Multi Gas Clip
Multi Gas Clip
Infrared
Pellistor
Quick-Start Guide
READ FIRST BEFORE OPERATION
The Multi Gas Clip is a personal safety device designed to detect the
presence of hydrogen sulfide (H2S), carbon monoxide (CO), oxygen
(O2) and combustible gases (LEL). Ensure that you have been
properly trained on the use of the detector and the appropriate
actions to take in the event of an alarm condition before operation.
Display Components
1
2
3
LCD
4
5
Description
1
Alarm Condition
2
Calibration/Test Mode
3
Battery Charge Level
4
Gas Readings
5
Gas Identifiers
The Gas Identifier (5) and the associated Alarm Condition (1)
will flash when a gas reading exceeds its alarm threshold.
Warning: Users must familiarize themselves with the
icons in both alarm and non-alarm states.
Warning: If the display is missing icons, or cannot be
clearly read, discontinue use and contact GCT.
Warning: The IR LEL sensor will not detect hydrogen (H2) or
acetylene (C2H2) gases however, due to their cross-sensitive
nature, the CO sensor will go into alarm well below 10% LEL.
Factory Default Alarms
Sensor
H2S
CO
O2
LEL
Low
10
35
19.5
10
High
15
200
23.5
20
TWA
10
35
STEL
15
50
Turning On the Detector
Press & hold down the button briefly to activate. Once activated, the
detector will begin to display each sensor’s readings. A chasing “0”
indicates the sensor is stabilizing:
Before each use, check that all
sensor and alarm ports are clear
of any debris or obstructions
Stabilization
Fully Operational
Turning Off the Detector
Press & hold down the button until the "OFF" countdown is complete.
During the countdown the detector will beep & flash 4 times then
shut down. Display will show:
Manually Calibrating the Detector
Continuously hold down the button through the "OFF" countdown &
the "CAL" countdown that follows it in order to start the manual
calibration mode. During these two countdowns the detector will beep
& flash 8 times with a pause between countdowns. Release button
when display shows "AUTO ZERO" followed by "APPLY GAS". Snap on
the MGC calibration cap & apply gas. Countdown display:
Note: Calibration can also be done using the MGC Dock or MGC Wall Mount Dock.
Battery Information
Low Battery Warning: With 20 minutes of run time left the
detector will beep, flash & display "LOW BAT". These same
warnings will continue until the detector performs an automatic
shutdown once the battery expires.
Charging: While charging, the battery icon will cycle and the
detector will display an approximate charge percentage. Full charge
occurs in 4-6 hours.
DO NOT charge the instrument in a combustible atmosphere.
DO NOT charge the instrument in temperatures above or below
the specified range of 0°C to 40°C.
Additional Questions?
For additional information on the Multi Gas Clip gas detector
regarding programing, calibration, etc. please refer to the
Multi Gas Clip User’s Manual. The manual and other helpful
videos & information are available for viewing or download at
www.gascliptech.com.
Contact Information
Gas Clip Technologies, Inc.
610 Uptown Blvd, Suite 4100
Cedar Hill, TX 75104 USA
Phone:
+1.972.775.7577
Fax:
+1.972.775.2483
Toll Free: 1.877.525.0808
E-mail:
info@gascliptech.com
Website:
www.gascliptech.com
Portable gas detectors you can count on
MGC QSG v2
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising