541A Walking Test AdapterSales - 1126.cdr
Walking Test Adapter Accessory for
Trek Model 541A Electrostatic Voltmeters
Walking Test Adapter Kit
The walking test adapter kit [CN 1K037] for the Model 541A series allow analysis of charge levels
accumulated on the human body per compliance with ANSI/ESD STM 97.2. and IEC Standard
61340-4-5.
The electrostatic charge and the ability for charge accumulation of a person walking on an ESD safe floor
with ESD safe footwear can be measured with a walking test adapter. This test, in addition to data from
standard resistance measurements, provide measurement data on the actual charge on a person while
walking. This test ensures ESD safe practices are being implemented properly and that charge is being
maintained under the maximum allowed voltages (typically less than 100V as per ANSI/ESD S20.202007, IEC 61340-5-1:2007, and IEC 61340-5-2:2007)
Model 541A with Probe
+100
Walking Test Case and Probe Holder Cap
Computer software captures real time data or saves it
for later analysis. Trek Model 541A electrostatic
voltmeters are microprocessor based electrostatic
voltmeters and utilize a miniature electrostatic field
chopper probe to provide highly accurate,
noncontacting, DC stable, spacing independent,
voltage measurements in either ionized or nonionized environments.
-100
A twenty by four (20 x 4) alphanumeric LCD screen
displays the present measured voltage, the positive
peak voltage value, the negative peak voltage value,
and additional menu information. Two specific
voltage detection ranges are available. The Model 541A-1 has a voltage measurement range of 0 to ±1
kV. The Model 541A-2 has a measurement range of 0 to ±100 V. Both instruments have exceptional 1%
accuracy and 0.1% resolution.
The Model 541A series includes independently programmable plus (+) and minus (-) voltage
threshold/alarm limits to provide audible, visual, and external warning signals. Audible and visible
alarms indicate the detection of programmed surface voltage limits. A relay contact output changes
state when alarm thresholds are met or exceeded. This signal can be connected to a process controlling
device or other equipment to monitor a particular environment. A voltage output monitor and the 4-20
mA current loop output can provide additional signal interfacing to facility monitoring equipment. An RS232 serial port can provide computer based control and monitoring.
TREK, INC. • 190 Walnut Street • Lockport, NY 14094 • 800-FOR TREK
716-438-7555 • 716-201-1804 (fax) • www.trekinc.com • sales@ trekinc.com
Measurement and Power SolutionsTM
Copyright Ó 2011 TREK, INC.