The Horiba F-7X BW is a pH/Water Quality Meter that offers a wide range of capabilities for measuring and analyzing water quality parameters. With its 8-pin mini DIN connector and optional USB or serial cables, it can easily connect to a PC for data transfer and analysis. The F-7X BW features multiple measurement modes, including pH, mV, ion, conductivity, salinity, resistivity, and TDS. It also has calibration modes for pH, ion, and conductivity, allowing users to ensure accurate and reliable measurements.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Manual
horiba F-7X BW Command Reference Manual | Manualzz

CODE:GZ0000438938

February, 2016 © 2016 HORIBA, Ltd.

pH/Water Quality Meter F-7X BW

Low-spec Command Reference

Preface

This manual describes the communication command list of the pH/Water Quality Meters with serial communication function, LAQUA-PH1100/PH1200/PH1300/PC1100/EC1100.

The contents of this manual are subject to change without notice.

Caution

Use the optional USB cable (part number: 3200373941) or serial cable (part number:

3014030151) to connect the instrument to a personal computer (referred to as PC in the rest of this document).

Make sure that the transfer format used in the instrument and a PC are the same.

When different transfer format is used, a communication error occurs and the online mode does not start up, and as a result RS-232C communication cannot be performed. Also, when the transfer format is changed, turn OFF the power of the instrument and PC and then reboot them.

The transfer format of the instrument is as follows.

 ・

Baud rate: 2400 bps

 ・

Character length: 8 bits

 ・ Parity: None

 ・

Stop bit: 1 bit

If you write the program for serial communication, at first, write the command to change the instrument to the online mode. By changing the instrument to the online mode, its operation keys except for key are locked and it changes to serial communication mode.

If the instrument’s power is turned OFF, reset the online mode.

If the instrument does not receive the command or occurs any errors after it requested the data, add the waiting time of a few seconds before request the data again.

If the instrument received the data continuously, the instrument does not response.

The instrument cannot be controlled by using the DCD, CTS, and DSR.

It is necessary to switch RTS to ON to perform communication. Set it 2.4 V above.

Pin assignment of the instrument and the external instrument are follows.

Instrument side (A connector MINI DIN 8 PIN)

2 TX

3 RX

4 CTS

5 COM

External instrument side (B connector DSUB 9 PIN)

2 TX (RX at an external instrument side)

3 RX (TX at an external instrument side)

5 COM

7 CTS (RTS at an external instrument side)

2 Miyanohigashi, Kisshoin Minami-ku, Kyoto 601-8510 Japan

http://www.horiba.com

For any questions regarding this product, please contact your local agency, or inquire from the HORIBA website.

(http://global.horiba.com/contact_e/index.htm)

・Command function list (control)

Item

Online/Offline

Potential follow-up stop pH measurement mode mV measurement mode

Ion measurement mode

Conductivity measurement mode

Salinity measurement mode

Resistivity measurement mode

TDS measurement mode

Calibration mode pH calibration start

Ion calibration start

Conductivity calibration start

Salinity calibration start

Calibration clear

Data IN

Command

Header Name

OL

C

(Control)

BR

PH

MV

IO

CO

SA

OH

TD

CM

CP

CI

CD

CS

CC

IN

Function

Changes between the online mode and the offline mode.

Releases the hold state and returns to instantaneous value display state.

Waits for the pH measurement.

Waits for the ORP measurement mode result.

Waits for the ion measurement.

Waits for the conductivity measurement.

Waits for the salinity measurement.

Waits for the resistivity measurement.

Waits for the TDS measurement.

Waits for the calibration.

Starts the calibration and inspection before use in the pH measurement mode or hold state.

Starts the ion calibration in the ion measurement mode or hold state.

Starts the conductivity calibration in the conductivity measurement or hold state.

Starts the salinity calibration in the salinity measurement or hold state.

Clears the calibration data in the measurement mode.

Stores the measurement data.

Response from pH meter

OK [CR][LF] or

ER, n [CR][LF] n = 1: n = 2: n = 3:

A non-existent command was entered.

The command was entered when the pH meter cannot accept it.

An unacceptable number was entered in the command.

2

●Online/Offline command

C, OL, x [CR][LF]

0 or 1    0: Offline   1: Online

Name (Online/Offline operation)

Header

When the instrument accepts the online command, it enters the online mode and the keys cannot be operated.

●Potential follow-up stop command

C, BR, x [CR][LF]

Channel (1 or 2)

Name (Potential follow-up stop)

Header

Releases the hold state and returns to instantaneous value display state.

This is invalid for auto stability mode.

●pH measurement mode command

C, PH, x [CR][LF]

Channel (1 or 2)

Name (pH measurement mode)

Header

When the instrument is in the online mode, this is valid for modes other than the calibration mode.

Waits for the pH measurement.

3

●mV measurement mode command

C, MV, x [CR][LF]

Channel (1 or 2)

Name (mV measurement mode)

Header

When the instrument is in the online mode, this is valid for modes other than the calibration mode.

Waits for the ORP measurement mode result.

●Ion measurement mode command

C, IO, x [CR][LF]

Channel (1 or 2)

Header

Name (Ion measurement mode)

When the instrument is in the online mode, this is valid for modes other than the calibration mode.

Waits for the ion measurement.

●Conductivity measurement mode command

C, CO [CR][LF]

Name (Conductivity measurement mode)

Header

When the instrument is in the online mode, this is valid for modes other than Calibration mode.

Waits for the conductivity measurement.

4

●Salinity measurement mode command

C, SA [CR][LF]

Header

Name (Salinity measurement mode)

When the instrument is in the online mode, this is valid for modes other than the calibration mode.

Waits for the salinity measurement.

●Resistivity measurement mode command

C, OH [CR][LF]

Header

Name (Ion measurement mode)

When the instrument is in the online mode, this is valid for modes other than the calibration mode.

Waits for the resistivity measurement.

●TDS measurement mode command

C, TD [CR][LF]

Name (TDS measurement mode)

Header

When the instrument is in the online mode, this is valid for modes other than the calibration mode.

Waits for the TDS measurement.

5

●Calibration mode command

C, CM, x, y [CR][LF]

0: Calibration mode end,  1: Calibration mode start

Channel (1 or 2)

Name

Header

Waits for the calibration at specified mode.

※ To return to the measurement mode without calibrating after entering the calibration mode, enter "0: Calibration mode end".

●pH calibration start command

C, CP, x, xxxxxx [CR][LF]

Calibration value to be set (when custom calibration is set)

Channel (1 or 2)

Header

Name

Starts the calibration when waiting for pH calibration.

Even when the setting is other than the custom calibration, it is necessary to enter the calibration value (It will not be used.).

ER, 2 is returned when the measurement value or temperature value is Or (Ur).

Setting range of calibration value

Specified by three decimal places.

0.000 to 14.000

※ Automatically returns to the measurement mode after the calibration is finished.

※ Automatically returns to the measurement mode even if a calibration error is generated.

※ A calibration error can be checked by the alarm inquiry (R, AL).

※ Calibration errors are released by the alarm clear (R, AR).

6

●Ion calibration start command

C, CI, x, yyyy, z [CR][LF]

Auxiliary unit of calibration value unit to be set.

0:-

1:μ

2:m

Follows the unit setting for mol/L or g/L.

Calibration value to be set

Channel (1 or 2)

Name

Header

Starts the ion calibration when waiting for ion calibration.

ER, 2 is returned when the temperature value is Or (Ur).

Calibration is possible when the measurement value is Or (Ur).

Setting range of calibration value

Unit: mg/L, μg/L, mmol/L, μmol/L

0.00 to 9.99

(Possible at [SP][SP]X or [SP]X.X.)

10.0 to 99.9

(Possible even at [SP]XX.)

[SP]100 to [SP]999

※ Automatically returns to the measurement mode after the calibration is finished.

※ Automatically returns to the measurement mode even if a calibration error is generated.

※ A calibration error can be checked by the alarm inquiry (R, AL).

※ Calibration errors are released by the alarm clear (R, AR).

7

●Conductivity cell constant calibration start command

C, CD, yyyyy, z [CR][LF]

Auxiliary unit of calibration value unit to be set.

0: -

1: μ

2: m

Follows the unit setting for S/m or S/cm.

No limit for input value when automatic calibration is ON.

Calibration solution concentration to be set

Even if the automatic calibration is set to ON, it is necessary to enter the calibration value (It will not be used).

No limit for input value when automatic calibration is ON.

Header

Name (Conductivity calibration start)

Starts the conductivity calibration when waiting for conductivity calibration.

ER, 2 is returned when the measurement value or temperature value is Or (Ur).

Setting range of calibration value

When the unit is μS/m: 0.001 to 2000.0

When the unit is mS/m or S/m: 0.001 to 200.0

※ Automatically returns to the measurement mode after the calibration is finished.

※ Automatically returns to the measurement mode even if a calibration error is generated.

※ A calibration error can be checked by the alarm inquiry (R, AL).

※ Calibration errors are released by the alarm clear (R, AR).

8

●Calibration clear command

C, CC, x [CR][LF]

Channel (1 or 2)

Name (Calibration clear)

Header

Clears the calibration data in the measurement mode.

●Data IN command

C, IN [CR][LF]

Header

Name (Data IN)

Stores the measurement data.

9

・Command function list (Request data)

Item

Request of the calibration history of pH

Request of the calibration history of ion

Request of the calibration history of conductivity

Request of the measurement value

Request of the clock data

Request of the number of stored memories

Request of memory data

Alarm inquiry

Clear alarm

Command

Header Name

PC

IC

R

(Request

Data)

CC

MD

OT

MC

MS

AL

AR

Function

Gets the latest calibration history of pH.

Gets the latest calibration history of ion.

Gets the latest calibration history of conductivity.

Gets the measurement value of specified channel.

Gets the clock data.

Gets the number of data stored in the memory.

Gets the memory data to be specified.

Gets the alarm code in the instrument.

Clears the alarm code in the instrument.

Response from pH meter

When it is OK:

Describes in each command.

or

ER, n [CR][LF] n = 1: n = 2: n = 3:

A non-existent command was entered.

The command was entered when the pH meter cannot accept it.

An unacceptable number was entered in the command.

10

●Request command and response of the calibration history of pH

Request command

R, PC, x [CR][LF]

Channel

Name (Request of calibration history of pH)

Header

11

Response from pH meter

RPC, x, x, x, x, Xxxxx.x, x, (⇒)

Header

Inspection before use 0: No 1: Yes

Asymmetry potential (7 digits including symbol)

±XXXX.X

Temperature setting 1 digit 0: ATC 1: MTC

Calibration result 0: good

Calibration points 1 digit  1 to 5

Channel 1 digit  1: CH1 2: CH2

(⇒)

Slope (sensitivity) (5 digits including decimal point and symbol)

For one-point calibration⇒First-point sensitivity: [SP]

For two-point calibration⇒First-point sensitivity: Potential diff. between 1st and 2nd solutions, second-point sensitivity [SP]

For three-point calibration⇒First-point sensitivity: Potential diff. between 1st and 2nd solutions, second-point sensitivity: Potential diff. between 2nd and 3rd solutions, third-point sensitivity [SP]

For four-point calibration⇒First-point sensitivity: Potential diff. between 1st and 2nd solutions, second-point sensitivity: Potential diff. between 2nd and 3rd solutions, third-point sensitivity: Potential diff. between 3rd and 4th solutions, fourth-point sensitivity [SP]

For five-point calibration⇒First-point sensitivity: Potential diff. between 1st and 2nd solutions, second-point sensitivity: Potential diff. between 2nd and 3rd solutions, third-point sensitivity: Potential diff. between 3rd and 4th sensitivity solutions, fourth-point sensitivity: Potential diff. between 4th and 5th solutions, fifth-point sensitivity [SP] xxxx, xx, xx, xx, xx, xx, xx.xxx, xx.xxx, xx.xxx, xx.xxx,

Xxxx.x, Xxxxx.x, xxx.x

Xxxx.x,

Xxxx.x,

Xxxx.x,

Xxxxx.x,

Xxxxx.x,

Xxxxx.x, xxx.x

x.xxx

First point calibration

Second point calibration

Third point calibration

[CR][LF]

Inspection data before use

Repeatability

(5 digits including decimal point and symbol)

0.000 to 9.999

Potential (7 digits including symbol) ±XXXX.X

Temperature (6 digits including decimal point and symbol) ±XXX.X

Minute 2 digits 00 to 59

Hour 2 digits 00 to 23

Day 2 digits 01 to 31

Month 2 digits 01 to 12

Year 4 digits A.D.

Calibration solution value (6 digits including decimal point and symbol)

0.000 to 14.000

Second 2 digits 00 to 59

12

Response from the instrument if it does not have the calibration data

RPC,************,x,0,3

Channel 1 digit  1:CH1 2:CH2

Display format is fixed. If no data exist, [SP] is displayed.

The number of transmitted calibration data is the number of calibration points.

Displayed calibration date and time is the latest calibration date and time.

When there are two or more calibration points, slope data is displayed and the slope data of the third point is a space.

If an inspection is carried out before use, its data will be transmitted after the calibration data is forwarded.

Slope data

For the slope data, the calibration efficiency, A X 100, between each point is output.

When it exceeds 999.9 or is a negative value, [SP][SP][SP][SP][SP] is output.

13

●Request command and response of the calibration history of ion

Request command

R, IC, x [CR][LF]

Channel

Name (Request of the calibration history of ion)

Header Collects the ion data under selection.

14

Response from pH meter

RIC, x, x, x, x, Xxxxx.x, X, X, (⇒)

Unit 0: g/L

1: mol/L

Ion type

0: -2  1: -1  2: +1  3: +2

Asymmetry potential (7 digits including symbol)

[SP][SP][SP][SP][SP][SP][SP] is output as the asymmetry potential of ion.

Temperature setting 1 digit 0: ATC 1: MTC

Calibration result 0: good

Calibration points 1 digit  1 to 5

Header

Channel 1 digit  1: CH1 2: CH2

Slope (sensitivity) (5 digits including decimal point and symbol)

For one-point calibration⇒First-point sensitivity: [SP]

For two-point calibration⇒First-point sensitivity: Potential diff. between 1st and 2nd solutions, second-point sensitivity [SP]

For three-point calibration⇒First-point sensitivity: Potential diff. between 1st and 2nd solutions, second-point sensitivity: Potential diff. between 2nd and 3rd solutions, third-point sensitivity [SP]

For four-point calibration⇒First-point sensitivity: Potential diff. between 1st and 2nd solutions, second-point sensitivity: Potential diff. between 2nd and 3rd solutions, third-point sensitivity: Potential diff. between 3rd and 4th solutions, fourth-point sensitivity [SP]

For five-point calibration⇒First-point sensitivity: Potential diff. between 1st and 2nd solutions, second-point sensitivity: Potential diff. between 2nd and 3rd solutions, third-point sensitivity: Potential diff. between 3rd and 4th sensitivity solutions, fourth-point sensitivity: Potential diff. between 4th and 5th solutions, fifth-point sensitivity [SP]

(⇒) xxxx, xx, xx, xx, xx, xx, xxxx xx.x

x.xx

xxxx x, Xxxx.x, Xxxxx.x, xxx.x

x, Xxxx.x, Xxxxx.x, xxx.x

x, Xxxx.x, Xxxxx.x, x, Xxxx.x, Xxxxx.x, x.xxx

First point calibration

Second point calibration

Third point calibration

[CR][LF]

Repeatability

Potential (7 digits including symbol)

±XXXX.X

Temperature (6 digits including decimal point and symbol)

±XXX.X

Auxiliary unit of calibration solution

0: -

1: μ

2: m

Calibration solution value (4 digits including decimal point and symbol)

0.000 to 9999

Second 2 digits 00 to 59

Minute 2 digits 00 to 59

Hour 2 digits 00 to 23

Day 2 digits 01 to 31

Month 2 digits 01 to 12

Year 4 digits A.D.

15

Response from the instrument if it does not have the calibration data

RIC,************,x,0,3

Channel 1 digit  1:CH1 2:CH2

16

●Request command and response of the calibration history of conductivity

Request command

R, CC [CR][LF]

Header

Name (Request of the calibration history of conductivity)

17

Response from pH meter

RCC, x, x, x, (⇒)

Temperature setting 1 digit 0: ATC 1: MTC

Header

Calibration result 0: good

Channel 1 digit  1: CH1 2: CH2

(⇒) xxxx, xx, xx, xx, xx, xx, xxx.x xxx.x, X, xxx.x, X,

X,

X,

Xxxx.x, Xxxxx.x,

X.XXX

Xxxx.x, Xxxxx.x,

X.XXX

xxx.x, X, X, Xxxx.x, Xxxxx.x,

X.XXX

xxx.x, X, X, Xxxx.x, Xxxxx.x,

X.XXX

xxx.x, X, X, Xxxx.x, Xxxxx.x,

X.XXX

[CR][LF]

Range 5 calibration data

Range 4 calibration data

Range 3 calibration data

Range 2 calibration data

Range 1 calibration data

Cell coefficiency

Potential (7 digits including symbol)

±XXXX.X

Temperature (6 digits including decimal point and symbol)

±XXX.X

Auxiliary unit

0: -

1: μ

Unit

2: m

0: S/m

1: S/cm

Calibration solution value (5 digits including decimal point and symbol)

0.000 to 199.9

Cell constant

Second 2 digits 00 to 59

Minute 2 digits 00 to 59

Hour 2 digits 00 to 23

Day 2 digits 01 to 31

Month 2 digits 01 to 12

Year 4 digits A.D.

Display format is fixed. If no data exist, [SP] is displayed.

Displayed calibration date and time is the latest calibration date and time.

For the calibration solution concentration, unit, auxiliary unit, temperature, and potential of non-calibration range, enter [SPA].

For the cell coefficiency of non-calibration range, the reference value before or after the calibrated range is displayed.

Response from the instrument if it does not have the calibration data

RCC,************,x,0,3

Channel 1 digit  1: CH1 2: CH2

18

●Request command and response of the measurement value

Request command

R, MD, x [CR][LF]

Channel

Header

Name (Request of the measurement value)

Requests the measurement value of the specified channel.

19

Response from pH meter

RMD, xxxx, xx, x, x, x, x, xxxx, xx, xx, xx, xx, xx, (⇒)

Header

Second 2 digits 00 to 59

Minute 2 digits 00 to 59

Hour 2 digits 00 to 23

Day 2 digits 01 to 31

Month 2 digits 01 to 12

Year 4 digits A.D.

Ion type 1 digit

For ion measurement

0: -2  1: -1  2: +1  3: +2

For other measurement parameter

[SP]

Measurement state 1 digit

0: Instantaneous value 1: Hold state 

2: Potential follow-up measurement

Measurement/calibration type 1 digit

0: Measurement 1: Calibration

Channel 1 digit

1: CH1 2: CH2

Measurement mode 1 digit

1: pH  2: Absolute mV  3: Relative mV  5: Ion  10: Conductivity  

12: Resistivity  11: Salinity  13: TDS

Sample ID 4 digits

※Enter spaces.

[SP][SP][SP][SP]

(⇒) Xxxxxxx, x, x, x, xxxxx, xxxxx, x [CR][LF]

Error state

2: Upper limit alarm

1: Lower limit alarm

0: No alarm

Potential 7 digits

(Leading zero suppression)

Temperature (6 digits including decimal point) -30.0 to 130.0

Right-justified

When the temperature value is Or (Ur), the character string

Or or Ur is displayed instead of an upper or lower limit value.

Temperature setting 1 digit

0: ATC  1: MTC

Unit 1 digit

(pH)

(ORP)

(Ion)

0: Non

0: mV

0: μg/L, 1: mg/L, 2: g/L, 3: mmol/L, 4: mol/L

(Conductivity) 0: S/m, 1: S/cm, 2: mS/cm-Fix

(Resistivity) 0: Ohm/m, 1: Ohm/cm

(Salinity) 0: PPT, 1: %

(TDS)

Auxiliary unit 1 digit

0: g/L

0: Non  1: μ  2: m  3: k  4: M

Measurement value max. 7 digits

20

※Matches the displayed digit of each component to the display specifications.

pH -2.000 to 16.000

ORP

Ion

-2000.0 to 2000.0

0.000 to 9999

Conductivity 0.000 to 1999

Resistivity

Salinity (%)

0.000 to 200.0

0.000 to 10.000

Salinity (PPT) 0.00 to 100.00

TDS 0.00 to 100

When the measurement value is Or (Ur), the character string Or or Ur is displayed instead of upper and lower limit values.

21

●Request command and response of the clock data

Request command

R, OT [CR][LF]

Header

Name (Request of clock data)

Requests the clock data.

Response from pH meter

ROT, xxxx, xx, xx, xx, xx, xx [CR][LF

Header

Second 2 digits 00 to 59

Minute 2 digits 00 to 59

Hour 2 digits 00 to 23

Day 2 digits 01 to 31

Month 2 digits 01 to 12

Year 4 digits A.D. (Leading zero suppression)

●Request command and response of the number of stored memories

Request command

R, MC [CR][LF]

Header

Name (Request the number of stored memories)

Requests the number of data stored in the memory.

Response from pH meter

RMC, xxx [CR][LF]

Number of memories

Header

22

●Request command and response of memory data

Request command

R, MS, xxx, x [CR][LF]

Header

Channel (1 or 2)

Memory number

Name (Request of the measurement value)

Requests the memory data to be specified.

23

Response from pH meter

RMS, xxxx, xxxx,(⇒)

Sample ID

Memory number

Header

(⇒) xx, x, x, x, x, xxxx, xx, xx, xx, xx, xx, Xxxxxxx, (⇒)

Measurement value max. 7 digits

※Matches the displayed digit of each

component to the display specifications.

pH

ORP

-2.000 to 16.000

-2000.0 to 2000.0

Ion 0.000 to 9999

Conductivity 0.000 to 1999

Resistivity 0.000 to 200.0

Salinity (%) 0.000 to 10.000

Salinity (PPT) 0.00 to 100.00

TDS 0.00 to 100

When the temperature value is Or (Ur), the character string Or or Ur is displayed instead of an upper or lower limit value.

Second 2 digits 00 to 59

Minute 2 digits 00 to 59

Hour 2 digits 00 to 23

Day 2 digits 01 to 31

Month 2 digits 01 to 12

Year 4 digits A.D. (Leading zero suppression)

For ion measurement 0: -2  1: -1  2: +1  3: +2

For other measurement parameter [SP]

Measurement state 1 digit

0: Instantaneous value

Measurement/calibration type 1 digit

0: Measurement

Channel 1 digit

1: CH1 2: CH2

Measurement mode 1 digit

1: pH  2: Absolute mV  3: Relative mV  5: Ion  10: Conductivity  12: Resistivity  11: Salinity  13: TDS

24

(⇒) x, x, x, Xxxxxx, Xxxxxxx, x [CR][LF]

Error state

2: Upper limit alarm

1: Lower limit alarm

0: No alarm

Potential 7 digits

(Leading zero suppression)

Temperature (6 digits including decimal point) -30.0 to 130.0

Right-justified

When the temperature value is Or (Ur), the character string Or or Ur is displayed instead of an upper or lower limit value.

Temperature setting 1 digit

0: ATC  1: MTC

Unit 1 digit

(pH)

(ORP)

(Ion)

0: Non

0: mV

0: μg/L, 1: mg/L, 2: g/L, 3: mmol/L, 4: mol/L

(Conductivity) 0: S/m, 1: S/cm, 2: mS/cm-Fix

(Resistivity) 0: Ohm/m, 1: Ohm/cm

(Salinity)

(TDS)

0: PPT, 1: %

0: g/L

Auxiliary unit 1 digit

0: Non  1: μ  2: m  3: k  4: M

25

●Alarm inquiry command and response

Request command

R, AL, x, y [CR][LF]

Request mode

0: Instrument alarm

1: pH

2: mV

3: Ion

4: Conductivity, salinity, resistivity

Channel

Header

Name (Alarm inquiry)

Requests the alarm code in the instrument.

Response from pH meter

RAL, x, y, zzzzzzzz [CR][LF]

Header

Error code

Request mode

Channel

Error code zzzzzzzz

0x00000001

0x00000002

0x00000004

0x00000008

0x00000010

0x00000020

0x00000040

0x00000080

0x00000100

0x00000200

0x00000400

Description

Internal memory error (instrument)

Lower battery error (instrument)

Electrode stability error (other than the instrument)

Asymmetry potential error (pH)

Sensitivity error (pH, ion)

Maximum calibration points exceeded (pH, ion)

Cannot identify standard solution (pH, conductivity)

Calibration interval error (pH)

Printer error (instrument)

Memory full (instrument)

Cell constant is out of range (conductivity)

26

●Alarm clear command and response

Request command

R, AR [CR][LF]

Header

Name (Alarm clear)

OK [CR][LF]

27