LI-610 Dew Point Generator Cannot Reach Below Ambient

LI-610 Dew Point Generator Cannot Reach Below Ambient
LI-610 Dew Point Generator Cannot Reach Below Ambient Temperatures
Occasionally, LI-610 Dew Point Generator Users will not be able to achieve a temperature set-point below ambient
temperatures. This can be caused by a combination of issues.
1.
2.
3.
4.
The Radiator and/or the Condenser Reservoir are low or empty.
Coolant pump is not on or is not working.
Blockages have formed in the cooling water path.
Peltier coolers are not receiving proper voltage.
Please see the following pictures and labels for descriptions on troubleshooting these issues:
1.
The Radiator and/or the Condenser Reservoir are low/empty.
a. Make sure the Radiator is full (pg 2-1 in Manual). Only use
distilled or de-ionized water (approximately 200 ml total). Add
15-20 drops (2 ml) of algicide. Fill until water level is visible in
the tube.
b. Make sure Condenser is full (pg 2-3 in Manual). Only use
distilled water (approximately 20-25 ml total), and make sure the
water level in the Condenser Reservoir is between „Max‟ and
„Min‟. Replace cap plug and start LI-610. Watch level to see if
more water needs to be added.
2.
Fill until water level is visible in the tube.
Coolant pump is not on or is not working.
a. Check to make sure that the coolant pump is working properly.
Disconnect the "To Cooler" tubing on the backside of the LI-610.
Have the “COOLER” switch on the front panel set to OFF. Power
ON the LI-610 and briefly toggle the "COOLER" switch to the ON
position. If the pump is working properly, water will discharge
from the "To Cooler" port. If water does not discharge, contact LICOR.
3.
Blockages have formed in the cooling water path.
a. Check for a blockage in the cooling water path. (See the
following pictures for an example of this blockage).Without
coolant circulating through the Peltier coolers to remove heat,
„Min‟ and „Max‟ Label
they cannot achieve a temperature target below ambient. There
are two Peltier coolers, one on each side of the condenser block.
There are water lines that connect the two coolers, and also run to and from the radiator. A blockage
can occur in any of the hose fittings that connect to the Peltier coolers. To remove the blockage,
disconnect either the "From Cooler" or "To Cooler" hoses from the back of the LI-610.Use the syringe
supplied with the LI-610 and seal it to the hose fitting as best you can. Then use the syringe to push
and pull water through the system until the water flows freely. This should dislodge any blockage that
LI-COR Biosciences
4421 Superior Street Lincoln, NE 68504
Toll-Free: 800-447-3576
Email: envsales@licor.com
www.licor.com
1
might inhibit coolant circulation. If a blockage is
detected, flush it out of the lines with water. The
Using the syringe, pull water up out of condenser to help free
blockages. Also push water down through condenser. Do
both tubes.
precipitates that cause these blockages are generally
formed during storage by a reaction between water,
air, and the brass hose fittings. These precipitates can
be prevented by keeping the radiator reservoir full of
de-ionized water during storage (if stored in
temperatures above freezing).
b. Reactions between water, air, and brass can build up
precipitates and cause blockages. Keeping the
Reservoir full can help prevent this. These blockages
can happen at several tube connectors (hose barbs)
located on the condenser. Inserting a thin, flexible (but
firm) wire can sometimes break through the blockages
and loosen the precipitates. These can be flushed out
afterwards with the syringe.
“To Cooler” tube
“From Cooler” tube
Precipitate
Blockage
4.
Peltier coolers are not receiving proper voltage.
a. If the above measures do not solve the problem, the Peltier coolers may not be powered properly.
Contact LI-COR for servicing this problem.
If none of the above suggestions allow for the LI-610 Dew Point Generator to reach be Temperatures below Ambient
Temperatures, contact LI-COR for servicing the LI-610.