G4 Multi Channel Weighing Instrument. RM Operating Instructions

G4 Multi Channel Weighing Instrument. RM Operating Instructions
GB
G4 Multi Channel
Weighing Instrument
Program versions 1.7.0.0 to 1.12.0.0
Operating instructions,
Quick installation
RM type
G4 Multi Channel Weighing Instrument
Contents
Introduction
General ............................................................ 1
Power supply .................................................... 1
Start-up ............................................................ 2
Maintenance ..................................................... 2
Safety information ............................................ 2
Service panel
General .............................................................3
Display ..............................................................3
Additional functions ...........................................4
Operation
Zero indication and zero adjustment .................5
Gross weight and net weight .............................5
Taring ................................................................6
Editing Preset Tare ...........................................7
Weight Printing ..................................................8
Level supervision ...............................................9
Accumulated Weights .....................................10
Flow display (program option) .........................12
Scale Batching (program option) .....................13
Installation
Mechanical installation ....................................15
Electrical installation ........................................15
VIEWPAN module ...........................................16
CPU unit ..........................................................17
WF IN, WF IN2 and HS WF2 ..........................18
Basic set-up
General ...........................................................21
Hardware configuration ...................................21
Security lock ....................................................21
Common parameters ......................................22
Clock set-up ....................................................24
Scale calibration ..............................................25
Data sheet calibration ......................................28
Deadweight calibration ....................................31
Zeroing of gross weight ...................................36
Appendix 1
Common set-up values
Appendix 2
Scale set-up values
Operating instructions, Quick installation
PRECAUTIONS
READ this manual BEFORE operating or servicing this instrument. FOLLOW these
instructions carefully. SAVE this manual for future reference.
!
WARNING
Only permit qualified personnel to install and service this instrument.
Exercise care when making checks, tests and adjustments that must
be made with power on. Failing to observe these precautions can
result in bodily harm.
DO NOT allow untrained personnel to operate, clean, inspect,
maintain, service, or tamper with this instrument.
USE IN HAZARDOUS LOCATIONS
See FM and cFM approvals.
THIS EQUIPMENT IS SUITABLE FOR USE IN CLASS I, DIVISION 2, GROUPS A, B,
C and D HAZARDOUS LOCATIONS or NON-HAZARDOUS LOCATIONS.
WARNING- EXPLOSION HAZARD.
NONINCENDIVE FIELD WIRING CONNECTIONS TO CLASS I, II, III, DIV 2,
GROUPS A, B, C, D, E, F, G WHEN INSTALLED PER DWG 401709.
DO NOT REMOVE OR REPLACE CONNECTORS OR DISCONNECT EQUIPMENT
WHEN A FLAMMABLE OR COMBUSTIBLE ATMOSPHERE IS PRESENT.
INDOOR USE ONLY.
USB CONNECTION NOT ALLOWED TO BE USED IN HAZARDOUS LOCATIONS.
ATTENTION-DANGER D'EXPLOSION.
RACCORDEMENT POUR ATMOSPHERE NON EXPLOSIVE POUR CLASS I, II, III,
DIV 2, GROUP A, B, C, D, E, F, G INSTALLE SUIVANT PLAN 401709.
NE PAS DECONNECTER OU REMPLACER LES CONNECTEURS OU
DEBRANCHER L'EQUIPEMENT EN PRESENCE D'ATMOSPHERE IMFLAMMABLE
OU EXPLOSIVE.
UTILISATION INTERIEURE UNIQUEMENT.
CONNEXION USB NON AUTORISEE EN ZONE DANGEREUSE.
INTENDED USE
The G4 Instrument family are multi channel measuring and control devices intended
for industrial systems. Its basic function is to convert the signals from transducers to
useful information. Transducer excitation is included as well as parameter controlled
signal processing, indication of output levels, error supervision and operation of
optional external equipment.
The instrument supports several types of communication interfaces. The instruments
are modular and can be equipped with different types of I/O units. There are
transducer interface modules, digital and analog input/output modules.
G4 Multi Channel Weighing Instrument
Introduction
General
G4 is a high performance multi channel weighing instrument intended for industrial
systems. Its basic function is to convert the signals from strain gauge transducers
to useful weight information. Several types of communication interfaces are
supported by G4, which makes it easy to integrate the instrument into industrial
processes.
G4, type RM, is easily mounted on a DIN rail. All connections and the service
panel are accessible from the front end. By a LCD-display and four function keys
at the service panel, weight values are presented and parameter values can be
viewed and edited.
The basic configuration of the instrument performs weighing and flow rate
measurement. All functions are controlled by set-up parameters.
‘Operating instructions, Quick installation’ for G4 type RM includes basic installation
and set-up information, required for correct measuring with the instrument.
This description deals with the following points:
• Weighing with G4 type RM.
• Flow rate measurement (program option).
• Scale Batching (program option)
• Quick installation.
• Data sheet calibration.
• Deadweight calibration in two points.
Additional installation and set-up of more instrument functions, not covered by this
description, can also be performed, for example:
• Complete installation.
• Complete set-up.
• Setting of communication parameters.
• Table calibration.
• Level supervision.
• Program option activation.
For a complete instrument description, refer to:
G4 Multi Channel Weighing Instrument
Technical Manual
RM type
Power supply
Power supply to the weight indicator should not be turned off during weekends and
over-night. Continuous power supply to electronics and transducers prevents moisture
condensation in the units.
1
Operating instructions, Quick installation
Start-up
As soon as power is connected to the instrument, it will display 'Vishay' during a
start-up period, then the text ‘Start up, please wait’ is displayed after some seconds.
Then G4 automatically switches to normal measuring operation.
If the instrument is set for operator start-up, the text
‘Manual start. Push any key!’ will be shown until a key is pushed.
If a warm up time has been set, the text ‘Warming up Please wait!’
will be displayed before the instrument switches to measuring operation.
If any error is detected, the start-up stops and an error message will be displayed.
Maintenance
The G4 instrument needs no maintenance performed by the end-user. Any service
or repair work must be performed by qualified personnel. Contact your supplier.
Cleaning
Before cleaning the G4, break the power connection to the instrument.
Use a soft cloth to clean the exterior of the instrument.
Safety information
Utilization.
Before connecting power to the instrument, check that all fixation screws at the modules
are tightened so that the instruments functional grounding by the housing is maintained.
The instrument may only be utilized for the measurement and control functions, described
in the Technical Manual for G4 Multi Channel Weighing Instrument, RM type. It is
especially important to adhere to the load limits of the input/output connectors.
We accept no responsibility for any damage arising from improper operation.
Any changes to the instrument, which causes any function changes, may only be
carried out by the manufacturer, or after discussion with and permission by the
manufacturer.
Meaning of symbols, used in this manual
Direct current.
!
2
Caution, risk of danger. Documentation needs to be consulted.
G4 Multi Channel Weighing Instrument
Service panel
General
The service panel VIEWPAN features a two-line, back lighted LCD-display and four
function keys. It can display the weight or flow value, or other selected information,
for one scale at a time.
It is also possible to connect a standard USB keyboard for PC to the USB connector
at the CPU module. The keyboard will work in parallel with the keys on VIEWPAN and
will make it easier to handle the configuration of the instrument, as it can be used as
an input for digits and characters.
The keyboard keys ' + ', ' – ', ' ↑ ' or 'Esc', and 'Enter' will correspond the VIEWPAN
keys ' + ', ' – ', ' ↑ ' , and ' ↵ '.
Function key 'F11' on the keyboard can be used to access the 'Main Menu'.
Display
During measuring operation the function
keys + and – are used to toggle G4 RM
between display of measured values for
the scales in the instrument.
As G4 RM is in measuring operation
the top line shows a weight or flow value.
The bottom line, to the left, indicates
the actual scale number (S1:, S2: etc.),
and information about the measured value
(G=gross weight, N=net weight, M=motion,
Z=zero, Flow=flow display).
VIEWPAN module
3
Operating instructions, Quick installation
Additional functions
During measuring operation the additional functions for a scale that are activated
in the parameter set-up can be displayed to the right on the bottom line.
Key ↑ toggles the display of additional functions 'on' and 'off'.
As the additional functions are displayed, the keys + and – will perform
scrolling of the additional functions that are available for the scale.
Possible functions are: Tare, B/N, Print, Zero, Levels, Pre.Tare, W/F.
See explanations of the functions below.
As key ↵ is pressed, the displayed additional function will be performed
and display of additional functions will be toggled 'off'.
Explanations of the additional functions
Tare: This function performs taring, which means storing of the actual gross weight
as auto tare value and switching over to display of net weight.
Net weight being the gross weight minus tare.
B/N: This function performs toggling between display of gross weight, indicated by G,
and net weight, indicated by N.
Net weight cannot be displayed if the tare value in use is zero.
Print: This function initiates a print out of the displayed weight value.
Zero: This function performs Zero adjustment, a zeroing that can be performed only if
the gross weight is displayed, stable, and close to the basic zero. Basic zero setting of
the gross weight must be performed in the calibration sequences.
Levels: This function gives the possibility to rapidly view and edit the switch-over value
for the Levels, assigned to the scale.
Function Levels can be used only if at least one of the 32 Levels is assigned for
the scale in sub menu Param. Set-up / Level Superv.
Pre.Tare: This function opens menu Preset Tare for the scale, where the value of
preset tare can be viewed and edited.
Preset Tare can be opened only for scales that have Tare Corr.Mode set to
Preset or Auto+Preset in the scale calibration.
W/F: This function performs toggling between display of the weight value and
and the calculated flow value. Flow display can be selected only if program option
'Flow rate' is activated and Flow Calc. is set to On in the scale calibration.
4
G4 Multi Channel Weighing Instrument
Operation
Zero indication and zero adjustment
A basic zero setting of the gross weight for unloaded scale is performed at installation
as a part of the calibration. When a ‘good zero’ is displayed, the letter ‘Z’ is shown
under the measurement value.
Minor correction of the zero value may be needed and can rapidly be performed.
Zero adjustment
If, for unloaded scale, the letter ‘Z’ is not shown, but the gross weight is close to zero,
a rapid zero adjustment can be performed.
The gross weight must be stable (letter ‘M’ not shown)
and the zeroing function must be enabled.
Press key ↑ .
Additional functions for the scale will appear
to the right on the bottom line.
Use key + or – to scroll to the Additional function
'Zero'.
000.2 kg
S1:G
000.2 kg
S1:G
Zero
Press key ↵ to perform zero adjustment.
Gross weight, adjusted to zero, is displayed.
000.0 kg
S1:G Z
Zero adjustment can only be performed if the accumulated deviation from the basic
zero setting in the last calibration is within -1 % to +3 % of the ‘Capacity’.
In other cases a new basic zero setting must be executed.
Gross weight and net weight
The gross weight is the total weight loaded on the scale after zero setting.
When gross weight is displayed the letter 'G' is shown under the measurement value.
The net weight is the difference between the gross weight and a tare value.
When net weight is displayed the letter 'N' is shown under the measurement value.
Net weight can not be displayed if the tare value is zero.
Switch-over between display of gross weight and net weight can be performed if
parameter 'Gross/Net key' is set to 'On'. This makes the additional function 'B/N'
available for the scale.
5
Operating instructions, Quick installation
When gross weight is displayed the letter 'G' is shown
under the measurement value.
Press key ↑ .
Additional functions for the scale will appear
to the right on the bottom line.
Use key + or – to scroll to 'B/N'.
Press key ↵ to change to display of net weight.
143.2 kg
S1:G
143.2 kg
S1:G
B/N
Note: If the tare value is zero, net weight cannot be displayed.
When net weight is displayed, the letter 'N' is shown
under the measurement value.
Press key ↑ .
Additional functions for the scale will appear
to the right on the bottom line.
Use key + or – to scroll to 'B/N'.
Press key ↵ to change to display of gross weight.
Gross weight is displayed.
103.2 kg
S1:N
103.2 kg
S1:N
B/N
143.2 kg
S1:G
Taring
Taring means storing of a tare value for the scale. Net weight will be calculated as
the difference between the gross weight and this tare value. If the tare value is zero,
net weight will not be displayed.
At default setting, taring can always be performed if parameter 'Tare key' is set to 'On'.
This makes the additional function 'Tare' available for the scale.
But the scale can also be set to allow taring only at stable weight (letter 'M' not shown).
In G4 two tare values can be stored, Auto tare and Preset tare.
Auto tare
By taring, the actual gross weight is stored as Auto tare value, and the scale will switch
to display of net weight 'zero'.
Press key ↑. Additional functions for the scale will
appear to the right on the bottom line.
Use key + or – to scroll to 'Tare'.
014.5 kg
S1:G
Tare
Press key ↵ to perform taring.
In this example 14.5 kg is stored as an auto tare
value.
6
000.0 kg
S1:N Z
G4 Multi Channel Weighing Instrument
Editing Preset tare
In this example Scale 1 is set for taring with ‘Preset tare’. The example shows editing
the value of ‘Preset tare’ from VIEWPAN.
Scale 1, set for preset taring,
is in normal weighing operation.
Press key ↑ to see Additional functions.
Use key + or – to find the additional
function 'Pre.Tare'.
Press key ↵ .
033.2 kg
S1:N
033.2 kg
S1:N
Pre.Tare
The value of Preset tare for the scale is shown.
Press key ↵ to edit the value.
Preset Tare 1
12.0
kg
If the Operator Lock is active, the Operator Code will be needed.
A cursor appears, meaning the value is ready
for editing. Use key + or – to edit one digit
at a time. Accept each digit by pressing key ↵ .
Preset Tare 1
000010.0
kg
Finish by pressing key ↵ for 1 second.
The edited value of preset tare
for Scale 1 is displayed.
Press key ↑ to return to normal weight operation.
Scale 1 displays the weight value.
Preset Tare 1
10.0
035.2 kg
S1:N
7
Operating instructions, Quick installation
Weight Printing
A printer can be used to print the displayed weight or the displayed flow rate. The
printer must be connected to one of the G4 Instrument serial communication ports and
the communication parameters must be correctly set.
This example shows how to make a print out of the displayed weight for Scale 1 (the
displayed scale) from VIEWPAN.
Scale 1 is selected and in normal weighing operation.
Press key ↑ to see Additional functions.
Use key + or – to find the additional function 'Print'.
Press key ↵ .
The displayed weight value is printed, indicated
by a 'P' in the display.
033.2 kg
S1:N
033.2 kg
S1:N
Print
033.2 kg
S1:N P
The print indicator ‘P’ is shown where the ‘Z’ (zero) or ‘M’ (motion) indications normally
appear. The print indication will appear even if no printer is configured and indicates
that the weight is accumulated.
Print out will be performed provided that:
8
•
The displayed weight is higher than the value of 'Min. Weight Print'. No print out
will be done if the weight is too low.
•
If 'Motion Check' is turned on, the displayed weigh must be stable ('M' not
shown).
If the weight is not stable ('M' shown), printing will be delayed and the 'P' will be
flashing. When the weight has become stable ('M' not shown) printing will be
performed.
G4 Multi Channel Weighing Instrument
Level supervision
With G4 it is possible to supervise different weighing signals at defined levels.
The instrument contains 32 supervising units, Level 1 to Level 32, which can be
connected to the scales by the parameter set-up.
All Levels, connected to a scale, can easily be shown by scrolling at the display.
Edit levels
The defined supervising levels for a scale are easily viewed and edited
from VIEWPAN at G4 RM.
Editing of levels is performed without interrupting the normal weighing operation.
Example: Scale 1, with some connected Levels,
is in normal weighing operation.
Press key ↑ to see Additional functions.
Use key + or – to find the additional
function 'Levels'.
Press key ↵ .
The value of the first Level for Scale 1 is shown.
Use key + or – to scroll all Levels for the scale.
Press key ↵ to edit a displayed value.
If the Operator Lock is active, the Operator Code
will be needed.
A cursor appears, meaning the value is
ready for editing.
Use key + or – to edit one cursor position at a time.
Accept the value of each position by pressing key ↵ .
123.4 kg
S1:G
123.4 kg
S1:G
Levels
Level 2
50.0
kg
Level 2
+000055.0
Finish by pressing key ↵ for 1 second.
The new value for the edited Level is displayed.
Press key ↑ to return to normal weight operation.
Level 2
55.0
kg
9
Operating instructions, Quick installation
Accumulated Weights
Every time a weight is printed the weight value is added to an accumulated value for
each scale. Accumulation is done even if no printer is set-up or connected.
Accumulated weights can be changed and set to zero from menu 'Accum. Weights'.
This menu is reached from the 'Main Menu'.
Zero setting of accumulated weights for all scales
The example below shows how to set the accumulated weights for all scales in the
instrument to zero.
Press the keys + and ↑ simultaneously
for 1 second to open 'Main Menu'.
123.4 kg
S1:G
The 'Main Menu' is displayed.
Use key + or – to find sub menu 'Accum. Weights'.
Press key ↵ .
Main Menu
Accum. Weights
Menu 'Accum. Weights' is displayed with the
accumulated weight for the first scale on the lower
row.
Accum.Weight,S1
3798.100
kg
Press the key + a number of times until
'Zero ALL values?' is shown.
Press key ↵ (Yes) to set the accumulated weights
for all scales to zero.
Zero ALL values?
No
Yes
If the Operator Lock is active, the Operator Code
will be needed.
A confirm question will be shown.
Press key ↵ (Yes) again to perform the zero setting.
Continue?
No
The instrument returns to accumulated weights display.
Accum.Weight,S1
0.000
kg
Press key ↑ several times to return to
weight display.
10
Yes
G4 Multi Channel Weighing Instrument
Edit (zeroing) of accumulated weight for one scale
The example below shows how to edit (zeroing) an accumulated weight for a scale.
Press the keys + and ↑ simultaneously
for 1 second to open 'Main Menu'.
123.4 kg
S1:G
The 'Main Menu' is displayed.
Use key + or – to find sub menu 'Accum. Weights'.
Press key ↵ .
Main Menu
Accum. Weights
Menu 'Accum. Weights' is displayed with the
accumulated weight for the first scale on the lower
row.
The keys + and – can be used to scroll the list
of available scales.
Accum.Weight,S1
3798.100
kg
Press key ↵ to edit a displayed value.
If the Operator Lock is active, the Operator Code
will be needed.
A cursor appears, meaning the value is
ready for editing.
Use key + or – to edit one cursor position at a time.
Accept the value of each position by pressing key ↵ .
Accum.Weight,S1
+0000000000.000
Finish by pressing key ↵ for 1 second.
The new value is displayed.
Press key ↑ several times to return to
weight display.
Accum.Weight,S1
0.000
kg
11
Operating instructions, Quick installation
Flow display (program option)
When option Flow display is activated for a scale, the additional function 'W/F' can be
shown in the scale display.
When 'W/F' is made visible during weight display, and key ↵ is pressed,
this will switch the scale over to display of flow value.
When 'W/F' is made visible during flow display, and key ↵ is pressed,
this will switch the scale over to weight display.
Example: Scale 1, with the Flow display activated,
is in normal operation, displaying
an increasing gross weight.
064.4 kg
S1:G M
Press key ↑ to see Additional functions.
Use key + or – to find the additional
function 'W/F'.
Press key ↵ .
The measurement value for Scale 1
changes from weight to flow.
As the scale is displaying the flow value,
press key ↑ to see Additional functions.
Use key + or – to find the additional
function 'W/F' .
Press key ↵ .
