Byron Medical PSI-TEC III PT-PLT-III-110, PT-PLT-III-220 Aspiration Platform Instruction Manual
PSI-TEC III PT-PLT-III-110, PSI-TEC III PT-PLT-III-220 are third generation aspiration systems with a high flow and powerful vacuum pump designed to evacuate fluid, soft tissue and general tissue removal from a surgical site. It also has a compressor used to drive pneumatic components, all on an integrated cart making it compact, mobile, attachments located at working level, and a pneumatically operated footswitch.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
Byron Medical’s
PSI-TEC III
Aspiration Platform
(with White Cap)
Instruction
Manual
PT-PLT-III-110
PT-PLT-III-220
Manufacturer: EU
Byron Mentor
800-777-3434
(31)
21-OM-0001 Rev D
Table of
Contents
I. Introduction
Symbol Definitions
Use of this Documentation
Warnings and Cautions
Operational Safety
Modifications
Damage or Loss in Shipment
Incorrect Items Shipped by Byron Medical
Policy on Returned Goods
Repair Program
Warranty Policy
II. Operation
General Description
Operative Controls and Components
System Set-up
Canisters Set-Up Illustration (See Page 7)
Operating Instructions
III. Maintenance
6
Cleaning Instructions
IV. Troubleshooting
Problems, Causes, and Solutions
Appendices
8
9
A. PSI-TEC III Aspiration Platform Specifications (115V/230V)
B. PSI-TEC III Replacement Parts List (115V/230V)
C. PSI-TEC III Hand Rail Parts List
D. PSI-TEC III Aspiration Platform Schematics (115V/230V)
10
11
19
20
2004 by Byron Medical
Byron Medical reserves all rights to this document. No part of this manual may be reproduced, stored, or transmitted in any form or by any means (electronic, mechanical, photocopying, recording, etc.) without prior written permission of Byron
Medical. Requests for permission to reproduce or distribute this document should be addressed to:
Byron Medical, Inc., 602 West Rillito Street, Tucson, Arizona 85705
3
3
2
2
4
4
2
2
1
2
5
5
5
AP
Footswitch
Fuse and Fuse Rating
Symbol Definitions
Compressed Air
Vacuum
Air Out
Danger: DO NOT use in the presence of flammable anesthetics
Attention, Consult users Manual
Risk of Electric Shock. DO NOT remove equipment cover
21-OM-0001 1 of 21 Rev D
I. Introduction
Use of This Documentation
This manual provides instruction for operation, maintenance, and troubleshooting procedures. Users should be thoroughly trained in using this product and applicable medical procedures.
Instruction manuals should be made available to the user(s) during the procedure. Follow all instructions contained in this manual pertaining to the devices used in the procedure, with particular attention given to the WARNINGS and CAUTIONS.
Warnings and Cautions
WARNINGS and CAUTIONS statements are placed throughout the text referring to procedures, practices, or conditions considered essential to the safety of personnel and patients and/or equipment and property. All WARNINGS and CAUTIONS should be reviewed prior to the use of this equipment. If you have any questions, please contact Byron Medical, (520) 573-0857 or (800)-777-3434.
Operational Safety
A thorough review of this entire instruction manual is essential prior to using this product. Byron
Medical has taken great care to ensure the safety of the patient and staff. However, features may not be readily discernible without reviewing this documentation. Use of this equipment should therefore not be under taken until the user(s) is fully familiarized with the instructions for assembly and operation. If you have any questions, contact Byron Medical, (520) 573-0857, (800) 777-3434, or FAX (520) 746-1757.
Modifications
Modifications of any kind NOT recommended and will void all warranties.
Damage or Loss in Shipment
Thoroughly inspect shipment immediately upon arrival. If goods are received short or in damaged condition, it is important that you notify the transportation company and insist on a notation of the loss or damage on the freight bill. Otherwise, it may be difficult to make a claim against the Transportation
Company.
If concealed loss or damage is discovered, retain all packaging materials, notify the transportation company immediately, and request an inspection. The agent will make an inspection and grant a concealed damage notation. A concealed damage report must be made within seven (7) days of shipment delivery. After seven (7) days, the transportation company reserves the right to refuse any claim for loss or damage.
21-OM-0001 2 of 21 Rev D
Incorrect Items Shipped by Byron Medical
Please check your shipment immediately for any shortage or incorrect items. If any discrepancies exist, notify Byron Medical, (520) 573-0857 or 800-777-3434, at once. Your prompt attention will ensure credit or exchange.
ALL DISCREPANCIES BETWEEN THE PACKING LIST AND THE PRODUCTS RECEIVED
MUST BE REPORTED WITHIN 48 HOURS OF RECEIPT OF THE PACKAGE TO QUALIFY
FOR A CREDIT.
Policy on Returned Goods
NO RETURNS OR EXCHANGES WILL BE ACCEPTED UNLESS YOU HAVE RECEIVED A
RETURN OF GOODS AUTHORIZATION NUMBER (RGA) FROM BYRON MEDICAL. THIS
NUMBER MUST BE ON THE RETURN BOX AND ALL WRITTEN COMMUNICATION.
Return numbers will expire if items are not returned within 30 days.
Credit will only be issued if the merchandise is in saleable condition. If the product has been used in surgery, the product must be CLEANED WITH A DISINFECTANT and a written statement confirming such must accompany the product(s).
Full credit will be issued for any item returned in 30 days that is in undamaged and saleable condition.
A restocking charge of 25% will be charged to any product returned between 31-90 days. A credit memo, based on the dollar amount, will be issued to the account.
OPENED PACKAGES OR BOXES containing disposable items are not returnable for credit or exchange. Each package or sales unit specifically states “Not Returnable if Package Seal is Broken.”
NO CREDIT WILL BE ISSUED ON ANY ITEM RETURNED AFTER 90 DAYS.
CREDIT MEMOS EXPIRE TWO YEARS AFTER DATE OF ISSUE.
21-OM-0001 3 of 21 Rev D
Repair Program
If repairs are necessary due to damage other than that incurred during initial shipment (see section labeled DAMAGE OR LOSS IN SHIPMENT), contact Byron Medical to return your unit. A Return of
Goods Authorization number must be obtained from Byron Medical’s Customer Service Department prior to returning any merchandise. When requesting a Return Goods Authorization number, please follow the return policy as listed under the section labeled POLICY ON RETURNED GOODS. Then, please carefully repack and return it post-paid to:
Byron Medical, Inc.
Attn: Repairs Department
RGA #:________________
602 West Rillito Street
Tucson, Arizona 85705
(520) 573-0857
(800) 777-3434
Repairs must be made by Byron Medical or by an approved authorized agent. Attempting repair without prior authorization nullifies all warranties.
Warranty Policy
Byron Medical products are manufactured for use only by qualified medical personnel who are trained in their use. This equipment carries a one-year warranty against defects from date of sale, which warrants it to be free from defects in material and workmanship. This warranty is valid only to the original purchaser, and will be voided if transferred to a third party.
Any Byron Medical product with such defects returned will be promptly repaired or replaced at no charge to the customer. However, you are responsible for the cost of shipping the product to us. The warranty does not apply to damage caused by misuse, mishandling, improper operation, excessive voltage, and/or abuse of the product. Repairs or modifications performed other than by Byron Medical or an approved authorized repair facility will nullify this warranty.
For details and warranty information, call Customer Service Department, (520) 573-0857 or
(800) 777-3434.
21-OM-0001 4 of 21 Rev D
II. Operation
General Description
The PSI-TEC III
Aspiration Platform is a third generation aspiration system with a high flow and powerful vacuum pump designed to evacuate fluid, soft tissue and general tissue removal from a surgical site. It also has a compressor used to drive pneumatic components, all on an integrated cart making it compact, mobile, attachments located at working level, and a pneumatically operated footswitch.
Operative Controls and Components
1. Rails (Assembly Required)
1
2. ON/OFF Switches (Vacuum and Air Pumps)
3. Vacuum Level Read-out (0-30 in of Hg)
4. Pneumatic Footswitch
5. Compressed Air (Return/In and Out)
5
3
2
System Set-Up
4
1. REMOVE ¾” Button Head Screws (8 each) from unit, and assemble rails to unit. a. Insert Flat Head
Screws (8 each), and t ighten screws with enclosed Hex
L-Key (See detailed view on page 19). b. Attach top rail c. While holding top rail, secure sides with d. Using Male Knobs (2 each), tighten IV pole mount
the Female Knobs (2each). to back of railing, and then insert IV bag holder.
21-OM-0001 5 of 21 Rev D
Figure 1
2. With unit in “OFF” position, attach pneumatic footswitch and power cord to back of unit, and then plug-in power cord to electrical outlet. (Figure 1)
Pneumatic
Footswitch
Power
Cord
3. Place canister bracket and dual-canister holder with canisters on rail. (Figure 2)
Figure 2
5. Attach disposable filter to connection port turning clock-wise until hand tight. (Figure 4)
6. Canister Set-Up, see page 5.
4. Insert IP Canisters and
finger tips push down
on liners. (Figure 3)
NOTE – Vacuum will not occur in the IP Canister system without BOTH liners installed.
Figure 4
Connection Port
Figure 3
7. Remove white cap by turning counter-clockwise .
Replace cap by placing it over the (AIR-OUT) port, then turn cap clockwise .
Operating Instructions
1. Switch “ON” aspirator vacuum and compressed air pumps , unit is ready for operation. Depressing the footswitch will activate or deactivate aspirator vacuum (suction) system.
NOTE – See INSTRUCTION MANUAL of applicable unit for proper connection and operation of pneumatic components with the PSI-TEC Systems.
21-OM-0001 6 of 21 Rev D
Canister Set-Up Illustration
Step 1 Attach disposable filter to connection port turning clock-wise until hand tight.
Step 2 Connect filter tubing from disposable filter to port A on CANISTER 2.
Step 3 Connect aspirator tubing from port B on CANISTER 2 to port C on CANISTER 1.
Step 4 Connect canister tubing E on CANISTER 2 to port D on CANISTER 1.
Step 5 Connect other canister tubing G on CANISTER 1 to port F on CANISTER 2.
Step 6 Using sterile technique, open tubing package (PT-5558 PSI-TEC
); connect one end to port H on CANISTER 1, and other end to the desired cannula instrumentation.
Disposable Filter
F
B
A
E
G
D
C
H
Canister 2
Canister 1
21-OM-0001 7 of 21 Rev D
CAUTION – STOP surgical aspiration PRIOR to removing liners from canisters.
WARNING – Single canister usage NOT recommended, as there is a greater probability of filter/ pump contamination.
2. (To CHANGE Filled Liner): a. Switch aspirator vacuum [ ] “OFF” when liner in CANISTER 1 is full (1500 - 1700cc). b. Disconnect aspiration tubing from port A on CANISTER 1. c. Disconnect canister tubing from port B on CANISTER
2, and then connect it to port A on CANISTER 1.
Capping the port prevents waste overflow of liner.
Canister LINER Removal
A
d. Remove liner, dispose liner and tubing according to your guidelines for HAZARDOUS WASTE
DISPOSAL.
B
6. (To REINSTALL New Liner): a. Place new liner in CANISTER 1, and with finger tips push down on lid of liner.
Canister 2
Canister 1
b. Connect loose end of canister tubing from CANISTER
1 to port B on CANISTER 2. c. Reconnect aspiration tubing to port A on CANISTER 1, switch “ON” aspirator vacuum.
III. Maintenance
Cleaning Instructions
The unit exterior should be wiped down with a damp soft cloth and germicide. DO NOT use cleaning solutions containing ALCOHOL or SOLVENTS.
WARNING – DO NOT IMMERSE IN LIQUID. STORE COVERED AND IN A CLEAN DRY
PLACE. CONTACT PROPER AUTHORIZED PERSONNEL FOR SERVICING.
DO NOT autoclave or ETO sterilize aspirator, canisters, liners, or single-use items.
21-OM-0001 8 of 21 Rev D
IV.
Troubleshooting
Problems, Causes, and Solutions
No vacuum
Vacuum Push Vacuum Switch [ ] “ON”.
Footswitch will not control aspirator.
Footswitch not connected to unit.
Connect footswitch to unit (rear).
System will not turn “ON”. Power cord not plugged-in to electrical outlet.
Plug-in power cord to electrical outlet.
Fuses blown on back of unit tripped located just above
Unplug power cord, replace blown fuse, reconnect power power cord (female connector) cord, switch on vacuum, and depress footswitch.
Low vacuum.
Footswitch not activated. Depress Footswitch.
System not receiving power. Plug Power Cord into electrical source.
IP Canister Liner(s) not properly sealed.
With vacuum [ ] “ON’ gently push down on lid.
Tubing improperly connected. Check all tubing at connection points for proper seal.
Aspirator stalls after quick restart.
No Compression
Aspiration pumps still under load (vacuum).
Release vacuum by venting with air (remove cannula from patient and restart).
System not receiving power. Plug Power Cord into electrical source
Ensure pneumatic component is inoperative. aspiration compressed air [ ] fittings. visible damage to connectors and tubing.
If still inoperative, Contact Byron Medical, (800) 777-3434 or (520) 573-0857 for technical support.
21-OM-0001 9 of 21 Rev D
APPENDIX - A
PSI-TEC III Aspiration Platform Specifications
PSI-TEC III
ASPIRATION PLATFORM 115V
PSI-TEC III
ASPIRATION PLATFORM 230V
Size
20.18 in (H) x 22.96 in (D) x 18.32 in (W)
51.26 cm (H)x 58.32 cm (D) x 46.53 cm (W)
20.18 in (H) x 22.96 in (D) x 18.32 in (W)
51.26 cm (H) x 58.32 cm (D) x 46.53 cm (W)
Weight
Power Cord
AC Power Input
Fuse
Maximum Vacuum
Maximum Flow Rate
Sound Level
Pump Type(s)
Maximum Compressor
Air Pressure
Internal Pressure
80 lbs. / 36.288 Kilograms
10 ft / 3 m Hospital Grade IEC
115 VAC, 60 Hz, 7.5 A
T 15 A 250 V
28 in of Hg. / 711 mm of Hg
4.75 CFM (Cubic feet per minute)
52 dB Typical
Two cylinder Piston/Single Piston
29 inHg
736.6 mmHg
100 psi MAX compressor output
Regulated to 60 PSI
250 psi Minimum
(250% Max Compressor Capacity)
Class 1 Type B
80 lbs. / 36.288 Kilograms
10 ft / 3 m Hospital Grade IEC
230 VAC, 50 Hz, 3.1 A
T 6.3 A 250V
28 in of Hg. / 711 mm of Hg
4.75 CFM (Cubic feet per minute)
52 dB Typical
Two cylinder Piston/Single Piston
29 inHg
736.6 mmHg
100 psi MAX compressor output
Regulated to 60 PSI
250 psi Minimum
(250% Max Compressor Capacity)
Class 1 Type B
Tank Capacity
IEC Classification
Environmental
Operating Temperature
Storage/Shipping
Temperature
Operating Humidity
10 deg to 40 deg C
54 deg to 104 deg F
-30 deg to 50 deg C
-22 deg to 122 deg F
30% to 75% relative humidity
(non-condensing)
10 deg to 40 deg C
54 deg to 104 deg F
-30 deg to 50 deg C
-22 deg to 122 deg F
30% to 75% relative humidity
(non-condensing)
21-OM-0001 10 of 21 Rev D
APPENDIX - B
PSI-TEC III 115V Platform
MODEL: PT-PLT-III-110
PARTS LIST
CMI NBR DESCRIPTION
2196 TIE, CABLE 4"
2520 SCREW, 4-40 X ¼ “ BUTTON HD MACHINE SLOTTED
2552 SCREW, 6-32 X ½ " PHILLIPS PAN HEAD MACHINE
2605 HEX NUT, 6-32 X ¼ " WIDTH
2620 HEX NUT, 8-32 X
2625
11
HEX NUT, 10-32 X
⁄
3 ⁄
32
8
"
" WIDTH
2665 LOCK WASHER, #6 INTERNAL TOOTH
2675 LOCK WASHER, #8 EXTERNAL TOOTH
2690 LOCKWASHER, 10 INTERNAL TOOTH
2695 LOCK WASHER, #10 SPLIT
6112 ADHESIVE
10176 FAN, 4” 110V MUFFIN
⁄
16
" SHRINK BLACK
17217 LOCK WASHER, #6 SPLIT
19771 PIPE JOINT COMPOUND
19861 POLY BAG, 6” X 9”
19974 POWER CORD, 125V 10A TO IEC
30400 BAG, CLEAR Z-LOCK 0.004M 12” X12”
32029 RECEPTACLE, 3 PIN
32030 PLUG, 3 PIN
32031 MALE TERMINAL PIN
32345 COMPRESSOR SIDE MTG BRKT(LH/RH)
32346 COMPRESSOR BASE PLATE PSI-TEC III
32347 VACUUM PUMP MTG PLATE (RH+LH)
32348 VACUUM PUMP MOTOR MOUNT
32351 AIR RECEIVER END CAP
32352 AIR RECEIVER TIE ROD
32353 AIR RECEIVER CYLINDER WALL
32357 RADIATOR BRACKET (RH + LH)
32361 ½ "-13 HEX NUT ZINC
32362 ¼ " LOCKWASHER ZINC
32364 3 ⁄
8
" LOCKWASHER ZINC
32365 ½ " LOCKWASHER ZINC
32367 ⅜ “ – 16 X ½ “ BUTTON HEAD SKT CAP
⁄
8
"
32388 ¼ " X
1
⁄
8
" ADAPTOR
32393 GAST VACUUM PUMP
UOM
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA 4
OZ 0.20
EA
IN
1
12
EA
OZ
EA
EA
2
0.60
1
1
EA
EA
EA
EA
1
4
4
20
EA 20
EA 2
EA
EA
1
1
4
1
2
4
1
2
1
2
26
4
4
4
1
Qty
27
1
6
2
4
10
5
9
10
3
3
21-OM-0001 11 of 21 Rev D
PSI-TEC III 115V Platform (Continued)
MODEL: PT-PLT-III-110
PARTS LIST
CMI NBR DESCRIPTION
32394 GAST COMPRESSOR PSI-TEC III
32395 AIR LOGIC PRESSURE SWITCH
32396 RADIATOR 6” X 6”
32397 120 VAC - HUMPHREY ¼ " VALVE
32399 ¼ " NPT FEMALE CHECK VALVE
32411 SWITCH, AIR ACTIVATED
32413 MOLEX 12 PIN FEMALE
32414 SHELL 12 PIN MALE
32415 CASTER, 4" PLATE MOUNT
32416 CASTER,LOCKING 4" PLATE MOUNT
32417 PSI-TEC III U SHAPED BAR
32420 IV POLE MOUNT
⁄
8
“- ¼ “ MALE ELBOW
32432 SCREW, 10-32 X ½ " BUTTON HEAD CAP
32434 ¼ " BRONZE MUFFLER
32435 MALE ¼ " - MALE ¼ " ELBOW
32437 MALE ¼ " - FEMALE
1
⁄
8
" ADAPTOR
32438 ¼ " - ½ " MALE ELBOW
32439
1
⁄
8
" - 1 ⁄
16
" RIGHT ANGLE ADAPTOR
32442 2 FOOT. FLEX HOSE WITH ½ " FEMALE
32446 BASE PLATE, PSI-TEC III
32448 SWITCH, 16 AMP LIGHTED ROCKER
32449 CONN MOLEX BLADE CRIMP
32450 TEE ¼ “ – ⅜ “ COMPRESSION
32451
32453
3
3
⁄
8
⁄
8
" STAINLESS STEEL TUBE 41” X 6
" COMP UNION
32454 ¼ " - ¼ " MALE TO MALE COUPLER
32455 RELAY MTG BLOCK WITH RETAINER
32456 TERMINAL,FORKED CRIMP #8
32457 GREEN/YELLOW 16 GAUGE WIRE
32458 BLUE 16 GAUGE WIRE
32459 BROWN 16 GAUGE WIRE
32465 VACUUM GAUGE 2 ½ " DIAMETER
32466
1
⁄
8
“ ID TUBING
32468 AIR OUTLET CONNECTOR
1
1
1
2
2
16
1
1
2
19
1
1
1
1
6.25
19.17
32.75
1
3
1
35
1
1
1
1
1
11
1
1
1
2
2
4
8
2
1
Qty
1
1
1
1
1
1
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
FT
FT
FT
EA
EA
EA
IN
EA
UOM
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
FT
EA
EA
21-OM-0001 12 of 21 Rev D
PSI-TEC III 115V Platform (Continued)
MODEL: PT-PLT-III-110
PARTS LIST
CMI NBR DESCRIPTION
32471 ½ " COMPRESSION TO ¼ " NPT
32472 ¼ "- ¼ " BULK HEAD ADAPTOR
32477 POWER ENTRY MODULE 110-220V 10A
32478 SCREW, ¼ "-20 X ¾ " SOCKET HEAD CAP
32481 FUSE, 15AMP FAST ACTING
32483 SCREW, ¼ "-20 X ½ " SOCKET HEAD CAP
32485 SCREW, 8-32 X 3
32486 O-RING
⁄
8
" PAN HEAD PHILLIPS
16
"-18 X 1" BUTTON HEAD
16
" ZINC
16
"-18 ZINC
32490 ADAPTOR, GAUGE VACUUM
32491 PLUG,
32493
1
⁄
8
" HEX PIPE
8
" QUICK RELEASE TO 3
32494
1
⁄
8
" QUICK RELEASE TO 10-32
