π H-4597, H-4598 OHAUS VALOR™ 3000 XTREME

π H-4597, H-4598 OHAUS VALOR™ 3000 XTREME
π H-4597, H-4598
1-800-295-5510
uline.com
OHAUS VALOR™ 3000
XTREME SCALES
OVERVIEW OF CONTROLS
CONTROL PANEL
CONTROL PANEL PARTS
2
1
OZ
3
#
DESCRIPTION
1
O/T On/Off Button
2
Mode/Unit Menu Button
3
Enter Cal Button
CONTROL FUNCTIONS
BUTTON
O/T
Mode/Unit
ON/Off
Menu
ON
PRIMARY FUNCTION
(Short Press)
If indicator is OFF, turns scale on.
If indicator is ON, sets zero.
Yes
MODE
ENTER
Steps through active units
and modes.
UNIT
Enters menu.
PAGE 1 OF 4
No menu function.
CAL
Initiates span calibration
process.
NO
MENU FUNCTION
(Short Press)
Stores 100% reference in
% mode.
Stores reference sample in
SG mode.
OFF
Turns the indicator off.
Cal
No
Tares weight of an item on pan.
SECONDARY FUNCTION
(Long Press)
Enter
Toggles through available
settings.
YES
Selects/accepts displayed
setting.
1114 IH-4597
OPERATION
MENU NAVIGATION
PERCENT MODE
TO ENTER THE MENU MODE
Use this mode to measure the weight of a sample as a
percentage of a reference weight.
1. Press and hold MODE/UNIT/Menu button until SEtUP is
displayed, then release it.
2. Press Yes to enter the displayed menu or press No to
advance to the next menu.
3. Once in the menu, press Yes to view the menu item
setting or press No to move to the next menu item.
4. When viewing the setting, press Yes to accept the
setting or press No to change the setting.
5. When End is displayed, press Yes to advance to the
next menu or No to return to the top of the current
menu.
INITIAL CALIBRATION
When the scale is first installed, it should be calibrated
to ensure accurate weighing results.
1. Press and hold the ENTER/Cal button until CAL is
displayed.
2. When the button is pressed again and released, the
display shows -C- while zero reading is stored. Next, the
display shows the calibration weight value in grams.
3. Place the specified calibration weight (sold
separately) on the pan. See table below for
appropriate weights.
1. Press MODE/UNIT/Menu until Set.rEF and the % icon
are displayed.
2. Press No to use the reference weight in memory and
proceed to step 6.
3. Press Yes to establish a new reference weight. Scale
display will now blink until reference is stored.
4. Put an empty container on the pan, then press
0/T/On/Off to tare the container weight.
5. Add the desired amount of material to the container,
then press ENTER/Cal to store the displayed weight.
The display shows 100%.
6. Replace the reference material with new material.
The display shows the percentage of the sample
compared to reference weight.
SPECIFIC GRAVITY MODE
Use this mode to measure the specific gravity of a
sample.
1. Press MODE/UNIT/Menu until Set.rEF and the S icon
are displayed.
2. Press No to use the reference weight in memory and
proceed to step 6.
3. Press Yes to establish a new reference weight. Scale
display will now blink until reference is stored.
CALIBRATION WEIGHTS
ULINE MODEL NO.
H-4597
H-4598
OHAUS MODEL NO.
V31XW3
V31XW6
Calibration Weights
3 kg
6 kg
4. The display shows -C- while the reading is stored.
When the display returns to the previous application
mode, scale is ready for use.
4. Put an empty container on the pan, then press
0/T/On/Off to tare the container weight.
5. Fill the container with water, then press ENTER/Cal
to store the displayed weight. The display shows the
water weight with the S icon blinking.
WEIGHING MODE
6. Replace the water with the same volume of sample
material, then press ENTER/Cal. The display shows
the specific gravity of the sample.
Use this mode to measure the weight of a sample in the
selected unit of measure.
FLUID OUNCE MODE
1. Press MODE/UNIT/Menu until the desired unit icon is
displayed.
Use this mode to measure the fluid ounce volume of a
sample.
2. Put an empty container on the pan, then press
0/T/On/Off to tare the container weight.
1. Press MODE/UNIT/Menu until the FL icon is displayed.
3. Add material to the container. The display shows the
weight of the material.
PAGE 2 OF 4
2. Put an empty container on the pan, then press
0/T/On/Off to tare the container weight.
3. Fill the container with sample material. The display
shows the volume of the sample.
1114 IH-4597
OPERATION CONTINUED
LEGAL FOR TRADE
Figure 1
When model V31XW3 (H-4597) or V31XW6 (H-4598) is
used in trade or a legally controlled application, it must
be set up, verified and sealed in accordance with local
weights and measures regulations. It is the responsibility
of the purchaser to ensure that all pertinent legal
requirements are met.
SETTINGS
Before verification and sealing, perform the following
steps:
1. Set the security switch on the underside of the scale
to the OFF position. (See Figure 1)
2. Set LFT to OFF under Setup menu.
3. Calibrate the scale.
Security Switch
ON
OFF
VERIFICATION
The local weights and measures official or authorized
service agent must perform the verification procedure.
SEALING
The local weights and measures official or authorized
service agent must apply a security seal to prevent
tampering with the settings. Refer to Figure 2 for sealing
methods.
Figure 2
WIRE SEAL:
Remove the spacers under the
cross-drilled screws. Reinstall
the screws and feed the wire
through the cross-drilled holes.
4. Enter the Units menu.
5. Set the desired units of measure to ON.
6. Exit the Units menu.
PAPER SEAL:
Place the paper seals over the
screws and sealing spacers.
7. Enter the Setup menu.
8. Set the desired Filter level.
9. Set the desired Power on Unit setting.
10. Set LFT to ON.
11. Set the Security Switch to ON. (See Figure 1)
12. Exit the Setup menu.
TROUBLESHOOTING
OPERATING ISSUE
CAUSES
RECOMMENDATIONS
Cannot turn on.
No power to scale.
Verify connections and voltage.
Poor accuracy.
Improper calibration.
Unstable environment.
Perform calibration.
Move scale to suitable location.
Cannot calibrate.
Unstable environment.
Incorrect calibration weight.
Move scale to suitable location.
Use correct calibration weight.
Cannot access mode.
Mode not enabled.
Enter menu and enable mode.
Cannot access unit.
Unit not enabled.
Enter menu and enable unit.
Err 8.1
Pan has load during power on.
Remove weight from pan and re-zero.
Err 8.2
Pan was removed prior to power on.
Install pan and re-zero.
OVER
Weight on pan exceeds capacity.
Remove weight from the pan.
UNDER
Pan was removed during weighing.
Reinstall pan.
Err 9
Factory calibration data corrupted.
Contact Ohaus or an authorized service agent.
Perform service calibration.
REF Err
Reference weight is too small.
Use larger sample.
LOWrEF
Reference weight is too low for accurate percent weighing.
Continue to weigh with less accurate results.
If the problem persists, or if the troubleshooting section does not resolve or describe your problem, contact Uline Customer Service
at 1-800-295-5510.
PAGE 3 OF 4
1114 IH-4597
TECHNICAL DATA
The technical data is valid under the following ambient
conditions:
• Ambient temperature: 32ºF to 104ºF (0°C to 40°C)
• Relative humidity: 10% to 95% relative humidity,
non-condensing
• Power: AC adapter 9VDC at 1.2A max. output
• Protection: Protected against dust and water
• Pollution degree: 2
• Installation category: Class III
• Height above sea level: Up to 13,100 ft. (4,000 m)
ULINE MODEL NO.
H-4597
H-4598
OHAUS MODEL NO.
V31XW3
V31XW6
Capacity (g)
3000
6000
Repeatability/Accuracy (g)
0.5
1
Linearity (g)
±1
±2
III
III
Approval Class
Ingress Protection
IP65
IP65
Weighing Units
Application Modes
Tare Range
g, kg, lb, oz, lb:oz
Weigh, Percent Weighing, Specific Gravity, Fluid Ounces
To capacity by subtraction
Stabilization Time
Power Requirements
Calibration
≤ 3 seconds
9 VDC at 1.2A max. AC adapter (supplied) or four C (LR14) batteries (not supplied)
Digital with external weight
Display Type
6-digit, 7-segment LCD with white LED backlight
Display Size
15mm / 0.6" digits
Keypad
Pan Size (mm/in)
3-button overlay
146 x 158 / 5.7 x 6.2
Net Weight (kg/lb)
1.4 / 3.1
Shipping Weight (kg/lb)
2.2 / 4.8
π
CHICAGO • ATLANTA • DALLAS • LOS ANGELES • MINNEAPOLIS • NYC/PHILA • SEATTLE • MEXICO • CANADA
1-800-295-5510
PAGE 4 OF 4
uline.com
1114 IH-4597