The measurement value for Scale 1
changes from flow to weight.
12
096.8 kg
S1:G M
W/F
2.15 kg/s
S1:Flow
2.15 kg/s
S1:Flow
2.15 kg/s
S1:Flow
W/F
131.1 kg
S1:G M
G4 Multi Channel Weighing Instrument
Scale Batching (program option)
The RM instrument is a complete multi scale batcher but it is not intended as an
operator interface for batching. It is possible to set-up the batching sequences
from the RM instrument front panel. However for control (start, stop, reset etc) of
the batching sequence should a supervising computer be used (PLC, SCADA
system etc).
An animated symbol is shown in the weight display, when batching is active
(running) for the selected scale.
Animated batching indicator symbol
Batched weights
For each scale accumulated weights (batched material) will be created.
In the menu ‘Batched Weights’, in the ‘Main Menu’, are the accumulated weights for
each configured scale to be found. Each batched weight can be changed, e.g. set to
zero (see ’Accumulated Weights’ above).
13
Operating instructions, Quick installation
14
G4 Multi Channel Weighing Instrument
Installation
Mechanical installation
G4 Multi Channel Weighing Instrument type RM is designed for mounting on a flat
surface by DIN rail, for example in a protective cabinet. To remove the instrument
from the rail, the black lever at the lower bottom side of the instrument must be
engaged.
All electrical connections to the instrument are made at the front side so enough
room for connectors and appropriate cable ducts should be arranged.
Electrical installation
The field wiring of the instrument shall be suitable to the environment
(e.g. chemically) in the end-user application.
!
Field wiring installation shall comply with any national regulations, hereunder
National Electrical Code (NEC) for US and/or Canadian Electrical Code for Canada.
•
A switch or circuit-breaker shall be included in the building installation.
•
The switch shall be in close proximity to the equipment and
within easy reach of the operator
•
The switch shall be marked as the disconnecting device for the equipment.
•
The equipment switch or circuit-breaker employed as disconnecting device
shall comply with relevant requirements of IEC 60947-1 and IEC 60947-3.
The power supply for the instrument should be made via an external DC source.
All electrical connections to the instrument, including connection to ground,
are made via plug-in terminal blocks. Shielded cables are needed, except for
the power input, and the cables should be routed so that electromagnetic
interference from power cables is avoided.
WARNING
Make sure that that the power to the instrument is turned off before:
- any modules are removed from or inserted in the instrument.
- any connections are connected to or disconnected from the instrument.
All modules should be regarded as ESD sensitive. Make sure that an ESD safe
environment is maintained when inserting modules, removing modules and
when handling modules separately from the instrument. Modules must be kept
in metallised ESD bags when not mounted in the instrument.
15
Operating instructions, Quick installation
VIEWPAN module
The output of the external dc supply must be rated 24 V
, ±15%
including fluctuations, min. 40 W. The supply must provide Double Insulation
between Mains parts and 24 V SELV or SELV-E circuit; and a limited-energy
circuit (max. available current 8 A). For the US market this energy limit
can be achieved with an ANSI/UL 248-14 fuse rated 5 A.
For other markets an IEC 60127 T type fuse rated 4 A may also be used.
!
The power supply unit, designed to supply the complete instrument, is incorporated
in the VIEWPAN module.
24 V
In
Terminals 1, 2, and 3.
Connect power to terminal 1 (positive) and
terminal 2 (0 V). To achieve functional
grounding, terminal 3 should be connected
to ground.
VIEWPAN
+24 V
Ground
+24 V
16
1
2
0V
Out
0V
Out
24 V
Terminals 4 (positive) and 5 can be used
to supply max. 100 mA to logics, like
outputs and inputs.
In
3
4
5
6
+24 VDC In
0V
Shield
+24 VDC Out
0V
Shield
G4 Multi Channel Weighing Instrument
CPU unit
External computing devices connected to the CPU communication interfaces
of the instrument have to comply with the standard, UL 60950.
The internal battery in the CPU module is to be used only in the equipment
where servicing of the battery circuit and replacement of the lithium battery
will be done by a trained technician.
!
COM1
RS-232 Serial communication.
This is a SELV/SELV-E circuit.
(DTE)
RS-232
9p 25p
COM1 can be used for serial communication
with computer/PLC (Modbus RTU) or a
printer.
Point to point communication, only one
G4 unit connected to the computer/PLC.
2
3
5
COM1 RS-232
7
3
TX
8
2
RX
9
7
0V
10
Shield
Connections are made to terminals
7 to 9. Shielded cable must be used.
Connect the shield to terminal 10.
COM2
RS-485 Serial communication for
2-wire or 4-wire with common 0 V.
This is a SELV/SELV-E circuit.
Communication port COM2 can be
used for serial communication with
computer/PLC (Modbus RTU) or a
printer.
Connections are made to terminals
1 to 5. Shielded cable must be used.
Connect the shield to terminal 6.
The communication lines must be
terminated at both ends.
If G4 is connected at the end of
the communication line, the switches
T2 and R2 must be set according to
the table below.
2-wire termination:
Both T2 switches ON,
both R2 switches OFF.
4-wire termination:
Both T2 switches ON
both R2 switches ON.
RS485
COM2
2-wire
1
2
3
4
5
6
RS485
TT+
RR+
T2
T2
R2
R2
+5
0
+5
0
0V
Shield
COM2
4-wire
1
2
3
4
5
6
TT+
RR+
T2
+5
T2
0
R2
R2
0V
Shield
17
+5
0
Operating instructions, Quick installation
WF IN, WF IN2 and HS WF2
!
The voltage levels on connectors of I/O modules shall not exceed
hazardous voltage levels of 30 Vrms, 42.4 Vpeak or 60 Vdc under
normal conditions. In wet locations these voltage levels shall
not exceed 16 Vrms, 22.6 Vpeak or 35 Vdc.
Transducer inputs
Terminals 17 – 23 (channel 1), 10 – 16 (channel 2). See next page.
Transducer connection should be handled with great care to achieve good
measuring data. Transducer integrated cables may not be shortened.
NOTE!
Transducer cables must be routed at least 200 mm away from 230/380 V, 50/60 Hz
power cables. By cables with other frequencies or high power, an even wider distance
is preferable.
4-wire connection should be used if the transducer integrated cable is long enough to
be connected directly to a transducer input. At 4-wire connection, some terminals must
be interconnected as shown in the figure on next page.
6-wire connection should be used if the integrated cable must be lengthened
or if several transducers should be connected to one transducer input.
The channel 1 cable shield must be connected to terminal 21 and
the channel 2 cable shield must be connected to terminal 14.
In WF IN and WF IN2 the shield terminals are internally connected to the G4 housing,
which is internally connected to earth via the power supply connector terminal 3 (Shield).
The shield shall not be connected at any other point.
In HS WF2 the transducer input channels are separately insulated by operational
insulation and the shields can be connected to the most convenient ground/earth point.
This can be the junction box when using multiple transducers or at the barrier ground
when using Ex zener barriers.
In the junction box SL-4 from Nobel Weighing Systems, see figure, all necessary
terminals are provided.
18
G4 Multi Channel Weighing Instrument
Transducer input, ch. 2
4-wire connection
Exc.+
10
Exc+
Exc.Sign.+
11
Sense+
Sign.-
Transducer
12
Exc-
13
Sense-
15
Sign+
16
Sign-
14
Junction box
Exc.+
6-wire connection
Shield
Transducer input, ch. 1
17
Exc.Sign.+
Transducer
18
Sign.-
19
20
Exc.+
Exc.-
22
Sign.+
Transducer
Sign.-
23
21
Exc+
Sense+
ExcSenseSign+
SignShield
A transducer may be connected directly to terminals at the transducer input.
For several transducers or long distances,
a junction box and lengthening cable is needed.
For a HS WF2 channel, the shield can be connected to
ground/earth at any point.
19
Operating instructions, Quick installation
20
G4 Multi Channel Weighing Instrument
Basic set-up
General
All operating functions in G4 are controlled by parameters, stored in the instrument
memory. The actual setting of the parameter values can be edited during normal
weighing operation, by the display and the keys on the service panel VIEWPAN.
WARNING. Changes done during editing of set-up parameters will affect the
behaviour of the instrument immediately. The user must take all necessary
precautions to prevent any undesired effects in process monitored or controlled
by the G4 instrument or a connected control system.
It is strongly recommended to activate the set-up lock in the instrument to prevent
any unauthorized changes of set-up parameters.
This section proposes a number of parameters to set, that will rapidly give
the instrument a useful presentation mode and basic measuring properties.
It contains setting of, for example, the measurement unit and resolution for
the weight value, and also some parameter settings for data sheet calibration
and deadweight calibration in two points.
If you are dealing with an already working instrument we recommend that you perform
a backup of the set-up parameters before you make any changes to the instrument.
When the set-up is finished, all parameter values should be noted (see appendix 1
and 2), or backed up in a file. Backup to an USB memory (or internal file) can be done
from menu 'Maintenance', sub menu 'Create Backup'.
Appendix 1 and 2 to this manual contains suitable forms to fill in for the instrument
hardware configuration and for the set-up parameters, covered by this manual.
Hardware configuration
The instrument G4, type RM, has room for one fieldbus module and three other
modules. The configuration of installed modules, used measuring channels, and
scale numbers is made in set-up parameter menu 'Hardware Config.'.
At delivery of an instrument, the hardware configuration normally corresponds to
the set-up. If any mismatch between installed modules and hardware configuration
occurs, an error message will be shown at instrument start-up.
Consult 'G4 Multi Channel Weighing Instrument Technical Manual RM type'
if you need to change the hardware configuration.
Security lock
With an instrument G4 it is possible to perform set-up operations during normal
operation, which may influence for example the performance of a scale. To prevent
these dangers, the instrument is equipped with security locks at two levels that can
be activated to protect from unauthorized access to editing of parameters and values.
Four-digit codes for the locks can be defined by the customer.
By default setting the code for both locks is: 1 9 3 7 .
21
Operating instructions, Quick installation
Common parameters
G4 has a number of parameters that are common for the whole instrument.
They are found in menu 'General', a sub menu to 'Param. Set-up'.
Press the keys + and ↑ simultaneously
for 1 second to open 'Main Menu'.
In 'Main Menu', use key + or – to find
sub menu 'Param. Set-up'.
Press key ↵ .
In 'Param. Set-up', use key + or – to find
sub menu 'General'.
Press key ↵ .
Sub menu 'General' is displayed with the first
parameter 'Language'. The keys + and –
can be used to scroll the list of parameters.
Some basic parameters are explained below.
123.4 kg
S1:G
Main Menu
Param. Set-up
Param. Set-up
General
General
Language
Press key ↵ to perform editing
of the displayed parameter.
The Set-up Code or the Operator Code may be needed to start editing.
Language
If parameter 'Language' is selected for editing, the actual language appears
on the bottom line with a flashing cursor.
Press key + or – to scroll the alternatives.
The selected language will be used for all text
at the display, and for all parameter names.
Language
English
Press key ↵ for 1 second to accept
the displayed alternative.
The selected language is displayed
without cursor.
Language
English
Now key + or – can be used to display another parameter.
22
G4 Multi Channel Weighing Instrument
Date Format
If parameter 'Date Format' is selected, the actual format appears on the bottom line
with a flashing cursor. Press key + or – to scroll the alternatives.
In the alternatives, YYYY stands for 'year',
MM stands for 'month', DD stands for 'day'.
Date Format
Press key ↵ for 1 second to accept the displayed
YYYY-MM-DD
alternative.
The selected date format is displayed without cursor.
Now key + or – can be used to display another
Date Format
parameter.
YYYY-MM-DD
Time Format
If parameter 'Time Format' is selected for editing,
the actual format appears on the bottom line
with a flashing cursor.
Press key + or – to scroll the alternatives.
Press key ↵ for 1 second to accept the displayed
alternative.
The selected time format is displayed without cursor.
Now key + or – can be used to display another
parameter.
Time Format
24 h
Time Format
24 h
Set-up Lock
For security reasons, we recommend setting of this
parameter to 'On'.
If parameter 'Set-up Lock' is selected for editing,
Set-up
the actual setting appears on the bottom line
Off
with a flashing cursor.
Press key + or – to scroll the alternatives.
Press key ↵ for 1 second to accept the displayed
Set-up
alternative.
The selected alternative is displayed without cursor.
On
If 'Set-up Lock' is set to 'On', parameter 'Set-up Code'
will appear. See below.
Now key + or – can be used to display another parameter.
Lock
Lock
Set-up Code
Set-up Code
If 'Set-up Code' is selected for editing, the actual code
appears on the bottom line with a flashing cursor.
1937
Use key + or – to edit the code, digit by digit.
Accept each digit by key ↵ .
Press key ↵ for 1 second to accept the displayed
Set-up Code
code.
****
The code, replaced by four asterisks, is displayed
without cursor.
Now key + or – can be used to display another parameter.
Press key ↑ twice to leave 'General' and return to the 'Main Menu', where a sub menu
is available for Clock Set-up of the instrument. See next page.
