OM16 milliohmmeter / microhmmeter is firstly a
designed for outdoor uses instrument but also in
workshop , and maintenance departments.
Easy to use, it performs measurements on
inductive and non-inductive resistances with a DC
current.
Its large interactive display informs in real time the
operator about the type of measurement, range,
calculation conditions (resistance calculated
according to a reference temperature), thresholds
status and values.
Designed for hard conditions:.
- Hard case (Pélicase type).
- Plastic washable keyboard.
- Key-locked measurement plugs.
Resistance is automatically calculated by the
instrument taking into account: measurement
value, ambient temperature, metal
temperature coefficient, reference
temperature
All parameters are programmable. (Ambient
temperature can be programmed or measured
thanks to the temperature sensor)
Why choosing an OM 16
•
4 wires measurement to provide highest
accuracy
•
Automatic measurement in non inductive
resistance mode. :automatic triggering when
continuity is established between the two
points :a single operator can perform
measurements.
•
EMF parasites are measured before each
measurement
•
Permanent measurement current for inductive
loads (coils, transformers, motors windings).
•
DC measurement current for non-inductive
resistances (earth bonding, coating, contact
resistances) with auto power off of the current.
•
•
2 programmable thresholds with LED indication
and display on the screen, light or loud beeper.
•
Memory: 1 000 measurements identified by
numbers.
Memory reading on the display, or via software
or printer.
•
Powered by NiMh batteries, quick charge
Display of temperature compensated resistance
value:
TECHNICAL CHARACTERISTICS
Range
Resolution
5 mΩ
25 mΩ
250 mΩ
2500 mΩ
25 Ω
250 Ω
2 500 Ω
0,1 µΩ
1 µΩ
10 µΩ
0,1 mΩ
1 mΩ
10 mΩ
100 mΩ
Accuracy (1 year)
23°C ± 5°C
0,05% + 0,5 µΩ
0,05 % + 3 µΩ
0,05 % + 30 µΩ
0,05% + 0,3 mΩ
0,05 % + 3 mΩ
0,05% + 30 mΩ
0,05% + 300 mΩ
Measuring current
Voltage drop
10 A
10 A
10 A
1A
100 mA
10 mA
1 mA
50 mV
250 mV
2,5 V
2,5 V
2,5 V
2,5 V
2,5 V
Accuracy given in % of reading + fixed value at 23°C ± 5°C
Ed 13/02/03
AOIP reserves the right to modify this document without prior notice
OM 16: ON-SITE MICROHMMETER
Temperature compensation
Manually keyed or measured by Pt100.
Resolution: 0,1°C, accuracy ± 0,5°C.
Sensor connected on the keyboard or through a cable.
Climatic environment according to CEI359
- Reference domain: 23 ± 5°C ( RH :45 to 75% w/o
condensing).
- Temperature Coefficient from 0 to 18°C and from 28 to
50°C: <10% of accuracy/°C.
- Functioning reference domain: 0 to 50°C (RH: 20 to 75%
w/o condensing).
- Limit functioning domain: - 10°C to + 55°C (RH: 10 to
80% w/o condensing).
- Storage temperature limits: - 40°C to + 60°C.
Environment
- Dimensions: 270 x 250 x 180 mm.
- Weight: 4Kg.
- Chocks and vibrations: EN61010-1.
- IP53 according to EN60529.
Battery system
- Batteries: NI/MH 8,5 AH.
- Charging time: 5H.
- Power supply: from 92 to 256 V (45 to 400 Hz).
- 5000 measurements at 10A
Safety
- According to EN 61010-1. Category II, pollution 2.
- 60 V
EMC
Radiant Emission:
EN 55022, class B; EN 61000-3-2; EN 61000-3-3
Immunity:
• EN 61000-4-2
• EN 61000-4-3
• EN 61000-4-5
• EN 61000-4-6
• EN 61000-4-11
• EN 61000-4-4
Measurement plugs
2 circular locking connectors.
Protection up to 250V.
Fuse protection.
RS232 Interface for printer, computer, trigger
•
ACCESSORIES
- Large Kelvin clip: AMT 006
Opening: 30 mm; cable of 5 m.
- Small Kelvin clip : AMT012
Opening: 12 mm; cable of 5 m.
- Compact test probe: AMT011
Length: 125 mm, cable of 5 m.
Diameter of rod: 4 mm.
-Test probe: AMT005
Length: 215 mm; cable of 5 m.
Diameter of rod: 3 mm.
-Trigger test probe: AMT013
Length: 215 mm; cable of 5 m.
Diameter of rod: 3 mm.
Clamps and test probes are sold by unit
-Temperature sensor: AMT014
Pt 100; Accuracy: ± 0,5°C.
-Cable for AMT 014: AMT015
Length 2m.
-Exploitation software: LOG OM
• PRINTERS
PX 58: Battery model
Paper larger 58mm
CX 85: DC power
model
Paper larger 85mm
(Serial cable supplied
In standard with
printers).
AOIP
BP 182
91133 RIS ORANGIS CEDEX France
Tel : +33 (0) 169 028 900
www.aoip.com
Ed 13/09/03
Accréditation
n° 2.1525
Température
n° 2.1524
Electricité-magnétisme
AOIP reserves the right to modify this document without prior notice
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising