Pulse Oximeter INV-430J/PE

Pulse Oximeter INV-430J/PE
Warning: Do not use the Oximeter in an MRI or CT environment.
This device is intended for non-medical use in healthy people to monitor
Warning: Do not operate the Oximeter if it is wet. Avoid moving the
their pulse and blood oxygen levels for sports and/or aviation only. Do not
oximeter from a cold to a hot and humid environment.
use for continuous monitoring.
Warning: Install the batteries properly before powering on the Oximeter
for normal use. Please remove the batteries if you are not planning to
Precautions
 Do not attempt to repair the Pulse Oximeter by yourself. Only
certified engineers should maintain and repair it.
 Change the contact position between the Oximeter probe and
the finger periodically if you are monitoring your SpO2 levels
and pulse rate for more than 2 hours.
 This product is not designed to be used by newborn babies.
 Seek for medical care if the measured value goes beyond the
normal range and you are sure that the instrument is not
malfunctioning.
 The pulse oximeter uses infrared light (invisible to your eyes) to
measure your SpO2 levels. Hence, please do not stare at the
light-emitting components of the Oximeter, as that could cause
harm and/or potentially blind your eyes.
 For details about clinical limitations and contraindications,
please carefully consult relevant medical literatures.
Warning: Close the battery cover when the instrument is in use.
Symbol Conventions
Symbol
Spot check pulse oximeter takes 1-2 reading every second and takes an
average of 3-6 readings before displaying the result. Hence, you will
use the Oximeter for a long time.
Pulse Oximeter INV-430J/PE
Limitations
have to wait at least 3-6 seconds before a result will be displayed on the
monitor. If the pulse oximeter fails to detect blood flow for the first few
seconds, then the results will be delayed accordingly.
The pulse oximeter does not work for people with naturally small fingers
Description
or low peripheral blood flow. This is because the pulse oximeter would
BF-type application part
not have enough blood to take a measurement. Please note that prolong
Caution: Please see this manual.
use of a pulse oximeter, hypotension, vasoconstriction, hypothermia and
%SpO2
Symbol of oxygen saturation
certain medications can lead to low or restricted blood flow. Nail polish
bpmPR
Symbol of pulse rate
can also impede the pulse oximeter from taking a reading.
Finally, it is generally accepted that the saturation percentage is
Temperature limitation
unreliable on the steep part (around 60 mm Hg) of the oxyhemoglobin
When end users abandon this product, they
must send the product to the collection place
for recycling.
dissociation curve.
Innovative features of the INV-430J/PE
The INV-430J/PE attempts to ameliorate some of the above problems by
1) Increasing the sensitivity of the sensor so that it can measure the
Manufacturer information
user’s SpO2 and Pulse Rate (PR) at blood perfusion Index (PI) as low as
Overview
0.2%.
SpO2 stands for peripheral capillary oxygen saturation. Oxygen
2) Comes with anti-motion technology so that the pulse oximeters can
The following factors may affect the accuracy of measurement:
saturation is defined as the ratio of oxyhemoglobin (HbO2) to the total
still function even when the finger moves.
◆ The Oximeter is used in an environment involving
concentration of hemoglobin (i.e. Oxyhemoglobin + reduced hemoglobin)
3) Using a self-adjusting smart spring system to fit the finger snugly but
high-frequency devices, such as high-frequency electric knives
present in the blood. It is an important physiological parameter involved
not so tight as to impede blood flow that might lead to a low PI.
and CT apparatuses.
in respiration and circulation. The Pulse Oximeter feature herein is small,
4) Using soft, hypoallergenic medical grade silicon in the finger chamber
portable, non-invasive and easy to use. The user only needs to insert a
so that the pulse oximeter will not cause discomfort.
finger into the chamber to measure his/her SpO2 and pulse rate.
4) Using solid ABS plastic to block ambient light from reaching the sensor
Working Principles, Expected Usage, and Applicable Scope
so that the signal to background ratio and the accuracy of the
Oxygenated blood absorbs light at 660nm (red light), whereas
measurement is dramatically increased.
deoxygenated blood absorbs light preferentially at 905nm (infra-red). A
5) Having a plethysmograph that allows the user to see and gauge the
pulse oximeter works by passing a beam of red
amount of blood detected by the pulse oximeter. This will allow the user
◆ The probe of the Oximeter is placed on the same arm that a
blood pressure cuff arterial duct or intravenous injection.
◆ The user suffers from hypotension, severe vascular atrophy,
severe anemia, or low oxygen.
◆ The user is in sudden cardiac arrest or shock state.
◆ The user is wearing nail polish or artificial nails.
Warning: Do not use the Oximeter in an environment with any
and infrared light through a pulsating capillary
Light beams
sensor
to know if the finger is properly inserted into the finger chamber and if the
flammable gases, flammable anesthetic, or other flammable substances.
bed and then measure the amount of red and
Warning: Keep unit and lanyard away from children as the included
infrared light emerging from the tissues via a
lanyard may present an entanglement or choking hazard to small
sensor. To improve accuracy, the NV-430J/PE
children. Adult supervision required; never leave children unattended
uses a proprietary algorithm to collect data from pulsatile arterial blood
Operation Guide
with unit or lanyard.
and ignores local noise from the tissues. The relative absorption of light
Switch on the power supply of the pulse Oximeter
Warning: Do not throw the batteries into fire, as that could cause an
by oxyhemoglobin (HbO) and deoxyhemoglobin is then processed
Stick one finger completely into the
explosion.
according to the Beer-Lambert's law and a qualitative measurement of
finger chamber of the Oximeter. The
Warning: Do not attempt to charge the included batteries, as that could
the users’ oxyhemoglobin status i.e. Oxygen saturation level (SpO2)
fingernail should be facing upward.
derived.
Release the clip and press the power
cause leakage, fire disaster, or even explosion. Dispose the used
batteries in accordance to the local laws and regulations.
pulse oximeter is taking the measurements properly (See below for
sensor
details)
key to power on the Pulse Oximeter.
The INV-430J/PE pulse oximeter
comes with anti-motion technology so that the pulse oximeter can
function even if there is finger movement. However, for best results, try to
keep your finger still during measurement. It is also not advisable to use
this instrument during sports activities as movement may lead to
inaccuracies. Once the reading stabilizes, read the measured values of
oxygen saturation and pulse rate on the OLED screen. The Oximeter will
battery faster.
V2.01.E
Alm setup
Alm
Beep
Demo
Restore
Brightness
Exit
Move "*" to the Exit option, and hold the function key to return to the
*
on
off
off
ok
4
V2.01.E
Sounds Setup *
SpO2 Alm Hi
100
SpO2 Alm Lo
94
PR Alm Hi
130
PR Alm Lo
50
+/+
Exit
monitoring interface.
About the Display
PRESS (<0.5 sec) the function key in measurement mode to change the
display mode. The six different modes are shown in the figure to the right.
Choose the mode that is the most convenient for you.
automatically shut down 10 seconds after you remove your finger.
Interface 1
If you do not insert your finger completely into the chamber
Interface 2
Plethysmograph
the measurement will be inaccurate.
Replace the batteries when the batteries are low and the
symbol
flickers on the screen.
Plethysmograph
Schematic Diagram of Display
The following figure shows the information display on the OLED screen
The INV-430J/PE comes with two invaluable features, the
of the Oximeter in normal detection state:
plethysmograph and the perfusion index, that help the user determine if
Symbol of oxygen saturation
the readings are accurate and reliable. The plethysmograph indicates the
Measured value of oxygen saturation
amount of blood flow detected by the pulse oximeter and the perfusion
Bar graph
index (PI) indicates the strength of your pulse. Each wave in the
Plethysmograph
plethysmograph (see above) corresponds to a heartbeat and the wave
Measured value of pulse rate
amplitude corresponds to the amount of blood detected by the pulse
PI
oximeter flowing through the blood vessels. The pulse oximeter is
Symbol of pulse rate
optimized when the height of the wave amplitude is consistent
Battery power indication
throughout as shown in the figure above. That is when you should take
Power-On Key/Function Key Operations
the reading. The pulse oximeter can function with a PI reading as low as
PRESS (<0.5 sec) the power-on/function key to turn on the pulse
0.2. However, if your PI is below 0.2%, this means that your blood
oximeter. Once it is turned on, HOLD (>0.5 sec) the power-on/function
perfusion is too low for a reliable read. Please warm your hands to
key to enter into the menu interface. PRESSING (<0.5 sec) on the
increase blood flow and retake your measurement to make sure that your
power-on/function key will allow you to to scroll through options while
finger is positioned right under the LED lights. Most of the time, a higher
HOLDING (>0.5 sec) it will select an item. "*" indicates the option you are
PI will give you a more reliable reading.
currently at (Please see figure below).
Alarm Range Setting
Alarm Sound Setting
To change the alarm range settings, HOLD the power-on/function key to
To turn the alarm on/off, PRESS the power-on key/function key to move
enter Menu Interface 1.With the "*" next to Alm Setup, HOLD the
Cleaning
To clean the instrument, power off and remove the batteries first. Then
clean the outer surface of the instrument (including the LED screen)
using a piece of dry soft cloth dipped with 75% medical alcohol. Do not
immerse the unit in alcohol.
Caution: Do not use any strong dissolving agent such as acetone.
Caution: Do not rub the body of the instrument using materials such as
steel wire balls or polished metal objects.
Caution: Ensure that there is no washing liquid on the surface of the
instrument.
Caution: Do not allow liquid to flow into the instrument during cleaning.
Caution: Do not immerse any part of the instrument into any liquid.
Disinfection
Before measurement with the instrument, wipe the rubber finger pad
inside the chamber using a piece of dry soft cloth dipped with 75%
medical alcohol. Clean the finger to be measured using the medical
alcohol for disinfection purposes before and after use.
Do not disinfect the instrument by using
high-temperature/high-pressure disinfecting gas.
"*" to Alm. HOLD the power-on key/functional key to switch the option
power-on key/function key to enter in Menu Interface 2. "*" should be at
from on to off or vice versa. When Alm is set to on and the measured
Sounds Setup. Press the power-on key/function key to move "*" to the
Maintenance
option you desire to change. SpO2 Alm Hi and SpO2 Alm Lo refer to the

values of the blood oxygen saturation (SpO2) and/or pulse rate (PR) go
beyond the stated upper or lower limit (See Alarm Range Setting to set
parameters for Alarm), the alarm will go off. When Alm is set to off, the
alarm is deactivated. The alarm will not go off regardless of your SpO2
and/or PR values.
upper and lower limit of the SpO2 levels that you wish the alarm to go off
at respectively. PR Alm Hi and PR Alm Lo refer to the upper and lower
limit of the Pulse Rate levels that you wish the alarm to go off at
Remove the batteries from the battery slot and properly store them
if you do not plan to use the Oximeter for a long period of time.

Replace the batteries if they are low on power.

Clean the Oximeter and the fingertip before use to ensure
accurate reading.
Brightness Setting
respectively.
To change brightness, PRESS the power-on key/function key to move "*"
While the "*" symbol stays on the +/- option, hold the function key to set
to brightness. HOLD the power-on key/function key to set the
the option to + or -. In + mode, select the corresponding upper or lower

Periodically check the Oximeter for damage.
brightness. There are 5 brightness settings. 5 is the brightest. The factory
limit option and hold the function key to increase the upper or lower limit;

Avoid using the Oximeter in an environment with flammable gases
default setting is 4. Please note that a brighter setting will drain the
in - mode, hold the function key to decrease the upper or lower limit.

Store the Oximeter between 14 and 122F (–10 to +50C) and at
humidity levels no greater than 93%.
or using it in an environment where the temperature or humidity is
excessively high or low.

Waterproof grade: IP22
Check the accuracy of the oxygen saturation and pulse rate
readings by using an appropriate calibration apparatus.
6.4 Storage and Transportation
Technical Specifications
1.
Dimensions: 62.0 mm (Width) × 37.0 mm (Depth) × 32.0 mm
(Height)
Weight: 42.5 g (including the height of the two AAA dry batteries)
2.
Peak wavelength range of the light emitted from the probe: red light
660 nm ± 3; infrared light 905 nm ± 5.
3.
Maximum optical output power of the probe: 1.2 mW for infrared light
(905 nm).
4. Normal working condition
Working Temperature
5C to 40C (41F to 104F)
Relative Humidity
15% to 80%, non-condensing
Atmospheric Pressure
70 kPa to 106 kPa
Rated Voltage
DC 3.0 V
5. Default values and conditions of alert
Parameter
Value
Upper limit: 100
Oxygen saturation
Lower limit: 94
Upper limit: 130
Pulse rate
Lower limit: 50
When the alert switch is on and the
actual measured value goes beyond
Alert condition
the preset alert parameter range, the
Oximeter gives an alert sound.
6. Technical parameters
Parameter
Value
Oxygen
35% to 100%
saturation
Display range
Pulse rate
25 bpm to 250 bpm
Oxygen
1%
saturation
Resolution
Pulse rate
1 bpm
Oxygen
±2% (70% to 100%)
Measurement
saturation
No requirement (≤ 69%)
precision
Pulse rate
±2 bpm
Oxygen
Upper limit: 50% to 100%
saturation
Lower limit: 50% to 100%
Alert range
Upper limit: 25 bpm to 250 bpm
Pulse rate
Lower limit: 25 bpm to 250 bpm
Oxygen
± 1% of the preset value
saturation
Alert error
The greater of ±10% of the preset
Pulse rate
value and ±5 bpm
PI
Weak PI
Min. 0.2%
Safety Type
Anti-electric-shock type: internal power supply device
Anti-electric-shock degree: BF-type application part
Running mode: continuous working
Innovo Groups
Phone: +1-858-888-9781
cs@innovogroups.com
http://www.innovogroups.com
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising