
on top side for meter DPA 08/10
isolated
power supply
case details
material of case
bezel
Short Form Data
4
5
6
7
8
9
input
power
supply
option
col
our
case
min/
max
ordering example
1
2
DPA
08
3
4 5 6 7 8
– 30 E 1 0 G –
Short Form Data
mA, mV
mA, V
Digital Panel Meters
for DC Current and
DC Voltage
(mA / V)
DPA 14
Input A
–
96 x 24 mm
–4 96 x 48 mm
–9 96 x 96 mm
Digits
30
31
40
41
50
Input
A
3 digits
3½ digits
4 digits
4½ digits
5 digits
Short Form Data
Digital Panel Meters
for DC Current and
DC Voltage
(mA / mV)
Display Span
–199 ...
999
–1999 ... +1999
–1999 ... 9999
–19999 ... +19999
–19999 ... 30000
DPA 14
Input D
–
96 x 24 mm
–4 96 x 48 mm
–9 96 x 96 mm
Digits
30
31
40
41
50
Input
D
Electrical Data
Electrical Data
measuring ranges
DC current
±20 mA
±200 mA
DC voltage
±2 V
±10 V
±40 V
±200 V
measuring principle
conversion rate
error, max.
auxiliary voltage
measuring ranges
DC current
±100 mA
±1 mA
DC voltage
±60 mV
±125 mV
±150 mV
±300 mV
measuring principle
conversion rate
error, max.
auxiliary voltage
power consumption
100 W
10 W
100 kW
560 kW
2.2 MW
12 MW
dual - slope integration
approx. 1 read per second
±0.01% of reading, ±1 count
DC 24 V (20 ... 30 V), electrically insulated or
AC 115/230 V, 50 Hz, electrically insulated
3 VA max.
digits
digits
digits
digits
digits
Display Span
–199 ...
999
–1999 ... +1999
–1999 ... 9999
–19999 ... +19999
–19999 ... 30000
DC Current 100 mA / 1 mA and
DC Voltage
60 / 125 / 150 / 300 mV
Auxiliary Voltage
1
24 V DC electr. insulated
2
115/230 V AC, 50 Hz electr. insulated
Options refer to last page
Display Color
R
red
G
green
DC Current 20 / 200 mA and
DC Voltage 2 / 10 / 40 / 200 V
Auxiliary Voltage
1
24 V DC electr. insulated
2
115/230 V AC, 50 Hz electr. insulated
Options refer to last page
Display Colour
R
red
G
green
input resistance
3
3½
4
4½
5
power consumption
input resistance
560 W
68 W
100 kW
220 kW
270 kW
560 kW
dual - slope integration
approx. 1 read per second
±0.01% of reading, ±1 count
DC 24 V (20 ... 30 V), electrically insulated or
AC 115/230 V, 50 Hz, electrically insulated
3 VA max.
DC
115 V
230 V
DC
AC
jumper when
measuring current
ordering example
230 V
plug - in terminal barrier strip
0V
plug - in terminal barrier strip
115 V
Connection
0V
Connection
24 V
display
7 - segment LED
digit height
14 mm
display span
see above
display colour
red or green
display
standard or reciprocal
overrange indication
center segments will flash
mean value set - up
adjustable from 1 up to 500 readings
round - off last digit
adjustable: 2nd, 5th, 10th steps, none
zero point, end value, polarity sign and decimal point programmable
0V
display
7 - segment LED
digit height
14 mm
display span
see above
display colour
red or green
display
standard or reciprocal
overrange indication
center segments will flash
mean value set - up
adjustable from 1 up to 500 readings
round - off last digit
adjustable: 2nd, 5th, 10th steps, none
zero point, end value, polarity sign and decimal point programmable
0V
Display
24 V
Display
AC
jumper when
measuring current
1
2
DPA
14
3
4 5 6 7 8
– 40 A 2 0 G – 4
9
–M
ordering example
1
2
DPA
14
3
4 5 6 7 8
– 41 D 1 0 R -
9
–M
Short Form Data
2x mA, V
DPA 14
Input B
–
96 x 24 mm
–4 96 x 48 mm
–9 96 x 96 mm
Short Form Data
A, V
Digital Panel Meters
for DC Current and
DC Voltage
(mA / V), 2 Inputs
Digits
30
40
50
Input
B
Digital Panel Meters
for AC Current and
AC Voltage
TRMS
Display Span
–199 ...
999
–1999 ... 9999
–19999 ... 30000
3 digits
4 digits
5 digits
2x DC Current 20 mA and
DC Voltage 10 V
Auxiliary Voltage
1
24 V DC electr. insulated
2
115/230 V AC, 50 Hz electr. insulated
Option
0
none
Display Colour
R
red
G
green
DPA 14
Input F
–
96 x 24 mm
–4 96 x 48 mm
–9 96 x 96 mm
Digits
30
40
50
Input
F
AC Current 3 / 6 A and
AC Voltage 2 / 100 / 250 / 500 V
Auxiliary Voltage
1
24 V DC electr. insulated
2
115/230 V AC, 50 Hz electr. insulated
Options refer to last page
Display Colour
R
red
G
green
Electrical Data
Electrical Data
measuring ranges
2x DC current
±20 mA
2x DC voltage
±10 V
change over
measuring principle
conversion rate
error, max.
auxiliary voltage
measuring ranges
AC current
3A
6A
AC voltage
2V
100 V
250 V
500 V
measuring principle
conversion rate
error, max.
auxiliary voltage
power consumption
input resistance
100 W
560 kW
24 V DC (opto - coupler)
dual - slope integration
approx. 1 read per second
±0.01% of reading, ±1 count
DC 24 V (20 ... 30 V), electrically insulated or
AC 115/230 V, 50 Hz, electrically insulated
3 VA max.
Display
power consumption
display
7 - segment LED
digit height
14 mm
display span
see above
display colour
red or green
display
standard or reciprocal
overrange indication
center segments will flash
wire -break detection
programmable for downscale detection
mean value set - up
adjustable from 1 up to 500 readings
round - off last digit
adjustable: 2nd, 5th, 10th steps, none
zero point, end value, polarity sign and decimal point programmable
Connection
Display Span
0 ... 999
0 ... 9999
0 ... 20000
3 digits
4 digits
5 digits
input resistance
20 mW
10 mW
10 kW
470 kW
1 MW
2.2 MW
dual - slope integration, true RMS
approx. 1 read per second
±0.2% of reading, ±1 count
DC 24 V (20 ... 30 V), electrically insulated or
AC 115/230 V, 50 Hz, electrically insulated
3 VA max.
Display
display
7 - segment LED
digit height
14 mm
display span
see above
display colour
red or green
display
standard or reciprocal
overrange indication
center segments will flash
mean value set - up
adjustable from 1 up to 500 readings
round - off last digit
adjustable: 2nd, 5th, 10th steps, none
zero point, end value and decimal point programmable
Connection
plug - in terminal barrier strip
Input 2
230 V
0V
DC
115 V
0V
AC
24 V
230 V
0V
DC
115 V
24 V
change over
Input 1
0V
plug - in terminal barrier strip
AC
jumper when
measuring current
ordering example
1
2
DPA
14
3
4 5 6 7 8
– 50 B 2 0 G – 4
ordering example
1
2
DPA
14
3
4 5 6 7 8
– 30 F 2 0 G – 4
9
–
Short Form Data
Short Form Data
mA, V
Digital Panel Meters
for Temperature
Pt 100 / Pt 1000
Display Span
–100.0 ... +199.9 °C
–200 ... +800 °C
–148 ... +392 °F
–328 ... +1472 °F
Input
M
for RTD Pt 100 Input
L
for RTD Pt 1000 Input
Auxiliary Voltage
1
24 V DC electr. insulated
2
115/230 V AC, 50 Hz electr. insulated
Options refer to last page
Display Colour
R
red
G
green
DPA 20
DPA 25
Input A
–
144 x 48 mm
–4 144 x 72 mm
Digits
30
31
40
41
50
Input
A
DC Current 20 / 200 mA and
DC Voltage 2 / 10 / 40 / 200 V
Auxiliary Voltage
1
24 V DC electr. insulated
2
115/230 V AC, 50 Hz electr. insulated
Options refer to last page
Display Colour
R
red
G
green
Electrical Data
Electrical Data
measuring ranges
measuring ranges
DC current
±20 mA
±200 mA
DC voltage
±2 V
±10 V
±40 V
±200 V
measuring principle
conversion rate
error, max.
auxiliary voltage
power consumption
Display
display
digit height
display span
display colour
overrange indication
zero adjustment
7 - segment LED
14 mm
see above
red or green
center segments will flash
possible
power consumption
7 - segment LED
DPA 20
20 mm
DPA 25
25 mm
display span
see above
display colour
red or green
overrange indication
center segments will flash
mean value set - up
adjustable from 1 up to 500 readings
zero point, end value, polarity sign and decimal point programmable
230 V
0V
115 V
0V
24 V
100 kW
560 kW
2.2 MW
10 MW
dual - slope integration
approx. 1 read per second
±0.01% of reading, ±1 count
DC 24 V (20 ... 30 V), electr. insulated or
AC 115/230 V, 50 Hz, electr. insulated
5 VA max.
display
digit height
plug - in terminal barrier strip
DC
100 W
10 W
Display
Connection
Connection
AC
plug - in terminal barrier strip
0V
Pt 100 / Pt 1000
4 - wire connection
input resistance
Pt 100 / Pt 1000
3 - wire connection
DC
AC
jumper when
measuring current
Pt 100 / Pt 1000
2 -wire connection
ordering example
1
2
DPA
14
3
4 5 6 7 8
– 31 M 2 0 G – 4
9
–M
ordering example
1
2
DPA
20
3
4 5 6 7 8
– 31 A 1 0 G –
230 V
measuring principle
conversion rate
error, max.
auxiliary voltage
for RTD input
resolution
–100.0...+199.9°C/–148...+392°F ±0.1 K
–200...+800°C/–328...+1472°F
±1 K
dual - slope integration
approx. 1 read per second
±0.01% of reading, ±1 count
DC 24 V (20 ... 30 V), electr. insulated or
AC 115/230 V, 50 Hz, electr. insulated
3 VA max.
Display Span
–199 ...
999
–1999 ... +1999
–1999 ... 9999
–19999 ... +19999
–19999 ... 30000
3 digits
3½ digits
4 digits
4½ digits *
5 digits *
* DPA 20 only
0V
115 V
Input M
Input L
–
96 x 24 mm
–4 96 x 48 mm
–9 96 x 96 mm
Digits
31
3½ digits
24 V
DPA 14
Digital Panel Meters
for DC Current and
DC Voltage
Short Form Data
Options
Digital Panel Meters
for AC Current /
AC Voltage
TRMS
Digital Panel Meters,
programmable
A, V
DPA 20
DPA 25
Input F
–
144 x 48 mm
–4 144 x 72 mm
Digits
Display Span
30
3 digits
0 ... 999
40
4 digits
0 ... 9999
50
5 digits *
0 ... 20000
* DPA 20 only
Input
F
AC Current 3 / 6 A and
AC Voltage 2 / 100 / 250 / 500 V
Auxiliary Voltage
1
24 V DC electr. insulated
2
115/230 V AC, 50 Hz electr. insulated
Options refer to right - side column
Display Colour
R
red
G
green
DPA 14
DPA 20
DPA 25
Input A/E
Input D
Input F
Input L
Input M
Electrical Data
measuring ranges
AC current
3A
6A
AC voltage
2V
100 V
250 V
500 V
measuring principle
conversion rate
error, max.
auxiliary voltage
power consumption
input resistance
Extras
20 mW
10 mW
10 kW
470 kW
1 MW
2.2 MW
dual - slope integration, true RMS
approx. 1 read per second
±0.2% of reading, ±1 count
DC 24 V (20 ... 30 V), electrically insulated or
AC 115/230 V, 50 Hz, electrically insulated
5 VA max.
Display
7 - segment LED
DPA 20
20 mm
DPA 25
25 mm
display span
see above
display colour
red or green
overrange indication
center segments will flash
mean value set - up
adjustable from 1 up to 500 readings
zero point, end value and decimal point programmable
analog output
0 ... 10 V /
0/4 ... 20 mA
6A
0V
115 V
230 V
3A
0V
24 V
plug - in terminal barrier strip
DC
AC
250V
baud rate (150 ... 19200 baud), parity,
data bits, device address, write direction
and protocol adjustable
serial output
RS232
1
2
DPA
25
3
1 control
output o.c.
2 control
outputs o.c.
input blanked
control/ display
hold
setpoints
DPA 14
10 V /
20 mA
0V
jumper when
measuring current
ordering example
S option
RS232 serial output
2 relays N.O. /N.C.
max. 200 V, max. 0.5 A, max. 10 W
max. 250 V, max. 4 A, max. 200 W
approx. 1 ms
adjustable, internal adjustment
small, single - digit 7 - segment LED
Connection
Connection
500V
R option
setpoints
capacity DPA 14
limit
DPA 20/25
response time
hysteresis
status indication
Further options
caption
please specify
extra lettering
please specify
logo on the lens
please specify
factory calibration
please specify
special measuring range on request
non - standard auxiliary voltage on request
enclosure code
IP 54 to IP 65 front
using rubber cover attachment 96x48
display
digit height
2V
100V
Options
0
none
1*
Analog Output 0 ... 10 V
(load ≥1000 W)
2*
Analog Output 0/4 ... 20 mA
(load ≤500 W)
4*
like 1, galvanically isolated
7*
1 Control Output, open collector
max. 30 V, max. 0.5 A, max. 12 W
8*
2 Control Outputs, open collector
max. 30 V, max. 0.5 A, max. 12 W
D*
Input Blanked Control
active high, 24 V
G*
Segment Test Input
H*
Display Hold Input
active high, 24 V
R
2 Setpoints
Relay N.O. /N.C.
S*
Rerial Output RS232
Additional Option
–M * Min./Max. Value Storage
* DPA 14 only (not with B input)
setpoints
DPA 20/25
4 5 6 7 8
– 40 F 2 0 G –4
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Telefax: 0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
– specifications subject to change without notice; date of issue 03/07 –
Product Guide
Digital Panel Meters,
Standard
96x48
96x96
✓
✓
✓
✓
✓
✓
✓
✓
for DC Current and Voltage
for AC Current and Voltage
for Standard Signals
96x48 mm or 96x96 mm Bezel
14 mm Digit Height
Red or Green LEDs
3½ Digits
Auxiliary Supply AC 115/230 V, 50 Hz
DM Series
711.U.001.03
General Data
Survey
Digital Panel Meters,
Standard
Digital Panel Meters,
Standard
Application
DMA 14
for standard applications in switchboard -, control equipment -,
process control -, and machine tool industry.
for mounting in
switchboards,
machine tools or
mosaic grid panels (optional)
for measuring
DC current or DC voltage,
AC current or AC voltage
connection
direct or for use on transducer, shunt,
transformer
Input A
Input D
Input E
Input C
Input F
–
96 x 48 mm
–9 96 x 96 mm
Mechanical Data
case details
material of case
bezel
terminals
enclosure code
panel fixing
mask lettering
operating elements
climatic suitability
operating
temperature range
storage
temperature range
relative humidity
dimensions (in mm)
bezel
panel cutout
panel thickness
mounting depth
bezel height
rectangular format
glass - fibre reinforced thermoplastics, black
black, mat or optionally
gray varnished (similar to RAL 7037) or
siliceous gray (similar to RAL 7032)
screw terminal barrier strip
IP 50 front
IP 20 rear of case
screw clamps
caption, custom logo or extra
lettering to special order
behind removable display lens;
for model DMA 14 – 31A20R – S3 on rear side
Input
A – S1 DC Current 20 / 200 mA and
DC Voltage 2 / 10 / 40 / 200 V
D – S1 DC Voltage
60 / 125 / 150 / 300 mV
E – S1 DC Current 20 / 200 mA and
DC Voltage 2 / 10 / 40 / 200 V
zero shifting
C – S1 DC Current 0/4 ... 20 mA and
DC Voltage 0 ... 10 V
F – S AC Current 3 / 6 A and
AC Voltage
2 / 100 / 250 / 500 V
A – S3 DC Voltage 0 ... 20 V unipolar
Display Colour
R
red
G
green
Case
–
–9
96 mm x 48 mm
96 mm x 96 mm
I
n
p
u
t
C
o
l
o
u
r
C
a
s
e
Measuring Ranges
Type
mA / V =
DMA
14
– 31
A
R
20 G – S1
–
–9
–20oC ... +23oC ... +70oC
mV =
DMA
14
– 31
D
R
20 G – S1
–
–9
75% annual average, non - condensing
DMA 14
DMA 14 ... –9
96 x 48
96 x 96
92+0.8 x 92+0.8
92+0.8 x 45.0+0.6
45
45
120
120
8
8
mA / V = zero point
DMA
14
– 31
E
R
20 G – S1
–
–9
standard signal
DMA
14
– 31
C
R
20 G – S1
–
–9
A/V~
DMA
14
– 31
F
R
20 G – S
–
–9
20 V=
DMA
14
– 31
A
R
20 G – S3
–
–9
0oC ... +23oC ... +50oC
Options
caption
extra lettering
logo on the lens
special measuring
range
factory calibration
non - standard auxiliary
voltage
enclosure code
please specify
please specify
please specify
on request
mounting in mosaic
grid panels
please specify type / manufacturer
of mosaic grid when ordering.
additional options
additional data
on request
on request
please specify
on request
IP 54 up to IP 65 meter front
using rubber cover attachment 96x48
mA, V
mV
Short Form Data
Short Form Data
mA, V
Digital Panel Meters
for DC Current and
DC Voltage
(mA, V / mV)
Digital Panel Meters
for DC Current and
DC Voltage
Zero Point Shifting
Digits
31
3½ digits
Input
A
DC Current 20 / 200 mA and
DC Voltage 2 / 10 / 40 / 200 V
D
DC Voltage
60 / 125 / 150 / 300 mV
DMA 14
Input A
Input D
–
96 x 48 mm
–9 96 x 96 mm
DMA 14
Input E
Input C
–
96 x 48 mm
–9 96 x 96 mm
Auxiliary Voltage
20
115/230 V AC, 50 Hz electr. insulated
Auxiliary Voltage
20
115/230 V AC, 50 Hz electr. insulated
Display Colour, Case
please refer to Survey
Display Colour, Case
please refer to Survey
Variant
– S1
adjustment by potentiometer and
jumper behind display lens.
Variant
– S1
adjustment by potentiometer and
jumper behind display lens.
Electrical Data
measuring ranges
A input
DC current
Electrical Data
input
resistance
110 W
10 W
±20 mA
±200 mA
DC Voltage
±2 V
±10 V
±40 V
±200 V
measuring principle
conversion rate
error, max.
auxiliary voltage
power consumption
Digits
31
3½ digits
Input
E
DC Current 20 / 200 mA and
DC Voltage 2 / 10 / 40 / 200 V
C
DC Current 0/4 ... 20 mA and
DC Voltage 0 ... 10 V
D input
DC voltage
input
resistance
60 kW
±60 mV
60 kW
270 kW
±125 mV
140 kW
1 MW
±150 mV
160 kW
4.7 MW
±300 mV
330 kW
dual - slope integration
approx. 1 read per second
±0.1% of reading, ±1 count
AC 115 / 230 V, 50 Hz, electrically insulated
4 VA max.
Display
measuring ranges
E input
DC current
±20 mA
±200 mA
DC voltage
±2 V
±10 V
±40 V
±200 V
zero point
measuring principle
conversion rate
error, max.
auxiliary voltage
power consumption
input
resistance
110 W
10 W
C input
DC current
0/4 ... 20 mA
input
resistance
100 W
DC voltage
160 kW
0 ... 10 V
160 kW
740 kW
2.2 MW
12 MW
for C input programmable from 2 up to 6 mA
dual - slope integration
approx. 1 read per second
±0.1% of reading, ±1 count
AC 115/230 V, 50 Hz, electrically insulated
4 VA max.
Display
plug - in terminal barrier strip
input A
input E
0V
230 V
plug - in terminal barrier strip
0V
Connection
115 V
Connection
AC
AC
jumper when
measuring current
jumper when
measuring current
input D
jumper when
measuring
current
1: 0 ... 20 mA
2: 4 ... 20 mA
0V
230 V
0V
115 V
input C
AC
AC
ordering example
230 V
end value
polarity sign
decimal point
blanked display
overrange indication
7 - segment LED, 14 mm
–1999 ... +1999 counts
for input E programmable
from –500 up to +500
programmable
only – or ± programmable
programmable
programmable for 1st digit (right - side)
polarity display will flash, remaining
numeric segments blanked
DMA
14
– 31
A
20 R – S1
230 V
end value
polarity sign
decimal point
blanked display
overrange indication
display
display span
zero point
115 V
7 - segment LED, 14 mm
–1999 ... +1999 counts
for input A fixed
for input D programmable from –50 up to +50
programmable
only – or ± programmable
programmable
programmable for 1st digit (right - side)
polarity display will flash, remaining
numeric segments blanked
115 V
display
display span
zero point
–
ordering example
DMA
14
– 31
C
20 R – S1
–9
Short Form Data
Short Form Data
A, V
V
Digital Panel Meters
for AC Current/Voltage,
TRMS
DMA 14
Input F
–
96 x 48 mm
–9 96 x 96 mm
Digits
31
3½ digits
Input
F
AC current 3 / 6 A and
AC voltage 2 / 100 / 250 / 500 V
Digital Panel Meters
for DC Voltage (20V)
Rear Potentiometer
DMA 14
Input A
–
96 x 48 mm
–9 96 x 96 mm
Auxiliary Voltage
20
230 V AC, 50 Hz
electrically insulated
Digits
31
3½ digits
Input
A
DC Voltage 0 ... 20 V unipolar
Auxiliary Voltage
20
230 V AC, 50 Hz
electrically insulated
Display Colour, Case
please refer to Survey
Display Colour, Case
please refer to Survey
Variant
–S
adjustment by potentiometer and
jumper behind display lens.
Variant
– S3
adjustment by potentiometer
on meter rear side
Electrical Data
Electrical Data
measuring ranges
AC current
3A
6A
AC voltage
2V
100 V
250 V
500 V
measuring principle
conversion rate
error, max.
auxiliary voltage
power consumption
measuring ranges
DC voltage
20 V
measuring principle
conversion rate
error, max.
auxiliary voltage
power consumption
input resistance
20 mW
10 mW
5 kW
270 kW
680 kW
1.3 MW
dual - slope integration, true RMS
approx. 1 read per second
±0.2% of reading, ±1 count
AC 115/230 V, 50 Hz, electrically insulated
4 VA max.
Display
display
display span
zero point
end value
polarity sign
decimal point
blanked display
overrange indication
7 - segment LED, 14 mm
0 ... 1999 counts
programmable from 0 up to 20
programmable
none
programmable
programmable for 1st digit (right - side)
polarity display will flash, remaining
numeric segments blanked
input resistance
530 kW
dual - slope integration
approx. 1 read per second
±0.1% of reading, ±1 count
AC 230 V, 50 Hz, electrically insulated
4 VA max.
Display
display
display span
zero point
end value
polarity sign
decimal point
blanked display
overrange indication
7 - segment LED, 14 mm
0 ... 1999 counts
fixed
programmable
none
programmable
programmable for 1st digit (right - side)
polarity display will flash, remaining
numeric segments blanked
Connection
plug -in terminal barrier strip
plug - in terminal barrier strip
0V
230 V
Connection
115 V
230 V
0V
AC
AC
jumper when measuring current
ordering example
DMA
14
– 31 F 20 R – S
–
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Telefax: 0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
ordering example
DMA
14
– 31 A 20 R – S3
–9
– specifications subject to change without notice; date of issue 01/07 –
Product Guide
Meter Accessories
ASK
WSK
Window Type CT’s and
Wound Primary Type CT’s
KBU
Split Core Current
Transformers
RH
HH
Insolating Spacers
mV
Shunts
kV
Voltage Dividers
V
AU
Rotary Cam Switches
AR
BA
Cover Frames
Blind Covers
Q
Terminal Safety Protection
SUS
LED Type
Switch Position Indicators
PI
PIR
Electromechanical
Switch Position Indicators
SM
U
PLC Fault Annunciators
Quartz Panel Watches
NT
=
DIN Rail Power Supplies
900.U.001.07
Short Form Data
Short Form Data
Window Type and
Wound Primary Type
Current Transformers
Split Core Current
Transformers
ASK 421.4
ASK 31.3
ASK 41.4
ASK 51.4
ASK 561.4
ASK 81.4
ASK 101.4
WSK 30
WSK 40
KBU 23
KBU 58
KBU 812
KBU 816
Application
Application
Window type CT’s (ASK) are mounted on busbars and are suitable for
primary currents from 30 A to 2500 A.
Wound primary CT’s (WSK) have a primary winding for lower primary
currents from 1 A to 30 A.
The KBU split core current transformers can be attached subsequently
to live wires.
The integrated KBU locking system allows simple mounting of the CT
via snap - in. By pressing a button the CT can be removed easily.
General Technical Data
General Technical Data
standards
standards
DIN 42 600 - 2, DIN EN 60 715,
DIN EN 60044 - 1, VDE 0414 - 44 - 1, VBG 4
material of case
polycarbonate, flame retardant,
self–extinguishing
mounting
push - in fixing feet,
or optionally on ASK
busbar clamps or
ASK31.3/41.4, WSK30/40 clamping attachment to 35 mm DIN rail
terminals
secondary terminals nickel - plated, with plus/
minus combination screws M5x10, integrated terminal cover
Window Type CT’s
Primary Current Ratings
Width of CT
ASK 421.4
30 – 500 A
71 mm
ASK 31.3
50 – 750 A
61 mm
ASK 41.4
50 – 1,000 A
71 mm
ASK 51.4
100 – 1,250 A
86 mm
ASK 561.4
200 – 1,250 A
86 mm
ASK 81.4
400 – 2,000 A
120 mm
ASK 101.4
500 – 2,500 A
130 mm
Wound Primary CT’s Primary Current Ratings
Width of CT
WSK 30
1 – 20 A
61 mm
WSK 40
1 – 30 A
71 mm
rated primary
1; 2.5; 5; 10; 15; 20; 25; 30; 40; 50; 60; 75; 80;
current
100 A and any decimal multiple up to 2500 A
as well as 1200; 1250; 1600 and 1800 A
rated secondary current
1 A or 5 A
rated output
1; 1.25; 1.5; 2.5; 3.75; 5; 7.5; 10; 15; 30; 45 VA
frequency range
50 ... 60 Hz, 162/3 Hz or 400 Hz on request
accuracy
classes 0.5 or 1
Special CT’s
summation, saturation, protective or tube type
CT’s, special CT’s suitable for H.R.C. fuse carriers or secondary
switchable C.T’s;
calibratable or calibrated C.T’s, with accuracy
classes 0.2; 0.5 and 0.5s on request
additional data
prices
refer to Data Sheet No. 500.D.001.##
refer to Price Sheet No. 555 – 563.P.001.##
material of case
attachment
terminals
Primary Ratings
KBU 23
KBU 58
KBU 812
KBU 816
rated continuous
current
rated peak current
rated excess factor
DIN 42 600 - 2,
DIN EN 60044 - 1, VDE 0414 - 44 - 1, VBG 4
polycarbonate, flame retardant,
self - extinguishing according to UL 94 V - 0
snap - in mounting and clamp screws
nickel - plated brass secondary terminals,
each with two plus/minus combination screws
rated primary current IN
100; 150; 200; 250; 300; 400 A
250; 300; 400; 500; 600; 750; 800; 1000 A
250; 300; 400; 500; 600; 750; 800; 1000;
1200; 1250; 1500 A
1000; 1200; 1500; 1600; 2000; 2500; 3000;
4000; 5000 A
ID = 1,0 . IN
Ith = 60 . IN (max. 1 s)
FS 5: up to 1500 A primary rated current
FS 10: 1600 A and higher prim. rated current
rated output
rated frequency
accuracy
rated secondary current INs
1 A or 5 A
1; 1.25; 1.5; 2.5; 3.75; 5; 7.5; 10; 15; 30 VA
50 Hz
classes 0.5 or 1
additional data
prices
refer to Data Sheet No. 566.D.001.##
refer to Price Sheet No. 564 – 566.P.001.##
Secondary Ratings
Short Form Data
Short Form Data
Insolating Spacers
Shunts
Class 0.5
RH
HH
60 mV
100 mV
150 mV
300 mV
General Technical Data
General Technical Data
material
The shunts herein referred to are manufactured with an accuracy class
0.5 according to DIN 43 703 in current ratings from 1 A up to 15,000 A
having a voltage drop of 60 mV or 150 mV. On special order with a
voltage drop and/or a rated current other than standard.
Format Version A
with insulating base
up to 25 A / 60, 100, 150 or 300 mV
optionally up to 150 A / 60 mV
for screw mounting (max. M8) or
clamping to 35 mm DIN rail
without insulating base 30 ... 150 A
Format Version B
L - profile end blocks
Format Version C
T - profile end blocks
material
resistance bars
manganin
end blocks
format version A
high conductivity brass
format version B
high conductivity brass/solid copper
format version C
solid copper
connections
threaded bolts
dimensions
complying with DIN 43 703
Rated Current
1; 1.2; 1.5; 2; 2.5; 3; 4; 5; 6; 8 A
and any decimal multiple of these numbers
Rated Voltage Drop 60 mV; 100 mV; 150 mV or 300 mV
accuracy
class 0.5
polyester, fibre- glass inforced,
red, self - extinguishing, halogen- free
insert nut galvanized steel
fire resistance
according to UL 94 (class V–0)
rated voltage
up to 1000 V AC (when used in excess voltage
categories I to IV according to IEC 60038)
breakdown voltage
min. 10 kV/mm
creep voltage strength CTI 600 (according to EN 60112:2003–03)
peak voltage strength >12 kV (according to IEC 61180–1:1994–09)
operating temperature range –40 ... +160°C
... RH ...
... HH ...
Type
body
Order
No.
Dimensions [mm]
Thread
Break
force [N]
Pulling
force [N]
Maximum
torque [Nm]
Weight
[g]
Quantity per
package [Pcs]
Options
rated voltage drop
rated current
purpose built shunts
other than standard on request
other than standard on request
up to max. 20,000 A
class 0.2
suitable for shunts 30 ... 150 A / 60 mV
others on request
on request
Accessories
cover
for shunts with insulating base
accuracy
insulating base
additional data
prices
refer to Data Sheet
refer to Price Sheet No. 666.P.101.##
additional data
refer to Data Sheet No. 800.D.001.##
dimensions and weights
prices
refer to Price Sheet No. 800.D.001.##
Short Form Data
Short Form Data
Voltage Dividers
1,000 V ... 10,000 V
Rotary Cam Switches
for AC Voltage
and AC Current
V
0
V
3
V 13
V 30
V 32
AU 11
AU 21
AU 31
AU 41
General Technical Data
General Technical Data
External multipliers (voltage dividers) are used in connection with moving - coil instruments in the measurement of DC voltages in electrical
installations connected to the ground.
construction
thermoplastic case containing cast resin
insulated film resistors of accuracy class 0.5
to DIN.
panel fixing
by two screws M4
terminals
screws M3
weight approx.
0.2 kg
DC voltage
1,000 V
1,500 V
2,000 V
2,500 V
3,000 V
4,000 V
5,000 V
6,000 V
10,000 V
sensitivity
2 kW/V
for meter movement
25 V, 250 mA
accuracy
class 0.5
The rotary cam switches comply with VDE 0660 and VBG 4.
additional data
prices
suitable moving - coil
instruments
refer to Data Sheet No. 806.D.001.##
refer to Price Sheet No. 806.P.001.##
PSQ 48, PQ 72/96/144 RS
(M–Series, 90o – Dial)
refer to Data Sheet No. 010.D.101.##
Voltmeter Changeover Switches
Model V 0
for phase voltages to neutral in 3 - phase 4 - wire
systems, switching positions: 0 – L1N – L2N – L3N
Model V 3
for delta voltages in 3 - phase 3 - wire systems,
switching positions: 0 – L1L2 – L2L3 – L3L1
Model V 13
for delta voltages and 1 phase voltage to neutral
in 3 - phase 4 - wire systems, switching positions:
L3L1 – L2L3 – L1L2 – 0 – L1N
Model V 30
for delta voltages and phase voltages to neutral
in 3 - phase 4 - wire systems, switching positions:
L3L1 – L2L3 – L1L2 – 0 – L1N – L2N – L3N
Model V 32
for delta voltages in two 3 - phase 3 - wire systems,
switching positions:
L3L1 – L2L3 – L1L2 – 0 – L1L2 – L2L3 – L3L1
Ammeter Changeover Switches
Model AU 11 single - pole with off - position, 1 current transformer
circuit, switching positions: 0 – 1
Model AU 21 single - pole with off - position, 2 current transformer
circuits, switching positions: 1 – 0 – 2
Model AU 31 single - pole with off - position, 3 current transformer
circuits, switching positions: 0 – 1 – 2 – 3
Model AU 41 single - pole without off - position, 4 current transformer
circuits, switching positions: 1 – 2 – 3 – 4
construction
suitable for switchboard mounting
panel thickness
1 ... 5 mm
terminals
M3.5x6 screws and wire clamps
cross - section
2x 2.5 mm2
of connection
operating voltage
AC 660 V
continuous current
25 A acc. to VDE 0660 section 107
load switching capacity 25 A acc. to VDE 0660 section 1
frequency
up to 3 kHz
product classification C3 acc. to VDE 0660
additional data
prices
refer to Data Sheet No. 812.D.001.##
refer to Price Sheet No. 812.P.001.##
AR
AR
AR
AR
AR
AR
BA
BA
BA
BA
Short Form Data
Short Form Data
Cover Frames
Glass-Inserted
Blind Covers
for DIN Cutouts
Terminal
Safety Protection
48
72
96
144
96x48
144x72
48
72
96
96x24
Q 48
Q 72
Q 96
Q 144
Application
Application
The glass - inserted cover frames AR 48/72/96/144/96x48/144x72 and
the blind covers BA 48/72/96/96x24 for clamp - fixing are used to cover
standard DIN - cutouts in switchgear panels.
Protective sleeves or full - sized rear covers provide protection against
accidental contact of meter terminals. They meet the requirements of
VBG 4 / DIN 57 106 and are safe against backhanded and fingertip
contact.
Mechanical Data
Cover Frames, glass - inserted
construction
case suitable for mounting in switchboards
or mosaic grid panels, stackable
material of case
pressed steel
window
glass, non - glaring glass or frosted glass
colour of bezel
black or gray
panel fixing
clamp - mounting or screw clamps
panel thickness
1 ... 15 mm
mounting
stackable next to each other
mounting depth
27 mm
Blind Covers
material
colour
panel fixing
panel thickness
mounting
self - extinguishing thermoplastics
black
clamp fixing
1 ... 4 mm
stackable next to each other
additional data
prices
refer to Data Sheet No. 919.D.011.##
refer to Price Sheet No. 919.P.011.##
Mechanical Data
material
for use on K–Series
full - sized rear cover
protective sleeves
for use on M–Series
full - sized rear cover
protective sleeves
SW 6
protective sleeves
SW 10
rubber nozzle
additional data
prices
moulded thermoplastics
suitable for meter - sizes
48 x 48, 72 x 72, 96 x 96, 144 x 144
suitable for meter terminals using hexagon
studs with wire clamps E3 and M4 screws
suitable for the most applied meters of sizes
72 x 72, 96 x 96, 144 x 144
suitable for voltmeters and ammeters
up to 4 A rated current
suitable for ammeters
exceeding 4 A up to 30 A rated current
suitable for meters with
screw terminals M5 or M6
refer to Data Sheet No. 919.D.101.##
refer to Price Sheet No. 919.P.101.##
Short Form Data
Short Form Data
LED Type Switch
Position Indicators
Electromechanical
Switch Position
Indicators
SUS–01
SUS–02
SUS–95
SUS–99
PI
PI
PI
PI
PIR
PIR
PIR
PIR
24
25
29
36
24
25
29
36
Application
Application
Switch position indicators are used to indicate the switching state in
electrical installations.
The SUS 01/02/95/99 switch position indicators are equipped with
LED’s of different colors. They can be used for operation in mimic circuit
diagrams of switch gears as well control panels and switchboards.
round square
with LED test front dimensions
SUS–01
*
Q
T
25 mm
SUS–02
*
Q
T
20 mm
SUS–95
*
Q
–
39 mm
SUS–99
*
Q
–
32 mm
The PI/PIR 24/25/29/36 switch position indicators are equipped with a
rotary magnet system. They can be used for operation in mimic circuit
diagrams of switch gears as well control panels, switchboards and mosaic technology.
General Technical Data
PI/PIR ... – 6 – SPAW
with amber/white
display field
LED colors
luminous power
life cycle
operating voltage
rated current
terminals
red, green, blue, white, yellow, orange
depending on LED’s used
min. 100,000 hours of operation
12 V AC, DC up to 230V AC, DC
max. 20 mA
SUS
SUS ... T
max. 1,5 mm2 connector blades
2,8x0,8x7,0
enclosure code
SUS
SUS ... T
case
IP 65
IP 65
terminals IP 20
IP 00
operating temperature –25 ... +60° C
panel cutout
SUS–01/95/99
SUS–02
ø 22 mm
ø 16 mm
cutout distance min.
SUS–01
SUS–02
SUS–95
SUS–95
30 mm
25 mm
40 mm
33 mm
panel thickness
max. 12 mm
additional data
prices
refer to Data Sheet
refer to Price Sheet No. 652.P.101.##
PI/PIR ... – 1
PI/PIR ... – 2
PI/PIR 29 – 1
PI/PIR ... – 3
PI/PIR ... – 4
PI/PIR ... – 5
PI/PIR ... – 7 – SPRG
with red/green
display field
General Technical Data
case format
round thermoplastic case
with round or square front - bezel, suitable for mounting in switchboards
(PI/PIR 25/29/36) or mosaic grid panels (PI/PIR 24)
material of case
polycarbonate UL 94 VO
position of use
any position permissible
terminals
screw terminals up to 1.5 mm2
with safety touch protection
enclosure code
IP 54
dimensions
(in mm)
front - bezel
case
mounting depth
panel cutout
panel thickness
operation voltage
PI 24
PI 25
PI 29
PI 36
PIR 24
PIR 25
PIR 29
PIR 36
V 24
V 25
Ø 29
V 36
Ø 21.8 mm
94 mm
ø22+0.5 mm
max. 12 mm
DC voltage (PI type) or
AC voltage (PIR type)
in the range of 24 V to 230 V
frequency range with AC voltage
40 Hz ... 10 kHz
permissible voltage variation
±20%
additional data
prices
refer to Data Sheet No. 644.D.101.##
refer to Price Sheet No. 664.P.101.##
Short Form Data
Short Form Data
PLC Fault Annunciators with Slide-In Dial
Quartz Panel Watches
SM 96/8
SM 96/16
U
U
U
U
48
72
96
144
Application
Application
Using the PLC fault annunciators messages or faults can be indicated.
They can be operated with 24V/7mA directly by the PLC and also without PLC. Other rates are possible on request.
SM 96/8
8 LED’s with caption fields
SM 96/16
16 LED’s with caption fields
The U 48/72/96/144 quartz panel watches are equipped with a precise
German clockwork.
A stepper motor and a high accurate quartz guarantee for a long life
cycle and operation stability.
During power failure, an alkaliline battery drives the watch for approx.
1/2 year.
The U 144, U 96, U 72 watches can be optionally supplied with radiocontrolled clockworks.
General Technical Data
indication
3–color LED’s (red, green, orange)
can be operated individually by terminals
LED diameter
5 mm
operation
24 V DC
other ratings see Options
All inputs have common ground (minus).
The LED’s can be tested via separate terminals rear accessible.
front dimensions
96mm x 96mm to DIN 43700
dimensions
96 mm x 96 mm
weight
520 g
Options
– common alarm for all messages
– common alarm only for red LEDs
– common alarm with or without latch
– flashing red LED’s
– operation with 24 V AC, other ratings on request
– power supply and operation with 24 V AC (with common alarm)
All options can be combined with each other.
additional data
prices
refer to Data Sheet
refer to Price Sheet No. 653.P.101.##
General Technical Data
stepper motor
frequency
quartz frequency
operation stability
time setting
power supply
battery
average power
consumption
operation indication
1 Hz
32.768 kHz
lower or less 1s/d to DIN 8325
via rear accessible knob
10 ... 30V DC
AA type IEC LR6 1,5V
1 mA at 24 V
green LED
dimensions (in mm)
front bezel
case
panel cutout
panel thickness
position of use
operation temperature
range
storage temperature
range
U 48
U 72
48
72
42.5
66
45+0,6
68+0,7
5 mm
any
–5°C ... +50°C
U 96
96
90
92+0,8
U 144
144
136
138+1
additional data
prices
refer to Data Sheet
refer to Price Sheet No. 653.P.101.##
–20°C ... +70°C
=
Short Form Data
DIN Rail
Power Supplies
NT 22.5
Application
Electrical Data
The NT 22.5 power supply can be used for multiple applications and
can be supplied with various output voltages (5 V, 12 V, 15 V, 24 V). It is
overload - and overtemperature - proof and has a function indication.
A green LED lights up if the output voltage is available.
The power supplies comply with safety requirements and are tested for
interference immunity.
Input
input voltage
frequency
General Technical Data
case details
projecting case clamping to DIN mounting rail
(to DIN EN 50 022 – 35)
material of case
ABS/PC black
self - extinguishing to UL rating 94 V–0
terminals
screw - terminals
wire cross - section
input 2.5 mm2 max. flex wire
output 2.5 mm2 max. flex wire or
1.5 mm2 max. solid wire
enclosure code
IP 40 case
IP 20 terminals to EN 60529
dielectric test
4 kV 50 Hz input to output
isolation
100 MW / 500 V DC
rated isolation voltage 600 V
class of protection
II to DIN EN 60601–1/ UL 60601–1
dimensions WxHxD
22.5 mm x 84 mm x 113 mm
weight
approx. 0.12 kg (10 W)
approx. 0.10 kg (5 W)
indication
green LED
lights if power is available
additional data
prices
refer to Data Sheet No. 067.##
refer to Price Sheet No. 067.##
WEIGEL - MESSGERÄTE GmbH
P.O.B.Ă720ā154 S DĆ90241 Nürnberg S Telephone: 0911ā/ā4ā23ā47Ćā0
Erlenstraße 14 S DĆ90441 Nürnberg S Fax:
0911ā/ā4ā23ā47Ćā39
Internet:
http://www.weigel-messgeraete.de
e-mail:
vertrieb@weigel-messgeraete.de
Output
output power/type
voltage
85 ... 264 V AC or 120 ... 370 V DC
47 ... 440 Hz
5 W or 10 W
available fixed voltages
5 V DC
12 V DC
15 V DC
80 mV
150 mV
150 mV
2.0%
1.0%
1.0%
residual ripple
voltage accuracy
overload protection
switch - off at >105% of rated power and
automatic switch - on after temperature drop
overvoltage protection
switch - off if output voltage is > 135% of UN
overtemperature protection
switch - off at Tj > 140°C and automatic
switch - on after temperature drop
24 V DC
240 mV
1.0%
Environmental
climatic suitability
operating
temperature range
storage
temperature range
Vibration
EMC emission
EMC immunity
to VDE/VDI 3540 sheet 2
–20 ... +70°C
–40 ... +85°C
10 ... 500 Hz, 2G 10 min./ 1 cycle,
period 60 min. to all 3 axis
to EN 55011 (CISPR11),
EN 55022 (CISPR22), class B
to EN 61000–4–2,3,4,5,6,8,11;
EN 50204, EN 55024, EN 60601–1–2
and EN 61204–3, crit. A
– specifications subject to change without notice; date of issue 03/10 –