MPM-U650 datasheet
MAGIC POWER TECHNOLOGY
MPM-U650 series Preliminary Specification
Preliminary
SPECIFICATION
For
SWITCHING POWER SUPPLY
M/N: MPM-U650 series
Revision Index
2016. Nov. 13
MAGIC POWER TECHNOLOGY
MPM-U650 series Preliminary Specification
FEATURES
 650W
medical power supply with force air
cooling
Peak power up to of 750W
 Single + multi output versions up to 5 outputs
included 5Vsb.
 Input to Output MOPP x 2; Input to Earth, 1 x
MOPP
 Utilizes a thermally efficient U channel chassis
design
 +5Vsb & remote on/off with IP<0.3W(TBD) at no
load standby
 Automatically fan speed control
 Design to meet medical safety regulations
1. Description
Model Name
MPM-U653
MPM-U654-1
MPM-U655
MPM-U656
Output
Voltage
12V
+5Vsb
19V
+5Vsb
24V
+5Vsb
48V
+5Vsb
Rated
Output Power
Peak
Output Power
Line
Regulation
Load
Regulation
Ripple &
Noise
Voltage
Setting
650W
5~10W (TBD)
650W
5~10W (TBD)
650W
5~10W (TBD)
650W
5~10W (TBD)
750W (Target)
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
(Note 1)
(Note 2)
750W (Target)
750W (Target)
750W (Target)
(Note 3)
(Note 3)
(Note 3)
Efficiency
(Note 4)
(TBD)
(TBD)
(TBD)
(TBD)
Note: 1) Rated output is 650W with a built in fan model. In a open frame type with suffix “U” model name, it can be 450W
(TBD) at 50℃ convection cooled environment.
2) The peak output only at input voltage more than 100VAC, and max. 1000mS(TBD).
3) See below Output Specification for detail test condition.
4) Initial setting voltage is at input 115VAC and all outputs at 60% rated load.
5) Total power is 650W.
2. Input Specification
Parameter
Input Voltage(Note 1)
Label Voltage
Input Frequency
Input Current
Conditions/Description
Continuous input range
Inrush Current
Cold start, 25°C
Power Factor
No Load Power
Consumption (Option)
@ 115 VAC
@ 230 VAC
Meets EN 61000-3-2 class D
This function exists only with optional +5Vsb,
model no. suffix "-SB"
Note 1) Output power derating to _TBD_ W when input voltage less than 90VAC.
REV. Nov. 3th 2016
Page 1 of 4
Min.
85
100
47
3.8
Nom.
115 / 230
50 / 60
Max.
264
240
63
8.5
20
40
0.3(TBD) 0.5(TBD)
Units
VAC
VAC
Hz
A
A
W
MAGIC POWER TECHNOLOGY
MPM-U650 series Preliminary Specification
3. Output Specification
Parameter
Efficiency
Conditions/Description
@ 115 VAC
Rated load,
@ 230 VAC
Min.
Nom.
88
90
Minimum load
Max.
89
91
0(Note 1)
Units
%
A
Ripple & Noise
(Pk-Pk)
Rated load, measured by a 20MHz
bandwidth limited oscilloscope and the each
output is connected with a 10µF Electrolytic
Capacitor and a 0.1µF Ceramic Capacitor
±1
%
Line Regulation
Rated load with ±10% changing in input
voltage at 115 / 230 VAC
±1
%
±1
%
Load Regulation
Hold-Up Time
Measured from 60% to 100% rated load and
from 60% to 20% rated load (60% ± 40%
rated load)
Rated load, input 115 / 230 VAC
16 Nom
20 Max
ms
Note: 1) There will be some provisos TBD.
4. Interface Signals and Internal Protection
Parameter
Remote Voltage Sense
Conditions/Description
Compensates for wire voltage drop.
PG signal: When power is turned on, the power good signal will go high 100ms to
Power Good (PG) / Fail (PF)
500ms after all output DC voltages are within regulation limits.
Signal
PF signal: The power fail signal will go low at least 1ms before the output voltages fall
below the regulation limits.
Automatically fan speed control by detecting temp. of unit. 2 step mode (50 / 100%
Fan Speed Control
fan speed).
Protected against output overload, the trigger point between 130%~150% of rated
Overload Protection
load.
Short Circuit Protection
Automatic recovery.
All output voltage except fan supply, latch off mode. The trigger point between
Overvoltage Protection
110%~130% of output voltage.
