GS-LXUV-UM-151
For GL100
GS-LXUV
Illumination / Ultraviolet Sensor
USER’S MANUAL
Thank you very much for buying this GRAPHTEC product.
This product is a measurement sensor (hereafter “module”) that
connects to the GL100-N/GL100-WL.
These directions describe preparations and cautions before
measurement.
To ensure safety, please read the operation instructions, etc.
For details on the warnings and how to handle this module, please
read the Quick Start Guide or USER'S MANUAL included on the
CD-ROM (included in the GL100 packaging)
2 How To Measure
1. Power supply
(Refer to Quick Start Guide or USER’S MANUAL.)
Connect this module while power is being supplied to the GL100 by a battery or USB cable.
2. Start-up and operation
(1) Screen display menu flow
After power-on, the GL100 is ready for operation by holding down [MENU] key.
When the module is connected, “Module Type Recognition” screen is displayed.
When the module is not connected, “Module Unconnected State” screen is
displayed.
Operate in accordance with the displayed instructions.
SENSOR ERROR!! Please connect the sensor QUIT key to Power Off
BAT LAN Module unconnected state
Recognition of module types
”Please wait” GL100-LX Sleeping!! ENTER key to start QUIT key to Power Off
BAT LAN GL100-LX Initializing!! Confirmation of the attached items.
•
User’s manual (this book): 1
If by any chance faults are found, please contact the store where you
bought the item.
* Please note that items mentioned in this book may change without
prior notice.
MANUAL-UV
This section describes the name and function of each part.
1. Hook portion
2. Illumination sensor
Hold down the [QUIT] key (approx. three seconds)
to put the module into standby state.
When running on batteries, the module will
automatically go into standby state after three
minutes of no operation.
Press the [ENTER] key while in standby state to
return to the free-running screen.
ILLUM: If the submenu shows
then there
are selections in those directions.
Numerical entry screen
<How to set>
Numbers can be inputted by increasing or decreasing the value with the
and
keys.
[AMP] 1/8
UV-A Range:30mW/cm 2 ILLUM. Range: 200000lx▽ 5. Cable packing
GL100 main module
1. Hook portion ..................
2. Illumination sensor ........
3. Ultraviolet sensor ..........
4. Connector .....................
Used to mount to a wall.
Sensor unit that measures illumination rays.
Sensor unit that measures ultraviolet rays.
Used to connect to the connector on the
GL100 module.
5. Cable packing ............... This packing is used when connecting the connector.
When measuring ultraviolet rays that have an effect on
the eyes or the skin, wear protective glasses, apply light
shielding, etc.
If the sensor unit becomes dirty, this may affect
measurements. Wipe it down with a soft cloth.
•
sure to handle the sensor unit in a way that it will
• Make
not crack. If the sensor unit is cracked it may affect
easurements. Replace the sensor if cracked.
After connecting the GL100 to modules or sensors, please always check/set the
time and date.
AMP input condition settings
Range
ILLUM
2000, 20000, 200000 lx
(3) DATA setting
Set the Sampling and Capture Mode those will be recorded to the data
recording media.
The recorded data’s size will be displayed in the information for the SD
card being recorded to. Please take note of it.
[DATA] 3/8
Sampling:1s▽ Capture MODE:CONT.▽ Capture DIST:SD Card▽
Free CAPA:498180 DATA recording condition setting
Sampling
Capture MODE
Capture DIST
500 ms, 1, 2, 5, 10, 20, 30 s,
1, 2, 5, 10, 20, 30, 60 min
CONT, 1 Hour, 24 Hour
Memory, SD card
Accessing SD card
Low battery
Alarm active
POWER(Green)
Power supplying
Wireless LAN connection
possible status
Select the conditions for beginning data recording after measurement starts.
Stop
: Pressing the [START/STOP] key on this module will start/stop recording.
: The recording will start with the trigger source conditions after
pressing the [START/STOP] key.
The recording will stop after pressing the [START/STOP] key.
: The recording will start after pressing the [START/STOP] key
and will be stopped with the trigger source conditions.
[TRIGGER] 3/8
TRIG Setting:Off▽ TRIG Source:Off▽ Access light
Flash once every 5 seconds
Flash once every 10 seconds
Flash once every 10 seconds
Flash once every 5 seconds
accessing an SD card, do not remove the SD card. The
• When
data may not write properly or the SD card may be damaged.
When
is displayed, replace the battery or
• connect“lowthebattery”
USB interface to supply power as soon as
possible. Caution: Batteries cannot be replaced when
recording data. Replace them after the recording has finished.
(2) Recording completion
• Press the [START/STOP] key to stop measuring.
• The screen display will change to the standby
screen display.
• Press [ENTER] key to change to the
free-running screen display.
GL100-LX Sleeping!! ENTER key to start If the CO2 sensor (GS-CO2) is included in the assembly, it
cannot be powered with batteries.
CO2 Sensor
(GS-CO2)
Branch adapter for GS
(GS-DPA)
* The Gl100’s power is
provided through a
USB cable
connection
GL100 main module
(1) Screen display menu flow
After connecting the power supply, connect this module and operate it in
accordance with the content displayed on the screen.
* Refer to “2 How To Measure” above and the USER'S MANUAL for the
CO2 sensor.
(2) Free-running screen
STOP ALM. 1:28
lx
ILLUM: 0 m UV-A: 0.016 W
PP
CO2: 736 m
BAT LAN SD S: 1.0s
Hold down the [QUIT] key (approx. three seconds)
to put the module into standby state.
Press the [ENTER] key while in standby state to
return to the free-running screen.
2. Combined illumination / ultraviolet sensor and temperature
and humidity sensor measurement
Composite measurement can be done by using the branch adapter for GS
(GS-DPA) and the temperature and humidity sensor (GS-TH) (each sold
separately).
Illumination /
Ultraviolet sensor
(GS-LXUV)
Temperature and
humidity sensor
(GS-TH)
QUIT key to Power Off
BAT LAN Check the recorded data with the application software included with this
module using the method below (for details, refer to the USER'S MANUAL).
(1) Connect the USB interface and check the online data
(2) Insert the SD card into PC and check the data directly
(3) Check the data directly from PC via wireless LAN
5 Specifications
Contents
Item
Measurement data
Measurement channel
Measurement range
Measurement accuracy
*1
Spectral sensitivity
(4) TRIGGER setting
Off
Start
Composite measurement can be done by using the branch adapter for GS
(GS-DPA) and the CO2 sensor (GS-CO2) (each sold separately).
4 How To confirm The Data
(2) AMP setting
The ultraviolet ray range is a fixed 30 mW/cm2. Please set the
measurement level to the illumination level.
4. Connector
STATUS (Orange)
Item selecting screen
Press the [MENU] key on the free-running screen to go to the setting screen.
<How to set>
Select the item with the directional keys (
) and press the [ENTER] key.
[Date/Time] 6/8
Type: YY:MM:DD▽ Date: 2014-01-04 Time: 03:18:26▽ 3. Ultraviolet sensor
The module’s status is shown with the lamp display.
(1) Setting screen operation
↑ 1s 2s 3s 6s 10s ↓ 1. Combined illumination / ultraviolet sensor and CO2 sensor
measurement
Illumination /
Ultraviolet sensor
(GS-LXUV)
3 Recording
3. Setting
[DATA] 3/8
Sampling:1s▽ Capture MODE:CONT.▽ Capture DEST:SD Card▽
Free CAPA:498180 1 Part Names
•
Module start-up
STOP ALM. 1:28
Off
Level Off
/ Mode Level
ILLUM Off
H Value setting
L
UV-A
Off
H
Value setting
L
After pressing [START] key, when the module is in awaiting recording start,
“ARMED” is displayed, and then when recording is started, "REC" is displayed.
When alarm occurs, “ALM” is displayed.
REC. ALM. 1:28
Current time
ILLUM: Note: The current time display can
lx
440 be switched to the elapsed
UV-A: time with the [QUIT] key when
m 0.030 W
recording.
BAT LAN SD S: 1.0s
When battery
Sampling interval
replacement
“SD” is displayed during accessing the SD card.
is required, “BAT”
is displayed.
LAN: displayed when the wireless LAN connection is enabled.
* You can switch to the Accumulation screen with the and keys when recording data.
<Operation>
Press [ENTER] key.
lx
440 UV-A: m 0.030 W
BAT LAN SD S: 1.0s
Alarm
Press the [START/STOP] key to start measuring with the set conditions.
Standby state
(2) Free-running screen
ALARM settings
[ALARM] 3/8
Alarm:Off▽ 6 Combined Measurement
(1) Recording
After opening the package, please confirm that there are no
problems (scratches and dirt) on the exterior before use.
604309092
<Operation>
Connect the module.
Confirmation of the exterior
(5) ALARM setting
Set the alarm information. The parameters will vary depending on the
setting range. Please set the number level.
TRIGGER capture condition settings
TRIG setting
TRIG Source
Off, Start, Stop
Off
Alarm
Date
Date, Time
Measurement resolution
Responsive
Accumulated function
Sampling interval
Illuminance, ultraviolet , accumulated illuminance, accumulated
ultraviolet intensity
* The accumulated illuminance and ultraviolet intensity are
displayed only when recording.
Illuminance: 1 channel
UV intensity: 1 channel
Illuminance: 0 to 200k lx
UV intensity: 0 to 30 mW / cm2
Illuminance: ±5% (at 25°C)
Ultraviolet intensity: ±5% (at 25°C)
Illuminance: 400 to 750 nm (peak 560 nm)
UV intensity: 260 to 400 nm (peak 370 nm)
Illuminance: 0.1 lx
UV intensity: 0.001 mW/cm2
Illuminance: 1 ms or less
UV intensity: 1 ms or less
Accumulated illuminance: 0 to 999,999 k lxh
Accumulated UV intensity: 0 to 999,999 mW/cm2h
0.5, 1, 2, 5, 10, 20, 30 sec.
1, 2, 5, 10, 20, 30, 60 min.
OFF / Level
approximate 20 cm
-10 to 50°C, 80% RH or less (non-condensing)
46 × 66 × 19 mm (not including protruding parts)
Alarm
Cable length
Usage environment
External dimensions
[W×D×H] (approximate)
Weight (approximate)
65 g
*1 This is the measurement accuracy when our standard light source is used and
compared with the calibrated test instrument.
Branch adapter for GS
(GS-DPA)
GL100 main module
(1) Screen display menu flow
After connecting the power supply, connect this module and operate it in
accordance with the content displayed on the screen.
* Refer to “2 How To Measure” above and the USER'S MANUAL for the
temperature and humidity sensor.
(2) Free-running screen
STOP ALM. 1:28
TEMP: +26.16℃
HUM.: +37.3%
DEWP: +10.30℃
PP
CO2: 736 m
BAT LAN SD S: 1.0s
Hold down the [QUIT] key (approx. three seconds)
to put the module into standby state.
When running on batteries, the module will
automatically go into standby state after three
minutes of no operation.
Press the [ENTER] key while in standby state to
return to the free-running screen.
< Extension cable >
The module can be used approx. 1.5 m away from the GL100 by using an
extension cable for GS (GS-EXC). However, you cannot connect and use multiple
extension cables.
CAUTION
It is not possible to be used by connecting two same sensors.
<Spectral Sensitivity Characteristic>
Illumination
1.0
Spectral
sensitivity
1.0
Ultraviolet
Spectral
sensitivity
400
560
Wavelength
750 (nm)
503-10 Shinano-cho, Totsuka-ku Yokohama
244-8503, Japan
260
370
Wavelength
400 (nm)
September 1, 2014