Cost-effective
source/measurement
solutions offer superior
performance and a
best-in-class graphical
user interface
• Integrated voltage/current 4-quadrant precision source and
measurement capabilities for easy and accurate I/V measurement
• Wide coverage up to 210 V, 3 A DC/10.5 A pulse
• 10 fA/100 nV minimum measurement resolution (6½ digits)
• 10 fA/100 nV minimum sourcing resolution (6½ digits)
• The 4.3” front panel color display supports both graphical and
numerical view modes
• High resolution arbitrary waveform generation (AWG) and list sweep
functions (10 μs minimum interval)
• High speed digitizing capability (maximum 100000 points/s sample
rate)
• Free application software to facilitate PC -based instrument control
• IVI-COM drivers, and SCPI supporting conventional SMU command set
for basic compatibility
• LXI Core Conformant, USB2.0, GPIB, LAN and digital I/O interface
Introduction
The Agilent B2900A Series of
Precision Source/Measure Units are
compact and cost-effective bench-top
Source/Measure Units (SMUs) with
the capability to output and measure
both voltage and current. An SMU
combines the capabilities of a current
source, a voltage source, a current
meter and a voltage meter along
with the capability to switch easily
between these various functions
into a single instrument. The Agilent
B2900A series of SMUs provide
best-in-class performance at a lower
price than ever before. They have
broad voltage (210 V) and current (3 A
DC and 10.5 A pulsed) sourcing capability, excellent precision (minimum
10 fA/100 nV sourcing and measuring
resolution ) and high measurement
throughput. They also support an
arbitrary waveform generation function. In addition, the Agilent B2900A
series possess a superior graphical
user interface with various viewing
modes that dramatically improve test
productivity, debug and characterization. The versatile integrated source
and measurement capabilities of the
Agilent B2900 series SMUs make
them an ideal choice for testing
semiconductors, active/passive components and a variety of other devices
and materials. The B2900 series
has a broad application range that
spans from R&D and education uses
to industrial development, test and
manufacturing. Moreover, they work
equally well as either standalone
or system components. The Agilent
B2900A series consists of four models, the B2901A, B2902A, B2911A and
B2912A, differentiated through their
available features (number of digits
displayed, measurement resolution,
minimum timing interval, supported
viewing modes, etc.) and by the
number of SMU channels (one or
two) they contain. This makes it easy
to select the exact price/performance
point to meet your testing needs.
Key Specifications of B2900A Series
Number of
channels
Voltage
Max output
Current
Measurement
B2902A
B2911A
B2912A
1
2
1
2
210 V
210 V
210 V
210 V
DC
3.03 A
3.03 A
3.03 A
3.03 A
Pulsed
10.5 A
10.5 A
10.5 A
10.5 A
31.8 W
31.8 W
31.8 W
31.8 W
Power
Source
B2901A
Max digits
Digits
5½
5½
6½
6½
Min resolution
Voltage
1 μV
1 μV
100 nV
100 nV
Current
1 pA
1 pA
10 fA
10 fA
Max digits
Digits
6½
6½
6½
6½
Min resolution
Voltage
100 nV
100 nV
100 nV
100 nV
Current
100 fA
100 fA
10 fA
10 fA
20 μs
20 μs
10 μs
10 μs
20 μs
(50000 pts/s)
20 μs
(50000 pts/s)
10 μs
(100000 pts/s)
10 μs
(100000 pts/s)
Min programmable interval for List sweep/
AWG waveform
Min trigger interval for digitizing
(Max sample rate)
Single view
View Modes
Dual view
Graph view
Roll view
2
Specifications
Specification conditions
The source and measurement accuracy are specified under the conditions
listed below.
1. Temperature: 23 °C ± 5 °C
2. Humidity: 30 % to 80 % RH
3. After 60 minutes warm-up
4. Ambient temperature change less than ±3 °C after self-calibration execution
5. Calibration period: 1 year
6. Measurement speed: 1 PLC (Power Line Cycle)
Maximum voltage and current
Max Voltage
Max Current
210 V
0.105 A
21 V
1.515 A2
6V
3.03 A2
200 V
1.515 A
6V
10.5 A
1
DC or Pulsed
Pulsed only1
Current [A]
10.5
Current [A]
3.03
1.515
1.515
0.1
-210
-21
-6
Voltage [V]
Voltage [V]
21
-0.1 6
-1.515
210
-200
-6
6
-1.515
200
-3.03
DC or Pulsed1
-10.5
Pulsed only1
1. See “Maximum pulse width and duty cycle” in Pulse Source Supplemental
Characteristics for applicable maximum voltage and current.
2. Max current limitation: For 21 V/1.515 A and 6 V / 3.03 A ranges, total max current is
limited by the table below for using 2 channels. Max current is not limited for using
1 channel only.
Maximum current limitation
Ch1 voltage
Ch2 voltage
Max total current limitation of Ch1 and Ch2
±(0 V< V ≤6 V)
±(0 V< V ≤6 V)
Ch1 current + Ch2 current ≤ 4 A
±(0 V< V ≤6 V)
±(6 V < V ≤21 V)
Ch1 current + Ch2 current x 1.6 ≤ 4 A
±(6 V < V ≤21 V)
±(0 V< V ≤6 V)
Ch1 current + Ch2 current x 0.625 ≤ 2.5 A
±(6 V < V ≤21 V)
±(6 V < V ≤21 V)
Ch1 current + Ch2 current ≤ 2.5 A
3
Source Specifications
Voltage source specifications
Programming Resolution
B2911A/B2912A
Accuracy
(% reading + offset)
Noise
(peak to peak)
0.1 Hz to 10 Hz 1
Max Voltage
(over range)
1 μV
100 nV
±(0.015 % + 225 μV)
≤ 10 μV
±210 mV
±2 V
10 μV
1 μV
±(0.02 % + 350 μV)
≤ 20 μV
±2.1 V
±20 V
100 μV
10 μV
±(0.015 % + 5 mV)
≤ 200 μV
±21 V
±200 V
1 mV
100 μV
±(0.015 % + 50 mV)
≤ 2 mV
±210 V
Range
B2901A/B2902A
±200 mV
1. Supplemental characteristics
Current source specifications
Range
B2901A/B2902A
B2911A/B2912A
Accuracy
(% reading + offset)
±10 nA2
---
10 fA
±(0.10 % + 50 pA)
Noise
(peak to peak)
0.1 Hz to 10 Hz 1
≤ 1 pA
±100 nA
1 pA
100 fA
±(0.06 % + 100 pA)
≤ 2 pA
±105 nA
±1 μA
10 pA
1 pA
±(0.025 % + 500 pA)
≤ 25 pA
±1.05 μA
Programming Resolution
Max Current
(over range)
±10.5 nA
±10 μA
100 pA
10 pA
±(0.025 % + 1.5 nA)
≤ 60 pA
±10.5 μA
±100 μA
1 nA
100 pA
±(0.02 % + 25 nA)
≤ 2 nA
±105 μA
±1 mA
10 nA
1 nA
±(0.02 % + 200 nA)
≤ 6 nA
±1.05 mA
±10 mA
100 nA
10 nA
±(0.02 % + 2.5 μA)
≤ 200 nA
±10.5 mA
±100 mA
1 μA
100 nA
±(0.02 % + 20 μA)
≤ 600 nA
±105 mA
±1 A
10 μA
1 μA
±(0.03 % + 1.5 mA)
≤ 70 μA
±1.05 A
±1.5 A
10 μA
1 μA
±(0.05 % + 3.5 mA)
≤ 100 μA
±1.515 A
±3 A
100 μA
10 μA
±(0.4 % + 7 mA)
≤ 120 μA
±3.03 A
3
±10 A
100 μA
10 μA
4
±(0.4 % + 25 mA)
1.Supplemental characteristics
2.10 nA range is not available for B2901A/B2902A.
