Pocket Conductivity Meter
How to use…
Pocket Conductivity Meter
INSTRUCTION SHEET For LaMotte Tracer Conductivity/TDS/Salinity
Testing Location - Field
This test using the meter should be performed in the field.
Materials
LaMotte Tracer (Gray w/ black trim)
SR-44W button batteries
Small wide mouthed sample containers marked A and B
Conductivity 1413 calibration solution (clear)
Testing Background
If properly taken care of, the Conductivity meters we have provided can quickly give accurate
numerical results.
Calibration Instructions
-LaMotte Tracer Conductivity/ Total Dissolved Solids/Salinity (gray, black trim)
1. Take off the bottom cap covering the electrodes.
2. Place the meter in 20 mL of 1413 µS/cm standard calibration solution. Turn the meter on by
pressing the ON/OFF button. The meter must be in conductivity mode (“µS” will be displayed above
the reading; to change modes, press the MODE button until it switches modes).
3. Press and hold the CAL button for ~2 seconds. “CAL” will appear on the bottom of the screen and
1413 will flash on the screen.
4. The device will automatically recognize and calibrate to the conductivity standard. 1413 will stop
flashing and the display will briefly read “SA” and “End”. (“SA” will not appear if the calibration
fails.)
5. Rinse the meter with distilled water, shake dry, and turn the meter off. Proceed to testing
instructions on next page.
LaMotte Tracer Conductivity/ Total Dissolved Solids/Salinity (gray, black trim)
1. Pour water sample into a small, wide mouth container that the meter can fit into. You only need 1
inch of sample water.
2. While the meter is OFF, remove the protective cap from the bottom. Immerse the bottom of the
meter 0.5 to 1 inches in water sample, just enough that the metal electrodes at the bottom are
submerged.
3. Press the ON/OFF button to turn on the meter- SELF CAL will flash on the display.
4. Press and hold the MODE button until you see the TDS on the bottom of the display and ppm in the
top left corner. (There should not be an “S” above the reading- that is the salinity mode and is not
used).
5. Allow the reading to stabilize, then record the TDS measurement on the data sheet.
6. You should also record conductivity: Check to see that meter is in CONDUCTIVITY MODE (µS should
be displayed above the reading). To change modes, press and hold the MODE button until the
correct units appear.
7. Allow the reading to stabilize, then record the CONDUCTIVITY on the data sheet.
8. Rinse the meter with distilled water, shake dry & press ON/OFF button to turn off and replace cap.
Disposal and Clean Up
Calibration solution should be replaced after 4 sampling events/dates. Calibration solution can be
flushed down the sink with plenty of water. Properly dispose of expired batteries; you can give them to
Creek Connections.
Safety Precautions
Normal safety precautions should be taken when handling the water sample, buffer solution and the
equipment to avoid breakage.
Trouble Shooting Tips
METER DOES NOT TURN ON
1. Check batteries. Make sure they are installed properly (see Changing Batteries instructions below).
2. If still not working, install new batteries (see Changing Batteries instructions below).
3. If the meter does not work after replacing batteries, recheck the direction of the batteries in the
meter. Reversed batteries in meter will result in failure of the meter to work. If the new batteries
are in the correct position, check to see if the red or black wires connecting to the battery
compartment are broken. If they are broken, you need to replace your meter.
4. If the batteries are new and the meter is still not working, the battery contacts may be corroded or
dirty. Check the metal contacts that the batteries fit between and look for rust or other oxidation.
Clean them with isopropyl alcohol. You may need to carefully scrap away rust or dirt with a flathead
screwdriver or pair of scissors. Replace batteries and try again.
5. If the meter still does not turn on, you need to replace your meter.
THE NUMBERS ARE FAINT OR DISAPPEAR
1. If the readout numbers are faint or disappear while meter is on and while it is IN a solution or if the
meter’s low battery indicator symbol appears, then the batteries need to be replaced (see Changing
Batteries Section below).
METER WILL NOT CALIBRATE
1. Did you take the cap off the meter?
2. For both types of meters, if the digital reading does not stabilize when calibrating… rinse the metal
electrodes with distilled water before calibrating. Make sure the meter is in new calibration solution
(this should be replaced after a 3 or 4 creek trips). Make sure the meter is actually IN the calibration
solution when turning the correct calibration screw on the back; numbers will go crazy if the meter
is on while in the air and not in a solution. Before removing from any solution or sample, shut the
meter off. After addressing these, if the reading still does not stabilize at all, the batteries may need
to be replaced (see Battery Changing Section below).
3. If air bubbles are trapped near electrodes, tap the meter or stir sample to release the air bubbles.
NUMBERS ON METER WILL NOT STABILIZE IN A SAMPLE
1. Did you take the cap off the meter?
2. Did you use fresh calibration solution?
3. Just wait. It takes some time for the meter to stabilize. After a few minutes, if the readout number
does not stabilize (it is “jumping around” still), shut off the meter, remove from the sample, and
then rinse the meter’s metal electrodes thoroughly with distilled water and try again.
4. If readout still does not stabilize, the batteries may need to be replaced (see Changing Batteries
Section below).
5. If the batteries are new, installed correctly, the contacts and wiring are in good condition, and the
readout still does not stabilize, then the electrodes may need cleaned more thoroughly. Long term
care of the meter states that you should soak the stainless steel electrodes periodically in isopropyl
alcohol for 10-15 minutes. This will help remove any buildup on the electrodes. Do this, rinse
electrodes with water then try your sample again later.
METER STILL NOT WORKING
If you have done all of the above and the meter is still malfunctioning, contact Creek Connections for a
replacement meter.
CHANGING BATTERIES
1. Open the top battery compartment lid carefully on top of the meter. Remove old batteries (give
them to your teacher for proper disposal).
2. Replace the batteries with new ones noting polarity (+ and -) inside the meter and on the batteries.
The red wire is the positive (+) side; black wire, negative (-).
What batteries to use: Tracer uses 1 SR-44W button battery.
LONG TERM CARE TIPS
1.
Make sure the meter is turned off after each use. DO NOT have meter turned on unless it is in a
solution, having it on when the electrodes are in the air is not good for the meter.
2.
Rinse the electrodes thoroughly with distilled water after each use. It is okay to store the meter
with some moisture in the electrode cap.
3.
To improve meter performance and accuracy, periodically rinse/soak the metal electrodes in
isopropyl alcohol for 10-15 minutes to help clean them.
4.
Do not expose meters to excess moisture. Store in a dry place. Do not drop meter in the creek or
leave out in rain.
5.
Make sure students CAREFULLY calibrate the meter.
This instructions sheet was adapted from the instructions for the LaMotte Tracer as well as from Creek
Connections staff observations.
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising