data sheet osl calibration kit || bn - SPINNER GmbH
DATA SHEET
OSL CALIBRATION KIT
~256
~70
|| BN ...
~336
Part number
BN 53 38 10
Connector type
7-16 plug, 7-16 socket
Impedance
50 Ω
Frequency range
0 - 7.5 GHz
Electrical data
Open
Short
1)
1)
Load
Frequency range
0 - 3 GHz
3 - 6 GHz
6 - 7.5 GHz
Phase error
≤ 0.5°
≤ 1°
≤ 1.5°
Offset length socket
23.47 mm
Offset length plug
18.47 mm
Frequency range
0 - 3 GHz
3 - 6 GHz
6 - 7.5 GHz
Phase error
≤ 0.5°
≤ 1°
≤ 1.5°
Offset length socket
26.6 mm
Offset length plug
21.6 mm
Frequency range
0 - 7.5 GHz
Return loss
≥ 44 dB
DC-resistance
50 Ω ±0.5
Material and surface
Inner conductor
CuBe age hardened
gold-plated
Outer conductor
copper alloy
gold-plated
Other metal parts
copper alloy
nickel-plated
Insulation
Standards
Operating temperature
Storage temperature
Template TD-00003A
Relative humidity
Weight
Product manual
Included in delivery
cross linked polystyrene
IEC 61169-4
+18 °C - +28 °C
2)
-40 °C - +70 °C
0 - 95% at 40 °C, non-condensing
approx. 1.8 kg
M36044
certificate of calibration, USB-Stick with calibration data and documentation,
aluminium storage case
SPINNER GmbH │ www.spinner-group.com │ [email protected]
Data subject to change without notice – Issue E
1/2
DATA SHEET
OSL CALIBRATION KIT
|| BN ...
1)
2)
The specifications for the opens and shorts are given as allowed deviation from the nominal model as defined in the
calibration data.
Temperature range within all components maintain conformance to their specification.
Calibration data
Calibration data in formats for the common VNAs are included in the kit. It includes individual calibration coefficients for
every kit to achieve the best possible performance.
Pin Depth Limits:
Pin depth is the distance between outer conductor mating plane and inner conductor mating plane. Positive values stand
for protrusion of the inner conductor, negative values for recession.
Load:
Connector Type
7-16
Typical Pin Depth
Measurement
Uncertainty
plug
socket
+1.73 to +1.75 mm
-1.79 to -1.81 mm
0.005 mm
0.005 mm
Connector Type
7-16
Typical Pin Depth
Measurement
Uncertainty
plug
socket
+1.62 to +1.77 mm
-1.77 to -1.92 mm
0.005 mm
0.005 mm
Ranges of
measurement 3)
+1.725 to +1.755 mm
-1.785 to -1.815 mm
Open, Short:
Ranges of measurement is the limit that could be measured with a suitable gauge due to the measurement
uncertainty. These values could still be within the specification. The measurement uncertainty is based on the
measurement with SPINNER gauges and the specified operating temperature. Deviation from these conditions may
cause higher measurement uncertainty.
Template TD-00003A
3)
Ranges of
measurement 3)
+1.615 to +1.775 mm
-1.765 to -1.925 mm
SPINNER GmbH │ www.spinner-group.com │ [email protected]