When the power supply operating over the temperature, the power supply will be shut
Over temperature Protection
down automatically to protect itself. Automatic recovery.
Remote on/off (optional)
The power supply will be turned on when the power On/Off pin is connected to
secondary GND. This function exists only with optional +5Vsb, model no. suffix "SB".
5. Other Options
Parameter
Connectors
Optional multi output
Optional Interface
Conditions/Description
Input: Terminal block or IEC 320/C14 AC inlet.
Output: Terminal block or equivalent. (TBD)
Optional 2 DC/DC converter with each max. 70W output and each max. 24VDC
output. Total output max. power is 650W
Only can be installed on MPM-U653, MPM-U654-1, MPM-U655.
Function TBD
6. Environment Specification
Parameter
Storage Temperature
Relative Humidity
Altitude
REV. Nov. 3th 2016
Conditions/Description
Non-condensing.
Operating
Non-operating
Page 2 of 4
Min.
-40
5
Nom.
Max.
+85
95
3K
4K
Units
°C
%RH
Meter
MAGIC POWER TECHNOLOGY
Operating Temperature
MPM-U650 series Preliminary Specification
Linearly derating from 50°C. Refer derating
curve below. (TBD)
-20
+70
°C
Max.
Units
Derating performance curves (TBD)
7. Safety Approvals, EMI and EMS Specification
Parameter
Approvals
Leakage Current(note 1)
Isolation
EMI (Note 2)
PFC
EMS (Note 2)
Conditions/Description
IEC 60601-1, 3rd edition
EN 60601-1, 3rd edition
ANSI / AAMI ES60601-1:2005
CAN / CSA-C22.2 No. 60601-1:08
N.C.
Earth
S.F.C
N.C.
Touch
S.F.C
N.C.
Patient
S.F.C
Input to Output (MOPP x 2)
Input to PE (MOPP x 1)
Output to PE
EN 60601-1-2
EN 55011
EN 61000-3-2
EN 61000-3-3
± 15 KV air discharge
IEC 61000-4-2
± 8 KV contact discharge
IEC 61000-4-3 10 V / m
IEC 61000-4-4 ± 2KV Line & PE
L - N: ± 1KV
IEC 61000-4-5
L / N - PE: ± 2KV
IEC 61000-4-6 10 V
IEC 61000-4-8 10 A / m
IEC 61000-4-11 Voltage dips >95%
Min.
Nom.
Design to meet
300
1000
100
300
100
300
4
1.5
1.5
B
B
D
A
A
A
A
A
A
A
A
μA
KVAC
Class
Criteria
A
Voltage dips 30%
A
Voltage dips 60%
B
Voltage interruptions >95%
B
Note: 1) according IEC60601-1 test method.
2) As a build-in type power supply, the power supply needs to be installed in a suitable enclosure to pass the EMI/EMC
tests. The final assembly has to comply with the valid EMI/EMC and safety.
REV. Nov. 3th 2016
Page 3 of 4
MAGIC POWER TECHNOLOGY
MPM-U650 series Preliminary Specification
8. Mechanical and Pin Assignment
Parameter
Dimension
Pin assignment
Conditions/Description
229 (L) x 127 (W) x 65 (H) mm, tolerance +/- 0.5mm.
CN1
CN2: main output “+”
CN2B: main output “-“
CN10: Pin 1: 0V
Pin 2: 5Vsb
Pin 3: PG/PF
Pin 4: Remote On/Off ;
CN5:
Pin 1:DC module 1 (+)
Pin 2: DC module 1 (+)
Pin 3: DC module 2 (+)
Pin 4: DC module 2 (+)
Pin 5: option -12V
Pin 6: 0V
Pin 7: 0V
Pin 8: 0V
Pin 9: 0V
Pin 10: 0V
Mechanical Drawing
1. With input terminal block, fan and cover type, estimated size 229 (L) x 127 (W) x 65 (H) mm
(TBD).
2. With input terminal block, open frame type, estimated size 229 (L) x 127 (W) x 65 (H) mm
(TBD).
3. With AC inlet, Fan and cover type Estimated size 229 (L) x 127 (W) x 65 (H) mm (TBD).
4. With AC inlet, open frame type Estimated size 229 (L) x 127 (W) x 65 (H) mm (TBD).
Fan: MF60101V1-10000-G99 (DC12V)
9. Packing Info.
(TBD)
10. Label
(TBD)
REV. Nov. 3th 2016