Operator`s Manual - Jamieson Equipment Co.

Operator`s Manual - Jamieson Equipment Co.
880 Performance Series
Panel Mount Controller/Indicator
Version 1.0
Operator’s Manual
152240
Contents
About This Manual ................................................................................ 1
1.0
Introduction............................................................................... 1
1.1
1.2
1.3
1.4
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operating Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Front Panel Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Indicator Operations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.4.1
1.4.2
1.4.3
1.4.4
1.4.5
1.4.6
1.4.7
1.4.8
1.4.9
1.4.10
1.4.11
1.4.12
1.4.13
1.4.14
2.0
2
3
3
6
Toggle Gross/Net Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Toggle Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Zero Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Acquire Tare . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Remove Stored Tare Value . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Preset Tare (Keyed Tare) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Print Ticket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Front Panel Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Displaying Audit Trail Information . . . . . . . . . . . . . . . . . . . . . . . 8
Setpoints . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Set Time and Date. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Display Accumulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Clear the Accumulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Display Tare. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Installation .............................................................................. 12
2.1 Unpacking and Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.2 Panel Mount Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.2.1
Mounting the Controller Assembly Remotely . . . . . . . . . . . . . 14
2.3 Cable Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.3.1
2.3.2
2.3.3
2.3.4
2.3.5
2.3.6
Load Cells . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Serial Communications . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
USB Device Communications – Port 2 (USBCOM). . . . . . . . .
USB Host Device. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Ethernet Communications . . . . . . . . . . . . . . . . . . . . . . . . . . .
15
15
16
16
16
16
2.4 CPU Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.5 Audit Trail and the Audit Jumper . . . . . . . . . . . . . . . . . . . . . . . . 17
2.5.1
Digital I/O. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.6 Option Cards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.7 Battery Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Technical training seminars are available through Rice Lake Weighing Systems.
Course descriptions and dates can be viewed at www.ricelake.com/training
or obtained by calling 715-234-9171 and asking for the training department.
© Rice Lake Weighing Systems. All rights reserved. Printed in the United States of America.
Specifications subject to change without notice.
Rice Lake Weighing Systems is an ISO 9001 registered company.
Version 1.0,
October 31, 2013
i
880 Operator’s Manual
3.0
Configuration .......................................................................... 19
3.1 Configuration Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.1.1
3.1.2
3.1.3
3.1.4
Revolution® Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . .
EDP Command Configuration . . . . . . . . . . . . . . . . . . . . . . . .
Front Panel Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . .
Alphanumeric Entry Procedure . . . . . . . . . . . . . . . . . . . . . . .
19
20
20
22
3.2 Menu Structures and Parameter Descriptions . . . . . . . . . . . . . . 22
3.2.1
3.2.2
3.2.3
3.2.4
3.2.5
3.2.6
3.2.7
3.2.8
3.2.9
3.2.10
4.0
Audit Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Setup Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Scale Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Feature Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Ports Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Print Format Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Setpoint Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Digital Input/Output Menu . . . . . . . . . . . . . . . . . . . . . . . . . . .
Analog Output Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Version Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
22
23
23
26
28
30
30
33
33
33
Print Formatting ...................................................................... 34
4.1 Print Formatting Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.2 Default Ticket Formats. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.3 Stream Serial Data Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.0
Specifications ......................................................................... 37
880 Limited Warranty.......................................................................... 39
Rice Lake continually offers web-based video training on a growing selection
of product-related topics at no cost. Visit www.ricelake.com/webinars.
Contents
ii
About This Manual
This manual is intended for use the operator of the 880 digital weight indicators and
applies to indicators using Version 1.0 of the 880 software.
The Technical/Service manual (PN 158387) that is referred to throughout this manual
and is available online.
Authorized distributors and their employees can view or download this
m a n u a l f r o m t h e R i c e L a k e We i g h i n g S y s t e m s w e b s i t e a t
www.ricelake.com.
Please leave this manual with the indicator when installation and configuration are
complete.
1.0 Introduction
The 880 is a single-channel digital weight indicator housed in a panel mount
enclosure. The front panel bezel can be sealed to a NEMA 4X/IP65 rating. The front
panel consists of a 6-button keypad; a large, seven-digit, 14-segment, vacuum
fluorescent display; two-character dot-matrix annunciator field; and 16-character dotmatrix prompt field. Features include:
• Drives up to 8 350or 16 700 load cells
• Supports four and six wire load cell connections
• Four configurable digital inputs or outputs
• Full duplex RS-232 or half duplex RS-485 communications at up to 115200
bps
• Ethernet TCP/IP interface for 10-Base T network communications
• USB interface for host (type A connection) or device
• Hardware slot for one option card
• Optional DeviceNet™ interface for communications network with AllenBradley® controllers1
