Agilent Technologies Infiniium 54810/15/20/25/35/45A Oscilloscopes

Agilent Technologies Infiniium 54810/15/20/25/35/45A Oscilloscopes
Agilent Technologies
Infiniium 54810/15/20/25/35/45A Oscilloscopes
Data Sheet
• 500 MHz, 1 GHz and 1.5-GHz
bandwidth models
• 2- or 4-channel models
• Up to 8-GSa/s sample rates
• Simple, analog-like front panel
• Familiar Windows ®98-based
graphical user interface
• Built-in information system
• LAN printer and file sharing
• Eye diagrams and mask testing
• VoiceControl Option
Agilent Infiniium oscilloscopes
combine a simple, analog-like front
panel, the graphical user interface of
a PC, and powerful connectivity
capabilities to make high-performance
features accessible and uncomplicated.
Infiniium oscilloscopes are designed to
help you get your measurement job
done faster.
Simple Analog-like Front Panel
Agilent Infiniium oscilloscopes give
you the simple, uncluttered front panel
of an analog scope for accessing basic
functions. Dedicated scale and
position knobs for each channel
provide intuitive operation. Trigger
LEDs show you trigger status at a
glance. To speed up measurements,
the QuickMeas key gives you instant
access to any four measurements.
Familiar Graphical User Interface
Agilent Infiniium scopes employ a
graphical user interface based on
Windows® 98. Windows’ familiar
interface means you won't have to
spend a lot of time learning and
relearning the scope. Pull-down
menus give you easy access to
advanced features. Dialog boxes
display all the choices you need to
make for measurement setups.
In addition, waveforms, markers, and
measurements can be easily positioned
on the display by using your mouse to
drag and drop.
Exceptional Connectivity
Agilent Infiniium scopes give you the
tools you need to document your work
and share your results. It takes just a
few mouse clicks to save waveforms
and screen bitmaps to hard drive,
floppy, or, when networked, to a drive
on your network. LAN connectivity
also lets you easily print your results
to networked printers, and makes it
simple to share your waveforms and
data with others.
The Performance You Need
And Agilent Infiniium scopes aren’t just
easier to use. These scopes provide the
high performance specifications and
features you need to get your
measurement job done faster.
With bandwidths to 1.5 GHz and
sample rates to 8 GSa/s, Infiniium can
view your fastest signals. Powerful
triggering and analysis makes it easy to
capture and analyze your waveforms.
Display and math features like colorgraded persistence and histograms let
you easily view and isolate signal
problems faster.
Performance to Get the Job Done
Agilent Infiniium scopes
combine ease-of-use with the
right specifications and a broad
feature set to help you get your
job done faster. Use the
information here to find the
scope that meets your signal
measurement needs, and
review the features on the next
two pages to see how Infiniium
makes its advanced power so
usable.
General-Purpose Versatility
The 500 MHz versions of Infiniium
provide exceptional performance and
capability for general purpose lab
measurements. Select a sample rate of
1 or 2 GSa/s based on your need for
single shot measurement accuracy, and
let your application needs and budget
determine your choice of a 2- or 4channel model. Whichever Infiniium
you select, you can be sure of getting a
powerful scope that makes it easier to
get your job done.
Ultimate Performance
The Infiniium family's 54845A provides
the performance you need for your
most demanding signals. 1.5 GHz
bandwidth and a sample rate of up to
8 GSa/s ensure fast, accurate capture
of your waveforms. Like the other
Infiniium models, the 54845A’s familiar
user interface makes it easy to take
advantage of every bit of the
performance inside.
For applications requiring more than
500 MHz of bandwidth, the new 54835A
offers four channels at 1 GHz, with
sample rates up to 4 GSa/s.
Model
54810A
54815A
54820A
54825A
54835A
Channels
2
4
2
4
4
Bandwidth
500 MHz
500 MHz
500 MHz
500 MHz
1 GHz
54845A
4
1.5 GHz
Sample Rate
1 GSa/s
1 GSa/s
2 GSa/s
2 GSa/s
4 GSa/s (2 channel mode)
2 GSa/s (4 channel mode)
8 GSa/s (2 channel mode)
4 GSa/s (4 channel mode)
Memory Depth
32K
32K
32K
32K
64K (2 channel mode)
32K (4 channel mode)
64K (2 channel mode)
32K (4 channel mode)
Infiniium models 54815A and 54825A
Infiniium models 54810A and 54820A
Infiniium models 54835A and 54845A
2
Features to Get the Job Done Faster
Drag and Drop Measurements
Just drag and drop an icon from the
measurement toolbar onto the portion
of the waveform you want to measure
and the measurement appears instantly
on screen.
Easy Zooming
Infiniium's graphical user interface
gives you a new, easier way to zoom.
Use the mouse to draw a box around
the section of the waveform that you
want to expand, then click inside the
box to zoom in on the area of interest.
You can also undo the zoom (up to 10
levels of undo available).
Built-in Information System
Infiniium's built-in information system
puts measurement assistance at your
fingertips in seven different languages
(English, French, German, Japanese,
Korean, Traditional Chinese and
Simplified Chinese). You'll no longer
have to look for the manual when you
need help setting up scope functions
or making complex measurements.
Setup Guides give you step-by-step
instructions for 24 different
measurements and procedures.
In addition, you'll find a thorough index
of help topics and context-sensitive
help available from the dialog boxes.
VoiceControl Option
If you’ve found that you don’t have
enough hands to probe fine pitch ICs
and control the front panel of your
scope at the same, Infiniium now offers
a solution. The VoiceControl option
allows you to speak natural English
language commands through the
supplied collar-mounted microphone,
allowing hands-free operation of the
front panel. VoiceControl is speaker
and gender independent and does not
require the scope to be "trained" to
recognize a particular user. For
complete information, visit our website
at www.agilent.com/find/infiniium or
order the Infiniium VoiceControl
datasheet (5968-6659).
Advanced Triggering
Agilent Infiniium scopes include
powerful Violation Triggering
technology, making it easier for you to
capture and view your toughest signals.
In addition to standard capabilities
such as edge, glitch, logic, state, and
video trigger, Violation Triggering adds
the ability to trigger on rise time, fall
time, setup and hold time, and runt
pulses. Runt pulse trigger is standard
3
on models 54835A/54845A, and
optional (opt. 015) on models
54810A/54815A/54820A/54825A.
Waveform Math with FFTs and Histograms
Agilent Infiniium scopes provide a wide
variety of math functions, including
Fast Fourier Transforms (FFTs) and
histograms. FFTs allow quick
identification of frequency components
within your signals, while histograms
provide a quantifiable way to analyze
signal variations over time. Both help
you better understand noise, jitter, and
other sources of problems within your
design.
Color-grade Persistence
The color-grade persistence display
mode provides a visual representation
of waveform behavior over time,
making it easy to locate and identify
signal anomalies. And you don’t lose
debugging power: all scope capabilities
are available in this mode.
Display Annotation
Add labels to the display to identify
waveforms, annotate setups, or provide
operator instructions. Infiniium’s
keyboard makes this an easy task.
Eye Diagram Measurements
Agilent Infiniium scopes provide eye
diagram measurements to make
general testing of communications
designs and systems even easier.
AutoMask and Mask Test Capability
AutoMask allows you to define a
tolerance limit and automatically create
a test envelope based on the waveform
on screen. Mask testing then provides
a pass/fail comparison of your
incoming signal to the test envelope.
Communication Mask Test Kit Option
Easily test your product’s conformance
to ANSI T1.102 and ITU G.703 standards
with Agilent Infiniium oscilloscopes’
Communication Mask Option. This
option provides a library of industry
standard waveform masks, and eases
the measurement task with pushbutton
alignment of the appropriate mask to
your signal and clear indication of
failure points on the waveform. A set of
“smart” termination adapters works
with the probe interface to assure errorfree measurements.
Internal Hard and Floppy Disk Drives
Saving your results is simple with
Infiniium’s data storage capabilities.
Use the internal 3.2 Gbyte hard drive or
the 120 MB LS-120 SuperDisk floppy
drive to store waveform data, screen
images, or scope setups. It’s easy to
share your information with others, as
well. Waveform data can be stored in
ASCII or Microsoft Excel formats, while
images can be stored as BMP, PCX, TIF,
GIF, EPS, or PS formats. Both allow
easy import into other programs for
documentation or further analysis.
Standard Interfaces
10/100 MBPS LAN, GPIB, RS-232, and
Centronics interfaces are standard
features4. LAN allows you to transfer
files or use network printers with the
same ease as when you do these tasks
on your PC. Infiniium scopes are fully
HP-IB programmable.
Printer Support
Agilent Infiniium oscilloscopes offer
full support of HP DeskJet and LaserJet
printers. Visit our website at
http://www.hp.com/info/infiniium21
for a list of tested printers.
1152A 2.5-GHz, 0.6-pF Active Probe
Use the 1152A active probe with
Infiniium model 54845A for the most
non-intrusive, faithful reproduction of
signals. Even without grounding, this
probe accurately reproduces highfrequency edges. Agilent's
breakthrough technology assures
minimum loading on your circuit.
High-Bandwidth Differential Probes
These high-bandwidth probes are
essential for viewing the high speed,
low level differential signals prevalent
in today’s designs. Both probes have
1:1 gain and additional attenuators for
larger signals. The 1154A also features
10X gain for mV measurements.
Simple Surface-Mount Probing
The Agilent Wedge Probe Adapter
provides accurate, mechanically noninvasive contact to 0.5 mm TQFP/PQFP
package leads. Infiniium's probes
include an adapter that connects to the
Wedge to make surface-mount probing
hassle-free.
Infiniium Accessories
A complete line of accessories is
available for Agilent Infiniium
oscilloscopes, including alternative
pointing devices, special probing
solutions, and more. For complete
information, visit our website at
http://www.agilent.com/find/infiniium,
or see the Agilent Infiniium Probes and
Accessories Product Overview
(5966-3543).
Performance Characteristics
* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period, and ±10°C (models 54810A/15A/20A/25A) or
±5°C (models 54835A/45A) from firmware calibration temperature.
Acquisition
Maximum Sample Rate
Real Time
Maximum Effective Sample Rate
Equivalent Time
Memory Depth
Memory Depth Modes
Auto:
Manual:
Sampling Modes
Real Time:
Equivalent Time:
Peak detect:
Filters:
9-bit Bandwidth Limit Filter:
(Sinx)/x Interpolation:
Averaging
Vertical
Number of Channels
Analog Bandwidth (-3dB)*
System Bandwidth with
1160A 10:1 passive probe
1161A 10:1 passive probe
1162A 1:1 passive probe
1163A 10:1, 500 Ω passive probe
1152A 2.5 GHz, .6 pF active probe
1153A 200 MHz differential probe
Real Time Bandwidth*
Rise Time1
Sensitivity2
1 MΩ :
50 Ω:
Expand about:
Input Impedance*
VSWR (50 Ω)
Input Coupling
Maximum Input Voltage*
1 MΩ:
50 Ω:
Hardware Bandwidth Limit (-3 dB)
Channel-to-channel isolation
(with channels at equal sensitivity)
54810A, 54815A, 54820A and 54825A
54835A
54845A
54810A/15A: 1 GSa/s on each channel
54820A/25A: 2 GSa/s on each channel
2 channel mode: 4 GSa/s
4 channel mode: 2 GSa/s
2 channel mode: 8 GSa/s
4 channel mode: 4 GSa/s
100 GSa/s
32,768 points/channel
500 GSa/s
2 channel mode: 65,536 points
4 channel mode: 32,768 points
Optimized for best combination of update rate and display quality.
Selectable from 16 to 32,768 points
Selectable
2 channel mode: from 16 to 65,536 points
4 channel mode: from 16 to 32,768 points
Successive single shot acquisitions.
Random Repetitive sampling (higher time resolution at faster sweep speeds)
Captures and displays narrow pulses or glitches n/a
1 ns or wider at sample rates of 250 MSa/s or less
BW = Sample Rate/20
On/Off selectable FIR digital filter. Digital signal processing adds points between acquired data
points to enhance measurement accuracy and waveform display quality. BW= Sample Rate/4
Selectable from 2 to 4096
54810A/20A: 2 (simultaneous acquisition)
54815A/25A: 4 (simultaneous acquisition)
500 MHz
4 (simultaneous acquisition)
500 MHz
–
25 MHz
500 MHz
500 MHz
200 MHz
54810A/15A: 250 MHz
54820A/25A: 500 MHz
54810A/15A: 700 ps (Equivalent Time)
1.4 ns (Real Time)
54820A/25A: 700 ps
–
500 MHz
25 MHz
1.0 GHz
1.0 GHz
200 MHz
50 Ω: 1.0 GHz (2 channel mode)
500 MHz (4 channel mode)
50 Ω: 350 ps
1M Ω: 700 ps
50 Ω: 1.0 GHz
50 Ω: 1.5 GHz
1MΩ: 500 MHz (with 1161A probe) 1 MΩ: 500 MHz (with 1161A probe)
1.5 GHz
1.3 GHz
50 Ω: 1.5 GHz (2 channel mode)
1.0 GHz (4 channel mode)
50 Ω: 233 ps
1 MΩ: 700 ps
1 mV/div to 5 V/div
2 mV/div to 2 V/div
1 mV/div to 5 V/div
2 mV/div to 1 V/div
Choose between expanding about center screen or ground. This applies to channels, memories or math functions.
1 MΩ ± 1% (≈8 pF), or 50 Ω ± 1%
1 MΩ ± 1% (≈12 pF), or 50 Ω ± 1.5%
n/a
dc to 500 MHz:
1.3
dc to 500 MHz:
1.3
500 MHz to 1 GHz: 1.5
500 MHz to 1 GHz: 1.5
1 GHz to 1.5 GHz:
1.75
dc, ac (7 Hz, available in 1 MΩ only)
dc, ac (7 Hz, available in 1 MΩ only)
± 250 V (dc + ac) [ac<10 kHz], CAT I
5 Vrms, CAT I
On/Off selectable, 30 MHz
dc to 50 MHz: 50 dB
50 MHz to 500 MHz: 40 dB
± 100 V (dc + ac) [ac<10 kHz], CAT I
5 Vrms, CAT I
n/a
dc to 100 MHz: 40 dB
100 MHz to 1 GHz: 30 dB
1 GHz to 1.5 GHz: 25 dB
4
Performance Characteristics
* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period, and ±10°C (models 54810A/15A/20A/25A) or
±5°C (models 54835A/45A) from firmware calibration temperature.
Offset Range
Dynamic Range
Full resolution channel scales
dc Gain Accuracy*2,3
Resolution2
Real Time
Equivalent Time
Offset Accuracy*2
dc Voltage Measurement Accuracy*2,3
Dual Cursor
Single Cursor
Example (Single cursor accuracy,
70 mV signal,10 mV/div, 0 offset)
AutoProbe Interface
Horizontal
Main Time Base Range
Horizontal Position Range (pre trigger)
Horizontal Position Range (post trigger)
Delayed Sweep Range
Delayed Sweep Delay Range
Resolution
Timebase Accuracy
Delta-t Accuracy*
Real Time mode4
Equivalent Time mode (≥ 16 avgs.)
Peak Detect mode
Example (Equivalent Time mode (≥ 16
avgs.), 9 ns signal, 1 ns/div, 1 channel)
Jitter, RMS
Trigger
Sensitivity*2
Internal (normal) dc to 100 MHz:
100 MHz to 500 MHz:
500 MHz to 1 GHz:
Internal (noise reject) dc to 100 MHz:
100 MHz to 500 MHz:
External (54810/20A) dc to 100 MHz:
100 MHz to 500 MHz:
Auxiliary (54815/25/35/45A)
dc to 500 MHz:
Maximum Input Voltage*
External (54810/20A)
Auxiliary (54815/25/45A)
Minimum Pulse Width (internal, external)
Level Range
Internal
External (54810/20A)
Auxiliary (54815/25/45A)
5
54810A, 54815A, 54820A and 54825A
Vertical Sensitivity
Available Offset
1 mV to 49 mV/div
±2V
> 50 mV to 249 mV/div
± 10 V
> 250 mV to 1.24 V/div
± 50 V
1.25 V to 5V/div
± 250 V
± 12 div from center screen
All volts/division settings >= 7 mV/div
54835A and 54845A
Vertical Sensitivity
1MΩ: 2 mV/div to 100 mV/div
>100 mV/div to 2 V/div
50 Ω: all
Available Offset
±4V
± 40 V
> ± 12 div
±8 div from center screen
10, 20, 50, 100, 200, 500, 1000 mV/div
(plus 2000 mV/div in 1 MΩ)
± 1.25% of full scale at full resolution channel scale ± 1% of full scale at full resolution channel scale
8 bits (0.4% of full scale), 12 bits or greater with sufficient averaging
8 bits (0.4% of full scale), 12 bits or greater with sufficient averaging
± (1.25% of channel offset + 2% of full scale) ± (1% of channel offset + 1% of full scale)
at full resolution channel scale
at full resolution channel scale
±[(dc gain accuracy)+(resolution)]
±[(dc gain accuracy)+(offset accuracy)+(resolution/2)]
Accuracy = ±[1.25%(80mV) + (1.25%(0)+2%(80 mV)) + Accuracy = ±[1%(80mV) + (1%(0 V)+1%(80 mV)) +
(.4%/2)(80 mV)] = ±2.8 mV
(.4%/2)(80 mV)] = ±1.76 mV
AutoProbe is an intelligent communication and power link between compatible probes and Infiniium scopes.
AutoProbe completely configures the scope for the attached probe. For instance, it identifies the probe type
and sets up the proper input impedance, attenuation ratio, probe power and offset range, as needed.
500 ps/div to 20 s/div
100 ps/div to 20 s/div
0 to -1 s or one full screen width, whichever is larger
0 to 1 s or one full screen width, whichever is larger
1 ps/div to current main time base setting 1 ps/div to current main time base setting
Within main time base acquisition record
10 ps
2 ps
50 ppm (.005%)
70 ppm (.007%)
±[(.005%)(delta-t) + (0.2)(sample period)]
±[(.005%)(delta-t) + (full scale/(2*memory depth)) + 60ps]
±[(.005%)(delta-t)+(1 sample period)]
Accuracy = ±[(.005%)(9 ns)+((10 ns)/(2*32,768))+ 60ps] =
±[(450 x 10-15)+(152 x 10-15)+(60 x 10-12] = 61 ps
n/a
±[(.007%)(delta-t)+(0.2)(sample period)]
±[(.007%)(delta-t)+(full scale/(2*memory depth))+ 30ps]
n/a
Accuracy = ±[(.007%)(9 ns)+((10 ns)/(2*65,536))+ 30ps] =
±[(630 x 10-15)+(76 x 10-15)+(30 x 10-12] = 31 ps
8 ps ± .005% (delay setting)
0.5 div
1.0 div
N/A
1.0 div
1.5 div
0.0225 x (signal range)
0.045 x (signal range)
0.5 div
1.0 div
1.5 div
N/A
300 mVp-p
1 MΩ: ±250 V (dc + ac) [ac<10 kHz],
CAT I, 50 Ω: 5 Vrms, CAT I
2.5 kΩ: ±15 V, CAT I
1 ns at > 1.0 div
300 mVp-p
n/a
2.5 kΩ: ±15 V, CAT I
500 ps at > 1.0 div
± 12 div from center screen
± 1 V, ± 5 V, ± 25 V
±5V
± 8 div from center screen
n/a
±5V
N/A
Performance Characteristics
* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period, and ±10°C (models 54810A/15A/20A/25A) or
±5°C (models 54835A/45A) from firmware calibration temperature.
Trigger (continued)
Sweep Modes
Trigger Coupling
Trigger Holdoff Range
Trigger Modes
Glitch
Pattern
State
Delay by Time
Delay by Events
Violation Trigger
Runt
Setup/Hold
Pulse Width
Transition
Accuracy (time) for glitch, pulse
width and time-qualified pattern
Video Triggering
Display
Display
Annotation
Active Display Area
Waveform Viewing Area
Display Resolution
Persistence
Color Graded Infinite Persistence
Display Update Rate
Measurement Conditions
Waveforms/sec
Vpp Measurements/sec
Measurements
Automatic Parametrics
Threshold Definition
Top-Base Definition
Statistics
Histograms
Eye Diagram Measurements
54810A, 54815A, 54820A and 54825A
54835A and 54845A
Auto, triggered, single
dc, ac(7 Hz), low frequency reject (50 kHz), high frequency reject (50 kHz)
60 ns - 320 ms
Edge, Glitch, Pattern, State, Delay by Time, Delay by Events, Violation (Runt, Setup/Hold Time,
Pulse width, Transition), Video, Line
Select positive or negative polarity, width. Captures glitches as narrow as 500 ps.
Select inputs as High, Low or X (don't care) to create pattern. Trigger when pattern is entered, exited,
present > t, present < t, or present over a range of time. Captures patterns as narrow as 500 ps.
Select one channel as clock, specify other inputs as High, Low or X. Logic Type: AND or NAND.
Setup time is 1 ns and hold time is 0 ns.
Time: 30 ns to 160 ms. The trigger is qualified by an edge. After the delay, a rising/falling edge on any
one selected input will generate the trigger.
Events: 1 to 16,000,000 rising or falling edges. The trigger is qualified by an edge.
After the delay, a rising/falling edge on any one selected input will generate the trigger.
Select Positive, Negative or either polarity, thresholds, time qualified. Captures runts as narrow as 500 ps.
Standard feature on models 54835A/54845A. Option 015 on models 54810A/54815A/54820A/54825A.
Modes: Setup, Hold or Setup and Hold. Select Clock, thresholds, setup and/or hold time. Edges must
be ≥ 1.5 ns. The difference in slew rates between clock and data must be less than a 2:1 ratio.
Triggers on pulse width > t, or < t. Captures pulses as narrow as 500 ps.
Select Rise Time or Fall Time, present > t or present < t, thresholds. Catures edges as fast as 800 ps.
1.5 ns - 20 ns: ±(20% setting + 500 ps)
20 ns - 160 ms: ±(3% setting + 2 ns)
525 lines/60 Hz (NTSC), 625 lines/50 Hz (PAL), 875 lines/60 Hz. Trigger on Field 1 or Field 2, any line.
User defined triggering: User can specify sync pulse level, width and polarity, edge number.
8.4 inch diagonal color active matrix LCD module incorporating amorphous silicon TFTs.
Up to 12 labels, with up to 100 characters each, can be inserted in the waveform display area.
Labels can be saved to and recalled from image files and setup files.
6.73"x 5.04" (33.92 sq. in.) 171 mm x 128 mm (21,888 sq. mm)
4.10"x 6.25" (25.6 sq. in.) 104 mm x 159 mm (16,536 sq. mm) in Full screen mode
640 pixels horizontally x 480 pixels vertically
Minimum, Variable (Up to 6 levels of gray scale, 100 ms to 40 s), Infinite, Color Graded
Previous sweeps are stored in differing colors based on the number of times each pixel is illuminated.
User can turn on or off, and can define colors to be used. This allows easy visual recognition of
waveform behavior over time.
Real Time mode, minimum persistence display mode, triggered sweep mode, no interpolation,
markers off, math off, connect the dots off, 1 channel acquisition, 50 ns/div, statistics on
>2,100
>130
27 automatic measurements: Vpp, Vmin, Vmax, Vavg, Vamptd, Vbase, Vtop, Vrms, Preshoot, Overshoot,
Vupper, Vmiddle, Vlower, Rise Time, Fall Time, Period, Frequency, Positive Width, Negative Width,
Duty Cycle, Delta Time, Tmax, Tmin, FFT Frequency, FFT Magnitude, FFT Delta Frequency,
FFT Delta Magnitude. Over HP-IB only: VTime, TVolts
Selectable 10%, 20%, 50%, 80%, 90% or Custom (in % or absolute voltage) Independent thresholds
can be set for each channel.
Standard or Custom (in absolute voltage)
On/Off selectable. Current measurement, mean and standard deviation
Use vertical and horizontal waveform markers to define region to analyze. Can use in vertical (for timing and
jitter measurements) or horizontal (noise and amplitude change) modes. Measurements included: mean,
standard deviation, peak-to-peak value, median, total hits, peak (area of most hits), and mean +/ 1,2, and 3 sigma.
Eye diagram display mode allows triggering on both the negative-going and positive going edges of a
signal. Eye diagram measurements include eye height, eye width, jitter, crossing percentage, Q factor,
and duty cycle distortion.
6
Performance Characteristics
* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period, and ±10°C (models 54810A/15A/20A/25A) or
±5°C (models 54835A/45A) from firmware calibration temperature.
Measurements (continued)
Mask Testing
Communications Mask Option
Measurement Toolbar
QuickMeas+
54810A, 54815A, 54820A and 54825A
54835A and 54845A
Allows pass/fail testing to user-defined or Agilent-supplied waveform templates. Up to 8 regions of polygons
allowed. Mask inversion inverts mask around baseline. Test modes include test forever, test to specified time or
event limit, and stop on failure. Auto-mask allows user to define tolerance range in time/voltage or percentage
and create a mask template from a captured waveform. Masks can be automatically aligned with the incoming
signal. All masks can be manually scaled via controls or the mouse.
Option 100 provides a set of ITU G.703 and ANSI T1.102 industry standard masks for compliance testing.
For a complete list of the masks provided, request Agilent Communications Mask Testing Option 100
technical data sheet (see page 8)
16 Drag and Drop automatic measurement icons.
Allows user to configure a button below the display for the following functions:
QuickMeas: activates 4 preselected automatic measurements. QuickPrint: prints out screen.
QuickScreen: saves screen image to file; multiple button pushes increments filename suffix (e.g. Data01, Data02, etc.).
Markers Modes
Waveform Math
FFT
Frequency Range5
Frequency Accuracy
Amplitude Display
Signal-to-noise ratio
Window Modes
Computer System/Storage
CPU
Disk Drives
File types: Waveforms
Images
Mouse
Waveform Memories
I/O
LAN
HP-IB
RS-232 (serial)
Centronics
Printers
PS/2 port
Keyboard port
Video Output
Auxiliary Output
General Characteristics
Temperature
Operating:
Nonoperating:
Humidity
Operating:
Nonoperating:
7
QuickWaveform: saves waveform data to a file on the disk drive; multiple button pushes increments filename suffix.
QuickSetup: loads in a specific instrument setup file
Manual Markers, Track Waveform Data, Track Measurements
4 functions f1-f4. Select from Add, Subtract, Multiply, Divide, Invert, Magnify, Vs, Min, Max, Integrate,
Differentiate, FFT Magnitude. Any of the Waveform Math functions (e.g., FFT) can be averaged.
54820A/54825A: dc to 1 GHz (Sample rate/2)
54810A/54815A: dc to 500 MHz (Sample rate/2)
54835A: 2 channel mode – dc to 2 GHz
4 channel mode – dc to 1 GHz
54845A: 2 channel mode – dc to 4 GHz
4 channel mode – dc to 2 GHz
(1/2 frequency resolution)+(5x10-5)(signal frequency) (1/2 frequency resolution)+(7x10-5)(signal frequency)
Power in dBm
70 dB at 32K memory depth. Noise floor varies with memory depth and with averaging.
Hanning, Flattop, Rectangular
AMD K6-2™ 400 MHz microprocessor
3.2 GByte internal hard drive. Storage capacity is limited only by disk space. 3.5" 120 MB LS-120 SuperDisk
floppy drive (reads/writes to std. 3.5” floppies). Store and recall setups, waveforms, and store screen
images to both the hard drive and the floppy drive.
Internal Y values; X and Y values in ASCII or Microsoft Excel formats
BMP, PCX, TIF, GIF, EPS or PS formats
Standard mouse supplied. Supports any Microsoft® mouse-compatible pointing device, serial or PS/2
4 nonvolatile waveform memories
Enables data/setup file transfers and use of network printers; supports Microsoft Networking;
10/100 MBPS operation that complies with IEEE 802.3 Ethernet and ISO/IEC 8802-3 Ethernet standards;
TCP/IP protocol and RJ-45 connector.
Fully programmable, complies with IEEE 488.2
2 ports: COM1, COM2. Printer and pointing device support
Printer support
Full support of HP DeskJet and LaserJet printers. Visit the Infiniium website at
http://www.agilent.com/find/infiniium
For PS/2 mouse
Standard keyboard for file naming and waveform annotation, scope control
15 pin VGA, full color
DC (±2.5v); square wave (715 Hz [±5%], 350 mVpp [±5%]); trigger output (350 mVpp [±5%], frequency
varies with occurance of trigger)
0°C to +50°C
-40°C to 70°C
10°C to +40°C
Up to 95% relative humidity (noncondensing) at +40°C
Up to 90% relative humidity at +65°C
Ordering Information
General Characteristics
Altitude
Operating:
Nonoperating:
Vibration
Operating:
Nonoperating:
Power
Weight
Dimensions (excluding handle)
Safety
Notes
www.agilent.com
54810A, 54815A, 54820A and 54825A
54835A and 54845A
Up to 4,600 meters (15,000 ft.)
Up to 15,300 meters (50,000 ft.)
Random vibration 5-500 Hz, 10 minutes per axis, 0.3g(rms)
Random vibration 5-500 Hz, 10 minutes per axis, 2.41g(rms);
resonant search 5-500 Hz, swept sine, 1 octave/minute sweep
rate, (0.75g), 5 minute resonant dwell at 4 resonances per axis.
Voltage: 100-240VAC, +/- 10%, Cat II, 47 to 440 Hz; Max power
dissipation: 390 W
Net: 10.6 kg (23.4 lb),
Net: 12.0 kg (26.5 lb),
Shipping: 13.6 kg (30 lb)
Shipping: 15 kg (33 lb)
Height: 216 mm ( 8.50 in); Width: 437 mm (17.19 in); Depth: 440 mm (17.34 in)
Meets IEC1010-1 +A2, CSA certified to C22.2 No.1010.1, Self certified to UL 3111
Rise time figures are calculated from tr = 0.35/bandwidth
2
54810A/15A/20A/25A: Magnification is used below 7 mV/div range. Below 7 mV/div, full scale is defined as 56 mV. 54845A: Magnification is
used below 10 mV/div range and between major attenuation settings. Full scale is defined as the major attenuator setting above an
intermediate setting. (Major settings 50 Ω: 10 mV, 20 mV, 50 mV, 100 mV, 200 mV, 500 mV, 1 V, 1 MΩ: all of the above plus 2 V)
3
54810A/15A/20A/25A: The dc gain accuracy decreases 0.08% of full scale per degree C from the firmware calibration temperature.
4
For bandwidth limited signals, tr >=1.4 x sample interval
5
FFT amplitude readings are affected by input amplifier roll-off; 54810A/15A/20A/25A: (-3 dB at 500 MHz, with amplitude decreasing as
frequency increases above 500 MHz). 54845A: (-3 dB at 1.5 GHz, with amplitude decreasing as frequency increases above 1.5 GHz).
1
Ordering Information
54810A Infiniium Oscilloscope
54815A Infiniium Oscilloscope
54820A Infiniium Oscilloscope
54825A Infiniium Oscilloscope
54835A Infiniium Oscilloscope
54845A Infiniium Oscilloscope
All of the above models include:
1 Mouse (C3751-60201), 1 Infiniium Mouse Pad (54810-85901), 1 Keyboard (E2610A),
1 User's Quick Start Guide, English language* (54810-97000), 1 Service Guide,
1 Programmer’s Guide, 1 Programmer’s Quick Reference Guide,
Information System in English, French, German, Japanese, Korean and Chinese (for Taiwan),
2 1160A 10:1 10 MΩ passive probes (54810A, 54820A), 4 1160A 10:1 10 MΩ
passive probes (54815A, 54825A), 4 1161A 10:1 10 MΩ passive probes (54635A, 54845A),
1 Accessory Pouch (54810-68701), 1 US power cord, Three-year warranty
* Other languages also available
Options
090 Delete standard probes
2 1160A probes for the 54810A/20A
4 1160A probes for the 54815A/25A
4 1161A probes for the 54835/45A
001* Additional standard probes
2 1160A probes for the 54810A/15A/20A/25A
2 1161A probes for the 54835/45A
002* Add 1 1162A 1:1 passive probe
003* Add 1 1163A 10:1 500 Ω, low C passive probe
006* Add 1 1152A 2.5 GHz, .6 pF active probe (for 54835/45A)
008* Add 1 1153A 200 MHz differential probe
009 Add 1 1154A 500 MHz differential probe
010 Add 1 1159A 1 GHz differential probe
015 Extended (runt) triggering (standard on 54835A/54845A)
100* Communications Mask Test Kit
includes mask library disk and set of termination adapters
200 VoiceControl option
1BP Mil Std 45662A and ANSI/NCSL Z-540 calibration with test data
1CL* Add 1 Keyboard (E2610A)
1CM*Add 1 Rackmount kit (E2609A)
UL5* Add 1 Touchpad pointing device (E2612A)
UL6* Add 1 Clip-on track ball pointing device (E2611A)
W50 Five-year, customer-return, repair coverage (additional 2 years)
* Multiple options can be ordered
For more information about Agilent
Technologies test and measurement
products, applications, services, and for a
current sales office listing, visit our web
site:
http://www.agilent.com/find/infiniium
You can also contact one of the following
centers and ask for a test and
measurement sales representative.
United States:
Agilent Technologies
Test and Measurement Call Center
P.O. Box 4026
Englewood, CO 80155-4026
(tel) 1 800 452 4844
Canada:
Agilent Technologies Canada Inc.
5150 Spectrum Way
Mississauga, Ontario
L4W 5G1
(tel) 1 877 894 4414
Europe:
Agilent Technologies
Test & Measurement
European Marketing Organisation
P.O. Box 999
1180 AZ Amstelveen
The Netherlands
(tel) (31 20) 547 9999
Japan:
Agilent Technologies Japan Ltd.
Measurement Assistance Center
9-1, Takakura-Cho, Hachioji-Shi,
Tokyo 192-8510, Japan
(tel) (81) 426 56 7832
(fax) (81) 426 56 7840
Latin America:
Agilent Technologies
Latin American Region Headquarters
5200 Blue Lagoon Drive, Suite #950
Miami, Florida 33126 U.S.A.
(tel) (305) 267 4245
(fax) (305) 267 4286
Australia/New Zealand:
Agilent Technologies Australia Pty Ltd
347 Burwood Highway
Forest Hill, Victoria 3131
(tel) 1-800 629 485 (Australia)
(fax) (61 3) 9272 0749
(tel) 0 800 738 378 (New Zealand)
(fax) (64 4) 802 6881
Asia Pacific:
Agilent Technologies
19/F, Cityplaza One, 1111 King’s Road,
Taikoo Shing, Hong Kong, SAR
(tel) (852) 2599 7889
(fax) (852) 2506 9233
Technical data is subject to change
Windows and MS-DOS are registered
trademarks of Microsoft Corporation.
© Agilent Technologies 2000
Printed in the U.S.A. 2/00
5968-8496EN