GCX M-series Articulating Arm Installation guide

GCX M-series Articulating Arm Installation guide

The M-series Articulating Arm is a mounting solution for Philips IntelliVue MP5/20/30/40/50/60/70, MX400/450/500/550/600/700/800 monitors on Dräger Apollo left side. This guide provides instructions on attaching the arm to the machine and mounting the monitor.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

GCX M-series Articulating Arm Installation Guide | Manualzz

3875 Cypress Drive Petaluma, CA 94954 800.228.2555 707.773.1100 Fax 707.773.1180 www.gcx.com

Installation Guide

Articulating Arm for Mounting Philips IntelliVue MP5/20/30/40/50/60/70, MX400/450/500/550/600/700/800 on Dräger Apollo Left Side

The purpose of this guide is to describe mounting of Arm and related parts on left side of machine.

WARNING: USE OF MOUNTING HARDWARE AND MONITORING COMPONENTS OTHER THAN

THOSE DESCRIBED IN THIS DOCUMENT MAY RESULT IN SERIOUS INJURY DUE TO TIPPING OF

THE ANESTHESIA MACHINE.

WARNING: ALL DEVICES AND MOUNTING EQUIPMENT SHOULD BE POSITIONED INWARD, CLOSE

TO THE APOLLO, DURING TRANSPORT.

Installer: When installation is completed, provide these instructions and all related documentation to the end user for future reference.

Parts Reference

The following parts and hardware are included in this installation kit (hardware not shown):

Item # Description Qty

1 M-series Articulating Arm 1

2 Nut Plate 2

3

4

5

6

M6 x 22mm Flat Head Socket Cap Screw (FHSCS)

M6 x 8mm Flat Head Machine Screw (FHMS)

M6 x 12mm FHMS

Down Post, 6''

2

1

2

1

6'' Down Post

7

8

9*

#10-32 x 3/8'' Socket Head Cap Screw (SHCS)

5/32'' Hex Wrench

Spacer

3

1

1

10* M6 x 16mm Pan Head Machine Screw (PHMS)

*See "Non-Table Mount" on p.4 for use of these parts.

2

Spacer

M-series

Articulating Arm

Table Mount

(Supplied by Philips)

DU-AG-0018-70 Rev B 12/11/13 GCX Corp Page 1 of 5

Tools Required: 5/32'' hex wrench (provided), Phillips screwdriver (not provided), 1/2'' socket wrench (not provided).

Attaching Down Post to M-Series Arm

1. Remove plastic bolt cap from bottom of Swivel Cup (not shown).

2. Using the 5/32'' hex wrench provided, fasten Post to Swivel Cup with three (3) #10-32 x 3/8'' SHCS as shown below.

M-Series Arm

Swivel Cup

#10-32 x 3/8'' SHCS (3)

6'' Post

DU-AG-0018-70 Rev B 12/11/13 GCX Corp Page 2 of 5

Mounting Arm in Left Side Accessory Track

1. Thread one (1) M6 x 22mm FHSCS a few turns into each Nut

Plate, leaving 15 mm of thread exposed per screw (right).

2. Insert Nut Plate/screw assemblies in top and bottom slots, holding bottom screw in place to prevent it from falling out of slot (right).

3. While holding bottom Nut Plate in position (below left), guide both Nut Plates into track and move entire assembly up track to desired mounting position.

4

.

Using the 5/32'' hex wrench provided, tighten bottom screw first, then top screw (below right).

Guide Nut Plates into Track

M6 x 22mm FHSCS (2)

Tighten Bottom

Screw First

Nut Plate (2)

Insert Nut Plates/Screws into Slots

Positioning Arm in Track

WARNING: Do not attempt to remove the Arm from the track while an instrument is mounted.

While *supporting the weight of the mounted instrument, loosen the two (2) flat head socket cap screws

(FHSCS) in the Slide. A 5/32'' hex wrench is provided for this task. Push up or pull down at the end of the arm nearest the Slide. Tighten screws to secure Arm in track.

* It may be necessary to dismount instrument before adjusting Arm in track.

FHSCS (2)

DU-AG-0018-70 Rev B 12/11/13 GCX Corp Page 3 of 5

Attaching Philips Table Mount to Mounting Adapter

Installation Note: Do not attach the Table Mount to the Arm/Mounting Adapter if you will be mounting a monitor equipped with the mounting interface shown in photo (right). See "Non-Table Mount" (bottom of page) for instructions specific to this type of mounting plate

Do not attach Table Mount if monitor is equipped with this mounting interface.

1. Fasten Philips-supplied Table Mount to Mounting Adapter on the Arm with one (1) M6 x 12mm FHMS and two (2) M6 x

8mm FHMS as shown below.

M6 X 12mm FHMS (1)

M6 X 8mm FHMS (2)

Table Mount

*Supplied by Philips

Mounting Adapter

2. Mount monitor in accordance with Philips Table Mount instructions or follow procedure below for Non-Table Mount or the MX400/450/500/550 Fixed Mount Adapter (PH-0073-15).

Non-Table Mount

1. Fit Spacer into recessed area on top of Mounting Adapter (below). Ensure Spacer remains between Mounting Adapter and monitor.

2. Insert two (2) M6 x 16mm PHMS through bottom of Mounting Adapter and thread into mounting holes in bottom of monitor (right).

Spacer

Mounting Adapter

Mounting the MX400/450/500/550 Monitor (Fixed Mount Adapter)

M6 x 16 mm PHMS (2)

Mount the monitor in accordance with separate Installation Guide DU-PH-0073-15 provided with the Mounting Adapter, ordered separately .

DU-AG-0018-70 Rev B 12/11/13 GCX Corp Page 4 of 5

M-Series Arm Operation and Adjustment

Refer to Installation Guide provided with the M-Series Articulating Arm (DU-WMM-0001).

Routine Maintenance

Periodically check all mounting hardware. Tighten as necessary for optimal operation.

Cleaning the Mounting Assembly

CAUTION: GCX makes no claims regarding the efficacy of the listed chemicals or processes as a means for controlling infection. Consult your hospital’s infection control officer or epidemiologist. To clean or sterilize mounted instruments or accessory equipment, refer to the specific instructions delivered with those products.

1. The mounting assembly may be cleaned with most mild, non-abrasive solutions commonly used in the hospital environment (e.g. diluted bleach, ammonia, or alcohol solutions).

2. The surface finish will be permanently damaged by strong chemicals and solvents such as acetone and trichloroethylene.

3. Do not use steel wool or other abrasive material to clean the mounting assembly.

4. Damage caused by the use of unapproved substances or processes will not be covered by warranty. We recommended that you test any cleaning solution on a small area of the arm that is not visible to verify compatibility.

5. Never submerge or allow liquids to enter the arm. Wipe any cleaning agents off of the arm immediately, using a waterdampened cloth. Dry the arm thoroughly after cleaning.

DU-AG-0018-70 Rev B 12/11/13 GCX Corp Page 5 of 5

advertisement

Key Features

  • Articulating Arm
  • Mounts Philips IntelliVue monitors
  • Dräger Apollo left side
  • Table Mount
  • Non-Table Mount
  • Down Post
  • Swivel Cup
  • Spacer
  • Mounting Adapter

Frequently Answers and Questions

How do I attach the down post to the M-series arm?
Remove the plastic bolt cap from the bottom of the swivel cup. Using the hex wrench provided, fasten the post to the swivel cup with three socket head cap screws.
How do I mount the arm in the left side accessory track?
Thread one flat head socket cap screw a few turns into each nut plate, leaving 15 mm of thread exposed. Insert the nut plate/screw assemblies in the top and bottom slots. Guide both nut plates into the track and move the assembly up to the desired position. Tighten the bottom screw first, then the top screw.
Can I remove the arm from the track while an instrument is mounted?
No, do not attempt to remove the arm from the track while an instrument is mounted.
How do I attach the Philips table mount to the mounting adapter?
Fasten the Philips-supplied table mount to the mounting adapter on the arm with one flat head machine screw and two flat head machine screws.
How do I clean the mounting assembly?
The mounting assembly may be cleaned with most mild, non-abrasive solutions commonly used in the hospital environment. Do not use steel wool or other abrasive material to clean the mounting assembly.