Data Sheet 20.2550
Page 1/10
dTRANS Az 01
µP Indicator/Controller
for analytical measurement
Type 202550
Panel-mounting housing to DIN 43 700
Type 202550 / …
Brief description
The instrument with 96mm x 48mm bezel and plug-in controller module visualizes and controls
variables in analytical measurement (pH value, redox voltage, conductivity, high-purity water,
oxygen, free chlorine, chlorine dioxide, ozone etc.).
The indicator has two analog and two logic inputs. The first analog input is suitable for
connecting a 0(4)—20 mA signal, which can be provided by any type of transmitter (including
2-wire types). The input signal is conditioned, as determined by the configuration, and
displayed. One special feature of the instrument is that the calibration procedures for pH, redox
and conductivity are internally programmed. The second analog input can be used to connect
Pt100 or Pt1000 resistance thermometers.
The instrument features two 4-digit 7-segment displays for indicating the main variable (red) and
the temperature (green). The temperature display is switched off in the default setting. A
separate temperature sensor (Pt100 or Pt1000) can be connected to the second analog input.
This can then be used to display the temperature of the medium and, if required, monitor it by
means of a limit switch. During programming, the displays provide comments on the inputs. A
great variety of control tasks can be handled by the various output options (relay contacts and
/ or analog outputs). The two relay “make” contacts that are provided on the instrument as
standard can be configured as a limit controller and / or pulse width or pulse frequency
controllers, or as a modulating controller. To obtain analog (continuous) controller outputs, the
optional analog outputs must be configured accordingly. All controller outputs can be
configured for a P, PI, PD or PID control action. In the basic version, the instrument provides two
relay “make” contacts and one logic output (0/5V). Two additional outputs can be, according to
choice, fitted with relay changeover contacts and / or analog outputs (process value output or
analog controller output), or with a serial interface (Profibus-DP or MODbus / Jbus protocol)
and /or a supply for 2-wire transmitters.
Block structure
5 outputs
2 analog inputs
Outputs 1+2
- relay
Input 2:
Temperature
manual entry or
Pt100 / Pt1000
Output 4
- logic 5 V or
- logic 12 V
2 logic inputs
for floating contacts
Functions:
- key inhibit
- range expansion (x10)
- freeze measurement
- alarm stop
- setpoint switching
- hold
- reset alarm time
Supply voltage
110 — 240 V AC +10%/-15%
48 — 63 Hz
20 — 53 V AC/DC 48 — 63 Hz
extra code / option
07.08 / 00381059
Transmitter / controller
Input 1:
standard signal 0(4) — 20 mA
Output 3
- relay or
- analog process value output or
- analog continuous controller output
or
- 18 V DC supply voltage output for
2-wire transmitter
Output 5
- Profibus-DP or
- serial interface RS422 / RS485
with Modbus/Jbus protocol or
- analog process value output or
- relay or
- analog continuous controller output
Type 202550 / …/640
Special features
❏ Panel-mounting unit,
just 96 x 48 x 110 mm
❏ Display of pH, mV, µS/cm, mS/cm, mg/l,
etc.
❏ 2 isolated process value outputs
0(4) — 20 mA / 0(2) — 10 V, freely
selectable and scalable for the main
variable or temperature, or proportional
controller (option)
❏ 2 relays as standard, freely programmable
as limit controller or P, PI, PID, PD controller
with pulse width or pulse frequency output,
or as modulating controller
❏ 2 logic inputs
❏ OPTION: Profibus-DP or serial interface
RS485 /422 with MODbus /Jbus protocol
❏ Protection IP65
❏ Wall-mounting housing, Protection IP65
Indicator/controller for
pH value or redox voltage
❏ Simple, guided calibration procedure
❏ Temperature compensation is possible
Indicator/controller for conductivity
❏ Calibration procedure for the relative cell
constant
❏ Calibration procedure for the temperature
coefficient of the measuring solution
❏ Temperature compensation is possible
Indicator/controller for free chlorine,
chlorine dioxide, ozone
❏ Connection of a sensor (according to Data
Sheet 20.2630, for example)
❏ Integrated calibration procedure
Universal indicator/controller
❏ Display range is freely scalable from
-1999 to 9999 digit
❏ Different calibration procedures
Approvals
JUMO GmbH & Co. KG• 36035 Fulda, Germany
Operation
Data Sheet 20.2550
Page 2/10
Indications / controls
For easy programming and operation, the
controller parameters and configuration data
are assigned to various levels.
(8)
(1)
K1
(7)
K2
K3
K4
(2)
(6)
(3)
°C
PGM
EXIT
(4)
(5)
Code words protect the levels from
unauthorized access.
Membrane keys ensure simple and userfriendly operation.
The two LED displays show the parameter
symbols and the corresponding values.
Operating level
The lower display, for example, shows the
symbol, the upper display the corresponding
value. Setpoints SPr1 and SPr2 can be
altered from the membrane keys.
(1) Status indicators (yellow)
for outputs 1 to 4
(6) 4-digit temperature display
(LED, green, 8mm high)
(2) Increment key for altering parameters
or manual operation of relay K2
(7) 4-digit process value indication
(LED, red, 13mm high)
(3) Decrement key for altering parameters
or manual operation of relay K1
(3) “CAL”: Initiating electrode calibration (1+ point or 2-point calibration)
(5)
(4) EXIT key to leave
the levels
(2)
+ Initiate manual operation or “hold”
(4)
(5) PGM key for selection of parameters and
confirmation of entries
Calibration options
Parameter level
The controller is adapted to the control loop at
this level. The appropriate parameters appear
here, with symbol and value. Only those
parameters will be indicated which
correspond to the configuration of the
controller (configuration level).
Configuration level
This level is used to adapt the controller to the
control task, or for adaptation of the inputs
and outputs.
07.08 / 00381059
If a transmitter / sensor without integral
calibration function is used as a signal source
(standard current signal), the dTRANS Az 01
can be configured accordingly, as indicator /
controller (for the pH value, for example). It
will then offer one of the following calibration
options, in accordance with the configuration.
Indicator/controller for pH value
The electrode parameters of a pH
combination electrode are subject to
manufacturing tolerances and variations
depending on usage. To compensate for
these changing electrode parameters, the
instrument offers two guided calibration
procedures:
1) 2-point calibration (standard)
This method of calibration makes a fresh
determination of the electrode zero and
slope using two solutions with known pH
values (e.g. buffer solutions).
It should be given preference!
2) 1-point calibration
In 1-point calibration, only the electrode
zero is freshly determined using a solution
with a know pH value (buffer solution).
Problems arising from an incorrect
electrode slope will not be detected by the
user!
This method should only be adopted in
cases where the electrode is not subject to
significant chemical and mechanical
influences.
In addition to the calibration procedures
described above, the instrument offers the
facility of manually entering and adjusting the
zero point and slope (as determined by a
laboratory, for example).
Indicator/controller for redox voltage
The electrode zero of a redox combination
electrode is subject to manufacturing
tolerances and variations depending on
usage.
The instrument provides the facility for a
guided calibration procedure – the 1-point
calibration – for the fresh determination of the
electrode zero, using a buffer solution or a
solution with a known redox voltage.
In addition, the instrument offers the facility of
manually entering or adjusting the electrode
zero (as determined by a laboratory, for
example).
Indicator/controller for electrolytic
conductivity
Calibration of the cell constant
Subject to manufacturing tolerances, the cell
constant of a conductivity cell may deviate
slightly from its nominal (printed) value. In
addition, the cell constant may change during
operation (due to deposits or wear, for
example). This results in a change of the
output signal from the cell. The instrument
offers the user the possibility of compensating
any deviation from the nominal value of the
cell constant through manual entry (range 80
— 120%) or automatic calibration of the
relative cell constant Krel.
