Signet 9900 Transmitter
Member of the SmartPro™ Family of Instruments
Features
• Multiple sensor types supported with one
instrument
• “Dial-type” digital bar graph
• NEW! View pH/ORP sensor calibration and
sensor usage information stored at the sensor
(requires compatible sensor and Preamp)
• Modules are field installable and replaceable
anytime
• Optional Relay Module for addition of two dry
contact relays
Panel Mount
Field Mount
• Optional H COMM Module for two-way
communication
• Optional Batch Module for Batch Control
The Signet 9900 Transmitter provides a single channel
interface for many different parameters including Flow,
pH/ORP, Conductivity/Resistivity, Salinity, Pressure,
Temperature, Level, Dissolved Oxygen, and other
sensors that output a 4 to 20 mA signal. The 9900
Transmitter can also be used as a Batch Controller when
a Batch Module is installed. New in Generation IV: view
pH/ORP sensor calibration and sensor usage information
stored at the sensor. This feature requires sensor with
memory chip and a 2751 Preamp.
The extra large (3.90” x 3.90”) auto-sensing backlit
display features “at-a-glance” visibility that can be
viewed at 4-5 times the distance over traditional
transmitters. The highly illuminated display and
large characters reduce the risk of misreading or
misinterpreting the displayed values. The display
shows separate lines for units, main and secondary
measurements as well as a “dial-type” digital bar graph.
The 9900 is offered in both panel or field mount versions.
Both configurations can run on 12 to 32 VDC power
(24 VDC nominal). The 9900 can also be loop powered
with compatible sensors.
Designed for complete flexibility, plug-in modules allow
the unit to easily adapt to meet changing customer
needs. Optional modules include Relay, Direct
Conductivity/Resistivity, H COMM, Batch, 4 to 20 mA
Output, and a PC COMM configuration tool. The unit
can be used with default values for quick and easy
programming or can be customized with labeling,
adjustable minimum and maximum dial settings, and
unit and decimal measurement choices.
• NEW! One 4 to 20 mA output in base unit. One
additional 4 to 20 mA available with optional
module.
• 4 to 20 mA input (with optional 8058 Signal
Converter)
• Warning and Relay LED indicators for
“at a glance” visibility
• Customizable features including label for
custom identification
• Optional PC COMM configuration tool for
configuration at a PC
MEASURING EQUIPMENT
77CJ
Applications
• Wastewater Treatment
• Reverse Osmosis
• Deionization
– Ultra Pure Water
– Two Bed System
– Mixed Bed System
• Chemical Manufacturing/Addition
• Metal and Plastic Finishing
• Fume Scrubber
• Cooling Towers
• Media Filtration
Specifications
General
Input Channels
One
Input Types Digital (S L)
3
Frequency
Measurement Types
Serial ASCII, TTL level, 9600 bps
Range
0.5 to 1500 Hz
Accuracy
0.5% of reading
Flow, pH/ORP, Conductivity/Resistivity, Salinity, Pressure, Temperature, Level, Dissolved Oxygen,
Batch or user-defined (via 8058)
Enclosure and Display
Case Material
PBT
Window
Shatter-resistant glass
Keypad
4 buttons, injection-molded silicone rubber seal
Display
Backlit, 7 and 14-segment
Update Rate
1s
LCD Contrast
5 settings
Indicators
“Dial-type” digital bar graph. LEDs for Open Collector, Relays and Warning Indicator
Enclosure Size
¼ DIN
Mounting
Panel
¼ DIN, ribbed on four sides for panel mounting clip inside panel, silicon gasket included
Field
Mounts to standard Signet field mount junction boxes. Optional angle adjustment adapter available.
Wall
Large enclosure (sold as an accessory) that encases the panel mount transmitter
Display Ranges
pH
0.00 to 15.00 pH
pH Temperature
-99 °C to 350 °C
ORP
-1999 to +1999 mV
Flow Rate
-9999 to 99999 units per second, minute, hour or day
Totalizer
0.00 to 99999999 units
Conductivity
0.0000 to 99999 μS, mS, PPM and PPB (TDS), kΩ, MΩ
-146 °F to 662 °F
Conductivity Temperature -99 °C to 350 °C
–146 °F to 662 °F
Temperature
-99 °C to 350 °C
–146 °F to 662 °F
Pressure
-40 to 1000 psi
Level
-9999 to 99999 m, cm, ft, in, %
Volume
0 to 99999 cm3, m3, in3, ft3, gal, L, lb, kg, %
Salinity
0 to 100 PPT
Dissolved Oxygen
PPM 0-50, % SAT 0-200, 0 to 999.9 TORR
Dissolved Oxygen
Temperature
-99 °C to 350 °C
–146 °F to 662 °F
Environmental
Ambient Operating Temperature
Backlit LCD
-10 °C to 70 °C
14 °F to 158 °F
Storage Temperature
-15 °C to 70 °C
5 °F to 158 °F
Relative Humidity
0 to 100% condensing for field mount; 0 to 95% non-condensing for panel mount
Maximum Altitude
4,000 m (13,123 ft)
Enclosure Rating
Designed to meet NEMA 4X/IP65 (front face only on panel mount); field mount is
100% NEMA 4X/IP65
Specifications (continued)
Electrical Requirements
Power to Sensors
Voltage
+4.9 to 5.5 VDC @ 25 °C, regulated
Current
1.5 mA max in loop power mode (up to 2.0 mA with 24 V @ 300 Ω max. loop impedance);
20 mA max when using DC power
Short Circuit
Protected
Isolation
Low voltage (< 48V AC/DC) to loop with DC power connected
No isolation when using loop power only
Terminal Blocks
Pluggable screw type
14 AWG max wire gauge
Input Power
DC
10.8 to 35.2 VDC, regulated
9900 without Relay Module
200 mA @ 10.8 VDC to 35.2 VDC
9900 with Relay Module
300 mA @ 10.8 VDC to 35.2 VDC
Overvoltage Protection
48 Volt Transient Protection Device
Current limiting for circuit protection
Reverse-Voltage Protection
Loop Power
No DC Power Input
Max. Loop Impedance
50 Ω @ 12 V
325 Ω @ 18 V
600 Ω @ 24 V
250 Ω @ 12 V
500 Ω @ 18 V
750 Ω @ 24 V
Dry-Contact Relays (2)
Open Collector (1)
Type
SPDT
N/A
Form
C
N/A
Max. Current Rating
5 A resistive
50 mA DC
Max. Voltage Rating
30 VDC or 250 VAC
30 VDC
Hysteresis
Adjustable (absolute in engineering units) (EUs)
Latch
Reset in test screen only
Delay
9999.9 seconds (max.)
Test Mode
Set On or Off
Cycle Time
99999 seconds (max.)
Maximum Pulse Rate
300 pulses/minute
With DC Power Input or with 2nd loop, all the time
Max. Loop Impedance
Relay Specifications
Proportional Pulse
400 pulses/minute
Volumetric Pulse Width
0.1 to 3200 s
Pulse Width Modulation
0.1 to 320 s
Input Types
Digital (S3L) or AC frequency
4 to 20 mA input via the 8058
pH/ORP input via the Digital (S3L) output from the 2750/2751 pH/ORP Sensor Electronics
Raw Conductivity/Resistivity input directly from Signet Conductivity/Resistivity electrodes via Direct Conductivity/
Resistivity Module or via 2850
Input Specifications
Digital (S3L)
Serial ACSII, TTL level, 9600 bps
Frequency Input
Sensitivity
80 mV @ 5 Hz, gradually increasing with frequency
Span
0.5 Hz to 1500 Hz @ TTL level input
Accuracy
± 0.5% or reading max error @ 25 °C
Resolution
1 μS
Repeatability
± 0.2% of reading
Specifications (continued)
Input Specifications continued
Power Supply
Rejection
±1 μA per volt
Short Circuit
Protected
Update Rate
(1/frequency) + 150 ms
Output Specifications
Current Output - One (1); Two (2) with 4 to 20 mA Output Module
Current Loop Output Standard
ANSI-ISA 50.00.01 Class H
Current Output
4 to 20 mA, isolated, fully adjustable and reversible
Span
3.8 to 21 mA
Zero
4.0 mA factory set; user programmable from 3.8 to 5.0 mA
Full Scale
20.00 mA factory set; user programmable from 19.0 to 21.0 mA
Accuracy
±32 μA max. error @ 25 °C @ 24 VDC
Resolution
6 μA or better
Temperature Drift
±1 μA per °C
Power Supply Rejection
±1 μA per V
Isolation
Low voltage (< 48 VAC/DC)
Voltage
12 to 32 VDC ±10%
Max. Impedance
(with DC power input)
250 Ω @ 12 VDC
500 Ω @ 18 VDC
750 Ω @ 24 VDC
Max. Impedance
(no DC power input)
50 Ω @ 12 VDC
325 Ω @ 18 VDC
600 Ω @ 24 VDC
Update Rate
150 mS nominal
Short circuit and reverse polarity protected
Adjustable Span
Reversible
Error Condition
Selectable error condition 3.6 or 22 mA
Actual update rate determined by sensor type
Test Mode
Increment to desired current (range 3.8 to 21.00 mA)
Shipping Weights
Base Unit
0.63 kg
1.38 lb
H COMM Module
0.16 kg
0.35 lb
Conductivity Module
0.16 kg
0.35 lb
Relay Module
0.19 kg
0.41 lb
Batch Module
0.16 kg
0.35 lb
4 to 20 Output Module
0.16 kg
0.35 lb
Standards and Approvals
CE, UL, CUL, FCC
RoHS Compliant, China RoHS
Manufactured under ISO 9001 and ISO 14001 for Environmental Management
and OHSAS 18001 for Occupational Health and Safety
Dimensions
99.06 mm
(3.90 in.)
44.45 mm
(1.75 in.)
29.97 mm
(1.18 in.)
107 mm
(4.21 in.)
99.06 mm
(3.90 in.)
91.44 mm
(3.60 in.)
54.10 mm
(2.13 in.)
8.13 mm
(0.32 in.)
3-9900.396
Angle Adjustment Adapter Kit
88.90 mm
(3.50
in.)
3.50
24.13 mm
(0.95 in.)
88.90 mm
(3.50 in.)
88.90
3.50mm
(3.50 in.)
25°
Ref. 44.58 mm
(1.77 in.)
Panel Mount
Field Mount - Pipe, Tank, Wall
Signet Model 9900 Transmitter
(Includes mounting bracket and panel gasket)
Signet Model 9900 Transmitter
with Junction Box (varies with sensor and installation)
+
+
3-8050
3-8051-X
3-8052
3-9900.396
(optional)
Signet Sensors - Flow, Level, Temperature, Pressure, DO
Use one input from sensor options below*
FLOW
515
2536
8510
8512
525
2000
2100
2507
2537
2540
2551
2552
2250
2350
Signet Fittings - See individual sensor data sheets
System Overview
Field Mount - Pipe, Tank, Wall
Signet Model 9900 Transmitter
with Junction Box (varies with sensor and installation)
+
+
3-8050
Signet Sensors - pH/ORP
Use one input from sensor options below*
With 2750/2751 Sensor Electronics
3-8052
3-9900.396
(optional)
Signet Wet-Tap Electrode Model 2756, 2757 and
3719 Wet-Tap with 2750/2751 Sensor Electronics
+
+
Signet Fittings - See individual sensor data sheets
All sold separately
Panel Mount
Signet Model 9900 Transmitter
(Includes mounting bracket and panel gasket)
2610
All sold separately
Panel Mount
Signet Model 9900 Transmitter
(Includes mounting bracket and panel gasket)
2450
Field Mount - Pipe, Tank, Wall
Signet Model 9900 Transmitter
with 3-9900.396 Angle Adapter and Junction Box
(varies with sensor and installation)
+
3-9900.396**
3-8050
3-8052
Signet Sensors - Conductivity/Resistivity and Salinity Electrodes
Use one input from electrode options below*
with Conductivity Module or 2850 Sensor Electronics
+
OR
Signet Fittings - See individual sensor data sheets
* See individual sensor datasheets for additional information
**3-9900.396 is required with the Conductivity Module and either 3-8050 or 3-8052 to provide sufficient clearance.
All sold separately
Plug in Modules
Optional modules are available to customize your 9900:
All modules come enclosed in a plastic cover. Modules are field installable and
replaceable any time.
Relay Module (Panel Installations only)
wer
DC Po
This module adds two programmable dry-contact relays
to the standard Open Collector output in the base unit.
Dry-contact relays are electromechanical switches with
a moving contact armature. They are suitable for many
general purpose applications, AC or DC, including loads
up to 250 V. Install RC Filter kits (3-8050.396) on relays
used to switch motor or inductive loads.
ltage
Loop Vo
3-9900.393
Direct Conductivity/Resistivity Module
wer
DC Po
Voltage
Loop
SHIELD
TEMP (WHT)
ISO GND (BLK)
SIGNAL (RED)
odule
MM M
H CO
The Direct Conductivity/Resistivity Module interfaces
Signet 2819-2823 and 2839-2842 Conductivity electrodes
directly to the 9900. The module also provides filtering
and conditioning. The 2850 Sensor Electronics can be
used in place of the Direct Conductivity/Resistivity
Module wired through the 9900 Digital (S3L) input.
le
Modu
/Res
Cond
3-9900.394
H COMM Module (HART®)
wer
DC Po
ge
Volta
Loop
The H COMM Module enables communication between
the 9900 and a HART®-enabled device. The HART
(Highway Addressable Remote Transducer) Protocol
superimposes digital signals on top of the 4 to 20 mA
analog signal.
odule
MM M
H CO
3-9900.395
Batch Module
The Batch Module adds batch capability to the 9900
Transmitter (Generation II and newer). It is compatible
with all Signet flow sensors. Up to 10 batch sizes can be
stored in one 9900 with customized names and K-Factors
available for each batch.
Refer to the Batch Control System datasheet for
further details.
3-9900.397
4 to 20 mA Output Module
The 4 to 20 mA Output Module adds a second 4 to 20 mA
Output to the 9900 Transmitter (Generation III and later).
Each of the outputs can be used to output the primary
and/or secondary measurement. Outputs have individual
settings available.
3-9900.398-1
Refer to the 4 to 20 mA Output Module manual
for further details.
Open Collector (R1)
Indicator LED
Mechanical Relay (R2)
Indicator LED
Backlight Sensor
(do not block)
Mechanical Relay (R3)
Indicator LED
Warning LED
Bar Graph
Units of Measure
(GPM, pH, sec, %, etc.)
Value
Label
Menu Indication
Menu Navigation Keys
ENTER
All possible segments shown in this illustration. The instrument’s software controls
which segments are shown at any particular time. Only the bar graph segment outline
and GF logo are visible when the unit is turned off.
9900 Module
H COMM
Relay
Conductivity/Resistivity
I
X
X
X
Batch
9900 Generation
II
III
X
X
X
X
4 to 20 mA Output
Sensor model
515/8510
525
2000
2100
2250
2350
2450
2507
2536/8512
2537-5
2540
2551
2552
2610-41
2610 + 8058
2724-2726
2734-2736
2750
2751
2756-2757
2764-2767
2774-2777
2819-2823
2839-2842
2850
4150 + 8058
X
X
X
X
X
IV
X
X
X
X
X
9900 Generation
I
X
X
X
X
X
X
X
X
X
X
X
X
X
II
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
III
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
IV
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
The Angle Adjustment Adapter Kit enables the 9900
transmitter to be mounted virtually anywhere. Field
Mount Installations with a Conductivity/Resistivity
Module require the Angle Adjustment Adapter Kit for
wiring clearance.
3-9900-1
(159 001 696)
Field Mount
3-9900-396
(159 001 701)
Angle Adjustment
Adapter Kit
3-8051 (159 000 187)
3-8051-1 (159 001 755)
3-8051-2 (159 001 756)
Flow Sensor
Integral Mounting Kit
Ordering Information
Mfr. Part No
Code
Description
9900 Base Unit - Single Channel, Multi-Parameter, 4 to 20 mA, Open Collector, DC power
3-9900-1P
159 001 695
9900 Panel Mount Transmitter
3-9900-1
159 001 696
9900 Field Mount Transmitter
3-9900-1BC
159 001 770
Batch Controller System
Optional Accessory Modules
Cond/Re
s Mod
ule
ule
M Mod
H COM
3-9900.393
159 001 698
Relay Module - 2 DCR (Dry-contact relays)
3-9900.394
159 001 699
Direct Conductivity/Resistivity Module
3-9900.395
159 001 697
H COMM Module
3-9900.397
159 310 163
Batch Module
3-9900.398-1
159 001 784
4 to 20 mA Output Module
Accessories and Replacement Parts
Mfr. Part No
Code
Description
6682-0204
159 001 709
Conductivity Module Plug, 4 Pos, Right Angle
6682-1102
159 001 710
DC Power Plug, 2 Pos, Right Angle
6682-1103
159 001 711
Relay Module Plug, 3 Pos, Right Angle
6682-1104
159 001 712
Loop Power Plug, 4 Pos, Right Angle
6682-3104
159 001 713
Freq/S3L Plug, 4 Pos, Right Angle
6682-3004
159 001 725
Terminal Block Plug
7310-1024
159 873 004
24 VDC Power Supply, 0.42 A, 10W
7310-2024
159 873 005
24 VDC Power Supply, 1.0 A , 24W
7310-4024
159 873 006
24 VDC Power Supply, 1.7 A, 40W
7310-6024
159 873 007
24 VDC Power Supply, 2.5 A, 60W
7310-7024
159 873 008
24 VDC Power Supply, 4.0 A, 96W
3-0251
159 001 724
PC COMM Configuration Tool
3-8050
159 000 184
Universal Mount Kit
3-8050.396
159 000 617
RC Filter kit (for relay use), 2 per kit
3-8051
159 000 187
Flow Sensor Integral Mounting Kit, NPT, Valox
3-8051-1
159 001 755
Flow Sensor Integral Mounting Kit, NPT, PP
3-8051-2
159 001 756
Flow Sensor Integral Mounting Kit, NPT, PVDF
3-8052
159 000 188
¾ in. Integral Mount Kit
3-8058-1
159 000 966
I-Go™ Signal Converter, wire-mount
3-8058-2
159 000 967
I-Go™ Signal Converter, DIN rail mount
3-9000.392-1
159 000 839
Liquid Tight Connector Kit, NPT (1 pc.)
3-9900.390
159 001 714
Standard Connector Kit, Right Angle, 9900 Transmitter
3-9900.391
159 001 715
Optional Connector Kit, In-Line, 9900 Transmitter
3-9900.392
159 001 700
Wall Mount Accessory Kit for 9900
3-9900.396
159 001 701
Angle Adjustment Adapter Kit (for Field Mounting)
3-9900.099 Rev H (7/14)
© Georg Fischer Signet LLC
3401 Aero Jet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057 • www.gfsignet.com • e-mail: signet.ps@georgfischer.com