PM180 Manual - Calex Electronics

PM180 Manual - Calex Electronics
PM180
6-Channel Modbus Pyrometer Hub
Operator’s Guide
The PM180 is a temperature indicator, data logger, alarm unit and configuration tool for Calex
infrared temperature sensors. It is compatible with all models in the PyroBus and PyroMiniBus
series, as well as all PyroMini and FibreMini models with Modbus output.
The PM180 functions as the Modbus Master on an RS485 network of up to 6 temperature sensors,
and can itself be connected as a slave device to another RS485 network via a second, isolated
Modbus interface. This allows multiple PM180 units to be multi-dropped to create a large network
of sensors and displays.
Optional alarm relay modules allow the PM180 to be connected to alarm equipment such as
sounders and beacons, and optional analogue output modules allow it to be connected to nonModbus instrumentation.
All the configurable parameters for the hub, the connected sensors and the optional output
modules are adjustable via the PM180’s built-in resistive touch screen interface, which can be
operated even with gloves on.
With an optional MicroSD Card inserted, the PM180 functions as a fully-configurable data logger.
SPECIFICATIONS
Display
2.83” (72 mm) resistive touch TFT, 320 x 240 pixels, backlit
Supply Voltage
10 to 30 V DC
Maximum Current Draw
100 mA
Ambient Temperature Range
0°C to 60°C
Relative Humidity
Maximum 95%, non-condensing
Configurable Parameters (global)
Temperature units, date and time, data logging, graph channels, alarm logging
Configurable Parameters (per channel) Signal processing, emissivity setting, reflected energy compensation, alarms, Modbus address
Alarm Configuration
12 alarms (2 per sensor) with adjustable level, individually configurable as HI or LO.
Temperature Units
°C or °F selectable
Temperature Resolution
0.1° below 1000°; 1° above 1000°
Signal Processing
Average, peak hold, valley hold, minimum, maximum
Display Sample Period
120 ms per device (720 ms in total for 6 devices)
Compatible Sensor Types
PyroBus (all models),
PyroMiniBus (all models),
PyroMini (-BB and -BRT models),
FibreMini (–BRT models)
Compatible Output Module Types
ICP DAS M-7061 12-channel relay output module
ICP DAS M-7061D As above, with LED indicators
ICP DAS M-7024 4-channel analogue output module, selectable V/mA
MECHANICAL
Dimensions
Weight
2
98(w) x 64(h) x 36(d) mm excluding cable glands
280 g
DATA LOGGING
With an optional MicroSD Card installed in the slot inside the PM180, data logging may be
manually started and stopped via a button on the temperature display screen, or scheduled to
begin at a pre-determined time via the Settings menu.
If the PM180 is connected via the slave interface to another Modbus network, logging may also be
started remotely by the Modbus Master on that network.
DATA LOGGING SPECIFICATIONS
Logging Interval
MicroSD Card
Internal Clock Battery
Variables Logged
File Format
Configurable Parameters
1 to 86,400 seconds (1 day)
Max. capacity: 32 GB (not included)
1 x BR 1225 3V (not included)
Target temperature, sensing head temperature, alarm events
.csv (can be imported to Excel)
Sample period, number of samples, scheduled start date and time
DIMENSIONS (MM)
98
86
64
Touch screen
72 mm (2.83")
20
Cable glands:
17 mm AF
Mounting
holes: use
M4 CSK
screws
36
48
36
18
3
PM180
Modbus
Slave
Modbus
Master
PWR+ PWR- RS+ RS-
0V* RS+ RS-
Removable
screw
terminals
RS485 +
RS485 Signal Ground*
* If using the Modbus
Slave interface, the 0V
terminal should be
connected to the signal
ground of this network's
Modbus Master (e.g. a
PLC or SCADA system)
RS485 +
RS485 +10 to 28 V DC
0V
Ensure PWR- of all sensors and GND of all output
modules on this network is connected to PWR- of
the PM180 Modbus Master interface
The PM180 has removable screw terminal blocks for the Modbus Slave and Modbus Master
interfaces.
− Connect the Master interface to the sensors and output modules. Be sure to check the power
supply requirements of each device before applying power.
− Optionally, connect the Slave interface to another Modbus network with its own Modbus
Master such as a PLC or SCADA system.
Isolation is provided between the Slave and Master interfaces.
MICROSD CARD AND BATTERY
The MicroSD Card and battery slots are located on the touch screen circuit board. Unscrew the lid
of the PM180 to access them.
The battery is optional. With a battery fitted, the internal clock will continue to run when the power
is off. Without a battery, the unit will request the date and time each time the power is cycled.
All other settings are stored in the unit’s permanent memory and will be preserved when it is
switched off, regardless of whether a battery is fitted.
4
PyroMiniBus
sensor
PyroBus
sensor
PyroMini or FibreMini
sensor
+24 V DC
+12 V DC
PASSWORD
The default password is 1234. The password may be changed via the touch screen interface.
USING THE TOUCH SCREEN INTERFACE
Visit www.calex.co.uk/PM180 for help on how to use the touch screen interface, including a
diagram showing how to navigate to each setting.
5
MODBUS OVER SERIAL LINE (RS485)
INTERFACE
Baud rate9600
Format
8 data, No parity, 1 stop bit
Reply delay (ms)20
SUPPORTED FUNCTIONS
Read register
0x03, 0x04
Write single register0x06
Write multiple register0x10
The list below includes all available addresses:
R = Read, W = Write
Address
0x0000
0x0001
L e n g t h Description
(words)
R/W
1
MODBUS slave address
R/W
7
Sensor identification string in ASCII
R
"PM180vx.xx" where x.xx is the firmware version
0x0008
2
Serial number
R
0x000A
2
Sample Period (1 to 86400)
R/W
0x000C
2
Number of Samples (0 to 86400)
R/W
0x000E
1
Data acquisition enabled (0 for disabled, 1 for enabled)
R/W
0x000F
1
Data acquisition start time: Hours (0 to 23)
R/W
0x0010
1
Data acquisition start time: Minutes (0 to 59)
R/W
0x0011
1
Data acquisition start time: Seconds (0 to 59)
R/W
0x0012
1
Data acquisition start time: Day (1 to 31)
R/W
0x0013
1
Data acquisition start time: Month (1 to 12)
R/W
0x0014
1
Data acquisition start time: Year (2012 to 2105)
R/W
0x0015
1
Alarm log settings
R/W
Bit 0 - Log trigger time
Bit 1 - Log while triggered
Bit 2 - Log acknowledge time
Bit 3 - Log reset time
0x0020
1
Remote request - Start logging
R/W
Write 1 - request start of data logging
Read 1 - request pending, Read 0 - no request pending
0x0021
1
Remote request - Stop logging
R/W
Write 1 - request termination of data logging
Read 1 - request pending, Read 0 - no request pending
0x0022
1
Remote request - Acknowledge alarms
R/W
Write 1 - request acknowledgement of alarms
Read 1 - request pending, Read 0 - no request pending
0x0023
1
Remote request - Reset alarms
Write 1 - request reset of alarms
Read 1 - request pending, Read 0 - no request pending
6
R/W
SENSOR SETTINGS ADDRESS SPACE
The settings of attached sensors can be read by adding the following offsets to the addresses specified by the sensor manufacturer:
Sensor index
Address offset
0
0x1000
1
0x1100
2
0x1200
3
0x1300
4
0x1400
5
0x1500
6
0x1600
See sensor manual for further details.
Notes:
1. For further information please refer to http://www.modbus.org/specs.php
2. Use address 255 to communicate with any connected unit (only one sensor connected)
3. Use address 0 to broadcast to all connected units (no response expected)
GUARANTEE
Calex guarantees each instrument it manufactures to be free from defect in material and workmanship
under normal use and service for the period of one year from the date of purchase. This guarantee
extends only to the original buyer according to Calex Terms and Conditions of Sale.
7
Issue E - October 15
Calex Electronics Limited
PO Box 2, Leighton Buzzard, Bedfordshire, England LU7 4AZ
Tel: +44 (0)1525 373178 Fax: +44 (0)1525 851319
Email: mail@calex.co.uk Online: http://www.calex.co.uk
8