3.10 A range is available only for pulse mode, not available for DC mode.
4.Measurement speed: 0.01 PLC
4
±10.5 A
Source Supplemental Characteristics
Temperature coefficient (0 °C to 18 °C and 28 °C to 50 °C):
±(0.1 x Accuracy specification) /°C.
Max output power and source/sink limits:
31.8 W
±6 V@ ±3.03 A, ±21 V@ ±1.515 A, ±210 V@ ±105 mA, four quadrant
source or sink operation.
Current limit/compliance:
Accuracy is same as current source. Minimum value is 1 % of range, or 1 nA
in 10 nA range.
Voltage limit/compliance:
Accuracy is same as voltage source. Minimum value is 1 % of range, or
20 mV in 200 mV range
Over range :
101 % of source range for 1.5 A and 3 A ranges. 105 % of source range other
than 1.5 A and 3 A ranges. No over range for 200 V range with current
exceeding 105 mA pulse only condition.
Over temperature protection:
Output turns off then resets at over temperature sensed internally.
Voltage output settling time:
Time required to reach within 0.1 % of final value at open load condition.
Step is 10 % to 90 % range
200 mV, 2 V Ranges: <50 μs
20 V Range: <110 μs
200 V Range:<700 μs
Slew rate:
≤ 0.36 V / µs, 20 V and 10 mA ranges, 10 MΩ load resistance
Current output settling time:
Time required to reach within 0.1 % (0.3 % for 3 A range)
of final value at short condition. Step is 10 % to 90 % range
10 nA, 100 nA Ranges: <10 ms
1 μA Range: <500 μs
10 μA, 100μA Ranges:<250 μs
1mA to 3 A Ranges:<80 μs
Noise 10 Hz to 20 MHz (V source): 3 mVrms, 20 V range.
V source overshoot:
< ±(0.1 %+10 mV). Step is 10 % to 90 % range, resistive load.
I source overshoot:
< ±0.1 % (<± 0.3 % for 3 A range). Step is 10 % to 90 % range, resistive load.
Voltage source range change overshoot:
≤ 250 mV. 100 kΩ load, 20 MHz bandwidth
Current source range change overshoot:
≤ 250 mV/R load , 20 MHz bandwidth
5
Pulse Source Supplemental Characteristics
Minimum programmable pulse width: 50 μs
Pulse width programming resolution: 1 μs
Pulse width definition:
The time from 10 % leading to 90 % trailing edge as follows.
Pulse width
Peak level
90%
Base level
10%
Maximum pulse width and duty cycle:
Pulsed
DC or
pulsed
Pulsed only
DC
Max
Voltage
Max Peak
Current
Max Base
Current
Pulse Width
Max Duty Cycle
Max
Voltage
Max
Current
210 V
0.105 A
0.105 A
50 μs to 99999.9 s
99.9999 %
210 V
0.105 A
1
1
21 V
1.515 A
1.515 A
50 μs to 99999.9 s
99.9999 %
21 V
1.515 A1
6V
3.03 A1
3.03 A1
50 μs to 99999.9 s
99.9999 %
6V
3.03 A1
200 V
1.515 A
50 mA
50 μs to 2.5 ms
2.5 %
180 V
1.05 A
50 mA
50 μs to 10 ms
2.5 %
6V
10.5 A
0.5 A
50 μs to 1 ms
2.5 %
1. Max current limitation: For 21 V/1.515 A and 6 V / 3.03 A ranges, total max current is limited by the table in page 3 for using
2 channels. Max current is not limited for using 1 channel only.
Minimum pulse width at the given voltage, current and settling conditions:
Source Value
Limit Value
Load
Source Settling
(% of range)
Min Pulse
Width
200 V
1.5 A
200 Ω
0.1 %
1 ms
6V
10.5 A
0.6 Ω
0.1 %
0.2 ms
1.5 A
200 V
65 Ω
0.1 %
2.5 ms
10.5 A
6V
0.5 Ω
0.1 %
0.2 ms
6
Measurement Specifications
Voltage Measurement Specifications
Measurement Resolution
B2901A/B2902A
B2911A/B2912A
Accuracy
(% reading + offset)
±200 mV
100 nV
100 nV
±(0.015 % + 225 μV)
±2 V
1 μV
1 μV
±(0.02 % + 350 μV)
±20 V
10 μV
10 μV
±(0.015 % + 5 mV)
±200 V
100 μV
100 μV
±(0.015 % + 50 mV)
Range
1. Supplemental characteristics
Current Measurement Specifications
Measurement Resolution
Range
B2901A/B2902A
1
B2911A/B2912A
Accuracy
(% reading + offset)
±10 nA
---
10 fA
±(0.10 % + 50 pA)
±100 nA
100 fA
100 fA
±(0.06 % + 100 pA)
±1 μA
1 pA
1 pA
±(0.025 % + 500 pA)
±10 μA
10 pA
10 pA
±(0.025 % + 1.5 nA)
±100 μA
100 pA
100 pA
±(0.02 % + 25 nA)
±1 mA
1 nA
1 nA
±(0.02 % + 200 nA)
±10 mA
10 nA
10 nA
±(0.02 % + 2.5 μA)
±100 mA
100 nA
100 nA
±(0.02 % + 20 μA)
±1 A
1 μA
1 μA
±(0.03 % + 1.5 mA)
±1.5 A
1 μA
1 μA
±(0.05 % + 3.5 mA)
±3 A
10 μA
10 μA
±(0.4 % + 7 mA)
10 μA
10 μA
±(0.4 % + 25 mA)3
2
±10 A
1.10 nA range is not available for B2901A/B2902A.
2.10 A range is available only for pulse mode, not available for DC mode.
3.Measurement speed: 0.01 PLC
7
Measurement Supplemental Characteristics
Temperature coefficient (0 °C to 18 °C and 28 °C to 50 °C):
±(0.1 x Accuracy specification) /°C.
Over range :
102 % of measurement range for 1.5 A and 3 A ranges. 106 % of measure
ment range other than 1.5 A and 3 A ranges.
Voltage measurement range change overshoot:
< 250 mV. 100 kΩ load, 20 MHz bandwidth
Current measurement range change overshoot:
<250 mV/R load , 20 MHz bandwidth
Derating accuracy for measurement speed less than 1 PLC :
Add % of range using the following table for measurement with PLC <1.
Derating accuracy with PLC setting< 1 PLC
Voltage Range
Current Range
0.2 V
2 V to
200 V
10 nA
100 nA
1 μA to
100 mA
1 A to
3A
0.1 PLC
0.01 %
0.01 %
0.1 %
0.01 %
0.01 %
0.01 %
0.01 PLC
0.05 %
0.02 %
1%
0.1 %
0.05 %
0.02 %
0.001
PLC
0.5 %
0.2 %
5%
1%
0.5 %
0.2 %
Triggering and Timing Specifications
Triggering
Digital I/O Trigger in to trigger out1: ≤ 5 μs
Digital I/O Trigger in to source change1: ≤ 5 μs
LXI Trigger in to source change1: Minimum 100 μs, Typical 200 μs, Maximum
Unknown
LXI Trigger in to measurement1: Minimum 100 μs, Typical 200 μs, Maximum
Unknown
Internal event to external LXI Trigger output1: Minimum 100 μs, Typical 200 μs,
Maximum Unknown
LAN[0-7] Event Receiving Latency1: Unknown
LAN[0-7] Event Sending Latency1: Unknown
Minimum trigger interval: 10 μs
1.Supplemental Characteristics
Timer
Trigger timing resolution: 1 μs to 100 ms
Timer accuracy: ±50 ppm
Arm/trigger delay: 0 to 100000 s
Arm/trigger interval:
B2901A/B2902A: 20 μs to 100000 s
B2911A/B2912A: 10 μs to 100000 s
Arm/trigger event: 1 to 100000
Time stamp:
TIMER value automatically saved when each measurement is triggered.
8
Other Supplemental Characteristics
Output Characteristics
Sensing Modes: 2-wire or 4-wire (Remote-sensing) connections
Low terminal connection: Chassis grounded or floating
Output connectors:
Banana jack. Triaxial connections are recommended for sourcing and
measuring less than 1 nA. A banana jack to triaxial adapter is available for
low current measurement.
Output location: Channel 1 at front, and channel 2 at rear.
Maximum load:
Normal mode: 0.01 μF
High Capacitance mode: 50 μF
DC floating voltage: Max ±250 V DC between low force and chassis ground.
Guard offset voltage (V source): <4 mV
Remote sense operation range:
Max voltage between High Force and High Sense = 3 V
Max voltage between Low Force and Low Sense = 3 V
Common mode isolation: >1 GΩ, <4500 pF
Maximum sense lead resistance: 1kΩ for rated accuracy
Sense input impedance: >10 GΩ
High Capacitance Mode
The high capacitance mode permits the measurement of devices with capacitances greater than the normal mode maximum load value of 0.01 μF. In high
capacitance mode the maximum allowed value of the load capacitance is 50 μF.
Voltage output settling time:
Time required to reach within 0.1 % of final value with 4.7 μF capacitive load
on a fixed range at specified current range and limit value.
200 mV, 2 V Ranges: 600 μs, at 1 A limit
20 V Range: 1.5 ms, at 1 A limit
200 V Range: 20 ms, at 100 mA limit
Current measurement settling time:
Time required to reach within 0.1 % of final value after voltage source is
stabilized on fixed range. Vout is 5 V unless noted.
1 μA Range: 230 ms
10 μA, 100 μA Ranges: 23 ms
1 mA, 10 mA Ranges: 0.23 ms
100 mA to 3 A Ranges: 100 μs
Mode change delay:
Delay into High Cap mode:
1 μA Range: 230 ms
10 μA, 100 μA Ranges: 23 ms
1 mA to 3 A Ranges: 1 ms
Delay out of High Cap mode:
All ranges: 10 ms
Noise 10 Hz to 20 MHz (20 V range): 4.5 mVrms
Voltage source range change overshoot (20 V range or below):
< 250 mV, 20 MHz bandwidth
High Capacitance mode working conditions:
V/I mode: Voltage source mode only
Range: Current measurement range is limited to fixed range only. 10 nA and
100 nA ranges are not available.
Current limit: ≥ 1 µA
9
Resistance Measurement
Resistance measurement can be performed under either auto or manual test
conditions. Auto resistance measurement is performed in current source and
voltage measurement mode. The total auto resistance measurement error is
shown in the table below. The total error of a manual resistance measurement
can be calculated using the voltage and current accuracy information as shown
below.
Source I mode, manual ohm measurement (4-wire):
Total error = Vmeas/Isrc = R reading x (gain error % of V range + gain error
% of I range + offset error of I range / Isrc value %) + (offset error of V
range / Isrc value)
Source V mode, manual ohm measurement (4-wire):
Total error = Vsrc/Imeas = 1/ [1/R reading x (gain error % of I range + gain
error % of V range + offset error of V range / Vsrc value %) + (offset error of
I range / Vsrc value)]
Measurement speed = 1 PLC
Applicable for temperature: 23 °C ± 5 °C
Example of total error calculation:
I Source value = 1 mA at 1 mA range
V measure range = 2 V range
Total error (% reading + offset) = (0.02 % + 0.02 % + 200 nA/1 mA) + (350
μV/1 mA) = 0.06 % + 0.35 Ω
Typical Performance of Auto Resistance Measurement (4-wire), 2 V range
Range
Resolution
Test Current
Current
Range
Total Error
(% reading + offset)
2Ω
1 μΩ
1A
1A
0.2 % + 0.00035 Ω
20 Ω
10 μΩ
100 mA
100 mA
0.06 % + 0.0035 Ω
200 Ω
100 μΩ
10 mA
10 mA
0.065 % + 0.035 Ω
2 kΩ
1 mΩ
1 mA
1 mA
0.06 % + 0.35 Ω
20 kΩ
10 mΩ
100 μA
100 μA
0.065 % + 3.5 Ω
200 kΩ
100 mΩ
10 μA
10 μA
0.06 % + 35 Ω
2 MΩ
1Ω
1 μA
1 μA
0.095 % + 350 Ω
20 MΩ
10 Ω
100 nA
100 nA
0.18 % + 3.5 kΩ
200 MΩ
100 Ω
10 nA
100 nA
1.08 % + 35 kΩ
System Speeds
Maximum Sweep Operation Reading Rates (reading/second) for 50 Hz
Measure
Speed
Measure To Measure To Source Measure Source Measure
Memory
GPIB
To Memory
To GPIB
< 0.001 PLC
20000
12500
19500
12500
0.01 PLC
4500
3950
4500
3950
0.1 PLC
500
490
500
490
1 PLC
49
49
49
49
Operation reading rate varies by number of sweep steps. Number of sweep steps is specified from 1 to 2500.
10
General Specifications
Environment
For use in indoor facilities
Temperature Range
Operating: 0 °C to +55 °C
Storage: -30 °C to +70 °C
Humidity Range
Operating: 30 % to 80 % RH, non-condensing
Storage: 10 % to 90 % RH, non-condensing
Altitude
Operating: 0 m to 2000 m
Storage: 0 m to 4600 m
Power Supply
AC voltage: 90 V to 264 V
Line frequency: 47 Hz to 63 Hz
Maximum volt-amps: 250 VA
Cooling
Forced air. Side intake and rear exhaust.
EMC
IEC61326-1/EN61326-1
AS/NZS CISPR 11
KC: RRA Notification amending Radio Waves Act Article 58-2
Safety
IEC61010-1/EN61010-1
CAN/CSA-C22.2 No. 61010-1-04, C/US
Certifications
CE, cCSAus, C-Tick, KC
Dimensions
Without handle & feet: 88 mm H x 213 mm W x 450 mm D
Bench Configuration (with handle & feet): 180 mm H x 260 mm W x 480 mm D
Weight
B2901A/B2911A: 5.0 kg
B2902A/B2912A: 6.4 kg
Operation and Functions
Front Panel Interface
4.3” Color TFT WQVGA (480x272, with LED backlight) provides a graphical user
interface (GUI) with the following features.
– 4-View Modes (Single, Dual, Graph, Roll view)
– Easy-to-use key operations
– Easy access to basic and advanced source/measurement capabilities
Single view
Single view provides basic and advanced settings and display capabilities for the
selected channel from the front panel of the instrument. No additional controller
or software is required.
Dual view
Dual view provides the basic settings and display capabilities for both channels
1 and 2. Up to 6½ digits can be displayed in dual view mode. This mode is available only for the B2902A/B2912A.
11
Operation and Functions cont’d
Graph View
Graph View displays measurement results on XY graphs (such as I-V and I-t/V-t
curves) on up to 2 channels. This is useful for quick evaluation of device characteristics, especially those obtained from sweep measurements.
Roll View
Roll view draws I-t or V-t curves similar to the curves drawn by a strip chart
recorder. Up to 1000 acquired data points can be displayed and updated while
the measurement is still in progress. Roll view’s continuous measurement capability is especially useful for monitoring low frequency phenomena. Roll view is
available only for the B2911A/B2912A.
Easy-to-use Key Operations
– 10-key, assist key and rotary knob for easy operation.
– 10-key enables direct input of alphanumeric values.
– Assist key guides operation on the front panel.
Easy Access to Source/Measurement Capabilities
Settings:
– V/I source mode
– Source value
– Limit (compliance) value
– Sweep parameters
– Pulse parameters
– Measure speed
– Range setting
– Trigger setting
Result Displays:
– Numeric results fully up to 6½ digits both in Single and Dual view modes
– Characterization in Graph View and Roll View
– Data list
Advanced Capabilities:
– Limit testing by user definable limits working with component handlers
– Trace memory for accumulating measurement (or math) result and
collecting their statistics
File Operations (USB memory):
– Save
– System Configuration
– Measurement/Math result
– Limit test result
– Trace data
– Graph screen dump
– Load
– System configuration
– List sweep data
12
Source/Measurement Capabilities
Sweep measurement
Number of steps: 1 to 2500
Sweep mode: linear, logarithmic (log) or list
Sweep direction: single or double
Type: DC, or pulse
Min programmable value to create list sweep waveform:
B2901A/B2902A: Min 20 μs with 1 μs resolution
B2911A/B2912A: Min 10 μs with 1 μs resolution
Digitizing/Sampling measurement
Digitizing/sampling measurement can be performed by appropriate setting of
measure speed and trigger interval.
Min trigger interval (Max speed of measurement):
B2901A/B2902A: 20 μs (50000 points/s)
B2911A/B2912A: 10 μs (100000 points/s)
Data Buffers
Data buffers are array of measurement elements. Each elements hold the
following values.
–Measured voltage
–Measured current
–Measured resistance
–Time stamp
–Measurement status
–Source setting
Max buffer size: 100000 points / channel
Program and Interface Capabilities
Programming Language
The Agilent B2900A series supports SCPI (Standard Commands for
Programmable Instruments).
Default Command Set:
Supports all of the advanced features of Agilent B2900A series.
Conventional Command Set:
Supports industry standard conventional SCPI command set for basic
compatibility
Program Memory
Program memory allows you to store long strings of SCPI command lines once
into the B2900A series’ volatile memory and then recall those strings multiple
times while the program is executing using a single SCPI command. By storing
the command strings in memory, the time that would have been spent sending
those same commands over a communication bus is eliminated. For tests
that utilize lots of repeated code (such as subroutines), program memory can
dramatically reduce test times.
Max number of characters of program name:
32, with alphabets, numbers, hyphens and underscores.
Max memory size: 100 kB (2500 lines typical)
LXI
LXI Core Conformant. The Agilent B2901A/B2902A/B2911A/B2912A follows
specified LAN protocols and adhere to LXI requirements such as a built-in Web
control server and IVI-COM driver.
13
Program and Interface Capabilities cont’d
Ethernet: 10/100Base-T
USB2.0: USB-TMC488 protocol (Rear x 1)
GPIB: IEEE-488.2 compliant
USB file system
USB 2.0 high-speed mass storage (MSC) class device (Front x 1)
Digital I/O interface
Connector: 25-pin female D.
Input/output pins: 14 open drain I/O bits
Absolute max input voltage: 5.25 V
Absolute min input voltage: - 0.25 V
Max logic low input voltage: 0.8 V
Min logic high input voltage: 2.0 V
Max source current: 1 mA, Vout = 0 V
Max sink current: 50 mA, Vout = 5 V
5 V power supply pin:
Limited to 600 mA, solid state fuse protected.
Safety interlock pin:
One active high pin and one active low pin. Activation of both pin enables
output voltage > 42 V.
Maximum Number of Simultaneously Triggered Units (using Digital I/O) 1: 8
1. Supplemental Characteristic
Furnished accessories
Power cable
USB cable
Quick Reference (English)
Product Reference CD-ROM (including PDF manuals, Agilent B2900A Quick I/V
Measurement Software and drivers)
Agilent I/O Library Suite
Software and drivers
Agilent B2900A Quick I/V Measurement Software
The Agilent B2900A series includes PC-based Agilent B2900A Quick I/V
Measurement Software. This powerful software makes it easy to quickly setup
and perform I/V measurements and to display the measurement data in a table
or graph without the need to perform any programming. The software allows
you to control up to four SMU channels over a GPIB or LAN connection, or one
B2900A series unit via a USB connection.
Operating environment:
Operating System:
Windows 7 (64bit/32bit), XP SP3 (32-bit).
Other requirements:
Microsoft .NET framework 4.0 or greater, and IO Libraries 16.0 or greater.
14
Furnished accessories cont’d
Agilent B2900A Graphical Web Interface
The Agilent B2900A Graphical Web Interface is a web browser based instrument
control panel. It enables you to set up and perform a measurement easily and
quickly from a web browser using the Agilent B2900A’s built-in web server. This
allows you to control one Agilent B2900A series unit over a LAN connection.
IVI-C or IVI-COM drivers
Compatible with Windows 7 (64bit/32bit), XP SP2 (32-bit) IO Libraries 16.0 or
greater. Supports Agilent VEE, Microsoft Visual Studio (Visual Basic, Visual C++,
Visual C#), National Instruments LabWindows and LabVIEW.
LabVIEW driver (VI)
National Instruments LabVIEW 7.0 or greater. LabView drivers are available at
NI.COM.
Ordering Information
Model Number
B2901A
Precision Source/Measure Unit, 1ch, 100 fA, 210 V, 3 A DC/10.5 A pulse
B2902A
Precision Source/Measure Unit, 2ch, 100 fA, 210 V, 3 A DC/10.5 A pulse
B2911A
Precision Source/Measure Unit, 1ch, 10 fA, 210 V, 3 A DC/10.5 A pulse
B2912A
Precision Source/Measure Unit, 2ch, 10 fA, 210 V, 3 A DC/10.5 A pulse
Options
Printed Manual (User’s Guide)
ABA
English
ABJ
Japanese
AB0
Traditional Chinese
AB2
Simplified Chinese
Calibration
A6J
ANSI Z540 compliant calibration
UK6
Commercial calibration certificate with test data
Rack Mount Kit
1CM
Rack mount kit
Accessories
Banana to Triaxial Adapter
N1294A-001 Banana - Triax Adapter for 2-wire (non Kelvin) connection
N1294A-002 Banana - Triax Adapter for 4-wire (Kelvin) connection
Interlock cable for 16442B
N1294A-011 Interlock cable for 16442B (1.5 m)
N1294A-012 Interlock cable for 16442B (3.0 m)
BNC Trigger Adapter
N1294A-031 GPIO-BNC Trigger Adapter
Test Fixture
N1295A
Device/Component Test Fixture
For details of differences, refer to the comparison table in the page 2.
15
myAgilent
myAgilent
www.agilent.com/find/myagilent
A personalized view into the information most
relevant to you.
www.lxistandard.org
LAN eXtensions for Instruments puts the
power of Ethernet and the Web inside
your test systems. Agilent is a founding
member of the LXI consortium.
Agilent Channel Partners
www.agilent.com/find/channelpartners
Get the best of both worlds: Agilent’s
measurement expertise and product
breadth, combined with channel partner
convenience.
Three-Year Warranty
www.agilent.com/find/ThreeYearWarranty
Agilent’s combination of product reliability
and three-year warranty coverage is another
way we help you achieve your business goals:
increased confidence in uptime, reduced cost
of ownership and greater convenience.
Agilent Advantage Services
www.agilent.com/find/AdvantageServices
Accurate measurements throughout the
life of your instruments.
Agilent Electronic Measurement Group
DEKRA Certified
ISO 9001:2008
Quality Management Sys
System
www.agilent.com/quality
If you need more source capabilities
or lower noise sourcing, the best
choice is Agilent Power Source
products.
Agilent B2961A/B2962A 6.5 Digit Low
Noise Power Source
www.agilent.com/find/
precisionSOURCE
Agilent B1505A Power Device
Analyzer/Curve Tracer (1500 A/10 kV)
www.agilent.com/find/B1505A
(BP-3-1-13)
Product specifications and descriptions
in this document subject to change
without notice.
© Agilent Technologies, Inc. 2013
Published in USA, April 25, 2013
5990-7009EN
Agilent B1500A Semiconductor
Device Analyzer
www.agilent.com/find/B1500A
DISTRAME S.A. - Parc du Grand Troyes - Quartier Europe Centrale - 40, rue de Vienne - 10300 SAINTE-SAVINE
Tél. : +33 (0)3 25 71 25 83 - Fax : +33 (0)3 25 71 28 98 - E-mail : infos@distrame.fr - Site internet : www.distrame.fr
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising