Sample Rack Usage

Sample Rack Usage
AU480 Sample Rack Usage
AU480
Racks
Racks are identified by a rack barcode label. The barcode label identifies
the type of sample (serum, urine, other-1 and other-2) designated for that
rack. Sample racks are configured in the AU software.
Racks are available in multiple colors to assist the operator in determining
rack function. Rack function is identified by the system by the placement
of magnets on the bottom of the rack.
Reagent Blanks (Blue)
•
Use any validated container to
place DI Water in position 1
Calibration (Yellow)
•
Multiple racks may be required for
multiple chemistries.
•
Use any validated container for
calibrators
Quality Control (Green)
•
Multiple racks may be required.
•
Use any validated container for
QC material
Note: Calibrators and QC may be barcoded or assigned to a rack and position.
AU480 Sample Rack Usage and Validated Sample containers
Version A, 072211
Page 1 of 3
Brea Training Center
Patient Samples
•
•
•
AU480
Racks
(Cont’d)
Patient samples can be placed in white, red or orange racks
Barcode Mode: Barcoded samples may be placed in any position
Sequential Mode: Samples without barcodes must be placed in
sequential order. Do not leave empty spaces in racks when using non
barcoded samples
Patient Samples (White)
•
Each sample type must be placed in
racks designated for that sample
type
− Routine Serum
− Routine Urine
− Routine Other 1
− Routine Other 2
•
Any validated container can be used
within a rack
•
Used with the Automatic Repeat
feature
Extra (Red)
•
Extra Samples: used to designate a
different sample category (user
defined)
•
Each sample type must be placed in
racks designated for that sample
type
Manual Repeats (Orange)
AU480 Sample Rack Usage and Validated Sample containers
Version A, 072211
•
If using the Manual Repeat feature,
you must place the sample in the
rack designated for that sample type
•
Any validated sample container may
be used in an orange rack
Page 2 of 3
Brea Training Center
AU480
Validated
Sample
Containers
Container
Placement
Description
•
Primary
Tubes
Rack/
STAT
table
•
•
Outside diameter:
11.5-16 mm
Inside diameter:
9-15 mm
Height :
55-102 mm
Part Number
N/A
2.0 mL Sample Cup
(Hitachi)
MU853200
Sample Cups, Glass
ZM006200
2.0 mL Sample Cup
(Hitachi)
MU853200
Sample Cups, Glass
ZM006200
2.0 mL Sample Cup
(Hitachi)
Dead volume =65 uL
(Do not use as insert
on the STAT table)
MU853200
Centrifichem Cups
Dead volume = 50 uL,
Evergreen
Scienific
P/N 1270093-010
Rack
Sample
Cups
STAT
table
16 mm
primary
tube
Nested
Cups
12/13 mm
primary
tube
EZ NEST Sample
Cups
(fits 16 x 100/
16 x 75 mm tubes)
EZ NEST Sample
Cups
( fits 13 x 100/
13 x 75 mm tubes)
Dead volume =50 uL
AU480 Sample Rack Usage and Validated Sample containers
Version A, 072211
1270016000
1270013000
Page 3 of 3
Brea Training Center
Photo
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising