armitage shanks markwik21 thermostatic basin mixer

armitage shanks markwik21 thermostatic basin mixer
ARMITAGE SHANKS
MARKWIK21 THERMOSTATIC
BASIN MIXER
page 1 of 11
Dimensions are nominal measurements only.
SPECIFICATIONS
Recommended use
Healthcare
Colour availability
Chrome
Pressure rating
Maximum continuous operating pressure 1000 kPa static, 500 kPa dynamic.
For best performance and longer life we recommend 100-500kPa.
Balanced hot/cold water pressure is preferred but will work on unbalanced pressure up to a ratio of 5:1
Temperature rating
Maximum continuous working temperature 40-41°C
Suitable Hot Water
Units
Storage tank: Yes a 20 kPa minimum pressure
Gravity Feed: Yes a 20 kPa minimum pressure
Water Efficiency
WELS approval not applicable for thermostatic products
Standards
AS 4032.1: 2005 Thermostatic Valves
11
CLEANING CHROME SURFACES
Continuous Flow: Yes a 20 kPa minimum pressure
CLEANING RECOMMENDATIONS
To see the complete Armitage Shanks range
go to www.reece.com.au/bathrooms
When cleaning chromed products use only a mild detergent, rinse & w
a soft cloth. ideally clean after each use to maintain appearance.
never use abrasive, scouring powders or scrapers. never use cleanin
containing alcohol, ammonia, hydrochloric acid, sulphuric acid, nitric a
phosphoric acid or organic solvents. use of incorrect cleaning products / methods may res
damage which is not covered by the manufacturer’s guarantee.
When cleaning chromed products use only a mild detergent, rinse & wipe
Outlet
cleaning.
a regular
basis
Outlet cleaning.
on On
a regular
basis the
outlet the
should be inspecte
dry with a soft cloth. Ideally clean after each use to maintain appearance.
to unscrew and remove the outlet, see section 9.7.
outlet
should be inspected & cleaned.
in areas where lime scale build-up is prevalent this should be avoid
Never use abrasive, scouring powders or scrapers. Never use cleaning
cleaning. if it should build up, it will have to be removed. an inhibite
To
unscrew and remove the outlet,
scale solvent can be used such as a kettle de-scaling solvent but
agents containing alcohol, ammonia, hydrochloric acid, sulphuric acid,
to follow
the manufacturer’s
guidelines. after de-scaling it is impo
see
section
6.1.
the parts thoroughly in clean water.
nitric acid, phosphoric acid or organic solvents. Use of incorrect cleaning
carefully
and lime
do notscale
use abrasive
materials
InClean
areas
where
build-up
is or scrapers.
products / methods may result in chrome damage which is not covered by
prevalent this should be avoided by regular cleaning.
the manufacturer’s guarantee.
continuing
If it should build up, it will have to be removed.
An improvement in
armitage shanks pursues
performance of its products.
inhibited proprietary scale solvent can be used
astherefore rese
thissuch
right is
a kettle de-scaling solvent but it is important specification
to follow without notice.
CUSTOMER CARE HELPLINE:
armitage shanks limited
CUSTOMER CARE FAx:
Ws15 4bt
the manufacturer’s guidelines. After de-scaling
it is
armitage
+44 (0)844 543 6170
near rugeley
important to rinse the parts thoroughly in clean
water.
staffordshire
Clean
carefully
and
do not use abrasive materials
englandor
+44
(0)844
543
6171
+44 (0)870 122 8822
scrapers. CARE EMAIL:
CUSTOMER
[email protected]
TM
WaterMark
AS 4032.1 WMK 25821
SAI Global
Disclaimer: Products in this specification manual must by regulation be installed by licensed and registered
trade people. The manufacturer/distributor reserves the right to vary specifications or delete models from
their range without prior notification. Dimensions and set-outs listed are correct at time of publication
however the manufacturer/distributor takes no responsibility for printing errors.
Tech Page Version 1
0
m
4.4
page 2 of 11
Water supply controlling devices (external)
ARMITAGE SHANKS
MARKWIK21 THERMOSTATIC BASIN MIXER
Pressure & temperature ranges of the incoming water supplies should comply with the limits specified
in table.1 (section 4.3).
note:
maximum
Important
Note recommended static pressure in as 3500.1.2 is 500 kpa.
to
avoid
exceeding
this
pressure,
installofa AS/NZS
suitable3500
pressure
reducingplumber.
valve - Your
prvmixer
(or pressure
Mixer
must
be installed
to the
requirements
by a qualified
comes tolimiting
you already factory
valve
plv)
on
both
hot
&
cold
incoming
water
supply
systems.
a
suitable
location
for
a
prvmixers
on the
assembled and tested. We do not recommend the dismantling of any internal part of the mixer. The
arehot
factory tested and
supply
may
on the
to the heating appliance.
sealed so
as be
to give
the cold
best inlet
performance.
supply pressures on both inlets should be reasonably balanced.
INSTALLATION INSTRUCTIONS
similarly, if the water supply temperature ranges do not conform to table.1 then suitable temperature
controlling devices should be installed to achieve this.
5
INSTALLATION GUIDE
Before connection, flush water through pipe-work to remove
all debris etc. to prevent damage to the valve mechanism.
then ensure Water supplies have been isolateD.
sealing
washer
1
2
Fig.3 Installation sequence shown
with clamping ring fixation
Mounting
1. remove the mixer mounting kit if already assembled to the
mixer tail. ensure the basin sealing washer is in place as shown.
this seal should locate into the base recess of the mixer. offer
the mixer towards the basin hole.
note: if preferred, the copper inlet pipes can be screwed into the
bottom of the mixer tail now, or fitted later from under the basin.
Feed the tail (& inlet pipes if fitted) into the basin hole. Allow the
mixer to rest on the basin seated on the rubber washer.
2. Partially fit the screws into the brass clamping ring. From below
the basin, fit the larger rubber washer, brass washer & clamping
ring onto the mixer tail as shown.
3. hand tighten the clamping ring onto the tail so that the screws
contact the brass washer which in turn pushes the rubber washer
until it makes contact with the underside of the basin. ensure the
mixer is positioned correctly, & then tighten the screws securely
with a slotted-drive screw driver.
6
Plumbers, please ensure a copy of the installation
instructions is left with the end user for future reference
H
C
COLD PIPE
TO FRONT
HOT PIPE
TO REAR
519
5.1
page 3 of 11
ARMITAGE SHANKS
MARKWIK21 THERMOSTATIC BASIN MIXER
5.2
PLUMBING OVERVIEW
once the mixer has been secured to the basin, consideration should be given to
installing & positioning of the service valves.
if not done already, screw the copper inlet pipes into the tail of the mixer. note
these Ø10mm inlet pipes can be trimmed if necessary.
To fit service valves:
Observe arrow markings on the service valves. Ensure water flows in the direction indicated.
With the flat seal in place, screw the inlet adaptors onto the service valves using a
24mm a/F spanner. the service valve can be held with grips or use an adjustable
spanner on the flats of the valve body.
slip the compression nuts & olives of the inlet adaptors onto the inlet pipes. push
the adaptors onto the pipes until they stop against the shoulder. slide the olives
into the top of the adaptors & slightly tighten the compression nuts (17mm a/F
spanner).
Orientate the service valve such that the filter cap & isolating screw are easily
accessible for future maintenance.
Fit pipe into the supply side of the service valve & make good the all the joints.
note: When the isolator screw slot is parallel to the valve body, the valve is open
& permits water to flow. To close the valve, rotate the isolator screw 90°.
Check that all joints are securely tightened, test for leaks.
Fig.4
DO NOT apply heat near this product. heat generated by
soldering could damage plastic parts and seals.
6
OPERATION
maximum blend
temperature (preset)
Fig.5
off
warm
Fig.5
cold
7
Plumbers, please ensure a copy of the installation
instructions is left with the end user for future reference
showing the handle control
positions. as the handle is rotated anti-clockwise from the off
position the delivered
water
progresses from cold through
warm to the preset max. temperature of approxximately
41°C.
page 4 of 11
ARMITAGE SHANKS
MARKWIK21 THERMOSTATIC BASIN MIXER
6.1
Outlet information
Table 3 shows the flow rate performance for the flow regulator outlet
q
3,8 l/min
ref 7C (fitted)
300kpa
Table 3 Flow rate data (Q=flow rate)
to replace/clean/service the outlet, use the outlet key supplied with the
product to unscrew the av housing. using the side of the key marked
“junior” locate the key into the inner ring of the housing and unscrew.
Change the outlet & re-secure the housing with the key, ensuring the seal
is in place.
ensure the outlet housing is adequately tightened to prevent leaks &
run back.
1
7
2
COMMISSIONING AND AUDITS
The following procedure should be conducted after installation to ensure the product is functioning
correctly.
7.1
7.1
Purpose
of commissioning
The Purpose
of Commissioning:
• To confirm the correct designation of product versus application.
• To confirm the correct supply water conditions for the product / installation.
• To adjust the mixed temperature if necessary to suit the water supply conditions of the installation.
• To check the product is performing properly.
• To start an audit log and record appropriate data.
This product is factory set at 40±1°C with supplies of circa; 300 KPa balanced pressure and
temperatures of 15°C cold & 65°C hot (i.e. an inlet ∆t of 50C° & a hot / mix ∆t of 24C°)
For supply requirements, see 4.0 Water Supply Conditions.
7.2
7.2
Commissioning
Process
decision
tree FC1)
Commissioning
Process
(see(see
decision
tree FC1)
• Establish that the supply conditions are within the requirements outlined above. If not investigate.
• When conditions are acceptable record them, together with hot and cold temperatures. Establish ∆t.
Check the outlet mix temperature is within 41°C±2°C, if not
(probably due to ∆t > 5C° difference to
factory setting) adjust appropriately. (see 9.1)
• If mixed temperature is correct undertake an 'in field' cold water isolation (CWI) test: (see 7.5 )
Pass – restore supply and record mixed temperature,
Fail – See 7.5 in-field 'Cold Water Isolation' test.
8
Plumbers, please ensure a copy of the installation
instructions is left with the end user for future reference
page 5 of 11
ARMITAGE SHANKS
MARKWIK21 THERMOSTATIC BASIN MIXER
7.2
cont. Commissioning process (decision tree FC1)
check the supply conditions
pressure 50 - 500 KPa & temperature <20 ºC & >55º
YES
NOTE:
Factory settings are:
Inlet ∆t = H - C = 65-15 = 50°C
Hot/Mix ∆t = 65-41 = 24°C
NO
record supply conditions
and ∆t for the hot & cold
investigate
NO
follow setting
or maintenance
instructions
MWT 41ºC (+0 / -2ºC)
re-set mixed to
41°C +0/-2°C
YES
Undertake in-field cold water
isolation test - see 7.5.
Archieved 41 ± 2°C
NOTES:
MWT - Maximum Working
Temperature
Commissioning; Supply Conditions
should be recorded
Yes
pass
accept
Figure 6
record mixed temperature
9
Plumbers, please ensure a copy of the installation
instructions is left with the end user for future reference
No
fail test
page 6 of 11
ARMITAGE SHANKS
MARKWIK21 THERMOSTATIC BASIN MIXER
7.3
7.3
AuditChecks
Checks
TMV‘s
Audit
onon
TMV's
The purpose of a performance audit of a product is:
 To check the product continues to perform properly.
 To flag the need for the product to be adjusted due to supply condition changes or mechanism
ageing.
 To ultimately identify the need for product maintenance.
Note - Additionally these audits facilitate regular verification of the supply conditions.
(See decision tree FC3)
Audit of Supplies To Fittings (FC3)
cold <20ºC
in 2 minutes
YES
hot>55ºC
in 1 minute
supply pressures
are appropriate
NO
record stable state supply
temperature (5 - 20ºC)
NO
investigate system issue
>65ºC
NO
YES
record stable state supply
temperature (55 - 65ºC)
YES
YES
is supply temperature
same as previous audit?
YES
implications to terminal fittings
NO
NO
insignificant change
∆t difference <5ºC
∆t>5ºC
temporary change
of known cause
NO
is supply temperature
same as previous audit?
long term or permanent
change - consider
recommissioning of thermo
fitting(s)
YES
no action
no action.
retest when
conditions normalise
go to field test audit 7.4
or commissioning 7.2
Figure 7
10
Plumbers, please ensure a copy of the installation
instructions is left with the end user for future reference
follow
commissioning
instructions see
YES
page 7 of 11
ARMITAGE SHANKS
MARKWIK21 THERMOSTATIC BASIN MIXER
7.4
7.4
Auditprocedure
procedure
decision
tree FC2)
Audit
(see(see
decision
tree FC2)
Fitting Field Test Audit (FC2)
 Verify the water supplies conditions are
similar to when the product was
commissioned i.e. inlet ∆t as previous + / 5C°
water supplies same as
previous audit or
∆t change >5ºC
Note: If inlet ∆t change is > 5C° consider the
cause of the change, If it is seen as a long
term or permanent change, recommission
the product. If however, it is only a temporary
change, retest when normal conditions
resume.
undertake the cold
water isolation test
see 7.5
 If inlet ∆t differs from the commissioning ∆t by
< 5C° carry out field 'cold water isolation'
(CWI) test (see 7.5 in field
Cold Water Isolation test failure).
 Pass - restore the cold supply and move on to
check the mix temperature.
pass
Note: If the product fails CWI test, refer to 7.5
and then recommission the valve.
restore cold
 Mixed Temperature is as commissioned ±2C°,
Accept and record the temperature.
Note: If mixed Deviation > ±2C° (1st
instance) reset mix as instructions and redo
the test from start. See 9.1 Adjustment of
the mix temperature.
fail
see 9
check mixed
temperature
 2nd & subsequent instances follow
maintenance instructions.
7.5
In-field 'Cold Water
Isolation´(CWI) test.
Locate the cold inline isolating
valve under the basin.
See Fig.11, section 10.
.
Using a flat-blade screw driver, the
screw can be rotated 90° anti-clockwise
to isolate the cold water supply.
same as at
commissioning
±2.0ºC
accept
& record
mixed temp.
deviation from
previous >+2.0ºC
1st instance
reset, follow
setting instructions
Figure 8
2nd instance
follow maintenance
instructions (section 9)
Before commencing the CWI test, ensure:
1. The water supply conditions are met either:
A. For commissioning a new product
go to commissioning FC1 (7.2)
B. Or the inlet ∆t is within ±5°C to when the
product was commissioned (see 7.4).
NOTE It is important that the hot temperature is
greater than 55°C
2. Mixed water outlet temperature is correct (see table 2, section 8).
11
Plumbers, please ensure a copy of the installation
instructions is left with the end user for future reference
page 8 of 11
ARMITAGE SHANKS
MARKWIK21 THERMOSTATIC BASIN MIXER
To perform a CWI test, operate the product by rotating the lever fully to the right.
Then conduct the following procedure:
1. Record the steady state temperature of both hot and cold water supplies. Note the ∆t.
2. Record the temperature of the mixed water at the outlet.
3. Isolate the cold water supply (by rotating the isolating screw 90° & monitor the flow of water
from the outlet.
If the flow ceases, CWI test passed:
1. Restore the cold water supply by rotating the isolator screw 90° clockwise.
2. Re-check the temperature of the stabilised mixed water at the outlet to ensure it is still correct.
Accept & record mixed temperature.
If there is an ongoing flow of water from the mixed water outlet, then 5 seconds after CWI collect the
discharging water into a measuring vessel for 60 seconds. To pass the CWI test the volume of
collected water should be less than 120ml.
If the product fails CWI test, see FC2 (see section 7.4). Follow product maintenance (see section 9)
and servicing (see section 8) instructions.
8
SERVICING
The need for servicing is normally identified as a result of the regular performance auditing.
Application
Maximum mixed water
temperature during
normal operation
Washbasin
Permitted maximum stabilised
temperature recorded during site
testing - excluding transient spikes
43ºC
41ºC
Table 2 A guide to maximum temperature sets
8.1
FREQUENCY OF REGULAR SERVICING
The purpose of servicing regularly is to monitor any changes in performance due to changes in either the
system or the product. This may highlight the need to adjust either the supply system or the product.
These products should be audited 6 to 8 weeks and again 12 to 15 weeks after commissioning. The
results are to be compared against original commissioning settings.
If there are no significant changes at the mixed temperature outlet then a 6 monthly servicing cycle may
be adopted. Otherwise, servicing checks should be carried out more frequently
(e.g. every 4 months).
Follow the recommended auditing and maintenance procedures detailed in sections 7 & 9.
During servicing, note the following:
1. Repeat the procedure of recording and checking supply temperatures. (The same type of measuring
equipment should be used)
2. If the temperature has changed significantly from the previously recorded valves, the
following should be checked:
a. All in-line or integral valve filters are clear of obstruction.
b. All in-line or integral check valves are clean and working properly to prevent backflow.
c. Any isolating valves are fully open.
d. The thermostat is free of debris
3. When satisfied with the mixed outlet temperatures re-record the temperatures.
12
Plumbers, please ensure a copy of the installation
instructions is left with the end user for future reference
page 9 of 11
ARMITAGE SHANKS
MARKWIK21 THERMOSTATIC BASIN MIXER
9
MAINTENANCE
When installed in a healthcare application, we recommend that the commissioning and maintenance
procedures, detailed under 7 & 8, be carried out.
9.1
Adjustment of the mix temperature
4mm
 Note: Cartridge is factory pre-set at 40±1°C at 300 Kpa.
retaining
screw
1. Move the lever to the fully on (max) position.
2. Remove the lever by prising out the cover button and
unscrewing the retaining screw using a 4mm Allen key.
3. Remove the upper Anticlockwise stop ring (slide off spindle).
4. Rotate spindle (using lever handle) to achieve correct mix
temperature. (40±1ºC) Rotate clockwise to reduce and
anticlockwise to increase
5. Refit the upper stop ring against cartridge stop face.
lever
handle
Note:1. This is the normal maximum temperature range for the
upper
product and as such there is a wide zone of spindle movement
stop ring
that gives this mix temperature. For best performance you
should refit the stop at the first point of anticlockwise rotation that lower
achieves the desired setting.
stop ring
Note:2. If you are unable to achieve 40±1°C Max and you have
the correct ∆t's it may be due to fine debris.
See 9.2 Removal and inspection of cartridge.
6. Upon successful completion of the temperature adjustment,
conduct the Cold Water Isolation (CWI) test (See 7.5) to
ensure product is working properly.
7. Refit the lever and record the mix temperature.
9.2
Removal and inspection of cartridge
cartridge
stop face
Figure 9
Adjusting the pre-set temperature
After checking that supply conditions are within the specified
parameters, if the fitting malfunctions or should the test
results fail to fall within the specified limits consider replacing the
cartridge with a new one. See 9.3 Important notes on debris.
Isolate the fitting by turning the isolating valves 90º anti-clockwise
(viewed facing the iso-valve).
Remove the lever handle as described in 9.1. Unscrew the cartridge
from the body with a 36mm a/f deep socket. Inspect cartridge for
damage.
Replace cartridge if necessary and reassemble cartridge into body.
cover button
screw
lever handle
sequential
cartridge
36mm A/F
(deep socket
required)
The replacement cartridge should be tightened to a torque of 15Nm.
(To re-secure the cartridge we recommend the use of Loctite 2400
or equivalent applied to the thread of the cartridge)
Replace lever handle and reinstate the supplies.
After fitting the new cartridge start the test procedure from the
section on commissioning.
 Redo the Cold Water Isolation (CWI) test (See 7.5).
Figure 10
Exploded view showing
sequential cartridge
 Once satisfied with the CWI test, re-check and if necessary, adjust the maximum mix temperature,
see (See 9.1).
 Record the mix temperature.
NOTE: During servicing (or replacing): the thermostatic cartridge fitted in this product does NOT
require lubrication.
13
Plumbers, please ensure a copy of the installation
instructions is left with the end user for future reference
page 10 of 11
ARMITAGE SHANKS
MARKWIK21 THERMOSTATIC BASIN MIXER
9.3
Important notes on debris
Although this product is protected by the service valve filters, debris can still find its way to the thermostat housing area. this can happen during servicing for example. remove cartridge (see section 9.2)
and carry out an inspection.
9.4
Thermostatic Cartridge Ageing
Following many years of normal service you may notice the following:
1. the need to carry out more frequent adjustment of mixed temperature.
2. the thermostatic element may not pass the CWi test.
these issues could be due to the ageing of the thermostat which loses some expansion capability over
time.
these are the principle objectives of testing, as they serve to indicate to maintenance staff the declining
performance capability of the thermostatic cartridge.
For this reason the audit testing flow chart highlights that 2nd Instance CWI test failure or 2nd instance
mixed deviation even with stable ‘as commissioned’ supply conditions and correct inlet supply ∆t’s, is
potentially the first indication of the need to replace the cartridge.
10
SERVICE VALVES
The service valves contain filter, check, isolating valve & 4lpm regulator.
isolation valves facilitate 4 activities:
1) Cleaning strainers & check valves
3) Disinfection
2) servicing the cartridge
4) audit cold water failure testing
Strainers
To ensure trouble free operation of the fitting, the strainer elements should be checked and
cleaned in accordance with the commissioning and servicing guide (see section 8).
to access the strainer element, simply close the isolating valve and unscrew the strainer cap.
The strainer element should be washed with clean water and refitted.
Circlip
flow regulator
check valve
to operate the
isolating valve rotate 90°
(illustrated closed)
strainer cap
strainer screen
direction
of flow
isolating valve
Fig.11 operating the isolating valve
Fig.12 exploded view of isolating valve
To achieve a suitable flow rate where supply pressures are very low it may be necessary to remove the
inline flow regulator.
NOTE: The flow regulator is retained within the valve body by a small circlip. A pair of circlip pliers will
be required to remove the regulator.
14
Plumbers, please ensure a copy of the installation
instructions is left with the end user for future reference
page 11 of 11
ARMITAGE SHANKS
MARKWIK21 THERMOSTATIC BASIN MIXER
11
SPARE PARTS
1B
3
4B
6
5
10
12
7c
11
11
9B
Fig.13 Exploded views
Ref.
Description
Part No.
1b
3
4b
5
5a
6
7C
9b
10
11
12
Cover cap, non-res, as logo
handle retaining screw
lever handle
thermostatic cartridge
sequential, thermostatic cartridge (10 bulk pack)
Couplers –copper tails pair
3.8 lpm laminar Flow insert with housing
Clamping kit & seal
10mm copper tails - pair
isolating valves-pair
Flow regulator
a962017aa
a961950nu
a961823aa
a962280nu
F960879nu
F961107nu
a860970aa
a963381nu
F961106nu
F961105nu
a962570nu
For more information on spare parts contact Armitage Shanks
Customer care, see details on back page.
15
Plumbers, please ensure a copy of the installation
instructions is left with the end user for future reference
YEAR
IT Y
E
5
TE
QU
AL
G UARA
N
REECE PRODUCT QUALITY GUARANTEE
You have purchased a quality product from Reece Pty Ltd
ABN 84 004 097 090 (“Reece”). This product is covered by a
5 year replacement product warranty and a 12 month warranty
over spare parts and labour.
This warranty does not include faults caused by:
5 YEAR WARRANTY
This warranty covers faults in the product construction, material
and assembly. Products which are within 5 years from the
date of purchase, found upon inspection by an authorised
Reece representative, to be defective in construction, material
or assembly, will be repaired or exchanged with an equivalent
product free of charge. Replaced items become Reece’s
property. This warranty also covers any spare parts included
under “Manufacturer’s Provisions” below.
• Installation or part installation by the purchaser or any person
other than a LICENSED PLUMBER
All replacement products will be available for collection without
charge to the customer at the nearest Reece branch to the
customer’s location, or elsewhere as agreed between the
customer and Reece.
Manufacturer’s Provisions
The following spare parts are covered by a 10 year warranty:
• Mixer Cartridge
Labour and Spare Parts
The labour for the replacement of products and spare parts
to which this warranty applies will be supplied by Reece or
relevant supplier using licensed plumbers engaged by Reece or
relevant supplier within 1 year from the date of purchase.
Spare parts other than those listed in the Manufacturer’s
Provisions above, which are within 1 year from the date of
purchase found upon inspection by an authorised Reece
representative to be defective in construction, material or
assembly, will be replaced free of charge.
• Unsuitable or improper use
• Incorrect installation or installation not in accordance with
the instructions provided
• Normal wear and tear
• Inadequate or complete lack of maintenance
• Chemical, electrochemical or electrical influences
• Harsh detergents or abrasive cleaners used on
product finishes
EXCLUSIONS
To the fullest extent permitted by law, Reece excludes all liability
for damage or injury to any person, damage to any property,
and any indirect consequential or other loss or damage.
CLAIM PROCEDURE
For all warranty queries customers are to contact the branch
where the product was purchased. These details can be found
on your purchase invoice.
Our goods come with guarantees that cannot be excluded
under the Australian Consumer Law. You are entitled
to a replacement or refund for a major failure and for
compensation for any other reasonably foreseeable loss or
damage. You are also entitled to have the goods repaired or
replaced if the goods fail to be of acceptable quality and the
failure does not amount to a major failure.
Warranty Conditions
This warranty will apply only under all of the
following conditions:
• The item has been installed by a licensed plumber
• Failure is due to a fault in the manufacture of the product
• Proof of purchase (including the date of purchase)
is provided
• The installation of the product is in accordance with the
instructions provided
MIXERS.C.5YR
This warranty does not cover products purchased as an
ex-display without being fully checked and tested for sale by
the manufacturer.
Don’t risk it, use a licensed plumber.™
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertisement