The Model 4320 4-in-1 Electrical Safety Tester is designed as a cost-effective
solution for manufacturers, who want to perform AC Hipot, DC Hipot, Insulation Resistance
and Ground Bond tests in a single instrument. This all-in-one test solution provides the
flexibility of an advanced test system without the high price. It provides only the features and
functionality needed to cost-effectively perform multiple electrical safety tests.
Features & Benefits
Feature
Benefit
4-in-1 Electrical Safety Tester
The 4320 combines AC Hipot, DC Hipot, Insulation
Resistance and Ground Bond tests in a single
instrument, which takes up less rack space and allows
a single DUT connection. It efficiently performs the 4
most common safety agency specified electrical safety
tests.
Liquid Crystal Display
Allows the user to clearly monitor the test parameters
as well as the test results.
Multiple Test Memories
The 4320 provides six test memories with six steps per
memory. This is beneficial for manufacturers testing
multiple products on one production line.
Continuous Duty Cycle
An advantage over other instruments in that the 4320
does not have any requirements to be periodically shut
down even when used at full output.
Low Current Sense
Monitors the maximum level of current flow, thus
ensuring that the Device Under Test (DUT) is properly
connected and that the test is properly being
performed.
No Load Setup
This provides the operator with an easy and safe way
to set trip currents and output voltages.
Electronic Ramp and Dwell
These electronic controls help to keep test results
consistent by ensuring that the test duration is
consistent.
Milliohm Offset Capability
This feature minimizes the effect of test lead or test
fixture resistance.
and Milliohm Trip Currents
Adjustable Output Current
The 4320 is versatile enough to
meet a wide variety of safety agency specifications.
Model 4320 Electrical Safety Tester
The 4320 includes a standard 9-pin interface providing the inputs and outputs
for all the basic remotes required for PLC relay control. It also provides rear panel outputs.
Model 4320 Specifications
INPUT
INSULATION RESISTANCE TEST MODE (Cont.)
115/230 VAC ± 10%, User selection
50/60 Hz ± 5%
6.3 Amp 250V Slo-Blo
VOLTAGE
FREQUENCY
FUSE
HI-LIMIT
LO-LIMIT
DELAY TIMER
DIELECTRIC WITHSTAND TEST MODE
OUTPUT RATING
OUTPUT ADJUSTMENT
HI-LIMIT
LO-LIMIT
FAILURE DETECTOR
OUTPUT RIPPLE
WAVE FORM
OUTPUT FREQUENCY
OUTPUT REGULATION
DWELL TIMER
RAMP TIMER
5 KV @ 20mA AC
6 KV @ 5 mA DC
Range:
Resolution:
Accuracy:
AC Range:
Resolution:
Accuracy:
DC Range:
Resolution:
Accuracy:
AC Range:
Resolution:
AccurÏacy:
DC Range:
Resolution:
Accuracy:
Audible and Visual
< 5% at 6KVDC/5 mA
Sinewave Distortion : < 2% THD
Range:
60 or 50 Hz, User Selection
± (1 % of setting + 5 volts) no load to full load
Range: 0, 0.2 - 999.9 sec, (0 = Constant)
0.1 sec increments
Range: 0 & 0.2 - 999.9 sec
0.1 sec increments
VOLTAGE DISPLAY
RESISTANCE DISPLAY
Visit us at
Range:
Resolution:
Accuracy:
Range:
Resolution:
Accuracy:
Range:
Resolution:
0 - 1000 MΩ (0 = Off)
1 - 1000 MΩ
0, 0.5 - 999.9 sec (0 = Constant)
0.1 sec/step
± (0.1% + 0.05 sec)
GROUND BOND TEST MODE
0 - 5 KV AC
0 - 6 KV DC
10 volt/step
± (2% of setting + 5 volts)
0.10 - 20.00 mA
0.01 mA/step
± (2% of setting + 0.02 mA)
0.00 - 5.00 mA
0.01 mA/step
± (2% of setting + 0.02 mA)
0.00 - 20.00 mA
0.01 mA/step
± (2% of setting + 2 mA)
0.00 - 5.00 mA
0.01 mA/step
± (2% of setting + 0.02 mA)
INSULATION RESISTANCE TEST MODE
OUTPUT VOLTAGE
Range:
Range:
Range:
Resolution:
Accuracy:
100 - 1000 Volts DC
1 volt/step
± (2% of reading + 5 volts)
0 - 1000 V
1 volt/step
± (2% of reading + 2 counts)
1 - 1000 MΩ (4 Digit, Auto Ranging)
500VDC
1000VDC
MΩ
MΩ
MΩ
0.01 1.00 - 40.00 1.00 - 80.00
0.1
35.0 - 999.9 75.0 - 999.9
OUTPUT VOLTAGE
OUTPUT FREQUENCY
OUTPUT CURRENT
CURRENT DISPLAY
HI-LIMIT
DWELL TIMER
MILLIOHM OFFSET
Range:
Range:
Accuracy:
Range:
Resolution:
AC 6 volts fixed
60 or 50 Hz, User Selection
±0.1%
3.00 - 30.00 Amps AC
0.1 Amp/step
± (2 % of Setting + 0.02 A)
Range:
0.00 - 30.00 Amps
Resolution: 0.1 Amp/step
Accuracy: ± (3 % of Reading + 0.1 A)
Range:
0 - 510 mΩ
0 - 510 mΩ for 0.0 - 10.0 Amps
0 - 200 mΩ for 10.1 - 25.0 Amps
0 - 150 mΩ for 25.1 - 30.0 Amps
Resolution: 1 mΩ/step
Accuracy: ± (2 % of Setting + 2 mΩ)
Range:
0, 0.5 - 999.9 sec (0 = Constant)
Resolution: 0.1 sec/step
Accuracy: ± (0.1% + 0.05 sec)
Max Offset Capability: 100 mΩ
Resolution:
1 mΩ / step
Accuracy:
± (2 % of Setting + 2 mΩ)
GENERAL SPECIFICATIONS
PLC REMOTE CONTROL
Input - Test, Reset, Interlock & Recall Memory 1 - 6
Output - Pass, Fail, Test-in-Process
MEMORY
Allows storage of up to 6 different test programs
with 6 steps per memory and single step mode.
SECURITY
Lockout capability to avoid unauthorized access to
test set-up programs.
CALIBRATION
Software and adjustments are made through
front panel.
MECHANICAL
Bench or rack mount with tilt up front feet.
DIMENSION
(W x H X D) 11 x 3.5 x 17 (280 x 89 x 430 mm)
WEIGHT
< 40 lbs. (18 kg)
WWW.HIPOT.COM or call us at 1-800-504-0055
28105 N. Keith Drive
Lake Forest, IL., 60045 U.S.A
Phone: 1-847-932-3662
Fax: 1-847-932-3665
E-mail: info@hipot.com
Web Site: www.hipot.com