Time Electronics - Caltest Instruments Ltd

Time Electronics - Caltest Instruments Ltd
Time Electronics
5030 Electrical Tester Calibrator
• RCD 3mA to 2500mA, 10ms to 2000ms
• Loop 50mΩ to 1.8kΩ
• Insulation up to 2GΩ / 1kV
• Continuity 0.1Ω to 10kΩ
• RS232 / USB Control
• Fast and intuitive user interface
• PC/laptop control via EasyCal software
DESCRIPTION
A precision instrument designed to calibrate RCD, loop, insulation, earth testers and multifunction installation testers. The 5030
accurately simulates RCD trip times and measures currents produced by RCD testers. It replicates loop impedance and auto
adjusts for local line impedance. It also provides insulation resistances and measures test voltages and currents.
Simple operation and automated calibration
User control is via the central navigation keypad. Scrolling though
menus and settings is intuitive and easy, with measurements and
settings shown on the large clear LCD display.
Alternatively connect the 5030 to a PC/laptop (via RS232 or USB)
installed with Time Electronics’ EasyCal software and automate the
calibration process. Increase speed of calibration and consistency of
results. Easily produce calibration certificates and reports to ISO 9001,
ISO 17025, and other international quality standards.
Regional Test Sockets and Fault Detection
For loop and RCD tests the electrical tester (device under test) connects directly to the dedicated front panel mains socket. The
5030 can be fitted with a number of regional type sockets (specified on order). In these test conditions, if the device under test is
faulty, the 5030 auto detects the fault, disconnects the output and warns the user. The 5030 is designed not to trip any RCDs on the
local supply.
Loop Impedance with Auto Local Loop Compensation
With 10 measurement points the 5030 covers a wide range of loop testers. The precision resistors that make up the loop calibration
function are high power and capable of withstanding up to 40A. An accurate automatic measurement of the local loop is made by
the 5030 and added to the resistor value to give the loop impedance value, allowing for precise loop impedance calibration.
Precise RCD trip times
Set between 10 and 2000ms, with an accuracy of 0.1ms. Predefined values are available for quick selection via the front panel.
Alternatively any value, with 1ms resolution, can be set via EasyCal calibration software. To avoid false trip times the 5030 incorporates
a ‘pre-test trigger delay’ setting. This feature is for use with RCD testers that produce a pre-test signal. A test current threshold setting
(0 to 100% of nominal current) is also user selectable.
RCD Current measurements
RCD current measurements are true RMS for AC, half wave rectified and DC. The ranges covered are from 6mA to 1000mA, with
multipliers of x0.5, x1, x2, and x5 up to a maximum of 2500mA.
Insulation Resistance and Test Voltage Measurement
The 5030 tests the functions of megohm meters using precision high value resistors up to 2GΩ, accurately measuring test voltages
up to 1kV under 0.5mA or 1mA test conditions. Short circuit current measurement is also available up to 3mA.
Continuity and Earth Resistance
The 5030 precision low ohm resistors allow calibration of continuity functions found on most multifunction testers and insulation
testers. Test voltages and currents are also measured.
Mains Voltage and Frequency
The local mains supply voltage and frequency is precisely measured by the 5030. This is used to cross reference the voltage
reading on the unit under test and confirm the instruments accuracy.
www.timeelectronics.com
5030 Specifications
TECHNICAL SPECIFICATIONS
Loop
Function
Range / Values
Resolution
Accuracy
Loop Impedance
Resistor Values
1800, 330.0, 180.0, 33.00, 18.00, 3.300, 1.800,
0.330, 0.150 & 0.050Ω
4 digit
±0.5% of displayed value ± 10mΩ
Local Loop Compensation
0 to 9.999Ω
0.001Ω
±0.5% of value ± 10mΩ
Test Current
40A max (200ms) / 50W max
–
–
Function
Range / Values
Resolution
Accuracy
1ms
±0.1ms
RCD
10, 20, 30, 40, 50, 60, 80, 100, 300, 500, 1000, 1500, 2000ms
Trip Time
(Fully variable from via remote commands - EasyCal software)
Current
6.000, 10.00, 30.00, 100.0, 300.0, 500.0, 1000mA
4 digit
±0.5% of reading
±1% with x5 multiplier
Current Multipliers
x0.5, x1, x2, x5
–
–
Maximum Current
2500mA
–
–
Waveforms
AC, DC & half wave rectified
–
–
Phase Detection
0° or 180°
–
–
Pre Trigger Delay
0 to 2000ms
–
–
Pre Trigger Threshold
0 to 100% of nominal current
–
–
Insulation (option ICO1)
Function
Resistance
Range / Values
Resolution
Accuracy
1MΩ to 2000MΩ
1MΩ
1% of value
50kΩ to 1990kΩ
50kΩ
1% of value
50.0 to 99.9V DC
0.1V
1% of reading
Test Voltage Measurement
@ 0.5mA or 1.0mA Load
1000 to 1200V DC
1V
1% of reading
Short Circuit Current
0.00 to 300.00mA DC
0.01mA
1% of range
Range / Values
Resolution
Accuracy
Continuity (option ICO1)
Function
0.1Ω to 100.0Ω
0.1Ω
1% of value
250Ω, 500Ω, 1.00kΩ, 2.50kΩ, 5.00kΩ & 10.0kΩ
3 digit
1% of value
0.0 to 50.0V DC
0.1V
0.5% of range
0 to 400mA DC
1mA
0.5% of range
1 watt maximum
–
–
Function
Range
Resolution
Accuracy
Line Voltage Measurement
200.0 to 260.0V RMS
0.1V
0.5% of reading
Line Frequency Measurement
45.00 to 65.00Hz
0.01Hz
0.1% of reading
Resistance
Test Voltage Measurement
(input resistance 10MΩ)
Test Current Measurement
(between 0.1Ω and 100Ω)
Power Dissipation
Voltage
GENERAL SPECIFICATIONS
Warm up.......................................................30 minutes to full accuracy
Settling Time................................................Less than 5 seconds
Standard Interfaces......................................RS-232 and USB
Temperature Performance...........................Operating: 10 to 35°C, Full Spec: 23°C +/- 5°C, Storage: -10°C to 50°C
Operating Humidity/Altitude........................ < 80% non condensing / Altitude: 0 to 3km. Non operating: 3km to 12km
Line Power...................................................220 - 240V AC 50Hz. Power Consumption 200W max
Dimensions / Weight....................................W430mm x H155mm x D255mm. Weight: 8kg
Supplied With...............................................User manual, RS-232 cable, USB adaptor/cable
ORDERING INFORMation
5030.............................................................Electrical Tester Calibrator
ICO1.............................................................Insulation and Continuity option
C201.............................................................Factory Calibration Certificate (NPL traceable)
C137.............................................................UKAS Calibration Certificate (ISO 17025)
ECFLA..........................................................EasyCal Software (see separate datasheet for options)
Full specifications are available on request. Due to continuous development Time Electronics reserves the right to change specifications without prior notice.
Time Electronics Ltd, Unit 11 Sovereign Way, Botany Industrial Estate, Tonbridge, Kent, TN9 1RH. United Kingdom.
T: +44 (0) 1732 355993 F: +44 (0) 1732 770312 E: mail@timeelectronics.co.uk W: www.timeelectronics.com
v1a 22/8/11