23
Operating instructions, Quick installation
Clock set-up
Date/time information always exists within G4 RM, but is not presented at VIEWPAN.
Setting of correct date and time is performed in 'Clock Set-up', a sub menu
to the 'Main Menu'.
Setting of date and time doesn't interrupt normal weighing operation.
(To open the 'Main Menu' from normal weight display,
press the keys + and ↑ simultaneously for 1 second.)
In 'Main Menu', use key + or – to find
sub menu 'Clock Set-up'.
Press key ↵ to enter the menu.
Parameter 'Date:' is displayed.
To edit the date, press key ↵ .
A flashing cursor appears on the bottom line.
Use key + or – to edit the date digit by digit.
Accept each digit by key ↵ .
Press key ↵ for 1 second to accept the displayed
date. The cursor disappears.
Now key + can be used to display next parameter.
Parameter 'Time:' is displayed.
To edit the time, press key ↵ .
Main Menu
Clock Set-up
Date:
2006-09-23
Date:
2006-09-24
Date:
2006-09-24
Time:
19:07:14
A flashing cursor appears on the bottom line.
Use key + or – to edit the time digit by digit.
Accept each digit by key ↵ .
(Display of time is halted during edit.)
Time:
19:09:24
Press key ↵ for 1 second
to accept the displayed time.
Time:
19:09:30
The cursor disappears.
Press the key ↑ to leave 'Clock Set-up' and get to the 'Main Menu'.
24
G4 Multi Channel Weighing Instrument
Scale calibration
All calibration set-up parameters are set individually for each scale. The scale number
is used to distinguish parameters for different scales. The following example shows a
set-up sequence for Scale 1, so all parameters will begin with 1: .
Individual parameters for the scales are found in menu 'Calibration',
a sub menu to 'Param. Set-up'.
('Param. Set-up' is a sub menu to the 'Main Menu'. To open 'Main Menu' from normal
weight display, press the keys + and ↑ simultaneously for 1 second.)
In 'Main Menu', use key + or –
to find sub menu 'Param. Set-up'.
Press key ↵ .
In 'Param. Set-up', use key + or –
to find sub menu 'Calibration'.
Press key ↵ .
Sub menu 'Calibration' contains settings
for the scales that are in use.
Use the keys + and – to find the desired
scale number, in this example Scale 1.
Main Menu
Param.Set-up
Param.Set-up
Calibration
Calibration
Scale 1
Press key ↵ .
As 'Scale 1' has been selected, the keys + and – can be used
to scroll the list of set-up parameters for the scale.
Some basic parameters are explained on the following pages.
Press key ↵ to perform editing of the displayed parameter.
The Operator Code or the Set-up Code may be needed to continue.
25
Operating instructions, Quick installation
1:Measurem. Unit
If this parameter is selected for editing, the actual measurement unit appears on
the bottom line with a flashing cursor. The measurement unit will be used for
the measured value and for related set-up parameters.
Press key + or – to scroll the alternatives.
Press key ↵ for 1 second to accept
the displayed alternative.
The selected measurement unit is displayed
without cursor.
Now key + or – can be used
to display another parameter.
1:Measurem. Unit
kg
1:Measurem. Unit
kg
1:Resolution
If this parameter is selected for editing, the actual resolution appears on the bottom
line with a flashing cursor. This parameter defines the smallest change in measured
value that will be shown, and the decimal point position for the scale.
0.1 means display of 0.0 – 0.1 – 0.2 – 0.3 etc.
0.2 means display of 0.0 – 0.2 – 0.4 – 0.6 etc
0.5 means display of 0.0 – 0.5 – 1.0 etc.
1:Resolution
0.1
Press key + or – to scroll the alternatives.
Press key ↵ for 1 second to accept the displayed
alternative. The selected resolution alternative is
displayed without cursor.
1:Resolution
0.1
Now key + or – can be used to display another parameter.
1:Capacity
If this parameter is selected for editing, the actual capacity with leading zeros appears
on the bottom line with a flashing cursor.
This parameter defines the nominal range for the scale. The value of ‘Capacity’ is
shown with measurement unit and decimal point position according to the settings
above.
Use key ↵ to position the cursor on a digit to edit.
Then use key + or – to edit the value of the digit,
and accept each digit by key ↵ .
Press key ↵ for 1 second to accept
the edited value for Capacity.
The edited value of capacity is displayed
without cursor.
Now key + or – can be used
to display another parameter.
26
1:Capacity
000300.0
kg
1:Capacity
300.0
kg
G4 Multi Channel Weighing Instrument
Scrolling the list of parameters for Scale 1 by
the key + will bring you to the last display:
'Scale 1
Scale 1 Calib.'
Scale 1
Scale 1 Calib.
Press key ↵ .
1:Calibr. Type
This display shows the current type of calibration
for the scale. The keys + and – can be used to
scroll the parameter values for this calibration.
1:Calibr. Type
Data Sheet
Press key ↵ to start a new calibration.
A cursor indicates that editing of calibration type is possible.
Three calibration types are available:
‘Data Sheet’ calibration for fast calibration when transducer data is available
and the installation is free from mechanical disturbances.
‘Deadweight’ calibration, the most accurate calibration type,
where known weights are used to give well defined load on the scale.
'Table' calibration for entry of recorded values from a previous calibration.
Use the keys + and – to find the type of
calibration to perform for the scale.
Press key ↵ for 1 second.
A question will be displayed.
Press key ↵ at 'Yes' to start the calibration.
1:Calibr. Type
Data Sheet
Start calib?
No
Yes
Performing a data sheet calibration is described on pages 28 – 31.
Performing a deadweight calibration is described on pages 31– 35.
27
Operating instructions, Quick installation
Data sheet calibration
This calibration method can be used when data sheets for the transducers
are available and no external forces influence the weighing installation.
If fixed supports are included, the load must be evenly distributed on all supports.
Data sheet calibration can be performed without any transducers connected,
but the transducers must be connected as the zero setting is performed.
A data sheet calibration for a scale should start with the scale parameters, described
on page 26. This example shows a Data Sheet calibration for Scale 1.
When data sheet calibration has been selected and you have accepted to start a new
calibration, the display will show the type of calibration that is selected for the scale.
Use keys + and – to find the parameter that should
be edited.
Press key ↵ and the parameter value will be
displayed with a cursor, ready to edit.
All parameters are explained below.
1:Calibr. Type
Data Sheet
1:Conv. Factor
This parameter defines the relationship between a weight value expressed in
transducer data sheet unit and the same weight value expressed in measurement unit.
The default value, 9.80665, can be used when transducers in N (Newton) are used for
weighing in kg.
If this parameter is selected for editing, the actual conversion factor appears
on the bottom line with a flashing cursor.
Use key ↵ to position the cursor on a digit to edit.
Then use key + or – to edit the value of the digit,
and accept each digit by key ↵ .
1:Conv. Factor
009.806650
Press key ↵ for 1 second to accept
the edited value for conversion factor.
The conversion factor is displayed without cursor.
Now key + or – can be used
to display another parameter.
1:Conv. Factor
9.80665
1:No of Transd
This parameter value should be set to the number of support points for the load, including
all transducers and fixed support points. The number will affect the number of parameters
'1:Rated Output' below. If this parameter is selected for editing, the number appears on
the bottom line with a flashing cursor.
Use the keys + and – to edit the number.
Press key ↵ for 1 second to accept number.
The number of transducers is displayed without
cursor.
Now key + or – can be used
to display another parameter.
28
1:No of Transd
2
1:No of Transd
2
G4 Multi Channel Weighing Instrument
1:Rated Load
Rated load for the transducer is a value, given in the data sheet. All the transducers
in the scale must have the same impedance and rated load. The rated load value,
expressed in the unit of the data sheet, should be entered for this parameter.
If this parameter is selected for editing, the actual rated load appears on
the bottom line with a flashing cursor.
Use key ↵ to position the cursor on a digit to edit.
Then use key + or – to edit the value of the digit,
and accept each digit by key ↵ .
1:Rated Load
001000.000000
Press key ↵ for 1 second to accept
the edited value for rated load.
The rated load is displayed without cursor.
Now key + or – can be used
to display another parameter.
1:Rated Load
1000.00
1:Rated Output 1, 2, etc.
The number of parameters 'Rated Output' corresponds to 'No of Transd.' above.
These parameter values should be set to the output signal values for the transducers,
given in the data sheets. For a fixed support point the value of the rated output should
be set to 0.00000 mV/V.
If this parameter is selected for editing, the actual rated output appears on the bottom
line with a flashing cursor.
Use key ↵ to position the cursor on a digit to edit.
Then use key + or – to edit the value of the digit,
and accept each digit by key ↵ .
1:Rated Output 1
2.03900
mV/V
Press key ↵ for 1 second to accept
the edited value for rated output.
The rated output 1 is displayed without cursor.
Now key + or – can be used
to display another parameter.
1:Rated Output 1
2.03900
mV/V
Enter the Rated Output value for all transducers and fixed support points in the same
way.
29
Operating instructions, Quick installation
1:Set Zero
This parameter is used to make the scale display gross weight zero when it is
unloaded. Make sure the scale is completely unloaded before this setting is performed.
Press key ↵ .
The current gross weight is displayed as a live weight.
Press key ↵ .
The value +000000.00 kg is proposed for '1:Set Zero'.
Press key ↵ for 1 second to accept 0.00 kg.
The zero setting is now finished.
Press key + to display next parameter.
1:Set Zero
0.00
kg
Gross Weight:
014.28
1:Set Zero
+000000.00 kg
1:Set Zero
0.00
kg
1:Zero Offset
This parameter shows the offset value, needed to make the scale display the gross
weight 'zero' for unloaded scale.
The value of '1:Zero Offset' should not be edited.
Press key + to see actual gross weight.
1:Zero Offset
14.28
kg
This display of live gross weight gives
the possibility to check the load on the scale
at any moment during the calibration.
Gross Weight:
xxx.xx
Press key + to see actual transducer signal.
This display of live transducer signal gives
the possibility to check the signal at any moment
during the calibration.
30
Transd. Signal:
x.xxxxx mV/V
G4 Multi Channel Weighing Instrument
Exit calibration.
To exit from calibration, press key ↑ twice.
A question, 'Save calib?', is displayed.
Answer Yes (key ↵ ) to keep the new settings
in the instrument memory, or
answer No (key – ) to exit calibration without saving
(all changes will be rejected).
Save calib?
No
Esc. Yes
In both cases the instrument switches to sub menu 'Calibration'.
Answer Esc. (key ↑ ) to continue the calibration for Scale 1.
Now it is possible to perform calibration of any remaining scales,
or to return to normal weight display by pressing key ↑ several times.
Deadweight calibration
This is the most accurate calibration method. It requires that known weights
to at least two-thirds of the scale capacity are available.
A deadweight calibration should start with the common parameters,
described on page 26.
An asterisk ( * ) will be shown at the calibration parameters that have not yet
been stored.
This example shows a two-point deadweight calibration for Scale 1.
When deadweight calibration has been selected
1:Calibr. Type
and you have accepted to start a new calibration,
Deadweight
the display will show the type of calibration that is
selected for the scale.
Use keys + and – to find the parameter that should be edited.
Press key ↵ and the parameter value will be displayed with a cursor, ready to edit.
All parameters are explained on the following pages.
31
Operating instructions, Quick installation
1:No of Cal. P
This parameter defines the number of calibration points. Up to six points can be
selected, parameters for load value and transducer signal will be displayed for
the selected number of calibration points.
In this example a two-point calibration is described.
If this parameter is selected for editing, the number
of calibration points appears on the bottom line
with a flashing cursor.
Use keys + and – to edit the number.
1:No of Cal. P
2
Press key ↵ for 1 second to accept the number.
The number of calibration points is displayed without
cursor.
Now key + or – can be used
to display another parameter.
1:No of Cal. P
2
1:Value Cal. P1
This parameter defines the load for the lowest calibration point. Normally the scale
should be unloaded and the parameter value set to 0 (zero). This weight value and
corresponding transducer signal value are automatically stored for the scale.
Press key ↵ .
The current gross weight is displayed as a live weight.
Press key ↵ .
The value +000000.0 kg, with a flashing cursor,
is proposed.
Press key ↵ for 1 second.
Parameter '1:Value Cal. P1' is displayed
without asterisk.
32
1:Value Cal. P1*
0.0
kg
Gross Weight:
014.28
1:Value Cal. P1*
+000000.0 kg
1:Value Cal. P1
0.0
kg
G4 Multi Channel Weighing Instrument
1:Value Cal. P2
This parameter, in a two-point calibration, defines the load for the highest calibration
point. Normally the scale should be loaded to at least two-thirds of the scale capacity.
The weight value and the corresponding transducer signal value are automatically
stored for the scale.
Press key ↵ .
1:Value Cal. P2*
500.0
kg
The current gross weight is displayed as a live weight.
Press key ↵ .
Gross Weight:
196.52
Change the proposed value to the value of the actual calibration load on the scale
for calibration point 2.
Use key ↵ to move the cursor step by step.
Use key + or – to edit one cursor position at a time.
Accept the value of each position by pressing key ↵ .
1:Value Cal. P2*
+000200.0 kg
Press key ↵ for 1 second to accept
the load value for point 2.
Parameter '1:Value Cal. P2' is displayed
without asterisk.
1:T.Signal P1
This parameter value shows the transducer signal
for point 1, a value that was automatically stored
at the lowest calibration point.
The parameter value can not be edited.
1:T.Signal P2
This parameter value shows the transducer signal
for point 2, a value that was automatically stored
at the highest calibration point.
The parameter value can not be edited.
1:Value Cal. P2
200.0
kg
1:T.Signal P1
0.42971
mV/V
1:T.Signal P2
0.67298
mV/V
33
Operating instructions, Quick installation
1:Set Zero
This parameter is used to make the scale display gross weight zero when it is
unloaded. Make sure the scale is completely unloaded before this setting is performed.
Press key ↵ .
1:Set Zero
0.00
kg
The current gross weight is displayed as a live weight.
Press key ↵ .
The value +000000.00 kg is proposed for '1:Set Zero'.
Press key ↵ for 1 second to accept 0.00 kg.
Gross Weight:
000.12
1:Set Zero
+000000.00 kg
The zero setting is now finished.
Press key + to display next parameter.
1:Set Zero
0.00
kg
1:Zero Offset
This parameter shows the offset value, needed to make the scale display the gross
weight 'zero' for unloaded scale. The value of '1:Zero Offset' should not be edited.
Press key + to see actual gross weight.
This display of live gross weight gives
the possibility to check the load on the scale
at any moment during the calibration.
1:Zero Offset
0.12
kg
Gross Weight:
xxx.xx
Press key + to see actual transducer signal.
This display of live transducer signal gives
the possibility to check the signal at any moment
during the calibration.
34
Transd. Signal:
x.xxxxx mV/V
G4 Multi Channel Weighing Instrument
Exit calibration.
To exit from calibration, press key ↑ twice.
A question, 'Save calib?', is displayed.
Answer Yes (key ↵ ) to keep the new settings
in the instrument memory, or
answer No (key – ) to exit calibration without saving
(all changes will be rejected).
Save calib?
No
Esc. Yes
In both cases the instrument switches to sub menu 'Calibration'.
Answer Esc. (key ↑ ) to continue the calibration for Scale 1.
Now it is possible to perform calibration of any remaining scales,
or to return to normal weight display by pressing key ↑ several times.
35
Operating instructions, Quick installation
Zeroing of gross weight
When the scale installation is calibrated, a zeroing is performed in order to make
the gross weight zero when the scale is unloaded. If mechanical equipment is added
later, this will affect the weight display for the scale, and a new zeroing of the gross
weight becomes necessary.
NOTE!
The scale must be unloaded when zeroing is performed.
Access ‘Set Zero' for any calibration type.
Navigate to sub menu 'Calibration' and select the
desired scale, in this example Scale 1.
The sequence is described on page 25.
Press key ↵ .
Calibration
Scale 1
Use key + to scroll the parameter list until
line 'Scale 1 Scale 1 Calib' is displayed.
Press key ↵ .
Scale 1
Scale 1 Calib
Use key + to scroll the parameter list until
parameter '1:Set Zero' is displayed.
Press key ↵ .
1:Set Zero
0.00
kg
The Set-up Code or the Operator Code may be
needed to continue.
Display of live gross weight gives the possibility
to check the load on the scale.
Press key ↵ .
The value +000000.00 kg is proposed for '1:Set Zero'.
Press key ↵ for 1 second to accept 0.00 kg.
The zero setting is now finished.
To exit the zero setting, press key ↑ twice.
A question, 'Save calib?', is displayed.
Gross Weight:
xxx.xx
1:Set Zero
+000000.00 kg
1:Set Zero
0.00
kg
Save calib?
No
Esc. Yes
Answer Esc. (key ↑ ) to continue the zero setting of
Scale 1.
Answer Yes (key ↵ ) to keep the new zero setting, or
answer No (key – ) to exit calibration without saving (all changes will be rejected).
In both cases the instrument switches to menu 'Calibration'.
Press key ↑ several times to return to weight display.
36
G4 Multi Channel Weighing Instrument
Common set-up values
Note: Only the set-up parameters covered in this manual are listed.
Location/Notes: .......................................................................................................
Software version: ........................................... Serial number: ..............................
Date: ...................................................
Hardware position
Module
Serial no.
Notes
Slot 1
........................
...........................
...........................
Slot 2
........................
...........................
...........................
Slot 3
........................
...........................
...........................
Slot 4
VIEWPAN
...........................
...........................
Fieldbus module
(at the CPU-unit)
........................
...........................
...........................
Parameter name
Default value
Set-up value
Language
English
...........................
Date Format
YYYY-MM-DD
...........................
Time Format
24 h
...........................
Set-up Lock
Off
...........................
Set-up Code
1937
...........................
Appendix 1
Common set-up values
Operating instructions, Quick installation
G4 Multi Channel Weighing Instrument
Scale set-up values
Note: Only the set-up parameters covered in this manual are listed.
Locations/Notes: .........................................................................................................
.....................................................................................................................................
Module: ................................... Ser. no.: ....................... Slot no.: ..........................
Parameter name:
Channel number: ..........................
...........................
Scale number:
..........................
...........................
Default values
X:Measurem. Unit
kg
...........................
...........................
X:Resolution
0.1
...........................
...........................
X:Capacity
500.0
...........................
...........................
X:Calibr. Type
Data sheet
...........................
...........................
X:Conv. Factor
9.80665
...........................
...........................
X:No of Transd.
3
...........................
...........................
X:Rated Load
2000.0
...........................
...........................
X:Rated Output 1
2.03900
...........................
...........................
X:Rated Output 2
2.03900
...........................
...........................
X:Rated Output 3
2.03900
...........................
...........................
X:Rated Output 4
2.03900
...........................
...........................
X:No of Cal. P
2
...........................
...........................
X:Value Cal. P1
0.0
...........................
...........................
X:Value Cal. P2
500.0
...........................
...........................
X:T.Signal P1
0.00000
...........................
...........................
X:T.Signal P2
1.66631
...........................
...........................
X:Zero Offset
0.00
...........................
...........................
Appendix 2
Scale set-up values
Operating instructions, Quick installation
Document no. 35132
Article no. 600 857 R8
© Vishay Nobel AB, 2017-01-26
Web site: www.blhnobel.com
Subject to changes without notice.