Factory FCA-X - Scale Center of Arizona

Factory FCA-X - Scale Center of Arizona
Factory FCA-X
53
288
107
144
Baugrösse EDE…
87
268
190
112
107
SW13
630
800
SW38
min.116
max.140
144
Baugrösse IGG…
190
134
: for use in Ex-Zone 1, 2 (ATEX) | DIV 1, 2 (FM, CSA)
: built-in calibration weight, isoTEST function
: high resolution, stable weighing results (time and temperature) thanks to the
monolithic weighing system, automatic reference weight optimization
: practical, combinable applications:
level 1: weighing, weighing in percent, net-total formulation, mass unit
conversion, weighing under ambient conditions (animal weighing), calculation
level 2: checkweighing, time-controlled measurement
level 3: totalizing, density determination, recalculation
: additional RS-485 interface (optional)
: backlit display
: ingress protection rate: IP65 (except FCA64EDE-HX = IP44)
: with verification of Intrinsic safety
General Information
Approval Europe (ATEX)
II 2 G EEx ib IIC T4
Approval USA/CND (FM/CSA)
CL1, DIV1. GP A, B, C, D, E, F, G T4 (CL1, Zone 1, GP IIA, IIB, IIC)
AC power source/power requirements
in safe areas: 110V... 220V, 24V
in hazardous Areas: 110V...230V, 24V, accumulator
48 – 63 Hz
Frequency
Ambient/Operating temperature range
0 ... +40 °C (273 ... 313 K, 32 °F ... 104 °F)
Power consumption
(when connecting via ING2)
Non-regulated input-voltage
Adaptation to ambient conditions and application requirements
Data output speed (depends on the filter level selected)
Selectable weight units
(models standard)
maximum 34 VA; average 18 VA
(display unit 12 to 15VA (average), max. 28VA)
10,5 V~ bis 30V~ (display unit)
by selection of 1 of 4 optimized filter levels
0.1 – 0.4 s
Grams, kilograms, carats, pounds, ounces, troy ounces, Hong Kong taels,
Singapore taels, Taiwanese taels, grains, pennyweights, milligrams
parts per pound, Chinese taels, momme, Austrian carats, tola, bath, mesghal
Selectable weight units (verifiable models)
Grams, kilograms
Built-in interface (display unit)
Built-in interface
RS-485 (to weighing platform) via round connector
RS-232 (to industrial printer YDP01IS-S0CE)
Model-specific information
Standard models
Model
Readability
Weighing range
Maximum load
Tare range (subtractive)
g
kg
kg
kg
FCA12EDE-PX
0.1/0.2
6/12
130
-12
FCA16EDE-HX
0.1
16
130
-16
FCA34EDE-PX
0.1/0.2/0.5
8/16/34
130
-34
FCA34EDE-HX
0.1
34
130
-34
Max. preload when starting
calibration/adjustment (isoTEST)
(scale must be zeroed)
Repeatability
Linearity
Sensity drift within
kg
ca. 10
ca. 19
ca. 21
ca. 21
≤±g
≤±g
0.05/0.05
0.2
0.05
0.2
0.05/0.05/0.1
0.2
0.1
0.2
+10 ... +30 °C£±/K-6
Response time (average)
≤±/K
s
2 x 10-6
1.5
2 x 10-6
1.5
2 x 10-6
1.5
2 x 10-6
1.5
External calibration weight
(of at least accuracy class…)
Platform dimensions (W x D x H)
Ingress protection rate
Net weight, approx.
g
10 (F1)
10 (F1)
10 (F2)
10 (F1)
mm
kg
300x400
IP65
16
300x400
IP65
16
300x400
IP65
16
300x400
IP65
16
g
kg
kg
kg
FCA64EDE-SX
1
64
130
-64
FCA64EDE-HX
0.1
64
130
-64
Max. preload when starting
calibration/adjustment (isoTEST)
(scale must be zeroed)
Repeatability
Linearity
Sensity drift within
kg
ca. 45
≤±g
≤±g
+10 ... +30 °C£±/K-6
Model
Readability
Weighing range
Maximum load
Tare range (subtractive)
FCA150IGG-HX
FCA300IGG-HX
ca. 45
1
150
600
150
ca. 120
2
300
600
300
ca. 300
0.3
1
0.1
0.5
1
4
2
8
≤±/K
3 x 10-6
3 x 10-6
Response time (average)
External calibration weight
(of at least accuracy class…)
s
g
1.5
10 (F2)
1.5
10 (F1)
2.5 · 10–6
1.5
50 (F2)
2.5 · 10–6
1.5
50 (F1)
Platform dimensions (W x D x H)
Ingress protection rate
Net weight, approx.
mm
300x400
IP65
16
300x400
IP44
16
800x600
IP67
800x600
IP67
70
70
kg
Models verified at the Factory by the Manufacturer
Model
Type designation
Accuracy class
Readability d
Weighing range
Verification scale interval e
Min. preload
Maximum load
Tare range (subtractive)
Max. preload when starting
calibration/adjustment (isoTEST)
(scale must be zeroed)
Application range acc. to CD
Response time (average)
Platform dimensions (W x D x H)
Ingress protection rate
Net weight, approx.
Model
Type designation
Accuracy class
Readability d
Weighing range
Verification scale interval e
Min. preload
Maximum load
Tare range (subtractive)
Max. preload when starting
calibration/adjustment (isoTEST)
(scale must be zeroed)
Application range acc. to CD
Response time (average)
Platform dimensions (W x D x H)
Ingress protection rate
Net weight, approx.
g
kg
g
g
kg
kg
g
s
mm
kg
g
kg
g
g
kg
kg
g
s
mm
kg
FCA12EDE-PXCE
FCA16EDE-HXCE
isoTEST in conjunction with BD BF
FCA34EDE-PXCE
FCA34EDE-HXCE
0.1/0.2
0.1
6/12
16
1
1
5
5
130
130
≤100% of maximum weighing range
ca. 10
ca. 19
0.1/0.2/0.5
8/16/34
1
5
130
0.1
34
1
5
130
ca. 21
ca. 21
5-12000
1.5
300 x 400
IP65
16
5-34000
1.5
300 x 400
IP65
16
5-34000
1.5
300 x 400
IP65
16
5-16000
1.5
300 x 400
IP65
16
FCA64EDE-SXCE
FCA150IGG-HXCE
isoTEST in conjunction with BD BF
FCA300IGG-HXCE
1
1
64
150
10
10
50
50
130
600
≤100% of maximum weighing range
ca. 45
ca. 120
20
300
20
1000
600
50-64000
1.5
300 x 400
IP65
16
1–300
1.5
800x600
IP67
70
0.05–150
1.5
800x600
IP67
70
ca. 300