Weighing Indicator
USER’S GUIDE
globalindustrial.com
globalindustrial.com
0713
11 Harbor Park Drive Port Washington, NY 11050
VI Calibration
Table of Contents
I
Main Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
1.
Connect the load cell properly, then turn on the indicator, press [UNIT] key during
startup, it will enter calibration mode and calibrate as follows:
Step
1
Operation
Display
Press [Tare] for division selection.
[d X]
II Display and Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . .4
2
1. Drawing of the indicator display and keyboard
3
Press [Tare] for DECIMAL
POINT selection.
[P X]
Set the full range.
[FULL]
2. Key functions
3. Connect the load cell to the indicator
4
Zero point calibration: Press [UNIT]
when the stable signal is on.
5
Full range point calibration: Press
[UNIT] when the value input is the
same as the loaded weight. Stable
signal must be on.
III Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-9
1. Power on and auto-zero setting
2. Zero function
[nOLOAD]
4. Counting mode
5. Accumulating mode
6. Tare function
2.
3.
7. RS-232 communication function
IV Maintenance and Warnings . . . . . . . . . . . . . . . . . . . .10
V Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
VI Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
globalindustrial.com
4.
Press [Tare] for digit bit selection.
Press [Zero ] for digit selection.
Press [UNIT] for confirm input.
Assure there is no load.
[End]
Press the calibration button under
the lead sealing board at the back
of the indicator.
Calibration
3. Unit setting
Select decimal point optional: 0~3,
press [UNIT] to confirm Example: 3.
While inputting the loaded weight,
[AdOLOAD] Press [Tare] for selection of the digit
bit; Press [Zero ] for digit selection.
When the input value is the
same as the loaded weight, press
[UNIT] when the stable signal is on.
6
7
Notes
Select division option
(1/2/5/10/20/50, press [UNIT] to
confirm Example: 20.
Button
Saves the calibration parameter and
sets back to the weighing status.
Attention: If switch is not pressed, all
parameters will not be saved.
Fast calibration for zero point and full range point: Press [UNIT] during startup,
it enters calibration mode.
Fast calibration for zero point: At any time before it shows [nOLOAD], press
[FUNC], it keeps the original division, decimal point, full range and enters the zero
point calibration mode. Press [ZERO] when the stable signal is on, it displays
[END] and keeps the original full range point calibration. Press the calibration
switch under the lead sealing board at the back of the indicator, this saves settings
and returns to weighing status.
Fast calibration for full range point: At any time before it shows [AdLOAD], press
[ACCU], it keeps the original division, decimal point, full range, zero point calibration
and enters full range point calibration mode. When it is finished, press the calibration
switch under the lead sealing board at the back of the indicator, this saves settings
and returns to weighing status.
globalindustrial.com
11
IV Maintenance and Warnings
To guarantee the clarity and life of the indicator, this device shouldn't be
used under direct sunlight and should be placed in room temperature area.
The indicator should be placed away from dense dust and moist environments.
Load cell should be securely connected to the indicator, and the system
should be positioned safely. They should be kept away from strong electric
and magnetic fields, corrosive objects, as well as flammable and explosive materials.
Read this manual carefully before operation
I Main Specifications
Do not use this device in an environment with flammable gas or steam.
Pressure vessel canned system is prohibited.
If placed in severe thunderstorm environments, a mounted lightning rod is
required to assure the safety of the operator and avoid damage to the indicator
and relevant devices.
Load cell and indicator are electrostatic sensitive devices; anti-static measures must
be performed during operation. Anti-static and anti-conductive matting is
required. It is prohibited to perform electric wielding and other strong electric
field operations.
It is prohibited to clean the case of indicator with strong corrosive chemicals
(such as, benzene and acidic lubricants)
Do not pour liquid or other conducting particles into the indicator, in order
to avoid electric hazards or damage to the indicator.
Power supply of the indicator and relevant devices must be turned off before
plugging or unplugging the connecting line of indicator and peripheral devices.
V Messages
Error Operation Messages
10
1.
Err 1
The AD value is too small when calibrated.
2.
Err 2
The zero point is out of range when calibrated.
3.
Err 3
The zero point is out of range upon starting.
4.
Err 4
The sample number inputted is zero when sampling in counting
5.
Err 5
The weight inputted is zero when scale is calibrated in
calibrating mode.
6.
Err 6
The unit weight is less than 0.25e when sampling in counting
7.
bAT-Lo
1. Model:
K3190 - A12 weighing indicator
2. Accuracy:
Grade III, n=3000
3. Sample Rate:
10x/second
4. Load cell sensitivity:
1.5 ~ 3mV / V
5. Scale interval:
1/2/5/10/20/50 for option
6. Display:
6 bits LCD, Six 1”H Digits
7. Scoreboard interface (optional):
Serial output mode: current loop
signal, transmission distance≤50m
8. Communication interface (optional):
RS232C; Baud rate 1200/2400/4800/9600
9. Power supply:
Battery DC6V/4AH
10. Operating temperature/humidity:
0 ~ 40°C; < 90%RH
11. Transporting temperature:
-20 ~ 50°C
Low power.
globalindustrial.com
globalindustrial.com
3
II Display and Keyboard
1. Drawing of the indicator display and keyboard
AC
Unit: Pound/KG
globalindustrial.com
1-800-645-1232
+ Out
- Out
RXD
Signal
Earth
TXD
(2-5)
COUNT
UNIT ACCUM
Lbs.
Kg.
UNIT
TARE
Count
Settings
Accum Off
STABLE
ZERO
Accum
Mode
Tare
Select
Zero
Edit
On
Off
B. Connect the scoreboard to indicator
1. The scoreboard interface adopts a 5-pin socket. Pins 3 and 4 in drawing
(2-5) define the leads.
Calibration
2. RS232 interface is used for scoreboard signal, which is transmitted in
serial binary code style.
Make sure that the communication interface output lead and computer
are connected correctly. If connection is unstable, risk of damage to
output port of instrument and input port of scoreboard will increase.
Only use provided connecting cable.
Professional computer technology and programming help is required.
Non-professional staff installation is prohibited.
LOAD CELL INTERFACE
RS232
BATTERY
POWER/ ON/OFF
DC
BACK VIEW
4
globalindustrial.com
globalindustrial.com
9
2. Key functions
6. Tare function
1. Place the load desired to be considered as tare and when indicator is in
weighing mode displaying positive weight and stable indicator is on, press [Tare]
key. The indicator will deduct the displayed weight value as tare weight, it then
displays net weight as “0”, and Tare sign indicator is on.
7. RS-232 communication function
A. Set parameters for RS-232 communication
There are 10 configurable modes from P1 to P10 (P11 and P12 are not applicable
for this unit).
1. When in weighing mode, press [Settings] button for approximately 5 seconds,
system will enter operator setting mode.
2. Press [Mode] to choose the mode and press [Select] to choose the parameter.
Parameters Description:
P1 – Unit:
1 = KG
2 = LB
P2 – Sleep mode:
1 = Turns off sleep mode
2 = Power off 10 minutes later
3 = Power off 20 minutes later
4 = Power off 30 minutes later
P3 – Bits per second:
1 = 9600
2 = 4800
3 = 2400
4 = 1200
P4 – Net weight or Gross weight:
1 = Net Weight
2 = Gross Weight
P5 – Output mode option:
1 = No transmission (RS232
communication stops)
2 = Continuous transmission
3 = Continuous transmission
when stable
4 = Command mode (Z: zero, T: tare,
R: transmit weight data one time)
5 = Current loop output
6 = Keep (Printer)
8
P6 – Backlight setting:
1 = No backlight
2 = Automatic backlight
3 = Keep lighting
P7 – Zero-tracking scope:
1 = 0.5e
2 = 1.0e
3 = 1.5e
4 = 2.0e
5 = 2.5e
6 = 3.0e
7 = 5.0e
P8 – Zero key scope:
1 = 2%FS
2 = 4%FS
3 = 10%FS
4 = 20%FS
P9– Zero scope upon starting:
1 = 2%FS
2 = 4%FS
3 = 10%FS
4 = 20%FS
P10 – digital filtering intensity:
1 = High
2 = Middle
3 = Low
P11 – N/A
P12 – N/A
globalindustrial.com
1. [Unit]: Change unit from “lb” to “kg”.
2. [Count/Settings/Accum OFF]: When in weighting mode, this button
activates counting mode. If button is pressed for about 5 seconds, the
system gets into settings. When in accumulating mode, this mode turns
this mode off.
3. [Accum/Mode]: When in weighting mode, this button adds weight to
the accumulated weight saved in internal memory. When in settings
mode, this button changes the mode to be set up.
4. [Tare/Select]: When in weighting mode, this button is used to determine
net weight. When in count mode, this button selects the digit to be edited.
5. [Zero/Edit]: When in weighing mode, this button resets scale to zero.
When in counting mode, this button edits the selected digit.
6. [On/Off]: When unit is off, this button turns the indicator on. When unit
is on, this button turns the indicator off.
3. Connect the load cell to the indicator
1. The 9-pin socket is used for the link-up of load cell,
as shown in graph 2-3.
2. When the shielded 4-pin cable is used, the indicator does not
have the function of long distance compensation.
3. Indicator must be connected to load cell. Shielded-cable of load cell
must be connected to ground wire. Do not insert or remove plug
when indicator is powered on.
4. Sensor and indicator are static sensitive devices; you must
adopt anti-static measures.
Indicator
6 EX+
EX-
S-
1
2
SH
7 S+
2 S-
3
4
5
6
7
8
9
EX+
S+
IN-
IN+
1 EX9 IN+
8 IN5 SH
globalindustrial.com
Load Cell
Excitation+
(Yellow)
Excitation(Black)
Load Cell Output Signal +
(Blue)
Load Cell Output Signal (Green)
Shield
5
III Operation
1. Power on and auto zero setting
1. Press [ON/OFF] for 3 seconds to power on.
2. The indicator will execute a “999999-000000”count down to self-check when
turned on. Then it will enter weighing mode.
3. When powered on and loading weight on scale deviates from zero point, but
still within zero set range, the indicator will set zero automatically. If out of
range, zero point adjustment is required: please recalibrate or reset system.
2. Zero function
5. Select another digit and
editas previously
explained.
7. Remove the pieces and place
another load to be counted. The
display will show the actual amount
of pieces.
6. When screen shows the
right amount of pieces on
the scale, press [Accum]
and display will show
current number of pieces.
When indicator is in weighing mode and “STABLE” indicator is on, the display
should show “0”. When no load is on the scale, press [Zero] to reset indicator to
“0” when needed. If [Zero] key is invalid, please recalibrate or reset zero
parameters. (Zero operation is only available when “STABLE” indicator is on.
3. Unit setting
When indicator is in weighing mode, press [UNIT] to select the unit desired to
be displayed (lb or kg).
• The indicator will show
above “UNIT” when number displayed is in lbs.
• The indicator will show
above “UNIT” when number displayed is in kg.
8. Press [Count] to go back to weighing mode.
NOTE: Accumulating mode is not available when in counting mode.
5. Accumulating mode (for weight only; max. 999999 kg or lb)
4. Counting Mode
1. Put weight on
scale. Display will
show actual weight.
1. When in weighing
mode and display
shows “0”, place a
known amount of
pieces on the scale.
2. Press [Count]. “COUNT” indicator will
appear on the display.
2. Press [Accum]. Display will show
accumulated total and “ACCUM” indicator
will appear on the display.
3. Replace the load with the new load to
be weighted.
4. Press [Accum] again to show
actual weight of the load in the
display, and press [Accum] again to
show total accumulated weight.
3.Press [Accum]. Screen
will show the number of
pieces to add or subtract.
6
4. Press [Select] to select the digit position wewant to
edit; and edit each digit pressing [Edit] until desired
value is shown.
globalindustrial.com
5. Repeat steps 3 and 4 until complete operation.
6. To erase total accumulated value, press [Accum] to show total accumulated
value and press [Accum Off] button.