Gel documentation system with computer and UPS • Gel

Gel documentation system with computer and UPS • Gel
1.
Gel documentation system with computer and UPS
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
Gel documentation system
Imaging: by scientific grade CCD camera, at least 1.4 mega pixels, 16bit
Dynamic range: >3 orders of magnitude
Illumination Area: at least 20 X 25cm
Lens: Motorized
Wheel: Manual/Motorized at least 3 position filter wheel.
Filters: at least 2 No. (Green/Red/Orange with further provision of up-gradation)
Exposure: Software controlled motorized control of aperture, zoom and focus
Transilluminator: Sliding/drawer/pull out type
Transilluminator power: software controlled
UV source: dual wavelength (254, 302 - 365 nm)
White light source: White light screen
Epi-illumination: Epi UV, epi-white and visible-blue
Safety: UV safety interlock
Applications: Suitable for EtBr, SYBR Green, GFP, SYBR Gold, Fluorescein, CY3, Rhodamine,
SYPRO, Texas Red, Hoechst, Cumarin
Software: Acquisition and Analysis Software for gels
Computer: Computer system complete with UPS (Branded computer with i7 processor, 4GB
RAM, 500 GB HDD, 19” TFT Monitor, DVD R/W, USB ports, Keyboard, optical mouse and
Compatible Windows -license)
Printer: Laser colour printer
Up-gradability: Chemiluminescence up-gradable
Operation: Electrical, 220-240V/50-60 Hz
Warranty: Two year from the date of installation
On line UPS of 3 KV with 1 hour backup
2.
Real time PCR
1.
1.1
1.2
Functional Requirements
The system is intended for real-time PCR (Polymerase Chain Reaction) detection and conventional
end point PCR analysis.
The system shall be able to run at least 6 different PCR protocols (up to 12 samples under each
protocol) at the same time.
2.
2.1
a.
b.
c.
d.
General Requirements
The system shall be stand-alone functionality and consist of the following:
Peltier-based thermal cycling system with 96-well optical blocks;
6.5" LCD touch screen and USB drive for easy set up and analysis, or
Windows based operating system for instrument control and data analysis; and
Software for instrument control, data collection and auto-analysis of sample files.
2.2
a.
b.
c.
d.
e.
f.
The system shall be able to support the following real-time and post quantitative PCR application:
Viral Load Analysis (HIV Viral Load Test);
SNP Genptyping;
MicroRNA Expression;
Translocation Analysis;
Gene Expression Profiling ;Viral Load Analysis;
Gene Detection.
2.3
a.
b.
c.
d.
The system shall designed to analyse 96 samples simultaneously in any of the following formats:
96-well plate with optical adhesive cover;
96-well plate with optical flat caps;
8-tube strips with optical flat caps;or
Individual tubes with optical flat caps.
2.4
The system shall be equipped with suitable light source and filter systems to enable multiple
wavelength detection, at least 5 channels with minimum sample and reagent usage.
2.5 The system's fluorescence detection chemistries shall include factory calibrated FAM/SYBR Green
I, HEX/JOE/VIC, TET, Cy5,Cy3,Alexa Fluor 350,NED/TAMRA and Texas Red/ROX dyes. The system's
melt curve or dissociation takes place in steps varying from 0.1 C/sec to 1 C/sec
2.6 The system shall be designed to complete a standard 40 cycle real-time PCR reaction using
fluorogenic 5' nuclease assay in a 96-well plate configuration in under 30 minutes
2.7 The system should allow the user to program different PCR programs running simultaneously, and
maximum temperature difference across the block should be more than 5 C.
2.8 The system shall have real-time quantitative PCR installation specifications which demonstrate the
ability to distinguish between 5,000 and 10,000 template copies with a 99.7% confidence level.
2.9 The system shall be designed to support reaction volumes of 10l to 50l.
2.10 The system shall provide specialised application specific software that is designed to collect and
analyze the fluorescence data for the applications of absolute quantitation, relative quantitation,
allelic discrimination/SNP(Single Nucleotide Polymorphism) detection and melt curve analysis.
2.11 The system software shall utilize a multi componenting algorithm designed to provide precise
deconvolution of multiple dye signals to enable the simultaneous detection of multiple
fluorophores.
2.12 The system shall be allowed to print with an external printer for hard copy output of results,
usage, quality control and calibration data.
2.13 The system shall allow to save, copy,export chart and text report data directly from display screen
to Microsoft Excel, Power Point, text and image files.
2.14 The system software shall be able to link up to four PCR system with one computer.
2.15 The system shall be well constructed with durable materials to withstand typical abuse and
cleaning.
2.16 The system shall be equipped with automatic self-diagnostic program upon start-up, which shall
detect and clearly indicate any defects or malfunction.
2.17 The system shall have connections that are secure enough to resist accidental disconnection.
2.18 The system shall be designed for easy installation, no required of calibration of optics and easy
access to serviceable parts.
2.19 The system shall be easy to clean and disinfect as appropriate.
2.20 The Tenderer shall provides comprehensive assay design and development guidelines for real-time
quantitative PCR and SNP genotyping assays
2.21 The Tenderer shall be able to supply all the necessary consumables to perform real-time PCR
diagnosis and quantitative PCR, etc.
The system shall be designed to permit future extension or upgrades, both hardware and
software, to more advanced system and capabilities.
Support reaction volumes
Temperature range
Temperature accuracy of set-point
Average ramp rate
Gradient operational range
Scan time, all channels
Scan time, single channel
Quantitative PCR Run Time
Sample Format
Block temperature
Type of detection
Excitation range
Emission range
:
:
:
:
:
:
:
:
:
:
:
:
:
1 to 50 l
0 –100 C
0.2 C at 90C
3.3 C/sec
30 –100 C
12 sec
3 sec
Fast <30 mins
Standard 96-well plates;
5C
Optical
450 to 750nm
450 to 750 nm
Standard Accessories
All standard accessories, required for normal operation, decontamination and maintenance, shall
be listed with itemised prices and included in the unit base price. Please submit list of all
consumables and their prices that used in analysing DNA and RNA samples.
•
•
•
•
•
•
•
Optional Accessories /System
All optional accessories/system shall be listed with itemised prices. Computer compatible with
software Intel core i5 processor 3.10 GHz,
1 TB SATA Hard Disk Drive,
4 GB DDR III RAM with Graphics Card,
Original Windows 7 OS
DVD R/W, 6 USB Ports (with at least 2 in front),
Optical Mouse,
Multimedia Key Board,
21" Bright view LED TFT Color Monitor and Speakers
Installation / Commissioning Requirements- Should be done with free of cost
3.
UV Visible spectrometer
1. Spectrophotometer with accessories (01 Unit)
1. Full spectrum 220-800 nm.
2. Measure 1 µl sample with high accuracy for nucleic acid, protein, microbial cell culture OD and
other general UV/VIS analysis
3. Computer compatible with software.
4. High absorbance range
5. Absorption precision should be 0.003.
6. Wavelength accuracy should be 1nm
7. 10 second measurement time
8. Large range of detection with detection of nucleic acid down to 2ng/microliter
9. Three years comprehensive warranty
10. Computer compatible with software
Intel core i5 processor 3.10 GHz, • 1 TB SATA Hard Disk Drive, • 4 GB DDR III RAM with Graphics
Card, • Original Windows 7 OS • DVD R/W, 6 USB Ports (with at least 2 in front), • Optical
Mouse, • Multimedia Key Board, • 21" Bright view LED TFT Color Monitor and Speakers Colour
laser printer (quote separately) • Equipment should work even without computer
Original illustrated catalogue for support of technical specifications/features List of users along
with phone & e-mail
Suitable online UPS (1 KVA)
4.
Trinocular microscope with infinity corrected optical system
1.
Mechanical rotatable stage for 2 slides with left and right hand control, course (15mm) and fine
adjustment (0.1mm) /rotation, minimum reading 1.0mm.
2. Light source 6V/30W halogen bulb, 220 - 240V, 50Hz.
3. Sextuple revolving nose piece.
4. Eye pieces with interpupillary distance adjustment, wide field 10x (Field view 22mm)
5.
P
lan achromatic objectives 4x (NA 0.1mm), 10x (NA 0.25), 20x (NA 0.40), Plan fluorite
40 x (NA 0.75), Plan fluorite 100x (NA 1.3).
6. Swing out Condenser (1.25 NA), along with phase contrast for 10x to 100x, iris diaphragm built in.
7. Dedicated digital CCD camera of minimum 5MP, binning option and approximate 20 FPS,
camera mount, good image resolution,
8. Camera software for multicolour imaging measurements and shading corrections etc.
9. Compatible Latest PC system, TFT Monitor 23” for viewing pictures and imaging.
10. Two year warranty, AMC.
5.
EXTRUDER-EXPANDER SYSTEM
Specification for supply, installation and commissioning of Pet Food Processing Plant (Semiautomatic-Extruder-Expander System), consisting of following main units and features:
• All the units must be compatible to each other as a complete system should be made with food
grade mirror finished stainless steel.
• Suitable for input raw material like: rice, maize and other grains, animal protein, fats etc, and having
followings main units.
1. Milling/Grinding Machine:
(a) Capacity: 100 to 120 kg/hr.
(b) Food grade steel.
(c) Electric prime mover: 15KW or above 3-phase, 50 cycle, motor.
(d) Essential running spares with price.
2. Pulvarizer
(a) Capacity: 120 kg/hr (minimum).
(b) Made of stainless steel (SUS 304/58E grade).
(c) Motor: 4 KW, 3-phase, 50 cycle (minimum).
3. Mixer
(a) Capacity: 140-150 kg/hr.
(b) Motor: 3 KW 3-phase, 50 cycle (minimum).
(c) Drum body and wings/ribbons of stainless steel (minimum).
(d) Essential running spares with price.
4. Screw Conveyor
(a) Capacity: 140-150 kg/hr.
(b) Conveying height: up to 2.5 meter (minimum).
(c) Motor: 1.1 KW, 3-phase, 50 cycle with speed control facility.
(d) Essential running spares with price.
5. Extruder
(a) Double screw extruder having feeding system, extruding system, cutting system, heating system,
transmission system and moderating system.
(b) Extrusion capacity: up to 75-120 kg/hr (minimum).
(c) Extrusion motor: 22 KW, 3-phase, 50 cycle (minimum).
(d) Heater: 7 heaters of 3KW each.
(e) Oil pump motor: 0.37 KW (minimum).
(f) Feeding motor: 1.1 KW (minimum).
(g) Screws should have self cleaning feature.
(h) Speed control through frequency modulation.
(i) Screw (special alloy steel).
(j) Forced lubrication for better performance longer life of machine.
(k) Self cleaning facility once work is done without opening the components.
(l) With dedicated and separate electric control panel for conveyor, feeder, extruder and cutting
system.
(m) Essential running spares with price.
6. Air conveyor
(a) Output capacity: up to 150 kg/hr (minimum).
(b) Motor: 1.1 KW, 3-phase, 50 cycle (minimum).
(c) Conveyor passage: sealed, stainless steel pipe line to avoid leakage.
(d) Air pressure: 1700 Pa (minimum).
(e) Air discharge: 18 M3/min (minimum).
(f) Conveying distance: up to 10 meter (minimum).
(g) Essential running spares with price.
7. Oven
(a) Minimum 3 layer drying oven type.
(b) Heating load: 25 KW (minimum).
(c) Blower motor load: 0.75 KW (minimum).
(d) Drying temperature: up to1500C (minimum).
(f) Output capacity: 150 kg/hr (minimum).
(g) Essential running spares with price.
8. Hoister
(a) Capacity: 150 kg/hr (minimum).
(b) Dimension: 2 x 0.6 x 1.6 meter (minimum).
(c) Power: 0.75 KW (minimum).
(d) Essential running spares with price.
9. Rolling & Flavouring Unit
(a) Capacity: 150 kg/hr (minimum).
(b) Flavour Container capacity: 20 lit (minimum).
(c) Discharge of flavour: up to 0.5 lit/min (minimum).
(d) Power: 105 KW(minimum).
10. Oil Sprayer
(a) Output capacity: up to 150 kg/hr (minimum).
(b) Oil Container capacity: 60 lit (minimum).
(c) Temperature range: Normal to 1000C.
(d) Oil spray rate: up to 6 liter/minute.
(e) Spray motor: 370W (minimum).
(f) Pellet transmission motor: 1.87 KW (minimum).
(g) Heaters.
(h) Essential running spares with price.
11. Packing Unit
(a) Weighing Unit.
(b) Sealing Unit.
12. (a) Electric power panel with electronic controlling unit with built in safety features for safe and
smooth operation of various machine & Units.
(b) Essential running spares with price.
6.
Upright Research Microscope with Digital Camera (1 Unit)
1. Infinity Corrected Optical System for Better Image contrast with feasibility of future Upgradation.
2. An ergonomic stand with suitable high contrast long life LED illumination with illumination
control & Abbe condenser.
3. Trinocular widefield Observation Tube with built-in diopter adjustment in one sleeve (ocular).
4. Eyepieces of at least 10X magnification.
5. Quintuple revolving nosepiece with Plan Achromatic objectives
6. 4X Plan Achromat; 10X Plan Achromat, Phase; 20X Plan Achromat; 40X Plan Achromat, Phase;
100X Plan Achromat, Oil
7. Ceramic coated Mechanical stage with dual slide holder
8. Universal condenser for Brightfield, Phase Contrast & Dark Field Applications.
9. Digital camera with 12.0 Mega Pixel or better resolution, LCD camera monitor complete with
image analysis software for arranging & measuring captured images, live preview, dynamic
image analysis, creating databases & Preparing reports
10. The microscope should be upgradable to Motorization, DIC & 100 watt fluorescence
• Original illustrated catalogue for support of technical specifications/features
• List of users along with phone & e-mail
7.
Bio Safety Cabinet Class II
Specifications:
1. Model: Self standing Class II
2. Stand: Stand with castor wheels
3. Cabinet Width: 4 ft
4. Workplace: Single piece stainless steel working area
5. Integrated air Filtration System with long life ULPA or HEPA filter for supply and exhaust with
99.9% efficiency
6. Controller: Microprocessor controller panel with LCD display
7. Lamps: Fluorescent and UV lamps
8. Utilities: Modular electric points 2 No., utility valves for gas and vacuum inside the chamber.
9. Noise level: Less than 66dBA.
10. Antimicrobial coating
11. Certification: NSF/EN certified.
12. Compatible UPS with at least 30 minute backup
13. Electrical operation: Single phase upto 220-240 volts/ 50-60 Hz
14. Warranty: Minimum 2 years from the date of installation
Optional:
1. AMC: AMC charges for additional three years must be given separately (under head “Optional”)
Additional:
1. Original illustrated Catalogue in support of technical specifications.
2. User list for the quoted model along with contact details including telephone no. and E. mail.
8.
ELISA Microplate Reader with accessories (automatic washer etc)
with computer attachment, UPS & softwear
•
•
•
•
•
•
•
•
Ability to read flat and U bottom 96 well plate
Wave length range 400-750 nm
Automatic calibration before reading
Plate shaking capability
Build in printer
Computer along with laser printer and software for data analysis, recording and output
Programmable ELISA plate washer with complete accessories.
Warranty: At least two years from the date of installation
9.
Orbital shaker
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
Bench top orbital shaker with temperature control range from 7°C below ambient to 50°C with
accuracy ±0.1°C.
It should have orbital shaking.
Speed should be 50-350 rpm with accuracy of ±2rpm
It should have acceleration/deceleration circuit to prevent sudden starts/stops.
It should have timer up to 99 Hrs.
It should display speed, running time, alarm etc on display screen.
It should have alarms alert when speed or temperature deviate from set points.
The system should have port for data logging.
It should have automatic restart after power interruption.
It should stop agitation when chamber is open.
It should have universal platform and clamps to holds various flask of different sizes.
It should be supplied with compatible stabilizer.
Operational voltage should be 220±20V
Should have two years standard warranty/from the date of installation
Should quote for the price of CMC/AMC per year after the expiry of standard warranty.
Should provide user list with phone no and e mail ID
Should provide brochure in original.
10.
Bedside monitor
1. High resolution TFT monitor
2. It should be Suitable for animals use
3. It should have the facilities for recording ECG, Heart rate, Respiration, Temperature , SPO2 and
4. Non invasive Blood pressure
5. Audible and Visual Alarms with adjustable alarm ranges
6. With attached PC for storage and retrieval of data
7. Graphic display of parameters
8. Option for electrical and battery operation with Rechargeable batteries and 0.5 KVA stabilizer
9. Built in printer
10. Accuracy 90-100 %
11. At least one year warranty
11.
Trolley based machine with drawers
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
Trolley based machine with drawers
Isoflurane and sevoflurane vaporizers
Oxygen Flow Meter
Flow rate of oxygen-0.1 L to 10 L
Oxygen flush up to 30 L or more
Gas pressure gauges
Safety valves
Pop off valve
Reservoir/rebreathing bag of 1 L
Soda Lime canister of 1L or more
Bains apparatus/circuit or latest breathing circuit
Provision for connecting to small and large oxygen cylinders including pin based cylinders
Provision for connecting to positive pressure ventilator
The machine should be supplied with following accessories:
(i) Small Animal Anaesthesia Mask of different sizes (Set of six sizes)
(ii) Rebreathing bag 1 and 2 Litre (2pc.each)
15. Minimum warranty period 2 years
12.
VENTILATOR FOR SMALL ANIMALS
1.
2.
3.
4.
Ventilator for small animals with respiration modes like IPPV, SIMV and manual respiration.
It should have tidal volume capacity of 50-1500 ml or more, for small to large body weight dogs
It should have both high and low pressure alarms
it should have monitors for tidal volume, MV, Peak pressure, PEEP, Breathing rate and I:E ratio
13.
High Speed Refrigerated Centrifuge
1. Bench top centrifuge for low volume and high volume applications.
2. Maximum Speed: RPM not less than 15000 rpm and / or RCF not less than 20,000 g
3. automatic LCD display
4. Two Rotors: for using 24 x 1.5/2.0 ml & 6 x 15.0 / 50 ml
5. Temp. range from –10⁰C to + 40⁰C
6. Automatic Rotor recognition to set maximum speed for safety.
7. Should have environment friendly refrigerant / CFC free.
8. Equipment safety : lid lock and imbalance recognition
9. Power supply requirement : 220-240 volts / 50-60Hz
10. Should be supplied with suitable voltage stabilizer (3KVA).
11. Warranty 2 years from the date of installation.
Additional:
1. Original illustrated Catalogue in support of technical specifications.
2. User list for the quoted model along with contact details including telephone no. and E. mail.
14
ROTARY EVAPORATOR WITH VACUUM AND CHILLER
A.
a)
b)
c)
ROTARY EVAPORATOR:
Rotation speed up to 280 rpm or better with microprocessor control.
Motorized lift with provisions for automatic lifting of the flask in case of power failure.
Heating bath with microprocessor controlled digital temperature display of both set and
Actual temperature simultaneously. Temperature range from ambient to 180°C with an accuracy of
±2°C or better.
d) Infrared communication between water bath and the evaporator for transferring the bath
temperature to the vacuum controller.
e) Automatic over heat cut-off protection with provision for wireless contacts on the heating bath
both for power supply and communication for better safety.
f) Provision for using evaporating flasks of 50-4000 ml (or more) capacity on the same joint adapter
without additional connections.
g) Vertical glass assembly. The supply must include one liter evaporating flask and receiving flask and
woulff bottle should be provided in standard scope of supply.
h) All display parameters like water bath temperature, vacuum etc. should have a separate electronics
control panel.
i) All glass components should be made of Borosilicate 3.3 glass, and cooling surface zone of
condenser should be 1460 cm2 or more.
j) All connections (for vacuum, water circulation etc.) should be at the lower part of the condenser.
B. VACUUM CONTROLLER
a) Modular vacuum controller with control knob and large graphic display screen.
b) Integrated solvent library of commonly used solvents for convenient setting of distillation
conditions.
c) Speed control for vacuum pump: measuring range: 1400 - 0 mbar; control range: 1100 - 1 mbar.
d) Integrated aeration valve and precision pressure sensor.
e) Timer function for process termination after pre-specified time
f) USB interface for data transfer to PC for storage, evaluation and optimization.
g) All controls for the rotary evaporator e.g., lifting /lowering of the evaporating flask, Rotation
start/stop should be controlled via the vacuum controller.
h) Provision for display of chiller temperature and control.
C. VACUUM PUMP
3
a) It should be a single stroke speed control vacuum pump with a flow rate of 1.8 m /h or better with
ultimate vacuum of less than 10 mbar.
b) It should have a two-stage diaphragm made of PTFE with provision for visual inspection of
diaphragms for checking solvent build up and contamination.
c) Provision for soundproofing with enclosed housing to minimize noise and vibrations
CHILLER
Compact and robust re-circulating chiller with a cooling capacity of 500 watts at 15°C: temperature
range: -10°C to 25°C deg C; pump capacity: 3 L/min at 0.6 bar; coolant: CFC-free. All components
like rotavapor, vacuum pump and chillers should be from same manufacturer. User list of at least 5
similar installations in Government institutes should be provided. This is certified that specifications
and requirement given above are need based for carrying out research work of the project and are
not biased to any manufacturer of this equipment
15
CO2 INCUBATOR
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
Microprocessor controlled, direct heat, air jacketed, stackable CO2 incubator.
Capacity : 170 Litres
Temperature control: From 10ºC below ambient to 50ºC.
Peltier control mechanism to bring the temperature below ambient temperature in order to
maintain the set temperatures especially during summer months
Temperature control accuracy : ±0.1ºC
Should be provided with six sided direct heating with fanless, gentle convection circulation
CO2 control range : from 0.2 to 20%
CO2 control accuracy and uniformity : ± 0.1%
Should be provided with infra red (IR) CO2 sensor and equipped with rapid recovery to set % CO2
At least 4 adjustable shelves and a water reservoir, both made up of stainless steel
Seam less stainless steel interior of the chamber
Temperature resistant inner glass door must be provided with the system
Should have independent door heaters
Should provide display of all the set and real time parameters
Alarm function and mal function parameters should be displayed on screen
Should essentially be equipped with a filter CO2 inlet
Equipment should have the facility for security system for over temperature cut off
Automatic re-setting of the pre-set parameters in case of power failure
Two CO2 cylinders with regulators should be essentially supplied with the system
Compatible 3 KVA servo voltage stabilizers must be provided
Power requirement of 220 Volts and 50 Hz
Warranty of 2 years on all parts of equipment should be essentially provided
Should provide minimum 10 satisfactory installation reports from government organizations along
with users list
Original literature/ manuals should be provided to support the technical specification of the model
quoted.
User list of at least 5 similar installations in Government institutes should be provided.
This is certified that specifications and requirement given above are need based for carrying out
research work of the project and are not biased to any manufacturer of this equipment
16
UV VIS Spectrophotometer :
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
Wavelength range: 200-830 nm
Spectral bandwidth: 1 nm or better
Detector: Dual silicon photodiode
Light source: Halogen and deuterium lamp (with automatic switching)/xenon flash lamp
Photometric system: Double beam optics
Photometric range: Absorbance up to 4 Abs and transmittance is 0 % to 100 %
Photometric accuracy: 0.002 or better
Wavelength increment: 1 nm
Wavelength accuracy: + 0.1 nm at D2 peak 656.1 nm and + 0.3 nm for entire range
Wavelength repeatability: + 0.1 nm
Wavelength scanning speed: up to 3000 nm/min or better
Noise level: < 0.00005 A or better
Stray light: <0.02 % or better
Should work in photometric, spectrum, quantitation, kinetics, and time scan mode with features for
DNA and protein quantitation.
15. Operation, data analysis and storage should be through the supplied PC (branded with i5 processor,
genuine windows 7 or better, 500 GB capacity, 8GB RAM, connectivity through USB port, with
appropriate software and a laser printer) or even as Standalone
16. Macro- and micro- cuvettes for measurement in visible and UV range
17. Along with compatible good quality online 3 KVA UPS
18. With minimum warranty of two years
User list of at least 5 similar installations in Government institutes should be provided.
This is certified that specifications and requirement given above are need based for carrying out
research work of the project and are not biased to any manufacturer of this equipment
17
WATER PURIFICATION SYSTEM
Water purification system consisting of separate pre-filtration unit, type II grade water production
system, reservoir and type I grade water production system with following specifications
o
The system should be able to provide pure and ultra pure water from the fed potable tap water
with water conductivity < 2000 μS/cm at 25°C, TOC <2000 ppb, Water Pressure 1 – 6 bar and pH 4
to 10 pH units
•
o
A pre-filtration:
A pre-filtration unit with 5 and 1 micron filter to remove the particulate matter and booster pump
to feed pressure.
•
o
Ist stage system (Analytical Grade Water System) for type II grade water:
Have pre-filter with anti-scaling and activated carbon for the removal of free chlorine and organics,
reverse osmosis for removal of ionic and inorganic impurities, provision of conductivity cell before
and after RO stage to measure the RO feed conductivity, RO efficiency and performance of the
system.
Pump with temperature feedback mechanism to maintain constant product flow rate.
Self regenerating electro-deionization principle (without softening cartridge) with carbon beads on
cathode to prevent scaling of EDI and consistent water quality.
Feed water handling of conductivity up to 2000 microS/cm, free chlorine up to 3 ppm and fouling
Index up to 12 without need of any extra carbon filter.
Provide first stage RO-DI water with a water flow rate of 3 Ltr/hr with inorganic ions and organics
removal up to 99%, resistivity >5 Mohm.cm, TOC <30 ppb.
UV lamp installed in the system to allow reduction of the level of bacteria and TOC
o
o
o
o
o
•
o
o
•
o
o
o
o
o
o
o
Reservoir:
Fifty liter capacity with sensor rod float switch programmed to have high and low level cut off based
on water level in tank.
Reservoir should posses a tank vent filter made of soda lime, activated carbon and Durapore
hydrophobic membrane to prevent contamination of stored water from air borne contaminants.
IInd stage system (Ultra Pure Grade Water System) for Type I Ultrapure water:
Ultra pure water production through two stage mixed bed ion exchange and activated carbon
cartridge.
Type II feed water specific cartridge for removal of trace contaminants.
Provision for recirculation of water to maintain high water quality during periods of non-use
Multi colour monitor for graphic display to show water quality parameters and other critical
information (resistivity, TOC level, alerts, water level in the reservoir, volume dispensed and other
alarms).
Product resistivity cell (cell constant 0.01/cm) should be present.
Facility for volumetric dispensing with minimum increments.
Ultrapure pyrogen/endotoxin free, RNAse and DNAse free water production possessing resistivity
≥18.2 MOhm.cm, conductivity 0.055 µS/cm, Micro organisms <0.1 cfu/ml, flow rate up to 1 Ltr/min
or better, TOC <5 ppb.
o
o
o
o
o
Variable water flow to deliver Type – I water at a flow rate of 2 L/min, controlled by plunger with an
option to have delivery of pure water from a 0.22micron filter
Consumables (One set of all filters, cartridges, sanitization tablets, etc.)
Suitable Voltage stabilizer should be provided with the system.
Stand for mounting the entire system on the wall to save the bench space and all accessories to
connect the system with tap water should be provided by the bidder
Warranty on the machine and all essential accessories for a period of at least two years from the
date of installation
User list of at least 5 similar installations in Government institutes should be provided.
18
AUTOMATIC DIETARY FBRE ANALYZER
Dietary Fiber Extraction System (6 Place): For determining dietary fiber by digestion with
thermostable enzymes followed by extraction (filtration and washing) in accordance with AOAC
enzymatic method 985.29.
A. Enzymatic Digestion system:
To perform enzymatic digestion of samples, and should facilitate the samples to be immersed in a
thermostatic water bath and stirred to ensure excellent thermoregulation and precision.
a)
b)
c)
An immersion heating head: temperature range ambient to 100°C with digital display and digital
timer
A Transparent tank for good visibility
A 6-place magnetic stirrer to ensure excellent thermoregulation through continuous and constant
sample mixing thereby preventing the sample from overheating
B. Dietary fiber extraction system (6 Places):
a) Peristaltic pump with high suction capacity for speed-up filtration
b) Electronic setting of counter pressure to unclog crucibles during filtration
c) Separate residues collection facility
d) Short of filtration time
Complete with compatible one set each of glass crucibles and beakers, voltage stabilizer of appropriate
capacity, and a branded cyclone mill for size reduction (grinding) of hard and soft fibrous feed/faeces
samples to match the prerequisite of sample preparation. All components with electrical compatibility
for 230 V/50 Hz. With minimum warranty of two years.
This is certified that specifications and requirement given above are need based for carrying out
research work of the project and are not biased to any manufacturer of this equipment
19
BIOSAFTY CABINET
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
22.
Self standing Biosafety cabinet Class II A2 type having 70:30 circulation (70% air re-circulation
and 30% exhaust)
Width of the cabinet - 4 feet.
Should be a NSF/ANSI-49 certified Biosafety cabinet (certificate to be provided with bid)
Must maintain Class 5/10 conditions as per ISO norms and should contain 99.99% supply and
exhaust efficient filters.(Either HEPA or ULPA)
Should have a microprocessor based control panel with display of airflow/air velocities in real
time and also display other parameters like UV light on/off, white light on/off, UV and filter usage
hours, remaining life of the filters/alarm for filter change, LPG port on/off etc.
The microprocessor controller must be located at an accessible position so as to help the user to
view display parameters from a seated working position in front of the cabinet
The cabinet should be essentially made of 304 stainless steel interior with curved edges at
corners and epoxy-coated steel exteriors
Must essentially have removable, seamless, perforated work surface with lift out knobs and two
adjustable arm rests made of stainless steel
The area beneath the work surface must be negatively pressurized to ensure that any
contaminations are quickly trapped and drawn in to the filters
Should essentially include a dual blower motor for simultaneously control of both inflow velocity
(100 FPM / 0.5 m/sec) and down flow velocity (80 FPM / 0.3 m/sec)
The cabinet must have smart system to automatically reduce fan/blower motor speed to 30%
when the front window sash is in closed position to ensure reduced energy consumption when the
cabinet is not in use.
Glare-free fluorescent lighting.
UV light fitted at appropriate position for disinfecting the work area of cabinet. UV light on/off
control and lamp usage timer should be present in the control panel
Front side of the cabinet should be made of at least 1/4" thickness tempered safety glass inclined
at 10o and sash opening of less than 30 inches for providing access to work area.
The interior of the front glass must be accessible for cleaning without requiring the user remove
or support the glass window.
Stainless steel cover to fully close the sash present in front side with two counterbalanced sash
handles and locking mechanism
Two electrical receptacles working on 220 volt models located one on each side of cabinet.
One Vacuum port and one LPG port present on the inner side walls of the cabinet.
The cabinet noise level must be less than 65 Db
Suitable online UPS with at least 30 minute backup should be provided essentially with the biosafety hood for facilitating uninterrupted work during power cut
Machine should operate at single phase, 220 volts and 50 Hz
Warranty of at least 2 years should be provided
User list of at least 5 similar installations in Government institutes should be provided.
20
-80oC DEEP FREEZE
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
Vertical (Upright) type deep freeze of Approx 550-570 liters capacity and operating temperature
of -80°C to -86°C.
Should have two compressors working in a cascade for maximum heat removal and quick
recovery to set temperature. Power consumption of <9 KWh/day, pull down time from ambient
to -85ºC should be less than 6 hours.
System should have 304L grade stainless steel interior and tough, powder coated exterior finish
constructed on steel gauge.
Should preferably have two independent lockable doors with five inner compartments.
Freezer should have insulated inner doors giving separate access to individual shelves with
provision for adjusting the height of shelves.
Fully programmable microprocessor controlled eye level display of temperature parameters.
Should have serial communication port for PC interface along with inbuilt systems monitoring
and reporting technology software
Thick insulation on cabinet door and lid to ensure uniform cooling and triple point silicon gasket
on the doors for tight sealing to avoid formation of ice crystals.
Built-in automatic high and low voltage compensator. Compressor should be able to run on any
voltage between 190 – 270V.
Audible and visible security alarms for deviation from set temperature, power failure, probe
failure, system failure, clean filter and door open.
Freezer must fulfill ISO 9001-safety requirements and have IEC 61010 electrical safety CE
certificate (certificate to be provided along with bid).
Freezer must use CFC-FREE, HCFC-FREE non flammable refrigerants, and refrigeration system
must be energy efficient.
One split air conditioner of at least two ton along with compatible stabilizer should be provided
to maintain the ambient room temperature at around 25ºC to maintain the ultra low
temperature (-80 to -85ºC) without overloading the compressor and to avoid over heating of
condensers and compressors in the summer when the ambient temperature shoots to > 40ºC.
A compatible servo voltage stabilizer should be provided with the instrument.
The instrument should operate on a single phase current with a voltage of 220 volts and 50Hz.
Accessories like racks for 2 inch boxes (10 nos.), boxes of 2 inch sizes (40 nos.), a pair of cryosafety gloves should be provided.
Warranty of at least two years from the date of installation.
Original literature/ manuals should be provided to support the technical specification of the
model quoted.
Should provide minimum 10 satisfactory installation reports from government organizations
along with users list.
21
RESEARCH MICROSCOPE: UPRIGHT LIGHT MICROSCOPE WITH FLUORESCENCE
AND DIGITAL IMAGING SYSTEM
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
Upright, bright field, phase contrast fluorescence microscope with trinocular observation tube,
digital camera and image analysis software.
Optical system: Universal infinity corrected optical system or equivalent.
Focus: Course and fine focus.
Illuminator: Kohler illumination. Pre-centered Halogen bulb of at least 12V, 100W or LED
Eyepieces of 10X magnifications with diopter adjustment facility with field of view of 22mm or
higher.
Septuple nosepiece with slots for DIC/polarizing attachments.
The objectives suitable for brightfield, phase contrast and fluorescence microscopy should be
Plan Achromat 4X, Plan Achromat 10X Phase, Plan Achromat 20X Phase, Universal Plan fluorite
40X Phase, Universal Plan fluorite 60X Phase oil, Universal Plan Apochromatic 100X Oil.
Mechanical stage with ceramic coating, with double slide holding capacity.
Universal Condenser with 5-6 positions and phase rings for all phase objectives.
Epi-fluorescence attachment: Epi-fluorescence with 100watt mercury illuminator. Fluorescence
turret with option for 5 to 6 filters cubes.
Fluorescence filters required: Narrow Band Pass filter for DAPI/Hoechst, Narrow Band Pass filter
for FITC/GFP, Narrow Band Pass filter for TRITC/Rhodamine/Alexa 546.
Photography attachment: Color cooled CCD camera with resolution of 5.0 million pixels or
better, minimum 8 bit output or higher and speed of minimum 25 frames per second. Computer
attachment by USB/Firewire should have peltier device for cooling upto 100c below ambient
temperature.
Image Analysis Software should have the following features:- Basic Image acquisition, image enhancement, image overlapping facility.
- Intensity measurements, line profile, spatial measurements such as length, width, area,
perimeter, etc.
- Microscope & camera should be controlled by the microscope manufacturer’s software for
better compatibility.
Data Processing Unit: Branded PC with i5 processor, minimum 4 GB RAM, 500 GB HDD or higher,
Original Windows 7/8 OS, DVD RW, at least 18.5″ TFT monitor, keyboard, mouse and 1 KVA
Offline UPS.
System should be supplied with a compatible stabilizer and online UPS.
Sole/exclusive distributor certificate of microscope & camera should be provided.
User list of at least 5 similar installations in Government institutes should be provided.
The microscope should be CE certified and CE certificate should be submitted along with the
technical bid.
Warranty of three years.
22
AIR CONDITIONER
Capacity 2.0 Ton; split-type, wall-mounted air conditioner, preferably with BEE star rating 3 or above;
with features like timer, auto-restart, and auto-swing along with compatible wallmounted voltage
stabilizer .
Installation at site Warranty at least 2 year.
23
Refrigerated incubator shaker
1.
2.
3.
4.
5.
6.
7.
8.
9.
It should be a bench top Orbital Shaker Incubator.
High quality Stainless steel
Temperature: Range from 5 ̊ C to 60̊C
Refrigeration-Hermetically sealed compressor uses CFC-free refrigerant
Control accuracy ± 0.1̊C by microprocessor PID controlled
Wave speed: minimum of 10 rpm to maximum 400 rpm
Shaking Platform size at least 18 x 18 inch
Display: LED for indication of speed, running time and temperature.
Alarms, audible & visible-Indicate when speed deviates more than 5 rpm or temperature more
than 1°C from setpoint, and when program or time has finished.
10. Glass wool/PUF insulation
Safety features:
1. It should have open lid cut out switch to stop platform motion.
Essential Accessories:
1. It should be supplied with the clamp/adaptors for flasks of 250 ml, 500 ml, 1000 ml and
15/50 ml tube racks.
2. It should be supplied with compatible stabilizer.
3. It should come with 2 years warranty
4. Unit should operate with 220-240 volts/ 50-60 Hz
Additional:
1. Original illustrated Catalogue in support of technical specifications.
2. User list for the quoted model along with contact details including telephone no. and E. mail.
24
The large animal metabolic cages
•
•
•
•
Replacement of existing iron sheet of the large animal metabolic cages (14 Nos.)of animal
nutrition shed with Stainless Steel(SS) Sheet of 22 gauge thickness.
Leak proof Urine trays (14 Nos.) of SS sheet 22 gauge thickness.
Replacement of existing iron hollow rods of large animal metabolic cages (14 Nos.) with suitable
SS hollow rods.
Repair, replacement and painting of existing iron frames of large animal metabolic cages (14
Nos.).
Note: Firms are requested to visit the A. N. Shed for getting the detailed information for replacement of
existing sheet, rods, urine trays etc. Moreover, Specifications are of generic nature.
25
Automatic Microprocessor Based Bomb Calorimeter
Automatic microprocessor based Bomb Calorimeter, Compressor based Chilling Unit along with
Automatic/manual Bomb Lift, Auto Oxygen Filling, Automatic Water Measurement and Dosing and
Auto-adjustment of Water Temperature.
1)
2)
3)
4)
5)
6)
7)
8)
9)
Working Principle -Iso-peribol
Analysis Time 5-10 minutes.
Type of Bomb SS Bomb
Measurement Range 10,000 cal/gm (Max.)
Measurement accuracy (Minimum)-99.7%
Reproducibility( error not more than) - 0.3%
Temperature Resolution - 0.001 to 0.0001oC
Permissible ambient Temperature - 5 to 40 oC
Operating Oxygen Pressure - 20-40 Kg/cm2
WARRANTY: 3 years
26
Farm and Field
A.
-
Rotavator
Two metres Gear type.
Multi-speed with side wheel.
48 Nos. Blade.
OR
B. Gyrovator
- Two metres Gear type.
- Multi-speed with side wheel.
- 60 Nos. Blade.
- Warranty: One year
27
Hydraulic Harrow
-
20 Discs off –set Disc Hydraulic Harrow in two gangs.
100 X 100 X 8 mm size angle iron at least 3400 mm long – 2 Nos.
Gang Beams – 2 Nos. made of MS Channel size 125 X 60 X 0.66 cm (Flenge thickness).
Bearings 4 Nos SKF/NBS, size 30209.
Wheel Rim, Tyre & Tubes- ADV Heavy duty rim of 5 mm thickness with 7.50-16 ADV Tyre & Tube
of Apolo/MRF/ Ceat – 2 each.
Jack (Hydraulic Cylinder) – Heavy duty
Disc Size – Front 600m dia square hole, thickness should not be less than 5mm.
Weight of Harrow should be 13.5 – 14.5 qtls.
Warranty of one year for its parts and workmanship.
28
Tractor
1234567-
Horse Power Capacity :
Engine
:
Number of cylinder :
Brake
:
Steering
:
Speed
:
60-65
Powerful Fuel Efficient Engine
04
Disc Brakes
Power Steering
Multi speed PTO with Reverse speed
Adjusted to Hydrolic Harrow (20 disc)/Cultivator (14-16 tines)/ Rotavator (48- 60 Nos
Blade)/Gyrovator/Hydrolic Trolly/ Thresher/Slacer/Shrubmaster
Warranty: At least 1year against manufacturing default free service at least 3 years.
29
Biosafety Cabinet (Class II)
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
Biosafety cabinet should be class II, type A2
Internal and external dimension (WxDxH in mm) should be at least 950x600x580 mm and
1049x750x1260, respectively
Cabinet should be fitted with HEPA filters
Average Airflow Velocity Inflow and down flow should be at least 0.50 m/s (>100 fpm) and 0.35
m/s (70 fpm), respectively.
Should be vibration free
Should be Fully compliant tested and certified U.S. Standard NSF 49 and European standard EN
12469
Cabinet should have Germicidal UV lamp
Cabinet Construction should be of corrosion proof stain resistant stainless steel interior (with
epoxy coated body)
Bio safety Cabinet must have internally electrical socket outlet, universal service fittings for gas,
air, and vacuum.
Workable power supply: 220-240V, 50/60Hz.
It should be provided with support stand, castor wheels, fixed foot or foot masters.
The product should have at least 10 installations in India in last 5 years.
The Instrument should be provided with uninterrupted power back-up at least for half an hour.
Technical features described above must be supported by original illustrated catalog, with list of
users of the product in India (with all contact detail: Name, affiliation, email, mobile numbers)
Warranty :A minimum of 2 years warranty from the date of installation
30
Integrated Semen straw filling-sealing and printing machine (0.25 ml)
1. Filling & sealing capacity – 12000-15000 straws per hour, four straws at a time.
2. Hopper capacity -2000-4000 straws
3. Have liquid level sensor to automatically stop the machine when semen is finished in the
reservoir.
4. Facility to automatically switch-on completely filled straws instead of 3-4 half filled straws at the
end of filling to avoid semen wastage in hoses and half filled straws.
5. In case of malfunction of machine, it should stop automatically
6. Standardization - ISO certified
7. MS Window compatible dedicated software.
8. Work as a filling & sealing machine only without changing any hardware
9. Ultrasonic sealing facility for the straws.
10. Suitable inkjet printer for permanent & legible printing
11. Automatic bull identification using Bar code.
12. Barcode printing with nozzle size : 40-50 micron, should have facility of 128 condensed bar code.
13. Straw printing should have facility for Barcode, date, bull no., year, batch, place etc. Also free
texts edit line.
14. Work as a printing machine only without changing any hardware.
15. Software for integrated data base for bull, file, administration for every breed.
16. Printer with efficient permanent viscosity checks facility.
17. Nozzle & other parts of the printer should not be chocked by ink due to unused period of
printing.
18. Printer with efficient automatic nozzle-close system, permanent viscosity checks, 24 hour ready
to use.
19. System software can communicate & regulate ultrasonic generator vacuum pump, printer &
filling- sealing machine with a single command of PC attached with machine.
20. Easy operation & maintenance
21. Easy to clean
22. All parts that require cleaning should be easily handled without tools.
23. Compact, user friendly & ergonomic.
24. Low noise.
25. Equipment must be quoted with all additional software.
26. Warranty: at least two years. Amount ( in Rs.) require for AMC for further next five years.
27. Free onsite training & installation
28. Satisfactory working report from users (India & abroad) to be attached
29. Supplier should have their service centre in India to provide after-sale services.
30. Accessory items: One split AC (2 tons) and Online UPS (5 KVA) with at least one hour power back
up on full load for running this instrument.
31
Lan Facility for left over buildings
Eligibility Criteria
• The bidder should be an exclusively IT company of repute having expertise, sound
financial base and capability to execute said work using state of the art technology with
best industry practices available in the global market. The firm should be in existence and
operating in India for the last Ten Years
• The firm/company should be ISO 9001 certified (Networking & Systems Integration). A
copy of the certificate must be attached, otherwise the bid will be considered as nonresponsive.
• The bidder should be in the business of supply, installation & commissioning of LANs for
at least five years. Proof of the same must be enclosed with the technical bid.
• The bidder should be a manufacturer, authorized distributor or large-scale system
integrator of the required networking equipment.
• Bidder should have proven experience of successfully completing at least five projectsof
LAN networking i.e. supply, installed in, testing and commissioning of LAN work with
value of each project being at least Rs 50 lakhs during last three financial years preferably
in Uttar Pradesh / Uttrakhand. Out of these two projects , one project should have been
done with similar components quoted in the offer.
• Proof of satisfactory completion along with scope of these works to be submitted with
the bid.
• An undertaking by the authorized person of the company (self-certificate) that the bidder
hasn’t been blacklisted by a Central /State Government, institution and there has been no
litigation with any Government Department/PSU/Corporation in Central/State
Government on account of similar services.
• Bidder/OEM must have their 24x7 support center in India preferably in NCR / Uttar
Pradesh. Details & proof of service facilities for Technical Support be attached.
• The bidder should have a minimum turnover of Rs. 5 crores (Rupees five crore only) in
each of the last three years on account of supply, installation and commissioning of
Networks. Relevant proof in the form of audited annual report or balance sheet duly
certified by Chartered Accountant certifying the turn over in the field of Networking
should be submitted.
• The bidder should have Authorization from concerned Original Equipment Manufacturers
(OEMs) for supply and support for all active and passive components quoted for at least
five years (Authorization Certificates to be attached).
• All Active Networking Equipment offered and quoted should be from single OEM. Make
and Model must be clearly mentioned in the technical bid. If they are not found in the
technical bid, the bid will be considered as non-responsive.
• Bidder must submit the MAF from the OEM’s of all Active Networking Equipment and
Passive Networking Equipment for participating in the bid and support commitment from
OEM for the Warranty period.
• The bidder shall not assign or sub contract, in whole or in part, its obligations to perform
under the Contract.
• All relevant documents must be enclosed with the tender document at the time of bid
submission. Committee will not entertain any documents submitted later by the bidder at
the time of Bid Evaluation.
•
OEM for active components should be Leaders (First Quadrant) in Gartner's Magic
Quadrant year 2013.
1. Intent of Specification
• Established in 1889, the Indian Veterinary Research Institute (IVRI) is one of the premier
research institutions dedicated to livestock research and development of the region. The
institute with faculty strength of more than 275 has a major mandate of research, teaching,
consultancy and technology transfer activities. The institute imparts quality post-graduate
education to students not only from various parts of the country, but also from overseas.
• Today, the institute with its deemed to be university status contributes immensely to human
resource development in the discipline of veterinary sciences with skills and knowledge
necessary for the challenges of the new millennium. It awards degrees to master and
doctoral programmes in more than 20 disciplines of veterinary and animal sciences, livestock
products technology, basic sciences and extension education.
• Institute is having a network of approximately 500 nodes, designed in star topology and it
includes Laptops, Printers, desktops and other network devices.
2. Scope of Work
• Scope of work comprises of supply, installation, configuration, testing and commissioning of
network components i.e. Active and Passive at various locations at IVRI Izatnagar.
• All switches should be IPV6 ready from day one and of reputed brand (Gartner Magic
quadrant)
• The scope of this tender is expansion of existing LAN infrastructure in IVRI Izatnagar.
• We need to connect Climatology Building, Animal Feed Store, Farm Section, LPM Section,
ATIC Building, Germ Plasm Centre, Human Hospital and Veterinary Polyclinic in star topology
from ARIS Cell as per the attached diagram.
• Farm Machinery and Power Section will be connected using UTP cabling from Farm Section
• Within building, connectivity will be done using UTP Cabling
• A sketch showing LAN Network at IVRI is enclosed as part of this Tender Document. As shown
in this sketch, it is intended to wire the network for about 100 nodes
• However, the alternate suitable proposal may be proposed by the bidder with full reasoning.
• The bidder shall also provide all required equipment which may not be specifically stated
herein but are required to meet the intent of ensuring completeness, maintainability and
reliability of the total system covered under this specification, including integration and
interoperability with the existing LAN.
• Bidder has to powering on equipment after ensuring correctness of terminations, interfaces
and power supply and making the system ready for testing and commissioning.
• Testing of LAN Cables after laying, terminations and ferruling at both the ends. All testing
tools and instruments shall be brought by the bidder and taken back after the testing. Cabling
work shall also be got certified from OEM.
• Configuration of the equipment as per the requirements of IVRI including Network
segmentation and Network Monitoring
• Field testing and commissioning of system, including integration with existing Network for
internet connectivity.
• Site acceptance tests to establish satisfactory performance of the equipment's as per specs.
• Assistance for familiarization and operation of the installed system & services for 3 months
after acceptance of system.
• In case, the quantity of cables or fixing wall mount sockets etc. exceeds or is less than the
quantity in bid price schedule, the payment for the executed quantity shall be paid on prorata basis, for the actual quantities for which the installation is carried out through the Bidder
on Certification by Site Engineer.
•
The successful bidder will be responsible for expansion of IVRI LAN asmentioned under
‘Scope of Work’ part and ensure enhanced smooth operation of the networkthereafter.
• After the expansion work, the expanded equipments have to be integrated with theIVRI
running network to ensure enhanced output.
• Any fault reported in the existingequipments/network due to this expansion work, has to be
taken care by the bidder
3. General Technical Requirements
• All active LAN components shall be of the same Make/manufacturer and shall be covered
under same back-up guarantee from the same OEM, to ensure full compatibility, interworking and inter-operability.
• Equipment furnished shall be complete in every respect with all mountings, fittings, fixtures
and standard accessories normally provided with such equipment's and/or needed for
erection, completion and safe operation of the equipment's as required by applicable codes
though they may not have been specifically detailed in the technical specification, unless
included in the list of exclusions. All similar standard components/parts of similar standard
equipment's provided, shall be inter-changeable with one another.
• The Bidder shall be responsible for providing all materials, equipment's, and services,
specified or otherwise, which are required to fulfill the intent of ensuring operability,
maintainability, and reliability of the complete equipment covered under this specification
within his quoted price. This work shall be in compliance with all applicable standards,
statutory regulations and safety requirements in force of the date of award of this contract.
• The bidder shall also be responsible for deputing qualified personnel for installation, testing,
commissioning and other services under his scope of work as per this specification. All
required tools and tackles for completing the scope of work as per the specification is also
the responsibility of the bidder.
• The bidder shall also be responsible for deputing qualified personnel for installation, testing,
commissioning and other services under his scope of work as per this specification. All
required tools and tackles for completing the scope of work as per the specification is also
the responsibility of the bidder.
• The Bidder shall furnish two complete sets of system and equipment instruction manuals and
detailed installation operating and maintenance documentation along with the CDs for the
same. Manuals shall describe system operation, shall include detailed system and
components description and shall cover the installation, operation, care and maintenance of
all system components including diagnostics. Complete parts lists including detailed
instructions for ordering shall be furnished. Complete documentation for operating, system
software, application programs, software selectable functions, shall be provided. After
installation "as built" drawings shall be provided by the bidder indicating the cable routes,
termination & termination details, actual configuration & parameters, etc.
Warranty and AMC
• The bidder warrants that the Goods supplied under this Contract are new, unused, of the most
recent or current models and that they incorporate all recent improvements in design and
materials unless provided otherwise in the Contract.
• This warranty shall remain valid for a period of a minimum 36 months after the Goods or any
portion thereof as the case may be, have been delivered, commissioned and accepted at the
final destination indicated in the Contract.
• The bidder will submit the rates of AMC for five year after the expiry of warranty period.
• The bidder shall post a well-qualified and experienced Service Engineer full time at the site at
his own expense(no extra payment will be made) and get the defects/bugs removed in the
system as detected by the Purchaser during the period of warranty.
Documentation
The bidder shall supply the following documents at the time of Complete set of Technical/Operation and
Maintenance Manual
• An inventory of items delivered.
• Rack wise inventory
• Node & connectivity details
• Switch installation Report
• UTP/OFC Wiring diagram
• Certification details for UTP & OFC connections Acceptance Test:
Schedule for Requirements
SNo
1
2
3
1
2
3
4
5
6
7
8
9
10
11
12
12
13
1
2
3
4
5
6
7
8
9
10
11
SPECIFICATIONS
ACTIVE COMPONENTS
Basic Layer3 Switch having minimum 24 Nos. Ethernet 10/100/1000 BaseT
and 2 nos. 1000BaseX SFP Ports
1000 Base LH/LX Transreceiver
1000 Base SX Transreceiver
PASSIVE COMPONENTS
6 Core SM Fiber Cable
Rack Mount 6 Port LIU Loaded with Couplers
Rack Mount 12 Port LIU Loaded with Couplers
Pigtals SM Simplex
SC-LC Fiber Patch Cord SM
SC-LC Fiber Patch Cord MM
Patch Panel Cat 6 Keystone -24 Port - loaded
I/O with Face Plate for Cat 6 Jack Keystone - loaded
UTP CAT 6 cable ( Box of 305 meters)
UTP CAT 6 Patch cord (1 Metre)
UTP CAT 6 Patch cord (2 Metre)
19" 12 U Rack with Cable Manager, Power Strip
HDPE Pipe 32 MM
Adaptor SC SM -ZR-SLEEVE Simplex
JOBWORK
Digging of Soft Soil
Digging of Hard Soil
Laying of Fiber Cable in HDPE Pipe
Root Marker Cemented
Splicing of Fiber
LIU Fixing and Dressing
UTP Cable Laying in Conduit with Material per meter cable used (without
Cable)
IO Fixing and Termination
Jack Panel Termination
Rack Fixing and Dressing including Fixing of Jack Panels, Cable Laying and
Ferruling etc.
Documentation and Configuration 1 Job
UOM
Qty
No.
15
Nos.
Nos.
15
15
Mtr.
Nos.
No.
Nos.
Nos.
Nos.
Nos.
Nos.
Nos.
Nos.
Nos.
3000
5
2
108
15
10
8
100
16
100
100
Nos.
Mtr.
No.
8
3000
108
Mtr.
Mtr.
Mtr.
Nos.
Nos.
Nos.
Mtr.
2400
600
2600
20
54
7
4800
Nos.
Nos.
Nos.
100
8
8
Job
1
Detailed Technical Specifications
Active Component
1. Detailed Technical Specifications of Basic Layer3 Edge Switch 24 Ports
a. Port Required
i. The switch should support a minimum of 24 nos. 10/100/1000 Ethernet Ports and 2 nos. 1G
SFP ports.
ii. The switch should have total 26 ports
b. Performance and Scalability
i. The switch should support Forwarding bandwidth of 108 Gbps
ii. The switch should support Full-duplex Switching bandwidth of 210Gbps
iii. The switch should support 64-Byte Packet Forwarding Rate of 70 Mpps
iv. The switch should support 1023 VLANs
v. The switch should support Maximum transmission unit (MTU) of 9198 bytes
vi. The switch should support 16000 Unicast MAC addresses
c. Stacking
i. The switch should support Stacking with dedicated stacking ports in addition to 26 ports.
ii. Stacking module should be Hot-swappable.
iii. The switch should able to stack eight switch in a single stack unit with stacking bandwidth of
80Gbps.
d. Power Supply
i. The switch should support an Redundant Power Supply
e. Layer-2 Features
i. The switch should support Automatic Negotiation of Trunking Protocol, to help minimize the
configuration & errors
ii. The switch should support IEEE 802.1Q VLAN encapsulation
iii. The switch should support Centralized VLAN Management. VLANs created on the Core
Switches should be propagated automatically
iv. The switch should support Spanning-tree PortFast and PortFast guard for fast convergence
v. The switch should support UplinkFast&BackboneFast technologies to help ensure quick
failover recovery, enhancing overall network stability and reliability
vi. The switch should support Spanning-tree root guard to prevent other edge swicthes
becoming the root bridge.
vii. The switch should support Per-port multicast storm control to prevent faulty end stations
from degrading overall systems performance
viii. The switch should support Local Proxy Address Resolution Protocol (ARP) working in
conjunction with Private VLAN Edge to minimize broadcasts and maximize available
bandwidth.
ix. The switch should support MVR (Multicast VLAN Registration)
f. L3 Features
i. The switch should support Inter-VLAN routing
ii. The switch should support IPv4 unicast Static Routing
iii. The switch should support 16 IPv4 Static routes
g. Configuration Operations
i. The switch should support configuration of the Software image and switch configuration
without user intervention
ii. The switch should support automatic configuration as devices connect to the switch port
iii. The switch should support system health checks within the switch
h. Quality of Service (QoS) & Control
i. The switch should support 4 egress queues per port to enable differentiated management
ii. The switch should support scheduling techniques for Qos
iii. The switch should support Weighted tail drop (WTD) to provide congestion avoidance
iv. The switch should support availability of up to 256 aggregate or individual polices per port.
i. Network security features
i. The switch should support IEEE 802.1x to allow dynamic, port-based security, providing user
authentication.
ii. The switch should support Port-based ACLs for Layer 2 interfaces to allow application of
security policies on individual switch ports.
j. IPv6 Features
i. The switch should be on the approved list of IPv6 Ready Logo phase II – Host
ii. The switch should support IPv6 unicast Static Routing
iii. The switch should support 16 IPv6 Static routes
iv. The switch should support IPv6 MLDv1 & v2 Snooping
2. 1000 Base LH/LX Transreceiver
i. Transreceiver Module should be of same make of switch and core switch
ii. Module should be compatible to the quoted switch and core switch
3. 1000 Base SXTransreceiver
i. Transreceiver Module should be of same make of switch and core switch
ii. Module should be compatible to the quoted switch and core switch
Passive Component
1. 6 Core SM Fiber Cable
Cable type
6 Core, Single mode, Armored, loose-tube, CST armour,
Gel Filled.
Fiber Type
9/ 125, Telcordia’s GR-20 and ITU-T 652.D Compliance.
Cable Construction will be BELLCORE GR 20 / IEC 794-1.
Attenuation
@1310nm < = 0.35 dB/Km
@1550nm < = 0.20 dB/Km
Coating / Cladding
non-circularity<= 12 microns.
Zero Dispersion Slope should be <= 0.086 ps / sqnm-km.
Max (chromatic) dispersion should be
<5.3 ps/nm-km @1270-1340 nm
<3.5 ps/nm-km @1285-1330 nm
<185 ps/nm-km @1550 nm
Tensile
rating --- 1000N
Armoring
Corrugated Steel tape Armor.
jacket
Inner:- High density polyethylene.
Outer :- High density polyethylene, anti - termite, anti rodent suitable for direct burial application.
Secondary Buffer Material should be Gel filled Loose Tube.
Min Bend 20 X Outer Diameter.
Test
Test (Must pass) : IEC794-1-E1 , IEC794-1-E2 , IEC794-1-E3
, IEC794-1-E4 , EIA-455-104 , IEC794-1-E7 , IEC794-1-E10 ,
IEC794-1-F1 , IEC794-1-F3 and IEC794-1-F5
Marking
Marking should be Identification marking at regular
intervals of 1 meter.
Fiber Core must be Raw fiber of corning. CORNING
marking should be visible on the OFC.
Resistance
Maximum Crush resistance- 44 N/mm
•
2. LIU,6 Port Rack Mount enclosure,
● 1U, 19 Inch Rack mountable Metallic Distribution Enclosure with Adaptor Plate and
Splicing Tray with air ventilation provision on the enclosure.
● Should have optimum air circulation (air rack version)
● 19 “ Rack Mountable Cabinet
● Complete Aluminium Alloy housing, fully powder coated
● Panel cover is of slide out for easy maintenance
● Rubber grommets are provided at the cable entry
● points for tight sealing.
● Reduced mounting depth of 210mm
● Universally compatible (eg ETSI std compliant)
● Cable entry from the rear or from the side
3. LIU,12 Port Rack Mount enclosure,
● 1U, 19 Inch Rack mountable Metallic Distribution Enclosure with Adaptor Plate and
Splicing Tray with air ventilation provision on the enclosure.
● Should have optimum air circulation (air rack version)
● 19 “ Rack Mountable Cabinet
● Complete Aluminium Alloy housing, fully powder coated
● Panel cover is of slide out for easy maintenance
● Rubber grommets are provided at the cable entry
● points for tight sealing.
● Reduced mounting depth of 210mm
● Universally compatible (eg ETSI std compliant)
● Cable entry from the rear or from the side
4. PIGTAIL SC SM SIMPLEX LENGTH- 1m
● Simplex LC/SC, 9/125 micron SM fiber, factory terminated ceramic connectors.
Qualifies as per ITU-T G652 D Fiber.
● Plastic molded plug type connectors
● Specially designed Low Smoke Zero Halogen Outer Sheath
● Standard 2.5mm ceramic ferrule
● Compact in size and easy to connect
● Recommended by EIA/TIA 568 Specifications
● Macrobending, Max. (100 turns) 0.05 dB (1,310/1,550nm@50mm) 0.05 dB
(1,625nm@60mm) Macrobending, Max. 0.05 dB@1,550nm (1 turns@32mm
mandrel).
●
Length 1 Mt.
5. Optical Fiber Patch Cords SM, - 2Mtr.
Type
Duplex LC/SC, 9/125 micron SM fiber, factory terminated
ceramic connectors. Qualifies as per ITU-T G652 D Fiber
Connector Type Plastic molded plug type connectors
Feature
Standard 2.5mm ceramic ferrule
Compact in size and easy to connect
Recommended by EIA/TIA 568
Macrobending, Max. (100 turns) 0.05 dB (1,310 / 1,550nm
Specification @50mm) 0.05 dB (1,625nm@60mm) Macrobending, Max.
0.05 dB@1,550nm (1 turns@32mm mandrel).
Length
2 Meter
6. Optical Fiber Patch Cords MM, - 2Mtr.
Type
•
•
•
•
Patch Cord MM patch cord LC-SC TYPE
1.6mm or 3mm simplex or Duplex Zipcord.
Fiber Patch cord must have Clear Curve Feature for
more flexibility.
Qualifies as per ITU-T G652 D Fiber
Outside Diameter
•
•
(Simplex): 1.6mm x 3.0mm
(Duplex): 1.6mm x 3.3mm
Minimum Cable
Retention Strength
•
1.6mm: 11.24 lbs (50 N)
Insertion Loss
•
Less than 0.5 dB for MM and 0.3 dB for SM
Length
2 Meter
7. Patch Panel Cat 6 UTP Modular -24 Port- Loaded 0.5U Height
Type 24-port, Unshielded Twisted Pair, Category 6, TIA / EIA 568
Port Keystone Krone type. Ports must be replaceable in 8 Port
Arrangement Module.
Category Category 6
Circuit Patch Panels Rear Cable Management without tie-wrap with
Identification simplified numbering labels
Scheme
Height . 5 U , to save the Rack space.
8. Information Outlet Key Stone with Face Plate (Single Port) and SMB
Features
●
●
●
●
●
●
Surface Mount Face Plate & Box with CAT6 Work
Area Data I/O Outlet (RJ45) adhering to ETL
Verification program for compliance with TIA568B.21
Category 6, TIA568.B.2-1 – 600MHz
Information Outlet should be UL Listed and I/O box
should be RoHS Compliant
568A/B configuration
Be constructed of high impact with color and icon
options for better visual identification. Integrated
hinged dust cover using collapsible angular spring
loaded shuttered technology.
Face plate should be Single Gang square plate and
Should have provision for Labe
Material
●
ABS / UL 94 V-0
Port
●
One
9. Category 6 UTP, 4 Pair Cable
Features
Category 6 Unshielded Twisted Pair 4 pair 100 Ohm cable
shall be compliant with ANSI/TIA/EIA-568-B.2-1 upto 600
Mhz.
Category 6 UTP cables shall extend between the work area
location and its associated telecommunications closet and
consist of 4 pair, 23 AWG, UTP Non Plenum cable jacket.
The 4 pair Unshielded Twisted Pair cable should be UL
Listed and ETL Certify.
Zero-Bit error throughput test as per IEEE standards
verified by ETL
The 4 pair Unshielded Twisted Pair cable should be RoHS
Compliant
Mechanical
Characteristics
Construction: 4 twisted pairs separated by internal X
shaped, 4 channel, full separator. Half shall not be
accepted.
Conductor Solid Copper
Insulator Polyethylene
Delay Skew : 35 ns MAX
Jacket PVC
Attenuation : 22.8dB/100m at 250MHz, 29.4dB/100m at
400MHz, 39dB/100m at 600MHz
10. and 11 Patch Cord Cat 6 UTP Blue 1m and 2m
Type
Unshielded Twisted Pair, Category 6, TIA / EIA 568-B.2
Conductor
●
24 AWG 7 / 32, stranded copper
Length
●
7-feet for workstation and 3feet for Jackpanel /
equipment
Plug Protection
●
Matching colored snag-less, elastomer polyolefin
boot
Warranty
●
25-year component warranty
Category
●
Category 6
Housing
●
Clear polycarbonate
Terminals
●
Phosphor Bronze, 50 micron gold plating over
selected area and gold flash over remainder, over 100
micron nickel under plate
Load bar
●
PBT polyester
Jacket
●
PVC
Insulation
●
Flame Retardant Polyethylene
12. 12 U Wall Mount Network Rack with Ass.
Size
12U Wall Mount
Door
Transparent Lockable Glass Door
Power Socket
3Ways 5A Power Socket 1 No.s
Cable Manager
1 Nos of 1U Size for Cabling Routing
32
Upgradation of existing network switches and OFC
Eligibility Criteria
• The bidder should be an exclusively IT company of repute having expertise, sound
financial base and capability to execute said work using state of the art technology with
best industry practices available in the global market. The firm should be in existence and
operating in India for the last Ten Years
• The firm/company should be ISO 9001 certified (Networking & Systems Integration). A
copy of the certificate must be attached, otherwise the bid will be considered as nonresponsive.
• The bidder should be a authorized distributor or large-scale system integrator of the
required networking equipment and he bidder should have Authorization from
concerned Original Equipment Manufacturers (OEMs) for supply and support for all active
and passive components quoted for at least five years (Authorization Certificates to be
attached).
• The bidder should be in the business of supply, installation & commissioning of LANs for
at least five years. Proof of the same must be enclosed with the technical bid.
• Bidder should have proven experience of successfully completing at least five projectsof
LAN networking i.e. supply, installed in, testing and commissioning of LAN work with
value of each project being at least Rs 50 lakhs during last three financial years preferably
in Uttar Pradesh / Uttrakhand. Out of these projects, three projects should have been
done with similar components quoted in the offer.
• Proof of satisfactory completion along with scope of these works to be submitted with
the bid.
• An undertaking by the authorized person of the company (self-certificate) that the bidder
hasn’t been blacklisted by a Central /State Government, institution and there has been no
litigation with any Government Department/PSU/Corporation in Central/State
Government on account of similar services.
• Bidder/OEM must have their 24x7 support center in India preferably in NCR / Uttar
Pradesh. Details & proof of service facilities for Technical Support be attached.
• The bidder should have a minimum turnover of Rs. 5 crores (Rupees five crore only) in
each of the last three years on account of supply, installation and commissioning of
Networks. Relevant proof in the form of audited annual report or balance sheet duly
certified by Chartered Accountant certifying the turn over in the field of Networking
should be submitted.
• All Active Networking Equipment offered and quoted should be from single OEM. Make
and Model must be clearly mentioned in the technical bid. If they are not found in the
technical bid, the bid will be considered as non-responsive.
• Bidder must submit the MAF from the OEM’s of all Active Networking Equipment and
Passive Networking Equipment for participating in the bid and support commitment from
OEM for the Warranty period.
• The bidder shall not assign or sub contract, in whole or in part, its obligations to perform
under the Contract.
• All relevant documents must be enclosed with the tender document at the time of bid
submission. Committee will not entertain any documents submitted later by the bidder at
the time of Bid Evaluation.
•
OEM for active components should be Leaders (First Quadrant)in Gartner's Magic
Quadrant year 2013.
Intent of Specification
• Established in 1889, the Indian Veterinary Research Institute (IVRI) is one of the premier
research institutions dedicated to livestock research and development of the region. The
institute with faculty strength of more than 275 has a major mandate of research, teaching,
consultancy and technology transfer activities. The institute imparts quality post-graduate
education to students not only from various parts of the country, but also from overseas.
• Today, the institute with its deemed to be university status contributes immensely to human
resource development in the discipline of veterinary sciences with skills and knowledge
necessary for the challenges of the new millennium. It awards degrees to master and
doctoral programmes in more than 20 disciplines of veterinary and animal sciences, livestock
products technology, basic sciences and extension education.
• Institute is having more than 25 buildings in its campus and all of the buildings are connected
using multimode and single mode OFC in star topology.
• Most of the network switches are L2 switches of Nortel Make.
Scope of Work
• The scope of this tender is to upgrade the existing LAN infrastructure in IVRI Izatnagar.
• Scope of work comprises of supply, installation, configuration, testing and commissioning of
active network components at various locations at IVRI Izatnagar
• We need the re-laying and termination of 6 Core Single Mode OFC from ARIS Cell to Y-Shape
building, Veterinary Polyclinic, Library, P & C Building, Animal Nutrition Division and Central
Office directly.
• All the existing active network components (Switches, fiber modules, fiber patch cord etc) of
ARIS Cell, Y-Shape building , Veterinary Polyclinic, Library, P & C Building, Animal Nutrition
Division,Central Office, CADRAD, BP Division, Standardization Division, JDR Office,
Biochemistry Building, KVK, Extension Education Building, and Parasitology Building are to be
replaced with 10Gbps uplinks and all the end user port should have connectivity of
10/100/1000 Mbps.
• The existing equipments which have its life will have to be effectively utilized in the network.
• Bidder has to provide powering on equipment after ensuring correctness of terminations,
interfaces and power supply and making the system ready for testing and commissioning.
• Testing of LAN Cables after laying , terminations and ferruling at both the ends. All testing
tools and instruments shall be brought by the bidder and taken back after the testing. Cabling
work shall also be got certified from OEM.
• Configuration of the equipment as per the requirements of IVRI including Network
segmentation and Network Monitoring
• Configuration of VLANs as per requirements of IVRI
• Field testing and commissioning of system, including integration with existing Network for
internet connectivity.
• Site acceptance tests to establish satisfactory performance of the equipment's as per specs.
• Assistance for familiarization and operation of the installed system & services for 3 months
after acceptance of system.
• In case, the quantity of cables or fixing wall mount sockets etc. exceeds or is less than the
quantity in bid price schedule, the payment for the executed quantity shall be paid on prorata basis, for the actual quantities for which the installation is carried out through the Bidder
on Certification by Site Engineer.
• The successful bidder will be responsible for upgradation of network components
asmentioned under ‘Scope of Work’ part and ensure enhanced smooth operation of the
networkthereafter. After the upgradation work, the upgraded equipments have to be
integrated with theIVRI running network to ensure enhanced output.
• Any fault reported in the existingequipments/network due to this upgradation work, has to
be taken care by the bidder
• A sketch showing Existing LAN Network at IVRI and proposed OFC re-laying diagram is
enclosed as part of this Tender Document.
• However, the alternate suitable proposal may be proposed by the bidder with full reasoning.
• The bidder shall also provide all required equipment which may not be specifically stated
herein but are required to meet the intent of ensuring completeness, maintainability and
reliability of the total system covered under this specification, including integration and
interoperability with the existing LAN.
General Technical Requirements
• All active LAN components shall be of the same Make/manufacturer and shall be covered
under same back-up guarantee from the same OEM, to ensure full compatibility, interworking and inter-operability.
• Equipment furnished shall be complete in every respect with all mountings, fittings, fixtures
and standard accessories normally provided with such equipment's and/or needed for
erection, completion and safe operation of the equipment's as required by applicable codes
though they may not have been specifically detailed in the technical specification, unless
included in the list of exclusions. All similar standard components/parts of similar standard
equipment's provided, shall be inter-changeable with one another.
• The Bidder shall be responsible for providing all materials, equipment's, and services,
specified or otherwise, which are required to fulfill the intent of ensuring operability,
maintainability, and reliability of the complete equipment covered under this specification
within his quoted price. This work shall be in compliance with all applicable standards,
statutory regulations and safety requirements in force of the date of award of this contract.
• The bidder shall also be responsible for deputing qualified personnel for installation, testing,
commissioning and other services under his scope of work as per this specification. All
required tools and tackles for completing the scope of work as per the specification is also
the responsibility of the bidder.
• The bidder shall also be responsible for deputing qualified personnel for installation, testing,
commissioning and other services under his scope of work as per this specification. All
required tools and tackles for completing the scope of work as per the specification is also
the responsibility of the bidder.
• The Bidder shall furnish two complete sets of system and equipment instruction manuals and
detailed installation operating and maintenance documentation along with the CDs for the
same. Manuals shall describe system operation, shall include detailed system and
components description and shall cover the installation, operation, care and maintenance of
all system components including diagnostics. Complete parts lists including detailed
instructions for ordering shall be furnished. Complete documentation for operating, system
software, application programs, software selectable functions, shall be provided. After
installation "as built" drawings shall be provided by the bidder indicating the cable routes,
termination & termination details, actual configuration & parameters, etc.
Warranty and AMC
• The bidder warrants that the Goods supplied under this Contract are new, unused, of the most
recent or current models and that they incorporate all recent improvements in design and
materials unless provided otherwise in the Contract.
• This warranty shall remain valid for a period of a minimum 36 months after the Goods or any
portion thereof as the case may be, have been delivered, commissioned and accepted at the
final destination indicated in the Contract.
• The bidder will submit the rates of AMC for five year after the expiry of warranty period.
• The bidder shall post a well-qualified and experienced Service Engineer full time at the site at his
own expense and get the defects/bugs removed in the system as detected by the Purchaser
during the period of warranty.
Documentation
The bidder shall supply the following documents at the time of Complete set of Technical/Operation and
Maintenance Manual
• An inventory of items delivered.
• Rack wise inventory
• Node & connectivity details
• Switch installation Report
• UTP/OFC Wiring diagram
• Certification details for UTP & OFC connections Acceptance Test:
Schedule for Requirements
SNo
1
2
3
4
5
Items Required
ACTIVE COMPONENTS
Layer3 Modular Core Switch Chassis with One supervisor switching line
card
Basic Layer3 Switch having minimum 48 Nos. Ethernet 10/100/1000
BaseT and 2 nos. 1000BaseX SFP Ports
Basic Layer3 Switch having minimum 24 Nos. Ethernet 10/100/1000
BaseT and 2 nos. 1000BaseX SFP Ports
1000 Base LH/LX Transreceiver
1000 Base SX Transreceiver
UOM
Qty
No.
1
No.
8
No.
8
Nos.
Nos.
15
8
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
1.
2.
3.
4.
5.
6.
7.
8.
PASSIVE COMPONENTS
6 Core SM Fiber Cable
Rack Mount 6 Port LIU Loaded with Couplers
Rack Mount 24 Port LIU Loaded with Couplers
Rack Mount 12 Port LIU Loaded with Couplers
Pigtals SM Simplex
SC-LC Fiber Patch Cord SM – 2mtr
Fiber Patch Cord MM – 2mtr
Adaptor SC SM -ZR-SLEEVE Simplex
HDPE Pipe 32 MM
19" 12 U Rack with Cable Manager, Power Strip
19" 42 U Open Rack with Cable Manager, Dual Power
Strip ( For Server and Network)
JOBWORK
Digging of Soft Soil
Digging of Hard Soil
Laying of Fiber Cable in HDPE Pipe
Root Marker Cemented
Splicing of Fiber
LIU Fixing and Dressing
Rack Fixing and Dressing including Fixing of Jack Panels,
Cable Laying and Ferruling etc.
Documentation and Configuration 1 Job
Mtr.
Nos.
No.
No.
Nos.
Nos.
Nos.
No.
Mtr.
Nos.
5000
4
2
2
108
12
15
108
3000
6
No.
1
Mtr.
Mtr.
Mtr.
Nos.
Nos.
Nos.
Nos.
2500
500
5000
20
108
8
6
Job
1
Detailed Technical Specifications
Active Component
4. Layer3 Modular Core Switch Chassis with One supervisor switching line card
a. Chassis specification and Feature
i. The chassis should be modular with minimum 7 slots, minimum 5 payload slots and minimum
2 supervisor switching line card slots.
ii. Switch should have dual hot-swappable power supplies, a hot-swappable fan tray with
redundant fans.
iii. Shall have Minimum 48 Nos. of 10/100/1000 Mbps TX Ports
iv. Shall have Minimum 24 Nos. of 1000 Base X Ports Distributed over 2 line cards
v. Shall have Minimum 8 Nos. of 10G SFP+/XFP Ports on the CPU or in line card
vi. It should support around five microseconds of latency to and from any 10 Gigabit Ethernet
port.
vii. Switch Must support IPv4 and IPv6 switching and routing in hardware
viii. PoE Plus capabilities on all ports in a line card simultaneously
ix. Supplied line card should have at least four built-in 10 GE SFP plus ports.
b. Performance
i. Switch must have minimum 900Gbps Switching capacity or more with 250 Mpps throughput
for IPv4 and 125 Mpps for IPv6 .
ii. Switch must support 48 Gbps Per-slot Switching Capacity
iii. Support for 256,000 ( IPv4 ) & 128,000 ( IPv6 ) entries in routing table (shared between
unicast and multicast)
iv. IPv6 support in hardware, providing wire-rate forwarding for IPv6 networks
v. Dual stack support for IPv4/IPv6 and Dynamic hardware forwarding-table allocations for ease
of IPv4-toIPv6 migration
vi. No performance penalty with advanced Layer 3 and Layer 4 services enabled
vii. Switch should supportminimum of 300 non-blocking GE ports. 100 GE of SFP plus ports,
viii. Switch Must be equipped with minimum 2GB RAM, 4 GB Flash / SDRAM
ix. Switch should support scalability upto 4000 virtual ports (VLAN port instances)
x. Switch should support minimum 55K MAC Address
xi. Bandwidth aggregation up to 16 Gbps through Port Aggregation.
xii. Hardware-based multicast management & Hardware-based ACLs
c. Layer 2 Features
i. The switch should support 8 bi-directional SPAN and/or RSPAN sessions
ii. IEEE 802.1Q VLAN encapsulation, Support for minimum 4096 VLANs per switch
iii. Support for Automatic Negotiation of Trunking Protocol, to help minimize the configuration
& errors.
iv. Centralized VLAN Management. VLANs created on the Core Switches should be propagated
automatically.
v. Spanning-tree PortFast and PortFast guard for fast convergence with STP instances of
minimum 2500.
vi. Switch should support 802.1d, 802.1s, 802.1w, 802.3ad, 802.3af Power over Ethernet (PoE)
vii. Spanning-tree root guard to prevent other edge swicthes becoming the root bridge.
viii. IGMP snooping v1, v2 and v3
ix. Port Aggregation across line cards
x. Link Aggregation Protocol (LACP)
xi. Support for Detection of Unidirectional Links and to disable them to avoid problems such as
spanning-tree loops
xii. Line card switchover should be in subsecond failover time, so that the Layer 2 Links need not
be renegotiated in case of failure of one of the CPU module.
xiii. The Switch should be able to discover the neighboring device of the same vendor giving the
details about the platform, IP Address, Link connected through etc, thus helping in
troubleshooting connectivity problems.
d. Layer 3 Features
i. IP routing protocols Open Shortest Path First [OSPF], Routing Information Protocol [RIP],
RIP2), Intermediate System to Intermediate System (IS-IS), Border Gateway Protocol Version
4 (BGP4) and Multicast Border Gateway Protocol (MBGP), ICMP Router Discovery Protocol,
Policy-Based Routing (PBR)
ii. Support for Router redundancy Protocol
iii. Multicast entries should be minimum of 65,000 ( IPv4 ) and 60,000 ( IPV6 ).
iv. The Switch should be Non-Stop Forwarding (NSF)-Aware for most of the dynamic routing
protocols.
v. Software routing of Internetwork Packet Exchange (IPX) and AppleTalk for legacy protocols
vi. IP multicast routing protocols (PIM, SSM, Distance Vector Multicast Routing Protocol
[DVMRP]), Pragmatic General Multicast (PGM)
vii. IGMP v1, v2, and v3, IGMP filtering on access and trunk ports
viii. Support for Virtual Route Forwarding feature to map VLANs to respective VRFs
ix. Support for IPv6 Protocol.
e. Quality of Service
i. QoS hardware entries should be minimum of 100,000.
ii. Per-port quality of service (QoS) configuration, Support for four queues per port in hardware
iii. Strict priority queuing,IP differentiated service code point (DSCP) and IP Precedence
iv. Classification and marking based on IP type of service (TOS) or DSCP, full Layer 3 and Layer 4
headers
v. Input and output policing based on Layer 3 and Layer 4 headers. Support for 1024 policers on
ingress and 1024 policers on egress configured as aggregate or individual
vi. Shaping and sharing output queue management, Support for Congestion avoidance feature
like DBL (Dynamic Buffer Limiting)
vii. No performance penalty for granular QoS functions
viii. Support for Automatic Quality of Service for easy configuration of QoS features for critical
applications.
f. Security Features
i. Support for Admission Control features for end-point security (PCs & Desktop Posture
Assessment) and thereby controlling the spread of viruses & worms
ii. RADIUS, which allow centralized control of the switch and restrict unauthorized users from
altering the configuration
iii. Standard and extended ACLs on all ports
iv. 802.1x user authentication (with VLAN assignment and Guest VLAN extensions)
v. Router ACLs (RACLs) on all ports (no performance penalty), VLAN ACLs (VACLs),Port ACLs
(PACLs)
vi. Private VLANs (PVLANs) on access and trunk ports
vii. Dynamic Host Configuration Protocol (DHCP) snooping and Option82 insertion
viii. Port Security, Secure Shell (SSH) Protocol versions 1 and 2
ix. Unicast MAC filtering, Unicast port flood blocking
x. Dynamic Address Resolution Protocol (ARP) inspection, IP source guard
g. Management Features
i. The switch should support Flexible NetFlow and IP SLA for enhanced visibility.
ii. Switch should have the feature of EnergyWise for simplified power management
iii.
iv.
v.
vi.
Configuration Rollback for improved configuration management
Single console port and single IP address to manage all features of the system
Software-configuration management, including local and remote storage
Optional Compact Flash memory card to store software images for backup and easy software
upgrades
vii. Manageability through common network-management software on a per-port and perswitch basis, providing a common management interface for routers, switches of the same
vendor
viii. Support for SNMP versions 1, 2, and 3
ix. Remote Monitoring (RMON) software agent to support four RMON groups (history, statistics,
alarms, and events) for enhanced traffic management, monitoring, and analysis. Support for
all nine RMON groups through the use of a mirrored port, which permits traffic monitoring of
a single port, a group of ports, or the entire switch from a single network analyzer or RMON
probe
x. Analysis support, including ingress port, egress port, and VLAN mirroring
xi. Support for sFlow / NetFlow or equivalent for Deep traffic & session flow analysis.
xii. Support for Port Mirroring, Remote Mirroring & ACL filtering for Port mirroring used for IDS
functionalities.
5. Basic Layer3 Switch having minimum 48 Nos. Ethernet 10/100/1000 Base T and 2 nos. 1000BaseX
SFP Ports
a. General Features
i. The switch should support a minimum of 48 nos. 10/100/1000 Ethernet Ports and 2 nos. 1G
SFP ports.
b. Performance and Scalability
i. The switch should support Forwarding bandwidth of 108 Gbps
ii. The switch should support Full-duplex Switching bandwidth of 216 Gbps
iii. The switch should support 64-Byte Packet Forwarding Rate of 130 Mpps
iv. The switch should support 1023 VLANs
v. The switch should support Maximum transmission unit (MTU) of 9198 bytes
vi. The switch should support 16000 Unicast MAC addresses
c. Stacking
i. The switch should support Stacking with dedicated stacking ports in addition to 26 ports.
ii. Stacking module should be Hot-swappable.
iii. The switch should able to stack eight switch in a single stack unit with stacking bandwidth of
80Gbps.
d. Power Supply
i. The switch should support an Redundant Power Supply
e. Layer-2 Features
i. The switch should support Automatic Negotiation of Trunking Protocol, to help minimize the
configuration & errors
ii. The switch should support IEEE 802.1Q VLAN encapsulation
iii. The switch should support Centralized VLAN Management. VLANs created on the Core
Switches should be propagated automatically
iv. The switch should support Spanning-tree PortFast and PortFast guard for fast convergence
v. The switch should support UplinkFast&BackboneFast technologies to help ensure quick
failover recovery, enhancing overall network stability and reliability
vi. The switch should support Spanning-tree root guard to prevent other edge swicthes
becoming the root bridge.
vii. The switch should support Per-port multicast storm control to prevent faulty end stations
from degrading overall systems performance
viii. The switch should support Local Proxy Address Resolution Protocol (ARP) working in
conjunction with Private VLAN Edge to minimize broadcasts and maximize available
bandwidth.
ix. The switch should support MVR (Multicast VLAN Registration)
f. L3 Features
i. The switch should support Inter-VLAN routing
ii. The switch should support IPv4 unicast Static Routing
iii. The switch should support 16 IPv4 Static routes
g. Configuration Operations
i. The switch should support configuration of the Software image and switch configuration
without user intervention
ii. The switch should support automatic configuration as devices connect to the switch port
iii. The switch should support system health checks within the switch
h. Quality of Service (QoS) & Control
i. The switch should support 4 egress queues per port to enable differentiated management
ii. The switch should support scheduling techniques for Qos
iii. The switch should support Weighted tail drop (WTD) to provide congestion avoidance
iv. The switch should support availability of up to 256 aggregate or individual polices per port.
i. Network security features
i. The switch should support IEEE 802.1x to allow dynamic, port-based security, providing user
authentication.
ii. The switch should support Port-based ACLs for Layer 2 interfaces to allow application of
security policies on individual switch ports.
j. IPv6 Features
i. The switch should be on the approved list of IPv6 Ready Logo phase II – Host
ii. The switch should support IPv6 unicast Static Routing
iii. The switch should support 16 IPv6 Static routes
iv. The switch should support IPv6 MLDv1 & v2 Snooping
6. Detailed Technical Specifications of Basic Layer3 Edge Switch 24 Ports (Schedule Item No.4)
a. Port Required
i. The switch should support a minimum of 24 nos. 10/100/1000 Ethernet Ports and 2 nos. 1G
SFP ports.
ii. The switch should have total 26 ports
b. Performance and Scalability
i. The switch should support Forwarding bandwidth of 108 Gbps
ii. The switch should support Full-duplex Switching bandwidth of 210Gbps
iii. The switch should support 64-Byte Packet Forwarding Rate of 70 Mpps
iv. The switch should support 1023 VLANs
v. The switch should support Maximum transmission unit (MTU) of 9198 bytes
vi. The switch should support 16000 Unicast MAC addresses
c. Stacking
i. The switch should support Stacking with dedicated stacking ports in addition to 26 ports.
ii. Stacking module should be Hot-swappable.
iii. The switch should able to stack eight switch in a single stack unit with stacking bandwidth of
80Gbps.
d. Power Supply
i. The switch should support an Redundant Power Supply
e. Layer-2 Features
i. The switch should support Automatic Negotiation of Trunking Protocol, to help minimize the
configuration & errors
ii. The switch should support IEEE 802.1Q VLAN encapsulation
iii. The switch should support Centralized VLAN Management. VLANs created on the Core
Switches should be propagated automatically
iv. The switch should support Spanning-tree PortFast and PortFast guard for fast convergence
v. The switch should support UplinkFast&BackboneFast technologies to help ensure quick
failover recovery, enhancing overall network stability and reliability
vi. The switch should support Spanning-tree root guard to prevent other edge swicthes
becoming the root bridge.
vii. The switch should support Per-port multicast storm control to prevent faulty end stations
from degrading overall systems performance
viii. The switch should support Local Proxy Address Resolution Protocol (ARP) working in
conjunction with Private VLAN Edge to minimize broadcasts and maximize available
bandwidth.
ix. The switch should support MVR (Multicast VLAN Registration)
f. L3 Features
i. The switch should support Inter-VLAN routing
ii. The switch should support IPv4 unicast Static Routing
iii. The switch should support 16 IPv4 Static routes
g. Configuration Operations
i. The switch should support configuration of the Software image and switch configuration
without user intervention
ii. The switch should support automatic configuration as devices connect to the switch port
iii. The switch should support system health checks within the switch
h. Quality of Service (QoS) & Control
i. The switch should support 4 egress queues per port to enable differentiated management
ii. The switch should support scheduling techniques for Qos
iii. The switch should support Weighted tail drop (WTD) to provide congestion avoidance
iv. The switch should support availability of up to 256 aggregate or individual polices per port.
i. Network security features
i. The switch should support IEEE 802.1x to allow dynamic, port-based security, providing user
authentication.
ii. The switch should support Port-based ACLs for Layer 2 interfaces to allow application of
security policies on individual switch ports.
j. IPv6 Features
i. The switch should be on the approved list of IPv6 Ready Logo phase II – Host
ii. The switch should support IPv6 unicast Static Routing
iii. The switch should support 16 IPv6 Static routes
iv. The switch should support IPv6 MLDv1 & v2 Snooping
7. 1000 Base LH/LX Transreceiver
i. Transreceiver Module should be of same make of switch and core switch
ii. Module should be compatible to the quoted switch and core switch
8. 1000 Base SXTransreceiver
i. Transreceiver Module should be of same make of switch and core switch
ii. Module should be compatible to the quoted switch and core switch
Passive Component
11. 6 Core SM Fiber Cable
Cable type
6 Core, Single mode, Armored, loose-tube, CST armour,
Gel Filled.
Fiber Type
9/ 125, Telcordia’s GR-20 and ITU-T 652.D Compliance.
Cable Construction will be BELLCORE GR 20 / IEC 794-1.
Attenuation
@1310nm < = 0.35 dB/Km
@1550nm < = 0.20 dB/Km
Coating / Cladding
non-circularity<= 12 microns.
Zero Dispersion Slope should be <= 0.086 ps / sqnm-km.
Max (chromatic) dispersion should be
<5.3 ps/nm-km @1270-1340 nm
<3.5 ps/nm-km @1285-1330 nm
<185 ps/nm-km @1550 nm
Tensile
rating --- 1000N
Armoring
Corrugated Steel tape Armor.
jacket
Inner:- High density polyethylene.
Outer :- High density polyethylene, anti - termite, anti rodent suitable for direct burial application.
Secondary Buffer Material should be Gel filled Loose Tube.
Min Bend 20 X Outer Diameter.
Test
Test (Must pass) : IEC794-1-E1 , IEC794-1-E2 , IEC794-1-E3
, IEC794-1-E4 , EIA-455-104 , IEC794-1-E7 , IEC794-1-E10 ,
IEC794-1-F1 , IEC794-1-F3 and IEC794-1-F5
Marking
Marking should be Identification marking at regular
intervals of 1 meter.
Fiber Core must be Raw fiber of corning. CORNING
marking should be visible on the OFC.
Resistance
Maximum Crush resistance- 44 N/mm
•
12. LIU,6 Port Rack Mount enclosure,
● 1U, 19 Inch Rack mountable Metallic Distribution Enclosure with Adaptor Plate and
Splicing Tray with air ventilation provision on the enclosure.
● Should have optimum air circulation (air rack version)
● 19 “ Rack Mountable Cabinet
● Complete Aluminium Alloy housing, fully powder coated
● Panel cover is of slide out for easy maintenance
● Rubber grommets are provided at the cable entry
● points for tight sealing.
● Reduced mounting depth of 210mm
● Universally compatible (eg ETSI std compliant)
Cable entry from the rear or from the side
13. LIU,24 Port Rack Mount enclosure,
● 1U, 19 Inch Rack mountable Metallic Distribution Enclosure with Adaptor Plate and
Splicing Tray with air ventilation provision on the enclosure.
● Should have optimum air circulation (air rack version)
● 19 “ Rack Mountable Cabinet
● Complete Aluminium Alloy housing, fully powder coated
● Panel cover is of slide out for easy maintenance
● Rubber grommets are provided at the cable entry
● points for tight sealing.
● Reduced mounting depth of 210mm
● Universally compatible (eg ETSI std compliant)
● Cable entry from the rear or from the side
14. LIU,12 Port Rack Mount enclosure,
● 1U, 19 Inch Rack mountable Metallic Distribution Enclosure with Adaptor Plate and
Splicing Tray with air ventilation provision on the enclosure.
● Should have optimum air circulation (air rack version)
● 19 “ Rack Mountable Cabinet
● Complete Aluminium Alloy housing, fully powder coated
● Panel cover is of slide out for easy maintenance
● Rubber grommets are provided at the cable entry
● points for tight sealing.
● Reduced mounting depth of 210mm
● Universally compatible (eg ETSI std compliant)
● Cable entry from the rear or from the side
15. PIGTAIL SC SM SIMPLEX LENGTH- 1m
● Simplex LC/SC, 9/125 micron SM fiber, factory terminated ceramic connectors.
Qualifies as per ITU-T G652 D Fiber.
● Plastic molded plug type connectors
● Specially designed Low Smoke Zero Halogen Outer Sheath
● Standard 2.5mm ceramic ferrule
● Compact in size and easy to connect
● Recommended by EIA/TIA 568 Specifications
● Macrobending, Max. (100 turns) 0.05 dB (1,310/1,550nm@50mm) 0.05 dB
(1,625nm@60mm) Macrobending, Max. 0.05 dB@1,550nm (1 turns@32mm
mandrel).
● Length 1 Mt.
●
16. Optical Fiber Patch Cords SM, - 2Mtr.
Type
Duplex LC/SC, 9/125 micron SM fiber, factory terminated
ceramic connectors. Qualifies as per ITU-T G652 D Fiber
Connector Type Plastic molded plug type connectors
Feature
Standard 2.5mm ceramic ferrule
Compact in size and easy to connect
Recommended by EIA/TIA 568
Macrobending, Max. (100 turns) 0.05 dB (1,310 / 1,550nm
Specification @50mm) 0.05 dB (1,625nm@60mm) Macrobending, Max.
0.05 dB@1,550nm (1 turns@32mm mandrel).
Length
2 Meter
17. Optical Fiber Patch Cords MM, - 2Mtr.
Type
•
•
•
•
Patch Cord MM patch cord LC-SC TYPE
1.6mm or 3mm simplex or Duplex Zipcord.
Fiber Patch cord must have Clear Curve Feature for
more flexibility.
Qualifies as per ITU-T G652 D Fiber
Outside Diameter
•
•
(Simplex): 1.6mm x 3.0mm
(Duplex): 1.6mm x 3.3mm
Minimum
Cable
Retention Strength
•
1.6mm: 11.24 lbs (50 N)
Insertion Loss
Less than 0.5 dB for MM and 0.3 dB for SM
32 upgradation of existing network switches and OFC
Eligibility Criteria
• The bidder should be an exclusively IT company of repute having expertise, sound
financial base and capability to execute said work using state of the art technology with
best industry practices available in the global market. The firm should be in existence and
operating in India for the last Ten Years
• The firm/company should be ISO 9001 certified (Networking & Systems Integration). A
copy of the certificate must be attached, otherwise the bid will be considered as nonresponsive.
• The bidder should be a authorized distributor or large-scale system integrator of the
required networking equipment and he bidder should have Authorization from
concerned Original Equipment Manufacturers (OEMs) for supply and support for all active
and passive components quoted for at least five years (Authorization Certificates to be
attached).
• The bidder should be in the business of supply, installation & commissioning of LANs for
at least five years. Proof of the same must be enclosed with the technical bid.
• Bidder should have proven experience of successfully completing at least five projectsof
LAN networking i.e. supply, installed in, testing and commissioning of LAN work with
value of each project being at least Rs 50 lakhs during last three financial years preferably
in Uttar Pradesh / Uttrakhand. Out of these projects, three projects should have been
done with similar components quoted in the offer.
• Proof of satisfactory completion along with scope of these works to be submitted with
the bid.
• An undertaking by the authorized person of the company (self-certificate) that the bidder
hasn’t been blacklisted by a Central /State Government, institution and there has been no
litigation with any Government Department/PSU/Corporation in Central/State
Government on account of similar services.
• Bidder/OEM must have their 24x7 support center in India preferably in NCR / Uttar
Pradesh. Details & proof of service facilities for Technical Support be attached.
• The bidder should have a minimum turnover of Rs. 5 crores (Rupees five crore only) in
each of the last three years on account of supply, installation and commissioning of
Networks. Relevant proof in the form of audited annual report or balance sheet duly
certified by Chartered Accountant certifying the turn over in the field of Networking
should be submitted.
• All Active Networking Equipment offered and quoted should be from single OEM. Make
and Model must be clearly mentioned in the technical bid. If they are not found in the
technical bid, the bid will be considered as non-responsive.
• Bidder must submit the MAF from the OEM’s of all Active Networking Equipment and
Passive Networking Equipment for participating in the bid and support commitment from
OEM for the Warranty period.
• The bidder shall not assign or sub contract, in whole or in part, its obligations to perform
under the Contract.
• All relevant documents must be enclosed with the tender document at the time of bid
submission. Committee will not entertain any documents submitted later by the bidder at
the time of Bid Evaluation.
• OEM for active components should be Leaders (First Quadrant)in Gartner's Magic
Quadrant year 2013.
Intent of Specification
• Established in 1889, the Indian Veterinary Research Institute (IVRI) is one of the premier
research institutions dedicated to livestock research and development of the region. The
institute with faculty strength of more than 275 has a major mandate of research, teaching,
consultancy and technology transfer activities. The institute imparts quality post-graduate
education to students not only from various parts of the country, but also from overseas.
• Today, the institute with its deemed to be university status contributes immensely to human
resource development in the discipline of veterinary sciences with skills and knowledge
necessary for the challenges of the new millennium. It awards degrees to master and
doctoral programmes in more than 20 disciplines of veterinary and animal sciences, livestock
products technology, basic sciences and extension education.
• Institute is having more than 25 buildings in its campus and all of the buildings are connected
using multimode and single mode OFC in star topology.
• Most of the network switches are L2 switches of Nortel Make.
Scope of Work
• The scope of this tender is to upgrade the existing LAN infrastructure in IVRI Izatnagar.
• Scope of work comprises of supply, installation, configuration, testing and commissioning of
active network components at various locations at IVRI Izatnagar
• We need the re-laying and termination of 6 Core Single Mode OFC from ARIS Cell to Y-Shape
building, Veterinary Polyclinic, Library, P & C Building, Animal Nutrition Division and Central
Office directly.
• All the existing active network components (Switches, fiber modules, fiber patch cord etc) of
ARIS Cell, Y-Shape building , Veterinary Polyclinic, Library, P & C Building, Animal Nutrition
Division,Central Office, CADRAD, BP Division, Standardization Division, JDR Office,
Biochemistry Building, KVK, Extension Education Building, and Parasitology Building are to be
replaced with 10Gbps uplinks and all the end user port should have connectivity of
10/100/1000 Mbps.
• The existing equipments which have its life will have to be effectively utilized in the network.
• Bidder has to provide powering on equipment after ensuring correctness of terminations,
interfaces and power supply and making the system ready for testing and commissioning.
• Testing of LAN Cables after laying , terminations and ferruling at both the ends. All testing
tools and instruments shall be brought by the bidder and taken back after the testing. Cabling
work shall also be got certified from OEM.
• Configuration of the equipment as per the requirements of IVRI including Network
segmentation and Network Monitoring
• Configuration of VLANs as per requirements of IVRI
• Field testing and commissioning of system, including integration with existing Network for
internet connectivity.
• Site acceptance tests to establish satisfactory performance of the equipment's as per specs.
• Assistance for familiarization and operation of the installed system & services for 3 months
after acceptance of system.
• In case, the quantity of cables or fixing wall mount sockets etc. exceeds or is less than the
quantity in bid price schedule, the payment for the executed quantity shall be paid on prorata basis, for the actual quantities for which the installation is carried out through the Bidder
on Certification by Site Engineer.
• The successful bidder will be responsible for upgradation of network components
asmentioned under ‘Scope of Work’ part and ensure enhanced smooth operation of the
networkthereafter. After the upgradation work, the upgraded equipments have to be
integrated with theIVRI running network to ensure enhanced output.
• Any fault reported in the existingequipments/network due to this upgradation work, has to
be taken care by the bidder
•
A sketch showing Existing LAN Network at IVRI and proposed OFC re-laying diagram is
enclosed as part of this Tender Document.
• However, the alternate suitable proposal may be proposed by the bidder with full reasoning.
• The bidder shall also provide all required equipment which may not be specifically stated
herein but are required to meet the intent of ensuring completeness, maintainability and
reliability of the total system covered under this specification, including integration and
interoperability with the existing LAN.
General Technical Requirements
• All active LAN components shall be of the same Make/manufacturer and shall be covered
under same back-up guarantee from the same OEM, to ensure full compatibility, interworking and inter-operability.
• Equipment furnished shall be complete in every respect with all mountings, fittings, fixtures
and standard accessories normally provided with such equipment's and/or needed for
erection, completion and safe operation of the equipment's as required by applicable codes
though they may not have been specifically detailed in the technical specification, unless
included in the list of exclusions. All similar standard components/parts of similar standard
equipment's provided, shall be inter-changeable with one another.
• The Bidder shall be responsible for providing all materials, equipment's, and services,
specified or otherwise, which are required to fulfill the intent of ensuring operability,
maintainability, and reliability of the complete equipment covered under this specification
within his quoted price. This work shall be in compliance with all applicable standards,
statutory regulations and safety requirements in force of the date of award of this contract.
• The bidder shall also be responsible for deputing qualified personnel for installation, testing,
commissioning and other services under his scope of work as per this specification. All
required tools and tackles for completing the scope of work as per the specification is also
the responsibility of the bidder.
• The bidder shall also be responsible for deputing qualified personnel for installation, testing,
commissioning and other services under his scope of work as per this specification. All
required tools and tackles for completing the scope of work as per the specification is also
the responsibility of the bidder.
• The Bidder shall furnish two complete sets of system and equipment instruction manuals and
detailed installation operating and maintenance documentation along with the CDs for the
same. Manuals shall describe system operation, shall include detailed system and
components description and shall cover the installation, operation, care and maintenance of
all system components including diagnostics. Complete parts lists including detailed
instructions for ordering shall be furnished. Complete documentation for operating, system
software, application programs, software selectable functions, shall be provided. After
installation "as built" drawings shall be provided by the bidder indicating the cable routes,
termination & termination details, actual configuration & parameters, etc.
Warranty and AMC
• The bidder warrants that the Goods supplied under this Contract are new, unused, of the most
recent or current models and that they incorporate all recent improvements in design and
materials unless provided otherwise in the Contract.
• This warranty shall remain valid for a period of a minimum 36 months after the Goods or any
portion thereof as the case may be, have been delivered, commissioned and accepted at the
final destination indicated in the Contract.
• The bidder will submit the rates of AMC for five year after the expiry of warranty period.
• The bidder shall post a well-qualified and experienced Service Engineer full time at the site at his
own expense and get the defects/bugs removed in the system as detected by the Purchaser
during the period of warranty.
Documentation
The bidder shall supply the following documents at the time of Complete set of Technical/Operation and
Maintenance Manual
• An inventory of items delivered.
• Rack wise inventory
• Node & connectivity details
• Switch installation Report
• UTP/OFC Wiring diagram
• Certification details for UTP & OFC connections Acceptance Test:
Schedule for Requirements
SNo
1
2
3
4
5
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
22.
9.
10.
11.
12.
13.
14.
15.
16.
Items Required
UOM
ACTIVE COMPONENTS
Layer3 Modular Core Switch Chassis with One supervisor switching line
card
Basic Layer3 Switch having minimum 48 Nos. Ethernet 10/100/1000
BaseT and 2 nos. 1000BaseX SFP Ports
Basic Layer3 Switch having minimum 24 Nos. Ethernet 10/100/1000
BaseT and 2 nos. 1000BaseX SFP Ports
1000 Base LH/LX Transreceiver
1000 Base SX Transreceiver
PASSIVE COMPONENTS
6 Core SM Fiber Cable
Rack Mount 6 Port LIU Loaded with Couplers
Rack Mount 24 Port LIU Loaded with Couplers
Rack Mount 12 Port LIU Loaded with Couplers
Pigtals SM Simplex
SC-LC Fiber Patch Cord SM – 2mtr
Fiber Patch Cord MM – 2mtr
Adaptor SC SM -ZR-SLEEVE Simplex
HDPE Pipe 32 MM
19" 12 U Rack with Cable Manager, Power Strip
19" 42 U Open Rack with Cable Manager, Dual Power
Strip ( For Server and Network)
JOBWORK
Digging of Soft Soil
Digging of Hard Soil
Laying of Fiber Cable in HDPE Pipe
Root Marker Cemented
Splicing of Fiber
LIU Fixing and Dressing
Rack Fixing and Dressing including Fixing of Jack Panels,
Cable Laying and Ferruling etc.
Documentation and Configuration 1 Job
Qty
No.
1
No.
8
No.
8
Nos.
Nos.
15
8
Mtr.
Nos.
No.
No.
Nos.
Nos.
Nos.
No.
Mtr.
Nos.
5000
4
2
2
108
12
15
108
3000
6
No.
1
Mtr.
Mtr.
Mtr.
Nos.
Nos.
Nos.
Nos.
2500
500
5000
20
108
8
6
Job
1
Detailed Technical Specifications
Active Component
9. Layer3 Modular Core Switch Chassis with One supervisor switching line card
a. Chassis specification and Feature
i. The chassis should be modular with minimum 7 slots, minimum 5 payload slots and minimum
2 supervisor switching line card slots.
ii. Switch should have dual hot-swappable power supplies, a hot-swappable fan tray with
redundant fans.
iii. Shall have Minimum 48 Nos. of 10/100/1000 Mbps TX Ports
iv. Shall have Minimum 24 Nos. of 1000 Base X Ports Distributed over 2 line cards
v. Shall have Minimum 8 Nos. of 10G SFP+/XFP Ports on the CPU or in line card
vi. It should support around five microseconds of latency to and from any 10 Gigabit Ethernet
port.
vii. Switch Must support IPv4 and IPv6 switching and routing in hardware
viii. PoE Plus capabilities on all ports in a line card simultaneously
ix. Supplied line card should have at least four built-in 10 GE SFP plus ports.
b. Performance
i. Switch must have minimum 900Gbps Switching capacity or more with 250 Mpps throughput
for IPv4 and 125 Mpps for IPv6 .
ii. Switch must support 48 Gbps Per-slot Switching Capacity
iii. Support for 256,000 ( IPv4 ) & 128,000 ( IPv6 ) entries in routing table (shared between
unicast and multicast)
iv. IPv6 support in hardware, providing wire-rate forwarding for IPv6 networks
v. Dual stack support for IPv4/IPv6 and Dynamic hardware forwarding-table allocations for ease
of IPv4-toIPv6 migration
vi. No performance penalty with advanced Layer 3 and Layer 4 services enabled
vii. Switch should supportminimum of 300 non-blocking GE ports. 100 GE of SFP plus ports,
viii. Switch Must be equipped with minimum 2GB RAM, 4 GB Flash / SDRAM
ix. Switch should support scalability upto 4000 virtual ports (VLAN port instances)
x. Switch should support minimum 55K MAC Address
xi. Bandwidth aggregation up to 16 Gbps through Port Aggregation.
xii. Hardware-based multicast management & Hardware-based ACLs
c. Layer 2 Features
i. The switch should support 8 bi-directional SPAN and/or RSPAN sessions
ii. IEEE 802.1Q VLAN encapsulation, Support for minimum 4096 VLANs per switch
iii. Support for Automatic Negotiation of Trunking Protocol, to help minimize the configuration
& errors.
iv. Centralized VLAN Management. VLANs created on the Core Switches should be propagated
automatically.
v. Spanning-tree PortFast and PortFast guard for fast convergence with STP instances of
minimum 2500.
vi. Switch should support 802.1d, 802.1s, 802.1w, 802.3ad, 802.3af Power over Ethernet (PoE)
vii. Spanning-tree root guard to prevent other edge swicthes becoming the root bridge.
viii. IGMP snooping v1, v2 and v3
ix. Port Aggregation across line cards
x. Link Aggregation Protocol (LACP)
xi. Support for Detection of Unidirectional Links and to disable them to avoid problems such as
spanning-tree loops
xii. Line card switchover should be in subsecond failover time, so that the Layer 2 Links need not
be renegotiated in case of failure of one of the CPU module.
xiii. The Switch should be able to discover the neighboring device of the same vendor giving the
details about the platform, IP Address, Link connected through etc, thus helping in
troubleshooting connectivity problems.
d. Layer 3 Features
i. IP routing protocols Open Shortest Path First [OSPF], Routing Information Protocol [RIP],
RIP2), Intermediate System to Intermediate System (IS-IS), Border Gateway Protocol Version
4 (BGP4) and Multicast Border Gateway Protocol (MBGP), ICMP Router Discovery Protocol,
Policy-Based Routing (PBR)
ii. Support for Router redundancy Protocol
iii. Multicast entries should be minimum of 65,000 ( IPv4 ) and 60,000 ( IPV6 ).
iv. The Switch should be Non-Stop Forwarding (NSF)-Aware for most of the dynamic routing
protocols.
v. Software routing of Internetwork Packet Exchange (IPX) and AppleTalk for legacy protocols
vi. IP multicast routing protocols (PIM, SSM, Distance Vector Multicast Routing Protocol
[DVMRP]), Pragmatic General Multicast (PGM)
vii. IGMP v1, v2, and v3, IGMP filtering on access and trunk ports
viii. Support for Virtual Route Forwarding feature to map VLANs to respective VRFs
ix. Support for IPv6 Protocol.
e. Quality of Service
i. QoS hardware entries should be minimum of 100,000.
ii. Per-port quality of service (QoS) configuration, Support for four queues per port in hardware
iii. Strict priority queuing,IP differentiated service code point (DSCP) and IP Precedence
iv. Classification and marking based on IP type of service (TOS) or DSCP, full Layer 3 and Layer 4
headers
v. Input and output policing based on Layer 3 and Layer 4 headers. Support for 1024 policers on
ingress and 1024 policers on egress configured as aggregate or individual
vi. Shaping and sharing output queue management, Support for Congestion avoidance feature
like DBL (Dynamic Buffer Limiting)
vii. No performance penalty for granular QoS functions
viii. Support for Automatic Quality of Service for easy configuration of QoS features for critical
applications.
f. Security Features
i. Support for Admission Control features for end-point security (PCs & Desktop Posture
Assessment) and thereby controlling the spread of viruses & worms
ii. RADIUS, which allow centralized control of the switch and restrict unauthorized users from
altering the configuration
iii. Standard and extended ACLs on all ports
iv. 802.1x user authentication (with VLAN assignment and Guest VLAN extensions)
v. Router ACLs (RACLs) on all ports (no performance penalty), VLAN ACLs (VACLs),Port ACLs
(PACLs)
vi. Private VLANs (PVLANs) on access and trunk ports
vii. Dynamic Host Configuration Protocol (DHCP) snooping and Option82 insertion
viii. Port Security, Secure Shell (SSH) Protocol versions 1 and 2
ix. Unicast MAC filtering, Unicast port flood blocking
x. Dynamic Address Resolution Protocol (ARP) inspection, IP source guard
g. Management Features
i. The switch should support Flexible NetFlow and IP SLA for enhanced visibility.
ii. Switch should have the feature of EnergyWise for simplified power management
iii.
iv.
v.
vi.
Configuration Rollback for improved configuration management
Single console port and single IP address to manage all features of the system
Software-configuration management, including local and remote storage
Optional Compact Flash memory card to store software images for backup and easy software
upgrades
vii. Manageability through common network-management software on a per-port and perswitch basis, providing a common management interface for routers, switches of the same
vendor
viii. Support for SNMP versions 1, 2, and 3
ix. Remote Monitoring (RMON) software agent to support four RMON groups (history, statistics,
alarms, and events) for enhanced traffic management, monitoring, and analysis. Support for
all nine RMON groups through the use of a mirrored port, which permits traffic monitoring of
a single port, a group of ports, or the entire switch from a single network analyzer or RMON
probe
x. Analysis support, including ingress port, egress port, and VLAN mirroring
xi. Support for sFlow / NetFlow or equivalent for Deep traffic & session flow analysis.
xii. Support for Port Mirroring, Remote Mirroring & ACL filtering for Port mirroring used for IDS
functionalities.
10. Basic Layer3 Switch having minimum 48 Nos. Ethernet 10/100/1000 Base T and 2 nos. 1000BaseX
SFP Ports
a. General Features
i. The switch should support a minimum of 48 nos. 10/100/1000 Ethernet Ports and 2 nos. 1G
SFP ports.
b. Performance and Scalability
i. The switch should support Forwarding bandwidth of 108 Gbps
ii. The switch should support Full-duplex Switching bandwidth of 216 Gbps
iii. The switch should support 64-Byte Packet Forwarding Rate of 130 Mpps
iv. The switch should support 1023 VLANs
v. The switch should support Maximum transmission unit (MTU) of 9198 bytes
vi. The switch should support 16000 Unicast MAC addresses
c. Stacking
i. The switch should support Stacking with dedicated stacking ports in addition to 26 ports.
ii. Stacking module should be Hot-swappable.
iii. The switch should able to stack eight switch in a single stack unit with stacking bandwidth of
80Gbps.
d. Power Supply
i. The switch should support an Redundant Power Supply
e. Layer-2 Features
i. The switch should support Automatic Negotiation of Trunking Protocol, to help minimize the
configuration & errors
ii. The switch should support IEEE 802.1Q VLAN encapsulation
iii. The switch should support Centralized VLAN Management. VLANs created on the Core
Switches should be propagated automatically
iv. The switch should support Spanning-tree PortFast and PortFast guard for fast convergence
v. The switch should support UplinkFast&BackboneFast technologies to help ensure quick
failover recovery, enhancing overall network stability and reliability
vi. The switch should support Spanning-tree root guard to prevent other edge swicthes
becoming the root bridge.
vii. The switch should support Per-port multicast storm control to prevent faulty end stations
from degrading overall systems performance
viii. The switch should support Local Proxy Address Resolution Protocol (ARP) working in
conjunction with Private VLAN Edge to minimize broadcasts and maximize available
bandwidth.
ix. The switch should support MVR (Multicast VLAN Registration)
f. L3 Features
i. The switch should support Inter-VLAN routing
ii. The switch should support IPv4 unicast Static Routing
iii. The switch should support 16 IPv4 Static routes
g. Configuration Operations
i. The switch should support configuration of the Software image and switch configuration
without user intervention
ii. The switch should support automatic configuration as devices connect to the switch port
iii. The switch should support system health checks within the switch
h. Quality of Service (QoS) & Control
i. The switch should support 4 egress queues per port to enable differentiated management
ii. The switch should support scheduling techniques for Qos
iii. The switch should support Weighted tail drop (WTD) to provide congestion avoidance
iv. The switch should support availability of up to 256 aggregate or individual polices per port.
i. Network security features
i. The switch should support IEEE 802.1x to allow dynamic, port-based security, providing user
authentication.
ii. The switch should support Port-based ACLs for Layer 2 interfaces to allow application of
security policies on individual switch ports.
j. IPv6 Features
i. The switch should be on the approved list of IPv6 Ready Logo phase II – Host
ii. The switch should support IPv6 unicast Static Routing
iii. The switch should support 16 IPv6 Static routes
iv. The switch should support IPv6 MLDv1 & v2 Snooping
11. Detailed Technical Specifications of Basic Layer3 Edge Switch 24 Ports (Schedule Item No.4)
a. Port Required
i. The switch should support a minimum of 24 nos. 10/100/1000 Ethernet Ports and 2 nos. 1G
SFP ports.
ii. The switch should have total 26 ports
b. Performance and Scalability
i. The switch should support Forwarding bandwidth of 108 Gbps
ii. The switch should support Full-duplex Switching bandwidth of 210Gbps
iii. The switch should support 64-Byte Packet Forwarding Rate of 70 Mpps
iv. The switch should support 1023 VLANs
v. The switch should support Maximum transmission unit (MTU) of 9198 bytes
vi. The switch should support 16000 Unicast MAC addresses
c. Stacking
i. The switch should support Stacking with dedicated stacking ports in addition to 26 ports.
ii. Stacking module should be Hot-swappable.
iii. The switch should able to stack eight switch in a single stack unit with stacking bandwidth of
80Gbps.
d. Power Supply
i. The switch should support an Redundant Power Supply
e. Layer-2 Features
i. The switch should support Automatic Negotiation of Trunking Protocol, to help minimize the
configuration & errors
ii. The switch should support IEEE 802.1Q VLAN encapsulation
iii. The switch should support Centralized VLAN Management. VLANs created on the Core
Switches should be propagated automatically
iv. The switch should support Spanning-tree PortFast and PortFast guard for fast convergence
v. The switch should support UplinkFast&BackboneFast technologies to help ensure quick
failover recovery, enhancing overall network stability and reliability
vi. The switch should support Spanning-tree root guard to prevent other edge swicthes
becoming the root bridge.
vii. The switch should support Per-port multicast storm control to prevent faulty end stations
from degrading overall systems performance
viii. The switch should support Local Proxy Address Resolution Protocol (ARP) working in
conjunction with Private VLAN Edge to minimize broadcasts and maximize available
bandwidth.
ix. The switch should support MVR (Multicast VLAN Registration)
f. L3 Features
i. The switch should support Inter-VLAN routing
ii. The switch should support IPv4 unicast Static Routing
iii. The switch should support 16 IPv4 Static routes
g. Configuration Operations
i. The switch should support configuration of the Software image and switch configuration
without user intervention
ii. The switch should support automatic configuration as devices connect to the switch port
iii. The switch should support system health checks within the switch
h. Quality of Service (QoS) & Control
i. The switch should support 4 egress queues per port to enable differentiated management
ii. The switch should support scheduling techniques for Qos
iii. The switch should support Weighted tail drop (WTD) to provide congestion avoidance
iv. The switch should support availability of up to 256 aggregate or individual polices per port.
i. Network security features
i. The switch should support IEEE 802.1x to allow dynamic, port-based security, providing user
authentication.
ii. The switch should support Port-based ACLs for Layer 2 interfaces to allow application of
security policies on individual switch ports.
j. IPv6 Features
i. The switch should be on the approved list of IPv6 Ready Logo phase II – Host
ii. The switch should support IPv6 unicast Static Routing
iii. The switch should support 16 IPv6 Static routes
iv. The switch should support IPv6 MLDv1 & v2 Snooping
12. 1000 Base LH/LX Transreceiver
i. Transreceiver Module should be of same make of switch and core switch
ii. Module should be compatible to the quoted switch and core switch
13. 1000 Base SXTransreceiver
i. Transreceiver Module should be of same make of switch and core switch
ii. Module should be compatible to the quoted switch and core switch
Passive Component
18. 6 Core SM Fiber Cable
Cable type
6 Core, Single mode, Armored, loose-tube, CST armour,
Gel Filled.
Fiber Type
9/ 125, Telcordia’s GR-20 and ITU-T 652.D Compliance.
Cable Construction will be BELLCORE GR 20 / IEC 794-1.
Attenuation
@1310nm < = 0.35 dB/Km
@1550nm < = 0.20 dB/Km
Coating / Cladding
non-circularity<= 12 microns.
Zero Dispersion Slope should be <= 0.086 ps / sqnm-km.
Max (chromatic) dispersion should be
<5.3 ps/nm-km @1270-1340 nm
<3.5 ps/nm-km @1285-1330 nm
<185 ps/nm-km @1550 nm
Tensile
rating --- 1000N
Armoring
Corrugated Steel tape Armor.
jacket
Inner:- High density polyethylene.
Outer :- High density polyethylene, anti - termite, anti rodent suitable for direct burial application.
Secondary Buffer Material should be Gel filled Loose Tube.
Min Bend 20 X Outer Diameter.
Test
Test (Must pass) : IEC794-1-E1 , IEC794-1-E2 , IEC794-1-E3
, IEC794-1-E4 , EIA-455-104 , IEC794-1-E7 , IEC794-1-E10 ,
IEC794-1-F1 , IEC794-1-F3 and IEC794-1-F5
Marking
Marking should be Identification marking at regular
intervals of 1 meter.
Fiber Core must be Raw fiber of corning. CORNING
marking should be visible on the OFC.
Resistance
Maximum Crush resistance- 44 N/mm
•
19. LIU,6 Port Rack Mount enclosure,
● 1U, 19 Inch Rack mountable Metallic Distribution Enclosure with Adaptor Plate and
Splicing Tray with air ventilation provision on the enclosure.
● Should have optimum air circulation (air rack version)
● 19 “ Rack Mountable Cabinet
● Complete Aluminium Alloy housing, fully powder coated
● Panel cover is of slide out for easy maintenance
● Rubber grommets are provided at the cable entry
● points for tight sealing.
● Reduced mounting depth of 210mm
● Universally compatible (eg ETSI std compliant)
Cable entry from the rear or from the side
20. LIU,24 Port Rack Mount enclosure,
● 1U, 19 Inch Rack mountable Metallic Distribution Enclosure with Adaptor Plate and
Splicing Tray with air ventilation provision on the enclosure.
● Should have optimum air circulation (air rack version)
● 19 “ Rack Mountable Cabinet
● Complete Aluminium Alloy housing, fully powder coated
● Panel cover is of slide out for easy maintenance
● Rubber grommets are provided at the cable entry
● points for tight sealing.
● Reduced mounting depth of 210mm
● Universally compatible (eg ETSI std compliant)
● Cable entry from the rear or from the side
21. LIU,12 Port Rack Mount enclosure,
● 1U, 19 Inch Rack mountable Metallic Distribution Enclosure with Adaptor Plate and
Splicing Tray with air ventilation provision on the enclosure.
● Should have optimum air circulation (air rack version)
● 19 “ Rack Mountable Cabinet
● Complete Aluminium Alloy housing, fully powder coated
● Panel cover is of slide out for easy maintenance
● Rubber grommets are provided at the cable entry
● points for tight sealing.
● Reduced mounting depth of 210mm
● Universally compatible (eg ETSI std compliant)
● Cable entry from the rear or from the side
22. PIGTAIL SC SM SIMPLEX LENGTH- 1m
● Simplex LC/SC, 9/125 micron SM fiber, factory terminated ceramic connectors.
Qualifies as per ITU-T G652 D Fiber.
● Plastic molded plug type connectors
● Specially designed Low Smoke Zero Halogen Outer Sheath
● Standard 2.5mm ceramic ferrule
● Compact in size and easy to connect
● Recommended by EIA/TIA 568 Specifications
● Macrobending, Max. (100 turns) 0.05 dB (1,310/1,550nm@50mm) 0.05 dB
(1,625nm@60mm) Macrobending, Max. 0.05 dB@1,550nm (1 turns@32mm
mandrel).
● Length 1 Mt.
●
23. Optical Fiber Patch Cords SM, - 2Mtr.
Type
Duplex LC/SC, 9/125 micron SM fiber, factory terminated
ceramic connectors. Qualifies as per ITU-T G652 D Fiber
Connector Type Plastic molded plug type connectors
Feature
Standard 2.5mm ceramic ferrule
Compact in size and easy to connect
Recommended by EIA/TIA 568
Macrobending, Max. (100 turns) 0.05 dB (1,310 / 1,550nm
Specification @50mm) 0.05 dB (1,625nm@60mm) Macrobending, Max.
0.05 dB@1,550nm (1 turns@32mm mandrel).
Length
2 Meter
24. Optical Fiber Patch Cords MM, - 2Mtr.
Type
•
•
•
•
Patch Cord MM patch cord LC-SC TYPE
1.6mm or 3mm simplex or Duplex Zipcord.
Fiber Patch cord must have Clear Curve Feature for
more flexibility.
Qualifies as per ITU-T G652 D Fiber
Outside Diameter
•
•
(Simplex): 1.6mm x 3.0mm
(Duplex): 1.6mm x 3.3mm
Minimum
Cable
Retention Strength
•
1.6mm: 11.24 lbs (50 N)
Insertion Loss
Less than 0.5 dB for MM and 0.3 dB for SM
33
High Speed Refrigerated Centrifuge
1. Microprocessor controlled.
2. Max speed 18000 rpm
3. Max. Volume 6x 250 ml
4. Temperature range -20 to 400C
5. Pre-selection of time 30 sec to 99 hours and continuous run
6. Large LCD display
7. Maintenance free induction drive
8. Motorize lid cock
9. Automatic rotor identification with over speed protection
10. Active imbalance identification
11. CFC free refrigeration
12. Permanent indication for actual &preset values
13. 10 accelerations / declaration
14. Quick key for short run
15. Stand still pre-cooling of chamber
16. Selection of both RPM and RCF
17. Audible indication of end of run
18. Self diagnostic for errors.
19. Specifications must be supported by original printed catalogue and list of users with detailed address
including phone, fax and email.
20. Warranty: Minimum 02 years from the date of installation Rotors:
21. Fix angle rotor for 12 x 1.5 / 2.0 ml with max speed upto 18000 rpm
22. Fix angle rotor for 8 x 50 ml with max speed up to 15000 rpm and adaptor for 15 ml falcon tubes
23. Swing out rotor 4 places, adaptors for 15 ml and 50 ml tubes with speed up to 4500 rpm.
24. Voltage Stabilizer-compatible
34
Specifications of Gradient PCR with UPS
• High performance modular Gradient Thermal Cycler Units with capability to drive additional satellite
thermal cycler modules.
• Assymetrical independently controllable universal dual block with 64 and 32 wells.
• Block should accommodate PCR tube strips, 0.2ml and 0.5ml PCR tubes or Eppendorf divisible PCR
plate
• Should have 12o C gradient range
• The blocks should have latest eight peltier based thermal elements for better uniformity and wider
gradient optimization temperature range
• Should use patented Triple Circuit Technology ensuring precise control of temperature.
• Internal auto validation before the start of the program
• Block Temp. Range: 4-99 deg C
• Temp. uniformity: better than +/- 0.4 deg C
• Temp. Gradient: Linear Gradient spread : 3-30 deg C
• Temp. Range for Gradient range: 30-99 deg C
• Lid temp. range: 37-110 deg C
• Heating Ramp Rate: Upto 3.5 deg C /sec
• Cooling Ramp Rate: Upto 2.5 deg C/ sec
• Traceability: Calibration using NIST traceable standards
• Lid descent and closing pressure must use Flexlid technology with Thermal sample
• Option to select Power On resume ON/OFF facility should be available (user selectable option)
• Communication interfaces : through USB ports, Ethernet, CAN in, CAN out
• Should have two USB ports: for Protocol transfer, Self-test, USB, Printer/mouse
• An intelligent program creation wizard driven software with the ability to automatically create a
protocol including all steps of PCR and number of cycles
• Should have Administrator and user login with or without PIN for enhanced security
• Possible to do TouchDown PCR, Incremental/Decremental, Automatic/Manual Hot Start
• Memory: more than 1000 programs & further extendable through memory sticks
• Should have log book function for error messages and new calibration
• Customized programming allows a maximum of 20 steps and 99 cycles
• Power save standby function
• System should capable of connecting Two additional Mastercycler nexus PCR Machines for ultimate
throughput
• GLP compliant record of each run should be recorded automatically
• Possible to drive the system through PC/Laptop
• Calibration according to NIST (USA), DKD/PTB (Germany) UKAS/NPL (UK), UL/cUL listed
• Should comply to RoHS (2011/65/EU)
• Should be supplied with suitable online UPS (2h backup) & Laptop Computer
• Maximum power consumption: 700 W
• Should come with a 2 year warranty
• UPS: Online/Compatible
35.
Portable Diagnostic Battery Operated Diagnostic Color Doppler Ultrasound Scanner with Printer
• Equipment should be single man usable, portable and lightweight exclusively for veterinary use on
Bovine, Equine, Porcine, Canine, Ovine & Feline
• Equipment should come with inbuilt color Doppler & pulse wave Doppler
• Equipment should work on applications viz. general, abdominal, gynecology, obstetrics, small parts,
superficial, musculoskeletal among others
• Equipment should be easy to carry and light weight, preferably between 3.0kg to 3.5kg rang including
weight of inbuilt, interchangeable and user extractable battery
• Equipment should have battery back-up of minimum 5 working hours
• Equipment should be useful for immediate use, boot up time of maximum 20 seconds including inbuilt
facility to restart without boot-up
• Equipment should have minimum 8" TFT LCD monitor with VGA Format Display size 640x480 pixels
• Equipment should have inbuilt video recording facility with recording area size 640 x 480 pixels
• Equipment should have adjustable tilt with stereo sound via inbuilt facility with integrated
loudspeaker
• Equipment should have digital brightness adjustment, MI (mechanical index) & TI (thermal index)
• Equipment should have following multi-frequency probes with facility to interchange the probe
without restarting the scanner to prevent any dynamic image loss:
o Wideband convex probe of central frequency 3.5 MHz, B mode frequencies of 2.5 MHz, 3.5 MHz & 5.0
MHz for applications on musculo-squelletic, cardiac examination on bovine
& equine, abdominal examination & parts on canine, porcine, ovine o Wideband Linear rectal probe
with central frequency 7.5 MHz, probe bandwidth of 5MHz to 10MHz, B mode frequencies of 5.0 MHz,
7.5 MHz & 10.0 MHz for applications on obstetrics & gynecology on large animals
• Equipment should have the following
1. Color Doppler and pulse wave Doppler as part of the standard configuration with color
Doppler having velocity, color invert with on/off feature
2. Color Doppler with automatic CFM focus, adjustable and selectable area of interest,
position, size of minimum 0–120 gain
3. Selectable and adjustable pulse wave gate position, gate size, frequency & gain from 0–
21
4. Adjustable baseline, adjustable audio, simplex mode and automatic measurements in
pulse wave mode with user convertible presets & protocols
5. Input and output options on the scanner device itself for external screen display, thermal
printer, DVI connection and USB port facility
6. Processing via minimum 32 true emission channels with minimum 64 reception channels
and multi-zone focusing and continuous dynamic focusing in receive with FREE inbuilt
ultrasound software upgrade (FPGA) for total warranty duration of the equipment
7. Alphanumeric physical keyboard, touch-pad, 10 related keys for easy access to operation
menus and audio adjustment
8. B (basic). M (motion), B+B, B+M & M operational modes with adjustable area region of
interest, zoom and full screen zoom function and minimum 5 user selectable focus points
for each focus
9. Cine loop, text & annotations, protocols, automatic and manual measurement and tables
10. Storing facility for clips and images with facility to transfer to PC, laptop and DICOM
feature supplied with minimum 32GB USB Stick
11. Data entry settings for patient name, breed, species, body marks, image storage in .jpeg
& .bmp file formats, video clip storing format
12. Individual user protocols, export feature for images, clips, reports, patient files to/review
from USB, archiving of patient files and reports on USB
13. Carry case with battery recharger, minimum 32GB memory stick, minimum 5 liters
ultrasound gel should be supplied with the machine
• Equipment should be backed by manufacturer certifications complying CE Marked to Council Directive,
Directive 93/42/EEC, ISO 13485 / ISO 14001 / ISO 9001 / EMAS / CE / UL.
Manufacturer certifications to be produced at the time of delivery/installation
• Equipment should be provided with full post sale service back–up onsite and free of charge facility to
train the user on equipment's working and use at the time of installation
36
Bench top incubator cum shaker
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
Bench top orbital shaker with temperature control range from 7°C below ambient to 50°C with
accuracy ±0.1°C.
It should have orbital shaking.
Speed should be 50-350 rpm with accuracy of ±2rpm
It should have acceleration/deceleration circuit to prevent sudden starts/stops.
It should have timer up to 99 Hrs.
It should display speed, running time, alarm etc on display screen.
It should have alarms alert when speed or temperature deviate from set points.
The system should have port for data logging.
It should have automatic restart after power interruption.
It should stop agitation when chamber is open.
It should have universal platform and clamps to holds various flask.
It should be supplied with compatible stabilizer.
Operational voltage should be 220±20V
Should have two years standard warranty/from the date of installation
Should quote for the price of CMC/AMC per year after the expiry of standard warranty.
Should provide user list with phone no and e mail ID
Should provide brochure in original.
37
Agriculture Tractor
Agriculture Tractor Horsepower -40 H.P., Three cylinder, Diesel, Direct injection (D.I.), water cooled,
Diesel fuel, complete with hydraulic, self starter, battery altemator, P.T.O.
38
MULTIDETECTION MICROPLATE READER
 Detection modes - Fluorescence, Time-Resolved Fluorescence (secondary mode),
Luminescence, UV-Visible Absorbance, ELISA
 Microplate types - 6- to 384-well plates; PCR plates
 Other labware- Take3 Micro-Volume Plate
 Temperature control- 4-Zone incubation with Condensation Control Ambient +4 °C to 50 °C;
variation+0.2 °C at 37 °C
 Software - Gen5 Data Analysis Software
 AlphaScreen/AlphaLISA :
Light source - Tungsten halogen; Sensitivity- 300 a mol of biotinylated-LCK-P peptide, 25 µL/well
in 384-well plate; Reading speed- 2 minutes for 96-well plate
 Absorbance :
Light source - Xenon flash; Wavelength selection - Monochromator; Wavelength range - 200 999 nm, 1 nm increments
 Fluorescence Intensity :
Light source- Tungsten halogen; Sensitivity - Top and Bottom: Fluorescein 5 pM (1 fmol/well 96well plate); Wavelength selection- Deep blocking filters; Wavelength range- 300 - 700 nm (200 850 nm option); Sensitivity -Multi-modemodels: 30amolATP (flash); Luminescence only: 10 amol
ATP (flash)
 Luminescence :
Wavelength range - 300 - 700 nm;
 Time-resolved Fluorescence (TRF)
Light source - Xenon flash; Wavelength selection – Monochromator;
 Reagent Dispensers :
Supported detection modes - All modes; Number - 2 syringe pumps; Dispense volume - 5-1000
µL in 1 µL increments ; Dead volume- 1 mL, 100 µL with back flush; Plate geometry- 6- to 384well plates; Dispense precision- <2% at 50 - 200 µL
 Physical Characteristics :
Connectivity- 1 USB, 1 RS232 for external PC control; Power- 100 - 240 Volts AC. 50/60 Hz;
Dimensions- 16"W x 15"D x 10"H (40.6 x 38 x 25.4 cm)
 Computer – Configuration – Core 2 Dual Processor, 1GB RAM, 160 GB
HDD, Keyboard, Mouse, serial / parallel / USB ports and 19”
TFT and Suitable Laser printer.
 ONLINE UPS (Compatible)
39
Freeze Drying Machine
1. Freeze Drying Machine with all essential installation and working accessories like Primary and
Secondary freeze drying assemblies, various rotors to handle different size ampoules/ vials,
rubber nipples to hold different size ampoules, secondary drying manifolds with not less than 48
ports and can extend up to 96 ports, essential adaptors & connectors, suitable Acrylic lid/ cover
02 nos., isolation valve, Quality control as per the standards of S.I. International system.
2. All Stainless steel smooth walled chambers, front mounted, bottom drains, CFC free refrigeration
system, ice condenser ≥ 5 litre capacity, Maximum ice capacity ≥ 4.5 kg; can freeze dry ampoules
(minimum 0.5 ml freeze drying volume) and vials (2.0 ml freeze drying volume). Entire
equipment should be preferably fitted on Movable Trolley. The equipment should maintain
constant vacuum during freeze drying
3. Ice removing capacity; in 12 hrs: ≥ 1.2 kg; in 24 hrs: ≥ 2.2 kg, operating temperature range minus
50 to minus 60 degree Celsius, Fitted with automatic timer to stop spinning after desired
condensation. Temperature display range ambient to minus 60 degree Celsius, Pressure Display
Range > 25μbar and should not be lesser than 50 mbar, weight ≤ 45 kg, supported with suitable
power motor capable to work under Indian electric supply conditions of 220-240 volts/ 50 Hz;
Single Phase. Pump should be preferably either oil free or rotary with two stages advance oil
filters, suitable for the aggressive solvents, pump displacement range ≥ 115 l/ min. Noise level ≤
45-50 dBA
4. Equipped with safety indicators, digital display for condenser temperature and vacuum pressure
preferably with colour coded bars to access real time data.
5. Capable to freeze dry live microbial cultures and the products having Sodium Azide as a
preservative.
6. Glassware accessories – vials 2.0 ml (2000 nos. with suitable stoppers/ caps and sealing
material), Ampoules – 0.5 ml (10,000 nos. with ampoules constrictors & with seal checking
electronic devise). Rubber accessories – 50 number of rubber nipples to hold different size
ampoules.
7. Should be supplied with suitable stabilizer and Service and maintenance warrantee of 03 years
from the date of installation.
8. Tender/ Quotations should be supported with Original brochure keeping details of machine;
accessories and manufacturer.
9. List of users along with their contact details where equipment had been supplied in past 5 years.
10. FOR Installation place/ venue.
40
THE THERMAL CYCLER
Touch Thermal Cycler with superb thermal performance for fast, reliable results. Fully modular platform
should support interchangeable reaction modules, including two optical modules for real time PCR that
swap in seconds without requiring tools. Touch screen with programming methods, including graphical
and automatic (Quick and easy protocol auto writer).
• Interchangeable reaction modules, including 5-colour optical, 4-colour optical, gradient-enabled
dual 48/48- well fast, gradient-enabled 96-well fast, gradient-enabled 96- deep well, and
gradient-enabled 384-well reactions modules.
• USB ports that support peripherals, such as storage device and mouse.
• Optical PC control and networking capability for up 32 systems enable the ultimate in high
throughput.

















Input power – upto 850 W max, programming options
Frequency- 50-60 Hz, single phase, Optional log-in required mode
8.5 inch LCD display and touch screen
Ports - 5 USB A, 1 USB B, Exportable run logs system error logs
Memory- >1,000 typical programs onboard; unlimited with USB flash drive, PC compatibility
Window XP or higher
Upgradable to real time PCR.
Temperature control modes - Calculated and block
PCR license- Yes
Reactions Modules- 96-Well Fast, 96-Deep Well, Dual 48/48 Fast, 384- Well
Sample capacity- 96 × 0.2 ml tubes or 1× 96-well plate , 2 × 48 × 0.2 ml tubes or 2 × 48-well plates,
1× 384-well plate
Maximum ramp rate- 5 0C/sec , 2.5 0C /sec, 4 0C /sec
Average ramp rate - 3.3 0C /sec, 2 0C /sec and 3 0C /sec
Temperature range- 0-100 0C
Temperature accuracy±0.20C of programmed target at 90 0C
Temperature uniformity- ± 0.4 0C well-to-well within 10 sec of arrival at 90 0C
Gradient range- 30-100 0C
Temperature differential range- 1-24 0C
41
Dual Slab
Dual Slab Chamber for slab gel electrophoresis system including 1 or 2 pressure platens + 1
safety cover + power leads , ready to run 1 or 2 gels.
•
Should facilitate high-resolution slab PAGE system for two gels for high resolution
separation of protein and nucleic acid in 2 gels of compact with simple operation.
•
One entire surface of gel sandwich should be in contact with buffer for rapid heat
dissipation and uniform gel temperature. This may afford relatively fast, high
resolution separation, while minimizing possibility of smiling effect, distortion or
diffusion of bands, without the requirement for cooling.
•
A gel should be held simply with unique pressure platen, without use of screws or
clamps, while eliminating the possibility of slippage or leakage. This chamber is
transparent, has no obstructions to visibility, and provides ease of access to sample
wells and observation of separation.
•
Purpose and application of the system include protein separation and identification,
determination of purity, estimation of molecular weight, examination of purified
product and expressed proteins, determination of enzyme form, SDS-PAGE, native
PAGE, DNA-PAGE.
•
Specifications include : Plate size 160 (W) x 160 (L) x 10.5-12 (T) mnm; Gel size 138
(W) x 130 (H) mm for Gel thickness 0.5-2 (D) mnm; Number of gels Up to 2 gels; Plate
setting with pressure platen; Temperature control Single face; Buffer volume 480 mL
for upper chamber, 450 mL for lower chamber, Cover lid, Safety lid with lead wires,
Material Acrylic plastic and stainless steel, except platinum for electrodes, glass for
plates, silicone rubber for gaskets and Teflon for combs; Dimensions 211 (W) x 196
(D) x 205 (H) mm.
Slab Gel Cast, 2-mm 12-well including a pair of glass plates, a gasket, and a comb, ready to
cast a 1 or 2 mm, 12-well gel. Slab gel cast is a gel casting apparatus to ensure leakfree casting of a quality gel. Should be supplied with 1 pair of 16 x 16 cm glass plates,
a U-shaped silicone rubber gasket and a 12- well Teflon comb, to readily cast a 1 or 2
mm thick gel. One of the glass plates, the Plain Glass Plate, is rectangular; another
glass plate the 1 mm or 2 mm Glass Plate, has 1 or 2 mm thick, permanently bonded
glass side-spacers on it, its top is notched, and the notch is bevelled. The glass plates
assembled with 1 or 2 mm U-shaped gasket put between them to seal the sides and
the bottom form a mold to cast a gel of a 13.8 x 13 cm effective area. The gel casting
apparatus consists of a base platen and a cover platen, both of which are made of
clear plastic. The base platen incorporates guide frames respectively for rectangular
glass plate, the gasket and the notched glass plate. So, just layering them along the
guides in the platen being layed aligns them accurately without the possibility of
slippage. Further overlaying cover platen and clamping them with built-in spring
clamps complete a gel cast. The cover platen incorporates 4 compression appliers
utilizing spring force, which exert the right amount of pressure to seal the mold and
eliminate leakage. The gel cast being stood is ready to introduce gel solution. The
base platen has a level mark for easily pouring an adequate volume of gel solution.
The precision components and parts offer the reliability for casting reproducible,
quality gels.
42
Livestock Sprinklers cooling system
(Specification for each unit)
Sl.
Name Article
No.
1
2
3
4
5
6
7
8
Fogger 4 way 28 LPH
(Naan Dan, Israel)
LLDPE Polytube 16 mm
P.V.C. 40 mm X 6 kg/cm2
Control valve 40 mm
Mono block Pump 1.5 H.P.
(Make C.G./Kirlosker) with Starter-control panel Electric
cable - 20 mtr (Pump Head 40 mtr/Discharge-1LPS)
Screen Filter 10m3/hr
Fitting & Accessories with safety covers for (monoblock
pump and line etc)
Complete Installation charges (Civil, electrical & plumbing
work as per site requirement) with freight
Our Requirement is for 16 Units of above specification
Quantity
(Approx)
40 No.
100 mtr
30 mtr
01 no.
1 No.
1
for each
unit
for each
unit
Cost (Rs) per
unit
(Approx.)
280/- per
fogger
10.50/mtr
42/- mtr
280/15000
7000/3500/4000/-
43
Rotavator/Gyrovator
C.
-
Rotavator
Two metres Gear type.
Multi-speed with side wheel.
48 Nos. Blade.
OR
D. Gyrovator
- Two metres Gear type.
- Multi-speed with side wheel.
- 60 Nos. Blade.
- Warranty: One year
44
Deep freezer (-800C)
Model: Vertical freezer
Capacity: >400 L capacity
Controller: Microprocessor controlled, fully programmable with memory backup
Display: Digital
Operating Temperature: < -86 °C
Construction: The inner body surface of stainless steel, polyurethane insulation, insulated inner
door and shelves
7. Safety: Audible and visible alarms for temperature variation, power failure, fault, battery low,
filter clean etc.
8. Compressor: Heavy duty, CFC/HCFC-Free non flammable refrigerants, energy efficient and
hermetically sealed cascade refrigeration system, Operating voltage (190-270 V).
9. Security: Security lock for unauthorized tampering and temp set points
10. Stabilizer: Suitable servo voltage stabilizer
11. Certification: Freezer must have ISO9001 and CE certified.
12. Gloves: A pair of suitable gloves of working with the -800C.
13. Users list: List of 10 recent installations with valid phone numbers of indenters.
14. Warranty: For freezer 2 years and for compressor 5 years from the date of installation.
15. Freezer Rack: Compatible with the model should be quoted as optional.
16. Electrical Operation: 220-240V/50-60Hz
17. Catalogue: Authentic company catalog of the model quoted.
1.
2.
3.
4.
5.
6.
To be quoted as Optional:
1. On-line UPS 5KVA with 02 hours backup
2. CO2 Backup system
45
Inverted Microscope with Fluorescent
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
Body: Microscope body with infinity optical corrected optical system, port for attaching digital camera, multiport option to add optical elements later, Shock less construction
Stage: Plain XY rectangular movable stage with microtiter plate and flask holders.
Condenser: Universal turret condenser with at least 5 positions.
Illumination system: ≥ 100 watt Pre-centred Halogen illumination
Eyepieces: 10X with F.O.V 22 or better and diopter adjustment facility, anti-fungus type.
Nosepiece: Sextuple Revolving Nosepiece.
Objectives: High performance long working distance Objectives suitable for Bright field/Phase Contrast/
DIC/fluorescence Observation with facility of cover glass correction; Achromat Phase 10 X (N.A. 0.25, W.D.
6.20 mm); Plan Fluor Long working Distance Phase 20X (N.A. 0.40, W.D. 8.2-6.9mm), Plan Fluor Long Working
Distance Phase 40x (N.A. 0.60, W.D. 3.6-2.8 mm).
DIC Attachment: DIC accessories i.e. Polarizer, Analyzer and DIC Modules for 20X & 40X objectives.
Observation: Binocular, inclined at 30 degrees.
Software: Suitable software for image acquisition and analysis capable Time Lapse Imaging, Z-Stack,
Annotation, Measurement, counting, ND Viewer, Large Image, Macro, Report Generator facility, Vector layer
& Multi-Dimensional File Format (ND format).
Data collection and processing unit: Branded data processor (Intel i5 or higher, 4 GB RAM, DVD Writer, 500
GB or higher HDD,TFT/LCD display latest operating system (license) and Multimedia Kit with antivirus
software (licence).
Printer: Colour laser printer (1200 X 1200 dpi resolution, 16 MB RAM).
Online UPS: Compatible online branded UPS with at least 1 hour backup
Warranty: At least 2 years from the date of installation of equipment.
Electrical operation: 220-240 volts/ 50-60 Hz
User List: Indian user’s list (minimum 10 recent installations) for the quoted model along with valid contact
details including telephone no. and E. mail of indenters.
Catalogue: Technical features described above should exclusively be supported by authentic original company
catalogue of the instrument/equipment quoted. Photocopy or computer print-out of catalogue will not be
accepted.
Compatibility: Microscope, camera and software should be compatible and provided by same manufacturer.
Optional: (prices to be quoted separately and essentially for each item):
1. Imaging system: Cooled Colour CCD camera with High quality 2/3” CCD chip/sensor, at least 5.0 mega Pixels
resolution in 12 bit per RGB Color channel suitable for handling bright field, fluorescence, DIC, darkfield
images, Having peltier device for cooling up to 10 degree Celsius below ambient temperature. Computer
attachment by USB / Firewire, Live display mode, Display Speed 4.0 frames/sec (at 2560 x 1920 resolution)
and Max 23 frame per / sec with roi & binning); binning modes: 2x2, 4x4, digital zoom: up to 16x.
2. Epi-fluorescence System: Fluorescence filter turret with at least 6 positions, >100 Watt Mercury Burner and
set of narrow band pass fluorescence filters for UV, Blue & Green Excitation (for FITC/GFP/ TRITC/Rhoda
mine/PI/Alexa Fluor/ DAPI/Hoechst).
3. AMC for next 03 years
46
Freeze Dryer
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
Model: Laboratory scale freeze dryer with vacuum pump and sealing device
Controller: Microprocessor controlled
Ice condensing capacity: Capable of removing at least 2.2 kg ice in 24 h.
Total Ice Capacity: > 3.5Kg
Condenser temperature: <-500C
Refrigeration: >1/3 Hp CFC free hermetically sealed compressor with suitable safety device
Display: Digital display of parameters i.e. ice condenser temp and vacuum
Ice condensation chamber: Made of high quality stainless steel inner condenser coils and drain
valve.
Manifolds: Primary and secondary drying manifold, 8 port column manifold, 16 port drum
manifold, vial cell stoppering shelf, along with suitable required no. connectors, adaptors, valves.
Vacuum Pump: High quality corrosion resistant rotary Vane Vacuum Pump with gas ballast and oil
mist filter and displacement capacity of at least 100 L/min and all necessary tubing’s. Ultimate
Pressure should be at least 1.9x10-3 mbar. Noise level: < 50dB
Sealing: Suitable sealing device for vials and ampoules.
Trolley: A trolley with wheels to keep the lyophilizer unit, vacuum pump and accessories.
Electrical requirement: 220-240V, 50Hz.
Certification: CE and ISO 9001.
Stabilizer: Servo voltage stabilizer compatible with the system.
Freeze drying ampoules: 2-5ml capacity (2000no.), 8 flasks (300ml).
Warranty: At least two years from the date of installation.
Brochure: Technical features should be supported by original company literature.
User list: List of ten recent installations with their contact number and mail IDs.
47
Spectrophotomètre
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
Model: Spectrophotometer Fully programmable, monochromator based system which supports
end point, kinetic and spectral scanning assays with capability for DNA/RNA quantification
Wavelength range: form 200-850 nm with 1nM increments
Photometric range: 0.005-2.0OD and should provide the absorbance precision: 0.003OD
Wavelength Accuracy: + 2nM
Wavelength Repeatability: + 0.2nM
Band Width: ≤ 2.5nm
Features: Path Length correction and spectral scanning
Software: facility for data analysis and sharing of data on PC.
Temperature control: Temperature controlled microplate chamber ambient +5 to >420C.
Shaking: Shaking/ mixing feature for microplate platform.
DNA/RNA/Protein quantification: Suitable plate for >16 samples (for 1-2ul volume).
Plate Format: For reading samples in 96 and 384 well plates
Operation: Standalone operation should be feasible
Data Processor: Suitable data processor complete with licensed anivirus (branded computer with
i5 or higher processor, 4GB RAM, 500 GB HDD, 21” TFT Monitor, DVD R/W, USB ports, key board,
optical mouse and Compatible Windows (license).
UPS: Compatible online UPS to provide at least one hour power backup
Printer: Suitable colour Laser printer
Warranty: Two years from the date of installation
User List: Authentic list of recent installations with contact address, and phone numbers.
Catalogue: Authentic original company catalog of the quoted euipment.
48
VIDEO INDIRECT OPHTHALMOSCOPE
1. Bionocular video indirect ophthalmoscope compatible with 4 mega pixel CCD Camera and PC
having image capture software and digital colour printer along with standard accessories.
2. It should have precision viewing upto 1.0mm approximately.
3. Should have focus adjustable for variable working distance
4. The accessories include spare bulbs, carrying case, rechargeable battery, transformer compatible
with voltage system, Table for PC and Voltage stabilizer
5. It should have 2 years warranty
49
Scanning colorimeter:
1. Application: For scanning color of meat product and measure Lightness (L), Redness (a),
Yellowness (b) values
2. Measurement principle: Dual beam spectrophotometer
3. Geometry: 45°illumination/0°viewing or diffuse/8°
4. Spectral range: 400-700nm
5. Light source: Xenon lamp
6. Operation: Easy to operate with thumb-tip navigation functions
7. Display: LCD
8. Parameters for display: Color data, color difference data, color plot, spectral data, spectral data
difference, spectral plot
9. Lamp Life: > 1 million flashes
10. Standards Conformance: CIE 15:2004, ISO 7724/1, ASTM E1164
11. Colorimetric Repeatability: ΔE*≤0.05 CIE L*a*b* on white tile
12. Color Scales: CIE L*a*b*
13. Data storage: 500-700 sample measurements
14. Standard accessories: including Nickel metal hydride batteries, Battery charger, Calibrated
instrument tile, Dust cover, Carrying case, USB flash drive, Users guide
15. Compatible voltage stabilizer: required
16. Warranty: 2 years
50
MAP Packing Machine with gas mixture
1. Applications: For modified atmosphere packaging of meat and meat products
2. Packaging type: Tray/pouches
3. Gas Flushing: Flush up to 3 gases (Nitrogen, carbon dioxide and oxygen)
4. Display: Digital LED/LCD Display
5. Control: Micro-processor based control
6. Sealing: Impulse sealing
7. Lid construction: Fully see through lid
8. Outer construction: Stainless steel contact parts
9. Control for process deviation: Emergency stop button
10. Jaw control: Independent control for individual jaws
11. Provision of wheels: Required for moving the equipment
12. Gas mixture: Gas mixture for mixing Nitrogen, carbon dioxide and oxygen
13. Accessories: Standard accessories
14. Compatible gas cylinders for Nitrogen, carbon dioxide and oxygen: Required
15. Warranty: 2 years from date of installation
51
Mass flow meter with Totalizer
Mass flow meter with totalizer for air, capacity 0 to 1000 litre/minute, temperature range10 -500C and
humidity range 0-100%. It should be fitted in a respiration chamber installed at IVRI Izatnagar in the line
dimension of 3-5 inches.
Warranty: 3 years
52
Sonicator
Sonicator probe type, 600 Watts (minimum) with sound abating chamber, 115 mm probe, Sample
volume 0.5-50 ml, sonicator tips suitable for 0.5-50 ml. Microcontroller based digital timer. Adjustable
pulse with on and off timer for 1.0 second to1.0 minute.
Warranty: 2 years
53
Gel Documentation System With Accessories
1. Gel documentation cabinet with 3 UV intensity (254, 302 and 356 nm) and white light illumination,
with covered camera housing.
2. Imaging by scientific grade high resolution CCD camera, atleast 4.0 mega pixels,16 bits
3. Motorized zoom lens.
4. Exposure: Software controlled, motorized control of aperture, zoom and focus
5. Should have dynamic flat fielding capacity
6. Transilluminator: Illumination area of atleast 20 x 25 cm /Sliding/ drawer/ pull out type with software
controlled power
7. White light converter screen
8. Suitable for EtBr, SYBR Green, GFP, SYBR Gold, Fluorescein, CY3, Rhodamine, SYPRO, Texas Red,
Hoechst, Cumarin
9. Acquisition and Analysis software for gels.
10. Up-gradable to Chemiluminescence
11. Branded Computer System (Processor Intel i3/i5/i7, 4 GB RAM, DVD Writer, 500 GB or higher HDD,
USB ports, keyboard, optical mouse and licensed compatible Windows software) and LCD monitor
from branded company
12. Online UPS (2KVA) from branded company
13. Two Year Warranty from the date of installation
Optional
1. UV torch (302 or 356 nm)
54
Cryostat Microtome Manual cutting


























Temperature range for specimen cutting -50°C and -55° with peltier cooling.
Cryo temperature down to -35°C.
LC display for indication of all data.
Preset and actual temperature and graphical indication.
Electronic control with user friendly touch pad keyboard.
Standby and sleep mode for energy saving.
Automatic evaporator defrosting and also manual defrost cycle activated when needed.
Defrost interrupt as per need.
Sliding window with heater and integrated adjustable fluorescent lamp for cryo-chamber.
Large, removable section waste tray.
Heavy duty rotary microtome with backlash and cross roller bearing stainless steel.
Knife carrier feed with precise stepping motor.
Section thickness setting from 1-100 microns with different increments.
Trimming thickness setting from 5-500 microns.
Specimen retraction of 20µm.
Motorized coarse feed with 3 speed selection.
Automatic approach system of specimen to knife edge.
Hand wheel break with activation at any position.
Disinfection through UV.
Disposable blade.
Specimen chuks.
Tissue freezing medium.
Cryostat oil, tools and instruction manuals.
Working at 220 – 240V, 50 Hz.
Warranty for 2 years from date of installation.
Compatible voltage stabilizer.
55
Headspace Gas analyzer:
Application: For measuring Oxygen and Carbon dioxide within the packaging headspace
Compatibility: Compatible for measurement in flexible packages/trays
Sensors: Zirconia sensor for Oxygen analysis and Infrared sensor for Carbon dioxide analysis
Oxygen measurement: 0.001 to 100%
Carbon dioxide measurement: 0.1 to 100%
Response time: <5 seconds for Oxygen and <10 seconds for Carbon dioxide
Accuracy O2: Above 10% = ±0.5% of reading, Below 10% = ±2% of reading
Accuracy CO2: ± 2% of reading
Calibrations: Automatic air calibration (taking O2 and CO2 as reference air) and automatic gas
calibration (taking pure O2 or CO2 as reference gas)
10. Alarms: Built‐in alarms for high and low gas concentrations.
11. Rugged sample wand: Designed to accept standard luer accessories and includes easily
replaceable needles and particulate filters.
12. Internal Sample Pump: Processor Controlled Diaphragm Type
13. Displays: Digital LED
14. Accessories: Standard accessories like two types of needles, needle adapter with tubing, self
adhesive septa and filter
15. Warranty: 2 years
1.
2.
3.
4.
5.
6.
7.
8.
9.
56
Spectrophotometer
1. Full spectrum 220-750 nm
2. Measure 1 ul sample with high accuracy for nucleic acid, protein, microbial cell culture
OD and other general UV/VIS analysis
3. Computer compatible with software.
4. High absorbance range
5. Absorption precision should be 0.003.
6. Wavelength accuracy should be 1nm
7. 10 second measurement time
8. Large range of detection with detection of nucleic acid down to 2ng/microliter
9. Three years comprehensive warranty.
10. Computer compatible with software
• Intel core i5 processor 3.10 GHz,
• 1 TB SATA Hard Disk Drive,
• 4 GB DDR III RAM with Graphics Card,
• Original Windows 7 processor
• DVD R/W, 6 USB Ports (with at least 2 in front),
• Optical Mouse,
• Multimedia Key Board,
• 21” Bright view LED TFT Color Monitor, Speakers
Original illustrated catalogue for support of technical specifications/features
List of users along with phone & e-mail
57
MICROTOME WITH Cryostat ACCESSORIES

High precision microtome with precision freezing chamber of upto -30 to -40OC temp

The microtome should contain all the essential parts assembled for cryosectioning

Specimen orientation with clearly defined zero position

Disposable blade holder

Frost-free, enclosed microtome assembly

Automatic shut off defrost: at -4o to -5OC chamber temperature

Cryo-embedding System

Diverse range of blades

The instruments should accompany initial complete cryosectioning materials (solution, blades
slides, stainer, embedding media(tissue freezing media) etc for atleast 100 samples

Three years warranty

Free installation

FOB,IVRI, Bareilly
Microtome:
Section thickness adjustment
: 2–60 μm
Total specimen feed
:20-25 mm
Vertical stroke
: 50-60 mm
Maximum specimen size
:55 x 55 mm or 50 x 80 mm
Specimen orientation
:8-10° (x, y, z-axis)
Electric coarse feed
Slow
:500-600 μm/s
Rapid
: 800-900 μm/s
Refrigeration system
: 50 Hz/60 Hz
Cryochamber
Temperature range
: 0 °C to –40 °C
Cooling time down to –30 °C max.
: 6-8 hrs.,
Defrost
:Automatic
Quick-freeze shelf
Maximum cooling
: –35 to - 40 °C
Number of freezing stations
: 10-15
Defrost
:Manual hot gas defrost, time-controlled
(duration 10-12 min.)
Optional to items

Auto disinfection chamber with Ultraviolet

Vacuum-assisted sectioning aid

Motor drive for specimen sectioning

Waste Removal system
58
STEREOZOOM MICROSCOPE BINOCULAR WITH INCIDENTAL












Trinocular
Digital photographic attachment
Interpupillary Distance : 50-75mm
Zoom range
:1 : 6-8
Objective Zoom Range : 0.8X – 5X
Working Distance
:100-120mm
Objectives
:1X
Total Magnification :7X – 60X
Designed with 5-7-position ratchet stop
Illumination
:Incident & Transmitted LED Light with adjustable brightness
Light source
:AC 100v-240V
Warranty
: Three Years on materials and workmanship.
59
ELECTRONIC BALANCE
1. Analytical balances for weighing up to 200-250 g maximum with 0.1 mg-0.5mg resolution ,
2. Weighing capacity Maximum 300gm and minimum 0.1 mg-0.5mg with increment of 0.1 mg
3. Equipped with a touch screen
4. The Automatic Leveling Function
5. Easy to read Large backlite LCD Display
6. Automatic External Calibration.
7. Good keys for precise activation of functions
8. Three years warranty
9. FOB , Bareilly
10. Free installation
11. Suitable voltage stabilizer to be provided essentially
60
Research Upright Microscope:
Model: TRINOCULAR Microscope complete with Opticsand imaging system
Adjustment: Fine and coarse focusing adjustment, anti-mold/anti-fungal treated optics.
ILLUMINATION SYSTEM: LED / 100w Halogen lamp source.
EYEPIECE: A pair of wide field eyepieces 10x (F.O.V: 20mm or better) with dioptre adjustment in
both the eyepiece. A 30° inclination of eyepiece tube
5. NOSEPIECE: Quintuple revolving Nosepiece.
6. STAGE: Rectangular stage with specimen holder, stage should be provided with upper limit stopper.
7. OBJECTIVES: Plan Achromat 4x , 10X, 20X, 40X and 100X (oil immersion)
8. CONDENSER: for Bright Field, Phase Contrast and Dark Field microscopy.
9. Epi-Fluorescence Attachment:At least 4 filter mountable, Hg/Xenon light source,
10. Imaging system: At least 10.0 mega pixel Cooled Colourcamera suitable for bright field and
fluorescent applications.
11. Software: Suitable software for image capturing and analysis.
12. Data collection and processing unit: Branded Desktop computer (Processor Intel i5, 4 GB RAM, DVD
Writer, 500 GB or higher HDD) with monitor (22 inch) and a branded UPS and Colour laser printer
(1200 X 1200 dpi resolution, 16 MB RAM).
13. Online UPS:Compatible online UPS with at least 60 minute backup
14. Warranty: At least 2 years from the date of installation of equipment.
15. Electrical operation: Single phase 220-240 volts/ 50-60 Hz
16. User List: Indian user’s list (minimum 10 installations) for the quoted model along with valid contact
details including telephone no. andE. mail of indenters.
17. Catalogue: Technical features described above should exclusively be supported by authentic original
illustrated company catalog of the instrument/equipment quoted. Photocopy or computer print-out
of catalogue will not be accepted.
1.
2.
3.
4.
61
Refrigerated Centrifuge
Complete high capacity multipurpose centrifuge based on Microprocessor, LCD display of speed, time
and temperature, automatic lidrelease, with rotors,adaptors and servovoltage stabilizer.
Speed: RPM: not less than 20000; RCF: not less than 30000 x g (with automatic LCD display/ conversion
of RPM/RCF values)
Rotors: Different rotors for using 1.5 ml, 15.0 ml and 50 ml centrifuge tubes.
Automatic rotor identification when spinning starts, automatic alarming for imbalance, over speed and
over temperature.
Lowest temperature range: At least -8°C (with facility for fast cool function)
Should have environment friendly refrigerant.
Voltage stabilizer: A compatible good quality 3KVA servo voltage stabilizer to be supplied with
centrifuge.
Warranty: At least two years
62
Small Animal Anaesthesia Machine
Tube flow meter, safety valves, gas pressure gauges, flow rate of O2- 0.1 L to 10 L/min or more, O2 flush
up to 30 L or more, soda lime canister of 1 L or more, latest breathing system/circuit (Bains
apparatus/circuit), reservoir bag-1 to 2L, isoflurane or sevoflurane vaporizers, machine should have
provision for attachment to pin based O2/gas cylinders and positive pressure ventilator, waste gas
scavenging system.
63
Inverted microscope with photographic attachment
1. Inverted microscope: Microscope body with infinity optical corrected optical system side port for
attaching digital camera upgradable to one or more additional port for other attachments. Shock less
construction with revolving or slider type Nosepiece to accommodate different type objectives at a time.
2. STAGE: Plain XY rectangular movable stage with microtiter plate and flask holders
3. CONDENSER: Long working distance turretor slider type at least 5 position condenser with High N.A 0.55
and W.D. 27mm.
4. ILLUMINATION SYSTEM: Halogen Illuminating pillar (≥100 watt) or LED based illumination
5. OBJECTIVES: High performance Long working distance plan apochromatic objectives suitable for Bright
field / Phase contrast/ fluorescence observation,4X, 10X, 20X, & 40X objectives
6. OBSERVATION: Binocular, inclined at 30 deg along with wide field eyepiece 10X F.N 22 mm
7. EYEPIECES 10X with F.O.V 22 or better or with integrated LCD display
8. Warranty of at least two years should be provided.
9. Suitable compatible UPS with at least 30 minute power back up should be provided essentially.
10. Should have provision for up gradation to≥100 watt fluorescence unit at later stage.
11. Should operate on single phase 220 volts and 50Hz power.
12. Indian user’s list for the quoted modelsalong with complete contact details including telephone no. & e.
mail should be provided.
13. Technical features described above should exclusively be supported by authentic original illustrated
company catalog of the instrument/equipment quoted. Photocopy or computer print-out of catalogue
may not be accepted.
14. Vendor’s business record is to be authenticated by producing Income Tax return for at least 3 consecutive
previous years.
Optional accessories (prices to be quoted separately for each item)
1. Digital Color Cooled Camera
• CCD Camera capable of handling bright field, fluorescence, DIC, darkfield images
• High quality 2/3” CCD sensor and at least 12 mega Pixels resolution in12 bit per RGB Color
channel
• Dust protected sensor with infrared cut-off filter. Exposure time up to 180 Seconds. Speed of 21
or more fps.
• Binning modes: 2x2, 4x4, digital zoom: upto 16x
• USB port for attaching camera onto desktop/laptop
• Should be separate modes for different microscopy techniques i.e Bright field, fluorescence, DIC,
Dark field images.
2. Suitable image analysis software
3. Data Processing Unit:
• Branded PC, Intel Core i7 CPU 3.40GHz. 2GB RAM With, DVD Writer, 320GB Or Higher Hard
Disk, 15.4” Wide screen, latest operating system and Multimedia Kit
• Branded color printer with more than 4500 dpi color printing for taking the prints of
photomicrographs
4. Epi fluorescence attachment
• With 6 positions filter block, for simultaneously attaching 6 fluorescence filters at a time.
Should provide high signal to noise ratio images& ≥ 100 watt fluorescence attachment.
• Band pass Fluorescent filters for FITC/GFP (Ex 465-495, DM 505, BA515-555), TRITC/Rhoda
mine/PI/Alexa Fluor (Ex 540/25, DM 565, BA605/55), DAPI/Hoechst (Ex 340-380, DM 400,
BA435-485).
• Objective lens with minimum autofluorescence & having features of uniform brightness in all
fields.
64
Macro Block Digester(12 Tube System)
Automatic auto sequencing twelve place PC compatible regular series Marco block digestion system
with auto lifting facility (Suitable for plants, soils water, food and feed samples).
Specification of Scrubber system:
Fully automatic, four stage. Acid neutralizer scrubber with micro intelligent micro-processor based with
digital display of time.
Warranty: Minimum 02 years from the date of installation
65
Defibrillator
1.
2.
3.
4.
5.
Device should be user friendly, automated compact size for external defibrillator.
Device should be safe to use for both operator & patient
Device shall provide audible prompts with text display for proper use.
Device shall deliver Biphasic Waveform, parameter as a function of patient impedance.
Device should perform Self patient analysis, Evaluates patient ECG, Signal quality, connection
impedance to amount of joule to be delivered for proper defibrillation. Analysis should be
according to the standard guide lines and recommendations.
6. Device should perform automated self teats for internal circuits, waveform delivery system and
pad cartridge and battery capacity on daily bases. It should be provide with indicator for the
same.
7. Device should provide with suitable battery with 3-4 years replacement warranty.
8. Device should provide with self adhesive pad, suitable for all age, size of patients with shelf life
of 2 years or more.
9. Device should provide with all standard accessories and at least on extra pad.
10. Device should have internal memory as well as option for external memory of event and ECG
data.
11. Device should have 5 years onsite warranty from date of installation following by AMC for next 2
years.
66
Micro volume spectrometer
# Path Length:
1 mm (or 10mm with optical cuvette holder)
# Sample volume:
1-3 l
# Built in xenon lamp and CCD array detector
# Wave length range should be 230 to 850 nm
# Minimum detection limit should be 2-5 ng/ l
# Reading time 2-5 seconds
# PC format software for the calculation of nucleic acids and protein.
# Should come with 3 years warranty and accessories like suitable UPS/Voltage stabilizer and data
workstation
67
Microscope
Fully functional Microscope for bright field, phase contrast, dark field and fluorescence
microscopy along with camera and image analysis software
[I] Microscope Body
Microscope body with universal infinity corrected optical system with port for attaching
camera. Revolving type sextuple nosepiece to accommodate atleast six different
objectives at a time and also equipped with slot for DIC slider/polarizer/analyzer
[II] STAGE
Mechanical XY rectangular coaxial stage with specimen holders and right handed low
drive control and torque adjustment mechanism
[III] CONDENSER
Universal turret or slider type condenser with NA 1.1 or better for different applications
like bright field, dark field, phase contrast and differential interference contrast
microscopy
[IV] ILLUMINATION SYSTEM
Halogen Illuminating pillar with pre-centered 12V, 100Watt halogen bulb and energy
saving motion sensors for switching off lamps during unattended operation.
[V] OBJECTIVES
High performance long working distance objectives [4X, 10X, 20X (Plan Achromat
Objectives), 40X (Semi-Apochromat Objective) and 100X (Plan Super Apochromat
Objective)] suitable for bright field, dark field, phase contrast, DIC and fluorescence
microcopy.
[VI] OBSERVATION / EYEPIECE
Wide field trinocular tube with 10X, anti-fungus type eyepiece along with dioptre
adjustment facility
[VII] Fluorescence attachment
1. With atleast eight cube filter turret for simultaneously attaching different
fluorescence filters at a time
2. 100 watt fluorescence attachment with mercury illumination.
3. Narrow band pass filters for UV, Blue, Green, fluorescence. Dedicated narrow
band pass filters for GFP and RFP to be provided essentially.
[VIII] DIGITAL COOLED COLOR CAMERA
1. Digital Cooled Color CCD Camera capable of handling brightfield, fluorescence,
DIC, and darkfield images
2. High resolution, atleast 17 mega Pixels resolution with pixel size of about 4.4um x
4.4um
3. Exposure Modes: manual and auto with exposure time from 1msec to 500
Seconds or more
4. High Speed (atleast 15 frames per second) for fast live images at full frame
resolution of 1600 x 1200.
5. USB port or fire wire cable for attaching camera onto desktop/laptop
6. All features of the camera should be software controlled for capturing images in
different microscopy techniques i.e Brightfield, fluorescence, DIC, darkfield
images
[IX] IMAGE ANALYSIS SOFTWARE
1. Original software for capturing live image and analysis of captured image must be
provided
2. Software should enable annotation, shading correction and image overlay
functions. There must be provision to merge the images for generating multi
fluorescence pictures
[X] DATA PROCESSING UNIT:
System should be supplied with latest configuration PC available in the market at
the time of supply with original windows based operating system to install the
software packages. Atleast 24” flat LED/LCD monitor along with accessories and
anti-virus software to be provided
OPTIONAL ACCESSORIES: (To be quoted separately with prices mentioned for each
item)
1. 2KVA online UPS with at least one hour power back
2. Separate narrow bandpass filter exclusively for viewing YFP and CFP to be quoted
as optional with prices mentioned separately
3. One spare set of 100W fluorescence lamp
4. Any other accessories that may enhance the performance of microscope
• Microscope, camera and software should be preferably from same manufacturer
for better compatibility
• Warranty of at least three years should be provided
• Should operate on single phase 220 volts and 50Hz power
68
Homogenizer high speed
Homogenizer high Speed
Speed: >25000rpm
Regulator: Electronic regulator
Display: Digital RPM display
Safety: Overload protection
Dispersing elements/Probes: Changeable Stainless steel probe/Dispersing elements for the volume
from 1-100ml
Disposable Dispersing elements/Probes: for 1-100ml (25 No.) to be provided along with the system
Holding stand: Compatible stand
Holding Clamp: Compatible clamp
Operation: Electrical, 220-240V/50-60Hz
Warranty: One year from the date of installation
69
Tissue Homogenizer
Volume: 1-1500ml
Speed: 3000-25000 rpm
Speed display: Digital LED
Speed control: Electronic
Dispersing element: Stainless steel(standard)
Extension arm length: 150-170mm
Accessories include: 1 ss dispersing element, 1 plate stand, 1 boss head clamp
Warranty: 3 years from date of installation.
70
Analytical Balance
1. Buit-in calibration made with auto calibration on tourch of a button
2. Capicity 150-210g with 90-120mm pan size.
3. Tare range: Full to capacity.
4. Readability: 0.001mg.
5. Avereage response time: 5 sec or better.
6. Wind draft shield; side & top opening shielding windows for easy pan access.
7. Backlit digital display with sealed key pad.
8. Operating voltage: 220-240V.
9. Ambient condition: 5-40ºC; upto 85% humidity.
10. Dust protection cover.
11. Suitable voltage stabilizer.
Original illustrated catalogue to support technical specification & User’s list
Two years on sitewarranty
Original: USB2/RS-232 port computer interface with easy data transfer to window ®XP/7OS or
better based PC.
71
UV-Vis spectrophotometer
1. Optical system:Double beam
2. Light source:Deuterium and Tungsten lamps
3. Photometric accuracy:±0.002 A at 1 A
4. Photometric stability:±0.0005 A at 500 nm
5. Photometric range:-3 to +3A
6. Photometric repeatability: ±0.001A
7. Spectral bandwidth: ≤1nm
8. Wavelength accuracy (nm):±0.15
9. Wavelength range (nm): ≤190 to ≥1100
10. Wavelength repeatability(nm): ±0.05
11. Scanning speed: ≥2400nm/min
12. Cell holder: ≥ 6 No.
13. Readout format: Quartz cuvette
14. Storage capacity: ≥200 results and ≥200 standard curves
15. Control:Capability to function either as stand alone or with PC control
16. Computer:Computer system complete with UPS (Branded computer with i3/i5/i7 processor, 4GB
RAM, 500 GB HDD, 19” TFT Monitor, DVD R/W, USB ports, Keyboard, optical mouse and
Compatible Windows -license)
17. Software:Suitable Data Analysis Software
18. Printer:Laser printer
19. Interfaces: USB and parallel printer interface
20. Accessories: including standard lamp and quartz cuvettes
21. Spare set of cuvettes: 6 No.
22. Compatible stabilizer: Required
23. Warranty: 3 years
72
Portable steam sterilizer
Portable steam sterilizer
Capacity > 50 litres
Pressure range: adjustable 15-20psi
Controller: Microprocessor
Purging: Auto
Exhaust: Auto
Display: LCD/LED
Application/Use: Programmable
Lid: Single lever/Radial locking
Safety: Overpressure valve, Low water detection
Buzzer: End of programme buzzer
Material internal: Steel
Material External: Steel /Powder coated steel
Carrier basket: Steel
Model: Vertical Floor
Certification: CE
Electrical Operation: 220-240V/ 50-60 hz
Warranty: One year from the date of installation
73
X-RAY MACHINE
High frequency X-Ray machine suitable for general Radiography & compatibility for Digital radiography.
1. The unit should be approved by AERB for radiation safety norms
2. High frequency X-Ray generator with Power output of generator should be 32 KW or more.
3. User friendly, Compact, Soft Touch Control panel for all operation of machine with all safety
norms.
4. One Dual Focus rotating anode X-ray tube thermally protected having focal spot:1.2 mm or less small Focus
2 mm or less Large Focus
5. Machine should be providing with motorized table, Bucky with tray and grid.
6. The company should be ISO certified. The company should provide layout plan and QE test
report for registration from AERB.
7. The company should have proven track record in Govt sector.Company should not be blacklisted
from any Government organization in India.
8. List of recent installation in Govt. Hospitals. list of govt installation along with purchase orders
may also be enclosed.
9. The machine should be supplied and installed at site with working compatibility on three phase
power supply.
10. Company should have well established after sales service network with call centre having 24X7
call login facility.Company should notify the complaint redressal policy.
11. Company should confirm the availability of spare parts for 10 years from date of supply of the
equipment.
12. In case of dealer /distributer, proper authorization from original manufacturer should be
enclosed.
13. The Machine should have 5year onsite warranty from date of Installation. The rate for CMC for 5
years, includingXray tube should be quoted separately.
14. Demo of machine and at least two training.
15. Minor electrical and civil work if required for installation of machine will be done by vendor.
74
CENTRIFUGE REFRIGERATED
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
Instrument: Fully functional refrigerated centrifuge with rotors
Controller: Microprocessor controller
Display: Broad LCD displaywith indication of Preset and Actual Values
Maximum speed: >18000 rpm
Maximum g-Force/RCF: >23000 g
Speed control range: 300rpm to max speed
Speed selection: Selection of speed in both rpm and g-force with increment of 10
Acceleration/De-acceleration selection: Minimum 9 acceleration and 9 deceleration rates
Option for un-braked de-acceleration: Yes
Pre-cooling program: Pre-cooling of rotor and chamber function
Refrigeration: CFC free refrigeration system
Pulse/short run/ continuous run option: Yes
Timer settings: 10 sec to 99 hours
Temperature Set range: -20oC to +40oC with 1oC increment
15. Program memory: For >30 runs.
16.
17.
18.
19.
20.
21.
22.
23.
i.
ii.
iii.
Maximum Volume: > 4x250ml
Motor: Maintenance free induction motor
Lid Locking: Motor driven lid lock
Rotor Identification: Automatic rotor identification
Noise level:<66 decibels
Safety features: Visual/audible alarm of malfunctions like lid open, rotor imbalance, over- speed, overtemperature etc.
Data communication:RS232interface
Rotors: To be provided with the machine (quote separately)
Fixed angle rotor for 24/30 X 1.5/2.0 ml tubes: max speed ≥ 18000 rpm and ≥ 23000g with Adaptor
for 0.2ml tubes (24/30no.)
Fixed angle rotor for at least 6/8x50 ml tubes with speed ≥ 15000 rpm and ≥ 18500g with adaptors for
15ml tubes (6/8no.) compatible for fixed angle rotor of 50 ml
Swing out Rotor for at least 8 x 50 ml and 12 x 15ml falcon tubes with speed >4500rpm
24. Stabilizer: Compatible servo controlled voltage stabilizer
Warranty: Minimum 02 years from date of installation
Electrical Operation: 220-240V/50-60hz
User List: Indian user list (more than 10) with contact no. and telephone no should be provided.
Authentic Original Catalogue:Technical features described above should be supported by the authentic
original company catalogue of the quoted instrument.
29. AMC: charges for additional three years must be given separately (under head optional).
25.
26.
27.
28.
75
Electronic Balance
Single Pan Balance
Weighing range= 1 mg to 200 gm
Repeatability= 0.001 gm, Less than 20 seconds
LCD display
Sensitivity= ± 2 ppm/ 0C
Pan size 100 mm maximum
Power source= 220/230 V AC
Warranty: Minimum 02 years from the date of installation
76
PRODUCT FORMING MACHINE
Automatic machine for forming products of patties form the mix of variable amount of
milk/coagulum and other food ingredients.
Forming of patties
Capacity to form about 800-1500 forms/hour
Approx. From specifications: Diameter 50-80mm.
Provision for the adjustable thickness and weight of the products
Speed regulator
All food contact area of high quality SS
Complete unit easy to operate and clean
Complete fittings and installation
Warranty: Two years guarantee/warranty
Photograph/diagramme: photograph of equipment along with all the detail
specifications should
be supplied user list: List of users with address, valid phone no. And email ids.
77
Autoclave Vertical
Autoclave Vertical
Capacity: > 75 litres
Pressure range: adjustable 15-20psi
Controller: Microprocessor
Purging: Auto
Exhaust: Auto
Display: LCD/LED
Application/Use: Programmable
Lid: Single lever/ Radial locking
Safety: Overpressure valve, Low water detection
Buzzer: End of programme buzzer
Material internal: Steel
Material External: Steel /Powder coated steel
Carrier basket: Steel
Model: Vertical Floor
Certification: CE
Electrical Operation: 220-240V/ 50-60 hz
Warranty: One year from the date of installation
78
Water Bath
Size Interior(cm); 60 Length X 30 Width X 20 Depth;(aprox)
Water bath of circulation water type;
Digital display of temperature
Body: Stainless steel, double walled;
Temperature range: 15 to 1000C
Microprocessor based digital temperature indicator.
Digital display of Set Value (SV) and process Value (PV)
Equipment must cover warranty of minimum 2 years from the date of installation
79
Water bath:
Capacity: 10 Litres
Temperature range: 5 to 1000C
Temperature control: Primary PID, Secondary: Hydraulic thermostat
Temperature sensor: 100 Pt
Display: LED 0C
Wetted material: Stainless steel
Lower level cut off: Adjustable
Working depth: 12-16cm
Temperature accuracy: ±2°C
Shaker facility: Required
Equipment must cover warranty of minimum 2 years from the date of installation
80
CO2 incubator
-
Microprocessor controlled CO2 incubator
Capacity= 150-200 Ltr
Temperature range= from at least 100C below ambient with inbuilt cooling to 50°C above ambient
with accuracy of ± 0.1°C.
It should have fan less all sided direct heating system.
Should have excellent uniformity and rapid recovery with no over shoot event.
It should have CO2 control range from 0-20% with accuracy and uniformity of ±0.1% and equipped
with rapid recovery system to maintain the CO2 concentration.
Should have minimum 4 adjustable shelves & water reservoir made up of stainless steel to achieve
at least 90-95 % RH.
Should be provided with front display of all the set and real time parameters.
It should have two level alarm systems including audio & screen displayed alarms for system status,
with programmable alarms for CO2 and temp set points, delays, duration etc.
Should be equipped with HEPA filter on CO2 inlet.
It should also have over temperature cut-out and automatic alarms set point reset.
It should have internal glass door
A voltage stabilizer (3KV) should be provided with CO2 incubator
CO2 cylinder and regulator= 2 Nos. each
Equipment must cover warranty of minimum 3 years from the date of installation
It should have ISO 9001 & CE Certification.
Operating voltage= 230 -240 Volt and 50 Hz.
Original illustrated catalogue for above technical specifications must be enclosed with list of users.
Optional =Split AC (1.5 ton) for better functioning of the equipment.
Optional= AMC/CMC may be covered at least for 2 years post warranty period
Optional= Door unit for individual chamber/ compartment may be quoted
81
Ultra Low temperature BOD incubator
1. Capacity:> 250 litres
2. SetTemperatureRange: 5°C to 60°C
3. Temperature Accuracy:± 0.5°C
4. Shelves:>3 No.
5. Controller: Microprocessor
6. Display: LCD
7. Safety: Over-temperature/overshoot controller
8. Fan: For forced air circulation
9. Illumination: door operated in-built bulb/tube
10. Material of construction Internal: Stainless steel
11. Material of construction External: Steel or Powder coated Steel
12. Doors: Double walled outer door with lock/ key and full view inner door
13. Movement:Castor Wheel
14. Model: Floor
15. Compressor: Hermetically sealed air cooling compressor system
16. Stabilizer: Suitable Servo controlled voltage stabilizer
17. Electric Operation: 220-240 V/50-60 Hz
18. Warranty: 2 years from the date of installation
82
Bench top Microscope Fluorescence base
Eyepiece: Widefield 10x adjustable
Objectives: (infinity corrected)
1. 4x 2. 10x 3. 40x 4. 100x phase contrast
Excitation Filters:
Blue Excitation Standard: 400nm-490nm
Green Excitation Standard: 510nm-550nm
Light Source: 50W halogen/mercury
Observation Head: Trinocular
Coaxial coarse with fine focusing controls
Camera: Vertical phototube fitted with compatible CCD camera(≥5M pixels)
Mechanical Body & software:
Co-axial focusing system, fine focusing scalez
Imaging software with JPG, TIFF format
Operating Power: 220-240 V AC with Frequency: 50 Hz
OPTIONAL ACCESSORIES (quoted separately):
1. Excitation filter standard: RED:600nm-650nm,Provision for installation of red filter be in
microscope.
2. Display 20 inch LED/LCD, resolution higher than 1366X768 Pixels
3. Desk Top computer with new specification &> 4 GB RAM
Equipment must cover warranty of minimum 2 years from the date of installation
83
LABORATORY ANIMAL METABOLIC CAGES FOR RATS
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
Metabolic cage for rats of different size and weights(150-300g)
Floor space around 450 cm2 or more and feeder opening >5 cm2.
Facilities for perfect (quantitative) separation of the faeces and urine through specially designed
PMP funnel and separation cone designed for automatic separation of faeces and urine for
accurate metabolic monitoring without urine wash-over.
Should be designed to reduce stress to the animal using drawer style design of feeder for easy
filling.
All parts below the cage floor should be removable without disturbing the animal.
Polycarbonate feeder specially designed to discourage animal from sleeping in the dish.
Stainless still water bottle for accurate measurement of fluid intake with support for waste
collection in graduated spillage collection tube.
Faeces and urine collection tubes should unlock with single twist from outside of the cage
without disturbing animal.
All cage components can be easily washed and autoclaved.
Stainless steel floor support with grid openings specific to animal comfort and size.
Cage components should be made up of PC and PMP.
Support Rack: Support rack for at least 12 cages, (dimension: vertical, 126 L x 50 W x 187 H cm or
higher) for holding 12 cages allowing minimizing movements of individual cages.
Two years warranty to cover the cages and rack.
Rates for AMC for next five years may also be provided.
ISO- Certification for design and production of laboratory animal equipments for biomedical
research
84
Vacuum Heating Oven
Microprocessor controlled Vacuum Oven with following features:
1. Bench top model convenient to place on work bench of 30 inch width
2. Fully programmable, jacket heating model to achieve uniform heat distribution
3. Temperature: ≥ 200℃
4. Chamber Volume: ≥ 50 Litre
5. Front opening door with Viton/fluorine rubber door gasket
6. Provision for heat tempered glass window in the front door for viewing inside chamber
7. Stainless steel shelf: at least 4
8. Provision for connecting inert gas like nitrogen
9. Digital display of sample temperature
10. Digital display of pressure and digital pressure controller with solenoid valve
11. Must have separate vacuum vent valve and release valve, stainless steel vacuum connections
and tubings
12. Digital timer to set desired time of run.
13. Maximum temperature deviation at 200oC : ≤4oC
14. Safety Features: Over temperature protection with audible and visual alarms, open door alert
etc.
15. Must have RS-232 interface for data logging applications.
16. Vacuum Pump with following features to be supplied
a) Oil free, chemically resistant vacuum pump with PTFE diaphragm
b) Pumping capacity of atleast 9 m3/h and achieving pressure of 2mbar
c) All tubings, vacuum port and clamps to connect it to Vacuum oven and anti-vibration
metal plate to accommodate vacuum pump on bench must be provided.
17. Electrical power requirement : 230V, 50Hz
18. Warranty :Atleast two years from the date of installation
19. Original catalogue/brochure showing details of quoted model must be provided
20. Any other accessory that may enhance the performance of this system may be quoted as
optional
21. Spare door gasket may be quoted as optional
85.
Refrigerated Centrifuge
1. Bench top refrigerated centrifuge
2. Maximum Speed: RPM not less than 12000 rpm and/or RCF not less than 15000 g.
3. Automatic LCD display
4. Rotor: 24 X 1.5/2.0 ml
5. Temp. range from 0⁰C to 40⁰C
6. Automatic Rotor recognition to set maximum speed for safety.
7. Fast pre-cooling and maximum temperature accuracy inside the rotor
8. Equipment safety : lid lock and imbalance recognition
9. Power supply requirement : 120-240 volts / 50-60Hz
10. Should be supplied with suitable servo voltage stabilizer (3 KVA).
11. Warranty 2 years from the date of installation.
12. Optional- other rotors for 15 ml & 50 ml capacity.
Additional:
1. Original illustrated Catalogue in support of technical specifications.
2. User list for the quoted model along with contact details including telephone no. and E.
mail.
86
Automated Plate Washer
Automated washer for washing the microplates used in ELISA, Cell based assays, magnetic beads and polystyrene
bead based assays with following features:
1. The automated washer must be capable of washing both solid bottom as well as filter bottom plates used
in ELISA, Cell Based assays, Polystyrene beads and Magnetic bead assays
2. Machine must be capable of standalone as well as PC controlled functioning with broad display, keypad
and on-board software for programming the washing parameters for ELISA, cell wash, biomagnetic
separation and vacuum filtration
3. Automated washing of magnetic beads in 96 well plate format (8 x 12 manifold) based on biomagnetic
separation using a strong magnet that pulls magnetic beads across flat bottom in each well. 96 well Flat
Magnet to be provided with system
4. Capable of vacuum filtration based washing of filter bottom plates. The vacuum filtration manifold to be
provided
5. Must be capable of washing the fragile cell layers in cell based assays with the help of angled dispensing
head for dispensing and aspiration applications. The angled dispensing head for 96 well format to be
provided essentially
6. Must be capable of automated cleaning of dispensing manifold to prevent salt/crystal using ultrasonic
bath method
7. There must be provision of variable intensity shaking of microplate platform with time programming
8. There must be provision for running pre-defined programmes for different washing applications and also
provision for developing user defined protocols
9. There must be provision for controlling the machine through external PC. The software for programming
washing parameters must be provided essentially.
10. All-in-one PC with latest configuration available in the market must be provided essentially along with the
system to install the control software
11. Must have safety features like waste and vacuum sensing, fluid flow detection, aerosol cover for the
microplates and overflow detection
12. There must be provision for selecting different vacuum level for vacuum filtration
13. Residual volume per well
: ≤2ul per well in case of solid bottom plates
14. Volume dispense range : ≤25 to ≥2000 µL/well (selectable in 1 µL increments)
15. Buffers
: Selection of 4 different wash buffers
16. Flow Rates
: Different flow rates including slow rate for cell washing
17. Wash Cycles
: From 1 cycle to ≥200 cycles
18. Waste System
: Must be supplied with 10L waste bottle with filters &
level sensing
19. Dispense Bottles
: Atleast 4litres dispense bottles (4 Nos)
20. Voltage
: 240V / 50Hz
21. Must be supplied with compatible vacuum pump and all accessories required for washing 96 well plates
by vacuum filtration
22. Must be provided with troughs for priming and rinsing operations
23. All tubings and filters for the dispense bottle, waste bottle and dispensing manifold has to be provided
essentially and one spare set of all filters to be provided essentially along with the system
24. Original brochure depicting the details of quoted model and accessories to be provided
25. Warranty
: Atleast three years from installation
Optional Accessories:
Any other accessory that may enhance the performance of the system should be quoted as optional with
price mentioned separately against each items
87
Spectrophotometer with Peltier controlled cell holder
Spectrophotometer with peltier controlled cell holder with following features:
1. The equipment must be capable of performing measurements as standalone system using local software
and also must be controlled through PC and external software for advanced analysis
2. The equipment must be supplied with factory fitted two single cell holders, one at the sample and other
at reference position to accommodate standard cuvettes
3. Additionally, the equipment must be essentially supplied with programmable Peltier Cell Holder with total
control of heating/cooling cycles (in the range of 5oC to 100oC with ramping rate of 0.1oC to 6oC/min) for
performing temperature dependent kinetic studies and thermal denaturation / renaturation assays
4. Must be supplied with software for controlling the equipment and performing routine nucleic acid and
protein concentrations and also for advanced analysis like melting experiments (Tm), enzyme kinetics and
difference UV. There must be provision for report generation in MS Word and Excel Worksheets and
presenting data as spectrum plots etc.
5. Equipment must be supplied with advanced kinetic analysis software for analyzing different enzyme
kinetic reactions and models like Michaelis-Menten, Hill plots and Lineweaver-Burke Equations etc.
6. Light Source
:
Xenon or Deuterium Halogen
7. Wavelength Range
:
≤190 to ≥1100 nm
8. Detector
:
Silicon photodiode or Photomultiplier tube
9. Variable Bandwidths
:
0.5, 1.0, 1.5, 2.0, 4.0 nm
10. Wavelength Accuracy
:
± 0.3 nm or better (in the range 190 to 900 nm)
11. Photometric Range
:
> 4 A or better
12. Scan Speed
:
From ≤1nm to ≥3500nm/min
13. Wavelength Reproducibility
:
SDEV of 10 measurements should be < 0.05
14. Equipment must be capable of performing single or multi-cell double-beam kinetic experiments. There
must be provision for accommodating multiple cuvettes in sample and reference compartments for
parallel kinetic experiments. The equipment must have holographic grating with atleast 1200lines/nm and
photometric stability of less than 5x10-4 absorbance units per hour.
15. Equipment must have provisions/ports for connecting accessories like peltier power supply unit, PC
through RS232, keyboard, mouse and external monitor etc.
16. The equipment must be supplied with 5KVA online UPS with atleast one hour power backup for
performing time dependent kinetic experiments, uninterruptedly in case of power faliure
17. Equipment must be supplied with all-in-one-PC with latest configuration available in the market at the
time of supply with wide LCD/LED screen and original windows operating software and genuine antivirus
protection. One branded laser printer must be supplied essentially for printing reports
18. Standard quartz cuvettes (two pair) and micro-volume quartz cuvettes (two Pairs) which can fit into
standard cuvette holder and peltier cell holder must be supplied essentially with the equipment
19. Warranty : Atleast three years from the date of installation on the machine and atleast five year on the
light source
20. Power Requirements: 240V and 50Hz
21. The bid must be supported with original product brochure of the quoted model
Optional Accessories (Prices to be quoted separately for each item)
1. Multiple cell holder for the reference and sample compartments
2. Micro cell holder for recording measurements of low volume samples
3. Fiber optic probe for performing absorption measurement of samples in any tube or container outside
the equipment chamber
4. Accessory for performing reflectance experiments
5. Any other accessory that may enhance the performance of the system may be quoted as optional