Keysight Technologies M9309A PXIe Vector Network

Keysight Technologies M9309A PXIe Vector Network
Keysight Technologies
M9309A PXIe Vector Network Analyzer
Synthesizer
187.5 MHz to 6 GHz
Data Sheet
02 | Keysight | M9309A PXIe Vector Network Analyzer Synthesizer - Data Sheet
Overview
The Keysight Technologies, Inc. M9309A PXIe vector network
analyzer synthesizer is a component for the M9485A PXIe vector
network analyzer. The M9485A provides a true multiport VNA
with up to 12-ports in one chassis.
For more information, see http://www.keysight.com/find/pxivna
Figure 1. M9309A PXIe VNA synthesizer
Technical Specifications and Characteristics
Block diagram
M9309A PXIe Synthesizer
External Trigger Trig 2 (front)
to PXI Backplane
6 GHz
Banded
Switched
Filter
Bank
Divide
by 1-2-4-8
PLL
850 MHz
Divide
by 1 or 2
100 MHz input signal
from M9300A PXIe
frequency reference
100 MHz In
100 MHz
Out
(+13 dBm Typical)
3 dB
(+13 dBm
Typical)
Figure 2. M9309A PXIe vector network analyzer synthesizer block diagram.
3 dB
3 dB
RF/LO Out
(RF Output: 187.5 MHz to 6 GHz)
(-10 dBm to +10 dBm)
RF output signal
03 | Keysight | M9309A PXIe Vector Network Analyzer Synthesizer - Data Sheet
Technical Specifications and Characteristics
Definitions for specifications
Temperatures referred to in this document are defined as follows:
–– Full temperature range = Individual module temperature
of ≤ 73 °C as reported by the module, and environment
temperature of 0 to 55 °C.
–– Controlled temperature range = Individual module
temperature of ≤ 50 °C as reported by the module, and
environment temperature of 20 to 30 °C.
Specifications describe the warranted performance of calibrated
instruments. Data represented in this document are specifications
under the following conditions unless otherwise noted.
–– Calibrated instruments have been stored for a minimum of
2 hours within the full temperature range
–– 45 minute warm-up time
–– Calibration cycle maintained
–– When used with Keysight M9300A frequency reference and
Keysight interconnect cables
Characteristics describe product performance that is useful in
the application of the product, but that is not covered by the
product warranty. Characteristics are often referred to as Typical
or Nominal values and are italicized.
–– Typical describes characteristic performance, which 80% of
instruments will meet when operated within the controlled
temperature range.
–– Nominal describes representative performance that is useful
in the application of the product when operated within the
controlled temperature range.
Recommended best practices in use
–– Use slot blockers and EMC filler panels in empty module slots
to ensure proper operating temperatures. Keysight chassis
and slot blockers optimize module temperature performance
and reliability of test.
–– Set chassis fan to high at environmental temperatures
above 45 °C.
Additional information
–– All graphs contain measured data from one unit and are
representative of product performance within the controlled
temperature range unless otherwise noted.
–– The specifications contained in this document are subject to
change.
–– The M9309A is only intended for use as part of the M9485A
PXIe vector network analyzer. It is not intended to be used as
a standalone instrument.
04 | Keysight | M9309A PXIe Vector Network Analyzer Synthesizer - Data Sheet
Technical Specifications and Characteristics
Frequency
Frequency
Range
187.5 MHz to 6 GHz
Resolution
Switching speed
0.01 Hz
1
< 100 µs, nominal
Amplitude
Amplitude
Settable range
-10 dBm to +10 dBm
Resolution
0.2 dBm, typical
Maximum output power
≤ 1 GHz
+3 dBm
> 1 GHz
+5 dBm
Power level accuracy
25 °C ± 5 °C
±1.5 dB
0 °C to 55 °C
≤ 3 GHz
±2 dB
> 3 GHz
±3 dB
Output impedence
50 Ω, nominal
1. Time for hardware to settle to within 1 ppm or 1 kHz, whichever is greater and amplitude to settle to within 0.2 dB.
05 | Keysight | M9309A PXIe Vector Network Analyzer Synthesizer - Data Sheet
Technical Specifications and Characteristics
Spectral purity
Phase noise
20 kHz offset
1 GHz
-119 dBc/Hz, typical
2 GHz
-114 dBc/Hz, typical
3 GHz
-109 dBc/Hz, typical
6 GHz
-105 dBc/Hz, typical
Harmonics
Range
≤ 0 dBm
300 MHz to 3 GHz
< -27 dBc
Figure 3: Single sideband phase noise from 100 Hz to 10 MHz offsets, at 1 GHz, 2 GHz, 3 GHz and 6 GHz RF frequencies.
06 | Keysight | M9309A PXIe Vector Network Analyzer Synthesizer - Data Sheet
Technical Specifications and Characteristics
Environmental and physical specifications
Operating and storage conditions
Temperature
Operating
0˚C to 55˚C
Non-operating (storage)
-40˚C to 70˚C
Humidity
1
Type-tested at 95%, +40˚C (non-condensing)
Altitude
Up to 15,000 feet (4,572 meters)
Connectors
RF Out, SMA female
Regulatory characteristics
EMC
Complies with the essential requirements of the European EMC Directive as well as current editions of the following standards (dates and editions are
cited in the Declaration of Comformity):
–– IEC/EN 61326-1
–– CISPR Pub 11 Group 1, class A
–– AS/NZS CISPR 11
–– ICES/NMB-001
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme a la norme NMB-001 du Canada.
Other
Warm-up time
45 minutes
Size
1 PXIe slot
Dimensions
Length
210 mm
Width
22 mm
Height
130 mm
Weight
0.535 kg (1.179 lbs)
Power drawn from chassis
≤ 22 W
System requirements
Topic
Operating systems
Windows 7 (32-bit and 64-bit)
Processor speed
1 GHz 32-bit (x86), 1 GHz 64-bit (x64). No support for Itanium 64.
Available memory
4 GB minimum
8 GB or greater recommended
Available disk space 2
1.5 GB available hard disk space, includes:
– 1 GB available for Microsoft .NET framework 3.5 SP1 3
– 100 MB for Keysight IO libraries suite
Video
Support for DirectX 9 graphics with 128 MB graphics memory recommended
(Super VGA graphics is supported)
Browser
Microsoft Internet Explorer 7.0 or greater
Keysight IO libraries
Version 16.3.16603.3 or later
1. Samples of this product have been type tested in accordance with the Keysight Environmental Test Manual and verified to be robust against the environmental stresses of storage, transportation and end-use; those stresses include but are not limited to temperature, humidity, shock, vibration, altitude, and
power line conditions. Test methods are aligned with IEC 60068-2 and levels are similar to MIL-PRF-28800F Class 3.
2. Because of the installation procedure, less memory may be required for operation than is required for installation.
3. NET framework runtime components are installed by default with Windows Windows 7. Therefore, you may not need this amount of available disk space.
07 | Keysight | M9309A PXIe Vector Network Analyzer Synthesizer - Data Sheet
Software
Instrument connection software
Keysight
IO library
The IO library suite offers a single entry point for connection to
the most common instruments including AXIe, PXI, GPIB, USB,
Ethernet/LAN, RS-232, and VXI test instruments from Keysight and
other vendors. It automatically discovers interfaces, chassis, and
instruments. The graphical user interface allows you to search for,
verify, and update IVI instrument and soft front panel drivers for
modular and traditional instruments. The IO suite safely installs in
side-by-side mode with NI I/O software.
Free software download at:
www.keysight.com/find/iosuite
Setup and Calibration Services
Calibration and traceability
Factory calibration
The M9309A PXIe VNA synthesizer ships factory calibrated with an ISO-9002,
NIST-traceable calibration certificate.
Included in base configuration
Calibration cycle
A one year calibration cycle is recommended.
Calibration sites
– At Keysight worldwide service centers
– On-site by Keysight
– By self-maintainers
For more information visit
www.keysight.com/find/infoline
N7800A calibration
and adjustment
software
The M9309A PXIe VNA synthesizer is supported by Keysight’s calibration and
adjustment software. This is the same software used at Keysight service centers to automate calibration. The software offers compliance tests for
ISO 17025:2005, ANSI/NCSL Z540.3-2006, and measurement uncertainty
per ISO Guide to Expression of Measurement Uncertainty.
Licensed software. For more information, visit
www.keysight.com/find/calibrationsoftware
Keysight calibration
status utility
The Keysight calibration status utility helps ensure your M9309A is calibrated
by managing the calibration interval and providing messages regarding
instrument and module calibration status.
Included in base configuration
08 | Keysight | M9309A PXIe Vector Network Analyzer Synthesizer - Data Sheet
Support and Warranty
Warranty
Global warranty
Keysight’s warranty service provides standard coverage
for the country where product is used.
All parts and labor necessary to return to full specified performance
Recalibration for products supplied originally with a calibration certificate
Return shipment
Included
Standard
Return to Keysight warranty – 3 years
15 days typical turnaround repair service
Included
R-51B-001-5Z
Return to Keysight warranty – 5 years
15 days typical turnaround repair service
R-51B-001-3X
Express warranty
3 years
The express warranty upgrades the global warranty to provide, for 3 years,
a 5 day typical turnaround repair service in the US, Japan, China and many
EU countries.
Optional
R-51B-001-5X
Express warranty
5 years
The express warranty upgrades the global warranty to provide, for 5 years,
a 5-day typical turnaround repair service in the US, Japan, China and many
EU countries.
Optional
Core exchange
program
Keysight’s replacement core exchange program allows fast and easy module repairs. A replacement
core assembly is a fully functioning pre-calibrated
module replacement that is updated with the defective module serial number,
allowing the replacement module to retain the original serial number.
For qualified self
maintainers in US
only
Self-test utility
A self-test utility runs a set of internal tests which verifies
the health of the modules and reports their status.
Included in base
configuration
Support
09 | Keysight | M9309A PXIe Vector Network Analyzer Synthesizer - Data Sheet
Configurations and Ordering Information
Ordering information
Warranty
Model
Description
Advantage services: Calibration and warranty
M9309A
PXIe VNA synthesizer: 187.5 MHz to 6 GHz
Includes return to Keysight warranty - 3 years
Keysight Advantage Services is committed to your success throughout
your equipment’s lifetime
Related products
Standard
Return-to-Keysight warranty - 3 years
R-51B-001-5Z
Return to Keysight warranty – 5 years
R-51B-001-3X
Express warranty – 3 years
Model
Description
R-51B-001-5X
Express warranty – 5 years
M9485A
PXIe vector network analyzer
N7800A
Calibration & adjustment software
M9018A
18-slot PXIe chassis
M9037A
PXIe embedded controller
M9389A
PXIe VNA source
M9300A
PXIe frequency reference
Accessories
Model
Description
Y1212A
Slot blocker kit: 5 modules
Y1213A
PXI EMC filler panel kit: 5 slots
Y1214A
Air inlet kit: M9018A 18-slot chassis
Y1215A
Rack mount kit for M9018A 18-slot chassis
10 | Keysight | M9309A PXIe Vector Network Analyzer Synthesizer - Data Sheet
myKeysight
www.keysight.com/find/mykeysight
A personalized view into the information most relevant to you.
www.pxisa.org
PCI eXtensions for Instrumentation (PXI) modular instrumentation delivers a
rugged, PC-based high-performance measurement and automation system.
Three-Year Warranty
www.keysight.com/find/ThreeYearWarranty
Keysight’s commitment to superior product quality and lower total cost
of ownership. The only test and measurement company with three-year
warranty standard on all instruments, worldwide.
Keysight Assurance Plans
www.keysight.com/find/AssurancePlans
Up to five years of protection and no budgetary surprises to ensure your
instruments are operating to specification so you can rely on accurate
measurements.
www.keysight.com/go/quality
Keysight Technologies, Inc.
DEKRA Certified ISO 9001:2008
Quality Management System
Keysight Infoline
www.keysight.com/find/service
Keysight’s insight to best in class information management. Free access to
your Keysight equipment company reports and e-library.
Keysight Channel Partners
www.keysight.com/find/channelpartners
Get the best of both worlds: Keysight’s measurement expertise and product
breadth, combined with channel partner convenience.
www.keysight.com/find/m9309a
For more information on Keysight
Technologies’ products, applications or
services, please contact your local Keysight
office. The complete list is available at:
www.keysight.com/find/contactus
Americas
Canada
Brazil
Mexico
United States
(877) 894 4414
55 11 3351 7010
001 800 254 2440
(800) 829 4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Other AP Countries
1 800 629 485
800 810 0189
800 938 693
1 800 11 2626
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
(65) 6375 8100
Europe & Middle East
Austria
Belgium
Finland
France
Germany
Ireland
Israel
Italy
Luxembourg
Netherlands
Russia
Spain
Sweden
Switzerland
United Kingdom
0800 001122
0800 58580
0800 523252
0805 980333
0800 6270999
1800 832700
1 809 343051
800 599100
+32 800 58580
0800 0233200
8800 5009286
800 000154
0200 882255
0800 805353
Opt. 1 (DE)
Opt. 2 (FR)
Opt. 3 (IT)
0800 0260637
For other unlisted countries:
www.keysight.com/find/contactus
(BP-03-20-15)
This information is subject to change without notice.
© Keysight Technologies, 2015
Published in USA, June 26, 2015
5992-0881EN
www.keysight.com
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertisement