PHABRIX SxA Operation Manual

Add to my manuals
37 Pages

advertisement

PHABRIX SxA Operation Manual | Manualzz

PHABRIX SxA

Operation Manual

RELEASE INFORMATION

Software Release: 0.9.2414

Date 8 December 2008

DISCLAIMER

Copyright © Phabrix Ltd. All rights reserved. Software products licensed are owned by Phabrix Ltd and are protected by international treaty provisions and national copyright laws.

No part of this document may be reproduced, stored in a retrieval system or transmitted in any form or by any means, electronic, mechanical, recorded or otherwise without the prior written consent of Phabrix Ltd.

The information in this document has been produced by Phabrix Ltd with care and is believed to be accurate. Phabrix Ltd does not assume responsibility for loss or damage resulting from errors, omissions or inaccuracies herein. This document is subject to change and revisions may be made and issued to include such changes.

PHABRIX LTD.

Blindmans Gate Cottage, Woolton Hill, Newbury, RG20 9XB, United Kingdom

Tel: +44(0)1635 255494 Web: www.phabrix.com

INDEX

Contents Page

Release information............................................................ 2

Disclaimer ........................................................................... 2

Phabrix Ltd. ......................................................................... 2

Index ................................................................................... 3

Getting Started with Phabrix SX ......................................... 6

Package Contents ........................................................... 6

General Safety ................................................................ 6

Avoiding Personal Injury ............................................. 6

Installation Environment ............................................ 6

Maintenance ............................................................... 7

Introduction ........................................................................ 8

Warranty ......................................................................... 8

Warranty exceptions .................................................. 8

Operating conditions .......................................................... 8

External Power Supply .................................................... 8

Maximum Input signals ................................................... 8

LCD Monitor .................................................................... 8

LCD Flicker ...................................................................... 9

Remote Network Operation ........................................... 9

Disposing of the unit ....................................................... 9

Specification ..................................................................... 10

LCD Display ................................................................... 10

SDI Inputs ...................................................................... 10

External Reference........................................................ 10

External Control ............................................................ 10

Headphone Output ....................................................... 11

Phabrix Sx Manual December 8, 2008

Contents Page

General .......................................................................... 11

Operation .......................................................................... 12

Turning the unit ON or OFF ........................................... 12

Unit Status Line ............................................................. 12

Using The Menus........................................................... 13

Main Menu ........................................................................ 13

Generator ...................................................................... 13

Analyser ........................................................................ 13

SIG info .......................................................................... 13

System ........................................................................... 14

Logging .......................................................................... 14

Speaker ......................................................................... 14

Signal Generator ............................................................... 14

Video Settings ............................................................... 14

GenLock Settings ........................................................... 15

Audio Settings ............................................................... 16

Master ....................................................................... 16

AES O/P ..................................................................... 16

Analyser ............................................................................ 17

Picture ........................................................................... 17

Audio Meters ................................................................ 18

Signal Data (engineering option) .................................. 19

Waveform Monitor ....................................................... 19

Waveform controls ................................................... 20

Setup Button/Dialog ................................................. 20

Vector Scope ................................................................. 21

Page 3

Signal Information ............................................................ 21

Video status .................................................................. 21

CRC/EDH Errors ......................................................... 21

EDH Data ................................................................... 21

PayLoad ID – SMPTE 352 .......................................... 22

Audio status .................................................................. 22

ANC status .................................................................... 23

System wide Settings ........................................................ 23

Memories ...................................................................... 23

Saving Memories ...................................................... 24

Recalling Memories .................................................. 24

Renaming Memories ................................................. 24

Adding more Memories ............................................ 24

Clearing Memories .................................................... 24

Export ........................................................................ 25

Command Scripts (cost Option) .................................... 25

Creating a command script: ...................................... 26

Adding steps to a command script ........................... 26

Editing Steps in a Command Script ........................... 26

Script Actions ............................................................ 26

Deleting steps in a command script .......................... 27

Deleting a command script: ...................................... 27

Renaming a command script: ................................... 27

Running a command script: ...................................... 27

Looping a command script:....................................... 27

Network Setup .............................................................. 28

TCP/IP setup ............................................................. 28

Remote web browser access .................................... 29

Phabrix Sx Manual December 8, 2008

Setting up a Windows(tm) PC for direct connection 29

Misc. .............................................................................. 30

Changing the date/time: ........................................... 30

Changing the date Format: ....................................... 30

Setting LCD Brightness .............................................. 31

Setting Screen Saver Mode ....................................... 31

Setting User Language .............................................. 31

Changing Options Security Code ............................... 31

Engineer ........................................................................ 31

Clear Memories ......................................................... 31

Default settings ......................................................... 32

Ext. Reference ........................................................... 32

Software Upgrade ..................................................... 32

Logging .............................................................................. 32

Events Log ..................................................................... 33

Events Log Setup ........................................................... 33

Audio Thresholds ...................................................... 34

Video status .............................................................. 34

Logging ...................................................................... 34

Speaker ............................................................................. 34

Remote File Access ........................................................... 35

FTP ................................................................................. 35

Unit file structure .......................................................... 35

Patterns ..................................................................... 36

Scripts ........................................................................ 36

Audio ......................................................................... 36

Setup ......................................................................... 36

Idents ........................................................................ 36

Page 4

Fonts ......................................................................... 36

File Formats ...................................................................... 37

Command Scripts .......................................................... 37

LOADMEM, ............................................................... 37

CHECKERRS, .............................................................. 37

PROMPT .................................................................... 37

CLEARLOG ................................................................. 37

DISABLEEVT ............................................................... 37

ENABLEEVT ............................................................... 37

CLEARERRORS ........................................................... 37

Phabrix Sx Manual December 8, 2008 Page 5

GETTING STARTED WITH PHABRIX SX

PACKAGE CONTENTS

The shipping box should contain:

1 black carrying bag containing:

Phabrix SX unit

Power Supply Unit

Mains lead

This Manual on a CD

GENERAL SAFETY

AVOIDING PERSONAL INJURY

This instrument is designed for use by qualified personnel only.

No user serviceable parts are provided. Units should be returned to your local PHABRIX agent for servicing.

The Operator should NOT remove the case from the unit.

Do not spill any liquid onto the unit or its power adaptor.

POWER SUPPLY

Make sure that the unit is connected to the correct power supply voltage. A power supply adaptor is supplied with the unit which may be connected to any AC power supply between 100 and 240VAC at 50-60Hz. Only the supplied power adaptor adaptor should be used with the unit. Do not use a damaged AC cable with the unit as it may cause a shock or fire hazard. Replacement AC cables are available from your local PHABRIX agent.

INSTALLATION ENVIRONMENT

OPERATING TEMPERATURE

The unit should only be operated between 0 and 40 °Centigrade. If the unit is operated at a higher temperature there is a possibility of a fire hazard. If the temperature is changed rapidly from a cold environment to a hot environment, moisture can be created internally which can cause malfunction or damage the unit. Allow the unit to sit for 30 minutes without power applied to reduce any possibility of condensation.

INPUT/OUTPUT TERMINALS

Do not connect the input or output BNC connectors to external power as this can damage the internal circuitry and cause the unit to work incorrectly.

Phabrix Sx Manual December 8, 2008 Page 6

WHEN NOT IN USE

Disconnect the unit from the power supply and AC power source when not in use.

MAINTENANCE

Wipe the case, and knobs gently with a soft cloth, lightly dampened with a neutral cleaning agent. A screen cleaning cloth may be used to clean the LCD. Do not apply force to the LCD when cleaning or it may be damaged.

Remove the power supply from the unit and turn OFF before cleaning. Do not allow any water or other liquid to enter the unit while cleaning.

Phabrix Sx Manual December 8, 2008 Page 7

INTRODUCTION

Thank you for purchasing a PHABRIX SX unit. Please read this manual carefully to ensure safe and correct operation. If, on reading this manual you still have questions on using the unit, please contact PHABRIX via the support web site at http://www.phabrix.com

. You may also register your product at this web site for a FREE extra 1 years warranty.

WARRANTY

The product has been designed and manufactured to be of the highest quality. However, should the unit develop a fault during the warranty period, please return to your local PHABRIX agent for repair.

WARRANTY EXCEPTIONS

PAHBRIX are not responsible for free service during the warranty period under the following conditions:

Incorrect voltage applied to unit.

Incorrect power adaptor used.

Fire, natural disaster.

Repair of instrument by non PHABRIX approved agent

Repair of damage caused by third party products.

Repair of damage caused by improper use.

Repair without proof of purchase.

OPERATING CONDITIONS

EXTERNAL POWER SUPPLY

DC-5V 5V

MAXIMUM INPUT SIGNALS

SDI input(s)

REF input:

LCD MONITOR

+/- 2V

+/- 2V

The LCD may have some pixels that are always turned ON or always turned OFF. This is normal and should not affect normal operation.

Phabrix Sx Manual December 8, 2008 Page 8

LCD FLICKER

The unit supports many video standards. The input SDI signal is displayed asynchronously and may flicker on the waveform display or picture display. The unit stores the input SDI signal internally, and then reads the internal frame using the LCD sync signal, which is asynchronous to the input SDI signal. LCD flickering may occur if a frame is skipped or repeated.

REMOTE NETWORK OPERATION

Remote network operation is only guaranteed when connected to a local machine.

DISPOSING OF THE UNIT

This product is subject to the European WEEE (Waste Electrical and Electronic Equipment) directive and should be disposed of according to the regulations of each country.

This unit contains a Lithium Polymer battery which should be disposed of correctly.

Phabrix Sx Manual December 8, 2008 Page 9

SPECIFICATION

LCD DISPLAY

Display Type

Display Format:

Backlight

Screen Saver

4.3 inch TFT colour

480 x 272 24 bits

Variable brightness

Reduces brightness after user adjustable time under battery operation.

SDI INPUTS

Supported standards

Connector

Input Impedance

525/59.94, 625/50,

720p/23.98,24/25/29.97/30/50/59.94/60

1035i/59.94/60

1080psF/23.98,24/25/29.97/30

1080i/50/59.94/60

1080p/23.98,24/25/29.97/30/50/59.94/60

BNC

75 ohm terminated

Input Return Loss >= 15dB (5MHz to serial clock frequency)

Maximum Input Voltage +/- 2V

EXTERNAL REFERENCE

Input Signal

Connector

Tri-level or Bi-Level (black burst) syncs 50/59.94/60Hz

BNC

Input Impedance 75 ohm terminated

Maximum Input voltage +/- 2V

EXTERNAL CONTROL

Ethernet

Ethernet Connector

USB

USB Connector

IEEE802.3 100Mb/s

RJ-45

USB 1.1 OTG (On the go - simulates flash disk)

Mini-AB

Phabrix Sx Manual December 8, 2008 Page 10

HEADPHONE OUTPUT

Connector

GENERAL

Miniature 3.5mm Stereo Jack

Environment

Operating Temperature 0-40 °C

Operating Humidity

<85% RH (no condensation)

Power Requirements

Dimensions

Weight

Accessories

AC 90-250V 50/60Hz 10W max

230 (L) x 93(H) x 45(D) mm

2kg

Instruction Manual on CD

AC Adaptor

Power Cord

Carry Case

Phabrix Sx Manual December 8, 2008 Page 11

OPERATION

TURNING THE UNIT ON OR OFF

To turn on your Phabrix SX press the red button at the top right hand side of the unit.

Once the system has started, pressing the button again will turn it off. If for some reason the unit stops responding, pressing and holding the button for a few seconds will turn it off.

The unit is battery operated and the battery should last for between 3 and 6 hours dependent on which video standard is being used and whether the input and outputs are active. If the unit is turned ON and the battery is almost discharged, it will turn itself OFF. If the battery level starts to get too low for operation a dialog will be displayed prompting the user to turn off the unit. The current settings will also be saved. The unit should then be connected to the power supply unit as soon as possible to recharge the battery. The battery should be fully charged in about 4 hours. The battery will charge faster if the unit is turned off while charging.

UNIT STATUS LINE

The top line of the menus shows the unit status and includes the Input Video status, Genlock status, battery status and current time. If the unit is being powered by the AC adaptor, the battery status will not be shown. If a command script is being run, this is shown on the top line to the left of the time as “Script”. A red “Log” is shown on the status line if there are any events in the event log.

The input video standard will be shown in red if the input signal has errors, but only if the analyser is connected to the input.

The output standard will be shown in red if the unit is generating errors.

Page 12 Phabrix Sx Manual December 8, 2008

USING THE MENUS

The Menu keys select which instrument is in use as well as selecting the options for the unit. The bottom of the LCD display shows the function of each Menu Key.

When in a specific function e.g. GENERATOR, ANALYSER etc, the cursor controls on the right hand of the unit select a field to edit. The current field has a red box around it. Pressing the OK button starts or ends editing of that field. Check boxes are always in edit mode and pressing the OK button inverts the current state. When editing some types of fields, the menu buttons at the bottom take on different functionality to allow cancelling edit mode or setting the default value for the field. When lists of options are displayed the menu buttons allow selection of the first/last item in the list as well as paging through the available options. If the list of items is small, the available items may be displayed on the menu buttons for instant access. The buttons along the bottom of the LCD display change the currently displayed page when not editing a data field. The left hand button always selects the top level menu and system overview page.

MAIN MENU

This page displays an overview for the system. It shows the video input and output standards and frame rate. The audio status for the input and generator shows which audio groups are present. If the generator is generating errors then the video standard is displayed in red. If the input detects errors, then the input standard will be displayed in red if the analyser is monitoring the input.

The menu buttons select the page to display as follows:

GENERATOR

This selects the generator page to allow setup of the generator output.

ANALYSER

This selects the analyser page to allow viewing and setup of the video and audio input signal.

SIG INFO

This selects the detailed signal information page which shows video and audio status for the analyser source.

Phabrix Sx Manual December 8, 2008 Page 13

SYSTEM

This allows setting up of system wide settings including memories and date/time.

LOGGING

This selects the logging page to view and setup the error logging system.

SPEAKER

This selects the speaker/headphone setup page.

SIGNAL GENERATOR

VIDEO SETTINGS

The unit supports all SD and HD SDI output standards including the 3GHz standards at 1080p/50/59/60. It will support Y,Cr,Cb formats as well as RGB formats.

Output On:

Standard:

Turns ON the output circuitry. Turn this OFF to save battery power if you are not using the video output.

Sets the video output format. Can select SD/720p/1035i/1080sf/1080i/1080p/3G-1080p.

Pattern:

<->:

Sets the frame rate for the current video output format. Only valid frame rates for the output standard may be selected.

This selects the video pattern that is output by the generator. Many standard patterns are provided by the unit. You may also download your own test patterns to the unit into the ‘Patterns’ directory and then select them using the

‘User File’ option in this field. See the ‘Unit File Structure - Patterns’ section for file formats supported. If a ‘User File’ has been selected, the ‘i’ button beside the file can be selected to get information on the file properties. Unsupported file types will be displayed as a black image.

This field, when checked, causes the test pattern selected to scroll horizontally. This helps check that a signal has not frozen due to signal loss. (Not implemented yet)

Phabrix Sx Manual December 8, 2008 Page 14

EDH:

Errors:

Ident:

If the output signal is SD (PAL-625 or NTSC-525) the insertion of EDH information may be turned on or off.

This field enables the insertion of CRC or EDH errors into the video signal. This allows checking of third party error detection circuitry. The number of errors inserted per second may be adjusted. (Currently error generation can only be turned on/off)

This section allows selection of either a user defined bitmap picture which has been downloaded into the ‘Idents’ directory of the unit or user defined text. For both types, the position of the ident can be set to one of Top-Left, Top-

Centre, Top-Right, Left-Centre, Centre-Screen, Right Centre, Bottom-Left, Bottom-Centre, Bottom-Right. When text ident is used, the font, font size and colour can be specified. Several fonts are built in to the unit, other fonts may be downloaded as required.

A smaller version of the video output signal is displayed on this page with any overlaid ident

GENLOCK SETTINGS

Source:

Delay:

The generator may be locked to an input reference which may be either the reference input which is a Bi-Level/Tri-

Level sync or may be locked to the video input. Alternatively, the generator may free run.

The delay of the video output may be adjusted by either lines and pixels or by micro-seconds. Adjustment of either field results in the alternative field being modified to match.

Page 15 Phabrix Sx Manual December 8, 2008

AUDIO SETTINGS

The unit can embed an audio signal on all 16 embedded audio outputs. If the SxA version has been purchased, the AES output source may also be set up.

Group n:

Source:

Each of the four groups may be separately enabled. When enabled, the source and level of each channel in a pair can be selected.

The source for each channel may be Silence, Fixed tone (a range of fixed frequencies), variable tone (where the frequency can be set on 1Hz steps from 20Hz to 20kHz), white noise, pulse tone or a user defined file downloaded into the ‘Audio’ directory on the unit. User defined files are looped and so should be set so that they start and finish on a zero crossing point of the signal. Dolby-E signals can also be used, downloaded into the ‘DolbyE’ directory.

Inv:

Link:

This check-box phase-inverts the audio signal to allow checking of third-party audio mixing.

This links the left and right channels of a pair so that changing the level on one channel changes the level on the other channel.

MASTER

The master level sets the 0dB level for all the embedded audio channels. Thus if the Master level is set to -18dB and group 1 pair 1 output is set to -2dB, the actual level output on that pair is -20dB. This feature allows all embedded outputs to be adjusted together as well as giving a simple method to change from -18dB to -20dB based standard levels.

AES O/P

This is only available on the SxA. The AES o/p control selects the source for the AES output of the unit. The source can be:

The AES input (via the built-in sample rate converter) to allow easy conversion to 48kHz sample rate AES signals.

A mirrored output of the specified embedded AES stream contained within the SDI output to allow easy generation of audio tones or white noise.

One of the embedded input pairs on the SDI input to allow use as a de-embedder.

Phabrix Sx Manual December 8, 2008 Page 16

Off, if the output is not required.

Note that the AES output is always at a 48kHz sample rate. The AES input may be at any sample rate from 32kHz to 192kHz.

ANALYSER

The signal analyser displays the video input or output signal it may also analyse a previously grabbed frame of video. On any page, if the page button is pressed again the video signal will be displayed in full screen mode. While in full screen mode, if the video input signal is not present, a message saying ‘No video detected’ will be shown. The source for the analyser is set by the field at the bottom left of the page. The current line and sample are the same for all pages of the analyser and therefore when selecting a sample on the Signal Data page, the cursor will show where that is on the ‘Full Picture’ page.

PICTURE

On this page, the video picture is displayed in a window as a down-converted display. Select ‘Pulse-X’ mode to change the picture to display the horizontal and or vertical blanking. Four modes of pulse cross are available:

OFF:

Normal picture display

H+V:

Display the horizontal and vertical blank interval for the signal

H:

Display the horizontal blanking interval.

V:

Display the vertical blanking interval.

Line:

Specify the current line for analysis.

Sample: Specify the current sample for analysis.

Field:

(Only for interlaced formats) This control selects the field for analysis. If the current line is in Field 1, changing the field will change to the same line in field 2. The field number will change to show the correct field for the current line number.

Aspect Ratio:

The aspect ratio of PAL (625 lines) may be set to be either Auto, 4:3 or 16:9 as required for the source signal. NTSC

(525) signals are always 4:3 aspect ratio and HD signals are always 16:9.

Zoom: If set to ‘Norm’ the picture displayed is a down-converted picture to fill the window or the full screen as requested. If set to

‘1:1’ then a pixel on the units LCD display will reflect a pixel in the video signal. In 1:1 mode, no filtering is used and by pressing ‘OK’ the cursor keys can scroll around the full picture.

Phabrix Sx Manual December 8, 2008 Page 17

ActPix: If checked, only active picture is displayed for picture view and waveform monitor, all other VBI data such as TRS words ANC data and audio is blanked.

Source: The analyser can view either the video SDI input or the Generator output so that comparisons can be made between input and output.

AUDIO METERS

This page shows up to 16 embedded audio channels. The Meter Setup button shows a dialog to set which meters are displayed, what their scale is set to and what labelling is used.

The scale used by the meters may be set to -18dB or -20dB according to the standard operating level used. The scales may be dBFs,

EBU, DIN, Nordic or VU.

Phabrix Sx Manual December 8, 2008 Page 18

SIGNAL DATA (ENGINEERING OPTION)

This page shows the video signal as a data stream in several formats as described below. In each mode, the data may be displayed in hexadecimal, decimal or binary formats. The binary format is not available in grid mode. The data may also be displayed as 10bit or 8bit format.

To simplify scrolling around the SDI signal view, use the cursor keys to move the red focus rectangle to the data display and press OK.

The focus rectangle should change to a blue colour and the cursor keys allow the display to be scrolled in any direction. The menu keys at the bottom allow the Line/Sample to be adjusted by 10 or 100 in any direction.

Grid: In this mode the data is shown in an X-Y format with lines and samples at the same time. Thus this is a representation of the picture but in data format. If the cursor keys are used to select the grid window and ‘OK’ pressed, the cursor keys will scroll around the window. The luma (Y) channel is shown in white whilst the Cr and Cb channels are shown in with a red or blue tint respectively.

Strm:

This is one of 3 modes for displaying the samples on the current line only. The cursor keys may be used to scroll the sample number when ‘OK’ has been pressed. The ‘info’ column shows the type of data being displayed.

AP

VBL

HBL

Active picture

Vertical blanking

Horizontal blanking

Comp: This is a similar to the Strm mode above but in a component mode with Cr and Cb in different columns. See Strm mode above for the info column description.

Split:

This mode shows the 2 streams split into 4 parts to show Y, Y’, Cr and Cb in separate columns. See Strm mode above for the info column description.

Source: The analyser can view either the video SDI input or the Generator output so that comparisons can be made between input and output.

WAVEFORM MONITOR

Phabrix Sx Manual December 8, 2008 Page 19

WAVEFORM CONTROLS

The line, sample and field controls all track the related controls on other pages.

The ‘All’ checkbox forces the waveform monitor to display all lines, overlaid on each other. Otherwise the specified line is displayed.

The waveform display format is set using the ‘Setup’ button. The waveform monitor displays the selected signal in one of 5 formats as set by the Mode control:

YCbCr

Y

Cb

Cr

GBR

Red

Green

Blue

The Y, Cr and Cb waveforms are shown as three separate waveforms in a Parade format

Only the luminance channel is shown

Only the blue chroma channel is shown

Only the red chroma channel is shown

The Green, Blue and Red waveforms are shown as three separate waveforms in a Parade format

Only the red channel is shown

Only the blue chroma channel is shown

Only the red chroma channel is shown

The next 2 controls show the vertical and horizontal magnifications. If either are set to a value other than x1, a vertical or horizontal offset will be shown on the bottom right edge of the waveform display. The offsets are set by moving the focus cursor to the waveform window and pressing OK. The menu keys can then be used to scroll around the window. The offset will be the same for all magnifications values so switching between x1 and x5 will not lose the offset used in x5 mode.

The vertical offset values are in signal level bits but displayed in the current scale units so switching between scales will keep the same offsets.

Source: The analyser can view either the video SDI input or the Generator output so that comparisons can be made between input and output.

SETUP BUTTON/DIALOG

The waveform monitor displays the analyser input in an oscilloscope like form.

The display may be restricted to a single line or all lines may be displayed at the same time. The display can be formatted as all streams (Luma and Chroma) or just a single stream. The streams may be in YCbCr or GBR formats.

The display may also be magnified to examine a specific part of the waveform.

Active Picture:

If checked, only active picture is displayed for picture view and waveform monitor, all other VBI data such as TRS words ANC data and audio is blanked. This control tracks the state of the ActPix control on the Picture tab.

Phabrix Sx Manual December 8, 2008 Page 20

Vertical Scale: The axes for the waveform monitor can be displayed either in percentages or in hex or decimal values as required.

VECTOR SCOPE

The vector scope may be set to show either the 100% bar positions or 75% positions. The colour bar position boxes will change according to the colour space for the current input video standard.

The ‘All’ checkbox forces the waveform monitor to display all lines, overlaid on each other. Otherwise the specified line is displayed.

SIGNAL INFORMATION

These pages give a detailed view of the status of the input (or generator output) signal. They are sub-divided into Video, Audio and

Ancillary (ANC) status.

VIDEO STATUS

CRC/EDH ERRORS

This displays the EDH or CRC status of the video input as appropriate for the video standard being monitored. The status of each

CRC/EDH count is displayed as the number of seconds since an error occurred. The individual CRC fields may be reset individually or together.

NOTE: In the current version of software, errors in ancillary packets are NOT checked. This will be added in a future version.

EDH DATA

If the signal is SD (PAL-625 or NTSC-525) the EDH values for both active picture (AP) and full-field (FF) are displayed for each field. To enable engineers checking EDH integrity, the EDH values calculated for active picture and full-field are also displayed.

Under normal conditions, the EDH-AP values should be constant, the full-field values may change if audio or other ancillary data is embedded in the SDI signal.

The EDH flags for active picture, full-field and ancillary data are also displayed for diagnostic purposes.

Phabrix Sx Manual December 8, 2008 Page 21

Edh: Error Detected Here:

This is set to 1 if a SDI error was detected. In the case of ancillary data, this means that one or more ANC data packets had an incorrect checksum.

Eda: Error Detected Already:

This is set to 1 if a SDI error was detected in the signal received by the previous device.

Idh: Internal error Detected Here:

This is set to 1 if a hardware error was detected in the previous device.

Ida: Internal error Detected Already: This is set to 1 if an idh flag was received by the previous device.

Ues: unknown error status: This is set to 1 if the previous device received an SDI signal from a device not supporting EDH

.

PAYLOAD ID – SMPTE 352

If the unit has detected a SMPTE 352 ancillary packet, it will be displayed here in hex and decoded format.

The display shows the transport media, frame rate and interlaced/progressive/segmented frame mode, as well as colour format, bit depth, dynamic range and channel number. See the SMPTE 352 specification for further details.

AUDIO STATUS

The Channel Status for the selected audio channel is displayed in decoded form as well as a hexadecimal dump of the bytes. The source may either come from the input signal or from the generator for comparison purposes.

Page 22 Phabrix Sx Manual December 8, 2008

ANC STATUS

The data contained in Ancillary packets may be displayed and decoded.

The data ID for the packet to be displayed should be selected using the DID field. A description of the DID is shown next to it and the header of the packet and checksum is displayed immediately below. The data packet is displayed as a hexadecimal list at the bottom of the page and may also be decoded into a text format.

If an error occurs in the packet, the header and checksum are displayed in red.

The display of data may be frozen by pressing the Freeze checkbox.

If the On Error checkbox is checked an error in the data packet will cause the display to be frozen and the Freeze checkbox will turn red.

Un-checking the freeze button will restore normal operation.

SYSTEM WIDE SETTINGS

MEMORIES

Phabrix Sx Manual December 8, 2008 Page 23

The current settings in the unit may be saved for future recall. The check boxes on the right-hand side determine what is saved in the memory. Non overlapping memories may then be combined together.

SAVING MEMORIES

Select the memory to save settings to by moving the cursor to the ‘Select Memory’ list and pressing ‘OK’. Use the up/down cursor keys to select the memory and then press ‘OK’. The name edit field next to the memory list will be updated with the name of the selected memory.

Specify which aspects of the unit are to be saved in the memory using the check boxes on the right hand side. If you want to change the name of the memory see the section ‘Renaming Memories’ below before saving.

Move the cursor to the ‘Save’ button and press ‘OK’. The text field under the Memory list will change to show the last Saved or Recalled memory.

RECALLING MEMORIES

Select the memory to be recalled by moving the cursor to the ‘Select Memory’ list and pressing ‘OK’. Use the up/down cursor keys to select the memory and then press ‘OK’. The name edit field next to the memory list will be updated with the name of the selected memory.

Move the cursor to the ‘Recall’ button and press ‘OK’. The text field under the Memory list will change to show the last Saved or

Recalled memory.

RENAMING MEMORIES

Select the memory to be renamed by moving the cursor to the ‘Select Memory’ list and pressing ‘OK’. Use the up/down cursor keys to select the memory and then press ‘OK’. The name edit field next to the memory list will be updated with the name of the selected memory.

Move the cursor to the edit field and press ‘OK’. The menu keys at the bottom of the screen will change to show text keys similar to a mobile phone. Pressing a key repeatedly will cycle through the characters for that key. Pressing a different key will move the cursor to the next position and insert the first key value for that key. The cursor left and right keys can be used to move the position of the text cursor which is where the next character will be inserted. The ‘Back Space’ key will delete the key to the left of the cursor. The ‘Cancel’ key will cancel edit mode and restore the original text. Once you have finished editing the name, press ‘OK’ to rename the memory.

ADDING MORE MEMORIES

To add a new memory, press the ADD button. This will use the current memory name and settings.

CLEARING MEMORIES

Select the memory to be cleared by moving the cursor to the ‘Select Memory’ list and pressing ‘OK’. Use the up/down cursor keys to select the memory and then press ‘OK’. The name edit field next to the memory list will be updated with the name of the selected memory.

Move the cursor to the ‘Clear’ button and press ‘OK’. The memory list will show the default name for that memory. A cleared memory cannot be recalled.

Phabrix Sx Manual December 8, 2008 Page 24

EXPORT

Memories can be exported to a single file for backup purposes and re-imported into the internal format. This allows different sets of memories to be used for different tasks, such as command scripts. Exported memories are in a text format and so may be edited on a

PC and downloaded to/from the backups directory on the unit via FTP or USB.

To export memories, select the Backup... button which will open a window to select the file to create. Existing files are shown and a new filename may be selected using the File Name edit box. Select the Save button to create the backup file. The filename created will have a .mem file extension.

To import memories, select the Restore... button which will open a window to select the file to import. Existing files are shown and one should be selected before the Load button is pressed. Select the Load button to import the backup file.

NOTE: When importing a backup file, all existing memories are deleted before the backup file is loaded so make sure that they are backed up first.

COMMAND SCRIPTS (COST OPTION)

Command scripts allow sequences of actions to be stored and recalled thus allowing often repeated test procedures to be run. Each step of a sequence has an associated action and time that the action is valid for. Command scripts are stored in the ‘Scripts’ directory on the unit and may be downloaded to the unit via the

Page 25 Phabrix Sx Manual December 8, 2008

USB port or via FTP transfers to the unit. See the section on Remote File access for further details. See also the section on file formats for details of command script formats

CREATING A COMMAND SCRIPT:

Move the cursor to the ‘New’ button and press ‘OK’. A new script will be created called ‘NewScriptn’ where ‘n’ is the next available script number. See also ‘Renaming a command script’.

ADDING STEPS TO A COMMAND SCRIPT

Move the cursor to the ‘Add Step’ button and press ‘OK’. A new step will be created with the default action of Recalling Memory 1 in 1 second. Multiple steps can be created at this stage and then the actions edited later.

EDITING STEPS IN A COMMAND SCRIPT

Select the step to be edited by moving the cursor to the step list and pressing ‘OK’ and then using the cursor keys to select the step.

Press ’OK’ again to select this step and update the action and time fields for the step. Move the cursor to the field to be edited and edit as required.

SCRIPT ACTIONS

Step Action is one of:

RECALL MEMORY,

Loads the memory specified in the Memory field.

CHECK INPUT ERRORS,

Check for errors in the step Time. If there are any errors an event is added to the event log and the sequence pauses with a message to continue or abort. (Optional? Checkbox to force abort on errors????? Which errors are checked? Use log setup to set errors???

PROMPT USER

This will put a short prompt on screen in a dialog asking the operator a question and waiting for OK to be pressed to continue.

CLEAR EVENT LOG

This will clear the event log which is useful when starting a test script so that the event log starts in a known state.

DISABLE EVENTS

This disables logging globally while a major change in output/input status is about to take place so that unwanted log events are not added to the log.

ENABLE EVENTS

Phabrix Sx Manual December 8, 2008 Page 26

This re-enables event logging after disabling it above.

CLEAR ERROR COUNTS

This clears all CRC/EDH error counts and should be run before checking for errors.

DELETING STEPS IN A COMMAND SCRIPT

Select the step to be deleted by moving the cursor to the step list and pressing ‘OK’ and then using the cursor keys to select the step.

Press ’OK’ again to select this step. Move the cursor to the ‘Delete Step’ button and press ‘OK’.

DELETING A COMMAND SCRIPT:

Move the cursor to the list of command scripts on the left hand side and select the script to be deleted as the current script. Then move the cursor to the ‘Delete’ button and press ‘OK’. The currently selected script is the one that is deleted.

RENAMING A COMMAND SCRIPT:

Move the cursor to the list of command scripts on the left hand side and select the script to be renamed as the current script. Move the cursor to the name edit field and press ‘OK’. The menu keys at the bottom of the screen will change to show text keys similar to a mobile phone. Pressing a key repeatedly will cycle through the characters for that key. Pressing a different key will move the cursor to the next position and insert the first key value for that key. The cursor left and right keys can be used to move the position of the text cursor which is where the next character will be inserted. The ‘Back Space’ key will delete the key to the left of the cursor. The ‘Cancel’ key will cancel edit mode and restore the original text. Once you have finished editing the name, press ‘OK’ to rename the script.

RUNNING A COMMAND SCRIPT:

Move the cursor to the list of command scripts on the left hand side and select the script to be run as the current script. Move the cursor to the ‘Run’ button and press ‘OK’. The Run button will now become a ‘Stop’ button to stop the running script.

LOOPING A COMMAND SCRIPT:

By default a script will run all steps and then stop running. The ‘Loop’ check box allows a script to loop until stopped. This feature allows testing of a system forever for heat-soak testing in a manufacturing environment or to check for system reliability over a period of time.

The Loop Count field shows how many loops have been run and how many hours and minutes the script has been running for.

Page 27 Phabrix Sx Manual December 8, 2008

NETWORK SETUP

TCP/IP SETUP

Phabrix SX is fully networkable and has a complete network interface to allow control of any unit from any other unit. This allows a remote generator to be controlled by a local analyser to check video links. All units on a network need an ‘address’ so that the other systems know how to talk to them. If you have a DHCP server on your Ethernet network, check the DHCP box and a TCP/IP address will be assigned to the unit automatically. If your network has ‘static’ (non changing) addresses, you will need to see your network administrator to get an address assigned to your unit and to get the Subnet mask and default gateway address.

Phabrix Sx Manual December 8, 2008 Page 28

REMOTE WEB BROWSER ACCESS

The Phabrix Sx may be controlled via your web browser. Operation has been tested using Firefox and Internet Explorer but should work with any browser which allows JavaScript.

To connect to the Phabrix Sx, find out its IP address and type it into the web browser.

You should see a screen like this:

Clicking on the buttons works exactly like the unit that you are connected to (note that auto-repeat doesn’t work due to browser speed limitations). The LCD screen on the browser image is updated every 10 seconds but may be refreshed manually by clicking the LCD area on the browser.

Screen dumps of the Phabrix Sx can be made using the browser’s or operating systems facilities. (On Windows(tm) Alt+Print Screen causes the web browser’s window data to be stored on the clipboard).

SETTING UP A WINDOWS(TM) PC FOR DIRECT CONNECTION

A Phabrix SX may be connected to a PC via your network or directly using an Ethernet cable between your PC and the SX. The Phabrix SX does not require a cross-over cable as it auto-senses the connection, thus any Ethernet cable may be used. The Phabrix SX should be set to a fixed IP address.

The PC’s network settings should be set to give a fixed IP address in the same range as the Phabrix Sx’s

IP address, but use a different address. To set up the PC, go to the Windows control panel and select

Network Connections and then select Local Area Connection and get its properties dialog.

Select Internet Protocol (TCP/IP) which is at the bottom of the list and click the ‘Properties’ button.

Page 29 Phabrix Sx Manual December 8, 2008

MISC.

A new dialog should appear. Set the IP address to a different value to that of your Phabrix

SX but keep the first 3 numbers the same. The Subnet mask should be the same as that on the Phabrix SX and should be as shown above.

The orange LED on the Phabrix SX Ethernet connector should light when connected and the green light should flash when network data is detected.

The Phabrix SX should be set with DHCP OFF and the other settings as shown. The DNS server setting MUST be set to the address of the PC that is directly connected for FTP and

Web access to work.

The IP Address can be any value within the valid Subnet region.

This page shows serial numbers, unit MAC address, version information and battery state. It also allows the date and time to be set and factory default settings to be recalled. The options security code is entered on this page if options are purchased.

CHANGING THE DATE/TIME:

Check the ‘Enable Change’ checkbox under the date and time fields and then use the cursor to move around the date and time. Press

‘OK’ to edit a field and again to complete the change. When all fields have been edited, uncheck the ‘Enable Change’ check box and the new date and time will be set.

CHANGING THE DATE FORMAT:

The date format used on logging screens etc may be set using the Format selection control. Three formats are available: Date-Month-

Year, Month-Date-Year and Year-Month-Date.

Phabrix Sx Manual December 8, 2008 Page 30

SETTING LCD BRIGHTNESS

Use the LCD brightness slider to set the LCD brightness, using a lower brightness will reduce the current drawn from the battery and thus lengthen battery life.

SETTING SCREEN SAVER MODE

When the unit is battery powered the screen saver will operate after a user defined time period and reduce the LCD brightness down to its lowest level. This will lengthen battery life. Pressing any key will restore the previously set LCD brightness.

SETTING USER LANGUAGE

The language used to display the menus in may be changed to one of the supported languages. (Currently only English is supported)

CHANGING OPTIONS SECURITY CODE

When new options are purchased for the unit a new Security Code will be supplied. This is specific to this unit and cannot be used on other units. The security code is entered using the Edit field below the System Option list and is edited by pressing OK and using the menu function keys. If an incorrect Security code is entered a dialog will be displayed. The new Security code will only be saved if it is valid, the old code will be used until a valid code is entered. Once a valid new code has been entered, the list of options provided by that code will be displayed. The options code is stored independently to memories and system settings.

ENGINEER

This page adds features for special use.

CLEAR MEMORIES

Check the ‘Enable’ checkbox above the ‘Clear ALL’ button. Press the ‘Clear ALL’ button and all the user memories will be deleted.

Phabrix Sx Manual December 8, 2008 Page 31

DEFAULT SETTINGS

Check the ‘Enable’ checkbox above the ‘Set Defaults’ button. Press the ‘Set Defaults’ button and the factory defaults will be recalled.

This will not affect the system Security Code or Free Run frequency.

EXT. REFERENCE

Press this button to calibrate the free running frequency using the External reference or Video Input as selected by the Generator –

Genlock page. If no input is present or the Generator is free running, an error dialog will be shown. The calibration will take a few seconds to take place and once completed, the calibration value will be shown. This reference value is not lost when memories are cleared or factory defaults recalled above.

SOFTWARE UPGRADE

The unit may download and install new versions of software when available. This process is a two stage process: Download the software and then Install it.

DOWNLOAD

Make sure that the network settings are correct and that the unit is connected to the internet via the Ethernet connection. Note that the unit should be connected to the AC adaptor so that the battery state does not affect the download process.

Pressing the Download button will cause the current release of software to be downloaded from the Phabrix Web Site. This will take a short time dependent on the connection to the Internet. Once the download has completed, the software will be checked for errors before being stored on the unit for future installation.

Multiple releases of software may be stored on the system so a unit may go back to a previous release if a test script requires it.

INSTALL

Note that the unit should be connected to the AC adaptor so that the battery state does not affect the installation process. The unit does NOT have to be connected to the Internet during installation.

Select the release of software to install using the field with releases listed. The largest number is the latest release. Select the ‘Install’ button and the installation will start. This process takes several minutes to decompress the software, extract the files and then reprogram the hardware and during this time the unit MUST be connected to the AC adaptor or corruption of the unit can occur which will stop it working. If an error is given during the installation, retry the installation and or download. Do NOT turn the unit off until an installation has completed correctly.

Once the installation has competed, the unit should be turned off and on again.

NOTE: As part of the installation procedure, all memories are archived to a backup file and then deleted. They can be restored from the

System-Memories page using the Restore button and selecting the ‘_Before_Upgrade’ archive.

LOGGING

Logging allows events to be detected and recorded for future examination. The events to be logged can be specified so that unwanted events do not appear in the event log. If a system has a problem with intermittent signals, the Phabrix Sx can be connected to that

Phabrix Sx Manual December 8, 2008 Page 32

source and can be left for several days to log any errors. At the end of that period the log can be examined and the time and date of each error noted.

The error log is saved when the unit is powered off and may be downloaded from the unit via FTP .

EVENTS LOG

The event log displays a list of events with a time stamp showing when they occurred. The event log may be cleared by pressing ‘Clear’ button on the Setup page. Pressing OK on the event log page allows the event log to be scrolled through to view all events in the list.

The currently selected event is shown on the event log header. The date format may be changed using the Date ‘Format’ control on the

System-Misc page.

EVENTS LOG SETUP

The events to be logged are set up on this page by checking the appropriate ‘Log’ checkboxes.

Phabrix Sx Manual December 8, 2008 Page 33

AUDIO THRESHOLDS

This section allows the thresholds for audio events to be set. If the audio level for a channel is higher than that specified for the Clip or

High fields for the number of video frames specified then an event will be added to the event log. If the audio level for a channel is lower than that specified for the Quiet or Muted fields, then and event will be added to the event log. Logging may be limited to specified audio pairs.

VIDEO STATUS

EDH/CRC

A log event will be added whenever a CRC/EDH error state changes. If the CRC/EDH state is correct, the event will show OK, else it will show FAIL. Separate Luma and Chroma CRC events may be shown. Note that CRC/EDH events may occur when an SDI signal is connected or removed.

INPUT STATUS

A log event will be added whenever the input video standard changes or if the input is lost or re-appears.

OUTPUT STANDARD

A log event will be added whenever the generator video standard changes.

LOGGING

This section controls ALL logging events

CLEAR

This allows the event log to be cleared.

ENABLE

This disables or enables logging of all events irrespective of that events individual Logging state.

SPEAKER

Page 34 Phabrix Sx Manual December 8, 2008

The unit contains a single loud speaker and associated stereo headphone socket which can be connected to any of the audio input or output channels or pairs.

The ‘Mute’ checkbox allows both signals to be muted.

A volume control is provided to adjust the level to headphones and speaker together. The balance control can be used to adjust the

Left-Right balance for the source being monitored.

The monitored source can be set to the input or output of the unit and the input and output sources can be set independently to either a stereo pair or single audio channel.

The speaker can be muted when headphones are plugged in by checking the ‘Mute Spkr if Headphones’ control.

REMOTE FILE ACCESS

Files may be uploaded or downloaded from/to the Phabrix SX by a remote PC using either the USB connection or over Ethernet.

FTP

You will need an Ethernet cable to connect the unit to a PC. If connecting directly to a PC, you will need a cross-over cable but if connecting via a router or switch, a normal Ethernet cable can be used.

To connect to the unit you will need to use a FTP client which may be a GUI based one such as FileZilla or that built into windows. To log onto the system you will need a login and password. The login for the system is ‘sxuser’ and the password is ‘phabrixsx’.

‘FileZilla’ is available for download from http://filezilla-project.org/download.php

Other FTP applications are available for other computer platforms.

Once you have connected to the unit, you should be able to see the unit file structure below:

UNIT FILE STRUCTURE

Several directories are provided for user files these are:

Phabrix Sx Manual December 8, 2008 Page 35

PATTERNS

This directory contains several sub-directories for the related video standards. Test pattern files should be placed in the correct directory for the video standard in use. i.e. if a 720p video standard is being used, only the test patterns in the 720yuv directory will be available. You may have a test pattern with the same name in different pattern directories but they should have the correct size and colour type (rgb/yuv) for the directory that they are in or they will be converted and may lose colour accuracy.

Patterns that are too large for the current standard will be clipped to fit; patterns that are too small will gain a black border and will be centred on the output image.

Test patterns are .pat files for the standard line based pattern files or user files. Supported user file formats are:

Windows bitmap files (.bmp)

10 bit video files (.dpx) We support 10bit RGB/YUV formats only in Left to Right orientation.

• targa (.tg) files which are compressed or uncompressed 8bit RGB files. We support Left to Right orientation only.

Note that bitmap files are RGB and will be converted to YUV internally as appropriate to the output standard. This will mean that not all valid YUV values can be output. If dpx YUV files are used then all valid values may be output.

SCRIPTS

Command scripts may be downloaded into this directory. See FileFormats-Command Scripts for more details.

AUDIO

.WAV audio files may be downloaded into this directory. They should be relatively short and edited so that they can be looped without clicking. The start and end audio levels should be at zero crossing point and have a similar slope.

SETUP

This directory contains the installer files for different releases of the software. Each release is comprised of a .tar.gz file and is self contained. If a release of software is copied onto the unit, the list of available releases will be updated within a minute.

IDENTS

This contains bitmap idents as used by the generator. Supported user file formats are: .bmp (Windows bitmap files), future formats will include: dpx files (10 bit video files) and targa (.tga) files

Note that bitmap files are RGB and will be converted to YUV internally as appropriate to the output standard. This will mean that not all valid YUV values can be output. If YUV files are used(.dpx), then all valid values may be output.

FONTS

The fonts available for use by the generator to add text idents may be enhanced by downloading new True Type fonts into this directory.

Phabrix Sx Manual December 8, 2008 Page 36

FILE FORMATS

COMMAND SCRIPTS

Command scripts are text files with an file extension of .cmd. They are formatted with one step per line. The line is formatted as

StepTime,StepAction,StepParameters

StepTime is a numeric value in seconds 0....n

StepAction is one of:

LOADMEM,

Loads the memory specified in the StepParameters field – must be a numeric value 1...num memories.

CHECKERRS,

Check for errors in StepTime seconds. If there are any errors an event is added to the event log and the sequence pauses with a message to continue or abort.

PROMPT

This will put a short prompt on screen in a dialog asking the operator a question and waiting for OK to be pressed to continue. The time field is ignored and the parameter following is the text displayed.

CLEARLOG

This clears the Event Log

DISABLEEVT

This disables Event logging

ENABLEEVT

This enables Event logging

CLEARERRORS

This clears all error counts (CRC/EDH etc.)

Page 37 Phabrix Sx Manual December 8, 2008