• Optional Ethernet/IP interface for Allen-Bradley PLC and other Ethernet/IP
master devices3
• Optional Profibus® DP interface for network communications2
• Optional Modbus
• Optional analog output module provides 0–10 VDC or 4–20 mA tracking of
gross or net weight values
• Optional four channel relay module, dry connect 3A @ 115VAC, 5A @
30VDC
• Available in 115 VAC and 230 VAC versions
1. DeviceNet™ is a trademark of the Open DeviceNet Vendor Association.
2. Profibus® is a registered trademark of Profibus International.
3. EtherNet/IP™ is a trademark of Open DeviceNet Vendor Association.
1
880 Operator’s Manual
1.1
Safety
Safety Symbol Definitions
Indicates a potentially hazardous situation that, if not avoided,
WARNING could result in serious injury or death, and includes hazards that
are exposed when guards are removed.
Indicates a potentially hazardous situation that, if not avoided,
CAUTION may result in minor or moderate injury.
Indicates information about procedures that, if not observed,
Important could result in damage to equipment or corruption to and loss of
data.
Safety Precautions
Do not operate or work on this equipment unless you have read
and understand the instructions and warnings in this manual.
Failure to follow the instructions or heed the warnings could
result in injury or death. Contact any Rice Lake Weighing
Systems dealer for replacement manuals. Proper care is your
responsibility.
Some procedures described in this manual require work inside
WARNING the indicator enclosure. These procedures are to be performed by
qualified service personnel only.
General Safety
WARNING
Failure to heed may result in serious injury or death.
DO NOT allow minors (children) or inexperienced persons to operate this unit.
DO NOT operate without all shields and guards in place.
DO NOT step on the unit.
DO NOT jump on the scale.
DO NOT use for purposes other than weight taking.
DO NOT place fingers into slots or possible pinch points.
DO NOT use any load-bearing component that is worn beyond 5% of the original
dimension.
DO NOT use this product if any of the components are cracked.
DO NOT exceed the rated load limit of the unit.
DO NOT make alterations or modifications to the unit.
DO NOT remove or obscure warning labels.
DO NOT use near water.
Before opening the unit, ensure the power cord is disconnected from the outlet.
Keep hands, feet and loose clothing away from moving parts.
Introduction
2
1.2
Operating Modes
The 880 has three modes of operation:
Normal mode
Normal mode is the weighing mode of the indicator. The indicator
displays gross or net weights as required, and annunciators to indicate
scale status and the type of weight value displayed.
Setup mode
Most of the procedures described in this manual require the indicator to
be in setup mode, including configuration and calibration.
To enter setup mode, remove the fillister head screw from the enclosure
backplate. Insert a screwdriver or a similar tool into the access hole and
press the setup switch once. The indicator display changes to show the
word CONFIG.
The 880 also has an Audit Trail that can track changes to configuration
and calibration, allowing the setup switch to be bypassed with Jumper J5
on the CPU board.
User Setup Mode
User setup mode (accessed by pressing the MENU key) is used
•
•
•
•
View the audit trail
Set the time and date
Access the accumulator
Change setpoint values
See Section 1.4.8 for more information about entering user setup mode.
1.3
Front Panel Display
Figure 1-1 shows the 880 front panel keys and the key functions assigned in normal
mode.
The numeric display consists of six large, 14-segment LED digits. If a negative
number is displayed the first LED is used to display “-”, reducing the number of
available digits to five.
The symbols on the keys in Figure 1-1 (representing up, down, enter, left, right)
describe the key functions assigned in setup mode. The keys are used to navigate
through menus, select digits within numeric values, and increment/decrement values.
See Section 3.1.3 for information about using the front panel keys in setup mode.
3
880 Operator’s Manual
Gross
Brutto
lb
Net
kg
T
PT
R1
R2
ZERO
R3
TARE
UNITS
MENU
PRINT
GROSS
NET
B/N
Figure 1-1. 880 Front Panel Display
Key
ZERO
PRINT
GROSS
NET
B/N
MENU
UNITS
TARE
Function
Sets the current gross weight to zero.
Also used to navigate to different menu levels or used to increment a
number when editing a value.
Sends “on-demand” print format out the serial port, provided the
conditions for standstill are met. PRINT may be displayed while the unit
prints.
Also used to navigate to different menus or to toggle to another digit when
editing a value.
Toggles between gross and net. If a tare value has been entered or
acquired, the net value is the gross weight minus the tare.
Gross mode is shown by the Gross/Brutto annunciator; net mode is
shown by the Net annunciator.
Also used to navigate to different menu levels or to decrement a number
when editing a value.
The MENU key is used to access the setup menu.
Also acts as the cancel key when editing parameter values.
Switches the weight display to an alternate unit, defined in the Format
menu, see Technical/Service manual for more information.
Units Available: lb, kg, oz, metric ton, ton, gram.
Also used to navigate to different menus or to toggle to another digit when
editing a value.
Performs one of several predetermined Tare functions dependent on the
mode of operation selected in the TAREFN parameter.
Also acts as enter for numeric or parameter entry.
Table 1-1. Key Functions
Introduction
4
Key
Function
Gross/Brutto LED
Gross weight mode (or Brutto in OIML mode)
Net LED
Net weight mode.
Gross
Brutto
Net
0
lb
kg
lb
kg
lb
kg
lb
kg
R1
Center of Zero LED
Indicates that the current gross weight reading is within +/0.25 display divisions of the acquired zero, or is within the
center of zero band.
A display division is the resolution of the displayed weight
value, or the smallest incremental increase or decrease that
can be displayed or printed.
Standstill LED
Scale is at standstill or within the specified motion band. Some
operations, including Zero, Tare and Printing, can only be done
when the standstill LED is on.
lb/kg LED
lb
Displays which unit of measure is being used.
The lb and kg annunciators indicate the units associated with
kg
the displayed value. If the displayed value is pounds, lb will be
lit. If the displayed value is kilograms, kg will be lit.
primary/secondary LED
primary
If the other units value is neither lb or kg, then lb will be lit for
secondary the units assigned as primary, and kg will be lit for the units
assigned as secondary.
lb/tn,t,oz,g,none LED
lb