Calibration of the temperature coefficient α
The conductivity of almost all solutions
JUMO GmbH & Co. KG • 36035 Fulda, Germany
depends on the temperature. To ensure
correct measurement, it is therefore
necessary to know both the temperature and
the temperature coefficient α [% per °C] of the
measuring solution. The temperature can
either be measured automatically, with a
Pt100 or Pt1000 temperature probe, or set
manually by the user. The temperature
coefficient can be determined automatically
by the instrument or entered manually, in the
range from 0 to 5.5 %/°C
Indicator/controller for free chlorine,
chlorine dioxide and ozone
(as per Data Sheet 20.2630)
The sensor slope is subject to manufacturing
tolerances and changes that develop during
operation.
The instrument offers the facility for a guided
calibration procedure - the 1-point calibration
- to freshly determine the sensor slope
through comparative measurement.
Additional functions of the
JUMO dTRANS Az 01
❏ Programmable response of the process
value output to underrange/overrange
On underrange or overrange, the process
value output can move to the following
operational states:
-4%, 0%, 100% or 110% freely selectable
Example: The instrument is programmed to
4 — 20 mA corresponding to
100 — 500 mV.
It can be set up so that, on falling below 100,
the output signal 4 mA (0%) is either
maintained or will jump to 3.84 mA (-4%). The
3.84 mA value can then be recognized as
“irregular” by a connected PLC.
❏ Bilinear output
This function divides the signal for the analog
process value output into two linear portions
(0 — 50% and 50 — 100% of the output
signal), with a knee-point at 50% of the output
signal. The knee-point of the characteristic
can be shifted along the dotted 50% line. The
factory setting of 50% produces a straightline characteristic.
.
The bilinear characteristic is used when the
“normal” measurement range is likely to be
frequently exceeded.
07.08 / 00381059
Data Sheet 20.2550
Example: The normal measurement range
spans 0—20µS/cm.
However, measurements of up to 80µS/cm
can also occur.
In this case, the range 0 — 100 µS/cm will be
selected, and the knee-point set at 20% of
this range (20% of 100 µS/cm corresponds to
20 µS/cm).
This results in measurements in the range
0 — 20 µS/cm being converted into an output
signal 0 — 10 mA.
Measurements in the range 20 — 100µS/cm
will be converted into an output signal 10 —
20 mA.
❏ The response of the controller relays to
“Hold” can be defined
“Hold” is initiated either manually, using the
keys, by a logic input, or by an alarm event.
The outputs of the relays K1 and K2 can move
to the following (programmable) states on
“Hold”:
0%
Relay de-energized
50% output
100% output
Output
accepted
For dynamic controllers,
50% of the maximum
pulse width or frequency
is produced
Relay is energized, or
maximum pulse width /
frequency
The present output
continues to be
produced
❏ In “Manual” mode, the relays K1 and K2
are operated manually, by using the
keys. Either key or switch operation can
be selected, by a setting at the parameter
level.
Key operation: The relay is switched as long
as the key is pressed (e.g. for manual dosing).
Switch operation: The first key stroke
switches the relay on, the second switches it
off again (toggle action), e.g. for emptying
large tanks.
❏ Simulation of the analog process value
output
The process value output (0/2 — 10 V or 0/
4 — 20 mA, depending on the setting) can be
switched in 10% steps from 0 — 100%, in the
manual mode.
Application: “Dry-run” commissioning of the
plant (without electrodes, fault search,
servicing).
❏ Controller output functions
Output 1 (relay): Switching, with pulse
frequency or pulse width action / limit
monitoring / switched off. Switching
function can be reversed.
MAX / MIN limit comparator.
Output 2 (relay): Switching, with pulse
frequency or pulse width action / limit
monitoring / MAX limit comparator for
temperature input / MIN limit comparator
for temperature input / switched off.
Switching function can be reversed.
MAX / MIN limit comparator.
Page 3/10
Output 3, relay or analog process value
output: “Hold” / alarm pulse contact; alarm
steady contact / MAX limit comparator /
MIN limit comparator / output of process
value (only for analog process value output)
/ output of temperature process value (only
for analog process value output) / analog
controller output (only for analog process
value output) / no function.
Output 4, logic output: “Hold“ / alarm pulse
contact / alarm steady contact / MAX limit
comparator / MIN limit comparator / no
function.
Output 5, relay or analog process value
output: “Hold” / alarm pulse contact; alarm
steady contact / MAX limit comparator /
MIN limit comparator / output of process
value (only for analog process value output)
/ output of temperature process value (only
for analog process value output) / analog
controller output (only for analog process
value output) / no function.
Limit comparator
(limit monitor)
A limit-monitoring function can be assigned to
the controller outputs 1 to 5 (depending on
the instrument version).
For each one, the direction of switching
(pulled in on going above /below a limit), pullin and /or drop-out delay, and a hysteresis
can all be defined.
Interface
The microprocessor indicator /controller can
be optionally fitted with an RS422 /RS485
interface. This is used for communication with
higher-level systems and integration into a
data network. The transmission protocol can
be either Profibus-DP or MODbus /Jbus.
Technical data
Inputs
Analog input 1
Input resistance approx. 40 Ω
Analog input 2
Resistance thermometer Pt100 or
Pt1000, in 2- or 3-wire circuit
-50 to +250•C
Measurement display in •C or •F (option)
Lead compensation, analog input 2
The lead resistance can be compensated in
software by a correction of the process value.
This is not required if the resistance
thermometer is connected in a 3-wire circuit.
When a resistance thermometer is connected
in a 2-wire circuit, lead compensation can be
provided by using an external compensation
resistor.
Functional description of logic inputs 1
and 2
The two standard logic inputs can be
operated by floating contacts (relays) from a
PLC, or by switches. The following functions
can be selected and assigned:
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Data Sheet 20.2550
Page 4/10
Key inhibit: The PLC or a key switch can be
used to lock the keys on the instrument, to
prevent unauthorized entries being made.
2.
Logic output, output 4
0/5V
Rload ≥ 250Ω
0/12V
Rload ≥ 650Ω (option)
Permissible
ambient temperature limits
-10 to +55°C
Setpoint switching: For comfortable process
control. As long as the logic input is not
operated, setpoint pair SPr1 and SPr2 is
active. If the appropriately configured logic
input is operated, then the second setpoint
pair is activated.
3.
Process value output, output 3 or 5
(option)
freely configurable:
0(2) — 10V
Rload ≥ 500Ω or
0(4) — 20mA
Rload ≥ 500Ω
electrically isolated from the inputs:
ΔU ≤ 30V AC or
ΔU ≤ 50V DC
Permissible storage temperature
-40 to +70°C
Freeze
measurement:
The
indicated
measurement and the process value output
no longer change.
Hold: This function can be used (for instance,
by a supervisory PLC) to put the instrument
into the secure “Hold” state. The response of
the controller to “Hold” is as previously
defined.
Hold reversed: The same function as for
“Hold”, but when the logic input is open.
Alarm stop: The alarm generation via the
configured output is reset or prevented, but
the alarm LED (e.g. K4) continues to blink as a
warning.
Deviation from characteristic of output
signal
≤ 0.25%, ± 50 ppm per °C
4.Relay, output 3 or 5 (option)
(only for instruments without a process
value output)
changeover contact
contact rating: 3A, 250V AC
with resistive load
contact life:
> 5x105 operations at rated load
5.
Reset alarm time: The alarm generation via
the configured output is prevented. The alarm
delay time is set to zero, but is restarted when
the logic input becomes inactive and the start
conditions are fulfilled once more.
Range expansion (x10): If only a small portion
of the measurement range is used, it may be
advantageous for the instrument to react to
0 — 10% of the process value by producing
0 — 100% of the output signal.
6.
Supply for
2-wire transmitter
(output 5, option)
18 V DC, max. 30 mA
Controller type
Limit controller and / or pulse width or pulse
frequency controller, proportional controller or
modulating controller, freely configurable and
selectable
Conductivity
0 — 9999 mS/cm or µS/cm
0 — 9.999 mS/cm or µS/cm
0 — 99.99 mS/cm or µS/cm
0 — 999.9 mS/cm or µS/cm
Control action
configurable as P, PI, PID or PD.
Sampling time
210msec
Universal display
-1999 to 9999 digit
-1.999 to 9.999 digit
-19.99 to 99.99 digit
-199.9 to 999.9 digit
Measurement circuit monitoring
Input 1: out-of-range
Input 2: out-of-range, probe short-circuit,
probe break
The outputs move to a defined (configurable)
status.
Deviation from characteristic
≤ 0.15% per 10 °C
Temperature display
-50 to +250•C (can be switched to •F)
Data backup
EEPROM
Deviation from characteristic
≤ 0.1% per 10 °C
Outputs
2 relay outputs, 1 logic output, 1 analog
process value output or 1 additional relay, and
1 serial interface are available.
07.08 / 00381059
Protocol
MODbus/Jbus or Profibus-DP
A/D converter
resolution > 15 bit
Redox voltage
-1999 to +1999 mV
Relay, output 1 / 2 (standard)
Make contact (n.o., can also
configured as n.c. break contact)
contact rating: 3A, 250V AC
with resistive load
contact life:
> 5x105 operations at rated load
Baud rate
4800/ 9600bps
General controller data
Display and control range
pH value
-1.00 to 14.00 pH
1.
Interface RS422 / RS485,
output 3 or 5 (option)
electrically isolated
be
Supply voltage
110 — 240 V AC +10%/-15%,
48 — 63 Hz or
20 — 53VAC/DC, 0 — 63/0 Hz
Power consumption
approx. 8V A
Electrical connection
via gold-plated faston connectors to
DIN 46 244/A; 4.8mm x 0.8mm
Permissible
ambient temperature
0 to +50°C
Climatic conditions
rel. humidity ≤ 75%, no condensation
Enclosure protection
to EN 60 529
front IP65 / rear IP20
Electrical safety
to EN 61 010
clearance and creepage distances for
- overvoltage category II
- pollution degree 2
Electromagnetic compatibility
to EN 61 326
radiated interference: Class B
interference immunity: to industrial
requirements
Housing
Panel-mounting housing in conductive plastic
to DIN 43 700, base material ABS
with plug-in controller module
Operating position
unrestricted
Weight
Switchgear cabinet unit (basic version)
approx. 320g
Wall-mounting housing (extra code /640)
approx. 1400g
Option
Wall-mounting housing
extra code /640
On request, the JUMO dTRANS Az 01 can be
supplied built into a surface-mounting housing. The housing is suitable for wall-mounting
or mounting on a DIN rail as per EN 50 022, 35
x 7.5 mm.
The housing is sturdy and provides IP67 protection for the built-in instrument. It is fitted
with six cable glands. Unused cable glands
can be tightly sealed using the blind grommets that are included in the delivery.
The electrical connection is made via screw
terminals.
(wire cross-section up to 2.5 mm2).
Standard accessories
– 2 fixing items (not with extra code /640
(wall-mounting housing))
– 1 Operating Instructions B20.2550.0.1
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Data Sheet 20.2550
Page 5/10
Parameters
Parameter
2
Value range
Comments
Alarm tolerance
pH / redox / conductivity1
AL1
0.00 — 99.99 pH
0000 — 9999 mV
0.000 — 99992
The alarm is only generated when the level
(setpoint + alarm tolerance) has been passed,
and the alarm delay time has elapsed (only
effective for pulse width /pulse frequency
controllers. It is internally fixed at 0 for limit
controllers.
Alarm delay
AL2
0 — 6000 sec
Delay time before the alarm contact is activated
Pb1
0.01 — 99.9 pH
1 — 9999 mV
1 — 99992
Proportional band 2
pH / redox / conductivity1
Pb2
0.01 — 99.9 pH
1 — 9999 mV
1 — 99992
Derivative time 1
dt1
Derivative time 2
dt2
Reset time 1
rt1
Reset time 2
rt2
Minimum ON time 1
(for limit controller or
pulse width controller)
or
minimum pulse width 1
(for pulse frequency controller)
tr1
Minimum ON time 2
(for limit controller or
pulse width controller)
or
minimum pulse width 2
(for pulse frequency controller)
tr2
Switching differential 1
pH / redox / conductivity1
HYS1
Switching differential 2
pH / redox / conductivity1
HYS2
Switching differential 3
pH / redox / conductivity1
HYS3
Switching differential 4
pH / redox / conductivity1
HYS4
Switching differential 5
pH / redox / conductivity1
HYS5
Pull-in delay 1
Ond1
Pull-in delay 2
Ond2
Pull-in delay 3
Ond3
Pull-in delay 4
Ond4
Proportional band1
pH / redox / conductivity1
1
Display
Pull-in delay 5
Ond5
Drop-out delay 1
Ofd1
Drop-out delay 2
Ofd2
Drop-out delay 3
Ofd3
Drop-out delay 4
Ofd4
Drop-out delay 5
Ofd5
Maximum pulse frequency 1
Fr1
Maximum pulse frequency 2
Fr2
Pulse period 1
Cy1
Pulse period 2
Cy2
Output level limit, output 1
Y1
Output level limit, output 2
Y2
Actuator time
tt
according to dimensional unit
unit as per configuration
07.08 / 00381059
0 — 9999 sec
Influences the P action of the controller
Influences the D action of the controller
If dt = 0, the controller has no D action.
Influences the I action of the controller
If rt = 0, the controller has no I action.
0.2 — 999.9 sec
Determined by the technical data of the dosing
device (solenoid valve, dosing pump)
0.01 — 99.9 pH
1 — 9999 mV
1 — 9999*
Defines the switch-off point for the control
contacts
0.2 — 999.9 sec
Delay time before the contact is activated
0.2 — 999.9 sec
Delay time until the contact moves back to the
initial position
0 — 150 pulse /min
Maximum frequency of pulses (operating a
dosing pump, for instance)
1.0 — 999.9 sec
The period in which a pulse is modulated
0 — 100%
The maximum output level for a pulse width /
pulse frequency controller
15 — 3000 sec
For modulating controller
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Data Sheet 20.2550
Page 6/10
Connection diagram
17
18
19
20
21
22
Rear view with
faston connectors
N
(L-)
L1
23 (L+)
TE
16
15
14
1
2
3
13
12
11
10
9
8
7
6
5
Wall-mounting housing
(extra code /640)
with terminal strip
4
Outputs
K
Terminal assignments
Relay 1
(K1)
Status indication
LED K1
1
23
22
Relay 2
(K2)
Status indication
LED K2
2
Relay 3
(K3)
Status indication
LED K3
3
Symbol
common
make (n.o.)
23
22
P
21
20
common
make (n.o.)
S
21
20
P
16
15
14
break (n.c.)
common
make (n.o.)
16
S
15
14
P
S
Ö
or
analog process value output
(electrically isolated)
15
14
–
+
Supply for 2-wire transmitter
3
15
14
+
-
Logic output 1
(K4)
Status indication
LED K4
4
19
17
–
+
14
15
+
-
14
15
+
Relay 4
(K5)
No status indication
5
3
2
1
break (n.c.)
common
make (n.o.)
17
19
+
-
3
2
1
P
analog process value output
(electrically isolated)
Supply for 2-wire transmitter
07.08 / 00381059
S
Ö
or
5
2
1
–
+
2
3
–
+
1
2
+
2
3
-
+
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Data Sheet 20.2550
Inputs
Terminal assignments
Input for standard signal
Ix (0(4)—20 mA)
7
8
Resistance thermometer
in 3-wire circuit
+
7
8
-
+
9
10
11
J
10
9
11
RxD
5
4
RxD +
RxD –
receive data
TxD
2
1
TxD +
TxD –
transmit data
GND
Serial interface RS485
(option)
+
GND
Serial interface
Profibus-DP
(option)
Symbol
9
10
11
Resistance thermometer
in 2-wire circuit
Serial interface RS422
(option)
Page 7/10
VP
3
GND
2
1
TxD/RxD +
TxD/RxD –
3
GND
4
supply voltage-plus (P5V)
5
receive data / transmit data
4
2
1
2
1
3
2
RxD/TxD-P 2
receive/transmit data-positive,
B conductor
RxD/TxD-N 1
receive/transmit data-negative,
A conductor
DGND
3
1
3
3
4
ground for data transmission
Logic input 1
13
19
13
19
Logic input 2
12
19
12
19
Supply voltage
see nameplate
07.08 / 00381059
AC/
DC
AC:
L1
N
TE
phase/line
neutral
technical earth
DC:
L+
L–
L1 N
L+ L- TE
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Dimensions
Panel cut-out to DIN 43 700
close mounting (minimum dimensions)
Type 202550/...
panel cut-out
to DIN 43 700
Option
Surface-mounting housing, extra code /640, Protection IP67
07.08 / 00381059
Data Sheet 20.2550
Page 8/10
JUMO GmbH & Co. KG• 36035 Fulda, Germany
Data Sheet 20.2550
Page 9/10
Type designation
(1)
202550
(2)
10
Basic type
*Generally
JUMO dTRANS Az 01
Microprocessor indicator / controller for analytical measurement
on all controllers of the
202550 series, the user can freely
select
the
following
configurations:
❏ Controller off
❏ Limit controller
❏ Pulse width controller with P,
PI, PD, PID control action
❏ Pulse frequency controller with
P, PI, PD, PID control action
❏ Modulating controller
Basic type extensions
Limit controller*
(3)
Inputs
660
0/4—20 mA front panel print: pH and mV, °C
661
0/4—20 mA front panel print: mV, °C
662
0/4—20 mA front panel print: mS/cm and µS/cm, °C
664
0/4—20 mA front panel print: none, °C
665
0/4—20 mA front panel print: mg/l, °C
(4)
000
no output
140
supply for 2-wire transmitter
310
relay, changeover contact
888
process value output, freely configurable
(5)
1
Output I
Output II
000
no output or interface
140
supply for 2-wire transmitter1
310
relay, changeover contact1
888
process value output, freely configurable1
(6)
Supply voltage
22
20 — 53 V AC/DC ±0%, 48 — 63/0 Hz
23
110 — 240 V AC +10%/-15%, 48 — 63 Hz
(7)
Interface
00
no serial interface
54
serial interface RS422/4851
64
serial interface Profibus-DP1
(8)
Extra codes
000
no extra codes
015
logic output 0/12 V DC,
instead of standard 0/5 V DC
640
surface-mounting housing for wall-mounting or DIN rail mounted, Protection IP67
(1)
(2)
(3)
Order code
202550
/
10
-
Order example
202550
/
10
-
660
(5)1
(4)
-
,
888
,
310
-
Optional accessories 1 (switchgear cabinet units)
Designation
Sales No.
Bracket for C-rail
70/00375749
Blank cover 96 x 48 mm
70/00069680
Optional accessories 2 (wall-mounting units)
Designation
Sales No.
Pole clamp, 60 mm dia. (clamping area: 50 - 70 mm dia.)
20/00437485
Pole clamp, 120 mm dia. (clamping area: 100 - 120 mm)
20/00437486
07.08 / 00381059
(7)1
(6)
23
-
(8)
/
00
/
000
If output II (4) = "310" “140” or
"888", then the interface option
(6) is not possible (or the other
way round)!
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Data Sheet 20.2550
Page 10/10
Application example
Indicator for pH
(1)Indicator / controller for pH, type 202550 with
integrated supply for a 2-wire transmitter
(2)2-wire transmitter, type 202701
(3)pH combination electrode, e.g. type 2GE-2G-U-1
(4)Compensation thermometer, type 2K-2
Advantages:
– long cable lengths can be covered, with low interference
– reduced wiring requirements, since supply for
2-wire transmitter is integrated in indicator
– calibration facility in indicator
– humidity problems reduced to a minimum during calibration
Indicator for free chlorine, chlorine dioxide or ozone
(1)Universal indicator / controller, type 202550 with
integrated supply for a 2-wire transmitter
(2)Measuring cell for free chlorine, type 202630
(3)Compensation thermometer, type 2K-2
07.08 / 00381059