YZ5F Slit Lamp
Instructions for use
1
Preface
Thank your for purchasing our YZ5F slit lamp . Please read this manual carefully for
the sake of your best use.
General Requirements for Safety
Please read carefully about following precautions to avoid unexpected personal
injury as well as the product being damaged and other possible dangers.
Precautions
Do not use this instrument in the environment prone to fire and blast or
where there is much dust and with high temperature. Use it in the room and
simultaneously be careful to keep it clean and dry.
2.
Check that all the wires are correctly and firmly connected before using.
Ensure that the instrument is well grounded.
3.
Please pay attention to all the ratings of the electrical connecting terminal.
4.
Only use fuse according to the specifications and ratings stipulated by our
product.
5.
Use the power cable supplied with this instrument.
6.
Don’t touch the surface of the lens and prism with hand or hard objects.
7.
Turn off the main power first before changing the main bulb and fuse.
8.
To prevent the instrument from falling down to floor , it should be placed on
the floor where the inclination angle is less than 10q.
9.
Turn off the power and cover the instrument with dust cover when it is not
in use .
10. In case there is any trouble, please first refer to the trouble shooting guide .
If it still can’t work , please contact the authorized distributor or our Repair
Department .
1.
THE SAFETY MARKS USED IN THIS INSTRUMENT
TYPE B
ATTENTION PLEASE
REFER TO THIS MANUAL
TERMINAL OF
THE PROTECTIVE GROUNDING
2
Contents
1. Nomenclature ----------------------------------------------------------------------------1
2. Assembly -----------------------------------------------------------------------------------3
2.1 Components ----------------------------------------------------------------------------3
2.2 Assembly Procedure -----------------------------------------------------------------5
2.3 Checking Procedure after Assembling----------------------------------------------9
3. Operation Procedures -------------------------------------------------------------- 10
3.1
3.2
3.3
3.4
3.5
Diopter Compensation and Pupil Distance Adjustment ------------------------- 10
Patient Position and Fixation Target ---------------------------------------------- 11
Base Operation----------------------------------------------------------------------- 11
Illumination Part Operation -------------------------------------------------------- 12
Fundus Observation with Hruby Lens -------------------------------------------- 14
4. Maintenance ---------------------------------------------------------------------------- 15
4.1
4.2
4.3
4.4
4.5
4.6
4.7
4.8
4.9
Replacing the Illumination Bulb--------------------------------------------------- 15
Replacing the Reflecting Mirror--------------------------------------------------- 15
Replacing the Fuse ------------------------------------------------------------------ 16
Replacing the Chin-rest Paper ----------------------------------------------------- 16
Adjusting the Tightness of the Slit Width Knob -------------------------------- 16
Adjusting the Inclination of the Illumination Part------------------------------- 16
Cleaning ------------------------------------------------------------------------------ 17
Protection ----------------------------------------------------------------------------- 17
Consumables ------------------------------------------------------------------------- 17
5. Trouble Shooting Guide ---------------------------------------------------------- 18
6. Responsibility -------------------------------------------------------------------------- 18
7. Transportation and Storage ------------------------------------------------------ 19
8. Optional Accessories ---------------------------------------------------------------- 19
8.1 10x Measuring Eyepiece ----------------------------------------------------------- 19
8.2 Applanation Tonometer ----------------------------------------------------------- 19
9. Specifications --------------------------------------------------------------------------- 20
3
Fig.1
4
1.Nomenclature
1 Joystick
Incline joystick to move the instrument slightly on the horizontal surface and
rotate it to adjust the elevation of the microscope.
2 Base Locking Screw
The base will be locked when fastening this screw.
3 Rail Cover
Protect the rail surface.
4 Base
Support the microscope and the illumination arms with the joystick controlling its
movement.
5 Work Table
6 Accessory Drawer
Store the focusing test rod and other accessories.
7 Brightness Control Switch
Three levels are available H ( High ), N ( Normal ), L ( Low ) . Avoid working
continuously at high setting, as the service life of the bulb will be shortened.
8 Main Power Switch
9 Pilot Lamp
10 Microscope Arm Locking Knob
Lock the rotational movement of the microscope arm.
11 Angle Mark Ring
Marks on the angle mark ring of the illumination arm which relates to the long
mark of the microscope arm represent the two arms’ angle. When the ‘0’ on the
ring relates to the short mark at one side of the operator, the right eyepiece may
be overspread, and the side of the patient the left eyepiece.
12 Location Roller
When it is in the middle, it stands for included angle of 0o between the
microscope arm and the illumination arm. And the right or left side the included
angle of 10o.
13 Microscope and Illumination Arm Couple Bolt
Fasten this bolt and the illumination arm and the microscope arm could be move in
couple state to rotate together. Loosen it and the illumination arm then can rotate
separately.
14 Chin-rest Elevation Adjustment Knob
Rotate the knob to adjust the elevation of the chin-rest
15 Hruby Lens Guide Plate
Also used as an assembly plate for the applanation tonometer.
16 Microscope Fixation Screw
5
17 Magnification Changer Lever
Push the lever to either side to select the desired magnification of the microscope.
18 10x Eyepiece
A 16x eyepiece is supplied as a standard accessory. Pull out the eyepiece for changing.
19 Diopter Adjustment Ring
Adjust the eyepieces diopter to obtain a sharp image before using the instrument.
20 Chin-rest
21 Fixation targets
Two kinds of fixation targets are available. One is for diopter adjustment to help
the patient to see the target clearly while the other is an illuminated fixed spot .
22 Forehead Belt
23 Slit Width Control Knob
The slit width is continuously adjustable within the range from 0 to 9mm. The
marks on the left knob stand for the approximate value of the width.
24 Illumination Inclination Lever
Four 5o inclination stops are available up to 20o.
25 Centering Knob
Loosening the knob allows the illumination light to be moved from the center of the
vision field for indirect retro-illumination. Fastening the knob brings the illumination
light back to the center.
26 Hruby Lens Holder
27 Hruby Lens
Used for observation of the fundus and the posterior segment of the vitreous
body.
28 Reflecting Mirror
Both long and short reflecting mirrors are provided. The long mirror is routinely
used for most examination procedures. The short mirror is used when the long
mirror interferes with the observation pathway, such as during funduscopy.
29 Horizontal Mark
When the horizontal center of the patient’s eye is in line with this mark, the
elevation of the microscope controlled by joystick is also in its center position.
30 Aperture and Slit Height Control Knob
Rotate this knob to adjust the spot and the slit height. Swing the knob
horizontally to revolve the slit.
31 Filter Selection Lever
There are four filters for selection
32 Slit Height and Aperture Display Window
33 Lamp Cap
6
2. Assembly
This section of the manual describes how to assemble YZ5F slit lamp. All parts
should be taken out with great care from the packing case before assembling.
2.1 Components------------------------------------------------------------------
Fig2-1
7
Fig.2-2
A
B
C
D
E
F
G
H
I
J
K
L
M
N
O
P
Q
R
S
T
U
V
W
X
Y
Z
Name
Quantity
Illumination Part
1
Microscope (with 10x magnification eyepieces)
1
Base Part
1
Head-rest Part
1
Breath Shield
1
Work Table with Power Box
1
Rail Cover
2
Input Power Cable
1
Hruby Lens
1*
Hruby Lens Guide Plate
1*
Spare Main Illumination Bulb
2
Chin-rest Paper
1
Focusing Test Rod
1
16x Eyepieces
2*
Dustproof Cover
1
Fixation Target
1
Spare Long Reflecting Mirror
1
Spare Short Reflecting Mirror
1
Protection Cap
1
Spare Fuse
2
Brush
1
Screw Driver with Wood Handle
1
Cross Screw Driver with Wood Handle
1
Watch Screw Driver (big)
1
Watch Screw Driver (small)
1
Spanner
1
(*Optionally available in some region.)
8
2.2 Assembly procedure ------------------------------------------------------Necessary tools are as follows:
Screw driver with wood handle
(V)
Cross screwdriver with wood handle (W)
Watch screwdriver
(X)
Spanner
(Z)
panel facing the operator, refasten the bolt
securely with the spanner (Fig.4).
1) Selecting Voltage and Fuse
Fig.4
Fig.3
— Check the setting on the voltage selector
located on the bottom of the power box
(Fig3). If it doesn’t match with the input
voltage, slide it to the proper position
with screwdriver (X).
— Open the fuse holder with screw driver
(W) and take out the fuse , check and
ensure that its rated value is
corresponding to the mains voltage:
110V----------------------1A
220V--------------------0.5A
It has been set to 220V, 0.5A before
leaving our factory.
3) Assembling the Head-rest Part (D)
—
Attention: Set the input voltage and
frequency of the instrument
according to that of the mains.
2) Assembling the Work Table (F)
— To attach the worktable on the YT2A
motorized instrument table, please
screw off four M8x20mm bolts with
spring washers with the spanner (Z).
— Lift the worktable to aim its screw hole at
the assembly hole of the instrument table.
Put down the worktable, with the power
Fig.5
Remove the four screws attached to
the chin-rest connection board with the
screw driver (W).
Fig.6
9
—
Put two cables in the gap between the
head-rest fixation plate and the
chin-rest connection board (Fig.5).
While ensuring they are not clamped,
retighten the previously removed
screws (Fig.6).
—
4) Assembling the base part (C) and the rail
cover (G)
—
—
—
—
—
Fig.7
Place the wheels of both sides of the base
(C) on the rail on the worktable (Fig.7).
Check whether the wheels can be rolled
steadily on the rail.
Remove four screws attached to the rail
with the screw driver (W).
Place the rail cover (G) to the rail, re-tighten
the previously removed screws.
mark and the limiting board (Fig.8).
Loosen the set screw in the illumination arm
with the screwdriver (X). Aim the assembly
hole of the illumination arm at the brass
shaft sleeve with care then put down, let the
shaft keeping close to the bottom surface
well and simultaneously the two red marks
stretch in one line (Fig.9).
After the two red marks accurately aligned,
re-tighten the set screw (Fig.9).
Fig.9
6) Assembling the microscope (B)
5) Assembling illumination part (A)
—
—
—
Fig.8
Loosen the illumination arm couple bolt
(13).
Rotate the brass shaft sleeve to make an
approximate angle of 45o between the red
—
10
Fig.10
There is a U-shape groove on the top of
the microscope arm and a screw hole
on the bottom of the microscope. Hold
the microscope to fix it to the arm with
the U-shape groove meeting the
location block of the arm well (Fig.10).
—
Fasten the fixation screw (16) with the
big screw driver (V) to fix the
microscope.
Attention: Avoid touching any lens
surface.
7) Assembling the breath shield (E)
9) Connecting plug
— Peel off the sticky tape attached to the
lamp cap, which ensures that the cap is
tightened to the lamp base during
shipping.
— Insert the plug on the top of the head-rest
part (D) into the socket of the lamp cap
(33) on the illumination part (A) (Fig.13).
Fig11
— Remove the breath shield fixation
screw from the microscope arm.
— Pass the removed screw through the
hole of the breath-shield then re-screw
it into the arm (Fig.11).
Fig.13
— Connect the two plugs below the
head-rest part with the corresponding
output socket of the power box. .
— Insert the plug of the input power cable
(H) into the input socket of the power
box. .
— Remove the cable clips from the
bottom of the work table with screw
driver (W) and wrap the output and
input cables respectively, then re-attach
them to the bottom of the work table
(Fig14).
8) Dismantle the illumination part
shipping pad
Fig12
— This pad is used to protect the slit
mechanism of the illumination part
during shipping.
— Remove the rubber band and gently
pull the pad out (Fig.12).
Fig.14
11
11) Assembling the chin-rest paper (L)
— Pull out the two fixing pins from the
chin-rest.
— Get rid of the paper package and let the
pins go through its holes.
— Insert the fixing pins into the holes
again (Fig.17).
10) Assembling the Hruby lens ( I ) and the
Hruby lens guide plate ( J )
— Insert the Hruby lens (I) into the Hruby
lens holder (26) on the head-rest part.
Be careful not to touch the lens surface
(Fig15).
Fig.15
— Place the Hruby lens guide plate (J)
into the main shaft hole of the base part
with the small end pointing to the
head-rest part (Fig16).
12) Placing spare parts
— Some spare parts could be stored in the
accessory drawer (6) (Fig.18).
Fig.16
Fig.18
Fig.17
12
2.3 Checking procedure after assembling ------------------------------------— Turn on the main power switch, and
the pilot lamp (9) will be lighted.
Open the slit width control knob (23)
to examine the illumination.
— Rotate the brightness control switch
(7) respectively at three positions
and the brightness should be
changed accordingly.
— Check the fixation target device to
con-firm it is lighting.
— Check whether all the moveable
parts such as aperture and slit height
control knob (30), filter selection
lever (31), and magnification change
lever (17) etc. could be operated
freely.
— After examining, turn off the main
power and cover the instrument with
the dust-proof cover (O).
1) Power plug
— This instrument supplies a 3-wire
cable. Please select a proper power
socket as matched.
— Ensure that the instrument is
grounded well.
Attention: Please uses the special
cable supplied with this
instrument.
2) The
power
box
and
the
illumination part
— When the main power switch (8) of
the power box is placed at ‘I’, it
turns on, and ‘O’ for turn off. The
main power switch should be set at
the ‘O’ position before connecting
the input cable with the power
socket.
13
3 Operation procedures
3.1 Diopter compensation and Pupil Distance adjustment...................................................
Use of the focusing text rod (M)
—
—
—
Fig.19
The rod is supplied as one of standard
accessories
for
confirming
the
microscope’s accurate adjustment. Insert
it into the main shaft hole with the flat
surface facing the objective lens
the
direction of the operator ( Fig.19) .
—
Fig.20
First, rotate the diopter adjustment
ring (19) counter clockwise down to
the end. (Fig.20).
Second, rotate the ring clockwise
until a sharp slit image appears on
the focusing text rod. .
Adjust another eyepiece in the same
procedure.
Record the diopter value on each
eyepiece for future reference.
4) Pupil distance adjustment
Separate the prism box of the
microscope with both hands to adjust
the P.D. until both eyes could see the
same image on the focusing test rod
through the eyepieces, and at the same
time a stereo vision will be obtained
(Fig.21).
Attention: After adjusting,
remember to take out the rod and
insert the protection cap.
2) Brightness adjustment
Switch on the main power switch and
set the brightness control switch (7) at
‘N’ position. Turn the slit width
control knob (23) to make the slit
width to be 2~3mm.
3) Diopter compensation
The focus of the microscope is
calibrated
according
to
the
emmetropia. If the operator is an
ametropia, he should adjust the
eyepiece diopter.
Suggest adjusting the diopter as
following procedures
Fig.21
Attention: While adjusting P.D.,
ensure that both eyepieces
are at the same height.
14
3.2 Patient position and fixation target----------------------------------------knob to adjust the diopter
compensation within the range from
-15D to +10D (Fig.23).
1) Positioning the patient’s head
Have the patient place his chin on the
chin-rest (20) and the forehead against
the forehead-rest belt (22). Adjust the
chin-rest adjustment knob (14) below
the chin-rest until the patient’s
canthus aligns with the horizontal
mark (29) (Fig.22).
Fig.23
— The fixation target with spot light is
especially for the patient whose
diopter exceeds -15D (Fig.23).
When changing, just loosen the
fixation screw, replace the fixation
target with the spot light source and
re-fasten the fixation screw.
Fig.22
2) Use of the fixation target
— For fixing the patient’s eyesight, just
make him look at the fixation target
(21) with the eye not to be examined.
To change fixing position, move the
lamp bar, as well as move the curved
lever around the forehead.
— The fixation target with diopter
compensation supplies a dot and
concentric circles target. Slide the
Fig.24
3.3 Base operation -----------------------------------------------------------------1) Horizontal rough adjustment
Keep the joystick (1) erect and move
the base (4) to make the microscope
move on the horizontal surface to
aim at the object roughly (Fig.25).
2) Vertical adjustment
Rotate the joystick to adjust the
microscope’s height until it aligns
with the target. Turn the joystick
clockwise to raise the microscope
and counter-clockwise to lower it.
15
3) Horizontal Fine adjustment
Tilt the joystick to make the
microscope move slightly on the
horizontal surface. While watching
through the eyepieces, tilt the joystick
to aim accurately at the object for a
sharp image.
4) Locking the base
When finishing the adjustment, fasten
the base locking screw (2) to lock the
base (4) and prevent it from sliding.
Fig.25
3.4 Illumination parts operation-----------------------------------------------1) Changing the slit width
Turn the slit width control knob (23)
and the slit width will be changed
from 0mm to 9mm. The slit becomes
a circle at the 9mm size. The width
value is indicated approximately by
the scale on the knob (Fig.26).
Fig.27
3) Rotating the slit image
Swing the aperture and slit height
control knob (30) horizontally to
revolve the slit image at any angle in
the vertical or horizontal direction.
The rotation angle of image is
indicated by the rotation angle scale
with small division for 5o and big
division for 10o (Fig.28).
Fig.26
2) Changing the aperture and slit
height
Turn the aperture and slit height
control knob (30) and 6 different
circular beams of light are available at
full aperture: 9, 8, 5, 3, 1, 0.2 dia
respectively. With a slit image, the slit
height can be changed continuously
from 1 to 9mm that is indicated
through the display window (32)
(Fig.27).
Fig.28
16
6) Reflecting mirror
Both short and long reflecting mirror
are available in this instrument. Use
the long mirror in normal examination.
When the angle between the
illumination part and the microscope
is within 3o to 10o, the examined
image might be obstructed. In this
case, use the short mirror. The short
mirror is also used when the
illumination part is inclined over 10o
(Fig.31).
4) Deflecting the illumination light
Loosen the centering knob (25) and
swing the slit width control knob (23)
back and forth so the light spot moves
away from the center of the
microscope vision field . It is mainly
used to examine the eyes by indirect
retro-illumination.
Fasten
the
centering knob and the slit light will
return to the center of the microscope
vision field (Fig.29).
Fig.31
7) Filter selection
Turn the filter selection lever (31) in
the horizontal surface to add four
different kinds of filters respectively
into the illumination pathway. Usually
the heat-absorbing filter is used so
that the patient may feel more
comfortable in long period of
examination (Fig.32).
Fig.29
5) Oblique illumination
Oblique illumination is used for
sectional or fundus examination by
using contact lens. Press down the
inclination lever so that the
illumination part may incline to 20o,
(5oof each division). Since the
illumination part may touch the
patient’s head, operate carefully
(Fig.30).
Fig.32
Fig.30
17
3.5 Fundus observation with Hruby lens (optional)-------------------------In routine using, the observation with
the slit lamp is limited within the
range from the cornea to the anterior
part of the vitreous body owing to
refraction effects of the cornea and
the crystalline lens. However with the
Hruby lens (I) in the front of the
microscope, the posterior part of the
vitreous body and fundus then can be
observed.
7) To examine different parts, either turn
the microscope and the illumination
arm or change the patient’s fixation
by manipulating the fixation target.
8) If the long mirror interference in the
examination, just replace it with the
short mirror.
9) After examination, move the Hruby
lens back to the original position on
one side of the chin-rest.
Operation procedure:
1) The pupil should be dilated for about
20 minutes.
2) Insert the Hruby lens guide plate (15)
into the main shaft hole of the
illumination and the microscope arm.
Attention: before moving the
Hruby lens to the right and
left, first have the patient’s
head move away from the
chin-rest to avoid his nose
touching the Hruby lens.
3) Pull out the Hruby lens holder from
one side of the head-rest as shown in
Fig.33. Move the Hruby lens holder
toward the operator so that it can slide
freely to the left and right below the
chin-rest. Insert the lower end of
Hruby lens lever into the groove on
the guide plate.
4) Move the focus of the illumination
light and the microscope to the
patient’s eye.
5) Move the lever to locate the Hruby
lens at the center of the vision field
and near the patient’s eye as shown in
Fig.33.
6) Move the lever to focus the Hruby
lens at the fundus, then adjust the slit
height and width to reduce the
unnecessary inter-ferrential light in
the vision field.
Fig.33
18
4
Maintenance
Attention: The replaced waste materials should be treated as industrial
rubbish.
4.1 Replacing the illumination bulb--------------------------------------------— Turn the main power switch (8) off.
— Pull out the plug connected to the
lamp house, rotate the lamp cap (33)
counter clockwise and pull it out from
the illumination part (A) (Fig.34).
the illumination may be uneven
(Fig.35).
Attention: The bulb is hot
— Place the lamp cap in the original
position and rotate it clockwise and
insert the connecting plugs.
— Turn on the main power switch and
check whether the new bulb works or
not.
Fig.34
— Take out the old bulb and replace it
with a new one. The groove in the
bulb fixation disc should be aimed at
the flange of the lamp base, otherwise
Fig.35
4.2 Replacing the reflecting mirror --------------------------------------------— Set the angle between the microscope
and the illumination arm to exceed
30o.
— Incline the illumination arm by more
than 10o.
the extended surface (Fig.36).
— Insert new long or short reflecting
mirror.
— When replacing the short mirror, just
push the bottom of the mirror by
using an object with a sharp end
(Fig.37).
Fig.36
— Remove the long mirror by holding
Fig.37
19
4.3 Replacing the fuse-------------------------------------------------------------— Turn off the main power switch (8)
and pull out the input cable from the
power socket.
— Screw off the fuse holder cover with
the screw driver (X) (Fig.38).
— Replace it with a new fuse, then
fasten the cover.
— The fuse specifications are as follows:
110V 1A,
250V
220V 0.5A,
250V
Attention: Please select the fuse of
same type, specification and rating.
Fig.38
4.4 Replacing the chin-rest paper ----------------------------------------------Fig.39
When the paper is exhausted, pull
up-wards two fixing pins of the
chin-rest and place a new package of
paper, then fix the fixing pins again
(Fig.39).
4.5 Adjusting the tightness of the slit width knob ---------------------------If the slit width control knob is too
loose, the slit width may be out of
control. Loosen the screw on the right
knob with the screw driver (Y), then
hold the left knob firmly with one
hand, while the other hand rotate the
right knob clock-wise to adjust its
tightness. When it is appropriate,
fasten the screw of the right knob
firmly again (Fig.40).
Fig.40
4.6 Adjusting the inclination of the illumination part ----------------------If the inclination mechanism of the
illumination part is too loose, fasten
the screws on both sides of the pivot
point with the screw driver (W) (Fig.
41).
20
Fig.41
4.7 Cleaning -------------------------------------------------------------------------with soft cloth dipped with soluble
detergent or water, sterilize with
medicinal alcohol.
1) Cleaning the lenses and mirrors
If any dust stick on the lenses or reflecting
mirrors, brush them with the brush (U) supplied
in the standard accessories. In case any dust still
remains, wipe it off with soft cotton dipped with
absolute alcohol.
Attention: Never scratch with
fingers or any other hard materials.
2) Cleaning the slide plate, rail and shaft
If the slide plate, rail and shaft are
dirty, the vertical and horizontal
movement will be unsteady. Wipe
them with clean soft cloth. (Fig.42)
3) Cleaning and sterilizing the plastic parts
Clean the plastic parts such as
chin-rest bracket, forehead-rest belt
Fig.42
Attention: Don’t wipe with any
corrosive detergent lest that the
surface should be damaged.
4.8 Protecting -----------------------------------------------------------------------There always are dusts and
physiological salt solution dropping
into the main shaft hole of the
illumination arm during the operation.
Please cover the main shaft hole with
the protection cap lest that the
instrument would be damaged. Take
off the cap when the guide plate needs
to be assembled ( Fig.43 .
Fig.43
4.9 Consumables -------------------------------------------------------------------Part name
Illumination bulb
YZ5F
Please specify names and
quantities when ordering
following consumables.
Slit
Long reflecting
mirror
Chin-rest paper
Lamp
Fuse 1A ( 110V )
0.5A ( 220V )
21
outlook
5 Trouble shooting guide
In case there is any trouble, please check according to the following table for
reference. If it still cannot work, please contact the Repair Department of Suzhou
Medical Instrument Factory or an authorized distributor.
Trouble
Possible cause
The cable isn’t connected
correctly with the power socket
The main power switch is on
‘ O ’ position
Remedy
Refer to
Connect the power cable
P7
correctly
Place the switch on ‘I ‘
P9
position
No
illumination The plug on the power box is Insert the plug firmly
loose
The plug on the lamp cap is Insert the plug firmly
loose
The bulb has burnt out
Change the bulb
P7
P7
P15
The fuse has blown
Change the fuse
P16
The bulb is not assembled
properly
The filter lever is in the middle
position or in the position of
grey filter.
