DFI 250 Quick Start Manual

DFI 250 Quick Start Manual
DFI 250
DIGITAL INDICATOR
QUICK START MANUAL
www.cooperinstruments.com
PH: (540) 349-4746 • FAX: (540) 347-4755
CONTENTS
1.0 INTRODUCTION.................................................................................................................................. 1
1.1 Approvals (for trade versions only) .................................................................................................1
1.2 Manuals ...............................................................................................................................................1
2.0 SHIPPING CONTENTS...................................................................................................................... 1
3.0 SPECIFICATIONS............................................................................................................................... 1
4.0 WARNINGS .......................................................................................................................................... 2
4.1 General................................................................................................................................................2
4.2 Configuration Issues.........................................................................................................................2
5.0 INSTALLATION ................................................................................................................................... 2
5.1 Electrical Safety .................................................................................................................................2
5.2 Panel Mount Template.......................................................................................................................2
5.3 Special Function Key.........................................................................................................................2
5.4 GSE-LINK............................................................................................................................................3
5.4.1 GSE-LINK Activation ..................................................................................................................3
6.0 CONNECTIONS................................................................................................................................... 3
6.1 Cable Connections.............................................................................................................................3
6.2 DC Power (DC PWR +, DC PWR –)...................................................................................................3
6.3 Load Cell Connection........................................................................................................................4
6.3.1 4-Wire Connection ......................................................................................................................4
6.3.2 6-Wire Connection ..........................................................................................................................4
6.4 Auxiliary Connections.......................................................................................................................4
6.4.1 Direct Personal Computer Link (RXD, TXD, GND)..................................................................4
6.4.2 Printer Connections (RXD/TXD, GND and DTR)......................................................................5
6.4.3 Remote Display (TXD, GND)......................................................................................................5
6.4.4 Remote Input ...............................................................................................................................5
6.4.5 Outputs.........................................................................................................................................5
6.5 Connecting Shields............................................................................................................................6
6.5.1 Cable Shield Connection and Earthing ....................................................................................7
6.6 Regulatory Sealing Requirements...................................................................................................7
6.6.1 Sealing..........................................................................................................................................7
7.0 INSTRUMENT SETUP........................................................................................................................ 7
7.1 Calibration Counter............................................................................................................................7
7.1.1 Trade Critical Settings................................................................................................................8
7.2 GSE-LINK............................................................................................................................................8
7.3 Access Full Setup ..............................................................................................................................8
7.4 Access Safe Setup.............................................................................................................................8
7.5 Exit Full or Safe Setup.......................................................................................................................8
7.6 Settings ...............................................................................................................................................8
8.0 ERROR MESSAGES........................................................................................................................12
8.1 Weighing Errors...............................................................................................................................12
8.2 Setup and Calibration Errors..........................................................................................................12
8.3 Diagnostic Errors.............................................................................................................................12
9.0 WARRANTY REPAIR POLICY.......................................................................................................13
CF 146
ii
V- Rev. 1.0 - 003R-620-100
1.0 INTRODUCTION
This manual contains information on the installation, calibration and setup of the instrument.
1.1 Approvals (for trade versions only)
• NSC approval (4000 divisions at 0.8µV/division).
• NMI approval (4000 divisions at 0.8µV/division).
• C-tick approved and CE approved.
1.2 Manuals
For more information on this instrument refer to the Reference Manual or Quick Start Manual.
2.0 SHIPPING CONTENTS
The following table identifies the items shipped with the indicator. Please check that your packing box contains the
specified items.
Shipped Items
• Indicator
• Operator Manual
• Quick Start Manual
• Trade Label (plastic)
• Panel Mount Template
• Function Key Overlay Stickers
Other Items (Optional)
• GSE-LINK Cable
• Power Supply
• Desk Mount with Battery Compartment
• U Bracket
3.0 SPECIFICATIONS
Performance
Resolution
Zero Cancellation
Span Adjustment
Stability/Drift
Excitation
A/D Type
A/D Conversion Rate
Operating Environment
Case Materials
Packing Weights
Digital
Display
Setup and Calibration
Digital Filter
Zero Range
Power Input
Standard Power Input
Variants
AC
Battery
Features
GSE-LINK Data Coupling
Correction
Outputs
Assignable Function Key
CF 146
Up to 30,000 divisions, minimum of 0.25µV/division, 20 updates/second (Trade
4000 divisions at 0.8µV/division)
±2.0mV/V
0.1mV/V to 3.0mV/V full scale
Zero: < 0.1µV/°C (+ 8ppm of deadload max)
Span < 8 ppm/°C, Linearity < 20ppm, Noise < 0.2µVp-p
5 volts for up to 4 x 350 or 8 x 700 ohm load cells (4-wire or 6-wire plus shield)
Maximum total load cell resistance: 1,000 ohms
24bit Sigma Delta with 8,388,608 internal counts
20Hz with FIR filtering > 80dB
Temperature: 14 to 122°F ambient
Humidity: <90% non-condensing
Storage: –4 to +122°F ambient
IP55 when panel mounted
ABS, Silicon Rubber, Nylon, Acrylic (no halogen used)
Basic Indicator: 0.75lb
LED Backlit LCD with six 0.8in high digits with units and annunciators
Full digital with visual prompting in plain messages
Sliding window average from 0.1 to 4.0 seconds
Adjustable from ±2% to ±20% of full capacity
12 to 24VDC (2.5 VA max) - ON/OFF key with memory feature
Wall Transformer: 110/240VAC 50/60Hz in 12VDC 0.5A out
4 x AA batteries (Alkaline or rechargeable NiMH, NiCad, etc.)
Infra-red Connector for optional GSE-LINK PC cable (to RS-232 PC port)
Five point linearity correction
RS-232 automatic transmit, network or printer outputs. Transmission rate: 2400,
4800 or 9600 baud
Unit switching, counting, manual hold, peak hold, live weight and totalizing
1
V- Rev. 1.0 - 003R-620-100
Drive Outputs
Battery Backed Clock Calendar
2 isolated transistor drive outputs (300mA total at 50VDC)
Battery life 10 years minimum
4.0 WARNINGS
4.1 General
• Indicator not to be subject to shock, excessive vibration or extremes of temperature (before or after installation).
• Inputs are protected against electrical interference, but excessive levels of electro-magnetic radiation and RFI
may affect the accuracy and stability.
• For full EMC or for RFI immunity, termination of cable shields and correct earthing of the instrument is
essential.
• Indicator and load cell cable are sensitive to excessive electrical noise. Install well away from any power or
switching circuits.
4.2 Configuration Issues
• Configuration and calibration can be performed from the front panel, using digital setup. When Full Setup is
used, all menu items are accessible and care must be taken to ensure no accidental changes are made to
calibration and trade settings.
• Enter a passcode to prevent unauthorized or accidental tampering. If the passcode is lost, GSE Scale Systems
should be contacted for further advice.
5.0 INSTALLATION
The following steps are required to set up the indicator.
• Inspect indicator to ensure good condition.
• Use connection diagrams to wire up load cell, power and auxiliary cables as required.
• Use the drill hole template provided for hole locations.
• Connect Power to indicator and press <POWER> key to turn the instrument On.
• Refer to the Instrument Setup section page 15 for information on configuring and calibrating the instrument.
• To turn instrument Off press and hold <POWER> key for three seconds (until display blanks).
5.1 Electrical Safety
• For your protection all electrical hardware must be rated for environmental conditions of use.
• Pluggable equipment must be installed near an easily accessible power socket outlet.
• To avoid the possibility of electric shock or damage to the instrument, always switch off or isolate the
instrument from the power supply before maintenance is carried out.
5.2 Panel Mount Template
Use the panel mount template for drill hole locations. The template indicates positions for the two 4mm mounting
screws through the panel. Also displayed on the template is the position of the rectangular hole that should be cut
to allow for the connection of cables. The drilling template supplied with the indicator allows for front or rear
machining of the panel.
5.3 Special Function Key
• The Special Function Key on the instrument ships as a blank key.
• If any of the special functions are to be used on the indicator it is important that the matching function key
overlay sticker (supplied) is applied to the keypad.
• Ensure keypad is clean and dry before affixing sticker.
CF 146
2
V- Rev. 1.0 - 003R-620-100
5.4 GSE-LINK
The optional GSE-LINK cable can be used to transfer setup and
calibration information from a PC (e.g. to be stored for later use and/or
transferred to other instruments). It can also be used to download software
upgrades to the instrument from a PC.
• Attach the GSE-LINK cable to the PC using the DB9 connector.
• Attach the GSE-LINK head to the left side of the instrument display using
the permanent magnet located within the head of the GSE-LINK.
WARNING: The GSE-LINK head contains a strong magnet and care should be taken with its proximity to
electronic media (e.g. credit cards, floppy disks, etc.) and/or other electronic instrumentation.
5.4.1 GSE-LINK Activation
A long press of the <GROSS/NET> key will toggle the GSE-LINK infrared communications On/Off.
When the GSE-LINK has been enabled the following will occur:
• The instrument briefly displays the prompt rin-L.
• The editing annunciators (i.e. GRP, ITM, etc.) will flash for up to five minutes while the instrument searches for
activity. During this period, the instrument also disables the RS-232 communications.
• Activity Located: If the instrument is successful in locating activity, the editing annunciators will continue to
flash during the entire period of communications.
• No Activity Located: If the instrument fails to locate activity, the GSE-LINK will be disabled and the editing
annunciators will stop flashing. The instrument will also revert back to the normal RS-232 communications (i.e.
the SERIAL:TYPE setting will be re-activated).
6.0 CONNECTIONS
6.1 Cable Connections
• All cable connections are made to the rear of the instrument using screwless terminals.
• Wires must be stripped of insulation by at least 10mm.
• To install, depress the orange lever beside the terminal required and push wire into the hole. Release the lever
and pull gently on the wire to ensure it is securely trapped in the terminal.
6.2 DC Power (DC PWR +, DC PWR –)
• The DC supply need not be regulated, provided that it is free of excessive electrical noise and sudden
transients.
• The instrument can be operated from a high quality AC wall transformer as long as there is sufficient capacity
to drive both it and the load cells.
CF 146
3
V- Rev. 1.0 - 003R-620-100
6.3 Load Cell Connection
The instrument may be connected for either 4-wire or 6-wire operation. For more information, refer to
BUILD:CABLE setting section 7.6.
6.3.1 4-Wire Connection
The minimum connectivity requirements are the connection of four wires (i.e. Excitation + and – along with Signal +
and –).
The BUILD:CABLE option must be set to 4 to allow for 4-wire connection.
6.3.2 6-Wire Connection
The excitation and signal lines are connected the same as for a 4-wire installation.
The BUILD:CABLE option must be set to 6 (the default) to allow for 6-wire connection.
6.4 Auxiliary Connections
This section provides diagrams to illustrate the terminal connections.
6.4.1 Direct Personal Computer Link (RXD, TXD, GND)
RS-232 - Instrument to PC using COM Port (DB9)
CF 146
4
V- Rev. 1.0 - 003R-620-100
RS-232 - Instrument to PC using COM Port (DB25)
6.4.2 Printer Connections (RXD/TXD, GND and DTR)
RS-232 – Instrument to Printer (DB25)
6.4.3 Remote Display (TXD, GND)
• Connect TXD to RXD and GND to GND on the remote display.
6.4.4 Remote Input
• The indicator requires a voltage free contact between TXD and RXD to enable the remote input (i.e.
SPEC:REM.FN).
Note: The remote input will not function when in setup or when using the GSE-LINK.
WARNING
The remote input is a
voltage free contact (e.g.
button, mechanical
relay). Connection of
any active circuitry may
damage the instrument.
6.4.5 Outputs
• Output drivers for the instrument are isolated open emitter transistor drives that are capable of driving up to a
total of 300mA.
CF 146
5
V- Rev. 1.0 - 003R-620-100
• This configuration allows for the direct connection of the instrument outputs to most types of PLC.
• The voltage applied to the COM terminal appears on the output lines (i.e. OUT1 and OUT2) when the outputs
are active (e.g. to connect to a PLC connect +24V to the common terminal). The outputs can then be
connected directly to PLC inputs so when the outputs are active the PLC will see a 24V signal.
• To drive external loads (e.g. relays), connect the relay coil positive supply to the output common and the output
line directly to one side of the relay coil.
• Connect the other end of the relay coil to the negative supply. It is recommended that fly-back diodes or
transient suppressors be fitted across relay coils to limit switching noise.
Outputs to Drive Relay
Outputs to Drive PLC
6.5 Connecting Shields
To obtain full EMC or for RFI immunity, cable shields MUST be connected to the earth lug on the rear of the
instrument.
This figure shows the connecting cables restrained using cable ties fastened around the cable strain relief anchors.
CF 146
6
V- Rev. 1.0 - 003R-620-100
6.5.1 Cable Shield Connection and Earthing
• Care should be taken when connecting shields to maximize EMC or RFI immunity and minimize earth loops
and crosstalk (interference) between instruments.
• For full EMC or for RFI immunity, termination of the cable shields at the earth lug is very important. The earth
lug of the instrument must be separately connected to ground potential via a reliable link.
• The instrument should only be connected to earth via a single reliable link to avoid earth loops.
• Where each instrument is separately earthed, interconnecting cable shields should be connected at one end
only.
• Caution: Some load cells connect the cable shield directly to the load cell (and therefore the scale base).
Connection of the load cell cable shield in this situation may be site specific.
6.6 Regulatory Sealing Requirements
To comply with regulatory sealing requirements for each instrument, (i.e. to ensure instruments are not accidentally
or deliberately tampered with), it is important that proper sealing procedures be adhered to.
6.6.1 Sealing
Affix sealing stickers to the rear of the instrument, over one or more screws in the locations indicated.
Also affix a sealing sticker over the load cell cable where the cable-tie strain relief is attached, as indicated.
Affix stickers in the locations indicated.
7.0 INSTRUMENT SETUP
7.1 Calibration Counter
The built-in calibration counter(s) monitor the number of times trade critical settings are altered. Refer to Trade
Critical Settings below for more information and to the OPTION:USE setting section 7.6.
The table below describes when the counter(s) will increment.
Industrial or OIML: The Calibration Counter increments when trade critical settings are changed. An example of
the counter is C.00019.
NTEP: Two counters display.
The Calibration Counter increments when trade critical settings in the Calibration (CAL) menu are changed. An
example of the counter is C.00010.
The Configuration Counter increments when other trade critical settings (i.e. not in the CAL menu) are changed.
An example of the counter is F.00009.
CF 146
7
V- Rev. 1.0 - 003R-620-100
7.1.1 Trade Critical Settings
Trade critical settings can affect calibration and/or legal for trade performance. In this document the
symbol
indicates the setting is trade critical. Each time a trade critical setting is altered, the calibration counter will be
incremented by one.
7.2 GSE-LINK
For information on setting up using the GSE-LINK cable refer to GSE-LINK section 5.4.
7.3 Access Full Setup
Full Setup provides access to configure and calibrate the instrument. All items in all menus will be enabled in Full
Setup.
WARNING: Care should be taken to avoid inadvertently altering the Build or Calibration settings.
• Ensure the instrument is on.
• Press and hold the <POWER> and <FUNCTION> keys together for two seconds.
7.4 Access Safe Setup
Safe Setup restricts access to the trade critical settings (marked with
• Ensure the instrument is on.
).
• Press and hold the <POWER> and <ZERO> keys together for two seconds.
7.5 Exit Full or Safe Setup
To save settings, exit setup and return to the normal weighing mode use one of the following methods:
• Method 1: Press <POWER> and <FUNCTION> keys together for two seconds.
• Method 2: Press <POWER> and <ZERO> together for two seconds.
• Method 3: Select - End - from the menus.
Warning: If the power is interrupted while in setup (i.e. by disconnecting the power cable or pressing the
<POWER> key), unsaved settings will be lost.
7.6 Settings
The following tables identify the settings available for the instrument.
CF 146
8
V- Rev. 1.0 - 003R-620-100
GROUP
(GRP)
BUILD
OPTION
CAL
CF 146
ITEM
(ITM)
SELECT
(SEL)
EDIT
(EDT)
OK
Underline = Defaults
000000, 00000.0, 0000.00, 000.000, 00.0000, 0.00000
000100 to 999999 Default = 003000
<SEL> changes position, <EDT> changes digit.
1, 2, 5, 10, 20, 50, 100
none, g, kg, lb, t
OFF, ON
4, 6
INDUST, OIML, NTEP
none, 0.2, 0.5, 1.0, 2.0, 3.0, 4.0
DP
CAP
Decimal Point Position
Maximum Capacity
RES
UNITS
HI.RES
CABLE
USE
FILTER
MOTION
Resolution(Count-By)
Weighed Units
Resolution x 10 Mode
4-Wire or 6-Wire
Scale Use
Reading Average
(time in seconds)
Motion Detection
INIT.Z
Z.TRAC
Z.RANGE
Initial Zero on Startup
Zero Tracking Sensitivity
Zero Operating Range
Z.BAND
Zero 'Dead' Band
ZERO
Zero Calibration Routine
Set between zero and full scale
(Current weight displays)
SPAN
Span Calibration Routine
Set between zero and full scale
(Current weight displays)
ED.LIN
Edit Linearization Points
Set between zero and full scale
L1. - - - Select Linearization point
1 to 5
(L1, L2, L3, L4, L5).
(Approx. % of full scale)
CLR.LIN
Clear Linearization Points
L1. - - - Select Linearization point
1 to 5
9
OFF, 0.5-1.0, 1.0-1.0
Default = 0.5 Graduations per Second
OFF, ON
OFF, SLOW (0.5 grads/sec), FAST (10 grads/sec)
-2_ 2, -1_ 3, -20_20
Default = -2_ 2 (–2% to +2%)
<SEL> changes position, <EDT> changes digit.
Default = 0
<SEL> key to start. Current weight displays.
Remove all weight.
<SEL>, <EDT> or <OK> starts routine (Z.in P displays).
<ITM> key to exit, <SEL>, <EDT> or <OK> to repeat routine.
<SEL> key to start. Current weight displays.
Add test weight.
<SEL> or <OK> to show calibration weight value.
Set correct weight. <SEL> changes position, <EDT> changes digit.
<ITM> or <OK> starts routine (S.in P displays).
<ITM> key to exit, <SEL>, <EDT> or <OK> to re-edit calibration weight and repeat
routine.
<SEL> to step through list of points.
<OK> to change selected point or <ITM> to exit.
After <OK> current weight displays.
Add test weight.
<OK> to enter corrected weight. <SEL> changes position,
<EDT> changes digit.
<OK> starts routine (L.in P displays).
<ITM> key to exit, <OK> to repeat routine.
<SEL> to step through list of points.
<OK> to select point to clear or <ITM> to exit.
After <OK> press <ITM> to choose Cont. Y (Yes) or Cont. N (No).
V- Rev. 1.0 - 003R-620-100
Accept
Accept
Accept
Accept
Accept
Accept
Accept
Accept
Accept
Accept
Accept
Accept
Accept
SPEC
DIR.ZER
(L1, L2, L3, L4, L5)
Direct Zero Calibration
(Current weight displays)
DIR.SPN
Direct Span Calibration
(Current weight displays)
FAC.CAL
SAFE.PC
Factory Calibration
Cont. N (No)
Cont. Y (Yes)
Safe Setup Passcode
FULL.PC
Full Setup Passcode
KEY.LOC
Front Panel Key Locking
P12345
(P for Power key. Other keys
numbered from the left, ie.
Zero=1.)
<FUNCTION> Key
Setting
Auto Power Off /
Battery Operation
KEY.FN
AUT.OFF
SERIAL
SET.PTS
CF 146
B.LIGHT
Backlight Operation
REM.FN
Remote Function
TYPE
BAUD
BITS
Serial Output Type
Serial Baud Rate
Serial Format Options
Position 1: Parity
Position 2: Data Bits
Position 3: Stop Bits
Position 4: DTR Handshake
ADDRES
Instrument Address
RST.CON
Reset Printed
Consecutive Numbers to 1
Cont. N (No)
Cont. Y (Yes)
Weight Source
SRC
10
<ITM> to return to CLR.LIN.
<OK> key to start. Current weight displays.
<OK> to enter direct zero setting (in mV/V). <SEL> changes position, <EDT>
changes digit.
<OK> key to store new zero calibration.
<ITM> to exit or <OK> to repeat operation.
<OK> key to start. Current weight displays.
<OK> to enter direct span setting (in mV/V). <SEL> changes position, <EDT>
changes digit.
<OK> key to store new span calibration.
<ITM> to exit or <OK> to repeat operation.
Cont. N
Warning: Choosing Cont. Y will restore default factory calibration in BUILD and
CAL menus.
(000000 no passcode). Set up to 6 digit passcode.
<SEL> changes position, <EDT> changes digit.
Activated only when FULL.PC is also set.
(000000 no passcode). Set up to 6 digit passcode.
<SEL> changes position, <EDT> changes digit.
P12345
Character indicates key is unlocked.
(–) Dash indicates key is locked.
<SEL> changes position, <EDT> changes digit.
Accept
NONE, TEST, COUNT, UNITS, HOLD, PEAK.H, LIVE.WT, SHOW.T
Accept
NEVER, 1, 5, 10 (time in minutes)
Default = Never powers off automatically
(Battery: powers down after 30 minutes)
ON, OFF
(Automatically turns indicator off after 10 seconds of inactivity)
NONE, KEY1 to KEY5, BLANK
(Activated only when SERIAL:TYPE is set to KEY)
NET, AUTO.1, AUTO.2, PRINT, MASTER, KEY
2400, 4800, 9600
N 8 1 - (Default Serial Format Options)
<SEL> changes position, <EDT> changes digit.
Parity: N None, O Odd, E Even
Data Bits: 7 or 8 data bits
Stop Bits: 1 or 2 stop bits
DTR: (-) DTR disabled or d DTR enabled
01 to 31
<SEL> changes position, <EDT> changes digit.
Cont. N
<ITM> to choose Cont. Y (Yes) or Cont. N (No)
Accept
OFF, GROSS, DISP (Displayed)
Accept
V- Rev. 1.0 - 003R-620-100
Accept
Accept
Accept
Accept
Accept
Accept
Accept
Accept
Accept
Accept
TARG.HI
FORMAT
YEAR
Target for Overweight
Setpoint 1 (Output 1)
Target for Underweight
Setpoint 2 (Output 2)
Date Format Setting
Year Setting
MONTH
Month Setting
DAY
Day Setting
HOUR
Hour Setting
MINUTE
Minute Setting
SCALE
Scale Base Test Display
FRC.OUT
Force Outputs
FACTRY
DEFLT
- END -
EXIT SETUP
Restore Factory Defaults
Cont. N (No)
Cont. Y (Yes)
Save settings and return to
normal weighing mode
TARG.LO
CLOC
TEST
CF 146
11
-99999 to 999999
Default = 000000
-99999 to 999999
Default = 000000
dd.mm.yy or mm.dd.yy
2000 to 2099
<SEL> changes position, <EDT> changes digit.
01 to 12
<SEL> changes position, <EDT> changes digit.
01 to 31
<SEL> changes position, <EDT> changes digit.
00 to 23
<SEL> changes position, <EDT> changes digit.
00 to 59
<SEL> changes position, <EDT> changes digit.
Display reads in milliVolts-per-Volt (factory calibrated to 0.1%)
Verify the correct load cell capacity and/or load cell wiring is used. Remove weight
from scale.
<EDT> advances through outputs (i.e. ON.1 and ON.2)
<OK> turns outputs off and exits test.
Default = OFF
Cont. N
Warning: Choosing Cont. Y will clear all stored data except BUILD and CAL
menus.
Accept
Accept
Accept
Accept
Accept
Accept
Accept
Accept
Accept
Accept
V- Rev. 1.0 - 003R-620-100
8.0 ERROR MESSAGES
8.1 Weighing Errors
• Check Setup = This item can be checked on site by service personnel.
Error
(U - - - - -)
Description
The weight is below the minimum allowable weight reading.
(O - - - - -)
The weight is above the maximum allowable weight reading.
Warning – overloading may damage mechanical scale elements.
(ZERO)
(ERROR)
The weight reading is beyond the limit set for Zero operation. The
operation of the <ZERO> key is limited in the setup during
installation. The indicator cannot be Zeroed at this weight.
Scale motion has prevented a <ZERO>, <TARE> or <PRINT>
operation from occurring on command.
(STABLE)
(ERROR)
Resolution
Increase the weight or decrease
the minimum allowable weight
reading.
Check the condition of load cell
connections. Check for
damaged load cell.
Increase the Zero Range
(Z.RANGE) or use the <TARE>
key instead.
Try the operation again once the
scale is stable.
8.2 Setup and Calibration Errors
Error
(ENTRY)
(DENIED)
(RES)
(LO)
(RES)
(HIGH)
(SPAN)
(LO)
Description
The instrument may be in Safe Setup
and an item that needs Full Setup has
been selected for editing.
When accessing setup, more than
three attempts have been made with
the incorrect passcode.
An attempt has been made to place a
linearization point below zero.
An attempt has been made to place a
calibration point too close to an
existing calibration point.
The scale build is configured for less
than 100 graduations.
The scale build is configured for more
than 30,000 graduations.
The load cell signal range (span) is too
small for these settings.
(SPAN)
(HI)
The load cell signal range (span) is too
large for these settings.
(ZERO)
(LO)
(ZERO)
(HI)
An attempt has been made to calibrate
zero below -2mV/V.
An attempt has been made to calibrate
zero above +2mV/V.
(LIN.PT)
(LO)
(PT.TOO)
(CLOSE)
Resolution
Access Full Setup to edit the item.
Turn the instrument off. When the instrument is turned back
on, enter the correct passcode to access setup.
Incorrect linearization point entered (must be between zero
and full scale).
Re-enter the calibration point. Points must be spaced by at
least 2% of full scale from each other.
Check the resolution (countby) and capacity settings.
Check the resolution (countby) and capacity settings.
Incorrect span weight entered (must be between zero and
full scale). Scale wiring incorrect. Wrong load cell capacity
(too large). Wrong or no calibration weight added to scale.
Incorrect span weight entered (must be between zero and
full scale). Scale wiring incorrect. Load cell capacity too small
for application.
Scale wiring incorrect.
Remove all weight from scale. Scale wiring incorrect.
8.3 Diagnostic Errors
• Check: This item can be checked on site by service personnel.
• Return for Service: The instrument must be returned to Cooper Instruments for factory service.
Error
(E0001)
(E0002)
(E0010)
(E0020)
CF 146
Description
The power supply voltage is too low.
The power supply voltage is too high.
The temperature is outside of allowable limits.
Scale build is incorrect. The number of graduations has been set
12
Resolution
Check supply
Check scale/cables
Check location
Fix up scale build
V- Rev. 1.0 - 003R-620-100
(E0100)
(E0200)
(E0300)
(E0400)
(E0800)
(E2000)
(E4000)
(E8000)
too low or too high.
The digital setup information has been lost.
The calibration information has been lost.
All setup information has been lost
The factory information has been lost.
The EEPROM memory storage chip has failed
ADC Out of Range Error. This may be caused from a broken load
cell cable.
The battery backed RAM data has lost data.
The FLASH program memory is incorrect
Re-enter setup
Re-calibrate
Enter setup and calibrate
Return for Service
Return for Service
Check BUILD:CABLE setting.
Check load cell cable, wiring, etc.
Re-enter setup
Return for Service
The E type error messages are additive. For example, if instrument is running off batteries and the temperature
drops, the battery voltage may be too low. The resulting error messages will be E0011 (0001 + 0010).
The numbers add in hexadecimal as follows:
1-2 -3 -4 -5 -6 -7 -8 -9 -A-B -C-D -E-F
(For example, 2 + 4 = 6, or 4 + 8 = C)
9.0 WARRANTY REPAIR POLICY
Limited Warranty on Products
Any Cooper Instruments product which, under normal operating conditions, proves defective in material or in
workmanship within one year of the date of shipment by Cooper will be repaired or replaced free of charge provided
that a return material authorization is obtained from Cooper and the defective product is sent, transportation
charges prepaid, with notice of the defect, and it is established that the product has been properly installed,
maintained, and operated within the limits of rated and normal usage. Replacement or repaired product will be
shipped F.O.B. from our plant. The terms of this warranty do not extend to any product or part thereof which, under
normal usage, has an inherently shorter useful life than one year. The replacement warranty detailed here is the
buyer’s exclusive remedy, and will satisfy all obligations of Cooper whether based on contract, negligence, or
otherwise. Cooper is not responsible for any incidental or consequential loss or damage which might result from a
failure of any and all other warranties, express or implied, including implied warranty of merchantability or fitness for
particular purpose. Any unauthorized disassembly or attempt to repair voids this warranty.
Obtaining Service under Warranty
Advance authorization is required prior to the return to Cooper Instruments. Before returning the item, contact the
Repair Department c/o Cooper Instruments at (540) 349-4746 for a Return Material Authorization number.
Shipment to Cooper shall be at buyer’s expense and repaired or replacement items will be shipped F.O.B. from our
plant in Warrenton, Virginia. Non-verified problems or defects may be subject to a $100 evaluation charge. Please
return the original calibration data with the unit.
Repair Warranty
All repairs of Cooper products are warranted for a period of 90 days from date of shipment. This warranty applies
only to those items that were found defective and repaired; it does not apply to products in which no defect was
found and returned as is or merely recalibrated. It may be possible for out-of-warranty products to be returned to
the exact original specifications or dimensions.
* Technical description of the defect: In order to properly repair a product, it is absolutely necessary for Cooper to
receive information specifying the reason the product is being returned. Specific test data, written observations on
the failure and the specific corrective action you require are needed.
CF 146
13
V- Rev. 1.0 - 003R-620-100