Alternate units that can be displayed include short tons (tn),
tn, t, oz, metric tons (t), ounces (oz), grams (g), or NONE (no units). If the
g or none displayed units is one of these alternate units, and the other
unit value is lb, then lb will be lit.
tn, T, oz, tn,t,oz,g,none /kg LED
g or none Alternate units that can be displayed include short tons (tn),
metric tons (t), ounces (oz), grams (g), or NONE (no units). If the
kg
displayed units is one of these alternate units, and the other
unit value is kg, then kg will be lit.
T LED
T
Indicates that a tare has been acquired and stored by the
system.
PT
PT LED
Indicates that a preset tare weight has been keyed in or
entered via the EDP serial port.
R2
R3
Bluetooth
Bluetooth communications enabled
R1, R2 and R3
Range settings for multi interval or multi range
Table 1-2. Annunciator Functions
5
880 Operator’s Manual
1.4
Indicator Operations
Basic 880 operations are summarized below:
1.4.1
Toggle Gross/Net Mode
1. Press
GROSS
NET
B/N
to toggle the display mode between gross and net.
Net mode is only available if a tare value has been entered or acquired,
Note the net value is the gross weight minus the tare. If no tare has been
entered or acquired, the display remains in gross mode.
A lit LED next to the Gross (Brutto in OIML) or Net on the display indicates the
current mode.
1.4.2
Toggle Units
UNITS
1. Press
to switch between primary and secondary units. The
current units LED will be lit.
1.4.3
Zero Scale
1. In gross mode, remove all weight from the scale and wait for the
LED to light.
2. Press
1.4.4
ZERO
. The
0
LED lights to indicate the scale is zeroed.
Acquire Tare
1. Place container on scale and wait for the
LED to light.
2. Press TARE
to acquire the tare weight of the container. Net weight
is displayed and the T LED lights to show the tare value was entered.
1.4.5
Remove Stored Tare Value
1. Remove all weight from the scale and wait for the
light.
2. The display should read zero and the
the scale if needed.
0
LED to
LED should be lit, zero
3. Press TARE
(or, in OIML mode, the ZERO key). Display shifts to
gross weight and the Gross LED is lit.
Note If keyed tare is allowed, press
To clear the tare press
TARE
TARE
to open the keyed tare prompt.
again.
Introduction
6
1.4.6
Preset Tare (Keyed Tare)
Note Tare mode muste be set to keyed or both for the preset tare caommand
to function.
1. With the scale empty and display showing zero weight, press
TARE
.
2. Display will show (000000); the focused digit will flash.
3. To edit the value:
• Press or to select the digit.
• Press or to increment or decrement the value.
• Press
TARE
• Press
or
• Press
TARE
to move to the decimal point entry.
to adjust the decimal point placement.
when the value is correct.
4. The display will change to the Net mode and the PT LED lights to
show the preset tare was entered.
Note Entering a keyed tare of zero will remove the stored tare value.
1.4.7
Print Ticket
1. Wait for the (
2. Press
1.4.8
to send data to a communications port.
Front Panel Setup
1. Press
•
•
•
•
•
•
7
PRINT
) LED to light.
MENU
to enter user setup mode. Use user setup to:
View audit trail information
Enter setup mode if audit trail is enabled
View or set time and date
View or clear the accumulator
Change setpoint values and enable/diable setpoints
View the current tare value
880 Operator’s Manual
1.4.9
Displaying Audit Trail Information
The Audit Trail Configuration and Calibration counters can be viewed
through the User Menu.
1. Press
2. Press
3. Press
4. Press
5. Press
6. Press
7. Press
MENU
, AUDIT is displayed.
to display the Legally Relevant Firmware version.
to display CALIB.
to view the Calibration Counter.
MENU
to return to CALIB.
to display CFG.
to view the Configuration Counter.
8. Press
MENU
to return to CFG.
9. Press
MENU
to return to the weigh mode.
1.4.10 Setpoints
Setpoints must be enabled in the setup mode to be accessible in the user setup
mode.
To enter the setup mode:
1. Remove the large fillister head screw from the back of the enclosure.
2. Insert a nonconductive tool into the access hole and press the setup
switch. Indicator display changes to show SCALE .
3. Press or until SETPTS is displayed.
4. Press , SP CFG is displayed.
5. Press , press or to desired setpoint number.
6. Press to enter setpoint settings.
7. Select the type by pressing or to desired setting, then press to
set the value. For complete list of options see Section 3.2.11.
8. When all settings have been made, press
weighing mode.
MENU
to return to
Note Setpoints are now accessible from the front panel menu.
Display or Edit Setpoint Value
1. Press
2. Press
3. Press
MENU
. AUDIT is displayed.
or until SETPTS is displayed.
and the first available setpoint number is displayed.
Introduction
8
4. Press
or
to toggle through each setpoint that is operator
accessible.
5. Press , VALUE is displayed.
6. Press again to display or edit the value. To edit the value:
• Press or to increment or decrement the value of the flashing
digit.
• Press or to select the digit to edit.
• Press
TARE
• Press
or
7. Press
TARE
to move to the decimal point entry.
to adjust the decimal point placement.
to accept the displayed value.
8. Repeat the above steps to set PREACT, if enabled
9. When all settings have been made, press
weighing mode.
MENU
to return to
Setpoint Value and Preact Value may be accessible from the front panel
Note in weigh mode.
Some indicator configurations may not allow setpoint values to be
changed through the front panel or may require a password to display or change
the setpoint value.
Turn Setpoint On or Off
To turn a setpoint off at the front panel:
1. Press
MENU
. AUDIT is displayed.
2. Press or until SETPTS is displayed.
3. Press and the first available setpoint number is displayed.
4. Press
or
to toggle through each setpoint that is operator
accessible.
5. Press , then press or to ENABLE/DISABLE.
6. Press , then press or to turn setpoint ON/OFF.
7. Press
TARE
to accept the setting.
8. Press
MENU
to return to weighing mode.
Some indicator configurations may not allow setpoints to be turned off
Note through the front panel or may require a password to turn the setpoint
on and off.
9
880 Operator’s Manual
1.4.11 Set Time and Date
1. Press
2.
3.
4.
5.
MENU
, AUDIT is displayed.
Press or until T&D is displayed.
Press , TIME is displayed.
Press , to enter time.
To edit the value ( 24 hour format):
• Press or to select the digit.
• Press or to increment or decrement the value.
6. Press
TARE
when the value is correct. DATE is displayed.
7. Press , to enter time.
8. To edit the value in specified format (MMDDYY, DDMMYY, or
YYMMDD):
• Press or to select the digit.
• Press or to increment or decrement the value.
9.
Press
10. Press
TARE
MENU
when the value is correct. TIME is displayed.
to return to weighing mode.
1.4.12 Display Accumulator
The accumulator must be enabled before use in either normal mode or
setpoint operations. Once enabled, weight (net weight if a tare is in the
system) is accumulated whenever a print operation is performed using the
PRINT key, digital input, setpoint PSHACC operation or serial command. The
scale must return to below the threshold value (except for the setpoint PSHACC
operation) before the next accumulation.
1. Press
MENU
2. Press
or
to enter the user setup mode, AUDIT is displayed.
until ACCUM is displayed.
Note ACCUM is only displayed if the accumulator is enabled.
3. Press
4. Press
, VIEW is displayed.
to view the current accumulator value.
5. While the accumulator value is displayed, press
value.
PRINT
to print the
Note The format of the print ouput can be configured using accumulator
print format. See Section 7.0.
Introduction 10
1.4.13 Clear the Accumulator
1. Press
2. Press
3. Press
4.
to enter the user setup mode, AUDIT is displayed.
MENU
or until ACCUM is displayed.
, then press or until CLR Y is displayed.
Press TARE
to clear the accumulator, CLEAR will display briefly
and display returns to CLR Y .
5. Press
MENU
to return to the weigh mode.
The PRINT key only performs one accumulation. Weight must return to
Note zero below the configured threshold value before another accumulation
is allowed.
Threshold is configured in the setup menu, see Section 3.2.3.
1.4.14 Display Tare
When a stored Tare value is displayed, the Gross and Net LEDs will be off
and the
will be lit. To display a stored tare:
1. Press
2. Press
3. Press
MENU
.
to TARE and press
MENU
to view the current tare value.
twice to return to weighing mode.
If there is no tare in the system, the value displayed will be zero and the Gross
and Net LED will be turned off.
See Section 9.12 for more information pertaining to the regulatory mode of
operation.
11 880 Operator’s Manual
2.0
Installation
This section describes procedures for connecting load cells, digital I/O, and serial
communications cables to the 880 indicator. Instructions for replacement of the CPU
board are included, along with assembly drawings and parts lists for the service
technician.
CAUTION
• Use a wrist strap to ground yourself and protect components from electrostatic
discharge (ESD) when working inside the indicator enclosure.
• This unit uses double pole/neutral fusing which could create an electric shock
hazard. Procedures requiring work inside the indicator must be performed by
qualified service personnel only.
• The supply cord serves as the power disconnect for the 880. The power outlet
supplying the indicator must be installed near the unit and be easily accessible.
2.1
Unpacking and Assembly
Immediately after unpacking, visually inspect the 880 to ensure all components are
included and undamaged. The shipping carton should contain the controller, display,
Operator's manual, CD and parts kit. If any parts were damaged in shipment, notify
Rice Lake Weighing Systems and the shipper immediately.
2.2
Panel Mount Installation
The controller can be mounted to the display DIN rail or mounted
Note remotely up to 250 ft away from display.
5.0"
6.0"
4.0"
3.0"
5.3"
Figure 2-1. 880 Panel Mount Dimensions
Use the Din Rail mount plate as a template (see Figure 2-2) to drill the mounting
holes in the panel for the stainless steel panel mount enclosure.
See Figure 2-1 for enclosure dimensions.
Installation 12
4.25"
0.88"
5/32" Ø
0.88"
3.38"
2.25"
4.00"
1.13"
7/8" Ø
Use DIN Rail mount plate
as a template for drilling
mounting holes in panel.
6.00"
Figure 2-2. Pattern for Cutting Holes
1.
2.
3.
Install the gasket on the display assembly. It must be seated correctly on
display before mounting to ensure tight seal.
Align the display assembly to the front and the DIN rail mounting plate to
the back of the panel with the holes cut in step 1.
Secure the display and mounting plate to the panel using the four 6-32 Kep
nuts (PN 14621) provided.
Ensure gasket is seated properly on
Important display assembly before mounting.
Panel for mounting
indicator
Din Rail
Mounting Plate
Display Assembly
Figure 2-3. Mounting Display Assembly
13 880 Operator’s Manual
6-32 Kep Nuts
4.
5.
6.
Connect the cable assembly to the controller assembly.
Place the controller assembly on the DIN rail as shown in Figure 2-4.
Secure to DIN rail by rotating the left and right hooks up connecting with the
bottom rail.
Connect Cable assembly to
controller assembly
Controller Assembly
Rotate hooks to secure
to bottom DIN rail
Figure 2-4. Install Controller Assembly
2.2.1
Mounting the Controller Assembly Remotely
To mount the controller assembly remotely a 6 pin connector (PN 153885) is
required. See Figure 2-5 for terminal location and pin assignments.
Controller Assembly can be mounted remotely on a standard 35 mm DIN
Note rail up to 250 ft from display.
J1
J2
J3
J4 J5 J6
Pin
Function
1
+6V
2
GND
3
-TX
4
+TX
5
-RX
6
+RX
When mounting the display
remotely from the controller
assembly, connect to this
terminal.
Figure 2-5. Mount Controller Assembly Remotely
Installation 14
2.3
Cable Connections
The 880 has three external connectors, a terminal connector for the power cord, and a
cutout for installed options. Enclosure disassembly is not required to make
connections to load cells, communications, digital inputs, and digital outputs. These
connectors are all externally mounted on the back of the controller.
2.3.1
Load Cells
To attach cable from a load cell or junction box, route the cable to the external
J1connector. Wire the load cell cable from the load cell or junction box to connector
J1 as shown in Table 2-1. If using 6-wire load cell cable (with sense wires), open the
unit and remove jumpers JP5 and JP6.
Note
For 4-wire installation, leave jumpers JP5 and JP6 on (see Figure 2-7).
Shield wire will attach to ground clamp on backplate.
Connector
Pin
Function
J1
1
+SIG
2
–SIG
3
+SENSE
4
–SENSE
5
+EXC
6
–EXC
For 6-wire load cell connections, remove jumpers JP5 and JP6.
Table 2-1. J1 Pin Assignments
2.3.2
Power Connections
Power connections to the 880 Panel Mount are shown below.
• Part number 152334 is used to connect AC power to the power supply board.
• Part number 15888 is used to connect DC power to the power supply board.
Attach the wires per the diagram shown below.
AC Power
DC Power
J1
J2
J3
J4 J5 J6
Pin
AC
DC
1
2
3
N
GND
L
+
GND
–
Figure 2-6. AC/DC Power Cord Connection Location
15 880 Operator’s Manual
2.3.3
Serial Communications
Connector J3 provides connections for the RS-232 or the two-wire RS-485. Table 2-2