Voltage selector is wrongly set
Assemble
the
bulb
properly
Set the filter lever to the
correct position
P15
P5
The fuse doesn’t comply with
the specification
Slit closes The slit width control knob is
automatically too loose
Set the voltage selector
correctly
Change the reflecting
mirror
Clean the surface with the
brush
Set the voltage selector
properly
Replace it with a suitable
fuse
Adjust the tightness of the
control knob
Fixation The output plug is loose
target is off
Insert the
firmly
Slit is
too dark
Fuse has
The coat of the reflecting mirror
is oxidized
Too much dust on the reflecting
surface
Voltage selector is wrongly set
blown
output
plug
P13
P15
P17
P5
P16
P16
P7
6 Responsibility
We will supply circuit diagram of the instrument, electric component list, drawing
annotation and calibration details according to the customer’s need for repair.
If there is any need for enquiry of relative information and relative service or some
questions, please contact with us directly or authorized distributors.
22
7 Transportation and storage
During the transportation, be careful to protect it from wetness, upside down and
violent vibration. The relative humidity should be 10% to 90%, and environment
temperature -25 to 40 .
This instrument should be stored in a well ventilated room without corrosive gas
where the relative humidity should be 10% to 80% and environment temperature -10
to 40 .
If the assembled instrument should be moved or transported in short distance, please
lock all the movable parts. Move this instrument carefully with hands pushing or
carrying its table. If for long distance transportation, please repack it with original
package.
8 Optional accessories (purchase in addition)
8.1 10x measuring eyepiece --------------------------------------------------Replace the common eyepiece with this one to measure the length and angle.
Scale specification
Length scale: 16mm
(0.5mm minimum graduations)
Angle scale: 360o
(5o minimum graduations)
Measuring parameters
Length scale:
To be used at 10x only
Diopter compensation:
-5D to +3D
Angle scale:
No limitation
Fig.44
8.2 Applanation tonometer --------------------------------------------------This YZ5F slit lamp could be equipped with YZ30R, Haag-Streit AG Model R-900 or
Model T-900 applanation tonometer for measuring the intraocular pressure. (Fig.44 and
Fig.45)
YZ30R Model R-900
Model T-900
Fig.45
Fig.46
23
9 Specifications
Microscope----------------------------------------------------------------------------------Type
Cross-angle stereo-vision
Model of magnifying
Change the objective for 2-grade magnification
Eyepiece
10x and 16x(optional)
Total magnification rate Objective *Eyepiece = magnification rate/vision field
1x
10x
10x
‡18mm
1.6x
10x
16x
‡ 14.5mm
1x
16x
16x
‡ 11.25mm
1.6x
16x
25.6x
‡ 9mm
Range of PD adjustment 10x eyepiece
55mm to 82mm
16x eyepiece
51mm to 78mm
Diopter adjustment
10x eyepiece
r8D
16x eyepiece
r10D
Illumination --------------------------------------------------------------------------------Slit projection magnification 2/3x
Slit width
continuous from 0mm to 9mm
(at 9mm , slit becomes a circle )
Slit height
continuous from 1mm to 8mm
Aperture diameter
‡9mm, ‡8mm, ‡5mm, ‡3mm,
‡2mm, ‡1mm,‡0.2mm
Slit angle
0o to 180ocontinuously adjustable from vertical to horizontal
Slit inclination
5o, 10o, 15o, 20o four steps
Filter piece
Heat-absorbing, grey, red-free, blue
Illumination bulb
12V30W halogen bulb
Movement base ----------------------------------------------------------------------------Fore and back movement
90mm
Left and right movement
100mm
Fine movement
15mm
Vertical movement
30mm
Chin-rest parts -----------------------------------------------------------------------------Vertical movement
80mm
Fixation target
Red LED
Hruby lens ----------------------------------------------------------------------------------Hruby lens
-58.7D (optionally available in some regions)
24
Power source -------------------------------------------------------------------------------Input voltage
100/220V ~r10%
Input frequency
50/60 Hzr1Hz
Input power
58VA
Output voltage
Illumination bulb 7.2V, 9.8V, 11.6V
Fixation target
7.2V
Electric safe standard
Conform to Standard IEC601 -1 , Class I Type B
Dimension and weight --------------------------------------------------------------------Packing box
720mm x 495mm x 480mm
Total weight
24 Kg
Net weight
21 Kg
— Subject to change in design or specifications without advance notice.