2010 bc.cdr - nice Neotech
Efficient
enlighten engineering
2010 BC
Infant Radiant Warmer
Features:
? Micro Controller based Electronics
? Controls: Servo Skin/Manual and Apgar
? Pre-warm: Use the equipment with 30% Heater Output
in Manual Mode.
? No need of calibration required for temperature probe.
? High Intensity Quartz Heating element for effective Radiation
(Ceramic heater Optional)
? Halogen Observation Lamp
? The Heater Module rotates 90° to both sides to facilitate
X-Ray Procedures.
? The Parabolic Reflector is designed to send uniform heat to
the bed surface.
? Memory Backup to restore set temperature and previous mode,
in case of power failure.
? Optimum Bed dimensions ensure optimum work space for
the nursing staff.
? Comprehensive Alarms and safety cut-offs.
? Programmable Mute time.
? Alert (Alarm) when sensor is detached from the patient's skin
(optional)
TECHNICAL SPECIFICATIONS
? Instant Warming and Uniform heating.
Controller
Electronics
Mode of operation
Skin mode control range
Manual mode timer
:
:
:
:
Apgar timer
:
Percentage of heater output
Memory back up
:
:
Programmable mute time
Safety heater cut off
:
:
Over temperature heater cut off
:
Temperature sensor (probe)
Accuracy of skin sensor
Accuracy of skin probe
Interchangeability
Calibration of skin sensor (probe)
:
:
Heater element
:
Power rating
Maximum heater wattage
Maximum power requirement
Examination light
X-ray facility
Mattress tilt
IV pole
Pump mounting pole
Castor
:
:
:
:
:
:
:
:
:
Monitor shelf
:
A15898
ISO 9001:2000
&
ISO 13485:2003
nice NeÖtech
Chennai
:
:
High speed micro controller
Servo skin, manual & apgar
o
o
30.0 C to 38.0 C
1 min to 15 mins (selectable)
15 mins to 30 mins (optional)
Alerts at 1, 5, 10, 15 & 20 mins
(selectable)
10% to 100% in 10% increments
Equipment restores the previous
settings
User can select the mute time
If the skin sensor is removed from
the baby's skin
o
If the skin temperature is >39 C in
any skin/manual mode
Thermistor sensor
± 0.1oC
±0.2 C
No need of calibration after
Changing new probe
Quartz Infrared heater
Ceramic heater (optional)
~230V AC, 50Hz, 700 W
600 Watts
700 Watts
12V, 50W Halogen lamp
Underneath the bed
± 30 degrees on both sides
Maximum load 2.2 Kg
Maximum load 2.2 Kg
4” castors 4 Nos. (2 with brake & 2
without brake
2 Nos. Fixed
o
Height adjustable heater module
Cabinet
Operating noise level
:
:
:
Available only in nice 2010
Available only in nice 2010BC
<50 dBA
Controller Alarms:
Check patient
:
Skin temperature high
Skin temperature low
Skin temperature probe fail
Over temperature
Heater fail
System fail
Power fail
:
:
:
:
:
:
:
If the skin sensor is removed from
the baby's skin
o
> 1.0 C from the set temperature
o
< 1.0 C from the set temperature
Skin Probe short or open
o
If skin temperature is >39 C
Heater defect
Electronic component defect
AC power interruption
Physical characteristics
Equipment
Weight
Mattress
Mattress from floor
Maximum load
Side panels
Maximum load for monitor shelf
: 85 cms(L) x70 cms(W)x185 cms(H)
: 70 Kgs (nice 2010BC)
: 73 cms (L) x 53 cms (W) x 5 cms (H)
: 95 cms
: 10 Kgs
: Aluminum hinged acrylic panels
: 4.5 Kgs
Heater module swivel
:
Environmental
Operating temperature
Storage temperature
Operating humidity
Storage humidity
Altitude
Quality/Test Approval
Quality
Electrical & product safety
±90°
0
o
: 15.0 C to 35.0 C
0
o
: -10.0 C to 60.0 C
: 15% to 90% RH, non condensing
: 15% to 90% RH, non condensing
: Sea level to 1-9 miles, 3 Km
:
:
:
ISO 9001:2000 & ISO 13485:2003
IEC 60601-1, IEC 60601-1-2
IEC 60601-1-4, IEC 60601-2-21
74-00-004, Rev 00
nice 2010 BC Infant Warmer with Bed & Storage Cabinet, nice 2010 B Infant Warmer with
Bed & Storage Tray, nice 2010 Mobile Free Standing Infant Warmer with Height Adjustment.
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising