Series 402 Hornet™ Hot Cathode Miniature-Ionization

Series 402 Hornet™ Hot Cathode Miniature-Ionization
InstruTech®
Series 402 Hornet™ Hot Cathode Miniature-Ionization Vacuum Gauge with Dual Convection
Hot Cathode Bayard-Alpert Ionization gauge with space-saving
built-in controller and display operates 3 different gauges
The first modular ionization vacuum gauge
capable of operating two convection gauges
-9
Full range measurements from 1 x 10 to 1,000 Torr plus
monitoring of your foreline
Low cost of ownership: Significant cost reduction in controller,
space, cabling costs and sensor replacement
Built-in bright digital OLED display with wide
viewing angle, RS485 digital interface, 3 setpoint relays
and 3 log-linear analog outputs
Dual hot filament design, rugged and compact metal
construction
IGM402 Ionization Gauge with Dual Convection
InstruTech IGM402 Hornet™ Modules
Technology
Low Cost of Ownership
Gauge The IGM402 Hornet ionization vacuum gauge module
provides the basic signal conditioning required to turn the gauge
into a complete measuring instrument. It incorporates numerous
design features to enhance performance and reduce cost. The
electrometer circuit auto zeroes to ensure that the readings are
not subject to temperature drift, eliminating the need for
unnecessary, expensive circuitry which further reduces the cost.
Controller The compact modular design with the built-in controller
and display operates three different gauges without requiring
expensive external controllers.
Multiple Gauges The IGM402 Hornet is capable of operating two
external convection vacuum gauges simultaneously.
Full Range Measurement The IGM402 Hornet can combine the
vacuum measurement from the ion gauge and a convection gauge
to provide full range measurements from 1 x 10-9 Torr to 1,000
Torr, or simply operate the ion gauge and the two convection
gauges as individual gauges.
Display The standard built-in bright OLED display provides a
convenient user interface for setup and operation of the vacuum
gauges. The display screen can show all three measurements on
the same screen or display them sequentially. Service screens
allow monitoring of filament operation. Error messages are
displayed for several common fault conditions.
Operation The operation of the gauge including degas, filament
on/off and emission current is set by the front panel push buttons,
digital inputs or RS485 commands.
Sensor For general vacuum applications, dual yttria coated
filaments are offered for use with air and inert gases such as N2,
argon, etc. Dual tungsten filaments are available for use with
gases that may not be compatible with yttria coated filaments.
Space No rack space required. The modular design negates the
need for expensive and limited rack space.
Cabling Cost The cabling cost to connect a nude/glass ionization
gauge to a rack-mount controller can be excessive and installation is
time consuming. With the IGM402 no ion gauge cable is required
and one 10 ft. convection gauge cable is included.
Sensor Replacement Many full range combination gauges provide
measurements from atmosphere to high vacuum with multiple
sensors built into one assembly. A sensor failure may require
replacement of the entire sensor assembly often approaching 50%
of the initial cost of the vacuum gauge itself.
The IGM402 Hornet provides an alternative to these other gauges
by combining the vacuum measurements from the ion gauge and
one of the convection gauges to provide a full range gauge. A
sensor replacement requires only the replacement of the damaged
sensor and not all the other sensors at once.
The IGM402 Hornet sensor assembly can be easily replaced in the
field.
Additional Point of Use In addition to the ion gauge and one
convection gauge, the IGM402 Hornet can provide vacuum
measurements from a second convection gauge without the need
for another expensive controller. This results in significant cost
savings for monitoring the foreline or an additional point of use.
Specifications
measurement range:
ionization (IG)
convection (CG)
used as a full range gauge (IG+CG)
accuracy - N2 (typical)
repeatability - (typical)
display
functionality
materials exposed to gases
sensitivity
x-ray limit
emission current
degas
internal gauge volume
overpressure protection
temperature
bakeout temperature
humidity
weight
housing (electronics)
mounting orientation
serial communications
analog outputs (3 total)
IG
IG+CG
CG
setpoint relays (3 total)
status outputs
input signal
filament selection
input power
connectors
convection gauge compatibility
CE compliance
environmental
Fitting
dimension A
NW16KF
1.45 in. (37mm)
NW25KF
1.45 in. (37mm)
NW40KF
1.45 in. (37mm)
1 1/3 in. Mini-CF 1.85 in. (47 mm)
2 3/4 in. Conflat® 1.70 in. (43 mm)
3/4 in. Tube
2.16 in. (55 mm)
1/2 in. VCR
2.58 in. (65 mm)
-9
-2
-9
Ordering Information
IGM402 Fittings / Flanges
NW16KF
NW25KF
NW40KF
1 1/3 in. Mini-CF / NW 16CF Mini- Conflat®
2 3/4 in. CF / NW35CF Conflat®
3/4 in. Tube (3/4 in. O.D. O-ring compression)
1/2 in. Cajon® 8VCR® female
1475 S. Fordham Street
Longmont, CO 80503
USA
-7
Part Numbers
With Yttria
Filaments
IGM402YBD
IGM402YCD
IGM402YDD
IGM402YED
IGM402YFD
IGM402YAD
IGM402YHD
With Tungsten
Filaments
IGM402TBD
IGM402TCD
IGM402TDD
IGM402TED
IGM402TFD
IGM402TAD
IGM402THD
Granville-Phillips® and Convectron® are registered trademarks of MKS Instruments, Andover, MA.
Conflat® is a registered trademark of Varian, Inc. / Agilent Technologies, Lexington, MA.
InstruTech®
-2
1 x 10 to 5 x 10 Torr / 1.3 x 10 to 6.7 x 10 mbar
/ 1.3 x 10 to 6.7 Pa
-4
-4
-2
1 x 10 to 1,000 Torr / 1.3 x 10 to 1,333 mbar
/ 1.3 x 10 Pa to 133 kPa
-9
-9
-7
1 x 10 to 1,000 Torr
/ 1.3 x 10 to 1,333 mbar
/ 1.3 x 10 Pa to 133 kPa
-8
-3
1 x 10 to 1 x 10 Torr; ± 15% of reading
-3
1 x 10 to 400 Torr; ± 10% of reading,
400 to 1,000 Torr; ±2.5% of reading
-8
-3
-3
1 x 10 to 1 x 10 Torr; ± 5% of reading , 1 x 10 to 1,000 Torr; ±2% of reading
bright OLED display, 3 digits plus 1 digit exponent, user-selectable Torr, mbar, or Pa
ionization gauge can operate up to 2 convection gauges
dual filaments: yttria coated iridium or optional tungsten
Ion collector: tungsten Grid: 304 Stainless Steel Others: 316/304 SS, glass, nickel
factory pre-set. Also user adjustable between 2 to 99
-10
-10
-8
< 5 x 10 Torr, < 6.7 x 10 mbar, < 6.7 x 10 Pa
100 µA, 4 mA
3 W, electron bombardment
3
3
1.0 in (16.4 cm )
-2
IG filament turns off at factory default of 5 x 10 Torr; also user adjustable below 50 mTorr
o
o
operating; 0 to + 40 C
storage; -40 to + 70 C
o
200 C (sensor only - electronics removed)
0 to 95% relative humidity, non-condensing
0.6 lb. (0.27 kg) with NW25 KF flange
aluminum extrusion
any
RS485 - User selectable ASCII protocol,
or BINARY protocol using InstruTech CRC8, minimum command interval: 50 ms for both protocols
one log-linear 0 to 9 Vdc, 1 V/decade , or
one log-linear 0.5 to 7 Vdc, 0.5 V/decade, and
two log-linear 1 to 8 Vdc, 1 V/decade or non-linear 0.375 to 5.659 Vdc
three single-pole, double-throw (SPDT), 1A at 30 Vdc resistive, or ac non-inductive
degas & filament on/off status via display messages, open collector transistor or RS485
degas and filament on/off & emission current are set by continuity to ground
using digital inputs, RS485 or manually using front panel push button
filament 1 or 2 selectable via front panel push buttons or RS485 commands
20 to 28 Vdc, 30 W protected against power reversal and transient over-voltages
(2) 9-pin D-Sub, (2) terminal blocks, (2) convection gauge connectors
InstruTech CVG101 Worker Bee™ or Granville-Phillips® 275 Convectron®
EMC Directive 2004/108/EC, EN61326-1, EN55011 Low Voltage Directive 2006/95/EC, EN61010-1
RoHS
Convection Gauge
Cable Assembly
HB431-1-3F (3 ft.)
HB431-1-10F (10 ft.)
HB431-1-25F (25 ft.)
HB431-1-50F (50 ft.)
>50 ft. Consult Factory
Replacement / Spare IG Sensor
Yttria
Tungsten
IG4YB
IG4TB
IG4YC
IG4TC
IG4YD
IG4TD
IG4YE
IG4TE
IG4YF
IG4TF
IG4YA
IG4TA
IG4YH
IG4TH
Swagelok®, Cajon®, VCR® are registered trademarks of the Swagelok Company, Solon, OH.
Phone +1-303-651-0551
Fax +1-303-678-1754
E-mail [email protected]
Web www.instrutechinc.com
© InstruTech - IGM402 updated 06/2017