Digital Weighing Indicator

Digital Weighing Indicator
Digital Weighing Controller
SI 4300
Instruction Manual
Digital Weighing Indicator
SI 4300
CONTENTS
1. Before Installation
---------------
3 page
2. Introduction
---------------
4 page
3. Specification
---------------
5 page
4. Installation
---------------
11 page
4-1. External Dimension & Cutting Size
---------------
11 page
4-2. Load Cell Installation
---------------
11 page
4-3. Installation Components
---------------
12 page
---------------
14page
5-1. Calibration
---------------
14 page
5-2. TEST Weight Calibration Mode
---------------
14 page
5-3. Simulation Calibration Mode(Without Test Weight)
---------------
17 page
5-4. Set-up
---------------
20 page
5-5. F-Function
---------------
21 page
---------------
35 page
6-1. Serial Interface (RS-232C)
---------------
35 page
6-2. Current Loop Interface
---------------
39 page
6-3. Print Interface (Centronics Parallel Interface)
---------------
40 page
6-4. Analog Output Interface (: 0~10V)
---------------
42 page
6-5. Analog Output Interface (4~20mA)
---------------
43 page
6-6. Serial Interface (RS-422 / 485)
---------------
44 page
6-7. BCD INPUT
---------------
45 page
6-8. BCD OOUTPUT
---------------
46 page
6-9. Command Mode
---------------
47 page
---------------
49 page
7-1. Load Cell Installation
---------------
49 page
7-2. Calibration Process
---------------
49 page
7-3. Indicator Error & Treatment
---------------
50 page
7-4. Indicator Test Mode
---------------
51 page
---------------
52 page
5. Set-up
6. Interface
7. Error & Treatment
Warrantee Certification
-2-
Digital Weighing Indicator
SI 4300
1. BEFORE INSTALLATION
1-1. Caution / Warning Marks
Warning
This mark warns the possibility to arrive death or serious injury in
case of wrongly used.
Caution
This mark cautions the possibility to arrive serious human body
injury or product lose in case of wrongly used.
1-2. Other Marks
Warning for Electric Shock or Damage.
Please do not touch by hand
Protective Ground(Earth) terminal
Prohibition of Operation process
1-3. Copy Rights
1).All Right and Authority for this Manual is belonged to Sewhacnm Co.,Ltd.
2).Any kinds of copy or distribution without Sewhacnm Co.,Ltd’s permission will be prohibited.
1-4. Inquiries
If you have any kinds of inquiries for this model, please contact with your local agent or Head Office.
Head Office : Sewhacnm Co.,Ltd.
Website : www.sewhacnm.co.kr
Email : sales@sewhacnm.co.kr
-3-
Digital Weighing Indicator
SI 4300
2. INTRODUCTION
2-1. Introduction
Thank you for your purchasing “SI 4300” Industrial Digital Weighing Controller.
This “SI 4300” model is advanced model for Check Weighing and Testing application, with powerful
communication performance.
With 2ports serial port interfaces and precise weighing control system, you can upgrade your weighing
process.
This “SI 4300” Weighing Controller is suitable for any kinds of Check weighing application with tow
different weighing modes (Photo Sensor mode and Hold input mode).
Enjoy your process with “SI 4300” Weighing Controller.
2-2. Cautions
1). Don’t drop on the ground or avoid serious external damage on item.
!
2). Don’t install under sunshine or heavy vibrated condition.
Cautions 3). Don’t install place where high voltage or heavy electric noise condition.
4). When you connect with other devices, please turn off the power of item.
5). Avoid from water damage.
6). For the improvement of function or performance, we can change item specification without prior
notice or permission.
7). Item’s performance will be up-dated continuously base on previous version’s performance.
2-3. Features
1). All Modules and Option Cards are isolated to maximize accuracy and performance.
2). External input terminal inside.(4pcs:Can be set by F11 mode)
3). By using “Photo-Coupler” on each module(Option, Analog board, In/Out), we improved “Impedance
problem”, “Isolation ability among inputs”, “Leading power problem”, and “Noise covering function”.
4). Data back-up function, when the sudden power off.
5). “Set value Error” check function added. – Check “Set values for each weighing mode”.
- If there is any wrong set value, “E” will be display and will not start weighing process
6). Polycarbonate film panel, strong for dust and water.
7). Weight Unit selection Function added. (“g”, “kg”, “t” selectable – F40)
8). Variable options(Order in advance, Refer Chapter 6. Interface) – “RS-232C” Standard installed.
9) 2port Serial Interface available – various applications are available.
10) Simulation Calibrate mode added (Can calibrate without Test weight)
-4-
Digital Weighing Indicator
SI 4300
3. SPECIFICATION
3-1. Analog Input & A/D Conversion
Input Sensitivity
0.2㎶ / Digit
Load Cell Excitation
DC 10V ( - 5V ~ + 5V )
Max Input Signal
Max3.2mV/V
[Zero] ±16PPM/℃
Temperature Coefficient
[Span] ±3.5PPM/℃
Input Noise
±0.3㎶ P.P
Input Impedance
Over 10㏁
A/D Conversion Method
Sigma-Delta
A/D Resolution(Internal)
520,000 Count(19bit)
A/D Sampling Rate
Max 500times / Sec
Non-Linearity
0.005% FS
Display Resolution(External)
1/20,000
3-2. Digital Part
Display
Parts
Specification
7Segments, 7digits VFD green Color
Main Display
Size :12.7(H) ×7.0(W)mm
Min. Division
×1, ×2, ×5, ×10, ×20, ×50
Max display value
+999,950
Under Zero value
"-" (Minus display)
Steady, Zero, Tare, Hold, Gross, Print, Comm.
" ▼" Condition display Lamp
kg, g, t / Over, Under, Pass
Red / Yellow-Green LED Display(3Ø )
Display
Status lamp
Number Key, Function, CAL. Lock key (14pcs)
Key
3-3. General Specification
Power Supply
AC110/220V±10%), 50/60Hz, about 30VA
Operating Temperature Range
-10℃ ~ 40℃
Operating Humidity Range
Under 85% Rh (non-condensing)
External Dimension
200mm(W)x105mm(H)x165mm(L)
Net Weight(kg)
About 2.3kg
Gross Weight(kg)
About 3.0kg
※ AC 110V, Power supply is an optional before ex-factory.
-5-
Digital Weighing Indicator
SI 4300
3-4. Option Card
Option No.1
Printer Interface : Centronics Parallel
Option No.2
Analog Output (0~10V or 0~5V)
Option No.3
Analog Output (4~20mA)
Option No.4
Serial Interface : RS-232C / 422 / 485
Option No.5
BCD INPUT (P/N change purpose)
Option No.6
BCD Output
Option No.7
Ethernet
※ Serial Interface (RS-232C) or Current Loop is Standard installed.
In the Optional Serial port, there is no Current Loop function.
3-5. Front Panel (Display / Key Pad)
Weight Unit LED
Part No.
Count
Weight Display
Over Range
setting
Status Lamp
Under Range
Setting
Function key
Calibration Lock KEY
3-5-1. Status Lamp (ANNUNCIATORS) : “▼” Lamp is “ON”.
Steady
When the weight is Steady, “▼” Lamp is turn on.
When the current weight is Zero, “▼” Lamp is turn on.
Zero
(Displayed weight is Zero, “▼” Lamp is turn on.)
Tare function is set, “▼” Lamp is turn on.
Tare
Hold
(Tare Reset  “▼” Lamp is turn off.)
When “Hold” key is input, “▼” Lamp is turn on.
When display Gross weight(Net weight + Tare Weight), “▼” Lamp is turn on.
Gross
Print
(Under F19-01 setting)
When print key input or Auto print, “▼” Lamp is turn on.
When indicator transfers or receives data from other devices, “▼” Lamp is turn on. (If the
Comm.
“▼” is off although there is some data transference, please check communication settings).
-6-
Digital Weighing Indicator
SI 4300
3-5-2. Key Operation
Make Weight value as Zero.
Under F08, you can set the Zero key operation range, as 2%, or 5%, or 10%, or 20% of
Max Capacity.
※ Under “Tare” key input, Zero key will not be activated.
Make Weight value as Zero, including Tare Weight.
Under F09, you can set the Tare key operation range, as 10%, 20%, 50%, or 100% of Max
Capacity.
Tare setting : Under F10-00 setting, “TARE” key input
Under F10-01 setting, “Tare”+ No. key + “Enter”
Remove set TARE value.
To Set the “HOLD” function or Rest the “Set HOLD” function
Under F21-02 setting, give the Check weighing start signal by pressing this key.
Under TARE Setting condition, can select “Gross weight” or “Net Weight” display
mode.
First input : Gross Weight will be displayed.
Second input : Net Weight will be displayed.
You can set each weighing process as a certain P/N.
Each weighing process will be saved with FINAL, PRE1, PRE2, and Free Fall set
value.(Max 50 kinds of P/N you can set)
And you can call certain P/N with each set value.
Check the “Accumulated weight” and “Count” for current Part No. or All Part No.s
Sub Total : F15-00 setting
Grand Total : F15-01 setting
Set the “Over” range of current Part No.’s Check weighing limit.
-7-
Digital Weighing Indicator
SI 4300
Set the “Under” range of current Part No.’s Check weighing limit.
1. Manual Print (F38-00 setting, under F35-01)
2. Manual weighing Data save for accumulated weighing count and weight.
(F01-00 / 02 setting)
3. Calibration mode
- Digit setting
Whenever pressing “0”key, digit will be change 1, 2, 5, 10, and 50.
- Decimal point position
Whenever pressing “0”key, decimal point will be change.
1. Modify the set value during setting process.
2. Calibration mode
- Move back to previous step.
F-function Mode.
- F-function Exit : Press “Clear” key, once.
- F-Test Mode Exit : Press “Clear” key, twice.
1. Save set value during setting process.
2. Calibration mode
- Save current setting and move to next step.
3. F-Function mode - Save current F-function setting, and move to next F-function
1. “F-TEST” Mode Entrance : Press “F” key for 5sec.
2. Under “F-function Mode”, Move to next Function or move to certain function No.(Press
function No. and press “F” key)
3. Function key (Refer “Function keys”)
Enter/Exit to “Calibration” mode.
-8-
Digital Weighing Indicator
SI 4300
※ “Function Keys”(Combined Key Function)
Reset the Count No. on the sub display window.
Only, reset the display – Original count data will be saved on the memory.
Print all P/Ns’ accumulated weighing count and weight.
(Grand-Total Print)
Print current P/N’s accumulated weighing count and weight.
(Sub-Total Print)
Delete all P/Ns’ accumulated weighing count and weight
(If you set F44-01, the data will be automatically deleted after “Grand-Total Print).
Delete current P/N’s accumulated weighing count and weight
(If you set F44-01, the data will be automatically deleted after “Sub-Total Print).
※ After Pressing
key, you have to input above function keys within 5sec. - After 5sec, the
key activation is loose
※ If you set “F51-01” you can check the
※ After Pressing “
”, “
”
/
key activation through Main display.
key, non-function keys are input, the “
activation will be loose.
-9-
”, “
”
key
Digital Weighing Indicator
SI 4300
3-6. Rear Panel
②
3
○
1
○
6
○
1
○
7
○
5
○
4
○
① POWER AC IN
- Power switch : Power on/off switch.
- Fuse : AC250V / 0.5A , φ5.25 , 20mm.
!
Warning
- AC IN : Available Input AC 110V / 220V.
※ The standard power supply is AC 220V(Fixed when ex-warehouse), if you want to have AC
110V, please inform in advance.
② Option Card 1
③ Option Card 2
※Option Card Connector installed for Optional Interface or Output.
(Printer I/F, Analog out, RS-422/485, or RS-232C(two port)
④ LOAD CELL Connector (N16-05)
⑤ SERIAL I/F
“RS-232C” or “CURRENT LOOP”(9Pin, D-Type Female) are built-in as standard
⑥ External Input : External control input for wired remote control.
Refer to F-Function F11 to select desired function mode.
Input signal
…………………………… Optical-Isolator
⑦ Relay Output Terminal : Set point(SP1, SP2, SP3, SP4) and Finish, Empty relay output.
(Refer “F21” setting.)
- 10 -
Digital Weighing Indicator
SI 4300
4. INSTALLATION
4-1. External Dimension & Cutting Size
External Dimension ,Cutting Size (unit : mm)
186
165
92
105
200
162
94
188
4-2. Installation Components
Power Cable
Communication Connector(D-SUB 9P)
- 11 -
Load-cell Cable
Digital Weighing Indicator
SI 4300
4-3. Load Cell Installation
4-3-1. Load Cell Connector Specification
RED(EXC+)
WHITE(EXC-)
GREEN(SIG+)
BLUE(SIG-)
YELLOW(SHLD)
4-3-2. Load Cell Installation
1) You can connect Max 8pcs of same capacity Load cells at once. (350Ω)
2) You have to make horizontal balance on the ground.
3) If you install more than 2pcs of Load cells, use Summing box and adjust output signal difference as
minimum. It can make wrong weighing process caused by each load cell’s variation.
4) If there is some temperature difference around Load cell, it can cause wrong weight measurement.
5) Don’t do Welding job or Arc discharge around installation place. But, there is no choice, please disconnect
power cable and Load cell cable.
6) If you measure static electricity material, please make earth between down part and up part of Load cell.
- 12 -
Digital Weighing Indicator
SI 4300
4-3-3. Formula to plan the precise weighing system
This “SI 4300” weighing controller’s Max input sensitivity is 0.2㎶ / Digit.
And for precise weighing system, the following formula must be satisfied.
!
Caution
Caution : “Input sensitivity” means Min. output voltage variation of weighing part to change 1digit. So,
please do not make large input voltage to make reliable weighing system.
Single Load cell use
EⅹBⅹD
A : Load cell capacity(kg)
A
B : Load cell Voltage(mV)
0.2㎶ ≤
EⅹBⅹD
Plural Load cells use
D : Digit
E : affirmation Voltage of Load cell
0.2㎶ ≤
AⅹN
N : Number of Load cell
Example1.)
Number of Load cell : 1pcs
Load cell capacity : 500kg
Load cell Voltage : 2mV/V
Digit : 0.05kg
Affirmation Voltage of Load cell : 5,000mV
Max Capacity of Weighing System : 300kg
Then, estimation result for this weighing system with formula,
5000ⅹ2ⅹ0.05
= 1 ≥ 0.2㎶
500
The calculated value is larger than 0.2㎶, so this
system has no problem.
Example2.)
Number of Load cell : 4pcs
Load cell capacity : 500kg
Load cell Voltage : 2mV/V
Digit : 0.10kg
Affirmation Voltage of Load cell : 5,000mV
Max Capacity of Weighing System : 1,000kg
Then, estimation result for this weighing system with formula,
5000ⅹ2ⅹ0.10
= 0.5 ≥ 0.2㎶
500ⅹ4
The calculated value is larger than 0.2㎶, so
this system has no problem.
※ According to “Resolution” or “Capacity”, it might not be calibrated like calculation.
- 13 -
Digital Weighing Indicator
SI 4300
5. SET-UP
5-1. Calibration
Calibration is the process of adjusting weight balance between “Real weight” on the load cell and
“Displayed weight of Indicator”. When you replace LOAD CELL or Indicator, you have to do Calibration
process once again
5-2. Test Weight Calibration Mode (Using Test weight)
Prepare the test weight as at least 10% of your weighing scale’s max capacity.
Remove “CAL-BOLT” on the indicator’s front panel and press “CAL - LOCK S/W” inside.
※Remark: If “P-W” is displayed, you should input the pass word to start calibration mode.
123
1. At normal mode, remove “CAL-BOLT” on the Front panel
5et-Cal
2. And press “CAL - LOCK S/W” inside.
Check the “SET-CAL. message on display.
※ To save the each step, press
3. If you press
key, and for the cancel or move back, press
key.
C999999
key, Calibration Mode starts.
After displaying “C999999”.
4. Input the max capacity of your weighing scale,
And press
C 20000
key.
Ex) Load cell CAPA : 20kg, division : 0.001
 Input 20000
5. Define the optimal position of decimal point.
Whenever you press
key, the location of decimal point
will be changed.
Ex) Load Cell CAPA : 20kg, division : 0.001kg

input 20.000
- 14 -
p 20.000
Digital Weighing Indicator
SI 4300
6. Press
key to save and move to next step.
p 20.000
7. Define the optimal Digit/Division value of weighing measurement.
Whenever you press
key, the Digit/Division value will
be changed in order of “1  2  5  10  20  50” .
Ex) Load cell CAPA : 20kg, division : 0.001

d
1
d
1
Input division “1”
“1”
8. press
key to save the Digit/Division value and move
to next step.
※ Caution : (Division value /Max capacity value) cannot over 1/20,000.
If the division is so small compare with max capacity, Error message “ Err 01 “ will be displayed
and move back to “CAPA” step again.
9. When you press
key, the indicator starts the dead process
to find “Zero” span.
10. Indicator will search “DEAD weight” during 5seconds.
After find optimal “Zero” span , step is automatically
Moves to next.
dead
Cal-1 0
※ Caution: At this step, if there is some force or vibration on weighing scale, and unstable condition will be
continued, “ErrorA” will be display, and “DEAD value” will not be calculated.
Please remove all the force or vibration and process it again.
l 20.000
11. Span Calibration mode starts..
- 15 -
Digital Weighing Indicator
SI 4300
12. Input the weight of your “Test weight”. And press
key.
Ex) Load Cell CAPA : 20kg, division 0.001
 Use test weight which is at least 10% of
l
2.000
max CAPA(20kg) = minimum 2kg of test weight is needed.
 Input test weight 2.000 to indicator.
13. When “UP” is displayed, load your test weight on the scale (weigh
bridge) Ex) Load Cell CAPA : 20kg, division 0.001
14. And press
Up
Up
key.
 Do not remove the test weight from weigh bridge.
15. Indicator will calculate span value during 5sec.
16. After finish calculation, span value will be displayed.
Please remove the test weight from weigh bridge.
Cal-2 0
0.629238
※ Caution : The “Test Weight’s value” must be at least 10% Max capacity of weighing scale.
“at least 10%” means to guarantee precise weighing process you have to make standard with at least
10% of the max capacity weight.
We programmed the calibration will not be done, when you load less than 10% of the max capacity.
17. Press
key to save all calibration process.
After then it resets automatically.
Now, fasten the Calibration Bolt on the front panel.
- 16 -
end
Digital Weighing Indicator
SI 4300
5-3. Simulation Calibration Mode (Calibrate without Test weight)
Through this “Simulation Calibration Mode” you can do simple calibration process without Test weight.
This calibration mode uses “Load cells’ max capacity” and “Rated output value(mV)”.
Simulation calibration’s degree of accuracy is lower than test weight calibration.
By simulation calibration’s characteristic, measured weight can be different with actual weight, according to
load cell’s actual output.
123
1. At normal mode, remove “CAL-BOLT” on the Front panel
5et-Cal
2. And press “CAL - LOCK S/W” inside.
Check the “SET-CAL. Message on display.
3. Please press
key, to start Simulation Calibration Mode.
※ To save the each step, press
4. Press
CellCal
key, for the cancel or move back, press
key to enter calibration mode.
After “CAPA” is displayed, Check the max Capacity of your load cell.
key.
C999999
(Refer the label on the load cell, or test report.)
5. After input max capacity of your load cell (at the label),
press
C
key
Ex) Load cell CAPA : 30kg, division : 0.01
3000
 Input 3000
In case of multiple pieces of load cells are installed, make sum of each load cell’s capacity and make setting
with max capacity. EX) There are 4pcs of load cells, and each load cell’s Max capa is1,000kg.
Then, total Max Capacitywill be 4,000kg(1,000 x 4) and you have to input 4,000.
- 17 -
Digital Weighing Indicator
SI 4300
6. Define the optimal position or decimal point
Whenever you press
key, the location of decimal point
p
30.00
p
30.00
d
1
d
1
will be changed.
7. Press
key to save Digit / Decimal point and
move to next step.
Ex) Load cell CAPA : 30kg, division : 0.01
 Input 30.00
8. Define the optimal Digit/Division value of weighing measurement.
Whenever you press
key, the Digit/Division value will
be changed in order of “1  2  5  10  20  50” .
Ex) Load cell CAPA : 30kg, division : 0.01  Input division “1”
9. press
key to save the Digit/Division value and
move to next step.
※ Caution : (Division value /Max capacity value) cannot over 1/20,000.
If the division is so small compare with max capacity, Error message “ Err 01 “ will be displayed
and move back to “CAPA” mode again.
10. 10. Under this step, measure the “DEAD Weight of Weighing Scale
When you press
key, the indicator starts the dead process
dead
to find “Zero” span.
11. Indicator will search “DEAD weight” during 5seconds.
After find optimal “Zero” span , step is automatically
Moves to next.
- 18 -
Cal-1 0
Digital Weighing Indicator
SI 4300
12. At this step input Max Output rate(mV) of load cell.
13. Input Load cell Output Rate(mV/V) (refer the load cell label)
Ex) Load cell Related output : 1.989 mV/V
Cell0Ut
o1.98900
※ Caution : Due to some variation between “Stated output rate” and “Real Output rate” of load
cell, there might be some weight difference after finishing calibration.
If you want to make more precise weighing process, please measure real output rate of load cell and input the
measured value. Then the weight measurement will be more precise than before.
14. After inputting R.O. value, press
0.087234
key.
Calculated “Span value” will be displayed.
15. Press
key to save all calibration process and fasten the
Calibration Bolt.
end
※ Caution : To process “Simulation Calibration” process, All indicator has its’ own standard value of 2mV gap.
So, if you replaced analogue board, you have to input standard value of 2mv gap.
And you can check the this 2mV gap value on F96.
(Normally, the gap value is between 200,000 ~400,000)
- 19 -
Digital Weighing Indicator
SI 4300
5-4. Set-up
Set-up means set the F-function and make SI 4300 weighing controller will perform more accuracy.
(Considering external / internal environmental condition)
5-4-1. Enter the Set-up Mode
key for 4sec. Then you can enter “F-Test” mode. Under this mode, press
1) Method : Press
No.1 key and enter the “F-function” mode.
5-4-2. F-Function Change
Under F-function mode, Whenever press
key, the Function No. will be increased one by one.
Increase to F-90 and return to F-01
If you move to certain function No., press f-function no. with number key and press
key.
Ex) If you want to call “F21-XX ” directly under “F-function mode”.
Press “
” and “
” key and press
key.
Then, you can call “F22-XX” directly.
5-4-3. F-Function Set Value Change
Under F-Function mode, input New set value with Number keys and press
If you don’t press
key to save.
key, the new set value will not be memorized.
Ex) If you want to change the “F01-01” to “F01-02”.
Under “F01-01” mode, press “
And press
” and “
” key.
key to save.
5-4-4. Exit “F-function” Mode
Under “F-function” mode, press
Under “F-Test” mode, press
key, you can move back to “F-Test” mode.
key once again, you can move back “Stand-by” mode.
- 20 -
Digital Weighing Indicator
SI 4300
5-5. F-Function
■ General Function Setting (● Factory default set value)
Weighing Data Save Method Selection
(Apply on Accumulated weighing count/weight)
●
0
Not Saved
1
Saved when “Print” , “Weight Judge Finish Relay output”
2
Save when “Print”
3
Automatic Save Mode - Save when “Weight Judge Finish Relay output”
F01
Weight-Back up selection
F02
●
0
Normal Mode
1
Weight Back up Mode
Motion Band Range setting
This is set “Steady” acceptable range of weighing part.
01
F03
06
∫
If there is vibration on weighing part, you can set this function and reduce the vibration
effect on weighing process.
50
1
: Weak vibration
~
50
: Strong Vibration
Zero Tracking Compensation Range setting
Due to external causes(Temperature, wind, and dust), there are small weight difference,
indicator will ignore the weight difference and display Zero.
00
F04
02
∫
For this compensation function, indicator will estimate the weight difference is over the
set range during fixed time period.
09
If there is large weight difference over set range within fixed time period, the “Zero” is
breaking and will find new zero point.
Auto Zero Range setting
Within the “Auto Zero” range, weighing part is steady, indicator will display current
00
F05
00
∫
weight as “Zero”
If the weighing part is not “Steady”, indicator will display current weight.
99
(Auto Zero Range : ± Set value + weight unit)
Digital Filter setting
F06
13
AB
A : Frequency Filter setting value (0~3)
If “B” set value is fixed, “A” set value
(0 : about 200Hz/sec, 3 : about 500Hz/sec)
is large, the indicator will response
B : Buffer Filter setting value (1~9)
more sensitive.
Zero /Tare key Operation mode selection
●
0
Activate when “Steady” condition, only
1
Always activated
F07
- 21 -
Digital Weighing Indicator
SI 4300
Zero key Operation Range selection
●
F08
0
Activated within 2% of Max Capacity
1
Activated within 5% of Max Capacity
2
Activated within 10% of Max Capacity
3
Activated within 20% of Max Capacity
4
Activated within 50% of Max Capacity
5
Activated within 100% of Max Capacity
6
Whenever Press “Zero” key (No Limit)
Caution : If you set over 20% , it may cause “CELL-Err” or inaccurate weighing value.
Tare key Operation Range selection
F09
●
0
Activated within 10% of Max Capacity
1
Activated within 20% of Max Capacity
2
Activated within 50% of Max Capacity
3
Activated within 100% of Max Capacity
“Key TARE” Operation Selection
●
0
Key TARE Function Not Use.
1
Key TARE Function Use
F10
External Input Selection
Set Value
Input 1
Input 2
Input 3
Input 4
0
TARE
TARE RESET
ZERO
PRINT
1
ZERO
TARE / RESET
PHOTO SENSOR
PRINT
2
ZERO
TARE / RESET
HOLD
HOLD/RESET
3
ZERO
TARE
TARE RESET
4
ZERO
PRINT
SUB TOTAL, PRINT
PHOTO SENSOR
5
RUN
STOP
PHOTO SENSOR
ZERO
●
F11
PRINT
“STEADY” condition check time setting
F12
03
01
During the set time period, estimate weighing part’s “STEADY” condition and display.
∫
If you set small value, indicator will take “STEADY” fast, if you set large value,
20
indicator will take “STEADY” slow.
Display Up-date rate selection (per 1 sed)
●
F13
0
238 times
5
31 times
1
102 times
6
26 times
2
64 times
7
23 times
3
47 times
8
20 times
4
34 times
9
18 times
- 22 -
Digital Weighing Indicator
SI 4300
“OVER” / “UNDER” Range Set value apply selection
●
0
Apply only certain P/N
1
Apply same set value to all P/N
F14
SUB/GRAND Total Display mode selection
●
Display Accumulated weighing count and weight of current P/N
0
(SUB TOTAL DATA Display)
Display Accumulated weighing count and weight of all P/N
1
F15
(GRAND TOTAL DATA Display)
2
Display weighing count of current P/N “OVER”,”UNDER” or “PASS”
3
Display weighing count of current all P/N “OVER”,”UNDER” or “PASS”
Minus(-) symbol display selection
●
0
Display (-) symbol on the display
1
Not used
F16
“NEAR ZERO” relay output mode selection
●
0
Display weight is Zero(Including “TARE” Zero) Near Zero relay output
1
Only Gross Zero(Net weight + TARE)  Near Zero relay output
F17
Equipment No. setting
F18
01
01~99
Equipment No. setting with No. key.
HOLD Selection
●
Sample HOLD : Hold the weight value, when the “HOLD” key input
0
Peak HOLD : Hold the “PEAK” value of weight, during the set time
F19
1
(Under F21-02 setting, must be set PEAK HOLD).
■ Relay Output Mode Setting
Weighing Mode selection
●
1
PHOTO SENSOR INPUT MODE
2
HOLD KEY INPUT MODE
3
SIMPLE COMPARISON MODE 1.
4
SIMPLE COMPARISON MODE 2.
F21
Relay output Mode(Each weighing Mode)
Weighing Mode
Output1
Output2
Output3
Output4
Output5
Output6
1
PHOTO SENSOR
UNDER
PASS
OVER
N.G
EMPTY
No using
2
HOLD MODE
UNDER
PASS
OVER
N.G
EMPTY
No using
3
SIMPLE 1..
UNDER
PASS
OVER
N.G
EMPTY
No using
4
SIMPLE 2.
UNDER
PASS
OVER
EMPTY
No using
- 23 -
No using
Digital Weighing Indicator
SI 4300
◆ Weighing Mode 1. PHOTO SENSOR INPUT MODE – F21-1 setting
Weight
OVER
Over Range
PASS
Under Range
UNDER
중량
과량
상한 설정 값
정량
하한 설정 값
미량
영점부근
판정 입력
t3
t4
출력 1
하한 출력
t3
t4
출력 3
상한 출력
t3
정량 출력
t6
t5
에러 출력
t6
t5
t4
출력 2
출력 4
출력 5
영점부근
Empty
Sensor IN
t3
t4
Output 1
UNDER
t3
t4
Output 3
OVER
t3
PASS
t6
t5
t6
t5
t4 Output 2
Output 4
ERROR
Output 5
EMPTY
1. Set Value Condition : Over Set value > Under Set value
※ If the setting conditions are not satisfied, “E” symbol displayed and you can process the weighing.
2. Weighing Judgment
After Photo Sensor Input, Indicator will delay the judgment during the “t3” set time.
After “t3” set time, indicator will judge the weight and output “Over” / “Under” / “Pass” relay.
During the “Weight Judgment Relay “ON” time, indicator will hold the weight display.
In case of “OVER” or “UNDER”, “Error” relay will be output during the “t5” time period, after “t6” delay
time period
3. External input : please set “F11-00”, only
4. Output Relay
Output
Contents
Output
Contents
Under(1)
Less than “Under” set, “ON”
N.G(4)
Under/Over relay output, “ON”
Over(3)
More than “Over” set, “ON”
Empty(5)
Within the Empty Range “ON”
Pass(2)
Under set ≤ Current Weight ≤ Over set
- 24 -
Digital Weighing Indicator
SI 4300
◆ Weighing Mode 2. (“HOLD” input Mode)
– Manual HOLD input / Manual HOLD Reset mode (Under F55-00 setting)
OVER
Weight
OVER Range
PASS
UNDER
Under Range
EMPTY
HOLD Input
t3
t4
Output 1
UNDER
t3
t4
Output
OVER
t3
PASS
t5
t5
t4 Output 2
Output 4
N.G
Output 5
Empty
1. Set Value Condition : Over Set value > Under Set value
※ If the setting conditions are not satisfied, “E” symbol displayed and you can process the weighing.
2. Weighing Judgment
After “HOLD(Sample/Peak)” key input, indicator will delay the judgment during “t3”.
After “t3” set time, indicator will judge the weight and output “Over” / “Under” / “Pass” relay.
During the “Weight Judgment Relay “ON” time, indicator will hold the weight display.
F19-00 (Sample HOLD) : “t3” setting is not working, and judge the weight at the input spot weight.
F19-01(Peak HOLD) : After input HOLD key, tracking the “Peak” weight during “t3” period and judge the
weight after finishing “t3” period.
HOLD Reset : After finishing Weight judgment, press “HOLD” key and reset.
3. External input Mode : F11-01, only
4. Output Relay
Output
Contents
Output
Contents
Under(1)
Less than “Under” set, “ON”
N.G(4)
Under/Over relay output, “ON”
Over(3)
More than “Over” set, “ON”
Empty(5)
Within the Empty Range “ON”
Pass(2)
Under set ≤ Current Weight ≤ Over set
- 25 -
3
Digital Weighing Indicator
SI 4300
◆ Weighing Mode 3. (“HOLD” input Mode)
– Auto HOLD input / Auto HOLD Reset mode (Under F55-01 setting)
OVER
Weight
OVER Range
PASS
UNDER
Under Range
EMPTY
HOLD Input
t3
t4
Output 1
UNDER
t3
t4
Output 3
OVER
t3
PASS
t5
t5
t4
Output 2
Output 4
N.G
Output 5
Empty
1. Set Value Condition : Over Set value > Under Set value
※ If the setting conditions are not satisfied, “E” symbol displayed and you can process the weighing.
2. Weighing Judgment – Please set “F19-01(Peak Hold).
After “HOLD(Peak)” key input, indicator will delay the judgment during “t3”.
After “t3” set time, indicator will judge the weight and output “Over” / “Under” / “Pass” relay.
※ Auto “HOLD” Input : The Weight is over than “Empty” Range, “HOLD” will be set automatically
and start tracking “Peak” value during “t3” period.
※ Auto “HOLD Reset” Input : After finishing Weight judgment, weigh will be lower than “Empty” Range,
“HOLD” will be reset automatically.
※ From Weight judgment relay output to Empty relay output, indicator will hold finished weight judgment
data on the main display.
4. Output Relay
Output
Contents
Output
Contents
Under(1)
Less than “Under” set, “ON”
N.G(4)
Under/Over relay output, “ON”
Over(3)
More than “Over” set, “ON”
Empty(5)
Within the Empty Range “ON”
Pass(2)
Under set ≤ Current Weight ≤ Over set
- 26 -
Digital Weighing Indicator
SI 4300
◆ Weighing Mode 3. (Simple Comparison Mode 1.)
Weight
OVER
Over Range
PASS
Under Range
UNDER
Empty
t3
t4
Output 1
UNDER
t3
t4
Output 3
OVER
t3
PASS
t5
t5
t4
Output 2
Output 4
N.G
Output 5
EMPTY
1. Set Value Condition : Over Set value > Under Set value
※ If the setting conditions are not satisfied, “E” symbol displayed and you can process the weighing.
2. Weighing Judgment
When “Current Weight” is Steady, after “t3” period, indicator judges the weight.
Simple Weight Comparison Mode.
3. Output Relay
Output
Contents
Output
Contents
Under(1)
Less than “Under” set, “ON”
N.G(4)
Under/Over relay output, “ON”
Over(3)
More than “Over” set, “ON”
Empty(5)
Within the Empty Range “ON”
Pass(2)
Under set ≤ Current Weight ≤ Over set
- 27 -
Digital Weighing Indicator
SI 4300
◆ Weighing Mode 4. (Simple Comparison Mode 2.)
Weight
Over Range
Under
Ragne
Empty
UNDER
Output 1
PASS
Output 2
OVER
Output 3
EMPTY
Output 5
1. Set Value Condition : Over Set value > Under Set value
※ If the setting conditions are not satisfied, “E” symbol displayed and you can process the weighing.
2. Weight Judgment
When “Current Weight” is Steady, Indicator will judge the Weight condition. – Limit Mode
UNDER : Empty < Current Weight < UNDER Range
PASS : UNDER Range≤ Current Weight ≤ OVER Range
OVER : OVER Range < Current Weight
3. Output Relay
Output
Contents
Output
Contents
Under(1)
Less than “Under” set, “ON”
PASS(2)
Under set ≤ Current Weight ≤ Over set
Over(3)
More than “Over” set, “ON”
Empty(5)
Within the Empty Range “ON”
- 28 -
Digital Weighing Indicator
SI 4300
Weight Judgment Relay Delay time(t3) setting
Delay time period setting, before Weight judgment.
01
F22
10
∫
99
Under F21-01 : Weight Judge after t3 set period, when “Photo Sensor” input
Under F21-02 : Weight Judge after t3 set period, when “HOLD” input
Under F21-03 : Weight Judge after t3 set period, when weight is “Steady”.
Under F21-04 : Not applied
Weight Judgment Relay “ON” time (t4) setting
After “t3” set period, judge the weight condition and output judgment relay during the
00
F23
10
∫
99
set period.
Under “00” setting : Judgment relay will out until the weight reaches to Empty Range.
Under “20” setting : Judgment relay will out during 2.0sec.
Weight Mode Selection (Absolute Weight / Positive weight mode)
●
0
Absolute Weight mode (Same replay output for “-“ weight)
1
Positive Weight mode (Relay output for only “+” weight)
F25
N.G Relay “Delay” time (t6) setting
In case of “UNDER”, or “OVER” weight, you can make some delay time before “N.G”
00
F26
00
∫
50
relay output...
Under “01” setting : N.G relay will delay during 0.1sec.
Under “20” setting : N.G relay will delay during 2.0sec.
N.G Relay “ON” time (t5) setting
In case of “UNDER” or “OVER” weight condition, “N.G” relay will be out with
01
F27
10
∫
99
“Weight judgment relay.
Under “01” setting : N.G relay will out during 0.1sec.
Under “20” setting : N.G relay will out during 2.0sec.
■ Communication Mode setting (Serial Port 1. - Standard installed port)
Parity Bit selection Mode
●
F30
0
No Parity
1
Odd Parity
2
Even Parity
Serial Communication Speed selection
F31
●
0
2,400bps
1
4,800bps
2
9,600bps
3
14,400bps
- 29 -
Digital Weighing Indicator
SI 4300
4
19,200bps
5
28,800bps
6
38,400bps
7
57,600bps
8
76,800bps
9
115,200bps
DATA Transference Method selection
●
0
Simplex Mode / Stream Mode
1
Duplex Mode / Command Mode
F32
Print port selection (Under F32-01 setting, only)
●
0
Same port as using for Command Mode.
1
The other port.
F33
“Check-Sum” detection selection (Under F32-01 setting, only)
●
0
Check-Sum data will not be included on transferred data.
1
Check-Sum data will be included on transferred data.
F34
Serial Port Application Selection (Under F32-00 setting, only)
●
0
DATA Transference purpose
1
Printing purpose (Serial Printer)
F35
DATA Transference Mode selection (Under F32-00, F35-00 setting, only)
●
F36
0
Stream Mode : Weighing Data will be transferred continuously.
1
Finish Mode : When Finish Relay output, only 1 time transferred.
2
Manual Mode : When “Print” key input, 1 time transferred.
DATA Transference Format selection(Under F32-00, F35-00 setting, only)
●
0
Format 1.
1
Format 2. (Format 1 + ID No.)
2
CAS Format
3
AND Format
F37
Print Mode selection (Under F32-00, F35-01 setting, only)
●
0
Manual Print : Whenever “Print” key input.
1
Auto Print : When Finish relay output, automatically print.
F38
Transferring DATA Byte selection
●
0
7 Byte data Transfer
1
8 Byte data Transfer
F40
- 30 -
Digital Weighing Indicator
SI 4300
■ Print Mode Setting (These settings will be apply to Serial and Parallel print)
Weight Unit selection
●
F41
0
kg
1
g
2
t
Print Format selection (If you install on Standard Serial Port)
●
Continuous Print
0
Serial No. and Weight will be printed continuously.
F42
Single Print
1
Date, Time, S/N, ID No. Weighing Data will be print
Print Format selection (If you install on Optional Serial Port)
●
Continuous Print
0
Serial No. and Weight will be printed continuously.
F43
Single Print
1
Date, Time, S/N, ID No. Weighing Data will be print
SUB/GRAND Total Data Delete selection
Manual Delete Mode
●
SUN Total Delete : “Clear” key + “P/N” key
0
GRAND Total Delete : “Clear” key + “S/N” key
F44
Automatic Delete Mode
1
After SUB/GRAND Total Print, Automatically Deleted.
Paper Withdraw Rate setting (After SUB/GRAND Total Print)
F45
03
00~09
Whenever set value increased, 1line will be added.
Paper Withdraw Rate setting (After Continuous/Single Print)
F46
03
00~09
Whenever set value increased, 1line will be added.
Printing Language Selection (If you install on Standard Serial Port)
●
0
KOREAN
1
ENGLISH
F47
Printing Language Selection (If you install on Optional Serial Port)
F48
●
0
KOREAN
1
ENGLISH
Minus(-) symbol Print selection
●
0
Print minus(-) symbol, if the weight is minus(-).
1
Ignore minus(-) symbol
F49
- 31 -
Digital Weighing Indicator
SI 4300
Parallel Print Port selection
●
F50
0
Parallel Port is not installed.
1
Share Standard Serial Port.
2
Share Optional Serial Port.
Function / Clear key Activation display selection
F51
●
0
Activation display not use
1
Activation display use
Auto Hold Reset selection
●
0
Auto HOLD Reset will not use
1
Auto HOLD Reset use (HOLD Reset at “Weight Judge relay output)
F52
Counting selection (On Sub Display window)
●
0
PASS Count display
1
Over Count display
2
Under Count display
3
N.G Count display (Under + Over)
F53
HIGH / LOW Letter display selection (On Sub Display window)
●
0
Letter not use
1
Letter will use (High relay : HI / Under relay : LO)
F54
Auto HOLD Tracking Function selection (Under F21-02 mode)
●
0
Auto HOLD Tracking not use
1
Auto HOLD Tracking use
F55
“EMPTY” Checking selection
●
0
EMPTY check
1
EMPTY not check
F56
Communication Interval Setting
●
0
Fast Speed (The interval is short)
1
Low Speed (The interval is long)
F57
4~20mA analogue Output Setting
●
0
Positive Output
1
Negative Output
F58
- 32 -
Digital Weighing Indicator
SI 4300
■ Other Setting
※ Under “Other setting mode”, you can not move to other function directly.
Press
key and move to F01 and move to other function No. directly.
EMPTY Range setting
You can set “EMPTY” Range.
X.X.X.X.X.X.
F80
(0.0.0.0.1.0)
Within set range, indicator will not display current weight and just display “Zero”.
“0.000” setting : When Net Zero, “Zero” status lamp and Near Zero relay will be
output.
“0.190” setting : Within 190, “Zero” Status lamp and Near Zero relay will be output.
SPAN Calibration Value Check
Span Calibration Value Check
Under F-function mode, enter “
F89
”, “
” key and press “
”.
X.X.X.X.X.X.
After checking the value and press “
” to exit
※ If you have difficulty to process Calibration again, the best way to matching the
net weight and display weight is doing Calibration process once again.
DATE Check / Change
F90
Check Current DATE data or you can Change to new date
TIME check / Change
F91
Check Current TIME data or you can Change to new date
Program & Hard ware Version Check
F98
Check the Program & Hard ware version (H/W : X.XX, S/W : X.XX.X)
Production DATE Check
F99
Check the Product’s Production Year and Month.
■ Communication Mode setting (Serial Port 2. - Optional Serial port)
Parity Bit selection Mode
●
F60
0
DATA Bit (8 Bit)
STOP Bit (1 Bit)
Parity Bit (Non)
1
DATA Bit (7 Bit)
STOP Bit (2 Bit)
Parity Bit (Non)
2
DATA Bit (7 Bit)
STOP Bit (1 Bit)
Parity Bit (Even)
3
DATA Bit (7 Bit)
STOP Bit (1 Bit)
Parity Bit (Odd)
4
DATA Bit (8 Bit)
STOP Bit (2 Bit)
Parity Bit (Non)
5
DATA Bit (8 Bit)
STOP Bit (1 Bit)
Parity Bit (Even)
6
DATA Bit (8 Bit)
STOP Bit (1 Bit)
Parity Bit (Odd)
- 33 -
Digital Weighing Indicator
SI 4300
Serial Communication Speed selection
●
0
2,400bps
1
4,800bps
2
9,600bps
3
14,400bps
4
19,200bps
5
28,800bps
6
38,400bps
7
57,600bps
8
76,800bps
9
115,200bps
F61
DATA Transference Method selection
●
0
Simplex Mode / Stream Mode
1
Duplex Mode / Command Mode
F62
Print port selection (Under F62-01 setting, only)
●
0
Same port as using for Command Mode.
1
The other port.
F63
“Check-Sum” detection selection (Under F62-01 setting, only)
●
0
Check-Sum data will not be included on transferred data.
1
Check-Sum data will be included on transferred data.
F64
Serial Port Application Selection (Under F62-00 setting, only)
●
0
DATA Transference purpose
1
Printing purpose (Serial Printer)
F65
DATA Transference Mode selection (Under F62-00, F65-00 setting, only)
F66
●
0
Stream Mode : Weighing Data will be transferred continuously.
1
Finish Mode : When Finish Relay output, only 1 time transferred.
2
Manual Mode : When “Print” key input, 1 time transferred.
DATA Transference Format selection(Under F62-00, F65-00 setting, only)
●
F67
0
Format 1.
1
Format 2. (Format 1 + ID No.)
2
CAS Format
Print Mode selection (Under F62-00, F65-01 setting, only)
●
0
Manual Print : Whenever “Print” key input.
1
Auto Print : When Finish relay output, automatically print.
F68
- 34 -
Digital Weighing Indicator
SI 4300
6. INTERFACE
6-1. Serial Interface (RS-232C)
RS-232C Serial Interface is sensitive/weak for electric Noise.
So, please isolate with AC power cable and use shield cable to reduce the electric noise effect.
6-1-1. Communication with PC(Personal Computer) or Other devices
2 : RxD
TxD :
3
3 : TxD
Personal Computer
(9pins Standard)
RxD :
2
5 : SG
SG :
SI 4300
5
6-1-2 Connection with External Display or other devices
1 : RxD
TxD : 3
2 : GND
GND :
5
SI 4100
SE 6125
(External Display)
6-1-3. Signal Format
① Type : EIA-RS-232C
② Communication Method : Half-Duplex, Full Duplex, Asynchronous
③ Serial Baud Rate : Selectable on “F-function31”
④ Data Bit : 8(No Parity mode, only)Bit – Refer “F30”.
⑤ Stop Bit : 1
⑥ Parity Bit : Non, Even, Odd (Selectable on “F-function 30”) - Refer “F30”
⑦ Code : ASCII
STX 02H
ETX 03H
CR 0DH
LF 0AH
- 35 -
Digital Weighing Indicator
SI 4300
⑧ Check-Sum (Error Detecting, “F-Function 34”)
LSB
1
MSB
+9V
1
2
3
4
5
0
0
6
-9V
Data Byte – Refer “F40”
Start Bit
Parity
Stop
Bit
Bit
6-1-4. Data Format(1) : ID Number will not be transferred. (Refer “F-function 37”)
,
Header 1.
,
Header 2.
Including Decimal
+/-
k
g
CR
LF
point
Data Byte (7 or 8 byte )
(F40 setting)
Unit
① Header 1. : OL : Over Load, Under Load
ST : Display weight “Steady”
US : Display “Un-Steady”
② Header 2. : NT : Net-Weight
GS : Net-Weight, under TARE
③ Data Bit(Number) 2B(H) : “+” Plus
2D(H) : “-“ Minus
2D(H) : “ “ Space
2E(H) : “.” Decimal Point
④ Unit : kg, g, t
6-1-5. Data Format(2) : ID Number + Data Transference (Refer “F-function 18, 37)
Including
,
,
,
+/-
Decimal
CR
point
ID Number
Data Byte
Header 1
Header 2
Unit
( 7 or 8 byte )
(F40 setting)
- 36 -
LF
Digital Weighing Indicator
SI 4300
① Header 1. : OL : Over Load, Under Load
ST : Display “Steady”
US : Display “Un-Steady”
② Header 2. : NT : Net-Weight
GS : Net-Weight, under TARE.
③ Data Bit(Number) 2B(H) : “+” Plus
2D(H) : “-“ Minus
2D(H) : “ “ Space
2E(H) : “.” Decimal Point
④ Unit : kg, g, t
6-1-6. Data Format(3) : CAS “CI5101A” Data Transference) – CAS 22byte Format
,
,
k
ID
Header 1
Lamp
CR
LF
Weight
Header 2
Data(8)
No.
g
Space
Display
Unit
①. Header 1. : OL : Over Load, Under Load
ST : Display “Steady”
US : Display “Un-Steady”
②. Header 2. : NT : Net-Weight
GS : Net-Weight, under TARE.
③. Lamp Display : Current Lamp Condition (ON/Off Data)
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
1
Steady
1
Hold
Print
Bit 2
Bit 1
Bit 0
Tare
Zero
Gross
Weight
④. Data Bit(Number) 2B(H) : “+” Plus
2D(H) : “-“ Minus
2D(H) : “ “ Space
2E(H) : “.” Decimal Point
⑤. Unit : kg, g, t
- 37 -
Digital Weighing Indicator
SI 4300
6-1-7. Data Format : AD – 4321 Data Transference) – AD – 4321
,
Header 1
,
18byte Format (Refer F-37)
+,- / Including Decimal point 8byte
Header 2
Data(8)
k
g
Weight
Unit
① Header 1. : OL : Over Load, Under Load
ST : Display “Steady”
US : Display “Un-Steady”
② Header 2. : NT : Net weight (Under Tare)
GS : Net weight (Under TARE reset)
③ Data Bit(Number) 2B(H) : “+” Plus
2D(H) : “-“ Minus
20(H) : “ “ Space
2E(H) : “.” Decimal Point
④ Unit : Kg, g, t
- 38 -
CR
LF
Digital Weighing Indicator
SI 4300
6-2. Current Loop Interface
“Current Loop” Interface is stronger for Electric Noise than “RS-232C” interface.
So, it can be used for long distance communication.(About 100m long distance).
※ Current Loop Interface supports, up to 9,600 Communication Speed, only.
6-2-1. Communication with Other Devices (Remote Display / External Display)
3 : RxD
TxD :
8
4 : RxD
TxD :
9
SI 4300
SE 6125
(External Display)
6-2-2. Current Loop Circuit
- 39 -
Digital Weighing Indicator
SI 4300
6-3. Print Interface (Option 01 : Centronics Parallel Interface)
This Print Interface Option is based on “Centronics Parallel Interface”, so this print interface can be
connected other printers using this communication method.
But, the print format is programmed based on our “SE7300”, and “SE7320” Industrial Printers, so you had
better to use these printer for convenience.
6-3-1. Connector Wire Connection
Pin
Signal
Contents
RE
Pin
Signal
Contents
RE
1
STROBE
STROBE signal
out
14
N.C
2
DATA0
Data(bit0) signal
out
15
N.C
3
DATA1
Data(bit1) signal
out
16
N.C
4
DATA2
Data(bit2) signal
out
17
N.C
5
DATA3
Data(bit3) signal
out
18
GND
GROUND
out
6
DATA4
Data(bit4) signal
out
19
GND
GROUND
out
7
DATA5
Data(bit5) signal
out
20
GROUND
out
8
DATA6
Data(bit6) signal
out
21
GROUND
out
9
DATA7
Data(bit7) signal
out
22
GROUND
out
10
ACK
Data Response
In
23
GROUND
out
11
BUSY
Busy signal
In
24
GROUND
out
12,13
N.C
GROUND
out
25
- 40 -
GND
Digital Weighing Indicator
SI 4300
6-3-2. Print Format (English)
Single Print Format
Continuous Print format
DATE :
2006-10-15
TIME :
10:20:30
ID_N :
01
STATE PART SERIAL WEIGHT
OVER 01 01 +1.000 kg
DATE :
DATE :
2006-10-15
TIME :
ID_N
10:20:30
:
01
STATE PART SERIAL WEIGHT
OVER
2006-10-15
TIME :
10:20:30
ID_N :
01
STATE PART SERIAL WEIGHT
PASS
02 +1.000 kg
01
01
+1.000 kg
PASS
01
02
+0.900kg
PASS
01
03
+0.910kg
PASS
01
04
+0.900kg
UNDER
01
05
+0.890kg
OVER
01
06
+1.000kg
Sub-Total Print Format
Grand-Total Print Format
SUB-TOTAL
GRD-TOTAL
DATE :
2006-10-15
DATE :
2006-10-15
TIME :
10:30:30
TIME :
10:40:30
ID_N :
01
ID_N :
01
PART:
10
T-COUNT :
2
OK-COUNT :
1
NG-COUNT :
1
T-WEIGHT :
PART
SERIAL
WEIGHT
10
2
2.100kg
T-PART :
2.100kg
T-COUNT :
2
T-OK-COUNT :
1
T-NG-COUNT :
1
T-WEIGHT :
- 41 -
1
2.100kg
Digital Weighing Indicator
SI 4300
6-4. Analog Output Interface (Option 02 : 0~10V Output)
This Option card converts weight value to Analog Voltage output(0~10V) and transfers to external
devices(Recorder, P.L.C), controlled by voltage output.
6-4-1. Specification
① Output Voltage : 0~10V DC output
② Accuracy : More than 1/1,000
6-4-2. Circuit
※ This Voltage output is proportioned on weight calibration and outputs 0~10V.
6-4-3. Output Adjustment
①This output is adjusted as when the weight is “Zero”, output is 0V and When the weight is “Full
capacity”, output is 10V.
②If you need additional adjustment, please adjust with “VR1(Zero)”, “VR2(Span) on the Analog Output
PCB.
※ Remark
This Analog option card converts Displayed weight value(Micro-process data) to analog value on
D/A Converter(Digital to Analog converter)
This D/A Converter has Max 1/4,000 accuracy, so this output is not suitable for high accuracy
application, like more than 1/3,000.
6-4-4. Connecter
9pin, “D-type” female
Terminal Block (3 pin)
1 : HI(+), 5 : (-)
※ For 0~5VDC or 1~5VDC analog output, please inform when you inquiry.
- 42 -
Digital Weighing Indicator
SI 4300
6-5. Analog Output Interface (Option 03 : 4~20mA Output)
This Option card converts weight value to Analog Electric Current output(4~20mA) and transfers to
external devices(Recorder, P.L.C), controlled by electric current output.
6-5-1. Specification
Output Current
4~20mA (Output Range : 2~22mA)
Accuracy
Over 1/1,000
Temperature Co-efficiency
0.01%℃
Max Loaded Impedance
500Ω MAX
※ When Weight display is “Zero”, 4mA current will be output, when Weight display is “Full
Capacity”, 20mA current will be output.
6-5-2. Circuit
※ “LO” terminal is not a “GND”, so this “LO” terminal
do not be connected with other “GND” terminal on other
devices.
6-5-3. Output Adjustment
① This output is adjusted as when the weight is “Zero”,
output is “4mA” and When the weight is “Full capacity”,
output is “20mA”.
② If you need additional adjustment, please adjust with “VR1(Zero)”, “VR2(Span) on the Analog
Output PCB.
※ Remark
This Analog option card converts Displayed weight value(Micro-process data) to analog value on
D/A Converter(Digital to Analog converter)
This D/A Converter has Max 1/4,000 accuracy, so this output is not suitable for high accuracy
application, like more than 1/3,000.
6-5-4. Connecter (9pin, “D-type” female)
9pin, “D-type” female
Terminal Block (3 pin)
1 : HI(+), 5 : (-)
- 43 -
Digital Weighing Indicator
SI 4300
6-6. Serial Interface (option 04 : RS-232C/422/485)
RS-422/485 serial interface is more stable for electric noise effect compare with other communication method,
using electric current difference.
But, install isolated place from Power cable or other electric cables and wires, and please use shielded cable for
better performance.
Recommendable communication distance is about 1.2km.
6-6-1. Signal Format
. Type : RS-422/485
②. Format : Baud Rate : Refer “F-function 36”.
Data Bit : 7 or 8(No Parity)
Stop : 1
Parity Bit : Even, Odd, No Parity (Selectable)
Code : ASCII (STX 02H, ETX 03H, CR 0DH, LF 0AH)
LSB
1
MSB
1
2
3
4
5
0
6
0
1.5V Electric
Potential Difference
Start 1 Bit
Parity Bit
Data Bit
Stop 1 Bit
6-6-2. Data Format
Same as RS-232C (Refer “6-1. Serial Interface”)
6-6-3. RS-422 Circuit (In case of RS-485, Use No6 and 7 pin - No.8 and 9 will be use)
D-SUB 9 pin
If you install additional RS-232C interface,
Terminal Block
refer “6-1. Serial Interface” section.
Terminal Block
1
2
3
4
RS-232
TX
RX
GND
GND
RS-485
RTX+
RTX-
NC
NC
RS-422
RXD+
RXD-
TXD+
TXD-
- 44 -
Digital Weighing Indicator
SI 4300
6-7. BCD Input ( Option 05) – Input for Part No. selection.
This “BCD interface” option card can be applied on PLC (Programmable Logic Controller), or Score Board
applications.
Each Input circuit is isolated with “Photo-Coupler”, from external devices electrically.
Wire Connection Diagram
- 45 -
Digital Weighing Indicator
SI 4300
6-8. BCD Output Interface( Option 06)
This “BCD interface” option card can be applied on PLC (Programmable Logic Controller), or Score Board
applications.( NPN TYPE)
Each Input circuit is isolated with “Photo-Coupler”, from external devices electrically.
- 46 -
Digital Weighing Indicator
SI 4300
6-9. Command Mode
Under “Command Mode”, Indicator will recognize the receipt of Order based on 02h(Header) and 03h(END)
signal, and transfers ACK(06)/ NAK(15).
6-9-1. Read Command (Standard Serial Port and Optional Port is same.)
P.C ->> SI 4300
Command
SI 4300 Response
Current Weight READ(7/8 byte, including Decimal point)
Current Weight
STX ID NO. RCWT ETX
-STX ID NO. RCWT ST/US,NT/GS, ±Current Weight (7/8byte)
Transfer
Weight unit(2byte) ETX
OVER Value
Over Set value READ(Including Decimal point 7/8byte)
STX ID NO. RHIG ETX
Transfer
Under Value
-STX ID NO. RHIG Over Value(7/8byte) ETX
Under Set value READ(Including Decimal point 7/8byte)
STX ID NO. RLOW ETX
Transfer
-STX ID NO. RLOW UNDER Value(7/8byte) ETX
Pass Weight
Pass Count READ(Including Decimal point 6byte)
STX ID NO. RPSN ETX
Count
Over Weight
-STX ID NO. RPSN Pass Count ETX
Over Count READ(Including Decimal point 6byte)
STX ID NO. ROSN ETX
Count
Under Weight
-STX ID NO. ROSN Over Count ETX
Under Count READ(Including Decimal point 6byte)
STX ID NO. RUSN ETX
Count
N.G(Over+Under)
-STX ID NO. RUSN Under Count ETX
N.G Count READ(Including Decimal point 6byte)
STX ID NO. RNSN ETX
Count
TIME setting
-STX ID NO. RNSN N.G Count ETX
TIME setting READ(6byte)
STX ID NO. RTIM ETX
Transfer
TARE set weight
STX ID No. RTIM time setting ETX
TARE set weight READ(7/8 byte, including Decimal point)
STX ID NO. RTAR ETX
Transfer
DATE setting
STX ID No. RTAR TARE Weight(7/8byte) ETX
DATE setting READ(6byte)
STX ID NO. RDAT ETX
Transfer
Serial Count
STX ID No. RDAT DATE setting ETX
Serial Count READ(6 byte)
STX ID NO. RSNO ETX
(Over+Pass+Under)
STX ID No. RSNO Serial Count ETX
SUB-TOTAL DATA READ
SUB-TOTAL
STX ID NO. RSUB ETX
-STX ID NO. RSUB P/N(2byte) Serial No.(6byte) Accumulated
READ
Weight(11byte) Weight unit(2byte) ETX
INTERNAL MEMORY READ
INDICATOR
-STX ID NO. RCWD DATE(6byte) TIME(6byte) P/N(2byte)
STX ID NO. RCWD ETX
MEMORY
S/N(6byte) TARE Weight(7/8byte) Current Weight(7/8byte)
READ
Weight unit(2byte) ETX
P/N
Part No. READ (2byte)
STX ID NO. RPNO ETX
READ
STX ID NO. RGRD ETX
GRAND-TOTAL
STX ID No. RPNO Part No.(2byte) ETX
GRAND-TOTAL DATA READ
- 47 -
Digital Weighing Indicator
SI 4300
READ
-STX ID NO. RGRD P/N(2byte) Serial No.(6byte) Accumulated
Weight(11byte) Weight unit(2byte)ETX
FINISH weight
FINISHED weight READ((7/8 byte, including Decimal point)
STX ID NO. RFIN ETX
STX ID NO. RWRS ETX
READ
-STX ID NO. RFIN FINISHED Weight(7/8byte) ETX
Current
Weight + External Input + Output Relay
Weight/External input /
Output Transfer
-STX IN NO. RWRS Weight(7/8byte), External input(4byte),
Output Relay(6byte) ETX
6-9-2. Write Command
P.C ->> SI 4300
Command
SI 4300
Response
STX ID NO. WHIG (Including Decimal point 7/8 byte) ETX
Over Range Setting
ACK or NAK
STX ID NO. WLOW (Including Decimal point 7/8 byte) ETX
Under Range Setting
ACK or NAK
STX ID NO. WDAT(6byte) ETX
Date Setting
ACK or NAK
STX ID NO. WTIM(6byte) ETX
Time Setting
ACK or NAK
STX ID NO. WSNO(6byte) ETX
Serial No. Setting
ACK or NAK
STX ID NO. WPNO(2byte) ETX
Part No. Setting
ACK or NAK
STX ID NO. WPSN (6byte) ETX
Pass Count Setting
ACK or NAK
STX ID NO. WOSN (6byte) ETX
Over Count Setting
ACK or NAK
STX ID NO. WUSN (6byte) ETX
Under Count Setting
ACK or NAK
STX ID NO. WPRT ETX
Print Command
ACK or NAK
STX ID NO. WSPR ETX
Sub-Total Print Command
ACK or NAK
STX ID NO. WGPR ETX
Grand-Total Print Command
ACK or NAK
STX ID NO. WTAR ETX
TARE Setting
ACK or NAK
STX ID NO. WTRS ETX
TARE Reset Setting
ACK or NAK
STX ID NO. WZER ETX
ZERO Setting
ACK or NAK
STX ID NO. WSTC ETX
Sub-Total Delete Command
ACK or NAK
STX ID NO. WGTC ETX
Grand-Total Delete Command
ACK or NAK
STX ID NO. WHOL ETX
HOLD Setting
ACK or NAK
STX ID NO. WHRS ETX
HOLD Reset Command
ACK or NAK
 How to Calculate Check sum.
Sum the value from “STX” to “ETX” and converts to ASCII(2byte) and transfer.
Convert the Sum value(HEX) to ASCII and transmit(28byte) .
ex) The sum HEX value from STX to ETX(02,30,31,52,43,57,54,03) is 1A6h.
Then, divide 1A6h by 100h(1A6h/100h). the rest of result is A6h.
Calculated remainder value is A6h, then convert A6h to ASCII, 41(A), 36(6), and transfer.
- 48 -
Digital Weighing Indicator
SI 4300
7. Error & Treatment
7-1. Load Cell Installation
Error
Cause
Treatment
Weight Value is
unstable
1.Load cell broken
2. Load cell isolation
resistance error
3. Weighing part touches
other devices or some weight
is on the weighing part
4. Summing Board Error
1. Measure
input/output resistance
of Load cell.
2.Measure Load cell
isolation resistance
3. Check attach point
with other devices.
1. Load cell Error
2. Load cell connection Error
1. Check Load cell connection
2. Measure Load cell Resistance
Weight Value is
increased regular rate, but
not return to “Zero”
Weight Value is
increased to under Zero
“UN PASS” display
“OL” or “UL” display
Load cell Output wire (SIG+,
SIG-) is switched
Load cell broken or Indicator
connection Error
Power was “ON” when some
weight is on the load cell?
1. Load cell broken or
Indicator connection Error
2. Loading over than Max
Capa.
Remark
1. Input Resistance of
“EXC+” and “EXC-“ is
about 400Ω. ±30
2. Output Resistance of
“SIG+“ and “SIG-” is
about 350Ω. ±3.5
3. Isolate Resistance is
more than 100MΩ
Make wire correction
Load cell Check
Load cell connection Check
Remove weight on the Load cell
1. Load cell Check
2. Load cell connection Check
3. Remove over loaded weight
7-2. Calibration Process
Error
Cause
Err 01
When Maxcapacity/digit value is over 20,000
Err 04
Standard weight value is over than Max Capa
Err 05
Standard weight value is less than 10% of
Max Capa
Err 06
1. Amp. Gain is too big
2. Sig+ and Sig- wire connection error
3. Test weight is not loaded
Err 07
1. Amp. Gain is too small
2. Sig+ and Sig- wire connection error
3. Test weight is not loaded
Err 08
Err A
Treatment
Re-input the Max Capacity, less than 20,000
(Max Capacity / Digit)
Re-input Standard weight value with Number keys,
under Max Capacity
Re-input Standard weight value with Number keys,
more than 10% of Max Capacity
Under “F-function” model, set value is “N.A”
When there is continuous vibration on the
weighing part,, indicator can not process
calibration any more.
Check standard weight’s weight with set value.
If there is difference between set value and real
weight, please re-input the value
(set value is too small)
Check standard weight’s weight with set value.
If there is difference between set value and real
weight, please re-input the value
(set value is too big)
Check the correct value and re-input
- Find vibration cause and remove
- Load cell check
- Load cell cable and connecting condition check
- 49 -
Digital Weighing Indicator
SI 4300
7-3. Digital Weighing Indicator
Error
Display
Cause
Treatment
1. Under “TEST” mode 1, check analogue
value. If you can not get any analogue value
1. Load cell Error
“CELLEr”
No.1
or
“--OL--”
or there is no change although adding load,
2. Load cell cable Error
please check load cell, load cell cable,
3.Load cell connection Error
connection conditions first.
4. A/D Board Error
2. Replace another load cell, and check the
5.It displays under 5000 or
indicator condition. If you have same
over520000.
problem, please replace new indicator and
check A/D board error.
1.If you set “Normal Mode”, please check
1. Power is ON, when some materials
weighing part empty or not before turn on
are on weighing part.
※ Under “Normal Mode”, if there are
the power. If there are some materials
in/on weighing part, please remove those
more than 20% loading of Max
capacity, “Un-Pass” display will be
No.2
“Un-Pass”
materials and turn on the power.
2. Please try to set F02-01(Back-up) mode so
appeared and indicator will stay
that the indicator
can remember first
until removing the load.
※Setting Back-up mode it can
memory empty value, and it
empty value.
※ Under “UNPASS”, please press
becomes set value without
key, then you can exit the mode you
displaying ”Un-pass”)
are.
In case, you set wrong set value for
PRE1, PRE2, FREE FALL or
No. 3
“E”
1.Please check each weighing mode’s setting
FINAL.
condition and make set properly.
Under this display, you canot process
weighing.
No.4
No.5
“FN-SET”
“P-Err”
1. When “FN-Memory” is defected
1. Please contact the distributor or Head
2. When the “FN-Memory” is empty.
Office.
Under Parallel Printer is connected and
1. Please check connection of the print cable.
installed.
2. Please check
1. Parallel printer interface is defected
※ If you only install “Parallel Print” option
or disconnected.
card, you can check to do.
the
trouble of print.
※ Under “CELL-Er”, Relay will not be Output, and Analogue Output(4~20mA/0~10V), either.
- 50 -
Digital Weighing Indicator
SI 4300
7-4. Indicator Test mode
Through this “Test Mode”, you can check basic conditions of Indicator.
This Test consists with total 7 tests.
7-4-1. Enter “Test Mode”
Press
key for 4sec, then display will show “F-Test”.
Under this display, press No.2 key and enter the “Test Mode”.
Under “Test Mode”, please choose each test and check the basic conditions of Indicator.
If you want to exit from each “Test Mode”, press
key.
7-4-2. Test Mode
Test Mode
Contents
Test 1.
Analogue
Value Test
Under “TEST” display, press No.1 key and Enter “TEST1” mode.
Under this mode, you can check the A/D value.
If the A/D value is close to 520,000, or there is no change although pressing or loading
some force on/in weighing part, please check load cell, load cell, cable, connecter, A/D
board.
Test 2.
Key test
Test 3.
Output Relay Test
Test 4.
External Input
Test
Under “TEST” display, press No.2 key and Enter “TEST2” mode.
Press each key, and check the pressed key is operated.
Under “TEST” display, press No.3 key and Enter “TEST3” mode.
This Test will be operated automatically from Relay1 to Relay6.
※ This test will operate automatically, so please remove all materials in/on
weighing parts.
If you can not remove materials, please remove relay terminals.
Under “TEST” display, press No.4 key and Enter “TEST4” mode.
If you press External input S/W, the External S/W No. will be displayed.
If the S/W No. is not displayed, please check connecting condition.
Test 5.
Communication Test
(Com. Port 1)
Under “TEST” display, press No.5 key and Enter “TEST5” mode.
After connecting No.2 and 3 pin of 9pin connector, you can test communication
condition, like TXD or RXD/TXD.
If there is an error in communication, “232-Err” will be displayed with 3times buzzer
sound. The communication is working properly, “232Pass” will be displayed with one
time buzzer sound.
Test 6.
Communication Test
(Com. Port 2)
Under “TEST” display, press No.6 key and Enter “TEST6” mode.
After connecting No.2 and 3 pin of 9pin connector, you can test communication
condition, like TXD or RXD/TXD.
If there is an error in communication, “232-Err” will be displayed with 3times buzzer
sound. The communication is working properly, “232Pass” will be displayed with one
time buzzer sound.
Test 7.
BCD IN Test
Test 8.
BCD OUT
This test is for “BCD Input”.
If you install “BCD IN” option card, you can test this option card operation through
this Test mode.
This test if for “BCD out”
Through this test mode, you can check operation of BCD output.
- 51 -
Digital Weighing Indicator
SI 4300
WARRANTEE CETIFICATION
This product is passed “Sewhacnm”s strict quality test.
If there is defect of manufacturing or abnormal detection within warrantee period, please contact our Agent or
Distributor with this Warrantee certificate.
Then, we will repair or replace free of charge.
WARRANTEE CLAUSE
1. The Warrantee period, we can guarantee, is one(1) year from your purchasing date
2. Warrantee Exception Clause
- Warrantee period is expired.
- Any kinds of Mal-function or defection caused by Modification or Repair without Sewhacnm’s permission.
- Any kinds of Mal-function, Defection, or External damage, caused by operator
- Any kinds of Mal-function, Defection, caused by using spare part from Non-Authorized Distributor or
Agent.
- Any kinds of Mal-function, Defection, caused by not following Warnings or Cautions mentioned on this
manual.
- Any kinds of Mal-function, Defection caused by “Force Majeur”, like Fire, Flood.
- Without presentation of this “Warrantee Certification”.
3. Other
- Any kinds of “Warrantee Certification” without authorized Stamp is out of validity
Manufacturer : SEWHACNM Co.,Ltd.
#504, 302Dong, 397, Seokcheon-ro, Ojeong-gu,
Bucheon-si, Gyeonggi-do, Korea
Product
Digital Weighing Indicator
Model
SI 4300
Serial No.
Tel : +82 70) 4754 6140
Fax :+82 32) 624 0065
AUTHORIZED
sales@sewhacnm.co.kr
STAMP
http://www.sewhacnm.co.kr
Made in KOREA
- 52 -
Digital Weighing Indicator
SI 4300
- 53 -
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project