Model 152-1 NewSales - 1334.cdr

Model 152-1 NewSales - 1334.cdr
Trek Model 152-1
Surface/Volume Resistance Meter
The Trek Model 152-1 Surface/Volume Resistance Meter is
designed to precisely measure surface or volume resistance
on a wide variety of conductive, dissipative, and insulative
materials. It features exceptional measurement accuracy and
wide measurement ranges. When used with Trek’s uniquelydesigned 152P-CR-1 concentric ring probe (shown to left), the
instrument provides consistent ease of operation even at very
high resistance values. The Model 152-1 is lightweight,
portable and operable via batteries or an AC line power source
with battery eliminator.
Key Specifications
! Measurement Range:
! Measurement Accuracy (of reading)
at 25° ±10° C and 20% to 70% RH:
– Point to Point Probe (152BP-5P):
– Concentric Ring Probe (152-CR-1):
– Two Point Resistance Probe (152P-2P):
! Probe Electrode Test Voltage:
! Test Limit Current:
10 3 to 1013 Ω
103 to 1012 Ω range, ±5%
1013 Ω range, ±8%
103 to 1013 Ω range, ±5%
103 to 1013 Ω range, ±10%
User selectable, 10 V or 100 V ±2%
Limited to less than 13 mA in 10 V range and 1.7 mA in 100 V range
Typical Applications Include
! Measuring surface or volume resistance on materials
· Conductive, dissipative, insulative
! Measuring in accordance with ANSI/ESD Standards for
· Garments (STM2.1)
· Work surfaces (S4.1)
· Flooring (S7.1)
· Footwear (STM9.1)
· Planar materials (STM11.11)
· Volume resistance (STM11.12 and IEC 61340-2-3)
· Seating (STM12.1)
· Two point resistance measurements (STM11.13)
· Floor materials/footwear (STM97.1)
Model
152AP-3 mm x 25.4 mm
Miniature Resistance
Probes
Model 152P-2P
Two Point Resistance
Probe
Model
152BP-5P
Point-to-Point
Probes
Features and Benefit
!
!
!
!
!
!
!
!
Complies to ANSI/ESD Association Standards
Concentric ring probe pre-amplifier eliminates interference and enables reliable operation at high resistance values
Exceptional accuracy, stability and repeatability
Wide measurement range (103 to 1013 Ω)
Elastomer electrodes for excellent surface contact
Optional accessories include a Walking Test Adapter and Test Plate set
NIST-traceable Certificate of Calibration provided with each unit
compliant
TREK, INC. • 190 Walnut Street • Lockport, NY 14094 • USA • 800-FOR TREK
716-438-7555 • 716-201-1804 (fax) • www.trekinc.com • sales@trekinc.com
Test Probes / Accessories (cont.)
Model 152-1 Specifications
Test Plate Set
(consists of two
separate plates)
Performance
Resistance Measurement
Range
103 to 1013 W
Resistivity
Resistance X Factor 10 = Resistivity
Measurement Accuracy
(of the reading) at 25 °C
and 20% to 70% RH
Point-to-point 5 lb probe (152BP-5P)
103 to 1012 Ω range, ±5%
1013 Ω range, ±8%
Concentric Ring Probe (152P-CR)
103 to 1012 Ω range, ±5%
1013 Ω range, ±8%
Two point resistance probe (152P-2P)
103 to 1011 Ω range, ±5%
1012 to 1013 Ω range, ±10%
Probe Electrode Test
Voltage
User selectable 10 V or 100 V, ±2%
Test Current Limit
Limited to less than 13 mA in the 10 V
range and less than 1.7 mA in the 100 V
range
The use of these plates is described in the
ESD STM 11.12 (IEC 61340-2-3) standard
Conductive Plate
(5” x 5” / 127 mm x 127 mm) A stainless steel
conductive plate with a mini banana plug
Insulative Plate
(5.4” x 5.4” / 137 mm x 137 mm) Acts as an
insulative surface
Walking Test Adapter
Kit
The Walking Test Adapter allows the analysis
of resistance levels on the human body (STM
97.1)
Mechanical
Dimensions
180 mm H x 100 mm W x 44 mm D
(7” H x 4” W x 1.75” D)
Weight
Approximately 0.5 kg (1 lb.) with battery
Operating Conditions
Temperature
15°C to 35°C (59 °F to 95 °F)
Relative Humidity
5% to 80%, non-condensing
Altitude
To 2000 m (6561.68 ft.)
Features
LCD Display
Three digits plus two digit exponent
(scientific notation)
Low Battery Indicator
LCD message for low battery
Test Voltage Range
Indicator
Indicates the test voltage selected, either
10 V or 100 V
Automatic Shutoff
If the unit is left idle for longer than 10
minutes, the unit automatically turns off
ANSI / ESD Association
Standards
The Model 152-1 conforms to ANSI / ESD
Association Standards for measuring
surface resistance and surface resistivity.
Please refer to the Applications section
on page 1 of this data sheet for more
information
Power Requirements
Test Probes/Accessories
Model 152BP-5P
Test Probes
Two (2) 9-Volt batteries (NEDA 1605 Alkaline,
or equivalent) provide approx 6 hrs of power
AC Line Operation
The use of an AC battery eliminator allows for
AC line operation. The eliminator output
connector is a female type 2.1 mm, DC power
plug
Included Accessories
Operator’s Manual
PN: 23426
Ground Cord
PN: N9044
Optional Accessories
Set of 2 (2.27 kg / 5 lb). Available for
performing resistance measurements
including ANSI/ESD STM 4.1 standards
(point-to-point or resistance to ground
measurement)
Model 152P-2P
Two-Point Resistance
Probe
Performs measurements on surface
areas too small to be measured with
conventional probes
Model 152P-CR-1
Surface/Volume Concentric
Ring Probe*
Measures surface and volume resistance
of materials as per IEC or ESDA
standards. A three (3) position switch on
the probe selects either SURFACE
distance or VOLUME resistance
measurements with either a GUARDED
or UNGUARDED outer electrode. Uses
an exclusive built-in pre-amplifier design.
Model 152APResistance Probes
(miniature, set of 2)
Battery Operation
Test Plate Set
PN: 17530
Universal Adapter
(90 V to 260 V)
PN: F5054R
Carrying Case
PN: 43378
Walking Test Adapter
Kit
Model 1K039
Notes
*The Model 152-CR-1 will operate with the previous Model 152
resistivity meter in “surface” mode, just as the Model 152-CR probe
did. The 152-CR will operate with the Model 152-1 Resistance Meter
with the measurement being in “ohms,” not “ohms/sq.”
3mm X 25.4mm. Can be handheld or
randomly positioned
Copyright © 2014 TREK, INC. All specifications are subject to change. 1435/JRB
TM
Measurement and Power Solutions
www.trekinc.com
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising