AC/DC POWER HiTESTER 3334, POWER HiTESTER 3333

AC/DC POWER HiTESTER 3334, POWER HiTESTER 3333
AC/DC POWER HiTESTER 3334
POWER HiTESTER 3333
By Popular Demand
Ideal for Meeting Energy Efficiency Standards
High Accuracy & Cost Performance All in One Device
Choose the 3334 for DC and Current/Power Integration Applications
3334:Compatible with the SPECpower®
benchmarking for server's power consumption
* SPECpower is a registered trademark of Standard
Performance Evaluation Corporation.
3334
n High Accuracy (±0.1% rdg. ±0.1% f.s. for 1 year)
n
Exceeds the 0.5% accuracy benchmark stipulated by international standards.
Extended Period of Guaranteed Accuracy of 3 Years Calibration expenses are reduced by a calibration interval six times that of our former models.
n Maximum Cost Performance
All the necessary functions and accuracy requirements are
provided in an easy-to-use, no-frills device.
3333
2
The HIOKI AC/DC POWER HiTESTER Solves All
3334 AC/DC POWER
All the Features for DC and Current/Power Integration Measurements
n
Complete Accuracy Over a Wide Input Range
1.00mA
0.150V
0.0000W
n
All Measurements Within this Range Fully Guaranteed for Accuracy
Current: 1mA to 30A, Voltage: 0.15V to 300V, Apparent Power: 0W to 9kW
■ Capture
Inrush Current with the Peak
Measurement Function
Measure AC or DC Loads
With a DC to 5kHz frequency bandwidth, all AC and DC measurement and AC/DC elements such as half-wave rectified values
can be tested reliably and accurately
[AC+DC Mode]: For half-wave rectified loads common in small
household appliances such as hair dryers
[DC Mode]: For pure DC loads in batteries
[AC Mode]: For loads in commercial power lines powering common household appliances
Measure for the Peak Value of Voltage and Current for Each Polarity Indepedently. Also measure the inrush current or surge current of electrical
equipment.
Measure simply by pressing the SHIFT+HOLD keys.
inrush current
Switch modes simply by pressing the DC/AC button on the panel
Measure the inrush current when copiers
and similar equipment are started
Half-wave rectified
0.0 A
peak current
All waves rectified
0.0 A
30.00A
300.0V
9.000kW
Some hair dryers operate by switching
the output by a half-wave rectified.
Measure the peak current of the standby
power of home entertainment devices
210mm
100mm
■ Intuitive
80% of Actual Size Shown
15mm
Small Footprint
Setting Procedures and Easy-to-Understand Displays
Both the 3333 and 3334 offer simple operating procedures and quick and easy-to-understand readings and alarm displays. Settings can be made
for obtaining the average of captured data (AVG), VT ratio (conversion ratio), CT ratio, GP-IB address, integration time (from 1 minute to 10,000
hours), and D/A Output Parameters. Information regarding the Power HiTESTER's currents status such as display hold, remote control settings, and
key lock (to prevent unauthorized reconfigurations) can be viewed at a glance.
*Easily test for over-consumption even when testing distorted waveforms that are commonly found in switching power supplies and similar devices by
monitoring for the [PEAK OVER] alarm, simply by setting for the alarm to activate and the display to light up when the input exceeds the range.
of your Energy Consumption Testing Needs
● Meet Industrial Standard Requirements for Test Accuracy
● Measure for Consumed Power
Also ideal for measuring the standby power and power consumption level of household appliances
■ AC/DC Current and Power Integration
Even measure the discharge level of each individual polarity of batteries
■ Accurate
Even for Waveforms
with Large Crest Factors
Reliably test waveforms with large crest factors (CF:peak value with
respect to the RMS value) such as pulsed systems
*Highest effective peak voltage and peak current values on the 3334
are 300% of the range. Accuracy is guaranteed for 1% to 100% of both
ranges.
300mA peak to peak
Measure the amount of solar generated power and how much is being
sold back to the power company
■ Universal
Devices that are highly vulnerable to power fluctuations such as copiers and cycle-controlled equipment can also be measured for integrated power
1mA rms
For example, in the 100mA range
where the
RMS value= 1mA
Peak value= 300mA,
CF=300
Even waveforms such as this can be
measured accurately with the 3334.
Power Supply
Compatible to 100 - 240V AC Power Supplies
■ Evaluate the power
consumption of your servers
Model 3334 is compatible with the SPECpower®
benchmarking criteria for evaluating the power consumption of servers.
l Supported by Ver.1.10 or later.
Link to SPECpower's® Website
http://www.spec.org/power_ssj2008/docs/devicelist.html
* SPECpower is a registered trademark of Standard
Performance Evaluation Corporation.
Rear Panel Design of Model 3334-01
with built-in GP-IB Interface
■ Analog
Output on All 4 Channels
•Simultaneously ouput the voltage, current
and active power values (DC ±2 V f.s., data
refreshed 5 times/second)
•Output the apparent power, power factor, or
integrated current/active power over an additional 1 channel
■ Waveform
Voltage
Current
Make a secure connection with the screw-type terminals
*Use a No.3 Phillips screwdriver
Active power
Actual Size
Integrated active power
Output over 3 Channels
Instantaneous waveforms of the
Waveform of a motor startup
measured voltage, current and
active power can be simultaneously output
•Output: 1 V f.s.
•Sampling speed: 74.4kHz
(at 50Hz: 1488 points/waveform)
(at 60Hz: 1240 points/waveform)
■ Easy-to-connect Terminals
Voltage
Current
Active power
■ Data
management with PC
Ask your distributor for more information
regarding the freeware for processing your
measurement data
Make full use of these interfaces
to increase efficiency
•RS-232C (3334)
•RS-232C, GP-IB (3334-01)
3
4
Fully Answering the Needs for a High Accuracy, Long-lasting, and User-
3333 AC POWER
Accuracy That Can Only Be Realized with a Digital Display
n Multifunctional
- 3 Meters In One
n Quick
Selectable simultaneous display of voltage, current, power
(active or apparent) or power factor
Display a: voltage, current or active power
Display b: current, active power or apparent power
Display c: active power, power factor, voltage or current
Press the SET key to display settings
Display a: AVG setting (number of samples to
average)
convenient for displaying severely
fluctuating values
Display b: VT (voltage transformer) ratio setting
Display c: CT (current transformer) ratio setting
To expand the measurement input
level, VT and CT ratios can be set for
transformer measurements. Primary
winding input values can be read directly.
Operation: Press the a, b or c key to switch between parameters
n Easy-to-see
and Simple Settings
display
Bright emissive numeric and unit symbol
display
Operation:Press NEXT or the a, b or c key to select a setting
item
Actual
Size
nCheck
nClearly visible instrument status
Warning lights indicate out-of-range input to Model
3333 when measuring voltage and current of distorted
waveforms, such as in power switching circuits.
Display hold, remote control and
key-lock states are indicated separately.
nEasy
range setting
Select current range simply by pressing the button
(accepts up to 30 AAC, including auto-ranging)
No voltage range setting required (accepts up to
300 VAC)
for Errors at a Glance
n Convenient
functions
1. Key Lock (SHIFT + KEY LOCK keys)
Useful for avoiding inadvertent operations on production
lines.
2. Data Print (SHIFT + DATA OUT keys)
Obtain hard copies of measurement data using the
optional Printer Model 9442.
5
Friendly Power Measuring Device for the Production and Inspection Lines
Model 3333
What are the advantages?
Measurement accuracy:
±0.5% rdg. or better
Model 3333 fully exceeds the accuracy level of traditional analog
meters that has an accuracy of only ±0.5% f.s.
Period of guaranteed accuracy
(Recommended calibration interval): 3 years
±0.5% f.s is assured for a full three years, reducing calibration costs
and production time losses
Easy Operation
Gone is the need to check for zero-position before measurement as
you would on traditional analog meters
Digital Display
Quickly grasp the measurement data at a glance
210
261
30
Facilitate reporting and data recording needs using your computer
100
Data management on a PC
7
15
Take care of a multitude of measurement needs with a single lowcost instrument
Cost-Performance
n Universal
Power Supply
Terminals
261
34
Screw-in terminal block affixes wires securely.
Because bad wiring connections can present a fire hazard,
the screw-type terminal block has been incorporated to
ensure secure wiring.
15
100
Compatible to 100 - 240V AC Power Supplies
n Easy-to-connect
7
210
*Be sure to use a No. 3 Phillips screwdriver (available as an optional
accessory) to loosen and secure the screws.
Actual Size
n Space-saving
footprint
Smaller installation space.
The installed footprint of the POWER HiTESTER
3333 is about 34% smaller than that of former models.
This size reduction makes the instrument especially easy
to install.
Rear Panel Design of Model 3333-01
with built-in GP-IB Interface
*Rack mounts for various installations also available on
special order. Please inquire for details.
160
n Three-channel analog output
100
Voltage, current and active power measurements
are simultaneously output as +2 VDC f.s. levels
(refreshed about five times per second).
15
n Connect
to Printer Model 9442
Use the optional Printer Model 9442 to print without
concern for troublesome settings.
n PC measurement and data management
m RS-232C interface built-in
m Select Model 3333-01 for additional built-in GP-IB interface
6
n3334 and 3333 Specifications
3334 (AC/DC)
3333 (AC)
● General Specifications
Measurable lines
Measurement parameters
Single-phase, 2-wire (AC/DC)
Single-phase, 2-wire (AC)
Voltage, current, active power, apparent power, power factor,
frequency, integrated current and active power, waveform
peak (voltage and current)
Voltage, current, active power, apparent power, power
factor
Measurement method
Simultaneous digital sampling of voltage and current, True RMS
Sampling Frequency
Approx. 74.4kHz
Measurement Range
Approx. 48kHz
Switch between auto-range or manual
Voltage
15.000/ 30.00/ 150.00/ 300.0V
200.0V
Current
100.00m/ 300.0m/ 1.0000/ 3.000/ 10.000/ 30.00A
50.00m/ 200.0m/ 500.0m/ 2.000/ 5.000/ 20.00A
Power 1.5000W to 9.000kW (refer to range composition table below) 10.000W to 4.000kW (refer to range composition table below)
Frequency bandwidth
DC, 45Hz to 5kHz
45Hz to 5kHz
Guaranteed at 23°C±5, max. 80%rh, sine wave input, power factor=1, in-phase voltage =0V
(accuracy specifications differ depending on usage period of 1 or 3 years)
● Accuracy
Warm-up time
3 minutes
Period of guaranteed accuracy
3 years (better accuracy specifications available for 1-year period)
Effective measurement range
10 minutes
Voltage, current:1% to 100% (Power: 0% to 100%)
Effect of power factor (at pf=0.5)
Temperature Coeffi cient
Maximum ±0.03%f.s./ºC
Values in the ( ) represent the effective measurement range
Measurements below 0.5% of the voltage or current range will be zero suppressed.
●Measurement ranges - Model 3334
Current
100.00mA
300.0mA
(1.00 to 100.00mA)
Voltage
15.000V
(0.150 to 15.000V)
30.00V
(0.30 to 30.00V)
150.00V
(1.50 to 150.00V)
300.0V
(3.0 to 300.0V)
1.5000W
(3.0 to 300.0mA)
(0.0000 to 1.5000W)
3.000W
(0.000 to 3.000W)
15.000W
(0.000 to 15.000W)
30.00W
(0.00 to 30.00W)
4.500W
(0.000 to 4.500W)
1.0000A
(0.0100 to 1.0000A)
15.000W
(0.000 to 15.000W)
9.000W
30.00W
(0.000 to 9.000W)
45.00W
(0.00 to 45.00W)
(0.00 to 30.00W)
150.00W
(0.00 to 150.00W)
90.00W
300.0W
(0.00 to 90.00W)
(0.0 to 300.0W)
Current
50.00mA
(5.00 to 75.00mA)
200.0V
(20.0 to 300.0V)
10.000W
(1.000 to 15.000W)
3.000A
(0.030 to 3.000A)
45.00W
(0.00 to 45.00W)
90.00W
(0.00 to 90.00W)
450.0W
(0.0 to 450.0W)
900.0W
(0.0 to 900.0W)
10.000A
(0.100 to 10.000A)
150.00W
(0.00 to 150.00W)
300.0W
(0.0 to 300.0W)
1.5000kW
(0.0000 to 1.5000kW)
3.000kW
(0.000 to 3.000kW)
200.0mA
(20.0 to 300.0mA)
40.00W
(4.00 to 60.00W)
500.0mA
(50.0 to 750.0mA)
100.00W
(10.00 to 150.00W)
2.000A
(0.200 to 3.000A)
400.0W
(40.0 to 600.0W)
5.000A
(0.500 to 7.500A)
1.0000kW
(0.1000 to 1.5000kW)
●Measurement accuracy - Model 3334
Frequency
Guaranteed
Voltage, current and active power
Current and active power
Period
(at less than 50% of input range)
(at 50% to 100% of input range)
±0.1%rdg.±0.2%f.s.
1 year
3 years
±0.1%rdg.±0.35%f.s.
1 year
±0.1%rdg.±0.1%f.s.
±0.2%rdg.
45 Hz ≤ f ≤ 66 Hz
3 years
±0.1%rdg.±0.2%f.s.
±0.3%rdg.
±0.1%rdg.±0.2%f.s.
±0.3%rdg.
1 year
66 Hz < f ≤ 1 kHz
3 years
±0.1%rdg.±0.35%f.s.
±0.45%rdg.
1 year
±3.0%f.s.
±3.0%rdg.
1 kHz < f ≤ 5 kHz
3 years
±4.5%f.s.
±4.5%rdg.
*Add ±50μA to the accuracy when measuring DC current
*Add (±50μA x voltage value) to the accuracy when measuring DC active power
Notes
DC
●Measurement accuracy - Model 3333
Frequency
45 Hz ≤ f ≤ 66 Hz
66 Hz < f ≤ 1 kHz
1 kHz < f ≤ 5 kHz
Accuracy not defined for current
input exceeding
20A
Values in the ( ) indicate accuracy when input exceeds 100% of range.
Guaranteed
Voltage, current and active power
Current and active power
Period
(input current 20 A or less)
(input current over 20 A)
1 year
3 years
1 year
3 years
1 year
3 years
30.00A
(0.30 to 30.00A)
450.0W
(0.0 to 450.0W)
900.0W
(0.0 to 900.0W)
4.500kW
(0.000 to 4.500kW)
9.000kW
(0.000 to 9.000kW)
Values in the ( ) represent the effective measurement range
Measurements below 1% of the voltage, current range will be zero suppressed.
●Measurement ranges - Model 3333
Voltage
Voltage, current, power: 10% to 150%
Maximum ±0.4%±rdg. (45 to 66Hz)
±0.1%rdg.±0.1%f.s. (±0.2%rdg.)
±0.1%rdg.±0.2%f.s. (±0.3%rdg.)
±0.1%rdg.±0.2%f.s. (±0.3%rdg.)
±0.1%rdg.±0.35%f.s. (±0.45%rdg.)
±3.0%f.s. (±3.0%rdg.)
±4.5%f.s. (±4.5%rdg.)
Notes
Accuracy not defined for current
input exceeding
20A
20.00 A
(2.00 to 30.00A)
4.000kW
(0.400 to 6.000kW)
●3334 and 3333 Arithmetic Expressions
Measurement
Formula
Parameters
Apparent Power
S=U×I
(S)
Power Factor (λ)
λ=P/S
(Sum of I from start of integration)
Integrated Current
(1 hour of data)
Integrated Active (Sum of P from start of integration)
Power
(1 hour of data)
*U=Tested Voltage Value, I=Tested Current Value, P=Tested
Active Power Value
Calculating precision is ±1dgt. against the results obtained
from each measured value
Current and active power integration available only on Model
3334.
7
3334 (AC/DC)
3333 (AC)
● Input
Input impedance
2.4 MΩ for voltage, 10 mΩ or better (50/60 Hz) for current
2.4 MΩ for voltage, 7 mΩ or better (50/60 Hz) for current
Maximum input voltage
300V, ±425Vpeak
300 Vrms, 425 Vpeak
Maximum input current
30 A, ±54.0Apeak *1
30 Arms, 42.5 Apeak
Maximum effective peak voltage
±300% of each voltage range, Within ±425Vpeak
Within 425Vpeak
Maximum effective peak current
±300% of each current range, Within ±54.0Apeak *1
±300% of each current range, Within ±42.5Apeak
Max. rated voltage to earth
300V (DC, 50/60Hz)
300V (50/60Hz)
voltage and current: 0.5% to 105% of range
active power: 0% to 110.25% of range
voltage and current: 1% to 152% of range
active power: 0% to 231.04% of range
● Display
Display indication range
Displacement power factor
0.000 to 1.000 (no polarity display)
Display refresh rate
Response time
approx. 5 times per second
within 0.5 s (time to rated accuracy after abrupt change in input [0 to 90% or 100 to 10% of range])
●Functions
Integration measurement
Wave peak measurement
Rectification method
Analog output
(D/A output)
Waveform output
Average function
VT or CT ratio
External Interfaces
Miscellaneous
No.of displayed digits: Six digits
Current Integration:
from 0.00000mAh, Polarity-independent integration and Sum value
Active power Integration:
from 0.00000mWh, Polarity-independent integration and Sum value
Integration time: 1 min to 10000 h
Measurement accuracy:
measurement accuracy of active power ±1dgt.
Maximum value of positive and negative waveform of voltage/current (up to 300% of full scale range)
Measurement accuracy: ±1.2%f.s. ("f.s." is 300% of each range)
Switchable between AC+DC(True RMS), DC(simple
AC(True RMS)
average display) and AC(True RMS)
Parameter output representation:
Parameter output representation:
voltage, current and active power (3 simultaneous channels)
voltage, current and active power (3 simultaneous channels)
D/A select an item from current integration, active power integration,
Voltage output: +2 VDC f.s. for each range
apparent power, power factor
Output accuracy: ±0.5% f.s. + individual measurement accuracy
Voltage output: ±2 VDC f.s. for each range
Output accuracy: ±0.5% f.s. + individual measurement accuracy
Parameter output representation:
voltage, current and active power (3 simultaneous channels)
Voltage output: 1 VDC f.s. for each range
Output accuracy: ±1.0% f.s. + individual measurement accuracy
Simple averaging of specified number of samples: 1, 2, 5, 10, 25, 50 or 100
VT ratios: 1, 2, 4, 10, 20, 30, 60, 100
VT ratios: 1, 2, 4, 10, 20, 30, 60, 100
CT ratios: 1,2,3,4,5,6,8,10,12,15,16,20,24,25,30,40,50,60,
CT ratios: 1,2,3,4,5,6,8,10,12,15,16,20,24,25,30,40,50,60,
75, 80,100,200,300,500,1000,2000,3000,5000,
75,80,100
10000
RS-232C interface: included as standard, Asynchronous communication method: full-duplex; Baud rate: 9600 bps (fixed)
GP-IB interface: Model 3334-01 only
GP-IB interface: Model 3333-01 only
IEEE-488.1 1987 compliant, IEEE-488.2 1987 reference IEEE-488.1 1987 compliant, IEEE-488.2 1987 reference
Display Hold (HOLD), Key Lock (KEYLOCK),
Display Hold (HOLD), Maximum value hold, Peak value hold,
Settings backup (preserves settings)
Key Lock (KEYLOCK), Backup function (preserves settings,
integration data)
●General Specifications
Safety
EMC
EN61010 Pollution Factor 2,
Measurement Category III (4000 V anticipated overvoltage)
EN61326, EN61000-3-2, EN61000-3-3
Operating environment
0 to 40 ºC, 80% RH or less, non-condensating
Storage environment
-10 to 50 ºC, 80% RH or less, non-condensating
Rated supply voltage
100 to 240 VAC, 50/60 Hz
Maximum rated power
Size and weight
20 VA
210 mm (8.27 in)W × 100 mm (3.94 in)H × 245 mm (9.65 in)D
(excluding feet and projections), 2.5 kg (88.2 oz)
*1 Supported by Ver.1.10 or later.
160 mm (6.30 in)W × 100 mm (3.94 in)H × 227 mm (8.94 in)D
(excluding feet and projections), 1.9 kg (67.0 oz)
■ Operate the Power HiTESTER from Your PC
■ Customized versions of
Model 3334 also available
on Special Order!
Data Management is as Easy as 1-2-3
■ RS-232C (built-in with the 3334 and 3333)
■ RS-232C, GP-IB (built-in with the 3334-01 and 3333-01)
1/10 of current ranges
Free RS-232C application for both models available from your authorized HIOKI distributor only.
Model 3334-20
•Features and Functions
Most appropriate for
standby power measurement
1. Operate the Power HiTESTER's keys on the PC as
you would on the actual unit
2. Further process test data on spreadhseet software
such as Excel
3334
Compatible to IEC62301
(standby power measurement)
l Current ranges 10.000mA to 3.000A
l Active power ranges
1.5000W to 450.0W (150V range)
3.000W to 900.0W (300V range)
3333
Please contact your HIOKI distributors
for further details.
Use of the software require a comprehensive understanding of the protocols and commands.
Support for modifications to the software not available.
-01 Model
-01 Model
Model : AC/DC POWER HiTESTER 3334
Model : POWER HiTESTER 3333
Model No. (Order Code) (Note)
3334
3334-01
3334-20
Model No. (Order Code) (Note)
3333
3333-01
(with GP-IB)
(Built to order model, 1/10 of current ranges)
(with GP-IB)
Accessories: Instruction manual ×1, Power cord ×1
Accessories: Instruction manual ×1, Power cord ×1
nOptions (Common to both Models 3334 and 3334-01)
RS-232C CABLE (9-pin to 9-pin, crossed cable/1.8m(0.07ft)) 9637
RS-232C CABLE (9-pin to 25-pin, crossed cable/1.8m(0.07ft)) 9638
GP-IB CONNECTOR CABLE (2m) 9151-02
No. 3 Phillips screwdriver
RS-232C CABLE 9637
GP-IB CONNECTOR CABLE 9151-02
No. 3 Phillips screwdriver
Print method : Thermal serial dot printing
Paper width : 112 mm(4.41ft)
Power supply : AC adapter 9443-02, or supplied nickel-metal hydride battery
Dimensions and weight : 160W(6.30")× 66.5H(2.62") × 17D(0.67") mm, 580g(20.5oz.)
■ Option Printer (For the 3333 and 3333-01)
PRINTER 9442
CONNECTION CABLE (for printer 9442) 9444
RECORDING PAPER 1196
AC ADAPTER (for printer 9442 operation in Europe, except Switzerland) 9443-02
CONNECTION CABLE 9444
AC ADAPTER 9443-02
When purchasing the Printer 9442, make sure you also purchase the Connection cable 9444
and AC adapter 9443-02 so that you can connect it to the 3333/3333-01.
Note: Company names and Product names appearing in this catalog are trademarks or registered trademarks of various companies.
HIOKI (Shanghai) SALES & TRADING CO., LTD.
TEL +86-21-63910090 FAX +86-21-63910360
http://www.hioki.cn / E-mail: info@hioki.com.cn
DISTRIBUTED BY
HIOKI SINGAPORE PTE. LTD.
TEL +65-6634-7677 FAX +65-6634-7477
HEADQUARTERS
E-mail: info-sg@hioki.com.sg
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 FAX +81-268-28-0568
HIOKI KOREA CO., LTD.
http://www.hioki.com / E-mail: os-com@hioki.co.jp TEL +82-2-2183-8847 FAX +82-2-2183-3360
E-mail: info-kr@hioki.co.jp
HIOKI USA CORPORATION
TEL +1-609-409-9109 FAX +1-609-409-9108
http://www.hiokiusa.com / E-mail: hioki@hiokiusa.com
All information correct as of Aug. 31, 2016. All specifications are subject to change without notice.
3334E11-68E
Printed in Japan
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising