iLog Thermocouple Data Logger

iLog Thermocouple Data Logger
407 – 15216 North Bluff Road
White Rock, BC
Canada V4B 0A7
Phone: 604.424.9092
Toll Free: 1.877.352.9158
Fax: 1.877.453.0658
Email: info@microedgeinstruments.com
Web: www.microedgeinstruments.com
iLog Thermocouple Data Logger
OVERVIEW
The iLog Thermocouple Data Logger is low
cost, high accuracy, battery powered, standalone internal temperature + external
thermocouple data logger. The logger records
up to 4 mega-byte of data and stores them in
non-volatile flash memory for later retrieval.
iLog Thermocouple Data Logger measures
wide range of temperatures and works with
all types of thermocouples (B, E, J, K, N, R,
S, T).
Its on-board temperature channel provides
environment monitoring and Cold-Junction
compensation.
Its aluminum enclosure makes it excellent in
the harshest industrial environment. Plug &
Play USB port and versatile custom equation
iLog Thermocouple Data Logger Specifications
Specifications
simplify communications and engineering
unit conversion.
16-bit ADC makes it well suited for science
and laboratory applications where precise and
accurate measurements are critical.
Simply plug the logger to computer's USB
port, and the software automatically
recognizes it and handles the configuration,
downloading, graph viewing and more...
FEATURES
High Data Resolution:
The 16-bit analog-to-digital converter meets
most high-resolution requirements.
Large Memory Size:
The 4-Mega-Byte Memory stores years of
measurements.
Page 1 of 5
Free Upgrade of Software:
Both firmware and PC software are free for
upgrade by couple of mouse clicks, keeping
the logger and software always up-to-date
with new features and bug fixes.
Multiple Communication Ports:
The iLog data loggers can be accessed via
USB, Serial Port, MODEM, or Ethernet
connections with auto baud rate of up to 115
kbps.
The iLog data logger notifies the alarm
condition over alarm terminal strips or
communication lines. (USB, Serial Port,
MODEM)
Excitation control turns on the power of
external transmitter/transducer only when the
logger is sampling.
Alarm & Excitation Control
Its on-board TTL serial port and USB
interfaces meet most communication
requirements.
TTL Serial Port
USB Port
Rugged Physical Design:
The rugged aluminum enclosure makes the
iLog data loggers perfect in the harshest
industrial environment.
12-Year Battery Life:
The internal lithium battery provides over 12
years of instantaneous logging operation
when sampling at an interval of one minute.
Aluminum Case
Fast Sampling Mode:
The iLog data loggers can log data with the
sampling interval as fast as 20 milliseconds,
replacing data acquisition devices.
Alarm and Excitation Output:
SITEVIEW SOFTWARE FEATURES
SiteView is a Windows-based application which works with iLog Series data loggers for
downloading, configuration, data analyzing and plotting.
Its user-friendly graphic interface plus powerful functionalities fit both novice and advanced users.
iLog Thermocouple Data Logger Specifications
Page 2 of 5
Besides basic data logger configuration, downloading, SiteView includes other powerful features
like:
 Multiple Communications Interfaces: USB, Serial Port, Ethernet…
 Custom Equation Editor with C# language meet any complicated measurement
requirements
 Multiple Data Views and Custom Formatting of Axis, Line, Annotation & Comments.
 Real-Time Chart View/Recording and User Calibration
 Export to CSV, TEXT, BMP, JPG, TIF, PNG, GIF file format.
iLog Thermocouple Data Logger Specifications
Page 3 of 5
SPECIFICATIONS
Product Identification
Product Name
Model
Inputs
Connections
Channels
Resolution
Accuracy
Input Impedance
Over-voltage protection
Alarms
Channel Alarms
Alarm Outputs
Alarm-On Delay:
Alarm Indicator
On-board Memory
Capacity
Data Retention
Sampling & Logging
Sampling Interval
Logging Mode
Logging Activation
Communications
Interface
Baud Rate
iLog Thermocouple Data Logger
iTC-80
Pluggable terminal block for thermocouple input, alarm and
excitation control outputs.
One on-board thermistor temperature (-40°C ~ 70°C, -40°F ~
158°F)
One external thermocouple channel:(-8mV ~ +73mV) for type
E,J,K,N T Thermocouple
0.0018%
Thermistor:± 0.2°C(0°C ~ 70°C, 32°F ~ 158°F)
Thermocouple: ± 0.1% ~ 0.2% + T/C Accuracy
>1 MOhms
up to -3VDC ~ +40VDC
Two editable alarm thresholds per channel.
ALARM1 & A2/EXT terminal strips can be configured as alarm
outputs.
Alarm-On: MOSFET(N-Channel) switch on.
Alarm-Off: MOSFET(N-Channel) switch off.
Max Power: 200mA @ 24VDC.
With purchase of SiteView software, the Site-Log can report alarm
status to host PC via USB, Modem or Ethernet Device Server.
Programmable 0 - 10 minutes delay with 1-minute increments.
On-board LED lights in red when in alarm condition.
4 Mega bytes (2 Mega measurements).
Over 20 years.
20 milliseconds to 12 hours user selectable. (external power
supply required if interval less than one second)
Stop recording or FIFO when memory is full.
Programmable instant, start delay or field push-button activation.
USB(USB cable included).
AUX(RJ11) for direct TTL level communications.
With purchase of Site USB DeviceServer, the Site-Log logger can
be connected to Ethernet for remote access.
Auto-detect baud rate from 2400 to 115200 bps on both USB and
AUX ports.
Battery
Power
Life Cycle
Software
Site View (Sold Separately)
Built-in 3.6V Lithium Battery.
12 years based on 1 minute sampling interval
Configuration, downloading, plotting, real-time view, custom
calibration and custom equation.
iLog Thermocouple Data Logger Specifications
Page 4 of 5
Software Requirements
Physical
Material
Dimension
Weight
Mounting
Others
LED Indicator
Computer with 1.0 GHz or faster processor
256 MB Memory or higher
1.0 GB of available hard-drive space or higher
Windows XP with SP2 or later, Vista, Window 7
At least one USB port or one COM port
Aluminum enclosure.
88 X 64.2 X 24 mm (3.46 X 2.53 X 0.95 inches)
200g.
Probe/Wall-mount holes for hanging/mounting.
Excitation Control
Operating Environment
Clock Accuracy
Approvals
Tri-Color LED: (can be disabled for power saving)
Normal Sampling: green when sampling
Alarm: red when sampling
Low Battery: amber when sampling.
A2/EXT terminal strip can be configured as excitation control
output for driving the power of connected devices.
Warm-up delay Interval settings: 10 to 240 seconds with 10second increments.
-40 ~ +70°C (-40°F ~ 158°F), 0~95%RH non-condensing.
± 1 minute per month.
CE, FCC
LOGGING CAPACITY TABLE
Sampling Interval
1 minute
10 seconds
Logging Capacity
727 days
121 days
Sampling Interval
1 second
100 ms
Logging Capacity
12 days
28 hours
ORDERING INFORMATION
Model
iTC-80
Description
iLog Thermocouple Data Logger
iLog Thermocouple Data Logger Specifications
Page 5 of 5
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising