2PTR2 Strain Gauge Signal Conditioner

2PTR2 Strain Gauge Signal Conditioner
2PTR2
2PTR2
Strain Gauge
Signal Conditioner
PRODUCT MANUAL
2PTR2Man-GB-rev1
Parker Hannifin AB, Parker SSD Drives
Head office:
Montörgatan 7, SE-302 60 Halmstad, Sweden
Tel: +46 35 177300
Fax: +46 35 108407
www.SSDdrives.se
Performance
Power supply:
Bridge supply
Power stability:
Bridge excitation:
Strain Gauge
AC
Signal
Conditioner
Frekvensomrik-
230V +10% -15%, 50-60Hz
230V +10% -15% will give less than
0.02% change in output voltage
6V or 10V /70mA selectable
Amplifier for loadcells and
pressure transducers
Temperature stability: 20ppm/°C
Input
Range:
The prime funcion of the strain gauge signal conditioner is to convert a dc signal from a strain gauge bridge
(e.g. loadcells or pressure transducers) into a high level
signal.
0-10mV to 0-10V
selectable in 19 steps
Output 1
0-10V
Summing amplifier for external
0-±10V summed to output 1.
Output 2
Range selectable:
0-20mA or 4-20mA
Meter indication
output
0-1mA unfiltered/filtered selectable
Enclosure
Two alternatives:
IP20 standard single rack
IP54 die cast Alu-box, EMC-proof
Ambient temperature: 0-50°C
Weight:
IP20
1.2 kg
IP54
1.65 kg
Trouble shooting
The instrument is provided with a socket, for the possibility
to connect a diagnostic instrument with 17 testpoints.
IP54
IP20
INPUT OPTIONS: 0-10mV TO 0-10V
VOLTAGE OUTPUT 1: 0-10VDC
STANDARDS
CURRENT OUTPUT 2, OPTIONS: 0-20mA, 4-20mA
The 2PTR2 meets the following standards when installed
in accordance with the relevant product manual.
EMC
EN61326-1:1997
EN61326/A1:1998
LVD
EN61010-1:1993 /A2:1995
EN61010-1/A2:1995
Web site:
www.SSDdrives.se
Mail address:
SEPARATE OUTPUT FOR INDICATION
EXCITATION VOLTAGE OPTIONS: STANDARD 6VDC
OR 10VDC OR TO CUSTOMER’S SPECIFICATION
POSSIBILITY TO CONNECT A DIAGNOSTIC UNIT
info.se.ssd@parker.com
2PTR2Man-GB-rev1
2PTR2
TECHNICAL SPECIFICATION
ORDERING CODE
OPTIONS
ORDERING CODE
IP20
IP20 enclosure
2PTR2-IP20
IP54
IP54 enclosure
2PTR2-IP54
DIMENSIONS
IP20 enclosure
220
56
Ø5
193
178
2PTR2
IP54 enclosure
171
137
55
25
272
245
Ø6
Web site:
www.SSDdrives.se
Mail address:
info.se.ssd@parker.com
2PTR2Man-GB-rev1
2PTR2
Web site:
www.SSDdrives.se
Circuit diagram
AC
Frekvensomrik-
Mail address:
info.se.ssd@parker.com
2PTR2Man-GB-rev1
Component layout
2PTR2
LK0
Web site:
www.SSDdrives.se
Mail address:
info.se.ssd@parker.com
2PTR2Man-GB-rev1
LK0
Web site:
www.SSDdrives.se
Mail address:
info.se.ssd@parker.com
5. Make a final check by loading at each end of the
roller. Apply 50% load and check that the output gives
you 5V.
4. Apply full load on the loadcells and calibrate pot. P1
until 10V reading on the output terminals.
3. Measure the ouput signal on terminals 11 and 12.
Unload the loadcells and set the ”zero” by pot.
P102. If P102 doesn’t give enough zero adjustment,
increase the SW101.
2. Measure the input signal at zero and full load and
calculate the difference in mV. Set the span on SW1SW3 according to the input selection table.
1. Check the excitation voltage on terminals 3 and 4 to
correspond to the loadcell ratings. Normally 10.0V is
factory set. Adjustable by pot. P101. (Put link LK101
for 6V operation.)
CALIBRATION OF 2PTR2 FOR LOADCELL
APPLICATION
2PTR2
Calibration procedure
AC
for loadcell application
Frekvensomrik-
2PTR2Man-GB-rev1
LK0
Web site:
www.SSDdrives.se
Mail address:
info.se.ssd@parker.com
5. Disable the calibration switch.
4. Measure the output signal on terminal 11 and 12.
Adjust P1 until reading 80% on the output terminals.
3. Put the calibration link LK0. Measure the input in mV
and read 80% of max input. Convert it to full load
input and set the span on SW1-3 according to the
input selection table.
2. Mesure the output signal on terminals 11 and 12
with unloaded transducer and set the ”zero” by pot.
P102. If P102 doesn’t give enough zero adjustment,
increase the SW101.
1. Check the excitation voltage on terminals 3 and 4
to correspond to the transducer ratings. Normally
10.0V is factory set. Adjustable by pot. P101. (Put link
LK101 for 6V operation.)
CALIBRATION OF 2PTR2 FOR PRESSURE TRANSDUCER APPLICATION
Calibration procedure
2PTR2
for pressure transducer application
2PTR2Man-GB-rev1
2PTR2
Web site:
www.SSDdrives.se
Connection diagram
AC
Frekvensomrik-
Mail address:
info.se.ssd@parker.com
2PTR2Man-GB-rev1
Block diagram & diagnostic
Web site:
www.SSDdrives.se
Mail address:
info.se.ssd@parker.com
2PTR2
2PTR2Man-GB-rev1
AC
Frekvensomrik-
Web site:
www.SSDdrives.se
Mail address:
info.se.ssd@parker.com
2PTR2Man-GB-rev1