shows the pin assignments.
Pin
RS-232
RS-485
1
2
3
GND
RX
TX
GND
B
A
Table 2-2. Serial Communications to Connector J7 Pin Assignments
For RS-232, all four switches on SW3 (see Figure 2-7) must be in the OFF
Note position.
For RS-485, all four switches on SW3 must be in the ON position.
2.3.4
USB Device Communications – Port 2 (USBCOM)
The USB Device Port is intended to be connected only to a PC. It will appear as a
Virtual Comm Port and be assigned a “COMx” (J4 micro USB connector, see
Figure 2-6) designation. Applications will communicate through the port like a
standard RS-232 communications port. See Technical/Service Manual for more
details.
2.3.5
USB Host Device
Not available at time of publication.
2.3.6
Ethernet Communications
The 880 supports Ethernet TCP/IP 10Base-T/100Base-TX communications using a
standard RJ45 connector (J6 – see Figure 2-6). Through the network, software
applications are able to communicate with the 880 using the EDP command set. See
Technical/Service Manual for more details.
Installation 16
2.4
CPU Board
J8
J7
Audit
Jumper
JP4
JP5
JP6
J1
J2
SW3
J3
J4
J5
J6
Figure 2-7. 880 CPU Board
Connectors
•
•
•
•
Load cell (J1)
DIO (J2)
Comm 1 (J3)
USB Micro Device (J4)
•
•
•
•
USB Host (J5)
Ethernet TCP/IP (J6)
Opt Header (J8)
Power Board (J7)
The Comm 1 port supports RS-232 or RS-485 communications; selectable with
switch SW3. The port is configured using the COM menu under Ports. See
Section 3.2.5.
2.5
Audit Trail and the Audit Jumper
The 880 includes an audit trail feature that keeps track of the number and the last date
of calibrations and Legal-for-Trade configuration changes. It is possible to setup the
880 to allow entry to the configuration and calibration menus using only the front
panel MENU key.
On the top of the CPU board is a 3-pin Jumper (JP4 - see Figure 2-7) that enables or
disables this feature.
• To use the audit trail and allow the use of the MENU key to enter the
configuration and calibration mode, place the jumper in the ON position.
• To prevent the use of the MENU key to enter the configuration and
calibration mode, instead requiring use of the externally sealable
configuration switch located inside the enclosure (see Figure 3-1), place the
Jumper in the OFF position.
The audit trail counters will operate in either position of the audit jumper.
17 880 Operator’s Manual
2.5.1
Digital I/O
Digital inputs can be set to provide many indicator functions, including all keypad
functions. Digital inputs are active (on) with low voltage (0 VDC), inactive (off) at 5
VDC. Use the DIG IN menu to configure the digital inputs.
Digital outputs are typically used to control relays that drive other equipment.
Outputs are designed to sink, rather than source, switching current. Each output is a
normally open collector circuit, capable of sinking 250 mA when active. Digital
outputs are wired to switch relays when the digital output is active (low, 0 VDC) with
reference to a 5 VDC supply.
Use the SETPTS menu (see Figure 3-16 on page 30) to configure digital outputs.
Table 2-3 shows the pin assignments for connector J2.
Connector
Pin
Signal
J2
1
2
3
4
5
6
+5VDC, 500mA max
GND
DIO1
DIO2
DIO3
DIO4
Table 2-3. J6 Pin Assignments (Digital I/O)
2.6
Option Cards
Connector J8 is reserved for option cards. Table 2-4 lists the options available for the
880 Indicator. Each kit includes instructions for installing and setting up the option.
Option
Addendum Part Number
Analog Output Card
Relay Board
EtherNet/IP
DeviceNet
Profibus DP
Modbus TCP
ProfiNet
131601
131602
156554
156558
156556
156557
156555
Table 2-4. Available Option Cards
2.7
Battery Replacement
When battery voltage gets low, the indicator display shows low bat. Replace the
battery when this warning is displayed. The battery should last about seven years.
Use the Revolution® configuration utility or EDP commands (see Section 6.1 on
page 64) to store a copy of the indicator configuration on a PC before attempting
battery replacement. If any data is lost, the indicator configuration can be restored
from the PC.
Danger of explosion if battery is incorrectly replaced. Replace
WARNING only with the same or equivalent type recommended by the
manufacturer.
Dispose of unused batteries according to the manufacturer’s instructions.
The battery should be considered bad and discarded at 2.9V.
Installation 18
3.0
Configuration
To configure the 880 indicator, the indicator must be placed in setup mode. The setup
switch is accessed through a small hole on the enclosure backplate, insert a
nonconductive tool into the access hole and press the configuration switch. Indicator
display changes to show CONFIG.
Use caution when inserting the nonconductive tool into the
Important backplate, press the tool in about 3/4 inch, using the board as a
guide, until the switch is engaged. Do not use excessive force, the
switch may be damaged.
Note If audit trail is enabled, press
Press
or
MENU
to access setup mode.
until SETUP is displayed, then press
to SCALE.
.
J1
J2
J3
J4 J5 J6
Setup
Switch
Access
Figure 3-1. Back View – Setup Switch Access
When the indicator is placed in setup mode, the word CONFIG is shown on the
display. The CONFIG menu is the first of ten top-level menus used for configuring
the indicator. Detailed descriptions of these menus are given in Section 3.2. When
configuration is complete, return to the CONFIG menu and press the (ZERO) key
to exit setup mode.
When configuration is complete, press
3.1
MENU
to return to the weigh mode.
Configuration Methods
The 880 indicator can be configured by using the front panel keys to navigate through
a series of configuration menus or by sending commands or configuration data to the
EDP port. Configuration using the menus is described in Section 3.1.3.
Configuration using the EDP port can be accomplished using either the EDP
command set described in the Technical Service manual or Version 3.0 or later of the
Revolution® configuration utility.
3.1.1
Revolution® Configuration
The Revolution configuration utility provides the preferred method for configuring
the 880 indicator. Revolution runs on a personal computer to set configuration
parameters for the indicator. When Revolution configuration is complete,
configuration data is downloaded to the indicator.
See Section 5.0 of the Technical/Service manual (PN 158387) for more
Note information on Revolution.
19 880 Operator’s Manual
3.1.2
EDP Command Configuration
The EDP command set can be used to configure the 880 indicator using either a
personal computer or terminal. Like Revolution, EDP command configuration sends
commands to the indicator EDP port; unlike Revolution, EDP commands can be sent
using any external device capable of sending ASCII characters over a serial
connection.
EDP commands duplicate the functions available using the indicator front panel and
provide some functions not otherwise available. EDP commands can be used to
simulate pressing front panel keys, to configure the indicator, or to dump lists of
parameter settings. See the Technical Service manual for more information about
using the EDP command set.
3.1.3
Front Panel Configuration
The 880 indicator can be configured using a series of menus accessed through the
indicator front panel when the indicator is in setup mode. Table 3-1 summarizes the
functions of each of the main menus.
AUDIT
SETUP
See
Figure 3-6
See
Figure 3-7
T&D
Displayed only
if audit trail
is enabled
SETPTS
TARE
See
Figure 3-16
000000
ACCUM
VIEW
CLR Y?
000000
ACCUM
Displayed only
if turned on in
Scale Menu
TIME
DATE
000000
000000
Figure 3-2. 880 Menu Layout
Menu
Menu Function
AUDIT
Audit Trail
Displays the legally relevant (LR) firmware version,
configuration count and calibration count.
SETUP
Setup
T&D
Time and Date
View and change time and date.
ACCUM
Accumulator
Displays the current accumulator value.
SETPTS
Setpoints
Configure setpoints.
TARE
Tare
Tare function. Enables or disables push-button and
keyed tare
Table 3-1. 880 Menu Summary
Configuration 20
T
ZERO
UNITS
MENU
TARE
PRINT
GROSS
NET
B/N
Figure 3-3. Five-Key Keypad Functions in Setup Mode
Four front panel keys are used as directional keys to navigate through the menus in
setup mode (see Figure 3-3). The UNITS ( ) and PRINT ( ) keys scroll left and right
(horizontally) on the same menu level; ZERO ( ) and GROSS/NET ( ) move up and
down (vertically) to different menu levels. The TARE key ( ) serves as an Enter key
for selecting parameter values within the menus.
Level 1
Parameter
Level 1
Parameter
Level 2
Parameter
Level 2
Parameter
Default value
Value
Value
When moving through values below the first menu level, press
above.
Value
to return to the level
Figure 3-4. Setup Mode Menu Navigation
To select a parameter, press or to scroll left or right until the desired menu group
appears on the display, then press to move down to the submenu or parameter you
want. When moving through the menu parameters, the default or previously selected
value appears first on the display.
To change a parameter value, scroll left or right to view the values for that parameter.
When the desired value appears on the display, press
to select the value and move
back up one level. To edit numerical values, use the navigation keys to select the digit
and to increment or decrement the value.
00 0 0 00
When editing numeric values, press or to change the selected digit.
Press
or
to increment or decrement the value of the selected digit.
After the number desired is displayed, press ENTER. The decimal point
will begin flashing if a decimal value is allowed. To move the decimal,
use or to move the decimal position. Press ENTER when done.
Figure 3-5. Editing Procedure for Numeric Values
21 880 Operator’s Manual
3.1.4
Alphanumeric Entry Procedure
Because of its limited keypad, the 880 will use the following scheme for
alphanumeric entry.
Level 1
Parameter
Level 1
Parameter
Level 2
Parameter
Level 2
Parameter
Display first 6
characters of format
Scroll left in format string
Press to insert a space
before the active character
Display and edit
active character and
ASCII value
Decrement ASCII value of active character
Scroll right in format string
Press
to accept the changes
and return to the level above.
Increment ASCII value of active character
Press
to accept the changes
and return to the level above.
Delete active
character
Table 3-2. Editing Procedure for Numeric Values
3.2
Menu Structures and Parameter Descriptions
The following sections provide graphic representations of the 880 menu structures. In
the actual menu structure, the settings you choose under each parameter are arranged
horizontally. To save page space, menu choices are shown in vertical columns. The
factory default setting appears at the top of each column and is bolded. Parameters
shown surrounded by a dotted-line box only appear under the special circumstances
explained under each box.
Most menu diagrams are accompanied by one or more tables that describe all
parameters and parameter values associated with that menu option. Default parameter
values are shown in bold type.
3.2.1
Audit Menu
AUDIT
LRV 1.00
SETUP
T&D
ACCUM
CALIB
CFG
number
number
SETPTS
TARE
Figure 3-6. Audit Menu Structure
AUDIT Menu
Parameter
LRV
CALIB
CFG
Choices
--000000
000000
Description
Legally relevant firmware version
Displays total calibration events.
Displays total configuration events.
Table 3-3. Audit Menu Parameters
Configuration 22
3.2.2
Setup Menu
SCALE
AUDIT
SETUP
T&D
ACCUM
SETPTS
TARE
FEATUR
PORTS
PFORMT
SETPTS
DIGIO
ALGOUT
VERS
Figure 3-7. Setup Menu Structure
3.2.3 Scale Menu
.... SETUP ....
SCALE
....
GRADS
SPLIT
10000
OFF
1-100000
2RNG
3RNG
2INTVL
FORMAT
ZTRKBN
ZRANGE
MOTBAN
OVRLOA
SSTIME
DSPRAT
0-100
0-100
0-100
FS+2%
1-65536
1-80
FS+ 1D
See
Figure 3-9
FS+ 9D
FS
3INTVL
SMPRAT
PWRUPM
ACCUM
THRESH
BOTH
GO
OFF
0-999999
NOTARE
DELAY
ON
DFSENS
DFTHRH
TAREFN
30HZ
LIGHT
1-99999
60HZ
MEDIUM
120HZ
HEAVY
PBTARE
KEYED
240HZ
480HZ
960HZ
7.5HZ
15HZ
Figure 3-8. Scale Menu Structure
23 880 Operator’s Manual
CALIBR
See
Figure 3-10
Format Menu
....
....
SETUP
....
SCALE
....
FORMAT
....
SEC
PRI
UNITS
DECPNT
DSPDIV
UNITS
1D
LB
88888.8
5D
KG
2D
KG
888888
1D
OZ
5D
OZ
888888
2D
TN
88.8888
TN
8.88888
T
888.888
T
88.8888
G
8888.88
G
888.888
NONE
88888.8
NONE
8888.88
LB
DECPNT
DSPDIV
888888
888880
8.88888
If SPLIT = OFF
....
....
SETUP
....
SCALE
....
FORMAT
....
Only available if
SPLIT = 3RNG or 3INTVL
DECPT1
DDIV1
MAX1
DECPT2
DDIV2
888888
888880
8.88888
5D
MAX2
DECPT3
DDIV3
1D
2000
88888.8
2D
weight
888888
888880
2D
MAX3
5D
5000
88888.8
5D
10000
LB
1D
weight
888888
1D
weight
KG
888880
2D
UNITS
OZ
TN
88.8888
8.88888
8.88888
888.888
88.8888
88.8888
T
8888.88
888.888
888.888
G
88888.8
8888.88
8888.88
NONE
If SPLIT = 2RNG, 3RNG, 2INTVL, or 3INTVL
Figure 3-9. Format Menu Structure
Configuration 24
Calibration Menu
....
....
SETUP
....
SCALE
....
CALIBR
....
WZERO
WVAL
WSPAN
Previous A/D raw
counts are shown
Press Enter
to calibrate zero
Press Menu
to cancel
Display and edit
test weight value
Previous A/D raw
counts are shown
Press Enter
to calibrate span
Press Menu
to cancel
Calibrating, please wait
REZERO
LAST
Calibrating, please wait
Calibrating, please wait
Calibrating, please wait
Press Enter to
compensate for
zero offset
Press enter to recall
the last established
pushbutton zero value
to allow calibration
without removing test
weights from the scale.
Press enter to
temporarily zero the
displayed weight of a
non-empty scale.
After span calibration
the previously
calibrated zero value
is used as the offset.
WLIN
Calibrating, please wait
POINT1
POINT2
POINT3
POINT4
POINT5
Same as POINT 1
WGT 1
CAL1
Display and edit
test weight value
Previous A/D raw
counts are shown
Press Enter
to calibrate span
Press Menu
to cancel
Calibrating, please wait
Figure 3-10. Calibration Menu Structure
25 880 Operator’s Manual
TEMP
3.2.4
Feature Menu
....
SETUP
....
CONFIG
FEATUR
REGION
....
UID
CONSC#
000000
CURVAL
RESVAL
USER
SETUP
0
0
0
0
See
Figure 3-12
KEYLCK
PASSWD
LOCALE
LATUDE
ELEVAT
OFF
0-90
-9999 to +9999
ON
Shown if LOCALE is ON
ZERO
GRSNET
UNITS
PRINT
TARE
MENU
UNLOCK
UNLOCK
UNLOCK
UNLOCK
UNLOCK
UNLOCK
LOCK
LOCK
LOCK
LOCK
LOCK
LOCK
Figure 3-11. Feature Menu Structure
Configuration 26
Region Menu
....
SETUP
FEATUR
....
....
REGION
REGULA
REGWOR
DECFMT
NTEP
GROSS
DOT
CANADA
BRUTTO
COMMA
TIME
INDUST
NONE
OIML
TFORMT
T SEP
SET
12 HOUR
COLON
XX.XXAM
24 HOUR
COMMA
DATE
DFORMT
D SEP
SET
MMDDY4
SLASH
000000
DDMMY4
DASH
Y4MMDD
SEMI
Y4DDMM
MMDDY2
See INDUST
submenu below
DDMMY2
Y2MMDD
Y2DDMM
SNPSHT
ZTARE
KTARE
MTARE
NTARE
CTARE
DISPLY
NO
NO
REPLAC
NO
YES
SCALE
YES
YES
REMOVE
YES
NO
NOTHIN
PRTMOT
PRTPT
OVRBAS
AUDAG
NO
NO
CALIB
NTEP
YES
YES
SCALE
CANADA
NONE
OIML
Figure 3-12. Region Menu Structure
27 880 Operator’s Manual
3.2.5
Ports Menu
....
SETUP
....
PORTS
COM
....
....
USBCOM
See
Figure 3-14
ETHNET
FLDBUS
DHCP
SWAP
DVCNET
PRFBUS
NONE
Address
Address
number
number
IPADRS
NETMSK
DFTGWY
DNSPRI
BYTE
BOTH
DNSSEC
PORT
RMOTIP
RMOTPT
MAC
ECHO
Figure 3-13. Ports Menu Structure
Configuration 28
Com and USBCOM Menu
....
....
SETUP
PORTS
....
....
USBCOM
COM
BAUD
BITS
STOP B
TERMIN
COMAND
9600
8NONE
1
CR/LF
ON
ON
STRIND
19200
7EVEN
2
CR
OFF
OFF
STRLFT
28800
7ODD
REMOTE
38400
TRIGGE
ECHO
RESPNS
57600
115200
Not available in USBCOM
1200
2400
4800
EOLDLY
TYPE
ADDRES
PRNMSG
0-255
232
0-255
OFF
485
Only Available
if TYPE is
set to 485
ON
Not available in USBCOM
TRIGGE
REMOTE
BAUD
BITS
STOP B
PRNMSG
SFMT
format
9600
8NONE
1
OFF
19200
7EVEN
2
ON
28800
7ODD
38400
57600
115200
Menu Structure if TRIGGE is set to REMOTE.
1200
2400
4800
Figure 3-14. Com Menu Structure
29 880 Operator’s Manual
3.2.6
Print Format Menu
....
SETUP
....
....
PFORMT
....
GFMT
NFMT
ACCFMT
SPFMT
HDRFMT
format
FMT
PORT
format
COM
USBCOM
OFF
Figure 3-15. Print Format Menu Structure
3.2.7
Setpoint Menu
.
SETUP
.
SETPTS
SP CFG
....
....
BATCHG
OFF
AUTO
MANUAL
SETPT1
SETPT2-20
OFF
GROSS
See Figure 3-17
NET
Menu Layout A
Same as
SETPT1
-GROSS
-NET
%REL
PAUSE
DELAY
See Figure 3-18
WAITSS
Menu Layout B
COUNTR
AUTJOG
TIMER
See Figure 3-19
CONCUR
Menu Layout C
Figure 3-16. Setpoint Menu Structure
Configuration 30
Setpoint Menu – Layout A
A
GROSS, NET, -GROSS, -NET, %REL
VALUE
TRIP
BANDVAL
HYSTER
PREACT
PREVAL
number
HIGHER
number
number
OFF
number
If TRIP=
INBAND or
OUTBAND
If TRIP=
HIGHER or
LOWER
LEARN
ON
LOWER
INBAND
OUTBAN
If PREACT is
ON or LEARN
PREADJ
PRESTA
PCOUNT
RELNUM
BATSEQ
CLRACM
number
number
number
1–20
OFF
OFF
ON
ON
If PREACT is LEARN
%RELSP Only
CLRTAR
PSHACM
PSHPRT
PSHTAR
OFF
OFF
OFF
ON
ON
ON
ONQUIE
WAITSS
SLOT
ACCESS
NAME
OFF
ON
6ALPHA
ON
HIDE
OFF
BRANCH
DIGOUT
SENSE
NONE
BIT1-4
NORMAL
0
SLOT 0
Based on which
DIGIO are config
as OUTPUT
INVERT
1-20
SLOT 1
If BATSEQ is ON
SLOT 1 - if RELAY option board is installed.
SLOT 0 - if any DIGIO are configured as OUTPUT.
Figure 3-17. Setpoint Menu Structure – Layout A
31 880 Operator’s Manual
Setpoint Menu – Layout B
B
PAUSE, DELAY, WAITSS, COUNTR, AUTJOG
VALUE
CLRACM
CLRTAR
PSHACM
PSHPRT
PSHTAR
number
OFF
OFF
OFF
OFF
OFF
ON
ON
ON
ON
ON
ONQUIE
WAITSS
DELAY, WAITSS and AUTJOG setpoints only.
ACCESS
NAME
OFF
6ALPHA
SLOT
NONE
SLOT 0
ON
SENSE
BIT 1-4
NORMAL
0
INVERT
1-20
Based on which DIGIO
are config as OUTPUT
SLOT 1
BRANCH
DIGOUT
COUNTR setpoint
only, if BATSEQ is ON
SLOT 1 - if RELAY option board is installed.
SLOT 0 - if any DIGIO are configured as OUTPUT.
Figure 3-18. Setpoint Menu Structure – Layout B
Setpoint Menu – Layout C
C
VALUE
number
TIMER and CONCUR
END
ACCESS
NAME
1
1
OFF
6ALPHA
1-20
1-20
ON
START
SLOT
DIGOUT
SENSE
NONE
BIT 1-4
NORMAL
SLOT 0
SLOT 1
Based on which DIGIO
are config as OUTPUT
INVERT
SLOT 1 - if RELAY option board is installed.
SLOT 0 - if any DIGIO are configured as OUTPUT.
Figure 3-19. Setpoint Menu Structure – Layout C
Configuration 32
3.2.8
Digital Input/Output Menu
....
SETUP
....
....
DIGIO
....
SLOT 0
BIT 1
BIT 2-4
OFF
Same as
SLOT 0
BIT 1
PRINT
ZERO
TARE
UNITS
DSPTAR
BATRUN
CLEAR
NT/GRS
BATSTR
BATSTP
DSPACC
CLRCN
BATPAS
OUTPUT
BATRST
Figure 3-20. Digital Input/Output Menu Structure
3.2.9
Analog Output Menu
The ALGOUT menu is used only if the analog output option is installed. If the analog
output option is installed, configure all other indicator functions and calibrate the
indicator before configuring the analog output. See Technical/Service Manual for
analog output calibration procedures.
.
SETUP
ALGOUT
....
....
OUTPUT
ERRACT
GROSS
0-10V
NET
0-20MA
4-20MA
ZEROSC
MODE
MIN
MINNEG
MAX
MAXNEG
TWZERO
TWSPAN
FULLSC
0
HOLD
number
OFF
10000
OFF
number
number
ON
number
ON
Figure 3-21. Analog Output Menu Structure
3.2.10 Version Menu
....
SETUP
....
....
VERS
....
SOFTWR
REG
V XX.XX
LRV X.XX
DEFALT
NO
YES
Figure 3-22. Version Menu Structure
33 880 Operator’s Manual
4.0
Print Formatting
The 880 provides five print formats. Formats GFMT and NFMT will be printed in a
ticket format based on the current mode of operation when the PRINT key is pressed
(see Table 4-2 on page 36). HDRFMT can be inserted into any other ticket format
using the <H1> formatting command. SPFMT is printed if PSHPRT is set to ON or
WAITSS in the setpoint configuration. The ACCFMT is printed if the accumulator is
embedded and the print key is pressed while viewing the accumulator value, or if a
setpoint PSHACM is set to ON. If PSHACM is set to ONQUIE, it will accumulate, but
not print.
Each print format can be customized to include up to 300 characters of information
(1000 for HDRFMT), such as company name and address, on print tickets. You can
use the indicator front panel (PFORMT menu), EDP commands, or the Revolution®
configuration utility to customize the print formats.
4.1
Print Formatting Commands
Table 4-1 lists commands you can use to format the 880 print formats. Commands
included in the format strings must be enclosed between < and > delimiters. Any
characters outside of the delimiters are printed as text on the ticket. Text characters
can include any ASCII character that can be printed by the output device.
Ticket Format
Command
Description
GFMT/
NFMT/
ACCFMT
SPFMT
<G>
Gross weight in displayed units (See
notes 1 and 2 below)


<N>
Net weight in displayed units (See notes
1 and 2)


<T>
Tare weight in displayed units (See notes
1 and 2)


<A>
Accumulated weight in displayed units


<AC>
Number of accumulator event (5-digit
counter)


<AT>
Time of last accumulator event


<AD>
Date of last accumulator event


<SCV>
Setpoint captured value

<STV>
Setpoint target value

<SPM>
Setpoint mode (gross or net label)

<SNA>
Setpoint name

<SN>
Setpoint number

<SPV>
Setpoint preact value

Table 4-1. Print Format Commands
Print Formatting 34
Ticket Format
Command
Description
GFMT/
NFMT/
ACCFMT
SPFMT
<TI>
Time


<DA>
Date


<TD>
Time and date


<UID>
Unit ID number (See note 3)


<CN>
Consecutive number (See note 3)


<H1>
Ticket header (HDRFMT)


<NLnn>
New line (nn = number of termination
(<CR/LF> or <CR>) characters) (See
note 4)


<SPnn>
Space (nn = number of spaces) (See
note 4)


<SU>
Toggle weight data format (formatted/
unformatted)
(See note 5)


<AN>
Alibi Ticket Number


<CR>
Carriage return character


<LF>
Line feed character


Note
1. Gross, net, and tare weights are 8 digits in length, including sign and decimal
point, followed by a space and a one- to five-digit units identifier. Total field
length with units identifier is 10-14 characters. Depending on what units are
configured, the units identifier will be lb, kg, oz, tn, t, or g.
2. Gross, net, tare, and accumulator weights can be printed in any configured
weight units by adding the following modifiers to the gross, net, tare, and
accumulator weight commands: /P (primary units), /D (displayed units), /S
(secondary units), /T (tertiary units). If not specified, the current displayed units
(/D) is assumed. Example: To format a ticket to show net weight in secondary
units, use the following command: <N/S>.
3. Unit ID and consecutive number (CN) fields are 1–6 characters in length, as
required.
4. If nn is not specified, 1 is assumed. Value must be in the range 1–99.
5. After receiving an SU command, the indicator sends unformatted data until
the next SU command is received. Unformatted data omits decimal points,
leading and trailing characters.
Table 4-1. Print Format Commands (Continued)
35 880 Operator’s Manual
4.2
Default Ticket Formats
Table 4-2 shows the default print formats for the 880 and lists the conditions under
which each print format is used. The HDRFMT format is used to specify header
information that can be used by the other ticket formats. The contents of the HDRFMT
format can be inserted into any other ticket format using the <H1> formatting
command.
Format
Default Format String
GFMT | FMT
GROSS<G><NL2><TD><NL>
Used When
Normal mode, no tare in
system
GFMT | PORT
Defines the communication port that the format will be sent to.
NFMT | FMT
GROSS<G><NL>TARE<SP><T><NL> Normal mode, tare in system
NET<SP2><N><NL2><TD><NL>
NFMT | PORT
Defines the communication port that the format will be sent to.
ACCFMT | FMT
ACCUM<A><NL><DA> <TI><NL>
Accumulator enabled and
displayed, or setpoint print
operation with
PSHACCM=ON
ACCFMT | PORT Defines the communication port that the format will be sent to.
SPFMT | FMT
<SCV><SP><SPM><NL>
Setpoint push print operation
(PSHPRNT=ON or WAITSS)
SPFMT | PORT
Defines the communication port that the format will be sent to.
In OIML and CANADA modes, the letters PT (preset tare) are
Note automatically inserted after the printed tare weight.
If the COM port is set to TYPE = RS485, the port will not perform a
demand print. See Technical/Service manual for more information.
Table 4-2. Default Print Formats
4.3 Stream Serial Data Format
If stream data transmission is configured for the COM and/or USBCOM port
(STRLFT or STRIND), by default the 880 sends data using the Consolidated Controls
serial data format shown in Figure 4-1.
<STX> <POL> <wwwwwww> <UNIT> <G/N> <S> <TERM>
ASCII 02
decimal
<CR> <LF>
or <CR>
G = Gross
N = Net
Polarity:
<space > = Positive
<–> = Negative
Weight: 7 digits, right-justified, dummy
zeroes, decimal point with no leading
zeroes except for leading zero immediately
preceding the decimal point. Leading
zeroes transmitted as spaces.
L = pounds
K = kilograms
T = tons (T, TN )
O = ounces
G = grams
<space > = NONE
Status:
<space> = valid
I = Invalid
M = Motion
O = Over/under range
Z = Center zero
Figure 4-1. Stream Serial Data Format
Note
Format can be changed, see Technical/Service Manual.
The character values can be changes for the stream format tokens, see
Table 6-9 on page 68.
If the COM port is set to TYPE = RS485, the port will not stream data, and
cannot be used in a local/remote application.
Print Formatting 36
5.0
Specifications
Power
Line Voltages
Power Consumption
Input Voltage – 100-240VAC, 9-36VDC
Input Frequency – 47-63Hz
AC: 15 watts
DC: 20 watts
Analog Specifications
Full Scale Input Signal -45 mV to +45 mV
Excitation Voltage
10 VDC ±,
8 x 350 or 16 x 700 load cells
Sense Amplifier
Differential amplifier with
4- and 6-wire sensing
Analog Signal Input Range
-45 mV to 45mV
Analog Signal Sensitivity:
0.3 μV/graduation minimum @ 7.5 Hz
1.0 μV/graduation typical @ 120 Hz
4.0 μV/graduation typical @ 960 Hz
A/D Sample Rate:
7.5 – 960Hz, software selectable
Input Impedance
200 M, typical
Noise (Usable Minimum
LSB)
0.3 V p-p
Internal Resolution
8 000 000 counts @ 23 usable bits, approximate
Display Resolution
100 000 dd
Input Sensitivity
10 nV per internal count
System Linearity
±0.01% of full scale
Temperature
Zero
±150 nV/°C, maximum
Span
±3.5 ppm/°C, maximum
Calibration Method
Software, constants stored in EEPROM
Common Mode
Voltage ±0.8V in unbalanced condition
Common Mode
Rejection120 dB minimum @ 50 or 60 Hz
Input Overload
± 12 V continuous, static discharge protected
EMI/RFI Protection
Signal, excitation, and sense lines protected by capacitor bypass
and filtering elements
Optional Analog
Output
Fully isolated, voltage or current output
Voltage output: 0 –10 VDC
Load resistance:1k minimum
Current output: 4–20 mA
External loop resistance: 500 maximum
Digital I/O
Four I/O Onboard
Primary keys, pseudo functions, batching functions
Serial Communications
RS-232
Full Duplex
37 880 Operator’s Manual
RS-485
USB
Half Duplex
USB Type A Connector 2.0
USB Micro A/B Connector 2.0
EtherNet TCP/IP
EtherNet
Operator Interface
Display
Keyboard
LED, Six 0.56 inch (14 mm),
14 Segment with Decimal or Comma
6-key membrane panel
Environmental
Operating Temperature 14°F to 104°F (–10 to +40°C) (legal-for-trade applications);
14°F to 122°F (–10 to +50°C)
(industrial applications)
Storage Temperature –25 to +70°C
Humidity
0–95% relative humidity
Enclosure
Enclosure Dimensions 6.0 in x 4.0 in x 4.95 in
(152 mm x 102 mm x 126 mm)
Weight
2.5 lb (1.2 kg)
Rating/Material
Display Bezel NEMA 4X, IP69K
O
NA
L CO N F E
R
EN
CE
t
t NA
TI
Certifications and Approvals
S
EI
UR
ES
ON W
G
HT
S AND ME
NTEP
CoC Number: 13-080
Accuracy Class III/IIIL nmax : 10 000
A
Measurement Canada
Approval:
Pending
Accuracy Class
III/IIIHD nmax : 10 000
UL
C
UL
®
File Number:E151461
US
LISTED
File Number: Pending
Specifications 38
880 Limited Warranty
Rice Lake Weighing Systems (RLWS) warrants that all RLWS equipment and
systems properly installed by a Distributor or Original Equipment Manufacturer
(OEM) will operate per written specifications as confirmed by the Distributor/OEM
and accepted by RLWS. All systems and components are warranted against defects in
materials and workmanship for two years.
RLWS warrants that the equipment sold hereunder will conform to the current written
specifications authorized by RLWS. RLWS warrants the equipment against faulty
workmanship and defective materials. If any equipment fails to conform to these
warranties, RLWS will, at its option, repair or replace such goods returned within the
warranty period subject to the following conditions:
• Upon discovery by Buyer of such nonconformity, RLWS will be given
prompt written notice with a detailed explanation of the alleged deficiencies.
• Individual electronic components returned to RLWS for warranty purposes
must be packaged to prevent electrostatic discharge (ESD) damage in
shipment. Packaging requirements are listed in a publication, Protecting
Your Components From Static Damage in Shipment, available from RLWS
Equipment Return Department.
• Examination of such equipment by RLWS confirms that the nonconformity
actually exists, and was not caused by accident, misuse, neglect, alteration,
improper installation, improper repair or improper testing; RLWS shall be
the sole judge of all alleged non-conformities.
• Such equipment has not been modified, altered, or changed by any person
other than RLWS or its duly authorized repair agents.
• RLWS will have a reasonable time to repair or replace the defective
equipment. Buyer is responsible for shipping charges both ways.
• In no event will RLWS be responsible for travel time or on-location repairs,
including assembly or disassembly of equipment, nor will RLWS be liable
for the cost of any repairs made by others.
These warranties exclude all other warranties, expressed or implied, including
without limitation warranties of merchantability or fitness for a particular purpose.
Neither RLWS nor distributor will, in any event, be liable for incidental or
consequential damages.
RLWS and buyer agree that RLWS’s sole and exclusive liability hereunder is limited
to repair or replacement of such goods. In accepting this warranty, the buyer waives
any and all other claims to warranty.
Should the seller be other than RLWS, the buyer agrees to look only to the seller for
warranty claims.
No terms, conditions, understanding, or agreements purporting to modify the terms of
this warranty shall have any legal effect unless made in writing and signed by a
corporate officer of RLWS and the Buyer.
© Rice Lake Weighing Systems, Inc. Rice Lake, WI USA. All Rights Reserved.
RICE LAKE WEIGHING SYSTEMS • 230 WEST COLEMAN STREET • RICE LAKE, WISCONSIN
54868 • USA
39 880 Operator’s Manual
© Rice Lake Weighing Systems
PN 152240
10/31/13
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising