Model T300 Gas Filter Correlation CO Analyzer

Model T300 Gas Filter Correlation CO Analyzer
Model T300
Gas Filter Correlation CO Analyzer
The Model T300 measures low ranges of carbon monoxide by
comparing infrared energy absorbed by a sample to that absorbed
by a reference gas according to the Beer-Lambert law.
Using a Gas Filter Correlation Wheel, a high energy IR light source is
alternately passed through a CO filled chamber and a chamber with
no CO present. The light path then travels through the sample cell,
which has a folded path of 14 meters. The energy loss through the
sample cell is compared with the span reference signal provided by
the filter wheel to produce a signal proportional to concentration,
with little effect from interfering gases within the sample. This
design produces excellent zero and span stability and high signal to
Ranges: 0-1 ppm to 0-1000 ppm,
user selectable
Dual ranges and auto ranging
Large, vivid, and durable color
graphics display with touch screen
interface
Ethernet, RS-232, and (optional)
USB com ports
Front panel USB connections for
peripheral devices and firmware
upgrades
8 analog inputs (optional)
noise ratio, allowing extreme sensitivity.
Adaptive signal filtering optimizes
response time
All T Series instruments offer an advanced color display, capacitive
Temperature & pressure
compensation
touch screen, intuitive user interface, flexible I/0, and built-in data
acquisition capability. All instrument set up, control and access to
stored data and diagnostic information is available through the front
panel, or via RS232, Ethernet, or USB com ports either locally or by
remote connection using the included APIcom™ software.
Comprehensive internal data
logging with programmable
averaging periods
Ability to log virtually any operating
parameter
Two-year warranty
Five-year warranty on GFC wheel
Free Customer Support by telephone and email for the life of the instrument
Model T300
Gas Filter Correlation CO Analyzer
Specifications
How to Order
Model T300 includes:
General
Ranges:
Min: 0-1 ppm Full scale
Max: 0-1,000 ppm Full scale (selectable,
dual ranges and auto ranging supported)
ppb, ppm, μg/m3, mg/m3 (selectable)
< 0.02 ppm (RMS)
< 0.5% of reading (RMS) above 5 ppm
0.04 ppm
< 0.1 ppm/24 hours
< 0.5% of reading/24 hours
10 seconds
< 60 seconds to 95%
1% of full scale
0.5% of reading
800 cm3/min ±10%
Measurement Units:
Zero Noise:
Span Noise:
Lower Detectable Limit:
Zero Drift:
Span Drift:
Lag Time:
Rise and Fall Time:
Linearity:
Precision:
Sample Flow Rate:
Electrical Specifications
Power Requirements:
Analog Output Ranges:
Recorder Offset:
100V-120V, 220V-240V, 50/60 Hz
10V, 5V, 1V, 0.1V (selectable)
±10%
1 x Ethernet: 10/100Base-T
2 x RS232 (300-115,200 baud)
2 x USB device ports
8 x opto-isolated digital outputs
6 x opto-isolated digital inputs
4 x analog outputs
1 x USB com port
1 x RS485
8 x analog inputs (0-10V, 12-bit)
4 x digital alarm outputs
Multidrop RS232
3 x 4-20mA current outputs
Optional I/O:
Operating Temperature Range:
Dimensions (HxWxD):
Weight:
5 - 40ºC
7” x 17” x 23.5” (178 x 432 x 597 mm)
40 lbs (18 kg)
50Hz
220V - 240V 60Hz
Select DC output voltage
10V
5V
1V
0.1V
Ambient zero and ambient span
Ambient zero and pressurized
span
Zero scrubber and pressurized
span
Zero scrubber and ambient span
Mounting Options:
Rack mount brackets with chassis
slides
Rack mount brackets only
Handle
I/O Options:
Physical Specifications
100V - 120V
Calibration Options:
Communication Specifications
Included I/O:
Two year warranty
Internal pump or external pump
(optional)
Dual ranges and auto ranging
47mm diameter particulate filter
8 isolated digital outputs
6 isolated digital inputs
RS-232 ports
Ethernet port
USB ports for peripheral devices
APIcom™ remote control
software
Select AC input voltage
4-20mA outputs (up to three
channels)
USB com port
8 Analog Inputs
Multi-drop RS232
RS485
Other Options:
Paramagnetic O2 Sensor
CO2 Sensor
Concentration alarm relays
Expendables kit
Certifications
US EPA:
MCERTS:
RFCA-1093-093
Sira MC 050069/04
The values expressed above are in accordance with EPA definitions.
All error specifications are based on constant conditions.
Specifications exceed US EPA and Eignungsgeprüft requirements.
Specifications subject to change without notice.
Printed documents are uncontrolled. SAL000053B (DCN 5880) T300/10.13.10
For more information about the Teledyne API
family of monitoring instrumentation products,
call us or visit our website at
9480 Carroll Park Drive
Ph. 858-657-9800
San Diego, CA 92121-5201
Fax 858-657-9816
Email api-sales@teledyne.com
w w w. t e l e d y n e - a p i . c o m
© 2010 Teledyne - Advanced Pollution Instrumentation, Inc.
8
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising