Nuvita 2090 digital thermometer User manual

Nuvita 2090 digital thermometer User manual

Below you will find brief information for digital thermometer 2090. The device is a non-contact infrared thermometer that can be used to measure both body and object temperatures. It features a backlit LCD display, memory function, and automatic shut-off. The device also has a beeper that can be turned on or off, and an alarm that can be set to alert the user if the temperature exceeds a certain threshold.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Nuvita 2090 User Manual | Manualzz

EN

Thermometre

No Contact

1 INTRODUCTION & CLASSIFICATION

This thermometer is classified as a Class II Medical Device, sorted as internally- powered equipment and type B application device. Water proof rating is: IPX0. It is prohibited to use this thermometer in flammable anesthetic gas or gas mixture of air and oxygen or nitrous oxide.

This is a continuous operation equipment.

LCD

 

Setting buttons

Measurement button

Probe

Battery lid

2 TECHNICAL PARAMETERS

Measurement method

Effective distance

Range

Accuracy

Resolution

Working condition

Storage condition

Power supply

Power consumption

Power level indicator

Memory

Display

Reading Scale

Automatic shut-off

Dimensions

Net weight

Standards

Human Body

Object

Non-contact

5cm~8cm

35.5

°C~42.9°C

(95.9

°F~109.2°F)

0 °C~118°C

(32 °F~244°F)

Human body

Object

±0.2

°C/0.4°F

±1.0

°C/1.8°F

0.1

°C/°F

10 °C~40°C (50°F~104°F)

RH≤80% Non-condensing

-25 °C~55°C (-13°F~131°F)

RH≤90% Non-condensing d.c. 3V (2x AA alkaline batteries)

When off

When measuring

≤10µW

≤30µW

Indication for low power level

Automatic recall of last reading

LCD back-lit

Celsius or Fahrenheit in 30 seconds

93mm x 153mm x 41mm

125g

EN60601-1

EN 12470-5

ASTM 1965-98

3 DISPLAY AND ICONS

FUNCTION

Battery Level

Beeper

Measurement mode

Reading scale

Reading display

Memory

 

ICON

 

Body temp

Object temp

 

 

Visible

Flashes

Invisible

Visible

Flashes

DETAILS

The battery is low, but the thermometer is functioning properly. Please replace battery as soon as possible.

The battery is exhausted and thermometer can not function properly. Please replace battery immediately.

Battery power is sufficient.

Beeper is on.

Beeper is off.

Body mode

Object mode

Celsius reading

Fahrenheit reading

Temperature value

Temperature value of last measurement

4

 

Menu

FUNCTION DEFINITION OF BUTTONS

®

Measurement mode

Settings

BUTTON

Body/object

Menu

+

-

DETAILS

To switch measurement mode between body and object

Activates menu and save settings

Increases parameter or select relevant parameter

Reduces parameter or select relevant parameter

5 SETTING

User can change reading scale between Celsius or Fahrenheit, switch on or switch off beeper, set alarm value, set deviation value and change measurement mode between human body or object. Please refer to below chart.

MENU FUNCTION

F-1

F-2

F-3

F-4

Reading scale setting

Beeper on/off

Alarm value setting

Deviation value setting

SAVE

+ DEFAULT REMARKS

°C °F °C

Off

Decrease

0.1°C

Decrease

On

Increase

0.1°C

Increase

On

38°C

0.1°C 0.1°C

°C

Save and automatically shut off

Effective range

37°C~42°C

Effective range

0°C~3°C

Measurement mode setting

When thermometer is on, it displays current measurement mode ( fig.8.1). Switch the “Body/Object” button to change measurement mode ( fig.8.2).

 

[fig 8.1]

 

[fig 8.2]

Reading scale setting: F-1

Activates setting menu by pressing “MENU” button. F-1 will be displayed. Press “+” to select Fahrenheit ºF as temp unit

(icon “°F” is flashing, fig.8.3); or press “-” to select Celsius ºC as temp unit (icon “°C” is flashing, fig.8.4).

 

[fig 8.3]

 

[fig 8.4]

Beeper setting: F-2

Press “MENU” again to save the setup and F-2 will be displayed (fig.8.5). Beeper is defaulted as “OPEN”. Press “+” to switch on beeper and

“OPEN” will be displayed (fig.8.6). Press “-” to switch off beeper and “CLOS” will be displayed

(fig.8.7).

 

[fig 8.5]

 

[fig 8.6]

 

[fig 8.7]

Alarm value setting: F-3

Press MENU again to save the setup and F-3 will be displayed (fig.8.8). Defaulted value

“38.0°C” will be displayed (fig.8.9). Press “+” or “-” to increase or decrease alarm value

(fig.8.10). Value will be increased or decreased by 0.1°C step by step.

 

[fig 8.8]

Deviation value setting: F-4

Press MENU again to save the setup and F-4 will be displayed (fig.8.11). Defaulted deviation value is 0.8°C (fig.8.12). When defaulted deviation value is flashing, press “+” or “-” to increase or decrease deviation value (fig.8.13).

Deviation value will be increased or decreased by 0.1°C step by step.

 

[fig 8.11]

 

 

[fig 8.9]

[fig 8.12]

 

[fig 8.10]

Save and exit (automatic shut off)

 

Press MENU to save the setup and “SAVE” will be displayed

(fig.8.14). All setup will be saved and thermometer will automatically shut off (fig.8.15).

After activating setup menu, users may select specific functions(F1/F2/F3/F4) directly for specific setting.

New setup will not be saved if the thermometer is shut off

[fig 8.14] unexpectedly during the procedure of setting.

After activating setup menu, the thermometer is unable to measure temperature. There will be no response when pressing measurement button.

!

Notice

Temperature reading under human body mode is obtained from dynamic compensation of environmental temperature and forehead surface temperature.

Object temperature mode is to test surface temperature of an object. The temperature obtained from the forehead under this mode is merely temperature of the skin surface but not core body temperature.

Deviation setting is able to adjust measurement value from 0°C to 3.0°C according to testing distance, surrounding temperature, skin difference, etc. Defaulted value is 0.8°C.

For example: if the tested body temperature is 36.2°C while the temperature is supposed to be 37.0°C, “F-4” to set up the value and adjust to 1.6°C, then the measurement value will be 37.0°C.

6 MEASUREMENT

6.1

Body temperature

Press measurement button to turn on thermometer and it displays boot screen (fig.9.1). After POST and two beeps, it will display value of last reading and be ready for measurement (fig.9.2).

Make sure the thermometer is under body mode.

   

Keep distance at 5cm to 8cm from upper eyebrows to the probe (fig.9.3). Press measurement button and when it gives a “beep” measurement is finished and value will be displayed (fig.9.4). If measurement value is exceeding alarm value (Defaulted value is

38°C), it gives 3 successive beeps as an indication.

 

[fig 9.1]

 

[fig 9.2]

After measurement, if the thermometer is idle for

30 seconds, it will display “OFF” (fig.9.5) and gives a

“beep” and shut off automatically.

[fig 9.3] [fig 9.4]

!

Notice

Keep the sensor and probe cavity clean before and after use.

To ensure the accuracy of measurement, it is recommended to start measurement ten minutes after carrying the thermometer to a new environment.

 

[fig 9.5]

Wait for 10 minutes to measure body temperature after measuring extremely high temperature or extremely low temperature.

Wait for 5 minutes to start a measurement when the measuring target (object or human) is from an environment with enormous difference in temperature.

Breeze, water, sweating, cosmetic on the forehead may affect measurement. Do not measure body temperature 30 minutes after exercise, bath or meals.

6.2

Object temperature

Press measurement button to turn on thermometer

(fig.9.6).

Make sure the thermometer is under object mode.

 

Keep vertical distance at 5cm to 8cm from object to measurement probe. Press measurement button and when it gives a “beep” measurement is finished and value will be displayed (fig.9.7).

[fig 9.6]

After measurement, if the thermometer is idle for 30 seconds, it will display “OFF” (fig.9.8) and gives a beep and shuts off automatically.

!

Notice

The value under this mode is object surface temperature instead of core temperature.

The defaulted infrared emissivity is 0.95. The reading will be deviated from the real temperature according to different emissivity. For example, the reading on stainless steel will be obviously lower than real temperature. BE CAUTIOUS FOR SCALDING.

 

 

[fig 9.7]

[fig 9.8]

6.3

Exceeding measurement range

Body mode: When measurement value is lower than 35.5ºC, it displays “Lo” (fig.9.9) and gives 4 successive beeps. When measurement value is higher than 42.9 ºC, it displays Hi (fig.9.10) and gives 4 successive beeps.

     

[fig 9.9]

 

[fig 9.10]

 

[fig 9.11]

Object mode: When measurement value is lower than 0ºC, it displays “Lo” (fig.9.11) and gives 4 successive beeps. When measurement value is higher than 118.0ºC, it displays “Hi” (fig.9.12) and gives 4 successive beeps.

!

Notice

[fig 9.12] [fig 9.13]

When surrounding temperature is lower than

10.0ºC or higher than 40.0ºC, it displays “Err” (fig.9.13). Measurement is not possible and hence accuracy is not assured.

7 BATTERY REPLACEMENT

Open the battery lid and take out the exhausted battery.

Put in 2x AA alkaline batteries and shut the battery lid.

After new battery is installed, thermometer will give 2 successive beeps. If no beeps are heard, check if the positive and negative pole are correct (see fig.10.1).

 

[fig 10.1]

!

Notice

Take out battery in case the thermometer is not used for long period. Do not discard the battery in a fire. Dispose of your batteries according to local regulations.

8 MAINTENANCE & TIPS

Make sure the sensor and probe cavity are clean otherwise accuracy will be affected.

Cleaning method for probe:

Use a cotton stick or soft cloth with water or alcohol to wipe the casing.

Use a cotton stick or soft cloth with alcohol to wipe the sensor surface or probe cavity gently. Do not use thermometer before alcohol is vaporized.

Read this manual book thoroughly before use. Make sure that the batteries are well installed.

It is not recommended to put the thermometer in any liquid or expose to strong sunlight or extremely low temperature.

Strong impact to the product will cause damage.

Do not dismantle this thermometer by yourself.

Keep the thermometer away from children’s reach.

Do not use the thermometer in cases of strong magnetic intereference.

The measurement results can fluctuate due to improper measurement procedures.

Please practice adequate measurement in order to improve your skill.

The measurement results cannot supersede a doctor’s diagnosis.

Special maintenance is not necessary for this thermometer. Please contact distributor or manufacturer in case of malfunction.

9 TROUBLESHOOTING

DESCRIPTION

LCD display

“Lo” or “Hi”

SOLUTION

Breeze, water, sweating, cosmetic on the forehead may affect measurement.

Check deviation value setting. Defaulted value is 0.8°C.

If the environmental temperature changes enormously or if the thermometer is used directly from high temperature to very low temperature, the measurement difference will happen. The thermometer should be kept in a relative stable environment for 10 minutes to obtain heat balance before starting a new measurement. Ensure measurement distance is

5cm to 8cm.

No response when pressing measurement button

No or improper display

No beeper

Take out and reassemble battery.

Check if the thermometer is under menu setup. In procedure of menu setting, thermometer is unable to measure and therefore no response.

Take out battery and install battery again.

Check if beeper is switched off.

10 QUALITY COMMITMENT AND AFTER-SALE SERVICE

This product benefits of a warranty of 24 months on material and manufacturing defects, starting from the date of purchase (see sales slip).

The warranty of 24 months does not include damages caused by the usual use of parties identified as “consumable” (e.g., batteries, brush heads or parts subject to usury).

The legal guarantee of 24 months is void if:

The product has undergone aesthetic damage due to improper use not in accordance with instructions in the manual.

This product has been modified and / or tampered with.

The cause of the failure was due to poor maintenance of the individual components and / or accessories and / or supplies (e.g. oxidation and / or scaling due to the retention of water or other liquids, sediment blocking the sensor, leak of corrosive liquid from batteries.)

The following is excluded from the legal guarantee of 24 months:

Costs related to replacements and / or repair of parts subject to wear or costs for ordinary maintenance of the product.

The costs and risks involved in transporting the product to and from the store where you purchased or otherwise authorized collection centre to receive the products under warranty.

Damage caused by or resulting from improper installation or improper use not in accordance with the directions in the instruction manual.

Damage due to natural disasters, accidental events or adverse conditions not compatible with the product.

Defects that have a negligible effect on product performance.

The manufacturer, distributor and all the parties involved in the sale do not assume any liability for losses and economic damage from any malfunction of the product. In accordance with current regulations the manufacturer, distributor and all the parties involved in the sale are not responding in any case for damages, including direct, indirect ones, loss of net income, loss of savings and additional damage and other details consequences going beyond the damage caused by the breach of warranty, contract, strict liability, wrongdoing or due to other causes, resulting from the use or inability to use the product and / or paper and electronic documents, including the lack of service.

For further information on the help service visit the website www.nuvitababy.

com

Shut off right after switching on

Check battery level or take out and install the battery again.

advertisement

Key Features

  • Non-contact measurement
  • Body and object temperature measurement
  • Backlit LCD display
  • Memory function
  • Automatic shut-off
  • Beeper
  • Alarm setting

Frequently Answers and Questions

What is the effective distance for measuring body temperature?
The effective distance for measuring body temperature is 5cm to 8cm from the upper eyebrows to the probe.
How do I change the measurement mode between body and object?
Press the "Body/Object" button to switch between body and object measurement mode.
Can I set an alarm value?
Yes, you can set an alarm value to alert you if the temperature exceeds a certain threshold.
How do I replace the batteries?
Open the battery lid and take out the exhausted battery. Put in 2x AA alkaline batteries and shut the battery lid.
How do I clean the sensor and probe cavity?
Use a cotton stick or soft cloth with water or alcohol to wipe the casing. Use a cotton stick or soft cloth with alcohol to wipe the sensor surface or probe cavity gently. Do not use the thermometer before the alcohol is vaporized.

Related manuals