⁄
8
⁄
8
" NPT
" ID X ¼ " OD SUPER THIN
32497 CLIP,RELAY RETENTION FOR #32404
32502 ATO MUFFLER, ¼ “ MALE
32507 FILTER, DUAL STAGE
UOM
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
FT
EA
EA
EA
32527 MACHINE SCREW, 8-32 X
32541 FOOTSWITCH
32542 1 ¼ " FLUTED PL KNOB
32543 1 ¼ " KNOB WITH STUD
⁄
8
" TRUS
32546 SCREW, 8-32 X ¾ “FLATHEAD
32555 CONNECTOR, 16 AWG 1.187 BLADE CRIMP
32556 COUPLER, “ TUBING TO
1
⁄
8
“ BARB
32566 LUG, SOLDER #6 FLAT LOCKING
32567 TUBING, “ SPLIT LOOM
5
32588 PSI-TEC III SIDE RAIL
32589 PSI-TEC III BACK RAIL
32591 5 ⁄
32
" HEX WRENCH, SHORT
32595 3” X 4” PLASTIC BAG
EA
EA
EA
EA
EA
EA
EA
EA
FT
EA
EA
EA
EA
EA
32599 SAFETY RELIEF VALVE, 100 PSI ¼ “ NPT
32600 AIR OUTLET SAFETY COVER
32602 CHAIN, #6 BALL STAINLESS
32603 CLIP, #6 BALL CHAIN STAINLESS
EA
EA
FT
EA
1
1
0.25
2
32605 TEE, ¼ “ NPT TO DUAL ¼ “ NPT FEMALE EA 1
32606
SCREW, 10-32 X
5 ⁄
8
“
BUTTON HEAD MACHINE SLOTTED
EA 2
32607 FUSE DRAWER, POWER INLET EA 1
32614 LOCKNUT, ¼ “-20 EA 12
3
1
2
1
1
1
4
2
1
1
3
1
2
3
1
1
3.25
2
1
1
8
4
1
2
28
2
2
4
Qty
1
2
1
1
6
2
21-OM-0001 13 of 21 Rev D
PSI-TEC III 115V Platform (Continued)
MODEL: PT-PLT-III-110
PARTS LIST
CMI NBR
32615 LOCKNUT, “-16
32616 LOCKNUT, ½ “-13
DESCRIPTION
32628 TY-RAP, 30 NYLON
32644 ¼ “-20 X 1” FLATHEAD COUNTER SINK
32666 PSI-TEC III SHIPPING CONTAINER
32680 ENCLOSURE, PLATFORM ALUMINUM
32694
SCREW, 6-32 X
5
⁄
8
“ FLAT HEAD MACHINE SCREW PHILLIPS
32695 SCREW, 8-32 X ½ ” PAN HEAD WITH WASHER
32696 SCREW,
⁄
16
32710 5 ⁄
16
16
“-14 X 1 ½ ” STAINLESS
32697 LOCKWASHER,
16
“ SPLIT
“ ID RUBBER
“- ¼ “ QUICK CONNECTOR COUPLING
32714 PSI-TEC III BACK PANEL
UOM
EA
EA
EA
EA
EA
EA
Qty
4
4
1
8
1
1
OZ 0.50
EA 2
EA
EA
10
1
EA
FT
EA
EA
1
0.10
1
1
21-OM-0001 14 of 21 Rev D
APPENDIX – B (Continued)
PSI-TEC III 230V Platform
MODEL: PT-PLT-III-220
PARTS LIST
CMI NBR DESCRIPTION
2196 TIE, CABLE 4"
2520 SCREW, 4-40 X ¼ “ BUTTON HD MACHINE SLOTTED
2552 SCREW, 6-32 X ½ " PHILLIPS PAN HEAD MACHINE
2605 HEX NUT, 6-32 X ¼ " WIDTH
2620 HEX NUT, 8-32 X
2625
11
HEX NUT, 10-32 X
⁄
3 ⁄
32
8
"
" WIDTH
2665 LOCK WASHER, #6 INTERNAL TOOTH
2675 LOCK WASHER, #8 EXTERNAL TOOTH
2690 LOCKWASHER, 10 INTERNAL TOOTH
2695 LOCK WASHER, #10 SPLIT
6112 ADHESIVE
⁄
16
" SHRINK BLACK
17217 LOCK WASHER, #6 SPLIT
19771 PIPE JOINT COMPOUND
19861 POLY BAG, 6” X 9”
30400 BAG, CLEAR Z-LOCK 0.004M 12” X12”
32029 RECEPTACLE, 3 PIN
32030 PLUG, 3 PIN
32031 MALE TERMINAL PIN
32345 COMPRESSOR SIDE MTG BRKT(LH/RH)
32346 COMPRESSOR BASE PLATE PSI-TEC III
32347 VACUUM PUMP MTG PLATE (RH+LH)
32348 VACUUM PUMP MOTOR MOUNT
32351 AIR RECEIVER END CAP
32352 AIR RECEIVER TIE ROD
32353 AIR RECEIVER CYLINDER WALL
32357 RADIATOR BRACKET (RH + LH)
32361 ½ "-13 HEX NUT ZINC
32362 ¼ " LOCKWASHER ZINC
32364 3 ⁄
8
" LOCKWASHER ZINC
32365 ½ " LOCKWASHER ZINC
32367 ⅜ “ – 16 X ½ “ BUTTON HEAD SKT CAP
⁄
8
"
32388 ¼ " X
1
⁄
8
" ADAPTOR
32395 AIR LOGIC PRESSURE SWITCH
32396 RADIATOR 6” X 6”
32397 120 VAC - HUMPHREY ¼ " VALVE
UOM
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA 4
OZ 0.20
IN
EA
12
2
OZ
EA
EA
EA
0.60
1
1
4
EA
EA
4
20
EA 20
EA 2
EA
EA
1
2
EA
EA
1
1
2
4
26
4
1
2
4
1
4
4
1
1
1
Qty
27
1
6
2
4
10
5
9
10
3
3
21-OM-0001 15 of 21 Rev D
PSI-TEC III 230V Platform(Continued)
MODEL: PT-PLT-III-220
PARTS LIST
CMI NBR DESCRIPTION
32399 ¼ " NPT FEMALE CHECK VALVE
32411 SWITCH, AIR ACTIVATED
32413 MOLEX 12 PIN FEMALE
32414 SHELL 12 PIN MALE
32415 CASTER, 4" PLATE MOUNT
32416 CASTER,LOCKING 4" PLATE MOUNT
32417 PSI-TEC III U SHAPED BAR
32420 IV POLE MOUNT
⁄
8
“- ¼ “ MALE ELBOW
32432 SCREW, 10-32 X ½ " BUTTON HEAD CAP
32434 ¼ " BRONZE MUFFLER
32435 MALE ¼ " - MALE ¼ " ELBOW
32437 MALE ¼ " - FEMALE
1
⁄
8
" ADAPTOR
32438 ¼ " - ½ " MALE ELBOW
32439
1
⁄
8
" - 1 ⁄
16
" RIGHT ANGLE ADAPTOR
32442 2 FOOT. FLEX HOSE WITH ½ " FEMALE
32446 BASE PLATE, PSI-TEC III
32448 SWITCH, 16 AMP LIGHTED ROCKER
32449 CONN MOLEX BLADE CRIMP
32450 TEE ¼ “ – ⅜ “ COMPRESSION
32451
32453
3
3
⁄
8
⁄
8
" STAINLESS STEEL TUBE 41” X 6
" COMP UNION
32454 ¼ " - ¼ " MALE TO MALE COUPLER
32455 RELAY MTG BLOCK WITH RETAINER
32456 TERMINAL,FORKED CRIMP #8
32457 GREEN/YELLOW 16 GAUGE WIRE
32458 BLUE 16 GAUGE WIRE
32459 BROWN 16 GAUGE WIRE
32465 VACUUM GAUGE 2 ½ " DIAMETER
32466
1
⁄
8
“ ID TUBING
32468 AIR OUTLET CONNECTOR
32471 ½ " COMPRESSION TO ¼ " NPT
32472 ¼ "- ¼ " BULK HEAD ADAPTOR
32477 POWER ENTRY MODULE 110-220V 10A
32478 SCREW, ¼ "-20 X ¾ " SOCKET HEAD CAP
16
6.25
19.17
32.75
1
3
1
1
2
2
1
2
19
1
1
35
1
1
2
1
1
6
1
1
1
1
1
1
1
2
2
1
1
1
1
11
Qty
1
1
4
8
1
1
EA
EA
EA
EA
EA
EA
EA
EA
EA
FT
FT
FT
EA
EA
EA
IN
EA
EA
EA
EA
EA
EA
UOM
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
FT
EA
21-OM-0001 16 of 21 Rev D
PSI-TEC III 230V Platform(Continued)
MODEL: PT-PLT-III-220
PARTS LIST
CMI NBR DESCRIPTION
32483 SCREW, ¼ "-20 X ½ " SOCKET HEAD CAP
32485 SCREW, 8-32 X
32486 O-RING
3 ⁄
8
" PAN HEAD PHILLIPS
16
"-18 X 1" BUTTON HEAD
16
" ZINC
16
"-18 ZINC
32490 ADAPTOR, GAUGE VACUUM
⁄
8
" HEX PIPE
32493
32494
1
⁄
8
" QUICK RELEASE TO 3 ⁄
8
" NPT
1
⁄
8
" QUICK RELEASE TO 10-32
⁄
8
" ID X ¼ " OD SUPER THIN
32497 CLIP,RELAY RETENTION FOR #32404
32498 VALVE, 320 ¼ “ FOR 240 VAC
⁄
8
“ FOR 220 VAC
32502 ATO MUFFLER, ¼ “ MALE
32504 COMPRESSOR, 230 VOLT GAST
32505 PUMP, 230 VOLT VACUUM GAST
32506 FAN, 4” 220/230V MUFFIN
32507 FILTER, DUAL STAGE
32527 MACHINE SCREW, 8-32 X
32541 FOOTSWITCH
32542 1 ¼ " FLUTED PL KNOB
32543 1 ¼ " KNOB WITH STUD
32546 SCREW, 8-32 X ¾ “FLATHEAD
32555 CONNECTOR, 16 AWG 1.187 BLADE CRIMP
32556 COUPLER, “ TUBING TO
32566 LUG, SOLDER #6 FLAT LOCKING
32567 TUBING, “ SPLIT LOOM
5 ⁄
32588 PSI-TEC III SIDE RAIL
32589 PSI-TEC III BACK RAIL
32591 5 ⁄
32
" HEX WRENCH, SHORT
32595 3” X 4” PLASTIC BAG
8
1
⁄
" TRUS
8
“ BARB
EA
EA
EA
EA
FT
EA
EA
EA
UOM
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
EA
FT
EA
EA
EA
EA
EA
1
2
1
1
3.25
2
1
1
2
1
1
1
1
1
Qty
28
2
2
4
8
4
3
1
2
1
1
1
4
2
1
1
3
1
2
3
32596 POWER CORD, 250V 10A EUROPEAN
32599 SAFETY RELIEF VALVE, 100 PSI ¼ “ NPT
32600 AIR OUTLET SAFETY COVER
32602 CHAIN, #6 BALL STAINLESS
EA
EA
EA
FT
1
1
1
0.25
32603 CLIP, #6 BALL CHAIN STAINLESS
32605 TEE, ¼ “ NPT TO DUAL ¼ “ NPT FEMALE
EA
EA
2
1
32606
SCREW, 10-32 X
5 ⁄
8
“
BUTTON HEAD MACHINE SLOTTED
EA 2
32607 FUSE DRAWER, POWER INLET EA 1
21-OM-0001 17 of 21 Rev D
PSI-TEC III 230V Platform(Continued)
MODEL: PT-PLT-III-220
PARTS LIST
CMI NBR
32614 LOCKNUT, ¼ “-20
32615 LOCKNUT, “-16
32616 LOCKNUT, ½ “-13
DESCRIPTION
32625 FUSE 6.3A 250V SLOW-BLOW GLOW
32628 TY-RAP, 30 NYLON
32644 ¼ “-20 X 1” FLATHEAD COUNTER SINK
32666 PSI-TEC III SHIPPING CONTAINER
32680 ENCLOSURE, PLATFORM ALUMINUM
32694
SCREW, 6-32 X
5
⁄
8
“ FLAT HEAD MACHINE SCREW PHILLIPS
32695 SCREW, 8-32 X ½ ” PAN HEAD WITH WASHER
32696 SCREW,
⁄
16
32710 5 ⁄
16
16
“-14 X 1 ½ ” STAINLESS
32697 LOCKWASHER,
16
“ SPLIT
“ ID RUBBER
“- ¼ “ QUICK CONNECTOR COUPLING
32714 PSI-TEC III BACK PANEL
UOM
EA
EA
EA
EA
EA
EA
Qty
12
4
4
2
1
8
EA
EA
1
1
OZ 0.50
EA 2
EA
EA
EA
FT
EA
EA
10
1
1
0.10
1
1
21-OM-0001 18 of 21 Rev D
APPENDIX - C
When Attaching Rails To PSITEC III
Replace The 3/4” Button Head Screws
With The 1” Flat Head Screw
1” Flat Head Screw (8)
D
APPENDIX - D
21-OM-0001 20 of 21 Rev D
APPENDIX – D (Continued)
21-OM-0001 21 of 21 Rev D
Byron Medical
602 West Rillito Street
Tucson, Arizona 85705
800-777-3434
Printed in the USA
advertisement
Key Features
- High flow vacuum pump
- Powerful vacuum pump
- Pneumatically operated footswitch
- Integrated cart
- Compact and mobile
- Attachments located at working level